Conventions and Symbols Used in These Operating Instructions1.1
Safety Precautions2
Overview3
Components3.1
Operation Keys3.2
Display Panel3.3
Setting up the Balance4
Unpacking and Delivery Inspection4.1
Installing the Components4.2
Selecting the Location and Leveling the Balance4.3
Selecting the Location4.3.1
Leveling the Balance4.3.2
Power Supply4.4
AC Operation4.4.1
Battery Operation4.4.2
Transporting the Balance4.5
Weighing Below the Balance4.6
Adjustment (Calibration)4.7
Adjustment with Internal Weight4.7.1
Adjustment with External Weight4.7.2
7
7
8
9
9
10
11
13
13
14
16
16
16
17
17
17
19
19
19
19
20
Weighing Made Simple5
Switching the Balance On and Off5.1
Performing a Simple Weighing5.2
Zeroing / Taring5.3
Switching Weight Units5.4
Recall / Recall Weight Value5.5
Weighing with the Weighing-in Aid5.6
Print / Transmit Data5.7
The Menu6
What is in the Menu ?6.1
Menu Operation6.2
Description of Menu Topics6.3
Main Menu6.3.1
Basic Menu6.3.2
Advanced Menu6.3.3
Interface Menu6.3.4
Application "Piece Counting"7
Application "Percent Weighing"8
Application "Check Weighing"9
Application "Statistics"10
22
22
23
23
24
24
24
24
25
25
26
27
27
27
29
32
35
37
39
41
ML Models3
Table of contents
Application "Formulation" (Net Total Formulation)11
Application "Totaling"12
Application "Dynamic Weighing"13
Application "Multiplication Factor Weighing"14
Application "Division Factor Weighing"15
Application "Routine Test"16
Application "Diagnostics"17
Repeatability Test17.1
Display Test17.2
Key Test17.3
Motor Test17.4
Balance History17.5
Calibration History17.6
Balance Information17.7
Service Provider Information17.8
Communication with Peripheral Devices18
Function PC-Direct18.1
Firmware (Software) Updates19
Operating Principle19.1
Update Procedure19.2
43
45
47
49
51
53
56
56
57
58
59
59
60
61
62
64
64
66
66
66
Error and Status Messages20
Error Messages20.1
Status Messages20.2
Cleaning and Service21
Interface Specification22
RS232C Interface22.1
MT-SICS Interface Commands and Functions22.2
Technical Data23
General Data23.1
Model-Specific Data23.2
Balances with Readability of 0.1 mg with Draft Shield23.2.1
Balances with Readability of 1 mg with Draft Shield23.2.2
Balances with Readability of 0.01 g23.2.3
Balances with Readability of 0.1 g23.2.4
Dimensions23.3
Balances with Readability of 0.1 mg with Draft Shield (235 mm)23.3.1
Balances with Readability of 1 mg with Draft Shield (235 mm)23.3.2
Balances with Readability of 1 mg with "Flex" Draft Shield (105 mm)23.3.3
Balances with Readability of 0.01 g with Square Weighing Pan and23.3.4
67
67
67
69
70
70
70
73
73
73
73
74
75
76
79
79
80
81
82
ML Models4
Table of contents
23.3.5
Accessories and Spare Parts24
Appendix25
Index26
Balances with Readability of 0.01 g / 0.1 g with Round Weighing
Pan
Balances with Readability of 0.1 g with Square Weighing Pan23.3.6
Accessories24.1
Spare Parts24.2
Conversion Table for Weight Units25.1
83
84
85
85
88
91
91
92
ML Models5
Introduction
1 Introduction
Thank you for choosing a METTLER TOLEDO balance.
The precision balances of the NewClassic line combine a large number of weighing possibilities with easy opera
tion.
These operating instructions apply to all balance models ML in the NewClassic line. However, the different models
have different charcteristics regarding equipment and performance. Special notes in the text indicate where this
makes a difference to operation.
