The titrators have been tested for the experiments and intended purposes documented in the
Tutorial and these Operating Instructions. However, this does not absolve you from the
responsibility of performing your own tests of the products supplied by us regarding their
suitability for the methods and purposes you intend to use them for. You should therefore
observe the following safety measures.
Measures for your protection
– Ensure that you plug the power cable supplied into a receptacle outlet that
is grounded! In the absence of grounding, a technical fault could be lethal.
– Switch the instrument off and disconnect the power cable before you open
Risk of electric
shock
the housin or change blown fuses! An electric shock could be lethal.
– Never work in an environment subject to explosion hazards! The housing
of the instrument is not gas tight (explosion hazard due to spark formation,
corrosion caused by the ingress of gases).
Risk of explosion
– Always test the titration vessel for firm seating in the titration head! If it falls
off, you could injure yourself if working with toxic titrants and solvents or
strong acids or bases.
Risk of corrosion
– When using chemicals and solvents, comply with the instructions of the
producer and the general lab safety rules! Additional safety precautions for
Karl Fischer titrations are described in Section 10.2.
Measures for operational safety
– Check the set operating voltage before you switch on the titrator (see
Section 11.1.5)! The instrument will be damaged if the operating voltage
does not match the line voltage.
Caution
– Use only fuses of the specified type if you need to change them!–Have
the instrument serviced only by METTLER TOLEDO Service!
– Always wipe off splashed liquids immediately! The instrument is not water-
proof.
– Exclude the following environmental influences:
• powerful vibrations,
• direct sunlight,
• atmospheric humidity greater than 80%,
• temperatures below 5 °C and above 40 °C,
• powerful electric or magnetic fields!
02/93METTLER DL77/DL70ES/DL67 1
11/97
Introduction
2
METTLER DL77/DL70ES/DL67
02/93
Introduction
Introduction
The DL77, DL70 ES, and DL67 METTLER titrators are microprocessor-controlled analytical
instruments that provide accurate and reproducible results thanks to their built-in intelligence.
With these titrators, you can perform end point, equivalence point and pH-stat titrations,
measure pH/mV and temperature, and determine TAN/TBN and p/m values. With the aid of
a polarization current source, you can determine water contents by the Karl Fischer method
(>2 mg H2O/sample). You can perform conductivity measurements and conductometric
titrations with an appropriate non-Mettler unit equipped with an analog output.
All titrators have a maximum of four inputs for electrodes, two for temperature sensors and
three 24-V outputs for stirrer, pump or valve attachments. They have a maximum of four
RS232C/CL interfaces, enabling you to connect a series of peripheral instuments:
- an attached METTLER balance transfers the sample weight automatically,
- a printer records the desired results,
- a color terminal serves as a second display and can be used for remote control, or a PC/robot
can interchange data with the titrator, and
- with an attached METTLER Sample Changer, each titrator is turned into a titration system
for the automatic analysis of whole series of samples.
How the three titrators differ from each other
DL77:Two titrations can be executed simultaneously and two sample changers can be
connected. Neither the DL70ES nor the DL67 allow this.
DL77/These titrators can control a maximum of four burette drives, the DL67 two.
DL70ES:
These instruments know the most important titrants and all METTLER sensors,
whereas the DL67 knows one titrant and one sensor.
With both titrators a maximum of 10 titration methods can be entered in a list and
processed in succession. With the DL67 one method can be entered at a time.
While a titration is running, you can develop and save a new method, or you can
enter sample data for the next titration method. This is not possible with the DL67.
02/93METTLER DL77/DL70ES/DL67 3
Introduction
Operating concept
What information will you find where?
1.The TUTORIAL, provided with the standard equipment, will help you to overcome any
inhibitions you may have with regard to the new instrument. You will get to know the
function keys, the keypad and the display. Using a stored method for an acid-base titration,
you will perform your first analysis.
2.These Operating Instructions provide a complete description of the concept and oper-
ating characteristics of the three titrators.
The operating concept and a compilation of the key combinations follow this introduction.
