1.2What is the Halogen Moisture Analyzer used for? ...................................................................................... 6
1.3Safety has priority .................................................................................................................................. 7
1.5Important information for these instructions ............................................................................................10
2Your first measurement in next to no time ...........................................................................................11
2.1Unpacking and checking the standard equipment ................................................................................... 11
2.2Selecting the location ...........................................................................................................................12
2.3Setting up, leveling and connecting to power supply ................................................................................ 12
2.4Preparing the printer ............................................................................................................................15
2.5Setting the date and time ......................................................................................................................16
2.6Your first measurement ........................................................................................................................18
3How to obtain the best results .............................................................................................................22
3.1Measurement principle of the Halogen Moisture Analyzer .........................................................................22
3.2Notes on adjustment of the balance and the heating module ....................................................................23
4.12AutoMet test measurement ....................................................................................................................47
5Methods facilitate your work ............................................................................................................... 51
5.1What is a method? ..............................................................................................................................51
5.2Recalling a method ............................................................................................................................. 52
5.3Naming, compiling or changing methods .............................................................................................. 52
5.4Method menu ..................................................................................................................................... 55
5.4.5.1Standby temperature ............................................................................................................................ 60
5.4.5.2Recording the settings ..........................................................................................................................62
5.4.6Drying temperature to 200 °C without time limitation ...............................................................................63
5.5Printing out methods ........................................................................................................................... 63
5.6Measured value journal ........................................................................................................................ 65
5.7Measured value statistics .....................................................................................................................66
Contents
6The menu (basic settings of the instrument) ........................................................................................69
6.2Adjusting the balance .......................................................................................................................... 73
6.3Testing the balance ............................................................................................................................. 74
6.4Adjusting the heating module ................................................................................................................ 75
6.5Testing the heating module ................................................................................................................... 77
6.6Resetting to the factory settings ............................................................................................................. 80
6.7Clearing the method memory ................................................................................................................81
6.8Setting the time ...................................................................................................................................81
6.9Entering the date ................................................................................................................................. 82
6.10Setting the vibration adapter ................................................................................................................. 82
6.11Selecting operating mode of the automatic sample chamber ....................................................................83
6.12Audio signal ....................................................................................................................................... 83
6.13Protecting your settings against change .................................................................................................84
6.14Superimposing or removing the function symbols ................................................................................... 84
6.15Switching the printer on or off ............................................................................................................... 85
6.16Entering company name for printed records ............................................................................................85
6.17Selecting the type of record ................................................................................................................... 86
6.18Defining the free print interval ................................................................................................................86
6.19Switching the statistics and journal functions on and off .......................................................................... 87
6.20Selective clearing of comment lines .......................................................................................................87
6.21Defining the test weight ........................................................................................................................88
6.22Tolerance for balance test .....................................................................................................................88
6.23Selecting the dialog language ...............................................................................................................89
7.2Changing the printer paper and ribbon ...................................................................................................92
7.3Replacing the power line fuses .............................................................................................................. 93
7.4Printer test ..........................................................................................................................................94
7.5Installing the built-in printer ..................................................................................................................94
7.6Dismantling and replacing the heating module .......................................................................................94
8If problems arise on occasion .............................................................................................................96
8.1Notes on the error messages ................................................................................................................96
8.2Meaning of the error messages ............................................................................................................. 97
9.5Switch-off criterion “Weight loss per unit of time” ................................................................................... 104
9.6Comments on certified version ............................................................................................................105
9.7Technical data .................................................................................................................................. 106
9.8MT-SICS Interface commands and functions .........................................................................................109
9.9Optional equipment, expendable material and spare parts ..................................................................... 111
Please read through this Section carefully, it contains important information for safe and economical operation of
your Moisture Analyzer.
1.1Introduction
Thank you for deciding to purchase a Halogen Moisture Analyzer from METTLER TOLEDO – you have made a wise
choice. Your Moisture Analyzer is fast and reliable.It offers a high level of operating convenience and useful
functions to facilitate determination of the moisture content of your samples
Behind your instrument stands METTLER TOLEDO, a leading manufacturer of not only balances and scales for the
lab and production, but also analytical measuring instruments. A customer service network covering the entire globe
with well trained personnel is at your service at all times, whether you are choosing accessories or require guidance
for a specific application to ensure optimum utilization of your instrument.
To ensure you make full use of the possibilities offered by your Moisture Analyzer, we advise you to read through
these operating instructions very carefully.
1.2What is the Halogen Moisture Analyzer used for?
Your Halogen Moisture Analyzer is used for determining the moisture content of almost any substance. The
instrument works on the thermogravimetric principle:At the start of the measurement the Moisture Analyzer
determines the weight of the sample, the sample is then quickly heated by the integral halogen heating module
and the moisture vaporizes.During the drying process the instrument continually measures the weight of the
sample and displays the reduction in moisture.Once drying has been completed, the moisture or solids content
of your sample is displayed as the final result.
Of decisive importance in practice is the rate of heating. In comparison with conventional infrared heating or the
drying oven method, for example, the halogen heating module of your instrument needs a shorter time to reach its
maximum heating power. It also allows use of high temperatures, an additional factor in shortening the drying time.
Uniform heating of the sample material ensures good repeatability of the drying results and makes it possible to
use a smaller amount of sample.
All parameters of a measurement (drying temperature, drying time, etc.) can be preselected. But your Moisture
Analyzer offers many other possibilities. To avoid exceeding the scope of this introduction, only a few of these are
listed here:
– In the built-in methods database 40 different methods can be individually optimized to suit your samples. They
are simple to activate at the touch of a key.
– You can set a reference value for which the optimum switch-off criterion is determined automatically in a test
measurement (AutoMet).
– You can choose between various types of result display at any time.
– Your settings and measurement results can be recorded, stored and protected (3-level password protection).
– Up to 4 comment lines can be entered per measurement (with optional use of a Barcode Reader).
– Thanks to the built-in rechargeable battery your valuable data remain stored even on power failure.
Getting to know your Moisture Analyzer
This wealth of functions notwithstanding, your Moisture Analyzer is very simple to operate. The status display (User
guide) guides you step by step through the measurement cycle. You always know which particular stage in a
measurement is currently being executed by the instrument and the next operating step. To exclude faulty handling
in routine work, the keypad can be locked (3-level password protection) to prevent access to all but the elementary
functions. The motorized, automatic sample chamber offers the operating convenience you are familiar with from
compact disk (CD) players. Moreover, the motorized sample chamber opens up new possibilities for automatic
operation of the instrument (loading by means of robots) when used with the LocalCAN universal interface installed
as standard.
Besides operating convenience, however, the quality of the measurement results is still what counts above all. The
built-in weighing cell meets the high standard which has made METTLER TOLEDO the worldwide market leader in
the field of high-resolution precision balances.
A brief word regarding standards, directives and procedures for quality assurance. The Moisture Analyzer conforms
with all common standards and directives. It supports stipulations, work techniques and result records as demanded
by all international quality assurance systems, e.g. GLP (Good Laboratory Practice), GMP (Good ManufacturingPractice). The instrument has a EC declaration of conformity and METTLER TOLEDO as the manufacturer has been
awarded ISO 9001 and ISO 14001 certification. This provides you with the assurance that your capital investment
is protected in the long term by a high product quality and a comprehensive service package (repairs, maintenance,
servicing, calibration service).
You will find further information in the following Sections of these operating instructions. Please read through this
information carefully to ensure you can use your instrument in an optimum and safe manner.
7
1.3Safety has priority
Your Moisture Analyzer employs state of the art technology and meets the latest demands
regarding instrument safety. This notwithstanding, improper operation can endanger personnel
and cause damage to tangibles. For safe and dependable operation, please comply with the
following instructions:
– The Moisture Analyzer is used for determining the moisture in samples. Please use the
instrument exclusively for this purpose. Any other type of use can endanger personnel and
damage the instrument or other tangibles.
– The Moisture Analyzer must not be operated in a hazardous environment and only under the
ambient conditions specified in these instructions.
– The Moisture Analyzer may be operated only by trained personnel who are familiar with the
properties of the samples used and with the handling of the instrument.
– Your Moisture Analyzer is supplied with a 3-pin power cable with an equipment grounding
conductor. Only extension cables which meet the relevant standards and also have an
equipment grounding conductor may be used. Intentional disconnection of the equipment
grounding conductor is prohibited.
Getting to know your Moisture Analyzer
8
The Halogen Moisture Analyzer works with heat!
– Ensure sufficient free space around the instrument to avoid heat accumulation and
overheating (approx. 1 m free space above the instrument).
– The vent over the sample must never be covered, plugged, taped over or tampered with in
any other way.
– Do not place any combustible materials on, under or next to the instrument when it is
connected to the power supply, since the area around the heating module becomes hot.
– Exercise caution when removing the sample. The sample itself, the automatic sample
chamber and any sample vessels used may still be very hot.
– During operation, you should never open the heating module as the ring-shaped heating
module or its protective glass can reach 400 °C! If you have to open the heating module on
occasion, disconnect the instrument from the power supply and wait until the heating module
has cooled down completely.
– No modifications must be made within the heating module. It is particularly dangerous to
bend any components or remove them, or to make any other changes.
Certain samples require special care!
With certain types of samples, there is a possibility of danger to personnel or damage to
tangibles through:
Fire or explosion:
– Flammable or explosive substances
– Substances containing solvents
– Substances which evolve flammable or explosive gases or vapors when heated.
With such samples, work at a drying temperature that is low enough to prevent the formation
of flames or an explosion and wear protective goggles. Should there be any uncertainty
regarding the flammability of a sample, always work with a small amounts of sample (max.
1 gram). In such cases, never leave the instrument unattended! In cases of doubt, perform
a careful risk analysis.
Poisoning, burning:
– Substances which contain toxic or caustic components. Such substances may be dried only
in a fume cupboard.
Corrosion:
– Substances which evolve corrosive vapors when heated (e.g. acids). In the case of such
substances, we advise you to work with small amounts of sample as the vapor can condense
on cooler housing parts and cause corrosion (if necessary, the heating module can be
changed very easily by the user, see Section 7.6).
Please note that the user always takes responsibility and assumes liability for damage
caused by use of the types of samples mentioned above!
Getting to know your Moisture Analyzer
– Never make any modifications or constructional alterations to the instrument and use only
original spare parts and optional equipment from METTLER TOLEDO.
– Your Moisture Analyzer is a rugged precision instrument – but you should still treat it carefully;
it will then thank you with many years of trouble-free operation.
– Please comply with all notes and instructions in these operating instructions. Keep the
instructions in a safe place where they are immediately to hand if any points are unclear.
If you lose these instructions, please contact your METTLER TOLEDO dealer for an
immediate replacement.
Moisture determination applications must be optimized and validated by the user according
to local regulations. Application-specific data provided by METTLER TOLEDO is intended for
guidance only.
9
1.4Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste.
This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which
you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content
of this regulation must also be related.
Thank you for your contribution to environmental protection.
Getting to know your Moisture Analyzer
10
1.5Important information for these instructions
These instructions guide you step by step through the operation of your Moisture Analyzer. The first two Sections
help you put the instrument into operation quickly, safely and properly and perform your first measurement within
a short space of time. In Sections 3 through 6 you become intimately acquainted with the wide range of functions
of your Moisture Analyzer. During this learning phase, you will find the table of contents and the detailed index in
Section 10 a valuable orientation aid. Sections 7 through 9 contain additional information on the maintenance of
your instrument, troubleshooting and the available options. As soon as you are familiar with your Moisture Analyzer,
you will find the illustrations at the beginning of these instructions and the associated references (in the key) useful
for quick access.
The following identifications and symbols are used in these instructions:
– Key designations are shown enclosed by twin angle brackets «». Keys with inscribed text are
shown with the actual inscription (e.g. «On/Off» or «Start»). For all keys with symbols,
designations are used in the text which describe the function of the particular key («Print»,
«Accept entry» or «Scroll down»).
– This symbol indicates safety and hazard instructions. If these are not complied with, injury
to the user, damage to your instrument or other tangibles and malfunctions can result.
– This symbol indicates additional information and directions which facilitate your handling of
the instrument and contribute to proper and economical use.
These instructions are also available in foreign languages. Should you require a set of
instructions in a different language, please contact your METTLER TOLEDO dealer. You will find
the address of your nearest dealer in the attached “Declaration of Conformity 11780294”.
Your first measurement in next to no time
11
2Your first measurement in next to no time
In the Section you will learn how to put your new Moisture Analyzer into operation and obtain measurement results
within a very short space of time.
2.1Unpacking and checking the standard equipment
Open the package and remove the instrument and the accessories. Check the completeness of the delivery. The
following accessories are part of the standard equipment of your new Moisture Analyzer:
–1 set of operating instructions, 1 set of short-form operating instructions
–1 Application brochure “Methods of moisture content determination”
–1 EC declaration of conformity (in separate brochure 11780294)
– Setup instructions
Remove the wrapping from the instrument.
Check the instrument for transport damage. Immediately inform your METTLER TOLEDO dealer if you have any
complaints or parts are missing.
Store all parts of the packaging. This packaging guarantees the best possible protection for the
transport of your instrument.
To prevent transport damage, the instrument is provided with a transport arrestment, which you
have to remove before putting the instrument into operation for the first time. The procedure for
removing the transport arrestment is explained in the description “Setup instructions 11780556”
enclosed with the package.
Your first measurement in next to no time
12
2.2Selecting the location
Your Moisture Analyzer is a precision instrument. An optimum location guarantees accuracy and dependability:
– Firm, horizontal location as free from vibrations as possible
– Avoid direct sunlight
– No excessive temperature fluctuations
– No powerful drafts
– Surroundings as free from dust as possible
– Adequate free space at the front for opening the sample chamber and
sufficient clearance around the instrument to allow warm air to dissipate.
– Sufficient distance from heat-sensitive materials in the vicinity of the
instrument.
2.3Setting up, leveling and connecting to power supply
Exact horizontal positioning and stable installation are prerequisites for repeatable results. To compensate small
irregularities or inclinations (±2%) at the location, the instrument can be leveled.
For exact horizontal positioning, the Moisture Analyzer has a level indicator
R
L
R
(level) and 3 leveling screws. As soon as the air bubble is precisely in the
L
middle of the cross hairs, the instrument is exactly horizontal. To level it,
proceed as follows:
Position your Moisture Analyzer at the selected location.
Screw in the right front leveling screw completely (it is not needed for
leveling). The instrument is now standing on the front left, nonadjustable
foot, as well as on the two rear leveling screws.
Turn the two rear leveling screws until the air bubble is located in the middle
of the level indicator. The arrows and the leveling screw designations (R =
right rear leveling screw, L = left rear leveling screw) on the cross hairs will
facilitate the setting:
Your first measurement in next to no time
13
R
L
indicates the right leveling screw must be turned in a clockwise direction.
Unscrew the front right leveling screw until it contacts the support and the
If the air bubble is located at the bottom left, for instance, the inscription
R
L
instrument is resting firmly on all 4 feet. Ensure that the air bubble remains
in the center of the level indicator.
R
L
R
Note: The instrument should be releveled each time its location is changed.
L
Connect the instrument to the power supply.
Warning
If the power cable supplied is not long enough, use only a 3-pin extension
cable with equipment grounding conductor!
For technical reasons, the halogen heating module is designed specifically
for a particular line voltage (110 V AC or 230 V AC). A heating module is
installed in the factory that is matched to the particular line voltage of the
country of destination. If you are not sure whether the heating module built
into your instrument is suitable for your local line voltage, check the voltage
data in the interior of the heating module before you connect the Moisture
Analyzer to the power supply!
Connection to a line voltage that is too high can lead to blowing of the fuses,
whereas a supply voltage that is too low will prolong the drying process.
On/Off
Ensure that the transport arrestment is removed. Instructions on how to
remove the transport arrestment can be found on the leaflet enclosed with the
package.
Switch the instrument on with the «On/Off» key.
Note: When putting into operation for the first time, leave the instrument
connected to the power supply for at least 5 hours to allow the built-in battery
to charge up! If the instrument is later disconnected from the power supply
for several months, the battery will become discharged. This leads to data
loss. To recharge the battery, leave the instrument connected to the power
supply for at least 5 hours.
Your first measurement in next to no time
14
Press the «Open/close auto sample chamber» key and ...
2
… the motorized, automatic sample chamber opens.
Always ensure sufficient free space in front of the instrument since the
emerging automatic sample chamber could knock over objects in its path
or push them off the bench.
Install the draft shield element (only one position possible) and then the
sample pan holder. Turn the sample pan holder until it engages. In the locked
position, the rear arm of the sample pan holder points exactly in the direction
of motion of the automatic sample chamber.
2
Press the «Open/close auto sample chamber» key and ...
