Setup
METTLER TOLEDO DL15 31
- Select "Buff. type" using the arrow keys. Then press Modify and select the buffer
type you are using.
- Select "Number of buffers " and press Modify.
- Add the desired number of buffer solutions to the buffer list using Add. You can
select these from a list and then press OK to include them in the buffer list. (You can
remove buffers again from the list by selecting them and pressing Delete.)
- When you have defined the desired number of buffer solutions, press OK to return to
the setup menu for the calibration.
- Select "Speed [%]" and enter the speed for the stirrer in percent.
- Select "Temp.sensor", press Modify and select the "PT1000" temperature sensor or
"Manual".
- If you have selected "Manual" for "Temp.Sensor", select "Temperature" and enter the
temperature in the corresponding units.
- Select "dE [mV]" and enter the millivolt number for the determination of the maxi-
mum drift dE/dt [mV/s] of the electrode potential.
- Select "dt [s]" and enter the time in seconds for the determination of the maximum
drift dE/dt [mV/s] of the electrode potential.
- Select "t(min) [s]" and "t(max) [s]" one after the other and define the time at which
the measured value at the earliest (tmin) and at the latest (tmax) is to be accepted.
- Select "Report" and select "none" or "printer" for the report printout using Modify.
- Start the sensor calibration by pressing Start.
- For the selected number of buffer solutions, a message appears each time reques-
ting you to prepare the corresponding buffer solution. You must confirm this by
pressing OK to start the measurement with the corresponding buffer solution.
- After all buffer solutions have been measured, a summary of the calibration results
appears on the display with the following information:
• Name of the sensor
• Calibration "OK" / "not OK"
• measured slope [mV/pH]
• measured zero point [pH]
If the measured slope is outside the limits defined for the sensor Min. [mV/pH] and
Max. [mV/pH], "Calibration not OK" appears and the calibration results are not accepted.
- If the measured slope lies within the limits defined for the sensor, the slope and zero
point for the sensor can be accepted by pressing Save. The date and time for the
calibration are automatically updated.
- If "Printer" was selected for the "Report" parameter, a report of the sensor calibration
is printed on the printer.
- Quit the display that follows by pressing Esc. in order to prevent the date and time
from being set to 00-00-0000 and 00:00.
• The sensor calibration can also be opened using the Sensor Help function key. (See:
"10.2.3 Calibrating a sensor")
• For storing into Setup, the slope is converted to 25 °C.