1.1 Conventions and Symbols Used in These Operating Instructions
Key designations are indicated by double angular brackets (e.g. «»).
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
This symbol indicates an automatic sequence.
These symbols indicate safety notes and hazard warnings which, if ignored, can cause personal
danger to the user, damage to the balance or other equipment, or malfunctioning of the balance.
This symbol indicates additional information and notes. These make working with your balance
easier, as well as ensuring that you use it correctly and economically.
ML Models7
Safety Precautions
2 Safety Precautions
Always operate and use your balance only in accordance with the instructions contained in this manual. The
instructions for setting up your new balance must be strictly observed.
If the balance is not used according to these Operating Instructions, protection of the balance may be
impaired and METTLER TOLEDO assumes no liability.
It is not permitted to use the balance in explosive atmosphere of gases, steam, fog, dust and
flammable dust (hazardous environments).
For use only in dry interior rooms.
Do not use sharply pointed objects to operate the keyboard of your balance! Although your bal
ance is very ruggedly constructed, it is nevertheless a precision instrument. Treat it with corre
sponding care.
Do not open the balance: It does not contain any parts which can be maintained, repaired, or
replaced by the user. If you ever have problems with your balance, contact your METTLER
TOLEDO dealer.
Use only balance accessories and peripheral devices from METTLER TOLEDO; they are optimally
adapted to your balance.
Use only the original universal AC adapter delivered with your balance.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to
countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point speci
fied for electrical and electronic equipment. If you have any questions, please contact the
responsible authority or the distributor from which you purchased this device. Should this device
be passed on to other parties (for private or professional use), the content of this regulation
must also be related.
Thank you for your contribution to environmental protection.
ML Models8
3 Overview
C
0.1 mg
C
1 mg
C
1 mg
C
0.01 g
3.1 Components
Overview
ML Models9
C
0.01g / 0.1g
C
0.1 g
11
12
13
10
14
C
5142
9
36
7
8
Overview
Name and Function of Components
1Display8Glass draftshield
2Operation keys9Handle for operation of the draft-shield door
3Model sticker (with approved models only)10 Socket for AC Adapter
4Level indicator11 RS232C serial interface
5Weighing pan 12 Kensington slot for anti-theft purposes
6Draft shield element13 Slot for second interface (optional)
7Leveling foot14 Product label
3.2 Operation Keys
ML Models10
Key Functions
Application Icons
Status Icons
Weighing-in aid
Weight Value Field
Unit Field
No.KeyPress briefly (less than 1.5 s)Press and hold (longer than 1.5 s)
1
●
To change display resolution (1/10d func
no function
tion) while application is running
2
●
To navigate back (scroll up) within menu
topics or menu selections
●
Decrease (numerical) parameters within
●
To select the weighing application
●
Decrease (numerical) parameters quickly
within menu and in applications
menu and in applications
3
●
To navigate forward (scroll down) within
menu topics or menu selections
●
To select assigned application and enter
ing the parameter settings of application.
Overview
4ON//OFF
5
6
7
8
9
●
Increase (numerical) parameters within
menu and in applications
●
Switch on
●
Zero/Tare
●
Cancel and to leave menu without saving
(one step back in the menu).
●
With entries: scroll down
●
To navigate through menu topics or menu
selections
●
To toggle between unit 1, recall value (if
selected), unit 2 (if different from unit 1)
and the application unit (if any)
●
To enter or leave menu selection (from / to
menu topic)
●
To enter application parameter or switch to
next parameter
●
To store parameter
●
Printout display value
●
Default application assignment: Piece
counting
●
Increase (numerical) parameters quickly
within menu and in applications
Balances with readability of 0.1 mg, with draft shield
(235 mm)
Place the following components on the balance in the
specified order:
Note: Push the side glass door (1) back as far as will
go.
●
Bottom plate (2)
●
Draft shield element (3)
●
Pan support (4)
●
Weighing pan (5)
Balances with readability of 1 mg, with draft shield
(235 mm)
Place the following components on the balance in the
specified order:
Note: Push the side glass door (1) back as far as will
go.
