Safety Notice
The contents of this manual have been checked against the hardware and software
described herein. Since deviations cannot be prevented entirely, we cannot guarantee
full agreement. However, the data in this manual is reviewed regularly and any
necessary corrections included in subsequent editions.
Faultless and safe operation of the product presupposes proper transportation, storage,
and installation as well as careful operation and maintenance. The seller of this
equipment cannot foresee all possible modes of operation in which the user may
attempt to utilize this instrumentation. The user assumes all liability associated with the
use of this instrumentation. The seller further disclaims any responsibility for
consequential damages.
Warranty
Products manufactured by Met One Instruments, Inc. are warranted against defects in
materials and workmanship for a period of (1) year from the date of shipment from the
factory. Offered products not manufactured by Met One Instruments, Inc. will be
warranted to the extent and in the manner warranted by the manufacturer of that
product.
Any product found to be defective during the warranty period will, at the expense of Met
One Instruments, Inc. be replaced or repaired and return freight prepaid. In no case
shall the liability of Met One Instruments, Inc. exceed the purchase price of the product.
This warranty may not apply to products that have been subject to misuse, negligence,
accident, acts of nature or that have or modified other than by Met One Instruments,
Inc. Consumable items such as bearings are not covered under this warranty.
Other than the warranty set forth herein, there shall be no other warranties, whether
expressed, implied or statutory, including warranties of fitness or merchantability.
Page 4 085-9800 Manual Rev A.doc
Page 4
Table of Contents
1. GENERAL INFORMATION ..................................................................................... 6
The 085 sensor is an extremely accurate microprocessor controlled relative humidity
and temperature sensor. The relative humidity sensor responds to the full range of 0 to
100% humidity. Response is linear with negligible hysteresis or temperature
dependence. The temperature sensor is a two-element composite thermistor type with
linear response over the range of –30°C to +50°C. The sensor is designed to be
mounted in a radiation shield when used outdoors. The exact model number you order
will determine the functions of your sensor. The following section describes the model
number feature assignments.
Figure 1 Model 085-35 Sensor
Page 6 085-9800 Manual Rev A.doc
Page 6
Figure 2 Model 085-6 Sensor
085-9800 Manual Rev A.doc Page 7
Page 7
1.2. 085 Options
The 085 options are defined by dash numbers as follows:
085-(Shield Option)
1.2.1. Radiation Shield Compatibility Option:
6 = Use with Model 076B or 077 radiation shield. This sensor comes with 8-inch long
pigtail wires for connection to the radiation shield junction box. Refer to the radiation
shield manual for sensor and cable connections.
35 = Use with Model 073B or 5980 radiation shield. This sensor has a circular
connector for use with Met One Instruments cable PN 2348.
2. SENSOR SITING
The EPA recommends sensor mounting in a radiation shield at a 2 meter height, ideally
over green mowed grass to minimize related terrestrial radiation errors on the
temperature and relative humidity readings. Typical installations may vary significantly
from these recommendations due to geographic limitations or specific monitoring
requirements.
Page 8 085-9800 Manual Rev A.doc
Page 8
3. INSTALLATION
If the sensor is to be mounted in a radiation shield, refer to the radiation shield manual
section for mounting details. Typical installations are shown below.
Sensors not installed in a radiation shield should be mounted in a representative
location having good airflow and shaded from sunlight or other heat radiation sources
that would affect measurement of relative humidity or temperature.
3.1. Mounting Instructions
Figure 2 Radiation Shield Installation
085-9800 Manual Rev A.doc Page 9
Page 9
3.2. Wiring Instructions
Sensor
Wire Color Description
Connection
Pin A White +10 to +18 VDC
Pin B Green Signal Ground
Pin C Yellow RH Analog Output
Pin D Black Temperature Common
Pin E Red Temperature Signal
Pin F No Connection No Connection
No Connection White/Brown Shield
Figure 3 Electrical Connector
View looking at connector pins.
Page 10 085-9800 Manual Rev A.doc
(Pins are also identified on connector).
Page 10
4. OPERATIONAL CHECK-OUT AND CALIBRATION
4.1. Relative Humidity Sensor Check-out
To verify correct wiring and test the basic sensor operation, blow on the sensor. The
moisture in your breath should cause the relative humidity reading to rise.
The relative humidity sensor has been calibrated at the factory. To check for proper
operation of the sensor it is advised that the output signal be checked against a local
weather service facility or a local relative humidity measuring device such as a
psychrometer. Due to normal atmospheric and geographical variations, the local
weather service data should be used only as a guideline. Ambient air relative humidity
can be expected to vary significantly over short distances and in brief periods of time.
