METREL MI 3100 SE EurotestEASI, MI 3100 S EurotestEASI Instruction Manual

Page 1
EurotestEASI
MI 3100 SE
MI 3100 s
Instruction manual
Version 1.3.2., Code no. 20 752 131
Page 2
Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) concerning safety and electromagnetic compatibility regulations
Manufacturer: METREL d.d.
Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si
© 2019 METREL
The trade names Metrel, Smartec, Eurotest, Autosequence are trademarks registered or pending in Europe and other countries.
No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
2
Page 3
MI 3100 s, MI 3100 SE EurotestEASI Table of contents
Table of contents
1 Preface ............................................................................................................................... 6
2 Safety and operational considerations ........................................................................... 7
2.1 Warnings and notes .................................................................................................... 7
2.2 Battery and charging ................................................................................................. 11
2.3 Standards applied ..................................................................................................... 13
3 Instrument description ................................................................................................... 14
3.1 Front panel ................................................................................................ ................ 14
3.2 Connector panel ........................................................................................................ 15
3.3 Back side .................................................................................................................. 16
3.4 Carrying the instrument ............................................................................................. 17
3.5 Instrument set and accessories ................................................................................. 18
3.5.1 Standard set MI 3100 s – EurotestEASI ................................................................ 18
3.5.2 Standard set MI 3100 SE – EurotestEASI ............................................................. 18
3.5.3 Optional accessories ............................................................................................. 18
4 Instrument operation ...................................................................................................... 19
4.1 Display and sound ..................................................................................................... 19
4.1.1 Terminal voltage monitor ....................................................................................... 19
4.1.2 Battery indication ................................................................................................... 19
4.1.3 Messages .............................................................................................................. 19
4.1.4 Results .................................................................................................................. 20
4.1.5 Sound warnings ..................................................................................................... 20
4.1.6 Help screens ......................................................................................................... 20
4.1.7 Backlight and contrast adjustments ....................................................................... 21
4.2 Function selection ..................................................................................................... 22
4.3 Settings ..................................................................................................................... 23
4.3.1 Memory (MI 3100 SE only) .................................................................................... 23
4.3.2 Language .............................................................................................................. 23
4.3.3 Date and time ........................................................................................................ 24
4.3.4 RCD testing ........................................................................................................... 24
4.3.5 Isc factor................................................................................................................ 25
4.3.6 Commander support (MI 3100 SE only) ................................................................ 26
4.3.7 Communication (MI 3100 SE only) ........................................................................ 26
4.3.8 Initial settings ........................................................................................................ 28
5 Measurements ................................................................................................................. 30
5.1 Voltage, frequency and phase sequence ................................................................... 30
5.2 Insulation resistance.................................................................................................. 32
5.3 Resistance of earth connection and equipotential bonding ........................................ 34
5.3.1 R LOWΩ, 200 mA resistance measurement .......................................................... 34
5.3.2 Continuous resistance measurement with low current ........................................... 35
5.3.3 Compensation of test leads resistance .................................................................. 36
5.4 Testing RCDs ............................................................................................................ 38
5.4.1 Contact voltage (RCD Uc) ..................................................................................... 39
5.4.2 Trip-out time (RCDt) .............................................................................................. 40
5.4.3 Trip-out current (RCD I) ......................................................................................... 40
5.4.4 RCD Auto-test ....................................................................................................... 42
5.5 Fault loop impedance and prospective fault current................................................... 44
5.6 Line impedance and prospective short-circuit current / Voltage drop ......................... 46
5.6.1 Line impedance and prospective short circuit current ............................................ 47
5.6.2 Voltage drop .......................................................................................................... 48
3
Page 4
MI 3100 s, MI 3100 SE EurotestEASI Table of contents
5.7 Earth resistance ........................................................................................................ 50
5.7.1 Standard earthing resistance measurement .......................................................... 51
5.8 PE test terminal ......................................................................................................... 53
5.9 PE conductor resistance (MI 3100 SE only) .............................................................. 55
6 Auto-sequences (MI 3100 SE only) ................................................................................ 57
7 Data handling (MI 3100 SE only) .................................................................................... 61
7.1 Memory organization ................................................................................................. 61
7.2 Data structure............................................................................................................ 61
7.3 Storing test results .................................................................................................... 63
7.4 Recalling test results ................................................................................................. 64
7.5 Clearing stored data .................................................................................................. 65
7.5.1 Clearing complete memory content ....................................................................... 65
7.5.2 Clearing measurement(s) in selected location ....................................................... 65
7.5.3 Clearing individual measurements ......................................................................... 66
7.5.4 Renaming installation structure elements (upload from PC) ................................ .. 67
7.5.5 Renaming installation structure elements with serial barcode reader or RFID reader 67
7.6 Communication ......................................................................................................... 68
7.6.1 USB and RS232 communication ........................................................................... 68
7.6.2 Bluetooth communication ...................................................................................... 69
8 Upgrading the instrument .............................................................................................. 70
9 Maintenance .................................................................................................................... 71
9.1 Fuse replacement ..................................................................................................... 71
9.2 Cleaning .................................................................................................................... 71
9.3 Periodic calibration ................................ ................................................................ .... 71
9.4 Service ...................................................................................................................... 71
10 Technical specifications ................................................................................................. 72
10.1 Insulation resistance.................................................................................................. 72
10.2 Continuity .................................................................................................................. 73
10.2.1 Resistance R LOW ............................................................................................ 73
10.2.2 Resistance CONTINUITY .................................................................................. 73
10.3 RCD testing ............................................................................................................... 74
10.3.1 General data ...................................................................................................... 74
10.3.2 Contact voltage RCD Uc .................................................................................... 74
10.3.3 Trip-out time ...................................................................................................... 75
10.3.4 Trip-out current .................................................................................................. 75
10.4 Fault loop impedance and prospective fault current................................................... 76
10.4.1 No disconnecting device or FUSE selected ....................................................... 76
10.4.2 RCD selected .................................................................................................... 76
10.5 Line impedance and prospective short-circuit current / Voltage drop ......................... 77
10.6 PE conductor resistance (MI 3100 SE only) .............................................................. 78
10.6.1 No RCD selected ............................................................................................... 78
10.6.2 RCD selected .................................................................................................... 78
10.7 Resistance to earth ................................................................................................... 79
10.7.1 Standard earthing resistance measurement – 3-wire measurement .................. 79
10.8 Voltage, frequency, and phase rotation ..................................................................... 80
10.8.1 Phase rotation ................................................................................................... 80
10.8.2 Voltage .............................................................................................................. 80
10.8.3 Frequency ......................................................................................................... 80
10.8.4 Online terminal voltage monitor ......................................................................... 80
10.9 General data ............................................................................................................. 81
4
Page 5
MI 3100 s, MI 3100 SE EurotestEASI Table of contents
Appendix A – Fuse table ........................................................................................................ 82
A.1 Fuse table – IPSC ..................................................................................................... 82
Appendix B – Accessories for specific measurements ....................................................... 86
Appendix C – Country notes .................................................................................................. 87
C.1 List of country modifications ...................................................................................... 87
C.2 Modification issues .................................................................................................... 87
C.2.1 AT modification - G type RCD............................................................................ 87
Appendix D – Commanders (A 1314, A 1401) ....................................................................... 89
D.1 Warnings related to safety .................................................................................. 89
D.2 Battery....................................................................................................................... 89
D.3 Description of commanders ....................................................................................... 90
D.4 Operation of commanders ......................................................................................... 91
5
Page 6
MI 3100 s, MI 3100 SE EurotestEASI Preface

1 Preface

Congratulations on your purchase of the Eurotest instrument and its accessories from METREL. The instrument was designed on a basis of rich experience, acquired through many years of dealing with electric installation test equipment.
The Eurotest instrument is a professional, multifunctional, hand-held test instrument intended to perform all the measurements on a.c. electrical LV installations.
The following measurements and tests can be performed:
Voltage and frequency, Continuity tests, Insulation resistance tests, RCD testing, Fault loop / RCD trip-lock impedance measurements, Line impedance / Voltage drop, Phase sequence, Earthing resistance tests, Pre-defined auto-sequences (MI3100 SE only).
The graphic display with backlight offers easy reading of results, indications, measurement parameters and messages. Two LED PASS/FAIL indicators are placed at the sides of the LCD. The operation of the instrument is designed to be as simple and clear as possible and no special training (except for the reading this instruction manual) is required in order to begin using the instrument. In order for operator to be familiar enough with performing measurements in general and their typical applications it is advisable to read METREL handbook Guide for testing and verification of low voltage installations.
The instrument is equipped with the entire necessary accessory for comfortable testing.
6
Page 7
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations

2 Safety and operational considerations

2.1 Warnings and notes

In order to maintain the highest level of operator safety while carrying out various tests and measurements, METREL recommends keeping your Eurotest instruments in good condition and undamaged. When using the instrument, consider the following general warnings:
General warnings related to safety:
The symbol on the instrument means »Read the Instruction manual with
special care for safe operation«. The symbol requires an action!
If the test equipment is used in a manner not specified in this user manual, the
protection provided by the equipment could be impaired!
Read this user manual carefully, otherwise the use of the instrument may be
dangerous for the operator, the instrument or for the equipment under test!
Do not use the instrument or any of the accessories if any damage is noticed! Consider all generally known precautions in order to avoid risk of electric shock
while dealing with hazardous voltages!
In case a fuse has blown follow the instructions in this manual in order to replace
it! Use only fuses that are specified!
Do not use the instrument in AC supply systems with voltages higher than
550 Va.c.
Service, repairs or adjustment of instruments and accessories is only allowed to
be carried out by a competent authorized personnel!
Use only standard or optional test accessories supplied by your distributor! Consider that protection category of some accessories is lower than of the
instrument. Test tips and Tip commander have removable caps. If they are removed the protection falls to CAT II. Check markings on accessories!
cap off, 18 mm tip: CAT II up to 1000 V cap on, 4 mm tip: CAT II 1000 V / CAT III600 V / CAT IV300 V
The instrument comes supplied with rechargeable Ni-MH battery cells. The cells
should only be replaced with the same type as defined on the battery compartment label or as described in this manual. Do not use standard alkaline battery cells while the power supply adapter is connected, otherwise they may explode!
Hazardous voltages exist inside the instrument. Disconnect all test leads, remove
the power supply cable and switch off the instrument before removing battery compartment cover.
All normal safety precautions must be taken in order to avoid risk of electric
shock while working on electrical installations!
7
Page 8
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
Warnings related to safety of measurement functions:
Insulation resistance
Insulation resistance measurement should only be performed on de-energized objects! Do not touch the test object during the measurement or before it is fully discharged! Risk
of electric shock!
When an insulation resistance measurement has been performed on a capacitive object,
automatic discharge may not be done immediately! The warning message and the actual voltage are displayed during discharge until voltage drops below 30 V.
Do not connect test terminals to external voltage higher than 600 V (AC or DC) in order
not to damage the test instrument!
Continuity functions
Continuity measurements should only be performed on de-energized objects! Parallel loops may influence on test results.
Testing PE terminal
If phase voltage is detected on the tested PE terminal, stop all measurements
immediately and ensure the cause of the fault is eliminated before proceeding with any activity!
Notes related to measurement functions: General
The indicator means that the selected measurement cannot be performed because
of irregular conditions on input terminals.
Insulation resistance, continuity functions and earth resistance measurements can only
be performed on de-energized objects.
PASS / FAIL indication is enabled when limit is set. Apply appropriate limit value for
evaluation of measurement results.
In the case that only two of the three wires are connected to the electrical installation
under test, only voltage indication between these two wires is valid.
Insulation resistance
The standard three-wire test lead, schuko test cable or Plug / Tip commanders can be
used for the insulation test with voltages ≤ 1kV.
If a voltage of higher than 30 V (AC or DC) is detected between test terminals, the
insulation resistance measurement will not be performed.
The instrument automatically discharge tested object after finished measurement. A double click of TEST key starts a continuous measurement.
Continuity functions
If a voltage of higher than 10 V (AC or DC) is detected between test terminals, the
continuity resistance test will not be performed.
Compensate test lead resistance before performing a continuity measurement, where
8
Page 9
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
Earth resistance
If voltage between test terminals is higher than 30 V the resistance to earth
measurement will not be performed.
If a noise voltage higher than approx. 5 V is present between the H and E or S test
terminals, ” (noise) warning symbol will be displayed, indicating that the test result may not be correct!
RCD functions
Parameters set in one function are also kept for other RCD functions! The measurement of contact voltage does not normally trip an RCD. However, the trip
limit of the RCD may be exceeded as a result of leakage current flowing to the PE protective conductor or a capacitive connection between L and PE conductors.
The RCD trip-lock sub-function (function selector switch in LOOP position) takes longer
to complete but offers much better accuracy of fault loop resistance (in comparison to the RL sub-result in Contact voltage function).
RCD trip-out time and RCD trip-out current measurements will only be performed if the
contact voltage in the pre-test at nominal differential current is lower than the set contact voltage limit!
The auto-test sequence (RCD AUTO function) stops when trip-out time is out of
allowable time period.
Z-LOOP
The low limit prospective short-circuit current value depends on fuse type, fuse current
rating, fuse trip-out time and impedance scaling factor.
The specified accuracy of tested parameters is valid only if the mains voltage is stable
during the measurement.
Fault loop impedance measurements will trip an RCD. The measurement of fault loop impedance using trip-lock function does not normally trip
an RCD. However, the trip limit may be exceeded if a leakage current flows to the PE protective conductor or if there is a capacitive connection between L and PE conductors.
Z-LINE / Voltage drop
In case of measurement of Z
with the instrument test leads PE and N connected
Line-Line
together the instrument will display a warning of dangerous PE voltage. The measurement will be performed anyway.
Specified accuracy of tested parameters is valid only if mains voltage is stable during the
measurement.
L and N test terminals are reversed automatically according to detected terminal voltage
(except in UK version).
Testing PE terminal
PE terminal can be tested in RCD, LOOP and LINE function selector switch positions
only!
For correct testing of PE terminal, the TEST key has to be touched for a few seconds. Make sure to stand on non-isolated floor while carrying out the test, otherwise test result
may be wrong!
PE conductor resistance (MI 3100 SE)
The specified accuracy of tested parameters is valid only if the mains voltage is stable
during the measurement.
9
Page 10
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
PE conductor resistance measurement will trip an RCD. The measurement of PE conductor resistance using trip-lock function does not normally
trip an RCD. However, the trip limit may be exceeded if a leakage current flows to the PE protective conductor or if there is a capacitive connection between L and PE conductors.
Autosequence tests (MI 3100 SE)
See notes above for selected tests in Auto-sequence.
10
Page 11
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
+
-
Symbols:
Indication of battery charging
Figure 2.3: Charging indication

