Mettler Toledo College B303, College B502, College B154, College B1302 DeltaRange, College B3001 Operating Instructions Manual

...
Operating instructions
METTLER TOLEDO B balance lines
•College-B
•CB
•GB
Short-form operating instructions
display appears Automatic procedure
Simple weighing
g
0.00
g
1182.03
g
1250.00
Taring
g
=00
g
=00
g
12+01
#
g
9%97
Calibration (adjusting)
Unit switching
Switching off
Switching on
CAL
------
2000.00
------
0.00
------
CAL donE
Cal/Menu
0.00
g
88888888
g
0.00
On
------
OFF
Off
On
#
ct
110.00
g
22.00
dwt
14.14
1
4.2.3 Unit 1, 2, 3 – selecting unit 1, 2, 3 with CB and GB balances ......14
4.2.4 A. oFF – setting automatic shutdown..........................................14
4.2.5 Selecting peripheral device........................................................15
4.2.6 Selecting data transfer mode .....................................................15
4.2.7 Selecting data transfer format ....................................................15
4.2.8 Setting baud rate (data transmission rate) ..................................16
4.2.9 Setting parity/bits .....................................................................16
4.2.10 Setting handshake ...................................................................16
5 Technical data and optional equipment ...................................17
5.1 Technical data.........................................................................17
5.2 Interface..................................................................................19
5.3 Optional equipment..................................................................19
6 Appendix .............................................................................. 20
6.1 Printout examples with LC-P45 and GA42 Printers ......................20
6.2 Preventive maintenance............................................................21
6.3 What if...?...............................................................................22
6.4 Index......................................................................................23
Contents
1 Getting to know the B balance lines ..........................................2
1.1 Introducing the B balance lines.................................................... 2
1.2 Layout of the B balances ............................................................3
1.3 The keys of the B balances (overview) .........................................4
2 Startup ....................................................................................5
2.1 Unpacking / Standard equipment.................................................5
2.2 Cautionary notes........................................................................ 5
2.3 Setting up .................................................................................6
2.4 Calibration (adjusting) ...............................................................8
3 Weighing ................................................................................9
3.1 On/off switching......................................................................... 9
3.2 Simple weighing........................................................................9
3.3 Taring..................................................................................... 10
3.4 METTLER TOLEDO DeltaRange balances.....................................10
3.5 Switching of weight units ..........................................................11
4 Menu .................................................................................... 12
4.1 Overview ................................................................................12
4.2 Description of the menu options ................................................13
4.2.1 Reset and printout of the balance setting..................................... 13
4.2.2 Adjusting the vibrations adapter on CB and GB balances..............13
2
1.1 Introducing the B balance lines
Three specially oriented B balance lines
We have derived three lines for special application fields from the standard range of the B balances: School balances (College-B), carat balances (CB) and gold balances (GB). These balances are geared to the special requirements of the respective segments with regard to type of weighing and functions.
METTLER TOLEDO DeltaRange balances have a movable fine range with 10 times smaller display increments (section 3.4).
All B balances are fitted with an RS232C interface as standard.
Notes
• Certified versions of the CB and GB balances are also available, please ask your METTLER TOLEDO dealer for details.
• If you wish to build on what you have learned about weighing in these operating instructions, you will find valuable tips in the booklet "Weighing the right way" 720906.
1 Getting to know the B balance lines
GB
College-B CB
College-B
College-B GB
3
1.2 Layout of the B balances
1 Keys 2 Display 3 Model plate
1)
4 Weighing pan/platform 5 Draft shield 6 Certification switch for weights and measures inspector and service engineer 7 Leveling feet 8 AC adapter socket
9 Provision for antitheft device 10 Leveling control 11 RS232C interface
Keys, operation and display are identical for all B balances.
1)
With details of: Max = maximum capacity
d = readability
*
Min = minimum capacity (recommended minimum load for certified balances)
*
e = verification scale interval (smallest display increment tested in certification) * Relevant only for certified balances
4
7
3 2
1
5 4
3
2 1
4
98
10
11
10 7
7
6
9
4
1.3 The keys of the B balances (overview)
B balances have two operator control levels: the weighing mode and the menu. The keys have different meanings, depending on the operator control level and how long a key is pressed.