… the motorized, automatic sample chamber closes.
Your first measurement in next to no time
15
2.4Preparing the printer
If your Moisture Analyzer is not equipped with a built-in printer, there is no need to read this Section. You can always
order the built-in printer at a later date and install it yourself.
You can prepare the built-in printer for use as follows:
Disconnect the instrument from the power supply.
Swing back the cover of the printer.
HR73
HalogenMoistureAnalyzer
HR73
Halogen Moisture Analyzer
HR73
Halogen Moisture Analyzer
Insert the ribbon cassette in the printer and press it down until you hear it click
into place.
EJECT
H
S
U
P
Tension the ribbon by turning the small wheel on the left of the cassette in
the direction shown by the arrow.
EJECT
H
S
U
P
Your first measurement in next to no time
16
Place the roll of paper in the paper compartment and feed the paper
horizontally through the slot at the back of the printing unit.
HR73
Halogen Moisture Analyzer
EJECT
H
S
U
P
Connect the instrument to the power supply and switch it on with the «On/
On/Off
Off» key. Press the «Paper feed» key until the leading edge of the paper exits
the printing unit at the top.
Lead the paper through the slot in the cover of the built-in printer and close
the cover.
HR73
Halogen M
oisture Analyzer
EJECT
H
S
U
P
2.5Setting the date and time
Your Moisture Analyzer prints out the date and time on every record. When you put your new instrument into operation
for the first time, you should enter the current date and the time. These settings are retained even if you disconnect
your instrument from the power supply.
In this Section you will enter the menu for the first time, although this is described in detail in a later section. Do not
worry if you have to press keys whose function you are not familiar with, simply follow the instructions step by step.
We describe the menu in detail in Section 6.
Note: Various languages are available for the dialog with your instrument and can be selected in the menu. You
will find further details on language selection in Section 6.23.
Press the «On/Off» key to switch the instrument on.
On/Off
After it has been switched on, the instrument performs a self-test. Wait until
the display shown opposite appears.
Your first measurement in next to no time
17
Menu
Menu
P∑SEÎNO
Menu
TIMË1‹04
Menu
TIMË1Æ28
Press the «Menu» key. You are now already in the menu.
Press the «Accept entry» or «Menu» key repeatedly until the display shown
1
opposite for entry of the time appears.
Important: Do not press any other key, otherwise you could possibly change
the factory settings of your instrument!
Use the «Scroll up» and «Scroll down» keys to set the time (00:00–23:59).
By pressing and holding the keys you can accelerate the change in the time.
1
Confirm your setting with the «Accept entry» key. The instrument stores the
time.
Reset
Menu
DAÎ2{0“03
Menu
DAÎ 0[0#0#
After confirmation of the time, the display for entry of the current date appears.
2
Use the «Scroll up» and «Scroll down» keys to set the date in the format “Day
– Month – Year” (DD.MM.YY). By pressing and holding the keys you can
accelerate the change in the date. Confirm your entry with the «Accept entry»
2
key and the instrument stores the date.
Note: With US English as the dialog language, the following applies to entry
of the date format “Month – Day – Year” (MM/DD/YY).
This completes your entry of the date and time, and you can now quit the
menu by pressing the «Reset» key. Your settings are stored and remain so
even in the event of a power failure. The Moisture Analyzer is now ready for
your first measurement.
Your first measurement in next to no time
18
2.6Your first measurement
After you have successfully put your new Moisture Analyzer into operation for the first time, you can immediately
perform your first measurement. In doing so, you will become familiar with the instrument and the status display
and at the same time perform a function check.
Please use the specimen sample supplied for your first measurement. This sample is an absorbent glass fiber filter.
During your first measurement, the instrument operates with the factory settings.
Your Moisture Analyzer has a novel graphical status display (User guide)
which you will become acquainted with in this section. The status display
informs you continuously about the current status of the instrument and
prompts you to execute the next operating step (shown flashing).
Switch the instrument on with the «On/Off» key.
On/Off
2
Initial status with closed automatic sample chamber.
Press the «Open/close auto sample chamber» key and …
… the motorized, automatic sample chamber opens.
Always ensure sufficient free space in front of the instrument since the
emerging automatic sample chamber could knock over objects in its path
or push them off the bench.
Status: Ready for taring
The status display flashes and prompts you to load the empty sample pan
and tare.
#
Your first measurement in next to no time
19
Place the empty sample pan in the sample pan handler (this is possible
without tilting the sample pan if you insert this in the pan handler from the
side directly below the round flange). Place the sample pan handler in the
automatic sample chamber. Ensure that the tongue of the pan handler fits
exactly in the slot of the draft shield element. The sample pan must lie flat
in the pan holder.
Note: We advise you to work with the sample pan handler at all times. The
pan handler is ergonomic, safe and provides protection against possible
burns due to the hot sample pan.
Press the «Tare (zero)» key. This sets the balance installed in the Moisture
Analyzer to zero. For taring, the automatic sample chamber closes briefly
then opens again.
Status: Ready for weighing
After taring, the status display flashes and prompts you to place the sample
in the sample pan.
Place the specimen sample in the sample pan.
Wet the specimen sample with a few drops of water.
Status: Ready for start
The status display flashes and prompts you to start the drying process.
Your first measurement in next to no time
20
Press the «Start» key and …
Start
… the motorized, automatic sample chamber closes. The instrument starts
the drying and measurement.
Status: Drying and measurement
You can follow the drying and measurement process in the display:
– The status display uses rising bubbles to symbolize the drying process.
– The current temperature in the heating module is displayed as well as the
104
≠30
-|84
– Further, the display shows information on the selected settings. You will
elapsed drying time and the current drying value.
find additional information on the meaning of the displays and the various
setting possibilities in Section 4. If your instrument has a built-in printer,
the measurement results will be printed out continually (at intervals of
30 seconds).
‹10
ENfl -1Ç52
At the end of the drying time, an audio signal sounds and the sample
chamber is automatically opened.
Warning: Sample pan and sample are still hot!
You can now read off the moisture content of your sample in the display.
59
Your first measurement in next to no time
21
Carefully remove the sample pan handler from the automatic sample
chamber.
Warning: As the pan and sample may still be hot, you should let these
cool down before removing the pan from the handler!
Reset
To remove the sample pan from the handler, lift the pan slightly from below
and pull it sideways out of the handler (if you no longer need the sample and
the pan, you can simply tilt the handler until the pan slides out).
Press the «Reset» key to delete the measurement result from the display. The
instrument is now ready for the next measurement. If you do not wish to
perform another measurement, press the «Open/close auto sample chamber» key. The automatic sample chamber closes.
Congratulations!
You have just performed your first measurement with your new Moisture
Analyzer. In the following section you will find important information on the
operating principle of your instrument, its calibration and optimum preparation of your samples.
Notes on the automatic sample chamber
– Before every automatic closure of the automatic sample chamber, an
audio signal sounds.
– For reasons of safety, the open automatc sample chamber is automati-
cally closed after 2 minutes. If a weight change occurs or a key is pressed
during this time, the 2 minute period starts over again.
How to obtain the best results
22
3How to obtain the best results
Following your first practical work with the Moisture Analyzer, in this Section you will find important information on
how to obtain optimum results. You will discover what parameters influence the measurement process and how
you can match the instrument optimally to your particular measurement task.
3.1Measurement principle of the Halogen Moisture Analyzer
Your instrument performs measurements based on the thermogravimetric principle, i.e. the moisture is determined
from the weight loss of a sample dried by heating.
In principle, your instrument thus comprises two instruments: a precision
120
balance and a heating module. In contrast to other thermogravimetric
methods (drying oven, infrared, microwave), the Halogen Moisture Analyzer
operates with a halogen heating module. This ensures fast heating of the
sample and thus guarantees rapid availability of the measurement results.
Irrespective of the measurement method, the quality of the measurement results stands or falls by the preparationof the sample and a correct choice of the important measurement parameters:
– Sample size
– Drying temperature
– Switchoff mode
– Drying time
Inappropriate setting of these parameters can cause the results to be incorrect or
misleading. For this reason, check that the results for each type of sample are what might
reasonably be expected.
You will find detailed information on the relationships between these parameters in the application brochure for
moisture determination (see Section 9.4), enclosed to your instrument.
In practice, however, not only the quality of the measurement results, but also the speed of the measurementprocess is important. Thanks to its drying principle (with the heat generated by a halogen radiator), the Halogen
Moisture Analyzer is very fast. You can increase the speed even further through optimum setting of the instrument.
The optimum drying temperature and the drying time are dependent on the nature and size of the sample and
on the desired accuracy of the measurement results. These can be determined only by experiment. The Halogen
Moisture Analyzer supports you in this task: It offers automatic performance and result recording of testmeasurements. The ideal settings can quickly be determined with AutoMet (see Section 4.12).
How to obtain the best results
23
3.2Notes on adjustment of the balance and the heating module
You should adjust your Moisture Analyzer regularly (e.g. every six months) if this is stipulated by your quality
assurance system (e.g. GLP, GMP, ISO 9001).
The balance and the integral heating module in your instrument can be adjusted using the appropriate accessories
(see Section 9.9).The heating module and balance can also be tested (to check the adjustment).The operator
can define a test weight or the test temperature, together with permissible tolerances for this test.A test report is
printed out showing the test result as “Passed” or “Failed” (see Section 6.3 and 6.5).
By adjusting the heating module you can ensure comparability with results from other instruments of the same
design.This is done by using an appropriate thermometer, which is provided in the form of a temperature
adjustment set with calibration certificate (see Section 9.9).This adjustment set comprises a black plate with a
thermometer, which allows a reproducible adjustment of the heating module.
A Moisture Analyzer is typically used in place of or in addition to the oven method.In an oven heat energy is
transferred by the flow of air, which establishes an equilibrium between the sample temperature and the ambient
temperature.This is not the case in a Moisture Analyzer.The actual sample temperature primarily depends on the
specific absorption properties of the sample (“dark” samples absorb more heat), which can change during the
measurement process.There can also be differences between the temperature at the surface of the sample and the
temperature inside the sample.The heat output is therefore not dependent on the true sample temperature but
instead is regulated by a temperature sensor underneath the halogen heating module.
For the reasons explained above, the temperature setting on the instrument will therefore by deviate to the true
sample temperature.By regularly adjusting the heating module, you will ensure a consistent and reproducible heat
output for the entire lifetime of your instrument.
Note: METTLER TOLEDO offers an adjustment service – please contact your local dealer.
– If you also use your instrument as a high precision lab balance, you should adjust the balance at regular intervals.
– If you replace the heating module, you must adjust it using the temperature adjustment set, which is available
as an accessory.
The procedure for adjusting the balance and heating module is described in Section 6.
3.3Optimum sample preparation
Preparation of the sample is decisive for the speed of the measurement process and the quality of the measurement
results.
Please note the following basic rules for the preparation of your sample:
The amount of sample you select should be as small as possible and only as large as necessary
Excessive amounts of sample require more time for drying and thus prolong the measurement process. If the amount
of sample is too small, the measurement result may possibly not be representative. The following always holds:
The greater the inhomogeneity of the sample, the larger the amount of sample needed to obtain a repeatable result
(please see also the list in Section 9.1).
How to obtain the best results
24
Distribute the sample evenly over the sample pan
You thus increase the surface area of the sample and facilitate heat absorption. The base of the pan should be evenly
covered.
With liquid, fat-containing, melting and highly reflecting samples, you should cover the sample with the glass fiber
filter available as optional equipment (HA-F1 214464, see Section 9.9). This also applies to samples which form
a skin on their surface when heated. The glass fiber filter ensures even and rapid heat distribution and prevents the
formation of a skin impervious to moisture on the sample surface.
3.4Further information on moisture determination
You will find further information on the moisture determination, the importance of the parameters and the preparation
of the samples in the Application Brochure “Methods of moisture content determination” dealing with moisture
determination (see Section 9.4).
Useful tips and a variety of example methods (comparison between Halogen Moisture Analyzer results and the oven
method) can be downloaded from www.mt.com/moisture.
If you require information on specific applications, your METTLER TOLEDO dealer will be pleased to help you.
Moisture determination applications must be optimized and validated by the user according
to local regulations. Application-specific data provided by METTLER TOLEDO is intended for
guidance only.
Practical application of your Moisture Analyzer
25
4Practical application of your Moisture Analyzer
In this Section we introduce you to the wide range of setting possibilities of your Moisture Analyzer and offer
information and tips for optimum setting of the parameters.
4.1Operating principles
You already performed your first measurement in Section 2. This measurement was carried out using the factory
settings. However, your instrument offers a number of setting possibilities to match the measurement process to
your specific requirements. For instance, you can preselect the drying temperature, the type of result display and
many other parameters.
Settings are made using the function keys. These keys are right at the top of
the keypad, directly below the display. The setting you select appears in the
display directly above the associated key.
95
123
%
95
125
987
654
You will learn the available setting possibilities and the function keys in detail
in the following Sections. Further, you will also discover how to set the drying
temperature and the drying time and how to work with the weighing-in aid.
Finally, you will perform a complete measurement with the specific settings
for your sample.
Numeric entries:
With some settings (e.g. the drying temperature), numeric entries are
required. There are two different ways of entering numerical values:
You can change the flashing displayed value with the «Scroll up» or «Scroll
down» key or ...
... you can key in the new value directly with the numeric keypad.
You can remove wrong numbers with the «Delete» key.
0
.
125
Practical application of your Moisture Analyzer
26
As soon as you have entered the desired value, press the «Accept entry» key.
The value will be stored.
Preliminary information on the methods:
All settings for a sample which you will learn in the following Sections can be collected in a so-called “Method”.
You have 40 methods freely available thus giving you the possibility of defining a method for a particular sample
and recalling it at any time. This eases the work load considerably as each time you switch to a different type of
sample there is no need to repeat all the parameter settings. The methods are retained when the instrument is
switched off. You will find information on the compilation and use of methods in Section 5.
4.2Selecting the drying program
This function key offers you four different drying programs for optimum matching of the drying characteristics to the
sample used.
Each time this key is pressed, the next program appears. After the fourth and
last program, the first reappears in the display. As soon as you have selected
the desired program, the setting is active and you do not have to confirm or
store it.
Standard drying
This drying program is set in the factory and is suitable for most samples.
The sample is heated to the drying temperature (set temperature) and held
constant at this temperature.
Rapid drying
This program is primarily suitable for samples with a moisture content over
30%. Following the start, the selected temperature is exceeded by 40% for
3 minutes to compensate the cooling due to vaporization and accelerate the
drying process. The drying temperature is then lowered to the set value and
maintained. You will find details on how to enter the drying temperature in
the Section 4.3.
Gentle drying
This program is suitable for the gentle drying of substances which tend to
form a skin (e.g. substances containing sugar). With this program the
temperature is continuously increased and attains the selected drying
temperature only after the elapse of the so-called "ramp time". The “ramp”,
i.e. the time that should elapse between the start of drying and attainment of
the final temperature is preselectable. The setting of the “ramp” is described
in the next Section 4.3.
Practical application of your Moisture Analyzer
27
Step drying
This program is suitable for the drying of substances composed of several
components which vaporize at different temperatures (e.g. ethereal oils).
With this program drying is performed stepwise, i.e. the sample is preheated
to a particular temperature (1st step) and kept at this temperature for the
selected time (hold time).
The temperature is then increased to the next value (2nd step) and again
maintained for a certain time. Finally, the temperature is raised to the
preselected drying temperature (set temperature) and held constant at this
value until the end of drying. With this program you can preselect the
temperature and time for each step. These settings are described in the
Section 4.3. If only one step is needed, the time of step 2 must be set to 0.
Step drying can also be useful for the rapid determination of substanceswith a high moisture content. Here, the 1st step is set higher than the final
temperature to compensate the cooling due to vaporization.
4.3Setting the drying temperature
After pressing this key you can enter the drying temperature (set temperature).
The type of entry depends on the selected drying program (see Section 4.2).
If you have selected one of the programs “Standard drying” or “Rapid drying”,
you need enter only the drying temperature. For the two programs “Gentle
drying” and “Step drying”, additional entries are required to define the ramp
or the steps. The different entry procedures are described below.
Note: The admissible input range for temperatures is 40°C–200 °C.
Selecting drying temperature for standard drying and rapid drying
Final temperature
Start of drying
You need enter only the desired final temperature.
Press the «Drying temperature» key and enter the desired drying temperature
(see Section 4.1).
– either with the «Scroll up» and «Scroll down» keys or
– directly via the numeric keypad
125
Press the «Accept entry» key.
Practical application of your Moisture Analyzer
28
Selecting drying temperature for gentle drying
Final
temperature
Ramp time
Start of drying
RAM∏‹00
RAM∏⁄30
For this temperature program you first define the so-called “ramp”, i.e. the
time that should elapse between the start of drying and attainment of the final
temperature. You then select the drying temperature (final temperature). The
ramp is started once the temperature reaches 50 °C.