2.2 Battery and charging

The instrument uses six AA size alkaline or rechargeable Ni-MH battery cells. Nominal operating time is declared for cells with nominal capacity of 2100 mAh. Battery condition is always displayed in the lower right display part. In case the battery is too weak the instrument indicates this as shown in figure 2.1. This indication appears for a few seconds and then the instrument turns itself off.
Figure 2.1: Discharged battery indication
The battery is charged whenever the power supply adapter is connected to the instrument. The power supply socket polarity is shown in figure 2.2. Internal circuit controls charging and assures maximum battery lifetime.
Figure 2.2: Power supply socket polarity
Warnings related to safety:
When connected to an installation, the instruments battery compartment can contain
hazardous voltage inside! When replacing battery cells or before opening the battery/fuse compartment cover, disconnect any measuring accessory connected to the instrument and turn off the instrument,
Ensure that the battery cells are inserted correctly otherwise the instrument will not
operate and the batteries could be discharged.
Do not recharge alkaline battery cells! Use only power supply adapter delivered from the manufacturer or distributor of the test
equipment !
Notes:
The charger in the instrument is a pack cell charger. This means that the battery cells
are connected in series during the charging. The battery cells have to be equivalent (same charge condition, same type and age).
If the instrument is not to be used for a long period of time, remove all batteries from the
battery compartment.
Alkaline or rechargeable Ni-MH batteries (size AA) can be used. METREL recommends
only using rechargeable batteries with a capacity of 2100mAh or above.
Unpredictable chemical processes can occur during the charging of battery cells that
have been left unused for a longer period (more than 6 months). In this case METREL recommends repeating the charge/discharge cycle at least 2-4 times.
If no improvement is achieved after several charge / discharge cycles, then each battery
cell should be checked (by comparing battery voltages, testing them in a cell charger,
11
Page 12
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
etc). It is very likely that only some of the battery cells are deteriorated. One different battery cell can cause an improper behaviour of the entire battery pack!
The effects described above should not be confused with the normal decrease of battery
capacity over time. Battery also loses some capacity when it is repeatedly charged / discharged. This information is provided in the technical specification from battery manufacturer.
12
Page 13
MI 3100 s, MI 3100 SE EurotestEASI Safety and operational considerations
Electromagnetic compatibility (EMC)
EN 61326
Electrical equipment for measurement, control and laboratory use – EMC requirements Class B (Hand-held equipment used in controlled EM environments)
Safety (LVD)
EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use – Part 1: General requirements
EN 61010-2-030
Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-030: Particular requirements for testing and measuring circuits
EN 61010-031
Safety requirements for electrical equipment for measurement, control and laboratory use – Part 031: Safety requirements for hand-held probe assemblies for electrical measurement and test
EN 61010-2-032
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-032: Particular requirements for hand-held and hand­manipulated current sensors for electrical test and measurement
Functionality
EN 61557
Electrical safety in low voltage distribution systems up to 1000 VAC and 1500 VAC – Equipment for testing, measuring or monitoring of protective measures
Part 1: General requirements Part 2: Insulation resistance Part 3: Loop resistance Part 4: Resistance of earth connection and equipotential bonding Part 5: Resistance to earth Part 6: Residual current devices (RCDs) in TT and TN systems Part 7: Phase sequence Part 10: Combined measuring equipment Part 12: Performance measuring and monitoring devices (PMD)
Reference standards for electrical installations and components
EN 61008
Residual current operated circuit-breakers without integral overcurrent protection for household and similar uses
EN 61009
Residual current operated circuit-breakers with integral overcurrent protection for household and similar uses
IEC 60364-4-41
Electrical installations of buildings Part 4-41 Protection for safety – protection against electric shock
BS 7671
IEE Wiring Regulations (17th edition)
AS/NZS 3017
Electrical installations – Verification guidelines

2.3 Standards applied

The Eurotest instruments are manufactured and tested in accordance with the following regulations:
Note about EN and IEC standards:
Text of this manual contains references to European standards. All standards of EN
6XXXX (e.g. EN 61010) series are equivalent to IEC standards with the same number (e.g. IEC 61010) and differ only in amended parts required by European harmonization procedure.
13
Page 14
MI 3100 s, MI 3100 SE EurotestEASI Instrument description
1
LCD
128 x 64 dots matrix display with backlight.
2
UP
Modifies selected parameter.
3
DOWN
4
TEST
TEST
Starts measurements.
Acts also as the PE touching electrode.
5
ESC
Goes one level back.
6
TAB
Selects the parameters in selected function.
7
Backlight, Contrast
Changes backlight level and contrast.
8
ON / OFF
Switches the instrument power on or off. The instrument automatically turns off 15 minutes after the last key was pressed.
9
HELP / CAL
Accesses help menus.
Calibrates test leads in Continuity functions.
Starts Z
REF
measurement in Voltage drop sub-function.
10
Function selector switch
Selects test / measurement function and settings.
11
MEM
Stores / recalls memory of instrument (MI 3100 SE only). No function (MI 3100 s).
12
Green LEDs Red LEDs
Indicates PASS / FAIL of result.

3 Instrument description

3.1 Front panel

Legend:
Figure 3.1: Front panel
14
Page 15
MI 3100 s, MI 3100 SE EurotestEASI Instrument description
1
Test connector
Measuring inputs / outputs.
2
Charger socket
3
PS/2 connector
(MI 3100 SE)
Communication with PC serial port Connection to barcode / RFID reader Connection of Bluetooth dongle
4
Protection cover
5
USB connector
(MI 3100 SE)
Communication with PC USB (1.1) port.

3.2 Connector panel

Figure 3.2: Connector panel
Legend:
Warnings!
Maximum allowed voltage between any test terminal and ground is 600 V! Maximum allowed voltage between test terminals on test connector is 600 V! Maximum short-term voltage of external power supply adapter is 14 V!
15
Page 16
MI 3100 s, MI 3100 SE EurotestEASI Instrument description
1
2
3
1
Battery / fuse compartment cover
2
Back panel information label
3
Fixing screws for battery / fuse compartment cover
1
Fuse F1
M 315 mA / 250 V
2
Fuses F2 and F3
F 4 A / 500 V (breaking capacity 50 kA)
3
Serial number label
4 Battery cells
Size AA, alkaline / rechargeable NiMH

3.3 Back side

Figure 3.3: Back panel
Legend:
Legend:
F
Figure 3.4: Battery and fuse compartment
16
Page 17
MI 3100 s, MI 3100 SE EurotestEASI Instrument description
3
1
2
R: 0.0 1999 Test current: max. 8.5mA Open-circuit voltage: 6.5V
Continuity 7mA
R: 0.18M 199.9M , U =50V ,100 , R: 0.12M 999M , U = 500V , 1kV U: 0V 1200V Nominal voltages: 100V , 250V , 500V , 1 kV Measuring current: min. 1mA at R =U 1k /V Short-circuit current: < 3mA
N
N
N N
Insulation resistance (EN 61557-2)
V 250V
Continuity
Tripping time
non-delayed (time-delayed) RCDs
1: 0ms 300ms (500ms) 2: 0ms 150ms (200ms) 5: 0ms 40ms (150ms), U : 0.0V 100.0V
C
Tripping current
I : 0.2 I 1.1 I AC ( t : 0ms 3 00ms, U : 0.0V 1 00.0V
C
N N N
1.5 I A)
Multiplier: 0.5, 1, 2, 5
Contact voltage
U : 0.0V 100 .0V R : 0.00 10.00k (R =U / I )
C S S C N
R : 0. 1999 I : 0. A . kAL-N(L) PSC17 20 1 4 Nominal voltage: 100V 440V/ 1 5Hz 500 Hz
Line impedance (EN 61557-3)
Fault loop (EN 61557-3)
R : 0. 1999 I : 0. A . kA
PFC 14 1 4
Nominal voltage: 100V 264V/ 1 5Hz 500 Hz
impedance
L-PE 17
Voltage, frequency
U: 0V 440V / f: 15Hz 500Hz
Phase rotation (EN 61557-7)
Nominal voltage: 100V 440V / 1 00 Results: 1.2.3 or 2.1.3
5Hz 5 Hz
RCD (EN 61557-6)
I : 10mA, 30mA, 100mA, 300mA, 500mA, 1A Nominal voltage: 100V 264V/ 15Hz 500Hz
R Low (EN 61557-4)
R: 0.12 1999
Test current: min. ±200mA a t 2
Open-circuit voltage: 6.5V 9.0V
Resistance to earth (EN 61557-5)
R: 0.04 9999 Open-circuit voltage: Short-circuit current
< 45V
: < 20mA
RMS
Ljubljanska 77 SI - 1354 Horjul Tel: +386 1 75 58 200 http://www.metrel.si
20 224 832
CAT III 600V
550V
1
Bottom information label
2
Neck belt openings
3
Handling side covers
The instrument hangs around operators neck only – quick placing and displacing.
Figure 3.5: Bottom
Legend:

3.4 Carrying the instrument

With the neck-carrying belt supplied in standard set, various possibilities of carrying the instrument are available. Operator can choose appropriate one on basis of his operation, see the following examples:
17
Page 18
MI 3100 s, MI 3100 SE EurotestEASI Instrument description
The instrument can be used even when placed in soft carrying bag – test cable connected to the instrument through the front aperture.

3.5 Instrument set and accessories

3.5.1 Standard set MI 3100 s – EurotestEASI

Instrument Soft carrying bag Schuko-plug test cable Test lead, 3 x 1.5 m Test probe, 3 pcs Crocodile clip, 3 pcs Set of carrying straps Set of Ni-MH battery cells Power supply adapter CD with instruction manual, and “Guide for testing and verification of low voltage
installations” handbook.
Short instruction manual Calibration Certificate

3.5.2 Standard set MI 3100 SE – EurotestEASI

Instrument Soft carrying bag Schuko-plug test cable Test lead, 3 x 1.5 m Test probe, 3 pcs Crocodile clip, 3 pcs Set of carrying straps RS232-PS/2 cable USB cable Set of Ni-MH battery cells Power supply adapter CD with instruction manual, “Guide for testing and verification of low voltage
installations” handbook and PC software EurolinkPRO.
Short instruction manual Calibration Certificate

3.5.3 Optional accessories

See the attached sheet for a list of optional accessories that are available on request from your distributor.
18
Page 19
MI 3100 s, MI 3100 SE EurotestEASI Instrument operation
Online voltages are displayed together with test terminal indication. All three test terminals are used for selected measurement.
Online voltages are displayed together with test terminal indication. L and N test terminals are used for selected measurement.
L and PE are active test terminals; N terminal should also be connected for correct input voltage condition.
Polarity of test voltage applied to the output terminals, L and N.
Battery capacity indication.
Low battery. Battery is too weak to guarantee correct result. Replace or recharge the battery cells.
Charging in progress (if power supply adapter is connected).
Measurement is running, consider displayed warnings.
Conditions on the input terminals allow starting the measurement; consider other displayed warnings and messages.
Conditions on the input terminals do not allow starting the measurement, consider displayed warnings and messages.
RCD tripped-out during the measurement (in RCD functions).
Portable RCD selected (PRCD).
Instrument is overheated. The measurement is prohibited until the temperature decreases under the allowed limit.
Result(s) can be stored. (MI 3100 SE only)

4 Instrument operation

4.1 Display and sound

4.1.1 Terminal voltage monitor

The terminal voltage monitor displays on-line the voltages on the test terminals and information about active test terminals in the a.c. installation measuring mode.

4.1.2 Battery indication

The battery indication indicates the charge condition of battery and connection of external charger.