Weighing mode (operation)
On
• Switch on
#
• Tare
• Switch
• Change settings
• Print
• Confirm settings
• Switch off
• Calibrate (adjust)
• Call menu
press and hold
press and hold key until the desired display appears
Menu (called up with
Cal
)
• Store and quit menu
press and hold
press briefly
press briefly
Off Cal
On
0/T
press key briefly
#
Off
On
Cal
Cal
#
• Abort
• Change settings
• Select menu options
Cal
5
2 Startup
2.1 Unpacking / Standard equipment
All B balances are supplied in an environmentally harmless package. The standard equipment of the B balances includes
AC adapter, to national codes,
Weighing pan,
operating instructions, to allow optimum utilization of the capabilities of your balance,
Description of interface commands (“MT-SICS Reference Manual, available in English only)
calibration weight, with CB balances and College B154 only,
draft shield, mounted with CB balances and College B154, for mounting by user with College B303,
carat pan, with CB balances.
Note
Calibration weights can be ordered from METTLER TOLEDO for all other B balances, see section 5.2.
2.2 Cautionary notes
• B balances may not be operated in hazardous areas.
1)
• Before attachment of the AC adapter, check whether the imprinted voltage value matches the local supply
voltage. If it does not, contact your local METTLER TOLEDO dealer.
1
)With the PS-EX2 power supply unit/AC adapter available as an option, all B balances can be used in
Ex zone 2 (section 5.3).
6
2.3 Setting up
The optimum location
The correct location makes an important contribution to the accuracy of the weighing results of high-resolution analytical and precision balances.
Hence, ensure a
• stable, vibration-free position as horizontal as possible. Avoid
• direct sunlight,
• excessive temperature fluctuations,
• drafts. The best position is on a stable bench in a corner protected against drafts as far as possible from doors, windows,
radiators or the ventilation slots of air conditioners.
Note
If vibrations can not be prevented, the balance can still provide accurate results if the vibration adapter is set accordingly, see section 4.2.2.
Antitheft device
B balances have provision for the attachment of an antitheft device. The following antitheft devices are available:
• Bolt and lock (not for GB16001) 229175
• Cable with lock (for all models) 590101
7
Leveling
B balances have a level control and adjustable leveling feet to compensate for slight irregularities in the weighing bench surface. The balance is exactly horizontal when the air bubble 1 is in the middle.
Procedure
Turn the two leveling feet 2 and 3 as shown in the diagram or described in the table so that air bubble 1 is in the middle. Air bubble at "12 o'clock" turn both leveling feet clockwise Air bubble at "3 o'clock" turn left leveling foot clockwise, right leveling foot counterclockwise Air bubble at "6 o'clock" turn both leveling feet counterclockwise Air bubble at "9 o'clock" turn left leveling foot counterclockwise, right leveling foot clockwise
Note
The balance must be releveled each time it is moved to a new location.
Connecting to the power supply
Before connection of the AC adapter, check that the imprinted voltage value matches the local supply voltage.
If this is not the case, please contact your local METTLER TOLEDO dealer.
Plug AC adapter into AC adapter socket of the balance and connect to the power supply.The balance performs a self-test. The test is finished when "OFF" appears.Press g briefly: balance is in operational readiness. Before any work is performed with the balance,
it should be calibrated (adjusted), (section 2.4).
Notes
To achieve accurate results with balances B154 and CB1503DR, these must be connected to the power supply for 20–30 minutes to warm up to the operating temperature. The PP-B10 PowerPack (rechargeable, external battery) can be used to operate all B balances independently of the power supply (section 4.2.4, 5.3).
1
23
Off
8
2.4 Calibration (adjusting)
To obtain accurate weighing results, the balance must be matched to the acceleration due to gravity at its location.
Calibration is necessary
• before the balance is used for the first time,
• at regular intervals during weighing operations,
• after a change in location.
Procedure
To obtain accurate results, the balance must be connected to the power supply and allowed to warm up to the operating temperature for 20–30 minutes before calibration.
