The METTLER TOLEDO DL31/DL38 titrators are microprocessor controlled analytical instruments which you
can use to perform Karl Fischer titrations simply, quickly and dependably.
You can attach a balance, the METTLER TOLEDO GA42 printer and a computer to the appropriate RS232C
interfaces as well as an EPSON- or HP-compatible printer to the Centronics interface and, at the DL38's TTLinterface, the METTLER TOLEDO oven sample changer Stromboli. The METTLER TOLEDO LabX titration
software allows you to control your titrator via a computer.
The titrators have a tutorial stored which quickly enables you to become acquainted with the operation and
method of functioning of the instrument.
The two titrators have the same operation. Differences in the hardware and software are mentioned in the
corresponding sections.
These Operating Instructions correspond to software version 2.0 or higher.
Where can you find what information?
These Operating Instructions and the Help texts integrated in the titrator provide a complete description of
the method of functioning of the two titrators. Explanations of parameters contained in the Help texts are
supplemented here when necessary.
On the Memo card you can find a short overview of menus and methods.
The Application brochure explains the theoretical fundamentals of the Karl Fischer titration, the control of
the titration in the DL31/DL38 and documents the METTLER methods stored in the DL38.
The Help texts in the titrator provide you with more information or refer you to the Operating Instructions or
the Application brochure.
You will find a detailed description of the communication between the titrator and a computer in the
Description of the computer interface.
METTLER TOLEDO DL31/DL385
Safety measures
2Safety measures
The titrators have been tested for the experiments and intended purposes documented in these Operating
Instructions. However, this does not absolve you from the responsibility of performing your own tests of the
products supplied by us regarding their suitability for the methods and purposes you intend to use them for.
You should therefore observe the following safety measures.
2.1Measures for your protection
– Ensure that you plug the power cable supplied into a receptacle outlet that is grounded! In
the absence of grounding, a technical fault could be lethal.
Risk of
electric shock
– Never work in an environment subject to explosion hazards! The housing of the instru-
ment is not gas tight (explosion hazard due to spark formation, corrosion caused by the
ingress of gases).
Risk of
explosion
– When using chemicals and solvents, comply with the instructions of the producer and
the general lab safety rules!
All Karl Fischer reagents are readily flammable and toxic.
– In the case of skin contact, immediately wash off with copious amounts of water!
– If eye contact occurs, rinse intensively with copious amounts of water and consult a
physician!
2.2Measures for operational safety
– Have the instrument serviced only by METTLER TOLEDO Service!
– Make sure that every connector is in the correct position! Do not damage the connector
pins of the magnetic stirrer cable! Pins damaged by faulty insertion can cause a short
circuit at the "Stirrer" output. This renders the titrator non-functional.
– Always wipe off splashed liquids immediately! The instrument is not water-proof.
Caution
– Exclude the following environmental influences:
• powerful vibrations,
• direct sunlight,
• atmospheric humidity greater than 80%,
• temperatures below 5 °C and above 40 °C,
• powerful electric or magnetic fields!
6METTLER TOLEDO DL31/DL38
3The titrator
Burette
The titrator
Titration stand
Titrant bottle
Arrestment knob
Pump knob
Keypad
Rear view
External stand (DV705)
Stirrer output
Electrode input (KF)
Solvent bottle
Waste bottle
Magnetic stirrer
Balance (RS232C)
GA42 printer (RS232C)
Computer (RS232C)
Stromboli, TBox, etc. (TTL I/O)
EPSON- and HP-compatible printers
(Centronics)
Pump unit
On/Off switch
Suction fitting
Output fitting
METTLER TOLEDO DL31/DL387
Socket for
external power pack
The titrator
3.1Putting into operation
– Plug the supplied external power pack into the titrator’s socket, connect to the power supply and switch
on the titrator: A self-test is performed before the name of the titrator appears (example DL38).
Hello!
F3
Arrow keys
(
)
9
8
7
]
[
4
5
6
1
2
3
CE
*
.
Del
0
Shift
Entry keys
<Reset> key
<i> key
Auxiliary function keys
Display
Reset
i
Burette
Stirrer
*
Results
Pump
METTLER TOLEDO
DL38
Version 2.0
Esc
F1
Setup
Method
F2
Run
Command keys
Menu keys
The command key <F3>, "Hello", the <i> key and all Menu and Auxiliary function keys can be activated
after the titrator has been switched on.
3.2The operating concept
– Press, e.g. the <Setup> key:
Title line
Selection bar
ResourcesSETUP
KF titrants
Br No. titrants
KF standards▼
PrintOK
Name of the key, in other words the displayed
list of resources belongs to the Setup menu
Commands
DL31: No "Br No. titrants"!
You can now also activate the arrow key (∇), the <Reset> key and the Command keys <F2>, "Print" and
<F3>, "OK".
▼The arrow in the display means that the list contains more resources. If you press the arrow
key (∇), the lines are scrolled upward, the selection bar is fixed. The commands you can
execute always refer to the line that is selected.
PrintPress <F2>: The list of KF titrants is printed out (if a printer is selected and defined).
OKPress <F3>: The list of KF titrants appears.
8METTLER TOLEDO DL31/DL38
The titrator
3.2.1The command keys
The commands which can be triggered with <F1>...<F3> change according to the selected menu item. The
following commands require an explanation:
Hello!When the instrument is delivered, the <F3> command in the initial mask "METTLER TOLEDO..."
is assigned to a Tutorial menu. This is deleted when you have performed the first titration with
the Tutorial.
Note You can call up the Tutorial at any time with the <i> key.
OKThis command is always a confirmation for what you
• wish to view,
• have viewed or entered,
• wish to adopt from a selection.
ModifyWhen this command appears a
• submenu can appear which can or must be modified,
• a selection menu appears from which you can or must accept values or names,
• an existing parameter value or name can be directly replaced.
If a value can be modified or entered only with the keypad, the "Modify" command is missing.
ABCThis command always appears when you can select letters or characters, e.g. for a name. In
this case, you can not enter numbers until you have confirmed "ABC" (press <F1>).
HelpInformation about the selected resource appears.
The <Esc> key
If you have made changes in the current menu or submenu, they are discarded, in other words
Esc
the original values/names remain in force. <Esc> always effects a return to the previous display.
The <Reset> key
The initial mask "METTLER TOLEDO..." reappears: <Reset> is used to terminate analyses or
Reset
other actions. Data that are not stored are lost.
METTLER TOLEDO DL31/DL389
The titrator
3.2.2The menu keys
In order to analyze a sample automatically, the required data must be stored. In the titrator, these data are
assigned to so-called task areas, the menus and can be accessed with the Menu keys. It is only the
coordination between these menus that makes the automatic analysis possible.
All resources required for the titration such as KF titrants and KF standards are stored in the
Setup
Setup menu and can be changed.
The sample is determined with the aid of a method. All methods necessary for the different
Method
analyses are stored in the Method menu and can be changed. All resources defined in the Setup
menu are available for these methods.
DL31: Only one method can be stored.
DL38: Fifty user methods can be stored.
The titration is performed in the Analysis menu. The method called up determines the analysis
Run
sequence.
NoteEach menu is subdivided into task areas (submenus) which are designated menus, lists or
masks in these Operating Instructions.
3.2.3The auxiliary function keys
To rinse the burette or recalculate results, the required commands are also assigned to menus. As they are
independent of a sample analysis, but can support it, we call them auxiliary functions. They are accessible
under the corresponding keys.
You can switch the stirrer on or off and change the stirring speed.
Stirrer
Pump
Burette
You can add or siphon off solvent.
You can rinse the burette and the tip of the dispensing tubing and add a specified volume.
*
Results
10METTLER TOLEDO DL31/DL38
You can view the result list of analyzed samples, perform recalculations, change the statistical
evaluation of a series (DL38 only!) and print out additional reports.
3.2.4The <i> key
The value of the current solvent capacity, the last drift determination, the last blank value
i
determination for DL38/Stromboli (METTLER method 910), the "Hello!" Tutorial, the KF
learn method and a menu for testing the titrator hardware are accessible with this key.
3.2.5The entry and arrow keys
The titrator
Shift
CE
Del
Shift
Shift
Shift
Shift
With the <Shift> key, you can activate the characters above the numbers.
CEYou delete all entered numbers/characters/letters.
DelYou delete the number/character/letter last entered.
+
7
+
4
+
+
You can move the cursor to the right or left (only if letters/characters are selected), see
example:
skips 3 lines of the display upward (page up),
skips 3 lines of the display downward (page down).
▲▼The arrow appears in the display when there are more lines to the list than are visible.
When you press the arrow keys (∇∆) next to the display, the lines are scrolled downwards or upwards. The selection bar is fixed. By pressing <Shift>, ∇ or <Shift>, ∆ the
lines are moved a page at a time. The commands that can be executed always refer to
the line highlighted by the selection bar.
METTLER TOLEDO DL31/DL3811
The titrator
3.3Changing the language
All texts in the titrator are stored in English, German, French, Spanish, Italian and Russian. If you have not
run the Tutorial and understand one of the available languages better than the one displayed, you should
select it:
Setup
ResourcesSETUP
KF titrants
Br No. titrants
KF standards▼
PrintOK
▼
ResourcesSETUP
Peripherals
Settings
OK
▼
SettingsSETUP
LanguageEnglish
User list
Date20-08-1998▼
ModifyOK
–Press the <Setup> key.
– Press the <Shift> and Arrow (∇) once, then press only the Arrow key
to select "Settings".
– Press <F3>, "OK".
– Press <F2> repeatedly until, e.g. "Español" appears.
▼
SettingsSETUP
LenguaEspañol
Lista usarios
Fecha20-08-1998 ▼
Modif.OK
12METTLER TOLEDO DL31/DL38
With "OK" or <Esc> the "Resources" mask reappears.
OKSpanish texts appear.
EscEnglish texts appear.
Setup (<Setup> key)
4Setup (<Setup> key)
In order to perform titrations, the titrator must know the titrants and their concentration. It must know the
settings of the attached instruments to transfer the data. In this menu you define and store these resources:
You set up your titrator.
When you press the <Setup> key, the list of resources appears:
The most commonly used titrants and KF standards are already stored in the titrator. You can not only delete
these or change their parameters, but also define new ones.
All titrants and KF standards needed for the METTLER methods are stored! If you delete
one of these resources, the error message that it is missing does not appear until the start
of the titration.
PrintDepending on your choice of resource, its list with the corresponding parameters will be printed
out.
OKThe list of the selected resources appears.
4.1KF titrants
With <F3>, "OK" the stored KF titrants appear.
1-comp 51 component reagent: 5 mg/mL
1-comp 21 component reagent: 2 mg/mL
2-comp 52 component reagent: 5 mg/mL
2-comp 22 component reagent: 2 mg/mL
Pyridine 5Pyridine (1 component reagent): 5 mg/mL
1-comp 5K1 comp. reagent for ketones: 5 mg/mL
...free line for a new titrant
METTLER TOLEDO DL31/DL3813
Setup (<Setup> key)
DeleteThe titrant is deleted.
