METREL EurotestXA MI 3105 Instruction Manual

Page 1
EurotestXA
MI 3105
Instruction manual
Version 2.4, HW 3; Code no. 20 751 009
Page 2
Distributor:
Manufacturer:
METREL d.d. Ljubljanska cesta 77 1354 Horjul Slovenia web site: http://www.metrel.si e-mail: metrel@metrel.si
Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) concerning safety and electromagnetic compatibility regulations
© 2006, 2007, 2008 METREL
No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
Page 3
MI 3105 EurotestXA Table of contents
3
1 Preface..................................................................................................................................... 6
2 Safety and operational considerations.................................................................................7
2.1 Warnings and notes ...........................................................................................................7
2.2 Battery and charging ........................................................................................................10
2.2.1 New battery cells or cells unused for a longer period.............................................. 11
2.3 Standards applied ............................................................................................................12
3 Instrument description......................................................................................................... 13
3.1 Front panel .......................................................................................................................13
3.2 Connector panel...............................................................................................................14
3.3 Back panel........................................................................................................................15
3.4 Bottom..............................................................................................................................16
3.5 Display organization.........................................................................................................17
3.5.1 Terminal voltage monitor......................................................................................... 17
3.5.2 Menu line................................................................................................................. 18
3.5.3 Message field .......................................................................................................... 18
3.5.4 Result field............................................................................................................... 19
3.5.5 Other messages ...................................................................................................... 19
3.5.6 Sound warnings....................................................................................................... 19
3.5.7 Help ......................................................................................................................... 19
3.5.8 Backlight and contrast adjustments......................................................................... 20
3.6 Carrying the instrument ....................................................................................................21
3.7 Instrument set and accessories........................................................................................21
3.7.1 Standard set ............................................................................................................ 21
3.7.2 Optional accessories ............................................................................................... 21
4 Instrument operation............................................................................................................ 22
4.1 Main menu........................................................................................................................22
4.2 Single test.........................................................................................................................22
4.3 Automatic testing..............................................................................................................24
4.3.1 Auto sequence number main menu ........................................................................ 26
4.3.2 Auto sequence set................................................................................................... 26
4.3.3 Test parameters in auto sequence.......................................................................... 27
4.3.4 Name and description of auto sequence................................................................. 27
4.3.5 Storing auto sequence settings (sequence, number, name)................................... 29
4.3.6 Pause flag and comments in auto sequence........................................................... 30
4.3.7 Setting pause flag and comments ........................................................................... 30
4.3.8 Building an auto sequence ...................................................................................... 32
4.4 Miscellaneous...................................................................................................................36
4.4.1 Language ................................................................................................................ 36
4.4.2 Supply system, Isc factor, RCD standard................................................................ 37
4.4.3 Memory ................................................................................................................... 39
4.4.4 Date and time .......................................................................................................... 39
4.4.5 Initial settings........................................................................................................... 40
4.4.6 Communication port ................................................................................................ 42
4.4.7 Locator .................................................................................................................... 42
4.4.8 Operator .................................................................................................................. 43
5 Measurements....................................................................................................................... 44
5.1 Insulation resistance.........................................................................................................44
5.2 Resistance to earth connection and equipotential bonding..............................................46
5.2.1 Continuity R200 mA measurement.......................................................................... 46
5.2.2 7 mA resistance measurement................................................................................ 48
5.2.3 Compensation of test leads resistance.................................................................... 49
5.3 Testing RCDs...................................................................................................................51
5.3.1 Contact voltage (RCD Uc)....................................................................................... 52
5.3.2 Trip-out time t .......................................................................................................... 53
5.3.3 Trip-out current........................................................................................................ 54
Page 4
MI 3105 EurotestXA Table of contents
4
5.3.4 RCD Autotest .......................................................................................................... 55
5.4 Fault loop impedance and prospective fault current.........................................................57
5.5 Line impedance and prospective short-circuit current......................................................59
5.6 Voltage, frequency and phase sequence.........................................................................61
5.7 Resistance to earth ..........................................................................................................63
5.7.1 Standard 3-wire measurement ................................................................................ 63
5.7.2 One clamp measurement ........................................................................................ 65
5.7.3 Two clamps measurement ...................................................................................... 66
5.7.4 Specific earth resistance measurement .................................................................. 67
5.8 Current .............................................................................................................................68
5.9 Sensors and adapters ......................................................................................................70
5.9.1 Illumination .............................................................................................................. 70
5.9.2 2 Ω line/loop impedance.......................................................................................... 72
5.10 PE test terminal................................................................................................................74
5.11 Locator .............................................................................................................................76
5.12 Varistor test ......................................................................................................................78
6 Data handling ........................................................................................................................ 80
6.1 Memory organization........................................................................................................80
6.2 Installation data structure .................................................................................................80
6.3 Storing test results............................................................................................................83
6.3.1 Saving results specialties ........................................................................................ 84
6.4 Recalling test results and parameters..............................................................................85
6.4.1 Recalling result........................................................................................................ 85
6.5 Clearing saved data .........................................................................................................86
6.5.1 Clearing specialties ................................................................................................. 87
6.6 Editing installation data structure .....................................................................................88
6.6.1 Adding new locations .............................................................................................. 88
6.7 Communication ................................................................................................................91
7 Maintenance .......................................................................................................................... 92
7.1 Replacing fuses................................................................................................................92
7.2 Cleaning ...........................................................................................................................92
7.3 Periodic calibration...........................................................................................................92
7.4 Service .............................................................................................................................92
8 Technical specifications....................................................................................................... 93
8.1 Insulation resistance.........................................................................................................93
8.2 Continuity .........................................................................................................................94
8.2.1 Resistance R200mA (L-PE, N-PE).......................................................................... 94
8.2.2 Resistance R7mA (L-PE, N-PE).............................................................................. 94
8.3 RCD testing......................................................................................................................94
8.3.1 General data............................................................................................................ 94
8.3.2 Contact voltage RCD-Uc ......................................................................................... 95
8.3.3 Trip-out time ............................................................................................................ 95
8.3.4 Trip-out current........................................................................................................ 96
8.4 Fault loop impedance and prospective fault current.........................................................96
8.4.1 No disconnecting device or FUSE selected ............................................................ 96
8.4.2 RCD selected .......................................................................................................... 97
8.5 Line impedance and prospective short-circuit current......................................................97
8.6 Voltage, frequency, and phase rotation............................................................................98
8.6.1 Phase rotation ......................................................................................................... 98
8.6.2 Voltage .................................................................................................................... 98
8.6.3 Frequency ............................................................................................................... 98
8.7 Online terminal voltage monitor........................................................................................98
8.8 Earth resistance ...............................................................................................................99
8.9 TRMS Clamp current......................................................................................................101
8.10 Illumination .....................................................................................................................101
Page 5
MI 3105 EurotestXA Table of contents
5
8.11 2 Ω line/loop impedance.................................................................................................102
8.11.1 High precision line impedance .............................................................................. 102
8.11.2 High precision fault loop impedance ..................................................................... 102
8.11.3 Contact voltage ..................................................................................................... 103
8.12 Varistor test ....................................................................................................................103
8.13 General data...................................................................................................................104
A Appendix A - Fuse table..................................................................................................... 105
B Appendix B - Accessories for specific measurements ................................................... 108
C Appendix C – Locator receiver R10K................................................................................109
C.1 Tracing principles ...........................................................................................................110
C.1.1 Positioning the receiver ......................................................................................... 110
C.1.2 Positioning current clamp ...................................................................................... 110
C.1.3 Positioning selective probe.................................................................................... 111
C.2 Detection distances for different connections.................................................................111
C.3 R10K power supply ........................................................................................................111
C.4 Maintenance...................................................................................................................111
D Appendix D - IT supply systems........................................................................................ 112
D.1 Standard references.......................................................................................................112
D.2 Fundamentals.................................................................................................................112
D.3 Measurement guides......................................................................................................113
D.3.1 MI 3105 test functions and IT systems.................................................................. 114
D.3.2 Voltage measurements ......................................................................................... 114
D.3.3 Line impedance ..................................................................................................... 114
D.3.4 RCD testing ........................................................................................................... 115
D.3.5 IMD testing ............................................................................................................ 115
D.3.6 First fault leakage current (ISFL)........................................................................... 117
D.4 Technical specifications .................................................................................................119
D.4.1 First fault leakage current ISFL ............................................................................. 119
D.4.2 Calibrated resistances for IMD testing................................................................... 119
E Appendix E - Reduced low voltage supply systems ....................................................... 120
E.1 Standard reference.........................................................................................................120
E.2 Fundamentals.................................................................................................................120
E.3 MI 3105 guides...............................................................................................................120
E.3.1 MI 3105 functions and reduced low voltage systems ............................................ 121
E.4 Technical specifications .................................................................................................123
E.4.1 RCD....................................................................................................................... 123
E.4.2 Fault loop impedance and prospective short-circuit current .................................. 125
E.4.3 Line impedance and prospective short-circuit current ........................................... 127
F Appendix F – Country notes .............................................................................................. 128
F.1 List of country modifications...........................................................................................128
F.2 Modification issues.........................................................................................................128
F.2.1 AT modification - G type RCD............................................................................... 128
F.2.2 ES modification - CONTINUITY LOOP Re............................................................ 129
F.2.3 IT modification - CONTINUITY LOOP Re ............................................................. 131
F.2.4 CH modification - Change L/N............................................................................... 133
Page 6
MI 3105 EurotestXA Preface
6
1 Preface
Congratulations on your purchase of the instrument and its accessories from METREL. The instrument was designed on basis of rich experience, acquired through many years of dealing with electric installation test equipment.
The multifunctional hand-held installation tester EurotestXA is intended for all tests and measurements required for total inspection of electrical installations in buildings. In general the following measurements and tests can be performed:
True rms voltage and frequency, phase sequence, Insulation resistance, Resistance to earth connection and equipotential bonding plus continuous
resistance measurement,
Line impedance, Loop impedance, RCD protection, Resistance to earth, Leakage and load currents, Testing of Insulation Monitoring Devices (IMDs), First fault leakage current, Illuminance measurements, 2 Ω line/loop impedance, Tracing the installation, Overvoltage protection devices, Specific earth resistance measurement.
Tests can be performed on the following supply systems:
TN / TT, IT, 110 V reduced low voltage (2 x 55 V), and 110 V reduced low voltage (3 x 63 V).
The high-resolution graphic display with backlight offers easy reading of results, indications, measurement parameters and messages. Operation is simple and clear – operator does not need any special training (except reading this instruction manual) to operate the instrument.
In order for operator to be familiar enough with measurements in general and their typical
applications it is advisable to read Metrel handbook Measurements on electric installations in theory and practice.
The instrument is equipped with all accessories necessary for comfortable testing. It is kept in a soft carrying bag together with all accessories.
Page 7
MI 3105 EurotestXA Safety and operational considerations Warnings and notes
7
2 Safety and operational considerations
2.1 Warnings and notes
In order to reach high level of operator’s safety while carrying out various tests and measurements using EurotestXA, as well as to keep the test equipment undamaged, it is necessary to consider the following general warnings:
Warning on the instrument means »Read the Instruction manual with
special care to safety operation«. The symbol requires an action!
If the test equipment is used in a manner that is not specified in this user
manual, the protection provided by the equipment might be impaired!
Read this user manual carefully, otherwise use of the instrument may be
dangerous for the operator, for the instrument or for the equipment under test!
Do not use the instrument and accessories if any damage is noticed! In case a fuse has blown follow the instructions in this manual to replace it! Consider all generally known precautions in order to avoid risk of electric
shock while dealing with hazardous voltages!
Do not use the instrument in supply systems with voltages higher than
550 V!
Service intervention or adjustment and calibration procedure is allowed to be
carried out only by a competent authorized person!
Use only standard or optional test accessories supplied by your distributor! Consider that older and some of new optional test accessories compatible
with this instrument meet overvoltage category CAT III / 300 V! It means that maximum allowed voltage between test terminals and ground is 300 V!
Instrument contains rechargeable NiCd or NiMh battery cells. The cells
should only be replaced with the same type as defined on the battery placement label or in this manual. Do not use standard alkaline battery cells while power supply adapter is connected, otherwise they may explode!
Hazardous voltages exist inside the instrument. Disconnect all test leads,
remove the power supply cable and switch off the instrument before removing battery compartment cover.
All normal safety precautions have to be taken in order to avoid risk of
electric shock when working on electrical installations!
Warnings related to measurement functions:
Insulation resistance
Do not touch the test object during the measurement or before it is fully
discharged! Risk of electric shock!
Automatic discharge of capacitive object will take some time after the finished
insulation resistance measurement. Warning message and actual voltage is displayed during discharging until voltage drops below 10 V. In no case you should disconnect test leads until tested object is completely discharged!
Page 8
MI 3105 EurotestXA Safety and operational considerations Warnings and notes
8
Notes related to measurement functions: General
Indicator means that the selected measurement can't be performed because of
irregular conditions on input terminals.
Insulation resistance, varistor test, continuity functions and earth resistance
measurements shall be performed on de-energized objects, i.e. voltage between test terminals should be lower than 10 V!
PASS / FAIL indication is enabled when limit is set to ON. Apply appropriate limit
value for evaluation of measurement results.
In case that only two of three wires are connected to test electrical installation, only
voltage indication between these two wires is valid.
Insulation resistance
When measuring insulation resistance between installation conductors all loads
must be disconnected and all switches closed!
The instrument automatically discharge tested object after finished measurement. Keep the TEST key pressed for continuous measurement.
Continuity functions
Parallel resistance paths and interfering currents in measured circuit will influence
the test result!
If necessary compensate test lead resistance before performing continuity
measurement, see 5.2.3.
Measurement of resistance of wire wound components like transformer or motor
windings is possible only in continuous function (R7mA) due to great influence of
the winding inductance.
RCD functions
Parameters set in one function are also kept for other RCD functions. The measurement of contact voltage will not trip-out RCD of tested installation if
selected rated test current is the same as rated IΔN of observed RCD. However, the RCD trip-out may occur and Uc measurement is affected because of PE leakage currents caused by appliances that are connected to the tested installation.
The RCD trip-out current test and Uc measurement could be affected as a result
potential fields of other earthing installations.
RCD trip-out current and time will be measured only if pretest of those functions
gives contact voltage lower than the selected conventional limit contact voltage.
L and N test terminals are reversed automatically according to detected terminal
voltage.
In case the RCD trips-out during safety pretests it is possible to continue
measurements just by recovering the RCD. Possible reasons for trip-out are incorrect RCD sensitivity (IΔN) selected or relatively high leakage currents in tested installations or defective RCD.
Page 9
MI 3105 EurotestXA Safety and operational considerations Warnings and notes
9
Z-LOOP
Fault loop impedance measurement trips-out the RCD. Use the Z-LOOP
Impedance, Protection: RCD option to prevent the trip-out.
Fault loop impedance function with selected RCD protection takes longer time to
complete but offers much better accuracy then R
L
sub-result in RCD: Uc function.
Specified accuracy of tested parameters is valid only if mains voltage is stable
during the measurement and no additional operating circuits are connected in parallel.
L and N test terminals are reversed automatically according to detected terminal
voltage.
Z-LINE
Measurement of Z
Line-Line
with the instrument test leads PE and N connected
together will generate warning of dangerous PE voltage when the TEST key is
touched but measurement is not prohibited.
Specified accuracy of tested parameters is valid only if mains voltage is stable
during the measurement and no additional operating circuits are connected in parallel.
L and N test terminals are reversed automatically according to detected terminal
voltage.
Earth resistance
High currents and voltages in earthing could influence the measurement results. High resistance of S and H probes could influence the measurement results. In this
case, indications Rp and Rc appear in the message field. There is no pass / fail indication in this case.
Resistance of E measuring wire is added to the measurement result of resistance to
earth. Use only standard test accessory without extension lead for E probe.
In two clamps test the distance between clamps should be at least 30 cm (see
figure 5.34).
In one clamp test the accuracy decreases as the ratio R / Re increases!
Line tracer
Receiver R10K should always be in IND mode when working with the MI 3105
instrument.
When dealing with complex installations (long conductors or more current loops
connected in parallel), it is advisable to disconnect parts of the installation that are not of interest at that moment. Otherwise, the test signal will spread all over the installation and the selectivity can fall to an unacceptable level.
Page 10
MI 3105 EurotestXA Safety and operational considerations Battery and charging
10
2.2 Battery and charging
The instrument uses six AA size alkaline or rechargeable Ni-Cd or Ni-MH battery cells. Nominal operating time is declared for cells with nominal capacity of 2100 mAh. Battery condition is always present on the display when the instrument is turned on. In case the battery is weak, the instrument indicates this as shown in figure 2.1. This indication appears for a few seconds and then the instrument is turned off.
Figure 2.1: Discharged battery indication
The battery is charged whenever the power supply adapter is connected to the instrument. Internal circuit controls charging assuring maximum battery lifetime. Power supply socket polarity is shown in figure 2.2.
+
-
Figure 2.2: Power supply socket polarity
The instrument automatically recognizes connected power supply adapter and controls charging.
Symbols:
Indication of battery charging
7.2 Battery voltage
7.2
Figure 2.3: Charging indication
Before opening battery / fuse compartment cover disconnect all measuring
accessories connected to the instrument and power off the instrument.
Insert cells correctly, otherwise the instrument will not operate and the battery could
be discharged.
Remove all battery cells from the battery compartment if the instrument is not used
for longer period.
Do not charge alkaline battery cells!
Page 11
MI 3105 EurotestXA Safety and operational considerations Battery and charging
11
Take into account handling, maintenance and recycling requirements that are
defined by related regulatives and manufacturer of alkaline or rechargeable batteries!
Use only power supply adapter delivered from manufacturer or distributor of the test
equipment to avoid possible fire or electric shock!
2.2.1 New battery cells or cells unused for a longer period
Unpredictable chemical processes can occur during charging of new battery cells or cells that were unused for a longer period (more than 3 months). Ni-MH and Ni-Cd battery cells are affected to capacity degradation (sometimes called as memory effect). As a result, the instrument operation time can be significantly reduced.
Recommended procedure for recovering battery cells:
Procedure Notes
Completely charge the battery.
At least 14h with in-built charger.
Completely discharge the battery.
Can be performed with normal work with the instrument.
Repeat the charge / discharge cycle for
at least two times.
Four cycles are recommended.
Complete discharge / charge cycle is performed automatically for each cell using external intelligent battery charger.
Notes:
The charger in the instrument is a pack cell charger. This means that the battery
cells are connected in series during the charging. The battery cells have to be equivalent (same charge condition, same type and age).
One different battery cell can cause an improper charging and incorrect discharging
during normal usage of the entire battery pack (it results in heating of the battery pack, significantly decreased operation time, reversed polarity of defective cell,…).
If no improvement is achieved after several charge / discharge cycles, then each
battery cell should be checked (by comparing battery voltages, testing them in a cell charger, etc). It is very likely that only some of the battery cells are deteriorated.
The effects described above should not be mixed with normal decrease of battery
capacity over time. Battery also loses some capacity when it is repeatedly charged / discharged. Actual decreasing of capacity, versus number of charging cycles, depends on battery type. It is provided in the technical specification from battery manufacturer.
Page 12
MI 3105 EurotestXA Safety and operational considerations Standards applied
12
2.3 Standards applied
The MI 3105 EurotestXA instrument is manufactured and tested according to the following regulations, listed below.
Electromagnetic compatibility (EMC)
EN 61326
Electrical equipment for measurement, control and laboratory use – EMC requirements Class B (Hand held equipment used in controlled EM environments)
Safety (LVD)
EN 61010 - 1
Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
EN 61010 - 031
Safety requirements for hand-held probe assemblies for electrical measurement and test
Functionality
EN 61557
Electrical safety in low voltage distribution systems up to 1000 V a.c. and 1500 V d.c. - Equipment for testing, measuring or monitoring of protective measures
Part 1 General requirements Part 2 Insulation resistance Part 3 Loop resistance Part 4 Resistance of earth connection and equipotential bonding Part 5 Resistance to earth Part 6 Residual current devices (RCDs) in TT and TN systems Part 7 Phase sequence Part 10 Combined measuring equipment
Other reference standards for testing RCDs
EN 61008
Residual current operated circuit-breakers without integral overcurrent protection for household and similar uses
EN 61009
Residual current operated circuit-breakers with integral overcurrent protection for household and similar uses
EN 60755 General requirements for residual current operated protective devices
EN 60364-4-41
Electrical installations of buildings - Part 4-41: Protection for safety -
Protection against electric shock BS 7671 IEE Wiring Regulations AS / NZ 3760 In-service safety inspection and testing of electrical equipment
Note about EN and IEC standards:
Text of this manual contains references to European standards. All standards of EN 6xxxx (e.g. EN 61010) series are equivalent to IEC standards with the same number (e.g. IEC
61010) and differ only in amended parts required by European harmonization procedure.
Page 13
MI 3105 EurotestXA Instrument description Front panel
13
3 Instrument description
3.1 Front panel
Figure 3.1: Front panel
Legend:
Switches the instrument power on or off.
1 ON / OFF
The instrument automatically turns off 15 minutes after the last key was pressed.
2 HELP Accesses help menus.
Adds new memory location.
3 F2
Confirmation of name entered in edit mode. Enters memory editing mode.
4 F1
Deletes character on the left in edit mode. 5 MEM Handling with memory. 6 ESC Exits selected and displayed option. 7 TAB Jumps between display windows.
Cursors Selection of tested function and its working parameters.
Initiates measurements.
8
Cursor keypad with TEST key
TEST
Acts also as the PE touching electrode.
9
BACKLIGHT, CONTRAST
Changes backlight level and contrast.
10 LCD 320 x 240 dots matrix display with backlight.
Page 14
MI 3105 EurotestXA Instrument description Connector panel
14
3.2 Connector panel
1
3
2
> 550V
4
5
6
Figure 3.2: Connector panel
Legend:
1 Test connector Measuring inputs / outputs, connection of measuring cables. 2 Charger socket Connection of power supply adapter.
3 PS/2 connector
Communication with PC serial port and connection to optional measuring adapters.
4 Protection cover
Protects from simultaneous access to test connector and power
supply adapter socket plus communication connectors. 5 USB connector USB (1.1) communication port. 6 Clamp connector Measuring input for current clamp.
Warnings!
Maximum allowed voltage between any test terminal and ground is 600 V! Maximum allowed voltage between test terminals is 550 V! Maximum short-term voltage of external power supply adapter is 14 V! Do not connect any voltage source on clamp connector sockets! It is
intended for connection of current clamp with current output only.
Maximum continuous current of current clamp input is 30 mA!
Page 15
MI 3105 EurotestXA Instrument description Back panel
15
3.3 Back panel
1
2
3
Figure 3.3: Back panel
Legend:
1 Battery / fuse compartment cover 2 Back panel information label 3 Fixing screws for battery / fuse compartment cover
5
6
4
1
2
3
F1
F3
F2
Fuse
Fuse
Fuse
SIZE AASIZE AA
SIZE AA
SIZE AASIZE AASIZ E AA
S/N XXXXXXX X
-
+
Figure 3.4: Battery and fuse compartment
Legend:
1 Fuse F1 T 315 mA / 250 V 2 Fuse F2 T 4 A / 500 V 3 Fuse F3 T 4 A / 500 V 4 Serial number label 5 Battery cells Size AA, alkaline / rechargeable NiMH or NiCd 6 Battery holder Can be removed from the instrument
Page 16
MI 3105 EurotestXA Instrument description Bottom
16
3.4 Bottom
3
1
2
R: 0.0 1999 Test current: max. 8.5mA Open-circuit voltage: 6.5V

