METREL MI 3108 EurotestPV Instruction Manual

Page 1
EurotestPV
MI 3108
Instruction manual
Version 1.2, Code no. 20 751 987
Page 2
Distributor:
Manufacturer: METREL d.d.
Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si
Mark on your equip ment cert ifies t hat this eq uipm ent me et s the req uire ment s of th e EU (European Unio n) concerning safety an d electromagneti c compatibility reg ul ati ons
© 2013 METREL
The trade names Metrel, Smartec, Eurotest, Autosequence are trademarks registered or pending in Europe and other countries.
No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
Page 3
MI 3108 EurotestPV Table of contents
Table of contents
1 Preface .................................................................................................................... 6
2 Safety and operational considerations ................................................................. 7
2.1 Warnings and notes .......................................................................................... 7
2.2 Battery and charging ....................................................................................... 12
2.3 Standards applied ........................................................................................... 14
3 Instrument description ......................................................................................... 16
3.1 Front panel ...................................................................................................... 16
3.2 Connector panel .............................................................................................. 17
3.3 Back side ......................................................................................................... 18
3.4 Carrying the instrument ................................................................................... 19
3.4.1 Secure attachment of the strap .......................................................................... 20
3.5 Instrument set and accessories ....................................................................... 21
3.5.1 Standard set MI 3108 ........................................................................................ 21
3.5.2 Optional accessories ......................................................................................... 21
4 Instrument operation ............................................................................................ 22
4.1 Display and sound ........................................................................................... 22
4.1.1 Terminal voltage monitor ................................................................................... 22
4.1.2 Battery indication ............................................................................................... 22
4.1.3 Messages .......................................................................................................... 22
4.1.4 Results .............................................................................................................. 23
4.1.5 Sound warnings ................................................................................................. 23
4.1.6 Help screens ..................................................................................................... 23
4.1.7 Backlight and contrast adjustments ................................................................... 24
4.2 Function selection ........................................................................................... 24
4.3 Instruments main menu ................................................................................... 26
4.4 Settings ........................................................................................................... 26
4.4.1 Memory ............................................................................................................. 27
4.4.2 Language .......................................................................................................... 27
4.4.3 Date and time .................................................................................................... 27
4.4.4 RCD standard .................................................................................................... 28
4.4.5 Isc factor ............................................................................................................ 29
4.4.6 Commander support .......................................................................................... 29
4.4.7 Communication ................................................................................................. 30
4.4.8 Initial settings ..................................................................................................... 31
4.4.9 Clamp Settings .................................................................................................. 33
4.4.10 Synchronization (A 1378 - PV Remote unit) ....................................................... 33
4.4.11 Solar settings ..................................................................................................... 35
5 Measurements – a.c. LV installatio ns ................................................................. 38
5.1 Voltage, frequency and phase sequence ........................................................ 38
5.2 Insulation resistance ........................................................................................ 40
5.3 Resistance of earth connection and equipotential bondi ng ............................. 42
5.3.1 R LOWΩ, 200 mA resistance measurement ...................................................... 42
5.3.2 Continuous resistance measurement with low current ....................................... 43
5.3.3 Compensation of test leads resistance .............................................................. 44
5.4 Testing RCDs .................................................................................................. 46
5.4.1 Contact voltage (RCD Uc) ................................................................................. 47
Page 4
MI 3108 EurotestPV Table of contents
5.4.2 Trip-out time (RCDt) .......................................................................................... 48
5.4.3 Trip-out current (RCD I) ..................................................................................... 48
5.4.4 RCD Autotest .................................................................................................... 50
5.5 Fault loop impedance and prospective fault current ........................................ 53
5.6 Line impedance and prospective short-circ ui t cur r ent / Volt ag e dr op .............. 55
5.6.1 Line impedance and prospective short circuit current ........................................ 56
5.6.2 Voltage drop ...................................................................................................... 57
5.7 Earth resistance .............................................................................................. 59
5.8 PE test terminal ............................................................................................... 61
6 Solar measurements - PV systems ..................................................................... 63
6.1 Insulation resistance on PV systems ............................................................... 63
6.2 PV inverter test ................................................................................................ 65
6.3 PV panel test ................................................................................................... 67
6.4 Measuring of environmental parameters ......................................................... 69
6.4.1 Operation with A1378 PV Remote Unit .................................................................... 71
6.5 Uoc / Isc test .................................................................................................... 71
6.6 I / V curve measurement ................................................................................. 73
7 Measurements - Power & Energy ........................................................................ 75
7.1 Power .............................................................................................................. 75
7.2 Harmonics ....................................................................................................... 76
7.3 Scope .............................................................................................................. 77
7.4 Current ............................................................................................................ 78
7.5 Energy ............................................................................................................. 79
8 Data handling ........................................................................................................ 81
8.1 Memory organization ....................................................................................... 81
8.2 Data structure .................................................................................................. 81
8.3 Storing test results ........................................................................................... 83
8.4 Recalling test results ....................................................................................... 84
8.5 Clearing stored data ........................................................................................ 85
8.5.1 Clearing complete memory content ................................................................... 85
8.5.2 Clearing measurement(s) in selected location ................................................... 85
8.5.3 Clearing individual measurements ..................................................................... 86
8.5.4 Renaming installation structure elements (upload from PC) .............................. 87
8.5.5 Renaming installation structure elements with serial barcode reader or RFID reader 87
8.6 Communication ............................................................................................... 88
8.6.1 USB and RS232 communication ....................................................................... 88
8.6.2 Bluetooth communication .................................................................................. 89
9 Upgrading the instrument .................................................................................... 90
10 Maintenance ...................................................................................................... 91
10.1 Fuse replacement ............................................................................................ 91
10.2 Cleaning .......................................................................................................... 91
10.3 Periodic calibration .......................................................................................... 91
10.4 Service ............................................................................................................ 91
11 Technical specifications................................................................................... 92
11.1 Insulation resistance, Insulation resistance of PV systems ............................. 92
Page 5
MI 3108 EurotestPV Table of contents
11.2 Continuity ........................................................................................................ 93
11.2.1 Resistance R LOWΩ ......................................................................................... 93
11.2.2 Resistance CONTINUITY .................................................................................. 93
11.3 RCD testing ..................................................................................................... 93
11.3.1 General data ...................................................................................................... 93
11.3.2 Contact voltage RCD-Uc ................................................................................... 94
11.3.3 Trip-out time ...................................................................................................... 94
11.3.4 Trip-out current .................................................................................................. 95
11.4 Fault loop impedance and prospective fault current ........................................ 95
11.4.1 No disconnecting device or FUSE selected ....................................................... 95
11.4.2 RCD selected .................................................................................................... 96
11.5 Line impedance and prospective short-circuit current / Voltage dr op .............. 96
11.6 Resistance to earth ......................................................................................... 97
11.7 Voltage, frequency, and phase rotation ........................................................... 97
11.7.1 Phase rotation ................................................................................................... 97
11.7.2 Voltage .............................................................................................................. 98
11.7.3 Frequency ......................................................................................................... 98
11.7.4 Online terminal voltage monitor ......................................................................... 98
11.8 TRMS Clamp current ....................................................................................... 98
11.9 Power tests ...................................................................................................... 99
11.10 PV tests ..................................................................................................... 100
11.10.1 Accuracy of STC data .................................................................................. 100
11.10.2 Panel, Inverter ............................................................................................. 100
11.10.3 I-V curve ...................................................................................................... 101
11.10.4 Uoc - Isc ...................................................................................................... 102
11.10.5 Environmental parameters ........................................................................... 102
11.10.6 Insulation Resistance of PV systems ........................................................... 103
11.11 General data .............................................................................................. 103
Appendix A - Fuse table ............................................................................................ 105
A.1 Fuse table - IPSC .......................................................................................... 105
Appendix B - Accessories for specific measurements .......................................... 108
Appendix C – Country notes .................................................................................... 110
C.1 List of country modifications .............................................................................. 110
C.2 Modification issues ............................................................................................ 110
C.2.1 AT modification - G type RCD................................................................................ 110
Appendix D – Commanders (A 1314, A 1401) ......................................................... 112
D.1 Warnings related to safety .......................................................................... 112
D.2 Battery ............................................................................................................... 112
D.3 Description of commanders ............................................................................... 112
D.4 Operation of commanders ................................................................................. 113
Appendix E – PV measurements - calculated values ............................................. 115
Page 6
MI 3108 EurotestPV Preface
1 Preface
Congratulations on your purchase of the Eurotest instrument and its accessories from METREL. The instrument was designed on a basis of rich experience, acquired through many years of dealing with electric installation test equipment.
The Eurotest instrument is a professional, multifunctional, hand-held test instrument intended to perform all the measurements on a.c. electrical LV installations and d.c. photovoltaic systems.
The following measurements and tests can be performed on a.c. electrical LV installations:
Voltage and frequency, Continuity tests, Insulation resistance tests, RCD testing, Fault loop / RCD trip-lock impedance measurements, Line impedance / Voltage drop, Phase sequence, Earthing resistance tests, Current measurements, Power, harmonics and energy measurements.
Measurements and tests on PV systems:
Voltages, currents and power in PV systems (Inverter and PV panels), Calculation of efficiencies and STC values in PV systems, Uoc / Isc measurements, Environmental parameters (Temperature and Irradiance), I-V curve test, Insulation resistance on PV systems.
The graphic display with backlight offers easy reading of results, indications, measurement parameters and messages. Two LED Pass/Fail indicators are placed at the sides of the LCD. The operation of the instrument is designed to be as simple and clear as possible and no special training (except for the reading this instruction manual) is required in order to begin using the instrument. In order for operator to be familiar enough with performing measurements in general and their typical applications it is advisable to read Metrel handbook Guide for testing and verification of low voltage installations.
The instrument is equipped with the entire necessary accessory for comfortable testing.
Page 7
MI 3108 EurotestPV Safety and operational considerations
2 Safety and operational considerations
2.1 Warnings and notes
In order to maintain the highest level of operator safety while carrying out various tests and measurements, Metrel recommends keeping your Eurotest instruments in good condition and undamaged. When using the instrument, consider the following general warnings:
General warnings related to safety:
The symbol on the instrument means »Read the Instruction manual
with special care for safe oper at io n«. The symbol requires an action!
If the test equipment is used in a manner not specified in this user manual,
the protection provided by the equipment could be impaired!
Read this user manual carefully, otherwise the use of the instrument may
be dangerous for the operator, the instrument or for the equ ipment under test!
Do not use the instrument or any of the accessories if any damage is
noticed!
Consider all generally known precautions in order to avoid risk of electric
shock while dealing with hazardous voltages!
If the 315 mA fuse blows follow the instructions in this manual in order to
replace it! Use only fuses that are specified!
Do not disassemble or repair the high breaking current fuse block! In case
it fails the entire block must be replaced with a new original one!
Do not use the instrument in AC supply systems with voltages higher than
550 Va.c.
Service, repairs or adjustment of instruments and accessories is only
allowed to be carried out by a competent authorized personnel!
Use only standard or optional test accessories supplied by your
distributor!
Consider that protection category of some accessories is lower than of the
instrument. Test tips and Tip commander have removable caps. If they are removed the protection falls to CAT II. Check markings on accessories! (cap off, 18 mm tip)…CAT II up to 1000 V (cap on, 4 mm tip)… CAT II 1000 V / CAT III 600 V / CAT IV 300 V
The instrument comes supplied with rechargeable Ni-MH battery cells. The
cells should only be replaced with the same type as defined on the battery compartment label or as described in this manual. Do not use standard alkaline battery cells while the power supply adapter is connected, otherwise they may explode!
Page 8
MI 3108 EurotestPV Safety and operational considerations
Hazardous voltages exist inside the instrument. Disconnect all test leads,
remove the power supply cable and switch off the instrument before removing battery compartment cover.
Do not connect any voltage source on C1 and P/C2 inputs. They are
intended only for connection of current clamps and sensors. Maximal input voltage is 3 V!
All normal safety precautions must be taken in order to avoid risk of
electric shock while workin g on el ect ri cal in stal lat io ns!
If the instrument is not in SOLAR operating mode the instrument displays a
warning if an external DC voltage of higher than 50 V is applied to the instrument. Measurements are blocked.
Warnings related to safety of measurement functions:
All PV functions
Use only dedicated accessories for testing on PV electrical installations.
Accessories for PV installations have yellow marked connectors. Appropriate warnings are displayed.
PV Safety probe A1384 has inbuilt protective circuit that safely disconnects the instrument from the PV installation in case of a failure in the instrument. PV test lead A1385 has integrated fuses that safely disconnects instrument from the PV installation in case of a failure in the instrument.
Do not use the instrument in PV systems with voltages higher than 1000 V
d.c. and/ or currents higher than 15 A d.c. ! Otherwise the instrument can be damaged.
PV sources can produce very high voltages and c urrents. Only skilled and
trained personnel should perform measurements on photovoltaic systems.
Local regulations should be considered. Safety precautions for working on the roof should be considered. In case of a fault in the measuring system (wires, devices, connections,
measuring instrument, accessories), presence of flammable gases, very high moisture or heavy dust an electrical arc can occur that will not extinguish by itself. Arcs can lead to fire and can cause heavy damage. Users must be skilled to disconnect the PV system safely in this case.
Insulation resistance, Insulation resistance of PV systems
Insulation resistance measurement should only be performed on de-energized
objects!
Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!
Page 9
MI 3108 EurotestPV Safety and operational considerations
When an insulation resistance measurement has been performed on a capacitive
object, automatic discharge may not be done immediately! The warning message
and the actual voltage are displayed during discharge until voltage drops
below 10 V.
Continuity functions
Continuity measurements should only be performed on de-energized objects! Parallel loops may influence on test results.
Testing PE terminal
If phase voltage is detected on the tested PE terminal, stop all measurements
immediately and ensure the cause of the fault is eliminated before proceeding with any activity!
Notes related to measurement functions: General
The indicator means that the selected measurement cannot be performed
because of irregular conditions on input terminals.
Insulation resistance, continuity functions and earth resistance measurements
can only be performed on de-energized objects.
PASS / FAIL indication is enabled when limit is set. Apply appropriate limit value
for evaluation of measurement results.
In the case that only two of the three wires are connected to the electrical
installation under test, only voltage indication between these two wires is valid.
Insulation resistance, Insulation resistance of PV systems
Insulation resistance:
If a voltage of higher than 30 V (AC or DC) is detected between test terminals, the insulation resistance measurement will not be performed.
Insulation resistance of PV systems:
Different pre-tests are carried out. If conditions are proper and safe the measurement will be be continued.
Otherwise or or message is displayed.
The instrument automatically discharge tested object after finished
measurement.
A double click of TEST key starts a continuous measurement.
Continuity functions
If a voltage of higher than 10 V (AC or DC) is detected between test terminals,
the continuity resistance test will not be performed.
Compensate test lead resistance before performing a continuity measurement,
where necessary.
RCD functions
Page 10
MI 3108 EurotestPV Safety and operational considerations
10
Parameters set in one function are also kept for other RCD functions! The measurement of contact voltage does not normally trip an RCD. However,
the trip limit of the RCD may be exceeded as a result of leakage current flowing to the PE protective conductor or a capacitive connection between L and PE conductors.
The RCD trip-lock sub-function (function selector switch in LOOP position) takes
longer to complete but offers much better accuracy of fault loop resistance (in comparison to the RL sub-result in Contact voltage function).
RCD trip-out time and RCD trip-out current measurements will only be performed
if the contact voltage in the pre-test at nominal differential current is lower than the set contact voltage limit!
The autotest sequence (RCD AUTO function) stops when trip-out time is out of
allowable time period.
Z-LOOP
The low limit prospective short-circuit current value depends on fuse type, fuse
current rating, fuse trip-out time and impedance scaling factor.
The specified accuracy of tested parameters is valid only if the mains voltage is
stable during the measurement.
Fault loop impedance measurements will trip an RCD. The measurement of fault loop impedance using trip-lock function does not
normally trip an RCD. However, the trip limit may be exceeded if a leakage current flows to the PE protective conductor or if there is a capacitive connection between L and PE conductors.
Z-LINE / Voltage drop
In case of measurement of Z
Line-Line
with the instrument test leads PE and N connected together the instrument will display a warning of dangerous PE voltage. The measurement will be performed anyway.
Specified accuracy of tested parameters is valid only if mains voltage is stable
during the measurement.
L and N test terminals are reversed automatically according to detected terminal
voltage (except in UK version).
Power / Harmonics / Energy / Current
Before starting any Power measurement the current clamp settings in Settings
menu should be checked. Select appropriate current clamp model and measuring range that are best fitted to the expected current values.
Consider polarity of current clamp (arrow on test clamp should be oriented
toward connected load), otherwise result will be negative!
PV measurements
A 1384 PV Safety Probe must be used for PANEL, UOC/ISC, I/V, INVERTER
(AC, DC) and ISO PV measurements.
A 1385 PV test lead must be used for INVERTER AC/DC measurements. Before starting a PV measurement the settings of PV module type and PV test
parameters should be checked.
Environmental parameters (Irr, T) can be measured or entered manually.
Page 11
MI 3108 EurotestPV Safety and operational considerations
11
Environmental conditions (irradiance, temperature) must be stable during the
measurements.
For calculation of STC results measured Uoc / Isc values, irradiance,
temperature (ambient or cell), and PV module parameters must be known. Refer to Appendix E for more information.
Always perform zeroing of DC current clamps before test.
Page 12
MI 3108 EurotestPV Safety and operational considerations
12
2.2 Battery and charging
The instrument uses six AA size alkaline or rechargeable Ni-MH battery cells. Nominal operating time is declared for cells with nominal capacity of 2100 mAh. Battery condition is always displayed in the lower right display part. In case the battery is too weak the instrument indicates this as shown in figure 2.1. This indication appears for a few seconds and then the instrument turns itself off.
Figure 2.1: Discharged battery indication
The battery is charged whenever the power supply adapter is connected to the instrument. The power supply socket polarity is shown in figure 2.2. Internal circuit controls charging and assures maximum battery lifetime.
+
-
Figure 2.2: Power supply socket polarity
Symbols:
Indication of battery charging
Figure 2.3: Charging indication
Warnings related to safety:
When connected to an installation, the instruments battery compartment can
contain hazardous voltage inside! When replacing battery cells or before opening the battery/fuse compartment cover, disconnect any measuring accessory connected to the instrument and turn off the instrument,
Ensure that the battery cells are inserted correctly otherwise the instrument will
not operate and the batteries could be discharged.
Do not recharge alkaline battery cells! Use only power supply adapter delivered from the manufacturer or distributor of
the test equipment !
Notes:
The charger in the instrument is a pack cell charger. This means that the battery
cells are connected in series during the charging. The battery cells have to be equivalent (same charge condition, same type and age).
Page 13
MI 3108 EurotestPV Safety and operational considerations
13
If the instrument is not to be used for a long period of time, remove all batteries
from the battery compartment.
Alkaline or rechargeable Ni-MH batteries (size AA) can be used. Metrel
recommends only using rechargeable batteries with a capacity of 2100mAh or above.
Unpredictable chemical processes can occur during the charging of battery cells
that have been left unused for a longer period (more than 6 months). In this case Metrel recommends to repeat the charge / discharge cycle at least 2-4 times.
If no improvement is achieved after several charge / discharge cycles, then each
battery cell should be checked (by comparing battery voltages, testing them in a cell charger, etc). It is very likely that only some of the battery cells are deteriorated. One different battery cell can cause an improper behavior of the entire battery pack!
The effects described above should not be confused with the normal decrease of
battery capacity over time. Battery also loses some capacity when it is repeatedly charged / discharged. This information is provided in the technical specification from battery manufactur er .
Page 14
MI 3108 EurotestPV Safety and operational considerations
14
2.3 Standards applied
The Eurotest instruments are manufactured and tested in accordance with the following regulations:
Electromagnetic compatibility (EMC)
EN 61326
Electrical equipment for measurement, control and laboratory
use – EMC requirements
Class B (Hand-held equipment used in controlled EM environments)
Safety (LVD)
EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements
EN 61010-2-030
Safety requirements for electrical equipment for measurement, control
and laboratory use – Part 2-030:
Particular requirements for testing
and measuring circuits
EN 61010-031
Safety requirements for electrical equipment for measurement, control
and laboratory use – Part 031: Safety requirements for hand-
held
probe assemblies for electr i c al measur ement and test
EN 61010-2-032
Safety requirements for electrical equipment for measurement,
control, and laboratory use - Part 2-
032: Particular requirements for
hand-held and hand-
manipulated current sensors for electrical test
and measurement
Functionality
EN 61557
Electrical safety in low voltage distribution systems up to 1000 VAC
and 1500 VAC –
Equipment for testing, measuring or monitoring of
protective measures
Part 1 General requirements
Part 2 Insulation resistance Part 3 Loop resistance Part 4 Resistance of earth connecti on and eq ui pot ential bonding Part 5 Resistance to earth Part 6 Residual current devices (RCDs) in TT and TN systems Part 7. Phase sequence Part 10 Combined measuring equipment
Part 12 Performance measuring and monitoring devices (PMD)
Reference standards for electrical installations and components
EN 61008
Residual current operated circuit-breakers without integral overcurrent protection for household and similar uses
EN 61009
Residual current operated circuit-breakers with integral overcurrent protection for household and similar uses
EN 60364-4-41
Electrical installations of buildings Part 4-41 Protection for safety – protection against electric shock
BS 7671
IEE Wiring Regulations (17th edition)
AS/NZS 3017
Electrical installations – Verification guidelines
Reference standard for photovoltaic systems
EN 62446
Grid connected photovoltaic systems – Minimum requirements for system documentation, commissioning tests and inspection
Page 15
MI 3108 EurotestPV Safety and operational considerations
15
Note about EN and IEC standards:
Text of this manual contains references to European standards. All standards of
EN 6XXXX (e. g. EN 61010) series are equivalent to IEC standards with the same number (e.g. IEC 61010) and differ only in amended parts required by European harmonization procedure.
Page 16
MI 3108 EurotestPV Instrument description
16
3 Instrument description
3.1 Front panel
Figure 3.1: Front panel
Legend:
LCD
128 x 64 dots matrix display with backlight.
UP
Modifies selected parameter.
DOWN
TEST
TEST
Starts measurements.
Acts also as the PE touching electrode.
ESC
Goes one level back.
TAB
Selects the parameters in selected function.
Backlight, Contrast
Changes backlight level and contrast.
8 ON / OFF
Switches the instrument power on or off.
The instrument automatically turns off 15 minutes after the last key was pressed
9 HELP / CAL
Accesses help menus.
Calibrates test leads in Continuity functions.
Starts Z
REF
measurement in Voltage drop sub-function.
10
Function selector
- RIGHT
Selects test function.
11
Function selector
- LEFT
12
MEM
Stores / recalls memory of instrument. Stores clamp and solar settings.
13
Green LEDs Red LEDs
Indicates PASS / FAIL of result.
Page 17
MI 3108 EurotestPV Instrument description
17
3.2 Connector panel
Figure 3.2: Connector panel (pic tur e of MI 31 08)
Legend:
Test connector
Measuring inputs / outputs
Charger socket
3 USB connector
Communication with PC USB (1.1) port.
Protection cover
5 C1
Current clamp measuring input #1
6 P/C2
Current clamp measuring input #2 Measuring input for exter nal pro bes
7 PS/2 connector
Communication with PC serial port
Connection to optional measuring adapters Connection to barcode / RFID reader
Connection to Bluetooth dongle
Warnings!
Maximum allowed voltage between any test terminal and ground is 600 V
a.c., 1000 V d.c.!
Maximum allowed voltage between test terminals on test connector is 600 V
a.c., 1000 V d.c.!
Maximum allowed voltage between test terminals P/C2, C1 is 3 V! Maximum short-term voltage of external power supply adapter is 14 V!
Page 18
MI 3108 EurotestPV Instrument description
18
3.3 Back side
1
2
3
Figure 3.3: Back panel
Legend:
Battery / fuse compartment cover
Back panel information label
Fixing screws for battery / fuse compartment cover
Figure 3.4: Battery and fuse compartment
Legend:
Fuse F1
FF 315 mA / 1000 V d.c. (Breaking capacity: 50 kA)
High breaking current fuse block
3 Serial number label
4 Battery cells
Size AA, alkaline / rechargeable NiMH
Battery holder
Can be removed from the instrument
Page 19
MI 3108 EurotestPV: Instrum ent description
19
3
1
2
R: 0.0 1999 Test current: max. 8.5mA Open-circuit voltage:6.5V
 
