Mettler Toledo HB43-S Operating Instructions Manual

Moisture Analyzer
HB43-S
Substance A
A Substance A
88 °C
-12.48
%
MC
Overview of your HB43-S Halogen Moisture Analyzer
1
21
20
19 17 1618
15
13
14
12
22
2
2
1
4
5
6
3
28
125 26
29
30
31
33
34 3632
23 24
27
31
11
8
10
7
9
35
Display, controls and connections of your HB43-S
No. Designation Info see
section
1 Display (example) all
2 Keypad all
3 Draft shield 2.3
4 Sample chamber 2.3/6.1
5 Heat shield 2.3
6 Sample pan support 2.3/2.4
7 Heating module 1.2/5.4
8 Inspection window and vent 5.4
9 Level indicator (level) 2.3
10 Temperature sensor 6.1
11 Halogen lamp protective glass 6.1
12 Key «Menu» 4/5
13 Key «Drying temperature» 4.4.4.1
14 Key «Ú» (Print) 4.4/5.1
15 Key «B» (for choosing method B) 4.3/4.4
16 Softkeys (context-dependent keys) 4.2
17 Key «Stop» (stop drying) 4.5
18 Key «Start» (start of drying) 5.6.8
19 Key «–>0/T<–» (Zero/Tare) 2.4
20 Key «A» (for choosing method A) 4.3/4.4
21 Key «ß» (for exiting the menu or
stopping drying)
4.2/4.5
22 Key «On/Off» 2.4/4.5
23 Active method (A or B) 4.3/4.5
24 Name of method 4.4.1
25 Display of drying time 2.4/4.4.8
26 Display of drying temperature 2.4/4.4.4
27 Display of drying result 2.4/4.5
28 Stability detector 7.1/7.2
No. Designation Info see
section
29 Symbol for calculated result 2.4
30 Status display ("User Guide") 2.4
31 Leveling screw 2.3
32 Antitheft device socket 2.3
33 Power supply receptacle 2.3
34 Power line fuse 6.2
35 Fan 1.3
36 RS232C interface connection 6.3/8.2
4 Contents
Contents
1 Getting to know your Moisture Analyzer ...................................................................................... 6
1.1 Introduction ................................................................................................................................ 6
1.2 What is the Halogen Moisture Analyzer used for ............................................................................. 6
1.3 Safety has priority ...................................................................................................................... 7
1.4 Important information about these instructions ............................................................................... 9
1.5 Disposal .................................................................................................................................... 9
1.6 Finding More Information ............................................................................................................. 9
2 Your first measurement in next to no time ................................................................................. 10
2.1 Unpacking and checking the standard equipment ......................................................................... 10
2.2 Selecting the location ................................................................................................................. 11
2.3 Setting up, leveling and connecting to power supply...................................................................... 11
2.4 Your first measurement .............................................................................................................. 13
3 How to obtain the best results .................................................................................................. 17
3.1 Measurement principle of the Halogen Moisture Analyzer ............................................................... 17
3.2 Notes on adjustment of the balance and the heating module .......................................................... 18
3.3 Optimum sample preparation ..................................................................................................... 19
4 Practical application of your Moisture Analyzer ......................................................................... 20
4.1 The operating concept and methods ............................................................................................ 20
4.2 The menu ................................................................................................................................. 21
4.3 Choosing and allocating methods from database ......................................................................... 21
4.4 Developing own methods and defining parameters ....................................................................... 23
4.4.1 Specifying name of method ........................................................................................................ 23
4.4.2 Specifying target weight of sample .............................................................................................. 24
4.4.3 Selecting the drying program ...................................................................................................... 25
4.4.4 Setting the drying temperature ..................................................................................................... 25
4.4.4.1 Setting the drying temperature directly .......................................................................................... 26
4.4.5 Temperature reduction ............................................................................................................... 26
4.4.6 Selecting the switch-off criterion .................................................................................................. 27
4.4.7 Setting time interval for free switch-off criterion .............................................................................. 29
4.4.8 Setting drying time for timed switch-off ........................................................................................ 29
4.4.9 Selecting the display mode ......................................................................................................... 29
4.4.10 Activating free factor .................................................................................................................. 31
4.4.11 Specifying free factor ................................................................................................................. 31
4.4.12 Choosing number of decimal places ........................................................................................... 32
4.4.13 Saving new method and exiting menu ......................................................................................... 32
4.4.14 Printing out records of methods .................................................................................................. 33
4.5 Performing a measurement ........................................................................................................ 33
4.6 Information on the measurement record ....................................................................................... 37
5 Basic settings of the instrument ................................................................................................ 38
5.1 Menu overview ......................................................................................................................... 38
5.2 Testing the moisture analyzer with SmartCal ................................................................................. 39
5.3 Adjusting the balance ................................................................................................................ 41
5.4 Adjusting the heating module ..................................................................................................... 43
5Contents
5.5 Selecting the dialog language ..................................................................................................... 45
5.6 System settings ......................................................................................................................... 46
5.6.1 Setting the time ......................................................................................................................... 46
5.6.2 Entering the date ....................................................................................................................... 46
5.6.3 Choosing time format ................................................................................................................ 47
5.6.4 Choosing date format ................................................................................................................ 47
5.6.5 Setting display contrast .............................................................................................................. 47
5.6.6 Setting display brightness .......................................................................................................... 47
5.6.7 How to activate the protection of menu settings ............................................................................ 48
5.6.8 Selecting the start mode for drying .............................................................................................. 48
5.6.9 Switching the printer on or off ..................................................................................................... 49
5.6.10 Defining the print interval ........................................................................................................... 49
5.7 Interface settings ....................................................................................................................... 50
5.7.1 Setting the baud rate (data transmission speed) ........................................................................... 50
5.7.2 Setting bit / parity ...................................................................................................................... 50
5.7.3 Setting the data flow control ("Handshake") ................................................................................. 51
5.8 Resetting to the factory settings ................................................................................................... 51
6 Servicing and replacing individual parts .................................................................................... 52
6.1 Cleaning the heating module and the sample chamber .................................................................. 52
6.2 Replacing the power line fuse ..................................................................................................... 53
6.3 Connecting to an external printer ................................................................................................. 54
7 If problems arise on occasion ................................................................................................... 55
7.1 Error messages ......................................................................................................................... 55
7.1.1 Input errors ............................................................................................................................... 55
7.1.2 Application errors ...................................................................................................................... 55
7.1.3 Function errors .......................................................................................................................... 56
7.2 What if...? ................................................................................................................................ 57
8 Further useful information ........................................................................................................ 59
8.1 Notes on interpretation of the measurement results and the ideal sample weight ............................... 59
8.2 RS232C interface ...................................................................................................................... 59
8.3 Application Brochure ................................................................................................................. 59
8.4 Switch-off criterion "Weight loss per unit of time" ........................................................................... 60
8.5 Technical data ......................................................................................................................... 61
8.6 MT-SICS Interface commands and functions ................................................................................. 64
8.7 Accessories, consumables and spare parts .................................................................................. 65
8.7.1 Accessories .............................................................................................................................. 65
8.7.2 Consumables ........................................................................................................................... 66
8.7.3 Spare parts ............................................................................................................................... 67
9 Index ...................................................................................................................................... 68
6 Getting to know your Moisture Analyzer
1 Getting to know your Moisture Analyzer
Please read through this section carefully, it contains important information for safe and economical operation of your Moisture Analyzer.
1.1 Introduction
Thank you for deciding to purchase a Halogen Moisture Analyzer from METTLER TOLEDO – you have made a wise choice. Your Moisture Analyzer is fast and reliable. It offers a high level of operating convenience and useful functions to facilitate determination of the moisture content of your samples.
Behind your instrument stands METTLER TOLEDO, a leading manufacturer of not only balances and scales for the lab and production, but also analytical measuring instruments. A customer service network covering the entire globe with well trained personnel is at your service at all times, whether you are choosing accessories or require guidance for a specific application to ensure optimum utilization of your instrument.
To ensure you make full use of the possibilities offered by your Moisture Analyzer, we advise you to read through these operating instructions very carefully.
The operating instructions are based on the initially installed firmware (software) version V 2.00.
1.2 What is the Halogen Moisture Analyzer used for
Your Halogen Moisture Analyzer is used for determining the moisture content of almost any substance. The instrument works on the thermogravimetric principle. At the start of the measurement the Moisture Analyzer determines the weight of the sample, the sample is then quickly heated by the integral halogen heating module and the moisture vaporizes. During the drying process the instrument continually measures the weight of the sample and displays the reduction in moisture. Once drying has been completed, the moisture or solids content of your sample is displayed as the final result.
Of decisive importance in practice is the rate of heating and even heating of the surface of the sample. In comparison with conventional infrared heating or the drying oven method, for example, the halogen heating module of your instrument needs a shorter time to reach its maximum heating power. It also allows use of high temperatures, an additional factor in shortening the drying time. Uniform heating of the sample material ensures good repeatability of the drying results and makes it possible to use a smaller amount of sample.
All parameters of a measurement (drying temperature, drying time, etc.) can be preselected. But your Moisture Analyzer offers many other possibilities. To avoid exceeding the scope of this introduction, only a few of these are listed here:
– The built-in database contains predefined drying methods for many products. You can easily adjust the parameters
of the individual methods to suit your needs.
The drying process can be adjusted to suit the type of sample
You can choose between various types of result display
Your settings and the measurement results can be recorded
Your settings are protected even if there is a power failure
The RS232C data interface fitted as standard allows your moisture analyzer to communicate with an external
printer or with a computer.
7Getting to know your Moisture Analyzer
This wealth of functions notwithstanding, your Moisture Analyzer is very simple to operate. The status display (User Guide) guides you step by step through the measurement cycle and you always know which particular stage in a measurement is currently being executed by the instrument and the next operating step. To exclude faulty handling in routine work, the keypad can be locked to prevent access to all but the elementary functions.
The Moisture Analyzer conforms with all common standards and directives. It supports stipulations, work techniques and result records as demanded by all international quality assurance systems, e.g. GLP (Good Laboratory Practice), GMP (Good Manufacturing Practice). The instrument has a EC declaration of conformity and METTLER TOLEDO as the manufacturer has been awarded ISO 9001 and ISO 14001 certification. This provides you with the assurance that your capital investment is protected in the long term by a high product quality and a comprehensive service package (repairs, maintenance, servicing, adjustment service).
1.3 Safety has priority
Your Moisture Analyzer employs state of the art technology and meets the latest demands regarding instrument safety. This notwithstanding, improper operation can endanger personnel and cause damage to tangibles. For safe and dependable operation, please comply with the following instructions:
– The Moisture Analyzer is used for determining the moisture in samples. Please use the instrument
exclusively for this purpose. Any other type of use can endanger personnel and damage the instrument or other tangibles.
– The Moisture Analyzer must not be operated in a hazardous environment and only under the
ambient conditions specified in these instructions.
– The Moisture Analyzer may be operated only by trained personnel who are familiar with the properties
of the samples used and with the handling of the instrument.
– Your Moisture Analyzer is supplied with a 3-pin power cable with an equipment grounding con-
ductor. Only extension cables which meet the relevant standards and also have an equipment grounding conductor may be used. Intentional disconnection of the equipment grounding conductor is prohibited.
Note: There are two versions of the instrument, 110 VAC and 230 VAC.
The Halogen Moisture Analyzer works with heat!
– Ensure sufficient free space around the instrument to avoid heat accumulation and overheating
(approx. 1 m free space above the instrument).
– The vent over the sample must never be covered, plugged, taped over or tampered with in any
other way. This also applies to the fan on the back of the instrument.
– Do not place any combustible materials on, under or next to the instrument when it is connected
to the power supply, since the area around the heating module becomes hot.
– Exercise caution when removing the sample. The sample itself, the sample chamber, the draft
shield, and sample container (if used) may still be very hot.
– Be very careful if you open the heating module during operation, because the circular heating
module or its protective glass can be as hot as 400 °C! For this reason, only ever touch the handles provided on the heating module.
– No modifications must be made within the heating module. It is particularly dangerous to bend
any components or remove them, or to make any other changes.
8 Getting to know your Moisture Analyzer
Certain samples require special care!
With certain types of samples, there is a possibility of danger to personnel or damage to tangibles through:
Fire or explosion
Flammable or explosive substances
Substances containing solvents
Substances which evolve flammable or explosive gases or vapors when heated.
With such samples, work at a drying temperature that is low enough to prevent the formation of
flames or an explosion and wear protective goggles. Should there be any uncertainty regarding the flammability of a sample, always work with a small amounts of sample (max. 1 gram). In such cases, never leave the instrument unattended! In cases of doubt, perform a careful risk analysis.
Poisoning, burning
– Substances which contain toxic or caustic components. Such substances may be dried only in a
fume cupboard.
