Mettler Toledo InTrac 784, InTrac 781 User guide

Page 1
InTrac78X
Operating Instructions
Page 2
04/2018 © Mettler-Toledo GmbH, Process Analytics
Subject to technical changes.
Printed in Switzerland. 30 283 187 B
Page 3
3
InTrac78X
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 4
4

Product and maintenance log

Product name:
Article no.: 30 261 244
Product identifier no.: InTrac78_ _ /_ _ _ _ /_ _ /_ _ _ /_ _ _ /_ _ _
Serial no.:
A logbook of activities should be maintained for the above-mentioned product, recording all data about events, adopted measures, manipulations etc. relative to the product, such as checking of the shipment, initial installation, service / mainte nance, repair work, startup and shutdown etc.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 5
5

How to use this instruction manual

This Instruction Manual is an integral part of the METTLER TOLEDO retractable housing InTrac78X
and contains notes and instructions that are important for safety and operation.
All persons working on or with the InTrac78X must have first read and understood the sections
appropriate to the work in hand.
Please read this instruction manual carefully before using the InTrac housing. Keep this document close to the unit, so that operating personnel may easily be able to refer to it at any time.
Caution! Please first read Chapter 1 “Introduction” on page 9 and Chapter 2 “Important
a
notes for housing uses in Ex-Classified area” on page 22.
Proprietary designations
The following are proprietary names and, for the sake of simplicity, will be mentioned in this instruction manual without the registration marking, e.g.
– InTrac, InPro, InFlow, EasyClean and InFit are registered trademarks of METTLER TOLEDO
Group.
®
– Viton – All other trademarks are the property of their respective holders.
Use of warnings and symbols
DANGER! Warning of a dangerous situation that can lead to extensive material damage,
a
CAUTION! Warning of a possible dangerous situation that can lead to light bodily harm
a
ATTENTION! Information referring to technical requirements. Non-adherence can lead to mal-
h
Explanation of housing designations
and Kalrez® are registered trademarks of DuPont.
todeath or grave bodily injury.
and / or material damage.
functions, uneconomic working and possibly also to loss of productivity. The manufacturer is not liable for any damages resulting from improper or inappropri­ate use.
®
.
The designation InTrac78X used in this Instruction Manual covers reference to the following housings:
InTrac781 for pH / redox electrodes, O
PG13.5 thread).
InTrac784 for pH / redox electrodes with liquid electrolyte e.g., InPro 2000 or 465 sensor series.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
and conductivity sensor (with diameter 12 mm and
2
Page 6
6

Contents

1 Introduction 9
1.1 Housing designations 10
1.2 Basic principles 10
1.3 Warning notices and symbols 11
1.4 Responsibilities, organizational measures 11
1.4.1 Responsibilities of operator 11
1.4.2 Responsibilities of personnel 11
1.4.3 Selection and qualications of personnel – basic duties 12
1.5 Product-specic hazards 12
1.5.1 Sensor design 12
1.6 Manipulation of and maintenance work on the housings 13
1.6.1 Polymer housings 14
1.6.2 Installation in pressurized systems 15
1.7 Installation in potentially explosive areas (hazardous areas) 16
1.8 Residual hazards 17
1.8.1 Rupture of pneumatic or hydraulic connections 17
1.8.2 Leaky connections 17
1.8.3 Malfunctions in overriding control and safety systems 17
1.8.4 Electricity or control (compressed) air failure 17
1.8.5 Medium residues 18
1.8.6 Manually operated housings 18
1.8.7 Pneumatically operated housing 19
1.8.8 Heat protection 20
1.8.9 External inuences 20
1.9 Emergency measures 20
1.10 Safety measures 20
1.11 Modications 21
2 ImportantnotesforhousingusesinEx-Classiedarea 22
2.1 Notes on operating instructions 22
2.2 Intended use 22
2.3 Safety instructions 23
2.4 Correct disposal of the housing 24
2.5 Use in Ex-zones 24
2.6 Ex-classication ATEX 24
2.6.1 Introduction 24
2.6.2 Special conditions 25
2.7 Ex-classication FM approved 26
2.7.1 Introduction 26
2.7.2 Special conditions 26
2.7.3 Applied standards 27
3 Productdescriptions 28
3.1 Scope of delivery 28
3.2 Packing 28
3.3 Checking the shipment 28
3.4 Product overview 28
3.5 Construction of the retractable housings 30
3.5.1 Operating modes of the retractable housings 30
3.6 Description of function of retractable housings 33
3.6.1 Manually operated versions (InTrac78XM) 34
3.6.2 Pneumatically operated versions (InTrac78X, type R or I) 34
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 7
Contents 7
4 Installationandstart-up 36
4.1 Preparation of the equipment 36
4.2 Fitting and installation work 37
4.2.1 Fitting the housing 37
4.2.1.1 Fitting via a ange 37
4.2.1.2 Fitting via NPT external thread 37
4.2.1.3 Fitting via a weld-in socket (with Ingold DN 25) 38
4.2.2 Adjusting the sensor protective cage 38
4.2.3 Attaching the ushing lines 39
4.2.3.1 Normal sensor cleaning 40
4.2.3.2 Backwash sensor cleaning 41
4.2.4 Installation for calibration of the measurement system with electrode / sensor installed 41
4.2.5 Pneumatic-connections (For InTrac78X, type R) 42
4.2.6 Pneumatic connection (For InTrac78X, type I) 42
4.2.7 Fitting the electrode / sensor 44
4.3 Startup procedures for housings 46
4.3.1 Startup of manually operated housings 46
4.3.2 Startup of pneumatically operated housings 47
4.4 Shutdown procedure for manually and pneumatically operated housings 47
4.5 Dismantling work 48
4.5.1 Removing the electrode / sensor 48
4.5.2 Removing the retractable housing 49
4.6 Use of retractable housings in conjunction with EasyClean 50
5 Operation 51
5.1 Important information for everyday operation 51
5.2 Inspection work in everyday operation 51
5.3 Cleaning the electrode / sensor 52
5.4 Calibrating the measuring system 52
6 Maintenance 53
6.1 Important information on maintenance 53
6.2 Topping up reference electrolyte (InTrac784) 53
6.3 Replacing seals in contact with the process medium (InTrac78X) 54
6.3.1 Flange 55
6.3.2 Ingold DN 25 55
7 Troubleshooting 57
8 Productspecications 59
8.1 Technical data 59
8.2 Spare parts list 60
9 Decommissioning,storage,disposal 61
9.1 Decommissioning 61
9.2 Storage 61
9.3 Disposal 61
10 Appendices 62
10.1 Electrode / sensor selection 62
10.2 Dimensional drawing 63
10.3 Product structure (InTrac78X) 66
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Transmitter InTrac 78X Printed in Switzerland 30 283 187 B
Page 8
8
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 9
Introduction 9

1 Introduction

The retractable housing InTrac78X is safe to operate. When used correctly, it represents no danger. – Before starting to use the housing, carefully read this instruction manual: the safety precautions
and warnings contained in it must be observed.
– The retractable housing has been tested in-house and dispatched ready for installation.
In addition to this instruction manual please also note the following:
• all local safety regulations concerning the execution of pneumatic and water installations.
• all instructions and warning remarks in the publications of the products that are used in con­junction with the retractable housing (electrodes, sensors, controls, etc.).
• all safety precautions for the plant into which the retractable housing will be installed.
• all instructions and warnings labeled on the retractable housing.
• all safety information relative to operation in potentially explosive atmospheres / hazardous areas (Ex classified zones).
The instruction manual contains the most important information for using InTrac78X efficiently
and in accordance with regulations. A basic condition for safe handling and operation without malfunctions is a knowledge of these safety instructions and observance of the further warnings in the instruction manual.
This instruction manual, and in particular the safety regulations, are intended for personnel en­trusted with the operation and maintenance of the retractable housing. It is assumed that these persons are familiar with the equipment in which the retractable housing is installed. Therefore, before any work is started with the retractable housing, this instruction manual must be read and understood by those persons involved.
The instruction manual must be stored where it is constantly accessible and available to any
person working with the InTrac78X.
On receipt of the shipment, check immediately:
• the retractable housing and accessories for any sign of transport damage. Report any damage immediately to the carrier and to your supplier.
• the type designation on the housing body.
• for completeness of the supply. Please notify your supplier immediately if the shipment is incomplete or in any way incorrect (see Chapter 3.1 “Scope of delivery” on page 28).
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 10
10 Introduction

1.1 Housing designations

Housing designation, product identifier no. as well as article no. and serial no. for clear identifica­tion when communicating with the manufacturer can be noted from the type plate.
ExampleInTrac781andInTrac784housingtypeplate
Note: The material number; (30261244) is the same for both InTrac781 and / or InTrac784
housings. However, the product Smart Configuration Key (SCK) and serial number (SNR) are different for every retractable housing.

1.2 Basic principles

The retractable housing InTrac78X is built in accordance with state-of-the-art technology and
recognized technical safety regulations.
However, the housing can be a source of risks and dangers:
– if the housing is operated by underskilled or unauthorized personnel. – if the housing is not used in compliance with regulations and / or stipulations for appropriate use.
InTrac78X may only be used in a technically correct condition for the purpose intended by the
supplier, with awareness by the user of safety and danger factors, taking the instruction manual and local safety regulations into consideration.
Malfunctions and damage that can affect the safety and function must immediately be remedied by the operator or an expert, and notified to the manufacturer in writing!
DANGER! A defective retractable housing must neither be installed nor put into operation.
a
Faulty containment and installation out of conformance with regulations and in­structions can lead to the escape of medium or to pressure surges (explosion), potentially harmful both to persons and to the environment.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 11
Introduction 11

1.3 Warning notices and symbols

The following symbols are used in this instruction manual to mark safety instructions:
DANGER! Warning of a dangerous situation that can lead to extensive material damage,
a
CAUTION! Warning of a possible dangerous situation that can lead to light bodily harm
a
ATTENTION! Information referring to technical requirements. Non-adherence can lead to mal-
h
todeath or grave bodily injury.
and / or material damage.
functions, uneconomic working and possibly also to loss of productivity.

