Mettler Toledo JP, JP105DUG Operating Instructions Manual

C
Jewelry Semi-Micro Balances
JP
Operating Instructions

Table of Contents

1 Introduction 3
2 Safety Information 4
2.2 Product specific safety notes ......................................................................................... 4
3 Design and Function 6
4 Installation and Putting into Operation 10
4.1 Unpacking and delivery inspection................................................................................. 10
4.7 Performing a simple weighing....................................................................................... 17
3.1.1 Components................................................................................................. 6
3.1.2 Operation keys.............................................................................................. 7
3.1.3 Display ........................................................................................................ 8
4.5.1 Switching on the balance ............................................................................... 12
4.5.2 Leveling the balance...................................................................................... 12
4.5.3 Left/right operating of the glass draft shield ...................................................... 13
4.5.4 Adjusting the balance .................................................................................... 13
4.6.1 Fully automatic adjustment FACT .................................................................... 14
4.6.2 Manual adjustment with internal weight ........................................................... 14
4.6.3 Manual adjustment with external weight .......................................................... 14
4.6.4 Customer fine adjustment............................................................................... 15
5 The Menu 21
5.3.1 Main menu................................................................................................... 23
5.3.2 Basic menu.................................................................................................. 24
5.3.3 Advanced menu............................................................................................ 25
5.3.4 Interface Menu .............................................................................................. 29
6 Applications 38
6.5 Application multiplication factor weighing ....................................................................... 47
6.6 Application division factor weighing ............................................................................... 49
6.7.1 Density determination of solids ....................................................................... 52
6.7.2 Density determination of liquids ...................................................................... 53
6.7.3 Formulae used to calculate density ................................................................. 54
6.9.1 Repeatability test........................................................................................... 58
6.9.2 Display test .................................................................................................. 58
6.9.3 Key test........................................................................................................ 59
Table of Contents 1Jewelry Semi-Micro Balances
6.9.4 Motor test..................................................................................................... 59
6.9.5 Balance history............................................................................................. 59
6.9.6 Calibration history ......................................................................................... 60
6.9.7 Balance information ...................................................................................... 60
6.9.8 Service provider information ........................................................................... 61
7 Communication with Peripheral Devices 62
7.1 USB - interface and installation...................................................................................... 62
7.2.1 PC-Direct via USB ......................................................................................... 63
7.2.2 PC-Direct via RS232 ..................................................................................... 64
7.2.2.1 Installing SerialPortToKeyboard software .................................................... 64
7.2.2.2 Settings on the balance............................................................................ 64
8 Error and Status Messages 66
9 Maintenance 68
9.2 Cleaning the glass draft shield....................................................................................... 68
10 Technical Data 71
10.4.1 RS232C interface .......................................................................................... 74
10.4.2 USB device................................................................................................... 75
10.4.3 Aux connection ............................................................................................. 75
10.4.4 MT-SICS interface commands and functions ..................................................... 75
11 Accessories and Spare Parts 76
Index 81
Table of Contents2 Jewelry Semi-Micro Balances

1 Introduction

Thank you for choosing a METTLER TOLEDO balance. The balance combines high performance with ease of use.
This document refers to the initially installed firmware (software) version V 2.00.
Finding more information
u www.mt.com/jewelry
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
u www.mt.com/contact
Conventions and symbols
Refers to an external document.
Key and / or button designations are shown in graphic or bold text (e.g.
Note
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
Elements of instructions
Prerequisites
§
1 Steps 2 ...
ð Intermediate results
ð Results
For useful information about the product.
).
Introduction 3Jewelry Semi-Micro Balances

2 Safety Information

Two documents named "User Manual" and "Operating Instructions" are available for this instrument.
The User Manual shows you how to start using the instrument.
The Operating Instructions contain a full description of the instrument and its use.
Keep both documents for future reference.
Include both documents if you transfer the instrument to other parties.
Only use the instrument according to the User Manual and the Operating Instructions. If you do not use the instrument according to these documents or if the instrument is modified, the safety of the instrument may be impaired and Mettler-Toledo GmbH assumes no liability.

