Metex 3200U User Manual

OPERATING INSTRUCTION MANUAL
MODEL 3200U
ULTRASONIC LEVEL & FLOW MONITOR
AquaMetrix Inc.
22-121 Granton Drive Richmond Hill, ON
Canada, L4B 3N4
Tel: (800) 742-1413
Fax: (905) 763-9480
www.aquametrix.com
AquaMetrix
Bench Test ·······················································3 Connections·······················································3 Keypad System ····················································4 Menu - Flow Chart ·················································4 Run·····························································6 Totalizer (Flow Mode)··············································6 Password ························································6 Units / Mode ······················································7 Flume Selection ···················································9 Calibration - for Level ·············································10 Calibration - for Open Channel Flow ··································11 4-20mA Current Loop Offset ········································12 Rejection Time ···················································12 Relay Parameters ·················································13 Special Functions ·················································15 Sensor Location - Tank Level········································17 Sensor Mounting/Location - Open Channel ·····························20 Enclosure Installation ··············································23 Error/Warning Messages ···········································24 Field Troubleshooting··············································25 Installation Considerations In Noisy Environments ·······················27 Customer Service ·················································29 Instrument Return Procedure ········································29 Appendix A - Options··············································30 Data Logger ·····················································32 RS232C Serial Output ·············································38 Appendix B - Applications Background································41 Conversion Guide ·················································42 Specifications ····················································43 Calibration Worksheet - Level Mode ··································46 Calibration Worksheet - Flow Mode ··································47
Ultrason
INDEX
IMPORTANT NOTE: This instrument is manufactured and calibrated to meet product specifications. Please read this manual carefully before installation and operation. Any unauthorized repairs or modifications may result in a suspension of the warranty.
Page 2
AquaMetrix
QUICK BENCH TEST:
Ultrason
Connect Sensor to the
T.DUCER terminals as shown below, then apply Power. When properly connected
a soft clicking can be heard from the sensor and figures will show on the large LCD display. Test operation of the Ultrason by holding the sensor steadily and aiming at a flat, stable target 12 to 28" (305 to 711 mm) away from the end of the sensor. Allow a few seconds for the Ultrason to lock onto the target before displaying its distance. The Ultrason will now display distance in ft or cm (factory calibration).
CONNECTIONS:
POWER INPUT: The standard model requires AC power input between 100-130VAC 50/60Hz (2 amp fuse is recommended). No adjustments are necessary for any voltage in this range. Optional 230VAC requires power input between 200-260VAC 50/60Hz. (See OPTIONS section of this manual for connection of optional 12VDC or 24VDC power input).
IMPORTANT NOTE
: To comply with CSA/NRTL standards, AC power input and relay connection
wires must have conduit entry to the instrument enclosure.
AMqua etrix
C
N.O
RELAY 1
100-130VAC (50/60 Hz)
OPTIONAL: 200-260VAC (50/60 Hz)
N.C
N.O
RELAY 2
C
C
N.O
N.C
N.C
RELAY 3
G
N
L
IMPORTANT: MUST CONNECT TO A GOOD GROUND (<1 Ohm) WITH 12 AWG CONDUCTOR
4\20
AC
T.DUCER
GND
RX
PU
TX
RS 232
*
N.O
RELAY 4
OPTIONAL
C
N.C
N.O
RELAY 5
C
*
C
N.O
N.C
GROUND STUD
IMPORTANT ATTACH CABLE SHIELDS (4-20mA, RS232) THIS END ONLY
N.C
RELAY 6
:
ALUMINUM CHASSIS
Page 3
AquaMetrix
KEYPAD SYSTEM
The Ultrason has a simple 3-button calibration system. Operating and calibration modes are shown on the 16-digit alphanumeric display. The keypad is used to move around the menu to calibrate the Ultrason, and to view operating mode and functions. A beep is sounded as each key is pressed. If the keypad is not used for 2 minutes, the Ultrason will automatically go to
RUN mode (scrolling display). Use the
keypad to explore the Menu and become familiar with its features.