Continu ity 7mA



R: 0.18M 199.9M , U =50V ,100 , R: 0.12M 999M , U = 500V , 1kV U: 0V 1200V Nominal voltages: 100V , 250V , 500V , 1kV Measuring current: min. 1mA at R =U 1k /V Short-circuit current: < 3mA
 

N
N
N N


Insulation resistance (EN 61557-2)
V 250V

Continu ity
Trippi ng tim e
non-delayed (time-delayed) RCDs
   
1: 0ms 300ms (500ms) 2: 0ms 150ms (200ms) 5: 0ms 40ms (150ms), U : 0.0V 100.0V
C
Trippi ng cur re nt
I : 0.2 I 1.1 I A C ( t : 0ms 300ms, U : 0.0V 100.0V


 
C
�� �
N N N

1.5 I A)
Multiplier: 0.5, 1, 2, 5

Contact voltage
U : 0.0V 100.0V R : 0.00 10.0 0k (R =U / I )
C S SCN
    
R : 0. 1999 I : 0. A . kA
L-N(L) PSC
17 20 1 4
Nominal voltage: 100V 440V/ 15Hz 500Hz


Line impedance(EN 61557-3)
 
Fault loop (EN 61557-3)
R : 0. 1999 I : 0. A . kA

PFC
14 14
Nominal voltage: 100V 264V/ 15Hz 500Hz

impedance
L-PE
17

Voltage, frequency
U: 0V 440V / f: 15Hz 500Hz

Phase rotation (EN 61557-7)
Nominal voltage: 100V 440V / 1 00 Results: 1.2.3 or 2.1.3

5Hz 5 Hz
RCD(EN 61557-6)
I : 10mA, 30mA, 100mA, 300mA, 500mA, 1A Nominal voltage: 100V 264V/ 15Hz 500Hz
 
R Low (EN 61557-4)
 

R: 0.12 1999 Test current: min. ±200mA at 2 Open-circuit voltage: 6.5V 9.0V
 
Resistance to earth (EN 61557-5)
R: 0.04 9999 Open-circuit voltage: Short-circuit current