3.The RS232C Interface Description, i.e., a detailed description of the communication bet-
ween titrator and computer, is provided with the standard equipment (since June, 1999).
Note: These Operating Instructions apply to software version 3.0 or 3.1.
The organization of the Operating Instructions
The organization is based on a modular principle. This allows supplementation or the interchange of individual sections or pages: new text sections carry the date of issue (in the footer
on every page).
The operating concept of the titrator
The operation of the titrator is menu driven. What does this mean?
The titrator handles various tasks:
It stores, for instance, titrant names with the corresponding concentrations; it knows that it must
terminate the titration after an equivalence point has been found; it rinses burettes and can
provide records of stored data on an attached printer.
We call the listing of these various tasks a menu. Each task is subdivided further. If you select
a task from the main menu you are shown a new set of tasks – the submenu – from which you
can select another task. We refer to this as a menu tree when the main menu branches into
submenus and these in turn into additional submenus.
4
METTLER DL77/DL70ES/DL67
05/99
02/93
Introduction
02/93METTLER DL77/DL70ES/DL67 5
Introduction
6
METTLER DL77/DL70ES/DL67
02/93
Introduction
Key combinations
Rapid menu change with key combinations
In the Tutorial, you have already become acquainted with several key combinations which you
can employ to obtain a certain menu immediately. Using these keys, you can reduce the
number of keystrokes before and during the titrations considerably. The key combinations
consist of the index key and a letter key which must be pressed simultaneously. Key combinations exist to
•jump from a submenu into the main menu or from a submenu into that of another branch,
or to
•trigger commands for the burette or the printer (see below)
If you press a key combination in a submenu, the DL70 stores all changes that you have made
up to this point in the menu. The selector bar can be on any line of the menu.
Key combinations for commands
+
U
+
Y
+
L
+
P
+
C
Rinse burette (see Section 4.1.1)
Rinse tip (see Section 4.1.2)
Trigger line feed on the printer (see Section 1.8.1)
Trigger form feed on the printer (see Section 1.8.1)
The current display is printed (copied!)
Key combinations for menu change
MAIN MENU
INSTALLATION
EDITOR
+
M
ANAL YSIS A
ANALYSIS B*
AUXILIARY FUNCTIONS
DOCUMENTATION
USER LEVEL
REMOTE CONTROL
In order to perform titrations the titrator must be acquainted with the titrants and their
concentration, the sensors with their possible unit of measurement, and the solvents that it can
dispense by means of pumps. It must know the burette drive on which the burette is located,
the input to which the sensor is connected and what output carries the stirrer. It needs the
names of the attached units such as a balance or printer in order to transfer data. In this menu
you enter and store the names of all chemical and mechanical resources: you install them.
The most common titrants, solvents and all METTLER sensors are already installed in the
titrator. Not only can you delete these resources or modify their parameters, but you also have
the possibility to install new ones.
Caution: All resources needed for the METTLER methods stored in the application data
base are installed accordingly. If you delete one of these, the titrator will wait until
start of titration of a METTLER method before outputting the error message that the
resource is not installed.
DL67: Only the titrant and the sensor for METTLER method M001 are stored in the DL67.
02/93METTLER DL77/DL70ES/DL67 1-3
Titrants
INSTALLATION
1.1Titrants
When you select this menu you receive the installed titrants with the specified parameters
concentration and burette drive. (You will find the menu tree depicted at the end of Section
1.1.2.)
NaOH0.1 mol/LDrive 3
HCl
HClO
4
etc...
DL67: The DL67 only has stored the titrant NaOH. You can, however, add titrants to the list
(see Section 1.1.3).
If you select, for instance, NaOH the list will be masked on the right by a selection menu
containing the following commands:
0.1 mol/LDrive 3
0.1 mol/LDrive 3
Delete
Modify
Add
1.1.1Delete
Position the selector bar on this command and confirm with RUN. The mask of the selection
menu disappears, the titrant NaOH is deleted.