4.1.3 Messages

In the message field warnings and messages are displayed.
19
Page 20
MI 3100 s, MI 3100 SE EurotestEASI Instrument operation
High electrical noise was detected during measurement. Results may be impaired.
L and N are changed.
Warning! High voltage is applied to the test terminals.
Warning! Dangerous voltage on the PE terminal! Stop the activity
immediately and eliminate the fault / connection problem before proceeding with any activity!
Test leads resistance in Continuity measurement is not compensated.
Test leads resistance in Continuity measurement is compensated.
High resistance to earth of test probes. Results may be impaired.
Measured signal is out of range (clipped). Results are impaired.
Fuse F1 is broken.
Measurement result is inside pre-set limits (PASS).
Measurement result is out of pre-set limits (FAIL).
Measurement is aborted. Consider displayed warnings and messages.
Continuous sound
Warning! Dangerous voltage on the PE terminal is detected.
HELP
Opens help screen.
UP / DOWN
Selects next / previous help screen.
ESC / HELP / Function selector switch
Exits help menu.

4.1.4 Results

4.1.5 Sound warnings

4.1.6 Help screens

Help menus are available in all functions. The Help menu contains schematic diagrams for illustrating how to properly connect the instrument to electric installation. After selecting the measurement you want to perform, press the HELP key in order to view the associated Help menu.
Keys in help menu:
20
Page 21
MI 3100 s, MI 3100 SE EurotestEASI Instrument operation
Click
Toggles backlight intensity level.
Keep pressed for 1 s
Locks high intensity backlight level until power is turned off or the key is pressed again.
Keep pressed for 2 s
Bar-graph for LCD contrast adjustment is displayed.
DOWN
Reduces contrast.
UP
Increases contrast.
TEST
Accepts new contrast.
ESC
Exits without changes.
Figure 4.1: Examples of help screens

4.1.7 Backlight and contrast adjustments

With the BACKLIGHT key backlight and contrast can be adjusted.
Figure 4.2: Contrast adjustment menu
Keys for contrast adjustment:
21
Page 22
MI 3100 s, MI 3100 SE EurotestEASI Instrument operation
Function selector switch
Selects test / measurement function, enters settings menu and selects auto-test mode.
UP / DOWN
Selects sub-function in selected measurement function.
TAB
Selects the test parameter to be set or modified.
TEST
Runs selected test / measurement function.
MEM
Stores measured results / recalls stored results. ( MI 3100 SE only)
ESC
Exits back.
UP / DOWN
Changes the selected parameter.
TAB
Selects the next measuring parameter.
Function selector switch
Toggles between the main functions.
MEM
Stores measured results / recalls stored results
Parameter
OFF
No limit values, indication: _ _ _.
ON
Value(s) – results will be marked as PASS or FAIL in accordance
with selected limit.

4.2 Function selection

For selecting test / measurement function, entering settings menu and auto-test mode (MI 3100 SE only) the FUNCTION SELECTOR SWITCH shall be used.
Function selector switch and keys:
Keys in test parameter field:
General rule regarding enabling parameters for evaluation of measurement / test result:
See chapter 5 Measurements for more information about the operation of the instrument test functions.
22
Page 23
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Options are:
recalling and clearing stored results (MI 3100 SE only), selection of language, setting the date and time, selection of reference standard for RCD tests, entering Isc factor, commander support (MI 3100 SE only), settings for Bluetooth communication (MI 3100 SE only), setting the instrument to initial values.
Figure 4.3: Options in
Settings menu
UP / DOWN
Selects appropriate option.
TEST
Enters selected option.
Function selector switch
Exits back to selected test / measurement function without changes.
In this menu the stored data can be recalled or deleted. See chapter 7 Data handling for more information.
Figure 4.4: Memory options
UP / DOWN
Selects option.
TEST
Enters selected option.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
In this menu the language can be set.
Figure 4.5: Language selection

4.3 Settings

Different instrument options can be set in the SETTINGS menu.
Keys:

4.3.1 Memory (MI 3100 SE only)

Keys:

4.3.2 Language

Keys:
23
Page 24
MI 3100 s, MI 3100 SE EurotestEASI Measurements
UP / DOWN
Selects language.
TEST
Confirms selected language and exits to settings menu.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
In this menu date and time can be set.
Figure 4.6: Setting date and time
TAB
Selects the field to be changed.
UP / DOWN
Modifies selected field.
TEST
Confirms new date / time and exits.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
In this menu the used standard for RCD tests can be set.
Figure 4.7: Selection of RCD test standard
UP / DOWN
Selects standard.
TEST
Confirms selected standard.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement without changes.
½I
N
*)
IN
2IN
5IN
General RCDs (non-delayed)
t> 300 ms
t< 300 ms
t< 150 ms
t< 40 ms
Selective RCDs (time-delayed)
t> 500 ms
130 ms < t< 500 ms
60 ms < t< 200 ms
50 ms < t< 150 ms

4.3.3 Date and time

Keys:
Note:
If the batteries are removed for more than 1 minute the set date and time will be lost.

4.3.4 RCD testing

Keys:
Maximum RCD disconnection times differ in various standards. The trip-out times defined in individual standards are listed below.
Trip-out times according to EN 61008 / EN 61009:
24
Page 25
MI 3100 s, MI 3100 SE EurotestEASI Measurements
½I
N
*)
IN
2IN
5IN
General RCDs (non-delayed)
t> 999 ms
t< 999 ms
t< 150 ms
t< 40 ms
Selective RCDs (time-delayed)
t> 999 ms
130 ms < t< 999 ms
60 ms < t< 200 ms
50 ms < t< 150 ms
½I
N
*)
IN
2IN
5IN
General RCDs (non-delayed)
t> 1999 ms
t< 300 ms
t< 150 ms
t< 40 ms
Selective RCDs (time-delayed)
t> 1999 ms
130 ms < t< 500 ms
60 ms < t< 200 ms
50 ms < t< 150 ms
½I
N
*)
IN
2IN
5IN
RCD type
IN [mA]
t
t
t
t
Note
I
10
> 999 ms
40 ms
40 ms
40 ms
Maximum break time
II
> 10 30
300 ms
150 ms
40 ms
III
> 30
300 ms
150 ms
40 ms
IV
S
> 30
> 999 ms
500 ms
200 ms
150 ms
130 ms
60 ms
50 ms
Minimum non-actuating time
Standard
½IN
IN
2IN
5IN
EN 61008 / EN 61009
300 ms
300 ms
150 ms
40 ms
EN 60364-4-41
1000 ms
1000 ms
150 ms
40 ms
BS 7671
2000 ms
300 ms
150 ms
40 ms
AS/NZS 3017 (I, II, III)
1000 ms
1000 ms
150 ms
40 ms
Standard
½IN
IN
2IN
5IN
EN 61008 / EN 61009
500 ms
500 ms
200 ms
150 ms
EN 60364-4-41
1000 ms
1000 ms
200 ms
150 ms
BS 7671
2000 ms
500 ms
200 ms
150 ms
AS/NZS 3017 (IV)
1000 ms
1000 ms
200 ms
150 ms
In this menu the Isc factor for calculation of short circuit current in Z-LINE and Z-LOOP measurements can be set.
Figure 4.8: Selection of Isc factor
Trip-out times according to EN 60364-4-41:
Trip-out times according to BS 7671:
Trip-out times according to AS/NZS 3017
*)
Minimum test period for current of ½IN, RCD shall not trip-out.
**)
Test current and measurement accuracy correspond to AS/NZS 3017 requirements.
**)
:
Maximum test times related to selected test current for general (non-delayed) RCD
Maximum test times related to selected test current for selective (time-delayed) RCD
Note:
Trip-out limit times for PRCD, PRCD-K and PRCD-S are equal to General (non-delayed)
RCDs.

4.3.5 Isc factor

25
Page 26
MI 3100 s, MI 3100 SE EurotestEASI Measurements
UP / DOWN
Sets Isc value.
TEST
Confirms Isc value.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
The support for remote commanders can be enabled or disabled in this menu.
Figure 4.9: Selection of commander support
UP / DOWN
Selects commander option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
In this menu the Bluetooth dongle A1436 can be initialized and device for scanning barcodes can be selected.
Figure 4.10: Communication menu
UP / DOWN
Selects option.
TEST
Confirms selected option.
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
Keys:
Short circuit current Isc in the supply system is important for selection or verification of protective circuit breakers (fuses, over-current breaking devices, RCDs). The default value of Isc factor (ksc) is 1.00. The value should be set according to local regulative. Range for adjustment of the Isc factor is 0.20 3.00.

4.3.6 Commander support (MI 3100 SE only)

Keys:
Note:
Commander disabled option is intended to disable the commander’s remote keys. In the
case of high EM interfering noise the operation of the commander can be irregular.

4.3.7 Communication (MI 3100 SE only)

Keys:
Initialization of the Bluetooth dongle
26
Page 27
MI 3100 s, MI 3100 SE EurotestEASI Measurements
UP / DOWN
Selects option.
TEST
Confirms selected option.
ESC
Exits back to Communication menu.
Function selector switch
Exits back to selected test / measurement function without changes.
The Bluetooth dongle A1436 should be initialized when it is used with the instrument for the first time. During initialization the instrument sets the dongle parameters and name in order to communicate properly with PC and other devices via Bluetooth.
Initialization procedure
Connect Bluetooth dongle A1436 to the instrument. Press RESET key on the Bluetooth dongle A1436 for at least 10 seconds. Select INIT. BT DONGLE in Communication menu and press the TEST key . Wait for confirmation message and beep. Following message is displayed if dongle was
initialized properly: EXTERNAL BT DONGLE SEARCHING OK!
Notes:
The Bluetooth dongle A1436 should always be initialized before first use with the
instrument.
If the dongle was initialized by another Metrel instrument it will probably not work
properly when working with the instrument again.
For more information about communication via Bluetooth refer to chapter 7.6
Communication and A1436 manual.
Selection of barcode scanner type
In this menu the barcode scanner type can be set. Options are:
serial barcode scanner and barcode scanner application on Android devices.
Keys:
Figure 4.11: Menu for scanner type
27
Page 28
MI 3100 s, MI 3100 SE EurotestEASI Measurements
In this menu the instrument settings, measurement parameters and limits can be set to initial (factory) values.
Figure 4.12: Initial settings dialogue
UP / DOWN
Selects option [YES, NO].
TEST
Restores default settings (if YES is selected).
ESC
Exits back to settings menu.
Function selector switch
Exits back to selected test / measurement function without changes.
Instrument setting
Default value
Language
English
Contrast
As defined and stored by adjustment procedure
Isc factor
1.00
RCD standards
EN 61008 / EN 61009
Commander (MI 3100 SE only)
A 1314, A 1401
Scanner type (MI 3100 SE only)
RS 232
Test mode:
Function
Parameters / limit value
Sub-function
INSTALLATION:
EARTH RE
No limit
R ISO
No limit Utest = 500 V
Low Ohm Resistance
R LOW
No limit
CONTINUITY
No limit, sound OFF
Rpe (MI 3100 SE only)
Rpe Rpe(rcd)
No limit No limit
Z - LINE
Fuse type: none selected
VOLTAGE DROP
ΔU: 4.0 %
Z
REF
: 0.00 Ω
Z - LOOP
Fuse type: none selected
Zs rcd
Fuse type: none selected
RCD
RCD t

4.3.8 Initial settings

Keys:
Warnings:
Customized settings will be lost when this option is used! If the batteries are removed for more than 1 minute the custom made settings will be
lost.
The default setup is listed below:
28
Page 29
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Nominal differential current: IN=30 mA RCD type: AC, non-delayed Test current starting polarity: (0) Limit contact voltage: 50 V Current multiplier: 1
AUTO SEQUENCES (MI 3100 SE only):
AUTO TT
FUSE: None selected Z
REF
: --­ΔU: 4.0 % RCD: 30 mA, AC, non-delayed, (0) Uc: 50 V
AUTO TN (rcd)
FUSE: None selected Z
REF
: --­ΔU: 4.0 % Rpe: No limit
AUTO TN
FUSE: None selected Z
REF
: --­ΔU: 4.0 % Rpe: No limit
Note:
Initial settings (reset of the instrument) can be recalled also if the TAB key is pressed
while the instrument is switched on.
29
Page 30
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.1: Voltage in single
phase system

5 Measurements

5.1 Voltage, frequency and phase sequence

Voltage and frequency measurement is always active in the terminal voltage monitor. In the special VOLTAGE TRMS menu the measured voltage, frequency and information about detected three-phase connection can be stored. Measurements are based on the EN 61557-7 standard.
Test parameters for voltage measurement
There are no parameters to be set.
Connections for voltage measurement
Figure 5.2: Connection of 3-wire test lead and optional adapter in three-phase system
Figure 5.3: Connection of plug commander and 3-wire test lead in single-phase system
30
Page 31
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the VOLTAGE TRMS function using the function selector switch.
Connect test cable to the instrument.
Connect test leads to the item to be tested (see Figure 5.2 and Figure 5.3).
Store voltage measurement result by pressing the MEM key (MI 3100 SE only).
Voltage measurement procedure
Measurement runs immediately after selection of VOLTAGE TRMS function.
Figure 5.4: Examples of voltage measurement in three-phase system
Displayed results for single phase system:
Uln ........... voltage between phase and neutral conductors
Ulpe ......... voltage between phase and protective conductors
Unpe ........ voltage between neutral and protective conductors
f ................ frequency
Displayed results for three-phase system:
U12........... voltage between phases L1 and L2
U13........... voltage between phases L1 and L3
U23........... voltage between phases L2 and L3
1.2.3 ......... correct connection – CW rotation sequence
3.2.1 ......... invalid connection – CCW rotation sequence
f ................ frequency
31
Page 32
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.5: Insulation resistance
Uiso
Nominal test voltage [50 V, 100 V, 250 V, 500 V, 1000 V]
Limit
Minimum insulation resistance [OFF, 0.01 M ÷ 200 M]
Select the R ISO function using the function selector switch.
Set the required test voltage.
Enable and set limit value (optional).
Disconnect tested installation from mains supply (and discharge insulation as required).
Connect test cable to the instrument and to the item to be tested (see Figure 5.6).
Press the TEST key to perform the measurement (double click for continuous
measurement and later press to stop the measurement).
After the measurement is finished wait until tested item is fully discharged.
Store the result by pressing the MEM key (MI 3100 SE only).

