Met One Instruments GT-521S User Manual

Met One Instruments, Inc
Corporate Sales & Service: 1600 NW Washington Blvd. Grants Pass, OR 97526 Tel (541) 471-7111 Fax (541) 471-7116
GT-521S
MANUAL
GT-521S Manual Rev D
Copyright Notice
Contact Information:
Tel: (541) 471 7111 Fax: (541) 471 7115 Web: www.metone.com Email: service@metone.com
Address:
Met One Instruments, Inc. 1600 NW Washington Blvd Grants Pass, Oregon 97526 U.S.A.
GT-521S Manual © Copyright 2014 Met One Instruments, Inc. All Rights Reserved Worldwide. No part of
this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any form by any means without the express written permission of Met One Instruments, Inc.
Technical Support
This manual is structured by customer feedback to provide the required information for setup, operation, testing, maintaining, and troubleshooting your GT-521S. Should you still require support after consulting your printed documentation, we encourage you to contact one of our expert Technical Service representatives during normal business hours of 7:00 a.m. to 4:00 p.m. Pacific Standard Time, Monday through Friday. In addition, technical information and service bulletins are often posted on our website. Please contact us and obtain a Return Authorization (RA) number before sending any equipment back to the factory. This allows us to track and schedule service work and to expedite customer service. Please have your instrument serial number available when contacting the manufacturer.
NOTICE
CAUTIONUse of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
WARNINGThis product, when properly installed and operated, is considered a Class I laser product. Class I products are not considered to be hazardous.
There are no user serviceable parts located inside the cover of this device. Do not attempt to remove the cover of this product. Failure to comply with this
instruction could cause accidental exposure to laser radiation.
Table of Contents
GT-521S Manual Rev D Page 1
1. Introduction ........................................................................................................... 4
2. Setup ...................................................................................................................... 4
2.1. Unpacking .................................................................................................................................. 4
2.2. Layout ........................................................................................................................................ 7
2.3. Default Settings ......................................................................................................................... 8
2.4. Initial Operation ......................................................................................................................... 8
3. User Interface ........................................................................................................ 8
4. Operation ............................................................................................................... 9
4.1. Power Up ................................................................................................................................... 9
4.2. Operate Screen ......................................................................................................................... 9
4.2.1. Sampling ........................................................................................................................... 10
4.2.2. Sample Status ................................................................................................................... 10
4.2.3. Sample History .................................................................................................................. 10
4.2.4. Warning / Error Messages ................................................................................................ 11
4.3. Sample Related Functions ...................................................................................................... 11
4.3.1. Starting/Stopping ............................................................................................................... 11
4.3.2. Real-Time Serial Output ................................................................................................... 11
4.3.3. Samples ............................................................................................................................ 11
4.3.4. Sample Time ..................................................................................................................... 11
4.3.5. Hold Time .......................................................................................................................... 11
4.3.6. Count Mode ....................................................................................................................... 12
4.3.7. Sample Timing .................................................................................................................. 12
5. Menu Selections .................................................................................................. 13
5.1. Edit Menu Items ....................................................................................................................... 13
5.2. Sample Setup Screen .............................................................................................................. 13
5.2.1. Location Number ............................................................................................................... 14
5.2.2. Sample Time ..................................................................................................................... 14
5.2.3. Hold Time .......................................................................................................................... 14
5.2.4. Samples ............................................................................................................................ 14
5.3. Settings Screen ....................................................................................................................... 14
5.3.1. Volume .............................................................................................................................. 14
5.3.2. Temperature ...................................................................................................................... 15
5.3.3. Count mode ....................................................................................................................... 15
5.4. Recall Data Screen .................................................................................................................. 15
5.5. Print Data Screen .................................................................................................................... 16
5.6. Memory Screen ....................................................................................................................... 16
5.6.1. View Available Memory ..................................................................................................... 17
5.6.2. Clearing Memory ............................................................................................................... 17
5.7. Count Alarms Screen .............................................................................................................. 17
5.7.1. Alarm Analog Output ......................................................................................................... 18
5.8. Serial ........................................................................................................................................ 18
5.8.1. Report Type ...................................................................................................................... 18
5.8.2. Baud .................................................................................................................................. 18
GT-521S Manual Rev D Page 2
5.8.3. Serial Output Mode ........................................................................................................... 19
5.9. Set Flow Screen ...................................................................................................................... 19
5.10. Set Sizes Screen ..................................................................................................................... 19
5.11. Set Clock Screen ..................................................................................................................... 20
5.12. Set Contrast Screen ................................................................................................................ 20
5.13. About Screen ........................................................................................................................... 20
6. Charging the Battery ........................................................................................... 21
7. Serial Communications ...................................................................................... 22
7.1. Serial Interface ........................................................................................................................ 23
7.2. Commands .............................................................................................................................. 24
7.3. Real Time Output .................................................................................................................... 24
7.4. Comma Separated Value (CSV) ............................................................................................. 25
7.5. Printer Format .......................................................................................................................... 25
8. Maintenance ........................................................................................................ 26
8.1. Service Schedule ..................................................................................................................... 26
8.2. Zero Count Test ....................................................................................................................... 26
8.3. Flow Rate Test......................................................................................................................... 26
8.4. Annual Calibration ................................................................................................................... 26
8.5. Flash Upgrade ......................................................................................................................... 27
9. Troubleshooting .................................................................................................. 27
10. Specifications ...................................................................................................... 28
11. Warranty / Service ............................................................................................... 29
GT-521S Manual Rev D Page 3
1. Introduction
The GT-521S is a full–featured, two-channel, battery operated, handheld particle counter. The two count channels can be configured to factory calibrated standard sizes or user defined custom sizes. It also has two alarm limits and various communication options. This instrument can store up to 8,000 sample events including data from the optional temperature / relative humidity probe (PN G3120). Sample history events can be viewed on the LCD display, printed on the optional printer (PN G3115) and downloaded to a computer.