PrintThe parameters of the titrant are printed out.
ModifyThe parameters of the titrant appear:
Name1-comp 5
Conc.[mg/mL]5.0000
Type1 component
Burette5 mL
Fill rate [%]100
Batch
Check after [day]0
Date00-00-0000
Time0:00
NamePress <F1>, "ABC" to change the name. The titrant with its parameters will be stored under this
name.
Conc.You can change the concentration or its range. You should change the concentration only if you
know the correct value!
TypeSelect one of the preset titrant types: 1 component, 2 components, Pyridine and Others. The
correct selection is important (see "Control parameters", Section 5.2).
BuretteSelect the burette (volume) which is used to dispense the titrant.
BatchPress <F1>, "ABC" to enter the batch number specified on the bottle.
Storage procedure
The titrant with its changed parameters is stored when you confirm the parameter list with <F3>, "OK". If
you press the <Esc> key, no changes will be made.
New KF titrant
You add a titrant to the titrant list when you confirm the free line with <F3>, "OK" The parameters always
appear, only the name of the titrant is missing.
– Change the parameters as described above.
The name of the new titrant must differ from those already stored, otherwise an existing
titrant will be overwritten!
NoteIn the DL31 you can store 10 titrants, in the DL38 20.
14METTLER TOLEDO DL31/DL38
Setup (<Setup> key)
4.2Br No. titrants (DL38 only)
With <F3>, "OK" the stored titrants for determination of the bromine number appear:
Br-/BrO
...free line for a new titrant
-
Bromide/bromate solution
3
DeleteThe titrant is deleted.
PrintThe parameters of the titrant are printed out.
ModifyThe parameters of the titrant appear:
NameBr-/BrO
Conc. [mol/L]0.05
Burette5 mL
Fill rate [%]100
Batch
-
3
NamePress <F1>, "ABC" to change the name. The titrant and its parameters will be stored under this
name.
BuretteSelect the burette (volume) which is used to dispense the titrant.
BatchPress <F1>, "ABC" to enter the batch number specified on the bottle.
Storage procedure
The titrant with its changed parameters is stored when you confirm the parameter list with <F3>, "OK". If
you press the <Esc> key, no changes will be made.
New Br No. titrant
You add a new titrant to the titrant list when you confirm the free line with <F3>, "OK". The parameters
always appear, only the name of the titrant is missing.
– Change the parameters as described above.
The name of the new titrant must differ from the one already stored, otherwise an existing
titrant will be overwritten!
NoteYou can store 5 titrants for the bromine number.
METTLER TOLEDO DL31/DL3815
Setup (<Setup> key)
4.3KF standards
With <F3>, "OK" the stored Karl Fischer standards appear which you can use to determine the concentration of the KF titrants.
di-Na tartratedi-sodium tartrate . 2 H2O
H2O standard 10Water standard : 10 mg/g
H2O standard 1Water standard: 1 mg/g
H2O standard 5Water standard: 5 mg/mL
H2O ISO 3696pure water following ISO standard
...free line for a new standard
DeleteThe standard is deleted.
PrintThe parameters of the standard are printed out.
ModifyThe parameters of the standard appear:
Name di-Na tartrate
H2O content15.660
Unit%
Batch
TypeMass
Weight [g]00.0500
NamePress <F1>, "ABC" to change the name. The standard and its parameters will be stored under
this name.
BatchPress <F1>, "ABC" to enter the batch number specified on the bottle.
TypeYou can select mass or volume. "Weight" changes accordingly to "Volume".
WeightOptimum values for a titrant concentration of 5 mg/mL and a burette volume of 5 mL are stored:
• 0.05 g for the tartrate standard,
• 0.75 g for the H2O standard 10,
• 3.5 g for the H2O standard 1,
• 0.01 g for the H2O standard ISO 3696,
Volume• 3 mL for the H2O standard 5.
Storage procedure
The standard with its changed parameters is stored when you confirm the parameter list with <F3>, "OK".
If you press the <Esc> key, no changes are made.
New KF standard
You can add a standard to the list of KF standards if you confirm the free line with <F3>, "OK". The
parameters always appear, only the name of the standard is missing.
– Change the parameters as described above.
The name of the new standard must differ from those already stored, otherwise an existing
standard will be overwritten!
16METTLER TOLEDO DL31/DL38
Setup (<Setup> key)
Note
• In the DL31 you can store 10 standards, in the DL38 20.
• The Application brochure offers tips on sample preparation, sampling and the optimum amount of sample.
4.4Main methods (DL38 only)
When the titrator is delivered, the <F3> command is assigned to the Tutorial "Hello!", in other words you
can select its sequence immediately. In this menu you can assign methods to command keys <F1>...<F3>
or delete "Hello!".
With <F3>, "OK", the following appears (example on delivery of the titrator):
Main methodsSETUP
F1:Free
ModifyOK
Main methodsSETUP
User methods
METTLER methods
OK
Main methodsSETUP
F1:Method 910
TitleStromboli
Delete ModifyOK
Main methodsSETUP
F2:Free
ModifyOK
With <F2>, "Modify", the following appears:
You can select a method from the METTLER methods (or from the user
methods if available), e.g. "910 Stromboli":
With <F3>, "OK", the following appears:
You can select a different method to configure the <F2> command (press
<F2>, "Modify").
If you confirm this display with <F3>, "OK", the following appears:
Main methodsSETUP
F3:Hello!
You can either delete "Hello!" with <F1> or overwrite it by a different
Delete ModifyOK
METTLER TOLEDO DL31/DL3817
method (press <F2>, "Modify").
Setup (<Setup> key)
4.5Peripherals
With <F3>, "OK" the names of the instruments which you can attach to the titrator via the RS232C interfaces, the Centronics interface, the stirrer output and the DV705 connection appear. You must define these
and possibly change the preset settings.
Printer
Balance
Titrn stand
Computer
If you have the METTLER TOLEDO oven sample changer Stromboli connected to the TTL I/O interface, no
further entries are necessary.
4.5.1Printer
With <F3>, "OK" the names of the printers which you can attach to the titrator appear.
HP
GA42
EPSON
None
– Select the attached printer.
– With an EPSON or HP printer change the type and possibly the settings with <F2>.
TypeSelect "PCL3+", "PCL3", "LQ/Stylus" or "LX/SX"
PaperSelect "Single sheet" or "Fanfold" (see "Feed")
FormatA4width = 21 cm, length = 29.7 cm
YesEach report starts on a new page; with "Paper: Single sheet" a footer is printed out
on each page.
NoThe individual reports are separated by a distance of 2 lines.
– If you select an EPSON- or HP-compatible printer, you must enter the following settings:
Standard settings for the printer are printed in bold
HP-compatibleEPSON-compatible
TypePCL3+ or PCL3LQ/Stylus or LX/SX
PaperSingle sheetSingle sheet or Fanfold
FormatA4 or 81/2 * 11"A4, 81/2 * 11" or 81/2 * 12"
Automatic feedYesYes or No
18METTLER TOLEDO DL31/DL38
Setup (<Setup> key)
4.5.2Balance
Unidirectional
Bidirectional
None
Unidirect. On request for the weight, the balance continuously transfers the current weight value.
Bidirect.
TitratorBalance
Data to titrator
On request for the weight, the balance transfers the current weight value when ordered by the titrator.
TitratorBalance
Command from titrator
Data to titrator
Configuration: "Send Cont."
Configuration: "Send Stable"
("Send on Transfer")
With METTLER TOLEDO balances AB, AG, AT and AX always use the setting "Bidirectional" and the following
settings on the balance:Baud rate: 2400, Data bits: 7, Parity: Even, Handshake: No.
4.5.3Titration stand
Internal
External
As an external titration stand you can attach the DV705 KF Titration Stand from METTLER TOLEDO whose
magnetic stirrer and pump are controlled via the DV705 connection ("External stand").
4.5.4Computer
Defined
Not defined
– If you attach a computer, you must select "Defined" and possibly change the settings.
Baud rateSelect 1200, 2400, 4800 or 9600.
ParitySelect "Even", "Odd" or "None".
Data bitsSelect 7 or 8 bits.
Stop bitsSelect 1 or 2 bit(s).
Character setSelect "ASCII" or "DL".
NoteIf you select "DL", you must specify 8 for the number of data bits!
Start/endSelect "'{'/CR" or "STX/ETX".
METTLER TOLEDO DL31/DL3819
Setup (<Setup> key)
Default settings for the computer
Baud rate4800
ParityEven
Number data bits8
Number stop bits1
Character setASCII
Start/end'{'/CR
Note
• The METTLER TOLEDO LabX titration software allows you to control your titrator via a computer.
•
You will find additional information on the communication between titrator and computer in the Description
of the computer interface.
4.6Settings
You can or must select or define the following settings for the titrator:
Language
User list
Date
Time
Titrator ID
Access for
Beep
Solv. capac.[mL]
Pumping time [s]
LanguageSelect English, German, French, Spanish, Italian or Russian.
User listPress <F2>, "Modify", and a list of defined names and an empty line ("…") appears
where you can enter a new name. Press "ABC" and you can enter your name. Erase a
selected name with "Delete".
You can then select your name for both the development of a method and the performance
of the analysis.
DateYou can set the date and select the format.
Day26
Month8
Year1998
Format31-08-1998
FormatSelect "31-08-1998", "08/31/1998", "31-Aug-1998" or "Aug/31/1998".
TimeYou can set the time and select the format.
Hour10
Minute20
Format24 h
FormatSelect "24 h" or "am/pm".
20METTLER TOLEDO DL31/DL38
Setup (<Setup> key)
Titrator IDTo enter an identification for your titrator, press <F1>, "ABC". The ID is always reported.
Access for
You can decide whether all persons who work with the titrator should also have access to
the Setup and Method menus, in other words have the right to modify resources or methods.
ExpertWhen the titrator is delivered, you have access to all menus. To change this,
select "Routine".
RoutineYou can
• in the Setup menu only print out the defined resources and view the list of
• in the Method menu only print out the list of methods (DL38 only) and
• in the Results menu not perform any recalculation and not change any
• under the <i> key, select neither the Tutorial, the KF learn titration nor the
Renewed access for experts is described on the last page of these Operating Instructions.
BeepYou can select:
Yes
Minimum
No
MinimumA beep sounds on transition from pretitration to standby, on display of the
results and in the case of error messages.
settings.
each method.
statistical data.
Hardware tests menu.
YesA beep also sounds after each keystroke.
Solv. capac.You can perform several titrations in the same solvent in succession, but its capacity is
limited. In the vicinity of the capacity limit, the reaction rate decreases.
