A logbook of activities should be maintained for the above-mentioned product, recording all data
about events, adopted measures, manipulations etc. relative to the product, such as checking of
the shipment, initial installation, service / mainte nance, repair work, startup and shutdown etc.
This Instruction Manual is an integral part of the METTLER TOLEDO retractable housing InTrac78X
and contains notes and instructions that are important for safety and operation.
All persons working on or with the InTrac78X must have first read and understood the sections
appropriate to the work in hand.
Please read this instruction manual carefully before using the InTrac housing. Keep this document
close to the unit, so that operating personnel may easily be able to refer to it at any time.
Caution! Please first read Chapter 1 “Introduction” on page 9 and Chapter 2 “Important
a
notes for housing uses in Ex-Classified area” on page 22.
Proprietary designations
The following are proprietary names and, for the sake of simplicity, will be mentioned in this
instruction manual without the registration marking, e.g.
– InTrac, InPro, InFlow, EasyClean and InFit are registered trademarks of METTLER TOLEDO
Group.
®
– Viton
– All other trademarks are the property of their respective holders.
Use of warnings and symbols
DANGER! Warning of a dangerous situation that can lead to extensive material damage,
a
CAUTION!Warning of a possible dangerous situation that can lead to light bodily harm
a
ATTENTION! Information referring to technical requirements. Non-adherence can lead to mal-
h
Explanation of housing designations
and Kalrez® are registered trademarks of DuPont.
todeath or grave bodily injury.
and / or material damage.
functions, uneconomic working and possibly also to loss of productivity. The
manufacturer is not liable for any damages resulting from improper or inappropriate use.
®
.
The designation InTrac78X used in this Instruction Manual covers reference to the following housings:
– InTrac781 for pH / redox electrodes, O
PG13.5 thread).
– InTrac784 for pH / redox electrodes with liquid electrolyte e.g., InPro 2000 or 465 sensor series.
– The retractable housing InTrac78X is safe to operate. When used correctly, it represents no danger.
– Before starting to use the housing, carefully read this instruction manual: the safety precautions
and warnings contained in it must be observed.
– The retractable housing has been tested in-house and dispatched ready for installation.
In addition to this instruction manual please also note the following:
• all local safety regulations concerning the execution of pneumatic and water installations.
• all instructions and warning remarks in the publications of the products that are used in conjunction with the retractable housing (electrodes, sensors, controls, etc.).
• all safety precautions for the plant into which the retractable housing will be installed.
• all instructions and warnings labeled on the retractable housing.
• all safety information relative to operation in potentially explosive atmospheres / hazardous
areas (Ex classified zones).
The instruction manual contains the most important information for using InTrac78X efficiently
and in accordance with regulations. A basic condition for safe handling and operation without
malfunctions is a knowledge of these safety instructions and observance of the further warnings
in the instruction manual.
This instruction manual, and in particular the safety regulations, are intended for personnel entrusted with the operation and maintenance of the retractable housing. It is assumed that these
persons are familiar with the equipment in which the retractable housing is installed. Therefore,
before any work is started with the retractable housing, this instruction manual must be read and
understood by those persons involved.
The instruction manual must be stored where it is constantly accessible and available to any
person working with the InTrac78X.
On receipt of the shipment, check immediately:
• the retractable housing and accessories for any sign of transport damage. Report any damage
immediately to the carrier and to your supplier.
• the type designation on the housing body.
• for completeness of the supply. Please notify your supplier immediately if the shipment is
incomplete or in any way incorrect (see Chapter 3.1 “Scope of delivery” on page 28).
Housing designation, product identifier no. as well as article no. and serial no. for clear identification when communicating with the manufacturer can be noted from the type plate.
Note: The material number; (30261244) is the same for both InTrac781 and / or InTrac784
housings. However, the product Smart Configuration Key (SCK) and serial number (SNR)
are different for every retractable housing.
1.2 Basic principles
The retractable housing InTrac78X is built in accordance with state-of-the-art technology and
recognized technical safety regulations.
However, the housing can be a source of risks and dangers:
– if the housing is operated by underskilled or unauthorized personnel.
– if the housing is not used in compliance with regulations and / or stipulations for appropriate use.
InTrac78X may only be used in a technically correct condition for the purpose intended by the
supplier, with awareness by the user of safety and danger factors, taking the instruction manual
and local safety regulations into consideration.
Malfunctions and damage that can affect the safety and function must immediately be remedied
by the operator or an expert, and notified to the manufacturer in writing!
DANGER!A defective retractable housing must neither be installed nor put into operation.
a
Faulty containment and installation out of conformance with regulations and instructions can lead to the escape of medium or to pressure surges (explosion),
potentially harmful both to persons and to the environment.
The following symbols are used in this instruction manual to mark safety instructions:
DANGER! Warning of a dangerous situation that can lead to extensive material damage,
a
CAUTION!Warning of a possible dangerous situation that can lead to light bodily harm
a
ATTENTION! Information referring to technical requirements. Non-adherence can lead to mal-
h
todeath or grave bodily injury.
and / or material damage.
functions, uneconomic working and possibly also to loss of productivity.
1.4Responsibilities, organizational measures
1.4.1Responsibilities of operator
– The operator is under obligation only to permit persons to work with InTrac78X retractable
housings, who are familiar with the basic requirements of work safety and accident prevention,
and who have been instructed in the handling of the housing. This instruction manual serves
as the basic document.
– In addition to the instruction manual there are also generally valid legal and other binding
regulations for work safety and accident prevention as well as for environmental protection, and
these must be provided by the operator and instructed to personnel using the housings.
– The operator / user must be fully aware of safety and potential danger factors during work with
the housings and this awareness must be checked by the operator at regular intervals.
– Measures must be taken to ensure that the retractable housings are only operated in a safe and
fully functional condition.
– If the housings are employed in hazardous areas, compliance with prevailing regulations is to
be ensured.
CAUTION!Before the retractable housing is put into operation, the operator must have al-
a
ready clarified that use of the housing in conjunction with the other associated
equipment and resources is fully authorized.
1.4.2Responsibilities of personnel
– All persons whose duty it is to operate the retractable housings are under obligation to read
Chapter 1 “Introduction” on page 9 and Chapter 2 “Important notes for housing uses in
Ex-Classified area” on page 22 as well as the warning notices in this instruction manual.
– In addition to the instruction manual, generally valid legal and other binding regulations for
work safety and accident prevention must be adhered to.
– Any method of working which is doubtful from a safety perspective and which exceeds the
operation according to regulations must be omitted.
– Avoid using high-pressure water jets to clean any parts and components of the housing.
ATTENTION! Before every start-up, the retractable housing must be checked for:
h
• damage to the connections, fastenings, etc.
• leakage
• perfect functioning
• authorization for use in conjunction with other plant equipment and resources.
DANGER! A defective retractable housing may neither be installed nor put into operation.
a
Faulty containment or installation out of conformance with regulations and instructions can lead to the escape of medium or to pressure surges (explosion),
potentially harmful both to persons and to the environment.
– Work on or with the retractable housings may only be carried out by reliable and appropriately
trained or instructed personnel. The personnel must have read this instruction manual in advance.
– Clear responsibilities / standard operation procedures (SOPs) must be established for the per-
sonnel entrusted with operation, service, repair, etc. of the housings.
– It must be ensured that only specifically assigned personnel may operate the housings.
DANGER! Incorrect manipulation or operation of the housings or non-observance of safety
a
regulations can lead to problems with the housings and to the escape of process
medium, thereby presenting a potential hazard to the environment, personnel
and material.
1.5 Product-specichazards
1.5.1 Sensor design
DANGER! The electrode or sensor may only be removed from a mounted retractable housing
a
This can endanger personnel and cause damage to the environment and material.
ATTENTION! If an electrode / sensor is broken, it must be replaced immediately.
h
when this is in the “Service” (retracted / withdrawn / maintenance) position, since
if the electrode / sensor is missing from the housing, there is potential path for the
process medium to escape to the environment.
Toxic or aggressive media can pose a threat to human life through poisoning,
chemical burns or scalding.