Continuity 7mA



R: 0.18M 199.9M , U =50V ,100 , R: 0.12M 999M , U = 500V , 1kV U: 0V 1200V Nominal voltages: 100V , 250V , 500V , 1kV Measuring current: min. 1mA at R =U 1k /V Short-circuit current: < 3mA
 

N
N
NN


Insulation resistance (EN 61557-2)
V250V

Continuity
Tripping time
non-delayed (time-delayed) RCDs
  
1: 0ms 300ms (500ms) 2: 0ms 150ms (200ms) 5: 0ms 40ms (150ms), U : 0.0V 100.0V
C
Tripping curre nt
I : 0.2 I 1.1 I AC ( t : 0ms 300ms, U : 0.0V 100.0V



C
 
NN N

1.5 I A)
Multiplier: 0.5, 1, 2, 5

Cont act v oltag e
U : 0.0V 100.0V R : 0.0 0 10.0 0k (R =U / I )
C
SSCN
 
R : 0. 1999 I : 0. A . kA
L-N(L) PSC
17 20 1 4
Nominal voltage: 100V 440V/ 15Hz 500Hz


Line impedance(EN 61557-3 )

Fault loop (EN 61557-3)
R : 0. 1999 I : 0. A .kA

PFC
14 1 4
Nominal voltage: 100V 264V/ 15Hz 500Hz

impedance
L-PE
17

Voltage, frequency
U: 0V 440V / f: 15Hz 500Hz

Phase rotation (EN 61557-7)
Nominal voltage: 100V 440V / 1 00 Results: 1.2.3 or 2.1.3

5Hz 5 Hz
RCD (EN 61557-6)
I : 10mA, 30mA, 100mA, 300mA, 500mA, 1A Nominal voltage: 100V 264V/ 15Hz 500Hz
 
R Low (EN 61557-4)
 

R: 0.12 1999 Test current: min. ±200mA at 2 Open-circuit voltage: 6.5V 9.0V
 
Resistance to earth (EN 61557-5)
R: 0.04 9999 Open-circuit voltage : Short-circuit current

