Metrel MI 2123 User guide

Insulation / Continuity Tester
MI 2123
User Manual
Version 1.1, Code No. 20 750 424
Distributor:
COSINUS Computermesstechnik GmbH
Lise-Meitner-Strasse 6
D-85521 Ottobrunn GERMANY
Tel: 089/665594-0
Fax: 089/665594-30 http://www.cosinus.de
Manufacturer:
web site:
http://www.metrel.si
e-mail: metrel@metrel.si
Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) concerning safety and electromagnetic compatibility regulations
© 2000..2007 METREL
No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
2
MI 2123 Insulation / Continuity Table of contents
1. General introduction--------------------------------------------------------------------------------4
1.1. Features---------------------------------------------------------------------------------------------4
1.2. Applied Standards--------------------------------------------------------------------------------4
2. Instrument description-----------------------------------------------------------------------------5
2.1. Instrument Casing--------------------------------------------------------------------------------5
2.2. Operator’s Panel----------------------------------------------------------------------------------5
2.3. Connectors-----------------------------------------------------------------------------------------6
2.4. Bottom Section------------------------------------------------------------------------------------7
2.5. Accessories ----------------------------------------------------------------------------------------7
3. Warnings and Instrument Messages ----------------------------------------------------------8
3.1. Warnings--------------------------------------------------------------------------------------------8
3.2. Instrument Messages----------------------------------------------------------------------------8
4. Measurements --------------------------------------------------------------------------------------10
4.1. Insulation Resistance--------------------------------------------------------------------------10
4.2. R ±200mA --------------------------------------------------------------------------------------12
4.3. Continuity----------------------------------------------------------------------------------------- 14
5. Operation with Results--------------------------------------------------------------------------- 16
5.1. Storing of Measurement Results ------------------------------------------------------------ 16
5.2. Storage Handling -------------------------------------------------------------------------------17
5.3. Data Transfer------------------------------------------------------------------------------------ 20
6. Maintenance ----------------------------------------------------------------------------------------- 21
6.1. Inspection-----------------------------------------------------------------------------------------21
6.2. Battery Replacement --------------------------------------------------------------------------21
6.3. Fuse replacement ------------------------------------------------------------------------------22
6.4. Cleaning-------------------------------------------------------------------------------------------22
6.5. Service --------------------------------------------------------------------------------------------22
7. Specifications--------------------------------------------------------------------------------------- 23
7.1. Measurements-----------------------------------------------------------------------------------23
7.2. General specifications-------------------------------------------------------------------------24
3
MI 2123 Insulation/Continuity Tester General introduction

1. General introduction

1.1. Features

The SMARTEC Insulation / Continuity Tester is a portable hand-held battery powered test instrument intended for testing of safety measures on mains installations. It operates on a SIMPLE and CLEAR basis. The equipment is designed and produced with the extensive knowledge and experience acquired through many years of dealing with electrical installation test equipment.
General functions of the SMARTEC Insulation / Continuity Tester:
- Insulation Resistance measurement,
- Continuity measurement using test current of 200 mA in accordance with EN 61557
standard,
- Continuity measurement (continuous measurement) using test current of up to 7 mA,
- AC / DC Voltage and Frequency measurement.
Features:
- Wide range of Insulation Resistance test voltages (50 up to 1000 V DC),
- Automatic discharge of tested item after measurement completion,
- Automatic test voltage polarity exchange in Continuity ± 200 mA function,
- Continuity 7 mA function for inductive test items,
- Memories,
- PC connection.
A custom designed LCD offers easy to read results. Operation is simple and clear, the operator does not need any special training (except reading and understanding this Users Manual) to operate the instrument. In order for the operator to be familiar with measurements in general, it is advisable to read the handbook Measurements on electric installations in theory and practice. The instrument allows storage of test results. Professional PC SW enables simple transfer of test results and other parameters in both directions between the tester and the PC.