●
Bottom plate (2)
●
Pan support (3)
●
Weighing pan (4)
ML Models14
C
2
3
4
1
5
C
5
4
3
2
1
C
2
1
Setting up the Balance
Balances with readability of 1 mg, with flex draft
shield (105 mm)
Place the following components on the balance in the
specified order:
a) Place bottom plate (1) (push and turn)
b) Place pan support (2) with weighing pan (3).
c) Caerfully unfold draft shield (4) and place it cor
rectly. Pay attention to the glass panels!
d) Place draft shield top (5)
Balances with readability of 10 mg with square
weighing pan and draft shield element
a) Remove the two screws (1)
b) Remove the plate (2) and retain it.
c) Place draft shield element (3) and fix it with the two
screws.
d) Place pan support (4) with weighing pan (5).
Balances with readability of 0.01 g / 0.1 g with round
weighing pan
Place the following components on the balance in the
specified order:
●
Pan support (1)
●
Weighing pan (2)
ML Models15
C
2
1
C
Setting up the Balance
Balances with readability of 0.1 g with square weigh
ing pan
Place the following components on the balance in the
specified order:
●
Pan support (1)
●
Weighing pan (2)
4.3 Selecting the Location and Leveling the Balance
Your balance is a precision instrument and will thank you for an optimum location with high accuracy and
dependability.
4.3.1 Selecting the Location
Select a stable, vibration-free position that is as horizontal as possible. The surface must be able to safely carry the
weight of a fully loaded balance.
4.3.2 Leveling the Balance
Observe ambient conditions (see Technical Data).
Avoid the following:
●
Direct sunlight
●
Powerful drafts (e.g. from fans or air conditioners)
●
Excessive temperature fluctuations
The balances have a level indicator and two or four adjustable leveling
feet to compensate for slight irregularities in the surface of the weighing
bench. The balance is exactly horizontal when the air bubble is in the
middle of the level glass.
Note: The balance should be leveled and adjusted each time it is moved
to a new location.
Balances with 2 leveling feet
Adjust the two front leveling feet appropriately until the air bubble comes
to rest exactly in the middle of the glass:
Air bubble at"12 o'clock"turn both feet clockwise
Air bubble at"3 o'clock"turn left foot clockwise, right
foot counterclockwise
ML Models16
4.4 Power Supply
Allow your balance to warm up for 30 minutes (0.1 mg models 60 minutes) to enable it to
adapt itself to the ambient conditions.
4.4.1 AC Operation
Setting up the Balance
Air bubble at"6 o'clock"turn both feet counterclockwise
Air bubble at"9 o'clock"turn left foot counterclockwise,
right foot clockwise
Balances with 4 leveling feet
a) First turn the two rear leveling feet all the way in.
b) Adjust the two front leveling feet as previously described.
c) Turn the rear leveling feet down onto the surface for extra stabilizing
safety, so the balance cannot tilt over under eccentric loads.
Your balance is supplied with an country-specific AC adapter or with a country-specific power cable. The power
supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz (for exact specifications, see section
"technical data").
First, check the local line voltage is in the range 100 - 240 VAC, 50/60 Hz and whether the
power plug fits your local power supply connection. If this is not the case, on no account connect the balance or the AC adapter to the power supply, but contact the responsible METTLER
TOLEDO dealer.
Important:
●
Before operating, check all cables for damage.
●
Guide the cables so that they cannot become damaged or interfere with the weighing
process!
●
Take care that the AC adapter cannot come into contact with liquids!
●
The power plug must be always accessible.
Connect the AC adapter to the connection socket on the back of your bal
ance (see figure) and to the power line.
4.4.2 Battery Operation
The Balance can also operates with batteries. Under normal operation conditions, the balance works independently
of the AC power line for about 8 to 15 hours (using alkaline batteries). Immediately after the AC power supply is
ML Models17
Setting up the Balance
interrupted e.g. by withdrawing the power plug or if there is a power failure, the balance switches automatically to
battery operation. Once the AC power supply is restored, the balance reverts automatically to AC operation.