Corrosion
– Substances which evolve corrosive vapors when heated (e.g. acids). In the case of such substances,
we advise you to work with small amounts of sample as the vapor can condense on cooler housing parts and cause corrosion.
Please note that the user always takes responsibility and assumes liability for damage caused by use of the types of samples mentioned above!
Never make any modifications or constructional alterations to the instrument and use only original
spare parts and accessories from METTLER TOLEDO.
– Your Moisture Analyzer is a rugged precision instrument – but you should still treat it carefully; it
will then thank you with many years of trouble-free operation.
– Please comply with all notes and instructions in these operating instructions. Keep the instructions
in a safe place where they are immediately to hand if any points are unclear. If you lose these instructions, please contact your METTLER TOLEDO dealer for an immediate replacement.
Moisture determination applications must be optimized and validated by the user according to local regulations. Application-specific data provided by METTLER TOLEDO is intended for guidance only.
9Getting to know your Moisture Analyzer
1.4 Important information about these instructions
These instructions guide you step by step through the operation of your Moisture Analyzer. The first two sections help you put the instrument into operation quickly, safely and properly and perform your first measurement within a short space of time. In sections 3 through 5 you become intimately acquainted with the wide range of functions of your Moisture Analyzer. During this learning phase, you will find the table of contents and the detailed index in section 9 a valuable orientation aid. Sections 6 through 8 contain additional information on the maintenance of your instrument, troubleshooting and the available options. As soon as you are familiar with your Moisture Analyzer, you will find the illustrations at the beginning of these instructions and the associated references (in the key) useful for quick access.
The following identifications and symbols are used in these instructions:
Key designations are shown enclosed by twin angle brackets «», e.g. «On/Off» or «Ú».
This symbol indicates safety and hazard instructions. If these are not complied with, injury to the
user, damage to your instrument or other tangibles and malfunctions can result.
– This symbol indicates additional information and directions which facilitate your handling of the
instrument and contribute to proper and economical use.
These instructions are also available in foreign languages. Should you require a set of instructions in a different language, please contact your METTLER TOLEDO dealer.
1.5 Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements. Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related. Thank you for your contribution to environmental protection.
1.6 Finding More Information
Internet http://www.mt.com/moisture
10 Your first measurement in next to no time
2 Your first measurement in next to no time
In the section you will learn how to put your new Moisture Analyzer into operation and obtain measurement results within a very short space of time.
2.1 Unpacking and checking the standard equipment
Open the package and remove the instrument and the accessories. Check the completeness of the delivery. The following accessories are part of the standard equipment of your new Moisture Analyzer:
80 aluminum sample pans
1 sample pan support
1 sample pan handler
1 specimen sample (circular, absorbent glass fiber filter)
1 draft shield
1 power cable
1 Operating instructions
1 Application brochure "Guide to Moisture Analysis"
1 List of predefined methods
1 EC declaration of conformity
Remove the wrapping from the instrument.
Check the instrument for transport damage. Immediately inform your METTLER TOLEDO dealer if you have any complaints or parts are missing.
Store all parts of the packaging. This packaging guarantees the best possible protection for the transport of your instrument.
11Your first measurement in next to no time
2.2 Selecting the location
Your Moisture Analyzer is a precision instrument. An optimum location guarantees accuracy and dependability:
2.3 Setting up, leveling and connecting to power supply
Exact horizontal positioning and stable installation are prerequisites for repeatable results. To compensate for small irregularities or inclinations (±2%) at the location, the instrument can be leveled.
Firm, horizontal location as free from vibrations as possible
Avoid direct sunlight
No excessive temperature fluctuations
No powerful drafts
Surroundings as free from dust as possible
Enough space around the instrument to prevent build-up of heat
Sufficient distance from heat-sensitive materials in the vicinity of the
instrument.
For exact horizontal positioning, the Moisture Analyzer has a level indicator (level) and 2 leveling screws. When the air bubble in the level indicator is exactly in the center, the instrument is standing perfectly horizontally. To level it, proceed as follows:
Position your Moisture Analyzer at the selected location.
Turn the two leveling screws until the air bubble is in the center of the level indicator.
Note: The instrument should be releveled each time its location is changed.
For protection against theft, your Halogen Moisture Analyzer is fitted with a socket for an antitheft device. Together with a lockable steel cable like those used to protect portable computers against theft, the socket can be used to secure the instrument. You can obtain a suitable antitheft cable from your METTLER TOLEDO dealer.
12 Your first measurement in next to no time
Connect the instrument to the power supply.
Warning If the power cable supplied is not long enough, use only a 3-pin extension cable with equipment grounding conductor!
Allow your instrument to warm up for 60 minutes. The instrument adapts
itself to the ambient conditions during this time.
For technical reasons, the halogen heating module is designed specifically for a particular line voltage (110 VAC or 230 VAC). A heating module is installed in the factory that is matched to the particular line voltage of the country of destination. If you are not sure whether the heating module built into your instrument is suitable for your local line voltage, check the voltage printed on the glass of the circular halogen lamp before you connect the Moisture Analyzer to the power supply!
Connection to a line voltage that is too high can lead to blowing of the fuses, whereas a supply voltage that is too low will prolong the drying process.
Please note that some parts of your instrument are always live when the instrument is connected to the power supply.
First lay the heat shield in the sample chamber. Then place the draft shield (can only be mounted in one position) and after that the sample pan support in position. Turn the sample pan support until it engages. In the engaged position the rear arm of the sample pan support lies exactly along the length of the instrument.
13Your first measurement in next to no time
2.4 Your first measurement
After you have successfully put your new Moisture Analyzer into operation for the first time, you can immediately perform your first measurement. In doing so, you will become familiar with the instrument and the status display and at the same time perform a function check.
Please use the specimen sample supplied for your first measurement. This sample is an absorbent glass fiber filter.
For your first measurement the instrument is set to the factory settings and operates in the automatic mode, which guides you through the entire measurement procedure without you having to press a single key. You work with the predefined method "A", which specifies all of the parameters for the drying process.
Press the «On/Off» key to switch the instrument on.
After it has been switched on, the instrument performs a self-test. Wait until the display shown opposite appears.
Your Moisture Analyzer has a graphical status display (User Guide) which you will become acquainted with in this section. The status display informs you continuously about the current status of the instrument and prompts you to execute the next operating step (shown flashing).
Initial status with heating module closed.
Open the heating module.
Status: Ready for taring
The status display flashes and prompts you to load the empty sample pan.
Place the empty sample pan in the sample pan handler (this is possible without tilting the sample pan if you insert this in the pan handler from the side directly below the round flange). Place the sample pan handler in the draft shield. Ensure that the tongue of the pan handler fits exactly in the slot of the draft shield. The sample pan must lie flat in the pan holder.
Note: We advise you to work with the sample pan handler at all times. The pan handler is ergonomic, safe and provides protection against possible burns due to the hot sample pan.
A Substance A
– – – °C
0.000 g
14 Your first measurement in next to no time
Close the heating module. This automatically sets the balance installed in the Moisture Analyzer to zero.
Status: Ready for weighing
After taring, the status display flashes and prompts you to place the sample in the sample pan.
The display shows the target weight for the sample that has been specified for the corresponding method (in this example for "Substance A"). Rather than having any effect on the measurement process, this target weight just serves as a weighing-in guide. To obtain the best possible reproducibility of results, the sample quantity weighed in should lie within the target weight tolerances.
Stability detector: The circle symbol to the left of the weight display repre­sents the stability detector. It is only visible while the weighing result has yet to stabilize. As soon as it disappears the weighing result is stable (see also sections 7.1 and 7.2).
Open the heating module.
Place the provided specimen sample in the sample pan.
Wet the specimen sample with a few drops of water, so that the displayed weight is at least 0.5 grams (required minimum weight of sample).
Note: The drying process cannot start until the minimum sample weight has been reached.
A Substance A
2.6 ... 3.5 g
0.000 g
+
15Your first measurement in next to no time
Status: Ready for start
The status display shows that the minimum sample weight has been reached and the instrument is ready for the drying process to start.
Close the heating module and the instrument then automatically begins the drying and measuring process.
Status: Drying and measurement
You can follow the drying and measurement process on the display:
The status display uses rising bubbles to symbolize the drying process.
The current temperature in the heating module is displayed (above 50 °C)
as well as the elapsed drying time and the current drying value.
– If there is an external printer connected to your instrument, the measurement
results will be printed out continually (at intervals of 30 seconds).
When drying is complete, an audio signal sounds and the bottom section of the display is shown in inverse (white font on dark background).
You can now read off the moisture content of your sample.
The star symbol to the left of the measurement result indicates that the result is a calculated result.
Open the heating module.
Carefully remove the sample pan handler from the sample chamber.
A Substance A
88 °C
-12.48
2:54
%
MC
A Substance A
105 °C 4:48
-25.63
%
MC
16 Your first measurement in next to no time
Warning: As the pan and sample may still be hot, you should let these cool down before removing the pan from the handler!
The Moisture Analyzer warns you that it is still hot: take note of the temperature display, which is continuously updated after the heating is switched off at the end of the measurement process. The status display with the symbol opposite also shows that the instrument is still hot. This "Hot Mode" display disappears when the temperature falls below 50 °C or when you start the next cycle (in this case the status display indicates the next step).
To remove the sample pan from the handler, lift the pan slightly from below and pull it sideways out of the handler (if you no longer need the sample and the pan, you can simply tilt the handler until the pan slides out).
Press the «ß» key to delete the measurement result from the display. The instrument is now ready for the next measurement.
Congratulations!
You have just performed your first measurement with your new Moisture Analyzer. In the following section you will find important information on the operating principle of your instrument, its adjustment and optimum preparation of your samples.
17How to obtain the best results
3 How to obtain the best results
Following your first practical work with the Moisture Analyzer, in this section you will find important information on how to obtain optimum results. You will discover what parameters influence the measurement process and how you can match the instrument optimally to your particular measurement task.
3.1 Measurement principle of the Halogen Moisture Analyzer
Your instrument performs measurements based on the thermogravimetric principle, i.e. the moisture is determined from the weight loss of a sample dried by heating.
In principle, your instrument thus comprises two instruments: A precision bal­ance and a heating module. In contrast to other thermogravimetric methods (drying oven, infrared, microwave), the Halogen Moisture Analyzer operates with a halogen heating module. This ensures fast heating of the sample and thus guarantees rapid availability of the measurement results.
Irrespective of the measurement method, the quality of the measurement results stands or falls by the preparation of the sample and a correct choice of the important measurement parameters:
Sample size
Drying temperature
Switch-off mode
Drying time
Inappropriate setting of these parameters can cause the results to be incorrect or misleading. For this reason, check that the results for each type of sample are what might reasonably be expected.
You will find detailed information on the relationships between these parameters in the application brochure for moisture determination, enclosed to your instrument.
In practice, however, not only the quality of the measurement results, but also the speed of the measurement process is important. Thanks to its drying principle (with the heat generated by a halogen radiator), the Halogen Moisture Analyzer is very fast. You can increase the speed even further through optimum setting of the instrument.
The optimum drying temperature and the drying time are dependent on the nature and size of the sample and on the desired accuracy of the measurement results. These can be determined only by experiment.
ïîð pÝ
18 How to obtain the best results
3.2 Notes on adjustment of the balance and the heating module
To obtain optimal results, when you install the Moisture Analyzer at the place where it will be used, you should use the METTLER TOLEDO adjustment set to adjust the heating module under the prevailing operating conditions (e.g. in a fume cupboard). This ensures that any differences from the factory adjustment that may result from the specific local conditions are corrected.
You should adjust your Moisture Analyzer regularly (e.g. every six months) at the place where it is used and under the prevailing operating conditions. This may also be stipulated by your quality assurance system (e.g. GLP, GMP, ISO 9001).
By adjusting the heating module you can ensure comparability with results from other instruments of the same design. This is done by using an appropriate thermometer, which is provided in the form of a temperature adjustment set with calibration certificate (see section 8.7). This adjustment set comprises a black plate with a thermometer, which allows a reproducible adjustment of the heating module.
The balance and the integral heating module in your instrument can be adjusted using the appropriate accessories (section 8.7).
A Moisture Analyzer is typically used in place of or in addition to the oven method. In an oven heat energy is transferred by the flow of air, which establishes an equilibrium between the sample temperature and the ambient temperature. This is not the case in a Moisture Analyzer. The actual sample temperature primarily depends on the specific absorption properties of the sample ("dark" samples absorb more heat), which can change during the measurement process. There can also be differences between the temperature at the surface of the sample and the temperature inside the sample. The heat output is therefore not dependent on the true sample temperature but instead is regulated by a temperature sensor underneath the halogen heating module.
For the reasons explained above, the temperature setting on the instrument will therefore deviate from the true sample temperature. By regularly adjusting the heating module, you will ensure a consistent and reproducible heat output for the entire lifetime of your instrument.