5.2 Insulation resistance

The insulation resistance measurement is performed in order to ensure safety against electric shock through insulation. Typical applications are:
insulation resistance between conductors of installation, insulation resistance of non-conductive rooms (walls and floors), insulation resistance of ground cables and resistance of semi-conductive (antistatic) floors.
Test parameters for insulation resistance measurement
Test circuits for insulation resistance
Figure 5.6: Connection of 3-wire test lead and tip commander
Insulation resistance measuring procedure
32
Page 33
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Figure 5.7: Example of insulation resistance measurement result
Displayed results:
R .............. insulation resistance
Um ........... test voltage (actual value)
33
Page 34
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.8: 200 mA RLOW Ω
Test
Resistance measurement sub-function [R LOWΩ, CONTINUITY]
Limit
Maximum resistance [OFF, 0.1 ÷ 20.0 Ω]
Buzzer On (sound if resistance is lower than the set limit value) or Off
HELP
Click
Calibrates test leads in Continuity functions.
Keep pressed for 1s
Enters Help screen

5.3 Resistance of earth connection and equipotential bonding

The resistance measurement is performed in order to ensure that the protective measures against electric shock through earth connections and bondings are effective. Two sub-functions are available:
R LOWΩ - Earth bond measurement according to EN 61557-4 (200 mA) and CONTINUITY - Continuous resistance measurement performed with 7 mA.
Test parameters for resistance measurement
Additional test parameter for Continuity sub-function
Additional key:

5.3.1 R LOWΩ, 200 mA resistance measurement

The resistance measurement is performed with automatic polarity reversal of the test voltage.
Test circuit for R LOWΩ measurement
Figure 5.9: Connection of 3-wire test lead plus optional extension lead
34
Page 35
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select R
LOWΩ
function using the function selector switch.
Set sub-function to R LOWΩ using UP / DOWN keys. Enable and set limit (optional). Connect test cable to the instrument. Compensate the test leads resistance (if necessary, see section 5.3.3 Compensation of
test leads resistance).
Disconnect from mains supply and discharge installation to be tested. Connect the test leads to the appropriate PE wiring (see Figure 5.9). Press the TEST key to perform the measurement. After the measurement is finished store the result by pressing the MEM button (MI 3100
SE only).
R LOWΩ measurement procedure
Figure 5.10: Example of RLOW result
Displayed result:
R .............. R LOWΩ resistance
R+ ............ result at positive polarity
R- ............. result at negative test polarity

5.3.2 Continuous resistance measurement with low current

In general, this function serves as standard -meter with a low testing current. The measurement is performed continuously without polarity reversal. The function can also be applied for testing continuity of inductive components.
Test circuit for continuous resistance measurement
Figure 5.11: Tip commander and 3-wire test lead applications
35
Page 36
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select R
LOW
function using the function selector switch.
Set sub-function to CONTINUITY using UP / DOWN keys. Enable and set the limit (optional). Connect test cable to the instrument. Compensate test leads resistance (if necessary, see section 5.3.3 Compensation of test
leads resistance).
Disconnect from mains supply and discharge the object to be tested. Connect test leads to the tested object (see Figure 5.11). Press the TEST key to begin performing a continuous measurement. Press the TEST key to stop measurement. After the measurement is finished store the result by pressing the MEM button (MI 3100
SE only).
Select R LOWΩ or CONTINUITY function. Connect test cable to the instrument and short the test leads together (see Figure
5.13).
Press the TEST key to perform resistance measurement. Press the CAL key to compensate leads resistance.
Continuous resistance measurement procedure
Figure 5.12: Example of continuous resistance measurement
Displayed result:
R .............. resistance

5.3.3 Compensation of test leads resistance

This chapter describes how to compensate the test leads resistance in both continuity functions,
R LOWΩ and CONTINUITY. Compensation is required to eliminate the influence of test leads
resistance and the internal resistances of the instrument on the measured resistance. The lead compensation is therefore a very important feature to obtain correct result.
symbol is displayed if the compensation was carried out successfully.
Circuits for compensating the resistance of test leads
Figure 5.13: Shorted test leads
Compensation of test leads resistance procedure
36
Page 37
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Figure 5.14: Results with old calibration values
Figure 5.15: Results with new calibration
values
Note:
The highest value for lead compensation is 5 . If the resistance is higher the
compensation value is set back to default value.
is displayed if no calibration value is stored.
37
Page 38
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.16: RCD tests
TEST
RCD sub-function test [Uc, RCDt, RCD I, AUTO]
IN
Rated RCD residual current sensitivity IN [10 mA, 30 mA, 100 mA, 300 mA, 500
mA, 1000 mA].
type
RCD type [AC, A, F]. starting polarity [ , , , ].
Characteristic and PRCD selection [selective ,general non-delayed , PRCD,
PRCD-K, PRCD-S].
MUL
Multiplication factor for test current [½, 1, 2, 5 I ].
Ulim
Conventional touch voltage limit [25 V, 50 V].

5.4 Testing RCDs

Various test and measurements are required for verification of RCD(s) in RCD protected installations. Measurements are based on the EN 61557-6 standard. The following measurements and tests (sub-functions) can be performed:
contact voltage, trip-out time, trip-out current and RCD auto-test.
Test parameters for RCD test and measurement
Notes:
Ulim can be selected in the Uc sub-function only. Selective (time-delayed) RCDs have delayed response characteristics. As the contact
voltage pre-test or other RCD tests influence the time delayed RCD it takes a certain period to recover into normal state. Therefore a time delay of 30 s is inserted before performing trip-out test by default.
Portable RCDs (PRCD, PRCD-K and PRCD-S) are tested as general (non-delayed)
RCDs. Trip-out times, trip-out currents and contact voltage limits are equal to limits of general (non-delayed) RCDs.
38
Page 39
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the RCD function using the function selector switch. Set sub-function to Uc using UP / DOWN keys. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the item to be tested (see Figure 5.17). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).
RCD type
Contact voltage Uc
proportional to
Rated I
AC
1.05IN
any
AC
21.05IN
A, F
1.41.05IN
30 mA
A, F
21.41.05IN
A, F
21.05IN
< 30 mA
A, F
221.05IN
Connections for testing RCD
Figure 5.17: Connecting the plug commander and the 3-wire test lead

5.4.1 Contact voltage (RCD Uc)

A current flowing into the PE terminal causes a voltage drop on earth resistance, i.e. voltage difference between PE equipotential bonding circuit and earth. This voltage difference is called contact voltage and is present on all accessible conductive parts connected to the PE. It shall always be lower than the conventional safety limit voltage. The contact voltage is measured with a test current lower than ½ I to avoid trip-out of the RCD and then normalized to the rated IN.
Contact voltage measurement procedure
The contact voltage result relates to the rated nominal residual current of the RCD and is multiplied by an appropriate factor (depending on RCD type and type of test current). The 1.05 factor is applied to avoid negative tolerance of result. See Table 5.1 for detailed contact voltage calculation factors.
Table 5.1: Relationship between Uc and I
39
N
Page 40
MI 3100 s, MI 3100 SE EurotestEASI Measurements
N
C
L
I
U
R
Select the RCD function using the function selector switch. Set sub-function to RCDt using UP / DOWN keys. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the item to be tested (see Figure 5.17). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).
RCD type
Slope range
Waveform
Notes
Start value
End value
AC
0.2IN
1.1IN
Sine
All models
A, F (IN 30 mA)
0.2IN
1.5IN
Pulsed
A, F (IN = 10 mA)
0.2IN
2.2IN
Loop resistance is indicative and calculated from Uc result (without additional proportional factors) according to:
.
Figure 5.18: Example of contact voltage measurement results
Displayed results:
Uc ........ contact voltage
Rl ......... fault loop resistance

5.4.2 Trip-out time (RCDt)

Trip-out time measurement verifies the sensitivity of the RCD at different residual currents.
Trip-out time measurement procedure
Figure 5.19: Example of trip-out time measurement results
Displayed results:
t ............ trip-out time
Uc ........ contact voltage for rated IN

5.4.3 Trip-out current (RCD I)

A continuously rising residual current is intended for testing the threshold sensitivity for RCD trip-out. The instrument increases the test current in small steps through appropriate range as follows:
40
Page 41
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the RCD function using the function selector switch. Set sub-function to RCD I using UP / DOWN keys. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the item to be tested (see Figure 5.17). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).
Maximum test current is I (trip-out current) or end value in case the RCD didn’t trip-out.
Trip-out current measurement procedure
Figure 5.20: Trip-out current measurement result example
Displayed results:
I ............ trip-out current
Uci ....... contact voltage at trip-out current I or end value in case the RCD didn’t trip
t ............ trip-out time
41
Page 42
MI 3100 s, MI 3100 SE EurotestEASI Measurements
HELP
Click
Toggles between top and bottom part of results field.
Keep pressed for 1s
Enters Help screen
RCD Auto-test steps
Notes
Select the RCD function using the function selector
switch.
Set sub-function to AUTO using UP / DOWN keys.
Set test parameters (if necessary).
Connect test cable to the instrument.
Connect test leads to the item to be tested (see Figure
5.17).
Press the TEST key to perform the test.
Start of test
Test with I
N
, 0 (step 1).
RCD should trip-out
Re-activate RCD.
Test with I
N
, 180 (step 2).
RCD should trip-out
Re-activate RCD.
Test with 5I
N
, 0 (step 3).
RCD should trip-out
Re-activate RCD.
Test with 5I
N
, 180 (step 4).
RCD should trip-out
Re-activate RCD.
Test with ½I
N
, 0 (step 5).
RCD should not trip­out
Test with ½I
N
, 180 (step 6).
RCD should not trip­out
Trip-out current test, 0 (step 7).
RCD should trip-out
Re-activate RCD.
Trip-out current test, 180 (step 8).
RCD should trip-out
Re-activate RCD.
Store the result by pressing the MEM key
(MI 3100 SE only).
End of test
Step 1
Step 2

5.4.4 RCD Auto-test

RCD auto-test function is intended to perform a complete RCD test (trip-out time at different residual currents, trip-out current and contact voltage) in one set of automatic tests, guided by the instrument.
Additional key:
RCD Auto-test procedure
Result examples:
42
Page 43
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Figure 5.21: Individual steps in RCD auto-test
Top
Bottom
Figure 5.22: Two parts of result field in RCD auto-test
Displayed results:
x1 ......... step 1 trip-out time (I=IN, 0º)
x1 ......... step 2 trip-out time (I=IN, 180º)
x5 ......... step 3 trip-out time (I=5IN, 0º)
x5 ......... step 4 trip-out time (I=5IN, 180º)
........ step 5 trip-out time (I=½IN, 0º)
........ step 6 trip-out time (I=½IN, 180º)
I .......... step 7 trip-out current (0º)
I .......... step 8 trip-out current (180º)
Uc ........ contact voltage for rated IN
Notes:
The auto-test sequence is immediately stopped if any incorrect condition is detected,
e.g. excessive Uc or trip-out time out of bounds.
Auto test is finished without x5 tests in case of testing the RCD types A and F with rated
residual currents of IN = 300 mA, 500 mA, and 1000 mA. In this case auto test result passes if all other results pass, and indications for x5 are omitted.
Tests for sensitivity (I , steps 7 and 8) are omitted for selective type RCD.
43
Page 44
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.23: Fault loop impedance
Test
Selection of fault loop impedance sub-function [Zloop, Zs rcd]
Fuse type
Selection of fuse type [---, NV, gG, B, C, K, D]
Fuse I
Rated current of selected fuse
Fuse T
Maximum breaking time of selected fuse
Lim
Minimum short circuit current for selected fuse

5.5 Fault loop impedance and prospective fault current

Fault loop is a loop comprised by mains source, line wiring and PE path to the mains source. The instrument measures the impedance of the loop and calculates the short circuit current. The measurement is covered by requirements of the EN 61557-3 standard.
Test parameters for fault loop impedance measurement
See Appendix A for reference fuse data.
Circuits for measurement of fault loop impedance
Figure 5.24: Connection of plug commander and 3-wire test lead
44
Page 45
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the Zloop function using function selector switch. Set sub-function to Zloop or Zs rcd using UP / DOWN keys. Select test parameters (optional). Connect test cable to the instrument. Connect test leads to the item to be tested (see Figure 5.17 and Figure 5.24). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).
Z
kUn
I
SC
SC
Un
Input voltage range (L-PE)
110 V
(93 V U
L-PE
134 V)
230 V
(185 V U
L-PE
266 V)
Fault loop impedance measurement procedure
Figure 5.25: Example of loop impedance measurement result
Displayed results:
Z ............... fault loop impedance
Isc ............ prospective fault current,
Lim ........... low limit prospective short-circuit current value
Prospective fault current ISC is calculated from measured impedance as follows:
where:
Un ......... Nominal U
voltage (see table below),
L-PE
ksc ........ Correction factor for Isc (see chapter 4.3.5 Isc factor).
Notes:
High fluctuations of mains voltage can influence the measurement results (the noise sign
is displayed in the message field). In this case it is recommended to repeat few
measurements to check if the readings are stable.
This measurement will trip-out the RCD in RCD-protected electrical installation if test Zloop is selected.
Select Zs rcd measurement to prevent trip-out of RCD in RCD protected installation.
45
Page 46
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.26: Line impedance
Figure 5.27: Voltage drop
Test
Selection of line impedance [Zline] or voltage drop [ΔU] sub-function
FUSE type
Selection of fuse type [---, NV, gG, B, C, K, D]
FUSE I
Rated current of selected fuse
FUSE T
Maximum breaking time of selected fuse
Lim
Minimum short circuit current for selected fuse.
ΔU
MAX
Maximum voltage drop [3.0 % ÷ 9.0 %].
HELP / CAL
Click
Measures Zref value for U function.
Keep pressed for 1s
Enters Help screen.