2. Setup
The following sections cover unpacking, layout and performing a test run to verify operation.
2.1. Unpacking
Unpack and inspect the contents of the shipping container. Standard items (included) are shown in Figure 1 – Standard Accessories. Optional accessories are shown in Figure 2 – Optional Accessories. Contact the supplier if any items are missing. Any damages incurred during shipping are the responsibility of the carrier. If any damage to the shipment is noticed before unpacking, a claim must be filed with the commercial carrier immediately. You should follow any special unpacking instructions provided by the carrier as you then carefully remove all items from the containers and inspect each component. It is recommended to document and photograph all damaged packages and items before, during, and after unpacking them. Contact Met One Instruments (see the Technical Support section at the beginning of this manual) to arrange for any replacement items needed.
GT-521S Manual Rev D Page 4
ATTENTION:
The included USB drivers must be installed before connecting the GT-521S USB port to your computer. If the supplied drivers are not installed first, Windows may install generic drivers that are not compatible with this product.
To install USB drivers: Insert the Comet CD. The install program should run automatically and display the screen below. If an AutoPlay pop-up window appears, select “Run AutoRun.exe”. Finally, select “USB Drivers” to start the install process.
Users using a 32 bit system should install the USB 32 bit drivers. Users using a 64 bit system should install the USB 64 bit drivers.
GT-521S Manual Rev D Page 5
GT-521S Standard Equipment
GT-521S
PN GT-521S
Battery Charger
PN 390031
Power Cord
PN 400113
Carrying Case
PN 8517
Calibration Certificate
PN GT-521S-9600
ISO-Kinetic Probe
PN G3110
Zero Count Filter
PN G3111
USB Cable PN 500787
Comet Software CD
PN 80248
Manual
PN GT-521S-9800
Printer
PN G3115
Flow Meter
PN 9801
Serial Cable
PN 3228
T/RH Probe
PN G3120
Figure 1 – Standard Accessories
Figure 2 – Optional Accessories
GT-521S Manual Rev D Page 6
2.2. Layout
Component
Description
Power Switch
Switch that turns the GT-521S on or off. Slide up (towards inlet nozzle) to turn on and slide down to turn off.
Charging Jack
Input jack for the battery charger. This connection charges the internal battery pack and provides continuous operating power for the unit.
USB Port
For USB serial communication
Serial Port
For serial communication (RS-232 or RS-485).
T/RH Connector
Mating connector for the optional external Temperature/RH sensor.
Inlet Nozzle
Ambient air inlet nozzle. Connect isokinetic probe to reduce turbulence in the air sample.
Charging Jack
Isokinetic Inlet Probe
T/RH Connector
USB Port
Serial Port
Power Switch
Inlet Nozzle (remove cap)
The following figure shows the layout of the GT-521S and provides a description of each of the components.
Figure 3 – GT-521S Layout
GT-521S Manual Rev D Page 7
2.3. Default Settings
Parameter
Value
Sizes
0.3 µm and 0.5 µm
Sample Location
1
Samples
Single
Sample Time
60 seconds
Sample Hold Time
0 seconds
Volume (concentration)
CF (particles/ft3)
Count Mode
Normal (cumulative)
Temperature Units
C
Baud Rate
9600
Serial Mode
RS-232
Alarm Limits
Disabled
Key
Description
Starts or stops a sample.  Starts the unit printing when in the Print Data Screen.  Recalls the selected data when in the Recall Data Screen.
Loads the Menu Screen when not editing a field.  Loads the Operate Screen when in the Menu Screen.  Cancels editing a field and return the field to the original value.
Loads the screen associated with menu item.  Begins editing of the selected item.  Stops editing a field and saves the changed value.
Navigates up/down when not editing (Menu).  Modifies field when editing.
Navigates right / left
Move through history buffer (Operate screen)
MENU
ESC
ENTER
STOP
START
The GT-521S comes with the user settings configured as follows.
2.4. Initial Operation
Before operating the GT-521S for the first time, it is recommended that the unit be fully charged. Information regarding charging the battery is found in section 6 of this manual.