Note: It is also possible to use rechargeable batteries. Charging batteries inside the balance is not possible.
battery fully When the balance is operating on its batteries, the battery symbol in the
display lights up. The number of segments that are lit is an indicator of
battery condition (3 = fully charged, 0 = discharged). When the batteries
2/3 full
are almost completely discharged, the battery symbol flashes.
1/3 full
battery empty
Inserting / Replacing Batteries
The balance must be disconnected from the power supply when carrying out all setup and
mounting work.
●
Make sure that the balance is off before removing or inserting batteries.
●
Do not place the balance on the pan support location bolt.
●
Battery Warnings: Read and follow all warnings and instructions supplied by the battery
manufacturer.
●
Do not mix different types or brands of batteries. Performance of batteries can vary very
greatly depending on the manufacturer.
●
If you don't operate the balance with batteries for an extended period, it is recommended to
remove the batteries from the balance.
Your balance uses 8 standard AA (LR6) batteries (alkaline batteries preferred)
a) Remove weighing pan, pan support and draft shield element or draft
shield "100 mm" if present.
b) Turn the balance carefully on its side.
c) Open and remove the battery-chamber cover.
d) Insert / replace the batteries with the correct polarity as shown in the
battery holder.
e) Insert and close the battery-chamber cover.
f) Turn the balance carefully to its normal position.
g) Reinstall all components in the reverse order.
ML Models18
A
Setting up the Balance
4.5 Transporting the Balance
Switch off the balance and remove the power cable and any interface cable from the balance. Refer to the notes in
Section "Selecting the location" regarding the choice of an optimal location.
Transporting Over Short Distances
For balances with a draft shield: Observe the following instructions to transport your balanceover a short distance to a new location: Never lift the balance using the glass draft shield.
The draft shield is not sufficiently fastened to the balance.
Transporting Over Long Distances
If you would like to transport or send your balance over long distances, use the complete original packaging.
4.6 Weighing Below the Balance
The balances are equipped with a hanger for carrying out weighings below the work surface (weighing below the
balance).
Attention:
●
Do not place the balance on the pan support location bolt.
a) Switch off the balance and remove the power cable and any
interface cable from the balance.
b) Remove weighing pan, pan support and draft shield element or
"Easy draft shield" if present.
c) Turn the balance carefully on its side.
d) Remove one of the caps (A) depending on the models.
e) Then turn the balance to its normal position and simply rein
stall all components in the reverse order.
4.7 Adjustment (Calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational
acceleration at its location. Adjusting is necessary:
●
before the balance is used for the first time.
●
at regular intervals during weighing service.
●
after a change of location.
4.7.1 Adjustment with Internal Weight
Note: On models with internal weight only (see technical data).
To obtain accurate results, the balance must be connected to the power supply for approxi
mately 30 minutes (0.1 mg balances 60 minutes) in order to reach operating temperature
before adjusting.
ML Models19
------
ADJ.DONE
0.00
g
ADJ.INT
-- Internal adjustment -
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
Temperature 22.5 °C
Diff 3 ppm
Adjustment done
-----------------------
-------
2000.00
g
ADJ.DONE
0.00
g
C
0.00
g
------
C
ADJ.EXT
Setting up the Balance
Sample adjustment printout using internal weight:
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) of advanced menu "ADJ.INT" must be selected.
a) Unload weighing pan
b) Press and hold « » to execute "Internal Adjustment".
The balance adjusts itself automatically. The adjusting is finished when
the deviation and the message "ADJ.DONE" appears briefly in the dis
play. The balance returns to the last active application and is ready for
operation.
4.7.2 Adjustment with External Weight
Note: Because of certification legislation, the approved models cannot be adjusted with an external weight.
To obtain accurate results, the balance must be connected to the power supply for approxi
mately 30 minutes (0.1 mg balances 60 minutes) in order to reach operating temperature
before adjusting.
