Operating Instructions for
Probe LE703
This Conductivity Measuring Cell utilizes the 4pole potentiometric method for measuring
conductivity, which incorporates a series of four
graphite poles embedded in the probe shaft.
This design completely eliminates the
polarization, which frequently occurs with the
2-plate amperometric method during measuring high conductivity solution.
Furthermore, combined temperature sensor
makes LE703 automatic temperature
compensating.
It is a general probe, good for lab use.
Order Number:
51340335
5. Allow sufficient time for the probe to stabi-
lize when measuring samples at different
temperatures.
6. The probe is not recommended for low
conductivity solutions. (
<10 S/
cm)
μ
7. Clean the probe with distilled water after
use.
8. For calibration and measurement proce-
dures, refer to the Instruction manual
supplied with conductivity meter.
Reagents and Supplies
1413 S/cm conductivity standard
μ
- 250 ml bottle, order No. 51300138
- Pack of 30 sachets, order No. 51302049
12.88 mS/cm conductivity standard
- 250 ml bottle, order No. 51300139
- Pack of 30 sachets, order No. 51302050
METTLER TOLEDO AG, Analytical
Sonnenbergstrasse 74
CH-8603 Schwerzenbach
Tel: +41 44 806 77 11
Fax: +41 44 806 73 50
http://www.mt.com
© Mettler-Toledo AG 2007
*12107277*
*12107277*
Operating Instructions
For optimum performance, use the following
operating procedures:
1. Remove all packaging material from probe
before use.
2. To prevent carry-over from solution to solu-
tion, rinse the probe with distilled water be-
tween measurements.
3. When measuring, make sure the solution is
above the cell chamber.
4. Ensure the cell chamber is free of bubbles
when measuring.
Precautions and Limitations
1. Do not expose the sensor to organic
solvents, either when cleaning or when
taking measurements.
2. Do not use the probe outside the recom-
mended temperature range.
3. Specification and performance of this probe
is guaranteed only when used with the ap-
propriate model of Mettler-Toledo conduc-
tivity meter.
Temperature Sensor
Cell Chamber