Complete the following steps to verify proper operation.
1. Slide the power switch up to turn on power.
2. Observe the Startup Screen for 2 seconds then the Operate Screen (Section
4.2).
3. Press Start / Stop key. The GT-521S will sample for 1 minute and stop.
4. Observe the counts on the display.
5. The unit is ready for use
3. User Interface
The GT-521S user interface is composed of a 7 button keypad and a LCD display. The following table describes keypad functionality. Note, some keys have more than one function.
GT-521S Manual Rev D Page 8
4. Operation
GT-521S
WWW.METONE.COM
18 AUG’06 11:23
0.3µ 0 CF
0.5µ 0 CF Location: 001
The following sections cover the basic operation.
4.1. Power Up
GT-521S power is controlled by a slide switch located on the right hand side of the unit. Move the power switch to the on position (towards the top of the case) to power up the unit.
The first screen shown on power up is the Startup Screen (Figure 4). This screen displays the product type and company website for 2 seconds before loading the Operate Screen.
Figure 4 – Startup Screen
4.2. Operate Screen
The Operate Screen displays the date/time, sample status, current sample data and previous sample data. Figure 5 shows the Operate Screen.
Figure 5 – Operate Screen w/o Temp/RH
The top line of the Operate Screen shows date and time or status/alarm messages depending upon the machine status. Lines 2 and 3 display the two count channels. Line 4 displays the location number or optional Temp/RH data. Use the up/down arrows to alternate between Location number and Temp/RH when the optional G3120 probe is attached.
Particle count units are user selectable. The selections include: Total counts (TC), particles per liter (/L), particles per cubic foot (CF) and particles per cubic meter (M3). Ambient temperature can be displayed in units of Celsius (ºC) or Fahrenheit (ºF). Both unit settings are discussed in section 5.2.4.
GT-521S Manual Rev D Page 9
4.2.1. Sampling
COUNTING... 52
0.3µ 29,780 CF
0.5µ 1,400 CF Location: 001
Status
Description
STARTING…
GT-521S is starting the sample and is waiting for the count system to initialize.
HOLDING…
GT-521S is taking multiple samples and is waiting for the hold time to finish.
STOPPING…
GT-521S is stopping the sample and is waiting for the count system to stop.
COUNTING… 10
GT-521S is sampling. The time remaining is displayed on the far right.
18 AUG’06 11:23
0.3µ 0 CF
0.5µ 0 CF
Location: 001
The Operate Screen displays current sample information when the unit is sampling (real time data). Concentration values (/L, CF, M3) are time dependent so these values may fluctuate early in the sample; however, after several seconds the measurement will stabilize. Longer samples (e.g. 60 seconds) will improve concentration measurement accuracy. Figure 6 shows the Operate Screen while sampling.
Figure 6 – Operate Screen Sampling
4.2.2. Sample Status
The top line of the Operate Screen displays the status of the GT-521S while the unit is sampling. The following table shows the various status messages:
4.2.3. Sample History
Sample history (previous data) can be viewed on the Operate Screen when the unit is stopped (not sampling). To view sample history, press Enter or from the Operate Screen. The unit will display the last sample event (newest record) and display in the upper right corner of the display (see Figure 7) to indicate history data, and displays the date and time stamp when the sample was collected. Press ◄ or ► to move through sample history one record at a time ( displays older events, displays newer events). Press the Enter key at any time to return to the Operate Screen. Press Start at any time to start a new sample.
Figure 7 – History Screen
GT-521S Manual Rev D Page 10
4.2.4. Warning / Error Messages
Display Message
Description
<COUNT ALARM>
Count alarm. The count is the alarm limit.
LOW BATTERY!
Low battery warning. Less than 15 minutes of normal operation remaining.
SENSOR ERROR!
Particle sensor error.
The GT-521S displays warning/error messages on the top line of the Operate Screen. These messages alternate with the normal header. The following table lists the warning/error messages:
4.3. Sample Related Functions
The following sub-sections cover sample related functions.
4.3.1. Starting/Stopping
To start or stop a sample, press the START/STOP key. A sample event can be manually started or stopped from either the Operate Screen or the menu.
4.3.2. Real-Time Serial Output
GT-521S provides real-time output on the serial port at the end of each sample. The format of the output is controlled by the report type setting (section 5.8.1).
4.3.3. Samples
The Samples setting controls the number of samples to take. Repeat configures the unit for continuous sampling. Single configures the unit for a single sample. 2-999 configures the unit for a number of samples. Samples are discussed in section 5.2.
4.3.4. Sample Time
The sample time determines the amount of time that counts are accumulated. The length of the sample is user settable from 1 – 9999 seconds and is discussed in section
5.2.
4.3.5. Hold Time
The hold time is used when Samples is set for more than one sample. The hold time represents the time from the completion of the last sample to the start of the next sample. The hold time is user settable from 0 – 9999 seconds and is discussed in section 5.2.