< 45V
: < 20m A
RMS
Ljubljanska 77 SI - 1354 Horjul Tel: +386 1 75 58 200 http://www.metrel.si
20 224 832
CAT III 600V
550V
Figure 3.5: Bottom
Legend:
Bottom information label
Neck belt openings
Handling side covers
3.4 Carrying the instrument
With the neck-carrying belt supplied in standard set, various possibilities of carrying the instrument are available. Operator can choose appropriate one on basis of his operation, see the following examples:
The instrument hangs around operators neck only - quick placing and displacing.
Page 20
MI 3108 EurotestPV: Instrum ent description
20
The instrument can be used even when placed in soft carrying bag – test cable connected to the instrument through the front aperture.
3.4.1 Secure attachment of the strap
You can choose between two methods:
Figure 3.6: First method
Figure 3.7: Alternativ e method
Please perform a periodical check o f the att achment.
Page 21
MI 3108 EurotestPV Instrument description
21
3.5 Instrument set and accessories
3.5.1 Standard set MI 3108
Instrument Soft carrying bag, 2 pcs PV Safety Probe Pyranometer PV Temperature probe AC/ DC current clamp Schuko-plug test cable Test lead, 3 x 1.5 m Test probe, 4 pcs Crocodile clip, 4 pcs Set of carrying straps PV MC 4 adapter male PV MC 4 adapter female PV MC 3 adapter male PV MC 3 adapter female RS232-PS/2 cable USB cable Set of NiMH battery cells Power supply adapter CD with instruction manual, and “Guide for testing and verification of low voltage
installations” handb ook.
Short instruction manual Calibration Certificate
3.5.2 Optional accessories
See the attached sheet for a list of optional accessories that are available on request from your distributor.
Page 22
MI 3108 EurotestPV Instrument operation
22
4 Instrument operation
4.1 Display and sound
4.1.1 Terminal voltage monitor
The terminal voltage monitor displays on-line the voltages on the test terminals and information about active test terminals in the a.c. installation measuring mode.
Online voltages are displayed together with test terminal indication. All three test terminals are used for selected measurement.
Online voltages are displayed together with test terminal indication. L and N test terminals are used for selected measurement.
L and PE are active test terminals; N terminal should also be connected
for correct input voltage condition.
4.1.2 Battery indication
The battery indication indicates the charge condition of battery and connection of external charger.
Battery capacity indication.
Low battery. Battery is too weak to guarantee correct result. Replace or recharge the battery cells.
Charging in progress (if power supply adapter is connected).
4.1.3 Messages
In the message field warnings and messages are displayed.
Measurement is running, consider displayed warnings.
Conditions on the input terminals allow starting the measurement; consider other displayed warnings and messages.
Conditions on the input terminals do not allow starting the measurement, consider displayed warnings and messages.
RCD tripped-out during the measurement (in RCD functions).
Instrument is overheated. The measurement is prohibited until the temperature decreases under the allowed limit.
Result(s) can be stored.
High electrical noise was detected during measurement. Results may be impaired.
Page 23
MI 3108 EurotestPV Instrument operation
23
L and N are changed.
Warning! High voltage is applied to the test terminals.
Warning! Dangerous voltage on the PE terminal! Stop the activity
immediately and eliminate the fault / connection problem before proceeding with any activity!
Test leads resistance in Continuity measurement is not compensated. Test leads resistance in Continuity measurement is compensated.
High resistance to earth of test probes. Results may be impaired.
Too small current for declared accuracy. Results may be impaired. Check in Current Clamp Settings if sensitivity of current clamp can be increased.
Measured signal is out of range (clipped). Results are impaired. Fuse F1 is broken.
External DC voltage is detected. Measurements in this operating mode are blocked.
4.1.4 Results
Measurement result is inside pre-set limits (PASS). Measurement result is out of pre-set limits (FAIL). Measurement is aborted. Consider displayed warnings and messages.
4.1.5 Sound warnings
Continuous sound
Warning! Dangerous voltage on the PE terminal is detected.
4.1.6 Help screens
HELP
Opens help screen.
Help menus are available in all functions. The Help menu contains schematic diagrams for illustrating how to properly connect the instrument to electric installation or PV system. After selecting the measurement you want to perform, press the HELP key in order to view the associated Help menu.
Keys in help menu:
UP / DOWN
Selects next / previous help screen.
ESC / HELP / Function selector
Exits help menu.
Page 24
MI 3108 EurotestPV Instrument operation
24
Figure 4.1: Examples of help screens
4.1.7 Backlight and contrast adjustments
With the BACKLIGHT key backlight and contrast can be adjusted.
Click
Toggles backlight intensity level.
Keep pressed for 1 s
Locks high intensity backlight level until power is turned off or the key is pressed again.
Keep pressed for 2 s
Bargraph for LCD contrast adjustment is displayed.
Figure 4.2: Contrast adjustment menu
Keys for contrast adjustment:
DOWN
Reduces contrast.
UP
Increases contrast.
TEST
Accepts new contrast.
ESC
Exits without changes.
4.2 Function selection
For selecting test / measurement function within each test mode the FUNCTION SELECTOR keys shall be used.
Keys:
Function selector
Selects test / measurement function.
UP/DOWN
Selects sub-function in selected measurement funct ion. Selects screen to be viewed (if results are split into more screens).
TAB
Selects the test parameter to be set or modified.
TEST
Runs selected test / measure me nt function.
MEM
Stores measured results / recalls stored results.
ESC
Exits back to main menu.
Page 25
MI 3108 EurotestPV Instrument operation
25
Keys in test parameter field:
UP/DOWN
Changes the selected parameter.
TAB
Selects the next measuring parameter.
Function selector
Toggles between the main functions.
MEM
Stores measured results / recalls stored results
General rule regarding enabling parameters for evaluation of measurement / test result:
Parameter
OFF
No limit values, indication: _ _ _.
ON
Value(s) – results will be marked as PASS or FAIL in
accordance with selected limit.
See Chapter 5 for more information about the operation of the instrument test functions.
Page 26
MI 3108 EurotestPV Instrument operation
26
4.3 Instruments main menu
In instrument’s main menu the test mode can be selected. Different instrument options can be set in the SETTINGS menu.
<INSTALLATION> a.c. LV installation testing <POWER> Power & Energy testing <SOLAR> PV systems testing <SETTINGS> Instrument settings
Figure 4.3: Main menu
Keys:
UP / DOWN
Selects appropriate opt ion.
TEST
Enters selected option.
4.4 Settings
Different instrument options can be set in the SETTINGS menu.
Options are:
Recalling and clearing stored results Selection of language Setting the date and time Selection of reference stand ar d f or
RCD tests
Entering Isc factor Commander support Setting the instrument to initial values Settings for Bluetooth com mu ni cat ion Settings for current clamps Menu for synchronization with PV
Remote unit
Settings for PV measureme nts
Figure 4.4: Options in Settings menu
Keys:
UP / DOWN
Selects appropriate opt ion.
TEST
Enters selected option.
ESC / Function selector
Exits back to main menu.
Page 27
MI 3108 EurotestPV Instrument operation
27
4.4.1 Memory
In this menu the stored data can be recalled or deleted. See chapter 8 Data han dl ing for more information.
Figure 4.5: Memory options
Keys:
UP / DOWN
Selects option.
TEST
Enters selected option.
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
4.4.2 Language
In this menu the language can be set.
Figure 4.6: Language selecti on
Keys:
UP / DOWN
Selects language.
TEST
Confirms selected language and exits to settings menu.
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
4.4.3 Date and time
In this menu date and time can be set.
Figure 4.7: Setting date and time
Keys:
TAB
Selects the field to be changed.
UP / DOWN
Modifies selected field.
TEST
Confirms new date / time and exits .
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
Warning:
If the batteries are removed for more than 1 minute the set date and time will be
lost.
Page 28
MI 3108 EurotestPV Instrument operation
28
4.4.4 RCD standard
In this menu the used standard for RCD
tests can be set.
Figure 4.8: Selection of RCD test
standard
Keys:
UP / DOWN
Selects standard.
TEST
Confirms selected standard.
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
Maximum RCD disconnection times differ in various standards. The trip-out times defined in individual standards are listed below.
Trip-out times according to EN 61008 / EN 61009:
½×I
N
*)
I
N
2×I
N
5×I
N
General RCDs (non-delayed)
t∆ > 300 ms t∆ < 300 ms t∆ < 150 ms t∆ < 40 ms
Selective RCDs (time-delayed)
t∆ > 500 ms 130 ms < t∆ < 500 ms 60 ms < t∆ < 200 ms 50 ms < t∆ < 150 ms
Trip-out times according to EN 60364-4-41:
½×I
N
*)
I
N
2×I
N
5×I
N
General RCDs (non-delayed)
t∆ > 999 ms t∆ < 999 ms t∆ < 150 ms t∆ < 40 ms
Selective RCDs (time-delayed)
t∆ > 999 ms 130 ms < t∆ < 999 ms 60 ms < t∆ < 200 ms 50 ms < t∆ < 150 ms
Trip-out times according to BS 7671:
½×I
N
*)
I
N
2×I
N
5×I
N
General RCDs (non-delayed)
t∆ > 1999 ms
t∆ < 300 ms t∆ < 150 ms t∆ < 40 ms
Selective RCDs (time-delayed)
t∆ > 1999 ms 130 ms < t∆ < 500 ms 60 ms < t∆ < 200 ms 50 ms < t∆ < 150 ms
Trip-out times according to AS/NZS 3017
**)
:
½×I
N
*)
I
N
2×I
N
5×I
N
RCD type
I∆N [mA]
t
t
t
t
Note
I
10
> 999 ms
40 ms
40 ms
40 ms
Maximum brea k time
II
> 10 ≤ 30
300 ms
150 ms
40 ms
III
> 30
300 ms
150 ms
40 ms
IV
S
> 30 > 999 ms
500 ms
200 ms
150 ms
130 ms
60 ms
50 ms
Minimum non-actuating time
*)
Minimum test period for current of ½×I∆N, RCD shall not trip-out.
**)
Test current and measurement accuracy correspond to AS/NZS 3017 requirements.
Page 29
MI 3108 EurotestPV Instrument operation
29
Maximum test times related to selected test current for general (non-delayed) RCD
Standard
½×I
N
I
N
2×I
N
5×I
N
EN 61008 / EN 61009
300 ms
300 ms
150 ms
40 ms
EN 60364-4-41
1000 ms
1000 ms
150 ms
40 ms
BS 7671
2000 ms
300 ms
150 ms
40 ms
AS/NZS 3017 (I, II, III)
1000 ms
1000 ms
150 ms
40 ms
Maximum test times related to selected test current for selective (time-delayed) RCD
Standard
½×I
N
I
N
2×I
N
5×I
N
EN 61008 / EN 61009
500 ms
500 ms
200 ms
150 ms
EN 60364-4-41
1000 ms
1000 ms
200 ms
150 ms
BS 7671
2000 ms
500 ms
200 ms
150 ms
AS/NZS 3017 (IV)
1000 ms
1000 ms
200 ms
150 ms
4.4.5 Isc factor
In this menu the Isc factor for calculation of
short circuit current in Z-LINE and Z-LOOP measurements can be set.
Figure 4.9: Selection of Isc factor
Keys:
UP / DOWN
Sets Isc value.
TEST
Confirms Isc value.
ESC
Exits back to settings menu.
Function selectors
Exits back to main menu without changes.
Short circuit current Isc in the supply system is important for selection or verification of protective circuit breakers (fuses, over-current breaking devices, RCDs). The default value of Isc factor (ksc) is 1.00. The value should be set according to local regulative. Range for adjustment of the Isc factor is 0.20 ÷ 3.00.
4.4.6 Commander support
T
he support for remote commanders can be
enabled or disabled in this menu.
Figure 4.10: Selection of commander
support
Page 30
MI 3108 EurotestPV Instrument operation
30
Keys:
UP / DOWN
Selects commander option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
Note:
Commander disabled option is intended to disable the commander’s remote
keys. In the case of high EM interfering noise the operation of the commander can be irregular.
4.4.7 Communication
In this menu the Bluetooth dongle A 1436 can be initialized and device for scanning
barcodes can be selected.
Figure 4.11: Menu for Bluetooth
communication
Keys:
UP / DOWN
Selects option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
Initialization of the Bluetooth dongle
The Bluetooth dongle A 1436 should be initialized when it is used with the instrument for the first time. During initialization the instrument sets the dongle parameters and name in order to communicate properly with PC and other devices via Bluetooth.
Initialization procedure
1. Connect Bluetoo th d ongle A 1436 to the instrument.
2. Press RESET key on the Bluetooth dongle A 1436 for at least 5 seconds.
3. Select INIT. BT DONGLE in Communication menu and press TEST.
4. Wait for confirmation message and beep. Following message is displayed if dongle was initialized properly:
EXTERNAL BT ‘DONGLE SEARCHING OK!
Notes:
The Bluetooth dongle A 1436 should always be initialized before first use with the
instrument.
If the dongle was initialized by another Metrel instrument it will probably not work
properly when working with the previous instrument again. Bluetooth dongle initialization should be repeated in that case.
Page 31
MI 3108 EurotestPV Instrument operation
31
For more information about communication via Bluetooth refer to chapter 8.6
Communications and A 1436 ma nual.
Selection of barcode scanner type
In this menu the barcode scanner type can be set. Options are:
Serial barcode scanner Barcode scanner application on Android devices (check list of approved Metrel
Android devices and applications)
Keys:
UP / DOWN
Selects option.
TEST
Confirms selected option.
ESC
Exits back to Settings menu.
Function selector
Exits back to main menu without changes.
Figure 4.12: Menu for Scanner type
4.4.8 Initial settings
In this menu the instrument settings, measurement parameters and limits can be set to initial (factory) values.
Figure 4.13: Initial settings
dialogue
Keys:
UP / DOWN
Selects option [YES, NO].
TEST
Restores default settings ( if YES is selected).
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
Warning:
Customized settings will be lost when this option is used! If the batteries are removed for more than 1 minute the custom made settings will
be lost.
Page 32
MI 3108 EurotestPV Instrument operation
32
The default setup is listed below:
Instrument setting
Default value
Language
English
Contrast
As defined and stored by adjustment procedure
Isc factor
1.00
RCD standards
EN 61008 / EN 61009
Commander
Enabled
Scanner type
RS232
Clamp settings
CLAMP 1 CLAMP 2
A1391, 40A A1391, 40A
Solar settings
See chapter 4.4.10 Solar Settings
Test mode: Function
Parameters / limit value
Sub-function
INSTALLATION:
EARTH RE
No limit
R ISO
No limit Utest = 500 V
Low Ohm Resistance
R LOW
No limit
CONTINUITY*
No limit
Z - LINE
Fuse type: none selected
VOLTAGE DROP
ΔU: 4.0 %
Z
REF
: 0.00 Ω
Z - LOOP
Fuse type: none selected
Zs rcd
Fuse type: none select ed
RCD
RCD t
Nominal differential current: I∆N=30 mA
RCD type: AC
non-delayed Test current starting polarity: (0°) Limit contact voltage: 50 V
Current multiplier: ×1
POWER:
CURRENT
C1
HARMONICS
U
I
U
h:1
ENERGY
I: 40A, U: 260A
SOLAR:
ISO PV
No limit Utest = 500 V
ENV.
Measured
I/V
Measured
INVERTER
AC/ DC
Page 33
MI 3108 EurotestPV Instrument operation
33
Note:
Initial settings (reset of the instrument) can be recalled also if the TAB key is
pressed while the instrument is switched on.
4.4.9 Clamp Settings
In Clamp settings menu the C1 and C2/P measuring inputs can be configured.
Figure 4.14: Configuration of current clamp
measuring inputs
Parameters to be set:
Model
Model of current clamp [A1018, A1019, A1391].
Range
Measuring range of current clamp [20 A, 200 A], [40 A, 300 A].
Selection of measuring parameters Keys
UP / DOWN
Selects appropriate option.
TEST