4.2. Limitations of RH Measurements at Below Freezing Temperatures
The sensor’s relative humidity output is referenced to saturated water vapor pressure
above liquid water. When the air temperature is below freezing, the sensor’s maximum
theoretical measurement range is limited as follows:
Air Temperature
(Deg C)
0 100
-5 96
-10 92
-15 88
-20 84
-25 80
-30 76
-35 72
-40 68
-45 64
-50 60
Maximum RH
(%)
085-9800 Manual Rev A.doc Page 11
Page 11
4.3. Temperature Sensor
Compare readings with a precision NIST-traceable temperature sensor. Calibration
verification should be performed using a Met One Instruments Model 5472 Thermal
Mass to assure that both instruments are at the same temperature. This is an
aluminum mass with drilled recesses for the 085 sensor and a standard temperature
sensor (NIST thermometer or RTD). The mass can be used in air, or can be partially
submerged in ice or an ice bath. The 085 sensor should not be submerged.
The voltage across the sensor is linear with temperature when a precision 18700
0.1% resistor (Met One Instruments part number 805020) is connected in series with
excitation voltage. This resistor is provided as part of all Met One Instruments
translators designed for use with this sensor and is included with Met One Instruments
data loggers configured for this sensor. For applications with other equipment, contact
Met One Instruments to obtain this special resistor. Maximum recommended excitation
is 3.2V. Higher excitation voltage will cause self-heating of the thermistors, and
resultant temperature measurement errors.
Reference Voltage -30.0 °C +50.0 °C
1.000 0.855 VDC 0.311 VDC
2.000 1.710 VDC 0.622 VDC
2.500 2.137 VDC 0.778 VDC
5.000 (DO NOT USE)
Signal Voltage vs. Temperature for Various Excitation Voltages
F
E
BLACK TEMPERATURE COMMON (CONNECT TO SIGNAL GROUND)
D
RED TEMPERATURE SIGNAL OUTPUT (CONNECT TO SUPPLIED NETWORK RESISTOR)
085 CONNECTOR
083E CONNECTOR
A
B
C
WHITE + 10 TO 18 VDC
GREEN GROUND
YELLOW RH OUTPUT
REFERENCE VOLTAGE
R1
R1
231000.1%
18700 0.1%
TO DATA LOGGER SIGNAL INPUT
Data Logger Connection
Page 16 085-9800 Manual Rev A.doc
Page 16
MAINTENANCE AND TROUBLE SHOOTING
4.6. General Maintenance Schedule
6 – 12 Month Intervals:
Inspect the sensor for proper operation per Section 4.0.
12 Month Interval:
Return the sensor to Met One Instruments for calibration.
Replace the two 720050 O-Rings.
Replace the 860014 Filter Membrane.
4.7. 085 Relative Humidity Sensor Maintenance and Calibration
WARNING: The sensor can be incorrectly calibrated or permanently damaged through
improper acts. Do not attempt a repair or calibration if you are unsure of the procedure.
Do not touch the sensor element if you do not know the correct procedure.
The instrument should operate for an extended period of time with a minimum of care or
maintenance.
The sensor should be re-calibrated every 12 months.
If parts or maintenance assistance are required, contact Met One Instruments. Obtain
shipping instructions and a return authorization (RA) before returning any unit.
4.8. Sensor Maintenance
4.8.1. Sensor Element
The RH sensor element is not user-replaceable (replacement requires sensor
recalibration). If the element becomes damaged, please send the 085 sensor to Met
One Instruments for repair. Please obtain shipping instructions and a return
authorization (RA) number before returning any unit.
Met One Instruments manufactures an RH Calibration Kit part number 10233 which can
be purchased to perform a calibration after the sensor element is replaced. Please
contact the Met One Service Department to purchase the calibration kit.
085-9800 Manual Rev A.doc Page 17
Page 17
4.8.2. Spare Parts
560014
820202
Thermistor
720050
Figure 5 Spare Parts
820202 RH SENSOR ELEMENT (Requires factory calibration)
560014 FILTER, MEMBRANE
720050 O-RING (Two required)
Page 18 085-9800 Manual Rev A.doc
Page 18
5. Operational Specifications
Relative Humidity
Model Number: 085
RH Sensing Element: Thin film polymer capacitor
Range: 0 to 100% RH
Temperature Operation Range: -30° C to +50° C (-22° F to 122° F)
Response Time: 10 sec. with 2 m/s aspiration
Accuracy: ± 2.0% from 0 to100% RH
Temperature Coefficient: Compensated internally
Output: 0 to 1 VDC Standard
0 to 5.0 VDC Optional
AutoMet Auto ID: No
Temperature
Temperature Sensor: Thermistor
Temperature Range: -30° C to +50° C (-22° F to 122° F)
Accuracy ±0.15° C (0.27° F)
Output: Resistive
AutoMet Auto ID: No
General
Input Power 10 to 18 VDC @ < 5 mA
Dimensions Length: 8.5 in (21.59 cm)
Diameter:0.75 in (1.91 cm)
085-9800 Manual Rev A.doc Page 19