1.2. Applied Standards

Instrument operation: IEC / EN 61557-1, IEC / EN 61557-2, IEC / EN 61557-4 DIN VDE 0413-1, DIN VDE 0413-4, DIN VDE 100 BS 7671 – 16th edition Electromagnetic compatibility (EMC): IEC 61326, Class B Safety: EN/IEC 61010-1 (instrument), EN/IEC 61010-2-31 (accessories)
4
MI 2123 Insulation/Continuity Tester Instrument description

2. Instrument description

2.1. Instrument Casing

The instrument is housed in a plastic casing, which maintains the protection class defined in general specifications. The casing consists of a main section, which includes the operator's panel and connectors and of a flexible cover.
Note! The cover is permanently fixed to the base section and cannot be detached.

2.2. Operator’s Panel

The operator’s panel consists of a custom LCD, a rotary switch, and a keyboard, see the figure below.
1
2
3
1
2
3
4
6
5
8
7
4
7865
Fig. 1. Front panel
5
MI 2123 Insulation/Continuity Tester Instrument description
Legend:
1..........Custom designed LCD.
2..........Up key to increase storage group.
3..........START key to start any measurement.
4..........DISP key to display subresults of selected function.
5..........MEM key to store and recall results.
6..........CLR key to erase memorized results.
7..........Down key to decrease storage group.
8..........Rotary switch to select measurement function.

2.3. Connectors

Use originl test accessories only! Max. allowed voltage between mains test terminals and ground is 300 V! Max. allowed voltage between mains test terminals is 600 V! Max. output voltage between test terminals is 1100 V!
1
123
Fig. 2. Connectors
Legend:
1..........Test connector.
2..........RS 232 connector to connect the instrument to PC.
3..........Battery charger connector.
Test connector is intended for connection of the test cable and is accessible only when the instrument cover is lifted up. RS 232 and battery charger connectors are accessible only when the cover of the instrument is closed. For safety reasons both connector fields cannot be accessible at the same time, the cover of the instrument selects the accessible one.
6
MI 2123 Insulation/Continuity Tester Instrument description

2.4. Bottom Section

1
2
DISCONNECT ALL TEST LEADS AND SWITCH OFF THE INSTRUMENT BEFORE REMOVING THE BATTERY COVER - HAZARDOUS VOLTAGE! DO NOT CHARGE IF BATTERY COMPARTMENT CONTAINS ALKALINE CELLS!
3
4
5
Legend:
1..........Plastic housing.
2..........Screw (4 pieces) to fix the battery cover.
3..........Battery compartment cover.
4..........Battery cover warnings.
5..........Trade mark label.
POWER SUPPLY: 4 × 1,2 V - NiCd, NiMH RECHARGEABLE (IEC LR 14) or 4 × 1,5 V ALKALINE BATTERIES (IEC LR 14)
VOR OFNEN DES BATTERIEFACHS ALLE ZULEITUNGEN VOM GERAT ENTFERNEN UND DAS GERAT AUSSCHALTEN - GEFAHRLICHE SPANNUNG! NICHT AUFLADEN WENN ALKALZELLEN VERWENDET SIND!
STROMVERSORGUNG:
oder 4×1,5V - ALKALZELLEN (IEC LR 14)
4×1,2V - NiCd, NiMH AUFLADBAR (IEC LR 14)
Fig. 3. Bottom section

2.5. Accessories

The instrument consists of standard and optional accessories. Optional accessories can be delivered upon request. See attached list for standard configuration and options or contact your distributor or see the METREL home page:
http://www.metrel.si .
7
MI 2123 Insulation/Continuity Tester Warnings and Messages

3. Warnings and Instrument Messages

3.1. Warnings

In order to reach highest level of operator’s safety while carrying out various measurements and tests using the SMARTEC Insulation / Continuity Tester, as well as to ensure the test equipment remains undamaged, it is necessary to consider the following general warnings:
If the test equipment is used in a manner not specified in this Users Manual,
the protection provided by the equipment may be impaired!
Do not use the instrument and accessories if any damage is noticed!
In case of blown fuse, follow the instructions in this Instruction Manual to
replace it!
Service intervention or calibration procedure is allowed to be carried out
only by a competent, authorised person!
Consider all generally known precautions, in order to avoid risk of electric
shock while dealing with electrical installations!
Use only standard or optional test accessories supplied by your distributor!
Disconnect all test leads and switch power off before opening Battery cover!
Do not charge when alkaline batteries are fitted!
!
symbol at the instrument means “Read the Users Manual with special
care!” The symbol requires an action!
symbol on LCD indicates that the measurement results show an irregular
!
condition of the item under test which may present a hazardous live voltage.