1.4 Responsibilities, organizational measures

1.4.1 Responsibilities of operator
The operator is under obligation only to permit persons to work with InTrac78X retractable
housings, who are familiar with the basic requirements of work safety and accident prevention, and who have been instructed in the handling of the housing. This instruction manual serves as the basic document.
– In addition to the instruction manual there are also generally valid legal and other binding
regulations for work safety and accident prevention as well as for environmental protection, and these must be provided by the operator and instructed to personnel using the housings.
– The operator / user must be fully aware of safety and potential danger factors during work with
the housings and this awareness must be checked by the operator at regular intervals.
– Measures must be taken to ensure that the retractable housings are only operated in a safe and
fully functional condition.
– If the housings are employed in hazardous areas, compliance with prevailing regulations is to
be ensured.
CAUTION! Before the retractable housing is put into operation, the operator must have al-
a
ready clarified that use of the housing in conjunction with the other associated equipment and resources is fully authorized.
1.4.2 Responsibilities of personnel
– All persons whose duty it is to operate the retractable housings are under obligation to read
Chapter 1 “Introduction” on page 9 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22 as well as the warning notices in this instruction manual.
– In addition to the instruction manual, generally valid legal and other binding regulations for
work safety and accident prevention must be adhered to.
– Any method of working which is doubtful from a safety perspective and which exceeds the
operation according to regulations must be omitted.
– Avoid using high-pressure water jets to clean any parts and components of the housing.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 12
12 Introduction
ATTENTION! Before every start-up, the retractable housing must be checked for:
h
• damage to the connections, fastenings, etc.
• leakage
• perfect functioning
• authorization for use in conjunction with other plant equipment and resources.
DANGER! A defective retractable housing may neither be installed nor put into operation.
a
Faulty containment or installation out of conformance with regulations and in­structions can lead to the escape of medium or to pressure surges (explosion), potentially harmful both to persons and to the environment.
1.4.3 Selectionandqualicationsofpersonnel–basicduties
– Work on or with the retractable housings may only be carried out by reliable and appropriately
trained or instructed personnel. The personnel must have read this instruction manual in advance.
– Clear responsibilities / standard operation procedures (SOPs) must be established for the per-
sonnel entrusted with operation, service, repair, etc. of the housings.
– It must be ensured that only specifically assigned personnel may operate the housings.
DANGER! Incorrect manipulation or operation of the housings or non-observance of safety
a
regulations can lead to problems with the housings and to the escape of process medium, thereby presenting a potential hazard to the environment, personnel and material.
1.5 Product-specichazards
1.5.1 Sensor design
DANGER! The electrode or sensor may only be removed from a mounted retractable housing
a
This can endanger personnel and cause damage to the environment and material.
ATTENTION! If an electrode / sensor is broken, it must be replaced immediately.
h
when this is in the “Service” (retracted / withdrawn / maintenance) position, since if the electrode / sensor is missing from the housing, there is potential path for the process medium to escape to the environment.
Toxic or aggressive media can pose a threat to human life through poisoning, chemical burns or scalding.
DANGER! A broken sensor jeopardizes process safety, since it is not then possible to achieve
a
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
reliable measurement results.
Page 13
Introduction 13
ATTENTION! The intelligent sensor locking system, modular manual drive train device and
h
Principally, personal protective equipment such as protective goggles and cloth-
Any manipulation of the electrode / sensor may only be carried out when the re-
“Service”or“Measure”positionoftheInTrac78XMhousings
the visualized sensor interlocking bolts in the manual retractable housings (InTrac78XM) makes it difficult for any unintentional removal of the electrode / sen-
sor from the housing when in the “Measure” position. This locking feature is basi­cally intended as an added safety feature for the device. However, the operator / user must be fully aware of safety and potential danger factors during work with the housings and this awareness must be checked by the operator at regular intervals.
ing must be worn.
tractable housing is in the “Service” (retracted / withdrawn / maintenance) position.
Note: The sensor interlocking bolts indicate the operation mode
(for manual retractable housing – InTrac78XM model only)
1.6 Manipulationofandmaintenanceworkonthehousings
ATTENTION! Before dismantling a retractable housing or commencing any maintenance work on
h
Manipulation of the housings may only take place after it is has been ensured that no process medium can escape through the housing in the event of incorrect manipulation.
For this reason, the complete system must be emptied and vented in advance (safe condition).
If the retractable housing has been in operation in conjunction with the automated cleaning and cali­bration system EasyClean, it must be ensured that before any manipulation of the housing, the Easy­Clean system has been completely switched off and isolated (power, compressed air and water).
It is principally necessary to wear personal protective outfit such as protective goggles and clothing.
Only such maintenance and repair work as is specifically described in this instruction manual may be performed on the retractable housing.
it, ensure that the equipment in which the retractable housing is installed is in a safe condition (depressurize, no explosion risk, empty, rinse, vent, etc.). Retractable housings may only be stripped down after having been completely dismounted.
Only original spare parts from METTLER TOLEDO may be used for replacing defective components.
DANGER! Non-compliance with the prescribed maintenance instructions can endanger per-
a
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
sonnel and the environment.
Page 14
14 Introduction
1.6.1 Polymer housings
ATTENTION! – Retractable housings made of polymer material do not have the same me-
h
DANGER! Polymer housings may not be used in applications subject to high mechanical
a
chanical properties as steel housings.
– Retractable housings made of polymer need more frequent servicing than steel
housings.
– Defective components may only be replaced by an authorized service center.
stress, as this could lead to collapse of the protective caging. In such case, pro­cess liquid would be able to access the flushing chamber, resulting in process medium mixing with cleaning medium. Please see Chapter 1.6.2 “Installation in pressurized systems” on page 15 for recommended maximum permissible temperature and pressure.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 15
Introduction 15
0
2
4
6
8
10
12
14
16
18
020406080100 120140 160
Max. Permissible Pressure (bar)
Max. Permissible Temperature (°C)
Maximum Housing Wetted Material Permissible Pressure and Temperature
PP
PVDH
PEEK Stainless steel/
DIN1.4404/ AISI316L DIN2.4602/ AISI Alloy C22
PVDF
1.6.2 Installation in pressurized systems
ATTENTION! The maximum permissible temperature and pressure specifications must not be
h
DANGER! If temperature and pressure limits are exceeded, there is a risk to the integrity of the
a
exceeded. The values depend upon the design and type of the retractable housing. The relative specifications are given on the type plate of the individual housings. Please note that the maximum permissible pressure and temperature may differ depending on the process medium. Please consult your local METTLER TOLEDO representative for more information.
system, thereby presenting a potential threat to human life and to the environment.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 16
16 Introduction

1.7 Installation in potentially explosive areas (hazardous areas)

ATTENTION!
h
– Please consult and follow the drawings and diagrams supplied with the measuring system. – Where appropriate, it should be considered to include the METTLER TOLEDO retractable hous-
ing InTrac78X and the process connections in your recurring pressure test program for the
complete plant as a whole.
– Before the retractable housing is put into operation, the operator must have already clarified
beyond doubt that use of the housing in conjunction with the other associated equipment and resources is fully authorized.
– Please note that for plastic type of retractable housing; not all metallic parts are electrically
connected. Hence, all metal part of retractable housing InTrac78X and the process connections
must be connected to the main potential equalization system of the plant (grounded). The grounding cable (see drawing) has to be installed by the operator.
– Potential electrostatic charging hazard to the housing flange and wetted parts. Proper dis-
charge is needed before carrying out housing installation or after service, and / or removing from the process.
– In special applications, the minimum conductivity of the medium (in contact with the housing
wetted part) shall be more than 1 nS / cm.
DANGER! Non-observance of legal regulations concerning use in hazardous areas can en-
a
ATTENTION! See also Chapter 2 “Important notes for housing uses in Ex-Classified area” on
h
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
danger human life and the environment.
page 22.
Page 17
Introduction 17

1.8 Residual hazards

ATTENTION! Despite all precautionary measures taken, residual hazards still remain.
h
1.8.1 Rupture of pneumatic or hydraulic connections
DANGER! The escape of aggressive medium can present a threat to human life and to the
a
environment.
1.8.2 Leakyconnections
– Connections can become loosened through the effects of vibration. – Never remove the lower and top housing clamp screws during operation. – Never remove the sensor / electrode when the housing is in the “Measure” position. Visually
check the housing position level and ONLY remove the sensor when it is fully retracted from the process or in the “Service” position.
– The connection between housing and process adaptor is a potential source of leakage.
ATTENTION! The connections between the housing and the process adaptor must be checked
h
DANGER! Leaky connections can allow process medium, cleaning solution or control (com-
a
regularly by the customer / operator, and kept in full working condition.
pressed) air to escape to the environment, presenting a hazard for persons and the environment.
1.8.3 Malfunctionsinoverridingcontrolandsafetysystems
ATTENTION! Problems occurring in the overriding control system can trigger off uncontrolled
h
insertion and retraction of the housing. This can in turn lead to a complete break­down or malfunctioning of the safety devices.
1.8.4 Electricity or control (compressed) air failure
DANGER!
a
– If there is power failure, information as to the end positions of the housing is no
longer available via the inductive position indication system.
– In addition, the overriding control system receives no reliable information con-
cerning exact positioning.
– If the control (compressed) air fails, it is possible that the insertion rod will not
reach the required end position.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 18
18 Introduction
DANGER! Before commencing any maintenance work on or removing an electrode / sensor
a
If the insertion rod comes to rest or is stuck between the defined “Measure” or
DANGER! The electrode / sensor may only be removed when the housing is in the “Service”
a
from a housing it is essential to visually check whether the housing is in the “Ser­vice”, “Measure” or any undefined intermediate position. The insertion rod must always be stationed in one of the two defined position:
– “Measure” (operating position), that is, in the fully inserted position. – “Service” (withdrawn position), that is, in the fully retracted position.
“Service” position, then no conforming measuring results can be ensured. Poten­tial medium leaking to the flushing connection is possible. Proper isolation of the process line, depressurized etc. is recommended before carrying out the re­pair / maintenance if the insertion rod is in any undefined position or other than “Service” position.
position, as otherwise there is a risk that process medium could escape to the environment. This can present a potential hazard to person, environment and material. Toxic or aggressive media can pose a threat to human life through poi­soning, chemical burns or scalding.
1.8.5 Mediumresidues
DANGER! When retracting the insertion rod from the process, small quantities of process
a
DANGER! If the insertion rod is not completely inserted into the process, there will be a path
a
medium will remain attached to the electrode / sensor and will thus enter the flush­ing chamber. If this medium is a toxic or environmentally damaging substance, then such contamination must be removed and disposed of in accordance with regulations!
for the process medium to enter the flushing chamber and thus to the flushing connection. This can lead to product loss or contamination.
1.8.6 Manuallyoperatedhousings
DANGER! The electrode or sensor may only be removed when the housing is in the “Service”
a
position and as soon as when the manual drive train is manually turned (anti­clockwise) until hand tight. As soon as the manual drive train is completely turned anti-clockwise, the housing will automatically self-lock. After that, the manual drive train cannot be further turned either clockwise or anticlockwise.
DANGER! Avoid excessive force or turning the manual drive train with any special tools.
a
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 19
Introduction 19
1.8.7 Pneumatically operated housing
ATTENTION! In the case of the pneumatically operated InTrac78X, failure of the compressed
h
DANGER! The electrode or sensor may only be removed when the insertion rod is fully
a
DANGER! Before the electrode / sensor is loaded into the pneumatic version InTrac78XR
a
air supply can result in the insertion rod being pushed to the “Service” position by the process pressure. This makes continuous measurement impossible.
retracted from the medium. It is essential to visually check whether the housing is in the “Service”, “Measure” or any undefined intermediate position. In addition, visually check if the sensor mounting thread is fully retracted from the process (see picture) before carrying out any maintenance work. Removing the sensor from the housing when the housing is not fully retracted from the process can result in process media escaping and presents a potential hazard both to persons and to the environment.
or InTrac 78XI, it must be ensure that the pneumatic valves that control the position of the housing are set to the “Service” position. Incorrect setting can cause the housing to insert into the process before the loading procedure has been completed, resulting in open access to the process medium. Process me­dium can thereby escape and present a potential hazard both to persons and to the environment.
“A” “B”
Note: “A”: “Measure” position
“B”: “Service” position
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 20
20 Introduction
1.8.8 Heat protection
DANGER! Take applicable protection measures prior to touching the holder / housing as parts
a
of the retractable holder may have reached process temperature and cause burns.
1.8.9 Externalinuences
ATTENTION! Objects falling on the housing can damage or destroy the unit, or cause leaks etc.
h
ATTENTION! Lateral forces may damage or destroy the unit.
h

1.9 Emergency measures

ATTENTION! Always observe and comply with local regulations!
h
ATTENTION! If the retractable housing has been in operation in conjunction with the automated
h
cleaning and calibration system EasyClean, it must be ensured that before any manipulations at the housing, the EasyClean system has been completely
switched off and isolated (power, compressed air and water).

1.10 Safety measures

ATTENTION! Always observe and comply with local laws and regulations! These are not an
h
DANGER! It is principally necessary to wear protective equipment such as protective goggles
a
ATTENTION! The operator is responsible for the instruction of personnel. Additional copies of
h
The operator is obliged to inform the supplier / manufacturer of the retractable
integral part of this instruction manual.
and protective clothing.
this instruction manual can be ordered from the equipment supplier. As an integral part of the retractable housing, this instruction manual must at all times be read­ily accessible to users at the point of operation of the housing.
housing immediately about any safety-relevant incidents, or observations made, during use of the housing.
DANGER! Incorrect manipulation and / or instruction errors can lead to potential hazards for
a
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
persons and for the environment.
Page 21
Introduction 21
ATTENTION! Before every start-up, the retractable housing must be checked for:
h
– damage to the connections, fastenings, etc. – leakage – defective cables and lines etc. – authorization for use of the housing in conjunction with the associated plant
resources.
– In the ex-classified area installation, strictly adhere and follow the instructions
as specified in Chapter 1.7 “Installation in potentially explosive areas (hazard­ous areas)” on page 16 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22.
DANGER! A defective housing must never be installed or put into operation. Poor contain-
a
ment, leaky connections etc. or non-compliant installation of the housings can lead to escape of process medium and thereby to a potential threat to life (incl. risk of explosion.).
1.11 Modications
ATTENTION! No attachments or modifications to the retractable housings are allowed.
h
DANGER! The manufacturer / supplier accepts no responsibility for any damage caused
a
by unauthorized attachments and alterations or for the incorporation of spare parts which are not of METTLER TOLEDO provenance. The risk is borne entirely by the operator.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 22
22 Important notes for housing uses in Ex-Classied area
2 ImportantnotesforhousingusesinEx-Classiedarea

2.1 Notes on operating instructions

These operating instructions contain all the information needed for safe and proper use of the housing.
The operating instructions are intended for personnel entrusted with the operation and mainte­nance of sensors and housings. It is assumed that these persons are familiar with the equipment in which the sensors and housings are installed.
See Chapter 1.3 “Warning notices and symbols” on page 11 for explanation.

