Mettler Toledo BBA462, BBK462, IND469, SQC16 Operating Instructions Manual

Operating Instructions
METTLER TOLEDO SQC16
Compact Scales BBA462 / BBK462 Terminal IND469
IND469
Rear of Scale
Scale specifications (example)
19 20 2221
Max1: 3kg Min1: 20g e1: 1g d1: 1g Max2: 6kg Min2: 40g e2: 2g d2: 2g
Keypad
1
- , / &
2
a b c
3
d e f
4
g h i
5
j k l
6
m n o
7
p q r s
8
t u v
9
w x y z
i
.
+ ...
C
0
23
24 25
28
27
26
Overview
111213
14
15
16
16
17
18
10
10a
10b
9
9a
9b
1
2
3
4
5
6
8
9
7
Overview
1 Display 2 On/Off key 3 Zero key 4 Tare key 5 Enter or print key 6 Command softkeys (6) 7 Keypad 8 Adjustable feet 9 Weighing pan
9a: wind protector
9b: shield 10 Power supply:
10a: power cord (scale w/o battery)
10b: AC adapter (scale with battery)
Rear of scale
11 Power supply 12 Model plate 13 Hole for anti-theft device 14 Spirit level (only on certified scales and those with
MonoBloc weighing cells)
15 PS2 connector for keyboard and/or BCR (Barcode
reader)
16 COM3 (RS232C interface) 17 COM2 (RS232C interface) 18 COM1 (RS232C interface)
Scale specifications
19 Maximum load 20 Minimum load 21 Verification scale interval (certified scale) 22 Max. resolution
Keypad 23 Shift key (ABCÆabcÆ123ÆABC …) 24 Database key 25 Info key 26 Special characters key 27 Clear key
28
Numeric keys
Contents METTLER TOLEDO SQC16
Contents Page
1 Setting up the scale .....................................................................................................8
1.1 Safety and environment...............................................................................................................8
1.2 Positioning and leveling the scale................................................................................................9
1.3 Connecting to power supply.........................................................................................................9
2 Basic functions .......................................................................................................... 10
2.1 Switching on and off and zeroing ...............................................................................................10
2.2 Setting date and time ................................................................................................................10
2.3 Language settings .....................................................................................................................11
2.4 Simple weighing .......................................................................................................................12
2.5 Weighing with tare ....................................................................................................................13
2.6 Recording weighing results........................................................................................................14
3 User setup and password ...........................................................................................15
3.1 Defining a user name ................................................................................................................15
3.2 Creating a password..................................................................................................................16
4 Getting to know the SQC16 software ........................................................................... 17
5 The SQC16 Application ..............................................................................................18
5.1 Overview of article definition .....................................................................................................20
5.2 Creating an article for the first time ...........................................................................................22
5.3 The database ............................................................................................................................25
5.3.1 Creating a new article .................................................................................................................26
5.3.2 The 'Search' softkey ....................................................................................................................27
5.3.3 Editing the parameters of an existing article...................................................................................27
5.3.4 Copying the parameters of an existing article to a new article..........................................................28
5.3.5 Deleting an article.......................................................................................................................28
5.3.6 Printing the parameters of an article .............................................................................................28
5.4 Sampling of articles ..................................................................................................................29
5.5 Test series ................................................................................................................................30
5.6 Taring .......................................................................................................................................31
5.6.1 Manual tare ...............................................................................................................................31
5.6.2 Tare series .................................................................................................................................32
5.6.3 Bulk tare....................................................................................................................................32
5.7 Density .....................................................................................................................................33
5.8 Printing/Clearing statistics.........................................................................................................33
METTLER TOLEDO SQC16 Contents
Contents Page
5.9
The System settings ..................................................................................................................34
5.9.1 Tolerance specifications ..............................................................................................................36
5.9.2 Global density............................................................................................................................36
5.9.3 Sample messages ......................................................................................................................37
5.9.4 Tare configuration.......................................................................................................................37
5.10 The Function settings .................................................................................................................38
5.10.1 Violation check...........................................................................................................................39
5.10.2 2nd Tolerance System ..................................................................................................................40
5.10.3 Shared Statistics.........................................................................................................................41
5.10.4 Test...........................................................................................................................................41
5.10.5 Supplement ...............................................................................................................................41
5.10.6 Adjust........................................................................................................................................41
5.11 The Report settings....................................................................................................................42
5.11.1 Statistics and Sample reports.......................................................................................................44
5.11.2 Define contents of the report.........................................................................................................45
5.11.3 x-chart (Mean value trace) ..........................................................................................................48
5.11.4 Marginals ..................................................................................................................................48
5.11.5 Linefeeds ...................................................................................................................................48
5.11.6 A4 Printout.................................................................................................................................49
5.12 Printing summary of database ...................................................................................................50
6 Special features of SQC16 .........................................................................................51
6.1 Working with barcodes ..............................................................................................................51
6.2 Individual tare sampling ............................................................................................................51
6.2.1 Preweighing and backweighing in two samples.............................................................................51
6.2.2 Preweighing and backweighing in one sample ..............................................................................53
6.3 Sampling with batch..................................................................................................................54
6.4 Transfer key ..............................................................................................................................55
7 Vision Setup..............................................................................................................57
7.1 Overview and operation.............................................................................................................57
7.2 Calling up the menu and entering the password .........................................................................58
7.3 Menu overview..........................................................................................................................59
7.3.1 Scale.........................................................................................................................................59
7.3.2 SQC16 ......................................................................................................................................60
7.3.3 Terminal ....................................................................................................................................61
7.3.4 Communication..........................................................................................................................62
7.3.5 Diagnostics................................................................................................................................63
Contents METTLER TOLEDO SQC16
Contents Page
7.4
Scale settings (SCALE)..............................................................................................................63
7.4.1 Adjust/calibrate (SCALE Æ Calibration).........................................................................................64
7.4.2 Display resolution and weighing unit (SCALE Æ Display)...............................................................64
7.4.3 Automatic zero point correction (SCALE Æ Tare)............................................................................65
7.4.4 Automatic zero point correction (SCALE Æ Zero)............................................................................65
7.4.5 Automatic save of tare and zero values (SCALE Æ Restart).............................................................66
7.4.6 Adaptation to environmental conditions and weighing mode (SCALE Æ Filter)..................................66
7.4.7 Automatic adjustment (SCALE Æ FACT)........................................................................................67
7.4.8 Minimum weight (SCALE Æ Min Weigh).......................................................................................68
7.4.9 Resetting scale to factory default settings (SCALE Æ Reset) ............................................................69
7.5 SQC16 settings (SQC16)...........................................................................................................69
7.5.1 User Login (SQC16 Æ User Login)...............................................................................................69
7.5.2 Auto Log Out (SQC16 Æ Auto Log Out) ........................................................................................70
7.5.3 Minimum Password Length (SQC16 Æ Minimum PW Length)........................................................70
7.5.4 User Setup (SQC16 Æ User Setup)...............................................................................................70
7.6 Terminal settings for device (TERMINAL Æ Device) ....................................................................71
7.6.1 Language settings (TERMINAL Æ Device Æ Language) .................................................................72
7.6.2 Sleep function (TERMINAL Æ Device Æ Sleep)..............................................................................72
7.6.3 Adjusting the contrast of the display (TERMINAL Æ Device Æ Contrast)...........................................72
7.6.4 Invert (TERMINAL Æ Device Æ Invert) ..........................................................................................73
7.6.5 Changing the size of weight display (TERMINAL Æ Device Æ Weight display) .................................73
7.6.6 Adjusting the date and time (TERMINAL Æ Device Æ Date Time) ....................................................73
7.6.7 Adjusting the date and time (TERMINAL Æ Device Æ Beep) ...........................................................74
7.7 Define supervisor password (TERMINAL Æ Access) ....................................................................74
7.8 Reset terminal settings to factory settings (TERMINAL Æ Reset) .................................................75
7.9 Communication settings (COMMUNICATION)..............................................................................75
7.9.1 Mode (COMMUNICATION Æ Mode)..............................................................................................76
7.9.2 Parameters (COMMUNICATION Æ Parameters) .............................................................................76
7.9.3 Printer type (COMMUNICATION Æ Printer type)..............................................................................77
7.9.4 Header (COMMUNICATION Æ Define Header)................................................................................77
7.9.5 Add LineFeed (COMMUNICATION Æ Add LineFeed)........................................................................77
7.9.6 Reset communication (COMMUNICATION Æ Comx Æ Reset Com)..................................................78
7.9.7 PS2 Settings (COMMUNICATION Æ PS2)......................................................................................78
7.10 Diagnostic settings (DIAGNOSTICS) ...........................................................................................79
7.10.1 Keyboard (DIAGNOSTICSÆ Keyboard Test)...................................................................................79
7.10.2 Display (DIAGNOSTICSÆ Display Test) ........................................................................................80
7.10.3 Serial Number 1 (DIAGNOSTICSÆ SNR1) .....................................................................................80
7.10.4 Serial Number 2 (DIAGNOSTICSÆ SNR2) .....................................................................................80
METTLER TOLEDO SQC16 Contents
Contents Page
7.10.5
List 1 (DIAGNOSTICSÆ List 1).....................................................................................................81
7.10.6 List 2 (DIAGNOSTICSÆ List 2).....................................................................................................81
7.10.7 Reset All (DIAGNOSTICSÆ Reset All) ............................................................................................81
8 Other important information .......................................................................................82
8.1 Error messages in display .........................................................................................................82
8.2 Messages on printer..................................................................................................................83
9 APPENDIX: Tolerance systems.................................................................................... 86
9.1 Entry of tolerances ....................................................................................................................86
9.2 Legal tolerance systems ............................................................................................................86
9.3 Free tolerance systems with one tolerance .................................................................................87
9.4 Free tolerance systems with two tolerances................................................................................87
9.5 Free tolerance systems with three tolerances .............................................................................87
10 APPENDIX: Optional equipment ..................................................................................88
10.1 LC-I/O Relay Interface................................................................................................................88
10.2 Attachment of peripheral devices ...............................................................................................89
10.3 Accessories...............................................................................................................................90
11 Technical data ........................................................................................................... 91
11.1 General data and delivered items ..............................................................................................91
11.1.1 BBA/BBK462 .............................................................................................................................91
11.1.2 IND469.....................................................................................................................................92
11.2 Dimensions...............................................................................................................................93
11.2.1 BBA/BBK462 .............................................................................................................................93
11.2.2 IND469.....................................................................................................................................94
11.3 Interface technical data .............................................................................................................94
11.3.1 BBA/BBK462 .............................................................................................................................94
11.3.2 IND469.....................................................................................................................................95
11.4 Interface commands ..................................................................................................................95
11.4.1 Preconditions.............................................................................................................................95
11.4.2 SICS Command set.....................................................................................................................95
11.5 Table of Geo Values ..................................................................................................................99
11.5.1 GEO VALUES 3000e, OIML Class III (European Countries) ..............................................................99
11.5.2 GEO VALUES 6000e/7500e OIML Class III (Height £ 1000 m) .....................................................100
11.6 Declaration of Conformity ........................................................................................................101
Chapter 1: Setting up the scale METTLER TOLEDO SQC16
8
1 Setting up the scale
Please read through these operating instructions carefully and adhere to them at all times. If you discover that materials are missing or that the wrong ones have been supplied, or if you have any other problems with your scale, please refer to the dealer and salesperson concerned, or if necessary to the METTLER TOLEDO representative responsible.
1.1 Safety and environment
Do not use the scale in hazardous areas (unless it is clearly identified
as being approved for these areas).
If the power cord is damaged, the scale must no longer be operated.
Therefore check the cable regularly and ensure that a free space of about 3 cm is left at the rear of the scale, so that the cable is not kinked too severely.
Never tamper with the retaining screws for the load plate support
underneath the weighing pan.
When the weighing pan is removed, never insert a solid object
underneath the load plate support.
Never open the scale by removing the screws in its base.
Use only approved accessories and peripherals.
Handle the scale with utmost care; it is a precision instrument. Blows on
the weighing pan must be avoided, and heavy overloads must not be placed on it.
Important instructions when using scales in the food sector: those parts
of the scale can come into contact with food products have smooth surfaces and are easy to clean. The materials used do not splinter and are free from contaminants. In food processing areas, it is recommended that a protective cover (accessory) is used. This must be cleaned regularly, just like the scale itself. Damaged or heavily contaminated protective covers must be replaced immediately.
When the scale is finally taken out of service, observe the current
environmental regulations. The scale is equipped with a battery that contains heavy metals and therefore must not be treated as normal refuse! Local regulations for disposing of environmentally hazardous substances must be complied with.
METTLER TOLEDO SQC16 Chapter 1: Setting up the scale 9
1.2 Positioning and leveling the scale
The correct location is a decisive factor in ensuring accurate weighing results.
- Choose a stable and vibration-free location (particularly important for
high-resolution scales using METTLER TOLEDO MonoBloc technology). Place the scale on a surface that is as horizontal as possible and strong enough to bear its weight when fully loaded.
- Check the ambient conditions.
- Avoid:
- Direct sunlight
- Strong drafts (e.g. from fans or air conditioning systems)
- Excessive temperature fluctuations.
Turn the adjustable feet so that the scale is horizontal. If a spirit level is
filled, the bubble must be located within the inner circle.
Note: The SQC16 has a special filter that accelerates certain procedures (zeroing, taring) in a non-tranquil environment. This inevitably means that there is a slight loss of accuracy with the results. For high-precision results, care must be taken to ensure as tranquil and stable an environment as possible, so that the filter is not activated.
Major changes in geographical location:
Every scale is set by the manufacturer to suit the local gravitational conditions (geographical adjustment value) in the geographical zone to which the instrument is supplied. If a major change of geographical location takes place, this setting must be adjusted by a service technician or a new setting made. Certified scales must, in addition, be recalibrated in accordance with national certification regulations.
1.3 Connecting to power supply
Before connecting the power supply plug, verify that the voltage stated on
the model plate is the same as the local power line voltage.
For maximum possible precision, adjust the scale after installing it
(Chapter
7.4.1). Note: Certified scales must be adjusted by an
authorized organization. Please consult your dealer.
Chapter 2: Basic functions METTLER TOLEDO SQC16
10
2 Basic functions
This Chapter describes how to switch the scale on and off, zero and tare it, weigh materials and record the results.
2.1 Switching on and off and zeroing
Briefly pressing « » key switches the scale on or off.
The scale carries out a display test, and then the software version is briefly displayed. Once the weight display appears, the scale is ready to operate and automatically zeroed.
Note: If necessary, the scale can be zeroed at any time with the line « » key.