3.2. Instrument Messages

Messages are generated on the LCD using special symbols and numeric segment combinations. The next figure shows all possible display segments and the table under the figure describes the messages.
Result value
Low battery Higher than
Warning!
Mains voltage present
at test terminals
Fig. 4. Display segments
Unit of displayed result
Operation with storage
Cursor to indicate compensated test leads
8
MI 2123 Insulation/Continuity Tester Warnings and Messages
Description of possible display messages:
>1999
FUS
SEr rES
End MEM
MEM
no MEM
rCL
Clr flashing
Clr / ALL alternating
Result over range. Blown fuse in Continuity functions only. Active serial communication. *Reset of the instrument (erased memories). All storage locations occupied. Operation with storage (store or recall). No result to be stored or no stored result to be recalled. Recall function is entered. Confirm or abort clearing of last stored result? Confirm or abort clearing of all stored results?
Table 1. Messages
Note!
* This indication appears after the first insertion of batteries, i.e. when there were no
batteries in the battery compartment for a specific period of time (a few hours), or if the microprocessor detects an irregular state in its memory or if reset of the instrument has been performed.
9
MI 2123 Insulation/Continuity Tester Measurements

4. Measurements

4.1. Insulation Resistance

Warnings!
Make sure that the item to be tested is disconnected (mains voltage
disconnected) before starting the Insulation Resistance measurement!
When measuring Insulation Resistance between conductors, all loads must
be disconnected and all switches closed!
Do not touch tested item while the test is in operation, risk of electric shock!
Do not connect test terminals to an external voltage higher than 600 V AC or
DC in order not to damage the test instrument!
In case of capacitive test item (capacitive compensation of reactive power,
long tested cable etc.), automatic discharge of the item may not be done immediately after finishing the measurement. Falling voltage will be displayed in that case – do not disconnect test leads until the voltage drops below 50 V!
The following figure shows connection of the instrument to tested item for testing of Insulation Resistance on electric installations. The measurement should be carried out between all conductor
s.
RCD
L1 L2 L3 N PE
PE N L3 L2 L1
Fig. 5. Connection of the instrument to item under test
There may be also other items to be tested such as motor windings, ground cables, antistatic floor coatings, walls etc., see the booklet
installations in theory and practice
Measurements on electric
10
MI 2123 Insulation/Continuity Tester Measurements
How to carry out the measurement:
- Connect test leads to item to be tested in accordance with figure 5.
- Select Insulation resistance function using appropriate test voltage by means of
rotary switch knob.
- Press the
release the key. With pressing the measurement starts to run.
- Read displayed result (Insulation Resistance).
- Check subresult (Actual test voltage) by pressing the
- The result can optionally be saved using the
instructions in paragraph 5.2. Storage handling.
START key and keep it pressed until the result has stabilised then
START key twice successively, continuous
DISP key.
MEM, Up and Down keys, see
M
Ω
V
Example of displayed Insulation Resistance
Fig. 6. Example of displayed result
Notes!
If an external voltage higher than 30 V AC/DC is present at test terminals, the
Insulation Resistance measurement will not be carried out after pressing START key but the voltage will be displayed marked with