2.2 Intended use

TheretractablehousingsInTrac78Xareintendedsolelyformeasurementtasksinconjunc­tion with thespecified METTLER TOLEDO electrodes/sensors, namely pH and redox (ORP)
combination electrodes as well as oxygen, CO2, conductivity or turbidity sensors. Use the housings only for this purpose.
Housings with the potentially explosive/hazardous areas.
Any use of these housings which differs from or exceeds the scope of use described in this instruc­tion manual will be regarded as inappropriate and incompatible with the intended purpose.
The manufacturer/supplier accepts no responsibility whatsoever for any damage resulting from such improper use. The risk is borne entirely by the user/ operator.
Other prerequisites for appropriate use include:
– Compliance with the instructions, notes and requirements set out in this instruction manual. – Correct maintenance of the housings. – Acceptance of responsibility for regular inspection, maintenance and functional testing of all
associated components, also including compliance with local operational and plant safety regulations.
– Operation in compliance with prevailing regulations concerning the environmental and operat-
ing conditions as well as with the admissible mounting positions.
– Compliance with all information and warnings given in the documentation relating to the prod-
ucts used in conjunction with the housings.
– Correct equipment operation in conformance with the prescribed environmental and opera-
tional conditions, and admissible installation positions.
– Consultation with METTLER TOLEDO Process Analytics in the event of any uncertainties.
or symbol on the type plate have received approval for operation in
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 23
Important notes for housing uses in Ex-Classied area 23
DANGER! The housing must be operated only with the specified electrodes/sensors. The
a
absence or the installation of an inappropriate electrode/sensor may adversely affect the resistance to pressure and temperature, the chemical resistance and the protection against explosion. Consequently, there can be leakage from the hous­ing and/or risk of explosion that may endanger persons and the environment.

2.3 Safety instructions

DANGER!
a
– The plant operator must be fully aware of the potential risks and hazards at-
tached to operation of the particular process or plant. The operator is respon­sible for correct training of the workforce, for signs and markings indicating sources of possible danger, and for the selection of appropriate, state-of-the-art instrumentation.
– It is essential that personnel involved in the commissioning, operation or main-
tenance of these housings or of any of the associated equipment (e.g. sensors, transmitters, etc.) be properly trained in the process itself, as well as in the use and handling of the associated equipment. This includes having read and un­derstood this instruction manual.
– The safety of personnel as well as of the plant itself is ultimately the responsibil-
ity of the plant operator. This applies in particular in the case of plants operating in hazardous zones.
– The housings and associated components have no effect on the process itself
and cannot influence it in the sense of any form of control system.
– Maintenance and service intervals and schedules depend on the application
conditions, composition of the sample media, plant equipment and signifi­cance of the safety control features of the measuring system. Processes vary considerably, so that schedules, where such are specified, can only be re­garded as tentative and must in any case be individually established and veri­fied by the plant operator.
– Where specific safeguards such as locks, labels, or redundant measuring sys-
tems are necessary, these must be provided by the plant operator. – A defective housing must neither be installed nor put into service. – Only maintenance work described in these operating instructions may be per-
formed on the housings. – When changing faulty components, use only original spare parts obtainable
from your METTLER TOLEDO supplier (see Chapter 8.2 “Spare parts list” on
page 60). – No modifications to the housings and the accessories are allowed. The manu-
facturer accepts no responsibility for damage caused by unauthorized modifi-
cations. The risk is borne entirely by the user. – Care must be taken during installation to avoid impacts or friction that could
create an ignition source. – Tampering and replacement with non-factory components may adversely affect
the safe use of the system. – Insertion or withdrawal of removable electrical connectors or modules is to be
accomplished only when the area is known to be free of flammable vapors. – WARNING – INTRINSICALLY SAFE APPARATUS CAN BE A SOURCE OF IGNITION
IF INTERNAL SPACINGS ARE SHORTED OR CONNECTIONS OPENED.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 24
24 Important notes for housing uses in Ex-Classied area
WARNING – DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE IS PRESENT. – WARNING – SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY. – WARNING – SUBSTITUTION OF COMPONENTS MAY IMPAIR SUITABILITY OF THE
EQUIPMENT
WARNING – FOR CONNECTION ONLY TO NON-FLAMMABLE PROCESSES.

2.4 Correct disposal of the housing

When the housing is finally removed from service, observe all local environmental regulations for proper disposal. See Chapter 9 “Decommissioning, storage, disposal” on page 61.

2.5 Use in Ex-zones

ATTENTION!
a
For an installation in Ex-zones please read the guide-lines following hereafter:
1258
Ex-classificationATEX:
II 1/2G c IIC TX Ga/Gb II 1/2D c IIIC TX Da/Db
Number of the test certificate:
SEV 13 ATEX 0161X
Ex-classificationFMapproved:
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 ºC
- 53 800 002; Entity
Project ID: 3021227
FM certificate number: FM16US0034X
FM18CA0021X
2.6 Ex-classicationATEX
2.6.1 Introduction
According to RL 94/9/EG (ATEX 95) Appendix l, InTrac 78X/*1/*2/*3/*4/*5/*6*7*8 housings are devices group ll, category 1/2G and according to RL 99/92/EG (ATEX 137) may be used in zones 0/1 or 0/2 and gas group IIC that are potentially explosive due to combustible substances in the temperatures classes T3 to T6.
For use/installation, the requirements of EN 60079-14 must be observed.
According to RL 94/9/EG (ATEX 95) Appendix l, InTrac 78X/*1/*2/*3/*4/*5/*6*7*8 housings are devices group Ill, category 1/2D and according to RL 99/92/EG (ATEX 137) may also be used in zones 20/21 resp. 20/22 that contain combustible dusts.
For use/installation, the requirements of EN 50281-1-2 must be observed.
The pneumatically operated housings with electrical (inductive) position indication of the probes
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 25
Important notes for housing uses in Ex-Classied area 25
may be employed in hazardous areas, Zone 1 and Zone 2 resp. Zone 21 and Zone 22, also in conjunction with separately certified, intrinsically safe, inductive proximity switches – e.g. Pepperl+Fuchs, Types NCB2*** – on condition that the applicable Gas Groups and Temperature Classes correlate with the actual inflammable materials present, and that the special conditions of the declaration are strictly adhered to.
2.6.2 Special conditions
The housings with pneumatic actuation position of the sensors with electrical feedback signal may be operated in hazardous areas Zone 1 and Zone 2 or Zone 21 and Zone 22 with separately certified intrinsically safe inductive proximity switches e.g. Pepperl + Fuchs types NCB2 ***– if the gas groups and temperature classes coincide with the used flammable substances and the spe­cial conditions of the Certificates are observed.
1. The maximum permissible ambient or process temperatures for Zone 0 (flammable gases or flammable liquids) shall be taken according to the following table:
Temperature class Max.environmentresp.mediatemperature
T6 68 °C T5 80 °C T4 108 °C T3 130 °C
The maximum permissible ambient or process temperatures must not exceed the aforementioned values and they will be found in this instruction manual, Chapter 8 “Product specifications” on page 59.
2. The maximum permissible surface temperature for Zone 20 (combustible dust) shall be taken accordingly to the following table:
Temperature class Max.environmentresp.mediatemperature
T 69 °C 68 °C T 81 °C 80 °C T 109 °C 108 °C T 131 °C 140 °C
The maximum permissible ambient or process temperatures must not exceed the aforementioned values and they will be found in this instruction manual, Chapter 8 “Product specifications” on page 59.
3. The metallic body of the housing type InTrac 78X*1/*2/*3/*4/*5/*6/*7/*8 has to be con­nected conductively to the equipotential system of the plant.
4. The housings type InTrac 78X*1/*2/*3/*4/*5/*6/*7/*8 are included in the periodic pressure testing of the system, where appropriate.
5. WARNING – POTENTIAL ELECTROSTATIC CHARGING HAZARD – by installation, use and
maintenance work, see instruction manual.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 26
26 Important notes for housing uses in Ex-Classied area
2.7 Ex-classicationFMapproved
IS/I,II,III/1/ABCDEFG/T6 Ta = 60 ºC
2.7.1 Introduction
The following FM control drawing and the standards listed in Chapter 2.7.3 “Applied standards” on page 27 must be observed, where applicable:
2.7.2 Special conditions
See Chapter 2.6.2 “Special conditions” on page 25.
approvedapproved
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 27
Important notes for housing uses in Ex-Classied area 27
2.7.3 Applied standards
United States Standards
Title Number Issue Date
Approval Standard for Electrical Equipment for Use in Hazardous (Classied) Locations – General Requirements
Approval Standard for Intrinsically Safe Apparatus and Associated Apparatus for Use in Class I, II & III, Division 1, Hazardous
(Classied) Locations
Approval Standard for Electrical Equipment for Mea­surement, Control and Laboratory Use
Explosive Atmospheres – Part 0: Equipment – General Requirements
Explosive Atmospheres – Part 11: Equipment Protection by Intrinsic Safety “i”
Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements
FM Class 3600 2011
FM Class 3610 2015
FM Class 3810 2005
ANSI/ISA-60079-0 (12.00.01) 2005
ANSI/ISA-60079-11 (12.02.01) 2009
ANSI/ISA-61010-1 (82.02.01) 2004
Canadian Standards
Title Number Issue Date
Explosive Atmospheres – Part 0: Equipment – General Requirements
Explosive Atmospheres – Part 11: Equipment Protection by Intrinsic Safety “i”
Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part 1: General Requirements
CAN/CSA-C22.2 No. 60079-0 2005
CAN/CSA-C22.2 No. 60079-11 2009
CAN/CSA-C22.2 No. 61010-1 2004
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 28
28 Product descriptions

3 Product descriptions

3.1 Scope of delivery

Standard supply of the retractable housing is made up of the following:
– One retractable housing finished assembled in accordance with description in order confirmation
– Cable grommet – Sensor protection rod (or position visual indication rod) – ForInTrac781Xonly – Sensor pressurized cylinder (inclu. cylinder glass), sensor pressure indicator and manual
pump – ForInTrac784Xonly – Hexagon key 2.5 mm – 4x spare screws M4x8 (DIN912) – 2x spacers for sensor – instruction manual – Quick set-up guide

3.2 Packing

The packing consists of cardboard and foam plastic.
Keep the packing for possible later use, such as for storage or transport of the housing. However, if you wish to dispose of the packing, please observe your local regulations in this respect.
ATTENTION! see Chapter 9 “Decommissioning, storage, disposal” on page 61.
h
3.3 Checkingtheshipment
When unpacking the shipment, examine carefully for signs of damage. Any damage noted must be reported immediately to the carrier and your supplier. Check that the shipment accords to the delivery papers and to your order.
ATTENTION! Damaged housings must not be installed or put into operation.
h

3.4 Product overview

The InTrac78X retractable housings combine rugged design with great versatility to meet the
demands of the harshest process condition in the chemical, petrochemical, power and utilities industries.
The retractable housing material is specially designed for a range of harsh applications. Wetted parts are available in different materials offering installation flexibility in many applications.
The retractable housing meets major international requirements including installation in Ex-clas­sified areas (ATEX and FM Approvals), where it can be safely installed.
The retractable housings are normally attached to tanks or piping with an applicable process con­nection (i.e. InFlow). In order to meet the various process requirements, the retractable housings are available in different versions, depending on:
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 29
Product descriptions 29
– which typeofelectrode/sensoristobeemployedInTrac781 is for pH / redox electrodes, O
and conductivity sensor (with diameter 12 mm and
2
PG13.5 thread). Depending on the product configuration, the InTrac781 can be operated with
sensor insertion length 225 mm (insertion depth approx. 80 mm) or 425 mm (insertion depth
approx. 280 mm). – InTrac784 is for pH / redox electrodes with liquid electrolyte e.g., InPro 2000 or 465 sensor
series. Depending on the product configuration, the InTrac784 can be operated with sensor
insertion length 250 mm (insertion depth approx. 80 mm) or 450 mm (insertion depth
approx. 280 mm).
Operating modes:
– M = manually operated version – R = pneumatically check back version – I = pneumatically operated version with inductive check back indicators
Depends on types of process connection, insertion depth (“H”) is approx.
– 80 mm – 280 mm
Exact dimension illustrations of all housings are to be found in the relative Appendix to this Manual.
Medium-wettedmaterials
Metal varieties:
– Stainless steel DIN 1.4404 / AISI 316 L – DIN 2.4602 / AISI Alloy C22 (e.g. Hastelloy)
Polymer varieties:
– Polypropylene (PP) – Full polyvinylidene fluoride (PVDF) version – Polyvinylidene Fluoride with combination of
AISI Alloy C22 for sensor tip protective cage (PVDH)
– Polyether ether ketone (PEEK)
Medium-wettedsealingmaterials(O-rings)
– FPM (Viton
®
) – EPDM – FFKM (Kalrez
®
)
Process connections
– Ingold DN 25 – Flange attachments (DIN, ANSI) – NPT external thread
Flushing connections
– G ¼" or ¹/
8" thread female
– ¼" NPT female – Without (blind plugs are included)
The exact version of the housing can be seen from the Smart Configuration Key (SCK) on the housing.
Example: InTrac781M/4404/VI/225/D04/G14
The product structure InTrac78X can be found in Chapter 10.3 “Product structure (InTrac78X)”
on page 66.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 30
30 Product descriptions