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) " of advanced menu ADJ.EXT" must be selected.
a) Have required adjustment weight ready.
b) Unload weighing pan.
c) Press and hold « » to execute "External Adjustment". The required
(predefined) adjustment weight value flashes in the display.
d) Place adjustment weight in center of pan. The balance adjusts itself
automatically.
e) When "0.00 g" flashes, remove adjustment weight.
The adjusting is finished when the deviation value and the message
"ADJ.DONE" appears briefly in the display. The balance returns to the last
active application and is ready for operation.
ML Models20
Sample adjustment printout using external weight:
-- External adjustment -
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
Temperature 22.5 °C
Nominal 2000.00 g
Actual 1999.99 g
Diff 5 ppm
Adjustment done
Signature
.......................
-----------------------
Setting up the Balance
ML Models21
On
Off
8888888
XXXXXXX
0.00
g
SHUTOFF
On
Off
Weighing Made Simple
5 Weighing Made Simple
This section shows you how to perform simple weighings and how you can accelerate the
weighing process.
5.1 Switching the Balance On and Off
This section shows you how to perform simple weighings and how you can accelerate the weighing process.
Switching On
a) Remove any load from weighing pan.
b) Press «On».
The balance performs a display test (all segments in the display light up
briefly), "WELCOME", Software version, Maximum load and Readability
appears briefly. (Startup "FULL" mode only)
The balance is ready for weighing or for operation with the last active
application.
Switching Off
Press and hold the «Off» key until "SHUTOFF" appears on the display.
Release the key.
After it has been switched off, your balance is in the standby mode. The display shows date,
time, maximum load and readability. As your balance needs no warm-up time in the standby
mode and is immediately ready for weighing.
When Quickstart "QUICK" (Advanced menu, topic "STARTUP") is selected: If you wish to perform
a weighing, you need now only place the sample on the weighing pan and the balance immedi
ately displays the result. There is no need to switch it on with the «On/Off» key.
Note:
●
Quickstart is not possible with approved balances.
●
Standby mode is available on line powered balances only.
ML Models22
5.2 Performing a Simple Weighing
1250.38
g
0.00
g
1182.03
g
C
#
0.00
g
121.01
g
#
0.01
g
0.00
g
C
C
121.01
g
0.00
g
0.00
g
95.97
g
121
.
01
g
-
Net
Net
Net
#
C
C
a) Press «» to zero the balance.
Note: If your balance is not in the weighing mode, first press and
hold the «» key until “WEIGH“ appears in the display. Release the
key. Your balance is in the weighing mode.
b) Place weighing sample on the weighing pan.
c) Wait until the instability detector " " disappears and the stability
beep sounds.
d) Read the result.
5.3 Zeroing / Taring
Zeroing
a) Unload the balance.
Weighing Made Simple
b) Press «
» to set the balance to zero. All weight values are
measured in relation to this zero point (see menu topic
"ZERO.RNG").
Note: Use the «» zeroing key before you start with a weighing.
Taring
If you are working with a weighing container, first set the balance to zero.
a) Place empty container on the balance. The weight is displayed.
b) Press «
» to tare the balance.
"0.00 g" and "Net" appears in the display. "Net" indicates that all weight
values displayed are net values.
Note:
●
If the container is removed from the balance, the tare weight will be
shown as a negative value.
●
The tare weight remains stored until the «» key is pressed
again or the balance is switched off.
ML Models23
0.78
oz
22.00
g
*
22.00
g
M
121.01
g
0.00
g
0.00
g
M
121.01
g
0.00
g
C
C
0%
100%
Weighing Made Simple
5.4 Switching Weight Units
The «» key can be used at any time to toggle between weight unit
"UNIT 1","RECALL" value (if selected), weight unit "UNIT 2" (if different
from weight unit 2) and the application unit (if any).
5.5 Recall / Recall Weight Value
Recall stores stable weights with an absolute display value bigger than 10d.