2.1 Definitions of signal words and warning symbols

Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following signal words and warning symbols:
Signal words
WARNING
A hazardous situation with medium risk, possibly resulting in death or severe injury if not avoided.
CAUTION
NOTICE
Warning symbols
Electrical shock
General hazard: read the Operating Instructions for information about the hazards and the resulting measures.
A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
A hazardous situation with low risk, resulting in damage to the instrument, other material damage, malfunctions and erroneous results, or loss of data.

2.2 Product specific safety notes

Intended use
This instrument is designed to be used in laboratories by trained staff. The instrument is intended for weighing purposes.
Any other type of use and operation beyond the limits of technical specifications without written consent from Mettler-Toledo GmbH is considered as not intended.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The instrument owner is responsible for the safety of all users of the instrument and third parties.
METTLER TOLEDO assumes that the instrument owner trains users to safely use the instrument in their workplace and deal with potential hazards. METTLER TOLEDO assumes that the instrument owner provides the necessary protective gear.
Safety Information4 Jewelry Semi-Micro Balances
Safety notes
WARNING
Danger of death or serious injury due to electric shock!
Contact with parts that carry a live current can lead to death or injury. 1 Only use the METTLER TOLEDO power supply cable and AC/DC adapter designed for
your instrument. 2 Connect the power cable to a grounded power outlet. 3 Keep all electrical cables and connections away from liquids and moisture. 4 Check the cables and power plug for damage and replace damaged cables and
power plugs.
NOTICE
Risk of damage to the instrument due to the use of unsuitable parts!
Using unsuitable parts with the instrument can damage the Instrument or cause it to malfunction.
− Only use parts from METTLER TOLEDO that are intended to be used with your
instrument.
NOTICE
Risk of damage to balance or software
In some countries, excessive mains voltage fluctuations and strong glitches may occur. This may affect the balance functions or damage the software.
− Use a voltage regulator for stabilizing.
Safety Information 5Jewelry Semi-Micro Balances

3 Design and Function

C
1
2
3
4
5
6
7
8
9
10
11
12
7
13 14
15
16
17
18

3.1 Overview

3.1.1 Components

Components overview legend
Display
1
Model sticker (with approved models only)
3
Draft ring
5
Handle/Coupling element for the operation of
7
Operation keys
2
Level indicator
4
Weighing pan
6
Glass draft shield
8
the draft shield doors
Handle for operation of the draft shield top
9
10
Top cover
door
Drip tray
11
Socket for AC/DC Adapter
13
Leveling feet
12
Aux (connection for "ErgoSens" or foot-
14
switch)
USB device interface
15
Kensington slot for anti-theft purposes
17
Design and Function6 Jewelry Semi-Micro Balances
RS232C serial interface
16
Product label
18

3.1.2 Operation keys

Weigh-Track
C
1 2 4 53 6
7
8
9
10
Terminal keys legend
No. Key Press briefly (less than 1.5 s) Press and hold (longer than 1.5 s)
1
2
3
4
5
6
To navigate back (scroll up) within
menu topics or menu selections
Decrease (numerical) parameters within
menu and in applications
To navigate forward (scroll down) within
menu topics or menu selections
Increase (numerical) parameters within
menu and in applications
Switch on
Zero/Tare
With entries: scroll down
To navigate through menu topics or
menu selections
To toggle between unit 1, recall value (if
selected), unit 2 (if different from unit 1) and the application unit (if any)
To change display resolution (1/10d
function) while application is running
Note: not available with approved models in selected countries.
To enter or leave menu selection (from /
to menu topic)
To enter application parameter or switch
to next parameter
To confirm parameter
To select the weighing application
Decrease (numerical) parameters
quickly within menu and in applications
To select assigned F1 application and
entering the parameter settings of appli­cation.
Default F1 application assignment:
Piece counting
Increase (numerical) parameters quickly
within menu and in applications
Switch off
To select assigned F2 application and
entering the parameter settings of appli­cation.
Default F2 application assignment:
Percent weighing
no function
To select assigned F3 application and
entering the parameter settings of appli­cation.
Default F3 application assignment:
Statistics
Design and Function 7Jewelry Semi-Micro Balances
No. Key Press briefly (less than 1.5 s) Press and hold (longer than 1.5 s)
Application Icons
Status Icons
Weight Value
Field
Weighing-in aid
Unit Field
Status Icons
7
8
9
10