< 45V
: < 20mA
RMS
Ljubljanska 77 SI - 1354 Horjul Tel: +386 1 75 58 200 http://www.metrel.si
20 224 83 2
CAT III 600V
550V
Figure 3.5: Bottom
Legend:
1 Bottom information label 2 Neck belt openings 3 Handling side covers
Page 17
MI 3105 EurotestXA Instrument description Display organization
17
3.5 Display organization
Figure 3.6: Typical single test
display
Menu line
Result field
Test parameter field
Message field
Terminal voltage monitor
Function tabs
3.5.1 Terminal voltage monitor
The terminal voltage monitor displays current voltages present on the test terminals. In its lower part, messages are displayed regarding the measured voltages and selected voltage
system (see 4.4.2 Supply systems).
Online voltage is displayed together with test terminal indication.
L and N test terminals are used for selected measurement.
L and PE are test terminals; N terminal should also be connected for reference in measuring circuit.
Polarity of test voltage applied to the output terminals.
Insulation test: two measuring terminals should be shorted.
,
Three-phase connection indicator.
TT / TN supply system.
IT supply system.
Reduced low voltage supply system.
Unknown supply system (atypical voltage at input terminals for selected supply system).
L – N polarity changed.
First fault in IT supply system. Check monitored voltages to fix the problem.
Page 18
MI 3105 EurotestXA Instrument description Display organization
18
Warning! Phase voltage on the PE terminal! Stop the activity immediately
and eliminate the fault / connection problem before proceeding with any activity!
3.5.2 Menu line
In the menu line, the name of the selected function is displayed. Additional information about active cursor / TEST keys and battery condition are shown.
Function name. Time.
Active keys on cursor / TEST keypad ( and TEST in this
example).
Battery capacity indication.
Low battery. Battery is too weak to guarantee correct result. Replace or recharge the battery cells.
Recharging in progress (if power supply adapter is connected).
3.5.3 Message field
In the message field, different warnings and messages are displayed.
Warning! High voltage is applied to the test terminals.
Measurement is running; consider displayed warnings.
Conditions on the input terminals allow starting the measurement (TEST
key); consider other displayed warnings and messages. Conditions on the input terminals do not allow starting the measurement
(TEST key), consider displayed warnings and messages.
Test leads resistance in CONTINUITY tests is not compensated, see
Chapter 5.2.3 for compensation procedure.
Test leads resistance in CONTINUITY tests is compensated.
RCD tripped-out during the measurement (in RCD functions).
Instrument is overheated, the temperature inside the instrument is higher than the safety limit, and measurement is prohibited until the temperature decreases under the allowed limit.
Fuse F1 has blown or not inserted (CONTINUITY and EARTH functions).
It is possible to store result(s).
High electrical noise during measurement. Results may be impaired.
Probe resistances Rc or Rp could influence earth resistance result.
Page 19
MI 3105 EurotestXA Instrument description Display organization
19
Low clamp current could influence earth resistance result.
Pause activated in auto sequence test. Follow required activity for paused test function.
3.5.4 Result field
Measurement result is inside pre-set limits (PASS).
Measurement result is out of pre-set limits (FAIL).
Measurement is aborted. Consider displayed warnings and messages.
3.5.5 Other messages
Hard Reset
Instrument settings and measurement parameters/limits are set to initial (factory) values; for more information refer to chapter 4.8.5.
Recalling original settings.
CAL ERROR!
Service intervention required.
3.5.6 Sound warnings
Periodic sound
Warning! Dangerous voltage on the PE terminal is detected. Refer to chapter 5.8 for more information.
3.5.7 Help
Key:
HELP
Opens help screen.
The help menus contains some basic schematic / connection diagrams to illustrate recommended connection of the instrument to the electrical installation and information about the instrument.
Pressing the HELP key in single test generates help screen for selected single test
function, while in other working menus the voltage system help is displayed first.
Keys in help menu: / Select neighbour help screen.
HELP
Rotates through help screens.
ESC
Exits help menu.
Page 20
MI 3105 EurotestXA Instrument description Display organization
20
Figure 3.7: Examples of help screen
3.5.8 Backlight and contrast adjustments With the BACKLIGHT key backlight and contrast can be adjusted.
Click
Toggle backlight intensity level.
Keep pressed for 1 s
Lock high intensity backlight level until power is turned off or the key is pressed again.
Keep pressed for 2 s Bargraph for LCD contrast adjustment is displayed.
Figure 3.8: Contrast adjustment menu
Keys for contrast adjustment:
Reduces contrast.
Increases contrast.
TEST
Accepts new contrast.
ESC
Exits without changes.
Page 21
MI 3105 EurotestXA Instrument description Carrying the instrument
21
3.6 Carrying the instrument
With the neck-carrying belt supplied in standard set, various possibilities of carrying the instrument are available. Operator can choose appropriate one on basis of his operation, see the following examples:
The instrument hangs around operators neck only - quick placing and displacing.
The instrument can be used even when placed in soft carrying bag – test cable connected to the instrument through the front aperture.
3.7 Instrument set and accessories
3.7.1 Standard set
Instrument Soft carrying bag Short instruction manual Product verification data Warranty declaration Declaration of conformity Universal test cable Three test tips Schuko plug commander
Three alligator clips Current clamp Power supply adapter CD with instruction manual,
handbook Measurements on electric installations in theory and practice, PC software
USB interface cable RS232 interface cable
3.7.2 Optional accessories
See the attached sheet for a list of optional accessories that are available on request from your distributor.
Page 22
MI 3105 EurotestXA Instrument operation Main menu & Single test
22
4 Instrument operation
4.1 Main menu
From the Main menu different instrument operation modes can be set.
 Single test menu (see 4.2),  Auto sequence menu (see 4.3),  Miscellaneous (see 4.4).
Figure 4.1: Main menu
Keys: / Select the mode.
TEST
Enters selected mode.
4.2 Single test
is intended to run individual test /
measurement functions.
Figure 4.2: Example of typical
Single test screen
Page 23
MI 3105 EurotestXA Instrument operation Single test
23
Keys in main single test screen:
/
Select test / measurement function:
<VOLTAGE> Voltage and frequency plus phase sequence. <CONTINUITY> Resistance to earth connection and equipotential
bonding.
<INSULATION> Insulation resistance. <Z-LINE> Line impedance. <Z-LOOP> Fault loop impedance. <RCD> RCD testing. <EARTH> Resistance to earth. <CURRENT> Clamp current. <SENSOR> Illumination. <VARISTOR TEST> Transient suppressor test.
The following functions are available only when the IT supply system is
selected (see chapter 4.4.2):
<IMD check> Insulation monitor device testing. <ISFL> Measurement of first fault leakage current.
/
Select sub-function in selected measurement function.
TEST
Runs selected test / measurement function.
TAB
Enters test parameters field.
ESC
Exits single test operation mode.
MEM
Stores measured results / recalls stored results.
Keys in test parameter field:
/ Select measuring parameter.  / 
Change the selected parameter.
TEST, TAB, ESC
Exit back to main single test screen.
General rule for enabling limits for evaluation of measurement / test result:
OFF
No limit comparison
Limit ON – enabled comparison Limit
ON
Limit Value – minimum / maximum limit value *
* Type of limit value depends on particular function.
See Chapter 5 for more information about operation of the instrument in single test
functions.
Page 24
MI 3105 EurotestXA Instrument operation Automatic testing
24
4.3 Automatic testing
is intended for automatic executing of predefined measurement sequences.
Figure 4.3: Typical auto
sequence screen
Auto sequence menu. Selected sequence number and (optional) name.
Sequence field.
Test parameter / auto sequence description field.
Saving and renaming options.
Running auto sequence:
Select auto sequence (see 4.3.2).
Connect the instrument to tested object as required for the first measurement in the
sequence.
Press TEST key. The sequence will pause at the functions marked with pause flag
II
.
Comments
regarding the paused function will be displayed (optional).
Press the TAB key to toggle between comments menu and auto sequence main
menu.
If the conditions at input terminals are valid, the test will proceed after pressing
the TEST key.
Press the F1 key to skip the paused function. The test will continue with the
next test (if any) or will end.
Press the ESC key to skip the remaining functions and finish the auto
sequence.
The set of measurements will be performed in sequential manner until the
conditions at input terminals are valid for each individual test. If not, the instrument will stop (the buzzer sounds). The auto sequence will proceed:
After correct conditions are restored at the input terminal (e.g. by reconnecting,
switching on the RCD).
If pressing the F1 key this function will be skipped. By pressing the ESC key to skip the remaining functions and finish the auto
sequence.
Results of a finished auto sequence can be viewed and stored. See chapter 6. for
more information.
Page 25
MI 3105 EurotestXA Instrument operation Automatic testing
25
Measurements are marked with one of the following symbol after finished test.
Measurement is finished and has failed. Measurement is finished and has passed. Measurement is finished. No comparison limit was applied. Measurement is not performed yet (during test) or was skipped.
Overall PASS result is reported if all performed tests passed.
Overall FAIL result is reported if one or more performed tests failed.
Figure 4.4: Waiting for right input condition to proceed
Figure 4.5: Overall PASS example
Figure 4.6: Overall FAIL example
Viewing auto sequence particular results:
After finished auto sequence press the key to move focus into sequence field.
Press TEST key. Result of selected function is displayed. Press the key (or ) to select the next function of the sequence. Repeat this part until all results are observed.
Viewing of the results is finished by pressing the key until selected sequence
number is focused or by pressing the ESC key.
Page 26
MI 3105 EurotestXA Instrument operation Automatic testing
26
4.3.1 Auto sequence number main menu
In the instrument up to 99 auto sequences can be stored.
#3 Auto sequence number. * Indicator that the preset sequence was changed and is
not stored yet, the auto sequence can be performed anyway.
TEST EXA1
Optional sequence name (see 4.3.4).
Indication of locked sequence (see 4.3.2).
4.3.2 Auto sequence set
Keys in main auto sequence menu:
TEST
Starts the selected test sequence.
 /  Select the test sequence number or measuring function (see 4.3.1).  /  Select individual sequence step / measuring function.
TAB
Enters test parameter field (see 4.3.3).
ESC
Exits auto sequence menu without changes.
F1
Enters editor for renaming selected test sequence and entering its
description (see 4.3.4).
Enters menu for
setting pause flag and comments (see 4.3.7).
F2
Saves entered test sequence (see 4.3.5).
MEM
Stores / recalls auto sequence results.
Function selection
Parameter selection
Figure 4.7: Examples of setting up auto sequence
For each of 6 predefined sequence steps any of the following measurement function can be selected: voltage, continuity, insulation, Zline, Zloop, RCD and earth. The field can also
be left empty (- - -).
Test parameters are applied to individual measurements as in the single test. The test parameter menu of selected measurement is available on the right side of the display.
The pause
II
flag holds the auto sequence until prosecution is confirmed with the TEST
key. It is recommended to use it if additional checks or reconnections have to be performed before performing the next measurement.
Page 27
MI 3105 EurotestXA Instrument operation Automatic testing
27
The key is indication of locked sequence. This indication appears at predefined sequences that were loaded into the instrument from PC. It is possible to modify locked auto sequences and run them. However, the modified sequence cannot be stored by overwriting.
Note:
It is recommended to save current auto sequence if modified or new prepared, to
keep it during manipulation.
4.3.3 Test parameters in auto sequence
Keys in test parameter menu (in auto sequence):
/ Select test parameter value or enable / disable parameter.  / Select test parameter.
TEST, TAB, ESC
Return to auto sequence main screen.
Whenever a new function is selected for auto sequence its test parameters should be verified and changed to appropriate values / states.
Test parameter merging
When the prepared sequence from section 4.3.2 contains selected at least two of Zline,
Zloop, or RCD, is possible to merge test parameters of one function to others of mentioned in the same sequence. Merged parameters are related to:
- fuse data, and
- RCD data, except start polarity of test current.
Additional key in main auto sequence menu with selected Zline, Zloop, or RCD:
F2
Merges test parameters.
Figure 4.8: Parameter merging possibility
4.3.4 Name and description of auto sequence F1
Enters test sequence name menu from auto sequence main menu.
Page 28
MI 3105 EurotestXA Instrument operation Automatic testing
28
Name and description for the selected auto sequence can be added or changed (optional) in this two level menu.
Figure 4.9: Auto sequence name
menu
Keys for 1st level:
/ Select between name and description field. TEST
Returns to auto sequence main menu.
F1
Enters editing of selected field (2nd level).
ESC
Returns to auto sequence main menu without changes.
Figure 4.10: Auto sequence name edit
menu
Figure 4.11: Auto sequence description edit
menu
Keys for 2nd level:
Highlighted key
Selected symbol or activity.
// / Select symbol or activity. TEST
Enters selected symbol or performs selected activity.
F1
Deletes last entered symbol in the name line.
F2
Confirms name and returns to 1
st
level of auto sequence name
menu.
ESC
Returns to 1st level of auto sequence name menu without changes.
20 characters is the maximum length of the auto sequence name. 100 characters is the maximum length of the auto sequence description.
Page 29
MI 3105 EurotestXA Instrument operation Automatic testing
29
4.3.5 Storing auto sequence settings (sequence, number, name) F2
Opens dialog for storing auto sequence settings in auto sequence main menu.
The dialog enables storing existing auto sequence settings into different location or overwriting existing.
Figure 4.12: Store dialog
Keys:
/
Select the auto sequence number.
TEST
Confirms the storing.
ESC
Returns to auto sequence main menu without changes.
Auto sequence settings are stored in nonvolatile memory. Stored auto sequence procedures remain in memory until the user changes them.
It is not possible to store any auto sequence in locked location. Locked auto sequence can be copied in an unlocked location. Stored sequence is unlocked in this case.
Figure 4.13: Store dialog for locked sequence
Figure 4.14: Failed storing
It is possible to unlock all locked sequences if necessary (see 4.4.5 for more information).
Page 30
MI 3105 EurotestXA Instrument operation Automatic testing
30
4.3.6 Pause flag and comments in auto sequence
The auto sequence holds if a pause flag is associated with the measurement and the pre­defined comment is displayed. When the input conditions are regular, the auto sequence
can be continued by pressing the TEST key.
Comment appears with the pause
Blinking pause flag in main screen
Figure 4.15: Examples of screens during the pause in auto sequence
Keys:
TAB
Toggles between comment screen and auto sequence main screen.
TEST
Continues with the paused test.
F1
Skip paused test.
ESC
Skip all tests and ends auto sequence.
4.3.7 Setting pause flag and comments
Operator of the instrument can prepare comments regarding the measurements. Warnings, reconnection hints or other useful remarks related to the test sequence can be applied this way.
F1
Enters pause set-up and comments menu for selected function in auto sequence main menu.
Set-up of comments is enabled if pause flag is set to ON.
Figure 4.16: Pause set-up menu
Page 31
MI 3105 EurotestXA Instrument operation Automatic testing
31
Keys:
/
Enable (ON) / disable (OFF) pause flag.
/ Select between pause flag and comments fields. TEST
Confirms pause and comment selection, and returns to auto sequence main menu.
ESC
Returns to auto sequence main menu without changes.
Comments set-up menu enables selection and editing of the pause comment.
Figure 4.17: Comments set-up
menu
Keys:
/ Select between setup of pause and comment.  / 
Select comment [--- (no comment), #1 ÷ #99].
F1
Enters edit comments menu for selected comment number.
TEST
Confirms pause and comment selection and returns to auto sequence main menu.
ESC
Returns to auto sequence main menu without changes.
Comments can be entered and edited in the Edit comments menu.
Maximum comment length: 250 characters (including space and new line characters).
Figure 4.18: Comments edit
menu
Keys:
Highlighted key
Selected symbol or activity.
// / Select symbol or activity. TEST
Enters selected symbol or performs selected activity.
F1
Deletes last entered symbol in the name line.
F2
Opens dialog for comment storing.
ESC
Deletes comment (immediately after entering the editor). Returns to auto sequence main menu without changes.
Page 32
MI 3105 EurotestXA Instrument operation Automatic testing
32
Storing comment opens dialog for storing into selected location.
Figure 4.19: Store dialog for
comment
Keys:
/ Select comment number. TEST
Confirms storing the comment and returns.
ESC
Returns to Edit comments menu.
Note:
It is not possible to overwrite comments associated to locked auto sequences.
4.3.8 Building an auto sequence
The instrument supports up to 99 auto sequences, each consisting of up to 6 steps. It is not necessary that all steps are enabled. The auto sequence can be prepared in the following ways:
By storing the existing auto sequence under another auto sequence number (see
4.3.5),
By changing an existing auto sequence and saving it under the same auto
sequence number (not possible for locked auto sequence),
By building a new auto sequence.
Building a new auto sequence
In the main menu (see 4.1) select auto sequence. Press the TEST key. Select auto sequence number (see 4.3.2).
Repeat until finished (maximum 6 steps):
Select auto sequence step (see 4.3.2). Select auto sequence function (see 4.3.2). Select auto sequence test parameters of the function (see 4.3.3).
Set / reset pause flag II and select or create new comment if necessary (see
4.3.7).
 Name (or rename) the auto sequence and enter its description (see 4.3.4).  Save prepared auto sequence (see 4.3.5).
Page 33
MI 3105 EurotestXA Instrument operation Automatic testing
33
Figure 4.20: Blank auto sequence
Example of building an auto sequence
A house installation wall socket protected with fuse (type gG, In = 6 A, td = 5 s) and RCD (type AC, IΔN = 30 mA) shall be tested. The following measurements must be performed:
 Equipotential bonding resistance of PE terminal to main PE collector (R 0.1 Ω),  Insulation resistances between L – N, L – PE and N – PE (U = 500 V, R 1 MΩ),  Voltages on the socket,  Line impedance with fuse verification,  RCD trip-out time at nominal current,  RCD trip-out time at increased current (5 x I
ΔN
). The name of test sequence number 10 is “Sock. 6A / 30mA(AC)”. Description of the test sequence is: “Verification of wall socket, protected with fuse and RCD”.
For the measurement the following conditions shall apply:
Equipotential bonding resistance and insulation resistance measurement has to be
performed on de-energized socket;
Equipotential bonding resistance test (see figure 5.6) should be performed with the
universal test cable and extension lead;
Insulation resistance test should be performed with the plug cable or commander
(see figures 5.2 and 5.3);
Other tests have to be applied on energized test socket with the plug cable or
commander (see figures 5.13, 5.22 and 5.26).
Example:
Item/keys
Chapter reference
Comment
Auto sequence, TEST
4.1
Selection of auto sequence operation in main menu.
/ 4.3.1 Selection of test sequence number 10. F1
4.3.4 Enter into sequence name editing menu.
F1
4.3.4 Enter the sequence name editor.
Sock. 6A / 30mA(AC)
4.3.4 Enter the name of auto sequence.
F2
4.3.4 Accept name and exit into sequence name editing menu.
4.3.4 Select description of test field.
F1
4.3.4 Enter the description of test editor.
Verification of wall socket, protected with fuse and RCD
4.3.4 Enter the description.
Page 34
MI 3105 EurotestXA Instrument operation Automatic testing
34
F2
Accept description and exit into sequence name
editing menu.
TEST
4.3.4 Exit sequence name editing menu.
4.3 Enter into sequence field.
/ 4.3.2 Select CONTINUITY. TAB
4.3.2 Enter test parameter selection mode.
TEST R200mA Limit ON Limit 0.1Ω
5.2
Set test parameters for equipotential bonding resistance.
TAB
4.3.2 Exit parameter mode.
F1
4.3.2 Set PAUSE (wait to prepare for measurement).
 /  4.3.7 Set PAUSE: ON. 
4.3.7 Select COMMENT.
4.3.7 Select COMMENT: #1.
F1
4.3.7 Enter Edit comment menu.
Disconnect mains, univ. cable + ext.
4.3.7 Enter the comment.
F2
4.3.7 Save the comment.
TEST
4.3.7 Store the comment to location #1.
4.3.7 Select COMMENT: #2.
F1
4.3.7 Enter Edit comment menu.
Commander
4.3.7 Enter the comment.
F2
4.3.7 Save the comment.
TEST
4.3.7 Store the comment to location #2.
4.3.7 Select COMMENT: #3.
F1
4.3.7 Enter Edit comment menu.
Connect mains
4.3.7 Enter the comment.
F2
4.3.7 Save the comment.
TEST
4.3.7 Store the comment to location #3.
4.3.7 Select COMMENT: #4.
F1
4.3.7 Enter Edit comment menu.
Turn ON RCD
4.3.7 Enter the comment.
F2
4.3.7 Save the comment.
TEST
4.3.7 Store the comment to location #4.
(3 x) 4.3.7 Select COMMENT: #1. TEST
4.3.7 Confirm selected pause and its comment.
4.3 Next step.
/ 4.3.2 Select INSULATION. TAB
4.3.2 Enter test parameter selection mode.
TEST ALL UISO 500 V Limit ON Limit 1MΩ
5.1 Setting test parameters for insulation resistance.
TAB
4.3.2 Exit parameter mode.
F1
4.3.2 Set PAUSE (wait to reconnect measuring leads).
 /  4.3.7 Set PAUSE: ON. 
4.3.7 Select COMMENT.
(2 x) 4.3.7 Select COMMENT: #2. TEST
4.3.7 Confirm selected pause and its comment.
4.3 Next step.
Page 35
MI 3105 EurotestXA Instrument operation Automatic testing
35
/ 4.3.2 Select VOLTAGE. F1
4.3.2 Set PAUSE (wait to connect mains voltage).
 /  4.3.7 Set PAUSE: ON. 
4.3.7 Select COMMENT.
(3 x) 4.3.7 Select COMMENT: #3. TEST
4.3.7 Confirm selected pause and its comment.
4.3 Next step.
/ 4.3.2 Select Z-LINE. TAB
4.3.2 Enter test parameter selection mode.
FUSE type gG FUSE I 6A FUSE T 5s
5.5
Set test parameters for line impedance and fuse test.
TAB
4.3.2 Exit parameter mode.
4.3 Next step.
/ 4.3.2 Select RCD. TAB
4.3.2 Enter test parameter selection mode.
TEST Tripout current Idn 30mA
type G
Ulim 50V
5.3
Test parameters for RCD trip-out current test
(results of this test are also contact voltage at IΔ and trip out time).
TAB
4.3.2 Exit parameter mode.
4.3 Next step.
F1
4.3.2 Set PAUSE (wait to activate RCD).
 /  4.3.7 Set PAUSE: ON. 
4.3.7 Select COMMENT.
(4 x) 4.3.7 Select COMMENT: #4. TEST
4.3.7 Confirm selected pause and its comment.
/ 4.3.2 Select RCD. TAB
4.3.2 Enter test parameter selection mode.
TEST Tripout time t Idn 30mA
type G MUL x5
Ulim 50V
5.3
Test parameters for RCD trip out time test at 5I
Δ
N
(result of this test is also contact voltage at I
Δ
N
).
TAB
4.3.2 Exit parameter mode.
(6 x) 4.3 Exit the editing of sequence field. F2
4.3.5 Store prepared test sequence.
TEST
4.3.5 Confirm storing.
Figure 4.21: Auto sequence screen of the example above
Page 36
MI 3105 EurotestXA Instrument operation Miscellaneous
36
4.4 Miscellaneous
Different instrument options can be set in the
menu.
Options are:
Selection of language, Selection of mains supply system, Recalling and clearing stored results, Setting date and time, Selection of communication port, Setting the instrument to initial values, Entering locator function, Selection of operator.
Figure 4.22: Options in Miscellaneous
menu
Keys:
/ / /
Selection of option.
TEST
Enters selected option.
ESC
Returns to the main menu.
4.4.1 Language
The instrument supports different languages.
Figure 4.23: Language selection
Keys:
/
Select language.
TEST
Confirms selected language and exits to settings menu.
ESC
Exits to settings menu without changes.
Page 37
MI 3105 EurotestXA Instrument operation Miscellaneous
37
4.4.2 Supply system, Isc factor, RCD standard In the Voltage system menu the following parameters can be selected:
Voltage system
Mains supply system type.
Set Isc factor
Correction factor for Isc calculation (ksc).
RDC testing
RCD normative reference.
Figure 4.24: System parameters
Keys:
/
Select option.
/
Change the option.
TEST
Confirms selected option.
ESC
Exits to settings menu with new setup.
Mains supply systems
The following supplying systems are supported:
TT / TN (earthed systems), IT (system insulated from earth),
110 V reduced low voltage (2 x 55 V center tap grounded), 110 V reduced low voltage (3 x 63 V three phase, star center grounded).
TN, TT and IT systems are defined in EN 60364-1 standard. 110 V reduced low voltage systems are defined in BS 7671.
See Appendix D for particular characteristics of IT supply system measurements and
instrument characteristics.
See Appendix E for particular characteristics of 110 V reduced low voltage supply systems
measurements and instrument characteristics.
Isc factor - ksc
Short circuit current Isc in the supply system is important for selection or verification of protective circuit breakers (fuses, over-current breaking devices, RCDs). The default value of ksc is 1.00. Change the value as required by local regulative for tested type of mains supply system. Range for adjustment of the ksc is 0.20 ÷ 3.00.
Page 38
MI 3105 EurotestXA Instrument operation Miscellaneous
38
RCD normative references
Maximum RCD disconnection times differ in various standards. The trip-out times defined in individual standards are listed below.
Trip-out times according to EN 61008 / EN 61009:
½×I
ΔN
*)
I
ΔN
2×IΔN 5×IΔN General RCDs (non-delayed)
t
Δ
> 300 ms tΔ < 300 ms tΔ < 150 ms tΔ < 40 ms
Selective RCDs (time-delayed)
t
Δ
> 500 ms 130 ms < tΔ < 500 ms 60 ms < tΔ < 200 ms 50 ms < tΔ < 150 ms
Trip-out times according to EN 60364-4-41:
½×I
ΔN
*)
I
ΔN
2×IΔN 5×IΔN General RCDs (non-delayed)
t
Δ
> 999 ms tΔ < 999 ms tΔ < 150 ms tΔ < 40 ms
Selective RCDs (time-delayed)
t
Δ
> 999 ms 130 ms < tΔ < 999 ms 60 ms < tΔ < 200 ms 50 ms < tΔ < 150 ms
Trip-out times according to BS 7671:
½×I
ΔN
*)
I
ΔN
2×IΔN 5×IΔN General RCDs (non-delayed)
t
Δ
> 1999 ms tΔ < 300 ms tΔ < 150 ms tΔ < 40 ms
Selective RCDs (time-delayed)
t
Δ
> 1999 ms 130 ms < tΔ < 500 ms 60 ms < tΔ < 200 ms 50 ms < tΔ < 150 ms
Trip-out times according to AS/NZ
**)
:
½×I
ΔN
*)
I
ΔN
2×IΔN 5×IΔN
RCD type
I
ΔN
[mA] tΔ t
Δ
t
Δ
t
Δ
Note
I
10
40 ms 40 ms 40 ms
II
> 10 30
300 ms 150 ms 40 ms
III > 30
> 999 ms
300 ms 150 ms 40 ms 500 ms 200 ms 150 ms
Maximum break time
IV
S
> 30 > 999 ms
130 ms 60 ms 50 ms Minimum non-actuating time
*)
Minimum test period for current of ½×IΔN, RCD shall not trip-out.
**)
Test current and measurement accuracy correspond to AS/NZ requirements.
Maximum test times related to selected test current for general (non-delayed) RCD Standard
½×IΔN I
ΔN
2×IΔN 5×IΔN
EN 61008 / EN 61009 300 ms 300 ms 150 ms 40 ms EN 60364-4-41 1000 ms 1000 ms 150 ms 40 ms BS 7671 2000 ms 300 ms 150 ms 40 ms AS/NZ (I, II, III) 1000 ms 1000 ms 150 ms 40 ms
Maximum test times related to selected test current for selective (time-delayed) RCD Standard
½×IΔN I
ΔN
2×IΔN 5×IΔN
EN 61008 / EN 61009 500 ms 500 ms 200 ms 150 ms EN 60364-4-41 1000 ms 1000 ms 200 ms 150 ms BS 7671 2000 ms 500 ms 200 ms 150 ms AS/NZ (IV) 1000 ms 1000 ms 200 ms 150 ms
Page 39
MI 3105 EurotestXA Instrument operation Miscellaneous
39
4.4.3 Memory
In this menu, the stored data can be recalled, viewed
and cleared. See chapter 6 Data handling for more
information.
Figure 4.25: Memory options
Keys: / Select option.
ESC
Exits this option.
TEST
Enters selected option.
4.4.4 Date and time
Date and time can be set in this menu.
Figure 4.26: Setting date and time
Keys:
Selects the field to be changed.
/ Modify selected field. ESC
Exits date and time setup without changes.
TEST
Confirms new setup and exits.
Page 40
MI 3105 EurotestXA Instrument operation Miscellaneous
40
4.4.5 Initial settings
Instrument settings and measurement parameters and limits are set to their initial values in this menu.
Figure 4.27: Initial settings dialogue
Keys:
TEST
Restores default settings.
ESC
Exits the menu without changes.
F2
Opens other settings menu.
Warning:
Custom made settings will be lost when this option is used!
The default setup is listed below:
Instrument setting Default value
Contrast As defined and stored by adjustment procedure Isc factor 1.00 Supply system TN / TT RCD standards EN 61008 / EN 61009 COM port RS 232 Language English
Function
Sub-function
Parameter / limit value
CONTINUITY R 200 mA
R LOWΩ
High limit resistance value: OFF
Continuity High limit resistance value: OFF
INSULATION Nominal test voltage: 500 V
Low limit resistance value: OFF
Selected test leads combination: LN Z - LINE Fuse type: none selected Z - LOOP Circuit breaker: Fuse
Fuse type: none selected RCD RCD t
Nominal differential current: I
ΔN
=30 mA RCD type: G Test current starting polarity: (0°) Limit contact voltage: 50 V Current multiplier: ×1
Page 41
MI 3105 EurotestXA Instrument operation Miscellaneous
41
Earth resistance 3-wire
3-wire Limit value: OFF One clamp Limit value: OFF Two clamps Limit value: OFF
Specific resistance Distance unit: m Current Limit value: OFF Sensors - illumination Limit value: OFF Adapters – 2 Ω line/loop impedance
mΩ L-N
Fuse type: none selected IMD test Limit value: OFF ISFL Limit value: OFF Varistor test Lo limit: 300 V
Hi limit: 400 V
Other settings
F2
Enters menu to unlock the protected auto sequences and comments and/or select distance unit for specific earth resistance measurement.
Unlock protected sequences or distance unit can be selected.
Figure 4.28: Other settings dialogue
Keys:
/
Select other settings item.
TEST
Enters selected item.
ESC
Exits the menu without changes.
Unlocking autotests and comments
Protection flag (key) for all default auto test sequences and associated comments will be cleared.
Figure 4.29: Other settings dialogue
TEST
Unlocks locked auto test sequences.
ESC
Exits the menu without changes.
Page 42
MI 3105 EurotestXA Instrument operation Miscellaneous
42
Units selection
Unit for specific earth resistance will be selected.
Figure 4.30: Other settings dialogue
/ Select distance units. TEST
Enters selected distance units.
ESC
Exits the menu without changes.
4.4.6 Communication port
The communication port (RS232 or USB) can be selected in this menu.
Figure 4.31: Communication port selection
Keys:
/
Select communication port.
TEST
Confirms selected port.
ESC
Exits without changes.
Note:
Only one port can be active at the same time.
4.4.7 Locator
This function enables tracing electrical lines.
Keys:
TEST
Starts locator function.
ESC
Exits miscellaneous menu.
See chapter 5.11 Locator for locator operation.
Page 43
MI 3105 EurotestXA Instrument operation Miscellaneous
43
4.4.8 Operator
This menu enables registering the operator of the instrument. Selected operator name appears on the bottom of the LCD during turning on of the instrument. It is also associated to stored measurement results. Up to 5 operators can be defined.
Figure 4.32: Operator menu
Keys:
/
Select operator.
TEST
Accepts selected operator.
ESC
Exits to miscellaneous menu without changes.
F1
Enters operators name edit menu.
Operators name can be entered or modified.
Maximum 15 characters can be entered for operator.
Figure 4.33: Operator name edit menu
Keys:
Highlighted key
Selected symbol or activity.
// / Select symbol or activity. TEST
Enters selected symbol or performs selected activity.
F1
Deletes last entered symbol in the name line.
F2
Confirms comment and returns to operator main menu.
ESC
Deletes operator (immediately after entering the editor). Returns to operator main menu without changes.
Page 44
MI 3105 EurotestXA Measurements Insulation resistance
44
5 Measurements
5.1 Insulation resistance
Insulation resistance measurement is performed in order to assure safety against electric shock through insulation. It is covered by the EN 61557-2 standard. Typical applications are:
Insulation resistance between conductors of installation, Insulation resistance of non-conductive rooms (walls and floors), Insulation resistance of ground cables, Resistance of semi-conductive (antistatic) floors.
See chapter 4.2 Single test for functionality of keys.
Figure 5.1: Insulation resistance
Test parameters for insulation resistance measurement
TEST Test configuration [L-N, L-PE, N-PE, ‘L-PE,N-PE’, ‘L-N,L-PE’, ALL] Uiso Test voltage [50 V, 100 V, 250 V, 500 V, 1000 V]
Limit
Minimum insulation resistance [OFF, 0.01 MΩ ÷ 200 MΩ, (‘L-PE,N-PE’, ‘L-N,L-PE’, ALL: 20 MΩ)]
Test circuits for insulation resistance
L1 L2
L3 N PE
mains voltage
switched off
closed switches
loads disconnected
N/L2
L
/
L
1
PE/L3
Figure 5.2: Connection of universal test cable for general insulation resistance
measurement (TEST: L-PE)
Page 45
MI 3105 EurotestXA Measurements Insulation resistance
45
L1 L2 L3 N PE
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
mains voltage
switched off
loads disconnected
Figure 5.3: Application of plug commander and / or universal test cable for insulation
resistance measurement (TESTS: ‘L-PE,N-PE’, ‘L-N,L-PE’, ALL)
Insulation resistance measuring procedure
Select the INSULATION function. Set test parameters. Enable and set limit value (optional). Disconnect tested installation from mains supply (and discharge tested insulation).Connect test cable to the instrument and tested item (see figures 5.2 and 5.3). Press the TEST key for measurement (keep pressing for continuous
measurement).
After the measurement is finished wait until tested item is discharged.
Store the result (optional).
Figure 5.4: Examples of insulation resistance measurement results
Displayed results:
Rln ..........Insulation resistance between L (+) and N (-).
Rlpe ........Insulation resistance between L (+) and PE (-).
Rnpe .......Insulation resistance between N (+) and PE (-).
Um ..........Test voltage(s) – actual value(s).
Note:
Follow the correct test wiring as indicated in terminal voltage monitor
when the particular insulation test is selected. If only two test wires are connected and L-N, L-PE or N-PE test is selected then technical specification for Insulation ALL applies.
Page 46
MI 3105 EurotestXA Measurements Continuity
46
5.2 Resistance to earth connection and equipotential bonding
The resistance measurement is performed in order to assure that protective measures against electric shock through earth bond connections are effective. Four subfunctions are available:
Earth bond resistance measurement according to EN 61557-4 (between N and PE
terminals, test current >200 mA),
Earth bond resistance measurement according to EN 61557-4 (between L and PE
terminals, test current >200 mA),
Continuous resistance measurement with lower test current (between N and PE
terminals, test current ca 7 mA),
Continuous resistance measurement with lower test current (between L and PE
terminals, test current ca 7 mA).
See chapter 4.2 Single test for functionality of keys.
Figure 5.5: Continuity