DANGER! A broken sensor jeopardizes process safety, since it is not then possible to achieve
the visualized sensor interlocking bolts in the manual retractable housings
(InTrac78XM) makes it difficult for any unintentional removal of the electrode / sen-
sor from the housing when in the “Measure” position. This locking feature is basically intended as an added safety feature for the device. However, the operator / user
must be fully aware of safety and potential danger factors during work with the
housings and this awareness must be checked by the operator at regular intervals.
ing must be worn.
tractable housing is in the “Service” (retracted / withdrawn / maintenance) position.
Note: The sensor interlocking bolts indicate the operation mode
(for manual retractable housing – InTrac78XM model only)
ATTENTION! Before dismantling a retractable housing or commencing any maintenance work on
h
Manipulation of the housings may only take place after it is has been ensured that no process
medium can escape through the housing in the event of incorrect manipulation.
For this reason, the complete system must be emptied and vented in advance (safe condition).
If the retractable housing has been in operation in conjunction with the automated cleaning and calibration system EasyClean, it must be ensured that before any manipulation of the housing, the EasyClean system has been completely switched off and isolated (power, compressed air and water).
It is principally necessary to wear personal protective outfit such as protective goggles and clothing.
Only such maintenance and repair work as is specifically described in this instruction manual
may be performed on the retractable housing.
it, ensure that the equipment in which the retractable housing is installed is in a safe
condition (depressurize, no explosion risk, empty, rinse, vent, etc.). Retractable
housings may only be stripped down after having been completely dismounted.
Only original spare parts from METTLER TOLEDO may be used for replacing defective components.
DANGER! Non-compliance with the prescribed maintenance instructions can endanger per-
ATTENTION! – Retractable housings made of polymer material do not have the same me-
h
DANGER! Polymer housings may not be used in applications subject to high mechanical
a
chanical properties as steel housings.
– Retractable housings made of polymer need more frequent servicing than steel
housings.
– Defective components may only be replaced by an authorized service center.
stress, as this could lead to collapse of the protective caging. In such case, process liquid would be able to access the flushing chamber, resulting in process
medium mixing with cleaning medium. Please see Chapter 1.6.2 “Installation in
pressurized systems” on page 15 for recommended maximum permissible
temperature and pressure.
Maximum Housing Wetted Material Permissible Pressure and Temperature
PP
PVDH
PEEKStainless steel/
DIN1.4404/ AISI316L
DIN2.4602/ AISI Alloy C22
PVDF
1.6.2Installation in pressurized systems
ATTENTION! The maximum permissible temperature and pressure specifications must not be
h
DANGER! If temperature and pressure limits are exceeded, there is a risk to the integrity of the
a
exceeded. The values depend upon the design and type of the retractable housing.
The relative specifications are given on the type plate of the individual housings.
Please note that the maximum permissible pressure and temperature may differ
depending on the process medium. Please consult your local METTLER TOLEDO
representative for more information.
system, thereby presenting a potential threat to human life and to the environment.
1.7Installation in potentially explosive areas (hazardous areas)
ATTENTION!
h
– Please consult and follow the drawings and diagrams supplied with the measuring system.
– Where appropriate, it should be considered to include the METTLER TOLEDO retractable hous-
ing InTrac78X and the process connections in your recurring pressure test program for the
complete plant as a whole.
– Before the retractable housing is put into operation, the operator must have already clarified
beyond doubt that use of the housing in conjunction with the other associated equipment and
resources is fully authorized.
– Please note that for plastic type of retractable housing; not all metallic parts are electrically
connected. Hence, all metal part of retractable housing InTrac78X and the process connections
must be connected to the main potential equalization system of the plant (grounded). The
grounding cable (see drawing) has to be installed by the operator.
– Potential electrostatic charging hazard to the housing flange and wetted parts. Proper dis-
charge is needed before carrying out housing installation or after service, and / or removing
from the process.
– In special applications, the minimum conductivity of the medium (in contact with the housing
wetted part) shall be more than 1 nS / cm.
DANGER! Non-observance of legal regulations concerning use in hazardous areas can en-
a
ATTENTION! See also Chapter 2 “Important notes for housing uses in Ex-Classified area” on
ATTENTION! Despite all precautionary measures taken, residual hazards still remain.
h
1.8.1Rupture of pneumatic or hydraulic connections
DANGER! The escape of aggressive medium can present a threat to human life and to the
a
environment.
1.8.2Leakyconnections
– Connections can become loosened through the effects of vibration.
– Never remove the lower and top housing clamp screws during operation.
– Never remove the sensor / electrode when the housing is in the “Measure” position. Visually
check the housing position level and ONLY remove the sensor when it is fully retracted from the
process or in the “Service” position.
– The connection between housing and process adaptor is a potential source of leakage.
ATTENTION! The connections between the housing and the process adaptor must be checked
h
DANGER! Leaky connections can allow process medium, cleaning solution or control (com-
a
regularly by the customer / operator, and kept in full working condition.
pressed) air to escape to the environment, presenting a hazard for persons and
the environment.
DANGER! Before commencing any maintenance work on or removing an electrode / sensor
a
If the insertion rod comes to rest or is stuck between the defined “Measure” or
DANGER! The electrode / sensor may only be removed when the housing is in the “Service”
a
from a housing it is essential to visually check whether the housing is in the “Service”, “Measure” or any undefined intermediate position. The insertion rod must
always be stationed in one of the two defined position:
– “Measure” (operating position), that is, in the fully inserted position.
– “Service” (withdrawn position), that is, in the fully retracted position.
“Service” position, then no conforming measuring results can be ensured. Potential medium leaking to the flushing connection is possible. Proper isolation of the
process line, depressurized etc. is recommended before carrying out the repair / maintenance if the insertion rod is in any undefined position or other than
“Service” position.
position, as otherwise there is a risk that process medium could escape to the
environment. This can present a potential hazard to person, environment and
material. Toxic or aggressive media can pose a threat to human life through poisoning, chemical burns or scalding.
1.8.5Mediumresidues
DANGER! When retracting the insertion rod from the process, small quantities of process
a
DANGER! If the insertion rod is not completely inserted into the process, there will be a path
a
medium will remain attached to the electrode / sensor and will thus enter the flushing chamber. If this medium is a toxic or environmentally damaging substance,
then such contamination must be removed and disposed of in accordance with
regulations!
for the process medium to enter the flushing chamber and thus to the flushing
connection. This can lead to product loss or contamination.
1.8.6Manuallyoperatedhousings
DANGER! The electrode or sensor may only be removed when the housing is in the “Service”
a
position and as soon as when the manual drive train is manually turned (anticlockwise) until hand tight. As soon as the manual drive train is completely turned
anti-clockwise, the housing will automatically self-lock. After that, the manual
drive train cannot be further turned either clockwise or anticlockwise.
DANGER! Avoid excessive force or turning the manual drive train with any special tools.
ATTENTION! In the case of the pneumatically operated InTrac78X, failure of the compressed
h
DANGER! The electrode or sensor may only be removed when the insertion rod is fully
a
DANGER! Before the electrode / sensor is loaded into the pneumatic version InTrac78XR
a
air supply can result in the insertion rod being pushed to the “Service” position by
the process pressure. This makes continuous measurement impossible.
retracted from the medium. It is essential to visually check whether the housing is
in the “Service”, “Measure” or any undefined intermediate position. In addition,
visually check if the sensor mounting thread is fully retracted from the process
(see picture) before carrying out any maintenance work. Removing the sensor
from the housing when the housing is not fully retracted from the process can
result in process media escaping and presents a potential hazard both to persons
and to the environment.
or InTrac 78XI, it must be ensure that the pneumatic valves that control the
position of the housing are set to the “Service” position. Incorrect setting can
cause the housing to insert into the process before the loading procedure has
been completed, resulting in open access to the process medium. Process medium can thereby escape and present a potential hazard both to persons and to
the environment.
DANGER! Take applicable protection measures prior to touching the holder / housing as parts
a
of the retractable holder may have reached process temperature and cause burns.
1.8.9Externalinuences
ATTENTION! Objects falling on the housing can damage or destroy the unit, or cause leaks etc.
h
ATTENTION! Lateral forces may damage or destroy the unit.
h
1.9 Emergency measures
ATTENTION! Always observe and comply with local regulations!
h
ATTENTION! If the retractable housing has been in operation in conjunction with the automated
h
cleaning and calibration system EasyClean, it must be ensured that before any
manipulations at the housing, the EasyClean system has been completely
switched off and isolated (power, compressed air and water).