2.2 Setting date and time
Time can be set in 24 hours format, and the date in European or US format.
Press the « » key continuously for about 2 seconds.
…appears on the screen and then briefly press the « » key again.
Select TERMINAL by pressing « » and press « ».
…appears on the screen. Select Device and press « ».
METTLER TOLEDO SQC16 Chapter 2: Basic functions 11
…appears on the screen. Select Date Time by pressing « » and press «
».
…appears on the screen. Select Date by pressing the « » and press «Edit».
…appears on the screen. Use the numeric keypad to enter the date (e.g.
11022005) and confirm with «OK». Incorrect inputs can be deleted with «Erase». «
» and « » can be used to move the cursor.
To set the correct time, do the same procedure but select Time instead of Date.
Note: Only the Supervisor has access rights to modify the Date/Time.
2.3 Language settings
Language can be set: English, German, French, Spanish or Italian.
Press the « » key continuously for about 2 seconds.
…appears on the screen and then briefly press the « » key again.
Select TERMINAL using « » and press « ».
…appears on the screen. Select Device and press « ».
…appears on the screen. On Language press «Edit».
…appears on the screen. Choose the desired language and then press «OK».
Chapter 2: Basic functions METTLER TOLEDO SQC16
12
2.4 Simple weighing
Place the object to be weighed on the scale.
The bar graph at the top of the display shows how much of the weighing range is being used and how much is still available (as % of total scale capacity).
Wait until the stability detector (a small ring at the left edge of the display) disappears, then…
… read the indicated net weight.
METTLER TOLEDO SQC16 Chapter 2: Basic functions 13
2.5 Weighing with tare
Place the empty weighing container or packaging on the scale.
T
Press the «
T
» key briefly to tare the scale.
The zero display and the 'Net' (net weight) symbol appear.
Place the material to be weighed in the container, then…
… read the net weight of the weighing sample.
Note: The tare weight is retained until either a new tare is determined, or the scale is zeroed or switched off.
Chapter 2: Basic functions METTLER TOLEDO SQC16
14
2.6 Recording weighing results
Press the « » key to send the current weighing result to the peripheral device (usually a printer) via a COM port which has to be set up as 'Printer'.
Please refer to Chapter
7.9 for instructions on configuring the interface(s).
METTLER TOLEDO SQC16 Chapter 3: User setup and password 15
3 User setup and password
To avoid incorrect operation of the scale in normal use, the vision setup menu can be protected with a password. The scale differentiates between users and a supervisor. When the scale leaves the factory, the entire menu can be accessed by anyone. We therefore recommend you to define your own supervisor password as soon as you set up the scale. This limits access by the users to a smaller number of vision setup menu items (settings for language, sleep mode, contrast, invert, weight display and date/time).
Note:
Please be aware that the term 'Supervisor' (valid for Vision Setup; i.e. general scale operations and SQC16 operator/administrator setup) is different from 'Operator' and 'Administrator' (valid for SQC16 specific operations).
There should be only one (unnamed) Supervisor who has access to the Supervisor password and therefore, can setup several (named) administrators and several (named) operators, including resetting of their passwords.
Please refer to Chapter
7.2 on how to navigate within the 'Vision Setup'-menu and thus reach the menu position 'Vision
Setup Æ TERMINAL Æ Access' (Chapter
7.7) for defining the Supervisor password.
3.1 Defining a user name
Press the « » key continuously for about 2 seconds.
…appears on the screen, enter Supervisor password (if any; Æ Chapter 7.7)
and then briefly press the «
» key again.
…appears on the screen.
Select SQC16 by pressing « » and then press « ».
If you want to work with operator login, User Login should be activated.
Select User Setup and press «Edit».
…appears on the screen.
Define the user (User name, Number, Access Rights) by pressing «Edit» again.
Chapter 3: User setup and password METTLER TOLEDO SQC16
16
Using the alphanumeric keypad of the scale, type the name of the user and then press «OK». Up to 16 different users can be defined. Specify also 'Number' and 'Access Rights' of user, i.e. whether only as operator (who
cannot enter «
» menu within SQC16 mode) or as administrator.
Note: Password menu here is only meant to delete existing passwords in case it was forgotten (the supervisor can erase passwords of operators and administrators). For defining Æ Chapter
3.1. Please see also Chapter 7.5 for
more detailed information about SQC16 vision menu.
3.2 Creating a password
Once operator and administrator names are defined, a password has to be defined, as soon as the specific user enters 'SQC16' for the first time.
In weighing mode, press «SQC16».
Select the user name and then press «OK».
Using the numeric keypad of the scale, type the password and then press «OK».
Note: The first time it will ask 'Create password' and then 'Re-type password'. Enter the same new password twice.
METTLER TOLEDO SQC16 Chapter 4: Getting to know the SQC16 software 17
4 Getting to know the SQC16 software
SQC16 is an application for the 4-Series Compact Scales / Terminals. It is a convenient SQC compact system.
This chapter describes working with the SQC16 application. You might consider consulting Chapter
0 first about how to specify the behavior of the scale during weighing to match the surroundings. Further, you can specify general options such as scale identification, date, time, etc.
The SQC16 compact system offers evaluation and monitoring possibilities to statutory requirements for the fields of filling process control. It is suggested to attach a suitable METTLER TOLEDO strip printer or A4/Report printer.
Note: For more information about the meaning of the expressions 'Adjustment', 'Batch Statistics', 'Global Density', 'Individual Tare', 'Plausibility Limits', 'Mean Value Requirements', 'Shared Statistics', 'Supplement', 'Test', 'Tolerance System', 'Violation Check', 'Weighing Mode' and '2
nd
Tolerance System' please consult the included CD and select
'Appendix' of the 'SQC16 Training'. Alternatively, open the file Appendix_E.pdf on the root of the CD.
Special features of the system:
Internal calibration weight available as option
Language a choice of 11 languages
Max. number of articles 300
Data backup/restore/edit Statistics printing convenient also under MS Windows 98/ME/NT/2000/XP
Bar code for simple and effortless article selection
Tolerance systems EU, Free1, Free2, Free3
Units g, kg, lb, oz, ml, l, fz
Statistics per article 2 sets of statistics closed manually, e.g. for hourly and daily statistics
Batch statistics printed and closed automatically at end of batch
A4/Letter report printing choose between the Complete or Compact printing
Histogram in sample record and statistics
Class table in sample record and statistics
Graphics x/R (mean value/range) or x/s (mean value/standard deviation) in sample record
and statistics
Additive or subtractive weighing for simple handling
Minimum nominal support checking of the nominal value (warning if less than 100 resolution steps are
defined)
Individual tare for tare weights with high standard deviation (i.e. wide scatter)
Mean tare enter manually or by weighing a tare series or bulk tare
Sample size max. 999 (Note: with individual tare, there are 50 pre-weighing values available
for each of max. 20 articles simultaneously)
Plausibility check selectable (for nominal: for each article / for tare: system wide)
Adjustment to get suitable adjustment messages for the filling machine
Violation check for alarms after various defined events (T1-, T2-Violators below defined amount,
etc.)
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
18
5 The SQC16 Application
Start
*
only if 'SQCSet/Config/ Function/Test' is active
Start
Edit
10 x T
Set nxT
Cancel
*
only if selected article has unit ml, l or fz
Tare Density * Weigh mode
Stat 1
Stat 2
Batch *
Close
Print/Clear
* only if Batch stats
are active for the selected article
Sample/Test *
Database Key
Articles
Ok
Cancel
Select Article
Options
Search *
* only when there are 5 or more articles in the database
Delete ClosePrintDefineNewCopy
Articles
Print
Catalog
Print System
Settings
Print
G. Dens.*
Reports
SQC16
Print/Clr
* only if selected article
has unit ml, l or fz
Reports
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 19
System
Function
Messages
Global
Density
Tare
Config.
Tolerance
A4 Printout
Linefeeds
Marginals
x - chart
Sample
Statistics
Report
Adjust
Test
Shared
Stats.
2nd Tol.
Sys.
Viol.
Check
Supplement
Shift key
Before sampling, it is necessary to define at least one article.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
20
5.1 Overview of article definition
Density *
Edit
Edit
Density Group *
A
B
Nominal
Edit
Edit
Tare
* only if unit is either ml, l or
fz and 'SQCSet/System/ Global Density' is active
* only if unit is either ml, l or
fz and the article has not been assigned to a density group
* only if 'SQCSet/Function/
Adjust' is active
Fast
Adjustment *
Normal
Slow
Minimum Step
Edit
Factor
Edit
Edit
Supplement *
* only if 'SQCSet/Function/
Supplement' is active
B
Tolerance System
EU
Free1
Free2
Free3
Tolerances
+/-
Active/Inactive
T1-
Edit
T1+*
Edit
* only if +/- is active
T2-*
Edit
* if tolerance system is not Free1
T2+*
Edit
* if tolerance system is not
Free1 and +/- is active
T3-*
Edit
* if tolerance system
is Free3
T3+*
Edit
* if tolerance system is
Free3 and +/- is active
C
Article Definition
Edit
No.
Edit
ID
Edit
Unit
g
kg
lb
oz
ml
l
A
Name
fz
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 21
Weighing Mode
Standard
Additive
Subtractive
Plausibility -
Edit
Plausibility +
Edit
Print Violations
Active/Inactive
Plausibility
Viol. Check T1 Critical
Edit
Viol. Check T1 Reject
Edit
<T2
Active/Inactive
Mean Tolerance
Mean Tol-
Edit
Mean Tol+
Edit
Violation Check *
* only if 'SQCSet/Function/
Violation Check' is active
Individual Tare
1,1..2,2
1,2..1,2
D
C
E
Active/Inactive
Batch
Active/Inactive
D
Edit
Sample Size
Active/Inactive
Print Ind. Values
E
Active/Inactive
Transfer Key
Select
Shared Statistics *
* only if 'SQCSet/Function/
Share' is active
Set as default
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
22
5.2 Creating an article for the first time
You must define at least the name and nominal fill quantity of one or more articles. Furthermore, you can define, e.g. the weighing mode, the weighing unit and the number of weighing per sample as well as the tolerance system to be used for the evaluation.
In weighing mode, press «SQC16».
If 'User Login' is activated then select the user name and press «OK».
Note: This step can be skipped if "Vision Setup" Æ "SQC16" Æ "User Log-in" is set to Off.
Using the alphanumeric keypad of the scale, type the password and then press «OK».
Press «Articles».
Since there is no existing database yet, press «Yes» to create one.
Using the keypad, type in the article name and then press «OK». Continue with the next parameter (Article No.) by pressing «
».
Press «Edit». Using the keypad, type in the article number and then press «OK». Continue with the next parameter (ID) by pressing «
».
Press «Edit». Using the keypad, type in the identification number and then press «OK». Continue with the next parameter (Unit) by pressing «
».
Press « » to change the unit. As soon as the choices appear, use « » to choose the desired unit and press «
». Continue with the next parameter
(Density if liquid was chosen, otherwise Nominal) by pressing «
».
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 23
Press «Edit». Using the keypad, type in the density weight value and then press «OK». This parameter appears only if Unit selected is ml (milliliter), l (liter) or fz (fluid ounce). Continue with the next parameter (Density Group) by pressing «
».
Press «Edit». Using the keypad, type in the number of density group and then press «OK». This parameter appears only if Unit selected is ml, l or fz and if "SQC Config" Æ "Function" Æ "Global Density" is active. Continue with the next parameter (Nominal) by pressing «
».
Press «Edit». Using the keypad, type in the nominal weight value and then press «OK». Continue with the next parameter (Tare) by pressing «
».
In order to work with reasonable weighing results, it is recommended not to enter Nominal values lower than 100 times the scale resolution. Nevertheless, the system allows entries with a minimum of 30 times the resolution.
Example BBK462SQC-3XS:
Scale resolution d = 0.01g Minimum recommended Nominal value = 100x0.01g = 1g Minimum allowed Nominal value = 30x0.01g = 0.3g
This warning message appears if the entered Nominal value is lower than 100 times the scale resolution.
By pressing «Yes», the minimum recommended Nominal value automatically suggested by the system, i.e. 10 ml, is being set. By pressing «No», the entered value is set as Nominal if it is not lower than 30 x resolution
NOTE: The minimum recommended Nominal value varies depending on the scale resolution and article unit.
This error message appears if the entered Nominal value is lower than 30 times the scale resolution.
Press «Edit». Using the keypad, type in the tare weight and then press «OK». Continue with the next parameter (Supplement) by pressing «
».
Press «Edit». Using the keypad, type in the supplement value and then press «OK». This parameter appears only if "SQC Config" Æ "Function" Æ "Supplement" is active. Continue with the next parameter (Adjustment) by pressing «
».
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
24
To select adjustment, press « ». This parameter appears only if "SQC Config" Æ "Function" Æ "Adjustment" is active.
Select the speed of adjustment (Fast, Normal or Slow) by pressing « ». Also, define the minimum step value by selecting the parameter Min. Step
and then pressing «Edit». Using the keypad, type in the minimum step value and then «OK». The parameter Min. Step sets the threshold, i.e. smaller adjustment messages will not be given.
Furthermore, define Factor using «Edit». Using the keypad, type in the factor value and then press «OK». The computed adjustment value will be multiplied with the factor value in order to determine the final adjustment message. Continue with the next parameter (Tol. System) by pressing «
».
Press « » to change tolerance system (EU, Free1, Free2 or Free3). Continue with the next parameter (Tolerance settings) by pressing «
».
Press « » to change the positive and negative tolerance settings. Continue with the next parameter (Violation Check) by pressing «
».
Press « » to select violation checking. This parameter appears only if "SQC Config" Æ "Function" Æ "Violation Check" is active. Continue with the next parameter (Individual Tare) by pressing «
».
Press « » to select individual tare mode. Continue with the next parameter (Weighing mode) by pressing «
».
Press « » to change weighing mode (Standard, Additive or Subtractive). After selecting, continue with the next parameter (Plausibility) by pressing «
».
Press « » to change plausibility settings. Continue with the next parameter (Sample Size) by pressing «
».
Change sample size (-n-) by pressing «Edit» and then type in the desired number of individual values per sample. Continue with the next parameter (Print Individual Value) by pressing «
».
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 25
Each individual value will be printed out in the sampling report if this function is selected. Press «
» to unselect (or « » to select). Continue with
the next parameter (Transfer Key) by pressing «
».
If selected (press « »), the weighing result will not be automatically sampled once stability is reached. You will have to press Accept for each individual value. Continue with the next parameter (Shared Statistics) by pressing «
».
