Mettler Toledo AB-S/FACT, PB-S/FACT, AB-S/PH, PB-S/PH, PB-L/PH Operating Instructions Manual

Operating instructions

METTLER TOLEDO B-S/FACT line of balances
• AB-S/FACT
• PB-S/FACT
off
O/T
On
off
O/T
METTLER TOLEDO
C
CLASSIC Plus
F
Cal/Menu
On
off
O/T
METTLER TOLEDO
C
Plus
CLASSIC
F
Cal/Menu
METTLER TOLE
DO
us
C
Pl
CLASS IC
F
Cal/Menu
On
off
O/T
METTLER TOLEDO
C
Plus
CLASSIC
F
Cal/Menu
www.mt.com/classic

Operating instructions in a nutshell

Press key briefly
Press and hold key
down until the desired display appears
automatic sequence
Switching on
On
88888888
0.00 g
Switching off
Off
OFF
Simple weighing
0.00 g
On
off
O/T
METTLER TOLEDO
C
• 1182.03 g
F
Cal/Menu
1250.00 g
Adjustment options
Cal/Menu
±
±
Taring
METTLER T
OLEDO
C
F
Cal/Menu
METTLER TOLEDO
C
F
Cal/Menu
O/T
METTLER TOLEDO
C
F
Cal/Menu
0.00 g
121.01 g
0.00 g
95.97 g
-121.01 g
O n
off
O/T
#
On
off
O/T
On
off
ŸÖEMU
CAL int
CAL E
AutoCal
FACt on
FACt off
Piece counting*
On
off
O/T
M
ETTLER
TO LED
O
C
F
Cal /Menu
#
Cal
On
o ff
O/T
METTLER TOLEDO
C
F
Cal
Cal
/Menu
F
±
On
o ff
O/T
METTLER TOLEDO
C
F
Cal /M
enu
Percent weighing*
On
o ff
O/T
METTLER TOLEDO
C
F
Cal /M enu
F
±
On
off
O/T
M ETTLER
T
OLEDO
C
F
Cal/Menu
Dynamic weighing*
Automatic start (Dyn A)
0.00 g
0.00 g
12.07 g
SEt 10
SEt 20
20
244
147.25 g
0.00 g
PCS
PCS
PCS
PCS
0.00 g
• 257.35 g
O n
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
‹ 260.33 g
Manual start (Dyn M)
0.00 g
0.00 g
• 257.35 g
O n
off
O/T
METTLER TO
LEDO
C
F
Cal/Menu
±
<
<
<
-- 3 --
<
-- 1 --
<
<
<
<
-- 3 --
<
-- 1 --
12.07 g
SEt100 %
100.0 %
Unit switching*
101.6 %
*These functions must be activated in the menu (section 4.3.4)
<
‹ 260.33 g
22.00 g
0.78
Free factor / selectable display increment*
O n
off
O/T
METTLER TOLEDO
C
F
C a l /M e
n u
F
±
±
** 12, 3 ...9, 0
±
±
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
oz
19.50 g
FAC ŸÖ
1. 0000
0. 0000
**
0. 0000
(2x)
0 9000
StEP
0 0900
0. 1000
StorEd
19.50 g
17.5
19.50 g
17.5
19.50 g
#
#

Contents

1 Getting to know B-S/FACT balances ...............................................4
1.1 Introducing the B-S/FACT line of balances .........................................4
1.2 Layout of B-S/FACT balances ..........................................................5
1.3 Overview of key functions ...............................................................6
2 Startup ........................................................................................7
2.1 Unpacking / standard equipment .....................................................7
2.2 Cautionary notes ...........................................................................7
2.3 Installing the draft shield ................................................................8
2.4 Setting up, leveling, preparations for weighing below the balance,
connecting to power supply ............................................................9
2.5 Adjusting (calibration) .................................................................11
3 Weighing ...................................................................................13
3.1 On/off switching .........................................................................13
3.2 Simple weighing ..........................................................................13
3.3 Taring ........................................................................................14
3.4 METTLER TOLEDO DeltaRange balances ........................................15
4 Menu .........................................................................................16
4.1 Overview ....................................................................................16
4.2 Menu operation ...........................................................................17
4.3 Description of menu options .........................................................18
5 Functions ...................................................................................25
5.1 Piece counting ............................................................................25
5.2 Percent weighing .........................................................................26
5.3 Dynamic weighing .......................................................................27
5.4 Weighing with free factor and/or selectable display increments .........29
5.5 Switching weight units .................................................................31
6 Technical data, optional equipment .............................................32
6.1 Technical data ............................................................................32
6.2 Interface .....................................................................................34
6.3 MT-SICS Interface commands and functions ...................................35
6.4 Optional equipment .....................................................................37
6.5 Dimensional drawings .................................................................38
7 Appendix ...................................................................................40
7.1 Typical printouts from METTLER TOLEDO
RS-P26 and LC-P45 printers ........................................................40
7.2 What if ...? .................................................................................41
7.3 Maintenance and cleaning ............................................................42
3