1.10 Safety measures
ATTENTION! Always observe and comply with local laws and regulations! These are not an
h
DANGER! It is principally necessary to wear protective equipment such as protective goggles
a
ATTENTION! The operator is responsible for the instruction of personnel. Additional copies of
h
The operator is obliged to inform the supplier / manufacturer of the retractable
integral part of this instruction manual.
and protective clothing.
this instruction manual can be ordered from the equipment supplier. As an integral
part of the retractable housing, this instruction manual must at all times be readily accessible to users at the point of operation of the housing.
housing immediately about any safety-relevant incidents, or observations made,
during use of the housing.
DANGER! Incorrect manipulation and / or instruction errors can lead to potential hazards for
ATTENTION! Before every start-up, the retractable housing must be checked for:
h
– damage to the connections, fastenings, etc.
– leakage
– defective cables and lines etc.
– authorization for use of the housing in conjunction with the associated plant
resources.
– In the ex-classified area installation, strictly adhere and follow the instructions
as specified in Chapter 1.7 “Installation in potentially explosive areas (hazardous areas)” on page 16 and Chapter 2 “Important notes for housing uses in
Ex-Classified area” on page 22.
DANGER! A defective housing must never be installed or put into operation. Poor contain-
a
ment, leaky connections etc. or non-compliant installation of the housings can
lead to escape of process medium and thereby to a potential threat to life (incl.
risk of explosion.).
1.11 Modications
ATTENTION! No attachments or modifications to the retractable housings are allowed.
h
DANGER! The manufacturer / supplier accepts no responsibility for any damage caused
a
by unauthorized attachments and alterations or for the incorporation of spare
parts which are not of METTLER TOLEDO provenance. The risk is borne entirely
by the operator.
These operating instructions contain all the information needed for safe and proper use of the
housing.
The operating instructions are intended for personnel entrusted with the operation and maintenance of sensors and housings. It is assumed that these persons are familiar with the equipment
in which the sensors and housings are installed.
See Chapter 1.3 “Warning notices and symbols” on page 11 for explanation.
combination electrodes as well as oxygen, CO2, conductivity or turbidity sensors. Use the
housings only for this purpose.
Housings with the
potentially explosive/hazardous areas.
Any use of these housings which differs from or exceeds the scope of use described in this instruction manual will be regarded as inappropriate and incompatible with the intended purpose.
The manufacturer/supplier accepts no responsibility whatsoever for any damage resulting from
such improper use. The risk is borne entirely by the user/ operator.
Other prerequisites for appropriate use include:
– Compliance with the instructions, notes and requirements set out in this instruction manual.
– Correct maintenance of the housings.
– Acceptance of responsibility for regular inspection, maintenance and functional testing of all
associated components, also including compliance with local operational and plant safety
regulations.
– Operation in compliance with prevailing regulations concerning the environmental and operat-
ing conditions as well as with the admissible mounting positions.
– Compliance with all information and warnings given in the documentation relating to the prod-
ucts used in conjunction with the housings.
– Correct equipment operation in conformance with the prescribed environmental and opera-
tional conditions, and admissible installation positions.
– Consultation with METTLER TOLEDO Process Analytics in the event of any uncertainties.
or symbol on the type plate have received approval for operation in
Important notes for housing uses in Ex-Classied area23
DANGER! The housing must be operated only with the specified electrodes/sensors. The
a
absence or the installation of an inappropriate electrode/sensor may adversely
affect the resistance to pressure and temperature, the chemical resistance and the
protection against explosion. Consequently, there can be leakage from the housing and/or risk of explosion that may endanger persons and the environment.
2.3 Safety instructions
DANGER!
a
– The plant operator must be fully aware of the potential risks and hazards at-
tached to operation of the particular process or plant. The operator is responsible for correct training of the workforce, for signs and markings indicating
sources of possible danger, and for the selection of appropriate, state-of-the-art
instrumentation.
– It is essential that personnel involved in the commissioning, operation or main-
tenance of these housings or of any of the associated equipment (e.g. sensors,
transmitters, etc.) be properly trained in the process itself, as well as in the use
and handling of the associated equipment. This includes having read and understood this instruction manual.
– The safety of personnel as well as of the plant itself is ultimately the responsibil-
ity of the plant operator. This applies in particular in the case of plants operating
in hazardous zones.
– The housings and associated components have no effect on the process itself
and cannot influence it in the sense of any form of control system.
– Maintenance and service intervals and schedules depend on the application
conditions, composition of the sample media, plant equipment and significance of the safety control features of the measuring system. Processes vary
considerably, so that schedules, where such are specified, can only be regarded as tentative and must in any case be individually established and verified by the plant operator.
– Where specific safeguards such as locks, labels, or redundant measuring sys-
tems are necessary, these must be provided by the plant operator.
– A defective housing must neither be installed nor put into service.
– Only maintenance work described in these operating instructions may be per-
formed on the housings.
– When changing faulty components, use only original spare parts obtainable
from your METTLER TOLEDO supplier (see Chapter 8.2 “Spare parts list” on
page 60).
– No modifications to the housings and the accessories are allowed. The manu-
facturer accepts no responsibility for damage caused by unauthorized modifi-
cations. The risk is borne entirely by the user.
– Care must be taken during installation to avoid impacts or friction that could
create an ignition source.
– Tampering and replacement with non-factory components may adversely affect
the safe use of the system.
– Insertion or withdrawal of removable electrical connectors or modules is to be
accomplished only when the area is known to be free of flammable vapors.
– WARNING – INTRINSICALLY SAFE APPARATUS CAN BE A SOURCE OF IGNITION
IF INTERNAL SPACINGS ARE SHORTED OR CONNECTIONS
OPENED.
24 Important notes for housing uses in Ex-Classied area
– WARNING – DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE IS PRESENT.
– WARNING – SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
– WARNING – SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY OF THE
EQUIPMENT
– WARNING – FOR CONNECTION ONLY TO NON-FLAMMABLE PROCESSES.
2.4Correct disposal of the housing
When the housing is finally removed from service, observe all local environmental regulations for
proper disposal. See Chapter 9 “Decommissioning, storage, disposal” on page 61.
2.5 Use in Ex-zones
ATTENTION!
a
For an installation in Ex-zones please read the guide-lines following hereafter:
1258
Ex-classificationATEX:
II 1/2G c IIC TX Ga/Gb
II 1/2D c IIIC TX Da/Db
Number of the test certificate:
SEV 13 ATEX 0161X
Ex-classificationFMapproved:
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 ºC
- 53 800 002; Entity
Project ID: 3021227
FM certificate number: FM16US0034X
FM18CA0021X
2.6Ex-classicationATEX
2.6.1Introduction
According to RL 94/9/EG (ATEX 95) Appendix l, InTrac 78X/*1/*2/*3/*4/*5/*6*7*8 housings are
devices group ll, category 1/2G and according to RL 99/92/EG (ATEX 137) may be used in zones
0/1 or 0/2 and gas group IIC that are potentially explosive due to combustible substances in the
temperatures classes T3 to T6.
For use/installation, the requirements of EN 60079-14 must be observed.
According to RL 94/9/EG (ATEX 95) Appendix l, InTrac 78X/*1/*2/*3/*4/*5/*6*7*8 housings are
devices group Ill, category 1/2D and according to RL 99/92/EG (ATEX 137) may also be used in
zones 20/21 resp. 20/22 that contain combustible dusts.
For use/installation, the requirements of EN 50281-1-2 must be observed.
The pneumatically operated housings with electrical (inductive) position indication of the probes
Important notes for housing uses in Ex-Classied area25
may be employed in hazardous areas, Zone 1 and Zone 2 resp. Zone 21 and Zone 22, also in
conjunction with separately certified, intrinsically safe, inductive proximity switches – e.g.
Pepperl+Fuchs, Types NCB2*** – on condition that the applicable Gas Groups and Temperature
Classes correlate with the actual inflammable materials present, and that the special conditions of
the declaration are strictly adhered to.
