Mettler Toledo College B303, College B502, College B154, College B1302 DeltaRange, College B3001 Operating Instructions Manual

...
Operating instructions
METTLER TOLEDO B balance lines
•College-B
•CB
•GB
Short-form operating instructions
display appears Automatic procedure
Simple weighing
g
0.00
g
1182.03
g
1250.00
Taring
g
=00
g
=00
g
12+01
#
g
9%97
Calibration (adjusting)
Unit switching
Switching off
Switching on
CAL
------
2000.00
------
0.00
------
CAL donE
Cal/Menu
0.00
g
88888888
g
0.00
On
------
OFF
Off
On
#
ct
110.00
g
22.00
dwt
14.14
1
4.2.3 Unit 1, 2, 3 – selecting unit 1, 2, 3 with CB and GB balances ......14
4.2.4 A. oFF – setting automatic shutdown..........................................14
4.2.5 Selecting peripheral device........................................................15
4.2.6 Selecting data transfer mode .....................................................15
4.2.7 Selecting data transfer format ....................................................15
4.2.8 Setting baud rate (data transmission rate) ..................................16
4.2.9 Setting parity/bits .....................................................................16
4.2.10 Setting handshake ...................................................................16
5 Technical data and optional equipment ...................................17
5.1 Technical data.........................................................................17
5.2 Interface..................................................................................19
5.3 Optional equipment..................................................................19
6 Appendix .............................................................................. 20
6.1 Printout examples with LC-P45 and GA42 Printers ......................20
6.2 Preventive maintenance............................................................21
6.3 What if...?...............................................................................22
6.4 Index......................................................................................23
Contents
1 Getting to know the B balance lines ..........................................2
1.1 Introducing the B balance lines.................................................... 2
1.2 Layout of the B balances ............................................................3
1.3 The keys of the B balances (overview) .........................................4
2 Startup ....................................................................................5
2.1 Unpacking / Standard equipment.................................................5
2.2 Cautionary notes........................................................................ 5
2.3 Setting up .................................................................................6
2.4 Calibration (adjusting) ...............................................................8
3 Weighing ................................................................................9
3.1 On/off switching......................................................................... 9
3.2 Simple weighing........................................................................9
3.3 Taring..................................................................................... 10
3.4 METTLER TOLEDO DeltaRange balances.....................................10
3.5 Switching of weight units ..........................................................11
4 Menu .................................................................................... 12
4.1 Overview ................................................................................12
4.2 Description of the menu options ................................................13
4.2.1 Reset and printout of the balance setting..................................... 13
4.2.2 Adjusting the vibrations adapter on CB and GB balances..............13
2
1.1 Introducing the B balance lines
Three specially oriented B balance lines
We have derived three lines for special application fields from the standard range of the B balances: School balances (College-B), carat balances (CB) and gold balances (GB). These balances are geared to the special requirements of the respective segments with regard to type of weighing and functions.
METTLER TOLEDO DeltaRange balances have a movable fine range with 10 times smaller display increments (section 3.4).
All B balances are fitted with an RS232C interface as standard.
Notes
• Certified versions of the CB and GB balances are also available, please ask your METTLER TOLEDO dealer for details.
• If you wish to build on what you have learned about weighing in these operating instructions, you will find valuable tips in the booklet "Weighing the right way" 720906.
1 Getting to know the B balance lines
GB
College-B CB
College-B
College-B GB
3
1.2 Layout of the B balances
1 Keys 2 Display 3 Model plate
1)
4 Weighing pan/platform 5 Draft shield 6 Certification switch for weights and measures inspector and service engineer 7 Leveling feet 8 AC adapter socket
9 Provision for antitheft device 10 Leveling control 11 RS232C interface
Keys, operation and display are identical for all B balances.