You should enter a value for the maximum titrant volume which is suitable for your
sample determinations. If this value is exceeded, the error message "Solvent capacity
exhausted" appears.
Only the titrant volume which is consumed during the analysis is summed. The volume
which is added, e.g. using the auxiliary function "Dispense" is not taken into account.
The difference between the defined value and the titrant consumption is shown under the
<i> key.
Pumping timeThe defined time is a safety criterion: the pump will shut off after this time if you, e.g.
leave the titrator while the pump is running.
METTLER TOLEDO DL31/DL3821
Methods (<Method> key)
5Methods (<Method> key)
The titrator automatically performs analyses with the help of predefined methods. On delivery of the instrument
in the DL31 the KF standard method specified by us
is stored. You can modify this method and save it
under a number.
MethodsMETHOD
KF standard method
HelpModify
▼
MethodMETHOD
No.0
of the method
HelpOK
When you have modified and saved the method,
the following appears (example):
MethodsMETHOD
in the DL38 the KF standard method specified by us
and 11 METTLER methods are stored. You can modify
these methods and save them as user methods.
MethodsMETHOD
User methods
METTLER methods
KF standard method
HelpPrintOK
A complete titration method comprises sample preparation, mixing and wait times, the actual titration,
result calculation and a report. These partial steps are defined in the titrator as parameters which determine
the progress of an analysis.
You modify a method
by changing its parameters.
22METTLER TOLEDO DL31/DL38
Methods (<Method> key)
5.1The KF standard method
You can use the KF standard method to develop your own method. In this you are guided from one parameter to another in the specified order. Exceptions are sample, control, termination and report parameters,
which are shown as a group in a mask. In the case of the DL38, this also applies to the calculation
parameters.
With the user and METTLER methods, all parameters appear as a list in a mask. Exceptions are
again the above-mentioned groups. The last parameter of a method "Author of the method" always appears
in its own mask.
The method is always shown with the following default parameters. The parameters which you can modify
are listed in Section 5.2.
No. of the method0
Title of the method
Sample parameters
ID
TypeMass
Minimum [g]0.0000
Maximum [g]5.0000
EntryBefore
Speed [%]35
Mix time [s]10
Titrant2-comp 5
Volume [mL]
for predisp.0.0000
(Wait time [s]
after predisp.0)
Auto startNo→ DL38 only
Control parameters
Current [µA]20
End point[mV]100
∆Vmin [µL]0.5
∆Vmax [µL]8.0
∆Vmax factor [%]30
StartCautious
Termination param.
Max. time [s]600
Drift stopRel.
Drift [µg/min]15
Vmax [mL]10.000
The METTLER methods 900 – 910 and specific parameters for automation with the METTLER TOLEDO oven
sample changer Stromboli are described in a separate chapter.
METTLER TOLEDO DL31/DL3823
Methods (<Method> key)
5.2KF method parameters
The following parameters can be changed. This will either cause some parameters to appear or disappear.
5.2.1No. (of the method)
The method is stored under this number. The numbers 900 – 999 are reserved for METTLER methods.
5.2.2Title (of the method)
You can give the method a specific name.
5.2.3Sample parameters
IDYou can enter an identification for the sample(s) now or just before the titration for every sample
("Request").
TypeSelect "Mass", "Fixed mass", "Volume", "Fixed volume", "Pieces" or "Fixed pieces".
Weightappears when "Fixed mass" is selected.
Volumeappears when "Fixed volume" is selected.
Numberappears when "Fixed pieces" is selected.
Densityappears when "Volume" or "Fixed volume" is selected.
Minimum/Maximum
appears when "Mass", "Volume" or "Pieces" is selected. The unit depends on the type of sample:
g, mL or pcs.
<F1>, "Help"For a fixed or variable amount of sample you can have the titrator calculate
and enter values if you know the approximate water content of your sample.
Entryof the amount of sample before or during the actual titration. "Entry" is faded out when "Fixed
mass", "Fixed volume" or "Fixed pieces" is selected.
DL38:
If you select "Yes" for the parameter "Auto start", on storage "Before" is automatically
set to "During".
5.2.4Speed
Stirring should guarantee quick and thorough mixing of the electrolyte. We rocommend the setting: 40%.
The magnetic stirring rod should not "jump". Only use the magnetic stirring rod included in the delivery.
5.2.5Mix time
This value for time is an empirical value and can be entered specific to the samples ("Request").
5.2.6Titrant
You can select from the titrants stored in the Setup menu.
24METTLER TOLEDO DL31/DL38
Methods (<Method> key)
5.2.7Volume (for predispensing)
A predispensing shortens the titration time and can be entered specific to the sample ("Request").
NoteIn the user and METTLER methods the parameter is called "Predis.".
5.2.8Wait time (after predispensing)
The parameter does not appear in the KF standard method if you have entered "0" for the predispensing.
This parameter always appears in the user methods.
5.2.9Auto start (DL38 only)
YesThe DL38 begins analysis as soon as a sample has been added. The time for adding a sample
may not exceed 30 seconds or the instrument will revert to its "Standby" mode. You should
select "Yes" for determinations in which, e.g. side-reactions could occur during the titration.
5.2.10 Control parameters
The following holds for all control parameters: You should not modify the preset default values for the
different reagent types until you have evaluated the titration curve of your sample (see Application brochure).
1 component2 componentsPyridine
Current20 µA20 µA20 µA
End point100 mV100 mV100 mV
Note∆Vmin and ∆Vmax depend on the burette volume; the values listed apply to the 5 mL burette.
In the Tutorial and KF standard method these default values are automatically entered for the selected titrant
and hence the type. This does not apply to user methods: When you select a titrant whose type differs from
the previous one, you should also adapt the control parameters.
CurrentSelect 1, 2, 5, 10, 15, 20, 30 or 40 µA for polarization current.
5.2.11 Termination parameters
To terminate the titration you must define "Vmax" and "Drift stop" ("Delay" respectively) as a condition; the
maximum time is an additional termination condition. The titration is terminated when the first of the
selected conditions is met.
METTLER TOLEDO DL31/DL3825
Methods (<Method> key)
Vmax
The maximum volume is intended as a safety precaution: It avoids dispensing an unnecessarily
large amount of titrant if the titration is faulty.
Drift stopSelect "Relative", "Absolute" or "No".
RelativeTermination when actual drift is less than the drift measured before the titration (D0)
plus the entered value (Di): D < D0 + Di.
AbsoluteTermination when the drift is less than the entered value (Di): D < Di.
No"Drift [µg/min]" disappears and "Delay [s]" appears.
DelayIf you enter "0" for the delay time, the titration is terminated as soon as the end point is reached.
5.2.12 Blank
For the entry of a correct value, you should perform a determination.
[µg]You can enter a value in the method.
RequestBefore each titration, the value is requested individually for each sample.
5.2.13 Drift
The drift value is needed for calculation of the result.
[µg/min]You can enter a value in the method.
RequestBefore each titration, the value is requested individually for each sample.
OnlineThe moisture which enters the titration vessel is always measured, evaluated and displayed in
the standby mode.
Determin. Some standard operating procedures require insertion of the value of a drift determination. If
you have performed a drift determination, the result is stored under the <i> key and used in the
calculation.
5.2.14 Result (DL31 only)
The water content is calculated with the appropriate formula by selection of the unit:
[mg], [µg], [%], [ppm], [mg/pc], [mg or µg] and [% or ppm].
[mg or µg], [% or ppm]
With the following limit values, the result is automatically set from one unit to the other:
mgµg%ppm
less than 0.1 → 99...less than 0.1 →999...
0.1.. ¨greater than 1000.1.. ¨greater than 1000
5.2.15 Calculation (DL38 only)
Calculation 1
R1Result of the first calculation. The water content is calculated with the appropriate formula by
selection of the unit: [mg], [µg], [%], [ppm], [mg/pc], [g/kg], [mL], [min], [ppm]ext.extr.,
[ppm]ext.soln., [%]ext.extr., [%]ext.soln..
26METTLER TOLEDO DL31/DL38
Methods (<Method> key)
[mL]Titrant consumption up to the end point.
[min]Duration of the titration.
[ppm]ext.extr., [ppm]ext.soln., [%]ext.extr., [%]ext.soln.
Special calculation for samples which are used after an external extraction or external
dissolution. The second and third calculation are needed (see Application brochure).
UnitYou can enter a nonselectable unit, e.g. "g" if you define a factor of 0.001 for the calculation of
the H2O content in "mg".
Statistics
To obtain the mean value "x", the standard deviation "s" and the relative standard deviation "srel"
of a sample series, you must select "Yes". You can terminate a sample series using the auxiliary
function "Results" (<Results> key).
Max. srelIf the entered value for the max. relative standard deviation is exceeded, a message appears.
Calculation 2 / Calculation 3
The available parameters are the same as for Calculation 1. Instead of a formula, you can select "No calculation".
The following units are missing: [ppm]ext.extr, [%]ext.extr., [ppm]ext.soln, [%]ext.soln.
NoteYou can not determine samples from an external extraction or dissolution in the routine mode
as you must enter the amount of sample for the extraction/dissolution and the amount of
solvent under calculation 2 and 3.
5.2.16 Standby
YesAfter a titration, the titration vessel is kept anhydrous. We recommend allowing the standby
titration to run even during lengthy pauses so that you can start the next titration at any time.
NoAfter a titration the start mask of the Analysis menu reappears.
[min]After the sample determination, a switch is made to the "Standby" mode. The subsequent rate
of consumption is displayed and printed out during the defined time.
5.2.17 Report
OutputSelect "Printer", "Computer" or "Print.+Comp." as output device.
TypeDL31:Select "GLP", "Short-form" or "None".
DL38:Select "GLP", "GLP+curves", "Short-form" or "None".
Curves can be printed out only on an EPSON or HP-compatible printer.
(GLP report: All data which are required by the rules of Good Laboratory Practice.)
5.2.18 Author of the method
The selection of your name documents that you are responsible for the development of the method.
METTLER TOLEDO DL31/DL3827
Methods (<Method> key)
5.3Storage procedures
The following always appears as the last parameter of a method:
MethodMETHOD
Author
of the method
HelpModifyOK
The modified method is stored when you confirm by pressing <F3>, "OK".
Examples:
DL31DL38
MethodsMETHOD
123 CaCl
KF standard method
HelpPrint Modify
2
MethodsMETHOD
4 H3PO
31 Oil
Delete Print Modify
4
If you have not entered any or a wrong method no., the following appears:
Error
No/invalid value
Min:1.00000
Max:899.000
OK
– Confirm the error message by pressing <F3>, "OK".
MethodMETHOD
No.0
of the Method
HelpOK
– Enter the new number and confirm by pressing <F3>, "OK". The method is stored.
28METTLER TOLEDO DL31/DL38
Methods (<Method> key)
If you have entered a method No. which already exists, the following appears:
DL38
Method 22METHOD
Method No. exists
Change number
Overwrite method
OK
Change number"No. of the method" reappears.
