Metex P-R575 User Manual

OPERATING INSTRUCTION MANUAL
COMBINATIONAL PH AND ORP PROBES
N116-27 REV. 5.0
MODEL 575 / 585
Tel: (800) 742-1413
(905) 954-0841
Fax: (905) 954-0415
www.aquametrix.com
P575 AND R575 SERIES & P585 AND R585 SERIES
pH AND ORP PROBES
INSTRUCTION MANUAL
1.0 GENERAL INFORMATION
This manual covers all AquaMetrix P/R575 and P/R585 Series conventional combination pH and ORP probes.
All mounting configurations are described. Consult factory before using the sensor in extremely strong solvents such as ethylene dichloride. Before placing the sensor into operation, remove the protective plastic cap. Store this cap for future
use. NOTE: If the sensor is removed from the process for an extended period of time, thoroughly clean
the sensor, put a piece of cotton ball with few drops of water into the protective cap and replace it on the sensor. This keeps the junction from drying out which causes slow response when put back into operation or causes permanent damage to the sensor. Sensors should not be left in dry lines or empty tanks for extended periods.
Do not store the sensors in a dry or humid location. When storing, check the protective cap regularly to make sure the cotton ball remains moist. Improper storage of sensors voids the warranty.
INSTALLATION DRAWINGS: Located at end of Manual N106-80 INSTALLIATION OF 575 / 585
SERIES COMBINATIONAL PROBES SUBERMISION
N106-81 INSTALLIATION OF 575 / 585 SERIES COMBINATIONAL PROBES INLINE
P / R 575 & 585 Page 1
2.0 SPECIFICATIONS
MEASURING RANGES:
pH: 0.01 to 14.00 pH ORP: -1000 to +1000 mV Consult factory for applications below 2.5 and above
12)
WETTED MATERIALS:
575 Series CPVC body, ceramic junction, glass electrode, RTV sealant (plus platinum for ORP)
585 Series Epoxy body, CPVC compression fitting, ceramic junction, glass electrode, RTV sealant (plus platinum for ORP)
TEMPERATURE LIMITS:
-5 to 80°C (23 to 176°F) MAXIMUM PRESSURE:
100 psig at 65°C TEMPERATURE COMPENSATION:
P/R585 and P/R575: Not provided P585K and P575K: Automatic
MAXIMUM FLOW RATE:
3 meters (10 ft.) per second
STABILITY:
pH: 0.05 pH / day
Non-cumulative
ORP: 3.0 mV / day
Non-cumulative SENSITIVITY:
pH: 0.01 pH unit ORP: 1.0 mV
SENSOR CABLE: 3 meters (10 ft.) coaxial, terminated with ring, terminal on centre wire, other wires are tinned.