1)
With details of: Max = maximum capacity
d = readability
*
Min = minimum capacity (recommended minimum load for certified balances)
*
e = verification scale interval (smallest display increment tested in certification) * Relevant only for certified balances
4
7
3 2
1
5 4
3
2 1
4
98
10
11
10 7
7
6
9
4
1.3 The keys of the B balances (overview)
B balances have two operator control levels: the weighing mode and the menu. The keys have different meanings, depending on the operator control level and how long a key is pressed.
Weighing mode (operation)
On
• Switch on
#
• Tare
• Switch
• Change settings
• Print
• Confirm settings
• Switch off
• Calibrate (adjust)
• Call menu
press and hold
press and hold key until the desired display appears
Menu (called up with
Cal
)
• Store and quit menu
press and hold
press briefly
press briefly
Off Cal
On
0/T
press key briefly
#
Off
On
Cal
Cal
#
• Abort
• Change settings
• Select menu options
Cal
5
2 Startup
2.1 Unpacking / Standard equipment
All B balances are supplied in an environmentally harmless package. The standard equipment of the B balances includes
AC adapter, to national codes,
Weighing pan,
operating instructions, to allow optimum utilization of the capabilities of your balance,
Description of interface commands (“MT-SICS Reference Manual, available in English only)
calibration weight, with CB balances and College B154 only,
draft shield, mounted with CB balances and College B154, for mounting by user with College B303,
carat pan, with CB balances.
Note
Calibration weights can be ordered from METTLER TOLEDO for all other B balances, see section 5.2.
2.2 Cautionary notes
• B balances may not be operated in hazardous areas.
1)
• Before attachment of the AC adapter, check whether the imprinted voltage value matches the local supply
voltage. If it does not, contact your local METTLER TOLEDO dealer.
1
)With the PS-EX2 power supply unit/AC adapter available as an option, all B balances can be used in
Ex zone 2 (section 5.3).
6
2.3 Setting up
The optimum location
The correct location makes an important contribution to the accuracy of the weighing results of high-resolution analytical and precision balances.
Hence, ensure a
• stable, vibration-free position as horizontal as possible. Avoid
• direct sunlight,
• excessive temperature fluctuations,
• drafts. The best position is on a stable bench in a corner protected against drafts as far as possible from doors, windows,
radiators or the ventilation slots of air conditioners.
Note
If vibrations can not be prevented, the balance can still provide accurate results if the vibration adapter is set accordingly, see section 4.2.2.
Antitheft device
B balances have provision for the attachment of an antitheft device. The following antitheft devices are available:
• Bolt and lock (not for GB16001) 229175
• Cable with lock (for all models) 590101
7
Leveling
B balances have a level control and adjustable leveling feet to compensate for slight irregularities in the weighing bench surface. The balance is exactly horizontal when the air bubble 1 is in the middle.
Procedure
Turn the two leveling feet 2 and 3 as shown in the diagram or described in the table so that air bubble 1 is in the middle. Air bubble at "12 o'clock" turn both leveling feet clockwise Air bubble at "3 o'clock" turn left leveling foot clockwise, right leveling foot counterclockwise Air bubble at "6 o'clock" turn both leveling feet counterclockwise Air bubble at "9 o'clock" turn left leveling foot counterclockwise, right leveling foot clockwise
Note
The balance must be releveled each time it is moved to a new location.
Connecting to the power supply
Before connection of the AC adapter, check that the imprinted voltage value matches the local supply voltage.
If this is not the case, please contact your local METTLER TOLEDO dealer.
Plug AC adapter into AC adapter socket of the balance and connect to the power supply.The balance performs a self-test. The test is finished when "OFF" appears.Press g briefly: balance is in operational readiness. Before any work is performed with the balance,
it should be calibrated (adjusted), (section 2.4).
Notes
To achieve accurate results with balances B154 and CB1503DR, these must be connected to the power supply for 20–30 minutes to warm up to the operating temperature. The PP-B10 PowerPack (rechargeable, external battery) can be used to operate all B balances independently of the power supply (section 4.2.4, 5.3).