1 Getting to know B-S/FACT balances

1.1 Introducing the B-S/FACT line of balances

4
AB-S/FACT
(0.1 mg / 0.01mg)
PB-S/FACT
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal/Menu
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal /Menu
PB-S/FACT (0.1/1 g)
Balance features
• The B-S/FACT balance line ranges from high-resolution analytical balances (AB-S/FACT) with a readability of 0.1 mg / 0.01 mg through to precision balances (PB-S/FACT) with a readability of 0.001g to 1g. The weighing ranges extend from 51g to 8.1kg.
FACT (Fully Automatic Calibration Technology), fully automatic, motorized adjustment (calibration) with internal weight.
• In addition to basic weighing operations such as weighing, taring and adjusting (calibration) the functions
Piece counting”, “Percent weighing”, “Free Factor” or “Dynamic weighing” (automatic or manual start)
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal/Menu
can be activated.
• METTLER TOLEDO DualRange balances have two ranges. These models also have a fine (semimicro) ran- ge from 0 to 31 g respectively 61 g. In this fine range the balance shows the result with a higher resolution, i.e. with one decimal place more.
• METTLER TOLEDO DeltaRange balances also have a movable fine range, with 10 times smaller display increments, over the the entire weighing range.
PB-S/FACT (0.01g)
• Several B-S/FACT balances are fitted with a glass draft shield in the factory; with other models a draft shield is available as an optional extra.
• All balances are fitted with an RS232C interface as standard.
• All models have an internal adjustment weight.
Note
All B-S/FACT balances are available as certified versions. Please ask your METTLER TOLEDO dealer for de-
On
off
O/T
METTLER TOLEDO
Plus
C
CLASSIC
F
Cal/Menu
PB-S/FACT (0.001g)
tails.

1.2 Layout of B-S/FACT balances

6
5 4
3
2
On
off
O/T
1
METTLER TOLE
DO
us
C
Pl
CLASS IC
F
Cal/Menu
8
5
12
RS232C
1 Keys 2 Display with backlight (exept ABxx5-S/FACT) 3 Model plate with the following data:
“Max”: maximum capacity “d”: readability “Min”: minimum capacity (recommended minimum load; only relevant for certified balances) “e”: verification scale interval (smallest display increment tested during certification; only relevant for
certified balances) 4 Draft shield element 5 Weighing pan 6 Draft shield (standard supply with models AB-S/FACT and PBxx3-S/FACT) 7 Leveling feet 8 Hanger opening for weighing below the balance (underside of balance) 9 AC adapter socket
7
10 RS232C interface 11 Lug for optional antitheft device 12 Leveling control
7
11
10
9
7
5

1.3 Overview of key functions

C
O/ T
On
Off
Cal/ Menu
C
O/ T
On
Off
Cal/ Menu
6
The balances have two operator control levels: the weighing mode and the menu. The function of each indi- vidual key depends on the operator control level and how long the key is pressed.
Weighing mode
0.01 g
Menu
ŸÖENV
Key functions in weighing mode
Press briefly
OnSwitch on
Press and hold down
Off • Switch off balance (standby mode)
# • Zero/tare
F
CCancel function
Switch
Change settings
FCall function
A function must be activated in the menu, otherwise “F nonE” appears in the dis­play
± • Transfer weighing data via
interface
Confirm settings
Cal/MenuAdjust (calibrate)
• Show menu (hold key down until MENU appears)
Key functions in menu mode
Press briefly
C • Close menu (without
Press and hold down
----------
saving changes)
F
Change settings
± • Select menu options
----------
Cal/Menu • Save changes and close menu