MENU - FLOW CHART
Ultrason
The following diagram shows part of the Ultrason Menu system. Arrows show the three directions to leave a box. Pressing a corresponding keypad arrow will move to the next box in the direction shown. Move the cursor (or underline) under numerals to increase or decrease the number with the ÈÇ keys.
At the bottom of each Menu column is a (even through power failure), move the cursor under pressed with the cursor under the Menu column.
Store? no changes will be stored and the system will return to the top of
Store? Yes box. To store the calibration values permanently
Yes and press the È or Ç key. If the È key is
Page 4
V 2.51
O
SLT32
SPECIAL FUNCTION
1234567890123456
Tag 01
VYiew Codes? es
23:39:51
Time
Date May 31/1999
30 s
LOE Time
Yes
Tot?
Reset
Cº
20.3
Temp
Cº
20.0
Min Temp
Cº
23.0
Max Temp
DISPLAY FCºº
0.00%
4.000
20.000
Simul
20mA ADJ
4mA ADJ
ISB?: oYesN
Yes
Store?
New Password: 00
Com 24 48 96 192
N MENU
RELAY PARAMETERS
CALIBRATION
Flow
Level
Temp
Pulse
R1 Function Off
1.333 ft
MinRg
HiAlm
LoAlm
R1 Mode Pump
R1on 0.00 C°
R1on 0.00 ft3
32.00 ft
MaxRg
R1 on 0.00ft3/s
R1 OFF 0.00 ft
R1 on 0.00 ft
° C
OFF
R1 0.00
0.00 %
100.00 %
20mA @
4mA @
Hold
On
R1 LOE Off
%
08 s
20
RejTime
Damping
Flow
Level
Temp
Pulse
R2 Function Off
Yes
Store?
HiAlm
LoAlm
R1 Mode Pump
R1on 0.00 C°
R1on 0.00 ft3
R1 on 0.00ft3/s
R1 OFF 0.00 ft
R1 on 0.00 ft
° C
OFF
R1 0.00
Hold
On
R1 LOE Off
Flow
Level
Temp
Pulse
Function Off
R3
HiAlm
LoAlm
R1 Mode Pump
R1on 0.00 C°
R1on 0.00 ft3
R1 on 0.00ft3/s
R1 OFF 0.00 ft
R1 on 0.00 ft
° C
OFF
R1 0.00
Hold
On
R1 LOE Off
(123456)
ALT
Yes
Store?
MIN
SENSOR
RANGE
OPTIONAL FEATURES
MAX LEVEL
MAX
RANGE
ZERO LEVEL
Ultrason - CALIBRATI
FLOW MODE ONLY
FLUME SELECTION
UNITS / MODE
PASSWORD: 00
K=
Custom
More
Sizes
Parshall
Sizes
V-Notch
cm %
Vol Hrt Flow
m
in
Level
ft
Range
n=
Yes
Store?
L
m3
VSG USMG
IG IMG
ft3
L
USG IG m3
Mx Vol xxxx.x ft3
ft3
123456
LoAlm
5.0 ft3/s
HiAlm
At:
Event
1
Time
Trend
Formatted
StartJan 01/2000
Flow
MAX
AVERAGE
TOTAL
Daily
Daily
Daily
TIME
Flow
MAX
Flow
MIN
Daily
StartJan 01/2000
03:02:16
Start
TIME
Flow
MIN
Log Site ID
Setup
s min hr d
Yes
Store?
DATA LOGGING
24 HR LOG
ONLY IF ENABLED
SHOWS 24 Hr FORMAT
STOP
RUN
1
Session No
->
01/2000
Jan
03:02:16
Start
Interval: 30 Sec
Isnterval: 24 Hr
Interval: 12 Hrs
->
31/1999
Days
Dec
255
5 Sec
2 Sec
Interval: 10 Sec
Interval:
Interval:
s Hr
Interval: 8
-
data
more
no
-
1 Sec
5 Min
Interval:
Interval: 30 Min
Interval: 10 Min
Interval:
s Hr
Interval: 4
2 Min
1 Min
Interval:
Interval:
Hr
Isnterval: 1
WrapAround? Yes
Yes
Reset Log?