2.6.2Special conditions
The housings with pneumatic actuation position of the sensors with electrical feedback signal may
be operated in hazardous areas Zone 1 and Zone 2 or Zone 21 and Zone 22 with separately
certified intrinsically safe inductive proximity switches e.g. Pepperl + Fuchs types NCB2 ***– if the
gas groups and temperature classes coincide with the used flammable substances and the special conditions of the Certificates are observed.
1. The maximum permissible ambient or process temperatures for Zone 0 (flammable gases
or flammable liquids) shall be taken according to the following table:
Temperature classMax.environmentresp.mediatemperature
T668 °C
T580 °C
T4108 °C
T3130 °C
The maximum permissible ambient or process temperatures must not exceed the aforementioned
values and they will be found in this instruction manual, Chapter 8 “Product specifications” on
page 59.
2. The maximum permissible surface temperature for Zone 20 (combustible dust) shall be
taken accordingly to the following table:
Temperature classMax.environmentresp.mediatemperature
T 69 °C68 °C
T 81 °C80 °C
T 109 °C108 °C
T 131 °C140 °C
The maximum permissible ambient or process temperatures must not exceed the aforementioned
values and they will be found in this instruction manual, Chapter 8 “Product specifications” on
page 59.
3. The metallic body of the housing type InTrac 78X*1/*2/*3/*4/*5/*6/*7/*8 has to be connected conductively to the equipotential system of the plant.
4. The housings type InTrac 78X*1/*2/*3/*4/*5/*6/*7/*8 are included in the periodic pressure
testing of the system, where appropriate.
5. WARNING – POTENTIAL ELECTROSTATIC CHARGING HAZARD – by installation, use and
The packing consists of cardboard and foam plastic.
Keep the packing for possible later use, such as for storage or transport of the housing. However,
if you wish to dispose of the packing, please observe your local regulations in this respect.
ATTENTION! see Chapter 9 “Decommissioning, storage, disposal” on page 61.
h
3.3 Checkingtheshipment
When unpacking the shipment, examine carefully for signs of damage. Any damage noted must
be reported immediately to the carrier and your supplier. Check that the shipment accords to the
delivery papers and to your order.
ATTENTION! Damaged housings must not be installed or put into operation.
h
3.4Product overview
The InTrac78X retractable housings combine rugged design with great versatility to meet the
demands of the harshest process condition in the chemical, petrochemical, power and utilities
industries.
The retractable housing material is specially designed for a range of harsh applications. Wetted
parts are available in different materials offering installation flexibility in many applications.
The retractable housing meets major international requirements including installation in Ex-classified areas (ATEX and FM Approvals), where it can be safely installed.
The retractable housings are normally attached to tanks or piping with an applicable process connection (i.e. InFlow). In order to meet the various process requirements, the retractable housings
are available in different versions, depending on:
pneumatic drive train with inductive check back indicator (type I)
6Housing clamp screw
7Flushing connection “Out”
8Flushing connection “In”
9Process ange connection
10Insertion rod with sensor protection cage
1
3
8
10
3.6Description of function of retractable housings
With the retractable housings InTrac78X, electrodes / sensors can be cleaned or replaced without
interrupting the ongoing process. In addition, with the aid of auxiliary equipment (EasyClean
system) it is possible to calibrate and / or backwash (when sensor retracted) and / or flush with the
electrode / sensor installed.
The manual version of InTrac78X equipped with the manual drive train. The manual drive train is
a mechanical rotary drive that converts a rotating motion into forwards / backwards movement of
the insertion rod. The sensor can be retracted from the process liquid to the cleaning chamber (for
maintenance) or likewise. State of the art drive train design allows operators to drive the sensor
against high process pressure without the needs to isolate the complete process line. Thanks to
manual drive train, the disturbance to the process is minimal.
To retract the sensor from the process, the “Measure” sensor interlocking bolt is pressed. Manually rotate the manual drive train anticlockwise until the “Service” sensor interlocking bolt pops
out. In this position it is safe to perform sensor maintenance. The reverse process is applied for
inserting the sensor back for operation.
On top of the sensor interlocking bolts, the level marking is visible on Sensor Position Rod (for
InTrac781M) or Sensor Pressurized Cylinder (for InTrac784M) to provide additional information
about the sensor status (either it is in “Measure” or “Service” position). It is necessary to ensure
the sensor position rod / sensor pressurized cylinder has fully reached the “Service” level marking
before carrying out any sensor maintenance.
ATTENTION! The sensor pressure indicator (InTrac784X) indicates the sensor’s internal pressure
h
(i.e. InPro2000 sensor). To ensure the sensor electrolyte continues to flow freely
from the sensor junction during operation, sensor internal pressure must be higher
than the process pressure. Please refer to sensor instruction manual for detail.
The compressed air connections are used to pressurize the piston of the insertion rod either from
above or from below. This allows the insertion rod and hence the electrode / sensor to be moved
either downwards (into the “Measure” position) or upwards (into the “Service” position). The piston remains pressurized with compressed air at the respective end position.
In housing versions equipped with a position indication system, a pneumatic (InTrac78XR) or
pneumatic with inductive (InTrac78XI) signal is triggered when the insertion rod is in the respec-
tive end position (“Measure” or “Service”). This signal is transmitted to a remote indicator (not
included with housing).
On top of the sensor position signal, the level marking both “Measure” and “Service” are visible on
the Sensor Position Rod and Sensor Pressurized Cylinder to provide additional sensor status information. It is necessary to ensure that the sensor position rod / sensor pressurized cylinder has
fully reached the “Service” level marking before carrying out any sensor maintenance.
Depending on the remote programmable logic controller (PLC) and the operating system (e.g.,
EasyClean), the InTrac78X (type R or I) is programmable for sensor backwash cleaning and
while the sensor is retracted from the process, for cleaning, or calibration with buffer (pH sensor).
ATTENTION! The housings are equipped with an intelligent sensor locking system: if no elec-
h
trode / sensor has been fitted in the housing, it is not possible to retract / send back
the insertion rod into the process.
CAUTION! The insertion and retraction speed of the insertion rod can be set as desired by
means of the adjustment screw on the throttle valve and by adjustment of the air
supply pressure. Please note that for proper operation the compressed air supply
must be at a higher pressure than the process pressure.
Page 35
Product descriptions 35
ATTENTION! The compressed air must be free of oil, dust and water. The recommended com-
h
CAUTION! For backwash cleaning, the cleaning liquid (from the flushing connection “IN”)
a
ATTENTION! The cleaning liquid pressure must not exceed 6bar. Install a pressure reducer if
h
ATTENTION! The sensor pressure indicator (InTrac784X) indicates the sensor internal pres-
h
DANGER! Process medium can present a hazard to human life and to the environment.
a
pressed air pressure is between 4to6bar.
must supply higher pressure than the process pressure. Insufficient cleaning liquid pressure may result in process media entering the flushing connections.
necessary.
sure. To ensure the sensor electrolyte continues to flow freely from the sensor
junction during operation, sensor internal pressure must be higher than the process pressure. Please refer to sensor instruction manual for details.
The retractable housings are mounted and fixed on a vessel (reactor, tank, pipe, etc.) either with flange
connection, or via a NPT thread. Various DIN or AISI nominated process connections are available.
ATTENTION! Attachment / gasket between the flange connection and / or threaded seal PTFE
h
CAUTION! In the case of inclined mounting, the angle of the housing must be equal to or
h
tape for the NPT connection (if applicable) is the responsibility of the customer.
In order to ensure optimum performance of the retractable housings at all time,
please pay attention to the following installation instructions:
– The retractable housing can be mounted vertically or in an inclined position.
greater than 15° above the horizontal.
admissible installation
– The retractable housing is to be mounted in such a position that there is always enough clear-
ance available for its correct functioning (correct “Measure” position in the sample medium) as
well as for maintenance work (checks, fitting and removal of the electrodes / sensors). The
relative dimensions can be seen on the drawings in the relative appendix to this instruction
manual, or in the specifications. It is recommended to have clearance of at least 0.5 to 1 m for
maintenance work purposes.
– Mounting of the retractable housing in exposed positions should be avoided. If this is not pos-
sible, relative measures to protect against damage or interference must be taken.
