Mettler Toledo MP120 User Manual

Measuring Samples
1. Select pH or mV mode
2. Measure sample
Example
Auto endpoint or
3. Rinse, blot and store electrode
n If you press mode to select pH or mV, the MP120 begins
measuring immediately, and you will not need to press read to start the sample measurement.
n The decimal point flashes during sample
measurement. When the electrode output has stabilized the stability indicator appears.
Manual endpoint - press read to endpoint Auto endpoint - the MP120 automatically endpoints
pH
6.84
ATC
25.0°c
Maintenance
General
The MP120 requires very little maintenance. Occasionally wipe the meter with a damp cloth. The casework is made of ABS/PC which is known to be affected by some organic solvents, including toluene, xylene and methyl-ethyl-ketone. It is good practice to wipe away any spillages as soon as they occur.
Every six months inspect and lightly grease the socket seals, and connector and battery ‘O’ rings with the grease supplied. Replace the ‘O’ rings if damaged.
CAUTION: To prevent permanent damage care should be taken to prevent liquid entering the MP120 when fitting electrodes or replacing batteries. If it does, remove the batteries and allow the meter to dry before using.
Replacing the Batteries
Replace batteries as soon as the low battery indicator appears. To prevent data loss always turn the MP120 off using the
mode key before replacing
the batteries. See the Installation section for details on installing batteries. Do not mix old and new batteries.
NOTE: Check the Program Menu settings after changing the batteries.
Electrode Maintenance
Refer to the electrode product insert for full details on maintaining your electrode.
Spares and Accessories
52000379 InLab 413 ‘3 in 1’ pH Electrode (IP67) 51302026 Holster 51302029 Field Case 51302028 Wrist Strap 51302033 Instrument Sealing Kit 51302027 Neck Strap 51300047 Guide to pH Measurement 51302069 pH 4.01 Buffer Sachets, 30/pack 51302047 pH 7.00 Buffer Sachets, 30/pack 51302070 pH 9.21 Buffer Sachets, 30/pack 51302034 30K NTC Probe (IP67) 51302036 Spear 30K NTC Probe (IP67) 52000397 InLab 438 Low Cost pH Electrode
Problem Solving
Err 1 - offset value (E0) out of range
Check correct buffer is used. Check mV reading for pH 7 buffer is 0 ± 30 mV. If it
is not, clean or replace the electrode.
Err 2 - slope out of range
Slope less than 85%, or not calculable.
+ Err 2 Electrode needs cleaning, conditioning
or replacing. Check correct buffers are used.
+ Err 2 Slope more than 105%. Check
calibration buffers.
NOTE: Where 0 mV/pH unit = 0%, and
59.16 mV/pH unit = 100%
Err 3 - pH buffer outside temperature limits
pH buffers must be between 5°C and 50°C for accurate temperature compensation.
Displayed (measurement out of range) Check electrode is connected and immersed in sample. Check wetting cap is removed.
Data Entry Errors
Entered temperature value changes to –5.0 or 105.0 ­the MP120 will not accept temperatures outside this range.
Dashes displayed instead of buffer value during cal buffer set up - that buffer has already been selected for a cal point. Cal points must have different buffer values.
Display/Controls Inactive
Low battery power or batteries fitted incorrectly ­remove batteries and replace correctly (with new batteries if necessary).
Operating Hints
n Remove the wetting cap from the end of the
electrode and the rubber cap from the fill hole (if fitted) before using the electrode.
n Calibrate using buffers with values that bracket
that of the sample. With a new electrode, or after maintenance, we recommend you use a buffer close to pH 7 for the first calibration point.
n For greatest accuracy, buffers and samples
should be at the same temperature.
n Do not use buffers after the expiry date. n When transferring the electrode from one
solution to another, rinse it with distilled water and blot dry with tissue paper - do not wipe the electrode as this may cause polarization and slow response.
n When you select the pH calibration buffers
(Program Menu), set the buffer you will use most frequently for 1-point calibrations as cal 1, and the buffer you will use most frequently for the second calibration point as cal 2.