Test parameters for resistance measurement
TEST Resistance measurement sub-function [R200mA NPE, R7mA NPE, R200mA
LPE, R7mA LPE]
Limit
Maximum resistance [OFF, 0.1 Ω ÷ 20.0 Ω]
5.2.1 Continuity R200 mA measurement
The resistance measurement is performed with automatic polarity reversal of the test voltage.
Page 47
MI 3105 EurotestXA Measurements Continuity
47
Test circuit for Continuity R200mA measurement
PCC1
PCC2
PCC3
PE/L3
L/L1
MPEC
MPEC....Main Potential Equilizing Collector
PCC....Protection Conductor Collector
extension lead
N/L2
Figure 5.6: Connection of universal test cable plus optional extension lead
Resistance to earth connection and equipotential bonding measurement procedure
 Select the CONTINUITY function.  Set sub-function R200mA (L-PE or N-PE).  Enable and set limit (optional).  Connect test cable to the instrument. Compensate test leads resistance (if necessary). Disconnect from mains supply and discharge tested installation. Connect test leads to the tested PE wiring (see figure 5.6).  Press the TEST key for measurement.  After the measurement is finished store the result (optional).
Figure 5.7: Example of continuity R200mA result
Displayed results:
R..............Main R200mA resistance (average of R+ and R- results),
R+............R200mA sub-resistance with positive voltage at N terminal,
R- ............R200mA sub-resistance with positive voltage at PE terminal.
Page 48
MI 3105 EurotestXA Measurements Continuity
48
5.2.2 7 mA resistance measurement
In general this function serves as standard Ω-meter with low test current. The measurement is performed continuously without pole reversal. The function can also be applied for testing continuity of inductive components.
Test circuit for continuous resistance measurement
Ry
S
zTx
N
/
L
2
P
E
/
L
3
L
/
L
1
Figure 5.8: Universal test cable application
Continuous resistance measurement procedure
Select the CONTINUITY function. Set sub-function R 7mA (L-PE or N-PE). Enable and set limit (optional). Connect test cable to the instrument. Compensate test leads resistance (if necessary). Disconnect from mains supply and discharge tested object. Connect test leads to the tested object (see figure 5.8). Press the TEST key for continuous measurement. Press the TEST key to stop measurement. After the measurement is finished store the result (optional).
Figure 5.9: Example of 7 mA resistance measurement
Displayed result:
R..............Resistance.
Page 49
MI 3105 EurotestXA Measurements Continuity
49
5.2.3 Compensation of test leads resistance
This chapter describes common principle for compensation of test leads resistance for both CONTINUITY functions. The compensation is required to eliminate the influence of test leads resistance plus internal resistances of the instrument. The lead compensation is
very important to obtain correct result. The compensation status ( / ) is indicated
in the message field.
Key:
F1
Enters test leads resistance compensation menu for any of mentioned functions.
See chapter 4.2 Single test for functionality of keys.
Figure 5.10: Test leads resistance
compensation menu
Keys:
TEST
Performs compensation
/ Sets function to be compensated
The instrument compensates following Continuity subfunctions.
Compensation NPE Same compensation for both 7 mA and 200 mA measurements.
Short N and PE terminals.
Compensation LPE Same compensation for both 7 mA and 200 mA measurements.
Short L and PE terminals
Page 50
MI 3105 EurotestXA Measurements Continuity
50
Circuits for compensating the resistance of test leads
N
/
L
2
P
E
/
L
3
N
/
L
2
P
E
/
L
3
extension lead
P
E
/
L
3
N
/
L
2
Figure 5.11: Shorted test leads-examples for N-PE
Compensation of test leads resistance procedure
Select the CONTINUITY function (any). Connect test cable to the instrument and short N/PE or L/PE terminals (see figure
5.11).
Press the F1 key to open test leads resistance compensation menu. Press the TEST key for measurement and compensation of test leads resistance. Press the ESC key to return to function menu.
Note:
20 Ω is limit value for test leads resistance compensation.
Page 51
MI 3105 EurotestXA Measurements RCD
51
5.3 Testing RCDs
Various test and measurements are required for verification of RCD(s) in RCD protected installations. Measurements are based on the EN 61557-6 standard.
The following measurements and tests (sub-functions) can be performed:
Contact voltage, Trip-out time, Trip-out current, RCD autotest.
See chapter 4.2 Single test for functionality of keys.
Figure 5.12: RCD test
Test parameters for RCD test and measurement
TEST RCD sub-function test [Tripout time t, Uc, AUTO, Tripout current].
Idn
Rated RCD residual current sensitivity I
ΔN
[10 mA, 30 mA, 100 mA, 300 mA,
500 mA, 1000 mA].
type
RCD type [ , ], test current waveform plus starting polarity [ , , ,
, ,
].
MUL Actual test current relative to rated Idn [½, 1, 2, 5]. Ulim Conventional touch voltage limit [25 V, 50 V].
The instrument is intended for testing of General (non-delayed) and
S
elective (time-
delayed) RCDs, which are suited for:
Alternating residual current (AC type, marked with symbol), Pulsating residual current (A type, marked with symbol). DC residual current (B type, marked with symbol).
Time delayed RCDs demonstrate delayed response characteristics. They contain residual current integrating mechanism for generation of delayed trip out. However, contact voltage pre-test in the measuring procedure also influences the RCD and it takes a period to recover into idle state. Time delay of 30 s is inserted before performing trip-out test to recover RCD after pretests.
Page 52
MI 3105 EurotestXA Measurements RCD
52
Circuits for testing RCD
L1 L2 L3 N
PE
R
E
Ro
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
Figure 5.13: Connecting the plug commander and the universal test cable
5.3.1 Contact voltage (RCD Uc)
A current flowing into the PE terminal causes a voltage drop on earth resistance, i.e. voltage difference between PE equipotential bonding circuit and earth. This voltage
difference is called contact voltage. The voltage is present on all accessible conductive
parts connected to the PE. It should always be lower than the conventional safety limit voltage. The contact voltage is measured with a test current lower than ½ IΔN to avoid trip-out of the RCD and then normalized to the rated IΔN.
Contact voltage measurement procedure
 Select the RCD function.  Set sub-function Uc.  Set test parameters (if necessary).  Connect test cable to the instrument. Connect test leads to the tested object (see figure 5.13).  Press the TEST key.  After the measurement is finished, store the result (optional).
Displayed contact voltage is proportional to the rated nominal residual current of the RCD and multiplied by appropriate factor. Common factor of 1.05 is applied to avoid negative tolerance of result, additional depend on RCD type and type of test current. See table 5.1 for detailed contact voltage calculation.
RCD type
Contact voltage Uc
proportional to
Rated I
ΔN
AC G
1.05×IΔN
AC
S
2×1.05×IΔN
any
A G
1.4×1.05×IΔN
A
S
2×1.4×1.05×I
ΔN
30 mA
A G
2×1.05×I
ΔN
A
S
2×2×1.05×I
ΔN
< 30 mA
B G
2×1.05×I
ΔN
B
S
2×2×1.05×IΔN
any
Table 5.1: Relationship between Uc and I
Δ
N
Page 53
MI 3105 EurotestXA Measurements RCD
53
Loop resistance result is indicative and calculated from Uc result (without additional
proportional factors) according to:
N
C
L
I
U
R
Δ
= .
Figure 5.14: Example of contact voltage measurement results
Displayed results:
Uc....... Contact voltage.
Rl........ Fault loop resistance.
5.3.2 Trip-out time t
Trip-out time measurement verifies the sensitivity of an RCD at different residual currents.
Trip-out time measurement procedure
Select the RCD function. Set sub-function Tripout time t. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the tested object (see figure 5.13). Press the TEST key. After the measurement is finished store the result (optional).
Figure 5.15: Example of trip-out time measurement results
Displayed results:
t .......... Trip-out time,
Uc....... Contact voltage for rated I
ΔN
.
Page 54
MI 3105 EurotestXA Measurements RCD
54
Note:
See 4.4.2 RCD normative reference for selection of appropriate standard test
conditions.
5.3.3 Trip-out current
A continuously rising residual current is intended for testing the threshold sensitivity for RCD trip-out. The instrument increases the test current in small steps through appropriate range as follows:
Slope range
RCD type
Start value End value
Waveform
AC
0.2×IΔN 1.1×IΔN
Sine A (IΔN 30 mA) 0.2×IΔN 1.5×IΔN A (IΔN = 10 mA) 0.2×IΔN 2.2×IΔN
Pulsed
B
0.2×IΔN 2.2×IΔN
DC
Maximum test current is IΔ (trip-out current) or end value in case the RCD didn’t trip-out.
Trip-out current measurement procedure
Select the RCD function. Set sub-function Tripout current. Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the tested object (see figure 5.13). Press the TEST key. After the measurement is finished, store the result (optional).
Trip-out
After the RCD is turned on again
Figure 5.16: Trip-out current measurement result example
Displayed results:
I .......... Trip-out current,
Uci ...... Contact voltage at trip-out current I, or end value in case the RCD didn’t trip,
t .......... Trip-out time.
Page 55
MI 3105 EurotestXA Measurements RCD
55
5.3.4 RCD Autotest
RCD autotest function is intended to perform complete RCD testing and measurement of belonging parameters (contact voltage, fault loop resistance and trip-out time at different residual currents) in one set of automatic tests, guided by the instrument. If any false parameter is noticed during the RCD autotest, then individual parameter test has to be used for further investigation.
RCD autotest procedure RCD Autotest steps Notes
 Select the RCD function.  Set sub-function AUTO.  Set test parameters (if necessary). Connect test cable to the instrument. Connect test leads to the tested object (see figure 5.13).  Press the TEST key. Start of test
Test with ½×I
ΔN
, 0° (step 1).
RCD should not trip-out
Test with ½×I
ΔN
, 180° (step 2).
RCD should not trip-out
Test with I
ΔN
, 0° (step 3).
RCD should trip-out
Re-activate RCD.
Test with I
ΔN
, 180° (step 4).
RCD should trip-out
Re-activate RCD.
Test with 5×I
ΔN
, 0° (step 5).
RCD should trip-out
Re-activate RCD.
Test with 5×I
ΔN
, 180° (step 6).
RCD should trip-out
Re-activate RCD. After the measurement is finished, store the result (optional). End of test
Result examples:
Step 1
Step 2
Page 56
MI 3105 EurotestXA Measurements RCD
56
Step 3
Step 4
Step 5
Step 6
Figure 5.17: Individual steps in RCD autotest
Displayed results:
.... Step 1 trip-out time (½×IΔN, 0º),
.... Step 2 trip-out time (½×IΔN, 180º),
... Step 3 trip-out time (IΔN, 0º), ... Step 4 trip-out time (IΔN, 180º), ... Step 5 trip-out time (5×IΔN, 0º), ... Step 6 trip-out time (5×IΔN, 180º),
Uc....... Contact voltage for rated I
ΔN.
Notes:
The autotest sequence is immediately stopped if any incorrect condition is detected,
e.g. excessive Uc or trip-out time out of bounds.
Auto test is finished without tests in case of testing the RCD type A with rated
residual currents of I
Δn = 300 mA, 500 mA, and 1000 mA. In this case auto test
result passes if all previous results pass, and indications and are omitted.
Page 57
MI 3105 EurotestXA Measurements Fault loop impedance
57
5.4 Fault loop impedance and prospective fault current
Fault loop is a loop comprising mains source, line wiring and PE return path to the mains source. The instrument has ability to measure impedance of mentioned loop and calculate short circuit current and contact voltage regarding the selected circuit breaker type. The measurement is covered by requirements of the EN 61557-3 standard.
See 4.2 Single test for active keys.
Figure 5.18: Fault loop
impedance
Test parameters for fault loop impedance measurement
Protection Selection of main protection device in fault loop [RCD, FUSE]* Fuse type Selection of fuse type [---, NV, Gg, B, C, K, D] ** Fuse I Rated current of selected fuse Fuse T. Maximum breaking time of selected fuse Isc_lim Minimum short circuit current for selected fuse combination.
See Appendix A for reference fuse data. * Select RCD to prevent trip-out of RCD in RCD protected installation. ** --- Means no fuse selected.
Circuits for measurement of fault loop impedance
L1 L2 L3 N
PE
R
E
Ro
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
Figure 5.19: Connection of plug cable and universal test cable
Page 58
MI 3105 EurotestXA Measurements Fault loop impedance
58
Fault loop impedance measurement procedure
 Select the Z-LOOP function.  Select test parameters (optional).  Connect test cable to the EurotestXA. Connect test leads to the tested object (see figure 5.18).  Press the TEST key.  After the measurement is finished store the result (optional).
Figure 5.20: Examples of loop impedance measurement result
Displayed results:
Z ..............Fault loop impedance,
ISC ............Prospective fault current,
R..............Resistive part of loop impedance,
Xl .............Reactive part of loop impedance.
Prospective fault current ISC is calculated from measured impedance as follows:
Z
kUn
I
SC
SC
×
=
where:
Un ........Nominal U
L-PE
voltage (see table below),
ksc ....... Correction factor for Isc (see chapter 4.4.2).
Un Input voltage (L-PE)
115 V
(100 V U
L-PE
< 160 V)
230 V
(160 V U
L-PE
264 V)
Notes:
High fluctuations of mains voltage influence the measurement results. The noise
sign is displayed in the message field in such case. Repeat the measurement.
Isc is not calculated in case the terminal voltage monitor does not detect voltage
state that corresponds to the selected supply system, indication
.
This measurement will trip-out RCD in RCD-protected electrical installation if FUSE is selected as breaking device instead of RCD.
Page 59
MI 3105 EurotestXA Measurements Line impedance
59
5.5 Line impedance and prospective short-circuit current
Line impedance is measured in loop comprising of mains voltage source and line wiring (L and N). It is covered by requirements of the EN 61557-3 standard.
See 4.2 Single test for keys functionality.
Figure 5.21: Line impedance
Test parameters for line impedance measurement
FUSE type Selection of fuse type [---, NV, Gg, B, C, K, D] * FUSE I Rated current of selected fuse FUSE T Maximum breaking time of selected fuse Isc_lim Minimum short circuit current for selected fuse combination.
See Appendix A for reference fuse data. *--- Means no fuse selected
Circuit for measurement of line impedance
L1 L2 L3 N
PE
R
E
Ro
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
N/L2
L
/
L
1
PE/L3
Figure 5.22: Phase-neutral or phase-phase line resistance measurement – connection of
plug commander and universal test cable
Line impedance measurement procedure
Select the Z-LINE function. Select test parameters (optional). Connect test cable to the instrument. Connect test leads to the tested object (see figure 5.22). Press the TEST key. After the measurement is finished, store the result (optional).
Page 60
MI 3105 EurotestXA Measurements Line impedance
60
Line to neutral
Line to line
Figure 5.23: Examples of line impedance measurement result
Displayed results:
Z ..............Line impedance,
ISC ............Prospective short-circuit current,
R..............Resistive part of line impedance,
XL............Reactive part of line impedance.
Prospective short circuit current is calculated as follows:
Z
kUn
I
SC
SC
×
=
where:
Un ........Nominal L-N or L1-L2 voltage (see table below),
ksc ....... Correction factor for Isc (see chapter 4.4.2).
Un Input voltage range (L-N or L1-L2)
115 V
(100 V U
L-N
< 160 V)
230 V
(160 V U
L-N
264 V)
400 V
(264 V < U
L-N
440 V)
Note:
High fluctuations of mains voltage can influence the measurement results. The
noise sign
is displayed in the message field in this case. Repeat the
measurement.
Isc is not calculated in case the terminal voltage monitor does not detect voltage
state that corresponds to the selected supply system, indication .
Page 61
MI 3105 EurotestXA Measurements Voltage, frequency, phase sequence
61
5.6 Voltage, frequency and phase sequence
Voltage and frequency measurement is always active in the terminal voltage monitor. In
the special voltage menu the measured voltage, frequency and information about
detected three-phase connection can be stored. Phase sequence measurement conforms to the EN 61557-7 standard.
See 4.2 Single test for keys functionality
Figure 5.24: Voltage in single
phase system
Test parameters for voltage measurement There are no parameters.
Circuits for voltage measurement
L3 L2 L1 N PE
N
/
L
2
N
/
L
2
L
/
L
1
L
/
L
1
P
E
/
L
3
P
E
/
L
3
result 1.2.3 res ult 2.1.3
Figure 5.25: Connection of universal test cable and optional adapter in three-phase system
L1 L2 L3 N
PE
R
E
Ro
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
N/L2
L
/
L
1
PE/L3
Figure 5.26: Connection of plug commander and universal test cable in single-phase
system
Page 62
MI 3105 EurotestXA Measurements Voltage, frequency, phase sequence
62
Voltage measurement procedure
Select the VOLTAGE function. Connect test cable to the instrument. Connect test leads to the tested object (see figures 5.25 and 5.26). Store current measurement result (optional).
Measurement runs immediately after selection of VOLTAGE function.
Figure 5.27: Examples of voltage measurement in three-phase system
Displayed results for single phase system:
Uln...........Voltage between phase and neutral conductors,
Ulpe.........Voltage between phase and protective conductors,
Unpe........Voltage between neutral and protective conductors,
f ...............frequency.
Displayed results for three-phase system:
U12..........Voltage between phases L1 and L2,
U13..........Voltage between phases L1 and L3,
U23..........Voltage between phases L2 and L3,
1.2.3 ........Correct connection – CW rotation sequence,
3.2.1 ........Invalid connection – CCW rotation sequence,
f ...............frequency.
Page 63
MI 3105 EurotestXA Measurements Resistance to earth
63
5.7 Resistance to earth