3.1.3 Display

Printout display value
Printout active user menu settings
Transfer data
Execute predefined adjusting
(calibration) procedure
Enter or leave menu (Parameter
settings)
Save parameters
Cancel and to leave menu without
saving (one step back in the menu).
no function
no function
no function
no function
Application icons
Menu locked Application statistics
Menu setting activated Application totaling
Application weighing Application multiplication factor
Application piece counting Application division factor
Application percent weighing Application density
While an application is running, the corresponding application icon appears at the top of the display.
Status icons
Indicates stored value (Memory) Applications diagnostics and routine test
Adjustments (calibration) started Acoustic feedback for pressed keys
activated
FACT activated Weighing range 1 (Dual Range models
only)
Service reminder Weighing range 2 (Dual Range models
only)
Weight value field and weighing-in aid
Indicates negative values Brackets to indicate uncertified digits
(approved models only)
Design and Function8 Jewelry Semi-Micro Balances
Weight value field and weighing-in aid
Indicates unstable values Marking of nominal or target weight
Indicates calculated values Marking of tolerance limit T+
Unit field
gram
g
kilogram
kg
milligram
mg
carat
ct
pound
lb
ounce
oz
ozt
GN
dwt
mom
msg
tlh
Marking of tolerance limit T-
troy ounce
grain
pennyweight
momme
mesghal
Hong Kong taels
tls
tlt
tola
baht
Singapore taels
Taiwan taels
tola
baht
Design and Function 9Jewelry Semi-Micro Balances

4 Installation and Putting into Operation

WARNING
Danger of death or serious injury due to electric shock!
The instrument must be disconnected from the power supply before performing all setup and assembly work.

4.1 Unpacking and delivery inspection

NOTICE
Risk of damage to the instrument due to the use of unsuitable parts!
Using unsuitable parts with the instrument can damage the Instrument or cause it to malfunction.
− Only use parts from METTLER TOLEDO that are intended to be used with your instrument.
1 Open the packaging and carefully remove all components. 2 Check the delivered items.
The standard scope of delivery contains the following items
Balance with draft shield
Weighing pan with pan support
Draft ring
Drip tray
Protective cover
AC/DC adapter
Power cable (country specific)
Operating Instructions or User Manual; printed or on CD-ROM, depending on country of use
Declaration of conformity

4.2 Selecting the location

A good location will enable the balance to work accurately and reliably. The surface must be able to safely take the weight of the balance when fully loaded. The following local conditions must be observed:
If the balance is not horizontal at the outset, it must be leveled during commissioning.
Operate the balance in indoor environments only and at an altitude of less than 4000 m above sea
level.
Before switching on the balance, leave the parts to reach room temperature (+5 to 40°C).
Ensure that the relative humidity is between 10% and 80% and non-condensing conditions are met.
The power plug must be easily accessible.
Installation and Putting into Operation10 Jewelry Semi-Micro Balances
No powerful vibrations
C
1
2
3
No direct sunlight
No excessive temperature fluctuations.
No powerful drafts.
No powerful electric or magnetic fields
Ensure that the surroundings as free from dust as
possible.

4.3 Installing components

Push the side glass doors back as far as will go and place the following components on the balance in the specified order:
1 Place the drip tray (1) into the correct position. 2 Place the weighing pan (3). 3 Place the draft ring (2).

4.4 Connecting the balance

WARNING
Danger of death or serious injury due to electric shock!
Contact with parts that carry a live current can lead to death or injury. 1 Only use the METTLER TOLEDO power supply cable and AC/DC adapter designed for
your instrument. 2 Connect the power cable to a grounded power outlet. 3 Keep all electrical cables and connections away from liquids and moisture. 4 Check the cables and power plug for damage and replace damaged cables and
power plugs.
NOTICE
Danger of damage to the AC/DC adapter due to overheating!
If the AC/DC adapter is covered or in a container, it is not sufficiently cooled and will overheat.
1 Do not cover the AC/DC adapter.
The balance is supplied with an AC/DC adapter and a country-specific power cable.
Install the cables so that they cannot be damaged or interfere with operation.
Insert the power cable in a grounded power outlet that is easily accessible.
2 Do not put the AC/DC adapter in a container.
Installation and Putting into Operation 11Jewelry Semi-Micro Balances
1 Connect the AC/DC adapter to the connection socket
On
Off
0.00
g
on the back of your balance (see figure) and to the
power line. 2 Screw the plug tight to the balance. ð The balance is ready for use.

