Essential Services for Dependable Performance of Your IND231/IND236 Weighing Terminal
Register your product
Contact METTLER TOLEDO for service
Installation, Configuration, Integration and Training
Initial Calibration Documentation
Periodic Calibration Maintenance
IND231/IND236
Weighing Terminal
Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper use of your
new equipment according to this Manual and regular calibration and maintenance by our factorytrained service team ensures dependable and accurate operation, protecting your investment.
Contact us about a service agreement tailored to your needs and budget. Further information is
available at www.mt.com/service
There are several important ways to ensure you maximize the performance of your investment:
1.
www.mt.com/productregistration so we can contact you about enhancements, updates and
important notifications concerning your product.
.
: We invite you to register your product at
2.
accuracy – an out of specification scale can diminish quality, reduce profits and increase
liability. Timely service from METTLER TOLEDO will ensure accuracy and optimize uptime and
equipment life.
a.
trained, weighing equipment experts. We make certain that your weighing equipment is
ready for production in a cost effective and timely fashion and that personnel are trained for
success.
b.
requirements are unique for every industrial scale so performance must be tested and
certified. Our calibration services and certificates document accuracy to ensure production
quality and provide a quality system record of performance.
c.
confidence in your weighing process and documentation of compliance with requirements.
We offer a variety of service plans that are scheduled to meet your needs and designed to
fit your budget.
: The value of a measurement is proportional to its
: Our service representatives are factory-
: The installation environment and application
: A Calibration Service Agreement provides on-going
English
interference, and (2) this device must accept any interference received, including
means, electronic or mechanical, including photocopying and recording, for any
purpose without the express written permission of METTLER TOLEDO.
U.S. Government Restricted Rights: This documentation is furnished with
Restricted Rights.
Copyright 2013 METTLER TOLEDO. This documentation contains proprietary
information of METTLER TOLEDO. It may not be copied in whole or in part
without the express written consent of METTLER TOLEDO.
METTLER TOLEDO reserves the right to make refinements or changes to the
product or manual without notice.
COPYRIGHT
METTLER TOLEDO® is a registered trademark of Mettler-Toledo, LLC. All other
brand or product names are trademarks or registered trademarks of their
respective companies.
METTLER TOLEDO RESERVES THE RIGHT TO MAKE REFINEMENTS OR
CHANGES WITHOUT NOTICE.
FCC Notice
This device complies with Part 15 of the FCC Rules and the Radio Interference
Requirements of the Canadian Department of Communications. Operation is
subject to the following conditions: (1) this device may not cause harmful
interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to Part 15 of FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment
is operated in a commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used in accordance
with the instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely to
cause harmful interference in which case the user will be required to correct the
interference at his or her expense.
Declaration of Conformity is located on the documentation CD.
RoHS Compliance Statement.
The majority of our products fall within categories 8 and 9. Those categories
currently do not fall within the scope of the Directive 2002/95/EG (RoHS) of
January 27, 2003. If our products are intended for use in other products
which themselves fall within the scope of the RoHS Directive, compliance
requirements have to be separately negotiated contractually.
Those products which fall within categories 1-7 and 10 will be in compliance
with the EU RoHS Directive from no later than July 1, 2006.
If it is not possible for technical reasons to replace any non-RoHS-compliant
substances in any of the above products as required, we plan to inform our
customers in a timely manner
Statement regarding harmful substances
We do not make direct use of harmful materials such as asbestos, radioactive
substances or arsenic compounds. However, we purchase components from
third party suppliers, which may contain some of these substances in very small
quantities.
English
Warnings and Cautions
• READ this manual BEFORE operating or servicing this equipment and FOLLOW these
instructions carefully.
• SAVE this manual for future reference.
WARNING
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT THE AC VERSION OF THE
IND231/IND236 TERMINAL TO PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE
GROUND PRONG.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL. EXERCISE CARE WHEN
MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON.
FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR
PROPERTY DAMAGE.
WARNING
THE IND231/IND236 IS NOT DESIGNED FOR USE IN AREAS CLASSIFIED AS HAZARDOUS
BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES. DO NOT INSTALL AN
IND231/IND236 INTO AN EXPLOSIVE ENVIRONMENT.
WARNING
WHEN THIS EQUIPMENT IS INCLUDED AS A COMPONENT PART OF A SYSTEM, THE
RESULTING DESIGN MUST BE REVIEWED BY QUALIFIED PERSONNEL WHO ARE FAMILIAR
WITH THE CONSTRUCTION AND OPERATION OF ALL COMPONENTS IN THE SYSTEM AND THE
POTENTIAL HAZARDS INVOLVED. FAILURE TO OBSERVE THIS PRECAUTION COULD RESULT IN
BODILY HARM AND/OR PROPERTY DAMAGE.
CAUTION
THE NiMH BATTERY USED IN THIS DEVICE MAY PRESENT A RISK OF FIRE OR CHEMICAL
BURN IF MISTREATED. DO NOT CRUSH, DISASSEMBLE, HEAT ABOVE 60°C OR INCINERATE.
REPLACE BATTERY WITH 30044650 ONLY. USE OF ANOTHER BATTERY MAY PRESENT A RISK
OF BURN, FIRE OR EXPLOSION.
CAUTION
USE CAUTION WHEN TESTING THE BATTERY. A LARGE AMOUNT OF CURRENT MAY BE
PRESENT IN THE BATTERY.
CAUTION
DISPOSE OF USED BATTERY PROMPTLY. KEEP AWAY FROM CHILDREN. DO NOT
DISASSEMBLE AND DO NOT DISPOSE OF IN FIRE.
Warnings and Cautions
CAUTION
BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR
INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER
AND WAIT AT LEAST THIRTY (30) SECONDS BEFORE ANY CONNECTIONS OR
DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE PRECAUTIONS COULD RESULT IN
DAMAGE TO OR DESTRUCTION OF THE EQUIPMENT AND/OR BODILY HARM.
NOTICE
DO NOT ATTEMPT TO CHARGE THE Ni-MH BATTERY IF THE BATTERY TEMPERATURE IS BELOW 0°C (32°F).
CHARGING IS NOT POSSIBLE AT OR BELOW THIS TEMPERATURE. DO NOT OPERATE THE BATTERY CHARGER
OUTSIDE ITS TEMPERATURE RANGE OF 0°C (32°F) TO 40°C (104°F).
NOTICE
TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE IND231/IND236 TERMINAL AND
WAIT AT LEAST 30 SECONDS BEFORE CONNECTING OR DISCONNECTING ANY HARNESS.
NOTICE
OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE DEVICES.
Disposal of Electrical and Electronic Equipment
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies
to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from
which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
Thank you for your contribution to environmental protection.
The IND231/IND236 industrial scale terminal provides a compact yet flexible
• Overview of IND231/IND236
• Use in Hazardous Areas
• Specifications
• Model Identification
• Inspection and Contents
Checklist
• Physical Dimensions
• Main PCB
• Scale Bases
• Options
• Display and Keyboard
solution for a variety of weighing needs. Available as either AC powered for
stationary applications or battery-powered for portable applications, this terminal is
at home in virtually any industrial environment.
Both 2mv/V and 3mv/V load cells are supported without the need for any
configuration change. The IND231/IND236 delivers precision measurement data
from grams to kilograms in a single, cost effective package.
Standard applications include basic weighing, animal weighing, Over/Under
checkweighing, counting and accumulation. Pre-defined application-specific print
templates including date and time stamp, provide transmission of weighing
transaction information to printers and PC software applications.