Resistance to earth is important for protection against electric shock. This function is intended for verification of earthing of house installation and other earthings, e.g., earthing for lighting. The measurement conforms to the EN 61557-6 standard. The following resistance to earth sub-functions are available:
Standard 3-wire, for standard resistance to earth measurements. One clamp, for measuring resistance to earth of individual earthing rods. Two clamps (also recommended in IEC 60364-6 for urban areas), for measuring
resistance to earth of individual earthing rods.
Specific earth resistance by using optional external adapter.
See 4.2 Single test for keys functionality.
Figure 5.28: Resistance to earth
Test parameters for earth resistance measurement
TEST
Test configuration [3-wire, one clamp, two clamps, ρ]
Limit
Maximum resistance [OFF, 1 Ω ÷ 5 kΩ, (2 clamps: 1 Ω ÷ 20 Ω)] If ρ selected: Distance Distance between probes [0.1 m ÷ 30.0 m] or [1 ft ÷ 100 ft]
5.7.1 Standard 3-wire measurement Circuits for 3-wire measurement
MPEC
Rc Rp R
E
d>5d
E
H
S
blue - N
b - Llack
green - PE
Figure 5.29: Resistance to earth, 3-wire measurement of PE grounding
Page 64
MI 3105 EurotestXA Measurements Resistance to earth
64
blue - N
black - L
green - PE
E
H
S
R3
E
R4
E
MPEC
d
>5d
R
E1
R2
E
RpRc
Figure 5.30: Resistance to earth, 3-wire measurement of lighting protection
Resistance to earth, 3-wire measurement procedure
Select the EARTH function. Select 3-wire measurement. Enable and set limit (optional). Connect test cable to the instrument. Connect test leads to the tested object (see figures 5.29 and 5.30). Press the TEST key. After the measurement is finished, store the result (optional).
Figure 5.31: Example of resistance to earth measurement results (3-wire)
Displayed results for earth resistance measurement:
R..............Earth resistance,
Rc............Resistance of S probe,
Rp............Resistance of H probe.
Page 65
MI 3105 EurotestXA Measurements Resistance to earth
65
5.7.2 One clamp measurement
The measurement supports testing of particular earthing branches in earthing system.
Circuit for one clamp measurement
blue - N
black - L
green - PE
E
H
S
R3
E
R4
E
MPEC
d
>5d
R
E1
R2
E
RpRc
Figure 5.32: Resistance to earth, measurement with a current clamp
Resistance to earth, one clamp measurement procedure
Select the EARTH function. Select one clamp measurement. Enable and set limit (optional). Connect test cable and clamp to the instrument. Connect test leads and clamp to the tested object (see figure 5.32). Press the TEST key. After the measurement is finished, store the result (optional).
Figure 5.33: Example of resistance to earth measurement results, one clamp
Displayed results for earth resistance measurement:
R..............Earth resistance of measured earthing branch,
Rc............Resistance of S probe,
Rp............Resistance of H probe,
Re............Resistance to earth of tested system.
Note:
Connect test clamp between E test terminal and ground, otherwise the parallel
resistance of all electrodes (R
E1 up to RE3) will be measured.
Page 66
MI 3105 EurotestXA Measurements Resistance to earth
66
5.7.3 Two clamps measurement
The measurement is intended for testing of particular earthing branches in earthing system, especially in urban areas. It is also required by IEC 60364-6:2006.
Circuit for two clamps measurement
blue - N
black - L
green - PE
R3
E
R4
E
MPEC
R
E1
R2
E
Figure 5.34: Resistance to earth, two clamps measurement of lighting protection
Resistance to earth, two clamps measurement procedure
Select the EARTH function. Select two clamps measurement. Enable and set limit (optional). Connect test clamps to the instrument. Connect test clamps to the tested object (see figure 5.34). Press the TEST key. After the measurement is finished, store the result (optional).
Figure 5.35: Example of resistance to earth measurement result, two clamps
Displayed results for earth resistance measurement:
R..............Earth resistance.
Note:
The distance between clamps should be at least 30 cm.
Page 67
MI 3105 EurotestXA Measurements Resistance to earth
67
5.7.4 Specific earth resistance measurement
The measurement is intended for measuring specific earth resistance by using special adapter A1199.
Circuit for specific earth resistance measurement
aaa
EES
S
H
a/20 max.
g
r
e
e
n
b
l
u
e
r
e
d
black
Figure 5.36: Specific earth resistance measurement with ρ-adapter
Specific earth resistance measurement procedure
Select the EARTH function. Connect ρ-adapter to the instrument.
Select ρ measurement.
Select distance unit (optional). Set distance (optional). Connect test leads of ρ-adapter to tested object (see figure 5.36) Press the TEST key. After the measurement is finished, store the result (optional).
Figure 5.37: Example of specific earth resistance measurement result
Displayed results for earth resistance measurement:
ρ Specific earth resistance.
Rc............Resistance of S probe,
ρ Rp Resistance of H probe.
Page 68
MI 3105 EurotestXA Measurements Current
68
Note:
Distance units can be selected in Miscellaneous/Initial settings/Other settings menu,
see 4.4.5.
5.8 Current
This function is intended for measurement of the electric current with current clamp. It is intended for measurement of leakage current and load current.
See chapter 4.2 Single test for functionality of keys.
Figure 5.38: Current
Test parameters for clamp current measurement Limit Maximum current [OFF, 0.1 mA ÷ 100 mA]
Test circuits for clamp current measurement
L1 L2
L3 N PE
L
e
a
k
a
g
e
c
u
r
r
e
n
t
L
o
a
d
c
u
r
r
e
n
t
Figure 5.39: Leakage and load current measurements
Page 69
MI 3105 EurotestXA Measurements Current
69
Current measuring procedure
 Select the CURRENT function.  Set test parameters.  Enable and set limit value (optional).  Connect current clamp to the instrument and tested item (see figure 5.39).  Press the TEST key to start the measurement.  Press the TEST key again to finish the measurement.  Store the result (optional).
Figure 5.40: Example of clamp current measurement result
Displayed result:
I...............Current.
Note:
Displayed current represents r.m.s. value for current clamp with ratio 1000:1. Use test clamp supplied by METREL or other with similar characteristics (current
output, ratio 1000:1, appropriate measurement range; consider error of test clamp when evaluating measured results)!
Current clamps Metrel A 1074 and A 1019 are suitable for use with the instrument
in range of 0.2 A
÷ 20 A. Below 0.2 A they can be used as indicator only. They are
not suitable for leakage current measurements.
The only Metrel current clamp, suitable for leakage current measurements, is
A 1018 (1000 A/1 A).
Page 70
MI 3105 EurotestXA Measurements Sensors - Illumination
70
5.9 Sensors and adapters
This function extends application range of the instrument by using external Metrel sensors and adapters. The probe is connected to the instrument via RS 232 interface. The instrument automatically recognizes connected probe.
5.9.1 Illumination
The measurement is performed with LUX meter type B or LUX meter type C probes in order to test and verify illumination.
See chapter 4.2 Single test for functionality of keys.
Figure 5.41: Insulation resistance
Test parameters for illumination measurement Limit Minimum illumination [OFF, 0.1 lux ÷ 20.0 klux]
Test setup for illumination measurement
RS 232
PS/2
Figure 5.42: Connection of LUX probe to the instrument
Page 71
MI 3105 EurotestXA Measurements Sensors - Illumination
71
Figure 5.43: LUX meter probe positioning
Illumination measuring procedure
Connect LUX probe to the instrument (see figure 5.42). Select the SENSORS function. Enable and set limit value (optional).
Power ON the LUX probe (ON/OFF key, green LED lits).
Press the TEST key for measurement. Press the TEST key to finish the measurement.
Power OFF the LUX probe.
Store the result (optional).
Figure 5.44: Example of illumination measurement result
Displayed results:
E .............Illumination.
Notes:
Take care of the LUX probe positioning. For accurate measurements make sure that the milk glass bulb is lit up without any
shadows cast by hand, body or other unwanted objects.
It is very important to know that it takes the time to get full power operation of the
artificial light sources (see technical data for light sources). Therefore, they should be switched on for at least that period before the measurements are proceeded.
Page 72
MI 3101 EurotestAT Measurements Adapters – 2 Ω impedance
72
5.9.2 2 Ω line/loop impedance
The measurement is performed with Impedance adapter A1143. It is automatically recognized in Z-LINE and Z-LOOP functions. With this adapter, low impedances up to 1999 mΩ can be measured. The measurement is covered by requirements of the EN 61557-3 standard.
See chapter 4.2 Single test for functionality of keys.
Figure 5.45: Impedance adapter
connected
Test parameters for 2 Ω line/loop impedance measurement
Function Z-LINE Test
Impedance function [mΩ L-N, mΩ L-L] Functions Z-LINE and Z-LOOP FUSE type Selection of fuse type [---, NV, Gg, B, C, K, D] * FUSE I Rated current of selected fuse FUSE T Maximum breaking time of selected fuse Isc_lim Minimum short circuit current for selected fuse combination.
See Appendix A for reference fuse data. *--- Means no fuse selected
Additional key:
F2
Toggles between result screens.
Test setup for 2
Ω line/loop impedance measurement
PS/2
RS 232
Figure 5.46: Connection of impedance adapter to the instrument
Page 73
MI 3101 EurotestAT Measurements Adapters – 2 Ω impedance
73
2 Ω line/loop impedance measuring procedure
Connect Impedance adapter to the instrument (see figure 5.46).  Select the functions Z-LINE or Z-LOOP.  Enable and set limit value (optional).  Power ON the Impedance adapter (ON/OFF key, green LED lits).  Connect Impedance adapter to tested installation.  Press the TEST key for measurement.  Store the result (optional).
Figure 5.47: Example of 2 Ω line/loop measurement results
Displayed results:
Z ........................ Line / loop impedance,
ISC ...................... Prospective short-circuit current,
R........................ Resistive part of line impedance,
Xl ....................... Reactive part of line impedance.
The following parameters are displayed in sub-screen for single-phase line impedance measurement:
IscMaxL-N ......... Maximum prospective short-circuit current.
IscMinL-N .......... Minimum prospective short-circuit current.
IscStd ................ Standard prospective short-circuit current.
When testing phase-to-phase line impedance the following parameters are displayed in sub-screen:
IscMax3Ph......... Maximum three-phase prospective short-circuit current.
IscMin3Ph.......... Minimum three-phase prospective short-circuit current.
IscMax2Ph......... Maximum two-phase prospective short-circuit current.
IscMin2Ph.......... Minimum two-phase prospective short-circuit current.
IscStd ................ Standard prospective short-circuit current.
The following parameters are displayed in sub-screen for loop impedance measurement:
IscMaxL-Pe ....... Maximum prospective fault current.
IscMinL-Pe ........ Minimum prospective fault current.
IscStd ................ Standard prospective fault current.
Ub...................... Contact voltage at maximum prospective fault current (Contact
voltage is measured against probe S terminal).
Page 74
MI 3101 EurotestAT Measurements PE test terminal
74
Notes:
For application and technical data of the Impedance adapter A1143 see its user
manual 20750859.
High fluctuations of mains voltage can influence the measurement results.
Check adapter indications in case the abort symbol appears after the start of
measurement.
5.10 PE test terminal
It can happen that a dangerous voltage is applied to the PE wire or other accessible metal parts. This is a very dangerous situation since the PE wire and MPEs are considered to be earthed. A common reason for this fault is incorrect wiring (see examples below).
When touching the TEST key in all functions that require mains supply the user
automatically performs this test.
Examples for application of PE test terminal
L1 N PE
Reversed phase and protection conductors!
T MOST DANGEROUS SITUATION!
HE
Figure 5.48: Reversed L and PE conductors (application of plug commander)
L1 N
Reversed phase and protection conductors!
MOST DANGEROUS SITUATION!
PE
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
Figure 5.49: Reversed L and PE conductors (application of universal test cable)
Page 75
MI 3101 EurotestAT Measurements PE test terminal
75
PE terminal test procedure
Connect test cable to the instrument. Connect test leads to the tested object (see figures 5.48 and 5.49). Touch PE test probe (the TEST key) for at least one second.
If PE terminal is connected to phase voltage the warning message is displayed,
instrument buzzer is activated, and further measurements are disabled in Z-LOOP and RCD functions.
Warning:
If line voltage is detected on the tested PE terminal, immediately stop all
measurements, find and remove the fault!
Notes:
In main and miscellaneous menus the PE terminal is not tested. PE test terminal does not operate in case the operator’s body is completely
insulated from floor or walls!
Page 76
MI 3101 EurotestAT Measurements Locator
76
5.11 Locator
This function is intended for tracing mains installation, like:
Tracing lines, Finding shorts, breaks in lines, Detecting fuses.
The instrument generates test signals that can be traced with the handheld tracer receiver R10K. See
Locator appendix for additional information.
Figure 5.50: Locator entry point
Parameters for locator There are no parameters.
Typical applications for tracing electrical installation
ON
LPEN
L
/
L
1
N
/
L
2
P
E
/
L
3
Figure 5.51: Tracing wires under walls and in cabinets
ON
Selective probe
Energized installation
Receiver R10K
Figure 5.52: Locating individual fuses
Page 77
MI 3101 EurotestAT Measurements Locator
77
Line tracing procedure
Select the LOCATOR function in MISC menu. Connect test cable to the instrument. Connect test leads to the tested object (see figures 5.51 and 5.52). Press the TEST key.
Trace lines with receiver (in IND mode) or receiver plus its optional accessory.
After tracing is finished press the ESC key to stop generating test signal.
LOCATOR
Figure 5.53: Locator active
Page 78
MI 3105 EurotestXA Measurements Varistor test
78
5.12 Varistor test
This test is performed to verify overvoltage protection devices. Typical devices are:
Metal oxide varistors, Gas arresters, Semiconductor transient voltage suppressors.
See chapter 4.2 Single test for functionality of keys.
Figure 5.54: Varistor test menu
Test parameters for varistor test
Lo limit Low limit DC threshold voltage [50 V ÷ 1000 V] Hi limit High limit DC threshold voltage [50 V ÷ 1000 V]
It = 1.00 mA
Threshold current
Test circuit for varistor test
L1 L2
L3 N PE
mains voltage
switched off
opened switches
N/L2
L
/
L
1
PE/L3
permanent
loads
disconnected
Figure 5.55: Connection of universal test cable for varistor test
Page 79
MI 3105 EurotestXA Measurements Varistor test
79
Varistor test procedure
Select the VARISTOR TEST function. Set test parameters.
Disconnect mains supply and consumers from tested overvoltage device.
Connect test cable to the instrument and tested item (see figure 5.55). Press the TEST key for measurement.
After the measurement is finished wait until tested item is discharged.
Store the result (optional).
Figure 5.56: Example of varistor test result
Displayed results:
U .............Measured threshold voltage at It (1 mA).
Uac .........Rated AC voltage.
Uac is calculated from U according to: Uac = U/1.6.
Page 80
MI 3105 EurotestXA Data handling
80
6 Data handling
6.1 Memory organization
The following data can be stored in instrument memory:
Auto sequence name, sequence, and function parameters, Auto sequence and single test results with belonging parameters, Installation structure with belonging data.
Stored data can be organized according to the installation structure of the tested object. Measured results can be stored into corresponding location of the structure.
6.2 Installation data structure
This functionality helps to organize operation with data in a simple and effective manner. The memory organization can be customized according to the actual structure of the tested electrical installation. Main benefits are:
Test results can be organized and grouped in a structured manner that equals the
structure of the tested electrical installation. If a test plan for verification of electrical installation is prepared it is possible to organize the data structure according to it. Each tested location place like room, floor, installation node, switchgear, etc. can be reflected as its own location in memory.
Simple browsing through structure and results. Test reports can be created with no or little modifications after downloading results
to a PC.
Test procedures can be prepared in advance on the PC and sent to the instrument. A new installation structure can be built on the instrument An existing structure can be upgraded on the instrument. A name can be assigned to each location.
The data structure can be accessed and updated in each of the three main memory menus (store, recall, clear memory), but also through tree structure view.
Basic view
Tree structure view
Figure 6.1: Example of data structure fields
Page 81
MI 3105 EurotestXA Data handling
81
Figure 6.2: Installation structure example as presented on PC
Legend:
Memory operation menu Installation data structure field Root level in the structure:
METREL d.d.: 1
st
level location name.
1/1: No. of selected / available locations on this level.
Sub-level (level 2) in the structure:
PRODUCT.: sub-location name. 2/5: No. of selected / available locations on this level.
Sub-level (level 3) in the structure:
3PH SOCKET: location name. 1/3: No. of selected / available locations on this level.
Results field – stored results in the selected location.
   