Press the «Drying temperature» key.
Entering the ramp time:
The instrument prompts you to define the ramp time. In the factory the ramp
time is set to 3 minutes.
Enter the desired value.
Press the «Accept entry» key.
Entering the drying temperature:
125
The instrument now prompts you to enter the drying temperature (final
temperature). Enter the desired temperature (e.g. 125 °C).
Press the «Accept entry» key.
Selecting drying temperatures for step drying
For this temperature program you define the temperature and the hold time
for each step.
Press the «Drying temperature» key.
Temp.
Step 1
Hold time
Step 1
Start of drying
Temp.
Step 2
Final
temperature
Hold time
Step 2
TEMπ∞100
TIM™∞⁄30
TEMπ”125
Practical application of your Moisture Analyzer
29
Entering the drying temperature and hold time for the first step:
The instrument prompts you to define the temperature for the first step. Enter
the desired temperature.
Press the «Accept entry» key.
The instrument now prompts you to enter the hold time for the first step. Enter
the desired time.
Press the «Accept entry» key.
Entering the drying temperature and hold time for the second step:
The instrument now requests entry of the temperature for the second step.
Enter the desired temperature.
TIM™”‹00
160
Press the «Accept entry» key.
Define the hold time for the second step.
Press the «Accept entry» key.
Entering the final temperature:
Finally, you enter the desired final temperature.
Press the «Accept entry» key.
Practical application of your Moisture Analyzer
30
Temperature limitation
°C
200
160
Start
1050
t (min.)
With regard to high temperatures, the moisture analyzer has several
protective mechanisms that work independently of each other:
– At temperatures above 160 °C a time limit becomes active. The higher the
temperature, the shorter the time until the instrument starts to lower the
temperature (envelope curve, see illustration opposite). Note: If necessary, this time-controlled lowering of the temperature can be deactivated in
the Methods menu (Section 5.4). If deactivated, temperatures from
160 °C - 200 °C are available without time limitation.
–A temperature switch independent of the normal black temperature sensor
over the sample (see Section 7.1) is triggered when it detects exceptionally high temperatures. This situation could arise, for example, if the
sample begins to burn or the heating module adjustment is defective. If
this happens, the heating module is tripped out by a mechanical switch,
and the instrument cannot be restarted. Please refer to Section 8.3 for
troubleshooting.
When working with temperatures over 180 °C we recommend that you
always wait 2 - 3 minutes between individual measurements, in order to
ensure good reproducibility of the readings.
Note: A measurement cannot be started until the drying temperature is
higher than the temperature in the measuring room.This avoids a false start
(Error 10, Section 8.2).
4.4Selecting the switch-off criterion
This function key offers you different switch-off criteria. A switch-off criterion defines when the instrument should end
the drying. Switch-off criteria save you having to keep checking your watch and stopping the drying manually.
The following settings can be selected for the switch-off criterion:
– “Weight loss per unit of time” (5 settings)
– “Free switch-off criterion” (if activated in the Methods menu,
Section 5.4)
– “Manual switch-off”
– “Timed switch-off”
– “Test measurement” for determination of the suitable switch-off criterion
Weight loss per unit of time
This switch-off criterion is based on a weight loss per unit of time. As soon as the mean weight loss is less that a
preset value during a specified time, the instrument considers drying as complete and automatically discontinues
the measurement process. During the drying, the time display shows you how long the measurement process has
been in progress. The switch-off criterion is inactive during the first 30 sec. You will find additional information in
Section 9.5.
Practical application of your Moisture Analyzer
31
The following 5 settings are available:
This setting is suitable for samples which dry very quickly (surface moisture)
or for (relatively inaccurate) fast measurements to determine a trend.
This setting is suitable for quick drying samples.
This is the factory setting. It is suitable for most types of samples.
This setting is suitable for samples which dry moderately quickly.
This setting is suitable for samples which dry very slowly (trapped moisture,
skin formation).
Note: If you do not know the behavior of a sample, perform a test measurement to determine the suitable switch-off criterion. The procedure is described later in this section under the title “Test measurement”.
Free switch-off criterion
This setting possibility is available only if you have activated it in the Methods menu (Section 5.4). There is one
free switch-off criterion available for each method. The free switch-off criterion is based on a user-defined mean
weight loss per unit of time.
You will find information on the activation and definition of the individual
switch-off criterion in Sections 5.4.1 and 9.5.
Note: The free switch-off criterion can be determined automatically with
AutoMet (Section 4.12).
Manual
With this switch-off criterion the measurement process continues until you stop it with the «Stop» key.
The time shown in the display tells you how long the current measurement
has been in progress.
Practical application of your Moisture Analyzer
32
Timed switch-off
With this switch-off criterion the measurement lasts until the preset drying time has elapsed (the time display
provides you with continuous information on the drying time).
If you select this criterion, you must then enter the desired drying time as
follows:
Press the «Drying time» key.
⁄30
Test measurement
Enter the desired drying time.
Press the «Accept entry» key.
Press “t” to perform a test measurement which will help you to determine
suitable switch-off criteria (Section 4.11).
Using AutoMet test measurement is a quick way of determining a Halogen
Moisture Analyzer method that corresponds most closely to your reference value
(target value) (Section 4.12).
Practical application of your Moisture Analyzer
33
4.5Selecting the display mode
With this function key you can select the desired type of result display. You also define what values are printed on
the records.
The following display modes are available:
Moisture content
The moisture content of the sample is displayed (and printed out) as a
%
percentage of the wet weight (= ww = initial weight = 100%). This is the
factory setting.
When the results are printed out, the moisture content is marked by “%MC”
(Moisture Content) (e.g. –11.35 %MC) and shown as a negative value.
WW
WW
%
100%
Before
drying
100%
Before
drying
After
drying
After
drying
DW
DW
MC
=
[%]
MC [0…100%] = – *
Wet weight WW – Dry weight DW
Wet weight WW
100%
During the drying the current measured value is continuously displayed in
percent.
Dry content
The dry content of the sample is displayed (and printed out) as a percentage
of the wet weight (= ww = initial weight = 100%).
When the results are printed out, the dry content is marked by “%DC” (Dry
Content) (e.g. 88.65 %DC).
DC
=
[%]
During the drying the current measured value is continuously displayed in
DC [100…0%] =*
Dry weight DW
Wet weight WW
100%
percent.
Weight in grams
%
The weight of the sample is displayed (and printed out) in grams. With this
setting, the Moisture Analyzer is used as a precision balance.
During the drying the current weight is continuously displayed in grams.
Practical application of your Moisture Analyzer
34
ATRO moisture content
The moisture content of the sample is displayed (and printed out) as a
%
percentage of the dry weight ( = DW = final weight = 100%).
When the results are printed out, the ATRO moisture content is marked by
“%AM” (ATRO Moisture Content) (e.g. –255.33 %AM) and shown as a
negative value
AM
=
WW
Before
drying
100%
After
drying
[%]
DW
=
During the drying, the current measured value is continuously displayed in
percent.
ATRO dry content (Wet weight)
The wet weight of the sample is displayed (and printed out) as a percentage
%
of the dry weight (= DW = final weight = 100%).
When the results are printed out, the ATRO dry content is marked by “%AD”
(ATRO Dry Content) (e.g. 312.56 %AD).
AM [0…-1000%] = – *
Wet weight WW – Dry weight DW
Dry weight DW
100%
=
[%]
AD
WW
Before
drying
100%
After
drying
DW
=
AD [100…1000%] =*
Wet weight WW
Dry weight DW
100%
During the drying, the current measured value is continuously displayed in
percent.
Comment on the ATRO display mode
If the current measured value in the ATRO display mode is greater or less than
the predefined limit value (i.e. greater than 999.99 %AD or less than
–999.99 %AM), a warning beep sounds and the instrument automatically
switches the display mode (from %AM to %MC and from %AD to %DC). A
display in the ATRO mode is no longer possible in this case, even if you have
started your drying operation in the %MC, %DC or “g” (grams) display
mode. If your measurement series contains such an inadmissible ATRO
measured value, the statistical evaluation (see Section 5.7) will also be
performed in the new, automatically selected mode.
Practical application of your Moisture Analyzer
35
4.6Defining the print interval
One point before we start: This setting possibility is available only if the built-in printer is activated in the menu
(see Section 6.15).
The print interval determines the printout frequency of the intermediate results of an ongoing measurement.
The following settings are available under this function key:
Printout at intervals
These 6 settings allow printout of the intermediate results at fixed, preset
intervals. You can thus trace the drying process using the printed record. The
following print intervals are available:
– Printout every 5 seconds
– Printout every 10 seconds
– Printout every 30 seconds (factory setting)
– Printout every minute
– Printout every 2 minutes
– Printout every 10 minutes
Free print interval
This setting possibility is available only if you have activated it in the menu.
You will find information on the activation and definition of the free print
interval in Section 6.18.
Manual printout
With this setting there is no automatic printout. However, you can print out
the intermediate results at any time with the «Print» key.
Practical application of your Moisture Analyzer
36
4.7Working with the weighing-in aid
The weighing-in aid can be defined for each method and facilitates weighing in of the sample to a preset weight
value (target value).This is particularly useful if you require all the samples you wish to process to have same weight
in order to improve the repeatability of the measurement results. In addition, the weighing-in aid can be configured
so that the drying process cannot be started if the weight of sample is outside the set tolerance. You are therefore
compelled to weigh in the correct quantity of sample (“Weighing-in aid active”).
Target weight
lower weight limit
Underweight
WGªBAÈOFF
WGªBAÈPAS
upper weight limit
Overweight
The weighing-in aid in the bottom right corner of the display comprises two
opposite facing bars and triangles with a minus or plus symbol.
The weighing-in aid is available only if it has been activated. If the minus
or plus symbol is visible after taring, the weighing-in aid is activated.
If you want to switch on the weighing-in aid, press the «Target weight
(Weighing-in aid)».
In the factory the weighing-in aid is switched off (“OFF”).
You can use the «Scroll up» and «Scroll down» keys to select either the
“Weighing-in aid passive” or “Weighing-in aid aktive” setting.
If the “Weighing-in aid passive” setting is selected, the weighing-in aid is
switched on and a target weight can be defined. With this setting the drying
process can be started even if the sample weight lies outside the set limits.
These are defined by the target weight and the sample weight tolerance as
a % of the target weight (see below).
WGªBAÈACT
TAR‚“500
With the weighing-in aid set at “Weighing-in aid active” the drying process
cannot be started if the sample weight is outside the set limits. You are
therefore compelled to weigh in the correct quantity of sample, and are
prevented from inadvertently starting the drying process with the wrong
sample weight.
After selecting the desired mode for the weighing-in aid, confirm your
selection by pressing the «Accept entry» key.
Enter the desired target weight (0.100 g to 80.000 g) via the numeric
g
keypad (Factory setting: 2.5 g) and confirm with the «Accept entry» key.
Practical application of your Moisture Analyzer
37
BA®LI˚ 10
%
keypad) to define the sample weight tolerance in the range 1% to 25%. The
sample weight tolerance (Factory setting: 10%) defines the lower and
upper sample weight limits. Confirm your entry with the «Accept entry» key.
The symbols of the weighing-in aid appear in the display after you have tared
the balance.
How the weighing-in aid operates
Before the start of weighing (standby mode)
Start of weighing (sample weight too low)
Lower weight limit reached (sample weight in tolerance range). The status
display (user guide) indicates with its flashing ‘rising bubble’ graphic that
the instrument is ready to run.
You can now use the «Scroll up» and «Scroll down» keys (or the numeric
Target weight reached (sample weight in tolerance range). The status
display (user guide) indicates with its flashing ‘rising bubble’ graphic that
the instrument is ready to run.
Upper weight limit reached (sample weight in tolerance range). The status
display (user guide) indicates with its flashing ‘rising bubble’ graphic that
the instrument is ready to run.
Upper weight limit exceeded (sample weight too high)
After initiation of the measurement with the «Start» key, the bars of the
weighing-in aid disappear.
Practical application of your Moisture Analyzer
38
4.8Comment lines
You can include up to 4 comment lines in your measurement records, e.g. in order to enter different sample IDs.
These comment lines are called C1 to C4 (see Section 6.20).The alphanumeric comment lines can be up to 20
characters in length and are method specific.To enter a comment:
Press the «Code» key. The input field for the record header comment appears.
The name of the comment line (C1 to C4) appears on the display in input
mode.
If a comment has been entered previously, it appears in the display. The
procedure for changing existing comments is described later. If you want to
enter a new comment, or completely overwrite an old one, proceed as
follows:
Press the «A…Z» key. A flashing “A” appears at the first input position on the
extreme left and signals that the instrument is ready for your entry.
You can now use the «Scroll up» and «Scroll down» keys to select the first
letter of the comment (uppercase letters only) or a special character (+, -,
etc.).
CA
A...Z
123
0
.
Confirm your selection with the «A…Z» key. The flashing “A” again appears,
but at the second position.
987
654
Enter a further character of the desired comment as described above.
Numbers can be entered via the numeric keypad.
Notes
–A space is symbolized by an underline character “_”.
– You can remove wrong characters with the «Delete» key and then correct
your entry.
Code
A...Z
Practical application of your Moisture Analyzer
39
Changing a comment line
If you want to change an existing comment (e.g. increasing a sample
number, entering a serial number), proceed as follows:
Press the «Code» key. The most recently saved comment appears (provided
that it has not been automatically cleared, see Section 6.20).
Press the «Delete» key. A flashing dash appears after the last character,
indicating input mode.
You can then add additional characters by pressing the «A…Z» key or a
number key on the numeric keypad.
By pressing the «Delete» key a second time you can delete the last character.
The second last character flashes and can be changed using the numeric
keypad or the «Scroll up» and «Scroll down» keys, or confirmed with the
«A...Z» key. Further characters can then be added as described above.
CHARGE1]0
Code
Code
As soon as you have finished entering your comment, press the «Accept
entry» key to quit the input mode (the flashing character will also be
accepted). From now on your comment will be automatically printed out on
each measurement record.
Defining different comment lines
Press the «Code» key once to enter comment C1. Press the «Code» key
repeatedly to enter comments C2, C3 and C4.
Automatic clearing of comment lines after each measurement
In the menu you can define which comment lines are to be cleared
automatically after each measurement (see Section 6.20).
Displaying comment line C1 during measurement
By pressing the «Code» key you can display the current comment C1 for 3
seconds during the drying process.
Note:Comment lines C1 to C4 are method specific.A general additional
output text for all methods (e.g. company name and department) can be
defined in the menu (Section 6.16).
The comment lines C1 to C4 can optionally be entered using a Barcode
Reader (Section 9.3).
Practical application of your Moisture Analyzer
40
4.9Information on the measurement record
If your instrument has a built-in printer and this is activated in the menu, the intermediate values will be printed out
at the preselected intervals and the final result recorded on completion of the measurement.
The extent of detail of the record depend on the settings selected in the menu (see Section 6.17). The illustrations
below show examples of a standard length measurement record (factory setting) and an extended measurement
record.
Structure of the standard-length measurement record
The standard-length measurement record contains the following data (from top to bottom):
A -MOISTURE DETERMINATION-
B COFFEE LTD
C PRODUCTION XY
D METTLER TOLEDO
Halogen MoistureAnalyzer
E Type:HR83
F SNR:1234567890
G SW:2.01
H Method:
38COFFEE POWDER
I C1:QUALITY CHECK
C2:BEFORE PACKING
C3:CHARGE 4711
C4:HANS MUSTER
J Switchoff mode F
1 mg /00035 sec
K Standard drying
L Drying temp.105 °C
M Disp.mode0...-100 %MC
N Start weight2.507 g
O 001:00 min-2.51 %MC
002:00 min-3.39 %MC
003:00 min-3.71 %MC
P Total time3:19 min
Q Dry weight2.412 g
R End result-3.79 %MC
S ----30.06.03---15.14----
A Record title
B Company name text (see Section 6.16)
C Department name text (see Section 6.16)
D Manufacturer and Designation of the instrument
E Type of the instrument
F Serial number of the instrument
G Version number of the software
H Number and Name of the selected method (see Section 5)
I Comment lines (e.g. sample ID) if code 1, 2, 3 or 4 was entered (see
Section 4.8)
J Selected switch-off criterion (or set drying time)
K Selected drying program
L Drying temperature (set temperature)
M Selected display mode
N Sample weight at start of drying
O Measured value at each print interval (the number of recorded measured
values depends on the selected print interval and the measurement time)
P Total time of drying
Q Dry weight of the sample in grams
R Final result in the selected display mode
S Date and time at the end of the measurement
Notes
– You will find information on the dimensions of the results (%MC, % DC,
%AM and %AD) in Section 4.5.