Have required calibration weight ready (section 5.1).Unload weighing pan.Press and hold
Cal
until "CAL" appears in the display, release key.
The required calibration weight value flashes in the display.Place calibration weight in center of pan. The balance calibrates itself.When "0.00" flashes, remove calibration weight.
The calibration (adjustment) is finished when "0.00 g" appears in the display. The balance is again in the weighing mode and ready for operation.
Note
• With certified balances, the calibration can be disabled after installation if required by the national certification
regulations.
• The adjustment can be terminated at any time using the
#
key. The following message appears:
Abort.
CAL
------
2000.00
------
0.00
------
CAL donE
Cal/Menu
0.00
g
9
3 Weighing
3.1 On/off switching
Switching on
Remove any load from weighing pan and press
On
briefly. The balance performs a display test. When zero is displayed, the balance is ready for operation.
88888888
g
0.00
On
------
Switching off
Press and hold
#
Off
On
until "OFF" appears in the display. Release key.
3.2 Simple weighing
Place weighing sample on the weighing pan.
g
0.00
g
1182.03
g
1250.00
Wait until the stability detector " " fades.
Read off result.
OFF
Off
On
#
10
3.3 Taring
Place empty container on the balance.
The weight is displayed.
Tare: press
#
briefly.
Add weighing sample to container, the net weight is displayed.
If the container is removed from the balance, the tare weight will be shown as a negative value. The tare weight remains stored until
#
is again pressed or the balance is switched off.
Note
With METTLER TOLEDO DeltaRange balances, the fine range with its 10 times smaller display increments is again available after every taring operation.
3.4 METTLER TOLEDO DeltaRange balances
METTLER TOLEDO DeltaRange balances have a movable fine range with 10 times smaller display increments. In this range there is always an additional decimal place in the display.
The balance operates in the fine range
• after switching on,
• after every taring operation. If the fine range is exceeded (section 5.1), the balance display automatically switches to larger display
increments.
g
=00
g
=00
g
12+01
#
g
9%97
-12+01
g
On
g
651.4
g
0.00
#
11
3.5 Switching of weight units
All B balances can display weight values in different weight units.
• With school balances (College-B), the fixed units gram (g) and newton (N*) are available. * As the balance is an instrument to determine mass, when school balances are used to determine force a
small measurement error arises. It depends on the geographic location as a constant conversion factor 1 kg = 9.80665 N is used.
• With carat and gold scales (CB, GB), 3 different units can be displayed in the weighing mode. These are freely selectable in the menu. The factory setting differs according to the country and the model.
Switching between unit 1, unit 2 and unit 3
Requirement
With CB and GB scales, different units must be activated in the menu (section 4) for Unit 1, 2 and 3.
ct
110.00
g
22.00
dwt
14.14
Press again, the display switches to the 3rd weight unit.
Note
With certified balances/scales, the unit selection can be blocked following installation if required by national legislation.
Press
, the display switches to the 2nd weight unit.
^
12
4 Menu
4.1 Overview
On the menu, you can select unit 1, 2 and 3, and you can adjust the vibration adapter and the automatic shutdown. With certified balances/scales, the unit selection can be blocked following installation if required by national legislation. The factory setting differs according to the country and the model.
Entry into menu
Press and hold
Cal
until "MENU" appears in the display. Release key, the 1st
menu option "rESEt" appears.
Select menu options
Press
briefly. Press key repeatedly to view the current balance settings.
Modify settings
Press
repeatedly until the desired setting appears.
Store settings
Press and hold
Cal
until "StorEd" appears. Release key, the balance returns
to the weighing mode.
Abort
Press
#
briefly. The balance returns to the weighing mode without
storing the changes.
Note
If no entry is made within 45 seconds, the balance returns to the weighing mode without storing the changes.
1)
only with CB and GB balances
2)
With certified balances, only the weighing units/functions allowed by the respective national weights and measures legislation may be selected.
3)
These menu options are shown only if your balance is equipped with an RS232C interface.
4)
These menu options are shown only if “HoSt” has been selected in menu option 4.2.5.