Enables changing data of selected parameter.
MEM
Saves settings.
ESC
Exits back to clamp settings menu.
Function selector
Exits back to main menu without changes.
Changing data of selected parameter Keys
UP / DOWN
Sets parameter.
TEST
Confirms set data.
ESC
Disable changing data of selected parameter.
Function selector
Exits back to main menu without changes.
Note:
Measuring range of the instrument must be considered. Measurement range of
current clamp can be higher than of the instrument.
4.4.10 Synchronization (A 1378 - PV Remote unit)
The main purpose of the synchronization is to get correct values of temperature and irradiance for calculation of STC measurement results. During the PV tests the displayed STC results are calculated on base of set or measured environmental data in the instrument’s Environmental menu. These values are not necessarily measured at the same time as other measurements.
Page 34
MI 3108 EurotestPV Instrument operation
34
Synchronization (of time stamps) enables to later update the PV measured results with enviromental data that were measured simultaneously with the A 1378 PV Remote unit. Stored STC values are then corrected accordingly.
Selecting this option will allow
synchronization of data between the instrument and PV Remote unit.
Figure 4.15: Synchronize menu
Data to be synchronized:
TIME
Instrument’s time and date will be uploaded to the PV Remote unit.
RESULT
Values of measured environmental parameters will be downloaded to the instrument. Saved STC results will be corrected accordingly.
Keys:
UP / DOWN
Selects data to be synchronized.
TEST
Synchronizes data. Follow the information on the LCD. If the
synchronization succeeded a confirmation beep will follow after
short connecting... and synchronizing... messages.
ESC
Exits back to settings menu.
Function selector
Exits back to main menu.
Connection for synchronization
Figure 4.16: Connection of the instruments during synchronization
Note:
Refer to A 1378 PV Remote unit user manual for more information.
Page 35
MI 3108 EurotestPV Instrument operation
35
4.4.11 Solar settings
In Solar settings parameters of PV modules and settings for PV measurements can be
set.
Figure 4.17: Solar settings
Keys:
UP / DOWN
Selects option.
TEST
Enters menu for changing parameter s .
ESC
Exits back to settings menu.
Function selector
Exits back to main menu without changes.
PV module settings
Parameters of PV modules can be set in this menu. A database for up to 20 PV modules can be created / edited. Parameters are used for calculation of STC values.
Note:
The database can be also created on the PC or mobile device and then sent to
the instrument. PCSW EurolinkPRO and some Android applications support this feature.
Figure 4.18: PV module settings menu
Parameters of PV module:
Module
PV module name
Pmax
1 W ..2000 W
Nominal power of PV module
Umpp
1.0 V .. 999 V
Voltage on maximum power point
Impp
0.01 A .. 15.0 A
Current on maximum power point
Uoc
1.0 V .. 999 V
Open circuit voltage of module
Isc 0.01 A .. 15.0 A
Short circuit current of modul e
NOCT
1.0 °C .. 99.0 °C
Nominal working temperature of PV cell
alfa
-5.00 mA/°C .. 300 mA/°C
Temperature coefficient of Isc
beta
-5.00 V/°C .. -0.001 V/°C
Temperature coefficient of Uoc
gamma
-5.00 %/°C .. 0.999 %/°C
Temperature coefficient of Pmax
Rs
0.01 Ω .. 9.99 Ω
Serial resistance of PV module
Page 36
MI 3108 EurotestPV Instrument operation
36
Selection of PV module type and parameters
Keys:
UP / DOWN
Selects appropriate opt ion.
TEST
Enters menu for changing type or parameters.
ESC, Function selector
Exits back.
MEM
Enters PV module type memory menu.
Changing a PV module type / parameter
Keys:
UP / DOWN
Sets value / data of parameter / PV module type.
TEST
Confirms set value / data.
ESC, Function selector
Exits back.
PV module type memory menu
ADD
Enters menu for adding a new PV module type.
OVERWRITE
Enters menu for storing changed data of selected PV module type.
DELETE
Deletes selected PV module type.
DELETE ALL
Deletes all PV module types.
Keys:
UP / DOWN
Selects option.
TEST
Enters selected menu.
Function selectors
Exits back to main function men u.
If Add or Overwrite is selected the menu for setting the PV module type name is displayed.
Figure 4.19: Setting name of PV module type
Keys:
/
Selects a character.
TEST
Selects the next character.
MEM
Confirms new name and stores it in the memory. Then returns to Module settings menu.
ESC
Deletes last letter. Returns to previous menu without changes.
If Delete or Delete all is selected a warning will be displayed.
Page 37
MI 3108 EurotestPV Instrument operation
37
Figure 4.20: Delete options
Keys:
TEST
Confirms clearing. In Delete all option YES must be selected.
ESC / Function selector
Exits back to main function menu without changes.
PV measurements settings
Parameters for PV measurements can be
set in this menu.
Figure 4.21: Selection of PV
measurement settings
Parameters for PV measurements:
Test std
Testing standard [IEC 60891, CEI 82-25]
Irr. Sens.
Type of irradiance measuring sensor [PV cell, Pyran.]
Irr. min.
Minimal valid solar irradiance for calculation [500 – 1000 W/m2]
T. sensor
Temperature for calcul at ion [Tamb, Tcell]
Mod.Ser.
Number of modules in serial [1 – 30]
Mod.Par.
Number of modules in parallel [1 – 10]
Selection of PV test parameters Keys:
UP / DOWN
Selects appropriate opt ion.
TEST
Enables changing data of selected parameter.
MEM
Saves settings.
ESC / Function selector
Exits back.
Changing data of selected parameter
Keys:
UP / DOWN
Sets parameter.
TEST
Confirms set data.
ESC / Function selector
Exits back.
Page 38
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
38
5 Measurements – a.c. LV installations
5.1 Voltage, frequency and phase sequence
Voltage and frequency measurement is always active in the terminal voltage monitor. In the special VOLTAGE TRMS menu the measured voltage, frequency and information about detected three-phase connection can be stored. Measurements are based on the EN 61557-7 standard.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.1: Voltage in single
phase system
Test parameters for voltage measurement There are no parameters to be set.
Connections for voltage measurement
Figure 5.2: Connection of 3-wire test lead and optional adapter in three-phase system
Page 39
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
39
Figure 5.3: Connection of plug commander and 3-wire test lead in single-phase system
Voltage measurement procedure
Select the VOLTAGE TRMS function using the function selector keys.
Connect test cable to the instrument. Connect test leads to the item to be tested (see figures 5.2 and 5.3).
Store voltage measurement result by pressing the MEM key (optional).
Measurement runs immediately after selection of VOLTAGE TRMS function.
Figure 5.4: Examples of voltage measurement in three-phase system
Displayed results for single phase system:
Uln ........... Voltage between phase and neutral conductors,
Ulpe ......... Voltage between phase and protective conductors,
Unpe ........ Voltage between neutral and protective conductors,
f ............... frequency.
Displayed results for three-phase system:
U12 .......... Voltage between phases L1 and L2,
U13 .......... Voltage between phases L1 and L3,
U23 .......... Voltage between phases L2 and L3,
1.2.3 ........ Correct connection – CW rotation seq uence,
3.2.1 ........ Invalid connection – CCW rotation sequence,
f ............... frequency.
Page 40
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
40
5.2 Insulation resistance
The Insulation resistance measurement is performed in order to ensure safety against electric shock through ins ulat ion. Typical applications ar e:
Insulation resistance between conductors of installation, Insulation resistance of non-conductive rooms (walls and floors), Insulation resistance of ground cables, Resistance of semi-conducti ve (antistatic) floors .
See chapter 4.2 Function selection for instructions on
key functionality.
Figure 5.5: Insulation resistance
Test parameters for insulation resistance measurement
Uiso
Test voltage [50 V, 100 V, 250 V, 500 V, 1000 V]
Limit
Minimum insulation resistance [OFF, 0.01 M ÷ 200 M ]
Test circuits for insulation resistance
Figure 5.6: Connections for insulation measurement
Page 41
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
41
Insulation resistance measuring procedure
Select the R ISO function using the function selector keys.
Set the required test voltage. Enable and set limit value (optional). Disconnect
tested installation from mains supply (and discharge insulation as
required).
Connect test cable to the instrument and to the item to be tested (see figure 5.6). Press the TEST
key to perform the measurement (double click for continuous
measurement and later press to stop the measurement).
After the measurement is finished wait until tested item is fully discharged.
Store the result by pressing the MEM key (optional).
Figure 5.7: Example of insulation resistance measurement resul t
Displayed results:
R.........................Insulation resistance
Um......................Test v oltag e – actual value.
Page 42
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
42
5.3 Resistance of earth connection and equipotential bonding
The resistance measurement is performed in order to ensure that the protective measures against electric shock through earth connections and bondings are effective. Two sub-functions are av ai l abl e:
R LOWΩ - Earth bond measurement according to EN 61557-4 (200 mA), CONTINUITY - Continuous resistance measurement performed with 7 mA.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.8: 200 mA RLOW Ω
Test parameters for resistance measurement
TEST
Resistance measurem ent sub-function [R LOWΩ, CONTINUITY]
Limit
Maximum resistance [OFF, 0.1 ÷ 20.0 ]
Additional test parameter for In Continuity sub-function
Buzzer On (sound if resistance is lower than the set limit value) or Off
5.3.1 R LOWΩ, 200 mA resistance measurement
The resistance measurement is performed with automatic polarity reversal of the test voltage.
Test circuit for R LOWΩ measurement
Figure 5.9: Connection of 3-wire test lead plus optional extension lead
Page 43
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
43
R LOWΩ measurement procedure
Select continuity function using the function selector keys.
Set sub-function to R LOWΩ. Enable and set limit (optional). Connect test cable to the instrument. Compensate the test leads resistance (if necessary, see secti on 5.3 . 3). Disconnect from mains supply and discharge installation to be tested. Connect the test leads to the appropriate PE wiring (see figure 5.9). Press the TEST key to perform the measurement. After the measurement is finished store
the result by pressing the MEM button
(optional)*.
Figure 5.10: Example of RLOW result
Displayed result:
R................R LOWΩ resistance.
R+..............Result at positive polarity
R-...............Result at negative test polarity
5.3.2 Continuous resistance measurement with low current
In general, this function serves as standard Ω-meter with a low testing current. The measurement is performed continuously without polarity reversal. The function can also be applied for testing continuity of inductive components.
Test circuit for continuous resistance measurement
Page 44
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
44
Figure 5.11: Tip commander and 3-wire test lead applications
Continuous resistance measurement procedure
Select continuity function using the function selector keys.
Set sub-function CONTINUITY. Enable and set the limit (optional). Connect test cable to the instrument. Compensate test leads resistance (if necessary, see section 5.3.3). Disconnect from mains supply and discharge the object to be tested. Connect test leads to the tested object (see figure 5.11). Press the TEST key to begin performing a continuous measurement. Press the TEST key to stop measurement.
After the measurement is finished, store the result (optional).
Figure 5.12: Example of continuous resistance measurement
Displayed result:
R............Resistance
5.3.3 Compensation of test leads resistance
This chapter describes how to compensate the test leads resistance in both continuity functions, R LOWΩ and CONTINUITY. Compensation is required to eliminate the influence of test leads resistance and the internal resistances of the instrument on the measured resistance. The lead compensation is therefore a very important feature to obtain correct result.
symbol is displayed if the compensation was carried out successfully.
Page 45
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
45
Circuits for compensating the resistance of test leads
Figure 5.13: Shorted test leads
Compensation of test leads resistance procedure
Select R LOWΩ or CONTINUITY function.
Connect test cable to the instrument and short the test leads together (see figure
5.13).
Press TEST to perform resistance measurement.
Press the CAL key to compensate leads resistance.
Figure 5.14: Results with old calibration
values
Figure 5.15: Results with new calibration
values
Note:
The highest value for lead compensation is 5 . If the resistance is higher the
compensation value is set back to default value.
is displayed if no calibration value is stored.
Page 46
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
46
5.4 Testing RCDs
Various test and measurements are required for verification of RCD(s) in RCD protected installations. Measurements are based on the EN 61557-6 standard. The following measurements and tests (sub-functions) can be performed:
Contact voltage, Trip-out time, Trip-out current, RCD autotest.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.16: RCD test
Test parameters for RCD test and measurement
TEST
RCD sub-function test [RCDt, RCD I, AUTO, Uc].
I∆N
Rated RCD residual current sensitivity IN [10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 1000 mA].
type
RCD type AC, A, F, B, B+, starting polarity [ , , , ,
,
],
selective
S
or general characteristic.
MUL
Multiplication factor for test current [½, 1, 2, 5 IN].
Ulim
Conventional touch voltage limit [25 V, 50 V].
Notes:
Ulim can be selected in the Uc sub-function only. Selective (time delayed) RCDs have delayed response characteristics. As the
contact voltage pre-test or other RCD tests influence the time delayed RCD it takes a certain period to recover into normal state. Therefore a time delay of 30 s is inserted before performing trip-out test by default.
Connections for testing RCD
Figure 5.17: Connecting the plug commander and the 3-wire test lead
Page 47
MI 3108 EurotestPV Measurements – a.c. LV installations
47
5.4.1 Contact voltage (RCD Uc)
A current flowing into the PE terminal causes a voltage drop on earth resistance, i.e. voltage difference between PE equipotential bonding circuit and earth. This voltage difference is called contact voltage and is present on all accessible conductive parts connected to the PE. It shall always be lower than the conventional safety limit voltage. The contact voltage is measured with a test current lower than ½ I∆N to avoid trip-out of the RCD and then normalized to the rated I∆N.
Contact voltage measurement procedure
Select the RCD function using the function selector keys.
Set sub-function Uc. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the item to be tested (see figure 5.17). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
The contact voltage result relates to the rated nominal residual current of the RCD and is multiplied by an appropriate factor (depending on RCD type and type of test current). The 1.05 factor is applied to avoid negative tolerance of result. See table 5.1 for detailed contact voltage calculation factors.