If selected (press « »), you will be asked to select the common article. With this function, you will be able to combine the sampling data of two or more articles into one common article, enabling you to monitor overall production.
In order to share an article's data with a common article, the following parameters have to be identical: Unit, Nominal, Tolerance System and Tolerances.
This parameter appears only if "System" Æ "Function" Æ "Shared Statistics" is active. When defining an article for the very first time, this parameter will not be available and will therefore jump to the next parameter, since there are no other existing articles yet to share statistics with. Continue with the next parameter (Batch) by pressing «
».
If selected (press « »), batch statistics are calculated and printed out for each batch. At the start of sampling, you will be asked to enter the batch name. Continue with the next parameter (Set as default) by pressing «
».
If selected (press « »), the parameter settings of this article will be set as default for defining new articles. The default article will appear within square brackets in the article list: [Article name].
Press Close (and confirm with Yes) to save the new article in the database. The article parameter will be printed.
5.3 The database
The following describes the use of the article database.
Briefly press the «» key.
…appears on the screen. You can now see the articles in your database.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
26
Scroll on the articles using « » and « », and press «OK» to choose the desired article to be used in the sampling. Press «Options» to go to the Options menu, or press «Search» to search articles in the database (see also Chapter
5.3.2). Otherwise, press «Cancel» to abort operation.
… appears on the screen if «Options» is pressed. This function is used to delete or copy the highlighted article or to change («Define») or print out its parameters. «New» will create a new article based on the default article.
5.3.1 Creating a new article
The following describes the operations for creating a new article in the database.
Briefly press the «
» key.
…appears on the screen. You can now see the articles in your database.
…appears on the screen. Press «Options».
…appears on the screen. Press «New».
…appears on the screen. Using the keypad, type in the name of the new article and then press «OK». In order to define the parameters of this new article, follow the same procedures as described in Chapter
5.2 for Article Number, ID, Unit, Density, Density Group, Nominal, Tare, Supplement, Adjustment, Tolerance System, Tolerances, Violation Check, Individual Tare, Weighing Mode, Plausibility, Sample Size, Print Individual Values, Transfer Key, Shared Statistics, Batch and Set as Default.
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 27
5.3.2 The 'Search' softkey
The following describes the operations for searching articles in the database.
Follow the first three steps as described in Chapter
5.3.
…when this screen appears, press «Search».
NOTE: Search button only appears when there are more than four articles in the database.
Press «Erase» to clear the article name and using the keypad, type in the first letter of the article you are searching for. The articles corresponding to the letter you typed in will be shown on the left side of the screen.
5.3.3 Editing the parameters of an existing article
The following describes the operations for editing the parameters of an existing article in the database.
Follow the first four steps as described in Chapter
5.3. Use the «Search» if necessary.
…when this screen appears, press «Define».
…appears on the screen. Change the parameters of the article name by pressing «Edit».
…appears on the screen. Press «Erase» to delete the existing article name. Change it by typing in a new name using the keypad. Press «
» and
«
» to move cursor from left to right and vice-versa. Press «OK» when
input is done. Otherwise, press «Cancel» to abort operation.
Continue editing the rest of the parameters by following the same procedures as described in Chapter
5.2.
Note: Once SQC16 has generated a statistics, you can no longer edit the following parameters of that article: Name, Unit, Nominal Tolerance system and Tolerance. If you wish to edit the parameters of an article that already has a statistics, you have to print and clear the statistics first by pressing «Print/Clr» in standby mode (see also Chapter
5.8).
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
28
5.3.4 Copying the parameters of an existing article to a new article
Sometimes, when creating a new article, it may be easier to copy the parameters of an existing article and then editing only the parameters that need to be changed. The following describes the operations for copying the parameters of an existing article on to a new one.
Follow the first four steps as described in Chapter
5.3. Use «Search» if necessary.
…when this screen appears, press «Copy». After copying and making necessary changes, you can select a particular article for you to set as default so that next time you copy, the default article will be selected first.
Using the keypad, type in the new article name and then press «OK». The parameters of the article chosen are now copied on to the new article. Continue with the other parameters by following the same procedures as described in Chapter
5.2.
5.3.5 Deleting an article
The following describes the operations for deleting an article from the database.
Follow the first four steps as described in Chapter
5.3.
…when this screen appears, press «Delete».
…appears on the screen. Press «Yes» to delete. Otherwise, press «Cancel» to abort operation.
Note: Articles that have existing statistics values cannot be deleted, until these statistics are cleared as described in Chapter
5.8.
5.3.6 Printing the parameters of an article
The following describes the operations for printing out the parameters of an article on an attached printer.
Follow the first four steps as described in Chapter
5.3. Use «Search», if necessary.
…when this screen appears, press «Print».
…appears on the screen. SQC16 will print out all the parameters of the chosen article.
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 29
5.4 Sampling of articles
The following describes the operations on how sampling of articles is performed.
Press «Start» to commence sampling.
…appears after placing the first item on the scale.
…appears after placing the second item on the scale.
…appears after placing the third item on the scale.
…appears after placing the fourth item on the scale.
…appears after placing the fifth item (default setting) on the scale.
…appears on the screen. A report will be printed out.
…appears on the screen. Press «Close» to close the Sample Report. Press «
» to scroll down to see the rest of the statistics.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
30
…end of statistics report. Press « » or «C» to close the end of sampling messages (see also «
»/System/Messages). Press «Close» to close the
Sample Report.
5.5 Test series
A test series is a sample for the simple determination of mean and standard deviation. Therefore, the test series results are not saved in the statistics of the article, but are merely printed out for testing purposes. It serves e.g. to set a filling machine after a product change.
Press «Test» to activate the test series function (if selected in System/Function/Test)
…appears on the screen. Press «Start» to commence test sampling.
…appears after placing the first item on the scale.
…appears after placing the second item on the scale.
…appears after placing the third item on the scale.
…appears after placing the fourth item on the scale.
…appears after placing the fifth item on the scale. You may go on until the maximum sampling of 999 is reached or press «End» to finish sampling.
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 31
…appears on the screen. A report will be printed out.
…appears on the screen. Press «Close» to close the Sample Report. Press «
» to scroll down to see the rest of the statistics.
Note: 'T' appears to indicate a test series within the mean value trace.
…end of statistics report. Press « » or «C» to close the end of sampling messages (see also «
»/System/Messages). Press «Close» to close the
Sample Report.
5.6 Taring
The following describes the three ways of entering tare values within SQC16.
Press «Tare» to activate the tare function.
5.6.1 Manual tare
Press «Edit» to change the tare weight of an article manually to a known value.
…appears on the screen. Using the keypad of the scale, type in the tare value and then press «OK».
…appears on the screen. SQC16 is now ready to start sampling using the new tare value.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
32
5.6.2 Tare series
A tare series is a procedure, by which individual tare items are placed on the scale, in order to determine the mean tare value to be used in the sampling of an article.
Press «Start» to commence tare series.
…appears after placing the first tare item on the scale.
…appears after placing the second sample on the scale.
…appears after placing the third sample on the scale.
…appears after placing the fourth sample on the scale.
…appears after placing the fifth sample on the scale.
A tare series does not end automatically. It needs to be ended by pressing «End».
…appears on the screen. SQC16 has generated a mean value and is now ready for sampling.
5.6.3 Bulk tare
The bulk tare softkey «10xT» is used to determine the mean tare value, based on the number of tare items in the settings. You can change the number of tare items by pressing «Set nxT». The default setting is n=10 (Æ «10xT»).
Press «10xT» to switch to bulk taring function.
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 33
…appears on the screen. Press «Start» to commence bulk taring.
Place 10 tare items on the scale and then press «Accept».
…appears on the screen. SQC16 has generated a mean value and is now ready for sampling.
5.7 Density
This function makes it possible for the user to conveniently change the density value, without having to go to the menu of the article definition. However, this function appears only if a liquid unit is selected in the article definition, i.e. "ml", "l" or "fz".
Press «Density».
Press «Erase» to delete the existing value and, using the keyboard, type in the desired value and then press «OK». Please refer to Chapter
5.9.2 Global
Density for more information.
5.8 Printing/Clearing statistics
The following describes the operations on how to print and/or clear statistics. Note the difference between «Print» (printing only) and «Prt/Clr» (print first and erase after printing). SQC16 will always print the statistics before permanently erasing it from the database. Of course, it should be checked first, if enough paper is available, otherwise the data will be lost. Consider backing up your data with BR16 PC-program before erasing important data.
Press «Print/Clr» while the article whose statistics should be printed (and cleared in case of Prt/Clr) is the current one.
Choose the statistics that you wish to delete («Stat1», «Stat2» or «Batch»).
Note: «Batch» appears only if "Batch" is activated under "Articles" Æ "Options" Æ "Define" Æ "Batch". See also Chapter
6.3.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
34
…appears on the screen. Press «Prt/Clr» if the selected statistics should be erased after printing. Press «Print» to leave data untouched after printing.
SQC16 prints out the selected statistics. Wait until printing is done.
…appears on the screen. Press «Close» to go back to sampling screen.
5.9 The System settings
You can press «
» then select System to set general SQC parameters (tolerance unit mode; global density; end of
sample messages, tare plausibilitiy), valid for all articles.
Press « ».
Three sub-menus appear on the screen, namely: System, Function and Report. The following is an overview of the System Settings menu.
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 35
System Settings
Tolerance Mode
Global Density
Selected
Not Selected
T (abs)
Selected
Not Selected
t (rel)
Selected
Not Selected
%
Selected
Not Selected
Unit
Selected
Not Selected
Messages
Sample1
Selected
Not Selected
Sample2
Selected
Not Selected
Stat1
Selected
Not Selected
Stat2
Selected
Not Selected
Batch
Selected
Not Selected
Tare Config
Tare Plaus. -
Tare Plaus. +
Press « » to enter the System settings sub-menu.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
36
The following selection appears:
"Tolerance" Define tolerance entries (menu)
"Global density" Global density (same for all articles)
"Messages" Define messages at the end of samples
"Tare Config" Define minimum and maximum tolerances in
percentage
5.9.1 Tolerance specifications
With the factory setting, the tolerances are entered and displayed relative to the nominal fill quantity and in the selected unit. If you wish to change this setting, select the parameter "Tolerance" and then press «
».
"T (abs)" Representation of the tolerances relative to zero
"t (rel)" Representation of the tolerances relative to the nominal fill
quantity
"%" Tolerance specifications in percentage
"Unit" Tolerance specifications in the unit applicable to the article
5.9.2 Global density
With Global density active, articles with liquid units can be one in one of among 30 Density Groups, selectable in the article definition menu. If the density of an article belonging to a specific Density Group was modified, all articles belonging to the same group will have their density parameters modified as well.
Press «
» or « » to select or unselect Global Density.
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 37
5.9.3 Sample messages
Selecting "Messages" will display relevant information at the end of the sampling, such as Article Name, Article ID, Date, Number of Samples, Total Number of Samples, Mean Value, Standard Deviation, Minimum Value, Maximum Value and Range.
Press «
» to activate sample messages.
Sample 1/2, Sample 2/2, Stat1, Stat2 and Batch are all selected by default. To deactivate, simply press «
».
Below are examples of sample messages that may appear on the display:
Sample 1/2
Sample 2/2
Statistics 1
Press «Close» to close the Sample Report.
5.9.4 Tare configuration
This function will allow to set a +/- plausibility for the individual values of a tare series. It refers to the current tare value.
Press «
» to enter the Tare Configuration menu.
Enter desired Minimum and Maximum Tolerances by pressing the «Edit»
and
typing in the value using the numeric keypad of the scale.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
38
5.10 The Function settings
You can use the menu option "«
»\Function" to switch several special SQC functions on or off.
The following is an overview of the Function Settings menu.
Function Settings
Violation Check
Selected
Not Selected Viol. Check Unit
Sample
Stat1
Stat2
Batch
2nd Tol. System
Shared Statistic
Test Supplement Adjust
%
PCS
Selected
Selected
Selected
Selected
Selected
Selected
Selected
Selected
Selected
Not Selected
Not Selected
Not Selected
Not Selected
Not Selected
Not Selected
Not Selected
Not Selected
Not Selected
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 39
Press « » to enter the Function settings sub-menu.
The following selection appears:
"Viol. Check" Check if a certain number of toleration violations
has been exceeded
"2
nd
Tol. System" Evaluate each sample of an article using an
additional, to be defined, 2
nd
tolerance system.
"Shared Statistics" Share a common set of statistics for various
articles
"Test" Sample without influence on the statistics, for
simple determination of mean value and standard deviation
"Supplement" May be necessary if a filling process is not stable
with time
"Adjust" To activate the adjustment algorithm with resulting
numeric messages that serve to optimize the filling process
5.10.1 Violation check
If you wish to have an alarm message when a certain number of tolerance violations has been exceeded, select the parameter "Viol. check".
Press «
» to activate Violation check.
The following selection appears:
"%" Check if unit should be percentage instead of PCS
"PCS" Check if unit should be pieces instead of %
Press «
» to check the desired unit.
"Sample" Select to monitor tolerance violations in Sample
"Stat1" Select to monitor tolerance violations in Statistics 1
"Stat2" Select to monitor tolerance violations in Statistics 2
"Batch" Select to monitor tolerance violations in Batch
Press «
» or « » to select or unselect any of the above 4 items.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
40
When no article has the Violation Check activated, you can select either "%" or "PCS" as its unit. Then you can select an article wherein you want to activate the Violation Check and define the values for VC T1 Critical, VC T1 Reject, <T2 and make the Mean Tolerance active or inactive.
"VC T1 Critical" If during the sampling, the specified amount of Tolerance Violators for T1 -/+ reaches the specified Critical values, an alarm will set off informing the user that there is a certain number of samples that are "T1 Critical (Pcs or %)".
"VC T1 Reject" If during the sampling, the specified amount of Tolerance Violators for T1 -/+ reaches the specified Reject values, an alarm will set off informing the user that there is a certain number of samples that are "T1 Reject (Pcs or %)", and the whole sampling will then be rejected.
"<T2" this refers to the amount of T2- violators; this function can be enabled or disabled.
"Mean Tolerance" The check for mean tolerances (
tm-
and
tm+
) is a check
on the deviation of the mean value from target value
5.10.2 2nd Tolerance System
If you desire evaluations of the same article using different tolerance systems, e.g. according to a free tolerance system to improve the filling accuracy, select the parameter "2
nd
Tol. System".
Press «
» to activate 2nd tolerance system, then the correlations and
relationships under the article definition.
