Mettler Toledo InPro 7100-VP, InPro 7108/25-VP, InPro 7108-VP/CPVC, InPro 7108-VP/PEEK, InPro 7108-VAR-VP Instruction Manual

...
InPro 7100-VP Series
4-electrode conductivity sensors
4-Pol-Leitfähigkeits-Messzellen
Sondes de conductivité à 4 électrodes
Instruction manual
Bedienungsanleitung
InPro 7100-VP Series
52 002 092
English Page 3
Deutsch Seite 19
Français Page 35
InPro 7100-VP Series 3
InPro 7100-VP Series
4-electrode conductivity sensors

Instruction manual

Subject to technical changes without prior notice.
© It is strictly forbidden to reprint this instruction manual or any parts thereof without the written permission of Mettler-Toledo GmbH, Process Analytics, Im Hackacker 15, 8902 Urdorf, Switzerland. No section or excerpt whatsoever may be reproduced or with the assistance of electronic systems be edited, duplicated or distributed, in particular in the form of photocopies, photographs, magnetic media or other recording methods. All rights reserved, especially the right of duplication and translation as well as in regard to patent and registration rights.
InPro is a registered trademark of the Mettler Toledo Group in Switzerland and a further twelve countries.
Mettler-Toledo GmbH, 8606 Greifensee, Switzerland
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
4 InPro 7100-VP Series

Contents

1 Product description 5
1.1 Introduction 5
1.2 Equipment and scope of delivery 5
1.3 Technical data 7
2 Safety 8
2.1 Application compatibility 8
2.2 Proper utilization 8
2.3 Safety measures 9
2.4 Use in Ex-zones 10
3 Initial start-up 12
3.1 Installation 12
3.2 Electrical connections 13
4. Maintenance and troubleshooting 14
4.1 Conditions of warranty 14
4.2 Maintenance 14
4.3 Troubleshooting 15
5 Removal from operation, storage, disposal 16
5.1 Removal from operation 16
5.2 Storage 16
5.3 Disposal 16
6 Dimensional drawings 17
7 Accessories 18
InPro 7100-VP Series © 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee 52 002 092 Printed in USA
InPro 7100-VP Series 5

1 Product description

1.1 Introduction

Thank you for buying this InPro® 7100-VP Series 4-electrode conductivity
sensor from METTLER TOLEDO. InPro 7100-VP Series 4-electrode conduc­tivity sensors are intended to be used to measure conductivity in medium to high conductivity water. The sensors are available in a wide selection of process connections to meet every application need.
Markings
The printed markings on each InPro 7100-VP Series 4-electrode conductivity
sensor contain the following information: Mettler Toledo Sensor manufacturer InPro 710X-VP Type of sensor Cell M: XXXXXX Cell constant Temp M: XXXXXX Temperature constant Serial No: XXXXXXXXX Serial number Ex-classification II I/2 G Ex ia IIB T3/T4/T5/T6 Ga/Gb
(plastic sensors ) Examination certificate number: SEV 14 ATEX 0129 X Notified body: Electrosuisse SEV, (No.: 1258)

1.2 Equipment and scope of delivery

Scope of delivery
The InPro 7100-VP Series sensors are delivered ready for use. Each sen-
sor is accompanied by this instruction manual, an individual «Certificate
of Quality», a Material Certificate following EN 10204 (3.1B) and a CE
Declaration of Conformity Certificate. Other Certificates may be included as
specified for the individual product description. Please check that the details
given in the Certificate of Quality match the sensor label. For each sensor, the
relevant cell constant has been de termined individually in solutions of known
conductivity at 18 – 25 °C (64 – 77 °F) during the manufacturing process,
and the value documented in the accompanying «Certificate of Quality».
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
6 InPro 7100-VP Series
4-electrode sensors are so-called diffuse-field sensors. The individually
determined cell constant was established under laboratory conditions without external influences.
Depending on the installation configuration, external influences (e.g. from
the pipe) can result in a shift in cell constant. If the application does not guarantee a minimum clearance of 30 mm (1.18") around the sensor, it is recommended to carry out recalibration of the sensor in installed position.
Packaging
The packaging consists of cardboard and plastic material. Keep the packaging for later use during storage or for transportation of
the sensor. Should you wish to dispose of the packaging material, please observe your local regulations as well as those data and instructions given in Chapter 5.3 of this manual.
Unpacking and inspection Please check the sensors immediately during unpacking in order to determi-
ne possible damage or missing items. Any irregularities should immediately be reportet to your carrier and to your supplier.
Each InPro 7100-VP Series sensor is equipped with an integral tempera ture
probe Pt 1000, IEC 751 Class A.
–1
The nominal cell constant is 0.25 cm
. The exact value is printed on the sensor label. For each sensor, the cell constant and temperature constant have been measured/established individually and the values have been documented in the accompanying Quality Certificate. All calibrations can be traced back to NIST and /or ASTM Standards.
Both the measuring range and the system measurement accuracy strongly
depend upon which type of transmitter is employed.
InPro 7100-VP Series © 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee 52 002 092 Printed in USA
InPro 7100-VP Series 7
••
••