RCD type
Contact voltage Uc
proportional to
Rated I∆N
AC
1.05×I∆N
any
AC
S
2×1.05×I∆N
A, F
1.4×1.05×I∆N
30 mA
A, F
S
2×1.4×1.05×I∆N
A, F
2×1.05×I∆N
< 30 mA
A, F
S
2×2×1.05×I∆N
B, B+
2×1.05×I∆N
any
B, B+
S
2×2×1.05×I∆N
Table 5.1: Relationship between Uc and I
N
Loop resistance is indicative and calculated from Uc result (without additional proportional factors) according to:
N
C
L
I
U
R
=
.
Figure 5.18: Example of contact voltage measurement results
Page 48
MI 3108 EurotestPV Measurements – a.c. LV installations
48
Displayed results:
Uc ........ Contact voltage.
Rl ......... Fault loop resistance.
5.4.2 Trip-out time (RCDt)
Trip-out time measurement verifies the sensitivity of the RCD at different residual currents.
Trip-out time measurement procedure
Select the RCD function using the function selector keys.
Set sub-function RCDt. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the item to be tested (see figure 5.17). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 5.19: Example of trip-out time measurement results
Displayed results:
t ........... Trip-out time,
Uc ........ Contact voltage for rated I∆N.
5.4.3 Trip-out current (RCD I)
A continuously rising residual current is intended for testing the threshold sensitivity for RCD trip-out. The instrument increases the test current in small steps through appropriate range as follows:
RCD type
Slope range
Waveform
Start value
End value
AC
0.2×I∆N
1.1×I∆N
Sine
A, F (I∆N 30 mA)
0.2×I∆N
1.5×I∆N
Pulsed
A, F (I∆N = 10 mA)
0.2×I∆N
2.2×I∆N
B, B+
0.2×I∆N
2.2×I∆N
DC
Maximum test current is I∆ (trip-out current) or end value in case the RCD didn’t trip-out.
Page 49
MI 3108 EurotestPV Measurements – a.c. LV installations
49
Trip-out current measurement procedure
Select the RCD function using the function selector keys.
Set sub-function RCD I. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the item to be tested (see figure 5.17). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 5.20: Trip-out current measurement result example
Page 50
MI 3108 EurotestPV Measurements – a.c. LV installations
50
Displayed results: I…….Trip-out current, Uci…Contact voltage at trip-out current I or end value in case the RCD didn’t trip, t…….Trip-out time.
5.4.4 RCD Autotest
RCD autotest function is intended to perform a complete RCD test (trip-out time at different residual currents, trip-out current and contact voltage) in one set of automatic tests, guided by the instrument.
Additional key:
HELP / DISPLAY
Toggles between top and bottom part of results field.
RCD autotest procedure
RCD Autotest steps
Notes
Select the RCD function using the function selector keys.
Set sub-function AUTO.
Set test parameters (if necessary).
Connect test cable to the instrument.
Connect test leads to the item to be tested (see figure
5.17).
Press the TEST key to perform the test.
Start of test
Test with I
N
, 0° (step 1).
RCD should trip-out
Re-activate RCD.
Test with I
N
, 180° (step 2).
RCD should trip-out
Re-activate RCD.
Test with 5×I
N
, 0° (step 3).
RCD should trip-out
Re-activate RCD.
Test with 5×I
N
, 180° (step 4).
RCD should trip-out
Re-activate RCD.
Test with ½×I∆N, 0° (step 5).
RCD should not trip­out
Test with ½×I∆N, 180° (step 6).
RCD should not trip­out
Trip-out current test, 0° (step 7).
RCD should trip-out
Re-activate RCD.
Trip-out current test, 180° (step 8).
RCD should trip-out
Re-activate RCD.
Store the result by pressing the MEM key (optional).
End of test
Page 51
MI 3108 EurotestPV Measurements – a.c. LV installations
51
Result examples:
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Figure 5.21: Individual steps in RCD autotest
Top
Bottom
Figure 5.22: Two parts of result field in RCD autotest
Page 52
MI 3108 EurotestPV Measurements – a.c. LV installations
52
Displayed results:
x1 ........ Step 1 trip-out time ( , IN, 0º),
x1 ........ Step 2 trip-out time ( , IN, 180º),
x5 ........ Step 3 trip-out time ( , 5×IN, 0º),
x5 ........ Step 4 trip-out time ( , 5×IN, 180º),
x½ ....... Step 5 trip-out time ( , ½×IN, 0º),
x½ ....... Step 6 trip-out time ( , ½×IN, 180º),
I ......... Step 7 trip-out current (0º),
I ......... Step 8 trip-out current (180º),
Uc ........ Contact voltage for rated IN.
Notes:
The autotest sequence is immediately stopped if any incorrect condition is
detected, e.g. excessive Uc or trip-out time out of bounds.
Auto test is finished without x5 tests in case of testing the RCD types A, F with
rated residual currents of In = 300 mA, 500 mA, and 1000 mA. In this case auto test result passes if all other results pass, and indications for x5 are omitted.
Tests for sensitivity (I ,, steps 7 and 8) are omitted for selective type RCD.
Page 53
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
53
5.5 Fault loop impedance and prospective fault current
Fault loop is a loop comprised by mains source, line wiring and PE return path to the mains source. The instrument measures the impedance of the loop and calculates the short circuit current. The measurement is covered by requirements of the EN 61557-3 standard.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.23: Fault loop impedance
Test parameters for fault loop impedance measurement
Test
Selection of fault loop impedance sub-function [Zloop, Zs rcd]
Fuse type
Selection of fuse type [---, NV, gG, B, C, K, D]
Fuse I
Rated current of selec ted f use
Fuse T
Maximum breaking time of selected fuse
Lim
Minimum short circuit current for selected fuse.
See Appendix A for reference fuse data.
Circuits for measurement of fault loop impedance
Figure 5.24: Connection of plug commander and 3-wire test lead
Page 54
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
54
Fault loop impedance measurement procedure
Select the Zloop or Zs rcd sub-function using the function selector keys and
/ keys.
Select test parameters (optional). Connect test cable to the instrument. Connect test leads to the item to be tested (see figures 5.17 and 5.24). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 5.25: Example of loop impedance measurement result
Displayed results:
Z .............. Fault loop impedance,
ISC ............ Prospective fault current,
Lim .......... Low limit prospective short-circuit current value.
Prospective fault current ISC is calculated from measur ed i mp eda nce as follows:
Z
kUn
I
SC
SC
×
=
where:
Un ........ Nominal U
L-PE
voltage (see table below),
ksc ....... Correction factor for Isc (see chapter 4.4.5).
Un
Input voltage range (L-PE)
110 V
(93 V ≤ U
L-PE
134 V)
230 V
(185 V ≤ U
L-PE
266 V)
Notes:
High fluctuations of mains voltage can influence the measurement results (the
noise sign is displayed in the message field). In this case it is recommended to repeat few measurements to check if the readings are stable.
This measurement will trip-out the RCD in RCD-protected electrical installation if test Zloop is selected.
Select Zs rcd measurement to prevent trip-out of RCD in RCD protected
installation.
Page 55
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
55
5.6 Line impedance and prospective short-circuit current / Voltage drop
Line impedance is measured in loop comprising of mains voltage source and line wiring. Line impedance is covered by the requirements of the EN 61557-3 standard. The Voltage drop sub-function is intended to check that a voltage in the installation stays above acceptable levels if the highest current is flowing in the circuit. The highest current is defined as the nominal current of the circuit's fuse. The limit values are described in the standard EN 60364-5-52.
Sub-functions:
Z LINE - Line impedance measurement according to EN 61557-3, ΔU – Voltage drop measure ment .
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.26: Line impedance
Figure 5.27: Voltage drop
Test parameters for line impedance measurement
Test
Selection of line impedance [Zline] or voltage drop [ΔU] sub-function
FUSE type
Selection of fuse type [---, NV, gG, B, C, K, D]
FUSE I
Rated current of selec ted f use
FUSE T
Maximum breaking time of selected fuse
Lim
Minimum short circuit current for selected fuse.
See Appendix A for reference fuse data. Additional test parameters for voltage drop measurement
ΔU
MAX
Maximum voltage drop [3.0 % ÷ 9.0 %].
Page 56
MI 3108 EurotestPV Measurements – a.c. LV installations
56
5.6.1 Line impedance and prospect ive short circuit current
Circuits for measurement of line impedance
Figure 5.28: Phase-neutral or phase-phase line impedance measurement – connection
of plug commander and 3-wire test lead
Line impedance measurement procedure
Select the Zline sub-function.
Select test parameters (optional). Connect test cable to the instrument. Connect test leads to the item to be tested (see figure 5.28). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 5.29: Examples of line impedance measurement result
Displayed results:
Z .............. Line impedance,
ISC ............ Prospective short-circuit current,
Lim .......... Low limit prospective short-circuit current value.
Prospective short circuit current is calculated as follows:
Z
kUn
I
SC
SC
×
=
where:
Un ........ Nominal L-N or L1-L2 voltage (see table below),
ksc ....... Correction factor for Isc (see chapter 4.5.5).
Page 57
MI 3108 EurotestPV Measurements – a.c. LV installations
57
Un
Input voltage range (L-N or L1-L2)
110 V
(93 V ≤ U
L-N
< 134 V)
230 V
(185 V ≤ U
L-N
266 V)
400 V
(321 V < U
L-L
485 V)
Note:
High fluctuations of mains voltage can influence the measurement results (the
noise sign is displayed in the message field). In this case it is recommended to repeat few measurements to check if the readings are stable.
5.6.2 Voltage drop
The voltage drop is calculated based on the difference of line impedance at connection points (sockets) and the line impedance at the reference point (usually the impedance at the switchboard).
Circuits for measurement of voltage drop
Figure 5.30: Phase-neutral or phase-phase voltage drop measurement – connection of
plug commander and 3-wire test lead
Voltage drop measurement procedure Step 1: Measuring the impedance Zref at origin
Select the ΔU sub-function using the function selector keys and  / keys.
Select test parameters (optional). Connect test cable to the instrument. Connect the test leads to the origin of electrical installation (see figure 5.30).
Press the CAL key to perform the measurement.
Step 2: Measuring the voltage drop
Select the ΔU sub-function using the function selector keys and  / keys.
Select test parameters (Fuse type must be selected). Connect test cable or plug commander to the instrument. Connect the test leads to the tested points (see figure 5.30). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Page 58
MI 3108 EurotestPV Measurements – a.c. LV installations
58
Step 1 - Zref
Step 2 - Voltage drop
Figure 5.31: Examples of voltage drop measurement result
Displayed results:
ΔU ........... Voltage drop,
ISC ............ Prospective short-circuit current,
Z .............. Line impedance at measured point,
Zref .......... Reference impedance
Voltage drop is calculated as follows:
[
]
100
)
(
%
=
N
N
REF
U
I
Z
Z
U
where:
ΔU ........ calculated voltage drop
Z………impedance at test point Z
REF
…...impedance at refer ence point IN………rated current of selected fuse UN…….nominal voltage (see table below)
Un
Input voltage range (L-N or L1-L2)
110 V
(93 V ≤ U
L-PE
< 134 V)
230 V
(185 V ≤ U
L-PE
266 V)
400 V
(321 V < U
L-N
485 V)
Notes:
If the reference impedance is not set the value of Z
REF
is considered as 0.00 Ω.
The Z
REF
is cleared (set to 0.00 Ω) if pressing CAL key while instrument is not
connected to a voltage source.
I
SC
is calculated as described in chapter 5.6.1 Line impedance and prospective
short circuit current.
If the measured voltage is outside the ranges described in the table above the
ΔU result will not be calculated.
High fluctuations of mains voltage can influence the measurement results (the
noise sign is displayed in the message field). In this case it is recommended to repeat few measurements to check if the readings are stable.
Page 59
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
59
5.7 Earth resistance
Earth resistance is one of the most important parameters for protection against electric shock. Main earthing arrangements, lightning systems, local earthings, etc can be verified with the earthing resistance test. The measurement conforms to the EN 61557­5 standard.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.32: Earth resistance
Test parameters for earth resistance measurement
Limit
Maximum resistance OFF, 1 ÷ 5 k
Connections for earth resistance measurement
Figure 5.33: Resistance to earth, measurement of main installation earthing
Page 60
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
60
Figure 5.34: Resistance to earth, measurement of a lighting protection system
Earth resistance measurements, common measurement procedure
Select EARTH function using the function selector keys.
Enable and set limit value (optional). Connect test leads to the instrument. Connect the item to be tested (see figures 5.33, 5.34). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 5.35: Example of earth resistance measurement result
Displayed results for ear th res is ta nc e me asurement:
R.............. Earth resistance,
Rp............ Resistance of S (potential) probe,
Rc ............ Resistance of H (current) probe.
Notes:
High resistance of S and H probes could influence the measurement results. In
this case, “Rp” and “Rc” warnings are displayed. There is no pass / fail indication in this case.
High noise currents and voltages in earth could influence the measurement
results. The tester displays the warning in this case.
Probes must be placed at sufficient distance from the measured object.
Page 61
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
61
5.8 PE test terminal
It can happen that a dangerous voltage is applied to the PE wire or other accessible metal parts. This is a very dangerous situation since the PE wire and MPEs are considered to be earthed. An often reason for this fault is incorrect wiring (see examples below). When touching the TEST key in all functions that requires mains supply the user automatically performs this test.
Examples for application of PE test terminal
Figure 5.36: Reversed L and PE conductors (plug commander)
L1 N
Reversed phase and protection conductors!
MOST DANGEROUS SITUATION!
PE
LPEN
L/L1
N/L2
PE/L3
Figure 5.37: Reversed L and PE conductors (application of 3-wire test lead)
Page 62
MI 3108 EurotestPV Measurements – a.c. LV installatio ns
62
PE terminal test procedure
Connect test cable to the instrument.
Connect test leads to the item to be tested (see figures 5.36 and 5.37). Touch PE test probe (the TEST key) for at least one second. If PE terminal is connected to phase volta
ge the warning message is displayed, instrument buzzer is activated, and further measurements are disabled in Zloop and RCD functions.
Warning:
If dangerous voltage is detected on the tested PE terminal, immediately stop all
measurements, find and remove the fault!
Notes:
PE test terminal is active in the INSTALLATION operating mode (except in the
VOLTAGE, Low ohm, Earth and Ins ul ati on func ti ons) .
PE test terminal does not operate in case the operator’s body is completely
insulated from floor or w al ls !
For operation of PE test terminal on commanders refer to Appendix D
Commanders.
Page 63
MI 3108 EurotestPV Solar measurements – PV systems
63
6 Solar measurements - PV systems
The following measurements for verification and troubleshooting of PV installations can be performed with the instrument:
Insulation resistance on PV systems PV inverter test PV panel test Environmental parameters Open voltage and short circuit test I-V curve test
6.1 Insulation resistance on PV systems
The Insulation resistance measurement is performed in order to ensure safety against electric shock through insulation between live parts on PV installations and earth. The measurement is carried out according to test method 1 in IEC / EN 62446 (test between panel / string / array negative and earth followed by a test between panel / string / array positive and earth).
See chapter 4.2 Function selection for instructions on
key functionality. The input voltage is displayed in order to check proper connection before carrying out the test.
Figure 6.1: Insulation resistance
Test parameters for insulation resistance measurement on PV systems
TEST
RISO - , RISO +
Uiso
Test voltage [50 V, 100 V, 250 V, 500 V, 1000 V]
Limit
Minimum insulation resistance [OFF, 0.01 M ÷ 200 M ]
Page 64
MI 3108 EurotestPV Solar measurements – PV systems
64
Test circuits for insulation resistance on PV systems
Figure 6.2: Connections for insulation resistance measurement on PV systems
Insulation resistance measuring procedure
Select the RISO - sub-function using the function selector keys and / keys.
Set the required test voltage. Enable and set limit value (optional). Connect PV safety probe to the instrument (see figure 6.2) Connect accessories to the PV system (see figure 6.2). Press the TEST
key to perform the measurement (double click for continuous
measurement and later press to stop the measurement).
After the measurement is finished wait until tested item is fully discharged. Store the result by pressing the MEM key (optional).
Select the RISO + sub-function using the  /  keys.  Reconnect DC+ lead (see figure 6.2). Press the TEST
key to perform the measurement (double click for continuous
measurement and later press to stop the measurement).
After the measurement is finished wait until tested item is fully discharged.
Store the result by pressing the MEM key (optional).
Figure 6.3: Examples of insulation resistance measurement result
Displayed results:
R.........................Insulation resistance
Um......................Test v oltag e – actual value.
Page 65
MI 3108 EurotestPV Solar measurements – PV systems
65
6.2 PV inverter test
The test is intended to check proper operation of the PV inverter. Following functions are supported:
Measuring of DC values at inverter’s input and AC values at inverter’s output. Calculation of the efficiency of the inverter.
See chapter 4.2 Function selection for instructions on key functionality. The input voltages are displayed in order to check proper connection before carrying out the test.
Figure 6.4: Examples of PV inverter test starting screens
Settings and parameters for PV inverter test
Input
Inputs/ Outputs being measured [ AC, DC, AC_DC]
Connections for PV inverter test
Figure 6.5: PV inverter test - DC side
Page 66
MI 3108 EurotestPV Solar measurements – PV systems
66
Figure 6.6: PV inverter test - AC side
Figure 6.7: PV inverter test - AC and DC sides
PV inverter test procedure
Select INVERTER sub-function using the function selector keys and  / keys.
Connect
PV safety probe and current clamp to the instrument (see figures 6.5
and 6.6) or
Connect
PV test lead A 1385 and current clamps to the instrument (see figure
6.7)
Connect accessories to the PV system (see figures 6.5 to 6.7). Check input voltages. Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 6.8: Examples of PV inverter test results screens
Page 67
MI 3108 EurotestPV Solar measurements – PV systems
67
Displayed results for PV inverter test:
DC column:
U.............. measured voltage at the input of the inverter
I ............... measured current at the input of the inverter
P .............. measured power at the input of the inverter
AC column:
U.............. measured voltage at the output of the inverter
I ............... measured current at the output of the inverter
P .............. measured power at the output of the inverter
η…......... .. calculated efficiency of the inverter
Notes:
With one current clamp the complete test can be performed in two steps. Input
should be set to DC and AC separately.
For the INVERTER AC/DC test fused test lead A 1385 must be used!
6.3 PV panel test
PV panel test is intended to check proper operation of PV panels. Following functions are supported:
measuring of output voltage, current and power of PV panel, comparison of measured PV output values (MEAS values) and calculated
nominal data (STC values)
comparison of measured PV output power (Pmeas) and theoretical output power
(Ptheo)
The PV panel test results are divided into three screens. See chapter 4.2 Function selection for instructions on key functionality. The input voltage is displayed in order to check proper connection before carrying out the test.
Figure 6.9: PV module test starting screens
Page 68
MI 3108 EurotestPV Solar measurements – PV systems
68
Connections for PV panel
Figure 6.10: PV panel test
PV panel test procedure
Select PANEL sub-function using the function selector keys.
Connect PV safety probe, current clamp(s) and sensors to the instrument. Connect the PV system to be tested (see figure 6.10). Check input voltage. Press the TEST key to perform the test.
Store the result by pressing the MEM key (optional).
Figure 6.11: Examples of PV measurement results
Displayed results are:
MEAS column
U.............. measured output voltage of the panel
I ............... measured output current of the panel
P .............. measured output power of the panel
STC column
U.............. calculated output voltage of the panel at STC
I ............... calculated output current of the panel at STC
P .............. calculated output power of the panel at STC
Pstc… ...... measured output power of the panel at STC
Pmax…... nominal output power of the panel at STC
η1….. ....... efficiency of the panel at STC
Page 69
MI 3108 EurotestPV Solar measurements – PV systems
69
Pmeas… . measured output power of the panel at momentary conditions Ptheo…. .. calculated theoretical output power of the panel at momentary
conditions
η2…...... ... calculated efficiency of the panel at momentary conditions
Notes:
Before starting the PV measurements settings of PV module type and PV test
parameters should be checked.
For calculation of STC results PV module type, PV test parameters, Uoc, Isc, Irr
and T (ambient or cell) values must be measured or be entered manually before the test. The results in ENV. and Uoc/Isc menus are considered. If there are no results in Uoc/Isc menu the instrument will consider results in I-V menu.
The Uoc, Isc, Irr and T measurements should be carried out immediately before
the PANEL test. Environmental conditions must be stable during the tests.
For best results PV remote unit A 1378 should be used.
6.4 Me a s uring of environmental parameters
Temperature and solar irradiance values must be known for:
Calculation of nominal values at standard test conditions (STC), Checking that environmental conditions are suitable for carrying out the PV tests.
The parameters can be measured or entered manually. The probes can be connected to the instrument or to the PV remote unit A 1378.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 6.12: Environmental
parameters screen
Test parameters for measuring / setting of environmental parameters
INPUT
Input of environmental data [ MEAS, MANUAL]
Page 70
MI 3108 EurotestPV Solar measurements – PV systems
70
Connections for measuring of environmental parameters
Figure 6.13: Measurement of environm ental par ameters
Procedure for measuring of environmental parameters
Select ENV. function and MEAS sub-function using the function selector keys
and  / keys.
Connect environmental probes to the instrument (see figure 6.13). Connect the item to be tested (see figure 6.13). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 6.14: Example of measured results
Displayed results for environmental paramete rs :
Irr………………...solar irradiance Tamb or Tcell…. temperature of ambient or PV cells
Note:
If the Irradiance result is lower than the set minimal value Irr min the STC results
will not be calculated (message is displayed).
Procedure for manual entering of environmental parameters
If the data is measured with other measuring equipment they can be entered manually. Select ENV. function and MANUAL sub-function using the function selector keys and Up/Down keys.
Page 71
MI 3108 EurotestPV Solar measurements – PV systems
71
Keys:
TEST
Enters menu for manual setting of environmental parameters.
Enters menu for changing selected par a me t er.
Confirms set value of parameter.
/
Selects environmental parameter. Selects value of parameter .
Function selector
Exits environmental menu and select PV measurement.
Figure 6.15: Example of manually entered results
Displayed results for environmental paramete rs :
Irr………………...solar irradiance Tamb or Tcell…. temperature of ambient or PV cells
Note:
Environmental parameters are cleared when the SOLAR test mode is exited.
6.4.1 Operation with A1378 PV Remote Unit
See PV Remote Unit User Manual.
6.5 Uoc / Isc test
The Uoc / Isc test is intended to check if protection devices in the d.c. part of the PV installation are effective. The measured data can be calculated to nominal data (STC values).
See chapter 4.2 Function selection for instructions on key functionality.
Figure 6.16: Uoc / Isc test
The input voltage is displayed in order to check proper connection before carrying out the test.
Page 72
MI 3108 EurotestPV Solar measurements – PV systems
72
Connection for Uoc / Isc test
Figure 6.17: Uoc / Isc test
Uoc / Isc test procedure
Select Uoc / Isc sub-function using the function selector keys and  / keys. Connect PV safety probe and sensors (optional) to the instrument. Connect the item to be tested (see figure 6.17). Check input voltage. Press the TEST key to perform the measurement. Store the result by pressing the MEM key (optional).
Figure 6.18: Example of Uoc / Isc measurement results
Displayed results for Uoc / Isc measurement: MEAS column
Uoc .......... measured open voltage of the panel
Isc............ measured short circuit current of the panel
STC column
Uoc .......... calculated open voltage at STC
Isc............ calculated short circuit current at STC
Page 73
MI 3108 EurotestPV Solar measurements – PV systems
73
Notes:
Before starting the PV measurements settings of PV module type and PV test
parameters should be checked.
For calculation of STC results correct PV module type, PV test parameters, Irr
and T (ambient or cell) values must be measured or be entered manually before the test. The Irr and T results in ENV. menu are considered. Refer to Appendix E for further information.
The Irr and T measurements should be carried out immediately before the Uoc /
Isc test. Environmental conditions must be stable during the tests.
For best results PV remote unit A 1378 should be used.
6.6 I / V curve measurement
The I / V curve measurement is used to check correct operation of the PV panels. Different problems on PV panels (failure of a part of the PV panel / string, dirt, shadow etc.) can be found.
Figure 6.19: I / V curve starting screens
The data to be measured is divided into three screens. See chapter 4.2 Function selection for instructions on key functionality.
Settings parameters for I / V curve test
1/3
Number of screen.
STC
Results (STC, measured, bot h) to be displ ayed.
Connection for the I / V curve test
Figure 6.20: I / V curve test
Page 74
MI 3108 EurotestPV Solar measurements – PV systems
74
I / V curve test procedure
Select I / V sub-function using the function selector keys and / keys.
Check or set PV module and PV testing parameters and limits (optional). Connect PV safety probe to the instrument. Connect environmental probes to the instrument (optional). Connect the item to be tested (see figure 6.20). Press the TEST key to perform the measurement.
Store the result by pressing the MEM key (optional).
Figure 6.21: Example of I / V curve results
Displayed results for I / V curve test:
Uoc .......... measured / STC open circuit voltage of the panel
Isc............ measured / STC short circuit current of the panel
Umpp ....... measured / STC voltage at maximal power point
Impp ........ measured / STC current at maximal power point
Pmpp ....... measured / STC maximal output power of the panel
Notes:
Before starting the PV measurements settings of PV module type and PV test
parameters should be checked.
For calculation of STC results correct PV module type, PV test parameters, Irr
and T (ambient or cell) values must be measured or be entered manually before the test. The Irr. and T results in ENV. menu are considered. Refer to Appendix E for further information.
The Irr. and T measurements should be carried out immediately before the I / V
curve test. Environmental conditions must be stable during the tests.
For best results PV remote unit A 1378 should be used.
Page 75
MI 3108 EurotestPV Measurements - Power & energy
75
7 Measurements - Power & Energy
1- phase power measurements and tests (sub-functions) can be performed with the EurotestPV instrument. Main features are:
Measurement of standard power parameters, Harmonic analysis of voltage and current, Displaying of voltage and current waveforms, Energy counting.
7.1 Power
The Power function is intended to measure the standard power parameters P, Q, S, THDU and PF.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 7.1: Power menu
Settings and parameters for Power test
There are no parameters to be set in this menu.
Connection for Power test
Figure 7.2: Power measurement
Page 76
MI 3108 EurotestPV Measurements - Power & energy
76
Power test procedure
Select POWER sub-function using the function selector keys and / keys.
Connect the voltage test leads and current clamp to the instrument. Connect
the voltage test leads and current clamp to the item to be tested (see
figure 7.2).
Press the TEST key to start the continuous measurement. Press the TEST key again to stop the measurement.
Store the result by pressing the MEM key (optional).
Figure 7.3: Power measurement results
Displayed results for the Power measurements:
P .............. active power
S .............. apparent power
Q ............. reactive power (capacitive or inductive)
PF…… .... power factor ( capacitive or inductive)
THDU……voltage total harmonic distortion
Notes:
Consider polarity and setup of current clamps (see chapter 4.4.8 Clamp settings). Results can also be stored while the measurement is running.
7.2 Harmonics
Harmonics are components of the voltage and currents signal with an integer multiple of the fundamental frequency. The harmonic values are an important parameter of power quality.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 7.4: Harmonics menu
Settings and parameters in Harmonics function
Input
Displayed parameters [ voltage U or current I]
h:0..h:11
Selected harmonic
Page 77
MI 3108 EurotestPV Measurements - Power & energy
77
Connection for the Harmonics measurement
(See figure 7.2)
Harmonics measurement procedure
Select HARMONICS sub-function using the function selector keys and  /
keys.
Connect voltage test leads and current clamp to the instrument. Connect
the voltage test leads and current clamp to the item to be tested (see
figure 7.2).
Press the TEST key to start the continuous measurement. Press the TEST key again to stop the measurement.
Store the result by pressing the MEM key (optional)
Figure 7.5: Examples of Harmonics measurement results
Displayed results for the Harmonics measurements:
Uh............ TRMS voltage of selected harmonic