Tested item is discharged automatically after finishing the measurement and the
actual voltage is displayed during discharge operation (only if high capacity is present).
Positive pole of test voltage is attached to red test terminal.
Actual test voltage
!
symbol.
11
MI 2123 Insulation/Continuity Tester Measurements
4.2. R ±200mA
Continuity of protection conductors is to be measured, before mains voltage is connected to tested installation (new or adapted installation). Max. allowed resistance value depends on power of connected loads, used installation system (TN, TT) etc.
Test current of at least 200 mA is required by the EN 61557 regulation which is why this function is to be used whenever tests are to be done in accordance with this standard.
For additional general information concerning Continuity measurement, refer to the handbook
Warning!
Make sure that the tested item is de-energised (mains voltage disconnected)
before starting the measurement! Fuse will blow if test probes are connected to mains voltage during the test.
How to carry out the measurement:
- Connect test leads to tested item in accordance with the figure below.
- Select R ±200mA function by means of rotary switch knob.
- Make sure test leads are compensated, see the instructions in the next chapter.
- Press the
function enables individual measurements (the START key should be pressed for each measurement).
- Read displayed result (Continuity), see figure 8.
- The result can optionally be saved using the
instructions in paragraph 5.2. Storage handling.
Measurements on electric installations in practice and theory.
START key and release it. To ensure the instrument’s protection this
MEM, Up and Down keys, see
RCD
L1 L2 L3 N PE
PE N L3 L2 L1
Fig. 7. Connection of test leads
12
MI 2123 Insulation/Continuity Tester Measurements
Ω
Comp
>
Fig. 8. Example of R
±
200mA
test result, test leads are compensated
Notes!
If an external voltage higher than 10 V AC/DC is present at test terminals, the
R ±200mA
voltage will be displayed, marked with
measurement will not be carried out after pressing START key, but the
symbol.
!
If test result is out of measurement range (open test leads), >1999 Ω message will
be displayed!
If indication FUS appears, replace blown fuse with new one. See the paragraph 6.3.
Fuse replacement.
Compensation of test leads’ resistance
In order to achieve high accuracy of test results it is advised to compensate resistance of the test leads before carrying out Continuity measurement. Especially it is advised to compensate test leads whenever the leads are changed for different ones.
How to compensate test leads:
- Select R ±200mA function.
- Short out test leads.
- Press the START key twice successively, total value of measured resistance
(resistance of test leads + internal resistance) is displayed for a while then displayed. Cursor symbol on display (see the fig. 4) shows compensation state. Resistance of test leads is therefore compensated and the instrument is ready for Continuity measurements.
Notes!
Max. resistance that can be compensated is 5 Ω. If total measured resistance is
higher than 5
Ω (e.g. open test leads) then any compensation previously stored is
ignored (compensation indicator disappears).
Compensation of test leads’ resistance can be done only in the R±200mA
The indicator for compensation of test leads appears only after compensation has
been performed and remains until the compensation is cancelled (e.g. by performing the compensation at open test leads).
Compensation of test leads’ resistance is stored and is used for both R ±200mA Continuity functions.
0.00 Ω is
function.
and
13
MI 2123 Insulation/Continuity Tester Measurements