1.1. IND231/IND236 Overview
1.1.1. Standard Features
• Easy to handle plastic enclosure design for the IND231, rugged stainless steel enclosure for
the IND236
• Supports one analog load cell platform with up to four 350Ω load cells
• Can be used as a remote display to show weight from a separate main terminal
• Large seven segment LCD display with white backlight and graphic icons for clear readability
in varying light conditions
• Powered by either 85–264 V AC or internal battery pack
• One standard serial port (COM1) for asynchronous, bidirectional communication
• Support for one of the following option boards:
COM2 electrically isolated RS232, RS422 and RS485 serial Interface
USB device Interface for connection to a PC
Discrete I/O interface
• Front panel key access to basic weighing functions – zero, tare, clear, configurable function
and print
• Selectable primary unit of measure including grams, kilograms, pounds and ounces
• Selectable second unit of measure including grams, kilograms, pounds and ounces
30094803 | 00 | 08/2013
METTLER TOLEDO IND231/IND236 Installation Manual
1-1
Introduction
Enclosure Type
Product Dimensions
(w × h × d)
Shipping Dimensions
(w × h × d)
Product Weight
Shipping Weight
Environmental Protection
Operating Environment
• Backup and restore of configuration and calibration settings, using InSite
• Automatic shutoff and backlight timeout features to help conserve energy and extend battery
life when using the battery powered version
1.1.2. IND231/IND236 Terminal Types
The terminal is available in the following four versions:
• IND231plastic housing, AC power (also designed for use with AA size alkaline cells)
• IND231plastic housing, Rechargeable Ni-MH Battery power
• IND236 stainless steel housing, AC power
• IND236 stainless steel, Rechargeable Ni-MH battery power
1.2. Use in Hazardous Areas
DO NOT USE THE IND231/IND236 TERMINAL IN AREAS CLASSIFIED AS HAZARDOUS
BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES. CONTACT AN AUTHORIZED
METTLER TOLEDO REPRESENTATIVE FOR INFORMATION ABOUT HAZARDOUS AREA
APPLICATIONS.
WARNING
®
SL tool
1.3. Specifications
The IND231 and IND236 terminal conforms to the specifications listed in Table 1-1.
Table 1-1: Terminal Specifications
Specifications IND231 IND236
Plastic, configurable as desk top or
column/wall mount enclosure
220 mm x 150 mm x 102 mm
(8.66 in. x 5.90 in. x 4.02 in.)
350 mm x 220 mm x 195 mm
(13.78 in. x 8.66 in. x 7.68 in.)
AC Version: 1.2 kg (2.6 lb)
Battery Version: 1.5 kg (3.3lb)
2.0 kg 3.0 kg
IP54 IP66/67
Operating temperature range: -10 °C to +40 °C
Storage temperature range: -20 °C to +60 °C
Relative Humidity: 10-95%, non-condensing.
Stainless steel, configurable as desk
top or column/wall mount enclosure
220 mm x 150 mm x 93.3 mm
(8.66 in. x 5.90 in. x 3.67 in.)
AC Version: 2.2 kg (4.9 lb)
Battery Version: 2.5 kg (5.5 lb)
1-2
METTLER TOLEDO IND231/IND236 Installation Manual
30094803 | 00 | 08/2013
English
Hazardous Areas
Power
Display
Weight Display
Scale Types
Number of Cells
Number of Scales
Analog/Digital Update Rate
Load Cell Excitation Voltage
Minimum Sensitivity
Keypad
Communication options
Application
Specifications IND231 IND236
The terminal cannot be operated in areas classified as Hazardous because of
combustible or explosive atmospheres in those areas. Contact an authorized
METTLER TOLEDO representative for information about hazardous area
applications.
AC version: Operates at 85–264 VAC, 49–61 Hz and includes a power cord
configured for the country of use.