5)
These menu options are shown only if “S.Stb” or “S.Cont” has been selected in menu option 4.2.6.
Unit 1
kg
g
Unit 1
m
Unit 1
mo
Unit 1
Unit 2
Unit 2
g
Unit 2
ozt
GN
tl
tl
Unit2 S
tl
Unit 2
m
Unit 2
mo
Unit2 H
Unit2 t
2
31
ÅoFF 2'
ÅoFF 5'
oz
Unit 2
ct
Unit 1
lb
Unit 2
7B-no
7B-E
8B-no
7B-odd
BD 2400
BD 4800
BD 9600
BD 600
BD 1200
BD 300
BD 150
HS OFF
HS SoFt
HS HArd
kg
Unit 2
S.STb
S.ConT
S. PŸ≈
3 Weighing unit 1
1), 2)
4 Weighing unit 2
1), 2)
2 Vibration adapter
1)
Weighing mode
Menu
6 Autom. shutdown
1 Reset
rESEt
LiST
ÅoFF -
7 Peripheral unit
3)
PòinTEò
11 Parity/bit
3)
10 Baud rate
3)
12 Handshake
3)
dwt
Unit 1
Unit 1
Unit 1
lb
Unit 1
oz
Unit 1
ozt
GN
dwt
Unit 2
Unit 2
ct
8 Send commands
4)
S.oFF
9 Send format
5)
S. SICS
Unit 1
kg
g
Unit 1
m
Unit 1
mo
Unit 1
ct
Unit 1
5 Weighing unit 3
1), 2)
dwt
Unit 1
Unit 1
Unit 1
lb
Unit 1
oz
Unit 1
ozt
GN
ÅoFF 10'
HoST
see key
3), 4), 5)
13
4.2 Description of the menu options
4.2.1 Reset and printout of the balance setting
Reset balance setting and functions to factory setting (rESEt)
Select "rESEt" and press and hold
Cal/Menu
until "r donE" is displayed. The balance is now reset to the factory setting and returns to the weighing mode. The factory setting differs according to the country and the model. _ 2 normal balance environment PrintEr Attachment to printer or host Unit 1 g bd2400 Transmission rate Unit 2 g 7b–E Character format Unit 3 g HS oFF Transmission protocol A. oFF – no automatic shutdown
Printig out the balance setting (LISt)
Select "LISt" and press and hold
Cal/Menu
until "StorEd" is displayed. The current balance setting is printed out and stored. The selection "LISt" appears only when a LocalCAN universal interface is installed (section 5.2, 6.1).
4.2.2 Adjusting the vibrations adapter on CB and GB balances
You can use the vibration adapter to match the balance to the ambient conditions.
_ 2 Setting with normal balance surroundings (factory setting).3 Setting with unstable balance surroundings. The balance operates slower but is less sensitive to
external influences (drafts, vibrations, etc.).
1 Setting with very stable balance surroundings. The balance operates very quickly but is sensitive to
external influences (drafts, vibrations, etc.).
_
2
3
1
_
2
rESEt
LISt
14
4.2.3 Unit 1, 2, 3 – selecting unit 1, 2, 3 with CB and GB balances
The following weight units can be selected. With certified balances/scales, the unit selection can be blocked following installation if required by national legislation.
Unit Conversion factor Comments
g gram kg kilogram 1 kg = 1000 g not with 1 mg balances lb pound 1 lb 453.59237 g not with 0.1 mg balances oz ounce 1 oz 28.349523125 g ozt troy ounce 1 ozt 31.1034768 g GN grain 1 GN 0.06479891 g not with 1 g balances dwt pennyweight 1 dwt 1.555173843 g ct carat 1 ct = 0.2 g mg milligram 1 mg = 0.001 g only with 0.1 mg and 1 mg balances mo momme 1 mo 3.749999953 g m mesghal 1 m ≈ 4.6083162 H tl Hong Kong taels 1 H tl 37.42900 g selectable only in unit 2 S tl Singapore taels 1 S tl 37.799366256 g selectable only in unit 2 t tl Taiwan taels 1 t tl 37.499995313 g selectable only in unit 2
4.2.4 A. oFF – setting automatic shutdown
The automatic shutdown appreciably extends the operating life in line-independent operation with the PP-B10 PowerPack. When the automatic shutdown is activated, the balance switches itself off if no weighing has been performed during the specified time. With the PowerPack the balance is then off, without the PowerPack it is on standby.