4.3. Continuity

Continuity function is intended to be used when there are inductive loads to be tested, for example motor windings, transformers etc. or if there is no need to follow prescribed standards. It can be used also as an ordinary multi-meter (resistance range). This function enables continuous measurements (the measurement is running continuously after pressing the START key).
Test current of only up to 7 mA is used in order to save the battery energy. For additional general information concerning Continuity measurement, refer to the
handbook
Warning!
Make sure tested item is de-energised (mains voltage disconnected) before
How to carry out the measurement:
- Connect test leads to tested item.
- Select Continuity function by means of rotary switch knob.
- Make sure test leads are compensated.
- Press the START key and release it, measurement starts to run continuously.
- Press the START key again to stop the measurement.
- Read displayed result (Continuity), see the figure below.
- The result can optionally be saved using the
instructions in paragraph 5.2. Storage handling.
Measurements on electric installations in practice and theory.
starting the measurement! Fuse will blow if test probes are connected to mains voltage during the test.
MEM, Up and Down keys, see
symbol.
!
14
>
Comp
Fig. 9. Example of Continuity test result, test leads are compensated
Notes!
If an external voltage higher than 10 V AC/DC is present at test terminals, the
Continuity measurement will not be carried out after pressing START key, but the
voltage will be displayed marked with
Compensation of test leads’ resistance can be carried out in R ±200mA function only.
If test result is out of measurement range (open test leads), >1999 Ω message will
be displayed!
If indication FUS appears, replace blown fuse. See paragraph 6.3. Fuse
replacement.
Positive pole of test voltage is attached to red test terminal.
MI 2123 Insulation/Continuity Tester Measurements
4.4. Voltage Measurement
How to carry out the measurement:
- Connect test leads to equipment under test.
- Select
- Press the
- Read displayed result (Voltage), see the figure below. Displayed voltage may be
exchanged with frequency by simply pressing the the measurement is running.
- Press the
- Check subresult (Frequency of displayed voltage) by pressing the DISP key.
- The result can optionally be saved using the MEM, Up and Down keys, see
instructions in paragraph 5.2. Storage handling.
VOLTAGE function by means of rotary switch knob.
START key, continuous measurement starts to run.
DISP key at any time, even whilst
START key again to stop the measurement.
V
Hz
Displayed result Frequency
Fig. 10. Example of voltage measurement result
Note!
Do not connect test leads to a voltage higher than 600 V AC, to avoid possible
damage to the instrument!
15
MI 2123 Insulation/Continuity Tester Operation with Results

5. Operation with Results

5.1. Storing of Measurement Results

The memory for storing of measurement results has a stacking group organisation. The operator has the possibility to organise 1999 groups. Each group has the possibility to stack any measured result. In ‘storing result’ mode the operator selects a group and then the result is stacked together with all its parameters into the selected group. The figure below shows a graphic presentation of result grouping.
Rm
Ri
R2 R1
G1 G2 G3 Gn G1999..... .....
Fig. 11. Storage locations organisation
A set of information stored on location Ri (main result, sub-results and parameters) depends on selected function. There can be many results stored on the same location.
Practical Example of Storage Organization
For example: There is a standard house installation that has to be verified with regard to various electrical characteristics. The test plan should be prepared before any measurement is taken. The plan is based on wiring schematics and contains group definitions e.g. floor, room, wall socket and the related measurements in each group. The groups are assigned with numbers from 1 to 1999. When the test plan is defined the operator starts to run the measurements from group to group. The instrument asks every time for the group when storing test results. The result can be simply stored into the appropriate group if the group number already corresponds to selected measurement place in test plan, or the group number can be changed first and then the result stored.
All results can be stored under a single group, e.g. No.: 1, if there is no need for grouping.
16
MI 2123 Insulation/Continuity Tester Operation with Results