Before you can define an article as a slave, you need to define an article as a master (reference) first. This master is a normal article. It has at least the following parameters defined:
a. Name b. Nominal value c. Tolerance system d. Tolerances
If you define a slave of a master, all data defined in the master-article will be copied to the slave. Now you can only change a few parameters on the slave-article:
a. Tolerance system b. Tolerances c. Violation check d. Individual values printout
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 41
5.10.3 Shared Statistics
If you wish to have a common set of statistics for various articles, select the parameter "Shared Statistics". To perform the shared statistics and for every common article, an article must be defined.
Press «
» to activate Share, then define the relationships under the article
definition.
5.10.4 Test
If you wish to perform sampling for simple determination of mean value and standard deviation, without influence on the statistics, select the parameter "Test".
Press «
» to activate Test.
5.10.5 Supplement
It may sometimes be necessary to switch on Supplement, if a filling process is not stable. To do so, select the parameter "Supplement".
Press «
» to activate Supplement.
5.10.6 Adjust
To work with adjustment messages, select the parameter "Adjustment". If selected, then the parameter 'Adjustment, as well as its sub parameters 'Speed', 'Min. Step' and 'Factor' will become available in the article definition of each article.
Press «
» to activate Adjustment.
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
42
5.11 The Report settings
This function allows you to generate reports according to your specific needs. If you wish to change this setting, select the parameter "Report".
Using « » and « », choose «Report». The following is an overview of the Report Settings menu.
Report Settings
See details in
Sample
Statistics
x - chart
Marginals
x
Header/
Footer
%
...
See details in
x
Header/
Footer
%
...
Individual
Chart
Statistics
Values
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Staf1 Info
Stat2 Info
Batch Info
Linefeeds
A4 Printout
Statistics
Edit
Sample
Edit
Code
Edit
Complete
Not
Selected
Selected
Compact
Not
Selected
Selected
x/s
x/R
Selected
Not
Selected
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 43
x - chartHistogram
%
Class Table
Tol.
Violators
Min/Max/
Range
Header/
Footer
Not
Selected
Selected
Header 1
Selected
Free
Edit
Not
Selected
Selected
Selected
Date
Time
Not
Selected
Footer 1
Footer 2
Header 2
Footer 3
Same as Header 1
hh:mm
hh:mm:ss
x
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Selected
Not
Selected
Press « » to enter the Report settings sub-menu.
The following selection appears:
"Statistics" Define the content of the statistics report
"Sample" Define the content of the sample report
"x-chart" R(ange) or s in the mean value trace
"Marginals" Switch reporting of marginals on or off
"Linefeeds" Define number of linefeeds at the end of report
"A4 printout" Switch report printing between Complete and
Compact
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
44
5.11.1 Statistics and Sample reports
Both statistics and sample reports menu contain the same parameters except for the "Individual Chart" and "StatVal" parameters which are only available in the sample reports. To change settings, select the parameter "Statistics" or "Sample" accordingly, then press «
».
Press «
» to change Header/Footer settings.
Using «
» and « », choose a header or a footer you wish to include in the
printout and then press «
».
…appears on the screen. Press «
» to activate type of header or footer. The parameter "Free" is used to enter a text, and the parameter "Date" and/or "Time" is used to print date and time.
"Free" Text entry using the alphanumeric/numeric keypad of
the scale. Maximum 24 characters are possible,
additional characters will be truncated. "Date" Switch date in the selected line on or off "Time" Format time printout in the selected line or switch off
If you wish to have the time in the selected line, you must define the format.
"hh:mm" Time printout in the selected line in hours and minutes "hh:mm:ss" Time printout in hours, minutes and seconds
Note: In one line, you can print only a free text or date and/or time.
Notes: It is possible to print out the identification of the scale in the header or footer of a report by entering:
Example: Header 3 Æ Free Æ .T.
METTLER TOLEDO SQC16
Example: Footer 2 Æ Free Æ .S.
SNR: 2511378
If the printer connected to the scale is an EPSON LX-300 printer, the report can only be customized by setting Header1, Header2 and Footer1 as "Free".
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 45
5.11.2 Define contents of the report
The following information can be printed out in the statistics and sample reports:
"Min/Max/Range" Minimum value/maximum value/range (=Max-Min) in the report
"Tolerance" Tolerance violators in the report
"Class table" Class table in the report
"Histogram" Histogram in the report
"x-chart" Mean value trace in the report
"%" x/s/R/Min/Max/R in percent in the report
"Individual Chart" Individual chart in the report (only in sample reports)
"StatVal" Statistics values in the report (only in sample reports)
Select or unselect the desired information using «
» or « ».
To print out minimum value, maximum value and difference in the report. The report will appear as follows:
Min 90.81% 72.65 g Max 100.27% 80.22 g R 9.46% 7.57 g
To print tolerance violators in the report. The report will appear as follows:
<T2- 0 0.00 % <T1- 1 20.00 % >T1+ 0 0.00 % >T2+ 0 0.00 %
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
46
To print class table in the report. The report will appear as follows:
<T2- 0 0.00 % <T1- 0 0.00 % <T1- 1 3.33 %
-4 1 3.33 %
-3 2 6.66 %
-2 3 9.99 %
-1 4 13.32 % +1 7 23.31 % +2 8 26.64 % +3 5 16.65 % +4 2 6.66 % >T1+ 0 0.00 % >T1+ 0 0.00 % >T2+ 0 0.00 %
To print the histogram in the report. A histogram is a graphical representation of the distribution. The area of the rectangle is proportional to the number of individual values in the classes.
The range from Nominal to T1 is divided into 4 classes, from T1 to T2 (if used) into 2 classes, and from T2 to T3 (if used) into one class. The histogram will appear as follows:
<T2
<T1 <T1 ▒▒
-4 ░░░
-3 ░░░░░░
-2 ░░░░░░░░░░░░
-1 ░░░░░░░░░░░░░░░ +1 ░░░░░░░░░░░░░░░░ +2 ░░░░░░░░░░░░░ +3
░░░░░░░
+4
░░░
>T1
▒▒
>T1 >T2
To print x-chart (mean value trace) in the report.
The standard deviation (s trace) or the range of the values (R trace) of the last 10 samples can be shown in the left trace (selectable in the menu "x- Chart", see below), the right trace shows the mean values of the last 10 samples.
s Chart x¯ Chart
│ * │ │ │ ░│ │ │ │ *│ │ │ │░ │ │ │ * │ │ ░ │ │ │ * │ │ │ ░ │ │
0 7.57 - Nom. +
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 47
To print x/s/Min/Max/R in percentage in the report. Mean value, standard deviation, minimum value, maximum value and
difference are also printed out relative to the nominal value. The report will appear as follows:
x¯ 96.68 % 77.350 g s 3.71 % 2.965 g Min 90.81 % 72.65 g Max 100.27 % 80.22 g R 9.46 % 7.57 g
The following information can be printed out only in the Sample report.
To print individual trace in the report.
The individual values of a sample are also shown graphically with the tolerances and violations. The report will appear as follows:
1 77.93 │ │ ░│ │ │ 2 76.61 │ │░ │ │ │ 3 72.65 │▒│ │ │ 4 79.34 │ │ │ │ 5 80.22 │ │ │ │
To print some statistics values in the sample report.
"Stat1" Summary information of Statistics1
"Stat2" Summary information of Statistics 2
"Batch" Summary information of Batch
Select or unselect the desired information using «
» or « ».
The report will appear e.g. as follows:
Stat.1: Sample# 3 x¯ 98.48% 78.784 g s 3.71% 2.965 g <T1- 1 6.67 %
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
48
5.11.3 x-chart (Mean value trace)
In addition to the mean value 'x', either the 'Range' (heaviest individual value minus the lightest individual value) or the standard deviation 's' will be shown.
Press «
»
… select either x/R (to show range) or x/s (to show standard deviation) by «
».
5.11.4 Marginals
Marginals are values less than the nominal value and greater than or equal to the negative tolerance limits. This can be switched on or off in the reports. They will only appear in the free tolerance systems.
Select or unselect Marginals using «
» or « ».
5.11.5 Linefeeds
For aesthetic purposes, empty lines (called linefeeds) may also be added at the end of the report.
Press «
» to add a linefeed in statistics, sample or code.
"Statistics" Linefeeds at the end of statistics reports (1..3..99). "Sample" Linefeeds at the end of sample reports (1..3..99). "Code" Linefeeds at the end of the entry of codes (1..3..99).
METTLER TOLEDO SQC16 Chapter 4: Chapter 5: The SQC16 Application 49
5.11.6 A4 Printout
The operator has the option whether to print in Complete or Compact. The Complete version prints the whole report, while the Compact version prints only a selection of the report, and the font is much smaller.
Press «
» to define the printing settings.
"Complete" Whole sample report will be printed, using normal font size "Compact" Selected parts of the sample report will be printed, using
smaller font size reports
…appears when "Complete" is selected.
This is an actual report printed in "Complete" mode.
…appears when “Compact” is selected. This is an actual report printed in "Compact" mode.
Note: The only added information not included in the compact printing are: Individual sample with charts, Tolerance System, Class Table, Violation Check, Histogram and s-chart/ x-chart.
Sample result Date: 20.01.2005 Time: 13:39 Operator: JA SAMPLE 01 Nominal 100.0 ml Tare (x,¯)14.8 g Density 1.2500 g/ml Qty Mean (x,¯) Std. Dev Min. Max. Range <T3- <T2- <T1- Marg >T1+ >T2+ >T3+ 10 100.14 ml 4.68 ml 93.00 ml 107.60 ml 14.60 ml 0 0 2 3 1 0 0
Stat1 From: 20.01.2005 - 13:39 To: 20.01.2005 - 13:40 Qty Mean (x,¯) Std. Dev Min. Max. Range <T3- <T2- <T1- Marg >T1+ >T2+ >T3+ 10 100.14 ml 4.68 ml 93.00 ml 107.60 ml 14.60 ml 0 0 2 3 1 0 0
Stat2 From: 20.01.2005 - 13:39 To: 20.01.2005 - 13:40 Qty Mean (x,¯) Std. Dev Min. Max. Range <T3- <T2- <T1- Marg >T1+ >T2+ >T3+ 10 100.14 ml 4.68 ml 93.00 ml 107.60 ml 14.60 ml 0 0 2 3 1 0 0
Batch <<<<<<<Batch>>>>>>>> From: 20.01.2005 - 13:39 To: 20.01.2005 - 13:40 Qty Mean (x,¯) Std. Dev Min. Max. Range <T3- <T2- <T1- Marg >T1+ >T2+ >T3+ 10 100.14 ml 4.68 ml 93.00 ml 107.60 ml 14.60 ml 0 0 2 3 1 0 0
Chapter 5: The SQC16 Application METTLER TOLEDO SQC16
50
5.12 Printing summary of database
This function allows you to print out the summary of all articles stored in the database.
Press «Print/Clr»
…appears on the screen. Press «Reports».
…appears on the screen. Press «PrtCat» to print out the catalog, «PrtSys» to print all parameter of 'System' or «PrtGlDens» to print a list with the values of the 30 density groups.
…appears on the screen. The attached printer will print out the following information of all articles stored in the database:
Article index
Article Name
Name of 2
nd
tol.sys article (only if defined)
Article number (only if defined)
Article ID (only if defined)
Nominal
Supplement (only if defined)
Target (only if supplement and adjustment are defined)
Tare
Density (only if unit is for liquid)
Tolerance system
Individual tare (only if active)
Common article (only if defined)
METTLER TOLEDO SQC16 Chapter 6: Special features of SQC16 51
6 Special features of SQC16
6.1 Working with barcodes
4100 4562
The SQC16 system can be equipped with a bar code reader. When the article data are defined, e.g. the following parameters can be read in: Article name, article number, identification, user name, text for header and footers in the printout.
If you wish to select the articles for sampling, using the bar code reader, you must define the number of the bar code (e.g. EAN) as article number.
To activate, the desired article can be selected directly by reading in the bar code.
In sampling mode, also the batch number can be read in with the bar code.
6.2 Individual tare sampling
There are two possible ways to work with individual tare:
-
1,1..2,2
Preweighing and backweighing are in two samples. First, all preweighings are done in one
sample and afterwards, the backweighings are done in the same order as the preweighings in a second sample.
-
1,2..1,2
Preweighings and backweighings are in one sample. After every preweighing, the
backweighing follows immediately.
Preweighings can be gross values (Æ backweighing: tare value) or tare values (Æ backweighing: gross value). Any defined mean tare value from the database will have its effect to calculate the net value as well.
Note: When you are done with preweighing but not with backweighing, you will not be able to delete the sampling unless you press «Start» and then «End».
6.2.1 Preweighing and backweighing in two samples
Activate individual tare sampling by defining an article with its "Ind. Tare" parameter selected. Press the «
» and select
the article you wish to use. Press "Options", then "Define" then press «
» until the parameter "Ind. Tare" is highlighted.
Press «
» and choose mode "1,1 .. 2,2" and then press « » to activate this individual tare mode.
".1" appears immediately after the article name, indicating that SQC16 is now ready for preweighing. Press «Start» to commence sampling.
Chapter 6: Special features of SQC16 METTLER TOLEDO SQC16
52
…appears on the screen. SQC16 is now ready to perform the preweighing.
First container was placed on the scale.
Continue procedure until the fifth container has been placed on the scale.
Fifth (which is here also the last) container was placed on the scale.
".2" appears immediately after the article name, indicating that SQC16 is now ready for backweighing. Press «Start» to commence sampling.
Place the first sample together with the first container on the scale.
Then, place second sample and second container on the scale. Continue procedure until the fifth sample has been placed.
…appears on the screen. Wait until the printout of the sample report is done.
…appears on the screen. SQC16 has generated a report. Press «Close» to close the Sample Report. Press «
» to scroll down to see the rest of the
statistics.
…end of statistics report. Press « » or «C» to close the end of sampling messages (see also «
»/System/Messages). Press «Close» to close the
Sample Report.
METTLER TOLEDO SQC16 Chapter 6: Special features of SQC16 53
6.2.2 Preweighing and backweighing in one sample
Activate individual tare sampling by defining an article with its "Ind. Tare" parameter selected. Press the «
» and select
the article you wish to use. Press "Options", then "Define", then press «
» until the parameter "Ind. Tare" is highlighted.
Press «
» and choose mode "1,2 .. 1,2" and then press again « » to activate this individual tare mode.
Note: When you are done with preweighing but not with backweighing, you will not be able to delete the sampling unless you press «Start» and then «End».
".1" appears immediately after the article name, indicating that SQC16 is now ready for preweighing. Press «Start» to commence sampling.
…appears on the screen. SQC16 is now ready to perform the preweighing.
Place the first container on the scale (In this example: Pre-weighing of tare values, not gross values).