pH
6.84
ATC
25.0°c
buffer
sample
Calibrating
1. 1-point calibration
Example
Auto endpoint or
2. Rinse electrode and blot dry
to return to sample measurement, or continue with step 3 for the second cal point
3. 2-point calibration - repeat steps 1 and 2 using second buffer
auto or manual endpoint
95 - 105% 90 - 94% 85 - 89% electrode in electrode electrode needs good condition needs cleaning conditioning
n When you press cal the pH buffer you selected for cal 1
(Program Menu) is displayed for 2 seconds, and then the MP120 starts measuring. If you want to use another buffer (choice of 3) press cal again. Press cal repeatedly to exit calibration routine.
n The decimal point flashes during cal measurement.
When the electrode output has stabilized the stability indicator appears. Manual endpoint - press read to endpoint Auto endpoint - the MP120 automatically endpoints
n When a 1-point calibration has endpointed you can
press mode to display the absolute mV value and temperature of the buffer; the display reverts to buffer and E
0
values after 2 seconds.
n When a 2-point calibration has endpointed you can
press mode to display the absolute mV value and temperature of the buffer; the display reverts to buffer and slope values after 2 seconds.
pH
cal
1
7.00
cal buffer reminder
pH
ATC
cal1
7.03
25.0°c
2
seconds
H2O
pH
4.00
97
cal2
% slope
H2O
pH
7.00
E0 mV
3
cal1
b697
Program Menu
The Program Menu allows you to set manual temperature compensation and pH calibration buffers. You can only enter the Program Menu if the current measurement has endpointed - press read if necessary.
Press and hold the mode key for 2 seconds to access the Program Menu - appears on the display.
Press mode to scroll through the options. Press read to exit the Program Menu at any time. If you press read when a value is flashing that value will not be entered.
You can enter temperature manually using
and
. The range is -5.0 to 105.0° C, with a
preset temperature of 25°C. (An ATC probe will override manual compensation.) Press mode to enter the value and move on. You can select three of four calibration buffers. The buffers are grouped in sets (b = 1, b = 2, b = 3); select your required set first using
and
:
set 1 = 7.00, 4.00, 10.01, 1.68 set 2 = 7.00, 4.01, 9.21, 1.68 set 3 = 6.86, 4.01, 9.18, 1.68
Press mode to enter the set and move on.
NOTE: Check buffer set 2 is selected to use the buffers supplied.
Choose three buffers (cal 1, cal 2, cal 3) from the set for calibration. For ease of use, set the buffers in the order that you will use them.
For example: Set 2 (factory settings) cal 1 = 7.00, cal 2 = 4.01 and cal 3 = 9.21
1.68 not selected. These can be rearranged in any order using
and
. Press mode to enter value and move on.
Press read to exit the Program Menu.
MP120 pH Meter
‘O’ rings
connector cover
connector cap
alignment
pH
7.00
ATC
25.0 °c
10
M
Display
Using the Memory
Entering a Reading into Memory
The MP120 can store up to 10 endpointed results. Press
M when the measurement has endpointed.
M 1 is displayed (or M 2 to M 10 if readings have already been stored).
M 10 indicates the memory is full.
Recalling Memory
You can only recall stored memories if the current measurement has endpointed.
Press R - the last stored memory is displayed. Press
or
to scroll through memories. RM 1 to RM 10 indicates which memory is being displayed. M 0 indicates no memories are stored.
Clearing the Memories
Press R then press ∧ or ∨ until M C is displayed. Press
mode to clear the memories, M 0 indicates the
memories have been cleared (press
read to exit without
clearing the memories).
Using Continuous Measurement
In normal mode, the MP120 will auto-off 10 minutes after the last key press, whether the unit is measuring or has endpointed. If you select continuous measurement mode, the MP120 will not auto-off.
To select continuous measurement mode press and hold
M for 2 seconds.
Buffer Tables
The MP120 automatically corrects for temperature using the values shown in the table.