Battery version :
Six "AA" size alkaline batteries or
NiMH rechargeable battery pack
(about 120hrs for one load cell)
40 mm height character, seven digits seven segment LCD with white backlight
including weight display, weight units, gross/net indication and symbols for
motion, center of zero and variable applications
Standard: One serial port (COM1) RS-232, 1,200 to 115,200 baud
Optional isolated serial port: (COM2) RS-232/422/485, 1,200 to
115,200 baud
Discrete I/O Interface
Optional Discrete I/O port: 2 inputs/4 outputs
USB Interface
Optional USB device port
Protocol
Serial Inputs: ASCII commands for CTPZ (Clear, Tare, Print, Zero), SICS
(most level 0 and level 1 commands)
Battery version :
NiMH rechargeable battery (about
120hrs for one load cell)
One
5 VDC
0.5 µV/e
30094803 | 00 | 08/2013
Zero, Tare, Print, X10, Date & Time, Basic weighing, Animal weighing,
Counting with APW enhancement, Over/Under checkweighing with 10 records
database, Accumulation, Remote display
METTLER TOLEDO IND231/IND236 Installation Manual
1-3
Introduction
Approvals
Accessories
Specifications IND231 IND236
Weights and Measures,
USA:
NTEP Class III/IIIL - 10,000d; Cert. #13-049,
Canada:
Class III - 10,000d; Class IIIHD - 20,000d,
Europe:
Class III, 2 x 3000e and 6000e; TC8351, T5976
OIML:
Class III, 2 x 3000e and 6000e; R76/2006-NL1-13.06
Product Safety
UL, cUL, CE
Wall mount/column mount bracket; desk bracket.
1.4. Model Identification
The IND231/IND236 model number, factory number and serial number are located on the data
plate of the terminal. Refer to Figure 1-1 to verify the configuration of the IND231/IND236 terminal
when it left the METTLER TOLEDO factory.
1-4
METTLER TOLEDO IND231/IND236 Installation Manual
Figure 1-1: IND231 Configuration Chart
30094803 | 00 | 08/2013
English
1.5. Inspection and Contents Checklist
Verify the contents and inspect the package immediately upon delivery. If the shipping container is
damaged, check for internal damage and file a freight claim with the carrier if necessary. If the
container is not damaged, remove the terminal from its protective package, noting how it was
packed, and inspect each component for damage.
If shipping the terminal is required, it is best to use the original shipping container. The terminal
must be packed correctly to ensure its safe transportation.
The package should include:
• IND231 or IND236 Terminal• User manual (or resource CD)
The physical dimensions of the IND231/IND236 enclosure are shown in Figure 1-2 and Figure 1-3
in mm and [inches].
30094803 | 00 | 08/2013
Figure 1-2: IND231 Enclosure Dimensions
METTLER TOLEDO IND231/IND236 Installation Manual
1-5
Introduction
Figure 1-3: IND236 Enclosure Dimensions
1.7. Main PCB
The IND231/IND236 terminal’s main printed circuit board (PCB) provides the analog load cell
scale interface, as well as the COM1 RS-232 serial port. The COM1 RS-232 serial port supports bidirectional communications at speeds up to 115200 bps. This port can be used for saving terminal
configuration data to a PC via the InSite
The main board also contains the DC power input connection, display interface, keypad interface,
and interface port for the options.
An AC/DC power board supplies the terminal with DC +12V/1A output.
An optional charging board is included with the rechargeable battery powered version.
®
SL tool.
1-6
METTLER TOLEDO IND231/IND236 Installation Manual
30094803 | 00 | 08/2013
English
1.8. Scale Bases
The IND231/IND236 terminal supports analog scale bases and provides 5 volts of excitation to
drive analog load cells. Up to four 350Ω load cells can be powered by the terminal.
A four- or six-wire load cell connection is provided, with sense lines to help maintain accuracy as
the load cell cable resistance changes due to variations in temperature.
1.9. Options
Three kinds of option interface can be mounted on COM2. The connection pin specifications are
printed on the PCB.
The following options are available for the IND231 and IND236.