Note: METTLER TOLEDO offers an adjustment service – please contact your local dealer.
If you replace the heating module, you must adjust it using the temperature adjustment set, which is available as an accessory.
The procedure for adjusting the balance and heating module is described in section 5.
19How to obtain the best results
3.3 Optimum sample preparation
Preparation of the sample is decisive for the speed of the measurement process and the quality of the measurement results.
Please note the following basic rules for the preparation of your sample:
The amount of sample you select should be as small as possible and only as large as necessary.
Excessive amounts of sample require more time for drying and thus prolong the measurement process. If the amount of sample is too small, the measurement result may possibly not be representative. It is always the case that the greater the inhomogeneity of the sample, the larger the amount of sample needed to obtain a repeatable result (please see also the information in section 8.1). Experience has shown a practical sample quantity go to be 3 – 5 g. You can use the target weights specified in the predefined methods as a reference for your own samples. Please note the supplied list of the predefined methods or look them up directly on the menu of your instrument (section 4.4.2).
Distribute the sample evenly over the sample pan.
You thus increase the surface area of the sample and facilitate heat absorption. The base of the pan should be evenly covered.
For liquid, fat-containing, melting and highly reflecting samples, you should use the glass fiber filters HA-F1 available as accessories (see section 8.7). This also applies to samples which form a skin on their surface when heated. The glass fiber filter ensures even and rapid heat distribution and prevents the formation of a skin impervious to moisture on the sample surface.
You will find further information on the moisture determination, the importance of the parameters and the preparation of the samples in the Application Brochure "Guide to Moisture Analysis" dealing with moisture determination (see section 8.3).
20 Practical application of your Moisture Analyzer
4 Practical application of your Moisture
Analyzer
In this section we introduce you to the wide range of setting possibilities of your Moisture Analyzer and offer information and tips for optimum setting of the parameters.
4.1 The operating concept and methods
You already performed your first measurement in section 2. This measurement was carried out using the factory settings. However, your instrument offers a number of setting possibilities to match the measurement process to your specific requirements.
The operating concept of your Moisture Analyzer is based on methods. A method contains all of the parameters for a drying process. These include:
The name of the method The target weight of the sample The drying program The drying temperature The switch-off criterion The type of display A free factor (optional)
Each of the two keys «A» and «B» on the keypad of your dryer can be allocated to one method. Pressing the required key loads all of the parameters of the corresponding method and drying is performed with the matching settings. The active method ("A" or "B") and its name are shown on the top line of the display.
Your Moisture Analyzer contains a database in which over 100 methods have already been predefined for different products. You can use one of the predefined methods directly, or adapt its parameters to your own requirements. You make these settings on the menu that we will introduce you to in the following sections.
21Practical application of your Moisture Analyzer
4.2 The menu
You can choose methods from the menu and change their parameters. The menu also contains other options for carrying out adjustments and for the basic settings of your Moisture Analyzer. You will familiarize yourself with these functions and settings in section 5. All menu settings are preserved even in the event of a power failure.
Press the «Menu» key to display the menu.
The menu options are shown in the top section of the display. The four fields at the bottom of the display contain the names of the "Softkeys", i.e. the keys whose function depends on the position on the menu. The four "Softkeys" are immediately underneath the display.
The option chosen from the menu is shown in inverse (white font on dark background). You can use the two arrow keys to select a different menu option by scrolling. If one of these two keys is no longer visible, you have reached the top or bottom of the list.
When you want to exit the menu, choose «Exit» (you can also exit the menu at any time by pressing the «ß» key).
4.3 Choosing and allocating methods from database
One method is allocated to each key «A» and «B» on the keypad. You can therefore quickly switch between two methods as part of the daily routine. On the menu you can choose from a database the methods that are to be al­located to the two keys.
Press the «Menu» key to display the menu.
In the following example we assume you want to allocate a method to the «A» key. Ensure that "Method A" is highlighted, then press «Sel.» to display the method database.
The database contains over 100 predefined methods, each of whose sets of parameters are designed to suit a particular product or group of products. The first four digits of the number of each method are based on the "NACE" code. "NACE" is a European Union system for classifying industrial sectors and product groups. The two digits after the dot are a consecutive numbering system. Further information on the "NACE code" and the preset parameters of the individual methods is to be found in the separate brochure supplied with your Moisture Analyzer. The two methods "Substance A" and "Substance B" contain standard parameters (basic settings).
Sel. Exit
Method A
Method B
SmartCal
Sel. Exit
Method A
Method B
Edit
1600.02 Sodium tartrat
0000.0A Substance A
0000.0B Substance B
22 Practical application of your Moisture Analyzer
Use the arrow keys to choose the required method and confirm with «=».
The next-higher menu level will then be displayed again. Press «Exit» to quit the menu.
You are now asked whether you want to save the change. Press «Yes» to save it. Choose «No» if you want to exit the menu without saving (in this case all of the changes you have made on the menu will be discarded).
If you have saved the new method allocation, the chosen method will then be made available when you press the «A» key.
If none of the predefined methods meet your requirements, you can define your own methods as described in the next Section.
If a printer is connected to your Moisture Analyzer, you can press the «Ú» key to print out all of the parameters of the two methods allocated to the «A» and «B» keys. Notes on the parameters are to be found in the next section.
Sel. Exit
Method A
Method B
A Poultry meat
– – – °C
0.000 g
Edit
1510.01 Egg yolk pow...
1510.02 Poultry meat
1510.03 Beef meat
No Yes
Save changes?
уууУЫМШСЬ РЯОЯУЫМЫОНуууу УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйиз НЙ отрр
ууууууууУЫМШСЬ Яуууууууу Т¿³» Р±«´¬®§ ³»¿¬ У»¬¸±¼ ЧЬ плпртро М¿®¹»¬ ©»·¹¸¬ нтррр ¹ Ь®§·²¹ °®±¹®¿³ НМЬ М»³°»®¿¬«®» пкрpЭ Н©·¬½¸у±ºº ³±¼» н Ь·-°´¿§ ³±¼» ыУЭ Ъ®»» º¿½¬±® Сºº
23Practical application of your Moisture Analyzer
4.4 Developing own methods and defining parameters
If none of the existing methods in the database meets your requirements, you can develop you own and allocate them to the «A» or «B» key. You can achieve this by choosing one of the existing methods and changing its parameters.
Important: Some methods are only saved until you allocate a new method to the corresponding key!
Press the «Menu» key to display the menu.
Choose the key to which the new method is to be allocated (in the opposite example the «B» key). Then press «Sel.» to display the method database.
Use the arrow keys to choose the method that is to form the basis of the new method. In order to minimize the number of changes you have to make, we recommend you choose a method whose parameters largely correspond to those of the new method to be developed.
Press «Edit» to display the parameters of the method.
Note: All of the ex works predefined methods have a number based on the "NACE code". When you change the parameters of such a method, an (un­numbered) copy of it is automatically created, with the original remaining unchanged. Copies can be changed as required.
The display shows a list of all of the parameters of the method: Name, tar­get weight, drying program, drying temperature, switch-off criterion, type of display and free factor. You will familiarize yourself with these parameters in the following sections.
Edit
Name Tomato juice
Target weight
Tomato juice
abc... A...
1 2
4.4.1 Specifying name of method
Use the arrow key to choose the parameter "Name" and press «Edit». There are two keys available for changing the name of the method:
1 Choose this option if you want to change the existing name.
2 Deletes the existing name. Choose this option if you want to enter a
completely new name.
Sel. Exit
Method A
Method B
SmartCal
Edit
1530.10 Tomato powd...
1530.11 Tomato juice
1530.12 Onion powder
24 Practical application of your Moisture Analyzer
Changing existing name
The cursor is positioned on the last character of the existing name and the «<–» key remains visible for a few seconds. If you want to delete part of the existing name, press this key immediately and delete the required characters. The «<–» key disappears after a few seconds and is replaced with the «–>» key. You can now enter the first new character: Scroll the list of characters upwards or downwards until the required character is displayed and press the «–>» key to move the cursor to the next position. The name of the method can be up to 20 characters long.
Entering new name
The existing name is deleted and the cursor is on the first character of the new name. Scroll the list of characters upwards or downwards until the required first character of the new name is displayed and press the «–>» key to move the cursor to the next position. Choose the second character from the list of characters. The name of the method can be up to 20 characters long.
As soon as you have entered all of the name of the method, press «=». The next-higher menu level will then be displayed again and you can recheck the name. In the opposite example the number "1" has been appended to the existing name.
4.4.2 Specifying target weight of sample
Use the arrow keys to choose the parameter "Target weight" and press «Edit».
Note: The target weight is just a guide to make it easier for you to weigh in the correct amount of the sample. Adherence to the target weight is not monitored. During weighing in, the target weight is displayed with a tolerance of 15% (e.g. "2.6 ... 3.5 g"). This tolerance is calculated by the Moisture Analyzer and cannot be changed by the user.
The current target weight flashes. You can use the «+» and «–» keys to change the weight in 0.5 g increments. The minimum value is 0.5 g, the maximum
54.0 g. Confirm the new target weight with «=». the next-higher menu level will then
be displayed again and you can recheck the new target weight.
Edit
Name
Target weight 3.000 g
Drying program
+
Target weight 3.500 g
d
Tomato juic e
f
%
A
B
Edit
Name Tomato juice1
Target weight
25Practical application of your Moisture Analyzer
4.4.3 Selecting the drying program
Use the arrow keys to choose the parameter "Drying program" and press «Edit».
Three different drying programs enable you to match the drying characteristic perfectly to the sample used.
Standard drying ("STD"): This drying program is suitable for most samples. The sample is heated to the drying temperature (set temperature) and held constant at this temperature.
Rapid drying: This program is primarily suitable for samples with a moisture content over 30%. Following the start, the selected temperature is exceeded by 40% for 3 minutes to compensate the cooling due to vaporization and accelerate the drying process. The drying temperature is then lowered to the set value and maintained.
LP16: This drying program corresponds to that of the earlier dryers, LP16 and LJ16. The chosen set temperature (see next section) corresponds to the final temperature of these models. If you choose this program, you cannot define any switch-off criterion (section 4.4.6), the switch-off characteristic corresponds to that of the LP16/LJ16. This program is primarily suitable for users of the earlier dryers who want to achieve a drying characteristic as similar as possible with the HB43-S.
Confirm the chosen drying program with «=». The next-higher menu level will then be displayed again and you can recheck the drying program.
Edit
Target weight
Drying program Rapid
Temperature
STD
Drying program Rapid
LP16
4.4.4 Setting the drying temperature
Use the arrow keys to choose the parameter "Temperature" and press «Edit».
The current drying temperature flashes. You can use the «+» and «–» keys to change the temperature in the range from 50 to 200 °C in 1 °C increments. Note: If you choose a drying temperature of over 160 °C, for reasons of backward compatibility with methods already developed for the Moisture Analyzer HB43, you can choose timed temperature reduction.
Confirm the new drying temperature with «=». The next-higher menu level with then be displayed again and you can recheck the new drying temperature.
Edit
Drying program
Temperature 130 °C
Switch-off mode
+
Temperature 131 °C
26 Practical application of your Moisture Analyzer
Press the «Drying temperature» key.
You can use the «+» and «–» keys to change the temperature in the range from 50 to 200 °C in 1 °C increments.
Confirm the new drying temperature with «=».
Press «Yes» to save the changes or «No» to exit the menu without saving. Note: If you do not want to save the changes, instead of the «No» you can also press the «ß» key.
4.4.5 Temperature reduction
This menu option is only available if you have chosen a drying tempera­ture of over 160 °C.
Use the arrow keys to choose the parameter "Temp. reduction" and press «Edit».
The following two settings are available.
On: This setting allows full backward compatibility with methods already developed for the HB43 with temperatures of over 160 °C. For temperatures over 160 °C temperature reduction occurs. The higher the temperature the less time it takes for the temperature to drop. The temperature is lowered continu­ously until after 1 hour it has reached 160 °C. The temperature of 160 °C is then maintained until the end of the measuring time.
Edit
Temperature
Temp. reduction On
Switch-off mode
Temp. reduction On
Off
4.4.4.1 Setting the drying temperature directly
If method "A" or method "B" is active, the drying temperature can be set directly using the «Drying temperature» key.
+
Temperature 105 °C
+
Temperature 120 °C
No Yes
Save changes?
27Practical application of your Moisture Analyzer
îðð
¬ ø³·²ò÷
ïêð
ïð ëð
Start
The diagram opposite shows the envelope of the temperature reduction. Off: The instrument heats to the chosen drying temperature (of over 160 °C) and maintains this temperature until the end of the measuring time.
Confirm the chosen setting with «=». The next-higher menu level will then be displayed again.
When working with temperatures over 180 °C we recommend that you always wait 2 - 3 minutes with open instrument lid between individual measurements, in order to ensure good reproducibility of the readings and avoid overheating of the instrument.