GT-521S Manual Rev D Page 11
4.3.6. Count Mode
COUNTING... 58
0.3µ 29,780 ΔCF
0.5µ 1,400 ΔCF Location: 001
Start
Time
Sample Time
Stop
Time
Start
Time
Sample Time Hold Time Sample Time
Stop
Time
The count mode is used to set whether counts are displayed as Cumulative (normal) or Differential. When in differential mode, the volume units will be preceded with the delta character Δ. Count mode is discussed in more detail in section 5.3.
Figure 8 – Differential Counts
4.3.7. Sample Timing
The following figures depict the sample timing sequence for both single and multiple samples. Figure 9 shows the timing for single samples. Figure 10 shows the timing for multiple samples.
Figure 9 – Single Sample
Figure 10 – Multiple Samples
GT-521S Manual Rev D Page 12
5. Menu Selections
Menu Item
Description
SAMPLE SETUP
View / change location number, sample time, hold time and samples.
SETTINGS
View / change volume (count units), ºC / ºF and Count mode (Normal or Difference)
RECALL DATA
Recall sample history based on time
PRINT DATA
Print stored data based on a time range.
MEMORY
View memory usage and clear memory.
COUNT ALARMS
Set count alarm limits for 2 particle sizes
SERIAL
Set output report type, baud rate and serial output mode
SET FLOW
Adjust the sample flow rate
SET SIZES
Set particle sizes
SET CLOCK
Set the date and time.
SET CONTRAST
Adjust the display contrast.
ABOUT
Display firmware version and serial number.
LOCATION: 001 SAMPLE TIME: 0060 HOLD TIME: 0000 SAMPLES: SINGLE
The Menu is accessible by pressing the Menu key on the Operate Screen. The table below shows the menu items. Press ▲ or ▼ to navigate to a menu item then press Enter to display a screen where you can view or change item setting(s).
5.1. Edit Menu Items
To change settings, press Menu to display the main menu, press ▲ or ▼ to navigate to the desired item and press Enter to display the item view/edit screen.
To edit pick list items (e.g. Settings Temp – C/F), press ▲ or ▼ to navigate to the item.
Press Enter to select the item. Press ▲ or ▼ to change the setting. Press ENTER to
save the setting or ESC to cancel and return to the previous value. To edit numeric values (e.g. Count Alarms - Alarm Limit), press ▲ or ▼ to navigate to
the item. Press Enter to select the item. Press ▲ or ▼ to increment or decrement a
value. Press ◄ or ► to select the next digit. Press ENTER to save the value or ESC to
cancel and return to the previous value.
5.2. Sample Setup Screen
Figure 11 shows the Sample Setup Screen. The four parameters are covered in the following sections.
Figure 11 – Sample Setup Screen
GT-521S Manual Rev D Page 13
5.2.1. Location Number
Selection
Description
REPEAT
Repeat configures the unit for continuous sampling.
SINGLE
Single configures the unit for a single sample.
002-999
Configures the unit to take N samples.
VOLUME: CF TEMP: C COUNT NORMAL
The location number is used to assign a unique number to a location or area. This important field is included in sample data records (display, printer and serial output).
5.2.2. Sample Time
The sample time determines the amount of time that counts are accumulated while the pump is running. The length of the sample is user settable from 1 – 9999 seconds.
5.2.3. Hold Time
The Hold time is the time between samples when sampling more than a single sample. The Hold time is user settable from 0 – 9999 seconds. The pump will remain on during the hold period if the Hold time is 60 seconds or less. If the Hold time is greater than 60 seconds, the pump will stop after each sample, and start a few seconds before the next sample. Hold times greater than 60 seconds will increase pump life.
5.2.4. Samples
The samples setting controls the number of samples to take as illustrated below.
5.3. Settings Screen
Figure 12 shows the Settings Screen. The four parameters are covered in the sections immediately following.
Figure 12 – Settings Screen
5.3.1. Volume
The GT-521S supports total counts (TC), particles per liter (/L), particles per cubic foot (CF), and particles per cubic meter (M3). Particle count information updates while the unit is sampling. Concentration values (/L, CF, M3) are time dependent so these values may fluctuate early in the sample; however, after several seconds the measurement will stabilize. Longer samples (e.g. 60 seconds) will improve concentration measurement accuracy. In total count mode, counts accumulate throughout the sample time.
GT-521S Manual Rev D Page 14
5.3.2. Temperature
Selection
Description
NORMAL
Displays cumulative counts on the screen and in reports.
DIFFERENCE
Displays differential counts on the screen and in reports.
COUNTING... 58
0.3µ 29,780 CF
0.5µ 1,400 CF Location: 001
COUNTING... 58
0.3µ 28,380 ΔCF
0.5µ 790 ΔCF Location: 001
RECALL DATA
01 JAN’00 00:00
The GT-521S displays temperature in Celsius (C) or Fahrenheit (F).
5.3.3. Count mode
The GT-521S can display counts in two different modes. These are Normal and Differential.
The Normal mode displays cumulative counts. For a given size the counts represent all particles that size and larger.