Note: You can also delete a titrant directly from the list by pressing the <–> (minus) key. You
can reinstall NaOH with the command Add (see Section 1.1.3).
1.1.2Modify
If you select this command you are shown the parameter mask of the titrant (see next page).
If you move the selector bar to one of these parameters you can modify its name or value.
Note: a. Default values are stored in the titrator for all parameters, e.g. for the concentration
of the titrant 0.1 mol/L, for the titer 1.0.
1-4
• These values will be overwritten once you have entered new ones.
• If you want to modify only one digit of a value, you must first indicate the digit with
the cursor (with → or ←) prior to entering the new one.
METTLER DL77/DL70ES/DL67
02/93
INSTALLATION
Titrants
Note: b. With many parameters the titrator has selection or recommendation menus in which
you need only select the values or names. If a parameter has one of these menus,
each time it is selected an arrow symbol (➡) appears at the extreme right in the
middle of the display. You can either accept the value or name following the
parameter or:
in the case of Recommendation,
– overwrite directly with a new entry or
– press SEL to select a new value or name from the recommendation menu that
appears.
in the case of Selection,
– press SEL to select a new value or name from the selection menu that appears.
You cannot enter the value or name yourself.
If only two names or values are possible for a parameter, these are toggled
1. You can overwrite the name or press SEL to call up the recommendation menu:
Recommendation menu: You can select a new titrant from the recommendation menu.
Position the selector bar on, for instance, HCl and confirm with RUN. The recommendation
menu disappears and HCl follows Name.
2. Enter the concentration [mol/L]
3. Enter the titer only if you know its value. If you determine the titer of the titrant with the
titrator, its value is entered here together with the date automatically (see
Titer
function,
Section 2.3.16).
4. Select the burette volume from the selection menu:
Selection menu: You press SEL, position the selector bar on, for instance, 5 mL in the
selection menu and confirm with RUN. The selection menu disappears and 5 mL now
follows burette volume.
5. Select the burette drive from the selection menu.
02/93METTLER DL77/DL70ES/DL67 1-5
Titrants
INSTALLATION
DL67: It is only possible to select one of two burette drives for this titrator. If more than two
drives have been installed, the titrator will recognize each of the first two (counting from
left to right).
6. You can neither delete nor overwrite the date. It refers to the titer determination of the titrant
using the method function
(see
Titer
function, Section 2.3.16).
Titer
and is entered here together with the time automatically
Note: As soon as you change a titrant parameter, the date and time are deleted.
The titrant with the modified parameters is installed when you quit the parameter mask with
EXIT. Here, the selector bar can be positioned on any parameter, but not on the title line.If you
confirm the title line with EXIT the following selection menu appears:
Save modifications?
Yes
No
If you confirm "Yes" with RUN, the titrator stores the modified values or names.
If you confirm "No" with RUN, the old values remain in force.
Note: If you press a key combination (<index + letter>) to quit the Installation menu, the
modified parameters are stored automatically if the selector bar is positioned on a
parameter line. If it is positioned on the title line, the selection menu "Save modifications?"
appears again (see above).
Enter new name or select
entry from recommendation
menu with SEL
Enter new concentration
NaOH
0.1
1.0
10 mL
Drive 3
(07-06-1992 12:20)
Titrant names
NaOH
HCl
HClO
4
etc.
Enter new titer or let it be
entered through titration
Select another burette
with SEL
Burette volume
1 mL
5 mL
10 mL
20 mL
Select another drive
with SEL
Burette drive
Drive 1
Drive 2
Drive 3
Drive 4
Date/time is entered only through titer determination by the titrator!
02/93METTLER DL77/DL70ES/DL67 1-7
Titrants
INSTALLATION
1.1.3Add
Select this command if you wish to add a titrant to the titrant list, for instance NaOH of
concentration 1.0 mol/L or with a different burette volume, or a new titrant. You can also add
a titrant directly to the list by pressing the <+> (plus) key. You are always shown the following
mask:
NameNaOH
Concentration [mol
/L]
0.1
Titer1.0
Burette volume10 mL
Burette driveDrive 3
Date/time00–00–0000 00:00
1. Select the titrant from the recommendation menu or enter the one you wish to install.
Note: If you install, for example, several NaOH solutions of the same concentration, you must
provide each name with a different flag to allow the titrator to distinguish between them,
e.g. NaOH/1.