To protect the instrument from overheating, it has an additional protective device: A temperature switch independent of the normal black temperature sensor over the sample is triggered when it detects exceptionally high temperatures. This situation could arise, for example, if the sample begins to burn or the heating module adjustment is defective. If this happens, the heating module is tripped out by a mechanical switch, and the instrument cannot be restarted. Reset the overload switch (see section 7.2).
This menu option is only available if you have chosen "LP16" as drying program (section 4.4.3).
Use the arrow keys to choose the parameter "Switch-off mode" and press «Edit».
A switch-off criterion defines when the instrument should end the drying. Switch-off criteria save you having to keep checking your watch and stopping the drying manually. The following settings can be selected for the switch-off criterion:
"Weight loss per unit of time" (5 settings) "Free switch-off" "Manual switch-off" "Timed switch-off"
Weight loss per unit of time
This switch-off criterion is based on a weight loss per unit of time. As soon as the mean weight loss is less than a preset value during a specified time, the instrument considers drying to be complete and automatically discontinues the measurement process. During the drying, the time display shows you how long the measurement process has been in progress. The switch-off criterion is inactive during the first 30 sec. You will find additional information in section
8.4. The following 5 settings are available:
1 This setting is suitable for samples which dry very quickly (surface moisture)
or for (relatively inaccurate) fast measurements to determine a trend.
2 This setting is suitable for quick drying samples.
3 This setting is suitable for most types of samples.
Edit
Temp. reduction
Switch-off mode 3
Display mode
2
Switch-off mode 3
4
4.4.6 Selecting the switch-off criterion
28 Practical application of your Moisture Analyzer
4 This setting is suitable for samples which dry moderately quickly.
5 This setting is suitable for samples which dry very slowly (trapped moisture,
skin formation).
Free switch-off criterion
If none of the five available "Weight loss per unit of time" switch-off criteria is suitable for your application, the Moisture Analyzer allows you to define a free switch-off criterion (1 mg/ 20 s to 1 mg/180 s). The free switch-off criterion is also based on the principle of weight loss per unit of time. As soon as this drops below the preset value, the measurement is automatically ended.
If you have chosen this setting, an additional menu option for setting the time interval becomes available (see section 4.4.7).
Manual switch-off
With this switch-off criterion the measurement process continues until you stop it with the «Stop» or «ß» key.
Timed switch-off
With this switch-off criterion the measurement lasts until the preset drying time has elapsed (the time display provides you with continuous information on the drying time).
If you have chosen this setting, an additional menu option for setting the drying time becomes available (see section 4.4.8).
Confirm the chosen switch-off criterion with «=». If you have chosen the free switch-off criterion or the timed switch-off, you can now specify the time interval or the drying time as described in the following sections.
Free
Switch-off mode Manual
Timed
Manual
Switch-off mode Timed
5
Switch-off mode Free
Manual
29Practical application of your Moisture Analyzer
4.4.7 Setting time interval for free switch-off criterion
Edit
Switch-off mode
Time interval 20 s
Display mode
+
Time interval 75 s
This menu option is only available if you have chosen the free switch-off criterion (section 4.4.6).
Use the arrow keys to choose the parameter "Time interval" and press «Edit».
The current time interval to which the weight loss of 1 mg relates flashes. You can use the «+» and «–» keys to change the time interval within the range from 20 to 180 seconds in 1 second increments (Factory setting: 20 seconds).
Confirm the new time interval with «=». The next-higher menu level will then be displayed again and you can recheck the time interval.
Edit
Switch-off mode
Time 10 min
Display mode
+
Time 15 min
4.4.8 Setting drying time for timed switch-off
This menu option is only available if you have chosen the timed switch-off
(section 4.4.6) as switch-off criterion.
Use the arrow keys to choose the parameter "Time" and press «Edit».
The current drying time flashes. You can use the «+» and «–» keys to change the drying time within the range from 1 to 480 minutes in 1 minute increments.
Confirm the new drying time with «=». The next-higher menu level will then be displayed again and you can recheck the drying time.
Edit
Switch-off mode
Display mode %MC
Free factor
g
Display mode %MC
%DC
4.4.9 Selecting the display mode
Use the arrow keys to choose the parameter "Display mode" and press «Edit».
Five different types of result display are available to you. The particular one chosen also determines which values are shown on the records printed out.
You will find a description of all five types of display below.
30 Practical application of your Moisture Analyzer
g: Weight in grams The weight of the sample is displayed (and printed out) in grams. During the
drying the current weight is continuously displayed in grams.
%MC: Moisture content The moisture content of the sample is displayed (and printed out) as a
percentage of the wet weight (= WW = initial weight = 100%). This is the factory setting. When the results are printed out, the moisture content is designated "%MC" (Moisture Content) (e.g. –11.35 %MC) and shown as a negative value. During the drying the current measured value is continuously displayed in percent.
%DC: Dry content The dry content of the sample is displayed (and printed out) as a percentage
of the wet weight (= WW = initial weight = 100%). When the results are printed out, the dry content is designated "%DC" (Dry Content) (e.g. 88.65 %DC). During the drying the current measured value is continuously displayed in percent.
%AM: ATRO moisture content The moisture content of the sample is displayed (and printed out) as a percentage
of the dry weight (= DW = final weight = 100%). When the results are printed out, the ATRO moisture content is designated "%AM" (ATRO Moisture Content) (e.g. –255.33 %AM) and shown as a negative value. During the drying the current measured value is continuously displayed in percent.
%AD: ATRO dry content (Wet weight)
The wet weight of the sample is displayed (and printed out) as a percentage
of the dry weight (= DW = final weight = 100%). When the results are printed out, the ATRO dry content is designated "%AD" (ATRO Dry Content) (e.g.
312.56 %AD). During the drying the current measured value is continuously displayed in percent.
MC [0-100%] = – *
100%
Wet weight WW – Dry weight DW
Wet weight WW
DC [1000%] = *
100%
Dry weight DW
Wet weight WW
AM [0…-1000%] = *
100%
Wet weight WW – Dry weight DW
Dry weight DW
AD [1001000%] = *
100%
Wet weight WW
Dry weight DW
Before drying
After drying
ïððû
ã Åû
Ã
WW
DC
DW
After drying
Before drying
ã
ã
Åû
Ã
ïððû
WW
AM
DW
Before drying
After drying
ã
ïððû
ã ÅûÃ
WW
AD
DW
After drying
Before drying
ïððû
ã Åû
Ã
WW
DW
MC
31Practical application of your Moisture Analyzer
Comment on the ATRO display mode
If the current measured value in the ATRO display mode is greater or less than the predefined limit value (i.e. greater than 999.99 %AD or less than –999.99 %AM), a warning beep sounds and the instrument automatically switches the display mode (from %AM to %MC and from %AD to %DC). A display in the ATRO mode is no longer possible in this case, even if you have started your drying operation in the %MC, %DC or g (grams) display mode.
4.4.10 Activating free factor
Use the arrow keys to choose the parameter "Free factor" and press «Edit». The free factor is deactivated in the factory.
The free factor allows conversion of the final result of a drying process. It can be used to compensate for systematic deviations from a reference result for example. Corrected final results only appear on the records printed out (not on the display) and are listed separately and without units.
Switch the free factor on if you want to correct the final result. Confirm your setting with «=». The next-higher menu level will then be displayed again, making available two additional menu options for defining the factor (see sections 4.4.11 and 4.4.12).
4.4.11 Specifying free factor
This menu option is only available if you have activated the free factor
(section 4.4.10).
The free factor is set to 1.000 in the factory. Press «Edit» to change this setting.
The current factor flashes. You can use the «+» and «–» keys to change the factor within the range from –10.000 to + 10.000 in increments of 0.001.
Confirm the new factor with «=». The next-higher menu level will then be displayed again and you can recheck the factor.
Edit
Display mode
Free factor Off
Off
Free factor On
Edit
Free factor
Factor 1.000
Output format
+
Factor 1.048
32 Practical application of your Moisture Analyzer
4.4.12 Choosing number of decimal places
This menu option is only available if you have activated the free factor
(section 4.4.10).
The final result that has been calculated and corrected by the free factor is displayed to 2 decimal places (factory setting). Press «Edit» to change this setting.
The current setting flashes. Use the arrow keys to choose whether you want to display the final result without ("X"), with one ("X.X"), two ("X.XX") or three ("X.XXX") decimal places.
Confirm the new number of decimal places with «=». The next-higher menu level will then be displayed again and you can recheck the setting.
When choosing the output format, take account of the fact that the signifi­cance of the decimal places depends on the sample quantity weighed in and the chosen "Free factor"!
4.4.13 Saving new method and exiting menu
After you have defined all of the parameters of the new method, press the « key repeatedly until you are asked whether you want to save the changes (each keystroke accesses the next-higher menu level). Note: Instead of the «=» key you can also press the «A» key from any position on the menu. This exits the menu immediately. You will also be asked in this case whether you want to save the changes. Press «Yes» to save the changes or «No» to exit the menu without saving. Note: If you do not want to save the changes, instead of the «No» you can also press the «ß» key. If you have saved the new method, it will be available to you when you press the corresponding key («A» or «B») .
Important: Some methods are only saved until you allocate a new method to the corresponding key!
Edit
Factor
Output format X.XX
X.X
Output format X.XX
X.XXX
No Yes
Save changes?
33Practical application of your Moisture Analyzer
4.4.14 Printing out records of methods
If from any position on the method definition menu you press the «Ú» key, the settings of the method are printed out. If you have made changes that have not yet been saved, you are asked whether you want to save them (the last set of parameters saved are always printed out). The Moisture Analyzer then reverts to the normal operating state.
Note: Even outside the menu you can print out a record of the methods with the «Ú» key, provided there is no measurement being carried out.
The sample record opposite is for a method defined by the user.
4.5 Performing a measurement
You are now familiar with all the method parameters of your Moisture Analyzer and have defined your own methods for your samples. The instrument is now ready for the determination of the moisture content of your own samples. In this section you will learn how to perform measurements, print out the measurement results and stop the measurement process.
Switch the instrument on with «On/Off» key.
Use the «A» or «B» key to choose the method you want to use to dry your sample.
The status display (User Guide) symbolizes the initial status of the instrument when the heating module is closed.
The instrument is set in the factory to operate in the automatic mode and you can perform the entire drying process without pressing any keys. You will find information on the manual mode in section 5.6.8.
Open the heating module.
The status display (User Guide) now prompts you to load the empty sample pan.
уууУЫМШСЬ РЯОЯУЫМЫОНуууу УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйиз НЙ отрр
ууууууууУЫМШСЬ Юуууууууу Т¿³» М±³¿¬± ¶«·½» ³±¼·º·»¼ М¿®¹»¬ ©»·¹¸¬ нтррр ¹ Ь®§·²¹ °®±¹®¿³ О¿°·¼ М»³°»®¿¬«®» пнр pЭ Н©·¬½¸у±ºº ³±¼» Ъ®»» п ³¹ с ни ­Ь·-°´¿§ ³±¼» ыУЭ Ъ®»» º¿½¬±® С² Ъ¿½¬±® птлпк С«¬°«¬ º±®³¿¬ ИтИИИ
уууоктпрторппуууплжрзууу
34 Practical application of your Moisture Analyzer
Position the empty sample pan in the sample pan handler. Place the sample pan handler in the draft shield. Ensure that the tongue of the sample pan handler lies exactly in the slot of the draft shield. The pan must lie flat in the pan holder.
Note: We advise you to work with the sample pan handler at all times. The pan handler is ergonomic, safe and provides protection against burns due to the hot sample pan.
Close the heating module. The built-in balance is then automatically set to zero.
Following taring, the status display prompts you to add the sample to the sample pan.
Open the heating module.
Add the sample to the sample pan. Please ensure that the sample is distributed evenly to obtain good analysis results.
The minimum sample weight required is 0.5 g. Please take account of the displayed target weight for your sample when weighing in. To obtain the best possible reproducibility of results, the sample quantity weighed in should lie within the target weight tolerances.
The status display indicates that you can now start the drying process.
As soon as the sample has been put in, close the heating module and the instrument will automatically start the drying and measuring process.
You can follow the measurement process on the display: The status display uses ascending bubbles to symbolize the drying process with the following values being continuously updated and displayed:
35Practical application of your Moisture Analyzer
current temperature in the heating module
elapsed time since the start of the measurement process
current result in the chosen display mode.
Your instrument is set in the factory so that the current intermediate result is printed out every 30 seconds if a printer is connected. If you want to, you can switch off printing out of records or set manual printing (using the «Ú») key . You will find information on the settings for the printer in section 5.6. You will find an explanation of the measurement record in the next section.
As soon as the switch-off criterion stipulated when defining the method is met (or the selected drying time has elapsed), a beep is given.