1.68 4.00 4.01 6.86 7.00 9.18 9.21 10.01
5°C 1.67 4.00 4.01 6.95 7.09 9.39 9.45 10.25 10°C 1.67 4.00 4.00 6.92 7.06 9.33 9.38 10.18 15°C 1.67 4.00 4.00 6.90 7.04 9.27 9.32 10.12 20°C 1.68 4.00 4.00 6.88 7.02 9.22 9.26 10.06
25°C 1.68 4.00 4.01 6.86 7.00 9.18 9.21 10.01
30°C 1.68 4.01 4.01 6.85 6.99 9.14 9.16 9.97 35°C 1.69 4.02 4.02 6.84 6.98 9.10 9.11 9.93 40°C 1.69 4.03 4.03 6.81 6.97 9.07 9.06 9.89 45°C 1.70 4.04 4.04 6.83 6.97 9.04 9.03 9.86 50°C 1.71 4.06 4.06 6.83 6.97 9.01 8.99 9.83
Installation
Input Connectors
If you ordered an IP67 electrode with the MP120, connect it as follows, otherwise follow step 2:
1. Make sure connector cap is loosened.
2. Connect the electrode BNC connector to the pH socket.
3. Lightly grease the ‘O’ rings with the grease supplied, then slide the waterproof cover over the socket. Make sure the cover is lined up correctly with the socket.
NOTE: Lightly greasing the cables will help removal of the connector cover in future.
4. Finger tighten the connector cap.
5. Repeat steps 1 - 4 connecting the phono connector to the ATC socket.
CAUTION: To avoid damaging the connectors always loosen the connector cap before removing the cover.
Installing the Batteries
1. Remove the hand grip, and unscrew the battery cover using a coin.
2. Insert the batteries, following the illustration.
3. Lightly grease the ‘O’ ring seal, and replace the battery cover and hand grip.
Fitting the Wrist Strap
pH socket
ATC
(automatic temperature
compensation) socket
pH socket
stability/auto endpoint indicator
low batt. indicator shown here
electrode condition shown here
hand grip coin battery
cover
’O‘ ring seal
Mettler-Toledo GmbH Analytical Sonnenbergstrasse 74 CH-8603 Schwerzenbach Switzerland Tel. (01) 806 77 11 Telefax (01) 806 73 50
Measurement Ranges
pH 0.00 to 14.00 mV ± 1999 Temp. -5.0 to 105.0°C
Resolution
pH 0.01 mV 1 Temp. 0.1
Relative Accuracy*
pH ± 0.01 mV ± 2 mV Temp. ± 0.2°C
Memory
10 memories
Isopotential Point
pH 7.00
Calibration Buffers
3 sets of 4 buffers
Calibration Points
2 of 3 selectable
Temperature Compensation
-5.0 to 105.0°C, automatic and manual
Specifications
Display
LCD
Auto Off
10 minutes from last keypress
Operating Conditions
Temperature5 to 40°C Humidity 5 to 80% (non condensing)
Size/Weight
3 1/4 x 7 3/4 x 1 3/4 inches (85 x 200 x 45 mm) 1lb (0.45kg)
Batteries
4 x AA/LR6 alkaline 1.5V d.c.
Regulatory Compliance
Complies with Part 15 of FCC Rules (Class A computing device) and European EMC Directives EN50081-1: 1992 and EN50082-1: 1992
Waterproof to IP67**
**when used as directed with electrodes with waterproof covers.
51709471 b697
BNC connector
MTC
cal 1
cal 2
cal 3
Controls
On: press any key. Off: from sample measurement press mode to step through to ‘off’.
Press Press & Hold & Release for 2 Seconds
starts/stops turns auto measurement endpoint on/off
starts calibration no action
selects pH/mV/off. enters Program Menu Displays mV and temperature of buffer after cals.
stores result in memory turns continuous (up to 10). Changes measurement value in Prog. Menu.
recalls result from memory. Changes no action value in Prog. Menu.
prog. menu and cont. read icons shown here
temperature, electrode offset/slope shown here
automatic or manual temp. compensation indicator
memory, cal and error number shown here
*± 1 least significant digit