– Depending on the settings in the Methods menu (Section 5.4), the data
printout can contain other information, which is described in the relevant
sections (Section 5.4.1 - 5.4.6).
Practical application of your Moisture Analyzer
41
Structure of the extended measurement record
In addition to the data provided in the standard-length measurement records (see Section 4.9), the extended
records contain the following information:
A Identification (consecutive number) of the balance adjustment, allows
assignment of the measurement record to the appropriate adjustment
record
B Identification (consecutive number) of the heating module adjustment,
allows assignment of the measurement record to the appropriate adjustment record
C Consecutive number of the measurement (corresponds to the sample size
or the number of measurement results in the statistics)
D Ramp time (time between the start of drying and attainment of the final
temperature)
E Moisture content
F Field for signature of the person who performed the Moisture determina-
tion.
Note: Depending on the settings in the Methods menu (Section 5.4), the data
printout can contain other information, which is described in the relevant
sections (Section 5.4.1 - 5.4.6).
Special events are recorded in the measurement record as
follows:
In the manual operating mode of the automatic sample cham-ber, the sample chamber was opened and closed during the
drying operation. When the automatic sample chamber is open,
drying is interrupted and then continued when the chamber is
closed.
Practical application of your Moisture Analyzer
42
Reset
Stop
007:30 min-24.73 %MC
008:00 min-24.55 %MC
>>>>>>>> ABORT <<<<<<<<<
----30.06.03---15:26----
Total time8:47 min
Dry weight1.916 g
End result-24.67 %MC
>>>>>>>> MANUAL <<<<<<<<
----30.06.03---15:27----
The «Reset» key has been pressed. The drying process is aborted
and the measurement result at the time of the abort is not
evaluated as it could be wrong.
The drying process has been stopped manually with the «Stop»
key. The elapsed drying time at the time of the stop and the
measurement result are recorded. Please note that with manual
stop the final result only shows the moisture content after the
manually determined total duration.
4.10Performing a measurement
You are now familiar with all the parameters of your Moisture Analyzer and have defined all values for your sample.
The instrument is now ready for the determination of your own samples. In this Section you will learn how to perform
measurements, print out the measurement results and stop the measurement process.
Switch the instrument on with «On/Off» key.
On/Off
The status display (User guide) symbolizes the initial status of the instrument
when the automatic sample chamber is closed.
Press the «Open/close auto sample chamber» key and …
2
… the automatic sample chamber opens.
Always ensure sufficient free space in front of the instrument since the
emerging sample chamber could knock over objects in its path or push them
off the bench.
Practical application of your Moisture Analyzer
43
The status display (User guide) now prompts you to load the empty sample
pan and tare the balance.
Position the empty sample pan in the sample pan handler. Place the sample
pan handler in the automatic sample chamber. Ensure that the tongue of the
sample pan handler lies exactly in the slot of the draft shield element. The
pan must lie flat in the pan holde.
Note: We advise you to work with the sample pan handler at all times. The
pan handler is ergonomic, safe and provides protection against burns due
to the hot sample pan.
#
Press the «Tare (zero)» key. This sets the integral balance to zero. For taring
the automatic sample chamber closes briefly then opens again.
Following taring, the status display prompts you to add the sample to the
sample pan.
Add the sample to the sample pan. If you are working with the weighing-in
aid, weigh the sample (minimum weight 0.1 g) using the bar display as
explained in the Section 4.7.
The status display now prompts you to start the drying process.
Start
Press the «Start» key and …
Practical application of your Moisture Analyzer
44
… the motorized, automatic sample chamber closes. The instrument starts
the drying and measurement.
You can follow the measurement process in the display: The status display
uses ascending bubbles to symbolize the drying process with the following
values being continuously updated and displayed:
– current temperature in the heating module
Note: You can use the «Drying temperature» key to display the preselected
drying temperature (current set temperature) for 2 seconds.
101
≠30
-|84
– elapsed time since the start of the measurement process
Note: If you have selected the switch-off criterion “Timed switch-off”, the
elapsed drying time will be displayed. You can display the preselected
drying time for 2 seconds with the «Drying time» key.
– current result in the preselected display mode
You can select a different display mode with the function key at any time
even during the measurement.
If your instrument has the built-in printer and this has been activated, the
measured values are printed out at the preselected intervals. You can use the
«Print» key to output the intermediate result manually on the printer at any
time. The units of the intermediate result are as defined in the preselected
display mode and can therefore be changed during the drying process by
pressing the «Display mode» function key. You will find an explanation of
the measurement record in the Section 4.9.
As soon as the preselected switch-off criterion is met (or the selected drying
time has elapsed), an audio signal sounds. The measurement process is
now ended and the sample chamber automatically opens.
Warning: Pan and sample may still be hot! Allow them to cool before you
remove the pan from the handler!
Stop
Practical application of your Moisture Analyzer
45
Note: You can also manually stop the measurement process prematurely
at any time by pressing the «Stop» key. With a measurement time of
minimum 30 seconds or more, the result will be printed out at the time of the
stop, transferred to the journal (see Section 5.6) and incorporated in the
statistics (see Section 5.7). If you abort a measurement process earlier or
by using the «Reset» key, the measurement result will not be displayed and
also not transferred to the journal or incorporated in the statistics.
‹10
ENfl -1Ç52
Reset
2
On/Off
59
You can now read off the measurement result in the display. You will find
information on the interpretation of the measurement results in Section 9.1.
The result and time display remain at their final values, whereas the
temperature continues to be updated.
Carefully remove the sample pan handler from the automatic sample
chamber.
Press the «Reset» key to delete the final result and the time display.
The display can also be cleared with the «Tare (zero)» key. This facilitates
efficient working if the next empty sample vessel has already been loaded.
If you do not wish to perform another measurement, switch the instrument
off with the «On/Off» key. After this operation, the sample chamber automatically closes.
Practical application of your Moisture Analyzer
46
4.11Test measurement
You can use this function only if your instrument is fitted with a printer!
Method:
34TOBACCO
C1:HAVANNA
C2:CHARGE 4711
C3:SYSTEM B
C4:H.MUSTER
Switchoff mode T
Standard drying
Drying temp.120 °C
Disp.mode0...-100 %MC
Start weight2.543 g
000:00 min-0.00 %MC
000:30 min-1.96 %MC
If you wish to work with the switch-off criterion “Weight loss per unit of
time” and are not familiar with the behavior of the sample, the test
measurement will help you select the suitable setting.
The diagram opposite illustrates the progress of a drying. The points at
which the individual switch-off criteria were reached (1–5) are marked.
MC (%)
-20
-15
-10
-5
0
0 1 2 3 4 5 6 7 8 9 10 11
t
2
1
Test measurement procedure
Press “t” («Switch-off criterion» function key) to perform a test measurement. You perform a test measurement just like any other measurement.
You will find further information on how to perform the measurement in
Section 4.10.
4
3
5
004:30 min-13.80 %MC
005:30 min-14.39 %MC
Switchoff mode 1
Time5:37 min
Result-15.14 %MC
-2.158 g
007:00 min-15.28 %MC
007:30 min-15.42 %MC
Switchoff mode 2
Time7:42 min
Result-15.81 %MC
-2.141 g
11:00 min-16.56 %MC
Switchoff mode 5
Time11:22 min
Result-16.91 %MC
-2.113 g
Total time12:33 min
Dry weight2.108 g
End result-17.11 %MC
Printing out the test measurement
During the test measurement a record is printed out which explains
exactly when and with what measurement result each switch-off criterion
was reached. The test procedure can be ended at any time with the «Stop»
key. Automatic switch-off follows after a measurement time of 8 hours.
Notes
– Test measurements are not entered in the journal or the statistics
(Sections 5.6 and 5.7).
– Depending on the settings in the Methods menu (Section 5.4), the data
printout can contain other information, which is described in the
relevant sections (Section 5.4.1 - 5.4.6).
Method:
34TOBACCO
C1:HAVANNA
C2:CHARGE 4711
C3:SYSTEM B
C4:H.MUSTER
Switchoff mode
Standard drying
Drying temp.120 °C
Disp.mode0...-100 %MC
Max. Time30:00 min
Target-16.80 %MC
Start weight2.543 g
000:00 min-0.00 %MC
000:30 min-1.96 %MC
007:30 min-15.28 %MC
007:30 min-15.42 %MC
Switchoff mode 2
Time7:42 min
Result-15.81 %MC
-2.141 g
008:00 min-16.12 %MC
AutoMet test measurement helps you to save time when developing
methods.It also enables you to obtain a result with your Moisture
Analyzer that corresponds most closely to the reference value for the
chosen temperature and sample quantity.
For AutoMet test measurement enter the reference value or reference
value by pressing the «Target» key.As soon as this value is reached, the
instrument automatically determines the optimum switch-off criterion for
the chosen conditions.This is a free switch-off criterion F (Section 5.5),
which after the measurement can be stored with the method by pressing
the «Accept entry» key.
Once you have defined the target value you can enter the maximum
length you require for the AutoMet test measurement.The range for the
switch-off criterion F that can be determined with AutoMet is restricted to
(1 mg/20 sec to 1 mg/180 sec).
AutoMet test measurement procedure
Press “t” («Switch-off criterion» function key) to perform the AutoMet test
measurement.Then press «Target» to define the reference value.
TRGÎ1]80Enter the reference value in %
You can enter the reference value in %MC, %DC, %AM or %AD
(depending on the setting).Confirm the reference value by pressing the
«Target» or «Accept entry» key.Then you can set the maximum
measuring time.
MA≈TIMË30Maximum time:10 to 480 min
Confirm the maximum measuring time by pressing the «Target» or
«Accept entry» key and the AutoMet test measurement will start.
Target
Switchoff mode 5
Time11:22 min
Result-16.91 %MC
2.113 g
Total time12:33 min
Dry weight2.108 g
End result-17.11 %MC
************************
AutoMet:
Target reached
Time8:22 min
Switchoff mode F
1 mg /00075 sec
************************
>>>>>>>> MANUAL <<<<<<<<
----30.06.03---15:50----
Printing out the AutoMet test measurement
The parameters for the active method are printed out in the record header.
The maximum time (e.g. 30 min) is also shown.The results for switch-
off criteria 1 to 5 and F (if defined in the Methods menu) are printed.The
value for F (1 mg / x sec) determined by AutoMet appears at the end of
the record. If the test measurement was stopped before the specified time,
the line MANUAL appears before the date.
Note: Depending on the settings in the Methods menu (Section 5.4), the
data printout can contain other information, which is described in the
relevant sections (Section 5.4.1 - 5.4.6).
Practical application of your Moisture Analyzer
48
There are three different scenarios for the AutoMet results:
Scenario 1: Switch-off criterion F within the range 1 mg / 20 – 180
seconds
As soon as the target value is reached, a “target” appears and is displayed
Target
until the test measurement is completed.
The AutoMet test measurement only suggests a switch-off criterion F if an
************************
AutoMet:
Target reached
Time8:22 min
Switchoff mode F
1 mg /00075 sec
************************
accurate moisture determination can be expected as a result of the drying
process.
AutoMet is therefore restricted to the range 1 mg / 20 seconds to 1 mg / 180
seconds.AutoMet simplifies method determination in particular for samp-
les whose drying curve tapers off slowly. This behavior is typical of organic
samples and is illustrated by the fine flour example shown in Figure A (at
the end of this section).
For samples whose drying curve changes very little once switch-off criterion
2 has been reached (1 mg / 20 seconds), we recommend using switch-off
************************
AutoMet:
Target reached
Time6:17 min
Switchoff mode F
1 mg /00>20 sec
Recommendation:
Switchoff mode 3
1 mg /00050 sec
************************
criterion 3. This scenario is illustrated in Figure B (at the end of this section).
The printout shows the measuring time until the target value is reached and
“1 mg/ > 20 sec” for switch-off criterion F. You can obtain the results for the
fixed switch-off criteria 1 to 5 from the printed test record.
The range for the switch-off criterion F that can be determined with
Since drying profiles can vary slightly even when the same method
parameters are used (e.g. depending on the sample distribution), we
recommend repeating the AutoMet test measurement (e.g. 3 sample
throughputs) when developing a method.The mean value can then be
stored as the method-specific switch-off criterion F (Section 5.4.1).This
procedure allows the result from the Moisture Analyzer to be matched to the
reference value as closely as possible.
Note: When performing the AutoMet test measurement, depending on the
curve profile, there may be slight differences between the actual switch-off
criterion F and the preset switch-off criterion.We therefore recommend
carrying out a follow-up measurement to confirm the suitability of the
switch-off criterion F determined with AutoMet.
************************
AutoMet:
Target reached
Time3:11 min
Switchoff mode F
1 mg /00<20 sec
************************
************************
AutoMet:
Target reached
Time17:33 min
Switchoff mode F
1 mg /0>180 sec
************************
Blank target
Practical application of your Moisture Analyzer
49
Scenario 2: Switch-off criterion F outside the range 1 mg / 20 – 180
seconds
If the target value is reached but a switch-off criterion cannot be determined
within the permitted range, a “blank target” symbol is displayed.
The target value was reached, but the method parameters were not chosen
correctly.The reproducibility of a moisture determination with switch-off
criteria of “1 mg / < 20 seconds” is low.If this message appears and the
actual moisture content was entered as the target value, the solution is
normally to lower the drying temperature.
The target value was reached, but the switch-off criterion that was determined is 1 mg / > 180 seconds and is therefore outside the permitted range.
If this message appears, in most cases the target value is almost reached
with switch-off criterion 5 (1 mg / 140 seconds).If the actual moisture
content was entered as the target value, raising the drying temperature may
reduce the measuring time in this case.
************************
AutoMet:
Target not reached
Please change parameters
************************
ƒSAVËYES
ƒSAVËNO
Scenario 3: Target value not reached
If the target value was not reached in the predefined time, no target is
displayed.If the actual moisture content was entered as the target value,
increasing the measuring time (“MAX.TIME”) or changing the method
parameters (e.g. raising the drying temperature or using a different drying
program) may help.
Saving switch-off criterion F
On completion of the AutoMet test measurement, the switch-off criterion
F that was obtained can be saved as a method parameter (Methods menu,
see Section 5.4),although it has to be confirmed.If a suitable switch-off
criterion was determined with AutoMet, at the end of the test measurement
the line “F.SAVE: YES” is displayed.Press the «Accept entry» key to save the
switch-off criterion F.
If you choose “NO” or press the «Reset» key, you can quit the AutoMet test
function without saving the switch-off criterion F.
Practical application of your Moisture Analyzer
50
5
4
3
2
1
F
Figure A
Drying profile for a test measurement with 2.5 g fine flour (standard drying, 100 °C).The fixed switch-off criteria
1 to 5 are shown, along with the switch-off criterion F (1 mg / 35 seconds) determined with AutoMet, with which
the exact target value (11.48 %) is reached.
4
3
1
2
5
Figure B
Drying profile for a test measurement with 2.5 g 90% (w/w) potassium chloride solution (standard drying,
160 °C).
The fixed switch-off criteria 1 to 5 are shown. Once switch-off criterion 2 is reached, the result changes
only slightly.In order to reach the target value (90%) AutoMet proposes switch-off criterion 3.
AutoMet test measurement is a quick way of determining your method parameters.However,
you must always personally check the suitability of switch-off criteria and other method
parameters for the intended purpose (e.g. by analyzing samples with different moisture
contents).
Note: Test measurements are not entered in the journal or the statistics (Sections 5.11 and
5.12).
Methods facilitate your work
51
5Methods facilitate your work
Methods simplify and accelerate your daily work. In this Section you will learn how you compile, print out and change
methods and how to use them. You will also become acquainted with the journal function and the statistics. You
will find an overview diagram of all menu options in Section 5.4.
5.1What is a method?
Perhaps you remember: In Section 4 you became acquainted with all the parameters and setting possibilities you
can use to match your Moisture Analyzer to a particular measurement task. If you have to change samples frequently
during your work, the methods save you having to reenter the parameters and preclude wrong settings. Methods
thus make it easier for you to switch between different types of samples.
A method contains all settings for a particular sample. It can be recalled at
Methods
a keystroke and your instrument immediately operates with the corresponding settings. You have 40 methods freely available.
Methods
Margarine 1
Drying prog.:Standard drying
Drying temp.:70 °C
Switchoff criterion:
Drying time:--------
Display mode result: MC (moisture content)
Print interval:30 sec.
Target weight:2.500 g
Loss < 1 mg/50 Sec.
The following parameters, which you have already learned in section 4, form
part of a method:
Each method has its own measured value journal and its own measured
4
value statistics.
You will find information on the journal and statistics in Sections 5.6 and
5.7.