Place the first sample, incl. container, on the scale. The container’s weight value becomes the tare weight and is subtracted from the gross weight. Actual weight is shown on the upper left-hand corner of the screen (Act).
Place the second container on the scale.
Place the second sample, incl. container, on the scale.
Continue procedure until the fifth container and fifth sample have been placed on the scale accordingly (in this example: -n- = 5).
…appears on the screen. Wait until the printout of the sample report is done.
…appears on the screen. SQC16 has generated a report. Press «Close» to close the Sample Report. Press «
» to scroll down to see the rest of the
statistics.
Chapter 6: Special features of SQC16 METTLER TOLEDO SQC16
54
…end of statistics report. Press « » or «C» to close the end of sampling messages (see also «
»/System/Messages). Press «Close» to close the
Sample Report.
6.3 Sampling with batch
Batch sampling means to have a batch identification for sampling data. Each batch has its own batch statistics.
Activate batch sampling by pressing the «
» and select the article you wish to use. Press "Options", then "Define", then
press «
» until the parameter "Batch" is highlighted. Press « » to activate batch sampling mode.
Press «Start» to commence sampling.
…appears on the screen. Using the keypad, type in the name of the batch and then press the «OK»
…appears on the screen. SQC16 is now ready for batch sampling.
Place first sample on the scale.
Continue procedure until the fifth sample has been placed on the scale
Continue procedure until the fifth sample has been placed on the scale
.…appears on the screen. Wait until the printout of the sample report is done.
METTLER TOLEDO SQC16 Chapter 6: Special features of SQC16 55
…appears on the screen. SQC16 has generated a report. Press «Close» to close the Sample Report. Press «
» to scroll down to see the rest of the
statistics.
…end of statistics report. Press « » or «C» to close the end of sampling messages (see also «
»/System/Messages). Press «Close» to close the
Sample Report.
Note: When a new batch name is defined, SQC16 will automatically print and then clear the previous batch statistics. You may then commence with a new batch statistics.
6.4 Transfer key
Using the transfer key enables you to control the weighing procedure especially for items that require filing or dosing.
Activate transfer key by pressing the «
» and select the article you wish to use. Press "Options", then "Define", then press
«
» until the parameter "Transfer key" is highlighted. Press « » to activate transfer key mode.
Press «Start» to commence sampling.
...appears on the screen.
Place first sample on the scale. Press «Accept» to accept the value.
...appears on the screen.
Chapter 6: Special features of SQC16 METTLER TOLEDO SQC16
56
Continue procedure until the fifth sample has been placed on the scale
.…appears on the screen. Wait until the printout of the sample report is done.
…appears on the screen. SQC16 has generated a report. Press «Close» . to close the Sample Report. Press «
» to scroll down to see the rest of the
statistics.
…end of statistics report. Press « » or «C» to close the end of sampling messages (see also «
»/System/Messages). Press «Close» to close the
Sample Report.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 57
7 Vision Setup
Vision Setup can be used to change the settings of the scale and to activate functions, thereby allowing the scale to be adapted to individual weighing needs.
Important: We recommend defining your own supervisor password (Æ Chapter
3.2).
7.1 Overview and operation
Standby / weighing mode
After switching on...
Long press on
Password
Type password, end with
SCALE
Chapter 7.3.1
SQC16
Chapter 7.3.2
TERMINAL
Chapter 7.3.3
COMMUNICATIONS
Chapter 7.3.4
DIAGNOSTICS
Chapter 7.3.5
Navigating through the Vision Menu is done through the six softkeys found at the bottom of the display. The softkeys change depending on the currently selected menu item to allow the user to perform the appropriate action for the selected item.
This softkey selects the previous item listed on the current menu.
This softkey selects the next item listed on the current menu.
This softkey selects the first item on the next “page” of the menu. If all the items for the current menu are visible (i.e. there are four or less items for the current menu), this softkey is not visible.
Chapter 7: Vision Setup METTLER TOLEDO SQC16
58
This softkey enters the submenu of the currently selected item. If the currently selected item does not have a submenu, this softkey is not visible.
This softkey leaves the current submenu and goes one level up. If the Vision Menu is already at the top level, pressing this softkey is similar to pressing «End» (see below).
This softkey allows the user to edit the value or setting of the currently selected item. If the currently selected item contains a submenu, this softkey is not visible.
Some menu items perform specific actions rather than editing. Pressing this softkey directs the Vision Menu to perform the action specified by the currently selected menu item.
This softkey can be pressed at any time to leave the Vision Menu. The Vision Menu will then prompt the user to save or discard the changes the user has made. The user may also press «Cancel» at this time to go back to the Vision Menu.
7.2 Calling up the menu and entering the password
Press « » key and hold it down until the prompt to enter the password appears.
User: No password required, just press « » key.
Supervisor: Enter password (sequence of keystrokes, Chapter
7.7)
immediately
and confirm with the « » key. Otherwise, after a few seconds, the scale returns to weighing mode. If an incorrect password is entered, the menu cannot be called up. Note: When the scale leaves the factory, no supervisor password is defined, so when the password is
requested, just press «
» key.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 59
7.3 Menu overview
7.3.1 Scale
Calibration
Start?
- 0 -
xxx g
xxx g
- done -
Display
Unit 1
g
kg
t
lb
oz
Resolution *
0.0005
0.01
Zero
2d
AZM
Off
0.5d
Restart
Off
On
Filter
Vibration
Low
Mid
High
Process
Universal
Dosing
Tare
Auto Tare
Off
On
Chain Tare
Off
On
AutoClr Tare
Off
On
Pushb. Tare
Disable
Active
*
Resolution depends
on the type of balance
1d
5d
10d
Stability
Fast
Standard
Precise
Temperature
Off
1 K
2 K
FACT
5 K
Reset
SURE?
Start?
Day / Time
Days
Time1
Time2
Time3
MinWeigh
Off
On
*
Only available in
Monobloc with
internal calibration
*
Only shows if
Min Weigh
value is set
*
Calibration weight depends on the type of balance
Edit Time1
Edit Days
Edit Time2
Edit Time3
Scale
Chapter 7: Vision Setup METTLER TOLEDO SQC16
60
7.3.2 SQC16
SQC16
User Login
Off
On
Auto Log Out
Off
On
Min. PW Length
0
1
2
3
4
5
6
User Setup
User1
User16
User
Number
Edit
Edit
Access Rights
Operator
Administrator
Password
Erase password
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 61
7.3.3 Terminal
Weight Display
Sleep
Contrast Invert Date Time
Access
Supervisor Pwd
Enter Pwd
Retype Pwd
Reset
Start?
Are you sure?
Off
1 minute
3 minutes
5 minutes
0
10
No
Yes
small
big
EU
Date Format
US
Date
13.12.02
Time
08.45.33
Beep
Off
On
TERMINAL
Language
Device
English
Deutsch
Italiano
Español
Français
Nederlands
Polski
Slovák
Slovenski
Czech
Magyar
15 minutes
30 minutes
Chapter 7: Vision Setup METTLER TOLEDO SQC16
62
7.3.4 Communication
COMMUNICATIONS
COM1
COM2
Mode
Printer
Auto Print
Continuous Wt
Dialog
2nd Display
COM3
Baud
300
600
1200
2400
4800
Parity
7-none
7-even
Handshake
No
XonXoff
Parameters Printer Type Add Line Feed
Reset Com
Start?
Are you sure?
0
9
LC-I/O
Keyboard
US International
UK
French
Swiss French
Belgian French
German
German Swiss
Italian
Spanish
Latin America
BCR
Not Used
Article Search
PS2
Define Header
Header Line 1
Header Line 2
Header Line 3
Header Line 4
Header Line 5
19200
38400
9600
57600
115200
COM 3
8-none
8-odd
8-even
7-odd
Sprinter 1
GA 42
RS-P42
LX-300
*COM 3 is not available in IND469SQC
GA 46
*COM 1 and COM 2 only
BCR
*COM 1 and COM 2 only
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 63
7.3.5 Diagnostics
DIAGNOSTICS
Keyboard Test Display Test
SNR1
SNR2 List1 List2 Reset All
Start?
PUSH 1
PUSH 25
Start?
<Display Test>
2564494
Start?
Printout
Start?
Are you sure?
7.4 Scale settings (SCALE)
This function block allows the user to change general scale functionality.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE' and how to navigate within the
'Vision Setup'-menu.
The screen on the right hand side shows the most important scale settings as a preview.
Enter the SCALE menu (« ») and choose one of the functions listed below.
Available functions and settings:
Adjust/calibrate the scale Æ Chapter
7.4.1
Display accuracy and weighing unit Æ Chapter
7.4.2
Tare settings Æ Chapter
7.4.3
Automatic zero point correction Æ Chapter
7.4.4
Automatic storage of tare and zero values Æ Chapter
7.4.5
Adaptation to environmental conditions and weighing mode Æ Chapter
7.4.6
Automatic adjustment(FACT) Æ Chapter
7.4.7
Minimum weight Æ Chapter
7.4.8
Reset 'SCALE' settings to factory settings Æ Chapter
7.4.9
Chapter 7: Vision Setup METTLER TOLEDO SQC16
64
7.4.1 Adjust/calibrate (SCALE Æ Calibration)
This function enables the scale to be adjusted/calibrated (weighing pan must be empty). Not available on certified scales!
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Calibration' and how to
navigate within the 'Vision Setup'-menu.
Press «Yes» and follow the instructions given on the screen.
After choosing the calibration weight, place the weight on the weighing pan and then press «OK».
It is recommended to use the maximum calibration weight from the list. If it is not possible to use the maximum weight, at least do not use less than one­third of the maximum load to ensure reliable weighing values.
Calibration is done. Continue with other functions or press «End» and confirm with «Yes» to save the changes.
Note: Abort calibration at any given time by pressing «Cancel».
7.4.2 Display resolution and weighing unit (SCALE Æ Display)
This function allows the user to change the weighing unit, and to set the resolution of the weight.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Display' and how to navigate
within the 'Vision Setup'-menu.
Press « » and select e.g. 'Unit 1', then «Edit» to change the unit.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 65
In order to change the settings of the resolution of the scale, select 'Resolution'.
Use « » and « » to select the desired resolution and press «OK».
7.4.3 Automatic zero point correction (SCALE Æ Tare)
This function allows the user to configure all the available tare function of the scale.
Please refer to Chapter
7.3.1 how to reach the menu position 'Vision Setup _ SCALE _ Tare' and how to navigate within
the 'Vision Setup'-menu.
Switch the available functions Auto Tare, Chain Tare, AutoClr Tare or
Pushb. Tare either On or Off.
Automatic tare automatically tares the scale once a load (>=9d) is placed on the weighing pan. If this is active, the icon
AT flashes on the lower right side of the display.
Chain tare allows several tare actions without clearing the tare memory. If this is not active, the tare memory has to be cleared by pressing the «C» softkey. The tare memory has to be cleared before a new tare can be performed.
Automatic clear tare automatically clears the tare memory once the load is removed from the weighing pan.
Push button tare enables/disables the use of the «ÆTÅ» key to perform manual taring.
7.4.4 Automatic zero point correction (SCALE Æ Zero)
With Auto Zero, small deviations in the weight (in the range of 50% of 1d) are automatically zeroed. Always active for certified scales!
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Zero' and how to navigate
within the 'Vision Setup'-menu.
Press « » to go to the Auto Zero settings.
Press «Edit» to assign the zero setting range or to switch the function off.
Chapter 7: Vision Setup METTLER TOLEDO SQC16
66
7.4.5 Automatic save of tare and zero values (SCALE Æ Restart)
This function allows the user to set the scale so that it is able to automatically save the tare and zero values after switching off or when a power outage occurs. Not available on certified scales. Automatic save switched off is the
factory setting.
Block can only be accessed by a supervisor.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Restart' and how to navigate
within the 'Vision Setup'-menu.
Press «Edit» to switch the automatic save switch function on or off.
7.4.6 Adaptation to environmental conditions and weighing mode (SCALE Æ Filter)
Vibration function allows the user to set the scale so that it is able to adapt itself to the existing environmental conditions.
Process function allows the user to set the weighing mode of the scale (weighing process adapter).
Stability function allows the use to adjust the weighing speed.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Filter' and how to navigate
within the 'Vision Setup'-menu.
Press « » to go to the vibration and process settings menu.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 67
Vibration:
Low: For very stable and stable environment. Scale operates very quickly but
is more sensitive to external influences.
Mild: For normal environment conditions. Scale operates at medium speed (factory setting).
High: For unstable environment. Scale operates more slowly but is less sensitive to external influences.
Process: Universal: Setting for all weighing types and normal weighing goods (factory
setting) Dosing: Setting for dispensing liquids or powdery substances.
Stability:
Fast: The scale operates very fast.
Standard: The scale operates of medium speed.
Precise: The scale operates with the greatest possible reproducibility.
The slower the scale works, the greater the reproducibility.
7.4.7 Automatic adjustment (SCALE Æ FACT)
FACT function allows the user temperature or time dependent adjustment. This menu item only appears on scales with an internal adjustment weight.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ FACT' and how to navigate
within the 'Vision Setup'-menu.
Press «
» to go to the FACT settings.
Press «Edit» to select the menu item temperature.
Use the «
» softkey to select the desired temperature difference for the automatic adjustment. Press «OK».
Chapter 7: Vision Setup METTLER TOLEDO SQC16
68
Defining up to 7 days of the week and up to 3 times for automatic adjustment.
- Select day of the week for automatic adjustment. 7 zeros appear in the display. The first zero stands for Monday, the second for Tuesday, the third for Wednesday etc.
- Use the key to go to the desired day of the week and enter 1. The display 0100100 means that Tuesday and Friday are selected as calibration days.
- Press «OK». Time 1 appears in the display.
- Enter the time(s) for the calibration (hours, minutes). The format for entering the time (EU or US) depends on the settings in the menu item TERMINAL-> Device.
- Press «OK».
Note: apart from the calibration days at least one time has to be defined in order to activate the time controlled calibration!
7.4.8 Minimum weight (SCALE Æ Min Weigh)
Min Weigh function allows the user to switch the minimum weight on and off. If the weight on the scale falls below the stored minimum value, an * appears on the display in front of the weight indicator. This menu item only appears if the service technician has saved a minimum weight.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Min Weigh' and how to
navigate within the 'Vision Setup'-menu.
Press «
» to go to the Min Weigh settings.
Press «Edit» to switch the minimum weight monitoring on/off.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 69
7.4.9 Resetting scale to factory default settings (SCALE Æ Reset)
This function resets the 'SCALE' block to its original factory settings. Can only be accessed by a supervisor.