Overwrite methodThe new or modified method is stored, that with
the same number deleted.
While you are modifying a user or a METTLER method, you can press the <Run> key (the <Setup> and
Auxiliary functions keys are blocked).
a. If you have not yet modified any parameter of the method, the start mask of the Analysis menu appears.
b. If you have already modified a parameter of the method, the mask "Author of the method" appears. The
method is stored with "OK" and the start mask of the Analysis menu is displayed.
METTLER TOLEDO DL31/DL3829
Methods (<Method> key)
5.4User and METTLER methods (DL38 only)
The following methods have been developed by METTLER TOLEDO and can be performed immediately (see
Application brochure).
If you wish to develop a bromine number determination to meet your requirements, you can modify
METTLER method 908 or 909 and store as user methods. The "909 Blank Br No." method's result serves
as blank value for the bromine determination.
No.908
TitleBromine No.
Sample parameters
IDRequest
TypeMass
Minimum [g]0.2000
Maximum [g]0.4000
EntryBefore
Speed [%]30
Mix time [s]120
TitrantBr-/BrO3Predisp. [mL] 0.0000
Wait time [s]0
Control parameters
Current [µA]2
End point [mV]150
∆Vmin [µL]1.0
∆Vmax [µL]20.0
∆Vmax factor [%] 100
StartNormal
Termination param.
Max. time [s]6000
Delay [s]15
Vmax [mL]20.000
BlankRequest
Calculation 1
R1 =x[Br No.]*f1
f1 =1.0000
Unit
Decimal places2
StatisticsYes
Max. srel [%]0.00
TypeSelect "Mass", "Fixed mass", "Volume" or "Fixed volume".
Weightappears when "Fixed mass" is selected.
Volumeappears when "Fixed volume" is selected.
Densityappears when "Volume" of "Fixed volume" is selected.
Min./Max. appear when "Mass" or "Volume" is selected. The unit depends
on the type of sample: [g] or [mL].
Entryof the amount of sample before or during the actual titration.
"Entry" is faded out when "Fixed mass" or "Fixed volume" is
selected.
No.909
TitleBlank Br No.
Sample parameters
IDRequest
TypeFixed volume
Density [g/mL] 1.000
Volume [mL]50
Speed [%]30
Mix time [s]120
TitrantBr-/BrO
Predisp. [mL] 0.0000
Wait time [s]0
Control parameters
R1Result of the first calculation. The bromine number or bromine index is calculated using the
appropriate formula by selection of [Br No.] or [Br index].
[mL]Titrant consumption up to the end point.
[min]Duration of the titration.
UnitBromine number and bromine index are quantities without dimension.
The bromine number is the amount of bromine in [g] reacting with 100 g of sample.
The bromine index is the amount of bromine in [mg] reacting with 100 g of sample.
Author"Author of the method" only appears if you change the method.
5.4.2METTLER method "910 Stromboli" (DL38 only)
If, according to your needs, you wish to develop an automated determination of water content of thermally
stable liquids or solids, then you can modify the METTLER method "910 Stromboli" and store it as a user
method.
No.910
TitleStromboli
Sample parameters
TypeMass
Minimum [g]0.0000
Maximum [g]5.0000
Speed [%]40
Mix time [s]300
Set temperature150
Titrant1-comp 5
Control parameters
Current [µA]5
End point [mV]100
∆Vmin [µL]0.5
∆Vmax [µL]4.0
∆Vmax factor [%] 100
StartCautious
Sample parameterThe sample parameters "ID" and "Enter" are not applicable,
Termination param.
Max. time [s]600
DriftstopRel.
Drift [µg/min]10
Vmax [mL]10.000
Mix timeThe time (in seconds) during which the water in the sample evaporates completely
and dissolves into the solvent in the titration vessel.
Auto start
This parameter does not apply because the sequence of the sample series is
controlled automatically.
Set temperatureEnter in °C at the desired temperature at which the water determinations are to be
performed. The set temperature can be selected between 50 °C to 300 °C. This
setting cannot be changed during a series. The set temperature is noted in the
titration protocol.
32METTLER TOLEDO DL31/DL38
Methods (<Method> key)
Blank
Drift
The blank [µg] is made up of the water content of the purge gas (time-dependent) and the
water content of the sample vessel (independent of time). The determination is performed as
a "normal" sample determination with the defined method parameters and, therefore, can be
integrated into the series’ sequence. The product of drift and duration (Drift * Duration) is
subtracted from the result, so that the blank is an absolute value independent of time. The
blank for the Stromboli can be viewed at any time by pressing the <i> key.
RequestThe value is retrieved prior to the series.
AutoThe value is determined automatically prior to the sample series ("Blank"
position on the sample rack). The result is stored under the <i> key
and automatically integrated into the calculations.
Determination
Note
The most recently stored value is used.
A one-time blank determination can be used for various sample series. In order
to do this, a method with the method parameter "Blank: Auto" is developed. In
the request for the sample parameters enter "0" for the number of samples. The
resulting blank value is stored and can be viewed at any time by pressing the
<i> key, i.e., it is available for all ensuing sample series.
The drift [µg/min] can be determined at any time in the STANDBY mode (press <F1>, "Drift")
or automatically prior to a sample series.
RequestThe value is retrieved prior to the series.
AutoThe value is determined automatically prior to the sample series ("Drift"
position on the sample rack). The result is stored under the <i> key
and automatically integrated into the calculations.
Determination
The most recently stored value is used.
StandbyThis parameter does not apply as no post consumption measurements are performed.
ProtocolThe results are printed out at the end of a series.
Author"Author of the method" only appears if you change the method.
At the beginning of a sample series, the following parameters need to be entered, provided you chose
"Request" in the method:
Mix time [s]
Blank [µg]
Drift [µg/min]
ID [Text]
No. of samples [max. 13]
Sample parameters (mass, volume, amount, etc.)
METTLER TOLEDO DL31/DL3833
Analysis (<Run> key)
6Analysis (<Run> key)
You perform titrations with a selected method. The required data are stored in the Method and Setup menu.
You prepare your sample, titrate and obtain the result.
While an analysis is running, all menus and auxiliary functions are blocked.
6.1Starting a method
Preparations
–Press the <Burette> key. Rinse the burette and fill ensuring that there are no air bubbles in the burette and
tubing.
– Press the <Pump> key and fill the titration vessel with approx. 40 mL solvent.
When you press the <Run> key, the start mask appears:
DL31DL38
StartRUN
Method No.22
Title
User
HelpOK
StartRUN
Method No.22
Title
User
Help ModifyOK
As method number
• "0" is entered if no method (DL31) or no user method (DL38) is stored.
• in the DL31 case always that of the stored method is automatically entered together with the title.
• in the DL38 case the number which you enter directly or accept it with <F2> from the METTLER or user
methods is entered.
Note
•You can select the user name. The name which was selected in the analysis last performed remains
stored.
• If you have defined an expiry date for the titrant and this has been passed, an appropriate message
appears before the pretitration starts. You should then first perform a concentration determination.
• DL38: The start mask always shows the number of the method with which an analysis was last performed, e.g. Method No. 22. If you delete this, "Method not available" appears when you confirm the
mask with "OK".
34METTLER TOLEDO DL31/DL38
Analysis (<Run> key)
e
r
StartRUN
• The titrator also accepts methods which are stored in the attached
computer: When you enter a number whose method is not stored in
Data transfer..
the titrator, the method is requested from the computer (see Description of the computer interface). If it is available, the pretitration is
started immediately following transfer; if it is not available, "Method
not available" appears.
6.1.1Menu sequence up to standby mode (KF methods)
Example:
StartRUN
Method No.22
TitleCoffe
User C. De Ca
Help ModifyOK
▼
DL31:"Modify", is missing.
The pretitration starts with <F3>, "OK": It is always executed with the
control parameters of the selected method. The length of the solid bar
reflects the addition of the titrant. The vertical bar represents the thresh-
PRETITRATIONRUN
old for the transition to the standby mode.
Help
STANDBYRUN
Active method:22
Potential [mV]74
Drift [µg/min]2→
DRIFT CONZSample
➞This arrow indicates that you can start a determination.
➘Wait, drift still decreasing!
➚Wait, drift increasing again!
Press <F1> to start the drift determination,
<F2> to start the concentration determination,
<F3> to start the sample determination.
NoteIf the pretitration lasts longer that 10 minutes, a message appears. After confirmation with "OK", a
pretitration again starts. In such a case, you should check whether, e.g. all connections of the
titration stand are tight.
METTLER TOLEDO DL31/DL3835
Analysis (<Run> key)
6.1.2Standby mode
During standby, the following menus are blocked or selectable:
Blocked
<Setup>, <Burette>, <Pump> and under the <i> key: "Hello!", "KF learn titration" and "Hardware
tests".
Selectable <Stirrer>, <Results>, <Method>, <Run> and <i> key (except
"Hardware tests"
).
"Hello!", "KF learn titration" and
NoteIf you press the <Run> key, the sample determination is started!
–To switch to the Analysis menu from the selected menu/selected auxiliary function, press the <Run> key.
Stirrer (<Stirrer> key)
You can change the speed. This is reported for the subsequent sample determination.
Results (<Results> key)
The menu is not shown until you have performed the first measurement or a result is available.
Method (<Method> key)
You can modify all methods. With the active method you can not modify the following parameters:
•Titrant
• Polarization current
• End point
The modified parameters are adopted for the next sample. For documentation conforming with GLP, you
should print out the method before you switch back to the Analysis menu.
Note (DL38 only)If you modify the parameters of the active method, a sample series will be terminated.
If you restart a method, the results of the current series are deleted.
36METTLER TOLEDO DL31/DL38
Analysis (<Run> key)
g
6.2Sample determination
With <F3>, "OK", or <Run> you start a sample determination. According to th selected method, the following
masks may appear.
In succession, specific parameters are requested if defined in the method as "Request".
Example:
The mask for the sample entry appears if a variable
amount of sample is defined with "Entry: Before".
ANALYSISRUN
Please add sample
0.3750 g -
0.8750
HelpOK
The mask for sample addition appears if a variable
amount of sample is defined with "Entry: During"
(example: mass).
Note
• If you take longer than 5 minutes for the entry of samples or sample-specific parameters, the titrator
switches back to the standby mode.
• If you add a sample while the titrator is in the standby mode without first pressing the <Run> key, the
titrator will begin a pretitration to titrate all the water in the sample. Immediately pressing the <Reset> key
and thereafter pressing the <Shift> and <Run> keys simultaneously can prevent this. By doing this, the
pretitration and standby mode is skipped and one of the above masks appears. In this manner your
sample is not "lost".