A. oFF – no automatic shutdown (factory setting) A. oFF 2 automatic shutdown after 2 minutes A. oFF 5 automatic shutdown after 5 minutes A. oFF 10 automatic shutdown after 10 minutes
Unit 2
Unit 2
Unit 2
Unit 2
g
Unit 2
kg
lb
oz
g
A. oFF
10'
A. oFF
5'
A. oFF
2'
A. oFF
15
Printer
HoSt
4.2.5 Selecting peripheral device
In this menu option you can select the desired peripheral device. The balance automatically stores the appropriate settings for every peripheral device.
Printer Attachment to a printer (e.g. METTLER TOLEDO GA42 Printer), (section 5.2).
Factory setting: bd 2400, 7b–E, HS oFF
Host Attachment to any peripheral device.
Factory setting: bd 9600, 8b–no, HS SoFt
4.2.6 Selecting data transfer mode
In this menu block you tell the balance how a value should be transferred to a peripheral device (e.g. computer). This menu option appears only if the setting “HoSt” has been selected in the menu option “Selecting peripheral device”.
S. oFF Data transfer mode switched off S. Stb The next possible stable value will be transferred after triggering of the Print/Transfer command. S. Cont All values will be automatically transferred.
4.2.7 Selecting data transfer format
With the “S. SICS” setting the data transfer formats described in MT-SICS are used. You will find the description in the enclosed description of the operating instructions (Reference manual MT-SICS).
With the “S. PM” setting the following data transfer formats of the PM balances are used. S. Stb: -----1.67890-g
S. Cont: S----1.67890-g
SD---1.39110-g
Note
If you wish to use other data transfer formats of the PM balances with your B-balance, please use the optional B-M emulation software which emulates all interface commands of the PM balances (see section 5.3).
The interface is unidirectional. Incoming interface commands in the setting “S. PM” are not processed further.
S. oFF
S. Stb
S. Cont
S. SICS
S. PŸ≈
16
4.2.8 Setting baud rate (data transmission rate)
The data transmission rate (baud rate) determines the speed of the transmission via the serial interface. The unit is baud (1 baud (bd) = 1 bit/second).
The following settings are available: 150 bd, 300 bd, 600 bd, 1200 bd, 2400 bd, 4800 bd and 9600 bd.
4.2.9 Setting parity/bits
In this menu option you can set the character format for the attached peripheral device. 7b–E 7 bits/even parity 7b–no 7 bits/no parity 8b–no 8 bits/no parity 7b–odd 7 bits/odd parity
4.2.10 Setting handshake
This menu option allows you match the data transmission to different serial receivers. HS oFF No handshake HS SoFt Software handshake (XON/XOFF) HS HArd Hardware handshake (DTR/CTS)
Note
If you have selected the setting “HS HArd”, the attached peripheral device must be switched on. If it is switched off, the balance is blocked.
4800
1200
2400
7b-odd
7b-E
7b-no
8b-no
HS HArd
HS oFF
HS SoFt
17
17
5 Technical data and optional equipment
5.1 Technical data Standard equipment
• Protective cover; with CB balances, the protective cover is part of the optional equipment
• AC adapter to national codes (Power supply: 115 V/230 V, -20/+15%, 50/60 Hz, 65/30 mA, 6 VA) (Power supply balance: 8–14.5 V, 50/60 Hz or
9.5–20 V = 1.5 VA)
• Device for weighing below the balance
With GB16001 the required hook is an option
• Carat pan with CB balances
• Calibration weight with CB balances and B154
• RS232C interface
Degree of protection
• Protects against dust and water.
• Pollution degree: 2
• Overvoltage category: II
Ambient conditions
The technical data are valid under the following ambient conditions:
• Ambient temperature 10 °C…30 °C
• Relative atmospheric humidity 15 %…85 %, noncondensing
• Height above sea level up to 4000 m
The operability is assured between ambient tem­peratures 5–40 ºC.