3.5 Construction of the retractable housings

3.5.1 Operating modes of the retractable housings
Each version of the retractable housing family InTrac 78X is available in three different types
(M,RandI):
InTrac781M
1
2
3
4
6
Pos Description
1 Cable grommet 2 Sensor position rod (with sensor status- “Service” / “Measure” position indications) 3 Manual drive train 4 Housing clamp screw 5 Sensor interlocking bolts 6 Flushing connection “Out” 7 Flushing connection “In” 8 Process ange connection 9 Insertion rod with sensor protection cage
5
7
8
9
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 31
Product descriptions 31
InTrac781(type“R”or“I”)
1
2
3
4 5
6
7
8
Pos Description
1 Cable grommet 2 Sensor position rod (with sensor status- “Service” / “Measure” position indications) 3 Pneumatic (type R) drive train or,
pneumatic drive train with inductive check back indicator (type I) 4 Housing clamp screw 5 Flushing connection “In” 6 Flushing connection “Out” 7 Process ange connection 8 Insertion rod with sensor protection cage
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 32
32 Product descriptions
InTrac784M
2
1
3
4
5
7
8
6
9
10
11
Pos Description
1 Cable grommet 2 Sensor pressure indicator 3 Sensor pressurized cylinder (with sensor status- “Service” / “Measure” position indications) 4 Glass cylinder 5 Manual drive train 6 Sensor interlocking bolts 7 Housing clamp screw 8 Flushing connection “In” 9 Flushing connection “Out” 10 Process ange connection 11 Insertion rod with sensor protection cage
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 33
Product descriptions 33
InTrac784(type“R”or“I”)
2
4
5
6
7
9
Pos Description
1 Cable grommet 2 Sensor pressure indicator 3 Sensor pressurized cylinder (with sensor status- “Service” / “Measure” position indications) 4 Glass cylinder 5 Pneumatic (type R) drive train or,
pneumatic drive train with inductive check back indicator (type I) 6 Housing clamp screw 7 Flushing connection “Out” 8 Flushing connection “In” 9 Process ange connection 10 Insertion rod with sensor protection cage
1
3
8
10

3.6 Description of function of retractable housings

With the retractable housings InTrac78X, electrodes / sensors can be cleaned or replaced without
interrupting the ongoing process. In addition, with the aid of auxiliary equipment (EasyClean system) it is possible to calibrate and / or backwash (when sensor retracted) and / or flush with the electrode / sensor installed.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 34
34 Product descriptions
3.6.1 Manuallyoperatedversions(InTrac78XM)
The manual version of InTrac78X equipped with the manual drive train. The manual drive train is
a mechanical rotary drive that converts a rotating motion into forwards / backwards movement of the insertion rod. The sensor can be retracted from the process liquid to the cleaning chamber (for maintenance) or likewise. State of the art drive train design allows operators to drive the sensor against high process pressure without the needs to isolate the complete process line. Thanks to manual drive train, the disturbance to the process is minimal.
To retract the sensor from the process, the “Measure” sensor interlocking bolt is pressed. Manu­ally rotate the manual drive train anticlockwise until the “Service” sensor interlocking bolt pops out. In this position it is safe to perform sensor maintenance. The reverse process is applied for inserting the sensor back for operation.
On top of the sensor interlocking bolts, the level marking is visible on Sensor Position Rod (for
InTrac781M) or Sensor Pressurized Cylinder (for InTrac784M) to provide additional information
about the sensor status (either it is in “Measure” or “Service” position). It is necessary to ensure the sensor position rod / sensor pressurized cylinder has fully reached the “Service” level marking before carrying out any sensor maintenance.
ATTENTION! The sensor pressure indicator (InTrac784X) indicates the sensor’s internal pressure
h
(i.e. InPro2000 sensor). To ensure the sensor electrolyte continues to flow freely
from the sensor junction during operation, sensor internal pressure must be higher than the process pressure. Please refer to sensor instruction manual for detail.
3.6.2 Pneumaticallyoperatedversions(InTrac78X,typeRorI)
The compressed air connections are used to pressurize the piston of the insertion rod either from above or from below. This allows the insertion rod and hence the electrode / sensor to be moved either downwards (into the “Measure” position) or upwards (into the “Service” position). The pis­ton remains pressurized with compressed air at the respective end position.
In housing versions equipped with a position indication system, a pneumatic (InTrac78XR) or pneumatic with inductive (InTrac78XI) signal is triggered when the insertion rod is in the respec-
tive end position (“Measure” or “Service”). This signal is transmitted to a remote indicator (not included with housing).
On top of the sensor position signal, the level marking both “Measure” and “Service” are visible on the Sensor Position Rod and Sensor Pressurized Cylinder to provide additional sensor status in­formation. It is necessary to ensure that the sensor position rod / sensor pressurized cylinder has fully reached the “Service” level marking before carrying out any sensor maintenance.
Depending on the remote programmable logic controller (PLC) and the operating system (e.g.,
EasyClean), the InTrac78X (type R or I) is programmable for sensor backwash cleaning and
while the sensor is retracted from the process, for cleaning, or calibration with buffer (pH sensor).
ATTENTION! The housings are equipped with an intelligent sensor locking system: if no elec-
h
trode / sensor has been fitted in the housing, it is not possible to retract / send back the insertion rod into the process.
CAUTION! The insertion and retraction speed of the insertion rod can be set as desired by
a
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
means of the adjustment screw on the throttle valve and by adjustment of the air supply pressure. Please note that for proper operation the compressed air supply must be at a higher pressure than the process pressure.
Page 35
Product descriptions 35
ATTENTION! The compressed air must be free of oil, dust and water. The recommended com-
h
CAUTION! For backwash cleaning, the cleaning liquid (from the flushing connection “IN”)
a
ATTENTION! The cleaning liquid pressure must not exceed 6bar. Install a pressure reducer if
h
ATTENTION! The sensor pressure indicator (InTrac784X) indicates the sensor internal pres-
h
DANGER! Process medium can present a hazard to human life and to the environment.
a
pressed air pressure is between 4to6bar.
must supply higher pressure than the process pressure. Insufficient cleaning liq­uid pressure may result in process media entering the flushing connections.
necessary.
sure. To ensure the sensor electrolyte continues to flow freely from the sensor junction during operation, sensor internal pressure must be higher than the pro­cess pressure. Please refer to sensor instruction manual for details.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 36
36 Installation and start-up

