Pôle Test et Mesure de CHAUVIN ARNOUX
6, avenue du Pré de Challes - F - 74940 ANNECY-LE-VIEUX
Tél. +33 (0)4.50.64.22.22 - Fax +33 (0)4.50.64.22.00
692917A01 - Ed. 02 - 09/18
English
1. PRECAUTIONS FOR USE
Failure to observe the safety instructions may cause an electric
shock, fire, explosion, or destruction of the instrument and of the
installations.
These safety instructions ensure the safety of persons and
correct operation of the device. If the tester is used in a way not
specified in this data sheet, the protection provided by the device
may be impaired;
This device complies with safety standard EN 61010-1 (Ed-2–
2001) for non current-carrying use, at an altitude below 2000 m,
indoors, with a level of pollution not exceeding 2.
• Do not use the instrument in an explosive atmosphere or in
the presence of flammable gases or fumes;
• Use the device only on non current-carrying circuits. Do not
apply voltages exceeding 50 V peak between the inputs or
with respect to earth;
• Do not use the instrument if it appears to be damaged,
incomplete, or not properly closed;
• Observe the environmental conditions of use;
• Do not modify the instrument and do not replace components
with "equivalents". Repairs and adjustments must be done by
approved qualified personnel;
• The tester must be warmed up for 30 seconds before use;
• In the presence of equipment emitting powerful electric fields,
the display may become unstable or erroneous.
* Replace the batteries as soon as the
symbol appears on
the display unit.
* If the tester is to be left unused for an extended period, remove
the batteries and store them separately.
Thank you for buying a TCX01 tester.
For best results from your device :
Read this user manual attentively ;
Observe the precautions for its use.
Meanings of the symbols used :
Risk of danger. The operator agrees to
refer to these instructions whenever this
danger symbol appears.
Earth
Indicates that the device is subject to the
waste segregation requirements of
directive 2002/96/EC.
The CE marking indicates compliance
with European directives.
2. DESCRIPTION
The TCX 01 tester is used for automatic measurements of
resistors and capacitors and for junction and continuity
tests on non current-carrying components or circuits.
The clip arrangement of the probe tips is especially well
suited to surface mounted components (SMC).
2.1 Presentation of the device
1. Measurement probe tip/cathode (-)
2. Measurement probe tip/anode (+)
3. “RANGE” key
4. “FUNC.” Key
5. LCD display unit
6. Battery compartment cover
2.2 LCD display unit
2.3 FUNC. key (Function key)
Press this key for more than one second to start up the tester; it
then switches to the search and automatic function recognition
mode .
Press this key repeatedly for less than one second to select the
desired measurement function.
Press this key for more than 2 seconds to switch the tester to
standby mode.
2.4 RANGE key (Change of range)
In automatic mode, press this key for less than 1 second to
switch the tester to manual mode.
In manual mode, press this key for more than 1 second to switch
the tester to automatic mode.
In manual mode, press this key for less than one second to
change the measurement range.
2.5 Measurement probe tips
F
or measurements of diodes or of polarized capacitors, the
polarity of the probe tips must be observed:
• + : positive probe tip (anode)
• - : negative probe tip (cathode)
3. TECHICAL CHARACTERISTICS
3.1 General characteristics
• Environmental conditions
Level of pollution: 2
Altitude < 2000 m
Operating temperature:
0~40°, (< 80% RH, without condensation)
Storage temperature:
-10~60°, (< 70% RH, without battery)
• Temperature coefficient:
0.1´(specified uncertainty)/°C (< 18° or > 28°)
• Max. voltage between terminals and with respect to earth:
50 VDC or 36 V RMS in AC
• Display refresh frequency: 3/sec.
• Display: 4 digits, with a maximum of 5999 points.
• Range overshoot indication: “OL” is displayed
• Spent battery indication:
The indication “ ” is displayed when the batteries are
spent and must be replaced.
• Automatic shutdown
If no key is pressed for 10 minutes, the tester automatically
switches itself off to save the batteries.
• Power supply: 2 1.5 V batteries (AG13 button)
• Dimensions: 181 (L) ´ 35 (W) ´ 20 (H) mm
• Mass: approximately 65 g (including batteries)
3.2 Measurement characteristics
Uncertainty: ± (% of reading + number of points) from 18°C to
28°C (64°F to 82°F) with relative humidity less tha n 80%.
NB: the number of points corresponds to the value of the
smallest (or last) significant digit.
NB : The number of points corresponds to the value of the
smallest (or last) significant digit.
3.2.1 Resistance
Range Resolution Uncertainty
600 Ω 0,1 Ω
6 kΩ 1 Ω
60 kΩ 10 Ω
600 kΩ 100 Ω
6 MΩ 1 kΩ
± (1.2 % of the reading