Work with methods is divided into two stages: in the setting stage you define
the parameters of the method and name the method. In the utilization stage
you perform routine measurements with different samples. When switching
from one sample to the next, you do not have to change any settings, you
need simply select the method suitable for the sample. You will find
information on how to use and set up of methods in the following Sections.
Methods facilitate your work
52
5.2Recalling a method
To activate all settings for a particular sample, simply call up the associated method. After recall of the method, all
appropriate settings become immediately active.
To recall an existing method (consult the next section for information on how
to compile methods), proceed as follows:
Press the «Method» key. The fixed method numbers (01-40) flash on the
Method
01
MARGARINE
02
SORP
screen and the name of the active method is displayed.
Now use the «Scroll up» and «Scroll down» keys to select the method. Then
press the «Accept entry» key. You can also enter the method number directly
via the numeric keypad.
All settings of the selected method are now active and you can immediately
perform your measurement.
During the measurement the number of the active method (e.g. method
number 02) is displayed. The method number and method name are printed
on the measurement record.
%
If you press the «Method» key during the measurement, the name of the
active method is displayed. If the name contains more than 10 characters,
you can view the rest of the name by pressing the «Method» key again.
Method
105
Ò10
02-|48
105
Ò10
SORP
5.3Naming, compiling or changing methods
When you put your Moisture Analyzer into operation for the first time, all 40 methods are identical and are allotted
the factory setting for all parameters. You can change the parameters of a method at any time (except when a
measurement is running or the settings are protected against change as described in Section 6.13).
The compilation of a new or the modification of an existing method comprises 3 steps:
– Selecting a method (01 - 40)
– Naming a method (optional)
– Selection of the settings
Method
A...Z
MARGARINE
GLUE TYPE B
A
01
05
Methods facilitate your work
53
Selecting a method
Press the «Method» key.
Now use the «Scroll up» and «Scroll down» keys to select the method you
wish to name, recompile or change. The illustrations show the method for
naming and compiling a new method which should be assigned the name
“Glue type B”.
Alternatively you can use the keys on the numeric keypad to select a method.
Naming a method
To name a method press the «A…Z» key. A flashing “A” appears at the first
input position on the extreme left and signals that the instrument is ready for
your entry. The procedure for changing an existing method name, without
needing to enter the new name completely, is described later.
A...Z
You can now use the «Scroll up» and «Scroll down» keys to select the first
G
letter of the designation (uppercase letters only) or a special character (+,
–, etc.).
Confirm your entry with the «A…Z» key. The flashing “A” reappears, but at
GA
the second position. Enter the additional characters of the desired method
name as described above.
Notes
–A space is symbolized by the underline character “_”.
– You can remove wrong characters or numbers with the «Delete» key and
then correct your entry.
–
Numbers can be entered directly via the numeric keypad.If you enter a
character via the numeric keypad, the instrument remains in numeric
input mode and a flashing dash appears at the next input position instead
of a flashing “A”. To switch back to the text mode, press the «A…Z» key
again.
If you want to change an existing method name (e.g. entering a serial
number), proceed as follows:
Press the «Method» key. The designation of the currently active method
Method
flashes in the display.
Methods facilitate your work
54
A...Z
GLUE TYPE B
Press the «Delete» key (to activate edit mode).A flashing underline
character is displayed after the last number or the last character flashes,
indicating that the system is ready for input.
You can then add additional characters by pressing the «A…Z» key or a
number key on the numeric keypad.
By pressing the «Delete» key a second time you can delete the last character.
The second last character flashes and can be changed using the numeric
keypad or the «Scroll up» and «Scroll down» keys, or confirmed with the
«A…Z» key. Further characters can then be added as described above.
As soon as you have finished entering the designation, press the «Accept
entry» key to quit the input mode (the flashing character is also accepted).
Selecting the settings
The selected method is now ready for the entry or change of the parameters.
Now use the appropriate keys to match the following parameters to your
measurement task:
– Drying program
– Drying temperature
– Switch-off criterion
– Drying time (dependent on switch-off criterion)
%
Method
Please consult Section 4 for the meaning of the individual parameters and details on how to proceed.
All settings are assigned to the selected method and automatically stored.
After you have completed all settings, your instrument is ready for measurement with the new or changed method.
Notes
– The methods can be protected in the menu against changes (see Section 6.13).
– We advise you to define a special method for test purposes, e.g. with the name “Test”. You can always change
the settings of this method to test the individual parameters without altering the methods you have already defined.
Menu
– Display mode
– Print interval
– Target weight for weighing-in aid
– Settings in the Methods menu (Section 5.4)
Methods facilitate your work
55
5.4Method menu
If you press the «Method» key followed by the «Menu» key, the Method menu opens (you will find additional
information on using the menu in Section 6).
You can define the following settings here: Free switch-off criterion, Free factor -MC, Free factor F.F.G, Method
compilation, High resolution (with standby temperature) and drying temperature up to 200 °C without time
limitation.
Method
Free switch-off criterion
(= switch-off criterion F)
Free factor -MC
(correction factor)
Free factor g
(calculated result)
Method compilation
Menu
ƒSO
ƒƒ-M
ƒƒ‚
AD∂˚
NONO
NO
NONO
YES
NONO
NO
NONO
YES
NONO
NO
NONO
YES
NO NO
NO
NO NO
YES
'W†M‚'W†M‚
'W†M‚1
'W†M‚'W†M‚
'SE100
±000
(0.500-2.000)
(1-100)
10
0101
01
0101
POSPOS
POS
POSPOS
NEG
02
03
…
40
(1-10 mg)
(5-180 sec)
High resolution
(0.0001 g)
Drying temperature up to
200 °C without time
limitation
RES
†EXÎ
STDSTD
STD
STDSTD
HIGH
NONO
NO
NONO
YES
ST∂BŸ
N0N0
N0
N0N0
YES
ßTEM∏
ßTIMË
ßSWO‡
4040
40
4040
(40 – 100 °C)
PERPER
PER
PERPER
VAR
NONO
NO
NONO
YES
3030
30
3030
(30 – 480 min.)
1Ò001Ò00
1Ò00
1Ò001Ò00
(0:00 – 23:59)
Methods facilitate your work
56
5.4.1Freely selectable switch-off criterion
The free switch-off criterion (= switch-off criterion F) is based on the principle of the mean weight loss per unit of
time.As soon as this drops below the preset value, the measurement is automatically ended.Alternatively the free
switch-off criterion can be determined automatically with the AutoMet function and stored with the method, see
Section 4.12 AutoMet test measurement.
Press the «Method» key followed by the «Menu» key to open the Method
Method
ƒSONO
ƒSOYES
' W†M‚1
Menu
menu.
The free switch-off criterion is deactivated (“NO”) as the factory setting.
If you wish to set a free switch-off criterion, use the «Scroll up» and «Scroll
down» keys to select “YES” and then press «Accept entry» to confirm.
A flashing 1 will appear in the display. Then use the «Scroll up» and «Scroll
down» keys or the numeric keypad to enter the weight loss (1 mg to 10 mg)
in 1 mg increments. Press the «Accept entry» key to confirm your input.
' SE100
Now enter the time in the flashing field.You can alter it in 5-second
increments using the «Scroll up» and «Scroll down» keys or enter any value
between 5 and 180 via the numeric keypad.Press the «Accept entry» key
to confirm your input.
Note: To perform a measurement with a free switch-off criterion, it must be
defined in the Method menu.“F” must also be selected as the switch-off
criterion (see Section 4.4).
Methods facilitate your work
57
5.4.2Free factor (-MC)
The free factor F.F.-MC (0.500 to 2.000) for the moisture content can be defined in the Method menu. It is used
to calculate a corrected final result (e.g. in order to compensate for systematic deviations from the reference result).
Press the «Method» key followed by the «Menu» key to open the Method
Method
ƒƒ-MNO
ƒƒ-MYES
ƒƒ-M±000
Menu
menu.
The free factor (-MC) is deactivated (“NO”) as the factory setting.
If you wish to set the free factor F.F.-MC, use the «Scroll up» and «Scroll
down» keys to select “YES” and then press «Accept entry» to confirm.
Now enter the factor in the flashing field.You can alter it using the «Scroll
up» and «Scroll down» keys or enter any value between 0.500 and 2.000
via the numeric keypad.Press the «Accept entry» key to confirm your input.
ƒƒ-MPOS
End result-22.90 %MC
************************
Factor (-%MC): 1.100
Calculated end result:
-25.19 %MC
************************
End result77.10 %DC
************************
Factor (-%MC): 1.100
Calculated end result:
74.81 %DC
************************
You can specify whether the factor has a positive (“POS”) or negative
(“NEG”) sign.
The final result is multiplied by the method-specific factor (0.500 to 2.000)
in %MC display mode and highlighted on the printout as the calculated final
result.
In %DC display mode, if you define a free factor (-MC), the final result %DC
is converted by means of the formula below and highlighted on the printout.
%DC = 100 + (factor x %MC)
Note:
In %AM, %AD or g display mode, the final result is not converted.
Methods facilitate your work
58
5.4.3Free factor (g)
The free factor F.F.G (1 to 100) can be defined regardless of the display mode setting - %MC, %DC, %AM, %AD
or g.
Press the «Method» key followed by the «Menu» key to open the Method
Method
ƒƒ‚NO
ƒƒ‚YES
ƒƒ‚10
Menu
menu.
The free factor F.F.G is deactivated (“NO”) as the factory setting.
If you wish to set the free factor F.F.G, use the «Scroll up» and «Scroll down»
keys to select “YES” and then press «Accept entry» to confirm.
Now enter the factor in the flashing field.You can alter it using the «Scroll
up» and «Scroll down» keys or enter any value between 1 and 100 via the
numeric keypad.Press the «Accept entry» key to confirm your input.
In %MC, %DC, %AM or %AD display mode, defining a F.F.G does not
Dry weight0.038 g
************************
Factor (g):20
Calculated end result:
0.760
************************
convert the final result.
5.4.4Compiling methods
You can compile two methods, which means that the second measurement is performed automatically after the
first one.The tare value from the first measurement is automatically used for the second.The dry weight from the
first measurement serves as the starting weight for the second measurement.
Press the «Method» key followed by the «Menu» key to open the Method
Method
AD∂˚NO
AD∂˚YES
Menu
menu.
Method compilation is deactivated (“NO”) as the factory setting.
If you wish to enable method compilation ADD.M:, use the «Scroll up» and
«Scroll down» keys to select “YES” and then press «Accept entry» to confirm.
AD∂˚04
Now enter the method you wish to add in the flashing field (enter the number
via the numeric keypad).Press the «Accept entry» key to confirm your input.
Method:
04RESIN-50
C1:CRUDE PRODUCT
C2:CHARGE 125
C3:SYSTEM B
C4:H.MUSTER
Switchoff mode 3
Standard drying
Drying temp.50 °C
Disp.mode0...-100 %MC
Start weight2.543 g
Total time2:44 min
Dry weight1.961 g
End result-22.90 %MC
----30.06.03---15:28----
Method (AddMet):
05RESIN-120
Switchoff mode F
1 mg /80 sec
Gentle drying
Drying temp.120 °C
Disp.mode0...-100 %MC
Start weight1.961 g
In the adjacent record you can see for example that method 04 (resin-50)
has been added to method 05 (resin-120).At the end of the measurement
the first method (resin-50) is activated again.
Note: Multiple compilations are not executed and a method cannot be
compiled with itself.
Total time5:33 min
Dry weight0.821 g
End result-41.87 %MC
----30.06.03---15:33----
Methods facilitate your work
60
5.4.5High resolution
Traditionally, moisture analyzers are operated at a maximum resolution of 0.01%. The following setting allows you
to determine moisture content using the resolution of the built-in analytical balance, which is ten times higher
(0.1 mg weight / 0.001% moisture). The high resolution is especially suitable for samples with low moisture content
that contain no other high-volatile substances besides water. Careful sample preparation is extremely important for
these kinds of samples! Note: High resolution is not available for the display modes “ATRO moisture content” and
“ATRO dry content” (Section 4.5).
Other optional settings become available after the high resolution is activated: You can activate the standby
temperature and define its parameters (temperature, duration and switch-off time).
Press the «Method» key followed by the «Menu» key to open the Methods
Method
RE˘STD
RE˘HIGH
Menu
menu. You can access the “RES: STD” menu item by repeatedly pressing the
“Menu” button.
The standard resolution is activated at the factory (“STD”).
If you wish to activate the high resolution, use the «Scroll up» and «Scroll
down» keys to select “HIGH” and then press «Accept entry» to confirm.
The initial weight and the final weight, as well as the result, are printed with
the high resolution. The final result with the high resolution is shown in the
display (the standard resolution is used while measuring).
5.4.5.1 Standby temperature
If necessary, you can activate the standby temperature to preheat the instrument after activating the high resolution.
Because of the short heat-up times with halogen moisture analyzers, no preheating is usually required for standard
applications. Under unusually heavy use, a preheated instrument can sometimes improve the reproducibility of the
results, since the instrument is in the same climatic state with every measurement. Note: For the highest accuracy,
a preheating period of 15 minutes is recommended for the first measurement (a shorter preheating period is shown
in the time window of the display and recorded on the printout, e.g. “Standby period 9 min.”).
The standby temperature is deactivated (“NO”) as the factory setting.
ST∂BŸNO
If you wish to activate the standby temperature, use the «Scroll up» and
ST∂BŸYES
ßTEM∏40
«Scroll down» keys to select “YES” and then press «Accept entry» to confirm.
Now enter the standby temperature in the flashing field. You can use the
«Scroll up» and «Scroll down» keys or the numeric keypad to select a
temperature within the range of 40 °C (factory setting) to 100 °C. Press the
«Accept entry» key to confirm your entry.
ßTIMËPER
ßTIMËVAR
ßTIMË30
Methods facilitate your work
61
Note: High standby temperatures can result in possible moisture loss while
weighing. You cannot select a standby temperature that is higher than the
final drying temperature. Otherwise, the message “Error 10” appears
(Section 8.2).
Setting the standby time allows you to define whether the instrument is
maintained at standby temperature continually or only for a limited amount
of time after measurement.
In the factory, the standby time is set so that the instrument is
permanently maintained at the standby temperature (“PER”).
If you wish to limit the standby time, use the «Scroll up» and «Scroll down»
keys to select “VAR” and then press «Accept entry» to confirm.
Now enter the desired standby time in the flashing field. You can use the
«Scroll up» and «Scroll down» keys or the numeric keypad to select the
standby time within the range of 30 minutes (factory setting) to 480 minutes.
Press the «Accept entry» key to confirm your entry.
ßSWO‡NO
ßSWO‡YES
ßSWO‡1ˇ00
You can now select whether you want to define a standby switch-off time.
This defines what time standby operation is cancelled. When this time is
reached, the instrument is no longer maintained at the standby temperature.
This prevents the instrument from continuing to heat after the working day
is over, for example.
In the factory, the standby switch-off time is deactivated (“NO”), i.e. the
instrument is permanently maintained at the standby temperature.
If you wish to activate the standby switch-off time, use the «Scroll up» and
«Scroll down» keys to select “YES” and then press «Accept entry» to confirm.
Now enter the standby switch-off time in the flashing field. You can use the
«Scroll up» and «Scroll down» keys or the numeric keypad to select a time
within the range of 0:00 - 23:59 (factory setting is 18:00). Press the «Accept
entry» key to confirm your entry.
Note
– The switch-off time is based on the moisture analyzer’s internal clock, and
we recommend that you check its setting.
- If you define both a standby time and a standby switch-off time, the
standby operation is switched off based on whichever condition is fulfilled
first.
Methods facilitate your work
62
5.4.5.2 Recording the settings
- You can cancel standby operation at any time by switching off the instrument. The next time the instrument is switched on, it continues working with the standby settings of the most recently active method.
– The status display opposite shows that the device is heating with the
standby temperature.
Method:
01
Standbytemp. 50 °C
Time 60 min
Switch-off 18:00
Disp. mode %MC
>>Standby period 11 min
Start weight 4.9989 g
0:30 min -0.44 %MC
1:00 min -0.45 %MC
Total time 1:35 min
Dry weight 4.9743 g
Moisture 0.0246 g
Method:
01
Standbytemp. 50 °C
Time 60 min
Switch-off 18:00
Switchoff mode 1
Standard drying
Drying temp. 105 °C
Disp.mode g
Print interval 0:30 min
The settings selected in the methods menu for
“High resolution” are documented in the printouts. The figure at left shows a section from the
data printout of a moisture measurement; the
figure at right shows a printout of the methods
setting (see also Section 5.5). The specific
information for “High resolution” is shown in
bold.