4 Installation and start-up

4.1 Preparation of the equipment

The retractable housings are mounted and fixed on a vessel (reactor, tank, pipe, etc.) either with flange connection, or via a NPT thread. Various DIN or AISI nominated process connections are available.
ATTENTION! Attachment / gasket between the flange connection and / or threaded seal PTFE
h
CAUTION! In the case of inclined mounting, the angle of the housing must be equal to or
h
tape for the NPT connection (if applicable) is the responsibility of the customer.
In order to ensure optimum performance of the retractable housings at all time, please pay attention to the following installation instructions:
– The retractable housing can be mounted vertically or in an inclined position.
greater than 15° above the horizontal.
admissible installation
– The retractable housing is to be mounted in such a position that there is always enough clear-
ance available for its correct functioning (correct “Measure” position in the sample medium) as well as for maintenance work (checks, fitting and removal of the electrodes / sensors). The relative dimensions can be seen on the drawings in the relative appendix to this instruction manual, or in the specifications. It is recommended to have clearance of at least 0.5 to 1 m for maintenance work purposes.
– Mounting of the retractable housing in exposed positions should be avoided. If this is not pos-
sible, relative measures to protect against damage or interference must be taken.
– Outdoor installation of the retractable housing is possible. However, avoid direct sunlight, rain,
water and / or excessive heat. The working ambient temperature is –10…70 °C.
ATTENTION! For outdoor installation of the InTrac78XM housing, the sensor interlocking bolts
h
should be installed facing a position where water splashes (when happen) shall not ingress into the drive train via the sensor interlocking bolts.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 37
Installation and start-up 37
4.2 Fittingandinstallationwork
CAUTION! In all installation work described below, ensure that the equipment in which the
a
CAUTION! For Ex-environment installation: Potential electrostatic charging hazard to the housing
a
housing is to be installed is in a non-hazardous condition (depressurized, empty, rinsed, vented, etc.).
flange and wetted parts. Proper discharge is needed before carrying out the housing installation or after service, and / or removing from the process. See Chapter “1.7 In­stallation in potentially explosive areas (hazardous areas)“ on page 16 and Chap­ter 2 “Important notes for housing uses in Ex-Classified area” on page 22.
4.2.1 Fitting the housing
Choose the installation point carefully to ensure the sensor is always filled with process medium even when sample flow is interrupted. Any air bubble in the medium may perturb the measure­ment. Consider installing proper air venting upstream of the housing.
CAUTION! The sensor protective cage (in “Measure” position) must have a clearance of at
a
least 3 – 5cm from the vessel / pipe / tank etc., and wall in the installation. Refer
respective housing insertion depth dimension for detail.
4.2.1.1 Fittingviaaange
1. Clean the sealing surfaces of the flange (housing and flange connection on vessel) and check for damage.
CAUTION! If the process media / reaction products are considered to be dangerous, it is im-
a
2. Use the appropriate flange gasket which is suitable for the process condition. Replace if necessary.
ATTENTION! In the case of housings made of special materials:
h
Check that the seal is present on the flange of the retractable housing and is not
3. Check the vessel to ensure that there are no obstacles in the insertion direction which could hinder the motion of the insertion rod.
4. Position the housing on the flange connection, align, and tighten evenly crosswise using the prescribed number of bolts and nuts.
perative that an embedded seal is used at the flange interface and / or a splash guard mounted. Fitting of the retractable housing with damaged flange connections is not allowed and can present a hazard to persons and / or lead to material damage.
damaged. Replace if necessary. Ensure correct quality and positioning of the seal.
4.2.1.2 Fitting via NPT external thread
1. Ensure the threads are free of particles or dirt.
2. Wind PTFE tape (or appropriate sealing tape which is suitable for the process condition) around the external male thread.
3. Screw the housing carefully into the female socket.
4. Check the Installation for leaks.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 38
38 Installation and start-up
11
4.2.1.3 Fittingviaaweld-insocket(withIngoldDN25)
1. Clean the centering spigot of the retractable housing and the bore of the weld-in socket, checking for damage
CAUTION! Fitting of the retractable housing with a damaged spigot or into a damaged weld-
a
2. Check the vessel / tank / piping etc. to ensure that there are no obstacles in the insertion direc­tion which could hinder the motion of the immersion tube.
3. Check the O-ring on the spigot for damage and replace it if necessary. Lightly smear the O-ring with grease. Ensure correct quality and positioning of the O-ring.
4. Position the housing on the weld-in socket and carefully insert the spigot into the bore.
5. Finally, tighten the ring nut until the connection is completely sealed.
CAUTION! Checking of the sealing / tightness of the process adaptor is the responsibility of
a
in socket is not allowed and can present a hazard to persons and / or lead to material damage.
the operator, who must also guarantee same by adopting appropriate measures. Additional safety provisions are necessary if the connection is subject to stress from vibrations.
4.2.2 Adjustingthesensorprotectivecage
A sensor protective cage is fitted to the lower end of the insertion rod. For visual inspection, a sym­bol is visible on the housing body and parallel to the sensor protective cage opening (see picture).
For maximum sensor tip protection against any particles in the medium, the sensor protective cage can be adjusted in respect of the medium flow.
1. Make sure the housing is fully retracted (or in “Service” position)
2. Loosen the housing clamp screws “1”.
3. Turn the cylinder drive train clockwise / anticlockwise until the preferable sensor protective cage position has been reached.
4. Tighten the housing clamp screws “1”.
DANGER! Before loosen the housing clamp screws, the pipe / vessel / tank etc must be iso-
a
lated (depressurize, empty, rinse, purge and vent, etc.).
11
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 39
Installation and start-up 39
4.2.3 Attachingtheushinglines
The InTrac781 / 784 is equipped with “In” and “Out” flushing connections. The flushing connection
labelled “In” is the inlet for the sensor cleaning liquid (i.e. clean water). The flushing connection labelled “Out” is the outlet for the cleaning medium drain out from housing during the sensor cleaning process.
To achieve the optimum sensor cleaning, the cleaning liquid inlet pressure is recommended to be at least 1 bar. The cleaning liquid pressure must not exceed 6bar. Install a pressure reducer if necessary.
During the insertion or retraction motion, if the insertion rod comes to rest in other than the defined end positions “Measure” and “Service”, process media can escape to the environment via the flushing chamber and the flushing lines. This could create a hazard for personnel, equipment and the environment.
When constructing the flushing water installation, use only materials which are resistant to corro­sion and chemicals. The cleaning medium must be considered beforehand. Ensure the cleaning medium will not react with the process medium or harm the housing wetted material.
Recommendations for proper operation and safe use
If cleaning of the sensor is not desired, the flushing connections (both “In” and “Out”) must be sealed by pegs, or pipe plugs as leakage of process liquid through the open flushing connections is possible. lf flushing lines must be equipped with check valves which may only be opened when the insertion rod is in the retracted position (“Service”). After maintenance work on the sensors, the valves must be closed.
Installing pipe plugs / pegs to flushing connection
When ordering InTrac781/ 784 housings, flushing connections ‘[G14] – G¼" thread female’ or
‘[N14] – ¼" NPT female’ [1] are delivered together with the housing. In cases where cleaning of
the sensor is not desired, the pipe plugs / pegs (sourcing locally) can be directly installed to the flushing connections [1]. Alternatively, the flushing connections [1] can be removed by turning
1
them anti-clockwise, and the pipe plugs / pegs G
In cases where a flushing connection ‘[G18] – G
8" can be directly installed to [2].
1
8" thread female’ is ordered, flushing connections
[1] are not included in housing delivery. Flushing connections can be directly installed to[2]. Ifcleaning of the sensor is not desired, the pipe plugs / pegs G1⁄8" can be directly installed to [2].
1
METTLER TOLEDO offer G
8" pipe plugs in different material. See spare part list for detail.
In cases where flushing connection ‘[00_] – Without (blind plugs included)‘ is ordered. Two blind
1
plugs G
ATTENTION! Avoid excessive force apply when installing pipe plugs / pegs to the housing.
h
8" are directly installed to [2] in the housing delivery.
Excessive forces can irreversibly damage the housing thread.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 40
40 Installation and start-up
2
1
Flushing lines and valves installation:
1. Install appropriate valves i.e. ball valve (and / or check valve if applicable) upstream of the flushing connection “In” and downstream of flushing connection “Out”. Note: The valves are not in the scope of delivery.
2. Connect the cleaning line to the flushing connection “In”.
3. Connect the drain line to the flushing connection “Out”.
4. Check tightness of the piping, valves connections.
5. When motorized valves are preferred, connect the cable to the operation system accordingly. Do the operating system set-up (if applicable).
2
1
DANGER! When retracting the insertion rod from the process, small quantities of process me-
a
dium will remain attached to the electrode / sensor and will thus enter the discharge line during flushing procedures. If the process medium contains toxic, caustic sub­stances damaging to the environment, then it is imperative to consult local regula­tions relative to the design and construction of effluent (treatment) installations.
4.2.3.1 Normal sensor cleaning
When the flushing lines are properly installed, the sensor can be retracted from the process and remains in the housing while it is cleaned with cleaning liquid. It is not necessary to dismantle the sensor and clean it manually in the lab.
1. Retract the sensor from the process until the housing is in the “Service” position.
2. Open both valves located upstream and downstream of the flushing connections “In” / “Out”.
3. Let the cleaning liquid clean the sensor for 2 minutes (or longer if appropriate).
4. Close both valves.
5. Insert the sensor into the process.
ATTENTION! The cleaning liquid pressure must not exceed 6 bar. Install a pressure reducer
h
ifnecessary.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 41
Installation and start-up 41
4.2.3.2 Backwashsensorcleaning
For special application where particles and dirt are present in the process medium, backwash cleaning is recommended. As soon as the housing start to retract from the process, backwash cleaning flushes out the dirt / particles which stick on the sensor tip / sensor protective cage. This allows the majority of the dirt to be flushed out into the process.
1. A valve upstream of the flushing connection “In” is open but the valve downstream of the flushing connection “Out” remains closed.
2. Retract the sensor from the process.
3. When the sensor is fully retracted (or in “Service” position), open the valve downstream of the flushing connection “Out”.
4. Let the cleaning liquid clean the sensor for 2 minutes (or longer when appropriate).
5. Close both valves.
6. Insert the sensor into the process.
CAUTION! For backwash cleaning, the cleaning liquid (from the flushing connection “IN”)
a
ATTENTION! The cleaning liquid pressure must not exceed 6 bar. Install a pressure reducer
h
must supply higher pressure than the process pressure. Insufficient cleaning liq­uid pressure may result in process media entering the flushing connections.
ifnecessary.
4.2.4 Installation for calibration of the measurement system with electrode/sensorinstalled
If the measurement system has to be cleaned and / or calibrated with the electrode / sensor installed, a distributor with three connections is attached to the flushing connection “In”.
Make three connections as shown in the following schematic diagram:
B
A
Note! When the housing is operated with EasyClean EC400, it is recommended to order the
h
InTrac housing with G ¼" thread female flushing connection.
C
• Connection “A” to the cleaning line inlet via a control valve.
• Connections “B” and “C” to the corre­sponding buffer solution supply, each via a non return valve.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 42
42 Installation and start-up
4.2.5 Pneumatic-connections(ForInTrac78X,typeR)
The InTrac78XR is operated with compressed air. The compressed air is used to drive the drive train of
the housing, so the sensor can retract or insert into the process. The extension of the cylinder of the drive train is fitted with four compressed air connections. For proper connection, the items below are required.
– Two pneumatic tubes with outter diameter 4 mm – Two pneumatic tubes with outter diameter 6 mm
The four compressed air connections are labelled numerically as below.
“1”: Air supply for “Service” position; pneumatic tube with outter diameter 6 mm – “2”: Air supply for “Measure” position; pneumatic tube with outter diameter 6 mm – “3”: “Service” position reply signal; pneumatic tube with outter diameter 4 mm – “4”: “Measure” position reply signal; pneumatic tube with outter diameter 4 mm
1
1
4
2
3
ATTENTION! – The pressure in the control air supply to the housing must reach 4 bars.
h
– The compressed air must be free of oil, dust and water. The recommended
compressed air pressure is between 4 to 6 bar.
– When no reply signal is required, the compressed air connections “3” and “4”
must be properly sealed.
CAUTION! The insertion and retraction speed of the insertion rod can be set as desired by
a
means of the adjustment screw on the throttle valve and by adjustment of the air supply pressure. Please note that for proper operation the compressed air supply must be at a higher pressure than the process pressure.
4.2.6 Pneumaticconnection(ForInTrac78X,typeI)
The InTrac78XI is operated with compressed air. There are two compressed air connections, and
two inductive check back sensor connections. The inductive check back sensor is installed to the housing prior to shipment. The connections are labelled numerically as below:
“1”: Air supply for “Service” position; pneumatic tube with outter diameter 6 mm – “2”: Air supply for “Measure” position; pneumatic tube with outter diameter 6 mm – “3”: “Service” position inductive check back sensor – “4”: “Measure” position inductive check back sensor
The compressed air is connected to connection “1” and “2” respectively via pneumatic tubes.
Two cable sockets (M12, 4-pin, NAMUR) are required to connect the inductive check back sensor
to the process control. The inductive check back sensor’s specification are as below:
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 43
Installation and start-up 43
Description Data
Switching element function NAMUR, NC Nominal voltage, U Switching frequency, f 0…2000 Hz Current consumption Measuring plate not detected ≥ 3 mA
Connection type Device connector M12
0
8.2 V (Ri approx. 1 k Ohm)
Measuring plate defected ≤ 1 mA
× 1, 4-pin
Connection:
Wire colors in accordance with EN60947-5-6
– Terminal 1 = Brown – Terminal 2 = Blue
For EasyClean 200 e only: For installation in ex-classified area, additional sensor output interface terminals (2 pcs.) are re-
quired to connect the inductive check-back from sensor (from InTrac 78XI housing) to the Easy­Clean 200 e system. The EasyClean 200 e system and the sensor output interface terminals must be installed outside the ex-classified area. The sensor output interface terminals (not in scope of delivery) can be purchased from Pepperl + Fuchs
®
; model number KCD2-E2L.
Please consult your local representative for more detail.
ATTENTION! – The pressure in the control air supply to the housing must reach 4 bars.
h
– The compressed air must be free of oil, dust and water. The recommended
compressed air pressure is between 4 to 6 bar.
CAUTION! The insertion and retraction speed of the insertion rod can be set as desired by
a
means of the adjustment screw on the throttle valve and by adjustment of the air supply pressure. Please note that for proper operation the compressed air supply must be at a higher pressure than the process pressure.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 44
44 Installation and start-up
4.2.7 Fittingtheelectrode/sensor
RetractablehousingsInTrac781
1
2
3
5 – 7 mm
1. Set the housing to the “Service” position. Visually check if the housing is in “Service” position (level mark from sensor position rod; “3”). For InTrac781M, it is possible to check if the sensor locking bolt is in the “Service” position.
2. Unscrew the cable antikink “1”.
3. Unscrew the cable grommet “2”.
4. Unscrew the sensor position rod “3”.
5. Check that the correct type of electrode / sensor has been selected (see Chapter 10.1 “Elec­trode / sensor selection” on page 62). For optimum sensor insertion length, the sensor washer must be of the range 5 to 7 mm. Sensors with a shorter washer must use the sensor spacer (included in the delivery).
6. Check the electrode / sensor for damage (e.g. breakage of electrode).
CAUTION! A damaged electrode / sensor must never be installed.
a
7. Check washers and O-rings on electrodes / sensors and replace if necessary.
8. Remove watering cap from the tip of the electrode / sensor and rinse electrode / sensor with water.
9. Carefully insert electrode / sensor into the immersion rod and screw in by hand as far as it will go.
DANGER! Under no circumstances whatsoever may tools be used.
a
10. Remove the cover cap of the electrode / sensor screw cap connection.
11. Run the sensor cable (open end) through the sensor position rod “3”, cable grommet “2”, and antikink protection “1”. Connect the plug to the electrode / sensor.
12. Mount sensor position rod “3”, follow by cable grommet “2” and tighten by hand. Finally, also hand-tighten the cable antikink protection “1”.
13. Connect the sensor open end to the transmitter.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 45
Installation and start-up 45
46
RetractablehousingInTrac784
23
CAUTION! Tightening of the ring nut “5” and the plastic adapter “4” (see diagram above) in
a
1. Set the housing to the “Service” position.
2. Unscrew the cable antikink protection “2”.
3. Unscrew the cable grommet “3”, the plastic adapter “4”, and finally “5”.
4. Check that the correct type of pH / redox electrode has been selected (see Chapter 10.1 “Electrode / sensor selection” on page 62).
5. Check the electrode for damage.
CAUTION! A damaged electrode must never be installed.
a
6. Remove the watering cap from the tip of the electrode, the stopper from the filling port, and the rubber band. Then rinse the electrode with water.
DANGER! After removing the stopper, do not tilt the electrode, as reference electrolyte may
a
7. Check the level of the reference electrolyte in the electrode and top up if necessary (see in­struction sheet of the relative electrode).
8. Carefully insert the electrode into the insertion rod until it comes up against the PTFE saddle.
the wrong order, can lead to breakage of the pH / redox electrode. It is therefore essential to follow the procedure described below:
spill out through the open filling port.
5
1
ATTENTION! If installing in an inclined position, ensure that the marking “Position electrode this
h
9. Check gasket “6” at upper part of housing for damage and replace if necessary. Mount up­per part of housing and tighten ring nut “5” by hand.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
side up” points upwards. In this position no reference electrolyte can flow out through the filling port (assuming that the fill level is correct).
Page 46
46 Installation and start-up
CAUTION! Ring nut “5” of upper part may be tightened only with plastic adapter “4”
a
10. Check condition of seal of plastic adapter “4” and replace if necessary. Mount plastic adapter “4” and tighten by hand.
11. Remove the cover cap of the plug connection of the electrode.
12. Run the sensor cable (open end) through the cable grommet “3” and antikink protection “2”. Connect the plug to the electrode.
13. Mount cable grommet “3” and tighten by hand. Finally, also hand-tighten the cable antikink protection “2”.
14. Connect the sensor open end to the transmitter.
15. Set the compensation pressure in the housing. The compensation pressure can be set with a bicycle pump via the valve assembly at the pressure gauge, or established by attaching a permanent (dry, oil-free and filtered) compressed air supply (using the pressure connection set included in the standard equipment supply).
ATTENTION! To ensure electrolyte flow from the reference electrode to the sample medium the
h
removed.
air pressure in the upper part of the housing must be at least 0.5 bars and maxi­mum 2 bars above that of the sample medium (take the hydrostatic pressure of the sample medium into account).
ATTENTION! If there is no (positive) pressure differential to the sample medium, no reliable
h
measurements can be taken.