Arrows point to existing non-displayed structure locations. Available memory information. No. of stored test results in selected location / No. of all stored test results (in complete structure). Option for opening the structure tree view.
Options for modification of the structure (see chapter 6.6).
Note:
Only three locations in the installation data structure field (placed horizontally) can
be displayed at the same time in the basic view.
Basic keys:
Page 82
MI 3105 EurotestXA Data handling
82
/ / /
Select the existing location.
Pressed for 2 s opens dialog box for adding a new location.
F1
Renames the current location.
F2
Enters installation structure tree screen.
ESC
Back to the last operation mode of the instrument.
Note:
The tree structure is limited to 2000 locations with 10 levels in depth, see figure 6.3.
Figure 6.3: Sub-levels depth definition
Figure 6.4 shows how individual structure elements are displayed on the instrument. The
outlook is the same for all three memory menus.
Page 83
MI 3105 EurotestXA Data handling
83
Figure 6.4: Data structure elements
6.3 Storing test results
After the completion of a single test or auto sequence the results and parameters are
ready for storing ( icon is displayed in the information field). Press the MEM key to store
the results.
See chapter 6.2 for definitions of displayed
fields.
Figure 6.5: Save test menu
Page 84
MI 3105 EurotestXA Data handling
84
Keys in save test menu - installation data structure field:
/ / /
Short press - select the location in structure of installation data field. Pressed for a few seconds in some cases - add a new location in the structure, see 6.6.1.
MEM
Saves test results to the last position in selected location and returns to the measuring menu.
TAB
Switches between results and structure data field, see 6.3.1.
ESC
Exits save test menu.
F1
Edits name of selected location (see 4.3.4).
F2
Enters installation structure tree view to select appropriate location.
Notes:
Press the MEM key twice to quickly store the results to pre-selected location.
By default it is offered to append the result to the existing results in the selected
location.
6.3.1 Saving results specialties
It is possible to overwrite existing result when storing new result.
Appending new result
Overwriting requires confirmation
Figure 6.6: Saving in result field
Keys in save test menu - results field:
/ Select stored test result.
TEST
Saves test result into selected line (confirmation is needed to overwrite an existing result).
ESC
Back to store test menu - installation data structure field.
Keys with open dialog:
/ Select YES / NO. TEST
Confirms selected option.
ESC
Cancels without changes.
For information about storing into a new non-existent location see 6.6.1.
Page 85
MI 3105 EurotestXA Data handling
85
6.4 Recalling test results and parameters
Press the MEM key in single or auto sequence menu when there is no result available for storing or select in MISC menu.
See chapter 6.2 for definitions of displayed fields.
Figure 6.7: Main recall menu
Keys in main recall memory menu:
/ / /
Short press - select the location in structure of installation data field. Pressed for a few seconds in some cases - add a new location in the structure, see 6.6.1.
TAB
Switches between results and structure data field.
ESC
Exits to the last state of the instrument.
F1
Edits name of selected location (for editing see 4.3.4).
F2
Enters installation structure tree view to select appropriate location.
6.4.1 Recalling result
Result field has to be selected.
Figure 6.8: Recall data menu
Keys in results field: / Select the stored data.
TEST
Opens selected stored item.
TAB, ESC
Back to recall memory main menu.
Page 86
MI 3105 EurotestXA Data handling
86
Key:
ESC
Back to recall memory main menu.
Figure 6.9: Stored single test
example
Keys:
/ Select stored data. TEST
Opens function result.
ESC
Back to recall memory main menu.
Key in open function result:
ESC
Back to observed auto sequence.
Figure 6.10: Auto sequence stored
example
6.5 Clearing saved data
From main menu select MISC menu and enter MISC: option (see 4.4.3).
In
select
option for erasing complete test
results memory.
Figure 6.11: Clear memory
Keys: / Select CANCEL / CLEAR.
TEST
Confirms selected option.
ESC
Cancels dialog without changes.
Page 87
MI 3105 EurotestXA Data handling
87
In select
option to delete particular result
or modify installation data structure.
Figure 6.12: Clear test menu
Keys: / Select location.
TEST
Opens dialog for clearing in installation data structure.
TAB
Moves focus into result field for selection of surplus result, see 6.5.1.
F2
Enters installation structure tree view to select appropriate location.
F1
Renames current location.
ESC
Back to the last mode of the instrument.
6.5.1 Clearing specialties
In the result field the particular stored test result can be cleared.
Selection of data for clearing
Dialog before clear
Figure 6.13: Clearing particular test
Keys:
/ Select stored test.
TEST
Opens dialog for clearing selected test.
ESC
Back to last mode of the instrument.
Keys in opened dialog:
/ Select YES / NO. TEST
Confirms selected option.
ESC
Cancels without changes.
Page 88
MI 3105 EurotestXA Data handling
88
Legend for clearing in installation data structure:
Results in current location. Results in sub-locations. Remove current location and its sub-locations.
Figure 6.14: Clear in installation
data structure menu
Keys: / / / Select option.
TEST
Confirms option.
ESC
Cancels dialog without changes.
6.6 Editing installation data structure
Installation data structure when once stored in the instrument can also be modified during use of the instrument. Editing possibilities are:
Adding new location – see 6.6.1,
Modifying the name of selected location,
Clearing location / tree structure, see 6.5.1.
The possibilities are accessible in save, recall and clear (partly) menus.
6.6.1 Adding new locations Note:
The structure can be expanded to 10 horizontal levels deep and with maximum
2000 storing locations.
Keys:
/ / /
Select the existing location.
Pressed for a few seconds in some cases - add a new location in the structure, see 6.6.1.
F2
Enters installation structure tree view to select appropriate location.
F1
Renames the current location.
ESC
Back to the last operation mode of the instrument.
(for 2 seconds)
Opens dialog box for adding new location at the same level. Active only if selected location is the last in the level.
Name of the new location: Same name as previous +1.
(for 2 seconds)
Opens dialog box for adding new location at the next sublevel. Active only if there are no sublevels at the selected location.
Name of the new location: Location
Page 89
MI 3105 EurotestXA Data handling
89
Keys in open dialog box: / Select YES / NO.
TEST
Confirms selected option.
ESC
Cancels dialog box without changes.
An example for opening a new location and storing a test result into the location is presented below.
Finished test with the results prepared for
saving is marked with
icon.
Figure 6.15: Test result prepared for saving
Key:
MEM
Enters the save test menu.
Figure 6.16: Save test menu
Keys:
F2
Changes structure view.
TEST
Confirms the new location.
F1
Enters name of the location.
Figure 6.17: Dialog box for new location
Page 90
MI 3105 EurotestXA Data handling
90
Enter name of the location.
Key:
F2
Confirms the name.
Figure 6.18: Entering name for the new
location
Key:
MEM
Saves results into the location.
Figure 6.19: Location prepared
Figure 6.20: Stored example
Page 91
MI 3105 EurotestXA Data handling
91
6.7 Communication
Stored results can be transferred to a PC. A special communication program on the PC automatically identifies the instrument and enables data transfer between the instrument and the PC. There are two communication interfaces available on the instrument: USB or RS 232 (for
selection see 4.4.6).
1
1
PS/2 - RS 232 cable
minimum connections: 1 to 2, 4 to 3, 3 to 5
PS/
2
for MI 3101
3
5
2
4
6
9 pin D female for PC
Figure 6.21: Interface connection for data transfer over PC COM port
How to transfer stored data:
In the MISC: select appropriate communication interface (USB / RS 232).
- RS 232 selected: connect a PC COM port to the instrument PS/2 connector
using the PS/2 - RS232 serial communication cable;
- USB selected: connect a PC USB port to the instrument USB connector using
the USB interface cable.
 Power on the PC and the instrument.  Run the program Eurolink.
The PC and the instrument automatically recognize each other. The program on the PC enables the following possibilities:
- Downloading data;
- Clearing storage;
- Changing and downloading user data;
- Preparing a simple report form;
- Preparing a file for exporting to a spreadsheet.
The program Eurolink is a PC software running on Windows 95/98, Windows NT, Windows
2000, Windows XP. Read the file README.TXT on CD for instructions about installing and running the program.
Note:
USB drivers should be installed on PC before using the USB interface. Refer to
USB installation instructions available on installation CD.
Page 92
MI 3105 EurotestXA Maintenance
92
7 Maintenance
Unauthorized person is not allowed to open the EurotestXA instrument. There are no user replaceable components inside the instrument, except three fuses and batteries under rear cover.
7.1 Replacing fuses
There are three fuses under back cover of the EurotestXA instrument.
F1
M 0.315 A / 250 V, 20×5 mm This fuse protects internal circuitry of continuity function if test probes are connected to the mains supply voltage by mistake during measurement.
F2, F3
F 4 A / 500 V, 32×6.3 mm General input protection fuses of test terminals L/L1 and N/L2.
Warnings:
Disconnect any measuring accessory and power off the instrument before
opening battery/fuse compartment cover, hazardous voltage inside!
Replace blown fuse with original type only, otherwise the instrument may be
damaged and/or operator’s safety impaired!
Position of fuses can be seen in Figure 3.4 in chapter 3.3 Back panel.
7.2 Cleaning
No special maintenance is required for the housing. To clean the surface of the instrument use a soft cloth slightly moistened with soapy water or alcohol. Then leave the instrument to dry totally before use.
Warnings:
Do not use liquids based on petrol or hydrocarbons! Do not spill cleaning liquid over the instrument!
7.3 Periodic calibration
It is essential that the test instrument is regularly calibrated in order technical specification listed in this manual is guaranteed. We recommend an annual calibration. Only an authorized technical person can do the calibration. Please contact your dealer for further information.
7.4 Service
For repairs under warranty, or at any other time, please contact your distributor.
Page 93
MI 3105 EurotestXA Technical specifications
93
8 Technical specifications
8.1 Insulation resistance
Insulation LN, LPE, NPE
Insulation resistance (nominal voltages 50 VDC, 100 V
DC
and 250 VDC)
Measuring range according to EN61557 is 0.25 MΩ ÷ 199.9 MΩ.
Measuring range (MΩ) Resolution (MΩ)
Accuracy
0.00 ÷ 19.99
0.01
±(5 % of reading + 5 digits)
20.0 ÷ 99.9 ±(10 % of reading)
100.0 ÷ 199.9
0.1
±(20 % of reading)
Insulation resistance (nominal voltages 500 VDC and 1000 VDC) Measuring range according to EN61557 is 0.15 MΩ ÷ 1000 MΩ.
Measuring range (MΩ) Resolution (MΩ)
Accuracy
0.00 ÷ 19.99
0.01
±(5 % of reading + 3 digits)
20.0 ÷ 199.9
0.1
200 ÷ 299
1
±(10 % of reading)
300 ÷ 1000
1
±(20 % of reading)
Insulation ALL and ‘L-PE,N-PE’, ‘L-N,L-PE’
Insulation resistance (nominal voltages 50 VDC, 100 VDC, 250 VDC, 500 VDC, 1000 VDC) Measuring range according to EN61557 is 0.34 MΩ ÷ 30.0 MΩ.
Measuring range (MΩ) Resolution (MΩ)
Accuracy
0.00 ÷ 19.99
0.01
20.0 ÷ 30.0
0.1
±(10 % of reading + 5
digits)
Voltage
Measuring range (V) Resolution (V) Accuracy
0 ÷ 1200
1
±(3 % of reading + 3 digits)
Nominal voltages .............................. 50 V
DC
, 100 VDC, 250 VDC, 500 VDC, 1000 VDC
Open circuit voltage .......................... -0 % / + 20 % of nominal voltage
Measuring current............................. min. 1 mA at RN=UN
×1 kΩ/V
Short circuit current........................... max. 0.6 mA
Specified accuracy is valid if universal test cable is used while it is valid up to 100 MΩ if tip commander is used. Specified accuracy is valid up to 100 MΩ if relative humidity > 85 %. In case the instrument gets moistened, the results could be impaired. In such case, it is recommended to dry the instrument and accessories for at least 24 hours. The error in operating conditions could be at most the error for reference conditions (specified in the manual for each function)
±5 % of measured value.
The number of possible tests............ > 1200, with a fully charged battery
Auto discharge after test.
Page 94
MI 3105 EurotestXA Technical specifications
94
8.2 Continuity
8.2.1 Resistance R200mA (L-PE, N-PE)
Measuring range according to EN61557 is 0.16 Ω ÷ 1999 Ω.
Measuring range R (Ω) Resolution (Ω)
Accuracy
0.00 ÷ 19.99
0.01
±(3 % of reading + 3 digits)
20.0 ÷ 199.9
0.1
200 ÷ 1999
1
±(5 % of reading)
2000 ÷ 9999
1 Indicator only
Measuring range R+, R- (Ω) Resolution (Ω)
Accuracy
0.00 ÷ 19.9
0.1
±(5 % of reading + 5 digits)
20.0 ÷ 199.9
0.1
200 ÷ 1999
1
±(10 % of reading)
2000 ÷ 9999
1 Indicator only
Open-circuit voltage.......................... 6.5 VDC ÷ 9 VDC
Measuring current............................. min. 200 mA into load resistance of 2 Ω
Test lead compensation.................... up to 20 Ω
The number of possible tests ........... > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.
8.2.2 Resistance R7mA (L-PE, N-PE)
Measuring range (Ω) Resolution (Ω)
Accuracy
0.0 ÷ 19.9
0.1
20 ÷ 1999
1
±(5 % of reading + 3 digits)
2000 ÷ 9999
1 Indicator only
Open-circuit voltage.......................... 6.5 VDC
÷ 9 VDC
Short-circuit current .......................... max. 8.5 mA
Test lead compensation.................... up to 20 Ω
8.3 RCD testing
8.3.1 General data
Nominal residual current ................... 10 mA, 30 mA, 100 mA, 300 mA, 500 mA,
1000 mA
Nominal residual current accuracy.... -0 / +0.1⋅IΔ; IΔ = IΔN, 2×IΔN, 5×IΔN
-0.1IΔ / +0; IΔ = 0.5×IΔN
AS / NZ selected: ± 5 %
Test current shape............................ Sine-wave (AC), pulsed (A), DC (B)
DC offset for pulsed test current ....... 6 mA (typical)
RCD type .......................................... G (non-delayed), S (time-delayed)
Test current starting polarity ............. 0 º or 180 º
Voltage range ................................... 50 V ÷ 264 V (14 Hz ÷ 500 Hz)
Page 95
MI 3105 EurotestXA Technical specifications
95
RCD test current selection (r.m.s. value calculated to 20ms) according to IEC 61009:
IΔN × 1/2 IΔN × 1 IΔN × 2 IΔN × 5 RCD IΔ
IΔN (mA)
AC A B AC A B AC A B AC A B AC A B
10 5 3.5 5 10 20 20 20 40 40 50 100 100
 