You can now read the measurement off the display. You will find information on the interpretation of the measurement results in section 8.1. The result and time display remain at their final values, whereas the temperature continues to be updated.
Open the heating module.
Carefully remove the sample pan handler from the sample chamber.
Warning: Pan and sample may still be hot! Allow them to cool before you remove the pan from the handler!
B Tomato juice 1
88 °C
-12.48
2:54
%
MC
A Tomato juice 1
105 °C 4:48
-25.63
%
MC
36 Practical application of your Moisture Analyzer
Press the «ß» key to delete the final result and the time display (alternatively, you can also use the «Stop» or «–>0/T<–» keys).
If you do not wish to perform another measurement, switch the instrument off with the «On/Off» key and close the drying unit.
Stopping measurement
You can manually stop the measurement process prematurely at any time by pressing the «Stop» key. After a measuring time of at least 30 seconds, the result will be displayed from the time when the process is stopped. If you stop the measurement process sooner, or terminate it with the «ß» key or by opening the heating module, the measurement result will not be displayed, instead a message to the effect that the measurement has been stopped will appear. This message must be confirmed by pressing «OK».
37Practical application of your Moisture Analyzer
4.6 Information on the measurement record
If your instrument is connected to a printer, and if the printer has been activated from the menu, the intermediate values will be printed out at the preselected intervals and the final result recorded on completion of the measurement. The following illustration shows a sample record. The record contains the following data:
A Record title
B Manufacturer and designation of the instrument
C Serial number of the instrument
D Version number of the software
E Chosen drying method
F Indicates a method developed by the user
G Target weight of the sample (weighing-in guide)
H Selected drying program
I Drying temperature (set temperature)
J Selected switch-off criterion (or set drying time)
K Selected display mode
L Free factor activated (factor and output format specified)
M Sample weight at start of drying
N Measured value at each print interval (the number of recorded measured
values depends on the selected print interval and the measurement time)
O Total time of drying
P Indication of dry weight (final weight)
Q Final result in the selected display mode
R Converted final result (only if the free factor has been activated)
S Line for entry of the sample ID
T Line for entry for comments
U Line for signature
V Date and time at the end of the measurement
Note: You will find information on the units of the results (%MC, %DC, %AM and %AD) in section 4.4.8.
Special events are recorded on the measurement record as follows:
You have stopped a measuring cycle prematurely after a measuring time of at least 30 seconds manually with the «Stop» key. The final result is printed out, but the record is specially labeled to indicate that measurement was aborted.
You have stopped a current measurement cycle after less than 30 seconds with the «Stop» key or in some other way (using the «ß» key or opening the heating module). No final result is printed out and the record is labeled as shown opposite.
A B C D E F G H I J K L
M
N
O P Q R
S
T
U
V
уУСЧНМЛОЫ ЬЫМЫОУЧТЯМЧСТу УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйи НЙ отрр Т¿³» М±³¿¬± ¶«·½» п ³±¼·º·»¼ М¿®¹»¬ ©»·¹¸¬ нтррр ¹ Ь®§·²¹ °®±¹®¿³ НМЬ М»³°»®¿¬«®» пнр pЭ Н©·¬½¸у±ºº ³±¼» н Ь·-°´¿§ ³±¼» ыУЭ Ъ®»» º¿½¬±® С² Ъ¿½¬±® птлпк С«¬°«¬ º±®³¿¬ ИтИИИ
Ч²·¬·¿´ ©¬т рткмл ¹
ржнр ³·² уктрл ыУЭ пжрр ³·² упмток ыУЭ пжнр ³·² уоотро ыУЭ ожрр ³·² уозтнр ыУЭ ожнр ³·² унктмн ыУЭ нжрр ³·² умнтпр ыУЭ нжнр ³·² умктио ыУЭ мжрр ³·² умйтпн ыУЭ
М±¬¿´ ¬·³» мжрз ³·² Ь®§ ©»·¹¸¬ ртнмп ¹ О»-«´¬ умйтпн ыУЭ Я¼¶т ª¿´«» уйптммз
Н¿³°´» ЧЬж ттттттттттттт
Э±³³»²¬ж ттттттттттттттт
Н·¹²¿¬«®»ж ттттттттттттт
уууоктпрторппууупкжмрууу
ввввввввУЯТЛЯФдддддддддд уууоктпрторппууупижрзууу
ввввввввввЯЮСОМддддддддд уууоктпрторппууупйжлиууу
38 Basic settings of the instrument
5 Basic settings of the instrument
In this section you will learn how to define the settings your instrument uses for operation. You will also learn how to adjust the built-in balance and the heating module. These settings and functions are available on the menu with which you have already become familiar in section 4.
5.1 Menu overview
Press the «Menu» key to call up the menu. The following menu options are available:
You have already become familiar with the two menu options for allocating methods to the «A» and «B» keys in section 4.
Press the arrow key to scroll the list of menu options.
With the two menu options "Weight adjust." and "Temperature adj." you can adjust the integrated balance and the heating module (sections 5.3 and
5.4).
Note: The number shows the next adjustment. In the opposite example the integrated balance had not yet been adjusted.
The menu option "Language" allows you to specify the language in which the Moisture Analyzer is to communicate with you (section 5.5).
The "System settings" contain basic settings for Moisture Analyzer operation (for example, date and time, pressure interval, etc.). You will find further information in section 5.6.
The menu option "Interface" allows you to specify the communication param­eters of the built-in RS232C interface (section 5.7).
The option "Factory reset" resets most of the menu settings to the factory settings (section 5.8).
Sel. Exit
Method A
Method B
Start Exit
SmartCal
Weight adjust. #1
Temperature adj.
Edit Exit
Temperature adj.
Language English
System settings
Sel. Exit
System settings
Interface
Factory reset
39Basic settings of the instrument
If you have chosen a menu option, by pressing the «Ú» key you can print out a corresponding record of the current settings. If you have made changes that have not yet been saved, you are asked whether you want to save them (the last set of parameters saved are always printed out). The Moisture Analyzer then reverts to the normal operating state.
The opposite example shows a sample record with the language, system and interface settings.
The following sections will acquaint you with the individual menu items. The order corresponds exactly to that in the menu.
ууууФЧНМ СЪ НЫММЧТЩНуууу УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйиз НЙ отрр Ф¿²¹«¿¹» Ы²¹´·-¸ Н§-¬»³ -»¬¬·²¹­ М·³» º±®³¿¬ омжУУ Ь¿¬» º±®³¿¬ ЬЬтУУтЗЗ Э±²¬®¿-¬ лр ы Ю®·¹¸¬²»-- зр ы У»²« °®±¬»½¬·±² Сºº Н¬¿®¬ ³±¼» Я«¬± Р®·²¬±«¬ С² Р®·²¬ ·²¬»®ª¿´ нр ­Ч²¬»®º¿½» Ю¿«¼®¿¬» зкрр Ю·¬ с Р¿®·¬§ исТ± Ш¿²¼-¸¿µ» И±²сИ±ºº
уууоктпрторппуууплжойууу
5.2 Testing the moisture analyzer with SmartCal
Proper operation of the halogen moisture analyzer can be quickly verified with a SmartCal test. SmartCal is a granular test material with a defined moisture value after a specified time at a defined drying temperature. SmartCal is available for four test temperatures. There are control limits for each test temperature. If the measured and normalized moisture value is within the control limits, the instrument has passed the functional test. If the value is outside the control limits, there may be a problem with the instrument or the test conditions may not have been met. Additional information on using SmartCal is available at www.mt.com/smartcal.
Choose the menu option "SmartCal" and press «Sel.».
Press «Edit». A list with the four test temperatures appears on the display.
Use the arrow keys to select the desired test temperature, then press «=».
Edit Exit
70 °C
Temperature 100 °C
130 °C
Sel. Exit
Method B
SmartCal
Weight adjust.
130 °C
Temperature 160 °C
40 Basic settings of the instrument
You will be prompted to place the sample pan in the instrument. Open the heating module. Position the empty sample pan in the sample pan handler. Close the heating module. The built-in balance is then automatically set to zero.
The instrument measures the tare weight.
Open the heating module. Distribute the sample evenly in the sample pan. After you close the heating module, the instrument automatically starts the drying process.
The instrument records the starting weight and starts drying.
After drying is finished, you can read the measured moisture content of the test material (not normalized). Confirm the result with «OK».
Enter the displayed moisture reading in the measurement record. Also enter the room temperature and relative humidity in the measurement record.
Place empty pan Close heating unit
Distribute SmartCal Close heating unit
SmartCal
- - - - -
- - - °C
SmartCal
160 °C
-10.95
10:00
%
MC
SmartCal
87 °C
3.77
0:25
%
MC
Result -10.95 %MC (not normalized)
OK
Record ambient
cond. °C and %rH
OK
41Basic settings of the instrument
Using the ambient conditions, normalize the displayed moisture reading and check the normalized value against the control limits. The limit values can be found in the SmartCal Short Operating Instructions (order number 30005792) or at www.mt.com/smartcal.
End the test with «OK». The moisture analyzer is now ready for other mea­surements.
If a printer is connected to your instrument and it is switched on (see section
5.6.9), a test record is printed out automatically after completion of the test. The following data is included in this record:
A Record title B Manufacturer and designation of the instrument C Serial number of the instrument D Version number of the software
E Chosen drying method F Target weight of the sample (weighing-in guide) G Selected drying program H Test temperature (set temperature) I Selected switch-off criterion K Duration of the test measurement L Selected display mode M Free factor deactivated N Sample weight at start of drying O Total time of drying P Indication of dry weight (final weight) Q Final result R Line for entry of the sample ID S Line for entry of the ambient temperature T Line for entry of the relative humidity U Line for entry of the normalized moisture reading V Line for signature W Date and time at the end of the measurement
5.3 Adjusting the balance
With this menu item you can adjust the balance of your instrument. Consult section 3.2 to learn when an adjustment is necessary. Before selecting the adjustment function, ensure that the sample pan is in position. Following a drying cycle, you should wait at least 30 minutes before undertaking an adjustment.
Choose the menu option "Weight adjust." and press «Start».
Start Exit
SmartCal
Weight adjust. #1
Temperature adj.
Normalize result
OK
уУСЧНМЛОЫ ЬЫМЫОУЧТЯМЧСТу
УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйи НЙ отрр Т¿³» Н³¿®¬Э¿´
Ì¿®¹»¬ ©»·¹¸¬ итлрр ¹ Ü®§·²¹ °®±¹®¿³ ÍÌÜ
Ì»³°»®¿¬«®» ïêð pÝ
Í©·¬½¸ó±ºº ³±¼» Ì·³»¼
Ì·³» ïð ³·²
Ü·-°´¿§ ³±¼» ûÓÝ Ú®»» º¿½¬±® Ѻº
ײ·¬·¿´ ©¬ò иткпм ¹
̱¬¿´ ¬·³» пржрр ³·² Ü®§ ©»·¹¸¬т йткйр ¹ Î»-«´¬ ø²±¬ ²±®³ò÷
упртзл ыУЭ
Н¿³°´» ЧЬж ттттттттттттт
Ы²ª·®±²³»²¬ ½±²¼·¬·±²-ж М»³°»®¿¬«®»тттттттттттpЭ
λ´ò ¸«³·¼·¬§òòòòтттты®Ш Ò±®³ò ®»-«´¬òтттттттыУЭТ
Н·¹²¿¬«®»ж ттттттттттттт
уууоктпрторппууупржноууу
A B C D E F G H I K L M
N
O P Q
R
S T U
V
W
42 Basic settings of the instrument
You are prompted to put the requested adjustment weight of 20 grams on the pan (the adjustment weight is available as accessories, see section 8.7). Place the requested adjustment weight in the middle of the sample pan.
Note: You can use the «ß» key to stop weight adjustment at any time.
After a short time you will be asked to remove the adjustment weight. Remove the weight from the sample pan.
The instrument will provide brief confirmation of the completion of adjustment and then return to the normal operating state.
The Moisture Analyzer is now ready for further measurements.
If you have connected a printer to your instrument and this is activated (see section 5.6.9), on completion of the balance adjustment an adjustment record will be automatically printed out with the following data:
A Record title B Manufacturer and designation of the instrument C Serial number of the instrument D Version number of the software E Sequence number of the balance adjustment F Date of the balance adjustment G Time of the balance adjustment H Line for entry of the number of the adjustment weight used I Adjustment weight used J Confirmation that the balance adjustment has been performed correctly K Field for signature of the person who performed the balance adjustment
Please enter the number of the adjustment weight used, sign the record and store it in a safe place. This assures traceability, one of the basic requirements of every quality assurance system.