Requirement: The function "RECALL" must be activated in the menu.
a) Load weighing sample. The display shows weight value and stores
stable value.
b) Remove weighing sample. When the weight is removed the Display
shows zero.
c) Press «». The display shows last stored stable weight value for 5
seconds together with asterisk (*) and Memory (M) symbols. After 5
seconds the display goes back to zero. This can be repeated unlimit
ed times.
Delete last weight value
As soon a new stable weight value is displayed, the old recall value
becomes replaced by the new weight value. When pressing «»,
the recall value is set to 0.
Note: If the power is switched off, the recall value is lost. The recall value
can not be printed.
5.6 Weighing with the Weighing-in Aid
The weighing-in aid is a dynamic graphic indicator which shows the
used amount of the total weighing range. You can thus recognize at a
glance whether the load on the balance approaches the maximum load.
5.7 Print / Transmit Data
Pressing the «» key transmits the weighing results over the interface
e.g. to a printer or a PC.
ML Models24
6 The Menu
6.1 What is in the Menu ?
The Menu allows you to match your balance to your specific weighing needs. In the menu you
can change the settings of your balance and activate functions. The main menu has 4 different
menus and these contains 27 different topics, each of which allows you various selection pos
sibilities. For Menu "PROTECT" see chapter "Description of menu topics" section "Main menu".
Note: See Quick Guide for the graphical overview of the menu (Menu Map) with all setting possibilities.
Menu "BASIC"
TopicDescription
DATESetting the current date.
TIMESetting the current time.
UNIT 1Specification of the 1st weight unit in which the balance should show the result.
UNIT 2Specification of the 2nd weight unit in which the balance should show the result.
KEY.BEEPSetting the key beep level.
STB.BEEPSetting the stability beep level.
RESETCall up of the factory settings.
The Menu
Menu "ADVANCE."
TopicDescription
SRV.ICONSwitching the service reminder (service icon) on or off.
ENVIRON.Matching the balance to the ambient conditions.
CALSettings for the type of adjustment (calibration).
DATE.FRMSetting the date format.
TIME.FRMPreselection of the time format.
RECALLSwitching the application "Recall" for storing stable weights on or off.
STARTUPSetting the mode which the balance powers up ("FULL" or "QUICK").
SHUTOFFSetting the time after which the balance should be switched off automatically.
B.LIGHTSetting the time after which the display backlight should be switched off automatically.
A.ZEROSwitching the automatic zero correction (Autozero) on or off.
ZERO.RNGSetting the zero limit of the zero/tare key.
ASSIGN:FSelection of assigned F key application and entering their parameter settings.
DIAGNOS.Starting a diagnostic application.
RESETCall up of the factory settings.
Menu "INT.FACE"
TopicDescription
RS232Matching the serial interface RS232C to a peripheral unit.
BAUDSetting the transfer speed of the serial interface RS232C.
BIT.PAR.Setting the character format (Bit/Parity) of the serial interface RS232C.
HD.SHKSetting the transfer protocol (Handshake) of the serial interface RS232C.
RS E.O.L.Setting the end of line format of the serial interface RS232C.
RS.CHARSetting the char set of the serial interface RS232C.
INTERVL.Selection of the time which the balance started an automatic printout.
RESETCall up of the factory settings.
ML Models25
406.2
g
ADVANCE
BASIC
TIME
DATE
ENVIRON.
STABLE
ENVIRON.
01.
2
500
01.
6
500
01.6500
The Menu
6.2 Menu Operation
In this Section you will learn how to work with the menu.
Select Menu
a) Press and hold « » to activate main menu. The first menu "BASIC"
is displayed (except menu protection is active).
b) Press «» repeatedly to change menu (Scrolling down/up «+» /
«–» keys).
c) Press «» to confirm the selection.
Note: The menu selection "BASIC", "ADVANCE." or "INT.FACE" can not be
saved. The selection "PROTECT" must be saved.
Select Menu Topic
Press «». The next menu topic appears in the display. Each time the
«» or the «+» key is pressed, the balance switches to the next menu
topic; the «–» key to the previous menu topic.