5.6 Line impedance and prospective short-circuit current / Voltage drop

Line impedance is measured in loop comprising of mains voltage source and line wiring. Line impedance is covered by the requirements of the EN 61557-3 standard. The Voltage drop sub-function is intended to check that a voltage in the installation stays above acceptable levels if the highest current is flowing in the circuit. The highest current is defined as the nominal current of the circuit's fuse. The limit values are described in the standard EN 60364-5-52.
Sub-functions:
Z LINE - Line impedance measurement according to EN 61557-3 and ΔU – Voltage drop measurement.
Test parameters for line impedance measurement
See Appendix A for reference fuse data. Additional test parameters for voltage drop measurement
Additional key:
46
Page 47
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the Zline function using the function selector switch. Set sub-function to Zline using UP / DOWN keys. Select test parameters (optional). Connect test cable to the instrument. Connect test leads to the item to be tested (see Figure 5.28). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).
Z
kUn
I
SC
SC
Un
Input voltage range (L-N or L1-L2)
110 V
(93 V U
L-N
134 V)
230 V
(185 V U
L-N
266 V)
400 V
(321 V U
L-L
485 V)

5.6.1 Line impedance and prospective short circuit current

Circuits for measurement of line impedance
Figure 5.28: Phase-neutral or phase-phase line impedance measurement – connection of plug
commander and 3-wire test lead
Line impedance measurement procedure
Figure 5.29: Examples of line impedance measurement result
Displayed results:
Z ............... line impedance
Isc ............ prospective short-circuit current
Lim ........... low limit prospective short-circuit current value
Prospective short circuit current is calculated as follows:
where:
Un ......... Nominal L-N or L1-L2 voltage (see table below),
ksc ........ Correction factor for Isc (see chapter 4.3.5 Isc factor).
47
Page 48
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the Zline function using the function selector switch. Set sub-function to ΔU using UP / DOWN keys. Select test parameters (optional). Connect test cable to the instrument. Connect the test leads to the origin of electrical installation (see Figure 5.30). Press the CAL key to perform the measurement.
Set sub-function to ΔU using UP / DOWN keys. Select test parameters (Fuse type must be selected). Connect test cable or plug commander to the instrument. Connect the test leads to the tested points (see Figure 5.30). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).
Note:
High fluctuations of mains voltage can influence the measurement results (the noise sign
is displayed in the message field). In this case it is recommended to repeat few
measurements to check if the readings are stable.

5.6.2 Voltage drop

The voltage drop is calculated based on the difference of line impedance at connection points (sockets) and the line impedance at the reference point (usually the impedance at the switchboard).
Circuits for measurement of voltage drop
Figure 5.30: Phase-neutral or phase-phase voltage drop measurement – connection of plug
commander and 3-wire test lead
Voltage drop measurement procedure Step 1: Measuring the impedance Zref at origin
Step 2: Measuring the voltage drop
48
Page 49
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Step 1 - Zref
Step 2 - Voltage drop
 
100
)(
%
N
NREF
U
IZZ
U
Un
Input voltage range (L-N or L1-L2)
110 V
(93 V U
L-N
134 V)
230 V
(185 V U
L-N
266 V)
400 V
(321 V U
L-L
485 V)
Figure 5.31: Examples of voltage drop measurement result
Displayed results:
ΔU ............ voltage drop
Isc ............ prospective short-circuit current
Z ............... line impedance at measured point
Zref .......... reference impedance
Voltage drop is calculated as follows:
where:
ΔU ........ calculated voltage drop
Z ........... impedance at test point
Z
...... impedance at reference point
REF
IN .......... rated current of selected fuse
UN ......... nominal voltage (see table below)
Notes:
If the reference impedance is not set the value of ZThe Z
is cleared (set to 0.00 Ω) if pressing CAL key while instrument is not connected
REF
is considered as 0.00 Ω.
REF
to a voltage source.
I
is calculated as described in chapter 5.6.1 Line impedance and prospective short
SC
circuit current
If the measured voltage is outside the ranges described in the table above the ΔU result
will not be calculated.
High fluctuations of mains voltage can influence the measurement results (the noise sign
is displayed in the message field). In this case it is recommended to repeat few
measurements to check if the readings are stable.
49
Page 50
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.32: Earth resistance
Test
Test configuration [EARTH RE]
Limit
Maximum resistance [OFF, 1 ÷ 5 k]
Select EARTH function using the function selector switch. Enable and set limit value (optional). Connect test leads to the instrument. Connect the item to be tested (see Figure 5.33 and Figure 5.34). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (MI 3100 SE only).

5.7 Earth resistance

Earth resistance is one of the most important parameters for protection against electric shock. Main earthing arrangements, lightning systems, local earthings, soil resistivity etc can be verified with the earthing resistance test. The measurement conforms to the EN 61557-5 standard.
The Earth resistance main function is 3-wire earth resistance test method for standard earth resistance tests with two earthing rods.
Test parameters for earth resistance measurement
Earth resistance measurements procedure
50
Page 51
MI 3100 s, MI 3100 SE EurotestEASI Measurements

5.7.1 Standard earthing resistance measurement

Connections for earth resistance measurement
Figure 5.33: Resistance to earth, measurement of main installation earthing
Figure 5.34: Resistance to earth, measurement of a lighting protection system
Figure 5.35: Example of earth resistance measurement result
Displayed results for earth resistance measurement:
R .............. earth resistance
Rp ............ resistance of S (potential) probe
Rc ............ resistance of H (current) probe
51
Page 52
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Notes:
High resistance of S and H probes could influence the measurement results. In this
case, “Rp” and “Rc” warnings are displayed. There is no PASS / FAIL indication in this case.
High noise currents and voltages in earth could influence the measurement results. The
tester displays the warning in this case.
Probes must be placed at sufficient distance from the measured object.
52
Page 53
MI 3100 s, MI 3100 SE EurotestEASI Measurements

5.8 PE test terminal

It can happen that a dangerous voltage is applied to the PE wire or other accessible metal parts. This is a very dangerous situation since the PE wire and MPEs are considered to be earthed. An often reason for this fault is incorrect wiring (see examples below). When touching the TEST key in all functions that requires mains supply the user automatically performs this test.
Examples for application of PE test terminal
Figure 5.36: Reversed L and PE conductors (plug commander)
Figure 5.37: Reversed L and PE conductors (application of 3-wire test lead)
53
Page 54
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Connect test cable to the instrument. Connect test leads to the item to be tested ( see Figure 5.36 and Figure 5.37). Touch PE test probe (the TEST key) for at least one second. If PE terminal is connected to phase voltage the warning message is displayed,
instrument buzzer is activated, and further measurements are disabled in Zloop and RCD functions.
PE terminal test procedure
Warning:
If dangerous voltage is detected on the tested PE terminal, immediately stop all
measurements, find and remove the fault!
Notes:
PE test terminal is active in all test and measurements except in the VOLTAGE, Low
ohm, Earth and Insulation functions.
PE test terminal does not operate in case the operator’s body is completely insulated
from floor or walls!
For operation of PE test terminal on commanders refer to chapter Appendix D –
Commanders.
54
Page 55
MI 3100 s, MI 3100 SE EurotestEASI Measurements
See chapter 4.2 Function selection for instructions on key functionality.
Figure 5.38: PE conductor resistance
Test
Selection of PE conductor resistance sub-function [Rpe,Rpe(rcd)]
Lim
Maximum resistance [OFF, 0.1 ÷ 20.0 ].

5.9 PE conductor resistance (MI 3100 SE only)

In TN system instrument measures the resistance of the protection conductor from the power transformer to the measurement site. In TT system the resistance of protection conductor from mains outlet to earth electrode and back to the power transformer via soil and the transformers earthing system is measured.
Test parameters for PE conductor resistance measurement
Circuits for measurement of PE conductor resistance
Figure 5.39: Connection of plug commander and 3-wire test lead
55
Page 56
MI 3100 s, MI 3100 SE EurotestEASI Measurements
Select the Rpe function using the function selector switch. Set sub-function to Rpe or Rpe(rcd) using UP / DOWN keys. Select test parameters (optional). Connect test cable to the instrument. Connect test leads to the item to be tested (see Figure 5.39). Press the TEST key to perform the measurement. Store the result by pressing the MEM key (optional).
PE conductor resistance measurement procedure
Figure 5.40: Examples of PE conductor resistance measurement result
Displayed results:
R .............. PE conductor resistance
Notes:
High fluctuations of mains voltage can influence the measurement results (the noise sign
is displayed in the message field). In this case it is recommended to repeat few
measurements to check if the readings are stable.
This measurement will trip-out the RCD in RCD-protected electrical installation if test Rpe is selected.
Select Rpe(rcd) measurement to prevent trip-out of RCD in RCD protected installation.
56
Page 57
MI 3100 s, MI 3100 SE EurotestEASI Autosequences
See chapter 4.2 Function selection for instructions on key functionality.
Figure 6.1: Main auto-sequence menu
UP / DOWN
Selects auto-sequence.
TEST
Enters selected auto-test sequence.
Function selector switch
Exits back to selected test / measurement function.
Figure 6.2: Auto-sequence menus
Figure 6.3: Editing parameters menus

6 Auto-sequences (MI 3100 SE only)

Auto-sequences are intended to perform automatic executing of pre-defined measurement sequences. Sequences are divided into three groups, each for selected supply system:
AUTO TT, AUTO TN (RCD) and AUTO TN.
The selected sequence is carried out in one set of automatic tests, guided by the instrument.
Keys in main auto-sequence menu
57
Page 58
MI 3100 s, MI 3100 SE EurotestEASI Autosequences
Key
Auto-sequence menu
Editing parameters menu
TAB
Enters viewing/editing test parameters.
Selects the test parameter to be set or modified.
UP / DOWN
Sets or modifies test parameters.
TEST
Runs selected auto­sequence.
Runs selected auto-sequence.
HELP / CAL Click
Switch between screens.
Measuring reference line impedance (when ZREF is selected).
HELP / CAL Keep pressed for 1 s
Enters help screens.
Enters help screens.
MEM
Stores auto-test results.
ESC
Exits back to previous menu.
Exits back to previous menu with saving changes.
Auto-sequence
Test /
measurement
Available editable parameters
AUTO TT
VOLTAGE Z LINE U* Zs rcd Uc
FUSE
fuse type, rated current, maximal braking time, minimal short-circuit current
ZREF
reference line impedance
U
voltage drop limit value
RCD
nominal current, RCD type, maximal contact voltage
AUTO TN (RCD)
VOLTAGE Z LINE U* Zs rcd Rpe(rcd)
FUSE
fuse type, rated current, maximal braking time, minimal short-circuit current
ZREF
reference line impedance
U
voltage drop limit value
RPE
maximal PE line resistance
AUTO TN
VOLTAGE Z LINE U* Z LOOP Rpe
FUSE
fuse type, rated current, maximal braking time, minimal short-circuit current
ZREF
reference line impedance
U
voltage drop limit value
RPE
maximal PE line resistance
Keys in auto-sequence and editing parameters menu
The following tests / measurements can be performed for selected autosequence. Parameters in each autosequence are user-defined as follows.
* applicable only if Z
REF
is set
58
Page 59
MI 3100 s, MI 3100 SE EurotestEASI Autosequences
Select the AUTO SEQUENCES function using the function selector switch. Select auto-sequence AUTO TT, AUTO TN (rcd), or AUTO TN. Select test parameters. Connect test cable to the instrument. Connect the test leads to the origin of electrical installation (see Figure 6.4 – step
1)(optional).
Press the CAL key to perform the Z
REF
measurement (optional).
Connect test leads to the item to be tested (see Figure 6.4step 2). Press the TEST key to start the auto-sequence. Store the result by pressing the MEM key (optional).
Step 1
Step 2
Step 3
Figure 6.5: Individual steps of AUTO TT auto-sequence
Figure 6.6: Example of AUTO TT auto-sequence results
Circuit for automatic measurement
Figure 6.4: Auto-sequence setup
Automatic measurement procedure
Displayed results during auto-sequence and saved results Voltage
Displayed results for single phase system:
Uln ........... voltage between phase and neutral conductors
Ulpe ......... voltage between phase and protective conductors
Unpe ........ voltage between neutral and protective conductors
f ................ frequency
Line impedance
Z ............... line impedance
Isc ............ prospective short-circuit current
59
Page 60
MI 3100 s, MI 3100 SE EurotestEASI Autosequences
Figure 6.7: Example of recalled AUTO TN auto-sequence results
Function
Results field
Left value on display
Right value on display
U
Voltage
Voltage between phase and neutral
Zln
Line impedance
Line impedance
Prospective short-circuit current
U*
Voltage drop
Voltage drop (if available)
Zs
Loop impedance
Loop impedance
Contact voltage (AUTO TT only)or Prospective fault current (AUTO TT excepted)
Zlp
Loop impedance
Loop impedance
Prospective fault current
Rpe
PE conductor resistance
PE conductor resistance
Lim ........... low limit prospective short-circuit current value
Loop impedance (Zs or Zs
RCD
)
Z ............... loop impedance
Isc ............ prospective fault current
Lim ........... low limit prospective fault current value
PE conductor resistance (Rpe or Rpe
RCD
)
R .............. PE conductor resistance
Displayed results once auto-sequence finished and recalled results:
Notes:
Before starting the auto-sequence, all settings of parameters should be checked. U measurement in each auto-sequence is enabled only if Z
60
REF
is set.
Page 61
MI 3100 s, MI 3100 SE EurotestEASI Data handling
Memory operation menu
Data structure field
1st level:
OBJECT: Default location name (object and its successive
number).
001: No. of selected element.
2nd level:
BLOCK: Default location name (block and its successive
number).
002: No. of selected element.
3rd level:
FUSE: Default location name (fuse and its successive
number).
003: No. of selected element.