Balance adjustment #1
0.000 g
Please remove weight
Balance adjustment #1
- - - - -
Balance adjustment done
уууЮЯФЯТЭЫ ЯЬЦЛНМУЫТМууу УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйи НЙ отрр
Я¼¶«-¬³»²¬ ²«³¾»® п Я¼¶т ¼¿¬» октпрторпп Я¼¶т ¬·³» пзжрл
Й»·¹¸¬ ЧЬж ттттттттттттт
Й»·¹¸¬ ортррр ¹
Û¨¬»®²¿´ ¿¼¶ò ¼±²»
Н·¹²¿¬«®»ж ттттттттттттт
ууууууууууЫТЬууууууууууу
A B C D
E F G
H
I
J
K
Balance adjustment #1
20.000 g
Please put weight on pan
43Basic settings of the instrument
5.4 Adjusting the heating module
With this menu item you can adjust the temperature control of the heating module. Consult section 3.2 to learn when an adjustment of the heating module is necessary. We advise you to wait 30 minutes after a drying cycle before performing the adjustment.
Choose the menu option "Temperature adj." and press «Start».
You will be prompted to remove the pan holder from the sample chamber.
Remove the pan support.
You will be prompted to insert the adjustment set.
Note: Your can press «C» to stop temperature adjustment at any time.
Insert the temperature adjustment set in the sample pan handler.
Place the sample pan handler in the sample chamber.
Note: The temperature adjustment set is available as accessories (see section
8.7).
Start Exit
Weight adjust.
Temperature adj. #1
Language
C
0:00
Dryer unit adj. #1
Insert adjustment kit
Dryer unit adj. #1
C
0:00
Remove pan holder
44 Basic settings of the instrument
Close the heating module to start the adjustment process, and the heating module adjustment will begin.
The heating module is heated to a temperature of 100 °C. You can follow this process on the temperature display. The instrument now waits 15 minutes until the temperature adjustment set shows the correct temperature, a beep is then given.
Read the temperature adjustment set through the inspection window of the heating module (the divisions correspond to 2 °C). Enter this temperature using the «+» and «–» keys. After entering the value, press «=». The temperature must be entered within 10 minutes after the start of the beep, otherwise the adjustment process will be terminated and an error message output.
As this adjustment is a two-point adjustment (adjustment of the temperature is defined by two points, namely 100 °C and 160 °C), the heating module now heats to the second temperature (160 °C) . Proceed exactly as you did for the first temperature. After you have confirmed your entry with «=», the adjustment is at an end and the instrument automatically quits the menu and returns to the standby mode.
When the adjustment is complete, you can open the heating module and remove the sample pan support and temperature adjustment set.
Warning: As the temperature adjustment set can still be hot, allow them to cool down before you remove them from the handler.
15:00 100 °C
+
Dryer unit adj. #1
Enter actual temp. 101 °C
C
22:44 160 °C
Dryer unit adj. #1
Heating to 160 °C...
C
5:35 100 °C
Heating to 100 °C...
Dryer unit adj. #1
45Basic settings of the instrument
If you have connected a printer to your instrument and activated it, (see section
5.6.9), on completion of the heating module adjustment an adjustment record will be automatically printed out with the following data:
A Record title B Manufacturer and designation of the instrument C Serial number of the instrument D Version number of the software E Sequence number of the heating module adjustment F Line for entry of the number of the temperature adjustment set used (serial
number of temperature adjustment set is printed on dial)
G Date of the heating module adjustment H Time of the heating module adjustment I Target and actual temperature for the first adjustment point J Target and actual temperature for the second adjustment point K Confirmation of the adjustment L Field for signature of the person who performed the heating module adjust-
ment.
Enter the number of the adjustment disk, sign the record and store it in a safe place. This assures traceability, one of the basic requirements of every quality assurance system.
5.5 Selecting the dialog language
You can choose the language in which your instrument is to communicate with you.
Choose the menu option "Language" and press «Edit».
The following languages are available:
English German ("Deutsch") French ("Français") Italian ("Italiano") Spanish ("Español") Portuguese ("Portuguese") Japanese ("Nihongo") Russian ("Russian")
Select the desired dialog language and then confirm your entry with «=».
Note: The instrument is preset to the language of the country of destina tion.
Edit Exit
Temperature adj.
Language English
System settings
уЬОЗЫО ЛТЧМ ЯЬЦЛНМУЫТМуу УЫММФЫО МСФЫЬС ШЮмнуН НТО понмлкйи НЙ отрр
ß¼¶«-¬³»²¬ ²«³¾»® ï
Ì»³°ò ®»º»®»²½» ×Üæ òòòò
Я¼¶т ¼¿¬» октпрторпп Я¼¶т ¬·³» оржоз М»³°т прр pЭж прп pЭ М»³°т пкр pЭж плз pЭ
Ü®§»® «²·¬ ¿¼¶ò ¼±²»
Н·¹²¿¬«®»ж ттттттттттттт
ууууууууууЫТЬууууууууууу
A B C D
E
F
G H I J
K
L
Language English
Deutsch
46 Basic settings of the instrument
5.6.2 Entering the date
Entering the date is only necessary the first time you put your instrument into operation. The current date is printed out on each report.
5.6 System settings
You use this menu option to specify the basic settings for your instrument.
5.6.1 Setting the time
Your Moisture Analyzer is fitted with a built-in clock. The current time and time (see next section) are printed out on each report. When you put your new instrument into operation for the first time (and at the changeover from summer to winter time), you should enter the current time, this setting is retained even if you disconnect your instrument from the power supply.
From the System settings menu choose the option "Time" and press «Edit».
You can use the «+» and «–» keys to set the time (in the chosen time format, see section 5.6.3). Note: Pressing the key and holding it down adjusts the time in larger increments.
Confirm the new time with «=». The next-higher menu level will then be displayed again.
Sel. Exit
Language
System settings
Interface
Time 16.43
Date
Edit
+
Time 17:12
+
Date 12.01.2007
Choose the menu option "System settings" and press «Sel.».
The display shows a list of all of the system settings. You will familiarize yourself with these settings in the following sections.
To change a setting, choose it with the arrow keys then press «Edit». With «=» you return to the next-higher menu level.
From the System settings menu choose the option "Date" and press «Edit».
You can use the «+» and «–» keys to set the date (in the chosen date format, see section 5.6.4). Note: Pressing the key and holding it down adjusts the date in larger increments.
Confirm the new date with «=». The next-higher menu level will then be displayed again.
47Basic settings of the instrument
From the System settings menu choose the option "Time format" and press «Edit». Use the arrow keys to choose the required time format:
Setting Display (example) 24:MM (factory setting) 16:42 12:MM 4:42 PM
24.MM 16.42
12.MM 4.42 PM
Confirm the new time format with «=». The next-higher menu level will then be displayed again.
5.6.4 Choosing date format
You can choose between the European (day.month.year = factory setting) and the American date format (month/ day/year).
From the System settings menu choose the option "Date format" and press «Edit».
Use the arrow keys to choose the required date format ("DD.MM.YY" = day. month.year, "MM/DD/YY" = month/day/year).
Confirm the new time format with «=». The next-higher menu level will then be displayed again.
5.6.5 Setting display contrast
You can adjust the contrast of the display of your instrument to suit your needs.
From the System settings menu choose the option "Contrast" and press «Edit».
You can use the «+» and «–» keys to set the contrast (range: 0 – 100). Confirm the new setting with «=». The next-higher menu level will then be
displayed again.
24:MM
Time format 12:MM
24.MM
5.6.3 Choosing time format
You can choose between different ways of displaying the time.
5.6.6 Setting display brightness
You can adjust the brightness of the display of your instrument to suit your needs.
From the System settings menu choose the option "Brightness" and press «Edit».
You can use the «+» and «–» keys to set the brightness (range: 0 – 100). Confirm the new setting with «=». The next-higher menu level will then be
displayed again.
+
Contrast 50
+
Brightness 90
Date format DD.MM.YY
MM/DD/YY
48 Basic settings of the instrument
5.6.7 How to activate the protection of menu settings
You can protect the menu settings against being changed unintentionally. The instrument is set in the factory to allow the menu settings to be changed (menu protection deactivated).
From the System settings menu choose the option "Menu protection" and press «Edit».
If you wish to protect the settings against change, switch Menu protection on ("On").
Confirm the setting with «=». The next-higher menu level will then be displayed again.
After Menu protection has been switched on, all of the menu settings are locked and the display shows the padlock symbol (instead of the «Edit» key). You can now only change the Menu protection setting.
5.6.8 Selecting the start mode for drying
You can select whether drying should be started automatically or manually. In the factory your instrument is already set to automatic drying. In automatic start mode the instrument tares, determines the initial (wet) weight and begins drying when the heating module is closed. This "SmartStart" function enables you to perform the entire drying process without pressing a single key. In manual start mode drying does not start automatically when the heating module is closed. When the «Start» key is pressed the initial (wet) weight, which is important for determining the moisture content, is recorded. This can be done with the heating module open or closed. Drying only begins when the heating module is subsequently closed. We recommend manual start mode for samples which contain highly volatile substances. In manual start mode you have time for further preparation of the sample (e.g. mixing the sample with silica sand, or distributing it evenly) because weight lost by evaporation during the preparation time is already recorded. In manual start mode you can even open the heating module while drying is taking place. If this is done, the drying process is not terminated (as it is in automatic mode) but only the heating is switched off until the heating module is closed again.
From the System settings menu choose the option "Start mode" and press «Edit».
Use the arrow key to choose the required operating mode. Confirm the setting with «=». The next-higher menu level will then be displayed
again.
Menu protection Off
On
Start mode Auto
Manual
49Basic settings of the instrument
5.6.9 Switching the printer on or off
You can switch the external printer on or off. In the factory this item is switched on.
From the System settings menu choose the option "Printout" and press «Edit».
Switch the printer on or off with the arrow key. Confirm the setting with «=». The next-higher menu level will then be displayed
again.
Note: The printer must be switched off if you want to connect the HB43-S to a computer via the serial interface. This prevents the HB43-S from sending characters to the computer in an uncontrolled manner. It permits data exchange to take place without interference using the MT-SICS interface commands (section 8.6).
When you switch the printer on, you must ensure that the interface settings on the menu of the instrument (section 5.7) match those of the printer.
When you have switched the printer on, an additional menu option is available, which will be explained in the following section.
Off
Printout On
5.6.10 Defining the print interval
This setting option is only available if the printer is activated (see section 5.6.9). The print interval determines
the printout frequency of the intermediate results of an ongoing measurement.
From the System settings menu choose the option "Print interval" and press «Edit».
Use the arrow keys to choose the required print interval. The three settings "30s" (factory setting), "1min" and "5min" allow automatic printout of the intermediate results at fixed intervals. This allows you to follow the drying process using the printed record. With the "Manual" setting there is no automatic printout. However, you can print out intermediate results at any time with the «Ú» key.
Confirm the setting with «=». The next-higher menu level will then be displayed again.
30s
Print interval 1min
5min
50 Basic settings of the instrument
5.7 Interface settings
A separate menu option is available for setting the communication parameters of the built-in serial RS232C inter­face.
Choose the menu option "Interface" and press «Sel.».
The display shows a list of all of the interface settings. You will familiarize yourself with these settings in the following sections.
To change a setting, choose it with the arrow keys then press «Edit». With «=» you return to the next-higher menu level.
Sel. Exit
System settings
Interface
Factory reset
Baudrate 9600
Bit / Parity
Edit
5.7.1 Setting the baud rate (data transmission speed)
The data transmission speed (baud rate) determines speed of transmission across the serial interface. The unit is the baud (1 baud (bd) = 1 bit/second).
From the Interface settings menu choose the option "Baud rate" and press «Edit».
Use the arrow keys to choose the required data transmission speed. The following settings are available: 1200 bd, 2400 bd (factory setting), 9600 bd, and 19200 bd.
Confirm the setting with «=». The next-higher menu level will then be displayed again.
1200
Baudrate 2400
9600
7/No
Bit / Parity 7/Even
7/Odd
5.7.2 Setting bit / parity
With this menu item you can set the character format for transmission across the serial interface.
From the Interface settings menu choose the option "Bit / Parity" and press «Edit».
Use the arrow keys to choose the required setting. The following settings are available:
Setting Effect
7/No 7 data bits, no parity 7/Even (factory setting) 7 data bits, even parity 7/Odd 7 data bits, odd parity 8/No 8 data bits, no parity
Note: For correct representation of special characters (umlauts, accents, etc.) on a printer you should use the setting "8/No".
Confirm the setting with «=». The next-higher menu level will then be displayed again.
51Basic settings of the instrument
5.7.3 Setting the data flow control ("Handshake")
With this menu option you can set data flow control ("Handshake").
From the Interface settings menu choose the option "Hand shake" and press «Edit».
Use the arrow keys to choose the required setting. The following settings are available:
Setting Effect
Off (factory setting) No data flow control Xon/Xoff Data flow control via software RTS/CTS Data flow control via hardware
Confirm the setting with «=». The next-higher menu level will then be displayed again.