Methods facilitate your work
63
5.4.6Drying temperature to 200 °C without time limitation
In the standard setting of your moisture analyzer, drying temperatures of 160 °C to 200 °C are possible for only
a limited period of time. After the temperature reaches this range, the instrument lowers the temperature back to
160 °C (this function is described in Section 4.3). For the sake of backward compatibility with previously developed
methods, this time limitation continues to be the standard setting of the instrument.
However, if needed, the temperature extension setting is available to you in the Methods menu, and allows drying
temperatures of up to 200 °C without time limitation.
Press the «Method» key followed by the «Menu» key to open the Methods
Method
†EXÎNO
†EXÎYES
Menu
menu.
Temperature extension is deactivated (“NO”) as the factory setting.
If you wish to activate the temperature extension, use the «Scroll up» and
«Scroll down» keys to select “YES” and then press «Accept entry» to confirm.
Note
– When working with temperatures over 180 °C, we recommend that you
always wait 2 to 3 minutes between individual measurements, in order
to ensure good reproducibility of the readings.
–A measurement cannot be started until the drying temperature is higher
than the temperature in the measuring room. Otherwise, the message
“Error 10” appears (Section 8.2).
Switchoff mode 1
Standard drying
Drying temp. 185 °C
No temp. lowering
Disp.mode g
Print interval 0:30 min
– In the data printouts the temperature extension is documented as shown
at left.
5.5Printing out methods
To ensure you do not lose your overview of the 40 methods, you can:
– print out a list of all methods or
– record the parameters of a particular method.
Note: Please note that these print functions are available only in the standby mode of the instrument and not during
Method:
34SODIUM TARTRATE
C1:CHECK
Switchoff mode 3
Gentle drying
Drying temp.150 °C
Disp.mode0...-100 %MC
Print interval 1:00 min
Target weight 2.500 g
00Active+/-10 %
----30.06.03---15:50----
Confirm your selection with the «Accept entry» key.
Press the «Print» key. A list of all parameters for the selected method is
printed. Please consult Section 4 for the meaning of the individual parameters and settings. Depending on the settings, your record may look slightly
different.
Methods facilitate your work
5.6Measured value journal
The journal function always provides you with an overview of the last 20 final results of each method.
Select the method whose measured value journal you wish to view.
MARGARINE
42 -2#59
Method
Journal
1∞40
Confirm your selection with the «Accept entry» key.
Press the «Journal» key and …
… the display shows the data of the measurement last performed. The
following data are shown for every measurement:
– the time of the measurement
– the consecutive number of the measurement (serial number)
– the final result (the asterisk symbol in the top left corner of the display
signals that the result has been calculated)
– the selected display unit, which can be changed. If “grams” have been
selected as the display unit, the dry weight is displayed.
You can use the «Scroll up» and «Scroll down» keys to view the data of the
other recorded measurements in turn.
Note: With the «Reset» key you can terminate the display of the measured
value journal at any time (the measured values will not be deleted).
If your instrument is equipped with a built-in printer, you can print out the
entire journal on the printer with the «Print» key. In addition to the data in the
display, the date of every measurement will be printed out.
Methods facilitate your work
66
CLEAÈNO
CLEAÈYES
After pressing the «Accept entry» key you are asked if you wish to clear the
measured value journal.
The suggested answer is “NO”. If you do not wish to clear the measured value
journal, press the «Accept entry» key now to confirm your selection and quit
the journal display. The journal will continue to be kept with the measured
values already recorded..
If you wish to clear the journal, use the «Scroll up» or «Scroll down» key to
select the response “YES”.
Then press the «Accept entry» key to confirm your selection. The existing
journal will be cleared and a new journal created.
Note: If you do not want measured data to be recorded in the instrument,
you can switch off the journal function in the menu (see Section 6.19).
5.7Measured value statistics
Your Moisture Analyzer keeps measured value statistics for each of the 40 methods. The measurement results of
each method are continuously recorded and transferred to the statistics. The statistics are updated until they are
cleared. After clearance, a new set of measured value statistics is created.
Select the method whose measured value statistics you wish to view.
Method
MARGARINE
Confirm your selection with the «Accept entry» key.
Press the «Stat.» key and ...
Stat.
* Stat. 1
… the display shows the first value of the statistics (the number of
measurement results).
˙124
You can use the «Scroll up» and «Scroll down» keys to view the additional
data of the measured value statistics in turn.
* Stat.1
˙124
* Stat.2
¤-2[03
%
Methods facilitate your work
67
The individual values have the following meaning:
– Number of measurement results recorded in the statistics (maximum
9999). The run number on the full records (see Section 6.17) corresponds to the number of measurement results (sample size).
– Mean value of all measurements in the selected display unit:
– Standard deviation in the selected display unit:
s =
1
n− 1
xi− x
()
∑
2
– Lowest recorded measurement result.
– Highest recorded measurement result.
– Absolute difference between highest and lowest result.
Note: You can press the «Reset» key to terminate the display of the measured
value statistics at any time (the measured values will not be deleted).
If your instrument is equipped with a built-in printer, you can print out all
statistical data of the current method on the printer with the «Print» key.
Methods facilitate your work
68
CLEAÈNO
CLEAÈ VALUE
CLEAÈYES
After pressing the «Accept entry» key, you are asked if you wish to clear the
measured value statistics.
The «Scroll up» or «Scroll down» key offers the following possibilities:
– “NO”: The measured value statistics will not be cleared, but continued with
the measured values already recorded.
– “Value”: The measured value last recorded in the statistics will be deleted
and the statistics will be continued with the remaining values. This allows
you to delete wrong measurements. The deletion must be performed
before the next value is recorded as only the last value is deleted!
– “YES”: All the existing measured value statistics will be completely cleared
and a new set of statistics created.
Select the desired option and then press the «Accept entry» key to confirm
your selection and quit the statistics display.
Note: If you do not want measured data to be recorded in the instrument,
you can switch off the statistics function in the menu (see Section 6.19).
The menu (basic settings of the instrument)
69
6The menu (basic settings of the instrument)
In this Section you will learn how to use the menu to define the settings your instrument uses for operation. You will
also find out how to adjust the integral balance and the heating module.The menu is a list of selection possibilities
which you can use to preselect the basic setting of your instrument.The menu comprises various options (you
will find an overview diagram of all menu options in Section 6.24).In addition to the settings (which are numbered
for clarity), certain functions (e.g. balance adjustment and heating module adjustment) can also be performed.
All menu settings are retained even in the event of a power failure.
Using the menu is simple:
Menu
P∑SEÎNO
Menu
Press the «Menu» key to call up the menu. The first menu option is password
definition (Section 6.1). All other menu items can be accessed by pressing
the «Menu» or «Accept entry» key repeatedly.
CENTRAL LABORATORY
10 RecordNormal
11 Free switchoffOff
12 Stat./JournalOn
13 Clear Code1-4
14 Test weight 50.000 g
15 Test Tolerance0.002 g
----30.06.03---15:56----
Reset
PHARMA LTD
You can use the «Scroll up» and «Scroll down» keys to switch between the
selection possibilities within the current menu items and hence select the
desired setting or execute the particular function.
Save changes
Once you have defined your setting in a menu option, press the «Accept
entry» key. The setting will be saved and the next menu option opened.
Do not save changes
If you close a menu option by pressing the «Reset» key, the changes made
in the current menu option will not be saved.
Print menu settings
If your instrument is equipped with a built-in printer, you can record all menu
settings. To do this simply press the «Menu» key followed by the «Print» key.
The following Sections will acquaint you with the individual menu items. The
order corresponds exactly to that in the menu.
The menu (basic settings of the instrument)
70
6.1Password protection
In this menu option you can protect your instrument against accidental changes to the menu and method settings
and method changes with a 3-level password (for details of how to protect the method settings without a password,
see Section 6.13).
Access authorizationPW-1
with password
PW-1, PW-2, PW-3PW-2
PW-3
Protected areaMethodsMenuMethods
settingssettingsselection
and
Journal/Statistic
Note: Password protection is limited to keyboard inputs (there is no protection via the LocalCAN universal
interface).
Set password
Password protection
P∑SEÎ
NONO
NO
NONO
YES
PW-1
PW-2
PW-3
OFFOFF
OFF
OFFOFF
ON
OFFOFF
OFF
OFFOFF
ON
OFFOFF
OFF
OFFOFF
ON
ABORT?ABORT?
ABORT?
ABORT?ABORT?
CLEAR?
ABORT?ABORT?
ABORT?
ABORT?ABORT?
CLEAR?
ABORT?ABORT?
ABORT?
ABORT?ABORT?
CLEAR?
Password PW-1
PW-1 protects all method settings, including the Method menu (see Section
%
5.4).In addition, the method memory can only be cleared and the factory
settings restored if PW-1 is not set.Also, if PW-1 is set (PW-1: ON), menu
option 6 (protecting method settings without a password) is not available (Error
12, see Section 8.2).Method settings can only be defined if PW-1 is set
to OFF. PW-1 can be reset by entering PW-1 (PW-1: OFF).
Note: The administrator (PW-1) requires only one password and can also
Method
➞
Menu
Method menu
(see Section 5.4)
use his/her password (level 1) to reset and redefine PW-2 and PW-3.
Menu
Journal
Method
Functions / settings
(apart from password)
Stat.
Selection
The menu (basic settings of the instrument)
71
Password PW-2
PW-2 protects all non-method-specific menu functions and menu settings,
along with the statistics and journal function, against unauthorized access.
You are prompts for PW-2 (E.PW2:) on opening the menu and also if you
press the Stat. or Journal keys. PW-2 can be reset by correctly entering
PW-1 or PW-2.
Note: To input for PW-2, both PW-1 and PW-2 are accepted.
Password PW-3
PW-3 ensures that only authorized persons can select methods. You are
prompts to enter PW-3 (E.PW3:) when you press the «Method» key.
PW-3 can be reset by correctly entering PW-1, PW-2 or PW-3.
Note: To input for for PW-3, PW-1, PW-2 and PW-3 are all accepted.
Menu
P∑SEÎNO
Menu
P∑SEÎYES
Menu
PW-∞OFF
Menu
PW-∞ON
Menu
PW-∞ ******
Menu
CH-∞
Set password (Example: PW-1)
The factory setting for this menu option is “NO” (no password protection).
To work with a password (PW), press the «Menu» key and use the «Scroll
up» and «Scroll down» keys to select “YES”.
Press the «Accept entry» key to
confirm your input.
In PW-1 select “ON”, press «Accept entry» to confirm this and enter a
numerical password (max. 6 digits, displayed as asterisks).
Press the
«Accept entry» key to confirm your input.
Once you have set the password you must enter it again (CH-1).Press the
«Accept entry» key to confirm your input.
The menu (basic settings of the instrument)
72
Deactivate password (Example: PW-1)
Menu
P∑SEÎNO
To deactivate or change the password (PW) as shown in this example, press
the «Menu» key and use the «Scroll up» and «Scroll down» keys to select
“YES”. Press the «Accept entry» key to confirm your input.
Menu
P∑SEÎYES
Menu
PW-∞ON
Menu
PW-∞OFF
Menu
PW-∞ ******
Menu
ABORT?
Menu
CLEAR?
Use the «Scroll up» and «Scroll down» keys to select “OFF”.Press the
«Accept entry» key to confirm your input.
Enter the password and press
«Accept entry».
When you press the «Accept entry» key, the prompt “ABORT?” or “CLEAR?”
appears. Choose “CLEAR” and press «Accept entry» to confirm. Press the
«Reset» key to leave the menu.
The menu (basic settings of the instrument)
73
6.2Adjusting the balance
In this menu item you can adjust the balance of your instrument. Consult Section 3.2 to learn when an adjustment
is necessary. Before selecting the adjustment function, ensure that the sample pan is in position. Following a drying,
you should wait at least 30 minutes before undertaking an adjustment.
Menu
∑CAˆNO
Menu
∑CAˆYES
52
∑CAˆ5≠000
The factory setting in this menu item is “NO” (no balance adjustment).
Note: The number above the word “NO” shows the number of adjustments
performed to date.
If you wish to adjust the integral balance, select “YES” (using the «Scroll up»
or «Scroll down» key).
Press the «Accept entry» key to start the adjustment process. The sample
chamber automatically opens.
The flashing display then prompts you to enter the adjustment weight (50 g).
Note: Ensure that your instrument is in a suitable location (see Section 2.2).
g
Place the requested adjustment weight in the middle of the sample pan. The
weight is automatically recorded.
Note: The 50 g adjustment weight is available as optional equipment (see
Section 9.9).
52
∑CAˆ≠000
The instrument stores the determined weight value and the display prompts
you to lift off the adjustment weight.
g
Remove the adjustment weight.
This concludes the adjustment of the balance and the instrument automatically quits the menu and returns to the standby mode. The Moisture Analyzer
is now ready for further measurements.
The menu (basic settings of the instrument)
74
A -- BALANCE ADJUSTMENT --
B 30.06.200314:48
C METTLER TOLEDO
Halogen MoistureAnalyzer
D Type:HR83
E SNR:1234567890
F SW:2.01
G Balance Adjustment159
H Weight ID:
........................
I Weight:50.000 g
J Adjustment performed
Signature:
........................
---------- END ---------
If your instrument is equipped with the built-in printer and this is activated
(see Section 6.15), on completion of the balance adjustment an adjustment
record will be automatically printed out with the following data:
A Record title
B Date and Time of the record printout of the balance adjustment
C Manufacturer and Designation of the instrument
D Type of the instrument
E Serial number of the instrument
F Version number of the software
G Identification (consecutive number) of the balance adjustment, allows
assignment of the measurement record to the appropriate adjustment
record
H Field for entry of the number of the adjustment weight used
I Adjustment weight used
J Field for signature of the person who performed the balance adjustment
Please enter the number of the adjustment weight used, sign the record and
store it in a safe place. This assures traceability, one of the basic requirements of every quality assurance system.
In the “W.CAL” menu you can print out the latest adjustment record at any
time by pressing the «Print» key.Only one original copy (with signature
field) can be printed, however.
6.3Testing the balance
In this menu option you can test the balance on your instrument.
Checking balance adjustment
Menu
The factory setting in this menu item is “NO” (no balance adjustment).
∑CAˆNO
Menu
∑CAˆTEST
∑CAˆ5≠000
To test (calibrate) the integral balance, use the «Scroll up» or «Scroll down»
keys to select “TEST”. The test weight to be used can be defined in the menu
(see Section 6.21).
Note:
Ensure that your instrument is in a suitable location (see Section 2.2).
Press the «Accept entry» key to start the test procedure.The sample chamber
opens automatically.
The flashing display then prompts you for the defined test weight.
Place the test weight on the balance. The rest of the procedure then continues
g
automatically.
∑CAˆ5≠001
The menu (basic settings of the instrument)
The display then shows you the recorded weight.
g
The test record is printed out automatically.
75
A ----- BALANCE-TEST -----
B 30.06.200314:48
C METTLER TOLEDO
Halogen MoistureAnalyzer
D Type:HR83
E SNR:1234567890
F SW:2.01
G Weight ID:
........................
H Set Weight:50.000 g
I Act. Weight: 50.001 g
J Tolerance: +/- 0.002 g
K Test result:
Passed
L Test performed
Signature:
........................
---------- END ---------
The balance test record contains the following information.
A Record title
B Date and Time of the record printout of the balance test
C Manufacturer and Designation of the instrument
D Type of the instrument
E Serial number of the instrument
F Version number of the software
G Field for entry of the number of the adjustment weight used
H Test weight used (= set weight)
I Recorded weight (= actual weight)
J Tolerance for the balance test (Section 6.22)
K Test result: “Passed” or “Failed”
L Field for signature of the person who performed the balance test
In the “W.CAL” menu you can print out the latest test record at any time by
pressing the «Print» key.Only one original copy (with signature field) can
be printed, however.The test record is only printed if the balance test was
performed after the last balance adjustment.
6.4Adjusting the heating module
In this menu item you can adjust the temperature control of the heating module. Consult Section 3.2 to learn when
an adjustment of the heating module is necessary. We advise you to wait at least 30 minutes after a drying operation
(or previous heating module adjustment) before performing the adjustment.
Menu
†CAˆNO
Menu
†CAˆYES
The factory setting in this menu item is “NO” (no heating module
adjustment).
Note: The number above the word “NO” shows the number of adjustments
performed to date.
If you wish to adjust the heating module, select “YES” (using the «Scroll up»
or «Scroll down» key).
The menu (basic settings of the instrument)
76
Press the «Accept entry» key to start the adjustment process. The sample
chamber automatically opens and you are prompted to insert the temperature adjustment set.
Note: The temperature adjustment set is available as an accessory (see
Section 9.9).
Remove the sample pan holder from the automatic sample chamber.
Place the temperature adjustment set (together with the sample pan handler)
in the sample chamber.