Ih ............. TRMS current of selected harmonic
THDU……voltage total harmonic distortion THDI…… . voltage total harmonic distortion
Notes:
Parameters (input and number of harmonic) can be changed and results can also
be stored while the measurement is running.
Displayed graph is auto-ranged.
7.3 Scope
The Scope function is intended to check the shape of voltage and current.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 7.6: Scope menu
Settings and parameters in Scope function
Input
Displayed parameters [ voltage U or current I or both]
Page 78
MI 3108 EurotestPV Measurements - Power & energy
78
Connection for the Scope measurement
(See figure 7.2)
Scope measurement pro cedu re
Select SCOPE sub-function using the function selector keys and Up/Down keys.
Connect voltage test leads and current clamp to the instrument. Connect the voltage test leads and current clamp to the item to be tested (see
figure 7.2).
Press the TEST key to start the continuous measurement. Press the TEST key again to stop the measurement.
Store the result by pressing the MEM key (optional)
Figure 7.7: Example of Scope measurement results
TRMS values of voltage and current are displayed.
Notes:
The parameter input can be changed and results can also be stored while the
measurement is running.
Displayed waveforms are auto-ranged.
7.4 Current
This function is intended for measurement of load and leakage currents with current clamps. Two independent measuring inputs are available.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 7.8: Current menu
Settings and parameters for current measurement
Input
Selected channel [C1, C2, both]
Page 79
MI 3108 EurotestPV Measurements - Power & energy
79
Connection for current measurement
Figure 7.9: Leakage and load current measurements
Current measurement procedure
Select CURRENT function using the function selector keys.
Select input channel (optional). Connect current clamp(s) to the instrument. Connect the clamp(s) to the item to be tested (see figure 7.9). Press the TEST key to start the continuous measurement. Press the TEST key again to stop the measurement.
Store the result by pressing the MEM key (optional).
Figure 7.10: Examples of current measurement result
Displayed results for Cur r ent meas ur e men t: I, I1, I2 …..…..Current
Note:
Channel C2 is intended for me asur i ng with clamps A 1391 only.
7.5 Energy
In this function consumed and generated energy can be measured.
See chapter 4.2 Function selection for instructions on key functionality.
Figure 7.11: Energy menu
Page 80
MI 3108 EurotestPV Measurements - Power & energy
80
Settings and parameters for the Energy measurement
I
MAX
Maximal expected TRMS current during measurement [ I
range
, I
range
/10,
I
range
/100]
U
MAX
Maximal expected TRMS voltage during measurement [ 260 V, 500 V]
Connection for the Energy measurements
(See figure 7.2)
Energy measurement procedure
Select ENERGY sub-function using the function selector keys and  / keys.
Connect the voltage test leads and current clamp to the instrument. Connect
the voltage test leads and current clamp to the item to be tested (see
figure 7.2).
Press the TEST key to start the measurement. Press the TEST key again to stop the measurement.
Store the result by pressing the MEM key (optional).
Figure 7.12: Example of Energy measurement results
Displayed results for the Energy measurements:
E+ ............ consumed energy (load)
E-............. generated energy (source)
P .............. momentary active power during energy measurement
t…………. time
Notes:
Consider polarity and setup of current clamps (see chapter 4.4.8 Clamp settings). I
MAX
and U
MAX
should be set high enough in order to avoid clamping of measured
signals. Clamping will results in wrong energy result.
If measured currents and voltages are lower than 20% of set I
MAX
, U
MAX
the
accuracy will be impaired.
Page 81
MI 3108 EurotestPV Data handling
81
8 Data handling
8.1 Memory organizatio n
Measurement results together with all relevant parameters can be stored in the instrument’s memory. After the measurement is completed, results can be stored to the flash memory of the instrument, together with the sub-res ul ts and function parameters.
8.2 Data structure
The instrument’s memory place is divided into 4 levels each containing 199 locations. The number of measurements that can be stored into one location is not limited.
The data structure field describes the location of the measurement (which object, block, fuse, connection) and where can be accessed. In the measurement field there is information about type and number of measurements that belong to the selected structure element (object and block and fuse and connection). The main advantages of this syste m are :
Test results can be organized and grouped in a structured manner that reflects
the structure of typical electrical installations.
Customized names of data structure elements can be uploaded from
EurolinkPRO PCSW.
Simple browsing through structure and results. Test reports can be created with no or little modifications after downloading
results to a PC.
Figure 8.1: Data structure and measurement fields
Page 82
MI 3108 EurotestPV Data handling
82
Data structure field
Memory operation menu
Data structure field
1st level:
OBJECT: Default
location name (object and its
successive number).
004: No. of selected element.
2nd level:
BLOCK
: Default location name (block and its
successive number).
001: No. of selected element.
3rd level:
FUSE: Default
location name (fuse and its successive
number).
002: No. of selected element.
4th level:
CONNECTION
: Default location name (connection
and its successive number).
003: No. of selected element.
No. of measurements in selected location
[No. of measurements in selected location and its sub-
locations]
Measurement field
Type of stored measurement in the selected location.
No. of selected test result / No. of all stored test results in selected location.
Page 83
MI 3108 EurotestPV Data handling
83
8.3 Storing test results
After the completion of a test the results and parameters are ready for storing ( icon is displayed in the information field). By pressing the MEM key, the user can store the results.
Figure 8.2: Save test menu
Memory available for storing results.
Keys in save test menu - data structure fiel d:
TAB
Selects the location element (Object / Block / Fuse / Connection)
UP / DOWN
Selects number of selected location element (1 to 199)
MEM
Saves test results to the selected location and returns to the measuring menu.
Function selector / TEST
Exits back to main function men u.
Notes:
The instrument offers to store the result to the last selected location by default. If the measurement is to be stored to the same location as the previous one just
press the MEM key twice.
Page 84
MI 3108 EurotestPV Data handling
84
8.4 Recalling test results
Press the MEM key in a main function menu when there is no result available for storing or select MEMORY in the SETTINGS menu.
Figure 8.3: Recall menu - installation
structure field selected
Figure 8.4: Recall menu - measurements
field selected
Keys in recall memory menu (installation structure field selected): TAB
Selects the location element (Object / Block / Fuse / Connection).
UP / DOWN
Selects number of selected location element (1 to 199)
Function selector / ESC
Exits back to main function men u.
TEST
Enters measurements field.
Keys in recall memory menu (measurements field):
UP / DOWN
Selects the stored measurement.
TAB / ESC
Returns to installation structure fi eld.
Function selector
Exits back to main function men u.
TEST
View selected measurement results.
Figure 8.5: Example of recalled measurement result
Keys in recall memory menu (measurement results are displayed)
UP / DOWN
Displays measurement results stored in selected location
MEM / ESC
Returns to measurements field.
Function selector / TEST
Exits back to main function men u.
Page 85
MI 3108 EurotestPV Data handling
85
8.5 Clearing stored data
8.5.1 Clearing complete mem ory content
Select CLEAR ALL MEMORY in MEMORY menu. A warning will be displayed.
Figure 8.6: Clear all memory
Keys in clear all memory menu
TEST
Confirms clearing of complete memory content (YES must be selected with / keys).
Function selector
Exits back to main function menu without changes.
Figure 8.7: Clearing memory in progress
8.5.2 Clearing measurement(s) in selected location
Select DELETE RESULTS in MEMORY menu.
Figure 8.8: Clear measurements menu (data structure field selected)
Keys in delete results menu (installation structure field selected):
TAB
Selects the location element (Object / Block / Fuse / Connection).
UP / DOWN
Selects number of selected location element (1 to 199)
Function selector
Exits back to main function menu.
ESC
Exits back to memory menu.
TEST
Enters dialog box for deleting all measurements in selected location and its sub-locations.
Page 86
MI 3108 EurotestPV Data handling
86
Keys in dialog for confirmation to clear results in selected location:
TEST
Deletes all results in selected location.
MEM / ESC
Exits back to delete results menu without changes.
Function selector
Exits back to main function menu without changes.
8.5.3 Clearing individual mea s urements
Select DELETE RESULTS in MEMORY menu.
Figure 8.9: Menu for clearing individual measurement (installation structure field
selected)
Keys in delete results menu (installation structure field selected): TAB
Selects the location element (Object / Block / Fuse / Connection).
UP / DOWN
Selects number of selected location element (1 to 199)
Function selector
Exits back to main function men u.
ESC
Exits back to memory menu.
MEM
Enters measurements field for deleting individual measurements.
Keys in delete results menu (measur em ents field selected):
UP / DOWN
Selects measurement.
TEST
Opens dialog box for confirmation to clear selected measurement.
TAB / ESC
Returns to installation st r uctur e fi el d.
Function selector
Exits back to main function menu without changes.
Keys in dialog for confirmation to clear selected result(s):
TEST
Deletes selected measurement result.
MEM / TAB / ESC
Exits back to measurements field without changes.
Function selector
Exits back to main function menu without changes.
Page 87
MI 3108 EurotestPV Data handling
87
Figure 8.10: Dialog for confirmation
Figure 8.11: Display after measurement
was cleared
8.5.4 Renaming installati on s tr uc t ure elements (upload from PC)
Default installation structure elements are “Object”, “Block”, “Fuse” and “Connection”. In the PCSW package Eurolink-PRO default names can be changed with customized names that corresponds the installation under test. Refer to PCSW Eurolink-PRO HELP for information how to upload customized installation names to the instrument.
Figure 8.12: Example of menu with customized installation structure names
8.5.5 Renaming installation structure elements with serial barcode reader or RFID reader
Default installation structure elements are “Object”, “Block”, “Fuse” and “Connection”. When the instrument is in the Save results menu location ID can be scanned from a barcode label with the barcode reader or can be read from a RFID tag with the RFID reader.
Figure 8.13: Connection of the barcode reader and RFID reader
How to change the name of memory location
Connect the barcode reader or RFID reader to the instrument .
In Save menu select memory location to be renamed.
A new location name (scanned from a barcode label or a RFID tag) will be
accepted by the instrument. A successful receive of a
barcode or RFID tag is
confirmed by two short confirmation beeps.
Note:
Use only barcode readers and RFID readers delivered by Metrel or authorized
distributor.
Page 88
MI 3108 EurotestPV Data handling
88
8.6 Communication
Stored results can be transferred to a PC. A special communication program on the PC automatically identifies the instrument and enables data transfer between the instrument and the PC. There are two communication interfaces available on the instrument: USB or RS 232. With the optional Bluetooth dongle A 1436 the instrument can communicate via Bluetooth too.
8.6.1 USB and RS232 communicat ion
The instrument automatically selects the communication mode according to detected interface. USB interface has priority.
Figure 8.14: Interface connection for data transfer over PC COM port
How to establish an USB or RS232 link:
RS-232 communication: connect a PC COM port to the instrument PS/2
connector using the PS/2 - RS232 serial communication cable;
USB communication: connect a PC USB port to the instrument USB connector
using the USB interface cable.
Switch on the PC and the instrument. Run the EurolinkPRO program. The PC and the instrument will automatically recognize each other.
The instrument is prepared to communicate with the PC.
The program EurolinkPRO is a PC software running on Windows XP, Windows Vista, Windows 7, and Windows 8. Read the file README_EuroLink.txt on CD for instructions about installing and running the program.
Note:
USB drivers should be installed on PC before using the USB interface. Refer to
USB installation instructions available on installation CD.
Page 89
MI 3108 EurotestPV Data handling
89
8.6.2 Bluetooth communic a t ion
How to configure a Bluetooth link between instrument and PC
For Bluetooth communication with PC a Standard Serial Port over Bluetooth link for Bluetooth dongle A 1436 must be con fig ur ed fir st .
Switch Off and On the instrument.
Be sure that the Bluetooth dongle A 1436 is properly initialized. If not the
Bluetooth dongle must be initialized as described in chapter 4.4.7 Communication.
On PC configure a Standard Serial Port to enable communication over Bluetooth
link between instrument and PC. No code for pairing the devices is needed.
Run the EurolinkPRO program. The PC and the instrument will automatically recognize each other.
The instrument is prepared to communicate with the PC.
How to configure a Bluetooth link between instrument and Android device
Switch Off and On the instrument.
Be sure that the Bluetooth dongle A 1436 is properly initialized. If not the
Bluetooth dongle must be initialized as described in chapter 4.4.7 Communication.
Some Android applications automatically carry out the setup of a Bluetooth
connection. It is preferred to use this option if it exists. This option is supported by Metrel's Android applications.
If this option is not supported by the selected Android application then
configure
a Bluetooth link via Android device’s Bluetooth configuration tool. No code for
pairing the devices is needed.
The instrument and Android device are ready to communicate.
Notes:
Sometimes there will be a demand from the PC or Android device to enter the
code. Enter code ‘NNNN’ to correctly configure the Bluetooth link.
The name of a correctly configured Bluetooth device must consist of the
instrument type plus serial number, eg. MI 3108-12240429D. If the Bluetooth dongle got another name, the configuration must be repeated.
Page 90
MI 3108 EurotestPV Upgrading the instrument
90
9 Upgrading the instrument
The instrument can be upgraded from a PC via the RS232 communication port. This enables to keep the instrument up to date even if the standards or regulations change. The upgrade can be carried with a help of special upgrading software and the communication cable as shown on Figure 8.14. Please contact your dealer for more information.
Page 91
MI 3108 EurotestPV Maintenance
91
10 Maintenance
Unauthorized persons are not allowed to open the EurotestPV instrument. There are no user replaceable components inside the instrument, except the battery and fuse under rear cover.