The Differential mode displays the number of particles between a given size and the size larger. When in differential mode, the volume units will be preceded the delta character Δ on the screen. For the serial output a lower case ‘d’ character is displayed in the header.
Figure 13 – Cumulative Counts
Figure 14 – Differential Counts
5.4. Recall Data Screen
Stored sample events can be viewed from the Operate Screen but this requires navigating one record at a time to reach a desired record. The Recall Data Screen provides a way to quickly navigate to a record based on time. Figure 15 shows the Recall Data Screen.
Figure 15 – Recall Data Screen
GT-521S Manual Rev D Page 15
To recall data, enter the desired date/time string and select the START/STOP button.
PRINT DATA LOCATION: 000 01 JAN00 00:00 18 AUG06 13:23
Setting
Description
LOCATION
The location ID of the sample events to print. Setting location to 000 prints all locations. Settable from 0 - 999
01 JAN00
The date/time to begin printing sample events from.
18 AUG06
The date/time to stop print samples at.
PRINTING STATUS SCANNING...15 PRINTING...10 FINISHED!
The unit will recall the data from the date/time entered (if an exact match is found) or the next most recent data available. The screen shown when the data is recalled is the Operate Screen. The unit will display “” on the upper right corner of the display to indicate history data. Use ◄ or ► to navigate through the stored sample events.
5.5. Print Data Screen
Stored sample events can be printed through the printer port within a user selected range. To print selected data, select PRINT DATA from the Menu. Figure 16 shows the Print Data Screen.
Figure 16 – Print Data Screen
Edit the location and time range to select which sample events to print. The following table describes settings.
After the print settings have been selected, press the Start/Stop button to begin printing the sample events. Once the Start/Stop button has been pressed the Printing Status Screen is displayed. Figure 17 shows the Printing Status Screen as it would look when finished.
Figure 17 – Printing Status Screen
Pressing the ESC button cancels the data printing and loads the menu. The format of the print is dependent upon the report type setting (Section 5.8.1)
5.6. Memory Screen
The GT-521S memory is composed of a single file which contains the data from sample events. Every time a sample is completed, the GT-521S stores that data into the memory. The GT-521S memory is circular, meaning when the memory is full, the unit will start overwriting the oldest saved samples with new samples. GT-521S provides the user the ability to view the memory usage as well as clear the memory.
GT-521S Manual Rev D Page 16
5.6.1. View Available Memory
FREE: 100% SAMPLES: 8000 PRESS ENTER TO CLEAR MEMORY!
COUNT ALARMS OFF SIZE ALARM LIMIT
0.3 00000000
0.5 00000000
Memory Screen is used to view available memory or to clear the memory. The Memory Screen is accessed by selecting MEMORY from the menu. Figure 18 shows the Memory Screen.
Figure 18 – Memory Screen
FREE shows the percent of space available for data storage. When 0% is displayed, memory is full and the oldest data will be overwritten by new data.
SAMPLES shows the number of samples which can be stored in memory before the memory is full. When 0 is displayed, memory is full and the oldest data will be overwritten.
5.6.2. Clearing Memory
To clear memory, press the ENTER key while viewing the Memory Screen. This will delete all the sample events in memory. A warning screen will be displayed to prevent accidental erasure.
5.7. Count Alarms Screen
Configure count alarm limits for two particle sizes. To disable count alarms, select OFF in the field next to the COUNT ALARM header. The alarm for a specific particle size can be disabled by setting the alarm value to zero. The alarm is active when the count is equal to or greater than the alarm limit. The maximum alarm limit value is 9,999,999. Figure 19 shows the Count Alarm Screen.
Alarm values do not change with the count units setting (TC, /L, CF, M3). In other words, a value of 1,000 will alarm at 1,000 counts or 1,000 particles per liter or 1,000 particles per cubic foot depending on the count unit setting.
Figure 19 – Set Alarms Screen
GT-521S Manual Rev D Page 17
5.7.1. Alarm Analog Output
Channel
Alarm Type
Analog Output
Channel 1
No Alarm
0.00 V
Channel 1
Hi Alarm
5.00 V
Channel 2
No Alarm
0.00 V
Channel 2
Hi Alarm
5.00 V
REPORT: CSV BAUD: 9600 SERIAL: RS-232
Along with the two channel alarms mentioned in section 5.7, there is also an alarm level analog output for each channel. The analog output is always active and it is not tied to the alarm logic being on or off. The table below shows the analog output levels for each alarm. The analog out is available on the DB-9 Serial connector on the right side of the instrument. See section 7.1 for details.
5.8. Serial
Configure the serial port settings.
Figure 20 – Serial Screen
5.8.1. Report Type
The Report setting determines the output format for the Serial Port. The choices are NONE, CSV and PRINTER.
When set to NONE, the unit will not automatically output the reading at the end of a sample to the Serial Port. CSV is a Comma Separated Values output format suitable for importing into a spreadsheet. PRINTER is the same format as the screen and optional external printer.