4.5 Setting up the balance

4.5.1 Switching on the balance

Before using the balance, it must be warmed up in order to obtain accurate weighing results. To reach operating temperature, the balance must be acclimatized and connected to the power supply for at least 60 minutes.
Switching on
− Press On.
ð The balance performs a display test. All
segments in the display light up briefly, WELCOME and software version. Maximum load and readability appears briefly. (Startup FULL mode only).
ð The balance is ready for weighing or for operation
with the last active application.
Legal-for-trade
Approved balances will execute an initial zero.

4.5.2 Leveling the balance

The balance has a level indicator and two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
The balance must be leveled and adjusted each time it is moved to a new location. To level it, proceed as follows: 1 Position your balance at the selected location. 2 Align the balance horizontally.
Installation and Putting into Operation12 Jewelry Semi-Micro Balances
C
È
3 Turning the two front leveling screws of the housing
until the air bubble is in the inner circle of the level indicator.
Example
Air bubble at 12 o'clock: turn both feet clockwise.
Air bubble at 3 o'clock: turn left foot clockwise, right foot counterclockwise.
Air bubble at 6 o'clock: turn both feet counterclockwise.
Air bubble at 9 o'clock: turn left foot counterclockwise, right foot clockwise.

4.5.3 Left/right operating of the glass draft shield

The glass draft shield of your balance can be adapted to the
environmental conditions and your weighing style, as well as to the
type of weighing and loading.
The position of the handles determines which door(s) of the draft
shield (left, right, or both) is/are opened.
Try various different combinations by moving the external handles
into the upper or lower position. We recommend you to set up the
glass draft shield so that it only opens on the side where the balance
is loaded. Your balance then works faster, because there are fewer
troublesome currents of air than when both doors of the draft shield
are opened together.

4.5.4 Adjusting the balance

To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at
its location and depending on the ambient conditions. After reaching the operation temperature, adjusting is
necessary.
before the balance is used for the first time.
when the balance was disconnected from the power or in case of power failure.
after a change of the location.
at regular intervals during weighing service.

4.6 Adjustment (calibration)

NOTICE
Before adjusting the balance, it must be warmed up.
Installation and Putting into Operation 13Jewelry Semi-Micro Balances

4.6.1 Fully automatic adjustment FACT

------
ADJ DONE
0.00
g
ADJ.INT
------
2000.00
g
ADJ DONE
0.00
g
C
0.00
g
------
C
ADJ.EXT
The factory setting is fully automatic adjustment FACT (Fully Automatic Calibration Technology) with the internal weight.
More information can be found in the section The Menu. The balance adjusts itself automatically:
after the warm-up phase on connection to the power supply.
when a change in the ambient conditions, e.g. the temperature, could lead to a noticeable deviation in
the measurement.
on a predefined time, see menu topic FACT.
time interval. (with OIML accuracy class II approved models).

4.6.2 Manual adjustment with internal weight

In the menu topic CAL (Adjustment) of advanced menu ADJ.INT must be selected.
§
Weighing pan is unloaded.
§
− Press to execute the internal adjustment.
ð The balance adjusts itself automatically. ð The adjusting is finished when the message ADJ
DONE appears briefly on the display. The balance
returns to the last active application and is ready for
operation.

4.6.3 Manual adjustment with external weight

Note
We recommend to disable FACT.
In the menu topic CAL (Adjustment) of advanced menu ADJ.EXT must be selected.
§
Required adjustment weight is ready.
§
Weighing pan is unloaded.
§
1 Press briefly to execute the external adjustment.
ð The required (predefined) adjustment weight
value flashes on the display.
2 Place adjustment weight in center of pan.
ð The balance adjusts itself automatically.
3 Remove adjustment weight, when 0.00 g flashes.
ð The adjusting is finished when the message ADJ
DONE appears briefly on the display. The balance
returns to the last active application and is ready for
operation.
Installation and Putting into Operation14 Jewelry Semi-Micro Balances