5.2. Storage Handling

The following keys are intended to be used for storage handling: There are the following possibilities of handling:
To store result (result
is displayed):
a) Press the
MEM key after completing the measurement. Last used
group will be offered automatically, see the figure below. b) Use c) Press the
and keys to change group number.
MEM key again to store result.
MEM, , and CLR.
To recall stored results:
MEM
Offered group number
Once stored -
no possibility
MEM
to store the same result again
Fig. 12. Group code and reaction to double storing of the same result
Notes!
- Press the START key or change rotary switch position to cancel
store procedure.
- Each result can only be stored once.
a) Press the MEM key just after changing rotary switch position, MEM
mark will appear on LCD and
rCL for a moment followed by the last
accessed group number. b) If there is no stored data
no and MEM will flash for a second
followed by selected function idle state display. c) Use the d) Press the
recall stored result. If there is no result stored in selected group
and e) Use the
and keys to change the group number.
MEM key again once appropriate group is selected to
no
MEM will flash followed by the group number.
DISP key to check sub-results and parameters.
f) To see previous or next stored results inside selected group press
key. Every entry is followed by the function code and then the
or
result. g) Press the
MEM key to change group number and go to b).
Note!
- Press the START key or change rotary switch position to cancel
recall procedure.
17
MI 2123 Insulation/Continuity Tester Operation with Results
Indication when entering recall
procedure
MEM
Last accessed group is offered
Storage buffer empty
Code of stored function (rotary
switch position 7 = R
MEM
Ω
Stored result (R±200mA)
MEM
Fig. 13. Displayed information during recall procedure
To clear complete Memory:
a) Exit storage operation (storing or recalling), press the CLR key and
keep it pressed until
release it. b) Press the
Clr and ALL messages start to alternate then
CLR key again to erase all stored results.
MEM
±200mA)
MEM
Fig. 14. Waiting for confirmation
Note! Press the procedure.
START key or change rotary switch position to cancel clear
18
MEM
MI 2123 Insulation/Continuity Tester Operation with Results
To clear only last stored result:
a) Exit storage operation (storing or recalling), press the CLR key, Clr
message appears, then release the key. b) Press the
CLR key again to erase last stored result.
MEM
Fig. 15. Waiting for confirmation
Note!
- Press the START key or change rotary switch position to cancel
clear procedure.
Stored Parameters
The following is a list of results, sub-results and parameters stored for each function. Function No Stored data Function No Parameters
RINS 50 V
1 Function code
Insulation resistance Actual test voltage
RINS 1000 V
5 Function code
Insulation resistance Actual test voltage
RINS 100 V
RINS 250 V
RINS 500 V
2 Function code
Insulation resistance Actual test voltage
3 Function code
Insulation resistance Actual test voltage
4 Function code
Insulation resistance Actual test voltage
CONTINUITY
R±200mA
VOLTAGE
6 Function code
Resistance
7 Function code
Resistance
8 Function code
Voltage Frequency
19
MI 2123 Insulation/Continuity Tester Operation with Results

5.3. Data Transfer

Stored results can be transferred to a PC. A special communication program has the ability to identify the instrument and download the data.
How to transfer stored data:
- Connect PC COM port to the instrument with the serial communication cable.
- Turn power on to the PC and the instrument.
- Run the
- The PC and the instrument automatically recognise each other.
- The program on the PC enables the following possibilities:
-- download data;
-- clear storage;
-- change and download user data;
-- prepare a simple report form;
-- prepare a file to import to a spreadsheet. The program
README.TXT for instructions on installing and running the program.
smartlink.exe. program
smartlink.exe is a Windows 95/98 based PC software. Read file
20
MI 2123 Insulation/Continuity Tester Maintenance
5

6. Maintenance

6.1. Inspection

To maintain operator’s safety and ensure reliability of the instrument it is good practice to inspect the instrument on a regular basis. Check that the instrument and accessories are not damaged. If any defect is found please consult service centre, distributor or manufacturer.

6.2. Battery Replacement

The battery symbol on the upper left corner of the LCD indicates the battery state (Ubat < 4.2 V). If a low battery state is indicated the batteries must be replaced to ensure accurate measurements. The instrument is switched off automatically when battery voltage drops down to 4.0 V;
bat indication is displayed before switching off.
Notes!
All four cells shall be replaced each time.
!
Nominal power supply voltage is 6 V DC. Use four 1.5 V alkaline cells, type IEC LR14 (dimensions: diameter = 26 mm, height = 50 mm).
Turn power off and disconnect any measurement accessory connected to
the instrument before opening battery cover.
M0.31
+
+
+
21
+
Fig. 16. Correct polarity of inserted batteries and fuse position
One set of full-capacity alkaline cells can supply the instrument for approx. 150 hours.
MI 2123 Insulation/Continuity Tester Maintenance
Rechargeable NiCd or NiMH battery can be used instead of alkaline. The instrument contains an additional connector for charging the batteries.
Notes!
Insert cells correctly, otherwise test instrument will not operate and battery could be
discharged, see the figure above for correct battery polarity!
If the instrument is not to be used for a long period of time remove all batteries from
the battery compartment.
To ensure the integrity of the stored results do not leave battery compartment empty
for more than 1 minute.
Warnings!
Do not charge when alkaline batteries are fitted!
Take into account handling, maintenance and recycling requirements as
defined by related standards and manufacturer of alkaline or rechargeable batteries.