METTLER TOLEDO DL31/DL3837
Analysis (<Run> key)
n
L
i
Sequence of the sample determination
Example (with "Entry: Before"):
The values for the expected water content and the limits for the weight are
adopted from the defined sample parameters of the method. The limits are
recalculated if you change the value for "H2O exp.".
*Appears only if a balance is defined.
If you have entered the weight value or have it transferred from a balance,
the mask shown appears.
Pressing <F3>, "OK", confirms addition of the sample and starts the analysis.
In the graph, the bar fills and empties in accordance with the drift. The
vertical bar represents the end point of the titration.
** Appears only if you have not yet entered the amount of sample. With <F1>, "Sample", the parameter
"Weight [g]" is displayed to enter the value.
During the titration you can switch between three different displays:
ANALYSISRUN
Drift:
9500 µg/min
Sample Cont. Graph
ANALYSISRUN
Content:
42.0012 mg
SampleConsum Graph
ANALYSISRUN
Consumpn: 1.2010 m
Duration: 0:31 m
Sample DriftGraph
38METTLER TOLEDO DL31/DL38
Analysis (<Run> key)
The titration is ended in accordance with the defined termination parameters. In the background the titrator
switches to the standby mode or the start mask. If you have not yet entered the amount of sample, the
parameter "Weight [g]" appears.
ReportRUN
Printing...
ResultRUN
Method 22
Sample 1
R1 = 1.98%
OK
The result is reported before it is displayed.
DL31: R = 1.98%
If you confirm "OK", either
• the standby mask (method parameter "Standby: Yes") appears or
• the start mask (method parameter "Standby: No"), or
• the post consumption mask (method parameter "Standby: [min]").
The result is now stored under the auxiliary function "Results" (<Results> key) so that you can, e.g.
recalculate it.
NoteIf, after the sample addition, the elapsing mixing time is shown and no amount of sample has been
entered, you can also press <F1>, "Sample", to enter the amount of sample.
METTLER TOLEDO DL31/DL3839
Analysis (<Run> key)
:
n
n
6.2.1Weight transfer from a balance
ANALYSISRUN
If you press <F2>, "Balance", in the corresponding mask, the value
displayed on the balance appears. It is transferred with <F3>, "OK".
> 1.3245 g
OK
6.2.2Post consumption
STANDBYRUN
Post consumption
Duration:0:34 mi
Drift:6 µg/mi
Stop
If you have selected "Standby [min]" in the method, the subsequent
measurement of the drift is performed directly after the sample determination and a drift value automatically printed out every minute (not on
HP-compatible printers).
6.2.3Calculations
DL31 (without Auto start) / DL38 ("Auto start: No")
If you press <F3>, "SAMPLE", or the <Run> key in the standby mode, measurement of the time needed for
calculation of the drift compensation starts. It is stopped as soon as the titration is at an end.
Sample
addition
Mix timeTitrationConfirm SampleRequestsPredisp.Wait time
Measured value acquisition
StandbyTime measuremnet (duration)
DL38 ("Auto start: Yes")
In the case of an automatic start of the titration, the titrator does not start the time measurement until you
have entered the sample.
Sample
addition
Mix timeTitrationConfirm SampleRequestsPredisp.Wait time
NoteFor the H2O content in µg, the formula is multiplied by 1000.
H2O [%]=
H2O [mg] * 0.1 [% * g/mg]
weight [g]
NoteFor the H2O content in ppm, the formula is multiplied by 10000,
for the H2O content in g/kg by 10.
H2O [mg/pc.] =
H2O [mg]
number [pc.]
These parameters are listed on every GLP report to allow you to check the result using the formula.
6.2.4Sample series with statistics (DL38 only)
If you have selected "Statistics: Yes" in the method under Calculation 1, 2 or 3, results and statistical data
of the sample or concentration determinations are stored under the auxiliary function <Results>.
The series is terminated if you
• enter the appropriate command,
• switch off the titrator or an attached Stromboli,
• modify the method parameters of the active method,
• select another method.
A sample series is also terminated if you
• select the concentration determination in the standby mode,
• change the current titrant in the Setup menu after a Reset.
A conc. series is also terminated if you
• select the sample determination in the standby mode,
• change the KF standard during the determinations,
• change the current standard in the Setup menu after a Reset.
The series is not terminated if you stop a determination with <Reset> and perform the next determination
with the same method.
NoteYou can perform a series with maximum 99 samples. The last 60 results are stored. (This is not the
case if operating a Stromboli!)
METTLER TOLEDO DL31/DL3841
Analysis (<Run> key)
After, e.g. the third sample determination, the following appears:
ResultRUN
Method 22
Sample 3
R1 = 1.91 %
xOK
▲
<F1>You can switch between the result and statistics displays.
StatisticsRUN
Method 22
R1 n = 3
x = 1.95 %
s = 0.021095 %▼
ResultOK
▼
NoteThe mean value is calculated after two samples, the absolute and relative standard deviation after
three samples.
6.3Concentration determination
For the correct determination of the water content of a sample, you should determine the concentration of
the titrant with Karl Fischer standards (see Application brochure).
The concentration is always determined with the control and termination parameters of the active method.
A predispensing and wait time defined in the method are not performed, a defined blank value is not
incorporated in the calculation and the selection of "Standby: No" is not taken into account. After a concentration
determination, a switch to the standby mode is always made to allow duplicate or multiple determinations.
You can select one of the standards stored in the Setup menu with <F2>,
"Modify".
A batch number defined in the Setup menu is entered here. If you change or
enter this for the first time, it will be adopted for the current concentration
determination.
Irrespective of the mixing time you have defined in the method, you can
enter this according to the standard.
This mask appears when "Entry: Before" is defined in the method. If
"Entry: During" is defined, the sample addition mask appears immediately.
HelpBal.*OK
▼
ANALYSISRUN
Please add sample
0.0500 g (approx
HelpOK
*Appears only if a balance is defined.
The amount of sample suggested in the previous mask is displayed.
The further progress corresponds to that of the sample determination.
METTLER TOLEDO DL31/DL3843
Analysis (<Run> key)
L
L
n
DL31
Each result is entered in the parameter list of the titrant, in other words the second overwrites the first etc.
ResultRUN
Method 22
Sample 2
Conc.= 4.9892 mg/m
OK
If you calculate the mean value of the results and then enter this in the list,
date and time are deleted; this means the shelf life date can not be checked.
DL38
The first result is entered in the parameter list of the titrant and is overwritten by the calculated mean value
of two then three results etc..
ResultRUN
Method 22
Sample 2
Conc.= 4.9892 mg/m
xOK
As in the sample determination, you can switch between the statistics and
result displays.
6.4Blank determination
If you have dissolved a sample in a solvent with a H2O-content unknown to you, run an analysis with only
the solvent and without the sample. The obtained value can be entered as a blank befor each titration
("Request") or can be entered automatically ("Determination"). This blank value is included in the calculations.
The blank value of the METTLER method "910 Stromboli" is stored under the <i> key.
6.5Drift determination
You must determine the drift when a value is required for the calculation of the water content which has been
obtained by a titration. With <F1>, "DRIFT", appears:
DRIFTRUN
The length of the solid bar represents the elapsed time.
Duration:4:24 mi
Help
After 5 minutes, the result is reported and displayed. At the same time, the value is stored under the <i> key.
44METTLER TOLEDO DL31/DL38
Analysis (<Run> key)
o
6.6Bromine number determination (DL38 only)
The course of the determinations corresponds to the sample determination of a KF titration. This also applies
to the performance of a sample series.
StartRUN
Methode No.123
Title Bromine N
UserR. Hube
Help ModifyOK
ANALYSISRUN
Weight [g]0
Minimum [g] 0.2000
Maximum [g] 0.4000
HelpBal.*OK
ANALYSISRUN
Consump:
2.3450mL
Sample**
ResultRUN
Method 123
Sample 2
Bromine No.= 4
OK
If method parameters are defined with "Request", they are displayed before
the sample entry mask appears (example "Type: Mass" and "Entry: Before").
*Appears only if a balance is defined.
When you have confirmed the sample addition, the titration starts.
** Appears only if you have not yet entered the amount of sample. With <F1>, "Sample", the parameter
"Weight [g]" is displayed to enter the value.
After the end of the titration, the result is reported before it is displayed.
If you confirm "OK", the start mask reappears.
NoteMolar mass Br2:159.8 [g/mol] = 0.1598 [g/mmol]
METTLER TOLEDO DL31/DL3845
Analysis (<Run> key)
6.7Sample determination with the oven sample changer Stromboli (DL38 only)
The oven sample changer Stromboli is used in combination with the DL38 titrator to automatically determine the water content of thermally stable liquids and solids. For this purpose, the sample is heated in the
oven and the evaporated water is transported into the titration vessel by means of a dry inert gas stream.
The DL38 controls all of Stromboli’s functions via the METTLER Method "910 Stromboli". Using this method
as point of departure, user methods can be developed allowing automatic determination of sample series
with up to 13 samples as well as drift and blank determinations. Prior to determination, Stromboli automatically inserts the sample vials on the sample rack into the oven.
The result of a sample determination with Stromboli is made up of the sample’s water content, blank value
(water content of an empty sample vial) and drift (sum of the water content of the rinsing vessel over the
measuring time "Duration").
The drift determination is automatically performed if "Drift: Auto" was selected in the method. Drift can be
determined manually at any time in the standby mode (press <F1>, "Drift"). In both cases the determination is performed with the sample vial in the "Drift" position. The vial is inserted into the oven and the
determination is performed for three minutes after the set temperature entered in the method has been
reached. The result is stored (i.e., it is available for subsequent sample series) and can be viewed by
pressing the <i> key (including date and time of determination).
The blank determination is automatically performed if "Blank: Auto" was selected in the method. The determination is performed with the sample vial in the "Blank" position (empty sample vial). The determination
is performed as a "normal" sample determination with the defined method parameters and, therefore, can
be integrated into the series’ sequence. The blank value is defined as sample "0" without any content
parameters and the default parameter "ID: Blank". The result of the blank determination is stored and can be
seen under sample number "0" in the titration report or viewed by pressing the <i> key (including date and
time of determination).
Sequence of a sample series
When starting a method from the idle state, the entered, desired temperature is conveyed to the oven. The
sample vial in position "Drift" is inserted into the oven and the oven heats to the set temperature during
pretitration, i.e., in the standby mode. The temperature display flashes on the Stromboli and the display
alternates between current temperature and set temperature. After reaching a constant drift at the set temperature, the sample series can be started from the standby mode by pressing <Run> or <F3>, "Sample".
ANALYSIS#01 RUN
ID
Weight [g]0
Minimum [g] 0.0000
Maximum [g] 5.0000
Bal.*StartOK
When starting a sample series the parameters are requested for which
"Request" has been selected in the method. The entries are stored by pressing
<F2>, "Start" and then the display for the next sample (#...) appears.