4.6.4 Customer fine adjustment

NOTICE
This function should be executed only by trained personnel.
The function customer fine adjustment ADJ.CUST.F allows you to adjust the value of the internal adjustment
weight with your own adjustment weight. The adjustable range of the adjustment weight is possible only in
a very small range. Customer fine adjustment impacts the function of internal adjustment. The customer fine
adjustment can be deactivated at any time.
Note
This feature is available on models with internal weight only.
Because of certification legislation, approved models cannot be adjusted with customer fine adjustment
(depending on selected countries' certification legislation).
Use certificated weights.
Balance and test weight have to be on operating temperature.
Observe the correct environmental conditions.
Installation and Putting into Operation 15Jewelry Semi-Micro Balances
Execute customer fine adjustment
200.0000
g
ADJ.DONE
0.00
g
------
------
g
g
200.0005
200.0005
C
C
ADJ.CUST.F
EXECUTE
SET REF.
ADVANCED
------
0.00
g
-
+
§
§
§
1 Select in the menu ADVANCED: ADJ.CUST.F
2 Confirm ADJ.CUST.F with
3 To carry out this operation select EXECUTE.
4 Start Adjustment with
5 Select the target adjustment weight.
6 Press and hold
7 Place required adjustment weight in center of pan. 8 Remove adjustment weight when zero is flashing.
9 Wait until ADJ DONE briefly appears.
ð The adjusting is finished when the message ADJ
ð If the error message WRONG ADJUSTMENT
The balance is under measuring condition.
Required adjustment weight is ready.
Weighing pan is unloaded.
.
.
ð SET REF. appears briefly.
ð The last saved value flashes on the display.
• For coarse setting, press
• For fine setting, press + to increase the value or
press - to decrease the value.
to confirm and execute
ADJ.CUST.F.
ð The required adjustment weight value flashes in
the display. This could take some time.
DONE appears briefly on the display. The balance
returns to the last active application and is ready for
operation
WEIGHT appears, the weight is not within the
allowed value range and could not be accepted.
ADJ.CUST.F could not be executed.
to change the value.
Storing the adjustment is not required.
Deactivate customer fine adjustment
1 Select in the menu ADVANCE.: ADJ.CUST.F.
2 Confirm ADJ.CUST.F with
3 To carry out this operation select RESET.
4 Start RESET by pressing
5 Select YES? and confirm with
ð The adjusting is finished when the message ADJ DONE appears briefly on the display. The balance
Note
ð NO? appears.
returns to the last active application and is ready for operation with initial adjustment.
.
.
.
Installation and Putting into Operation16 Jewelry Semi-Micro Balances