School balances (College-B) B154 B303 B502 B1302 B3002 B3001
DeltaRange
Readability 0.1 mg 0.001 g 0.01 g 0.01 g 0.01 g*/0.1 g 0.1 g Maximum capacity 151 g 310 g 510 g 1310 g 600 g*/3100 g 3100 g Repeatability (s) 0.1 mg 0.001 g 0.01 g 0.01 g 0.01 g*/0.1 g 0.1 g Linearity +/– 0.3 mg 0.003 g 0.02 g 0.02 g 0.02 g*/0.1 g 0.1 g Calibration weight 100 g 200 g
1)
500 g
1)
1000 g
1)
2000 g
1)
2000 g
1)
Calibration weight with certified balances
2)
2)
2)
2)
2)
2)
Overall dimensions (WxDxH) in mm 190x290x339 190x290x204 190x290x62 Weighing pan in mm ø 80 ø 100 180x166 Effective height above weighing pan 232 mm 118 mm Net weight (with packaging) 5 kg (8.1 kg) 3 kg (5 kg) 2.5 kg (4.5 kg) * Values in fine range (DeltaRange)1) Optional equipment
2)
Not available as certified version
18
Carat balances CB203 CB603 CB1503
DeltaRange
Readability 0.001 ct / 0.01 ct / 0.01 g 0.001 ct / 0.01 g 0.001 ct* / 0.01 ct / 0.01 g Maximum capacity 210 ct / 610 ct / 122 g 610 ct / 122 g 500 ct* / 1600 ct / 320 g Repeatability (s) 0.001 ct / 0.005 ct / 0.001 g 0.001 ct / 0.001 g 0.001 ct* / 0.005 ct / 0.003 g Linearity +/– 0.002 ct / 0.005 ct / 0.01 g 0.002 ct / 0.01 g 0.002 ct* / 0.005 ct / 0.01 g Calibration weight 100 g (500 ct) 100 g (500 ct) 200 g (1000 ct)
Calibration weight with certified balances 100 g (500 ct) 100 g (500 ct) 200 g (1000 ct) Overall dimensions (WxDxH) in mm 190x290x252 190x290x252 190x290x339 Weighing pan in mm ø 80 ø 80 ø 80 Effective height above weighing pan 159 159 232 Net weight (with packaging) 3.7 kg (6.5 kg) 3.7 kg (6.5 kg) 4.6 kg (7.5 kg)
Gold balances GB802 GB1302 GB2002 GB3002 GB3002 GB1501 GB3001 GB6001 GB16001
DeltaRange
Readability 0.01 g 0.01 g 0.01 g 0.01 g 0.01 g*/0.1 g 0.1 g 0.1 g 0.1 g 0.1 g Maximum capacity 810 g 1310 g 2100 g 3100 g 600 g*/3100 g 1510 g 3100 g 6100 g 16100 g Repeatability (s) 0.01 g 0.01 g 0.01 g 0.01 g 0.01 g*/0.05 g 0.05 g 0.05 g 0.05 g 0.5 g Linearity +/– 0.02 g 0.02 g 0.02 g 0.03 g 0.02 g*/0.1 g 0.1 g 0.1 g 0.1 g 0.5 g Calibration weight 500 g
1)
1000 g 1)2000 g 1)2000 g 1)2000 g
1)
1000 g 1)2000 g 1)2000 g
1)
8000 g
1)
– with certified balances 800 g 1)1300 g 1)2000 g 1)3000 g 1)3000 g
1)
1500 g 1)3000 g 1)6000 g
1)
16000 g
1)
Overall dimensions (WxDxH) in mm 190x290x62 190x290x82 381x321x92 Weighing pan in mm 180x166 175x166 349x232 Net weight (with packaging) 2.5 kg (4.5 kg) 2.7 kg (4.2 kg) 6.8 kg (8.3 kg) * Values in fine range (DeltaRange)
1)
Optional equipment
* Values in fine range (DeltaRange)
19
19
5.3 Optional equipment Calibration weights
Available as OIML weights (E1) (E2, F2 with certifi­cate) for further details, see METTLER TOLEDO weights brochure 721192
or as calibration weights (not OIML)
Draft shields
CB (all models) 224371 B154 224370 B303 224372
Note: CB and B154 draft shields can also be used
for B303 models.