2 Startup

2.1 Unpacking / standard equipment

On
off
O/T
METTLER TOLE
DO
C
F
C al
/M enu

2.2 Cautionary notes

All B-S/FACT balances are supplied in environmentally compatible packaging.
The standard equipment for every balance comprises:
AC adapter, to national codes
AC adapter with country-specific power cable (Models AB135-S/FACT / AB265-S/FACT)
Transparent plastic protective cover, fitted, to keep your balance clean at all times
Operating instructions, to ensure optimum utilization of your balance›s capabilities
CE declaration of conformity (in separate brochure 11780294)
AB-S/FACT with readability of 0.01/ 0.1 mg: Weighing pan ø 80 mm with draft shield element
• Draft shield “0.1 mg” (237 mm) PB-S/FACT with readability of 0.001 g: Weighing pan ø 100 mm with draft shield element
Draft shield “mg” (165 mm) PB-S/FACT with readability of 0.01 g: Weighing pan ø 180 mm with pan support
Retaining ring for protective cover, fitted PB-S/FACT with readability of 0.1/1 g: Weighing pan 180 x 180 mm with pan support
• Base plate and intermediate plate for protective cover,fitted
• B-S/FACT balances must not be operated in hazardous areas with the standard-supply AC adapter.
• Before connecting the AC adapter, verify that the voltage printed on it corresponds to the local mains voltage. If this is not the case, please contact your local METTLER TOLEDO dealer.
• B-S/FACT balances may only be used indoors in a dry environment.
• For use with an certified (CSA or equivalent) power supply, which must have a limited and SELV circuit output.
7

2.3 Installing the draft shield

8
Draft shield “0.1 mg” (237 mm):
ABxx5-S/FACT
1 2
1
2
On
off
O/T
METTLER TOLEDO
C
F
Cal
/Menu
On
off
O/T
1
On
off
O/T
METTLER TOLEDO
C
F
Cal
/Menu
2
METTLER TOLEDO
C
F
Cal/Menu

2.4 Setting up, leveling, preparations for weighing below the balance, connecting to power supply

The optimum location
The correct location makes an important contribution to the accuracy of the weighing results of high-resolution analytical and precision balances.
Hence, ensure a
• stable, vibration-free position as horizontal as possible
Avoid
• direct sunlight
• excessive temperature fluctuations
• drafts
The best location is on a stable bench in a corner protected against drafts, as far away as possible from doors, windows, radiators or the louvers of air conditioners.
Antitheft device
B-S/FACT balances are equipped with a lug for the optional antitheft device.
The antitheft device (cable with lock) is suitable for all models. It is available from METTLER TOLEDO unter the order number 590101.
9
10
Leveling
B-S/FACT balances have a spirit level and two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
Procedure
Adjust the two leveling feet appropriately until the air bubble comes to rest exactly in the middle of the glass: Air bubble at “12 o›clock” turn both feet counterclockwise
Air bubble at “3 o›clock” turn left foot clockwise, right foot counterclockwise Air bubble at “6 o›clock” turn both feet clockwise Air bubble at “9 o›clock” turn left foot counterclockwise, right foot clockwise
Note
The balance should be leveled and adjusted (Section 2.5) each time it is moved to a new location.
Preparations for weighing below the balance
To carry out weighing operations below the balance, the special cover on the underside of the balance must be slackened (care: when doing this the balance may only be laid on its side, not turned upside down!), turned through 180° and retightened. This exposes the opening for the hanger, making weighing below the balance possible.
Connecting the power supply
Before connecting the AC adapter, check that the voltage printed on it corresponds to the local mains
voltage. If this is not the case, please contact your local METTLER TOLEDO dealer.
Plug the AC adapter into the AC adapter socket on the balance, and connect to the power supply. The balance performs a self-test. This test is finished when “OFF” appears. Press the «On» key briefly: the balance is in operational readiness. Before any work is performed with the
balance, it must be adjusted (Section 2.5).
Notes
To achieve accurate results with analytical balances (AB-S/FACT), these must be connected to the power supply for at least 60 minutes to warm up to operating temperature before the first weighing operation is carried out. An optional AccuPac B-S (rechargeable external battery) can be used to operate all B-S/FACT balances inde­pendently of the mains power supply