1.3 Technical data

Measuring principle 4-electrode sensor 4-electrode sensor 4-electrode sensor 4-electrode sensor Electrode material SS 316LSS 316L SS 316L SS 316L
Body material CPVC PEEK PEEK PEEK RTD Built-in Pt1000 Built-inPt1000Built-in Pt1000 Built-in Pt1000 Sensor diameter See drawing See drawing See drawing See drawing Insertion length 28 mm 28 mm 40/65mm25mm
Max.sensor length 151mm 126.7 mm 123 /148mm 105 mm
Process connection –1"NPT –1"NPT –Cap nut G1
Cellconstant nominal 0.25cm Cellconstant accuracy ±1.0 %±1.0%±1.0%±1.0% System accuracy
Working conditions
Max. pressure 7 bar 17 bar 17 bar 17 bar at 25°C (77°F) (100 psi) (250 psi) (250 psi) (250 psi) Max. pressure –7bar 7 bar 7 bar at 95°C (203°F) Measuring –10…80°C –10…140°C temperature range (14…176°F)(14…284°Fc)(14…284°Fc)(14…284°Fc)
Temperature range N/A –10…140°C (sterilization) (14…284°Fc)(14…284°Fc)(14…284°Fc) Temperature accuracy ± 0.25°C±0.25 °C ± 0.25 °C ±0.25°C at 25°C (77°F) (± 0.5 °F)(± 0.5°F)(±0.5 °F)(±0.5 °F)
Design
Temperature Pt 1000 Pt1000 Pt1000Pt1000 compensation IEC class A IEC class A IEC class A IEC class A Cableconnection VarioPin (IP68) VarioPin (IP 68) VarioPin (IP68) Vario Pin (IP 68) Wetted parts: –MetalsSS 316LSS 316LSS316LSS 316L
–Plastics CPVC PEEK (FDA) PEEK (FDA) PEEK (FDA) – O-rings N/A N/A EPDM (FDA) N/A
Certificates
Cellconstant •• Materialcertificate EN10204 3.1• •• • Materialspecification 2.1• •• • ATEX ••
a The system accuracy depends
on the measuring range and the particular transmitter.
InPro 7108-VP/ InPro 7108-VP/ InPro 7108/25/*-VPInPro 7108-TC-VP CPVC PEEK InPro7108-VAR-VP
or HA-C22 or HA-C22
(1.10")(1.10") (1.57/2.56") (0.98")
(5.96")(4.99") (4.86 /5.84") (4.14")
1
–1"NPT Conduit – Tri-Clamp 2"
–1
a
±5.0 %or better±5.0%or better ± 5.0 %or better±5.0% or better
b
–(100 psi) (100 psi) (100 psi)
–1
0.25 cm
c
Sterilizable Sterilizable Sterilizable
c
–1
0.25 cm
–10…140°Cc–10…140°C
–10…140°Cc–10…140°C
or HA-C22 or HA-C22
b Accordingto PED
(Pressure Equipment Directive, 97/23/EC, Art. 3, Sec. 3).
c Shortterm 150°C (302°F).
4"–Tri-Clamp 1.5"
–Tuchenhagen-
Varivent DN 40–DN 125
–1
0.25 cm
c
c
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
8 InPro 7100-VP Series
Practical measuring range
Transmitter InPro 7108 InPro 7100 InPro 7100 i
M100 0.02 – 500 M300 0.02 – 650 0.02 – 400 – M300 ISM 0.02 – 500 M400 4-w / 2-w 0.02 – 650 0.02 – 400 0.02 – 500 M700 0.02 – 500 0.02 – 400 – M800 0.02 – 650 * 0.02 – 400 0.02 – 500 System accuracy
± 5 % ± 5 % ± 5 %
All values in mS / cm. * M800 1-channel only.
Subject to technical changes

2 Safety

2.1 Application compatibility

The wetted material parts of the sensor (several different materials come
into contact with the sample medium) can under some circumstance be incompatible with the particular composition of the process medium and/or of the operating conditions. Responsibility to verify application compatibil ity lies wholly with the user.
The compatibility of different types of material are outlined on http://www.coleparmer.com/techinfo/chemcomp.asp. Mettler-Toledo GmbH, Process Analytics accepts no responsibility whatso-
ever for the correctness or accuracy of such details.