5.8 Resetting to the factory settings
You have the option of resetting the basic settings of the instrument to the factory settings.
Choose the menu option "Factory reset" and press «Start». You are asked whether you really want to activate the factory settings.
Warning: If you answer the question with "Yes" all of the individual settings will be replaced with the factory settings! However, time, date, dialog language, contrast, brightness and method parameters will not be reset.
You can deliberately reset the method parameters by allocating a method from the method database to the corresponding method key.
After being reset to the factory settings, the instrument automatically returns from the menu to the normal operating state.
Handshake Off
Xon/Xoff
Start Exit
Interface
Factory reset
52 Servicing and replacing individual parts
6 Servicing and replacing individual parts
In this section you will learn how to keep your Moisture Analyzer in good condition and how to replace expendable parts.
6.1 Cleaning the heating module and the sample chamber
To obtain precise measurement results, we recommend you to clean the temperature sensor and the protective glass of the halogen lamp regularly. Please note the following directions for cleaning your instrument:
Disconnect the instrument from the power supply before cleaning.
Open the heating module.
The draft shield, sample pan support and heat shield can be removed for cleaning without tools.
To dismantle the sample chamber you need a normal, commercially available screwdriver.
Use a lint-free cloth for cleaning.
Clean the exterior of the instrument with a mild cleaning agent. Although the housing is extremely rugged and resistant to solvents, never use abrasive cleaning agents or solvents!
Ensure that no liquid enters the interior of the instrument.
53Servicing and replacing individual parts
To dismantle the heating module the four screws (1) must be unscrewed. A screwdriver is needed for this purpose (Torx T-20).
Following this, the cover (2) of the heating module can be pulled off toward the back.
The gold-plated reflector (3) and the protective glass of the halogen lamp (4) can be pulled upward and out of the guide rails of the heating module.
Clean the reflector and the protective glass of the halogen lamp with a mild cleaning agent.
Carefully remove any deposits from the black temperature sensor (5).
Avoid touching the round halogen lamp. If you have to remove any splashes, deposits, or spots of grease from the halogen lamp, it is advisable to use a weak organic solvent such as, for example ethanol.
After the temperature sensor has been cleaned, we recommend you adjust the heating module (see section 5.4).
6.2 Replacing the power line fuse
If the display of your instrument remains "dark" after switching on, this is most probably because the instrument's power line fuse is defective.
To change the fuse, proceed as follows:
Disconnect the instrument from the power supply.
The power line fuse is located on the back of the instrument. Using a screwdriver, turn the fuse holder anticlockwise and pull it out of the instrument.
Check the condition of the fuse. Replace a blown fuse with a fuse of the same type and with the same rated value (5 x 20 mm, T6.3 H 250 V)
.
Do not use a fuse of a different type or rated value, or short out (bridge) the fuse, because this can put your safety at risk and damage the instrument!
54 Servicing and replacing individual parts
6.3 Connecting to an external printer
To prepare the printer for use, proceed as follows:
Disconnect the instrument from the power supply.
Connect the printer. The HB43-S is fitted with a 9-pin (f) socket for this purpose. The METTLER TOLEDO RS-P42 and RS-P26 plain paper printers are provided with suitable connection cables.
Connect the instrument and the printer to the power supply.
The following settings are recommended for optimum function (Moisture Analyzer and printer):
Baud rate: 9600
Bits/parity: 8/No
Handshake: Xon/Xoff
Make sure the parameters of the printer and the HB43-S are set to the same values (see sections 5.7.1 and 5.7.2, and the printer’s operating instructions).
If your printer has several character sets (for example, RS-P26), choose the IBM/DOS set.
55If problems arise on occasion
7 If problems arise on occasion
In this section you will learn how errors can appear during operation of your Moisture Analyzer and how you can rectify these errors.
7.1 Error messages
Your instrument distinguishes between the three different types of errors explained in the following sections.
7.1.1 Input errors
If you press a key that is not in use (for example, the «Ú» key with the printer switched off) or whose use is not permitted in the current operating state, your instrument will indicate this input error with a short beep.
7.1.2 Application errors
An application error occurs when the instrument cannot perform a procedure or a corrective action is necessary as the limits of a value range have been violated or because a general operating error exists. Application errors are reported by your instrument with a beep. Furthermore, the message "E" ("Error") appears in the display followed by the error number and error message text. Before you can continue working, you must delete the error message with «OK» or with the «ß» key. You will find a list of all application errors below.
Display Cause Remedy
E1 Weight unstable
No stability during taring or ad­justment
Ensure stable ambient conditions and an optimum location. Also take care that no part of the sample or the sample pan touch the draft shield or the sample pan handler. Highly volatile substances in the sample also prevent a stable weighing result being detected
E2 Wrong adjustment weight
Either no weight, or the wrong weight, has been placed on the sample pan during adjustment. This message is also displayed if you do not remove the weight when prompted to do so by your instrument.
Repeat the adjustment process and load the required adjustment weight
E3 Sample weight too low
Sample weight below 0.5 g Weigh in a sample of al least 0.5 g
56 If problems arise on occasion
Display Cause Remedy
E4 Missing tare weight
Sample pan has not been tared before the start of drying
Tare sample pan
E8 Temp. entry missing
During adjustment of the heating module, the wait time of 10 min­utes for entry of the temperature value has been exceeded
Repeat heating module adjustment and enter temperature values before elapse of the wait time
7.1.3 Function errors
Function errors indicate that a program or hardware error exists. In such a case, disconnect the instrument from the power supply. Should the error reappear after reconnection to the power supply, contact your METTLER TOLEDO dealer to arrange an appointment for diagnosis and repair. Note the error number in order to facilitate the work of the service engineer.
Display Cause Remedy
E9 Function error + code
Undefined condition Pull out the power supply plug and allow the
instrument to cool for at least 5 minutes. If this error occurs repeatedly, please contact your METTLER TOLEDO dealer and inform him of the displayed code
E11 Adjustment data lost
The heating module and balance adjustment data, and the method and menu settings, have been lost. Cause: Power interruption while being stored
Readjust heating module and balance and reset.
57If problems arise on occasion
7.2 What if...?
the display remains "dark" after switching on?
no line voltage power cable not connected blown power line fuse instrument faulty
Ensure that the instrument is connected to the power supply and that power is actually supplied. Check the power line fuse of the instrument and replace if necessary (see section 6.2). If the instrument still refuses to function, contact your METTLER TOLEDO dealer.
after switching on "0.000" flashes on the display?
The sample pan holder is not installed. Install the sample pan holder.
the symbol of the stability detector is continuously lit up immediately after the start?
As soon as the symbol of the stability detector fades, the weighing result is stable and is accepted as a "wet weight". If the symbol does not fade, your instrument is probably at an unsuitable location (vibrations, shocks, powerful drafts, etc.). Seek a more suitable location.
Samples containing readily volatile substances may never reach stability owing to continuous evaporation. In In this case you must select manual operating mode for the drying unit (see section 5.6.8).
the printer that is connected does not print?
Ensure that the printer is activated on the menu (see section 5.6.9) and that the ribbon and paper are correctly installed. Check the transmission rate, bit/parity and handshake (see section 5.7).
incorrect characters are printed?
Change the bit/parity setting of your printer and HB43-S to "8/No". Check to see if both instruments have the same baud rate setting and use the same character set (see sections 5.7 and 6.3).
no drying time can be entered in the menu?
Entry of the drying time is possible only if you have selected the switch-off criterion "Timed switch-off", with all other switch-off criteria this input possibility is not available (see sections 4.4.6. and 4.4.8).
Menu settings cannot be changed?
You have protected the menu settings against change (see section 5.6.7).
the measurement takes too long?
You have selected an unsuitable switch-off criterion (see section 4.4.6).
An excessive amount of sample can also be the cause of slow drying, likewise samples which tend to form a skin which hinders vaporization.
Perform experiment at higher temperature.
58 If problems arise on occasion
the instrument does not heat following the start?
The heating module is overheated and the thermal overload protection has responded. The instrument is equipped with thermal overload protection (bimetallic sensor) which switches off the heating element if overheating occurs. This can be caused, for example, by a burning sensor.
Note: If this is not possible, there may be a fault (e.g. a faulty halogen heating module). In such a case, contact your METTLER TOLEDO dealer.
After the temperature switch is reset or a heating module replaced (section 6.1) we recommend that you readjust the heating module of your Halogen Moisture Analyzer (section 5.4).
the measurement results are not repeatable?
– The samples are not homogeneous, i.e. they have different compositions. The more inhomogeneous a sample,
the larger the amount of sample needed to obtain a repeatable result.
– You have selected a drying time that is too short (for the "Timed switch-off" criterion). Extend the drying time or
select a suitable switch-off criterion "Weight loss per unit of time".
– The sample does not become completely dry (e.g. owing to skin formation). Dry the sample with the aid of glass
fiber discs (see section 3.3).
You have selected a temperature that is too high and the sample has oxidized. Lower the drying temperature. The sample boils and the splashed drops continuously change the weight. Lower the drying temperature. Insufficient heating power because the protective glass of the halogen lamp is dirty. Clean the protective glass of
the halogen lamp (see section 6.1).
– The temperature sensor is contaminated or faulty. Clean the temperature sensor (see section 6.1) or have it re-
placed by a service engineer.
The support on which the instrument is standing is not sufficiently stable. Use a stable support. The surroundings are very unstable (vibrations etc.).
Reset the overload protector Warning: The instrument may still be hot!
1. Disconnect the instrument from the power supply.
2. Use a screwdriver to remove the cover.
3. Use a finger to reset the overload switch.
4. Replace the cover and reconnect the instrument to the power supply.
59Further useful information
8 Further useful information
8.1 Notes on interpretation of the measurement results and the ideal sample weight
The accuracy of the measurement results depends on the wet weight and the original moisture of the sample. The relative accuracy of the measurement result improves with increasing wet weight. While the moisture of the sample is fixed, the weight of the sample can frequently be determined by the user. However, with increasing weight the drying process is lengthened. We thus advise you to select the weight of your sample to obtain the repeatability
you require.
The target weights specified in the predefined methods (section 4) generally lead to adequate repeatability.
You will find further information in the application brochure (section 8.3).
8.2 RS232C interface
So that your Moisture Analyzer can be connected to a peripheral device (e.g. printer or computer) it is fitted with an RS232C interface as standard.
9-pin socket
To adapt the HB43-S to the peripheral device the interface parameters can be set on the menu (see section 5.7).
The many ways in which the HB43-S Moisture Analyzer can document results can only be fully utilized if a printer is connected, for example the METTLER TOLEDO RS-P26 or RS-P42. The printed results are a decisive contribution to simplifying working procedures in accordance with GLP/GMP.
For interface commands please refer to the "Reference Manual METTLER TOLEDO Standard Interface Command Set" 11781024, available from your METTLER TOLEDO dealer or downloadable from the Internet (www.mt.com/ moisture). More Information is found in section 8.6.
ïë
ç ê
묆
ÜßÌß
̨Ü
×Ò
ÍÙÒÜ
ÑËÌ
uШЯТЬНШЯХЫe
8.3 Application Brochure
The application brochure for moisture determination from METTLER TOLEDO (order number: 11796096, www. moisture-guide.com) contains a great deal of useful information for optimum utilization of your Moisture Analyzer. One brochure is included with your HB43-S. To order additional copies, please contact your METTLER TOLEDO dealer.
Example methods and applications can be found at www.mt.com/moisture and www.mt.com/moisture-methods.
Moisture determination applications must be optimized and validated by the user according to local regulations. Application-specific data provided by METTLER TOLEDO is intended for guidance only.
60 Further useful information
¹
¹
¹
¹
¬
ø
÷
¬
øïð -÷
¬
øïìð -÷
¬
øëð -÷
Weight of sample
Switch-off criterion 1 Switch-off criterion 3 Switch-off criterion 5 Time
Weight of sample
equal
area
=
= Mean weight loss per unit of time
8.4 Switch-off criterion "Weight loss per unit of time"
With the switch-off criterion "Weight loss per unit of time", drying is automatically ended as soon as the mean weight
factory, each with a fixed weight loss per unit of time. Furthermore, the switch-off criterion "Free" allows you to define the weight loss per unit of time yourself.
The following applies to the individually selectable levels:
Switch-off criterion 1 1 mg 10 seconds
Switch-off criterion 2 1 mg 20 seconds
Switch-off criterion 3 1 mg 50 seconds
Switch-off criterion 4 1 mg 90 seconds
Switch-off criterion 5 1 mg 140 seconds
Switch-off criterion "Free" 1 mg 20 seconds to 180 seconds
The graph below exemplifies the mode of operation of the switch-off operation (not to scale).