1.9.1. Isolated Serial interface RS232/RS422/RS485
This optional port provides RS-232 and RS-422/485 communication which can be configured in
setup. The port is bidirectional and can be used for various functions such as print, auto print,
Toledo continuous output or SICS communications.
The COM2 port is galvanically isolated for both RS-232 and RS-485, to provide surge voltage
protection.
The RS-485 connection can be used as an RS-422 transmit only, when sending continuous output
to a scoreboard or remote display.
The interface RS422/485 is required for data transmission with higher baud rate over longer
distances up to 500 m.
Communication settings can be configured in the terminal setup menu.
1.9.2. Discrete I/O
The discrete I/O interface option provides four dry-contact relay outputs. The relay contacts will
switch up to 30 volts DC or 250 volts AC at 1A.
The two inputs are switch selectable as either active (for simple pushbutton control) or passive (for
connection to devices that supply their own power for the inputs).
1.9.3. USB
The USB port provided is a UART-USB device bridge acting as a virtual COM port, and is used for
communication of serial data to devices such as a PC. The port is bidirectional and can be
configured for various functions such as print, auto print, Toledo continuous output or SICS
communications. USB external keyboards and bar code scanners are not supported.
30094803 | 00 | 08/2013
METTLER TOLEDO IND231/IND236 Installation Manual
1-7
Introduction
User’s Guide
Configuration.
Configurable
1.10. Display and Keyboard
IND231/IND236 uses a transflective type segment LCD display with a white backlight. The main
character height is 40 mm. The front panel, including display and keypad, is shown in Figure 1-4.
The only keypad difference between IND231 and IND236 is the name on the upper right corner of
the terminal.
Function keys
Print Key
ON/OFF key
Figure 1-4: IND231 Front Panel Layout
1.10.1. Display Layout
At the top of display, a single system line displays terminal status such as range number, center of
zero, checkweighing status, counting, accumulation and auto accumulation.
Below the system line is the weight display block. During normal, basic weighing operation, the
terminal display shows the gross or net weight in the larger 40 mm (1.57 in.) size. When one of
the applications is running, prompting messages or menu index characters are also shown in 40
mm (1.57 in.) height. At the right of weight display is a vertical column displaying the gross and
net symbols, average piece symbol, and weight units. The star symbol, minus symbol and
dynamic symbol appear in a column to the left of the weight display block.
At the bottom of the normal basic weighing display is a block used for assigning functions to
and
. For display and keypad operation during setup, refer to Chapter 3 in the
,
1.10.2. Front Panel Keys
The IND231/IND236 terminal operator interface provides a total of seven membrane keys. The print
key and five scale function keys (three with fixed function, two with configurable functions) are
positioned under the display.
Arrows on the first four keys indicate their use in menu navigation.
1-8
METTLER TOLEDO IND231/IND236 Installation Manual
30094803 | 00 | 08/2013
English
User’s Guide
Operation
The ON/OFF key is located at the bottom left of the display. These keys are used to enter the setup
menu, to navigate and select setup elements as described in Chapter 2 of the
.
,
30094803 | 00 | 08/2013
METTLER TOLEDO IND231/IND236 Installation Manual
1-9
Introduction
1-10
METTLER TOLEDO IND231/IND236 Installation Manual
30094803 | 00 | 08/2013
English
2 Installation
This chapter covers
• Opening the Enclosure
• Environmental Protection
• Mounting the Terminal
• Installing Cables and
Connections
• Capacity Label Instructions
• Closing the Enclosure
• Sealing the Enclosure
This chapter provides installation instructions for the IND231/IND236 terminal.
Please read these procedures thoroughly before beginning installation.
This section includes details about opening and mounting the enclosure, and
explains how to install cables and PCBs. Instructions for labeling, closing and
sealing the enclosure are also provided.
2.1. Opening the Enclosure
BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR INTERCONNECTING
WIRING BETWEEN ELECTRONIC EQUIPMENT ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY (30)
SECONDS BEFORE ANY CONNECTIONS OR DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE
PRECAUTIONS COULD RESULT IN DAMAGE TO OR DESTRUCTION OF THE EQUIPMENT AND/OR BODILY HARM.