Pressing <F2>, "Start" begins the sample series even if not all of the sample
data have been entered. The analysis will only begin when the oven has
reached the set temperature. Until the set temperature has been reached
(approx. 1-2 min), the message "Waiting for Stromboli" appears.
46METTLER TOLEDO DL31/DL38
Analysis (<Run> key)
s
e
No.IDStatu
0 BlankDone
1Activ
2Ready
3Ready
ModifyOK
▼
Sample 1RUN
ID
Weight [g]0
Minimum [g] 0.0000
Maximum [g] 5.0000
HelpBal.*OK
*Appears only if a balance is defined.
By pressing <F1>, "Sample" during a series, the current status (ready,
active or done) of the samples is displayed. Pressing <F2> "Modify" allows you to alter or complete your entry. After each sample the result is
calculated and reported, and the sample series is continued without displaying the results. If a printer is installed, the result is printed out after each
sample.
At the end of a sample series the results of the final sample and the series’
statistics (press <F1>, "x") are displayed. In standby mode, the results
and statistics can be viewed by pressing the <Results> key.
Note
• If the input of the sample specific parameters takes longer than 5 minutes, the titrator will return to
Standby mode. The inputs to this point will then be lost. With a sample series and a Stromboli method,
the sample data for the first sample can be entered and the analysis started. The sample data for the
remaining samples can be entered while sample 1 is being analyzed by pressing <F1>, "Sample".
• If an analysis is terminated by the termination parameter "Max. time" even though the entire amount of
water has not been titrated yet, then the Stromboli’s sample rack returns to its basic position ("Drift"
position below the oven) and remains there even during pretitration of the next sample. The Stromboli
only moves on to the next sample after the DL38 has gone into standby mode.
• In the following cases the sample series will be interrupted:
•
When one of the following messages appears: "Weight [g] 0", "No titration!", "Capacity exhausted!",
"No drift determination performed!", "Blank determination not yet performed!", "CONC determination due!".
•When the calculated value for "srel" is above the value entered in the method.
In order to continue the series, the message on the display has to be confirmed by pressing <F3>, "OK",
on the titrator.
• At the beginning of the next series, results and statistics of a series are automatically erased as soon as
the message "Series aborted!" is confirmed by pressing <F3>, "OK".
• As soon as a Stromboli is attached to a DL38 and switched on, each titration is started automatically,
regardless of whether "Auto start" has been set to "Yes" or "No". The Stromboli must be switched off for
a KF method without Stromboli!
METTLER TOLEDO DL31/DL3847
Analysis (<Run> key)
t
5
e
6.8KF learn titration
If you are unsure what parameters should be entered for your sample, select the learn titration stored under
the <i> key. The titrator calculates the range for the amount of sample you need to enter as well as control
and termination parameters, stores these and prints them out after the titration.
The learn titration is blocked when an analysis or auxiliary function is running. The learn titration cannot be
used with a Stromboli.
If you confirm "KF learn titration" with "OK", the following appears:
If you enter a number under which a method is already stored, an appropriate message appears:
DL31:
Although the method is overwritten, you should enter a different
number (documentation of the method!).
DL38:
The method is stored as a new user method with a different number.
DL31:Instead of "Calculation 1", "Result" appears.
Correct selection of the titrant used is important: The control depends on
the titrant type!
The pretitration starts with <F3>, "OK".
▼
STANDBYKF-L
Active method 55
Potential [mV]87
Drift [µg/min]2→
HelpSAMPLE
48METTLER TOLEDO DL31/DL38
–
Wait until the drift is completely stable before you confirm <F3>, "SAMPLE"!
Analysis (<Run> key)
n
e
The masks for the entry of mixing time and sample ID appear before the amount of sample is requested.
AnalysisKF-L
Weight [g]0
HelpBal.*OK
▼
ANALYSISKF-L
Add sample or
confirm additio
1.2345 g
HelpOK
▼
ResultKF-L
R1 = 0.61% !
Calculated param
ters are stored
OK
▼
STANDBYRUN
Active method:55
Potential [mV]87
Drift [µg/min]2→
DRIFT CONC. SAMPLE
*Appears only if a balance is defined.
Appears if you have already entered the amount of sample.
Otherwise, "Please add sample" appears.
After the end of the titration, the result and the "learned" parameters are
printed out before the result is displayed. The result is always marked
with an exclamation mark (!).
The standby mask of the Analysis menu with the active method number
appears so that you can determine the next sample with the "learned" method.
The following parameters are calculated:
• Minimum and maximum for the amount of sample to be entered,
• ∆Vmin, ∆Vmax, and ∆Vmax factor (control parameters),
With the aid of the Symbol keys you can execute functions which are independent of a titration method, but
which can support or expand it. While an analysis is running, all auxiliary functions are blocked; in standby
mode (KF titrations) you can activate only the functions of the <Stirrer> or <Results> key.
7.1<Stirrer>
You can switch the stirrer on and off and change its speed. Start and stop the stirring with <F3>.
StirrerSTIRRE
Speed [%]35
Star
StirrerSTIRRE
Speed [%]35
StartThe stirrer stirs at the defined rate.
To change the speed in the standby mode in KF titrations,
–press the <Stirrer> key, change the speed and confirm with "OK": The
stirrer stirs at the new speed.
OK
– Press the <Run> key to return to the Analysis menu.
7.2<Burette>
The following functions are possible under this menu.
Rinse
Rinse tip
Dispense
Rinse
Start and stop the rinsing process with <F3>.
BuretteBURETTE
Rate
With <F2> you can change the rate at which the burette should
be filled. For the two-component reagents we advise always set-
Rinse (rate: 100
Rinse tip
Dispense: 0.010
Help Modify Star
Start
ting 60%. They outgas easier than the one-component reagents.
The titrator delivers 1 burette volume. The piston of the burette
moves below the zero position to ensure that air bubbles "hidden"
in the stopcock are siphoned up and then expelled.
– Always place the dispensing tubing in a waste beaker!
Rinse tip
Start5% of the burette volume are delivered to eliminate any diffusion of solvent in the dispensing
tubing tip.
NoteYou should also place the dispensing tubing in a waste beaker during this operation to ensure the
titrated solvent is not "overtitrated".
Dispense
StartYou can dispense the displayed volume immediately. With <F3>, "Stop", you can stop the
dispensing.
<F2>With <F2> this mask appears in which you can modify the dispensing parameters:
BuretteBURETTE
Volume [mL]0.010
Burette5 mL
Fill rate [%]100
Star
7.3<Pump>
With the aid of the pump you can add solvent to the titration vessel and siphon it off.
PumpPUMP
To empty, press <F3>
To fill, press <F3>
and press and
blue pump but
Star
PumpPUMP
Emptying/filling...
Stir Stop
You switch the pump on and off with <F3>.While the pump is running, you can switch on the
stirrer with <F2>, "Stir".
Filling
– Push the dispensing tubing up to the 60 mL mark of the titration vessel so that the solvent does not
splash during filling. The solvent starts to flow into the titration vessel as soon as a sufficiently high
vacuum which siphons up the solvent has built up in the titration vessel.
– Add approx. 40 mL to the titration vessel (the platinum pins of the electrode must be immersed in the
solvent!), release the pump button and press <F3>.
– Pull up the tubing.
Emptying
– Push the suction tubing to the base of the titration vessel so that the solvent can be siphoned off.
– Pull up the tubing when the pump is running as far as it will go, then press <F3>.
• The pump switches off automatically when the pumping time defined in the Setup menu is reached.
• Each pumping operation resets the solvent capacity to the value defined in the Setup menu. At the same
time, the value is stored under the <i> key.
• In the case of the pump of the attached DV705 Titration Stand, you must press the lower arrow key
during emptying and the upper arrow key during siphoning up.
7.4<Results>
The following functions are available in this menu:
All results
Recalculation
Calculations (DL38 only)
Statistics (DL38 only)
Report
Statistics series (DL38 only)
All resultsis accessible when one result is available.
Statisticsare accessible when 2 results of a sample series are available.
Reportis accessible when a drift determination has been performed or a result is available.
Statistics series is accessible when the first result of a sample series is available.
In order to change, e.g. results of KF determinations in the standby mode,
–press the <Results> key, change the result under "Recalculation" and confirm with "OK".
– Press the <Run> key to switch back to the Analysis menu.
7.4.1All results
You can view the results generated in the titration of a sample or sample series. You can also print out the
results. The data in the result memory are deleted when you switch off the titrator.
DL31:The results of the last 20 determinations are stored.
DL38:You can perform a sample series with 99 samples. The results of the last 60 samples remain
stored, in other words the results of sample 1 are eliminated when you start to titrate sample 61.
Statistical data of the last 60 samples are always calculated.
Note
• As "Result" the sample whose determination was terminated with the error message "No titration!" is also
stored.
•A result is also stored if you terminate a determination with <Reset>. The result is calculated with the
titrant volume consumed up to that time.
The sample with the changed result is marked by an asterisk (*).
Note
•
The values obtained in the determination always remain stored so that you can also nullify changes later.
• Under "All results" you can print out the recalculated results. Here, the changed values and the original
are reported.
• DL38: The changed results are also taken into account for the statistical data.
• DL38: In KF conc. determinations the calculated result or its mean value is not entered in the parameter
list of the KF titrant if you change a value, e.g. after three determinations. However, if you perform a conc.
determination after this change, the recalculated result is taken into account in the statistics calculation.
At the end of every determination, you can print out an (additional) data report or send it to the computer.
ReportRESULTS
OutputPrint.+Com
TypeGLP
Help ModifyOK
NoteIf you have changed the result of this last determination beforehand, it will be printed out marked
with an asterisk (*).
7.4.4Calculations (DL38 only)
At the end of every titration you can perform additional calculations for this sample.
CalculationsRESULTS
R =x [mg]*
f =1.0000
Unitmg▼
Help ModifyOK
With "OK" the result is displayed and can then be printed out.
7.4.5Statistics (DL38 only)
If you have selected "Statistics: Yes" in the method under calculation 1, 2 or 3, in a series of 2 samples the
mean value will be calculated, in one comprising more that 2 samples the absolute and relative standard
deviation. You can print out these results here and change the statistical evaluation by eliminating outliers.
The DL38 does not incorporate results in the statistics which were obtained after a learn titration, a "Reset",
and the following messages:
• No titration!,
• End point not reached! Termination through Vmax,
• CONC value not stored! Conc. range exceeded,
• CONC value not stored! Termination through Vmax.
Using the result list (press <F1>) you decide what sample result is an
outlier and then delete it.
If you delete "Sample 1" with <F1>, "Undo" is superimposed to allow
you nullify the command. At the same time, sample 1 is marked as
"deleted".
Confirm "OK" to show the following:
▼
StatisticsRESULTS
Method 11
R1
n =
2
x = 0.70 %
ResultPrintOK
(Example: Sample 1 is deleted)
The statistical data are calculated with the changed number of samples.