– Outdoor installation of the retractable housing is possible. However, avoid direct sunlight, rain,
water and / or excessive heat. The working ambient temperature is –10…70 °C.
ATTENTION! For outdoor installation of the InTrac78XM housing, the sensor interlocking bolts
h
should be installed facing a position where water splashes (when happen) shall
not ingress into the drive train via the sensor interlocking bolts.
CAUTION! In all installation work described below, ensure that the equipment in which the
a
CAUTION! For Ex-environment installation: Potential electrostatic charging hazard to the housing
a
housing is to be installed is in a non-hazardous condition (depressurized, empty,
rinsed, vented, etc.).
flange and wetted parts. Proper discharge is needed before carrying out the housing
installation or after service, and / or removing from the process. See Chapter “1.7 Installation in potentially explosive areas (hazardous areas)“ on page 16 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22.
4.2.1Fitting the housing
Choose the installation point carefully to ensure the sensor is always filled with process medium
even when sample flow is interrupted. Any air bubble in the medium may perturb the measurement. Consider installing proper air venting upstream of the housing.
CAUTION! The sensor protective cage (in “Measure” position) must have a clearance of at
a
least 3 – 5cm from the vessel / pipe / tank etc., and wall in the installation. Refer
respective housing insertion depth dimension for detail.
4.2.1.1 Fittingviaaange
1. Clean the sealing surfaces of the flange (housing and flange connection on vessel) and
check for damage.
CAUTION! If the process media / reaction products are considered to be dangerous, it is im-
a
2. Use the appropriate flange gasket which is suitable for the process condition. Replace if necessary.
ATTENTION! In the case of housings made of special materials:
h
Check that the seal is present on the flange of the retractable housing and is not
3. Check the vessel to ensure that there are no obstacles in the insertion direction which could
hinder the motion of the insertion rod.
4. Position the housing on the flange connection, align, and tighten evenly crosswise using the
prescribed number of bolts and nuts.
perative that an embedded seal is used at the flange interface and / or a splash guard
mounted. Fitting of the retractable housing with damaged flange connections is not
allowed and can present a hazard to persons and / or lead to material damage.
damaged. Replace if necessary. Ensure correct quality and positioning of the seal.
4.2.1.2 Fitting via NPT external thread
1. Ensure the threads are free of particles or dirt.
2. Wind PTFE tape (or appropriate sealing tape which is suitable for the process condition)
around the external male thread.
3. Screw the housing carefully into the female socket.
1. Clean the centering spigot of the retractable housing and the bore of the weld-in socket,
checking for damage
CAUTION! Fitting of the retractable housing with a damaged spigot or into a damaged weld-
a
2. Check the vessel / tank / piping etc. to ensure that there are no obstacles in the insertion direction which could hinder the motion of the immersion tube.
3. Check the O-ring on the spigot for damage and replace it if necessary. Lightly smear the
O-ring with grease. Ensure correct quality and positioning of the O-ring.
4. Position the housing on the weld-in socket and carefully insert the spigot into the bore.
5. Finally, tighten the ring nut until the connection is completely sealed.
CAUTION! Checking of the sealing / tightness of the process adaptor is the responsibility of
a
in socket is not allowed and can present a hazard to persons and / or lead to
material damage.
the operator, who must also guarantee same by adopting appropriate measures.
Additional safety provisions are necessary if the connection is subject to stress
from vibrations.
4.2.2Adjustingthesensorprotectivecage
A sensor protective cage is fitted to the lower end of the insertion rod. For visual inspection, a symbol is visible on the housing body and parallel to the sensor protective cage opening (see picture).
For maximum sensor tip protection against any particles in the medium, the sensor protective
cage can be adjusted in respect of the medium flow.
1. Make sure the housing is fully retracted (or in “Service” position)
2. Loosen the housing clamp screws “1”.
3. Turn the cylinder drive train clockwise / anticlockwise until the preferable sensor protective
cage position has been reached.
4. Tighten the housing clamp screws “1”.
DANGER! Before loosen the housing clamp screws, the pipe / vessel / tank etc must be iso-
a
lated (depressurize, empty, rinse, purge and vent, etc.).
The InTrac781 / 784 is equipped with “In” and “Out” flushing connections. The flushing connection
labelled “In” is the inlet for the sensor cleaning liquid (i.e. clean water). The flushing connection labelled
“Out” is the outlet for the cleaning medium drain out from housing during the sensor cleaning process.
To achieve the optimum sensor cleaning, the cleaning liquid inlet pressure is recommended to be at
least 1 bar. The cleaning liquid pressure must not exceed 6bar. Install a pressure reducer if necessary.
During the insertion or retraction motion, if the insertion rod comes to rest in other than the defined
end positions “Measure” and “Service”, process media can escape to the environment via the
flushing chamber and the flushing lines. This could create a hazard for personnel, equipment and
the environment.
When constructing the flushing water installation, use only materials which are resistant to corrosion and chemicals. The cleaning medium must be considered beforehand. Ensure the cleaning
medium will not react with the process medium or harm the housing wetted material.
Recommendations for proper operation and safe use
If cleaning of the sensor is not desired, the flushing connections (both “In” and “Out”) must be
sealed by pegs, or pipe plugs as leakage of process liquid through the open flushing connections
is possible. lf flushing lines must be equipped with check valves which may only be opened when
the insertion rod is in the retracted position (“Service”). After maintenance work on the sensors,
the valves must be closed.
Installing pipe plugs/ pegs to flushing connection
When ordering InTrac781/ 784 housings, flushing connections ‘[G14] – G¼" thread female’ or
‘[N14] – ¼" NPT female’ [1] are delivered together with the housing. In cases where cleaning of
the sensor is not desired, the pipe plugs / pegs (sourcing locally) can be directly installed to the
flushing connections [1]. Alternatively, the flushing connections [1] can be removed by turning
1
them anti-clockwise, and the pipe plugs / pegs G
In cases where a flushing connection ‘[G18] – G
⁄8" can be directly installed to [2].
1
⁄8" thread female’ is ordered, flushing connections
[1] are not included in housing delivery. Flushing connections can be directly installed to[2].
Ifcleaning of the sensor is not desired, the pipe plugs / pegs G1⁄8" can be directly installed to [2].
1
METTLER TOLEDO offer G
⁄8" pipe plugs in different material. See spare part list for detail.
In cases where flushing connection ‘[00_] – Without (blind plugs included)‘ is ordered. Two blind
1
plugs G
ATTENTION! Avoid excessive force apply when installing pipe plugs / pegs to the housing.
h
⁄8" are directly installed to [2] in the housing delivery.
Excessive forces can irreversibly damage the housing thread.
1. Install appropriate valves i.e. ball valve (and / or check valve if applicable) upstream of the
flushing connection “In” and downstream of flushing connection “Out”. Note: The valves are
not in the scope of delivery.
2. Connect the cleaning line to the flushing connection “In”.
3. Connect the drain line to the flushing connection “Out”.
4. Check tightness of the piping, valves connections.
5. When motorized valves are preferred, connect the cable to the operation system accordingly.
Do the operating system set-up (if applicable).
2
1
DANGER! When retracting the insertion rod from the process, small quantities of process me-
a
dium will remain attached to the electrode / sensor and will thus enter the discharge
line during flushing procedures. If the process medium contains toxic, caustic substances damaging to the environment, then it is imperative to consult local regulations relative to the design and construction of effluent (treatment) installations.
4.2.3.1 Normal sensor cleaning
When the flushing lines are properly installed, the sensor can be retracted from the process and
remains in the housing while it is cleaned with cleaning liquid. It is not necessary to dismantle the
sensor and clean it manually in the lab.
1. Retract the sensor from the process until the housing is in the “Service” position.
2. Open both valves located upstream and downstream of the flushing connections “In” / “Out”.
3. Let the cleaning liquid clean the sensor for 2 minutes (or longer if appropriate).
4. Close both valves.
5. Insert the sensor into the process.
ATTENTION! The cleaning liquid pressure must not exceed 6 bar. Install a pressure reducer
For special application where particles and dirt are present in the process medium, backwash
cleaning is recommended. As soon as the housing start to retract from the process, backwash
cleaning flushes out the dirt / particles which stick on the sensor tip / sensor protective cage. This
allows the majority of the dirt to be flushed out into the process.