Note: When using the optional external printer, set the report type to PRINTER.
5.8.2. Baud
Use the Baud Rate selection to set the serial communications baud rate. GT-521S communicates at baud rates from 300 – 38400. The default baud rate is 9600.
GT-521S Manual Rev D Page 18
5.8.3. Serial Output Mode
Serial Setting
Description
RS-232
RS-232/USB serial communication with CSV style output. Records automatically transmitted after each sample.
RS-485
RS-485 communication with CSV style output.
SET FLOW < > TO ADJUST ENTER TO SAVE
███████
SET SIZES CH1: 00.3µ CH2: 00.5µ
The Serial Output Mode setting controls the behavior of the GT-521S serial output. The modes are RS-232 and RS-485. The following table lists the Serial Output settings and describes their meanings.
5.9. Set Flow Screen
The GT-521S has a factory calibrated flow rate of 0.1 CFM (2.83 LPM). Use the following procedure to adjust the flow rate when a periodic flow rate check (Section 8.3) indicates a flow rate error greater than ± 5%.
1. Connect a reference flow meter to the inlet fitting on the top of the unit. (PN 9801)
2. Access the Set Flow Screen (Figure 21) by pressing Menu then select Set Flow. The pump will start automatically when you enter the Set Flow Screen and stop when you leave the screen.
3. Press ◄ or ► to obtain a reading of 0.1 CFM (2.83 LPM) ± 5% on the reference flow meter.
4. Press Enter to save the setting or ESC to cancel the change.
Figure 21 – Set Flow Screen
5.10. Set Sizes Screen
The GT-521S has factory calibrated particle sizes (for example, 0.3, 0.5, 1.0, 5.0 and 10 µm). These standard sizes will support most applications and provide the best size accuracy (± 10%). This unit also supports custom sizes. These sizes are configured using the Set Sizes Screen (Figure 22). Custom sizes are interpolated using the standard size calibration curve. Therefore, the size accuracy for custom sizes is somewhat reduced (± 15%).
Figure 22 – Set Sizes Screen
GT-521S Manual Rev D Page 19
The unit sorts sizes from small to large after each size change. Duplicate sizes are not
SET CLOCK DATE: 18 AUG’06 TIME: 11:25:36
Date / Time Formats
Date
dd mmm’yy
dd=day, mmm=month, yy=year
Time
hh:mm:ss
hh=hours, mm=minutes, ss=seconds
SET CONTRAST < > TO ADJUST ENTER TO SAVE
███████
GT-521S
80203 R0.1.0
www.metone.com
allowed. Any attempt to set the two sizes to the same value will result in a “DUPLICATE SIZES!” warning message.
5.11. Set Clock Screen
To set the date and time select SET CLOCK from the menu. Figure 23 shows the Set Clock Screen and the following table describes the date / time formats.
Figure 23 – Set Clock Screen
5.12. Set Contrast Screen
Press ◄ or ► to improve display quality. Press Enter to save the setting or ESC to
cancel the change. Figure 24 shows the Set Contrast Screen.
Figure 24 – Set Contrast
5.13. About Screen
Figure 25 shows the About Screen. The About Screen shows the firmware version and programmable logic versions on the second line. Press ▲ or ▼ to toggle between the two version numbers.
Figure 25 – About Screen
GT-521S Manual Rev D Page 20
6. Charging the Battery
Caution:
The provided battery charger is designed to work safely with this device. Do not attempt to connect any other charger or adapter to this device. Doing so may result in equipment damage.
To charge the battery, connect the battery charger to an AC power outlet and the DC barrel connector to the socket on the right side of the GT-521S. The battery charger is universal and will work with power line voltages of 100 to 240 volts, 50 to 60 Hz. A discharged battery pack will take approximately 2.5 hours to fully charge.
When fully charged the battery inside the GT-521S will power the unit for about 6 hours of continuous sampling. Under normal operation, the battery will power the unit for about 24 hours. For continuous operation, operate the unit with the battery charger attached. Charge the battery before storing the GT-521S. Storing a discharged battery will degrade its performance.
GT-521S Manual Rev D Page 21
7. Serial Communications
GT-521S provides serial communications via the USB and DB-9 connectors located on the right hand side of the unit. The following sections discuss the various serial communications available with GT-521S.
ATTENTION:
The included USB drivers must be installed before connecting the GT-521S USB port to your computer. If the supplied drivers are not installed first, Windows may install generic drivers that are not compatible with this product.
To install USB drivers: Insert the Comet CD. The install program should run automatically and display the screen below. If an AutoPlay pop-up window appears, select “Run AutoRun.exe”. Finally, select “USB Drivers” to start the install process.
Users using a 32 bit system should install the USB 32bit drivers. Users using a 64 bit system should install the USB 64 bit drivers.
WARNING:
The optional RS-232 serial cable (PN 3228) is custom built to function with the GT-521S. The use of any other serial cable can cause damage!