Protective covers (set of 3) B154, CB1503 228183 B303, CB203, CB603 224007 B502–3001, GB1302–3001 224008 GB6001 228182 GB16001 230018
Dust covers for B154 238465 AC adapters
Euro (230 V) 228063 Euro (ground contact) (230 V) 228198 UK (240 V) 228066 US (120 V) 228064 Japan (100 V) 228065 Australia (bench version) (240 V) 228190
Power supply unit
for hazard zone 2 for all models PS-EX 2
PowerPack
Line-independent, rechargeable exter­nal power source, for 10 hours weighing operation (see section 4.5) PP-B10
Connection plate PowerPack/balance
long, for B, GB 230166 short, for GB16001 230168
Hook for weighing below the balance
needed only for GB16001 21301097
Density kits for B154, CB
• for determination of solids 33360
• for determination of liquids
with displacement body 33360+210260
Transport case
• for all models (except GB16001) and
B models (except B154) accommo­dation for balance and PowerPack 224009
• for GB16001, accommodation for
balance and PowerPack 230031
Antitheft devices
• Cable and lock (for all models) 590101
• Bolt and lock (not for GB16001) 229175
B-M emulation 21301730 Software EPROM for the use of B balances in sys­tems together with METTLER TOLEDO PM balances
RS232C interface and interface accessories
Every B balance is fitted with an RS232C interface for attachment to a peripheral device (e.g. printer or PC with a 9-pin male connector). Matching to a different device can be carried out in the menu (sections
4.2.5–4.2.8). You will find a detailed description of the available
interface commands in the enclosed brochure “Ref­erence Manual MT-SICS, available in English only).
The wide range of features of the B balances regarding docu­mentation of the results can not be exploited to the full until a printer, e.g. the GA42 or LC­P45 from METTLER TOLEDO is attached. The printed results make a decisive contribution to a simple way of working in
compliance with GLP/GMP.
Printers with normal paper
• Application printer with adjustment, statistics and multiplication function as well as time and date, 24 characters LC-P45
• Simple printer, 24 characters, without time/date GA42
Cables and cabling accessories
• RS9-RS25: (m/f), length 2 m 11101052
• RS9-RS9: (m/f), length 1 m 11101051
• RS9-RS9: (m/m), length 1 m 21250066
PG-S 16
15
96
IN
DATA
OUT
IN
GND
OUT
«HAND-SHAKE»
5.2 Interface
20
6 Appendix
6.1 Printout examples with LC-P45 and GA42 Printers
Functions triggered by balance Functions triggered by printer
1)
1)
possible only with LC-P45
Function: Calibration
-- BALANCE CALIBRATION -
Date: .................
Time: .................
METTLER TOLEDO Balance Type: CB203 SNR: 1116150017
Weight ID: .............
Weight: 500.000 ct
Ext. calibration done
Signature:
........................
--------- END ----------
Function: Verification With automatic insertion of date and time, with fixed weight (calibration weight)
----- BALANCE TEST -----
08.04.97 15:14:22
METTLER TOLEDO Balance Type: CB203 SNR: 1126203803
Weight ID: .............
Target: .............
Actual: 500,001 g
Diff: .............
External test done
Signature:
........................
--------- END ----------
Function: Statistics
08.04.97 15:18:55 ID 45.698-3 SNR: 1126203803
1 100.00 g 2 100.01 g 3 100.01 g 4 100.00 g 5 100.00 g n 5 x 100.004 g s 0.005 g srel 0.01 % min. 100.00 g max. 100.01 g dif. 0.01 g
--------- END ----------
Function: Multiplier With rounding to the nearest 5
08.04.97 15:21:50 ID 45.698-3 SNR: 1126203803
Factor 1.650
99.99 g * 164.9835
Function: List Printout of the current balance settings
--------- LIST ---------
Date: ..................