1. A valve upstream of the flushing connection “In” is open but the valve downstream of the
flushing connection “Out” remains closed.
2. Retract the sensor from the process.
3. When the sensor is fully retracted (or in “Service” position), open the valve downstream of
the flushing connection “Out”.
4. Let the cleaning liquid clean the sensor for 2 minutes (or longer when appropriate).
5. Close both valves.
6. Insert the sensor into the process.
CAUTION! For backwash cleaning, the cleaning liquid (from the flushing connection “IN”)
a
ATTENTION! The cleaning liquid pressure must not exceed 6 bar. Install a pressure reducer
h
must supply higher pressure than the process pressure. Insufficient cleaning liquid pressure may result in process media entering the flushing connections.
ifnecessary.
4.2.4Installation for calibration of the measurement system with
electrode/sensorinstalled
If the measurement system has to be cleaned and / or calibrated with the electrode / sensor
installed, a distributor with three connections is attached to the flushing connection “In”.
Make three connections as shown in the following schematic diagram:
B
A
Note! When the housing is operated with EasyClean EC400, it is recommended to order the
h
InTrac housing with G ¼" thread female flushing connection.
C
• Connection “A” to the cleaning line inlet
via a control valve.
• Connections “B” and “C” to the corresponding buffer solution supply, each
via a non return valve.
The InTrac78XR is operated with compressed air. The compressed air is used to drive the drive train of
the housing, so the sensor can retract or insert into the process. The extension of the cylinder of the drive
train is fitted with four compressed air connections. For proper connection, the items below are required.
– Two pneumatic tubes with outter diameter 4 mm
– Two pneumatic tubes with outter diameter 6 mm
The four compressed air connections are labelled numerically as below.
– “1”: Air supply for “Service” position; pneumatic tube with outter diameter 6 mm
– “2”: Air supply for “Measure” position; pneumatic tube with outter diameter 6 mm
– “3”: “Service” position reply signal; pneumatic tube with outter diameter 4 mm
– “4”: “Measure” position reply signal; pneumatic tube with outter diameter 4 mm
1
1
4
2
3
ATTENTION! – The pressure in the control air supply to the housing must reach 4 bars.
h
– The compressed air must be free of oil, dust and water. The recommended
compressed air pressure is between 4 to 6 bar.
– When no reply signal is required, the compressed air connections “3” and “4”
must be properly sealed.
CAUTION! The insertion and retraction speed of the insertion rod can be set as desired by
a
means of the adjustment screw on the throttle valve and by adjustment of the air
supply pressure. Please note that for proper operation the compressed air supply
must be at a higher pressure than the process pressure.
The InTrac78XI is operated with compressed air. There are two compressed air connections, and
two inductive check back sensor connections. The inductive check back sensor is installed to the
housing prior to shipment. The connections are labelled numerically as below:
– “1”: Air supply for “Service” position; pneumatic tube with outter diameter 6 mm
– “2”: Air supply for “Measure” position; pneumatic tube with outter diameter 6 mm
– “3”: “Service” position inductive check back sensor
– “4”: “Measure” position inductive check back sensor
The compressed air is connected to connection “1” and “2” respectively via pneumatic tubes.
Two cable sockets (M12, 4-pin, NAMUR) are required to connect the inductive check back sensor
to the process control. The inductive check back sensor’s specification are as below:
Switching element functionNAMUR, NC
Nominal voltage, U
Switching frequency, f0…2000 Hz
Current consumptionMeasuring plate not detected ≥ 3 mA
Connection typeDevice connector M12
0
8.2 V (Ri approx. 1 k Ohm)
Measuring plate defected ≤ 1 mA
× 1, 4-pin
Connection:
Wire colors in accordance with EN60947-5-6
– Terminal 1 = Brown
– Terminal 2 = Blue
For EasyClean 200 e only:
For installation in ex-classified area, additional sensor output interface terminals (2 pcs.) are re-
quired to connect the inductive check-back from sensor (from InTrac 78XI housing) to the EasyClean 200 e system. The EasyClean 200 e system and the sensor output interface terminals must
be installed outside the ex-classified area. The sensor output interface terminals (not in scope of
delivery) can be purchased from Pepperl + Fuchs
®
; model number KCD2-E2L.
Please consult your local representative for more detail.
ATTENTION! – The pressure in the control air supply to the housing must reach 4 bars.
h
– The compressed air must be free of oil, dust and water. The recommended
compressed air pressure is between 4 to 6 bar.
CAUTION! The insertion and retraction speed of the insertion rod can be set as desired by
a
means of the adjustment screw on the throttle valve and by adjustment of the air
supply pressure. Please note that for proper operation the compressed air supply
must be at a higher pressure than the process pressure.
1. Set the housing to the “Service” position. Visually check if the housing is in “Service” position
(level mark from sensor position rod; “3”). For InTrac781M, it is possible to check if the
sensor locking bolt is in the “Service” position.
2. Unscrew the cable antikink “1”.
3. Unscrew the cable grommet “2”.
4. Unscrew the sensor position rod “3”.
5. Check that the correct type of electrode / sensor has been selected (see Chapter 10.1 “Electrode / sensor selection” on page 62). For optimum sensor insertion length, the sensor
washer must be of the range 5 to 7 mm. Sensors with a shorter washer must use the sensor
spacer (included in the delivery).
6. Check the electrode / sensor for damage (e.g. breakage of electrode).
CAUTION! A damaged electrode / sensor must never be installed.
a
7. Check washers and O-rings on electrodes / sensors and replace if necessary.
8. Remove watering cap from the tip of the electrode / sensor and rinse electrode / sensor with water.
9. Carefully insert electrode / sensor into the immersion rod and screw in by hand as far as it
will go.
DANGER! Under no circumstances whatsoever may tools be used.
a
10. Remove the cover cap of the electrode / sensor screw cap connection.
11. Run the sensor cable (open end) through the sensor position rod “3”, cable grommet “2”,
and antikink protection “1”. Connect the plug to the electrode / sensor.
12. Mount sensor position rod “3”, follow by cable grommet “2” and tighten by hand. Finally,
also hand-tighten the cable antikink protection “1”.
13. Connect the sensor open end to the transmitter.
side up” points upwards. In this position no reference electrolyte can flow out
through the filling port (assuming that the fill level is correct).
Page 46
46 Installation and start-up
CAUTION! Ring nut “5” of upper part may be tightened only with plastic adapter “4”
a
10. Check condition of seal of plastic adapter “4” and replace if necessary. Mount plastic
adapter “4” and tighten by hand.
11. Remove the cover cap of the plug connection of the electrode.
12. Run the sensor cable (open end) through the cable grommet “3” and antikink protection
“2”. Connect the plug to the electrode.
13. Mount cable grommet “3” and tighten by hand. Finally, also hand-tighten the cable antikink
protection “2”.
14. Connect the sensor open end to the transmitter.
15. Set the compensation pressure in the housing. The compensation pressure can be set with
a bicycle pump via the valve assembly at the pressure gauge, or established by attaching a
permanent (dry, oil-free and filtered) compressed air supply (using the pressure connection
set included in the standard equipment supply).
ATTENTION! To ensure electrolyte flow from the reference electrode to the sample medium the
h
removed.
air pressure in the upper part of the housing must be at least 0.5 bars and maximum 2 bars above that of the sample medium (take the hydrostatic pressure of
the sample medium into account).
ATTENTION! If there is no (positive) pressure differential to the sample medium, no reliable
h
measurements can be taken.
4.3Startup procedures for housings
4.3.1Startup of manually operated housings
Before startup, all fitting and installation work (see Chapter 4.2 “Fitting and installation work” on
page 37) must have been completed!
Each time before startup, check the measuring system. Inspect the electrode / sensor assembly
and examine for leaks from housing and apparatus (see also Chapter 7 “Troubleshooting” on
page 57). Do not commence operation until the measuring system has been checked and any
necessary corrective action taken.
Before startup of a housing in hazardous area, it is to be clarified beyond doubt that it is permitted
to use the housing in question in conjunction with the other associated plant resources. Read and
fully understand Chapter 1.7 “Installation in potentially explosive areas (hazardous areas)” on
page 16 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22
before carrying out the start-up.