Please refer to Chapter
7.3.1 on how to reach the menu position 'Vision Setup Æ SCALE Æ Reset' and how to navigate
within the 'Vision Setup'-menu.
Press «Yes» to reset the scale settings. 'SURE?' will appear, press «Yes» to confirm. The scale settings are now reset to its original factory settings.
7.5 SQC16 settings (SQC16)
This function block allows the user to change the SQC16 application specific settings of the scale. Can only be accessed by a supervisor.
Please refer to Chapter
7.3.2 on how to reach the menu position 'Vision Setup Æ SQC16' and how to navigate within the
'Vision Setup'-menu.
The screen shows the most important application settings as a preview.
Enter the SQC16 menu (« ») and choose one of the functions listed below.
Available functions and settings:
User Login Æ Chapter
7.5.1
Auto Log Out Æ Chapter
7.5.2
Minimum Password Length Æ Chapter
7.5.3
User Setup Æ Chapter
7.5.4
7.5.1 User Login (SQC16 Æ User Login)
If this function is turned off, SQC16 will allow the application to operate without the user having to log in.
Please refer to Chapter
7.3.2 on how to reach the menu position 'Vision Setup Æ SQC16 Æ User Login' and how to
navigate within the 'Vision Setup'-menu.
Press «Edit» to switch User Login on or off.
Confirm selection by pressing «OK».
Chapter 7: Vision Setup METTLER TOLEDO SQC16
70
7.5.2 Auto Log Out (SQC16 Æ Auto Log Out)
This function, if activated, automatically logs you out after each sampling procedure.
Please refer to Chapter
7.3.2 on how to reach the menu position 'Vision Setup Æ SQC16 Æ Auto Log Out' and how to
navigate within the 'Vision Setup'-menu.
Press «Edit» to switch on or off.
Confirm selection by pressing «OK».
7.5.3 Minimum Password Length (SQC16 Æ Minimum PW Length)
With this function, you can specify the minimum length of the passwords to be defined. If the user defines a password that has less than the specified number of digits, SQC16 will give a message that the password is invalid.
Please refer to Chapter
7.3.2 on how to reach the menu position 'Vision Setup Æ SQC16 Æ Minimum PW Length' and
how to navigate within the 'Vision Setup'-menu.
Press «Edit» to change the minimum length of password.
Select the desired number of length and confirm selection by pressing «OK».
7.5.4 User Setup (SQC16 Æ User Setup)
This function allows you to define and set up a maximum of 16 users that can work with SQC16 if “User Login” (Chapter
7.5.1) is switched on. You can define their name, ID number, type of access rights and reset their passwords from this function.
Please refer to Chapter
7.3.2 on how to reach the menu position 'Vision Setup Æ SQC16 Æ User Setup' and how to
navigate within the 'Vision Setup'-menu.
Press «Edit» to set up a new user or modify the settings of an existing user.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 71
Use « » and « » to select the user you wish to modify. To define a new user, select “<EMPTY>” from the user list and press «Edit» to modify the settings of the selected user.
Use « » and « » to select the parameter you wish to modify and press «Edit».
User User name (must be unique) Number User ID number Access Rights Select whether the user has Administrator or Operator
access rights. An Operator cannot enter the «
».
Password Reset the user’s password.
7.6 Terminal settings for device (TERMINAL Æ Device)
This function block allows the user to change display and peripheral oriented settings of the scale. Only the “Device” block is available to the user.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device' and how to
navigate within the 'Vision Setup'-menu.
The screen shows the most important terminal settings as a preview.
The screen shows a preview of the most important settings of the active item, e.g. those of device if 'Device' is active (i.e. highlighted).
Enter the TERMINAL menu (« ») and choose one of the functions listed below.
Available functions and settings:
Language settings Æ Chapter
7.6.1
Sleep function Æ Chapter
7.6.2
Contrast Æ Chapter
7.6.3
Invert Æ Chapter
7.6.4
Weight display Æ Chapter
7.6.5
Date and time Æ Chapter
7.6.6
Beep Æ Chapter
27.6.7
Chapter 7: Vision Setup METTLER TOLEDO SQC16
72
7.6.1 Language settings (TERMINAL Æ Device Æ Language)
This function allows the user to change the language settings of the scale.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Language' and
how to navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Language settings menu.
Use « » and « » to select the desired language setting and press «OK».
7.6.2 Sleep function (TERMINAL Æ Device Æ Sleep)
This function is useful to enhance the lifetime of the display backlight.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Sleep' and how
to navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Sleep settings menu.
The display backlight will turn itself off automatically when the selected time elapses.
7.6.3 Adjusting the contrast of the display (TERMINAL Æ Device Æ Contrast)
This function allows the user to adjust the contrast of the screen display.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Contrast' and
how to navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Contrast settings menu.
Select the degree of contrast using « » or « ». Press «OK» to confirm selection.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 73
7.6.4 Invert (TERMINAL Æ Device Æ Invert)
This function allows the user to select either a white or a black background of the screen display.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Invert' and how
to navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Invert settings menu.
Choose the desired setting (Yes or No) and then press «OK» to confirm selection.
7.6.5 Changing the size of weight display (TERMINAL Æ Device Æ Weight display)
This function allows the user to select either a small or a big weight display on the screen.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Weight display'
and how to navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Weight Display settings menu. Default setting is 'Small'.
Choose the desired size and then press «OK» to confirm selection.
7.6.6 Adjusting the date and time (TERMINAL Æ Device Æ Date Time)
This function allows the user to set date and time of the scale.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Date Time' and
how to navigate within the 'Vision Setup'-menu.
Press « » to go to the Date and Time settings menu.
Press «Edit» to change format either from US or EU format. Select 'Date' to change the date and 'Time' to adjust the internal clock of the scale. Press «OK»
to confirm selection.
Chapter 7: Vision Setup METTLER TOLEDO SQC16
74
7.6.7 Adjusting the date and time (TERMINAL Æ Device Æ Beep)
This function allows the user to switch the beep on or off, that is appearing on each key press.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Device Æ Beep' and how
to navigate within the 'Vision Setup'-menu.
Press the «Edit» key to go to the Beep settings menu.
Choose the desired setting (On or Off) and then press «OK» to confirm selection.
7.7 Define supervisor password (TERMINAL Æ Access)
This function allows the user to change the supervisor password of the scale. Can only be accessed by a supervisor.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Access' and how to
navigate within the 'Vision Setup'-menu.
Press « »
… and then press «Edit» to enter a new password.
Enter the password using the keypad, press «OK» to confirm. 'Retype Pwd' appears. Retype the new password and press «OK» again.
In case you forgot the password, you may press «
», « »,
«
», « » keys in sequence to access the Vision Setup.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 75
7.8 Reset terminal settings to factory settings (TERMINAL Æ Reset)
This function resets the 'TERMINAL' block to its original factory settings. The supervisor password ('TERMINAL Æ Access') will not be reset, only “Device” block.
Please refer to Chapter
7.3.3 on how to reach the menu position 'Vision Setup Æ TERMINAL Æ Reset' and how to
navigate within the 'Vision Setup'-menu.
Press «Yes» to reset the terminal settings. 'SURE?' will appear, press «Yes» to confirm. The terminal settings are now reset to its original factory settings.
7.9 Communication settings (COMMUNICATION)
This function block allows the user to change the peripheral settings of the scale. Differences will appear depending on the option pack installed. Can only be accessed by a supervisor.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION' and how to navigate
within the 'Vision Setup'-menu.
The screen shows the most important communication settings as a preview.
The screen shows a preview of the most important settings of the active item, e.g. those of COM1 if 'Com1' is active (i.e. highlighted).
Note: COM3 is not available for IND469SQC. COM2 is used for ETHERNET or WLAN (IND469 only) in case such option is installed.
Enter the COMMUNICATION menu (« »), select a port and choose one of the functions listed below.
Available functions and settings:
Mode Æ Chapter
7.9.1
Parameters Æ Chapter
7.9.2
Printer type Æ Chapter
7.9.3
Define Header Æ Chapter
7.9.4
Add Linefeed Æ Chapter
7.9.5
Reset 'COMMUNICATION' settings to factory settings Æ Chapter
7.9.6
PS2 settings Æ Chapter
7.9.7
Chapter 7: Vision Setup METTLER TOLEDO SQC16
76
7.9.1 Mode (COMMUNICATION Æ Mode)
This function enables the user to set the input/output mode of a COM port.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ Mode' and how to
navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Mode settings menu.
The mode setting can be changed to either Printer, Auto Print, Continuous
Weight, Dialog, 2
nd
Display, or LC-I/O. In this example, Com1 was set to
Printer.
Press «End» and confirm with «Yes» to save the changes.
7.9.2 Parameters (COMMUNICATION Æ Parameters)
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ Parameters' and
how to navigate within the 'Vision Setup'-menu.
The current Baud Rate, Parity and Handshake settings are displayed on the screen. Press «
» and …
… «Edit» to change settings of Baud, Parity or Handshake.
Baud: Parity: Handshake:
300 7-none No 600 7-odd XonXoff 1200 7-even 2400 8-none 4800 8-odd 9600 8-even 19200 38400 57600 (Only for COM3) 115200 (Only for COM3)
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 77
7.9.3 Printer type (COMMUNICATION Æ Printer type)
This function allows you to choose the type of printer to be used in printing out reports.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ Printer Type' and
how to navigate within the 'Vision Setup'-menu.
Press «Edit» to go to the Printer Type settings menu.
Choose the desired type of printer and then press «OK».
7.9.4 Header (COMMUNICATION Æ Define Header)
This feature is basically used to add header lines on the printouts during weighing mode. Up to 5 lines can be added. If defined for an A4/Report printer, line 1,2 and 3 will be used for header information in SQC mode print-outs as well.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ Define Header'
and how to navigate within the 'Vision Setup'-menu.
… when this screen appears, press « » enter the header settings menu.
Press «Edit» to edit a header line.
Using the keypad, type a header line text. Up to a maximum of 24 characters are allowed. Press «OK» when done. Identification information '.T.' and '.S.' (see
5.11.1) can be used as well.
7.9.5 Add LineFeed (COMMUNICATION Æ Add LineFeed)
This feature is used to add an empty Linefeed on the printout right after each report. The default is '4', meaning 4 empty lines are generated right after each report, to make it more convenient for the user to detach the printed output from the printer.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ Add LineFeed' and
how to navigate within the 'Vision Setup'-menu.
Chapter 7: Vision Setup METTLER TOLEDO SQC16
78
… when this screen appears, press «Edit» to change the Add LineFeed settings menu.
Use « » and « » to choose the desired number of lines and then press «OK» to confirm changes.
7.9.6 Reset communication (COMMUNICATION Æ Comx Æ Reset Com)
This function resets the 'COMMUNICATION' block to its original factory settings.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ Comx Æ Reset
Com' and how to navigate within the 'Vision Setup'-menu.
Press «Yes» to reset the settings of the current Com port. 'SURE?' will appear, press «Yes» to confirm. The chosen Com port is now reset to its original factory settings.
7.9.7 PS2 Settings (COMMUNICATION Æ PS2)
This function allows setting up of the peripherals that are attached via the PS2 Interface.
Please refer to Chapter
7.3.4 on how to reach the menu position 'Vision Setup Æ COMMUNICATION Æ PS2' and how to
navigate within the 'Vision Setup'-menu.
Press « » to go to the PS2 settings menu.
Select the item you wish to configure.
Keyboard
It is possible to attach a PS2 keyboard to the scale via the PS2 Interface. This function enables you to specify the layout of the keyboard that you are using.
Note: When a PS2 keyboard is attached, it is possible to work with SQC16 using only this keyboard. The keys F1-F6 simulate the softkeys, F9-F11 simulate the top 3 function keys of the scale. You can press «Enter» and «Esc» keys to simulate the «OK» and «Cancel» softkeys respectively. The «Backspace» key simulates the «C» key.
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 79
BCR
Aside from a keyboard, it is also possible to attach a barcode reader via the PS2 Interface. This function enables you to specify the designation of the data coming from the barcode reader used for Direct Input (i.e. when the scale is in weighing or SQC mode).
Not Used Ignore data coming from the barcode reader. Article search Use the data for searching through the article database.
Note: Whenever the scale is asking for user input, the barcode reader can always be used to scan in data in the context of the user input.
7.10 Diagnostic settings (DIAGNOSTICS)
This function block allows the user to verify if all keys are functioning properly. Can only be accessed by a supervisor.
Please refer to Chapter
7.3.5 on how to reach the menu position 'Vision Setup Æ DIAGNOSTICS' and how to navigate
within the 'Vision Setup'-menu.
Enter the DIAGNOSTICS menu (« ») and choose one of the functions listed below.
Available functions and settings:
Keyboard Test Æ Chapter
7.10.1
Display Test Æ Chapter
7.10.2
Serial Number 1 Æ Chapter
7.10.3
Serial Number 2 Æ Chapter
7.10.4
List1 Æ Chapter
7.10.5
List2 Æ Chapter
7.10.6
Reset All Æ Chapter
7.10.7
7.10.1 Keyboard (DIAGNOSTICSÆ Keyboard Test)
Please refer to Chapter
7.3.5 on how to reach the menu position 'Vision Setup Æ DIAGNOSTICS Æ Keyboard Test' and
how to navigate within the 'Vision Setup'-menu.
In order to check if all the keys are functioning properly, press «Yes» to start the keyboard test. Note: The test cannot be cancelled after starting ('End' has no function during test).
Chapter 7: Vision Setup METTLER TOLEDO SQC16
80
Press all 25 keys in sequence. If a key is functioning, the scale jumps to the next key. The keys are numbered as follows:
5 6 7 8 9 10
432
1
23
20
17
14
11
24
21
18
15
12
25
22
19
16
13
7.10.2 Display (DIAGNOSTICSÆ Display Test)
Please refer to Chapter
7.3.5 on how to reach the menu position 'Vision Setup Æ DIAGNOSTICS Æ Display Test' and how
to navigate within the 'Vision Setup'-menu.
In order to check if the screen is functioning properly, press «Yes» and…
… appears on the screen.
7.10.3 Serial Number 1 (DIAGNOSTICSÆ SNR1)
Please refer to Chapter
7.3.5 on how to reach the menu position 'Vision Setup Æ DIAGNOSTICS Æ SNR1' and how to
navigate within the 'Vision Setup'-menu.