You can delete several sample results and print out the result each time.
As long as the result list of this sample series is stored, you can nullify the deletion operation.
7.4.6Statistics series (DL38 only)
With the aid of this function you can terminate a sample series. This deletes all data in the result memory,
a subsequent analysis with the same method restarts with "Sample 1".
Stat.seriesRESULTS
End series?
No
Yes
OK
METTLER TOLED O DL31/DL3855
Information menu (<i> key)
8Information menu (<i> key)
The measured values and menus below are stored under the <i> key:
Solv. capac.The solvent capacity defined in the Setup menu is entered as an initial value. After each
titration, the difference between the initial value and the amount of titrant consumed is
displayed. If "0" is reached or exceeded, an error message appears. As soon as you
switch on the pump, the value defined in the Setup menu is reentered.
In bromine number determinations, the solvent capacity is set to "0".
Drift
Here, the value of the drift determination last performed is stored and automatically adopted
for the drift compensation if you have defined "Drift: Determination" as a method parameter.
BlankThe value of the last blank determination perfomed is stored here and automatically adopted
for calculations (only for method with Stromboli).
KF learn titr.The course of this titration is described in Section 6.
Hello!As you get to know the titrator step by step, the procedure is not described.
Hardware testsThe functions of this menu which you can use to test the hardware of the titrator are
described in Section 9.
56METTLER TOLEDO DL31/DL38
Error messages, malfunctions and hardware tests
x
x
x
r
L
D
9Error messages, malfunctions and hardware tests
9.1Error messages from the titrator
Here, the error messages which refer to this section or are not self-explanatory are explained.
AAfter the titrator has been switched on, the following errors can be displayed:
• ERROR
• RAM x test ERROR(Q)"x" represents a number
• ERROR –> HW version
• ERROR –> DCB
• ERROR –> application
• No application found
The messages refer to a faulty hardware component.
– Press any key: The titrator attempts to continue the start process; this may be successful, but does not
guarantee 100% functioning.
Measures
1. Note the displayed text.
2. Note the instrument number and software version of the titrator (shown on every report).
3. Please then inform METTLER TOLEDO service of these points and the attached peripheral units with the
defined configuration.
B.If the following three displays appear, parts of the hardware may be faulty.
Internal error:
xxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxx
xxxxxxxxxxxxxxxxxxx
SYSTEM IS BLOCKED!
See manual: Erro
OS ERROR: x
P1: xx
P2: xx
D1: xx
D2: xx
SYSTEM IS BLOCKED!
Please call METT
TOLEDO service
ERROR: x
SYSTEM IS BLOCKE
Please call METTLER
TOLEDO service
"x":represents letters and/
or numbers.
Measures
1. Note the text or error (e.g. "ERROR: 1000") and the operation that triggered it.
2. Note the instrument number and software version of the titrator (shown on every report).
3. Then switch off the titrator.
4. Please inform METTLER TOLEDO service of these points and the attached peripheral units with the
defined configuration.
METTLER TOLEDO DL31/DL3857
Error messages, malfunctions and hardware tests
C.If the following warning appears, you should regard the executed action as failed and repeat it.
Warning: x
Press any key
to continue
"x":represents a numberl
Measures
1. Press any key: The titrator can continue to function.
2. On no account consider the result of a determination terminated in this manner!
3. If the warning appears frequently, you should take the measures described under point B.
D.Faulty data deleted!
The titrator has stored only part of a method if, e.g. the power fails when the method is being stored. It
deletes this method completely. It is also possible that it has deleted several methods.
Measures
1. Confirm the message.
2. Check whether your methods have been deleted, if so reenter them.
The titrator has stored only parts of parameters of a resource if, e.g. the power failed during storage of the
parameters. All data stored in the Setup menu are deleted.
Measures
1. Confirm the message. The titrator now loads the default parameters of the Setup menu, in other words all
data stored on delivery of the titrator.
2. Switch the titrator off and on.
3. Check whether the default parameters are again available and reenter the names and values stored
before the failure.
If you receive this error message frequently, the battery is discharged! This must be changed by
METTLER TOLEDO service.
E.Method not stored!
The titrator can no longer store a method if the user data memory is full.
Measure
1. Confirm the message.
2. Delete one or more methods from the memory or transfer to the computer.
3. Reenter the new method and save it.
58METTLER TOLEDO DL31/DL38
Error messages, malfunctions and hardware tests
9.2Other errors and malfunctions
The following collection of errors and malfunctions which are not reported by the titrator will help you to
rectify the fault yourself in many cases so that you do not have to rely on the METTLER TOLEDO service
(specified in the Table as "Service").
You can also first check some malfunctions with the hardware test (see Section 9.3) before you contact the
service.
ErrorPossible causeRectification
No display on titratorInstrument not attachedAttach to power supply, otherwise contact service.
to power supply.
Several points of theContact service, see also Section 9.3.1.
display are missing
The display and theContact service, see also Section 9.3.1.
key do not match
Transmission error toPeripheral device faultyCheck attached devices for correct functioning,
attached peripheral deviceor switched off.see also Section 9.3.2.
Device (printer, balance,Instrument not switched on;Switch on instrument;
computer) does not reactwrong settings and/or configurationsettings and configuration must match,
(switch settings wrong);see Sections 4.5 and 9.3.2;
wrong cable.cable see Section 11.2.
Pump does not functionSuction tubing not connected,Connect suction tubing and check connections,
titration stand leaks;see Section 10.3;
pump faulty.check pump, see Section 9.3.3.
TTL inputs or outputsInputs and/or outputs faulty.Check inputs/outputs, see Section 9.3.3.
do not function
Burette and stopcockBurette drive faulty.Contact service.
do not move to zero
position on switching on
Magnetic stirring barCable of magnetic stirrer notCheck magnetic stirrer, see also 9.3.3;
does not rotateplugged in properly;change magnetic stirrer;
wrong magnetic stirring bar.use bar supplied, see Section 11.
Wrong potential valuesElectrode faulty or contaminated.Check electrode (see electrode specifications bulletin);
use new electrode.
No dispensing,Burette tip blocked.Clean burette tip.
titrant expelled from
stopcock or piston
METTLER TOLEDO DL31/DL3859
Error messages, malfunctions and hardware tests
9.3Hardware tests
The menu "Hardware tests" is stored under the <i> key and you can use to check whether, e.g. the stirrer
and pump of the titrator are functioning correctly. If you order the connector set for the interface check, you
can also test the interface and the TTL inputs and outputs (DL38 only) of the titrator.
9.3.1User interface
Display test
5 displays appear in succession, 4 with different patterns, the last without a pattern. This allows you to
detect faulty pixels of the display. The test mode automatically switches off after approx. 20 seconds.
Keypad test
When you press a key, a numeric code assigned to this key appears which allows you to recognize a faulty key
(see following Table).
Run ............212 ................ 2................... 33
Beep tests
You test the internal beeper.
9.3.2Peripheral interfaces
Interface test
– Before the test
– plug the test socket 51107288 into the balance, computer or GA42 interface.
– plug the Centronics cable into the Centronics interface and the test plug 51107287 into the Centronics
socket of the cable.
After the test phase, either "Test successful" or "Test failed" appears.
60METTLER TOLEDO DL31/DL38
Error messages, malfunctions and hardware tests
Balance test
The menu option is faded out if no balance is defined in the Setup menu.
• The weight displayed on the balance must appear in the titrator display.
Printer test (GA42, EPSON or HP)
The appropriate printer must be attached. If the error message "Printer: No connection" appears, you can
quit this error message with <Esc>: With "OK", the titrator attempts to set up the connection to the printer
again.
• The letters and characters available in the titrator must be printed out.
9.3.3Inputs and outputs
TTL-I/O (DL38 only)
– Before the test, plug test plug 51107723 into the TTL-I/O interface.
After the test phase, either "Test successful" or "Test failed" appears.
Internal stirrer
You test the rotational movement and change in the speed of the magnetic stirring bar.
External stirrer
You test the rotational movement and change in the speed of the magnetic stirring bar with the attached
DV705 Titration Stand.
Internal pump
You check whether the pump functions:
•A "noise" must be heard.
• It must be able to dispense and siphon off when solvent and waste bottle are correctly attached.
External pump
You check whether the pump of the attached DV705 Titration Stand functions (see "Internal pump").
NoteFor dispensing, you must always press the lower button (⇓ arrow), for siphoning off the upper
button (⇑ arrow).
METTLER TOLEDO DL31/DL3861
Installation and maintenance
10Installation and maintenance
10.1Equipping the titration stand
On delivery, the titration vessel comprises O-ring, cover and threaded ring. The seal is fastened in the
titration stand. If you have to change the cover or replace the O-ring, the illustration will serve as an aid. The
titration arm can be turned in both directions.
Silicon tubing (133 mm)
Threaded ring with lug
Connector
Stopper ST 7.5
3-hole adapter
O-ring
Titration vessel
titration arm
1. Carefully slide the magnetic stirring bar supplied into
the titration vessel.
2. Insert the titration vessel with the lug of the threaded
ring into the middle groove of the titration arm and
push downward towards the stirrer. Fasten the screw.
3. Place the three-hole adapter, the ST stopper and the
connector in the cover openings.
4. Fill a drying tube with molecular sieves, close with
the cover and "push" into the titration stand to ensure tightness. Connect the cover and the connector
by silicon tubing.
Cover and drying tube
Seal
ScrewGroove of
Titration
stand
62METTLER TOLEDO DL31/DL38
Installation and maintenance
10.2Installing the burette
1. Screw the burette with holder onto the titrant bottle (see DV100x Operating Instructions).
– Never inhale the vapors of Karl Fischer reagents and avoid skin contact. The reagents are
toxic.
2. Place the drying tube holder in the opening of the burette holder, fill a drying tube with molecular sieves
and position in the holder.
3. Push the burette onto the titrator: The slot of the driver cam and piston rod must be parallel to the insert.
(If this is not the case, press the <Reset> key.)
4. Turn arrestment knob in the opposite direction to the arrow and when pushing in check whether the
piston rod of the burette is correctly positioned in the push rod.
5. Turn the arresting knob in the direction of the arrow to fasten the burette.
6. Place the dispensing tubing in the cover opening.
– When you rinse burette and tubing, place the dispensing tubing in a waste beaker;
do not dispense in the titration vessel!
Driver cam
Push rod of
the burette drive
Hole for dispensing tubing
Arrestment knob
METTLER TOLEDO DL31/DL3863
Installation and maintenance
10.3Installing the solvent and waste bottles
Screw sleeve
Socket
Screw cap
Gasket
Dispensing or
suction tubing
Adjusting sleeve
1. Open the solvent bottle, lay in the gasket and screw
on the screw cap (you may need an adapter, see
DV100x Operating Instructions).