Key
Switch-off criterion 1 (rapid availability of the result, suitable for determination of a trend)
Switch-off criterion 3 (standard setting)
Switch-off criterion 5 (suitable for precision measurement)
61Further useful information
8.5 Technical data
Please note that the Moisture Analyzer will undergo continuous further development in the interest of the users. METTLER TOLEDO thus reserves the right to change all technical data at any time and without prior notification.
Dryer
Heating module: Halogen ring-shaped radiator
Temperature range: 50–200 °C
Temperature increment: 1 °C
Heating module adjustment: with temperature adjustment set HA-TC or HA-TCC
Balance
Minimum sample weight: 0.5 g
Maximum sample weight: 54 g
Balance adjustment : with external weight, 20 g ± 0.1 mg
Stability detector: with symbol in display
Balance resolution: 1 mg
Result resolution: 0.01% Repeatability (sd) with 2 g sample: 0.10%
Repeatability (sd) with 10 g sample: 0.015%
Data
Time, date system clock, fail safe
Operational settings: read-only memory, fail safe
Switch-off criteria: 5 levels, manual, timed, free
Drying programs: Standard drying, fast drying, LP16 mode
Drying time: manual (unlimited), timer: 1 minute to 480 minutes
Reset protection: by locking the menu settings
Evaluation
Display modes: 5 modes: moisture content (MC), dry content (DC), weight (g),
ATRO moisture content (AM), ATRO dry content (AD)
Records: Using external printer (optional)
62 Further useful information
Hardware
Audio signal: Provided
Data interface: Built-in RS232C interface
Inspection window: in heating module
Leveling: 2 leveling screws and level indicator Display: Backlit matrix display
Status display (User Guide): integrated in display
Sample pan, ø: 90 mm
Thermal overload protection: bimetallic-element switch in heating module
Dimensions (w x h x d): 23 x 15 x 36 cm (see also following dimension diagram)
Weight, ready to measure: 4.3 kg
Admissible ambient conditions
Use only in closed rooms
Height above mean sea level: Up to 2000 m
Temperature range: 5 °C to 40 °C
Atmospheric humidity: 80% rh @ + 30 °C
Warm-up time: At least 60 minutes after connecting the instrument to the power
supply; when switched on from standby-mode, the instrument is ready for operation immediately.
Voltage fluctuations: –15%+10%
Overvoltage category: II
Pollution degree: 2
Power load: Max. 450 W during drying process
Current consumption: 4 A or 2 A, according to the heating module
Power supply voltage: 100 VAC – 120 VAC or 200 VAC – 240 VAC, 50/60 Hz
(the voltage is determined by the heating module)
Power line fuse: 1 piece, 5 x 20 mm, T6.3 H 250 V
63Further useful information
Dimensions
ÝÔßÍÍ×Ý Ð´«-
íïçïç
íêð
íçð
64 Further useful information
8.6 MT-SICS Interface commands and functions
Many of the instruments and scales used have to be capable of integration into a complex computer or data acquisition system. To enable you to integrate instruments into your system in a simple manner and utilize their capabilities to the full, most instrument functions are also available as appropriate commands via the data interface. All new METTLER TOLEDO instruments launched on the market support the standardized command set "METTLER TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depend on the functionality of the instrument.
Basic information on data interchange with the instrument
The instrument receives commands from the system and acknowledges the command with an appropriate response.
Command formats
Commands sent to the instrument comprise one or more characters of the ASCII character set. Here, the following must be noted:
a space (ASCII 32 dec., in this description represented as ñ).
RLF (ASCII 13 dec., 10 dec.).
The characters CRLF, which can be inputted using the Enter or Return key of most entry keypads, are not listed in this description, but it is essential they be included for communication with the instrument.
Example: S – Send stable weight value
Command Í Send the current stable net weight value.
Response НсНсЙ»·¹¸¬К¿´«»сЛ²·¬
Current stable weight value.
Íñ× Command not executable (instrument is currently executing another command, e.g. taring, or timeout as stability was not reached).
Íñõ Instrument in overload range.
Íñó Instrument in underload range.
Example
Command Í Send a stable weight value.
Response НсНссссснртрррс¹ The current, stable weight value is 30.000 g.
For the MT-SICS commands available for your instrument and further information please refer to the Reference Manual "MT-SICS for Moisture Analyzers 11781024" downloadable from the Internet under www.mt.com/moisture.
65Further useful information
8.7 Accessories, consumables and spare parts
If you require other spare parts for your Moisture Analyzer, please contact your METTLER TOLEDO dealer.
8.7.1 Accessories
Description Part No. Adjustment weight
Adjustment weight 20 g (Class F1 (Adjustment of balance)
00158640
Temperature adjustment set
Temperature adjustment set, HA-TC (Adjustment of heating module)
00214455 Calibrated temperature adjustment set, HA-TCC (including test certificate) (Adjustment of heating module))
00214528
Recalibration from HA-TCC, HA-TCCRe (including test certificate) (Recalibration with certificate)
00214534
Sample pan
Reusable sample pan (steel, height: 6 mm), HA-DR1, Set of 3 pieces
00214462
Reusable sample pan (steel, height: 15 mm), 1 piece 00013954
Transport case 11113855 Sample pan handler
Sample pan handler, HA-PH, 3 pieces 00214526
Protective cover
Chemically resistant protective cover, 2 pieces
11113363
Printer
Printer RS-P42 Printer RS-P26
00229265
12120788
Cables for RS232C Interface
RS9 – RS9 (m/f): Connection cable for PC or printer RS232C, length = 1 m 11101051
Antitheft device
Steel cord
11600361
66 Further useful information
Description Part No. SmartCal
cSmartCal, 24 tests 30005791 cSmartCal, 12 tests 30005793 SmartCal, 24 tests 30005790 SmartCal, 12 tests 30005792
StarterPac SmartCal
StarterPac cSmartCal, 12 tests 30005918 StarterPac SmartCal, 12 tests 30005917
8.7.2 Consumables
Description Part No. Printer paper
Printer paper, Set of 5 rolls 00072456 Paper roll (self-adhesive) for printer, Set of 3 rolls 11600388
Ribbon
Ribbon (cassette, black) for printer, Set of 2 pieces 00065975
Sample pan
Aluminum sample pan ø 90 mm, HA-D90, Set of 80 pieces 00013865 Reinforced aluminum sample pan ø 90 mm, 1 piece 11113863
Glass fiber filter
Glass fiber filter (for liquids), HA-F1, Set of 100 pieces 00214464
67Further useful information
8.7.3 Spare parts
Pos Description Part No.
1 Sample pan handler 11113873 2 Sample pan support 00214642 3 Draft shield 00214368 4 Heat shield 11113360 5 Leveling screw 11101302
68 Index
9 Index
Data interface 62 Date 46, 61 Date format 47 Decimal places 32 Dialog language 45 Dimensions 63 Display 62 Display brightness 47 Display contrast 47 Display mode 29, 61 Disposal 9 Draft shield 34, 52, 55, 67 Dry content 30 Drying 48 Drying and measurement 15 Drying process 59 Drying program 25, 61 Drying tempera-
ture 17, 25, 26, 58
Drying
time 17, 28, 29, 57, 58, 61
E
Equipment 10 Error message 44, 55 Error message text 55 Error number 55 Explosion 8 Extension cable 12
F
Factory settings 51 Fire 8 Free factor 31 Free space 7 Free switch-off 27 Free switch-off criterion 28 Function errors 56 Fuse 53 Fuse holder 53
A
Accessories 65 Accuracy 59 Adjusting the balance 41 Adjusting the heating module 43 Adjustment 18, 55 Adjustment process 44 Adjustment record 42, 45 Adjustment weight 42, 55, 65 Air bubble 11 Antitheft device 65 Application Brochure 19, 59 Application error 55 ATRO dry content 30 ATRO moisture content 30 Audio signal 62
B
Balance adjustment 42 Baud rate 50, 54, 57 Beep 44 Bimetallic sensor 58 Bit / parity 50, 57 Burning 8
C
Cables for RS232C Interface 65 Character format 50 Chemically resistant protective
cover 65 Cleaning 52 Cleaning agent 52, 53 Communication parameters 50 Computer 49, 59 Connection cable 54 Consumables 66 Corrosion 8 Current consumption 62
D
Database 20 Data flow control 51
G
Glass fiber filter 13, 19, 66 GLP 7, 18, 59 GMP 7, 18, 59 Good Laboratory Practice 7 Good Manufacturing Practice 7 Guide rails 53
H
Halogen heating module 6 Halogen lamp 53 Halogen Moisture Analyzer 6, 17 Handshake 51, 57 Heat 7 Heat accumulation 7 Heating element 58 Heating module 6, 7, 12, 13,
14, 15, 17, 18, 27, 33, 34, 35, 43, 44, 48, 53, 58, 61
Heating module adjustment 27,
56, 61
Heat shield 52, 67
I
Initial status 13 Input errors 55 Inspection window 44, 62 Instrument safety 7 Interface 49, 50 Interface commands 64 ISO 9001 7, 18 ISO 14001 7
L
Language 45 Level 11 Level indicator 11 Leveling 11, 62 Leveling screw 67 Leveling screws 11 Line voltage 12, 57 Liquid 52
69Index
LJ16 25 Location 11, 55, 57 LP16 25
M
Manual switch-off 27, 28 Measure-
ment 10, 13, 20, 33, 57 Measurement parameters 17 Measurement principle 17 Measurement process 36 Measurement record 35, 37 Measurement result 59 Measurement results 35 Measuring cycle 37 Menu 38, 57, 59 Menu overview 38 Method 20, 23 Moisture 6, 17, 59 Moisture content 15, 25, 30 Moisture determination 59 MT-SICS 59, 64 MT-SICS interface commands 49
N
NACE code 21, 23
O
On/Off 33 Operating concept 20 Overheating 7, 58
P
Package 10 Packaging 10 Padlock symbol 48 Pan holder 13, 34, 43 Paper 57 Paper roll 66 Parameter 33 Peripheral device 59 Personnel qualification 6 Poisoning 8 Power cable 7, 12, 57
Power line fuse 53, 57, 62 Power supply 11, 12, 52, 56 Power supply voltage 62 Printed results 59 Printer 45, 49, 54, 57, 59,
61, 65 Printer paper 66 Printer RS-P26 54, 59 Printer RS-P42 54, 59 Print interval 49 Printout 49 Protective glass 7, 52, 53, 58 Protect the settings 48
Q
Quality assurance system 18,
42, 45
R
Rapid drying 25 Ready for start 15 Ready for taring 13 Ready for weighing 14 Reconnection to the power
supply 56 Record 37, 41, 42, 45, 61 Reference Manual 59 Reflector 53 Repeatability 59, 61 Reset 16, 36, 55 Reset protection 61 Resetting the thermal overload
protector 58 Result display 29 Reusable sample pan 65 Ribbon 57, 66 RS232C interface 50, 59
S
Safety 7 Sample 13, 19, 28, 33, 48,
55, 58, 59 Sample chamber 35, 43, 52 Sample pan 13, 19, 34, 55,
62, 65, 66
Sample pan handler 13, 34, 35,
43, 55, 65, 67 Sample pan support 52, 67 Sample weight 55, 59, 61 Self-test 13 Servicing 52 Setting up 11 Signal 55 SmartCal 39, 66 SmartStart 48 Softkeys 21 Solvent 52, 53 Spare parts 67 Specimen sample 13, 14 Stability 55 Stability detector 57, 61 Standard equipment 10 Start 48, 57, 58 StarterPac SmartCal 66 Start mode 48 Status display 7, 13, 33, 62 Stop 28, 36, 37 Stopping measurement 36 Switch-off criterion 27, 28, 57,
60, 61 Switch-off criterion "Weight loss
per unit of time" 58 Symbol of the stability detector 57 System settings 46
T
Target weight 24 Taring 55 Technical data 61 Temperature 35, 44, 57, 58 Temperature adjustment set 18,
43, 44, 61, 65 Temperature entry 56 Temperature increment 61 Temperature range 61 Temperature reduction 26 Temperature sensor 27, 52, 53,
58 Test record 41
Thermal overload protection 58 Thermal overload switch 58 Thermogravimetric principle 6 Time 46, 61 Timed switch-off 27, 28, 57 Time format 47 Time interval 29 Tolerance 24 Trained personnel 7 Transport case 65 Two-point adjustment 44
U
Unpacking 10 User Guide 13, 33, 62
V
Vapors 8 Vent 7
W
Weight 55 Weight in grams 30 Weight loss per unit of
time 27, 60
Leer
Mettler-Toledo AG Laboratory & Weighing Technologies
CH-8606 Greifensee, Switzerland Tel. +41 (0)44 944 22 11 Fax +41 (0)44 944 30 60 Internet: www.mt.com
Subject to technical changes. © Mettler-Toledo AG 12/2011 11780961A en
öïïéèðçêïö
GWP
®
– Good Weighing Practice
The global weighing guideline GWP® reduces risks associated with your weighing processes and helps to
www.mt.com/GWP