The serial number of the scale is displayed
7.10.4 Serial Number 2 (DIAGNOSTICSÆ SNR2)
'SNR2' cannot be chosen within SQC16 (no analog option support)
METTLER TOLEDO SQC16 Chapter 7: Vision Setup 81
7.10.5 List 1 (DIAGNOSTICSÆ List 1)
Please refer to Chapter
7.3.5 on how to reach the menu position 'Vision Setup Æ DIAGNOSTICS Æ List 1' and how to
navigate within the 'Vision Setup'-menu.
Press «Yes» to print out the current settings of the whole Vision Setup menu.
7.10.6 List 2 (DIAGNOSTICSÆ List 2)
'List 2' cannot be chosen within SQC16 (no analog option support)
7.10.7 Reset All (DIAGNOSTICSÆ Reset All)
This function resets all blocks of the whole Vision Setup Menu to its original factory settings. The supervisor password ('TERMINAL Æ Access') will not be reset.
Please refer to Chapter
7.3.5 on how to reach the menu position 'Vision Setup Æ DIAGNOSTICS Æ Reset All' and how to
navigate within the 'Vision Setup'-menu.
Press « » to go to the List Memory settings menu and press «Yes» to reset all.
Chapter 8: Other important information METTLER TOLEDO SQC16
82
8 Other important information
This Chapter gives information on error messages and instructions for cleaning your scale. It also includes the declaration of conformity and technical data.
8.1 Error messages in display
Overload
Reduce the load on the scale or the preload.
Underload
Place weighing pan on the scale and ensure that it can move freely.
Weight reading does not stabilize
Ensure a tranquil environment.
Ensure that the weighing pan is free to move.
Change the setting of the vibration adapter (Chapter
7.4.6)
Not possible to zero scale
Ensure that the zeroing is only carried out in the permissible range and not under overload or underload conditions.
Timeout occurred
The scale was not able to attain a stable weight value within the time out period (15 seconds) after «Start» was pressed.
Value is out of range
The scale is in an underload or overload condition. Thus, it is not possible to perform sampling.
No calibration/adjustment
Disconnect the power cord plug and then plug it in again. If the error message reappears, calibrate/adjust the scale (Chapter
7.4.1). If this does
not help, contact your dealer or local representative.
Cannot save article
No nominal or tolerance values are defined. Define nominal and tolerance values in article definition.
METTLER TOLEDO SQC16 Chapter 8: Other important information
83
EAROM checksum error
Disconnect the power cord plug and then plug it in again. If the error message reappears, contact your dealer or local representative.
8.2 Messages on printer
1 48.20 │ │ ░│ │ │ Not plaus: 33.10 g 2 50.91 │ │ │ │ 3 50.91 │ │ │ │ 4 48.37 │ │ ░│ │ │ 5 50.90 │ │ │ │
Not plausible
Weight of sample is not within the plausible limits.
1 48.20 │ │ ░│ │ │
-------- ABORT ---------
Abort
«Abort»
softkey was pressed during sampling.
<T2- 0 0.00 % <T1- 1 20.00 % >T1+ 0 0.00 % >T2+ 0 0.00 % Report cancelled.
Report Cancelled
«Cancel»
softkey was pressed during printing of sampling report.
-------- SAMPLE --------
----20.03.80 - 03:21----
-- THE QUALITY COMPANY ­Operator RIZA
Name AAA
IndTare <--­Nominal 100.00 g
Preweigh values rejected
-------- ABORT ---------
Preweigh values rejected
«End»
softkey was pressed before the first sample was weighed.
1 99.83 │ │ │ │ 2 97.28 │ │░ │ │ │ Skipped 4 99.83 │ │ │ │ 5 50.90 │ │ │ ░│ │
Skipped
«Skip»
softkey was pressed during backweighing. If a tare sample is not
successfully carried out, it can be skipped by pressing the «Skip»
softkey.
Therefore, its corresponding preweigh values will be omitted.
Chapter 8: Other important information METTLER TOLEDO SQC16
84
T1 Critical 3.00 % T1 Reject 2.00 % <T2 Viols 1
T1 Critical Violator
Number or percentage of critical amount of T1- violators in the sample.
T1 Rejected Violator
Number or percentage of rejected amount of T1- violators in the sample.
<T2 Violators
Number of <T2 violators in the sample.
Stat1 T1 Critical 3.00 % Stat1 T1 Reject 2.00 % Stat1 <T2 Viols 1
Stat1 T1 Critical Violator
Number or percentage of critical amount of T1- violators in Stat1.
Stat1 T1 Rejected Violator
Number or percentage of rejected amount of T1- violators in Stat1.
Stat1 <T2 Violators
Number of <T2 violators in Stat1.
Stat2 T1 Critical 3.00 % Stat2 T1 Reject 2.00 % Stat2 <T2 Viols 1
Stat2 T1 Critical Violator
Number or percentage of critical amount of T1- violators in Stat2.
Stat2 T1 Rejected Violator
Number or percentage of rejected amount of T1- violators in Stat2.
Stat2 <T2 Violators
Number of <T2 violators in Stat2.
Batch T1 Critical 3.00 % Batch T1 Reject 2.00 % Batch <T2 Viols 1
Batch T1 Critical Violator
Number or percentage of critical amount of T1- violators in Batch.
Batch T1 Rejected Violator
Number or percentage of rejected amount of T1- violators in Batch.
Batch <T2 Violators
Number of <T2 violators in Batch.
Stat1: Sample 1 x 93.57 % 93.570 g
<TM-: x-Nom -6.430 g
s 7.32 % 6.850 g <T1- 3 60.00 %
Mean Tolerance Violator
Difference: Mean value minus Nominal value
METTLER TOLEDO SQC16 Chapter 8: Other important information
85
Signature:
........................
----------03:51--------­Batch cleared
New batch statistics
-------- SAMPLE --------
----20.03.80 - 03:51----
Batch cleared
When changing batch name at the start of sampling, statistics of old batch will be printed and cleared. A new batch statistics will be generated.
Signature:
........................
----------03:51--------­Stat1 cleared
Stat1 cleared
«PrtClr»
softkey was pressed in standby mode. All contents of Stat1 will be
deleted.
Signature:
........................
----------03:51--------­Stat2 cleared
Stat2 cleared
«PrtClr»
softkey was pressed in standby mode. All contents of Stat2 will be
deleted.
Stat1: Samples 2 x 96.74 % 48.370 g s 0.01 % 0.007 g
Δx = 1.64 g
----------03:56---------
Adjustment message
An adjustment message was printed due to steadily increasing or decreasing mean values over several samples.
Chapter 9: Appendix: Tolerance systems METTLER TOLEDO SQC16
86
9 APPENDIX: Tolerance systems
The tolerance system depends on the legal regulations applicable to the filling operations. All tolerance systems can be plus/minus or minus systems. Toggle with "+/-" in the article definition under "Tolerances".
9.1 Entry of tolerances
Entry of tolerances as a difference to the nominal value (N), in absolute value or in %, selectable in the menu "System" under "Tolerance Mode".
Example: N = 200 g, lower limit = 195 g
Entry in "Tolerances" Entry in "Tolerance Mode"
5 g t(rel) and Unit
2.5 % t(rel) and % 195 g T(abs) and Unit
97.5 % T(abs) and %
T1-or %
<T1-
t1-or %
200g
0g
195g
9.2 Legal tolerance systems
EU +/-
European Union
and Switzerland
Tolerances:
t1–, t1+, t2+ selectable, t2– fixed:
t2– := 2*t1–
Default values: t1– statutory, t1+ := t1– t2+ := t2–
T2+ T1+
T1-
T2-
>T1+
>T2+
<T1-
<T2-
t2+
t1+
t1-
t2-
N
EU -
European Union
and Switzerland
Tolerances: t1– selectable, t2–
t2– := 2*t1–
Default values: t1– statutory
<T1-
t2-
t1-
T1-
T2-
N
<T2-
METTLER TOLEDO SQC16 Chapter 9: Appendix: Tolerance systems
87
9.3 Free tolerance systems with one tolerance
Free 1+/-
Free tolerance system, e.g. for USA handbook 133 test plans A and B
Canada
Australia
Tolerances:
t–, t+ selectable.
default value: t+ := t–
>T+
.
<T-
t+
t-
N
T+
T-
Marg.
Free 1-
Free tolerance system, e.g. for USA handbook 133 test plans A and B
Canada
Australia
Tolerances:
t– selectable, (t–:=t1–)
.
<T-
t-
N
T
-
Marg.
9.4 Free tolerance systems with two tolerances
Free 2+/-
Free tolerance system
Tolerances:
t1+, t2+, t1–, t2– selectable.
Default values:
t2– := 2*t1–,
t+ := t–.
T2+
T1+
T1-
T2-
>T1+
>T2+
<T1-
<T2-
t2+
t1+
t1-
t2-
N
Marg.
Free 2-
Free tolerance system
Tolerances:
t1–, t2– selectable.
Default values:
t2– := 2*t1–
T1-
T2-
<T1-
<T2-
t1-
t2-
N
Marg.
9.5 Free tolerance systems with three tolerances
Free 3+/-
Free tolerance system
Tolerances:
t1+, t2+, t3+
t1–, t2–, t3– selectable.
Default values:
t2– := 2*t1–,
t3– := 3*t1–,
t+ := t–.
T2+
T1+
T1-
T2-
>T1+
>T2+
<T1-
<T2
-
t2+
t1+
t1-
t2-
N
T3+
T3-
t3-
t3+
>T3+
<T3-
Marg.
Free 3-
Free tolerance system
Tolerances:
t1–, t2–, t3– selectable.
Default values:
t2–:=2*t1–
t3–:=3*t1–
T1-
T2-
<T1-
<T2-
t1-
t2-
N
T3
-
t3-
<T3-
Marg.
Chapter 10: Appendix: Optional equipment METTLER TOLEDO SQC16
88
10 APPENDIX: Optional equipment
10.1 LC-I/O Relay Interface
The LC-I/O Relay Interface allows adjustment messages to be relayed to an attached machine and peripheral devices such as warning lamps, control motors or valves to be controlled and switched by the SQC16.
Output signals: The outputs are controlled as follows:
Function Outpu
t
On period
Violator check T1: critical 1
until confirmed with «
» key
Violator check T1: rejected 1
flashes until confirmed with «
» key
0.5 s/0.5 s
Violator check TM+ 2
until confirmed with «
» key
Violator check TM- 2
flashes until confirmed with «
» key
0.5 s/0.5 s
Adjustment- 3 adjustment * factor 1) 10 … 65000 ms, steps of 1 ms
Adjustment+ 4 adjustment * factor 1) 10 … 65000 ms, steps of 1 ms
Tolerance violator T1- 5 until sample removed
Tolerance violator T2- 6 until sample removed
Tolerance violator T3- 5+6 until sample removed
Tolerance violator T1+ 7 until sample removed
Tolerance violator T2+ 8 until sample removed
Tolerance violator T3+ 7+8 until sample removed
1)
See also "Adjustment" and "Factor" in Chapter 5.2
Example: Adjustment = 0.45 g Factor = 1.0 t = 0.45 sec Adjustment = 3.52 oz Factor = 0.1 t = 0.352 sec
Input signals: No input signals are relayed to SQC16.
METTLER TOLEDO SQC16 Chapter 10: Appendix: Optional equipment
89
10.2 Attachment of peripheral devices
The following peripheral devices can be controlled:
Bar-code reader and/or keyboard (PS2, fixed)
Host PC for Backup/Restore/Edit program BR16 (COM1, 3)
Strip printer (COM2/RS-P42)
A4/Report printer
Relay interface LC-I/O
Boldface = standard connection
The following devices (max. 3 at the same time) can be controlled via COM ports:
LC-I/O If the relay interface has to be controlled, it must be attached to one of the scale's COM
ports and that port must be set to "LC-I/O".
Host (PC) If the host has to be controlled, it must be attached to one of the scale's COM ports and
that port must be set to "Dialog".
Strip printer If a strip printer has to be controlled, it must be attached to one of the scale's COM ports
and that port must be set to "Printer" with the printer type set to the appropriate printer.
A4/Report printer (Epson LX-300) If an A4/Report printer has to be controlled, it must be attached to one of the scale's
COM ports and that port must be set to "Printer" with the printer type set to "LX-300".
Note: The bar-code reader and the keyboard must be plugged to the PS/2 connector.
Chapter 10: Appendix: Optional equipment METTLER TOLEDO SQC16
90
10.3 Accessories
BBA462, BBK462 (IND469, GA46 Æ CD / Datasheet) accessories:
Printer with normal paper
Strip Printer RS-P42, incl. RS232 cable RS-P42 229265
Strip Printer 'Sprinter 1', EURO version
Strip Printer 'Sprinter 1', UK version
RS232 cable for 'Sprinter 1' 1.8m (25/9-pin D­Sub, m/m crossover)
21253399
21253745
21253677
Accessories for RS-P42 or Sprinter 1
Paper rolls, 5 pcs
Ribbon cartridges, black, 2 pcs
72456
65975
Auxiliary display
Auxiliary display (not incl.: RS232 cable 410024) RS-PD/PASM 21302875
Barcode scanner (PS/2)
For attachment of barcode reader DATALOGIC
DLC7070-M1
21900880
Barcode scanner (PS/2Y)
For attachment of barcode reader and keyboard DATALOGIC
DLC7070-M1
21900881
Barcode scanner (wireless)
For attachment of barcode reader DATALOGIC
Gryphon M100-CS
21900949
Protective covers (1 incl.)
Protective cover for small platform scale
21203207
Protective cover for large platform scale
21203206
Antitheft device
Mechanical antitheft device 229175
Operating instructions
German
21901253
(1 incl.)
English
21901254
French
21901255
Italian
21901256
Spanish
21901257
Relay interface, digital outputs
8x220V (not incl.: RS232 cable 410024) LC-I/O 21202217
Keyboard
PC compatible Mini-Keyboard (US Layout) 21900944
Protective cover
For keyboard 21900944 21900945
PC program
BR16 Backup/Restore/Edit SQC16 Data
RS232 cable 1.8m (9-pin D-Sub, m/f, parallel)
BR16 21901246
410024
METTLER TOLEDO SQC16 Chapter 11: Technical data
91
11 Technical data
11.1 General data and delivered items
Standard delivery package:
Complete scale
Operating instructions
Quick Guide (with included CD)
11.1.1 BBA/BBK462
Applications
Weighing SQC
Settings Vibration adapter
Weighing process adapter Automatic zero correction Power-saving shutoff Display backlighting
Display Active point-matrix display, 35 mm high, with CFL backlit (235 x 64 pixel)
Interface
3 built-in RS232C interface (COM1, COM2, COM3; see
11.3) PS2 interface for keyboard and barcode reader Optional interfaces (Ethernet; instead of COM2)
Ambient conditions
Use Altitude Temperature range BBA4.. Temperature range BBK4.. Over voltage category Contamination level Relative humidity
Indoor use only up to 2000 m –10 .. +40 °C / 14 .. 104 °F +10 .. +30 °C / 50 .. 86 °F II 2 Maximum relative humidity 80 % for temperatures up to 31 °C / 88 °F, decreasing linearly to 50 % relative humidity at 40 °C / 104 °
Mains connection
Direct connection to the mains (MAINS supply voltage fluctuations up to ±10% of the nominal voltage):
230 V, 50 Hz, 70 mA 240 V, 50 Hz, 70 mA 120 V, 60 Hz, 90 mA 100 V, 50/60 Hz, 90 mA
For battery operation:
Connection via mains adapter: 90 – 264 V, 47 – 63 Hz, 300 mA In feed on the unit: 24 V, 1.3 A
Chapter 11: Technical data METTLER TOLEDO SQC16
92
Battery operation
If the voltage supply is interrupted, the unit automatically switches over to battery operation.