Change Settings in a Selected Menu Topic
a) Press «».The display shows the current setting in the selected
menu topic. Each time «» or «+» is pressed, the balance switch
es to the next selection; press «–» to the previous selection. After the
last selection, the first is shown again.
b) Press «», the selected setting is accepted but not yet executed.
The settings are executed only after "SAVE:YES" has been confirmed.
Change Settings in a Submenu Selection
The same procedure as for menu topics.
Input Principle of Numerical Values
a) Press «» for input of numerical values.
b) Press «» to select a digit or a value (depending on the applica
tion). The selected digit or the selected value is blinking.
c) For changing digits or values, press «+» to scroll up or «–» to scroll
down.
d) Press «
» to confirm the input.
ML Models26
SAVE:YES
SAVE:NO
The Menu
Saving Settings and Closing the Menu
a) Press and hold « » to leave menu topic.
b) Press «» to execute "SAVE:YES". Changes are saved.
c) Press «» to execute "SAVE:NO". Changes are not saved. To toggle
between "SAVE:YES" and "SAVE:NO" press «».
Cancel
For leaving menu topic or menu selection without saving press «C» (one
step back in the menu).
Note: If no entry is made within 20 seconds, the balance reverts to last active application mode. Changes are not
saved. If changes are made, the balance asks "SAVE:NO".
6.3 Description of Menu Topics
In this Section you will find information regarding the individual menu topics and the available selections.
6.3.1 Main Menu
Selecting the menu.
"BASIC"The small "BASIC" menu for simple weighing is displayed.
"ADVANCE."The extended "ADVANCE." menu for further weighing settings is dis
played.
"INT.FACE"The menu "INT.FACE" for all interface parameter settings for periph
eral devices e.g. printer is displayed.
"PROTECT"Menu protection. Protection of balance configurations against
unmeant manipulation.
"OFF"Menu protection is off. (Factory setting)
"ON"Menu protection is on. The menu BASIC, ADVANCE. and INT.FACE
are not displayed.
Note:
●
The menu selection "BASIC", "ADVANCE." or "INT.FACE" can not be saved.
●
To activate "PROTECT" "ON" or "OFF", this selection must be saved.
6.3.2 Basic Menu
"DATE" – Date
Setting the current date according to date format.
Note: A reset of the balance will not change this setting.
"TIME" – Time
Setting the current time according to time format
"+1H"Set the current time forwards by 1 hour (to adjust summer or winter
time). (Factory setting)
ML Models27
The Menu
"-1H"Set the current time backwards by 1 hour (to adjust summer or
winter time).
"SET.TIME"Enter the current time.
Note: A reset of the balance will not change this setting.
"UNIT 1" – Weight Unit 1
Depending on requirements, the balance can operate with the following units (depending on the model)
●
Only those weight units allowed by the appropriate national legislation are selectable.
●
With approved balances, this menu topic has a fixed setting and cannot be changed.
●
Conversion table for weight units see chapter Appendix.
Units:
1)
g
kg
mg
2)
3)
GramdwtPennyweight
KilogrammomMomme
MilligrammsgMesghal
ctCarattlhTael Hong Kong
lbPoundtls
4)
Tael Singapore
ozOunce (avdp)tltTael Taiwan
oztOunce (troy)tolaTola
GNGrainbahtBaht
1)
factory setting
2)
not with 0.1 mg and 1 mg balances
3)
with 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
"UNIT 2" – Weight Unit 2
If it is required to show the weighing results in weighing mode in an additional unit, the desired second weight unit
can be selected in this menu topic (depending on the model). Units see "UNIT 1".
Note: Only those weight units allowed by the appropriate national legislation are selectable.
"KEY.BEEP" – Key Beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the set
ting.
"MED"Medium level (Factory setting)
"HIGH"High level
"OFF"Beep switched off
"LOW"Low level
"STB.BEEP" – Stability Beep
If the unstable symbol disappears, the stability beep becomes active. This menu topic allows you to preselect the
volume of the stability beep.