2.2 Proper utilization

METTLER TOLEDO InPro 7100-VP Series sensors are intended solely for
the precise measurement of conductivity in aqueous solutions in industrial applications.
Any other use, or any operation over and above that intended by the
InPro 7100-VP Series © 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee 52 002 092 Printed in USA
InPro 7100-VP Series 9
manufacturer, are not recommended and can lead to harm or injury to material/equipment and persons. This is also relevant for applications which do not comply with the technical data labeled on the sensor. For any damage arising from such misuse, the user assumes full and sole responsibility.

2.3 Safety measures

The InPro 7100-VP Series sensors have been manufactured in line with
state-of-the-art technology and in accordance with accepted technical safety regulations. Nevertheless, the sensors can still represent a source
of risk and danger: – if the sensors are operated by insufficiently trained personnel, – if the sensors are employed incorrectly or not as intended by the manufacturer, – if the sensors are not regularly maintained or serviced.
Local legislation and regulations must be observed at all times. Such
stipulations do not form an integral part of this instruction manual. It is necessary to use protective gear, including gloves, for persons coming
in contact with the fluid. It is recommended to use gloves when handling
the electrodes of the sensor to limit contamination of the sensor. The user is responsible for the instruction and training of personnel. Addi-
tional copies of the instruction manual can be ordered from your supplier.
This instruction manual is an essential element of the sensor equipment
and must at all times be readily available to operators directly at the mea-
surement site. Before the sensor is removed from the process/process adapter, it must be
ensured that the process pressure has been reduced to a safe level and the
process temperature lowered to a safe range. Any escape of hot process
fluid under pressure can cause damage to material/equipment or injury to
persons. No modification whatsoever may be carried out on the sensors. Any
unauthorized modification or manipulation of the sensors results in imme-
diate expiration of the full scope of warranty granted by the manufacturer.
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
10 InPro 7100-VP Series

2.4 Use in Ex-zones

Note! For intended installation in an Ex-classified area, please observe the fol-
lowing guidelines:
Ex- classification: II 1/2 G Ex ia IIB T3/T4/T5/T6 Ga/Gb (plastic sensors)
II 1/2 G Ex ia IIC T3/T4/T5/T6 Ga/Gb (metallic sensors/metallic
sensors with plastic surface)
Marking and number of the test certificate: SEV 14 ATEX 0129 X
Introduction
1. The conductivity sensors made of plastic are according to Directive 94/9/EC (ATEX 95) Annex I equipment group II, category 1/2G after RL 99/92/EC (ATEX 137) in the zones 0/1 or 0/2 or 1/2 and the gas groups IIA and IIB, explosion hazards due to flammable substances in the field of temperature classes T1 to T6, may be used.
2. The conductivity sensors made of metal or metal coated with plastic to Directive 94/9/EC (ATEX 95) Annex I to equipment group II, cate­gory 1/2G after RL 99/92/EC (ATEX 137) in zones 0/1 or 0/2 or 1/2 and gas groups IIA, IIB and IIC, explosion hazards due to flammable
3. For use/installation, the requirements according to EN 60079-14
substances in the field of temperature classes T1 to T6, may be used.
must be observed.
InPro 7100-VP Series © 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee 52 002 092 Printed in USA
InPro 7100-VP Series 11
Electrical ratings
Conductivity measuring circuit and temperature sensing circuit in type of
protection intrinsic safety Ex ia IIC only for connection to a certified intrin­sically safe circuit.
Maximum values: Ui 16 V li 190 mA Pi 200 mW
The effective internal inductance and capacitance are negligible. The values above apply, each as the sum of all the individual circuits of
the associated intrinsically safe supply and evaluation unit.
Special conditions for safe use
1. The relationship between the maximum permissible ambient or media temperature and temperature class is shown in the following table:
For Ui ≤ 16 V, Ii ≤ 190 mA, Pi 200 mW.
Conductivity measuring circuit and temperature measuring circuits:
Temperature class Maximum ambiant or media temperature
T6 51 °C T5 63 °C T4 T3 143 °C
91 °C
2. For installation, the system drawings and wiring diagrams of the mea­surement systems should be observed.
3. The capacitance and inductance of the connecting cable has to be con­sidered.
4. The metal body of the conductivity sensors or the safety weld-in socket
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
12 InPro 7100-VP Series
or the fitting InFit 76*-*** or InTrac 7**-*** or other appropriate fitting must be electrically connected to the equipotential bonding system.
5. The metal body of the conductivity sensors or the safety weld-in socket or the fitting InFit 76*-*** or InTrac 7**-*** or other appropriate fitting is optionally included in the routine pressure test of the system.