2. Loosen the screw sleeve and slide over one of the
pump tubing. Push this dispensing tubing to the base
of the bottle and screw fasten the sleeve. (There is
an O-ring in the socket which ensures tightness.)
3. Push the other end of the dispensing tubing into the
opening on the left of the three-hole adapter and fasten with the adjusting sleeve.
4. Perform steps 1 and 2 with the waste bottle supplied. Push the suction tubing to just shortly beneath
the screw cap and fasten the sleeve.
5. Push the other end of the suction tubing into the
opening on the right of the three-hole adapter and
fasten with the adjustment sleeve so that you can
easily push the tubing to the base of the titration
vessel and pull it up again.
Teflon seal
Output
fitting
Suction
fitting
6. Fill two drying tubes with molecular sieve and place
them in the screw cap of the bottles.
7. Push the silicon tubing (450 mm) over the drying
tube cover of the waste bottle and connect to the
suction fitting of the pump.
8. You can connect the output fitting to tubing to lead
off the solvent vapor to a suitable place.
Check all drying tubes and screw fittings once again for
firm seating to assure the tightness of the system!
64METTLER TOLEDO DL31/DL38
10.4Attaching the magnetic stirrer, electrode and devices
Installation and maintenance
Balance (RS232C)
External stand (DV705)
Stirrer output
Electrode input (KF)
Pump unit
Suction fitting
Output fitting
Socket for
external power pack
Magnetic stirrer
On delivery, the magnetic stirrer is pushed into the slot for the
burette drive. Attach its cable to the stirrer output. The arrow on
the Mini-DIN connector must point towards the screw.
GA42 printer (RS232C)
Computer (RS232C)
Stromboli, TBox, etc. (TTL I/O)
EPSON- and HP-compatible printers
(Centronics)
On/Off switch
Magnetic stirrer
– Make sure that every connector is in the correct position! Do not damage the connector
pins of the magnetic stirrer cable! Pins damaged by faulty insertion can cause a short
circuit at the "Stirrer" output. This renders the titrator non-functional.
Electrode
Unscrew the protective cover of the DM143-SC and connect the electrode to the triaxial cable. Plug the cable
into the KF input and insert the electrode in the corresponding cover opening of the titration vessel.
Balance, printer, computer, TBox DR42
Attach all peripheral devices to the same plug connector (phase) to which the titrator is connected!
Otherwise, equalizing currents could flow across the cables.
Attach the devices used to the corresponding connectors/sockets on the rear side of the titrator.
METTLER TOLEDO and SARTORIUS balances are compatible with these titrators.
METTLER TOLEDO DL31/DL3865
Installation and maintenance
10.5Using TTL inputs and outputs (DL38 only)
The TTL-I/O socket has two inputs and four outputs (In 1, In 2 and Out 1...Out 4). The outputs are occupied
by the titrator for commands.
If you have the METTLER TOLEDO oven sample changer Stromboli connected to the TTL I/O interface, no
further entries are necessary.
Inputs
A level change from "high" to "low" is detected at the inputs.
In 1In the Run menu
• the pretitration in KF titrations or the titration in bromine determinations is started,
• the sample determination in KF titrations is started (always corresponds to the <F3> key),
• the actual analysis is started (corresponds to confirming the sample addition),
• the result display is confirmed.
In 2corresponds to the <Reset> command on the keypad.
Outputs
At output 1 a pulse is given with a length of 100 – 200 ms. The other outputs assume either an active
("low") or inactive ("high") condition.
Out 1outputs a pulse before the display of "Add sample" and before display of the result list.
Out 2is inactive when an analysis can be started.
Out 3is active in the case of an error message; the condition does not change until the message has
been confirmed with "OK".
Out 4is active only during the mix time defined in the method.
For this condition, the connection of the TBox DR42 is suitable, it can be used to control an
homogenizer (see TBox Operating Instructions).
10.6Maintenance and servicing
Cleaning
–You can clean the titrator with a cloth moistened with alcohol.
–To clean the titration vessel, turn the titration arm to the right and unscrew the vessel from the cover.
–To clean the cover, loosen the fastener screw, take the titration vessel with cover up and out and unscrew
cover.
–You should clean the electrode when their response time deteriorates owing to certain Karl Fischer
reagents and samples (the end point drifts!): Either place in an ultrasound bath with deionized water for
a few minutes or for 60 seconds in a chromic acid bath and then rinse with water or ethanol.
66METTLER TOLEDO DL31/DL38
Installation and maintenance
O-rings
You should replace the O-rings of the screw caps when the system is no longer tight:
– Loosen the screw sleeve, take out the pump tubing and remove the O-ring using tweezers (see also
illustration in Section 10.3).
You should replace the O-ring of the threaded ring for the same reason:
– Loosen the fastener screw of the titration vessel, take titration vessel with cover up and out and unscrew
(see illustration in Section 10.1).
Dispensing and suction tubing
You should replace the pump tubing when the Teflon seal is no longer properly tight (wear and tear).
Molecular sieve
If the drift of the KF solution remains too high, you should fill the drying tubes with fresh molecular sieve.
You can regenerate molecular sieve in a drying oven at 160 – 300 °C for at least 24 hours. As the molecular
sieve used on the waste bottle also contains SO2, you should rinse it with deionized water before regeneration.
Pump unit
If the diaphragm pump no longer functions, you can replace it:
– Pull out the pump unit and insert a new one.
Pump button
If the pump button becomes too brittle, you should replace it:
– Remove the button and place a new one over the ring.
Magnetic stirrer
If the magnetic stirrer is faulty, you can replace it:
– Swing out the titration arm and push the motor out of the burette slot.
– Push in the new magnetic stirrer and attach its cable to the stirrer output.
Battery
The battery installed in the titrator has a service life of approx. 5 years. The battery may be changed only by
METTLER TOLEDO service.
Disposal
Contact your METTLER TOLEDO dealer if you have to dispose of the titrator.
METTLER TOLEDO DL31/DL3867
Standard and optional equipment
11Standard and optional equipment
11.1Standard equipment
1 KF titratoraccording to order
1 Set Operating Instructions (DL31/DL38)according to order
1 Application brochureaccording to order
1 Memo cardaccording to order
1 CD Software LabX lightaccording to order
1 Set Operating Instructions (LabX light)according to order
1 Keypad cover
You can order each part listed with an order number from METTLER TOLEDO.
Note
The standard and optional equipment for the 5 mL burette are listed in the Operating Instructions
for the interchangeable burrettes enlosed in the burette package.
(the input signal for the titrator must be at least 150 ms)
OutOpen collector
• Vce max. = 24 V
• Ic max. = 20 mA
GNDSignal ground, galvanically connected to titrator electronics
1
Data ready
2
3
4
5
6
7
8
9
Data accepted
10
Printer busy
11
End of paper
12
Printer active
13
Automatic line feed
14
Printer error
15
Printer initialized
16
Printed selected
17
18
19
20
21
22
Signal ground, galvanically connected to titrator
electronics
23
24
25
Titrator
METTLER TOLEDO DL31/DL3877
Technical data
Declaration of System Validation
We herewith inform you that the products/systems
DL31, DL38
including software and accessories were developed, tested and successfully validated according to the
international ISO9001:1994 based Life cycle rules of Mettler-Toledo GmbH, Analytical.
Life cycle checkpoint details were reviewed and approved by the Project Supervisory Group (PSG). The
products/systems were tested to meet functional and performance specifications and release criteria at
release to shipment. In order to support GLP and validation requirements, we will make the following
documents available to an authorized, governmental or regulatory agency for inspection.
• Performance Specifications
• Software Specifications
• Quality Plan
• Project Management System
• Test Plan
• Customer Requirements
• Review Reports
• Source Code
Mettler-Toledo GmbH, Analytical will maintain possession of all documents and their reproductions and
may require a non-disclosure agreement to be provided by those requiring access to these documents.
Schwerzenbach,Dr. Bernhard GrobDr. Rolf Rohner
August 1998General ManagerMarketing Manager
Business Unit AnalyticalTitration
ISO 9001 certificate for METTLER TOLEDO
The Mettler-Toledo GmbH company, Greifensee, was examined and evaluated in 1991 by the Swiss
Association for Quality and Management Systems (SQS), and was awarded the ISO 9001 certificate.
This certifies that Mettler-Toledo GmbH, Greifensee, has a quality management system that conforms
with the international standards of the ISO 9000 series.
Repeat audits are carried out by the SQS at intervals to check that the quality management system is
operated in the proper manner and is continuously updated in relation to changes brought about.
78METTLER TOLEDO DL31/DL38
Product Life Cycle Model
Ideas
Technical data
Project Study
Project Start
Phase:
Basic Know how
Phase:
Feasibility
Phase:
Project Release
Product Guidelines
Performance Specifications
Quality Plan
Documentation Plan
Project Plan
Review Report
Software Specifications
Test Plan
Review Report
Confirmation
Review Report
Phase:
Project End
Usage
Readiness
Review Report
Customer Complaints (CRIS)
Customer Requirements
Phase out
METTLER TOLEDO DL31/DL3879
80METTLER TOLEDO DL31/DL38
13Index
Symbole
<F1>...<F3> 9
Index
A
Access for expert/routine 21
All results 52
Amount of sample (calculating) 24
Analyse (Run) 34
Arrow keys 11
Author 27
Auto start 25
Auxiliary function keys 10
Auxiliary functions 50
TBox DR42 (attaching) 65
Technical data 73
Termination parameters (Br No.) 31, 32
Termination parameters (KF) 25
Time 20
Titrant (method parameter) 24
Titration stand 19
Titration stand (equipping) 62
Titrator (overview) 7
Titrator ID 21
TTL I/O socket 77
TTL inputs & outputs 66
Tubing (replacing) 67
Tutorial 9, 56
Type (sample) 24
U
User interface (checking) 60
User list 20
W
Wait time 25
Waste bottle (installing) 64
Weight transfer 40
82METTLER TOLEDO DL31/DL38
e
e
METTLER TOLEDO DL31/DL38 KF titrators
Switching from routine to expert mode
SettingsSETUP
Titrator ID▲
Acces forRoutin
BeepY es
Solv. capac.[mL]80▼
HelpO K
SettingsSETUP
Titrator ID▲
Acces forRoutin
BeepY es
Solv. capac.[mL]80▼
HelpModifyO K
Select this line in the Setup menu.
–While holding down the <Shift> key, press key <1> three times:
"Modify" is assigned to <F2> allowing you to select "Expert".
– Confirm <F3>, "OK", to save the change.
✃
METTLER TOLEDO DL31/DL3883
84METTLER TOLEDO DL31/DL38
To protect your METTLER TOLEDO product’s future:
METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of
all METTLER TOLEDO products for years to come. Please send for full details about our attractive
terms of service.
Thank you.
Printed on 100% chlorine-free paper, for the sake of our environment.
Subject to technical changes and to the availability of