Time: ..................
METTLER TOLEDO Balance Type: CB203 SNR: 1126203803
SW-Ver.: 1.70 2.0 Vibr.: 2 Unit 1: g Unit 2: g Unit 3: g A.Off: ­Output: Printer Baud: 2400 Bit: 7 Parity: even Handshake:off
--------- END ----------
Function: Calibration With automatic insertion of date and time
-- BALANCE CALIBRATION -
08.04.97 15:13:37
METTLER TOLEDO Balance Type: CB203 SNR: 1126203803
Weight ID: .............
Weight: 500,000 ct
Ext. calibration done
Signature:
........................
--------- END ----------
21
6.2 Preventive maintenance
Servicing
Regular servicing of your balance by a service engineer extends its life. Ask your METTLER TOLEDO dealer for details of the servicing possibilities.
Cleaning
The balance housing and the weighing pan are made of high-grade, resistant materials. All usual cleaning agents can thus be used.
Note
After working with chemicals, it is advisable to wash or clean the weighing pan and the baseplate (if draft shield fitted). Although all materials are of high quality, corrosion may occur if corrosive substances are stored for a lengthy period of time (and on exclusion of air, e.g. by a coat of grease) on chrome steel.
22
6.3 What if...?
Error/error message
Cause
Overload
Underload
Wait for stability before pressing key.Ensure more stable ambient conditions.
Remove sample from weighing pan, rezero (tare).
If overload appears when balance switched on, recalibrate (readjust) balance.
Rectification
Remove sample from weighing pan, rezero.Check whether weighing pan is positioned properly.
If underload appears when balance switched on, recalibrate (readjust) balance.
No stability in taring or calibration (adjusting)
No or wrong calibration weight on pan
Place required calibration weight in center of pan.
Mount correct weighing pan.
Wrong or missing weighing pan
Abort of the adjustment using the
#
key.
Balance/scale software not sufficiently up-to-date for operation with LC-B interface.
To change the balance/scale software 299702 (Order
No. 600150), ask your METTLER TOLEDO dealer.
Error 5
Error 1
ñ------ó
ò------ô
Error 2
Abort
0
23
6.4 Index
AC adapter................................................. 5, 7
Adjustment..................................................... 8
Ambient conditions ....................................... 17
Antitheft device ............................................... 6
Automatic shutdown ..................................... 14
Battery operation .......................................... 19
Calibration Adjustment
Carat scales....................................... 2, 11, 18
Cautionary notes ............................................ 5
Certified balances ........................................... 3
Certified balances ........................................... 3
Certified version.............................................. 2
Cleaning...................................................... 21
Connecting to power supply......................... 5, 7
Connection ................................................ 5, 7
Degree of protection ...................................... 17
DeltaRange .................................................... 2
DeltaRange balances .................................... 10
Draft shield .................................................. 19
Error/error message ...................................... 22
Factory setting.............................................. 13
Fine range ................................................... 10
Gold scales........................................ 2, 11, 18
Hazard zone .................................................. 5
Interface ................................................ 16, 19
Leveling......................................................... 7
Menu ...................................................... 4, 12
Operation....................................................... 4
Optional equipment....................................... 19
Power supply ................................................. 7
Power supply units ....................................... 19
PowerPack ............................................ 14, 19
Preventive maintenance................................. 21
Printers.................................................. 19, 20
Printout examples......................................... 20
Protective covers .......................................... 19
RS232C interface.......................................... 19
School balances................................. 2, 11, 17
Service ........................................................ 21
Setting up ...................................................... 6
Stability detector ............................................. 9
Standard equipment....................................... 5
Standard equipment...................................... 17
Switching off .................................................. 9
Switching on .................................................. 9
Taring ......................................................... 10
Technical data .............................................. 17
Unstable balance environment ....................... 13
Vibration adapter .......................................... 13
Weighing....................................................... 9
Weighing mode.............................................. 4
Weighing pan ................................................ 3
Weight units........................................... 11, 14
24
Printed on 100 % chlorine-free Paper, for the sake of our environment.
Subject to technical changes and to the availability
of the accessories supplied with the instruments.