4.7 Performing a simple weighing

1250.38
0.00
1182.03
C
¢¢¢¢¢¢¢¢
¢¢¢¢¢¢
#
0.00
121.01
g
¢¢¢¢
#
0.01
g
0.00
g
C
C
121.01
g
0.00
g
0.00
g
95.97
g
#
C
The weighing application allows you to perform simple weighings and how you can accelerate the weighing process.
If your balance is not in the weighing mode, press and hold the
key down until WEIGHING appears in
the display. Release the key. Your balance is in the weighing mode and set to zero.
1 Press to tare the balance.
2 Place the sample on the weighing pan.
3 Wait until the instability detector
disappears and
the stability beep sounds.
4 Read the result.
Zeroing
Use the
zeroing key before you start with a weighing. 1 Unload the balance.
2 Press
to zero the balance.
ð All weight values are measured in relation to
this zero point.
Taring
If you are working with a weighing container, first set the balance to zero. 1 Place empty container on the weighing pan.
ð The weight is displayed.
2 Press
to set the balance to zero.
ð 0.00 g appears in the display.
3 Place weighing sample into the weighing container. ð The result appears in the display.
METTLER TOLEDO DualRange balances
METTLER TOLEDO DualRange balances have two ranges. These models have a fix fine (semi-micro) range between 0 g and maximum capacity, fine range. In this fine range the balance shows the result with a higher resolution, i.e. with one decimal place more.
Installation and Putting into Operation 17Jewelry Semi-Micro Balances
Switching weight units
0.78
oz
22.00
g
¢¢¢¢
*
22.00
g
¢¢¢¢
M
121.01
g
¢¢¢¢
0.00
g
0.00
g
M
121.01
g
0.00
g
C
C
¢¢¢¢¢¢¢¢
0%
100%
key can be used at any time to toggle between weight unit UNIT 1, RECALL value (if selected) and
The weight unit UNIT 2 (if different from weight unit 1) and the application unit (if any).
− Press to set weight unit or recall value.
Recall / recall weight value
Recall stores stable weights with an absolute display value bigger than 10d.
Function RECALL is in the menu activated.
§
1 Load weighing sample.
ð The display shows weight value and stores
stable value.
2 Remove weighing sample.
ð The display shows zero.
3 Press
.
ð The display shows last stored stable weight
value for 5 seconds together with asterisk (*) and memory (M) symbols. After 5 seconds the display goes back to zero. This can be repeated unlimited times.
Delete last weight value
As soon a new stable weight value is displayed, the old recall value becomes replaced by the new weight value.
− Press
.
ð The recall value is set to 0.
If the power is switched off, the recall value is lost. The recall value can not be printed.
Weighing with the weighing-in aid
The weighing-in aid is a dynamic graphic indicator which shows the used amount of the total weighing range. You can thus recognize at a glance when the load on the balance approaches the maximum load.
Print / transmit data
Press the
key to transmit the weighing results over the interface e.g. to a printer or a PC.
Installation and Putting into Operation18 Jewelry Semi-Micro Balances
Switching off
SHUTOFF
On
Off
C
− Press and hold the Off key until SHUTOFF appears
on the display. Release the key.
ð Balances switch into standby mode.
Note
After switching on from standby mode, your balance needs no warm-up time and is immediately ready
for weighing.
Standby mode is not possible with approved balances (only available in selected countries).
If your balance has been switched off after a preselected time, the display is dimly lit and shows date,
time, maximum load and readability.
If your balance has been switched off manually, the display is off.
To completely switch off mains operated balances, they must be disconnected from the power supply.

4.8 Transporting the balance

1 Press and hold the Off key.
2 Disconnect the balance from the power supply. 3 Disconnect all interface cables.
Transporting over short distances
To move the balance over a short distance to a new location, follow the instructions below.
NOTICE
Risk of damage to the glass draft shield
Glass parts on the balance can become damaged.
− Do not lift the balance by the glass draft shield. The draft shield is not sufficiently fastened to the balance to allow for this.
1 Hold the balance with both hands as shown. 2 Carefully lift the balance and carry it to its new
location.
Installation and Putting into Operation 19Jewelry Semi-Micro Balances
Transporting over long distances
1
2
We recommend using the original packaging for transportation or shipment of the balance or balance components over long distances. The elements of the original packaging are developed specifically for the balance and its components and ensure maximum protection during transportation.

4.9 Weighing below the balance

Your balance is equipped with a hanger for performing weighing operations below the work surface (weighing below the balance).
1 Press and hold the Off key.
2 Disconnect the balance from the power supply. 3 Disconnect all interface cables.
4 Push the side doors and the top door of the glass draft shield completely to the back. Important: top
cover must be closed. 5 Remove weighing pan, draft ring and drip tray. 6 Carefully tip the balance over backwards, until it is lying on its back. 7 Remove the cap (1) and retain it. The hanger (2)
for weighing below the balance is easily accessible
now. 8 Carefully turn the balance to its normal position and
reinstall all components in the reverse order.
Installation and Putting into Operation20 Jewelry Semi-Micro Balances