3 Initial start-up

3.1 Installation

The sensors are to be mounted in such a way that the sample medium
flows directly into the sensor through the orifice at the tip and re-emerges through the vent holes. Any other installation position of the sensors can create the risk of formation of airlocks, or of contamination through depos­its of solid matter.
Flow direction
Flow direction
T-piece
> R 30
Sensor
Flow of medium must be accross the face of the sensor. Air bubbles and
solid matter deposits are to be avoided. A minimum clearance of 30 mm (1.18") between sensor and pipe wall must be maintained (see drawing above).
Vertical mounting (top entry) in a pipe is recommended only if the pipe is
full and no air bubbles are able to develop in the flow.
In the event of side mounting (side entry) of the sensor, vertical upward
flow of sample process medium must be ensured.
InPro 7100-VP Series © 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee 52 002 092 Printed in USA
InPro 7100-VP Series 13

3.2 Electrical connections

All of the InPro 7100-VP Series sensors can be connected to the asso ciated
transmitter using the appropriate multiwire cable from METTLER TOLEDO.
Colors and function of the cable strands (supplied separately)
4-electrode sensor Color
Cnd outer 1 white/blue
Cnd inner 1 white
Cnd inner 2 blue
Cnd outer 2 black
RTD green
RTD sense red
RTD ret / GND Bare shield
Connection to the relative transmitter must be followed according to the
wiring diagram (see respective transmitter instruction manual for detail).
METTLER TOLEDO supplies the cables. Please refer to chaper 7 “Accesso-
ries” for ordering information
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
14 InPro 7100-VP Series

4. Maintenance and troubleshooting

4.1 Conditions of warranty

METTLER TOLEDO guarantees the quality of materials and workmanship
within a narrow range of manufacturing tolerances, so that the product purchased is free from any substantial deviations from material and manufacturing quality standards. The warranty is valid for a period of one year from date of delivery (ex-works). If within this warranty period, any repair or replacement should become necessary, and such cause is not due to misuse or incorrect application, please return the sensor, freight pre-paid, to your appropriate METTLER TOLEDO agency. Repair work will be carried out free of charge. Final decision on whether the defect is due to a man­ufacturing error or to incorrect operation of the sensor by the customer is made at the discretion of the Customer Service department of METTLER TOLEDO. After expiration of the warranty period, faulty sensors will be repaired or replaced on an exchange basis against payment of the costs involved.

4.2 Maintenance

Dirty or contaminated sensors can deliver incorrect measurement values.
If fouling is presumed, the sensor is to be removed from operation and the electrodes as well as the insulation between the electrodes cleaned with a soft cloth. Do not clean electrodes with an abrasive paper or pad. Suitable cleaning solutions are mild detergents or strongly diluted acids (< 0.5 % by wt.) such as nitric acid. Never use clean ing agents that
are not compatible with the material to be cleaned. When handling acids, precautionary measures are to be taken. The sensor must be thoroughly flushed with distilled or distilled water
prior to re-installation. Following cleaning and re-installation into the process, it can take from
several minutes up to several hours until the sensor delivers the originally
measured value.
InPro 7100-VP Series © 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee 52 002 092 Printed in USA
InPro 7100-VP Series 15

4.3 Troubleshooting

Error Possible cause Corrective action
No signal on display Electrical connections either missing
or incorrectly in place
Check all connections and associated cabling
Sensor is not in contact with the sample medium
No temperature signal Temperature probe is not connected
Transmitter does not support Pt1000 RTD
Incorrect or unstable measurement reading
> R 30
Sensor is contaminated/fouled
Sensor is installed too close to pipe wall
Check installation for air pockets
Check all connections and associated cabling
Replacement of transmitter or sensor necessary
Clean the sensor (electrodes)
Ensure that a minimum distance of 30 mm (1.18") between sensor tip and pipe wall is maintained (see drawing)
Ensure that a minimum distance of 30 mm (1.18") between
sensor tip and pipe wall is maintained.
© 02 / 2017 Mettler-Toledo GmbH, CH - 8606 Greifensee InPro 7100-VP Series Printed in USA 52 002 092
16 InPro 7100-VP Series

5 Removal from operation, storage, disposal

5.1 Removal from operation

The sensor is only conditionally subject to aging. When used as intended
and appropriately maintained and serviced, the lifetime of the sensor can
extend to several years. Before removing the sensor from the process/process adapter, it must be
ensured that the process pressure has been reduced to a safe level and
the process temperature has lowered to a safe range. Any escape of hot
process medium under pressure can cause damage to material/equipment
or injury to persons. After removal from the process the sensor should first be flushed with
distilled water. If the sensor incurs a defect, it can not be repaired and must be disposed
under observance of prevailing local regulations.

5.2 Storage

If the sensor is not use, it can be stored dry. However, it has to be recon-
ditioned in the process fluid accordingly prior to renewed operat ion. This
procedure may take several hours to complete.

5.3 Disposal

The user is responsible for the proper disposal of the conductivity sensor. The
sensor contains electronic components that must be disposed of correctly
in order to avoid hazards to people or to the environment.