CAUTION
The front panel of the IND231/IND236 terminal is locked in place by four screws which attach it to
the rear housing of the enclosure. To access the terminal’s PCB in order to install options, connect
internal wiring and set switches, separate the front panel from the enclosure as follows:
1. For IND231, use a T-20 torx screwdriver to loosen three of the screws. Use a slotted
screwdriver to loosen the sealing screw, indicated in Figure 2-1.
2. For IND236, use a 10mm socket wrench to loosen three of the screws. Use a 5mm Allen Head
insert screwdriver to loosen the sealing screw, indicated in Figure 2-2. Note that the screws are
captive, and remain attached to the rear cover of the IND236.
30094803 | 00 | 08/2013
Figure 2-1: IND231 Screws (left) and IND236 Sealing Screw (right)
METTLER TOLEDO IND231/IND236 Installation Manual
2-1
Installation
Sealing Screw
Figure 2-2: Opening the Enclosures – IND231 (left) and IND236 (right)
3. Separate the front housing from the rear housing, as shown in Figure 2-3.
Figure 2-3: Removing the Cover
2.2. Environmental Protection
The IND231/IND236 terminal is designed for standard industrial use. IND231 has been tested and
found to meet IP54 standards. The IND236 meets the requirements of IP66/IP67.
2-2
METTLER TOLEDO IND231/IND236 Installation Manual
WARNING
THE IND231/IND236 IS NOT DESIGNED FOR USE IN AREAS CLASSIFIED AS HAZARDOUS
BECAUSE OF COMBUSTIBLE OR EXPLOSIVE ATMOSPHERES. DO NOT INSTALL AN
IND231/IND236 INTO AN EXPLOSIVE ENVIRONMENT.
30094803 | 00 | 08/2013
English
2.3. Mounting the Terminal
The terminal can be placed on a desktop, attached to a vertical surface or fastened to a column.
Mount the terminal where viewing is optimal and the terminal keypad is easily accessible.
2.3.1. IND231Mounting
2.3.1.1. Desktop Mounting
The IND231 has two self-adhesive rubber feet (Figure 2-4) attached to the bottom of the housing to
prevent sliding.
Figure 2-4: IND231 – Rubber Feet
2.3.1.2. Wall Mounting& Column Mounting
One mounting bracket and two tightening knobs are included with the IND231. These can be used
to mount the terminal either to a column or to a vertical surface.
1. Use two bracket hand knobs to attach the brackets to the bottom of the terminal, as shown in
Figure 2-5.
Figure 2-5: Bracket and Tightening Knobs
2. The bracket allows a wide range of adjustment, as shown in Figure 2-6.
30094803 | 00 | 08/2013
METTLER TOLEDO IND231/IND236 Installation Manual
2-3
Installation
Figure 2-6: Adjusting the Bracket for Mounting on Wall or Column
3. Mark the position of the bracket mounting holes on the surface to which the terminal will be
mounted using the dimensions shown in Figure 2-7, or by holding the terminal up to the
surface and marking through the bracket holes.
Figure 2-7: Hole Pattern Template for Mounting Brackets
4. The hardware to mount the terminal bracket is not included with the terminal—it must be
supplied locally. Ensure that the mounting hardware is capable of supporting the weight of the
terminal, which is approximately 2.0 kg (4.4 lb). Using the locally supplied hardware, secure
the terminal bracket to the surface.
2.3.2. IND236 Mounting
2.3.2.1. Desktop Mounting
When the IND236 terminal will be mounted on a flat surface, an optional desktop mounting bracket
(PN: 22021070) is available for purchase.