10.1 Fuse replacement
There is a fuse under back cover of the EurotestPV instrument.
F1
FF 315 mA / 1000 V d.c. , 32×6 mm (Breaking capacity: 50 kA) This fuse protects internal circuitry for continuity functions if test probes are
connected to the mains supply voltage by mistake during measurement. Position of fuse can be seen in Figure 3.4 in chapter 3.3 Back panel.
Optional accessory A 1385 PV test lead has replaceable fuse in each test lead.
FF 315 mA / 1000 V d.c. , 32×6 mm (Breaking capacity: 50 kA)
Warnings:
Disconnect all measuring accessory and switch off the instrument
before opening battery / fuse compartment cover, hazardous voltage inside!
Replace blown fuse with original type only, otherwise the instrument or accessory
may be damaged and/or operator’s safety impaired!
10.2 Cleaning
No special maintenance is required for the housing. To clean the surface of the instrument or accessory use a soft cloth slightly moistened with soapy water or alcohol. Then leave the instrument or accessory to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons! Do not spill cleaning liquid over the instrument!
10.3 Periodic calibration
It is essential that the test instrument is regularly calibrated in order that the technical specification listed in this manual is guaranteed. We recommend an annual calibration. Only an authorized technical person can do the calibration. Please contact your dealer for further information.
10.4 Service
For repairs under warranty, or at any other time, please contact your distributor.
Page 92
MI 3108 EurotestPV Technical specifications
92
11 Technical specifications
11.1 Insulation resistance, Insulation resistance of PV systems
Insulation resistance (nominal voltages 50 VDC, 100 V
DC
and 250 VDC)
Measuring range according to EN 61557 is 0.15 MΩ ÷ 199.9 MΩ.
Measuring range (MΩ)
Resolution (MΩ)
Accuracy
0.00 ÷ 19.99
0.01
±(5 % of reading + 3 digits)
20.0 ÷ 99.9
0.1
±(10 % of reading)
100.0 ÷ 199.9
±(20 % of reading)
Insulation resistance (nominal voltages 500 VDC and 1000 VDC) Measuring range according to EN 61557 is 0.15 MΩ ÷ 1 GΩ.
Measuring range (MΩ)
Resolution (MΩ)
Accuracy
0.00 ÷ 19.99
0.01
±(5 % of reading + 3 digits)
20.0 ÷ 199.9
0.1
±(5 % of reading)
200 ÷ 999
±(10 % of reading)
Voltage
Measuring range (V)
Resolution (V)
Accuracy
0 ÷ 1200
±(3 % of reading + 3 digits)
Nominal voltages ............................... 50 VDC, 100 VDC, 250 VDC, 500 VDC, 1000 VDC
Open circuit voltage ........................... -0 % / +20 % of nominal voltage
Measuring current .............................. min. 1 mA at RN=UN×1 kΩ/V
Short circuit current ........................... max. 3 mA
The number of possible tests ............ > 1200, with a fully charged battery
Auto discharge after test. Specified accuracy is valid if 3-wire test lead is used while it is valid up to 100 MΩ if tip commander is used. Specified accuracy is valid up to 100 Mif relative humidity > 85 %.
In case the instrument gets moistened, the results could be impaired. In such case, it is recommended to dry the instrument and accessories for at least 24 hours. The error in operating conditions could be at most the error for reference conditions (specified in the man u al for each f unction) ±5 % of measured value.
Page 93
MI 3108 EurotestPV Technical specifications
93
11.2 Continuity
11.2.1 Resistance R LOWΩ
Measuring range according to EN 61557 is 0.16 Ω ÷ 1999 Ω.
Measuring range R (Ω)
Resolution (Ω)
Accuracy
0.00 ÷ 19.99
0.01
±(3 % of reading + 3 digits)
20.0 ÷ 199.9
0.1
±(5 % of reading)
200 ÷ 1999
Open-circuit voltage ........................... 6 .5 VDC ÷ 9 VDC
Measuring current .............................. min. 200 mA into load resistance of 2
Test lead compensation ..................... up to 5
The number of possible tests ............ > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.
11.2.2 Resistance CONTINUITY
Measuring range (Ω)
Resolution (Ω)
Accuracy
0.0 ÷ 19.9
0.1
±(5 % of reading + 3 digits)
20 ÷ 1999
Open-circuit voltage ........................... 6 .5 VDC ÷ 9 VDC
Short-circuit current ........................... max. 8.5 mA
Test le ad compensation ..................... up to 5
11.3 RCD testing
11.3.1 General data
Nominal residual current (A,AC) ........ 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
Nominal residual current accuracy ..... -0 / +0.1I; I = IN, 2×IN, 5×IN
-0.1I / +0; I = 0.5×IN
AS/NZS selected: ± 5 %
Test current shape ............................. Sine-wave (AC), pulsed (A, F), smooth DC (B, B+)
DC offset for pulsed test current ........ 6 mA (typical)
RCD type ........................................... (non-delay ed) , S (ti m e-delayed)
Test current starting polarity ............. 0 º or 180 º
Voltage range .................................... 93 V ÷ 134 V (45 Hz ÷ 65 Hz)
185 V ÷ 266 V (45 Hz ÷ 65 Hz)
Page 94
MI 3108 EurotestPV Technical specifications
94
IN × 1/2
IN × 1
IN × 2
IN × 5
RCD I
IN (mA)
AC
A,F
B, B+
AC
A,F
B,B+
AC
A,F
B,B+
AC
A,F
B,B+
AC
A,F
B,B+
10 5 3.5 5 10
20
20
20
40
40
50
100
100
30
15
10.5
15
30
42
60
60
84
120
150
212
300
100
50
35
50
100
141
200
200
282
400
500
707
1000
300
150
105
150
300
424
600
600
848
n.a.
1500
n.a.
n.a.
500
250
175
250
500
707
1000
1000
1410
n.a.
2500
n.a.
n.a.
1000
500
350
500
1000
1410
n.a.
2000
n.a.
n.a.
n.a.
n.a.
n.a.
n.a.
n.a. ..................................................... not applicabl e
AC type .............................................. sine wave test current
A, F types…… ................................... pulsed current
B, B+ types ........................................ smooth DC current
11.3.2 Contact voltage RCD-Uc
Measuring range according to EN 61557 is 20.0 V ÷ 31.0V for limit contact voltage 25V Measuring range according to EN 61557 is 20.0 V ÷ 62.0V for limit contact voltage 50V
Measuring range (V)
Resolution (V)
Accuracy
0.0 ÷ 19.9
0.1
(-0 % / +15 %) of reading ± 10 digits
20.0 ÷ 99.9
0.1
(-0 % / +15 %) of reading
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free of interfering voltages.
Test current ...................................... max. 0.5×I∆N
Limit contact voltage ......................... 25 V, 50 V
Specified accuracy is valid for complete operating range.
11.3.3 Trip-out time
Complete measurement range corresponds to EN 61557 requirements. Maximum measuring times set according to selected reference for RCD testing.
Measuring range (ms)
Resolution (ms)
Accuracy
0.0 ÷ 40.0
0.1
±1 ms
0.0 ÷ max. time *
0.1
±3 ms
* For max. time see normative references in chapter 4.4.4 RCD standard – this specification applies to max. time >40 ms.
Test current ...................................... ½×I∆N, I∆N, 2×I∆N, 5×I
N
5×I∆N is not available for I∆N=1000 mA (RCD type AC) or I
N
300 mA (RCD types A, F,
B, B+). 2×I∆N is not available for I∆N=1000 mA (RCD types A, F) or I
N
300 mA (RCD types B,
B+). 1×I∆N is not available for I∆N=1000 mA (RCD types B, B+).
Specified accuracy is valid for complete operating range.
Page 95
MI 3108 EurotestPV Technical specifications
95
11.3.4 Trip-out current
Trip-out current Complete measurement range corresponds to EN 61557 requirements.
Measuring range I∆
Resolution I∆
Accuracy
0.2×I∆N ÷ 1.1×I∆N (AC type)
0.05×I∆N
±0.1×IN
0.2×I∆N ÷ 1.5×I∆N (A type, I∆N ≥30 mA)
0.05×I∆N
±0.1×IN
0.2×I∆N ÷ 2.2×I∆N (A type, I∆N <30 mA)
0.05×I∆N
±0.1×IN
0.2×I∆N ÷ 2.2×I∆N (B type)
0.05×I∆N
±0.1×IN
Trip-out time
Measuring range (ms)
Resolution (ms)
Accuracy
0 ÷ 300
±3 ms
Contact voltage
Measuring range (V)
Resolution (V)
Accuracy
0.0 ÷ 19.9
0.1
(-0 % / +15 %) of reading ± 10 digits
20.0 ÷ 99.9
0.1
(-0 % / +15 %) of reading
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free of interfering voltages. Trip-out measurement is not available for I∆N=1000 mA (RCD types B, B+). Specified accuracy is valid for complete op erating range.
11.4 Fault loop impedance and pr ospe c tive fault current
11.4.1 No disconnecting device or FUSE selected
Fault loop impedance Measuring range according to EN 61557 is 0.25 Ω ÷ 9.99kΩ.
Measuring range (Ω)
Resolution (Ω)
Accuracy
0.00 ÷ 9.99
0.01
±(5 % of reading + 5 digits)
10.0 ÷ 99.9
0.1
100 ÷ 999
± 10 % of reading
1.00 k ÷ 9.99 k
10
Prospective fault current (calculated value)
Measuring range (A)
Resolution (A)
Accuracy
0.00 ÷ 9.99
0.01
Consider accuracy of fault
loop resistance
measurement
10.0 ÷ 99.9
0.1
100 ÷ 999
1.00 k ÷ 9.99 k
10
10.0 k ÷ 23.0 k
100
The accuracy is valid if mains voltage is stabile during the measurement.
Test current (at 230 V) ...................... 6.5 A (10 ms)
Nominal voltage range ...................... 93 V ÷ 134 V (45 Hz ÷ 65 Hz)
185 V ÷ 266 V (45 Hz ÷ 65 Hz)
Page 96
MI 3108 EurotestPV Technical specifications
96
11.4.2 RCD selected
Fault loop impedance Measuring range according to EN 61557 is 0.46 Ω ÷ 9.99 kΩ.
Measuring range (Ω)
Resolution (Ω)
Accuracy
0.00 ÷ 9.99
0.01
±(5 % of reading + 10 digits)
10.0 ÷ 99.9
0.1
100 ÷ 999
± 10 % of reading
1.00 k ÷ 9.99 k
10
Accuracy may be impaired in case of heavy noise on mains voltage. Prospective fault current (calculated value)
Measuring range (A)
Resolution (A)
Accuracy
0.00 ÷ 9.99
0.01
Consider accuracy of fault
loop resistance
measurement
10.0 ÷ 99.9
0.1
100 ÷ 999
1.00 k ÷ 9.99 k
10
10.0 k ÷ 23.0 k
100
Nominal voltage range ...................... 93 V ÷ 134 V (45 Hz ÷ 65 Hz)
185 V ÷ 266 V (45 Hz ÷ 65 Hz) No trip out of RCD.
11.5 Line impedance and prospective short-circuit current / Voltage
drop
Line impedance Measuring range according to EN 61557 is 0.25 Ω ÷ 9.99kΩ.
Measuring range (Ω)
Resolution (Ω)
Accuracy
0.00 ÷ 9.99
0.01
±(5 % of reading + 5 digits)
10.0 ÷ 99.9
0.1
100 ÷ 999
± 10 % of reading
1.00 k ÷ 9.99 k
10
Prospective short-circuit current (calculated value)
Measuring range (A)
Resolution (A)
Accuracy
0.00 ÷ 0.99
0.01
Consider accuracy of line
resistance measureme nt
1.0 ÷ 99.9
0.1
100 ÷ 999
1.00 k ÷ 99.99 k
10
100 k ÷ 199 k
1000
Test current (at 230 V) ...................... 6.5 A (10 ms)
Nominal voltage range ...................... 93 V ÷ 134 V (45 Hz ÷ 65 Hz)
185 V ÷ 266 V (45 Hz ÷ 65 Hz) 321 V ÷ 485 V (45 Hz ÷ 65 Hz)
Page 97
MI 3108 EurotestPV Technical specifications
97
Voltage drop (calculated value)
Measuring range (%)
Resolution (%)
Accuracy
0.0 ÷ 99.9
0.1
Consider accuracy of line
impedance
measurement(s)*
Z
REF
measuring range……………………… 0.00 Ω ÷ 20.0 Ω
*See chapter 5.6.2 Voltage drop for more information about calculation of voltage drop result
11.6 Resistance to earth
Measuring range according to EN61557-5 is 2.00 Ω ÷ 1999 Ω.
Measuring range (Ω)
Resolution (Ω)
Accuracy
0.00 ÷ 19.99
0.01
±(5% of reading + 5 digits)
20.0 ÷ 199.9
0.1
200 ÷ 9999
Max. auxiliary earth electrode resistance RC ... 100×RE or 50 k (whichever is lower)
Max. probe resistance RP ................................ 100×RE or 50 k (whichever is lower)
Additional probe resistance error at R
Cmax
or R
Pmax.
±(10 % of reading + 10 digits)
Additional error
at 3 V voltage noise (50 Hz) ............................ ±(5 % of reading + 10 digits)
Open circuit voltage ......................................... < 15 VAC
Short circuit current .......................................... < 30 mA
Test voltage frequency .................................... 125 Hz
Test voltage shape .......................................... sine wave
Noise voltage indication threshold ................... 1 V (< 50 , worst case)
Automatic measurement of auxiliary electrode resistance and probe resistance. Automatic measurement of voltage noise.
11.7 Voltage, frequency, and phase rotation
11.7.1 Phase rotation
Nominal system voltage range ......... 100 VAC ÷ 550 VAC
Nominal frequency range .................. 14 Hz ÷ 500 Hz
Result displayed ............................... 1.2.3 or 3.2.1
Page 98
MI 3108 EurotestPV Technical specifications
98
11.7.2 Voltage
Measuring range (V)
Resolution (V)
Accuracy
0 ÷ 550
±(2 % of reading + 2 digits)
Result type ........................................ True r.m.s. (trms)
Nominal frequency range .................. 0 Hz, 14 Hz ÷ 500 Hz
11.7.3 Frequency
Measuring range (Hz)
Resolution (Hz)
Accuracy
0.00 ÷ 9.99
0.01
±(0.2 % of reading + 1 digit)
10.0 ÷ 499.9
0.1
Nominal voltage range ...................... 10 V ÷ 550 V
11.7.4 Online terminal voltage monitor
Measuring range (V)
Resolution (V)
Accuracy
10 ÷ 550
±(2 % of reading + 2 digits)
11.8 TRMS Clamp current
Instrument
Maximum voltage on C1 and P/C2 measuring inputs…3 V
Nominal frequency ............................ 0 Hz, 40 Hz ÷ 500 Hz
AC current clamp A1018
Range = 20 A
Measuring range (A)
Resolution (A)
Accuracy*
0.0 m ÷ 99.9 m
0.1 m
±(5 % of reading + 5 digits)
100 m ÷ 999 m
1 m
±(3 % of reading + 3 digits)
1.00 ÷ 19.99
0.01
±(3 % of reading)
Range = 200 A
Measuring range (A)
Resolution (A)
Accuracy*
0.00 ÷ 0.09
0.01
indicative
0.10 ÷ 19.99
0.01
±(3 % of reading + 3 digits)
20.0 ÷ 199.9
0.1
±(3 % of reading)
AC current clamp A1019
Range = 20 A
Measuring range (A)
Resolution (A)
Accuracy*
0.0 m ÷ 99.9 m
0.1 m
indicative
100 m ÷ 999 m
1 m
±(5 % of reading )
1.00 ÷ 19.99
0.01
±(3 % of reading)
Page 99
MI 3108 EurotestPV Technical specifications
99
Range = 200 A
Measuring range (A)
Resolution (A)
Accuracy*
0.00 ÷ 0.09
0.01
indicative
0.10 ÷ 1.99
0.01
±(5 % of reading + 3 digits)
2.00 ÷ 19.99
0.01
±(3 % of reading + 3 digits)
20.0 ÷ 199.9
0.1
±(3 % of reading)
AC / DC current clamp A1391
Range = 40 A
Measuring range (A)
Resolution (A)
Accuracy*
0.00 ÷ 1.99
0.01
±(3 % of reading + 3 digits)
2.00 ÷ 19.99
0.01
±(3 % of reading)
20.0 ÷ 39.9
0.1
±(3 % of reading)
Range = 300 A
Measuring range (A)
Resolution (A)
Accuracy*
0.00 ÷ 19.99
0.01
indicative
20.0 ÷ 39.9
0.1
40.0 ÷ 299.9
0.1
±(3 % of reading + 5 digits)
* Accuracy at operating conditions for instrument and current clamp is given.
11.9 Power tests
Measurement characteristics
Function symbols
Class according to
IEC 61557-12
Measuring range
P E 2.5
5 % ÷ 100 % I
Nom
(1)
Q 2.5
5 % ÷ 100 % I
Nom
(1)
S 2.5
5 % ÷ 100 % I
Nom
(1)
PF 1 - 1 ÷ 1
f
0.05
40 Hz ÷ 60 Hz
I, I
Nom
1.5
5 % ÷ 100 % I
Nom
U
1.5
110 V ÷ 500 V
Uhn
2.5
0 % ÷ 20 % U
Nom
THDu 2.5
0 % ÷ 20 % U
Nom
Ihn
2.5
0 % ÷ 100 % I
Nom
THDi 2.5
0 % ÷ 100 % I
Nom
(1)
– I
Nom
depends on set current sensor type and selected current range:
- A 1018, A1019 (20 A or 200 A),
- A 1391 (40 A or 300 A)
Page 100
MI 3108 EurotestPV Technical specifications
100
Note:
Error of external voltage and current transducers is not considered in this
specification.
Power (P, S, Q)
Measuring range is from 0.00 W (VA, Var) to 99.9 kW (kVA, kVar)
Power factor
Measuring range is from – 1.00 to 1.00
Voltage harmonics
Measuring range is from 0.1 V to 500 V
Voltage THD
Measuring range is from 0.1 % to 99.9 %
Current harmonics and Current THD
Measuring range is from 0.00 A to 199.9 A
Energy
Measuring range is from 0.000 Wh to 1999 kWh Measurement is performed continuously without gaps.
Notes:
Error of external voltage and current transducers is not considered in this
specification.
Accuracy values for Energy are valid if I > 0.2 I
MAX
. I
MAX
is set in ENERGY
measuring menu.
11.10 PV tests
11.10.1 Accurac y of STC data
Accuracy of STC values is based on accuracy of measured electrical quantities, accuracy of environmental parameters, and entered parameters of PV module. See Appendix E: PV measurements – calculared values for more information about calculation of STC values.
11.10.2 Panel, Inverter DC Voltage
Measuring range (V)
Resolution (V)
Accuracy
0.0 ÷ 199.9
0.1
± (1.5 % of reading + 5 digits)
200 ÷ 999
±1.5 % of reading
DC Current
Measuring range (A)
Resolution (mA)
Accuracy
0.00 ÷ 19.99
10
±(1.5 % of reading + 5 digits)
20.0 ÷ 299.9
100
±1.5 % of reading
DC Power
Measuring range (W)
Resolution (W)
Accuracy
0 – 1999
± (2.5 % of reading + 6 digits)
Loading...