7 Data handling (MI 3100 SE only)

7.1 Memory organization

Measurement results together with all relevant parameters can be stored in the instrument’s memory. After the measurement is completed, results can be stored to the flash memory of the instrument, together with the sub-results and function parameters.

7.2 Data structure

The instrument’s memory place is divided into 4 levels each containing 199 locations. The number of measurements that can be stored into one location is not limited.
The data structure field describes the location of the measurement (which object, block, fuse and connection) and where can be accessed. In the measurement field there is information about type and number of measurements that belong to the selected structure element (object and block and fuse and connection). The main advantages of this system are:
Test results can be organized and grouped in a structured manner that reflects the
structure of typical electrical installations.
Customized names of data structure elements can be uploaded from EurolinkPRO
PCSW.
Simple browsing through structure and results. Test reports can be created with no or little modifications after downloading results to a
PC.
Data structure field
Figure 7.1: Data structure and measurement fields
61
Page 62
MI 3100 s, MI 3100 SE EurotestEASI Data handling
4th level:
CONNECTION: Default location name (connection and its
successive number).
004: No. of selected element.
No. of measurements in selected location [No. of measurements in selected location and its sub­locations]
Type of stored measurement in the selected location.
No. of selected test result / No. of all stored test results in selected location.
Measurement field
62
Page 63
MI 3100 s, MI 3100 SE EurotestEASI Data handling
Figure 7.2: Save test menu
Memory available for storing results.
TAB
Selects the location element (Object / Block / Fuse / Connection)
UP / DOWN
Selects number of selected location element (1 to 199)
MEM
Saves test results to the selected location and returns to the measuring function.
ESC / TEST
Exits back to measuring function without saving.
Function selector switch
Switches to other test / measuring function or settings menu without saving.

7.3 Storing test results

After the completion of a test the results and parameters are ready for storing ( icon is displayed in the information field). By pressing the MEM key, the user can store the results.
Keys in save test menu - data structure field
Notes:
The instrument offers to store the result to the last selected location by default. If the measurement is to be stored to the same location as the previous one just press
the MEM key twice.
63
Page 64
MI 3100 s, MI 3100 SE EurotestEASI Data handling
Figure 7.3: Recall menu - installation structure
field selected
Figure 7.4: Recall menu - measurements field
selected
TAB
Selects the location element (Object / Block / Fuse / Connection).
UP / DOWN
Selects number of selected location element (1 to 199).
ESC
Exits back to measuring function or memory menu.
Function selector switch
Switches to other test / measuring function or settings.
TEST / MEM
Enters measurements field.
UP / DOWN
Selects the stored measurement.
TAB / ESC
Returns to installation structure field.
Function selector switch
Switches to other test / measuring function or settings.
TEST / MEM
View selected measurement results.
Figure 7.5: Example of recalled measurement result
UP / DOWN
Displays measurement results stored in selected location.
MEM / ESC
Returns to measurements field.
TEST
Returns to installation structure field.
Function selector switch
Switches to other test / measuring function or settings.

7.4 Recalling test results

Press the MEM key in every measuring function when there is no result available for storing or select MEMORY in the SETTINGS menu.
Keys in recall memory menu (installation structure field selected):
Keys in recall memory menu (measurements field selected):
Keys in recall memory menu (measurement results are displayed)
64
Page 65
MI 3100 s, MI 3100 SE EurotestEASI Data handling
TEST
Confirms clearing of complete memory content (YES must be selected with UP / DOWN keys).
ESC
Exits back to memory menu without changes.
Function selector switch
Switches to test / measuring function without changes.
TAB
Selects the location element (Object / Block / Fuse / Connection).
UP / DOWN
Selects number of selected location element (1 to 199)
Function selector switch
Switches to test / measuring function.
ESC
Exits back to memory menu.
TEST
Enters dialog box for deleting all measurements in selected location and its sub-locations.
TEST
Deletes all results in selected location.

7.5 Clearing stored data

7.5.1 Clearing complete memory content

Select CLEAR ALL MEMORY in MEMORY menu. A warning will be displayed.
Figure 7.6: Clear all memory
Keys in clear all memory menu
Figure 7.7: Clearing memory in progress

7.5.2 Clearing measurement(s) in selected location

Select DELETE RESULTSin MEMORY menu.
Figure 7.8: Clear measurements menu (data structure field selected)
Keys in delete results menu (installation structure field selected):
Keys in dialog for confirmation to clear results in selected location:
65
Page 66
MI 3100 s, MI 3100 SE EurotestEASI Data handling
MEM / ESC
Exits back to delete results menu without changes.
Function selector switch
Switches to test / measuring function without changes.
TAB
Selects the location element (Object / Block / Fuse / Connection).
UP / DOWN
Selects number of selected location element (1 to 199)
Function selector switch
Switches to test / measuring function.
ESC
Exits back to memory menu.
MEM
Enters measurements field for deleting individual measurements.
UP / DOWN
Selects measurement.
TEST
Opens dialog box for confirmation to clear selected measurement.
TAB / ESC
Returns to installation structure field.
Function selector switch
Switches to test / measuring function.
TEST
Deletes selected measurement result.
MEM / TAB / ESC
Exits back to measurements field without changes.
Function selector switch
Switches to test / measuring function without changes.
Figure 7.10: Dialog for confirmation
Figure 7.11: Display after measurement was
cleared

7.5.3 Clearing individual measurements

Select DELETE RESULTS in MEMORY menu.
Figure 7.9: Menu for clearing individual measurement (installation structure field selected)
Keys in delete results menu (installation structure field selected):
Keys in delete results menu (measurements field selected):
Keys in dialog for confirmation to clear selected result(s):
66
Page 67
MI 3100 s, MI 3100 SE EurotestEASI Data handling
Figure 7.12: Example of menu with customized installation structure names
Figure 7.13: Connection of the barcode reader and RFID reader
Connect the barcode reader or RFID reader to the instrument. In Save menu select memory location to be renamed. A new location name (scanned from a barcode label or a RFID tag) will be accepted by
the instrument. A successful receive of the barcode or RFID tag is confirmed by two short confirmation beeps.

7.5.4 Renaming installation structure elements (upload from PC)

Default installation structure elements are »Object«, »Block«, »Fuse« and »Connection«. In the PCSW package Eurolink-PRO default names can be changed with customized names that corresponds the installation under test. Refer to PCSW Eurolink-PRO HELP for information how to upload customized installation names to the instrument.
7.5.5 Renaming installation structure elements with serial barcode
reader or RFID reader
Default installation structure elements are »Object«, »Block«, »Fuse« and »Connection«. When the instrument is in the SAVE RESULTS menu location ID can be scanned from a barcode label with the barcode reader or can be read from a RFID tag with the RFID reader.
How to change the name of memory location
Note:
Use only barcode readers and RFID readers delivered by METREL or authorized
distributor.
67
Page 68
MI 3100 s, MI 3100 SE EurotestEASI Data handling
RS-232 communication: connect a PC COM port to the instrument PS/2 connector using
the PS/2 - RS232 serial communication cable;
USB communication: connect a PC USB port to the instrument USB connector using the
USB interface cable.
Switch on the PC and the instrument. Run the EurolinkPRO program. The PC and the instrument will automatically recognize each other. The instrument is prepared to communicate with the PC.

7.6 Communication

Stored results can be transferred to a PC. A special communication program on the PC automatically identifies the instrument and enables data transfer between the instrument and the PC. There are three communication interfaces available on the instrument: USB, RS 232 and Bluetooth.

7.6.1 USB and RS232 communication

The instrument automatically selects the communication mode according to detected interface. USB interface has priority.
Figure 7.14: Interface connection for data transfer over PC COM port
How to establish an USB or RS-232 link:
The program EurolinkPRO is a PC software running on Windows XP, Windows Vista, Windows 7, and Windows 8. Read the file README_EuroLink.txt on CD for instructions about installing and running the program.
Note:
USB drivers should be installed on PC before using the USB interface. Refer to USB
installation instructions available on installation CD.
68
Page 69
MI 3100 s, MI 3100 SE EurotestEASI Data handling
Switch Off and On the instrument. Be sure that the Bluetooth dongle A1436 is properly initialized. If not the Bluetooth
dongle must be initialized as described in chapter 4.3.7 Communication (MI 3100 SE only).
On PC configure a Standard Serial Port to enable communication over Bluetooth link
between instrument and PC. Usually no code for pairing the devices is needed.
Run the EurolinkPRO program. The PC and the instrument will automatically recognize each other. The instrument is prepared to communicate with the PC.
Switch Off and On the instrument. Be sure that the Bluetooth dongle A1436 is properly initialized. If not the Bluetooth
dongle must be initialized as described in chapter 4.3.7 Communication (MI 3100 SE only).
Some Android applications automatically carry out the setup of a Bluetooth connection. It
is preferred to use this option if it exists. This option is supported by Metrel's Android applications.
If this option is not supported by the selected Android application then configure a
Bluetooth link via Android device’s Bluetooth configuration tool. Usually no code for pairing the devices is needed.
The instrument and Android device are ready to communicate.

7.6.2 Bluetooth communication

How to configure a Bluetooth link between instrument and PC
For Bluetooth communication with PC a Standard Serial Port over Bluetooth link for Bluetooth dongle A 1436 must be configured first.
How to configure a Bluetooth link between instrument and Android device
Notes:
Sometimes there will be a demand from the PC or Android device to enter the code.
Enter code ‘NNNN’ to correctly configure the Bluetooth link.
The name of correctly configured Bluetooth device must consist of the instrument type
plus serial number, eg. MI 3100SE-12240429D. If the Bluetooth dongle got another name, the configuration must be repeated.
69
Page 70
MI 3100 s, MI 3100 SE EurotestEASI Upgrading the instrument

8 Upgrading the instrument

The instrument can be upgraded from a PC via the RS232 communication port. This enables to keep the instrument up to date even if the standards or regulations change. The upgrade can be carried with a help of special upgrading software and the communication cable as shown on
Figure 7.14. Please contact your dealer for more information.
70
Page 71
MI 3100 s, MI 3100 SE EurotestEASI Maintenance

9 Maintenance

Unauthorized persons are not allowed to open the Eurotest instrument. There are no user replaceable components inside the instrument, except the battery and fuses under rear cover.

9.1 Fuse replacement

There are three fuses under back cover of the Eurotest instrument.
F1
M 0.315 A / 250 V, 205 mm This fuse protects internal circuitry for continuity functions if test probes are connected to the mains supply voltage by mistake during measurement.
F2, F3
F 4 A / 500 V, 326.3 mm (breaking capacity: 50 kA) General input protection fuses of test terminals L/L1 and N/L2.
Position of fuses can be seen in Figure 3.4 Battery and fuse compartment in chapter 3.3 Back side.
Warnings:
Disconnect all measuring accessory and switch off the instrument before opening battery
/ fuse compartment cover, hazardous voltage inside!
Replace blown fuse with original type only, otherwise the instrument or accessory may
be damaged and/or operator’s safety impaired!