4.3 Startup procedures for housings

4.3.1 Startup of manually operated housings
Before startup, all fitting and installation work (see Chapter 4.2 “Fitting and installation work” on page 37) must have been completed!
Each time before startup, check the measuring system. Inspect the electrode / sensor assembly and examine for leaks from housing and apparatus (see also Chapter 7 “Troubleshooting” on page 57). Do not commence operation until the measuring system has been checked and any necessary corrective action taken.
Before startup of a housing in hazardous area, it is to be clarified beyond doubt that it is permitted to use the housing in question in conjunction with the other associated plant resources. Read and fully understand Chapter 1.7 “Installation in potentially explosive areas (hazardous areas)” on page 16 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22 before carrying out the start-up.
Proceed as follows for startup:
1. For InTrac78XM start-up, press the “Service” interlock bolt continuously and turn the man­ual drive train clockwise carefully.
2. The “Measure” interlock bolt pops up when the housing is fully inserted into the process.
ATTENTION! Only turn the manual drive train by hand. No excessive force should be applied
h
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
when turning the manual drive train.
Page 47
Installation and start-up 47
4.3.2 Startup of pneumatically operated housings
Before startup, all fitting and installation work (see Chapter 4.2 “Fitting and installation work” on page 37) must have been completed!
Each time before startup, check the measuring system, inspect the electrode / sensor assembly and examine for leaks from housing and apparatus (see also Chapter 7 “Troubleshooting” on page 57). Do not commence operation until the measuring system has been checked and any necessary corrective action taken.
Before startup of a housing in a hazardous area, it must be established beyond doubt that it is permitted to use the housing in question in conjunction with the other associated plant resources. Read and fully understand Chapter 1.7 “Installation in potentially explosive areas (hazardous areas)” on page 16 and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22 before carrying out the start-up.
Proceed as follows for startup:
1. Open the air supply for the control system. Check:
– the insertion and retraction motion of the insertion rod, as well as the status of the indica-
tors for the respective end positions.
– the control air system for tightness. Immediately seal any leaking connections.
2. Drive the insertion rod into the “Service” position and open flushing water supply (at least
1bar). Check flushing water system for tightness. Immediately seal any leaking connections.
3. If the system is equipped for calibration of the measuring system, move the housing into the “Service” position and perform a calibration (see Chapter 5.4 “Calibrating the measuring system” on page 52). Check:
– for correct calibration procedure. – the buffer solution systems for tightness. Immediately seal any leaking connections.
On successful completion of all function checks, the equipment in which the housing is installed can be started up.
CAUTION! Following startup of the equipment, particular care must be taken in any manipu-
a
lations at the retractable housing. Therefore it is essential to pay close attention to the information in Chapter 5.1 “Important information for everyday operation” on page 51.

4.4 Shutdown procedure for manually and pneumatically operated housings

1. Move housing to the “Service” position.
2. Open flushing water supply, flush electrode / sensor and close flushing water supply again.
3. Isolate the pipe / tank / vessel etc, from the housing (depressurize, empty, rinse, purge and
vent, etc.).
4. Close control air feed to the housing (except InTrac78XM).
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 48
48 Installation and start-up
4.5 Dismantlingwork
4.5.1 Removingtheelectrode/sensor
CAUTION! Removal and installation of the electrodes / sensors may only be performed in the
a
InTrac781retractablehousings
“Service” position of the housing.
1 2
3
1. Set the housing to the “Service” position.
ATTENTION! Visually check the sensor position rod “3” and ensure the housing is fully re-
h
2. Open flushing water supply, flush electrode / sensor and close flushing water supply again.
3. Unscrew cable antikink “1”.
4. Unscrew cable grommet “2”.
5. Unscrew sensor position rod “3”.
6. Disconnect the sensor cable from the electrode / sensor.
7. Unscrew electrode / sensor and carefully remove from insertion rod.
tracted (at “Service level” mark).
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 49
Installation and start-up 49
InTrac784retractablehousing
CAUTION! If the plastic adapter “4” and the ring nut “5” are loosened in the wrong order, this
a
1. Set the housing into the “Service” position.
2. Open flushing water supply, flush electrode / sensor and close flushing water supply again.
3. Depressurize the upper part of the housing by slightly loosening the valve assembly at the
pressure gauge or by interrupting the compressed air supply. Retighten the valve assembly.
4. Remove cable antikink “2” and cable grommet “3”.
5. Disconnect the sensor cable from the electrode.
6. Remove plastic adapter “4”.
7. Undo ring nut “5” and remove upper part “7”.
8. Carefully withdraw the electrode from the insertion rod.
can lead to breakage of the pH / redox electrode. It therefore essential to following the procedure described below.
23 475
1
ATTENTION! Specific information on the electrode (calibration / alignment with the measuring
h
system, storage, etc.) is to be found in the relative documentation for the electrode or measuring system.
4.5.2 Removing the retractable housing
1. Shut down the housing (see Chapter 4.4 “Shutdown procedure for manually and pneu-
matically operated housings” on page 47).
2. Remove the electrode / sensor (see Chapter 4.5.1 “Removing the electrode / sensor” on
page 48).
3. Dismantle flushing water and control air lines.
CAUTION! Close the connection in order to prevent ingress of dirt.
a
4. Undo ring nut or flange connection and carefully remove housing.
CAUTION! Never place or support the housing on the front ends of the insertion rod (risk of
a
damage).
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 50
50 Installation and start-up
w
4.6 UseofretractablehousingsinconjunctionwithEasyClean
The retractable housing can be employed in conjunction with METTLER TOLEDO’s cleaning and
calibration system, EasyClean. This enables full automation of the measuring system. Current regulations together with the connection instructions given in the relative Instruction Manual, as well as directives governing pneumatic and water connections are to be fully observed at all times.
CAUTION! In the case of installations intended for operation in hazardous areas, only Easy-
a
Clean 400X systems may be used in conjunction with a InTrac78X housing.
Distributed Control System (DCS)
Transmitter M700 (X)EasyClean 400 (X)
Pressurized water Pressured air
Media adapter
ith dosing pumps for calibration buffer and cleaner (3.5 l)
Power in
Interconnection cable
InTrac housing and pH electrode
Media connection
5 m (10 m optional)
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 51
Operation 51

5 Operation

5.1 Important information for everyday operation

CAUTION! No attempt may be made to move the insertion rod into the “Measure” position
a
During operation,
never loosen the lines of the pneumatic system or of the flushing and buffer solutions. – never remove fastening components (screw / bolts of flange, ring nut, etc.).
If any malfunctions occur during operation, the equipment in which the housing is installed must first be made safe before any corrective measures are taken.
For all work on the equipment during everyday operation, wear the stipulated protective clothing (protective goggles, gloves, breathing apparatus, etc).
5.2 Inspectionworkineverydayoperation
without an electrode / sensor properly installed!
The following inspection work should be performed in everyday operation:
• Check all systems such as air, flushing water and buffer solutions for leaks. Refill buffer solu­tions if necessary.
• Visually check fastenings (ring nut, flange, NPT-thread) of the housing at the vessel for firm seating and leaks.
• Check the condition of the electrode / sensor. A faulty or damaged electrode / sensor must be replaced immediately.
• Check the inspection window for leaking (picture).
Additionalchecktothehousingwithpressurecompensation(InTrac784):
• Check functioning of pressure gauge.
• Check air pressure in upper part (pressure gauge). The air pressure must be at least 0.5 bars and maximum 2 bars above that of the sample medium (take hydrostatic pressure of the sample medium into account) to ensure the flow of electrolyte from the reference electrode to the sample medium.
ATTENTION! The desired overpressure is set with a bicycle pump via the valve assembly at the
h
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
pressure gauge or established by means of a compressed air supply.
Page 52
52 Operation
• Check level of the reference electrolyte. The level of the reference electrolyte steadily sinks due to outflow through the diaphragm. If the liquid level has sunk to a level below the mouth of the bulb (reservoir) of the pH / redox electrode, then the reference electrolyte must be topped up (see Chapter 6 “Maintenance” on page 53).
5.3 Cleaningtheelectrode/sensor
The electrode / sensor must be cleaned before removal, before calibration of the measurement system or at regular intervals during operation (depending on the process medium). Proceed as follows:
1. Move the housing into the “Service” position.
2. Open both valves locatged upstream and downstream of the flushing connections “In” / ”Out”.
3. Let the cleaning liquid clean the sensor for 2 minutes (or longer if appropriate).
4. Close flushing water supply (or both valves).

5.4 Calibrating the measuring system

The frequency of calibration of the measurement system depends on the electrode / sensor type and the sample medium. To calibrate the measurement system, proceed as follows:
1. Retract the insertion rod into the “Service” position.
2. Open flushing water supply, flush electrode / sensor and then close flushing water supply
again (see Chapter 5.3 “Cleaning the electrode / sensor” on page 52).
3. Remove electrode / sensor (procedure, see Chapter 4.5.1 “Removing the electrode / sensor”
on page 48).
ATTENTION! This is only necessary if your installation does not allow calibration of the mea-
h
4. Perform calibration in accordance with the operating instructions for the respective elec-
trode / sensor and transmitter.
5. Re-install electrode / sensor (procedure, see Chapter 4.2.7 “Fitting the electrode / sensor” on
page 44).
ATTENTION! This is only necessary if your installation does not allow calibration of the
h
6. Open flushing water supply, flush the electrode / sensor and then close flushing water sup-
ply again.
7. Set the insertion rod into the “Measure” position.
surement system with the electrode / sensor installed.
measurement system with the electrode / sensor installed.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 53