GT-521S Manual Rev D Page 22
7.1. Serial Interface
Pin
Function
Communication Type
1
Not Used
2 TX
RS-232
3
RX
RS-232
4
Not Used
5
Ground
RS-232 & RS-485
6
Analog Out 1
DAC
7
Analog Out 2
DAC
8
A+
RS-485
9
B-
RS-485
1 2 3 4 5
6 7 8 9
Mini USB
The serial interface to the GT-521S is a standard 9 pin (DB-9) connector. It is located on the right hand side of the instrument as shown below.
Communication with the GT-521S through the RS-232 DB-9 port requires a custom serial cable (PN3228). Standard serial cables will not work and may cause damage to the instrument if connected.
Below is a description of the DB-9 connector:
GT-521S Manual Rev D Page 23
7.2. Commands
Settings (must match computer settings):
Baud Rate = 9600 (default, see Section 5.8.2 to change this) Parity = None Stop Bits = 1
Command
Description
?,H
Displays the help menu
1
Returns the units settings information
2
Returns all available records from the data file
3
Returns all records since last ‘2’ or ‘3’ command.
4 n
Returns the last n records. If n is omitted this returns the last record.
D
Date (mm/dd/yy) Set Date
T
Time (hh:MM:ss) Set Time
C
Clear data
S
Start a sample
E
End a sample
ST
Sample Time
RV
Show software revision.
ID
Set/Get the Location ID. Range 1-999.
FAx
Alarm Limit setting where x=1 or 2 for Alarm 1 or 2.
SE
Serial Mode (0=RS-232, 1=RS-485)
SF
Alarms on/off where 0=Off, 1=On
SH
Set Hold Time in seconds
SN
Set Number of Samples (0=Repeat, 1=Single, 2-999=N samples)
SR
Set Report Mode (0=NONE, 1=CSV, 2=Printer)
SS
Read serial number
CU
Count Units (0=CF, 1=/L, 2=TC, 3=M3)
TU
Temperature Units (0=C, 1=F)
RZ
Returns Channel Size information.
DT
Sets Date/Time without user interaction (mm-dd-yy hh:MM:ss)
OP
Operational Status. S=Stop, R=Running, H=Hold.
CS
Set Channel Sizes (2 Channel Sizes) CS sz1 sz2
CM
Set Count Mode 0=Normal, 1=Difference
GT-521S provides serial commands for accessing stored data and settings. They are not case sensitive and are terminated by a carriage return. These commands are all available via the USB, RS-232 and RS-485 hardware interfaces. The serial settings (baud rate, parity and stop bits) in the counter must match the serial settings in the computer for proper communication regardless of the hardware interface type.
The following table lists the available commands:
7.3. Real Time Output
The real time output occurs when the unit finishes a sample. The output format is either a comma separated value (CSV) or a printer style depending on the Report Type setting (see section 5.8.1).
GT-521S Manual Rev D Page 24
7.4. Comma Separated Value (CSV)
CSV Fields
Field
Parameter
Example Value
Notes
1
Date and Time
2017-03-23 09:21:29
2
Channel 1 Size
00.3
3
Channel 1 Count (TC, /L, CF, M3)
00084140
Note 2
4
Channel 2 Size
00.5
5
Channel 2 Count (TC, /L, CF, M3)
00008680
Note 2
6
Temperature (C,F)
+022
Note 2 & Note 3
7
RH (%)
033
Note 3
8
Location
001
9
Sample Time (0-9999 seconds)
0060
10
Status Bits (see below)
000
Note 5
11
Check sum
*03414
Status Bits
Notes (for table above):
Bit
Value
Condition
0
OK (no alarms or errors)
1. CSV header included for multiple record transfers like All Data (2) or New Data (3) or last n data (4 n).
0 1 Count alarm size 1
2. Units determined by product setting.
1 2 Count alarm size 2
3. Temperature and RH will be null (,,) if Temp/RH probe is not attached
2 4 Not used
4. Alarm sizes will be null (,,) if alarms are disabled.
3 8 Not used
5. Status bits combinations are possible. For example, 17 (00010001B) = Low Battery and Size 1 alarm.
4
16
Low battery
5
32
Sensor error
6 64
Not used
7
128
Not used
CSV Header (Note 1):
Time,Size1,Count1(CF),Size2,Count2(CF),AT(C),RH(%),Location,Seconds, Status If differential Mode: Time,Size1,Count1(dCF),Size2,Count2(dCF),AT(C),RH(%),Location,Seconds, Status
CSV Example Record:
2017-03-23 09:21:29,00.3,00084140,00.5,00008680,+022,033,001,0060,000,*03414 Note: If COUNT is set to NORMAL data will be reported as cumulative counts. If COUNT is set
to DIFFERENCE data will be reported as differential counts regardless of setting when data was collected.
7.5. Printer Format
The printer output format is 9 lines by 26 characters (including T/RH if attached).