Net weight
Model Without battery With battery
With internal calibration
weight (without battery) BBA4.. – .. SM 4,6 kg 5,3 kg – BBA4.. – .. LA 8,2 kg 8,9 kg – BBK4.. – .. XS 4,9 kg 5,6 kg 5,4 kg BBK4.. – .. SM 4,7 kg 5,4 kg 5,2 kg BBK4.. – .. LA 10,5 kg 11,2 kg 11,7 kg
Protection Class (IEC 529, DIN 40050, EN60529)
IP43 (not with Ethernet interface)
Resolution of the analog second scale interface
300.000 points in non certified configuration 7.000 points in certified configuration
11.1.2 IND469
Applications
Weighing SQC
Settings
Automatic zero compensation mode during switching on and during operation Filter for adapting to the environmental conditions (vibration adapter) Filter for adapting to the weighing mode, e.g. dispensing (weighing process adapter) Switch-off function, sleep mode for power-operated devices, energy saving mode for storage battery operation Display illumination
Display
Active point-matrix display, 35 mm high, with CFL backlit (235 x 64 pixel)
Keyboard
Tactile-touch membrane keypad Scratch-resistant labeling
Housing
Stainless steel 1.4301 or AISI 304
Net weight
IND469 with AC power Supply unit
approx. 2.6 kg
IND469 with storage battery approx. 3.2 kg
Protection type (DIN 40050)
IP69k
Mains connection
Direct connection to power supply (supply voltage fluctuation not exceeding ±10 % of the rated voltage)
IND469 weighing terminal: Rated voltage 100 ... 240 VAC / 47 ... 63 Hz / 300 mA IND469xx weighing terminal: Rated voltage 230 VAC ±10 % / 47 ... 63 Hz / 300
mA
BBA469 compact scale: Rated voltage 100 ... 240 VAC / 47 ... 63 Hz / 300 mA
METTLER TOLEDO SQC16 Chapter 11: Technical data
93
Storage battery operation Supply at device: 24 VDC / 1.0 A
If the supply voltage is interrupted, the scale switches automatically over to storage battery operation
Ambient conditions
Application Height Temperature range Class III Temperature range Class II Over voltage category Degree of soiling Relative humidity
In interiors up to 2000 m –10 .. +40 °C / 14 .. 104 °F 0 .. +40 °C / 32 .. 104 °F II 2 Up to max. 80%, non condensing
Interface
2 built-in RS232C interface (COM1, COM2; see
11.3) PS2 interface for keyboard and barcode reader Optional interfaces (Ethernet or WLAN; instead of COM2)
11.2 Dimensions
11.2.1 BBA/BBK462
A B C D E F G H I K L
XS 335 265 100 240 200 46 276 208 216 165 165
SM 335 265 100 240 200 46 276 208 216
LA 370 360 115 350 240 52 310 304 310
1)
All dimensions in mm
* with adjustable feet fully screwed in
Chapter 11: Technical data METTLER TOLEDO SQC16
94
11.2.2 IND469
Dimensions in mm
11.3 Interface technical data
11.3.1 BBA/BBK462
The scale is provided with an EIA RS-232C (CCITT V24/V.28) voltage-controlled interface as standard. Maximum cable length is 15 m. All interfaces are in the form of a 9-pin D-sub female connector. Instructions for configuring the interfaces are given in Chapter
7.9.2.
Interface
1 (standard) 2 (standard) 3 (standard)
Type
RS232C RS232C RS232C
Pin 1 --- --- ---
Pin 2 TxD TxD TxD
Pin 3 RxD RxD RxD
Pin 4 --- --- ---
Pin 5 GND GND GND
Pin 6 --- --- ---
Pin 7 --- --- ---
Pin 8 --- --- ---
Pin assignment
Pin 9 VCC VCC VCC
TxD: Transmit data RxD: Receive data
GND: Signal ground VCC: Power supply +5V
METTLER TOLEDO SQC16 Chapter 11: Technical data
95
11.3.2 IND469
The device is equipped with 2 communication interfaces. The following combinations are possible:
COM1 COM2 Standard RS232 RS232 Ethernet RS232 Ethernet WLAN RS232 WLAN
11.4 Interface commands
Your scale can be configured and operated, and can communicate with a PC via RS232C interface.
11.4.1 Preconditions
The following preconditions must be fulfilled to achieve communication between the scale and a PC:
The scale must be connected to the RS232C interface of the PC using the necessary cable (e.g. 410024).
The scale interface must be set in “Dialog” mode (see Chapter
7.9.1).
A terminal program must be available on the PC (e.g. “Hyper Terminal”).
The communication parameters (protocol, bits and parity, data transfer rate) must be set using the same values in
the terminal program and in the scale (see Chapter
7.9.2).
11.4.2 SICS Command set
Your scale basically supports the Mettler Toledo Standard Interface Command Set (MT-SICS). The SICS command set used are “Level 0”, “Level 1”, and some “Level 2” commands. Detailed information on the interface commands is given in the “MT SICS Reference Manual” (No. 705184, only available in English).
Notes:
Every command line must be terminated with <CR><LF> (corresponding to the “Enter” or “Return” key on the PC
keyboard). The command is then executed immediately. To correct a line, this needs to be entered again completely.
For commands with parameters, the “_” symbol signifies an empty space, and in the examples given, serves solely
to clarify the syntax.
For commands that require text parameters, the quotation marks must be entered, as they indicate to the scale that
they enclose a text string and not another parameter.
Chapter 11: Technical data METTLER TOLEDO SQC16
96
The following lists the supported MT-SICS commands:
I0 Inquiry of all implemented MT-SICS commands. I1 Inquiry of MT-SICS level and MT-SISCS versions. I2 Inquiry of scale data. I3 Inquiry of scale software version. I4 Inquiry of serial number. S Send stable weight value. SI Send weight value immediately irrespective of scale stability. SIR Send weight values repeatedly irrespective of scale stability. Z Zero the scale. ZI Zero the scale immediately irrespective of scale stability. @ Reset the scale to condition after switching on but without a zero setting being performed. D Write text into scale display. (E.g. D_“text”). DW Switch back to weight display after D command. K Configure key control.
SR
Send current stable weight value and then continuously on weight change equal to or greater than the preset value. (E.g. SR_10.00_g).
T
Tare the scale.
TA
Inquire or preset the tare weight value.
TAC
Clear the tare value. TI Tare immediately regardless of whether the current value is stable or not. C2 Initiate calibration. I10 Inquire or set the scale ID. I11 Inquiry of scale type. DAT Inquire of set the date on the scale. P100 Print out text on the printer. (E.g. P100_“text”). P101 Print out current stable weight value. P102 Print out current weight value irrespective of scale stability. PWR Switch the scale on or off. ST
Send stable weight after pressing the «
» key. TIM Inquire or set the time on the scale. SU Send stable weight value with currently displayed unit. SIU Send weight value with currently displayed unit immediately irrespective of scale stability. SIRU Send weight value with currently displayed unit immediately and repeat. SRU Send stable weight value with currently displayed unit and repeat on weight change greater than or equal to
preset value.
Aside from the standard MT-SICS commands, SQC16 also provides the following commands for working with the scale: M01 Inquire or set the weighing mode ('Vision Setup Æ SCALE Æ Filter Æ Process': Chapter
7.4.6).
M01 Inquiry of weighing mode. M01_0 Set weighing mode to “Universal”. M01_1 Set weighing mode to “Filling”.
METTLER TOLEDO SQC16 Chapter 11: Technical data
97
M03 Inquiry or setting of AutoZero ('Vision Setup Æ SCALE Æ Zero Æ Auto Zero': Chapter
7.4.4).
M03 Inquiry of AutoZero mode. M03_0 Set “AutoZero” to off. M03_1 Set “AutoZero” to on.
M09 Inquiry or setting of display contrast ('Vision Setup Æ TERMINAL Æ Device Æ Contrast': Chapter
7.6.3)
M09 Inquiry of display contrast. M09_x Set the display contrast to x% where x can have a value of 0-100.
M14 Inquiry of available languages
M14
--> M14_B_x_"<First Language>" ... M14_A_x_"<Last Language>"
where x (all these terms are in English):
0 = English 1 = German 2 = French 3 = Spanish 4 = Italian 8 = Polish 10 = Czech 11 = Hungarian 12 = Slovak 13 = Slovenian 14 = Dutch
Example:
M14
--> M14_B_0_"English" M14_B_1_"German" M14_B_2_"French" M14_B_3_"Spanish" M14_B_4_"Italian" M14_B_8_"Polish" M14_B_10_"Czech" M14_B_11_"Hungarian" M14_B_12_"Slovak" M14_B_13_"Slovenian" M14_A_14_"Dutch"
M15 Inquiry/setting of language (see also chapter
7.6.1).
M15 Inquiry of actual language.
M15_x Setting of language.
x: Number according to available languages (see command M14).
Examples: M15
--> M15_A_0 <English is currently set>
M15_1
--> M15_A <German is set now>
Chapter 11: Technical data METTLER TOLEDO SQC16
98
M16 Inquiry or setting of the sleep status ('Vision Setup Æ TERMINAL Æ Device Æ Sleep': Chapter
7.6.2)
M16 Inquiry of “Sleep” status. M16_x Set the “Sleep” status to the value of x where x can be: 0 = Off 1 = 1 minute 2 = 3 minutes 3 = 5 minutes
M19 Inquiry of adjustment weight. M21 Inquiry or setting of “Unit1” ('Vision Setup Æ SCALE Æ Display': Chapter
7.4.2) and display unit.
M21 Inquiry of “Unit1” and display unit. M21_Des_x Set the “Des” (designation) unit to the value of x where Des: x: 0 = Unit 1 0 = g 1 = Display unit 1 = kg 2 = t 7 = lb 8 = oz
I31 Inquire or define the record header for printouts in 'Weighing mode'. For SQC, see special header and footer
lines in system setup.
I31_x Inquiry the definition for the x header line. I31_x_”text” Define the text for a specific header line where:
x 1..5 text String of characters with a maximum of 24 characters.
Example: I31_1_“Mettler Toledo GmbH”
I31_2_“Heuwinkelstrasse” I31_3_“CH-8606 Naenikon” I31_4_“Telefon 01/944 22 11” I31_5_“Internet www.mt.com”
The header can be viewed in 'Vision Setup Æ COMMUNICATION Æ Define Header'.
METTLER TOLEDO SQC16 Chapter 11: Technical data
99
11.5 Table of Geo Values
For weighing instruments verified at the manufacturer’s, the geo value indicates the country or geographical zone for which the instrument is verified. The geo value set in the instrument (e.g."Geo 18") appears briefly after switch-on or is specified on a label. Table GEO VALUES 3000GEO 3000e shows the geo values for European countries. Table GEO VALUES 6000e/7500e shows the geo values for different gravitation zones.
11.5.1 GEO VALUES 3000e, OIML Class III (European Countries)
Geographical latitude Geo Value Country
46°22' - 49°01' 18 Austria 49°30' - 51°30' 21 Belgium
41°41'- 44°13' 16 Bulgaria 42°24' - 46°32' 18 Croatia 48°34' - 51°03' 20 Czech Republic 54°34' - 57°45' 23 Denmark 57°30' - 59°40' 24 Estonia 59°48' - 64°00' 64°00' - 70°05'
25*
26
Finland
41°20' - 45°00' 45°00' - 51°00'
17
19*
France
47°00' - 55°00' 20 Germany 34°48' - 41°45' 15 Greece 45°45' - 48°35' 19 Hungary
51°05 - 55°05' 22 Ireland 63°17' - 67°09' 26 Iceland 35°47' - 47°05' 17 Italy 55°30’ - 58°04' 23 Latonia 49°27' - 50°11' 20 Luxembourg 47°03' - 47°14' 18 Liechtenstein 53°54' - 56°24' 22 Lithuania 50°46' - 53°32' 21 The Netherlands 57°57' - 64°00' 64°00' - 71°11'
24*
26
Norway
49°00' - 54°30' 21 Poland 36°58' - 42°10' 15 Portugal 43°37' - 48°15' 18 Romania 55°20' - 62°00' 62°00' - 69°04'
24*
26
Sweden
45°49' - 47°49' 18 Switzerland 47°44' - 49°46' 19 Slovak Republic 45°26' - 46°35' 18 Slovenia 36°00' - 43°47' 15 Spain 35°51' - 42°06' 16 Turkey 49°00' - 55°00' 55°00' - 62°00'
21*
23
Great-Britain
*factory setting
Chapter 11: Technical data METTLER TOLEDO SQC16
100
11.5.2 GEO VALUES 6000e/7500e OIML Class III (Height £ 1000 m)
Geographical latitude Geo Value
00°00’ - 12°44’ 5 05°46’ - 17°10’ 6 12°44’ - 20°45’ 7 17°10’ - 23°54’ 8 20°45’ - 26°45’ 9 23°54’ - 29°25’ 10 26°45’ - 31°56’ 11 29°25’ - 34°21’ 12 31°56’ - 36°41’ 13 34°21’ - 38°58’ 14 36°41’ - 41°12’ 15 38°58’ - 43°26’ 16 41°12’ - 45°38’ 17 43°26’ - 47°51’ 18 45°38’ - 50°06’ 19 47°51’ - 52°22’ 20 50°06’ - 54°41’ 21 52°22’ - 57°04’ 22 54°41’ - 59°32’ 23 57°04’ - 62°09’ 24 59°32’ - 64°55’ 25 62°09’ - 67°57’ 26 64°55’ - 71°21’ 27 67°57’ - 75°24’ 28 71°21’ - 80°56’ 29 75°24’ - 90°00’ 30