Maintenance 53

6 Maintenance

6.1 Important information on maintenance

Establish a service plan adapted to your process.
ConsultyourlocalMETTLERTOLEDOrepresentativefordetails.
DANGER! The information and instructions given in Chapter 1 “Introduction” on page 9
a
CAUTION! Maintenance and service work on the housings may only be carried out by ap-
a
CAUTION! It is possible that the process medium could harm the environment and your
a
and Chapter 2 “Important notes for housing uses in Ex-Classified area” on page 22 must be fully adhered to.
propriately trained personnel. Only the maintenance and repair work described in the following Sections may be performed on the retractable housings.
health (toxic, caustic, corrosive, etc.). Therefore, ensure that the equipment is in a non-hazardous condition before you start with any maintenance work.
DANGER! Before startup, the housing must be checked to ensure that:
a
– the right types of O-ring have been selected and that they are undamaged and
correctly positioned (flushing chamber, insertion rod).
ATTENTION! Only original spare parts from METTLER TOLEDO may be used, otherwise all
h
guarantees become automatically invalid.
6.2 Toppingupreferenceelectrolyte(InTrac784)
The work described below applies only for pH / redox electrodes with liquid reference electrolyte.
Inorder to top up the reference electrolyte, proceed as follows:
1. Remove electrode (see Chapter 4.5.1 “Removing the electrode / sensor” on page 48).
ATTENTION! The reference electrolyte may not be topped up with the electrode installed.
h
2. Top up reference electrolyte (Electrolyte No. see marking “Refill” on the electrode).
CAUTION! Take care not to exceed the maximum filling level.
a
3. Re-install the electrode (see Chapter 4.2.7 “Fitting the electrode / sensor” on page 44).
Do not allow any spillage of reference electrolyte to remain in the housing. Wash out and dry the housing.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 54
54 Maintenance
6.3 Replacing seals in contact with the process medium (InTrac78X)
Depending on the application, it is a must to replace the seal at least every year. With aggres- sive media, the seals may need to be changed at correspondingly shorter intervals. The seal
condition may visually inspect via the housing inspection windows. The seals must be replaced when there is leaking visible on the inspection windows. Frequent insertion and retraction motion of the insertion rod can also have an influence on the necessary maintenance intervals.
ATTENTION! Recommendations by METTLER TOLEDO concerning maintenance intervals are
h
To replace the seals, proceed as follows:
1. Set the housing to the “Service” position.
ATTENTION! Visually check the sensor position rod and ensure the housing is fully retracted
h
based solely on experience gained in standard applications and in no way bind­ing or an admission of any guarantee liability whatsoever on the part of the man­ufacturer / supplier. Depending upon the degree of aggressiveness of the process medium, the necessary maintenance intervals to support smooth operation of the equipment may be correspondingly shorter.
(at “Service level” mark).
2. Perform shutdown procedure as described in Chapter 4.4 “Shutdown procedure for manu-
ally and pneumatically operated housings” on page 47
3. Removing the sensor / electrode as described in Chapter 4.5.1 “Removing the electrode / sen-
sor” on page 48
4. Remove the retractable housing as described in Chapter 4.5.2 “Removing the retractable
housing” on page 49
CAUTION! Never place or support the housing on the front ends of the insertion rod (risk of
a
5. Loosen the four housing clamp screws.
ATTENTION! Usually the housing clams remain on the housing drive train and should be eject-
h
6. Remove the drive train “1” and cleaning chamber “2” from the process flange connection
“3” as shown.
7. There are a total of seven seals have to be replaced (arrows A to G in picture). Replace all
seven seals with new ones.
damage).
ed by gently pushing it with the small slot screw drive (flat-end).
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 55
Maintenance 55
6.3.1 Flange
G
E
F
• Seal “A”: diameter 18.72 × 2.62 mm
• Seal “B”: diameter 10.77 × 2.62 mm
• Seal “C” and “D”: diameter 21.95 × 1.78 mm
• Seal “E”: scraper (diameter 19 × 6 × 1 mm)
• Seal “F”: diameter 21.89 × 2.62 mm
• Seal “G”: diameter 29.82 × 2.62 mm
6.3.2 Ingold DN 25
DC AB
32 1
G
DC AB
E
F
32 1
• Seal “A”: diameter 18.72 × 2.62 mm
• Seal “B”: diameter 10.77 × 2.62 mm
• Seal “C” and “D”: diameter 21.95 × 1.78 mm
• Seal “E”: scraper (diameter 18 × 6 × 1 mm)
• Seal “F”: diameter 18.77 × 1.78 mm
• Seal “G”: diameter 21.95 × 1.78 mm
ATTENTION! Attention! When installing seals “E”, and “F” follow these instruction. If the seals
h
are installed in the wrong order, it can lead to leaking of media into the housing.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 56
56 Maintenance
8. Carefully install seal “F” to the housing
9. Slighly bend the seal / scraper “E” as pictured.
10. Gently insert the seal “E” into the cleaning chamber.
ATTENTION! Never bend the seal / scraper “E” with excessive force. By doing this, the seal will
h
11. Install the drive train “1” and cleaning chamber “2” to the process flange connection “3”.
Mount the housing clamps and fasten tightly with the four screws.
ATTENTION! Usually the housing clamps need to be gently pushed, so they can fit between the
h
12. Install the housing to the process line / vessel / tank etc as described in Chapter 4.2 “Fitting
and installation work” on page 37
13. Fitting the electrode / sensor as described in Chapter 4.2.7 “Fitting the electrode / sensor” on
page 44.
14. Follow start up procedure as described in Chapter 4.3 “Startup procedures for housings” on
page 46.
ATTENTION! Details of replacements and examination of the seals are to be documented in the
h
CAUTION! Only perform the dismantling work described and instructed in this Chapter and
a
be irreversibly damaged.
drive train “1” and the flange connection “3”. The clamps should never block the inspection windows.
maintenance log.
replace only the seals listed above. Check that the correct types of seals have been selected, examine for any signs of damage and for correct seating.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 57

Troubleshooting 57

7 Troubleshooting
In this chapter you will find an overview of possible malfunctions which could appear during opera­tion of the retractable housings, their cause, as well as a guide to appropriate remedial measures.
DANGER! It is essential to comply with the safety instructions given in Chapter 1 “Introduc-
a
Malfunction Cause Remedial action
Insertion rod cannot
beinserted.
Insertion rod remains stuck between the end positions “Service” and “Measure”.
Sample medium dis-
charges via ushing
lines during normal operation.
Sensor breaks fre­quently.
Sample medium escapes from the inspection windows.
Sample medium
escapes at the ange
connection / fastenings.
Sample medium escapes through the NPT thread.
InTrac784: No pres­sure in upper part (pressure compensa­tion) or drops rapidly.
There is no electrode / sensor installed in the housing . Install an electrode / sensor. Loose sensor. Tighten sensor. InTrac78X type R and I: Wrong pneumatic tube con-
nection.
InTrac78X type R and I: compressed air pressure toolow.
The sensor position rod / sensor pressurized cylinder is not fully fastened.
InTrac78X type R and I: No or insufcient control air
pressure . InTrac78X type R and I: Leak in control air system. Check / make leakproof control air system. Deposits of sample medium lead to seizing of inser-
tion rod. Drive unit defective. Contact your local METTLER TOLEDO representative. The housing is stuck in between “Service” and “Mea-
sure” positions.
Seals of ushing chamber are defective. Remove housing and replace seals (see Chapter 6
Process liquid contains solids. Adjust the sensor protective cage (see Chapter 4.2.2
Sensor is not correctly installed. See Chapter 4.2.7 “Fitting the electrode / sensor” on
Seals need to be replaced. Replacing seals (see Chapter 6.3 “Replacing seals
Flange connection unevenly positioned or not fas­tened tightly enough.
Flange gasket defective. Check gasket and replace if necessary Thread insufciently sealed. Seal using PTFE tape / appropriate sealing tape
Compensation pressure not (correctly) set. Set compensation pressure with pump or check com-
Leak in upper part. Check screw ttings and seals Broken electrode. Replace electrode
tion” on page 9 and Chapter 2 “Important notes for housing uses in Ex­Classified area” on page 22.
Check pneumatic connection, Chapter 4.2.5 “Pneu-
matic-connections (For InTrac78X, type R)” on page
42. Pressure must between 4 and 6 bar. The compressed
air pressure must higher than the process pressure. Check.
Fasten the sensor position rod / sensor pressurized cylinder.
Check control air system / adjust pressure.
Remove housing and clean (see Chapter 6 “Mainte­nance” on page 53).
Follow troubleshooting as above, “Insertion rod remains stuck between the end positions “Service” and “Measure”.
“Maintenance” on page 53).
“Adjusting the sensor protective cage” on page
38).
page 44.
in contact with the process medium (InTrac78X)” on
page 54).
Check positioning of ange connection and / or tighten
the fastenings
pressed air supply
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 58
58 Troubleshooting
Malfunction Cause Remedial action
Incorrect measurement
values / uctuations in
measurement data.
InTrac78XI end posi­tion is not indicated.
InTrac78X type R and I:
Compressed air emitted at inspection windows.
Defective electrode / sensor or transmitter. Check electrode / sensor and transmitter. Replace or
repair as necessary.
Improperiate installation point. Ensure the sensor tip always in contact with
medium. Air bubble may perturb the measurement. See Chapter 4.2.1 “Fitting the housing” on page
37. Power failure / loosen cable connection. Check power supply / check cable connection. Defect in inductive indicator. Air supply failure in
pneumatic indicator. Drive train defective. Contact your local METTLER TOLEDO representative.
DANGER! Breakage of the cleaning chamber can lead to mixing of process medium and
a
Contact your local METTLER TOLEDO representative.
cleaning medium.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 59
Product specications 59
8 Productspecications

8.1 Technical data

Ambient conditions
Temperature SS316L, alloy C-22: –10 to 70 °C;
PP, PVDF, PEEK, PVDH: 0 to 70 °C
Operating mode manual or pneumatic or with inductive check back
Weight InTrac781 (80 mm insertion length) approx. 4.0 kg
InTrac784 (80 mm insertion length) approx. 4.5 kg
Dimension See dimension drawing (Chapter 10.2)
Material Refer to Chapter 10.3 “Product structure (InTrac78X)” on page
66.
Cleaning liquid Max. 6 bar
Control air for Free of oil, dust and water. 4 – 6 bar pneumatic operation
(InTrac78X,typeRandI)
Pocess particle size Less than 1 mm diameter
Process condition See diagram “Maximum Process Permissible Pressure and
Temperature.
Maximum Housing Wetted Material Permissible Pressure and Temperature
18
16
14
12
10
8
6
4
Max. Permissible Pressure (bar)
2
0
020406080100 120140 160
Max. Permissible Temperature (°C)
PP
PVDH
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
PEEK Stainless steel/
PVDF
DIN1.4404/ AISI316L DIN2.4602/ AISI Alloy C22
Page 60
60 Product specications

8.2 Spare parts list

Spare parts
ForInTrac78Xwhichinoperatewithsensorlength220/225/250mm (with flange or NPT connection)
Designation Order no.
®
Set of seals FPM (Viton Set of seals FFKM (Kalrez Set of seals EPDM 30 283 339
ForInTrac78Xwhichinoperatewithsensorlength220/225/250mm (with Ingold DN 25 connection)
Designation Order no.
Set of seals FPM (Viton Set of seals FFKM (Kalrez Set of seals EPDM 30 360 072
) 30 283 337
®
) 30 283 338
®
) 30 335 858
®
) 30 360 071
ForInTrac78Xwhichinoperatewithsensorlength420/425/450mm (with flange or NPT connection)
Designation Order no.
®
Set of seals FPM (Viton Set of seals FFKM (Kalrez
) 30 283 340
®
) 30 283 341
Set of seals EPDM 30 283 342
Blind plugs for flushing connections
Designation Order no.
Blind plugs 2 × G¹⁄
Blind plugs 2 × G¹⁄
Blind plugs 2 × G¹⁄
Blind plugs 2 × G¹⁄
Blind plugs 2 × G¹⁄
8" (1.4301) 30 327 316 8" (2.4602) 30 327 317 8" (PVDF) 30 327 318 8" (PP) 30 327 319 8" (PEEK) 30 327 390
Replacement of all other parts of the housings may only be carried out by appropriately trained personnel. Please contact your local METTLER TOLEDO representative.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 61
Decommissioning, storage, disposal 61

9 Decommissioning, storage, disposal

 PLEASEREFERTOChapter1.6“Manipulationofandmaintenanceworkonthehousings”on
a
page 13!
Decommissioning may only be carried out by persons with appropriate training or by skilled technicians.

9.1 Decommissioning

Proceed as described in Chapter 4.5.2 “Removing the retractable housing” on page 49.

9.2 Storage

Store the InTrac78X in a dry place. Before storage, the housing must peroperly cleaned and dry.

9.3 Disposal

Waste electrical prouducts should not be disposed of with household waste.
It is recommended that the operator disposes of the device in accordance with local
regulations. The operator must deliver the device either to a licensed private or
public disposal company, or dispose of it himself in accordance with prevailing
regulations. Waste is to be recycled or disposed of without causing any risk to hu­man health, and without using procedures or methods that might harm the environ­ment. Check with your Local Authority or retailer for recycling advise.
Sorting
Sorting into waste groups takes place when dismantling the device. The groups are listed in the current European Waste Catalogue. This catalog is valid for all wastes, whether intended for dis­posal or for recycling.
The packaging is made up of the following materials:
– cardboard – foam plastic
Depends on the housing configuration, in general the housing may be made of the following or a combination of the following materials:
– Stainless steel (DIN 1.4404 / AISI 316 L, and / or DIN 2.4602 / AISI Alloy C22). – Polypropylene (PP) – Polyvinylidene fluoride (PVDF) – Polyether ether ketone (PEEK) – For housing InTrac 78X[I] (Inductive check back) only:
Contain electrical or electronic in the inductive check back sensor.
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 62
62 Appendices

10 Appendices

10.1 Electrode/sensorselection
This is a brief overview of pH and redox electrodes suitable for use with specific media. For further information on the extensive range of electrodes and sensors available from METTLER TOLEDO, please contact your local METTLER TOLEDO representative
SuitableElectrodes/Sensors Sensor length (mm) mm mm
pH / ORP InPro2000 250 450
InPro2000 i InPro3100 225 425
InPro3100 i (SG)
InPro3250 (SG)
InPro3250 i / SG
InPro3251
InPro3252
InPro3253 (SG)
InPro3253i / SG
InPro3300
InPro4260
InPro4260 SG
InPro4262
InPro4260 i / SG
InPro4800-1-2-i / SG
InPro4800 (SG) InPro4802 (SG)
InPro 4881 i / SG 425 InPro 4281 i / SG DXK pH-Redox-Reference El.425
DPA HA-405-DPA-SC-S8
Pt 4805-DPA-SC-S8
DPAS 405-DPAS-SC-K8S
Pt 4805-DPAS-SC-K8S Pt
DO / GPO InPro6860 i 220 420
InPro6980 i
InPro6800 / 6850 i
InPro6950 (i) / InPro6900 (i)
InPro6000 G Series
CO
2
Conductivity InPro7001-VP 225
InPro5000 (i) 220
InPro7100 (i) 425
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 63
Appendices 63
A3
A3
A2
A1
B
A3
A2
A1
B
A3
A3
A2
A1
B
A3