6.3. Fuse replacement

The instrument contains a fuse to protect R can blow if an external voltage is connected to test terminals during the measurement. See the fig. 16. for fuse location.
Warning!
Replace blown fuse with prescribed one only (M0.315A) otherwise the instrument may be damaged.
±200mA and Continuity functions. The fuse

6.4. Cleaning

Use a soft cloth slightly moistened with soapy water or spirit to clean the surface of the instrument and leave the instrument to dry totally, before using it.
Notes!
Do not use liquids based on petrol or hydrocarbons!
Do not spill cleaning liquid over the instrument!

6.5. Service

Repairs under or out of warranty time: Please contact your distributor for further information.
Note!
Unauthorised person is not allowed to open the instrument!
22
MI 2123 Insulation/Continuity Tester Specifications

7. Specifications

7.1. Measurements

Insulation resistance
Measuring range Riso (Un 500V): 0.12MΩ ÷ 20GΩ
Display range Riso
Ω)
(M
Un: 500V, 1000V
0.00 ÷ 19.99
20.0 ÷ 199.9 200 ÷ 1999
2.00k ÷ 19.99k
Measuring range Riso (Un
Display range Riso
Ω)
(M
Un: 50V, 100V, 250V
0.000 ÷ 1.999
2.00 ÷ 19.99
20.0 ÷ 199.9
Nominal output voltages: 50V, 100V, 250V, 500V, 1000V Output voltage tolerance: +20%, -0% Nominal output current: Short circuit current: < 3mA
R±200mA
Measuring range R: 0.08 Ω ÷ 2000 Ω
Display range R
(
Ω)
0.00 ÷ 19.99
20.0 ÷ 199.9 200 ÷ 1999
Open-terminal test voltage: 4 - 7 Vd.c. Test current for R Compensation of test leads (up to 5 Automatic polarity exchange: yes Measurement mode: single measurement
2Ω: > 200 mA
Resolution
(MΩ)
0.01
0.1 1
10
< 500V): 0.012MΩ ÷ 200MΩ
Resolution
(MΩ)
0.001
0.01
0.1
1mA
Resolution
(Ω)
0.01
0.1 1
Ω): yes
±(3% of reading + 3 Digits)
±(10% of reading)
±(5% of reading + 3 Digits)
±(2% of reading + 2 Digits)
Accuracy
Accuracy
Accuracy
±3% of reading
23
MI 2123 Insulation/Continuity Tester Specifications
Continuity
Display range R
(
Ω)
0.0 ÷ 199.9 200 ÷ 2000
Resolution
(Ω)
0.1 1
Accuracy
±(5% of reading + 3 Digits)
Open-terminal test voltage: 4 - 7 V DC Short-circuit test current: up to 7 mA Measurement mode: continuous measurement
Voltage U
Display range U
(V)
0 ÷ 600
Resolution
(V)
1
Accuracy
±(3% of reading + 3 Digits)
Nominal frequency range DC, 45 – 65 Hz

7.2. General specifications

Power supply 6V DC (4 × 1.5V alkaline battery IEC LR14)
or 4.8V DC (4 × 1.2V NiCd, NiMH rechargeable
battery IEC LR14) Battery charging 1.5 hours for full charge (using fast charger) Auto power off yes, approximately 10 min of no activity
Dimensions (w × h × d) 15.5 × 9.5 × 19 cm Weight (without accessory, with battery) 1.3 kg
Display custom LCD Memory approx. 1000 measurements Computer connection RS 232 (9600 baud, no parity, 8 bit data,
1 stop bit) Protection classification Class II (double insulation)
Over-voltage category CATIII 300V Pollution degree 2 Protection degree IP 54 Working temperature range 0 Reference temperature range 10 Max. humidity 85 % RH (0 Reference humidity range 40
÷ 40 °C
÷ 30 °C
÷ 40°C)
÷ 60 % RH
24
Loading...