5 The Menu

5.1 What is in the menu ?

The menu allows you to match your balance to your specific weighing needs. In the menu you can change the settings of your balance and activate functions. The main menu has 4 different menus and these contains different topics, each of which allows you various selection possibilities. Menu item PROTECT see chapter Description of menu topics >
Main menu.
Menu BASIC
Topic Description
DATE
TIME
UNIT 1
UNIT 2
KEY BEEP
STAB.BEEP
RESET
Menu ADVANCED
Topic Description
WEIG.MODE
ENVIRON.
CAL
ADJ.CUST.F
FACT
FACT PRT.
DATE.FORM
TIME.FORM
RECALL
SHUTOFF
BCKLIGHT
DISPLAY
AUTOZERO
ZERO RNG
LANGUAGE
ASSIGN:F1
ASSIGN:F2
ASSIGN:F3
DIAGNOSE
SERV.ICON
SRV.D.RST
Setting the current date.
Setting the current time.
Specification of the 1st weight unit in which the balance should show the result.
Specification of the 2nd weight unit in which the balance should show the result.
Setting the key beep level.
Setting the stability beep level.
Call up of the factory settings.
Adapting the balance to the weighing mode.
Matching the balance to the ambient conditions.
Settings for the type of adjustment (calibration).
Executing customer fine adjustment.
Settings for fully automatic balance adjustment based on a selected time.
Switching the automatic FACT printout on or off.
Setting the date format.
Preselection of the time format.
Switching the application recall for storing stable weights on or off.
Setting the time after which the balance should be switched off automatically.
Setting the time after which the display backlight should be switched off automatically.
Adjusting the brightness and contrast of the display.
Switching the automatic zero correction (Autozero) on or off.
Setting the zero limit of the zero/tare key.
Setting the preferred language.
Selection of assigned F1 key application and entering their parameter settings.
Selection of assigned F2 key application and entering their parameter settings.
Selection of assigned F3 key application and entering their parameter settings.
Starting a diagnostic application.
Switching the service icon (service reminder) on or off.
Reset service date and hours (service reminder).
The Menu 21Jewelry Semi-Micro Balances
Menu INT.FACE
406.2
g
¢¢¢¢
ADVANCED
BASIC
TIME
DATE
Topic Description
RS232
HEADER
SINGLE
SIGN.L
LINE.FEED
ZERO PRT.
COM.SET
BAUDRATE
BIT/PAR.
STOPBIT
HD.SHAKE
RS.TX.E.O.L.
RS CHAR
USB
USB COM.S.
USB E.O.L.
USB CHAR
INTERVAL
ERGOSENS
Matching the serial interface RS232C to a peripheral unit.
Setting the header for printout of individual values.
Setting the information for printout of individual values.
Setting the footer for printout of individual values.
Setting line feed for printout of individual values.
Setting the auto print function for printing zero.
Setting the data communication format of the serial interface RS232C.
Setting the transfer speed of the serial interface RS232C.
Setting the character format (Bit/Parity) of the serial interface RS232C.
Setting the character format (stop bit) of the serial interface RS232C.
Setting the transfer protocol (Handshake) of the serial interface RS232C.
Setting the end of line format of the serial interface RS232C (outgoing data).
Setting the char set of the serial interface RS232C.
Matching the USB interface to a peripheral unit.
Setting the data communication format of the USB interface.
Setting the end of line format of the USB interface.
Setting the char set of the USB interface.
Selection of the time interval for the simulated print key press.
Settings for external key e.g. METTLER TOLEDO ErgoSens

5.2 Menu operation

In this section you will learn how to work with the menu.
Select menu
1 Press to activate main menu.
ð The first menu BASIC is displayed (except
2 Press
down/up + / – keys).
3 Press
Note
The menu selection BASIC, ADVANCED or INT.FACE can not be saved. The selection PROTECT must be saved.
Select menu topic
1 Press .
ð The next menu topic appears in the display.
2 Press
next menu topic.
3 Press – key to return to the previous menu topic.
menu protection is active).
repeatedly to change menu (Scrolling
to confirm the selection.
or the + key, the balance switches to the
The Menu22 Jewelry Semi-Micro Balances
Change settings in a selected menu topic
ENVIRON.
STABLE
ENVIRON.
+ 01.2500
+ 01.6500
+ 01.6500
SAVE:YES
SAVE:NO
The >> flashing symbol in the display indicates selectable options available.
1 Press .
ð The display shows the current setting in the
selected menu topic.
2 Press
next selection.
3 Press – key to return to the previous selection.
ð After the last selection, the first is shown again. 4 Press
For store the setting, see section Saving settings and
closing the menu.
Change settings in a submenu selection
The same procedure as for menu topics.
Input principle of numerical values
1 Press for input of numerical values. 2 Press
the application).
ð The selected digit or the selected value is
3 Press + to scroll up or – to scroll down for
changing digits or values. 4 Press
or the + key, the balance switches to the
to confirm the setting.
to select a digit or a value (depending on
blinking.
to confirm the input.
Saving settings and closing the menu
1 Press briefly to leave menu topic.
2 Press
ð Changes are saved.
3 Press
ð Changes are not saved.
4 Press
SAVE:NO.
Cancel
− Press C for leaving menu topic or menu selection
without saving.
Note
If no entry is made within 30 seconds, the balance reverts to last active application mode. Changes are not saved. If changes are made, the balance asks SAVE:NO.
to execute SAVE:YES.
to execute SAVE:NO.
to toggle between SAVE:YES and

5.3 Description of menu topic

In this section you will find information regarding the individual menu topics and the available selections.