GT-521S Manual Rev D Page 25
8. Maintenance
Time Period
Item
Manual Section
Weekly
Zero Count Test
8.2
Monthly
Flow Rate Test
8.3
Yearly
Annual Calibration
8.4
Due to the nature of the instrument, there are no customer serviceable components in the GT-521S. The case of the GT-521S should never be removed or opened for any reason. Opening or removing the case of the GT-521S voids the warranty and
may result in exposure to laser radiation, which can cause eye injury.
8.1. Service Schedule
Although there are no customer serviceable components in the GT-521S, there are service items which ensure the proper operation of the instrument. The table below shows the service schedule for the GT-521S.
8.2. Zero Count Test
Air leaks or debris in the particle sensor can cause false counts which may result in significant count errors when sampling in clean environments. Perform the following zero count test weekly to ensure proper operation:
1. Attach zero count filter to the inlet nozzle (PN G3111).
2. Configure the unit as follows: Samples = SINGLE, Sample Time = 60 seconds, Volume = Total Count (TC)
3. Start and complete a sample.
4. The smallest particle size should have a count <= 1.
8.3. Flow Rate Test
The flow rate test verifies the sample flow rate is within tolerance. The reference flow meter must be non-loading because the vacuum pump can be loaded down by external restrictions. Met One Instruments sells a suitable flow meter (PN 9801). The flow rate test follows:
1. Connect a ±3% reference flow meter to the sample inlet nozzle.
2. Start a 5 minute sample.
3. The flow meter reading after ~3 minutes should be 0.1 CFM (2.831 LPM) ±5%.
4. Adjust the flow rate if the error is > ±5% (see section 5.8).
8.4. Annual Calibration
The GT-521S should be sent back to Met One Instruments yearly for calibration and inspection. The annual calibration cannot be performed by the customer because this calibration requires proprietary software and equipment and a skilled technician. Met One Instruments maintains a calibration facility for calibrating particle counters according to industry accepted methods such as ISO and NIST. The annual calibration also includes inspection and preventative maintenance to improve product reliability.
GT-521S Manual Rev D Page 26
8.5. Flash Upgrade
Symptom
Possible Cause
Solution
Display does not turn on
Low Battery Defective Battery
Charge battery Send to service center
Pump does not turn on when a sample is started
Low Battery Defective pump
Charge battery Send to service center
Keypad does not work
Loose connector Internal hardware failure
Send to service center
Sample result is lower than normal
Flow rate is low Optics may be contaminated
Perform flow rate test Send to service center
Sample result is higher than normal
Flow rate is high Air leak in unit Optics may be contaminated
Perform flow rate test Perform zero test Send to service center
Battery does not hold a charge
Defective or worn out battery Defective charger
Send to service center
GT-521S is firmware upgradeable via the serial connection using a Met One Instruments flash burn program. Binary files and the flash program must be provided by Met One Instruments.
9. Troubleshooting
The following section covers some common failure symptoms, causes and solutions. It is important to note that there are no customer serviceable components in this product. The GT-521S case should never be removed or opened for any reason. Opening or
removing the case will void the warranty and may result in exposure to laser radiation, which can cause eye injury.
GT-521S Manual Rev D Page 27
10. Specifications
Performance
Particle Size Range
0.3 µm – 10 µm, 2 count channels
Calibrated Sizes
0.3 µm, 0.5 µm, 1.0 µm, 2.0 µm, 5.0 µm and 10 µm
User Size Settings
0.1 µm steps from 0.3 µm - 2.0 µm
0.5 µm steps from 2.0 µm – 10 µm
Concentration Range
0 – 3,000,000 particles per cubic foot (105,900 particles/L)
Accuracy
± 10% to calibration aerosol
Sensitivity
0.3 µm
Flow Rate
0.1 cfm (2.83 lpm)
Sample Time
Adjustable: 1 to 9999 seconds
Hold Time
Adjustable: 0 to 9999 seconds
Electrical
Light Source
Laser Diode, 90mW, 780 nm
Battery
7.4V Li-ion battery pack.
Battery Life
24 hours Intermittent operation, 6 hours continuous operation
Battery Charge Time
Fully changed in 2.5 hours
AC Adapter/Charger
Li-ion battery charger, 100 – 240 VAC, 50/60Hz
Communications
USB, RS-232 or RS-485
Standards
Meets ISO 21501-4
Interface
Display
16 character x 4 line LCD
Keyboard
7 key membrane type
Physical
Height
6.25” (15.9 cm)
Width
4.00” (10.2 cm)
Depth
2.12” (5.4 cm)
Weight
2.00 lbs (0.91 kg)
Environmental
Operating Temperature
0º C to +50º C
Storage Temperature
-20º C to +60º C
Accessories
Supplied
Operation Manual USB Cable Comet Software AC Adapter / Battery Charger Iso-kinetic Sample Probe Carrying Case Zero Particulate Filter (PN G3111)
Optional
RH & Temperature Probe (PN G3120) Flow Meter (PN 9801) Portable Printer (PN G3115) Custom Serial Cable (PN 3228)
GT-521S Manual Rev D Page 28
11. Warranty / Service
GT-521S Manual Rev D Page 29
Loading...