9.2 Cleaning

No special maintenance is required for the housing. To clean the surface of the instrument or accessory use a soft cloth slightly moistened with soapy water or alcohol. Then leave the instrument or accessory to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons! Do not spill cleaning liquid over the instrument!

9.3 Periodic calibration

It is essential that the test instrument is regularly calibrated in order that the technical specification listed in this manual is guaranteed. We recommend an annual calibration. Only an authorized technical person can do the calibration. Please contact your dealer for further information.

9.4 Service

For repairs under warranty, or at any other time, please contact your distributor.
71
Page 72
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range (M)
Resolution (M)
Accuracy
0.00 19.99
0.01
(5 % of reading + 3 digits)
20.0 99.9
0.1
(10 % of reading)
100.0 199.9
(20 % of reading)
Measuring range (M)
Resolution (M)
Accuracy
0.00 19.99
0.01
(5 % of reading + 3 digits)
20.0 199.9
0.1
(5 % of reading)
200 999
1
(10 % of reading)
Measuring range (M)
Resolution (M)
Accuracy
0.00 19.99
0.01
(5 % of reading + 3 digits)
20.0 199.9
0.1
(5 % of reading)
200 999
1
(20 % of reading)
Measuring range (V)
Resolution (V)
Accuracy
0 1200
1
(3 % of reading + 3 digits)

10 Technical specifications

10.1 Insulation resistance

Insulation resistance (nominal voltages 50 VDC, 100 V
Measuring range according to EN 61557 is 0.15 M 199.9 M.
Insulation resistance (nominal voltage 500 VDC)
Measuring range according to EN 61557 is 0.15 M 999 M.
Insulation resistance (nominal voltage 1000 VDC)
Measuring range according to EN 61557 is 0.15 M 999 M.
Voltage
and 250 VDC)
DC
Nominal voltages .................................... 50 VDC, 100 VDC, 250 VDC, 500 VDC, 1000 VDC
Open circuit voltage ................................ -0 % / +20 % of nominal voltage
Measuring current ................................... min. 1 mA at RN=UN1 k/V
Short circuit current ............................... max. 3 mA
The number of possible tests ................. > 1200, with a fully charged battery
Auto discharge after test. Specified accuracy is valid if 3-wire test lead is used while it is valid up to 100 M if tip commander is used. Specified accuracy is valid up to 100 M if relative humidity > 85 %. In case the instrument gets moistened, the results could be impaired. In such case, it is recommended to dry the instrument and accessories for at least 24 hours. The error in operating conditions could be at most the error for reference conditions (specified in the manual for each function) 5 % of measured value.
72
Page 73
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range R ()
Resolution ()
Accuracy
0.00 19.99
0.01
(3 % of reading + 3 digits)
20.0 199.9
0.1
(5 % of reading)
200 1999
1
Measuring range R+, R- ()
Resolution ()
Accuracy
0.0 199.9
0.1
(5 % of reading + 5 digits)
200 1999
1
Measuring range ()
Resolution ()
Accuracy
0.0 19.9
0.1
(5 % of reading + 3 digits)
20 1999
1

10.2 Continuity

10.2.1 Resistance R LOW

Measuring range according to EN 61557 is 0.16  1999 .
Open-circuit voltage ................................ 6.5 VDC 9 VDC
Measuring current ................................... min. 200 mA into load resistance of 2
Test lead compensation .......................... up to 5
The number of possible tests .................. > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.

10.2.2 Resistance CONTINUITY

Open-circuit voltage ................................ 6.5 VDC 9 VDC
Short-circuit current ................................ max. 8.5 mA
Test lead compensation .......................... up to 5
73
Page 74
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
IN × 1/2
IN × 1
IN × 2
IN × 5
RCD I
IN (mA)
AC
A, F
AC
A, F
AC
A, F
AC
A, F
AC
A, F
10
5
3.5
10
20
20
40
50
100
30
15
10.5
30
42
60
84
150
212
100
50
35
100
141
200
282
500
707
300
150
105
300
424
600
848
1500
n.a.
500
250
175
500
707
1000
1410
2500
n.a.
1000
500
350
1000
1410
2000
n.a.
n.a.
n.a.
Measuring range (V)
Resolution (V)
Accuracy
0.0 19.9
0.1
(-0 % / +15 %) of reading ± 10 digits
20.0 99.9
0.1
(-0 % / +15 %) of reading

10.3 RCD testing

10.3.1 General data

Nominal residual current (A,AC) ............. 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
Nominal residual current accuracy .......... -0 / +0.1I; I = IN, 2IN, 5IN
-0.1I / +0; I = 0.5IN
AS/NZS selected: ± 5 %
Test current shape .................................. Sine-wave (AC), pulsed (A, F)
DC offset for pulsed test current ............. 6 mA (typical)
RCD type ................................................ non-delayed, S (time-delayed), PRCD, PRCD-K,
PRCD-S
Test current starting polarity .................. 0º or 180º
Voltage range ......................................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz)
n.a. ......................................................... not applicable
AC type................................................... sine wave test current
A, F types…… ........................................ pulsed current

10.3.2 Contact voltage RCD Uc

Measuring range according to EN 61557 is 20.0 V 31.0V for limit contact voltage 25V Measuring range according to EN 61557 is 20.0 V 62.0V for limit contact voltage 50V
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free of interfering voltages.
Test current ........................................... max. 0.5IN
Limit contact voltage .............................. 25 V, 50 V
Specified accuracy is valid for complete operating range.
74
Page 75
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range (ms)
Resolution (ms)
Accuracy
0.0 40.0
0.1
1 ms
0.0 max. time *
0.1
3 ms
Measuring range I
Resolution I
Accuracy
0.2IN 1.1IN (AC type)
0.05IN
0.1IN
0.2IN 1.5IN (A type, IN≥30 mA)
0.05IN
0.1IN
0.2IN 2.2IN (A type, IN<30 mA)
0.05IN
0.1IN
0.2IN 2.2IN (B type)
0.05IN
0.1IN
Measuring range (ms)
Resolution (ms)
Accuracy
0 300
1
3 ms
Measuring range (V)
Resolution (V)
Accuracy
0.0 19.9
0.1
(-0 % / +15 %) of reading 10 digits
20.0 99.9
0.1
(-0 % / +15 %) of reading

10.3.3 Trip-out time

Complete measurement range corresponds to EN 61557 requirements. Maximum measuring times set according to selected reference for RCD testing.
* For max. time see normative references in chapter 4.3.4 RCD testing. This specification applies to max. time > 40 ms.
Test current ........................................... ½IN, IN, 2IN, 5I
5IN is not available for IN=1000 mA (RCD type AC) or I
N
300 mA (RCD types A, F).
N
2IN is not available for IN=1000 mA (RCD types A, F) Specified accuracy is valid for complete operating range.

10.3.4 Trip-out current

Trip-out current
Complete measurement range corresponds to EN 61557 requirements.
Trip-out time
Contact voltage
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free of interfering voltages. Specified accuracy is valid for complete operating range.
75
Page 76
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range ()
Resolution ()
Accuracy
0.00 9.99
0.01
(5 % of reading + 5 digits)
10.0 99.9
0.1
100 999
1
10 % of reading
1.00 k 9.99 k
10
Measuring range (A)
Resolution (A)
Accuracy
0.00 9.99
0.01
Consider accuracy of fault
loop resistance measurement
10.0 99.9
0.1
100 999
1
1.00 k 9.99 k
10
10.0 k 23.0 k
100
Measuring range ()
Resolution ()
Accuracy
0.00 9.99
0.01
(5 % of reading + 10 digits)
10.0 99.9
0.1
100 999
1
10 % of reading
1.00 k 9.99 k
10
Measuring range (A)
Resolution (A)
Accuracy
0.00 9.99
0.01
Consider accuracy of fault
loop resistance measurement
10.0 99.9
0.1
100 999
1
1.00 k 9.99 k
10
10.0 k 23.0 k
100

10.4 Fault loop impedance and prospective fault current

10.4.1 No disconnecting device or FUSE selected

Fault loop impedance
Measuring range according to EN 61557 is 0.25  9.99k.
Prospective fault current (calculated value)
The accuracy is valid if mains voltage is stabile during the measurement.
Test current (at 230 V) ........................... 6.5 A (10 ms)
Nominal voltage range ........................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz)

10.4.2 RCD selected

Fault loop impedance
Measuring range according to EN 61557 is 0.46  9.99 k.
Accuracy may be impaired in case of heavy noise on mains voltage.
Prospective fault current (calculated value)
Nominal voltage range ........................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz) No trip out of RCD.
76
Page 77
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range ()
Resolution ()
Accuracy
0.00 9.99
0.01
(5 % of reading + 5 digits)
10.0 99.9
0.1
100 999
1
10 % of reading
1.00 k 9.99 k
10
Measuring range (A)
Resolution (A)
Accuracy
0.00 0.99
0.01
Consider accuracy of line
resistance measurement
1.0 99.9
0.1
100 999
1
1.00 k 99.99 k
10
100 k 199 k
1000
Measuring range (%)
Resolution (%)
Accuracy
0.0 99.9
0.1
Consider accuracy of line
impedance measurement(s)*
10.5 Line impedance and prospective short-circuit current /
Voltage drop
Line impedance
Measuring range according to EN 61557 is 0.25  9.99k.
Prospective short-circuit current (calculated value)
Test current (at 230 V) ........................... 6.5 A (10 ms)
Nominal voltage range ........................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz)
321 V 485 V (45 Hz 65 Hz)
Voltage drop (calculated value)
Z
measuring range ..................................... 0.00 Ω 20.0 Ω
REF
*See chapter 5.6.2 Voltage drop for more information about calculation of voltage drop result.
77
Page 78
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range ()
Resolution ()
Accuracy
0.00 19.99
0.01
(5 % of reading + 5 digits)
20.0 99.9
0.1
100.0199.9
0,1
10 % of reading
200 1999
1
Measuring range ()
Resolution ()
Accuracy
0.00 19.99
0.01
(5 % of reading + 10 digits)
20.0 99.9
0.1
100.0 199.9
0,1
10 % of reading
200 1999
1

10.6 PE conductor resistance (MI 3100 SE only)

10.6.1 No RCD selected

PE conductor resistance

10.6.2 RCD selected

PE conductor resistance
Accuracy may be impaired in case of heavy noise on mains voltage.
Nominal voltage range ........................... 93 V 134 V (45 Hz 65 Hz)
185 V 266 V (45 Hz 65 Hz) No trip out of RCD.
78
Page 79
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range ()
Resolution ()
Accuracy
0.00 19.99
0.01
(5% of reading + 5 digits)
20.0 199.9
0.1
200 9999
1

10.7 Resistance to earth

10.7.1 Standard earthing resistance measurement – 3-wire
measurement
Measuring range according to EN61557-5 is 2.00  1999 .
Max. auxiliary earth electrode resistance RC ...........100RE or 50 k (whichever is lower)
Max. probe resistance RP .......................................100RE or 50 k (whichever is lower)
Additional probe resistance error at R
Additional error at 3 V voltage noise (50 Hz) ...........(5 % of reading + 10 digits)
Open circuit voltage ................................ ................<30 VAC
Short circuit current ................................................< 30 mA
Test voltage frequency ...........................................125 Hz
Test voltage shape .................................................sine wave
Noise voltage indication threshold ..........................1 V (< 50 , worst case)
Automatic measurement of auxiliary electrode resistance and probe resistance. Automatic measurement of voltage noise.
Cmax
or R
.(10 % of reading + 10 digits)
Pmax.
79
Page 80
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications
Measuring range (V)
Resolution (V)
Accuracy
0 550
1
(2 % of reading + 2 digits)
Measuring range (Hz)
Resolution (Hz)
Accuracy
0.00 9.99
0.01
(0.2 % of reading + 1 digit)
10.0 499.9
0.1
Measuring range (V)
Resolution (V)
Accuracy
10 550
1
(2 % of reading + 2 digits)

10.8 Voltage, frequency, and phase rotation

10.8.1 Phase rotation

Nominal system voltage range ............... 100 VAC 550 VAC
Nominal frequency range ....................... 14 Hz 500 Hz
Result displayed .................................... 1.2.3 or 3.2.1

10.8.2 Voltage

Result type ............................................ True r.m.s. (TRMS)
Nominal frequency range ....................... 0 Hz, 14 Hz 500 Hz