2. Enter the concentration [mol/L].
3. Enter the titer only if you know its value. If you determine the titer with the titrator, its value
together with the date is entered here automatically (see
Titer
function, Section 2.3.16).
4. Select the burette volume from the selection menu.
5. Select the burette drive from the selection menu.
6. The date together with the time for the newly installed titrant is entered here automatically
only after the titer has been determined (see
Titer
function, Section 2.3.16).
The titrant with the appropriate parameters is installed when you quit the parameter mask with
EXIT (see the appropriate description at the end of Section 1.1.2).
1-8
METTLER DL77/DL70ES/DL67
02/93
INSTALLATION
Sensors
1.2Sensors
When you select this menu you are shown the installed sensors with the specified parameters
unit of measurement and the sensor input. (You will find the menu tree depicted at the end of
Section 1.2.2.)
DG111-SCpHSensor 1(Combined pH electrode – aqueous medium)
DG101–SCpHSensor 1(Combined pH electrode for small volumes in a
small titration vessel – aqueous medium)
DG113–SCmVSensor 1(Combined glass electrode with movable sleeve
frit – nonaqueous medium)
DG114–SCmVSensor 1(Combined glass electrode with movable sleeve
frit – aqueous medium)
DG115–SCmVSensor 1(Combined glass electrode with sleeve frit –
aqueous medium)
DM140–SCmVSensor 2(Combined platinum ring electrode – redox
titrations)
DM141–SCmVSensor 2(Combined silver ring electrode – argentometry)
DM142mVSensor 1(Double-pin platinum electrode – voltametry)
DP550% TSensor 2(Phototrode – transmission measurements at
550 nm)
DP660% TSensor 2(Phototrode – transmission measurements at
660 nm)
DL67: Only sensor DG111-SC is stored in the DL67. You can, however, add sensors to the
list (see Section 1.2.3).
Note: There is no need to install a reference electrode as it is part of the installed measuring
electrode at input sensor 1. The input for the reference electrode is marked (see Section 11.1.4).
If you select DG111-SC, for instance, the list is masked on the right by a selection menu
containing the following commands:
Delete
Modify
Add
02/93METTLER DL77/DL70ES/DL67 1-9
Sensors
INSTALLATION
1.2.1Delete
Position the selector bar on this command and confirm with RUN. The mask of the selection
menu disappears, the DG111-SC sensor is deleted.
Note: You can also delete a sensor directly from the list by pressing the <–> (minus) key. You
can reinstall the DG111-SC with the Add command (see Section 1.2.3).
1.2.2Modify
If you select this command you are shown the parameter mask of the sensor in which you can
change the parameter values:
NameDG111-SC
Unit of measurementpH
Sensor inputSensor 1
Zero point [Unit]7.0
Slope [mV/Unit]–59.16
Temperature [°C]25.0
Date/time(e.g.: 02–06–1992 10:15)
1. Select the sensor name from the recommendation menu or enter the one you wish to install
in place of the DG111-SC sensor.
2. Select the unit of measurement suitable for the sensor from the selection menu. Within a
method you can later again choose between the selected unit of measurement and "mV"
in the functions Measure, Titration and pH/mV-stat (see Section 2.3.4/ 12/13).
mV:Either there is no other unit of measurement suitable for the sensor, or you
require only mV.
If you select the unit mV for a sensor, the calibration parameters zero point,
slope and temperature are ignored by the titrator.
pH:After you select pH you then enter the calibration parameters for a pH electrode
or you let the titrator do it (see notes a. and b. on page 1-12).
pM:M represents any cation. After you select pM you then enter the calibration
parameters for a pM electrode or you let the titrator do it (see notes a. and b.
on page 1-12).
pX:X represents any anion. After you select pX you then enter the calibration
parameters for a pX electrode or you let the titrator do it (see notes a. and b.
on page 1-12).