Note: Ensure that your instrument is in a suitable location (see Section 2.2).
Press the «Start» key to initiate the adjustment process. The automatic
Start
sample chamber closes and the heating module adjustment starts.
†CAˆ 100
†CAˆ 104
†CAˆ156
100
⁄30
The heating module is heated to a temperature of 100 °C. You can follow this
process in the temperature display. The instrument waits for 15 minutes until
the temperature adjustment set is displaying the temperature correctly.At the
end of this time a continuous audio signal sounds.
Read the temperature on the temperature adjustment set through the
inspection window in the heating module and use the «Scroll up» and
«Scroll down» keys (or the numeric keypad) to enter the value (NB: 1
graduation mark = 2 °C). After entering the value, press the «Accept entry»
key. The temperature must be entered within 10 minutes of the audio signal
sounding, otherwise the adjustment process will be terminated and an error
message outputted.
As this adjustment is a two-point adjustment (adjustment of the temperature
is defined by two points, namely 100 °C and 160 °C), the heating module
now heats to the second temperature (160 °C) . Proceed exactly as you did
for the first temperature. After you have confirmed your entry with the «Accept
entry» key, the adjustment is at an end.
On completion of the adjustment, the sample chamber automatically opens
and you can remove the sample pan handler with the temperature adjustment set.
A - DRYER UNIT ADJUSTM. -
B 30.06.200314:50
C METTLER TOLEDO
Halogen MoistureAnalyzer
D Type:HR83
E SNR:1234567890
F SW:2.01
G Dryer Unit adjustm.43
H Adjustment setID:
........................
I Temperature 100°C: 101°C
J Temperature 160°C: 162°C
K Adjustement performed
Signature:
........................
---------- END ---------
The menu (basic settings of the instrument)
77
Warning: The temperature adjustment set may still be hot.You should
therefore allow it to cool down first before removing it from the sample pan
handler.
On completion of the heating module adjustment, the instrument automatically quits the menu and returns to the standby mode. It is now ready for
further measurements.
The adjustment record is printed out automatically:
A Record title
B Date and Time of the record printout of the heating module adjustment
C Manufacturer and Designation of the instrument
D Type of the instrument
E Serial number of the instrument
F Version number of the software
G Identification (consecutive number) of the heating module adjustment,
allows assignment of the measurement record to the appropriate adjustment record
H Field for entry of the number of the temperature adjustment set used
(number is printed on the set)
I Set and actual temperature for the first adjustment point
J Set and actual temperature for the second adjustment point
K Field for signature of the person who performed the heating module
adjustment
Enter the number of the temperature adjustment set, sign the record and store
it in a safe place. This assures traceability, one of the basic requirements of
every quality assurance system.
In the “T.CAL” menu you can print out the latest adjustment record at any time
by pressing the «Print» key.
Only one original copy (with signature field) can
be printed, however.
6.5Testing the heating module
In this menu option you can test the temperature control on the heating module.
Menu
†CAˆNO
Menu
†CAˆTEST
Check the heating module adjustment with TEST
The factory setting in this menu item is “NO” (no heating module
adjustment).
To check the last heating module adjustment at 100 °C and 160 °C, use
the «Scroll up» or «Scroll down» keys to select “TEST”.Press the «Accept
entry» key to confirm your selection.
The menu (basic settings of the instrument)
78
Menu
†TOˆ3
†TESÎ
Start
A --- DRYER UNIT TEST ---
B 30.06.200314:48
C METTLER TOLEDO
Halogen MoistureAnalyzer
D Type:HR83
E SNR:1234567890
F SW:2.01
G Adjustment set ID:
........................
H Set Temperature: 100 °C
I Act. Temperature: 101 °C
J Tolerance:+/- 3 °C
K Set Temperature: 160 °C
L Act. Temperature: 161 °C
Tolerance:+/- 3 °C
M Test result:
Passed
N Test performed
Signature:
........................
---------- END ---------
Choose the permissible tolerance (as defined by your company’s quality
management) for the heat output.With the HA-TCC temperature adjustment
set, the HR83 can maintain a tolerance of +/- 3°C.
Press the «Accept entry» key to start the test procedure.The sample chamber
opens automatically and you are prompted to insert the temperature
adjustment set.
Note: Ensure that your instrument is in a suitable location (see Section 2.2).
Press the «Start» key to initiate the test procedure.The sample chamber
closes and the test for the two-point heating module adjustment begins.
Now proceed as for the standard two-point heating module adjustment.
The test record is printed out automatically.
The heating module test record contains the following information.
A Record title
B Date and Time of the record printout of the heating module test
C Manufacturer and Designation of the instrument
D Type of the instrument
E Serial number of the instrument
F Version number of the software
G Field for entry of the number of the temperature adjustment set used
(number is printed on the set)
H Set temperature (first temperature) for the test adjustment
I Actual temperature (first temperature) for the test adjustment
J Tolerance (this value enables the test to be graded as “Passed” or
“Failed”)
K Set temperature (second temperature) for the test adjustment
L Actual temperature (second temperature) for the test adjustment
M Test result: “Passed” or “Failed”
N Field for signature of the person who performed the heating module test
In the “T.CAL” menu you can print out the latest test record at any time by
pressing the «Print» key.Only one original copy (with signature field) can
be printed, however.The test record is only printed subsequently if the
heating module test was performed after the last heating module adjustment.
The menu (basic settings of the instrument)
79
Check the heating module adjustment with TEST-1
The purpose of this menu option is to check the heat output between 50 °C
and 180 °C.The temperature for TEST-1 can be freely chosen within this
range.
Menu
†CAˆNO
Menu
†CAˆTEST1
†CAˆ65
†TOˆ3
Start
A -- DRYER UNIT TEST 1 --
B 30.06.200314:48
C METTLER TOLEDO
Halogen MoistureAnalyzer
D Type:HR83
E SNR:1234567890
F SW:2.01
G Adjustment set ID:
........................
H Set Temperature:65 °C
I Act. Temperature: 66 °C
J Tolerance:+/- 3 °C
K Test result:
Passed
L Test performed
Signature:
........................
---------- END ---------
The factory setting in this menu item is “NO” (no heating module
adjustment).
To test the heating module, use the «Scroll up» or «Scroll down» keys to
select “TEST 1”.Press the «Accept entry» key to start the test procedure.
Enter the target temperature you require (e.g. 65 °C) and press «Accept
entry». Then define the temperature tolerance and press «Accept entry»
again.
The sample chamber automatically opens and you are prompted to insert
the temperature adjustment set. Remove the sample pan holder from the
automatic sample chamber and insert the temperature adjustment set in the
sample pan handler. Place the sample pan handler in the automatic sample
chamber.
Press the «Start» key to initiate the test procedure.The sample chamber
closes and the test for the heating module adjustment begins.When the
target temperature is reached (after a waiting time of 15 min) a continuous
audio signal sounds.The actual temperature can be entered over a period
of no more than 10 minutes.
To do this, read the temperature on the temperature adjustment set through
the inspection window in the heating module and use the «Scroll up» and
«Scroll down» keys (or the numeric keypad) to enter the value (NB: 1
graduation mark = 2 °C).Press the «Accept entry» key to confirm your input.
The temperature must be entered within 10 minutes of the audio signal
sounding, otherwise the adjustment process will be terminated and an error
message outputted.
The test-1 record is printed out automatically.
The heating module test-1 record contains the following information.
A Record title
B Date and Time of the record printout of the heating module test
C Manufacturer and Designation of the instrument
D Type of the instrument
E Serial number of the instrument
F Version number of the software
G Field for entry of the number of the temperature adjustment set used
(number is printed on the set)
H Set temperature for the test adjustment
The menu (basic settings of the instrument)
80
I Actual temperature for the test adjustment
J Tolerance (this value enables the test to be graded as “Passed” or
“Failed”)
K Test result: “Passed” or “Failed”
L Field for signature of the person who performed the heating module test
In the “T.CAL” menu you can print out the latest test record at any time by
pressing the «Print» key.
be printed, however.The test record is only printed subsequently if the
heating module test-1 was performed after the last heating module adjustment.
6.6Resetting to the factory settings
Only one original copy (with signature field) can
In this menu item you have the possibility the reset the basic settings of the instrument to the factory setting.
Warning! With the exception of the time, date and dialog language, all
individual settings (incl. the method memory)
you have made in the
menu will be lost and replaced by the factory settings! Journal and
statistics data will be deleted.
Menu
FC†SEÎNO
The factory setting in this menu item is “NO”. If you really wish to delete
your individual menu settings, select “YES” and confirm this with the «Accept
entry» key.
Menu
FC†SEÎYES
Menu
ABORT?
For reasons of safety, you are again asked if you really wish to delete your
data. You can abort the procedure at this point or delete the data, i.e. reset
to the factory settings. Confirm your selection with the «Accept entry» key.
Menu
SET?
The menu (basic settings of the instrument)
81
6.7Clearing the method memory
The method memory contains the individual settings for your measurements (switch-off criterion, display mode,
print interval, etc.) and the results of your measurements (statistics, journal). This menu item allows the method
memory to be cleared. Your settings will be replaced by the factory settings and statistics and journal will be cleared.
Warning! If the method memory is cleared, you will lose all individual
settings with the exception of the settings in the menu!
Menu
µCLÈNO
Menu
The factory setting in this menu item is “NO”. If you really wish to clear the
method memory, select “YES” and confirm this with the «Accept entry» key.
µCLÈYES
For reasons of safety, you are again asked whether you really wish to clear
ABORT?
the method memory. You can abort the procedure at this point or clear the
memory. Confirm your selection with the «Accept entry» key.
CLEAR?
6.8Setting the time
In this menu item you enter the current time. This is necessary when putting your instrument into operation for the
first time and for possible corrections (e.g. changeover from summer to winter time).
Menu
1
TIMË1‹04
Menu
1
TIMË1Æ28
Set the time with the «Scroll up» and «Scroll down» keys (or via the numeric
keypad) (Range 00:00–23:59).
Confirm your setting with the «Accept entry» key and the instrument stores
the time.
The menu (basic settings of the instrument)
82
6.9Entering the date
In this menu item you enter the current date. This is necessary when putting the instrument into operation for the
first time and for possible corrections.
Menu
DAÎ2{0“03
2
Set the date with the «Scroll up» and «Scroll down» keys (or via the numeric
keypad) in the format “Day – Month – Year” (DD.MM.YY). Confirm your entry
with the «Accept entry» key and your instrument stores the date.
Menu
DAÎ0[0#03
2
Note: If US English is used as the dialog language, the entry has the format
“Month – Day – Year” (MM/DD/YY).
6.10Setting the vibration adapter
You can use the vibration adapter to match your instrument to the ambient conditions at its location.
Menu
VIBRÅLOW
Menu
VIBRÅME∂
Menu
VIBRÅHIGH
3
The following settings are available:
Setting for very stable ambient conditions. The instrument operates quickly
but is sensitive to vibrations.
3
Setting for normal ambient conditions. This is the factory setting.
3
Setting for unstable ambient conditions. The instrument operates more
slowly but is less sensitive to vibrations.
Select the appropriate setting and confirm your selection with the «Accept
entry» key.
The menu (basic settings of the instrument)
83
6.11Selecting operating mode of the automatic sample chamber
In this menu item you can choose whether the automatic sample chamber should operate automatically or
manually. Your instrument is set in the factory to the automatic operating mode. We advise you to use the manual
operating mode for samples which contain readily volatile substances. In contrast to the automatic operating mode,
in the manual operating mode the sample chamber does not automatically close when the «Start» key is pressed.
However, the initial weight (wet weight) important for the determination of the moisture content is recorded. In the
manual operating mode you have time for further preparation of the sample (e.g. mixing with quartz sand or even
distribution of the sample) while weight losses due to evaporation during the preparation time are measured from
the outset. As soon as the sample is ready for drying, press the key «Open/close auto sample chamber». The
automatic sample chamber closes and drying starts. In the manual operating mode you yourself can open the
sample chamber during a drying operation e.g. in standby mode (Section 5.4.5). In contrast to the automatic mode,
drying will not be stopped but simply interrupted until the automatic sample chamber is closed again.
Menu
STARÎAUTø
4
The factory setting in this menu item is “AUTO.” (automatic operating
mode). If you wish to switch to the manual operating mode, select “MAN.”
and confirm this with the «Accept entry» key.
Menu
4
STARÎMA~
6.12Audio signal
Many processes are signalled by an audio signal (e.g. end of drying, entries, error messages, etc.). In this menu
item you can select whether the audio signal should be loud, soft or switched off.
Menu
5
BEE∏OFF
Menu
5
BEE∏LOW
Menu
5
BEE∏HIGH
The following settings are available:
No audio signal
Soft signal (Factory setting)
Loud signal
Select the desired setting and confirm with the «Accept entry» key.
Note: The alarm clock symbol in the display indicates that an audio signal
has been activated in the menu.
The menu (basic settings of the instrument)
84
6.13Protecting your settings against change
You can protect the settings of your methods parameters (e.g. switch-off criterion, drying temperature, etc., see
Section 4) and the settings in the Methods menu (Section 5.4) against change by locking the appropriate keys for
entries. This also protects the statistics and the journal against clearance. Alternatively you can work with a
password which protects your method settings if the system is restored to factory settings (see Section 6.1).
Menu
KE¥PROÎ OFF
Menu
KE¥PROÎON
6
The instrument is set in the factory to allow the parameters to be changed
(“PROT: OFF”).
6
If you wish to protect the settings against change, selection “PROT: ON” and
confirm this with the «Accept entry» key.
The display shows the padlock symbol and the following settings can then
no longer be changed:
– Switch-off criterion
– Drying program
– Display mode
– Print interval
– Drying temperature
– Drying time
– Target weight of the weighing-in aid
– All settings in the Methods menu (Section 5.4)
If you press one of the locked keys, an audio signal and an error message
(see Section 8.2) draw your attention to the nonavailability of the corresponding function.
6.14Superimposing or removing the function symbols
In this menu item you can select whether the function symbols should always be visible or only when needed. In
routine operation, the display is easier to survey if the symbols are faded out.
Menu
7
SYMBOˆON
Menu
7
SYMBOˆOFF
In the factory the symbols are switched on (“ON”).
If you wish to switch off the symbols, select “OFF” and confirm this with the
«Accept entry» key.
Note: If you switch off the function keys, they will be displayed for around
30 seconds when the instrument is switched on and then faded out. If one
of the function keys is pressed, the function symbols will be shown
immediately and then faded out again 30 seconds after a function key has
last been pressed.
6.15Switching the printer on or off
In this menu item you can switch the built-in printer on or off.
Menu
8
PRINÎON
In the factory this item is switched on (“ON”).
The menu (basic settings of the instrument)
85
Menu
PRINÎOFF
8
If you wish to switch the printer off, select “OFF” and confirm this with the
«Accept entry» key.
6.16Entering company name for printed records
In this menu item you can define whether the name of your company (or any other text, e.g. the name of the
department) should appear at the very top of the printout in the record header. There are two lines available.
Menu
9
COMPANŸ OFF
Menu
9
COMPANŸON
Menu
A
9
A...Z
In the factory this item is switched off (“OFF”).
If you wish to print your company name, select “ON” and confirm this with
the «Accept entry» key.
Now enter the name: press the «A…Z» key. A flashing “A” appears at the first
input position on the extreme left and signals that the instrument is ready for
your entry.
Menu
M
Menu
MA
You can now use the «Scroll up» and «Scroll down» keys to select the first
letter of the name (uppercase letters only) or special characters (+, –, etc.).
9
9
A...Z
Confirm your selection with the «A…Z» key. The flashing “A” now reappears,
but at the second position.
Enter the remaining characters of the desired name (up to 20 characters) as
described above.
Notes
–A space is symbolized by the underline character “_”.
– You can remove wrong characters with the «Delete» key and then correct
your entry.
The menu (basic settings of the instrument)
86
–
Numbers can be entered directly via the numeric keypad.If you enter a
character via the numeric keypad, the instrument remains in numeric
input mode and a flashing dash appears at the next input position instead
of a flashing “A”. To switch back to the text mode, press the «A…Z» key
again.
As soon as you have finished entering the name, press the «Accept entry»
SAMPLELTD
key.
If you wish you can enter a department name, for example, in a second line:
CENTRAL LABORATORY
press the «Accept entry» key.
The inputted name appears with immediate effect on the record assuming
you have selected the full record (see next Section).
6.17Selecting the type of record
In this menu item you decide the length and hence the extent of detail of the measurement records.
Menu
10
RCORflNORµ
Menu
10
RCORflFULL
In the factory setting, normal records (“NORM.”) are printed out. You will
find the normal record explained in Section 4.9.
If you wish for more detailed measurement records, select “FULL” and
confirm this with the «Accept entry» key.