2-4
METTLER TOLEDO IND231/IND236 Installation Manual
Figure 2-8: IND236 with desktop mounting Brackets
30094803 | 00 | 08/2013
English
Figure 2-9: IND236 desktop Bracket Dimensions
2.3.2.2. Wall Mounting & Column Mounting
One mounting bracket is included with the IND236. This can be used to mount the terminal either to
a column or to a vertical surface. The bracket allows a wide range of adjustment, as shown in
Figure 2-10.
Figure 2-10: IND236 Mounted to Column (left) and Wall (right)
Figure 2-11: IND236 Wall and Column Bracket Dimensions
2.4. Installing Cables and Connectors
Information for installing cables and connectors for the IND231/IND236 terminal is provided in this
section, including the following:
• Connectors and Cable Glands
• Enclosure Opening Assignments
30094803 | 00 | 08/2013
METTLER TOLEDO IND231/IND236 Installation Manual
2-5
Installation
• PCB Identification
• Main Board Wiring Connections
• AC Power Connection
• Charging Board and Rechargeable Battery Pack Connections
• Analog Load Cell connection
• COM1 Serial Port Connection
• Wiring Connections for Options
2.4.1. Connectors and Cable Glands
The IND231 uses standard (IEC320 C14) power socket and DSUB-9 connectors for RS-232
communication. Cable glands are used for load cell and option connections.
The IND236 terminal is designed to withstand severe wet environments. However, care must be
taken when installing cables and/or connectors that enter the terminal enclosure. To ensure a
watertight seal:
• Before connecting wires, pass the cables through an appropriately sized cable gland. Figure
2-12 shows a cable gland with its clamping nut removed.
Figure 2-12: Cable Gland
• Depending upon the diameter of the cable to be installed, select (Table 2-1) one of the rubber
grommets (if required) to properly seal around the cable.
Table 2-1: Gland, Grommet and Cable Sizes
IND231
Grommet
None 4–8 mm 5–10 mm
Small grommet 3–6.5 mm
When making cable terminations inside the enclosure, ensure that the cable length from the
Load Cell Cable Gland
Cable Diameter
IND236
and Option Cable Gland
4.5–6.8 mm
terminal strip/connector to the terminal housing is sufficient so that no strain is placed on the
connector assembly when the housing is in the fully open position.
2-6
METTLER TOLEDO IND231/IND236 Installation Manual
30094803 | 00 | 08/2013
English
1
2
3
4
After making the wiring connections described in the next section, ensure that the nut on the
cable gland is tightened properly to seal around the cable. The seal must be watertight.
2.4.1.1. Enclosure Opening Assignments
The Figure 2-13 shows the openings in the IND231 enclosure.
1 2 3 4
Figure 2-14 shows the openings in the IND236 enclosure and Table 2-2 indicates the assignment
of each opening.
Figure 2-13: IND231 Enclosure Openings
1 4
2 3
Figure 2-14: IND236 Enclosure Openings
Table 2-2: Connectors and Cable Gland Assignments
Assignment
Position
Standard connection Optional connection
AC Power Cord -
None (option port) Isolated RS-232/422/485 USB Discrete I/O
COM1 (RS-232) -
Load Cell Cable -
30094803 | 00 | 08/2013
METTLER TOLEDO IND231/IND236 Installation Manual
2-7
Installation
This is the only method
to change scale parameters in an approved terminal.
Entering Setup Mode
Scale |
Type Approval
User’s Guide, Configuration
Option board
2.4.2. PCB Identification
The IND231/IND236 terminal circuit boards all mount on the PCB support assembly. The AC
connection board is only needed for the IND236. The charging board is only used for terminals
powered by a rechargeable battery pack. The option boards all mount in the same location. Figure
The Figure 2-16 shows the assignment of connectors on the main board. Be sure to attach cables
to the correct socket.
AA Cell Enclosure
DC power
Charging board socket
COM1 RS-232
Metrology
switch
Keypad
Load Cell
The metrology switch is a momentary contact switch. When it is pressed, the terminal enters setup
at the F1 Scale settings menu, where scale parameters can be configured.