10.8.3 Frequency

Nominal voltage range ........................... 10 V 550 V

10.8.4 Online terminal voltage monitor

80
Page 81
MI 3100 s, MI 3100 SE EurotestEASI Technical specifications

10.9 General data

Power supply voltage ............................. 9 VDC (61.5 V battery or accu, size AA)
Operation ............................................... typical 20 h
Charger socket input voltage ................. 12 V 10 %
Charger socket input current .................. 400 mA max.
Battery charging current ........................ 250 mA (internally regulated)
Measuring category ............................... 600 V CAT III
300 V CAT IV
Protection classification ......................... double insulation
Pollution degree ..................................... 2
Protection degree .................................. IP 40
Display .................................................. 128x64 dots matrix display with backlight
Dimensions (w h d) .......................... 23 cm 10.3 cm 11.5 cm
Weight .................................................. 1.3 kg, without battery cells
Reference conditions
Reference temperature range ................ 10 C 30 C
Reference humidity range ...................... 40 %RH 70 %RH
Operation conditions
Working temperature range ................... 0C 40 C
Maximum relative humidity .................... 95 %RH (0C 40 C), non-condensing
Storage conditions
Temperature range ................................ -10C +70 C
Maximum relative humidity .................... 90 %RH (-10C +40 C)
80 %RH (40C 60 C)
Communication transfer speed
RS 232 .................................................. 57600 baud
USB ....................................................... 256000 baud
Size of memory (MI 3100 SE only) ........ up to 1800 measurements
The error in operating conditions could be at most the error for reference conditions (specified in the manual for each function) +1 % of measured value + 1 digit, unless otherwise specified in the manual for particular function.
81
Page 82
MI 3100 s, MI 3100 SE EurotestEASI Appendix A – Fuse table
Rated
current
(A)
Disconnection time [s]
35m
0.1
0.2
0.4
5
Min. prospective short- circuit current (A)
2
32.5
22.3
18.7
15.9
9.1
4
65.6
46.4
38.8
31.9
18.7
6
102.8
70
56.5
46.4
26.7
10
165.8
115.3
96.5
80.7
46.4
16
206.9
150.8
126.1
107.4
66.3
20
276.8
204.2
170.8
145.5
86.7
25
361.3
257.5
215.4
180.2
109.3
35
618.1
453.2
374
308.7
169.5
50
919.2
640
545
464.2
266.9
63
1217.2
821.7
663.3
545
319.1
80
1567.2
1133.1
964.9
836.5
447.9
100
2075.3
1429
1195.4
1018
585.4
125
2826.3
2006
1708.3
1454.8
765.1
160
3538.2
2485.1
2042.1
1678.1
947.9
200
4555.5
3488.5
2970.8
2529.9
1354.5
250
6032.4
4399.6
3615.3
2918.2
1590.6
315
7766.8
6066.6
4985.1
4096.4
2272.9
400
10577.7
7929.1
6632.9
5450.5
2766.1
500
13619
10933.5
8825.4
7515.7
3952.7
630
19619.3
14037.4
11534.9
9310.9
4985.1
710
19712.3
17766.9
14341.3
11996.9
6423.2
800
25260.3
20059.8
16192.1
13545.1
7252.1
1000
34402.1
23555.5
19356.3
16192.1
9146.2
1250
45555.1
36152.6
29182.1
24411.6
13070.1

Appendix A – Fuse table

A.1 Fuse table – IPSC

Fuse type NV
82
Page 83
MI 3100 s, MI 3100 SE EurotestEASI Appendix A – Fuse table
Rated
current
(A)
Disconnection time [s]
35m
0.1
0.2
0.4
5
Min. prospective short- circuit current (A)
2
32.5
22.3
18.7
15.9
9.1
4
65.6
46.4
38.8
31.9
18.7
6
102.8
70
56.5
46.4
26.7
10
165.8
115.3
96.5
80.7
46.4
13
193.1
144.8
117.9
100
56.2
16
206.9
150.8
126.1
107.4
66.3
20
276.8
204.2
170.8
145.5
86.7
25
361.3
257.5
215.4
180.2
109.3
32
539.1
361.5
307.9
271.7
159.1
35
618.1
453.2
374
308.7
169.5
40
694.2
464.2
381.4
319.1
190.1
50
919.2
640
545
464.2
266.9
63
1217.2
821.7
663.3
545
319.1
80
1567.2
1133.1
964.9
836.5
447.9
100
2075.3
1429
1195.4
1018
585.4
Rated
current
(A)
Disconnection time [s]
35m
0.1
0.2
0.4
5
Min. prospective short- circuit current (A)
6
30
30
30
30
30
10
50
50
50
50
50
13
65
65
65
65
65
15
75
75
75
75
75
16
80
80
80
80
80
20
100
100
100
100
100
25
125
125
125
125
125
32
160
160
160
160
160
40
200
200
200
200
200
50
250
250
250
250
250
63
315
315
315
315
315
Fuse type gG
Fuse type B
83
Page 84
MI 3100 s, MI 3100 SE EurotestEASI Appendix A – Fuse table
Rated
current
(A)
Disconnection time [s]
35m
0.1
0.2
0.4
5
Min. prospective short- circuit current (A)
0.5 5 5 5 5
2.7
1
10
10
10
10
5.4
1.6
16
16
16
16
8.6
2
20
20
20
20
10.8
4
40
40
40
40
21.6
6
60
60
60
60
32.4
10
100
100
100
100
54
13
130
130
130
130
70.2
15
150
150
150
150
83
16
160
160
160
160
86.4
20
200
200
200
200
108
25
250
250
250
250
135
32
320
320
320
320
172.8
40
400
400
400
400
216
50
500
500
500
500
270
63
630
630
630
630
340.2
Rated
current
(A)
Disconnection time [s]
35m
0.1
0.2
0.4
Min. prospective short- circuit current (A)
0.5
7.5
7.5
7.5
7.5
1
15
15
15
15
1.6
24
24
24
24
2
30
30
30
30 4
60
60
60
60
6
90
90
90
90
10
150
150
150
150
13
195
195
195
195
15
225
225
225
225
16
240
240
240
240
20
300
300
300
300 25
375
375
375
375
32
480
480
480
480
Fuse type C
Fuse type K
84
Page 85
MI 3100 s, MI 3100 SE EurotestEASI Appendix A – Fuse table
Rated
current
(A)
Disconnection time [s]
35m
0.1
0.2
0.4
5
Min. prospective short- circuit current (A)
0.5
10
10
10
10
2.7
1
20
20
20
20
5.4
1.6
32
32
32
32
8.6
2
40
40
40
40
10.8
4
80
80
80
80
21.6
6
120
120
120
120
32.4
10
200
200
200
200
54
13
260
260
260
260
70.2
15
300
300
300
300
81
16
320
320
320
320
86.4
20
400
400
400
400
108
25
500
500
500
500
135
32
640
640
640
640
172.8
Fuse type D
85
Page 86

MI 3100 s, MI 3100 SE EurotestEASI Appendix B – Accessories for specific measurements

Function
Suitable accessories (Optional with ordering code A….)
Insulation resistance
Test lead, 3 x 1.5 m Tip commander (A 1401)
R LOWΩ resistance Continuity
Test lead, 3 x 1.5 m Tip commander (A 1401) Test lead, 4 m (A 1012)
Line impedance Voltage drop Fault loop impedance PE resistance
Test lead, 3 x 1.5 m Plug commander (A 1314) Mains measuring cable Tip commander (A 1401) Three-phase adapter with switch (A 1111)
Earth connection resistance (MI 3100 SE only)
Test lead, 3 x 1.5 m Plug commander (A 1314) Mains measuring cable Tip commander (A 1401)
RCD testing
Test lead, 3 x 1.5 m Plug commander (A 1314) Mains measuring cable Three-phase adapter with switch (A 1111)
Earth resistance - RE
Test lead, 3 x 1.5 m Earth test set, 3-wire, 20 m (S 2026) Earth test set, 3-wire, 50 m (S 2027)
Phase sequence
Test lead, 3 x 1.5 m Three-phase adapter (A 1110) Three-phase adapter with switch (A 1111)
Voltage, frequency
Test lead, 3 x 1.5 m Plug commander (A 1314) Mains measuring cable Tip commander (A 1401)
Auto sequences (MI 3100 SE only)
Test lead, 3 x 1.5 m Plug commander (A 1314) Mains measuring cable Tip commander (A 1401)
Appendix B – Accessories for specific
measurements
The table below presents recommended standard and optional accessories required for specific measurement. Please see attached list of standard accessories for your set or contact your distributor for further information.
86
Page 87
MI 3100 s, MI 3100 SE EurotestEASI Appendix C – Country notes
Country
Related chapters
Modification type
Note
AT
5.4, C.2.1
Appended
Special G type RCD
TEST
RCD sub-function test [Uc, RCDt, RCD I, AUTO]
IN
Rated RCD residual current sensitivity IN [10 mA, 30 mA, 100 mA, 300 mA, 500
mA, 1000 mA].
type
RCD type [AC, A, F]. starting polarity [ , , , ].
Characteristic and PRCD selection [selective , general non-delayed , delayed
, PRCD, PRCD-K, PRCD-S].
MUL
Multiplication factor for test current [½, 1, 2, 5I ].
Ulim
Conventional touch voltage limit [25 V, 50 V].

Appendix C – Country notes

This appendix C contains collection of minor modifications related to particular country requirements. Some of the modifications mean modified listed function characteristics related to main chapters and others are additional functions. Some minor modifications are related also to different requirements of the same market that are covered by various suppliers.

C.1 List of country modifications

The following table contains current list of applied modifications.

C.2 Modification issues

C.2.1 AT modification - G type RCD

Modified is the following related to the mentioned in the chapter 5.4:
Added G type RCD, Time limits are the same as for general type RCD, Contact voltage is calculated the same as for general type RCD.
Modifications of the chapter 5.4: Test parameters for RCD test and measurement
* Model MI 3102 BT only
Notes:
Ulim can be selected in the Uc sub-function only. Selective (time delayed) RCDs and RCDs with (G) - time delayed characteristic
demonstrate delayed response characteristics. They contain residual current integrating mechanism for generation of delayed trip out. However, contact voltage pre-test in the measuring procedure also influences the RCD and it takes a period to recover into idle
state. Time delay of 30 s is inserted before performing trip-out test to recover type RCD after pre-tests and time delay of 5 s is inserted for the same purpose for type
RCD.
87
Page 88
MI 3100 s, MI 3100 SE EurotestEASI Appendix C – Country notes
RCD type
Contact voltage
Uc
proportional to
Rated I
AC
,
1.05IN
any
AC
21.05IN
A, F
,
1.41.05IN
30 mA
A, F
21.41.05IN
A, F
,
21.05IN
< 30 mA
A, F
221.05IN
Modification of the chapter 5.4.1
Table C.1: Relationship between Uc and I
Technical specifications remain the same.
N
88
Page 89
MI 3100 s, MI 3100 SE EurotestEASI Appendix D – Commanders

Appendix D – Commanders (A 1314, A 1401)

Note:
Commanders A 1314 and A 1401 are applicable on MI 3100 SE only.

D.1 Warnings related to safety

Measuring category of commanders
Plug commander A 1314 ...... 300 V CAT II
Tip commander A1401
(cap off, 18 mm tip) .............. 1000 V CAT II / 600 V CAT II / 300 V CAT II
(cap on, 4 mm tip) ................ 1000 V CAT II / 600 V CAT III / 300 V CAT IV
Measuring category of commanders can be lower than protection category of the
instrument.
If dangerous voltage is detected on the tested PE terminal, immediately stop all
measurements, find and remove the fault!
When replacing battery cells or before opening the battery compartment cover,
disconnect the measuring accessory from the instrument and installation.
Service, repairs or adjustment of instruments and accessories is only allowed to be
carried out by a competent authorized personnel!

D.2 Battery

The commader uses two AAA size alkaline or rechargeable Ni-MH battery cells. Nominal operating time is at least 40 h and is declared for cells with nominal capacity of 850 mAh.
Notes:
If the commander is not used for a long period of time, remove all batteries from the
battery compartment.
Alkaline or rechargeable Ni-MH batteries (size AA) can be used. Metrel recommends
only using rechargeable batteries with a capacity of 800 mAh or above.
Ensure that the battery cells are inserted correctly otherwise the commander will not
operate and the batteries could be discharged.
89
Page 90
MI 3100 s, MI 3100 SE EurotestEASI Appendix D – Commanders
Figure D.1: Front side tip commander (A 1401)
Figure D.2: Front side plug commander (A 1314)
Figure D.3: Back side
1
TEST
TEST
Starts measurements.
Acts also as the PE touching electrode.
2
LED
Left status RGB LED
3
LED
Right status RGB LED
4
LEDs
Lamp LEDs (Tip commander)
5
Function selector
Selects test function.
6
MEM
Store / recall / clear tests in memory of instrument.
7
BL
Switches On / Off backlight on instrument
8
Lamp key
Switches On / Off lamp (Tip commander)
9
Battery cells
Size AAA, alkaline / rechargeable NiMH
10
Battery cover
Battery compartment cover
11
Cap
Removable CAT IV cap (Tip commander)

D.3 Description of commanders

Legend:
90
Page 91
MI 3100 s, MI 3100 SE EurotestEASI Appendix D – Commanders
Both LED yellow
Warning! Dangerous voltage on the commander’s PE terminal!
Right LED red
Fail indication
Right LED green
Pass indication
Left LED blinks blue
Commander is monitoring the input voltage
Left LED orange
Voltage between any test terminals is higher than 50 V
Both LEDs blink red
Low battery
Both LEDs red and switch off
Battery voltage too low for operation of commander
Connect commander to the instrument. Connect commander to the item to be tested (see figure D.4). Touch PE test probe (the TEST key) on commander for at least one second. If PE terminal is connected to phase voltage both LEDs will light yellow, the warning
message on the instrument is displayed, instrument’s buzzer is activated, and further measurements are disabled in Zloop and RCD functions.

D.4 Operation of commanders

PE terminal test procedure
Figure D.4: Reversed L and PE conductors (application of plug commander)
91
Loading...