Proceed as follows for startup:
1. For InTrac78XM start-up, press the “Service” interlock bolt continuously and turn the manual drive train clockwise carefully.
2. The “Measure” interlock bolt pops up when the housing is fully inserted into the process.
ATTENTION! Only turn the manual drive train by hand. No excessive force should be applied
Before startup, all fitting and installation work (see Chapter 4.2 “Fitting and installation work” on
page 37) must have been completed!
Each time before startup, check the measuring system, inspect the electrode / sensor assembly
and examine for leaks from housing and apparatus (see also Chapter 7 “Troubleshooting” on
page 57). Do not commence operation until the measuring system has been checked and any
necessary corrective action taken.
Before startup of a housing in a hazardous area, it must be established beyond doubt that it is
permitted to use the housing in question in conjunction with the other associated plant resources.
Read and fully understand Chapter 1.7 “Installation in potentially explosive areas (hazardous
areas)” on page 16 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on
page 22 before carrying out the start-up.
Proceed as follows for startup:
1. Open the air supply for the control system. Check:
– the insertion and retraction motion of the insertion rod, as well as the status of the indica-
tors for the respective end positions.
– the control air system for tightness. Immediately seal any leaking connections.
2. Drive the insertion rod into the “Service” position and open flushing water supply (at least
1bar). Check flushing water system for tightness. Immediately seal any leaking connections.
3. If the system is equipped for calibration of the measuring system, move the housing into the
“Service” position and perform a calibration (see Chapter 5.4 “Calibrating the measuring
system” on page 52). Check:
– for correct calibration procedure.
– the buffer solution systems for tightness. Immediately seal any leaking connections.
On successful completion of all function checks, the equipment in which the housing is installed
can be started up.
CAUTION! Following startup of the equipment, particular care must be taken in any manipu-
a
lations at the retractable housing. Therefore it is essential to pay close attention to
the information in Chapter 5.1 “Important information for everyday operation” on
page 51.
4.4Shutdown procedure for manually and
pneumatically operated housings
1. Move housing to the “Service” position.
2. Open flushing water supply, flush electrode / sensor and close flushing water supply again.
3. Isolate the pipe / tank / vessel etc, from the housing (depressurize, empty, rinse, purge and
vent, etc.).
4. Close control air feed to the housing (except InTrac78XM).
The retractable housing can be employed in conjunction with METTLER TOLEDO’s cleaning and
calibration system, EasyClean. This enables full automation of the measuring system. Current
regulations together with the connection instructions given in the relative Instruction Manual, as
well as directives governing pneumatic and water connections are to be fully observed at all times.
CAUTION! In the case of installations intended for operation in hazardous areas, only Easy-
a
Clean 400X systems may be used in conjunction with a InTrac78X housing.
Distributed Control System (DCS)
Transmitter M700 (X)EasyClean 400 (X)
Pressurized water
Pressured air
Media adapter
ith dosing pumps
for calibration buffer
and cleaner (3.5 l)
CAUTION! No attempt may be made to move the insertion rod into the “Measure” position
a
During operation,
– never loosen the lines of the pneumatic system or of the flushing and buffer solutions.
– never remove fastening components (screw / bolts of flange, ring nut, etc.).
If any malfunctions occur during operation, the equipment in which the housing is installed must
first be made safe before any corrective measures are taken.
For all work on the equipment during everyday operation, wear the stipulated protective clothing
(protective goggles, gloves, breathing apparatus, etc).
5.2 Inspectionworkineverydayoperation
without an electrode / sensor properly installed!
The following inspection work should be performed in everyday operation:
• Check all systems such as air, flushing water and buffer solutions for leaks. Refill buffer solutions if necessary.
• Visually check fastenings (ring nut, flange, NPT-thread) of the housing at the vessel for firm
seating and leaks.
• Check the condition of the electrode / sensor. A faulty or damaged electrode / sensor must be
replaced immediately.
• Check the inspection window for leaking (picture).
• Check air pressure in upper part (pressure gauge). The air pressure must be at least 0.5 bars
and maximum 2 bars above that of the sample medium (take hydrostatic pressure of the
sample medium into account) to ensure the flow of electrolyte from the reference electrode to
the sample medium.
ATTENTION! The desired overpressure is set with a bicycle pump via the valve assembly at the
pressure gauge or established by means of a compressed air supply.
Page 52
52 Operation
• Check level of the reference electrolyte. The level of the reference electrolyte steadily sinks due
to outflow through the diaphragm. If the liquid level has sunk to a level below the mouth of the
bulb (reservoir) of the pH / redox electrode, then the reference electrolyte must be topped up
(see Chapter 6 “Maintenance” on page 53).
5.3 Cleaningtheelectrode/sensor
The electrode / sensor must be cleaned before removal, before calibration of the measurement system
or at regular intervals during operation (depending on the process medium). Proceed as follows:
1. Move the housing into the “Service” position.
2. Open both valves locatged upstream and downstream of the flushing connections “In” / ”Out”.
3. Let the cleaning liquid clean the sensor for 2 minutes (or longer if appropriate).
4. Close flushing water supply (or both valves).
5.4Calibrating the measuring system
The frequency of calibration of the measurement system depends on the electrode / sensor type
and the sample medium. To calibrate the measurement system, proceed as follows:
1. Retract the insertion rod into the “Service” position.
2. Open flushing water supply, flush electrode / sensor and then close flushing water supply
again (see Chapter 5.3 “Cleaning the electrode / sensor” on page 52).
3. Remove electrode / sensor (procedure, see Chapter 4.5.1 “Removing the electrode / sensor”
on page 48).
ATTENTION! This is only necessary if your installation does not allow calibration of the mea-
h
4. Perform calibration in accordance with the operating instructions for the respective elec-
trode / sensor and transmitter.
5. Re-install electrode / sensor (procedure, see Chapter 4.2.7 “Fitting the electrode / sensor” on
page 44).
ATTENTION! This is only necessary if your installation does not allow calibration of the
h
6. Open flushing water supply, flush the electrode / sensor and then close flushing water sup-
ply again.
7. Set the insertion rod into the “Measure” position.
surement system with the electrode / sensor installed.
measurement system with the electrode / sensor installed.
DANGER! The information and instructions given in Chapter 1 “Introduction” on page 9
a
CAUTION! Maintenance and service work on the housings may only be carried out by ap-
a
CAUTION! It is possible that the process medium could harm the environment and your
a
and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page
22 must be fully adhered to.
propriately trained personnel. Only the maintenance and repair work described in
the following Sections may be performed on the retractable housings.
health (toxic, caustic, corrosive, etc.). Therefore, ensure that the equipment is in
a non-hazardous condition before you start with any maintenance work.
DANGER! Before startup, the housing must be checked to ensure that:
a
– the right types of O-ring have been selected and that they are undamaged and
6.3Replacing seals in contact with the process medium
(InTrac78X)
Depending on the application, it is a must to replace the seal at least every year. With aggres-
sive media, the seals may need to be changed at correspondingly shorter intervals. The seal
condition may visually inspect via the housing inspection windows. The seals must be replaced
when there is leaking visible on the inspection windows. Frequent insertion and retraction motion
of the insertion rod can also have an influence on the necessary maintenance intervals.
ATTENTION! Recommendations by METTLER TOLEDO concerning maintenance intervals are
h
To replace the seals, proceed as follows:
1. Set the housing to the “Service” position.
ATTENTION! Visually check the sensor position rod and ensure the housing is fully retracted
h
based solely on experience gained in standard applications and in no way binding or an admission of any guarantee liability whatsoever on the part of the manufacturer / supplier. Depending upon the degree of aggressiveness of the process
medium, the necessary maintenance intervals to support smooth operation of the
equipment may be correspondingly shorter.
(at “Service level” mark).
2. Perform shutdown procedure as described in Chapter 4.4 “Shutdown procedure for manu-
ally and pneumatically operated housings” on page 47
3. Removing the sensor / electrode as described in Chapter 4.5.1 “Removing the electrode / sen-
sor” on page 48
4. Remove the retractable housing as described in Chapter 4.5.2 “Removing the retractable
housing” on page 49
CAUTION! Never place or support the housing on the front ends of the insertion rod (risk of
a
5. Loosen the four housing clamp screws.
ATTENTION! Usually the housing clams remain on the housing drive train and should be eject-
h
6. Remove the drive train “1” and cleaning chamber “2” from the process flange connection
“3” as shown.