2.5 Adjusting (calibration)

To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location. Adjusting is necessary:
• before the balance is used for the first time
• at regular intervals during weighing service
• after a change of location

2.5.1 Fully automatic adjustment (calibration) FACT

The factory setting is fully automatic adjustment FACT (Fully Automatic Calibration Technologiy) with the internal weight (see also Section 4.3.3). In this setting, you have no need worry about adjusting your balance. The balance adjusts itself automatically
• after the warm-up phase on connection to the power supply,
• when a change in the ambient conditions, e.g. the temperature could lead to a noticeable deviation in the
measurement.

2.5.2 Manual adjustment (calibration)

To obtain accurate results, the balance must be connected to the power supply for 30 minutes (AB-S/FACT analytical balances 60 minutes) in order to reach operating temperature before adjusting.
Cal/Menu
CAL int
à
------
à
CAL donE
à
0.0000 g
Adjusting with internal weight
To carry out this operation, in the second menu option (Adjustment) select “CAL int” (=factory setting)
(Section 4.1).
Cal
Unload weighing pan Press and hold the «Cal/Menu» key down until “CAL” appears in the display, then release key. The balance adjusts itself automatically.
The adjusting is finished when the message “CAL done” appears briefly in the display, followed by
“0.0000 g”. The balance is again in weighing mode and ready for operation.
11
Cal/Menu
Cal/Menu
On
off
O/T
C
On
off
O/T
C
12
Adusting with external weight
CAL E
à
To carry out this operation, in the second menu option (Adjustment) select “CAL E” Have required adjustment weight ready Unload weighing pan Press and hold the «Cal/Menu» key down until “CAL” appears in the display, then release key. The re-
CAL
à
quired adjustment weight value flashes in the display.
Place adjustment weight in center of pan. The balance adjusts itself automatically. When “0.00 g” flashes, remove adjustment weight.
2000.00
Cal
à
METTLER TOLEDO
F
Cal/Menu
------
0.00
Cal
à
Cal
à
METTLER TOLEDO
F
Cal/Menu
------
Cal
à
The adjusting is finished when the message “CAL done” appears briefly in the display, followed by
“0.00 g”. The balance is again in weighing mode and ready for operation.
Notes
• Because of certification legislation, the certified models cannot be adjusted with an external weight.
• The adjustment procedure can be terminated at any time with the «C» key. The message “Abort” appears
briefly to confirm that adjustment has been canceled, and the balance reverts to weighing mode.
CAL donE
à
0.00 g

3 Weighing

3.1 On/off switching

On
0.00 g
Off

3.2 Simple weighing

0.00 g
88888888
à
OFF
à
Switching on
Remove any load from weighing pan and press «On» key briefly. The balance performs a display test (all segments in the display light up briefly). When zero is displayed, the balance is ready for operation.
Switching off
Press and hold the «Off» key down until “OFF” appears in the display. Release the key.
Place weighing sample on the weighing pan.
On
off
O/T
METTLER TOLEDO
C
F
Cal /Menu
• 1182.03 g
Read the result.
Wait until the stability detector “
” disappears.
1250.00 g
13

3.3 Taring

On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
#
0.00 g
121.01 g
0.00 g
14
Place empty container on the balance.
The weight is displayed.
Tare: press the «#» key briefly.
95.97 g
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
-121.01 g
On
off
O/T
METTLER TOLEDO
C
F
Cal/Menu
Add weighing sample to container. The net weight is now displayed.
If the container is removed from the balance, the tare weight will be shown as a negative value. The tare weight remains stored until the «#» key is pressed again or the balance is switched off.
Note
With METTLER TOLEDO DeltaRange balances (next Section), the fine range with its 10 times smaller display increments is available again after every taring operation.