30 15 10.5 15 30 42 60 60 84 120 150 212 300
 
100 50 35 50 100 141 200 200 282 400 500 707 1000
 
300 150 105 150 300 424 600 600 848 n.a. 1500 n.a. n.a.
 
500 250 175 250 500 707 1000 1000 1410 n.a. 2500 n.a. n.a.
 
1000 500 350 500 1000 1410 n.a. 2000 n.a. n.a. n.a. n.a. n.a.
n.a.
n.a..................................................... not applicable
AC type............................................. sine wave test current
A type…… ........................................ pulsed current
B type................................................ smooth DC current
8.3.2 Contact voltage RCD-Uc
Measuring range according to EN61557 is 20.0 V ÷ 33.0V for limit contact voltage 25V Measuring range according to EN61557 is 20.0 V ÷ 66.0V for limit contact voltage 50V
Measuring range (V) Resolution (V) Accuracy
0.0 ÷ 19.9
(-0 % / +15 %) of reading ± 10 digits
20.0 ÷ 99.9
0.1 (-0 % / +15 %) of reading
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free of interfering voltages.
Test current....................................... max. 0.5×IΔN
Limit contact voltage ......................... 25 V, 50 V
Specified accuracy is valid for complete operating range.
8.3.3 Trip-out time
Complete measurement range corresponds to EN 61557 requirements. Maximum measuring times set according to selected reference for RCD testing.
Measuring range (ms) Resolution (ms) Accuracy
0 ÷ 40 *
0.1
±1 ms
0 ÷ max. time *
0.1
±3 ms
* For max. time see normative references in 4.4.2 – this specification applies to max. time
>40 ms.
Test current....................................... ½×IΔN, IΔN, 2×IΔN, 5×I
ΔN
5×IΔN is not available for IΔN=1000 mA (RCD type AC) or I
ΔN
300 mA (RCD types A, B).
2×IΔN is not available for IΔN=1000 mA (RCD type A) or I
ΔN
300 mA (RCD type B).
1×IΔN is not available for IΔN=1000 mA (RCD type B). Specified accuracy is valid for complete operating range.
Page 96
MI 3105 EurotestXA Technical specifications
96
8.3.4 Trip-out current
Trip-out current Complete measurement range corresponds to EN 61557 requirements.
Measuring range IΔ Resolution
Accuracy
0.2×IΔN ÷ 1.1×IΔN (AC type) 0.05×IΔN ±0.1×IΔN
0.2×IΔN ÷ 1.5×IΔN (A type, IΔN 30 mA) 0.05×IΔN ±0.1×IΔN
0.2×IΔN ÷ 2.2×IΔN (A type, IΔN <30 mA) 0.05×IΔN ±0.1×IΔN
0.2×IΔN ÷ 2.2×IΔN (B type) 0.05×IΔN ±0.1×IΔN
Trip-out time
Measuring range (ms) Resolution (ms) Accuracy
0 ÷ 300
1
±3 ms
Contact voltage
Measuring range (V) Resolution (V) Accuracy
0.0 ÷ 19.9
0.1
(-0 % / +15 %) of reading ± 10 digits
20.0 ÷ 99.9
0.1 (-0 % / +15 %) of reading
The accuracy is valid if mains voltage is stabile during the measurement and PE terminal is free of interfering voltages. Trip-out measurement is not available for IΔN=1000 mA (RCD type B). Specified accuracy is valid for complete operating range.
8.4 Fault loop impedance and prospective fault current
8.4.1 No disconnecting device or FUSE selected
Fault loop impedance Measuring range according to EN61557 is 0.25 Ω ÷ 19999 Ω.
Measuring range (Ω) Resolution (Ω)
Accuracy
0.00 ÷ 9.99
0.01
10.0 ÷ 99.9
0.1
100 ÷ 19999
1
±(5 % of reading + 5 digits)
Prospective fault current (calculated value)
Measuring range (A) Resolution (A) Accuracy
0.00 ÷ 9.99
0.01
10.0 ÷ 99.9
0.1
100 ÷ 999
1
1.00k ÷ 9.99k
10
10.0k ÷ 23.0k
100
Consider accuracy of fault
loop resistance
measurement
The accuracy is valid if mains voltage is stabile during the measurement.
Test current (at 230 V)...................... 6.5 A (10 ms)
Nominal voltage range ...................... 30 V ÷ 500 V (14 Hz ÷ 500 Hz)
Page 97
MI 3105 EurotestXA Technical specifications
97
8.4.2 RCD selected
Fault loop impedance Measuring range according to EN61557 is 0.46 Ω ÷ 19999 Ω.
Measuring range (Ω) Resolution (Ω)
Accuracy *
0.00 ÷ 9.99
0.01
±(5 % of reading + 10 digits)
10.0 ÷ 99.9
0.1
±10 % of reading
100 ÷ 19999
1
±10 % of reading
* Accuracy may be impaired in case of heavy noise on mains voltage
Prospective fault current (calculated value)
Measuring range (A) Resolution (A) Accuracy
0.00 ÷ 9.99
0.01
10.0 ÷ 99.9
0.1
100 ÷ 999
1
1.00k ÷ 9.99k
10
10.0k ÷ 23.0k
100
Consider accuracy of fault
loop resistance
measurement
Nominal voltage range ...................... 50 V ÷ 500 V (14 Hz ÷ 500 Hz)
No trip out of RCD. R, XL values are indicative.
8.5 Line impedance and prospective short-circuit current
Line impedance Measuring range according to EN61557 is 0.25 Ω ÷ 19.9 kΩ.
Measuring range (Ω) Resolution (Ω)
Accuracy
0.00 ÷ 9.99
0.01
10.0 ÷ 99.9
0.1
100 ÷ 999
1
1.00k ÷ 9.99k
10
10.0k ÷ 19.9k
100
±(5 % of reading + 5 digits)
Prospective short-circuit current (calculated value)
Measuring range (A) Resolution (A) Accuracy
0.00 ÷ 0.99
0.01
1.0 ÷ 99.9
0.1
100 ÷ 999
1
1.00k ÷ 99.99k
10
100k ÷ 199k
1000
Consider accuracy of line
resistance measurement
Test current (at 230 V)...................... 6.5 A (10 ms)
Nominal voltage range ...................... 30 V ÷ 500 V (14 Hz ÷ 500 Hz)
R, XL values are indicative.
Page 98
MI 3105 EurotestXA Technical specifications
98
8.6 Voltage, frequency, and phase rotation
8.6.1 Phase rotation
Nominal system voltage range.......... 100 VAC ÷ 550 VAC
Nominal frequency range.................. 14 Hz ÷ 500 Hz
Result displayed ............................... 1.2.3 or 3.2.1
8.6.2 Voltage Measuring range (V) Resolution (V) Accuracy
0 ÷ 550
1
±(2 % of reading + 2 digits)
Result type........................................ True r.m.s. (trms)
Nominal frequency range.................. 0 Hz, 14 Hz ÷ 500 Hz
8.6.3 Frequency
Measuring range (Hz) Resolution (Hz) Accuracy
0.00 ÷ 999.99
0.01
±(0.2 % of reading + 1 digit)
Nominal voltage range ...................... 10 V ÷ 550 V
8.7 Online terminal voltage monitor
Measuring range (V) Resolution (V) Accuracy
10 ÷ 550
1
±(2 % of reading + 2 digits)
Page 99
MI 3105 EurotestXA Technical specifications
99
8.8 Earth resistance
Earth resistance three-wire method
Measuring range according to EN61557 is 0.67 Ω ÷ 9999 Ω
Measuring range (Ω) Resolution (Ω)
Accuracy
0.00 ÷ 19.99
0.01
20.0 ÷ 199.9
0.1
±(3 % of reading + 3 digits)
200 ÷ 1999
1
± 5 % of reading
2000 ÷ 9999
1
± 10 % of reading
Additional spike resistance error if
Rc max. or Rp max. is exceeded ...... ±(5 % of reading + 10 digits)
Rc max. ............................................ 100 RE or 50 kΩ (whichever is lower)
Rp max. ............................................ 100 RE or 50 kΩ (whichever is lower)
Automatic test of probe resistance.......... yes
Additional error
at 3 V noise (50 Hz) .......................... ±(5 % of reading +10 digits)
Automatic test of voltage noise ......... yes
Noise voltage indication threshold .... 1 V (<50 Ω, worst case)
Open-terminal test voltage................ 40 VAC
Test voltage frequency...................... 125 Hz
Short-circuit test current.................... < 20 mA
Earth resistance one clamp method
Measuring range (Ω) Resolution (Ω)
Accuracy
0.00 ÷ 19.99
0.01
20.0 ÷ 199.9
0.1
±(3 % of reading + 3 digits)
200 ÷ 1999
1
± 5 % of reading
2000 ÷ 9999
1
± 10 % of reading
Additional spike resistance error if
Rc max. or Rp max. is exceeded ...... ±(5 % of reading + 10 digits)
Rc max. ............................................ 100 R
E
or 50 kΩ (whichever is lower)
Rp max. ............................................ 100 RE or 50 kΩ (whichever is lower)
Automatic test of probe resistance.......... yes
Resistance ratio induced error .......... 2 % x R/Re
Additional error
R and Re, at 3 V noise (50 Hz)..... ±(5 % of reading +10 digits)
R, 2 A noise (50 Hz)................... ±(10 % of reading +10 digits)
Automatic test of voltage noise ......... yes
Noise voltage indication threshold .... 1 V (<50 Ω, worst case)
Open-terminal test voltage................ 40 V
AC
Test voltage frequency...................... 125 Hz
Short-circuit test current.................... <20 mA
Low clamp current indication ............ yes
Noise current indication .................... yes
Additional clamp error has to be considered.
Page 100
MI 3105 EurotestXA Technical specifications
100
Earth resistance two clamps method Measuring range (Ω) Resolution (Ω)
Accuracy*
0.00 ÷ 19.9
0.01
±(10 % of reading + 10 digits)
20.0 ÷ 30.0
0.1
±(20 % of reading )
30.1 ÷ 39.9
0.1
±(30 % of reading )
* Distance between test clamps >30 cm.
Additional error
at 3 A / 50 Hz noise into 1 Ω ............. ±(10 % of reading)
Test voltage frequency...................... 125 Hz
Noise current indication .................... yes
Low clamp current indication ............ yes
Additional clamp error has to be considered.
Specific earth resistance Measuring range (Ωm) Resolution (Ωm)
Accuracy
0.0 ÷ 99.9
0.1
100 ÷ 999
1
1.00k ÷ 9.99k
0.01k
10.0k ÷ 99.9k
0.1k
>100k 1k
See accuracy note
Measuring range (Ωft) Resolution (Ωft)
Accuracy
0.0 ÷ 99.9
0.1
100 ÷ 999
1
1.00k ÷ 9.99k
0.01k
10.0k ÷ 99.9k
0.1k
>100k 1k
See accuracy note
Principle:
ρ = 2·π·distance·Re,
with Re as measured resistance in 4-wire method.
Accuracy note:
Accuracy of the specific earth resistance result depends on measured resistance
Re and is as follows:
Measuring range (Ω)
Accuracy
1.00 ÷ 1999 ±5 % of measured 2000÷ 19.99k ±10 % of measured >20k
±20 % of measured
Additional error
See Earth resistance three-wire method.
Loading...