1-10
METTLER DL77/DL70ES/DL67
02/93
INSTALLATION
Sensors
%T:After you select %T you then enter the calibration parameters of a phototrode
(unit of measurement: transmission, see
Operating Instructions "Phototrode"
A:After you select A you then enter the calibration parameters of a phototrode
(unit of measurement: transmission. The absorption (A = -log T) is calculated,
see
Operating Instructions "Phototrode"
).
µS/cm:After you select µS/cm you then enter the calibration parameters of the conduc-
tivity cell.
mS/cm:After you select mS/cm you then enter the calibration parameters of the conduc-
tivity cell.
Note: Conductivity measurements and conductivity titrations can be performed using a
conductometer equipped with an analog output.
3. Select the sensor input from the selection menu.
).
Notes: a. In the standard configuration of the titrator you have the sensor 1 and sensor 2 input
available. You should select the sensor 1 input for all glass electrodes as this has
the highest input resistance (see
Technical data
, Section 11.2.1)
Rear view of the titrator
.
, Section 11.1.4 and
b. For technical reasons, only the "low resistance" sensor input 2 may be selected for
phototrodes (see
Section 11.2.1)
Rear view of the titrator
.
, Section 11.1.4 and
Technical data
Caution: The RS option has an additional "low resistance" sensor input, the temperature
option an additional high resistance sensor input (see Sections 11.1.6 and 11.1.7).
Depending on the installation manner, the "low resistance" input may, for example,
be either sensor 3 or sensor 4! (See
Technical data
, Section 11.2.1).
Rear view of the titrator,
Section 11.1.4 and
4. Enter the zero point of the sensor:
The zero point of a sensor is the measured value at which it displays 0 mV potential (zero
point of electrode assembly).
• For the zero point of a pH electrode this is pH0 with unit pH.
,
• For the zero point of an ion-selective electrode this is pM0 with the unit pM, or pX0 with
the unit pX.
• The zero point of the phototrode is normally 0% T (100% T = 1000 mV).
• The zero point of a conductivity cell is normally 0 µS/cm or 0 mS/cm.
02/93METTLER DL77/DL70ES/DL67 1-11
Sensors
INSTALLATION
5. Enter the slope of the sensor.
The slope of a sensor is the potential change in mV per unit of measurement.
• For the slope of a pH electrode the unit is: mV/pH.
• For the slope of an ion-selective electrode the unit is: mV/pM or mV/pX.
• The slope of a phototrode is normally 10 mV/ %T.
• For the slope of a conductivity cell the unit is: mV/µS*cm
-1
or mV/mS*cm-1.
6. Enter the temperature: The actual calibration temperature is either
• defined by you when starting a calibration method (see Section 3.1) or
• measured automatically if you have attached and installed a temperature sensor (see
Sections 1.3, 2.3.2 and 2.3.17).
7. You can neither delete nor overwrite the date. It refers to the calibration of the sensor using
the method function
(see
Calibration
Calibration
and is entered here together with the time automatically
function, Section 2.3.17).
Notes: a. The calibration parameter values of the titrator installed in the factory are theoretical
values for a new sensor. You have to perform a calibration of your sensor if you wish
to determine accurate values (see
Calibration
function, Section 2.3.17).
b. When a pH, pM or pX sensor is calibrated the calibration parameters (zero point,
slope, temperature) are entered here automatically.
c. As soon as you change a sensor parameter, the date and time are deleted.
Caution: Do not transfer the calibration data obtained for glass electrodes attached to sensor
input 1 (or for the temperature option sensor input) to the electrodes you attach to
sensor input 2 (or the sensor input of the RS option)! Recalibrate these sensors to
obtain correct values!
The sensor with the modified parameters is installed when you quit the parameter mask with
EXIT (see the appropriate description at the end of Section 1.1.2).