Specifications for the design of the measurement record can be found in
Section 4.9.
6.18Defining the free print interval
You became familiar with the preselected print intervals in Section 4.6. If the preselected values are not suitable for
your application, you can define a free print interval in this menu item.
Menu
11
ƒPR†I˙OFF
Menu
11
ƒPR†I˙ON
In the factory the free print interval is deactivated (“OFF”).
If you wish to define a free print interval, select “ON” and confirm this with
the «Accept entry» key.
Menu
11
PR†I˙1fi00
Now enter the desired print interval using the «Scroll up» and «Scroll down»
keys or the numeric keypad (between 5 seconds and 60 minutes). Confirm
your entry with the «Accept entry» key.
The free print interval is now available with its own symbol under the function
key «Print interval».
The menu (basic settings of the instrument)
87
6.19Switching the statistics and journal functions on and off
If you do not wish to use the statistics and journal functions (see Section 5), you can switch them off in this menu
option.This makes it easier to use the instrument if you are working with only one type of sample and therefore
do not require the features of the statistics and journal functions.
Menu
12
ST/JOÙON
Menu
12
ST/JOÙOFF
Menu
ABORT?
Menu
CLEAR?
The statistics and journal functions are switched on as the factory
setting (“ST/JOU: ON”).
To switch off the statistics and journal functions choose “ST/JOU: OFF” and
confirm by pressing the «Accept entry» key.
When you press the «Accept entry» key, the prompt “ABORT?” or “CLEAR?”
appears. Choose “CLEAR” and press «Accept entry» to confirm.
The features of the statistics and journal functions are no longer available
to you.
6.20Selective clearing of comment lines
In this menu you can clear the existing comment lines (see Section 4.8).In routine operation it is often the case
that individual comment lines can be retained for several measurements.On the other hand you need to prevent
comment lines (compulsory measurement inputs) from being accidentally included from an old measurement.You
can therefore activate automatic clearing after each measurement for all or individual comment lines.
To clear comment lines after each measurement, use the «Scroll up» and
«Scroll down» keys to select “YES”. Press the «Accept entry» key to confirm
your setting.
You can then specify which comment lines you wish to be cleared after each
measurement. Use the «Scroll up» and «Scroll down» keys to choose the
comment lines you wish to be cleared. The options are:
Press the «Accept entry» key to confirm your setting.
The menu (basic settings of the instrument)
88
6.21Defining the test weight
You learnt about checking the balance adjustment in Section 6.4.In this menu option you can specify the target
weight for the test.
Menu
TES†Á5≠000
14
g
The factory setting in this menu option is “50.000 g”.
To define a test weight, choose menu option 14.You can then define a new
test weight between 0.100 g and 80.000 g via the numeric keypad and
press the «Accept entry» key to confirm.
6.22Tolerance for balance test
In this menu option you can define the tolerance range.This value is used to calculate whether the balance test
is “Passed” or “Failed” (difference calculation).
Menu
TES†Î ≠002
15
The factory setting in this menu option is “0.002 g”.
g
To define a tolerance range, choose menu option 15.
You can then define
a new tolerance range between 0.001 g and 0.010 g via the numeric
keypad and press the «Accept entry» key to confirm.
6.23Selecting the dialog language
In this menu item you specify the dialog langauge.
Menu
DEUTSCH
Select the desired dialog language and then confirm your entry with the
«Accept entry» key.
The menu (basic settings of the instrument)
89
Menu
FRANCAIS
Menu
ITALIANO
Menu
ESPANOL
Menu
RUSSIAN
Menu
ENGLISH EU
Menu
ENGLISH US
The following languages are available:
– German (“Deutsch”)
– French (“Français”)
– Italian (“Italiano”)
– Spanish (“Español”)
– Russian (“Russian”)
– British English (“English EU”)
– American English (“English US”)
Note: If you select American English (“English US”) as the dialog language,
the date format is changed and appears on all records in the American
notation (month/day/year).
Saves the menu item and skips to the next menu
item
For information on the separate Method menu, please refer to Section 5.4.
Select settings
Servicing and replacing individual parts
91
7Servicing and replacing individual parts
In this Section you will learn how to keep your Moisture Analyzer in good condition and how to replace expendable
parts. You also find information on how to check the printer and change the printer paper and ribbon.
7.1Cleaning
To obtain precise measurement results, we advise you to clean the temperature sensor and the protective glass of
the heating modules at regular intervals. Please note the following directions for cleaning your instrument:
Disconnect the instrument from the power supply before cleaning.
Open the automatic sample chamber by pulling it out.
Remove the draft shield element and the sample pan holder before cleaning.
Use a lint-free cloth for cleaning.
Clean the exterior of the instrument and the drying compartment with a mild
cleaning agent. Although the housing is extremely rugged and resistant to
solvents, never use abrasive cleaning agents or solvents!
Ensure that no liquid enters the interior of the instrument.
To clean the temperature sensor and the protective glass of the heating module,
you should first open the heating module as described in Section 7.6.
Carefully remove any deposits from the black temperature sensor.
Servicing and replacing individual parts
3
92
To clean the protective glass of the heating module, dismantle the heating
module (see Section 7.6), place it on a flat work surface and clean the
protective glass with a commercial glass cleaner.
The air inlet of the fan is located at the rear of the instrument and its exterior
should be cleaned from time to time to free it from any dust deposits.
After the temperature sensor and/or the protective glass have been cleaned,
we recommend adjusting the heating module (Section 6.4).
7.2Changing the printer paper and ribbon
Apart from the need to replace the printer paper and ribbon on occasion , the built-in printer requires no maintenance.
Changing the printer paper
The printer paper is changed as follows:
Open the cover of the built-in printer.
Pull any remaining paper upwards out of the printing unit.
HR7
Insert a new roll of paper in the paper compartment and feed the paper
horizontally through the slot at the back of the printing unit.
HR73
Halogen Moisture Analyzer
EJECT
H
S
U
P
Press the «Paper feed» key until the leading edge of the paper exits the
printing unit at the top.
Lead the paper through the slot in the cover of the built-in printer and close
the cover.
HR73
Halogen Moisture Analyzer
EJECT
H
S
U
P
HR73
Halogen Moisture Analyzer
HR73
Halogen Moisture Analyzer
Servicing and replacing individual parts
93
Replacing the ribbon
When the print quality deteriorates, replace the ribbon as follows:
Open the cover of the built-in printer and press the projection marked “PUSH”
on the right of the ribbon cassette.
EJECT
H
S
U
P
Pull the ribbon cassette up and out of the instrument and insert the new
cassette. Thread the paper between the ribbon and the cassette. Press the
cassette down as far as it will go until you hear it click into place.
EJECT
H
S
U
P
Tension the ribbon by turning the small wheel on the left of the cassette in
the direction shown by the arrow. After you have tensioned the ribbon, close
the cover of the printer.
Used ribbons must be disposed of in accordance with applicable
HR73
Halogen Moisture Analyzer
EJECT
H
S
U
P
customer and national regulations.
7.3Replacing the power line fuses
If the display of your instrument remains “dark” after switching on, in all probability the power line fuses of the
instrument are blown.
To change the fuses, proceed as follows:
Disconnect the instrument from the power supply.
The two power line fuses (for neutral and phase) are located at the rear of
the instrument. Turn the two fuse holders with a screwdriver to the left and
pull the fuse holders out of the instrument.
Check the condition of both fuses. Replace blown fuses by those of the same
type with the same rated value (5 x 20 mm, T6, 3H 250 V).
Servicing and replacing individual parts
94
The use of fuses of a different type or with different values as well as the
bridging of fuses is not allowed and can hazard your safety and lead to
instrument damage!
7.4Printer test
Your instrument has a function to check the proper functioning of the built-in printer.
Ensure that the instrument is switched off.
Press and hold the «Paper feed» key while simultaneously switching the
On/Off
On/Off
instrument on with the «On/Off» key. The printer prints out its complete
character set.
To stop the test, switch the instrument off with the «On/Off» key.
7.5Installing the built-in printer
If your instrument is not already fitted with a built-in printer, you can order a printer as an option and retrofit it at
any time. You can also change the built-in printer yourself without any problems should this be necessary. The
procedure for installing the built-in printer is described in the 11780558 you received with the printer.
7.6Dismantling and replacing the heating module
If a fault arises, the complete heating module can be replaced. We also advise you to open or dismantle the heating
module to clean the protective glass of the heating module (see Section 7.1).
Disconnect the instrument from the power supply and be sure to allow the
heating module to cool down for 10 minutes before you start the work!
The heating module is released by withdrawing the automatic sample
chamber 2-3 cm.
Servicing and replacing individual parts
95
Swing the heating module upward.
Disengage the hinge of the heating module by turning the two yellow levers
90° forward.
Remove the heating module (pull forward).
Install the (new or cleaned) heating module in the instrument and lock the
hinge with the two yellow levers.
After a heating modules has been cleaned or replaced, we recommend readjusting the heating module of your Halogen Moisture Analyzer (Section 6.4).
Defective parts must be disposed of in accordance with applicable
customer and national regulations.
If problems arise on occasion
96
8If problems arise on occasion
In this Section you will learn how errors can appear during operation of your Moisture Analyzer and how you can
rectify these errors.
8.1Notes on the error messages
Your instrument distinguishes between the three different types of errors explained below.
Input errors (key not active or can not be pressed in current operating
status) are reported by your instrument with a short audio signal (if
activated in the menu) without an error message in the display.
An application error occurs when the instrument can not perform a
ERROÈ 3
procedure or a corrective action is necessary as the limits of a value range
have been violated or because a general operating error exists. Application errors are reported by your instrument with an audio signal. Further,
the message “ERROR” appears in the display followed by the error
number. You will find a list of all application errors in the next Section.
ERROR:3
Sample weight out if tol
----30.06.03---15:28----
ERROÈ 3
SYSTEM-ER®
If your instrument is equipped with a built-in printer, you can print the error
message in clear text with the «Print» key.
Before you can continue working, you must delete the error message with
Reset
5
the «Reset» key.
System errors indicate that a program or hardware error exists. The
message “SYSTEM-ERR.” appears in the display and the error number is
shown in the top right corner. In such a case, disconnect the instrument
from the power supply. Should the error reappear after reconnection to the
power supply, contact your METTLER TOLEDO dealer to arrange an
appointment for diagnosis and repair, Note the error number in order to
facilitate the work of the service engineer.
If problems arise on occasion
97
8.2Meaning of the error messages
In this Section you will find all error messages which appear in the display, the text of the error record on the printer,
the reason for the error message and directions on how to rectify the error.
Display:Error: 1
ERROÈ 1
ERROÈ 2
Printout:Weighing result unstable
Cause:No stability during taring or adjustment
Remedy:Ensure stable ambient conditions and an optimum location.
Also take care that no part of the sample or the sample pan
touch the draft shield element or the sample pan holder. Highly
volatile substances in the sample also prevent a stable
weighing result being detected.
Display:Error: 2
Printout:Wrong calibration weight
Cause:No or wrong adjustment weight loaded
Remedy:Load required adjustment weight
ERROÈ 3
ERROÈ 4
ERROÈ 5
Display:Error: 3
Printout:Sample weight out of tol
Cause:Sample weight below 100 mg or outside the set limits with the
weighing-in aid active (Section 4.7).
Remedy:Weigh in a sample of al least 100 mg, adjust the sample
weight so that it is within the set limits or deactivate the
weighing-in aid.
Display:Error: 4
Printout:Missing tare weight
Cause:Sample vessel has not been tared
Remedy:Tare sample vessel
Display:Error: 5
Printout:Entry inadmissible
Cause:Lower/ upper limit of input range violated.
For the «Target» key:
function key) is not activated.Taring or weighing in has not
yet been performed.
Remedy:Enter value in valid range.
For the «Target» key:
(Section 4.12).Tare and weigh in sample.
Test measurement (switch-off criterion
Activate test measurement (“t”)
If problems arise on occasion
98
ERROÈ 6
ERROÈ 7
ERROÈ 8
Display:Error: 6
Printout:Not activated in menu
Cause:Keypad has been locked in menu, printer deactivated in menu
Remedy:Remove keypad locking, activate printer in menu
Display:Error: 7
Printout:Timer not activated
Cause:The selected switch-off criterion does not allow entry of a time
Remedy:Select “Timed switch-off” as switch-off criterion
Display:Error: 8
Printout:Entry missing
Cause:In the heating module adjustment, the wait time of 10 minutes
for entry of the temperature value has been exceeded
Remedy:Repeat heating module adjustment and enter temperature
values before elapse of the wait time
ERROÈ 9
ERROÈ 10
ERROÈ 11
Display:Error: 9
Printout:Function error
Cause:Undefined condition
Remedy:Press «Reset» key or briefly switch instrument off then on again
with the «On/Off» key. If this error appears in the calibration,
please contact your METTLER TOLEDO dealer
Display:Error: 10
Printout:Temperature > Start-Temp.
Cause:Sample chamber has not cooled down sufficiently or the
standby temperature (Section 5.4.5) was set higher than the
drying temperature
Remedy:Allow the instrument to cool down to the starting temperature
or reduce the standby temperature
Display:Error: 11
Printout:PW:
Cause:Password was entered incorrectly
Remedy:Enter password correctly
Wrong entry
ERROÈ 12
£-----$
!-----¨
RAM LOST
If problems arise on occasion
99
Display:Error: 12
Printout:Password protection PW-1
Cause:The method parameters are protected by a password (PW-1)
Remedy:Reset password PW-1, change the methods and set a new
PW-1
Display:I_ _ _ _ _I
Cause:Underload
Remedy:Load the sample pan holder
_ _ _ _ _
Display:I
Cause:Overload
Remedy:Decrease the weight of your sample
Display:RAM LOST
Cause:The installed battery is discharged (the equipment was discon-
nected from the power supply for a longer periode) or the
instrument is faulty
Remedy:Charge battery (connect instrument to the power supply for
5 hours) and then reenter all settings. If the error reappears,
contact your METTLER TOLEDO dealer.
I
8.3What if...?
… the display remains “dark” after switching on?
– no line voltage
– power cable not connected
– blown power line fuses
– instrument faulty
Ensure that the instrument is connected to the power supply and that power supply is actually supplied. Check the
power line fuses of the instrument and replace if necessary (see Section 7.3). If the instrument still refuses to
function, contact your METTLER TOLEDO dealer.
… after switching on “0.000” flashes in the display?
The sample pan holder is not installed. Install the sample pan holder.
… the function symbols disappear from the display after a certain time
You have deactivated the symbols in the menu (see Section 6.14).
If problems arise on occasion
100
… the symbol of the stability detector is continuously lit up immediately after the start?
As soon as the symbol of the stability detector fades, the weighing result is stable and is accepted as a “wet weight”.
If the symbol does not fade, your instrument is probably at an unsuitable location (vibrations, shocks, powerful
drafts, etc.). Seek a more suitable location.
Samples containing readily volatile substances may never reach stability owing to continuous evaporation. In this
case, you must select manual operation for the automatic sample chamber (see Section 6.11).
… the built-in printer does not print?
Ensure that the printer is activated in the menu (see Section 6.15) and that the ribbon and paper are correctly
installed. If the printer still refuses to print, perform a printer test (see Section 7.4). If the test is unsuccessful, contact
your METTLER TOLEDO dealer.
… no drying time can be entered?
Entry of the drying time is possible only if you have selected the switch-off criterion “Timed switch-off”, with all other
switch-off criteria this input possibility is not available (see Section 4.4). If you attempt to enter a drying time, the
instrument reports “Error: 7” (see Section 8.2).
… certain keys are inactive?
You have protected the parameters (switch-off criterion, print interval, etc.) in the menu against change (see Section
6.13). If you attempt to change a parameter, the instrument reports “Error: 6” (see Section 8.2).
… an error message appears when I press certain keys?
You have entered an illegal value or you have activated password PW-1 in the menu, see Error messages (see
Section 8.2).
… the measurement takes too long?
You have selected an unsuitable switch-off criterion. The suitable switch-off criterion is simple to determine by a test
measurement. You will find information on how to perform a test measurement in Sections 4.4 and 4.11.
An excessive amount of sample can also be the cause of slow drying, likewise samples which tend to form a skin
which hinders vaporization.
Perform experiment at higher temperature.
… the weighing-in aid is not visible?
The weighing-in aid is available only if you have activated (see Section 4.7).
… the automatic sample chamber does not close after pressing «Start»?
You have selected the manual operating mode in the menu (see Section 6.11).
… the device switches off?
The method you are using called for a standby temperature, and you have activated the standby time or standby
switchoff time (Section 5.4.5.1).
… the device begins to heat up when it is switched on or when the method is changed?
The method you are using called for you to activate the standby temperature (Section 5.4.5.1).
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.