5.3.1 Main menu

Selecting the menu.
BASIC The small BASIC menu for simple weighing is displayed.
The Menu 23Jewelry Semi-Micro Balances
Note
The menu selection BASIC, ADVANCED or INT.FACE can not be saved.
To activate PROTECT ON or OFF, this selection must be saved.

5.3.2 Basic menu

DATE – Date
Setting the current date according to date format.
Note
A reset of the balance will not change this setting.
ADVANCED The extended ADVANCED menu for further weighing settings is
displayed.
INT.FACE The menu INT.FACE for all interface parameter settings for
peripheral devices e.g. printer is displayed.
PROTECT
Menu protection. protection of balance configurations against unmeant manipulation.
OFF Menu protection is off. (Factory setting)
ON Menu protection is on. The menu BASIC, ADVANCED and
INT.FACE are not displayed. This is indicated with
in the
display.
TIME – Time
Setting the current time according to time format
+1H
Set the current time forwards by 1 hour (to adjust summer or winter time). (Factory setting)
-1H
Set the current time backwards by 1 hour (to adjust summer or winter time).
SET TIME
Enter the current time.
Note
A reset of the balance will not change this setting.
UNIT 1 – Weight unit 1
Depending on requirements, the balance can operate with the following units (depending on the model)
Legal-for-trade
Only those weight units allowed by the appropriate national legislation are selectable.
With approved balances, this menu topic has a fixed setting and cannot be changed.
Units:
g
kg
mg
ct
lb
oz
ozt
GN
1)
2)
3)
Gram
Kilogram
Milligram
Carat
Pound
Ounce (avdp)
Ounce (troy)
Grain
dwt
mom
msg
tlh
tls
tlt
tola
baht
Pennyweight
Momme
Mesghal
Tael Hong Kong
4)
Tael Singapore
Tael Taiwan
Tola
Baht
The Menu24 Jewelry Semi-Micro Balances
1)
factory setting
2)
not with 0.01 mg, 0.1 mg and 1 mg balances
3)
with 0.01 mg, 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
UNIT 2 – Weight unit 2
If it is required to show the weighing results in weighing mode in an additional unit, the desired second weight unit can be selected in this menu topic (depending on the model). Units see UNIT 1. Select NO, if you do not want to use UNIT 2.
Legal-for-trade
Only those weight units allowed by the appropriate national legislation are selectable.
KEY BEEP – Key beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the setting.
MED Medium level (Factory setting)
HIGH
OFF
LOW
High level
Beep switched off
Low level
STAB.BEEP – Stability beep
If the unstable symbol disappears, the stability beep becomes active. This menu topic allows you to preselect the volume of the stability beep.
LOW Low level (Factory setting)
MED
HIGH
OFF
Medium level
High level
Beep switched off
RESET – Reset balance settings
This menu topic allows you to call-up the factory settings.
To toggle between YES? and NO? press
Note
A reset of the balance will not change the DATE, TIME and ZERO RNG settings.

5.3.3 Advanced menu

WEIG.MODE – Weighing mode settings
This setting can be used to to adapt the balance to the weighing mode.
UNIVERS. For all standard weighing applications. (Factory setting)
DOSING
ENVIRON. – Environment settings
This setting can be used to match your balance to the ambient conditions.
STANDARD
UNSTABLE
(or + or –).
For dosing liquid or powdery products. With this setting, the balance responds very quickly to the smallest changes in weight.
Setting for an average working environment subject to moderate variations in the ambient conditions. (Factory setting)
Setting for a working environment where the conditions are continuously changing.
The Menu 25Jewelry Semi-Micro Balances
Loading...