Sensor parameters
Name
Unit of measurement
Sensor input
Zero point [unit]
Slope [mV/unit]
Temperature [°C]
Date/time
Enter new name or select entry from
recommendation menu with SEL
DG111-SC
pH
Sensor 1
7.0
-59.16
25.0
(02-06-1992 10:15)
Selection
Delete
Modify
Add
The sensor is
deleted
Sensor names
DG111-SC
DG101-SC
DG113-SC
etc.
Select another unit of measurement
with SEL
Select another input with SEL
Enter new zero point or let it be
entered through calibration
Enter new slope or let it be entered
through calibration
Enter new temperature
Date/time is entered only through calibration by the titrator!
Units of meas.
mV
pH
pM
etc.
Sensor inputs
Sensor 1
Sensor 2
Sensor 3
Sensor 4
02/93METTLER DL77/DL70ES/DL67 1-13
Sensors
INSTALLATION
1.2.3Add
Select this command if you wish to add a sensor to the sensor list, for instance a
DG111-SC sensor with different calibration parameters or a sensor not yet installed. You can
also add a sensor directly to the list by pressing the <+> (plus) key. You are always shown the
following mask:
NameDG111-SC
Unit of measurementmV
Sensor inputSensor 1
Zero point [Unit]7.0
Slope [mV/Unit]–59.16
Temperature [°C]25.0
Date/time00–00–0000 00:00
1. Select the sensor name from the recommendation menu or enter the name of the sensor
you wish to install.
Note: If you install, for example, several DG111-SC sensors, you must provide each name
with a different flag to allow the titrator to distinguish between them, e.g. DG111/2.
2. Select the unit of measurement suitable for the sensor from the selection menu.
3. Select the sensor input from the selection menu.
4. Enter the zero point of the sensor.
5. Enter the slope of the sensor.
6. Enter the temperature.
(see explanation of the parameters and note under
Modify
).
7. The date together with the time for the newly installed sensor is entered here automatically
only after the sensor has been calibrated (see
Calibration
function, Section 2.3.17).
The sensor with the appropriate parameters is installed when you quit the parameter mask with
EXIT (see the appropriate description at the end of Section 1.1.2).
1-14
METTLER DL77/DL70ES/DL67
02/93
INSTALLATION
Temperature sensors
1.3Temperature sensors
When you select this menu you are shown the installed temperature sensors with the specified
parameters sensor type and sensor unit:
You can neither delete a temperature sensor nor add a new one to the list. You can, however,
install several Pt100 or Pt1000 at the same temperature sensor input (Temp 1 or Temp 2), for
example.
If you select TEMP A, for instance, you are shown the parameter mask of the Pt sensor in which
you can change the parameter values or names:
NameTEMP A
Sensor typePt100
Sensor inputTemp 1
Zero point [°C]0.0
Date/Time00–00–0000 00:00
1. You can not change the sensor name: The line will be skipped when the selector bar is
moved.
2. Select the sensor type with SEL: "Pt100" or "Pt1000" (see Section 11.1.7).
3. Select the sensor input with SEL: "Temp 1" or "Temp 2" (see Section 11.1.4).
4. Enter the zero point of the Pt sensor.
The zero point of a Pt sensor is the deviation of a measured value from a reference value,
such as 0 °C for an ice bath.
5. You can neither delete nor overwrite the date. It refers to the calibration of the Pt sensor
using the auxiliary function
Calibration temperature sensors
and is entered here together
with the time automatically (see Section 4.7).
Notes: a. The installed zero point is the theoretical value for a new Pt sensor. You have to
perform a calibration of your sensor if you wish to determine an accurate value.
When calibrating, the zero point is automatically entered here together with the date
(see auxiliary function
Calibration temperature sensors
, Section 4.7).
b. As soon as you change a parameter of the temperature sensor, the date and time
are deleted.
02/93METTLER DL77/DL70ES/DL67 1-15
Temperature sensors
INSTALLATION
The Pt sensor with the modified parameters is installed when you quit the parameter mask with
EXIT (see the appropriate description at the end of Section 1.1.2).