"LOW"Low level (Factory setting)
"MED"Medium level
"HIGH"High level
ML Models28
"OFF"Beep switched off
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up the factory settings.
To toggle between "YES?" and "NO?" press «» (or «+» or «–»).
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
6.3.3 Advanced Menu
"SRV.ICON" – Service Reminder
This menu topic allows you to switch the service reminder "
"ON"Service reminder " " switched on.You will be informed after one
Year or 8000 operating hours to call service for recalibration. This
will be indicated by the flashing service icon: " ". (Factory set
ting)
"OFF"Service reminder " " switched off.
"ENVIRON." – Environment Settings
This setting can be used to match your balance to the ambient conditions.
" on or off.
The Menu
"STD."Setting for an average working environment subject to moderate
variations in the ambient conditions. (Factory setting)
"UNSTAB."Setting for a working environment where the conditions are continu
ously changing.
"STABLE"Setting for a working environment which is practically free from
drafts and vibrations.
"CAL" – Adjustment (calibration)
In this menu topic you can preselect the function of the « » key. Your balance can be adjusted with internal or
external weights by pressing the « » key. If you have attached a printer to your balance, the data of the adjustment
(calibration) are printed out.
"ADJ.OFF"The adjustment is switched off. The « » key has no function.
"ADJ.INT"Internal adjustment: adjustment is performed at a keystroke with
the built-in weight (depending on the model, see technical data).
"ADJ.EXT"External adjustment: adjustment is performed at a keystroke with a
selectable external weight.
Note: This function is not available for approved balances.
"200.00 g"Defining the external adjustment weight: define the weight of the
external adjustment weight (in grams).
Factory setting: depends on the model.
"DATE.FRM" – Date Format
This menu topic allows you to preselect the date format.
The following date formats are available:
Display examplesPrinting examples
ML Models29
The Menu
"DD.MM.Y"01.02.0901.02.2009
"MM/DD/Y"02/01/02/01/2009
"Y-MM-DD"-02-012009-02-01
"D.MMM.Y"1.FEB.091.FEB.2009
"MMM.D.Y"FEB.1.09FEB 1 2009
Factory setting: "DD.MM.Y"
"TIME.FRM" – Time Format
This menu topic allows you to preselect the time format.
The following date formats are available:
"24:MM"15:04
"12:MM"3:04 PM
"24.MM"15.04
"12.MM"3.04 PM
Factory setting: "24:MM"
Display examples
"RECALL" – Recall
This menu topic allows you to switch the "RECALL" function on or off. When it is switched on recall stores the last
stable weight if the absolute display value was bigger than 10d.
"OFF"
"ON"
Note: The recall value is displayed with an asterisk and cannot be printed.
"STARTUP" – Startup Mode
You can set your balance such that it either immediately starts from the standby mode when you load a weight or it
must be switched on with the «ON/OFF» key after which it then performs a display test.
Note: This topic is not visible with approved balances.
"QUICK""Quickstart": The balance can be started directly from the standby
"FULL"Start with display test: You must switch on the balance with the
"RECALL" switched off (Factory setting)
"RECALL" switched on
mode and is immediately ready for weighing. You can load the
weight in the standby mode and the balance immediately shows
the current weighing result. This is the Factory settingNote: Standby mode is available on line powered balances only.
«ON/OFF» key. After it has been switched on, it performs a display
test for approx. 2 sec. in which all display elements lights up, it
shows "WELCOME", software version, maximum load and read
ability. The balance is ready for weighing.
"SHUTOFF" – Automatic Shutoff
If the automatic shutoff function is activated, the balance automatically switches itself off after a preselected time of
inactivity (i.e. with no key being pressed or changes of weight occurring etc.) and is switched to the standby
mode.
"A.OFF:10' " Automatic shutoff after 10 minutes of inactivity. (Factory setting)
ML Models30
Loading...
+ 70 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.