7. There are a total of seven seals have to be replaced (arrows A to G in picture). Replace all
seven seals with new ones.
damage).
ed by gently pushing it with the small slot screw drive (flat-end).
9. Slighly bend the seal / scraper “E” as pictured.
10. Gently insert the seal “E” into the cleaning chamber.
➡
ATTENTION! Never bend the seal / scraper “E” with excessive force. By doing this, the seal will
h
11. Install the drive train “1” and cleaning chamber “2” to the process flange connection “3”.
Mount the housing clamps and fasten tightly with the four screws.
ATTENTION! Usually the housing clamps need to be gently pushed, so they can fit between the
h
12. Install the housing to the process line / vessel / tank etc as described in Chapter 4.2 “Fitting
and installation work” on page 37
13. Fitting the electrode / sensor as described in Chapter 4.2.7 “Fitting the electrode / sensor” on
page 44.
14. Follow start up procedure as described in Chapter 4.3 “Startup procedures for housings” on
page 46.
ATTENTION! Details of replacements and examination of the seals are to be documented in the
h
CAUTION! Only perform the dismantling work described and instructed in this Chapter and
a
be irreversibly damaged.
drive train “1” and the flange connection “3”. The clamps should never block the
inspection windows.
maintenance log.
replace only the seals listed above. Check that the correct types of seals have
been selected, examine for any signs of damage and for correct seating.
In this chapter you will find an overview of possible malfunctions which could appear during operation of the retractable housings, their cause, as well as a guide to appropriate remedial measures.
DANGER! It is essential to comply with the safety instructions given in Chapter 1 “Introduc-
a
MalfunctionCauseRemedial action
Insertion rod cannot
beinserted.
Insertion rod remains
stuck between the end
positions “Service” and
“Measure”.
Sample medium dis-
charges via ushing
lines during normal
operation.
Sensor breaks frequently.
Sample medium
escapes from the
inspection windows.
Sample medium
escapes at the ange
connection / fastenings.
Sample medium
escapes through the
NPT thread.
InTrac784: No pressure in upper part
(pressure compensation) or drops rapidly.
There is no electrode / sensor installed in the housing . Install an electrode / sensor.
Loose sensor.Tighten sensor.
InTrac78X type R and I: Wrong pneumatic tube con-
nection.
InTrac78X type R and I: compressed air pressure
toolow.
The sensor position rod / sensor pressurized cylinder
is not fully fastened.
InTrac78X type R and I: No or insufcient control air
pressure .
InTrac78X type R and I: Leak in control air system.Check / make leakproof control air system.
Deposits of sample medium lead to seizing of inser-
tion rod.
Drive unit defective.Contact your local METTLER TOLEDO representative.
The housing is stuck in between “Service” and “Mea-
sure” positions.
Seals of ushing chamber are defective.Remove housing and replace seals (see Chapter 6
Process liquid contains solids.Adjust the sensor protective cage (see Chapter 4.2.2
Sensor is not correctly installed.See Chapter 4.2.7 “Fitting the electrode / sensor” on
Seals need to be replaced.Replacing seals (see Chapter 6.3 “Replacing seals
Flange connection unevenly positioned or not fastened tightly enough.
Flange gasket defective.Check gasket and replace if necessary
Thread insufciently sealed.Seal using PTFE tape / appropriate sealing tape
Compensation pressure not (correctly) set.Set compensation pressure with pump or check com-
Leak in upper part.Check screw ttings and seals
Broken electrode.Replace electrode
tion” on page 9 and Chapter 2 “Important notes for housing uses in ExClassified area” on page 22.
Check pneumatic connection, Chapter 4.2.5 “Pneu-
matic-connections (For InTrac78X, type R)” on page
42.
Pressure must between 4 and 6 bar. The compressed
air pressure must higher than the process pressure.
Check.
Fasten the sensor position rod / sensor pressurized
cylinder.
Check control air system / adjust pressure.
Remove housing and clean (see Chapter 6 “Maintenance” on page 53).
Follow troubleshooting as above, “Insertion rod
remains stuck between the end positions “Service”
and “Measure”.
“Maintenance” on page 53).
“Adjusting the sensor protective cage” on page
38).
page 44.
in contact with the process medium (InTrac78X)” on
page 54).
Check positioning of ange connection and / or tighten
Replacement of all other parts of the housings may only be carried out by appropriately trained
personnel. Please contact your local METTLER TOLEDO representative.
Decommissioning may only be carried out by persons with appropriate training or by skilled
technicians.
9.1 Decommissioning
Proceed as described in Chapter 4.5.2 “Removing the retractable housing” on page 49.
9.2 Storage
Store the InTrac78X in a dry place. Before storage, the housing must peroperly cleaned and dry.
9.3 Disposal
Waste electrical prouducts should not be disposed of with household waste.
It is recommended that the operator disposes of the device in accordance with local
regulations. The operator must deliver the device either to a licensed private or
public disposal company, or dispose of it himself in accordance with prevailing
regulations. Waste is to be recycled or disposed of without causing any risk to human health, and without using procedures or methods that might harm the environment. Check with your Local Authority or retailer for recycling advise.
Sorting
Sorting into waste groups takes place when dismantling the device. The groups are listed in the
current European Waste Catalogue. This catalog is valid for all wastes, whether intended for disposal or for recycling.
The packaging is made up of the following materials:
– cardboard
– foam plastic
Depends on the housing configuration, in general the housing may be made of the following or a
combination of the following materials:
– Stainless steel (DIN 1.4404 / AISI 316 L, and / or DIN 2.4602 / AISI Alloy C22).
– Polypropylene (PP)
– Polyvinylidene fluoride (PVDF)
– Polyether ether ketone (PEEK)
– For housing InTrac 78X[I] (Inductive check back) only:
Contain electrical or electronic in the inductive check back sensor.
This is a brief overview of pH and redox electrodes suitable for use with specific media. For further
information on the extensive range of electrodes and sensors available from METTLER TOLEDO,
please contact your local METTLER TOLEDO representative
Mettler-Toledo (M) Sdn Bhd
Bangunan Electroscon Holding, U 1-01
Lot 8 Jalan Astaka U8 / 84
Seksyen U8, Bukit Jelutong
MY - 40150 Shah Alam Selangor
Phone +60 3 78 44 58 88
e-mail
MT-MY.CustomerSupport@mt.com
1258
Mexico
Mettler-Toledo S.A. de C.V.
Ejército Nacional #340
Polanco V Sección
C.P. 115 60
MX - México D.F.
Phone +52 55 1946 0900
e-mail mt.mexico@mt.com
Norway
Mettler-Toledo AS
Ulvenveien 92B
NO - 0581 Oslo Nor way
Phone +47 22 30 44 90
e-mail info.mtn@mt.com
Philippines
Mettler-Toledo Philippines Inc.
6F NOL Towers, Commerce Ave.
Madrigal Business Park
Ayala Alabang
Muntinlupa 1780 Philippines
Phone +63 2 528 8920
e-mail
MT-PH.CustomerSupport@mt.com
Poland
Mettler-Toledo (Poland) Sp.z.o.o.
ul. Poleczki 21
PL - 02-822 Warszawa
Phone +48 22 440 67 00
e-mail polska@mt.com
Mettler-Toledo Türkiye
Haluk Türksoy Sokak No: 6 Zemin ve 1.
Bodrum Kat
34662 Üsküdar - Istanbul, TR
Phone +90 216 400 20 20
e-mail sales.mttr@mt.com
USA
METTLER TOLEDO
Process Analytics
900 Middlesex Turnpike, Bld. 8
Billerica, MA 01821, USA
Phone +1 781 301 8800
Freephone +1 800 352 8763
e-mail mtprous@mt.com
Vietnam
Mettler-Toledo (Vietnam) LLC
G Floor, SCS Building, Plot T2-4
D1 Street, Saigon Hi-tech Park
Tan Phu Ward, District 9
Ho Chi Minh City, Vietnam
Phone +84 28 73 090 789
e-mail
MT-VN.CustomerSupport@mt.com