Menu tree
Temperature sensors
TEMP A
TEMP B
TEMP C
TEMP D
Pt100
Pt100
Pt1000
Pt1000
Temp 1
Temp 2
Temp 1
Temp 2
Temperature sensors
Name
Sensor type
Sensor input
Zero point [°C]
Date/time
The name cannot be changed
Select the other type with SEL: Pt1000
Select the other input with SEL: Temp 2
Enter new zero point or let it be entered
through calibration!
TEMP A
Pt100
Temp 1
0
00-00-0000 00:00
1-16
Date/time is entered only through
zero-point calibration by the titrator
METTLER DL77/DL70ES/DL67
02/93
INSTALLATION
Auxiliary reagents
1.4Auxiliary reagents
When you select this menu you receive the installed auxiliary reagents with the specified
parameters dispensing rate and auxiliary output. An auxiliary reagent is a solvent that is
dispensed using a time-controlled device, e.g. a diaphragm pump or an electromagnetic valve.
(You will find the representation of the menu tree at the end of Section 1.4.3).
H2O250 mL/minAux.2
CH3OH250 mL/minAux.2
CHCl
3
etc.
DL67: Only auxiliary reagent H2O is stored in the DL67. You can, however, add auxiliary rea-
gents to the the list (see Section 1.4.3).
If you select H2O, for instance, the list is masked on the right by a selection menu containing
the following commands:
250 mL/minAux.2
Delete
Modify
Add
1.4.1Delete
Position the selector bar on this command and confirm with RUN. The mask of the selection
menu disappears, the auxiliary reagent H2O is deleted.
Note: You can also delete an auxiliary reagent directly from the list by pressing the <–> (minus)
key. You can reinstall H2O with the command Add (see Section 1.4.3).
1.4.2Modify
If you select this command you are shown the parameter mask of the auxiliary reagent in which
you can change the parameter values:
1. Select the name of the auxiliary reagent from the recommendation menu or enter the name
of the reagent you wish to install in place of H2O.
02/93METTLER DL77/DL70ES/DL67 1-17
Auxiliary reagents
INSTALLATION
2. Enter the dispensing rate [mL/min] of the device.
Notes: a. The titrator uses the dispensing rate to calculate the volume to be dispensed for
time-controlled pumps or electromagnetic valves (see functions Pump and Rinse,
Sections 2.3.8 and 2.3.9).
b. The dispensing rate of each device must be determined experimentally in advance
for each solvent:
– Add the particular solvent using the dispensing device to a measuring cylinder
within 1 minute (stopwatch) and note the volume.
– Repeat this procedure for, e.g. periods of 20, 30 and 40 seconds.
– Use the different volumes to calculate the mean value per minute and enter this
value.
3. Select the auxiliary output from the selection menu.
Note: The titrator has three auxiliary outputs (Aux. 1, 2 and 3). One auxiliary output of the
titrator is normally assigned to the stirrer, e.g. Aux. 1 (see Section 1.7). You should thus
select only Aux. 2 and/or Aux. 3 for this auxiliary reagent (see
Technical data
, Section
11.2.2).
If you have attached the sample changer (ST20A or ST20), this unit offers you two additional outputs called "RINSE" and "DOSE".
The auxiliary reagent with the modified parameters is installed when you quit the parameter
mask with EXIT (see the appropriate description at the end of Section 1.1.2).
1.4.3Add
Select this command when you wish to add an auxiliary reagent to the list. You can also add
an auxiliary reagent directly to the list by pressing the <+> (plus) key. You are always shown
the following mask:
1. Select the name from the recommendation menu or enter the reagent you wish to install.
Note: If you install several auxiliary reagents with the same name, you must provide each one
with a different flag to allow the titrator to distinguish between them, e.g. H2O/2.
2. Enter the dispensing rate [mL/min] of the device.
1-18
METTLER DL77/DL70ES/DL67
02/93
Loading...
+ 320 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.