2-8
METTLER TOLEDO IND231/IND236 Installation Manual
sections in Chapter 3 of the
Display
Figure 2-16: Main Board Connections
Refer to the
Backlight
and
, for more details.
30094803 | 00 | 08/2013
English
2.4.4. AC Power Connection
Figure 2-17 shows an AC Power board without its protective cover.
AC Lines input
DC lines output
Figure 2-17: Power Board
An external power cord with a standard power socket (IEC320 C14) is supplied for the IND231.
A permanently attached line cord supplies AC power to the AC version of the IND236 terminal. The
ground wire has a loop terminal for connection to the ground connection inside the terminal (Figure
2-18).
Power cord ground
wire connected to
grounding point
Connection board
Power board
(protective cover
removed)
Figure 2-18: IND236 AC Power Connection Board and Grounding
In the IND236, an interface PCB transfers power from the line cord to the power board (Figure
2-18).
The terminal requires 85 to 264 VAC (at 176 mA maximum) with a line frequency of 49 to 61 Hz
of power. It is internally fused at 3.15 amps, 250 volts.
The integrity of the power ground for equipment is important for both safety and dependable
operation of the terminal and its associated scale base. A poor ground can result in an unsafe
condition should an electrical short develop in the equipment. A good ground connection minimizes
extraneous electrical noise pulses. The IND231/IND236 should not share power lines with noisegenerating equipment. To confirm ground integrity, use a commercial branch circuit analyzer. If
adverse power conditions exist, a dedicated power circuit or power line conditioner might be
required.
30094803 | 00 | 08/2013
WARNING
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT TO PROPERLY GROUNDED
OUTLET ONLY. DO NOT REMOVE THE GROUND PRONG.
METTLER TOLEDO IND231/IND236 Installation Manual
2-9
Installation
2.4.5. Analog Load Cell Connection
NOTICE
TO AVOID DAMAGE TO THE PCB OR LOAD CELL, REMOVE POWER FROM THE IND231/IND236 TERMINAL AND WAIT
AT LEAST 30 SECONDS BEFORE CONNECTING OR DISCONNECTING ANY HARNESS.
Load cell connections are made to the load cell connector on the main board, indicated in Figure
2-16.
The IND231/236 terminal is designed to power up to four 350-ohm load cells (or a minimum
resistance of approximately 87 ohms). To confirm that the load cell load for this installation is
within limits, the total scale resistance (TSR) must be calculated, as follows:
TSR =
Load Cell Input Resistance (Ohms)
Number of Load Cells
Before connecting the load cells, ensure that the TSR of the load cell network to be connected to the
IND231/IND236 has a resistance greater than the minimums listed above. If the resistance is
below the minimum, the IND231/IND236 will not operate properly.
In addition, the maximum cable distance must be reviewed. Table 2-3 provides recommended
maximum cable lengths based on TSR and cable gauge.
Table 2-3: Recommended Maximum Cable Lengths
TSR (Ohms)
350 243/800 610/2000 1219/4000
87 (4-350 Ω cells)
24 Gauge
(meters/feet)
60/200 182/600 304/1000
20 Gauge
(meters/feet)
16 Gauge
(meters/feet)
The IND231/IND236 terminal is designed to support both 2mV/V and 3mV/V load cells from the
same circuitry. A load cell output rating selection jumper is not required.
Table 2-4 and Figure 2-19 show the terminal definitions for the analog load cell terminal strip. Note
that when using four-wire load cells, jumpers must be placed between the +Excitation and +Sense
terminals and between the −Excitation and −Sense terminals.
2-10
METTLER TOLEDO IND231/IND236 Installation Manual
Table 2-4: Load Cell Termination Pin Assignments
Pin Signal
1 +EXC
2 +SEN
3 +SIG
4 Shield
5 -SIG
6 -SEN
7 -EXC
30094803 | 00 | 08/2013