10.2 Dimensional drawing

InTrac781
InTrac781(operationmodeR,MandI)
A2
B
A1
A2
A1
B
A2
B
A1
InTrac781R InTrac781M InTrac781I
Sensor length [mm] 225 450 225 450 225 450 A1 [mm] 208 208 208 208 208 208 A2 [mm] 518 518 518 518 518 518 Recommended space
clearance for sensor maintenance;
590 790 590 790 590 790
A3 [mm] B [mm] 98 98 79 79 131 131
Legend: R: Pneumatic check back M: Manual I: Pneumatic, Inductive check back
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 64
64 Appendices
A3
A2
A1
B
A3
A2
A1
B
A3
A2
A1
B
InTrac784
InTrac784(operationmodeR,MandI)
A3
A2
A1
B
A3
A2
A1
B
A3
A2
A1
B
InTrac784R InTrac784M InTrac784I
Sensor length [mm]
250 450 250 450 250 450
A1 [mm] 219 219 219 219 219 219 A2 [mm] 618 618 618 618 618 618 Recommended
space clear­ance for sensor
850 1050 850 1050 850 1050 maintenance; A3 [mm]
B [mm] 98 98 79 79 131 131
Legend: R: Pneumatic check back M: Manual I: Pneumatic, Inductive check back
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 65
Appendices 65
E1
E2
D2
D1
E1
E1
OP
E2
E2
D2 D2
D1 D1
E1
E1
OP
E2
E2
D2 D2
D1 D1
E1
OP
E2
D2
D1
InTrac78Xwithvariousprocessconnections
ProcessconnectionInTrac781/784(independentofoperationmode)
Flange AISI 316 L Flange AISI Alloy C22 NPT thread Ingold DN 25
OP
D2
D1
InTrac781 / 784
(R, M or I)
D2
D1
InTrac781 / 784
(R, M or I)
E1
E2
D2
D1
InTrac781 / 784
(R, M or I)
E1
E2
D2
E1
E2
D1
InTrac781 / 784
(R, M or I) Sensor length [mm] 225 / 250 425 / 450 225 / 250 425 / 450 225 / 250 225 / 250 E1 [mm] 44 244 39 239 7 51 E2 [mm] 80 280 75 275 43 87 OP [mm] 25 D1 [mm] 31 36 31 36 31 25 D2 [mm] 19 19 19 19 19 18
Legend: R: Pneumatic check back M: Manual I: Pneumatic, Inductive check back
E1
E2
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 66
66 Appendices
10.3 Productstructure(InTrac78X)
Sensor type
7 8 1 Solid [781] solid electrolyte, DO, Cond 7 8 4 Liquid [784] liquid electrolyte (InPro2000, 465)
Operating mode and indicators
M Manual [M] R Pneumatic check back [R] I Pneumatic, Inductive check back [I]
Material,medium-wettedparts
4 4 0 4 1.4404 [4404] C 2 2 _ 2.4602 / Alloy C22 [C22_] P P _ _ PP [PP__] P V D F PVDF [PVDF] P E E K PEEK [PEEK] P V D H PVDF / Alloy C22 [PVDH]
SealingMaterial(wettedsealings)
V I Viton [Vi] K A Kalrez [Ka] E P EPDM [EP]
Sensor length
2 2 5 225 mm [225] 4 2 5 425 mm [425] 2 5 0 250 mm [250] 4 5 0 450 mm [450]
Process connection
D 0 0 Ingold DN 25 [D00] D 0 2 DN32 PN16 [D02] D 0 3 DN40 PN16 [D03] D 0 4 DN50 PN16 [D04] D 0 6 DN80 PN16 [D06] A 0 1 A150-1¼ [A01] A 0 2 A150-1½ [A02] A 0 3 A150-2 [A03] A 0 4 A150-3 [A04] N 0 2 NPT 1¼ [N02]
Flushing connections
0 0 _ Without (blind plugs
included) [00_]
1
G 1 8 G G 1 4 G¼" thread female [G14] N 1 4 ¼" NPT female [N14]
⁄8" thread female [G18]
InTrac 7 8 / / / / /
7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
PleasecontactyourlocalMETTLERTOLEDOrepresentativeformoreoptions.
Housing InTrac 78X © 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland 30 283 187 B Printed in Switzerland
Page 67
Appendices 67
© 04/2018 Mettler-Toledo GmbH, CH-8606 Greifensee, Switzerland Housing InTrac 78X Printed in Switzerland 30 283 187 B
Page 68
Sales and Service:
METTLERTOLEDOMarketOrganizations
Australia
Mettler-Toledo Limited 220 Turner Street, Port Melbourne, VIC 3207 Australia Phone +61 1300 659 761 e-mail info.mtaus@mt.com
Austria
Mettler-Toledo Ges.m.b.H. Laxenburger Str. 252/2 AT - 1230 Wien Phone +43 1 607 4356 e-mail prozess@mt.com
Brazil
Mettler-Toledo Ind. e Com. Ltda. Avenida Tamboré, 418, Tamboré BR - 06460- 000 Baru eri / SP Phone +55 11 4166 7400 e-mail mtbr@mt.com
Canada
Mettler-Toledo Inc. 2915 Argentia Rd #6 CA - ON L5N 8G6 Mississauga Phone +1 800 638 8537 e-mail ProInsideSalesCA@mt.com
China
Mettler-Toledo International Trading (Shanghai) Co. Ltd. 589 Gui Ping Road, Cao He Jing CN - 200233 Shan ghai Phone +86 21 64 85 04 35 e-mail ad@mt.com
Croatia
Mettler-Toledo d.o.o. Mandlova 3, HR - 10000 Zagreb Phone +385 1 292 06 33 e-mail mt.zagreb@mt.com
Czech Republic
Mettler-Toledo s.r.o. Trebohost icka 2283 / 2 CZ - 100 00 Praha 10 Phone +420 226 808 150 e-mail sales.mtcz@mt.com
Denmark
Mett ler-Toledo A /S Naverland 8 DK - 2600 Glost rup Phone +45 43 27 08 00 e-mail info.mtdk@mt.com
France
Mettler-Toledo Analyse Industrielle S.A.S. 30, Boulevard de Douaumont FR - 75017 Paris Phone +33 1 47 37 06 00 e-mail mtpro-f@mt.com
ISO
9001
certified
ISO
14001
certified
Germany
Mettler-Toledo GmbH Prozeßanalytik Ockerweg 3, DE - 35396 Gießen Phone +49 641 507 444 e-mail prozess@mt.com
Great Britain
Mettler-Toledo LTD 64 Boston Road, Beaumont Leys GB - Leices ter LE4 1AW Phone +44 116 235 7070 e-mail enquire.mtuk@mt.com
Hungary
Mettler-Toledo Kereskedelmi KFT Teve u. 41 HU - 1139 Budap est Phone +36 1 288 40 40 e-mail order.mt-hu@mt.com
India
Mettler-Toledo India Private Limited Amar Hill, Saki Vihar Road, Powai IN - 400 072 Mumbai Phone +91 22 4291 0111 e-mail sales.mtin@mt.com
Indonesia
PT. Mettler-Toledo Indonesia GRHA PERSADA 3rd Floor Jl. KH. Noer Ali No. 3A Kayuringin Jaya Kalimalang, Bekasi 17144, ID Phone +62 21 294 53919 e-mail mt-id.customersupport@mt.com
Italy
Mettler-Toledo S.p.A. Via Vialba 42 IT - 20026 Novate Mil anese Phone +39 02 333 321 e-mail customercare.italia@mt.com
Japan
Mettler-Toledo K.K. Process Division 6F Ikenohata Nisshoku Bldg. 2-9-7, Ikenohata, Taito-ku JP - 110-00 08 Tokyo Phone +81 3 5815 5606 e-mail helpdesk.ing.jp@mt.com
Malaysia
Mettler-Toledo (M) Sdn Bhd Bangunan Electroscon Holding, U 1-01 Lot 8 Jalan Astaka U8 / 84 Seksyen U8, Bukit Jelutong MY - 40150 Shah Alam Selangor Phone +60 3 78 44 58 88 e-mail MT-MY.CustomerSupport@mt.com
1258
Mexico
Mettler-Toledo S.A. de C.V. Ejército Nacional #340 Polanco V Sección C.P. 115 60 MX - México D.F. Phone +52 55 1946 0900 e-mail mt.mexico@mt.com
Norway
Mettler-Toledo AS Ulvenveien 92B NO - 0581 Oslo Nor way Phone +47 22 30 44 90 e-mail info.mtn@mt.com
Philippines
Mettler-Toledo Philippines Inc. 6F NOL Towers, Commerce Ave. Madrigal Business Park Ayala Alabang Muntinlupa 1780 Philippines Phone +63 2 528 8920 e-mail MT-PH.CustomerSupport@mt.com
Poland
Mettler-Toledo (Poland) Sp.z.o.o. ul. Poleczki 21 PL - 02-822 Warszawa Phone +48 22 440 67 00 e-mail polska@mt.com
Russia
Mettler-Toledo Vostok ZAO Sretensky blvd. 6/1 Office 6 RU - 101000 Mos cow Phone +7 495 621 56 66 e-mail inforus@mt.com
Singapore
Mettler-Toledo (S) Pte. Ltd. Block 28 Ayer Rajah Crescent # 05-01 SG - 139959 S ingapore Phone +65 6890 00 11 e-mail mt.sg.customersupport@mt.com
Slovakia
Mettler-Toledo s.r.o. Hatt alova 12 / A SK - 831 0 3 Bratislava Phone +421 2 4444 1221 e-mail predaj@mt.com
Slovenia
Mettler-Toledo d.o.o. Pot heroja Trtnika 26 SI -1261 Ljubljan a-Dobrunje Phone +386 1 547 49 05 e-mail darko.divjak@mt.com
South Korea
Mettler-Toledo (Korea) Ltd. 1 & 4 F, Yeil Building 21 Yangjaecheon- ro 19-gil SeoCho-Gu, Seoul 06753 Korea Phone +82 2 3498 3500 e-mail Sales_MTKR@mt.com
Spain
Mettler-Toledo S.A.E. C / Miguel He rnández, 6 9 – 71
ES - 08908 L’Hospitalet de Llobregat
(Barcelona) Phone +34 902 32 00 23 e-mail mtemkt@mt.com
Sweden
Mettler-Toledo AB Virkesvägen 10 Bo x 92161 SE - 12008 Stockholm Phone +46 8 702 50 00 e-mail sales.mts@mt.com
Switzerland
Mettler-Toledo (Schweiz) GmbH Im Langacher, Postfach CH - 8606 Gre ifensee Phone +41 44 944 47 47 e-mail ProSupport.ch@mt.com
Thailand
Mettler-Toledo (Thailand) Ltd. 272 Soi Soonvijai 4 Rama 9 Rd., Bangkapi Huay Kwang TH - 10320 B angkok Phone +66 2 723 03 00 e-mail MT-TH.CustomerSupport@mt.com
Turkey
Mettler-Toledo Türkiye Haluk Türksoy Sokak No: 6 Zemin ve 1. Bodrum Kat 34662 Üsküdar - Istanbul, TR Phone +90 216 400 20 20 e-mail sales.mttr@mt.com
USA
METTLER TOLEDO Process Analytics 900 Middlesex Turnpike, Bld. 8 Billerica, MA 01821, USA Phone +1 781 301 8800 Freephone +1 800 352 8763 e-mail mtprous@mt.com
Vietnam
Mettler-Toledo (Vietnam) LLC G Floor, SCS Building, Plot T2-4 D1 Street, Saigon Hi-tech Park Tan Phu Ward, District 9 Ho Chi Minh City, Vietnam Phone +84 28 73 090 789 e-mail MT-VN.CustomerSupport@mt.com
Management System certified according to
ISO 9001 / ISO 14001
Subject to technical changes. 04/2018 © METTLER TOLEDO. All rights reserved. Printed in Switzerland. 30 283 187 B
Mettler-Toledo GmbH, Process Analytics Im Hackacker 15, CH-8902 Urdorf, Switzerland Tel. +41 44 729 62 11, Fax +41 44 729 66 36
www.mt.com/pro
Loading...