26640 Hrs Left
Yes
*** STORING ***
Store?
RUN
FLOW MODE ONLY
C
RELAYS:
Tot:
E
AquaMetrix
o
RUN
Ultrason
A scrolling display shows the units selected from the
UNITS/MODE column, the mode of operation (Range, Level or Flow), the full scale value for the large numeric display, Totalizer
value (in
FLOW mode), and air temperature at the sensor location.
RUN
EC
FLOW MODE ONLY
Tot:
RELAYS:123456
PASSWORD: 00
ECHO CONFIDENCE
EC
Echo Strength is automatically adjusted by the Ultrason in response to operating conditions in the applications.
From RUN use È to get to the Echo Confidence display,
EC.
Minimum
Marker
Current
Echo Strength
Arr
w
TOTALIZER
The Totalizer display will only be enabled when the Ultrason is calibrated in
FLOW mode. From RUN
use È or Ç keys to display Tot: value. The Totalizer value is updated every 2 seconds with flow volume > 1 litre (0.264 USG). The display will show up to 10 digits and then restart at 0 automatically.
Units: ft³, USG, IG, L
T: 9999999999
TOT: 0
Units = m³
Press È or Ç keys to return to
RUN.
The totalizer can be reset to zero by going Æ to
T: 0283168470
SPECIAL FUNCTION and È to Reset Tot?.
TOT: 0
RELAY STATUS DISPLAY
Press È from Engerized relays will be displayed in reverse type
Tot: to RELAYS 1,2,3.
RELAYS:
PASSWORD
The password (a number from
CALIBRATION menu.
From
RUN (scrolling display) press Æ to PASSWORD.
00 to 99) prevents unauthorized access to the
Press Æ to place the cursor under the digits and È or Ç to change the number and then Æ to proceed to
CALIBRATION.
PASSWORD: 00
PASSWORD: 00
PASSWORD: 01
Factory default Password is and È to New Password.
00. A new password can be stored by going Æ to SPECIAL FUNCTION
Page 6
AquaMetrix
UNITS / MODE
Ultrason
Press Æ to
DATA LOGGING
Range Level
UNITS/MODE, then È . Press Æ to move the cursor under the required measurement units:
ft - feet in - inches
m - meters
cm - centimeters
% - percent
UNITS / MODE
ft in m cm %
Vol Hrt Flow
ft3 USG IG m3 L
Mx Vol xxxx.x ft3
Store? Yes
ft3 USG USMG
IG IMG m3 L
s min hr d
FLUME SELECTION
Press È to
Range mode displays distance from the sensor to the target or liquid surface like a tape measure.
Range Level > Vol Hrt Flow.UseÆ to select the operating mode of the Ultrason.
Range mode is useful to measure the exact distance from the sensor to the zero level during calibration, or to monitor “outage” or free space in a tank.
Level mode can be used to measure tank level in linear units, or “Head” (height of flow) in an open
channel for comparison with a flume manufacturer’s flow tables.
Vol Volume mode displays tank inventory in engineering units like gallons or liters. Hrt Horizontal Round Tank mode sets the Ultrason to calculate and display volume units in a
horizontal round tank.
Flow mode is for open channel flow through a flume or weir.
From
Vol Hrt or Flow press È to make your selection. (Range or Level mode will bypass the
Volume units selection menu).
Vol Hrt or Flow modes will give you the additional choice of
volumetric units:
ft3 - cubic feet
USG - US gallons
USMG - US million gallons (FLOW only)
IG - Imperial gallons
IMG - Imperial million gallons (FLOW only)
m3 - cubic meters
L - liters
Press Æ to position the cursor underneath your selection and press È to make your selection.
Page 7
AquaMetrix
Vol or Hrt modes will also prompt you to enter a maximum volume. Press È from MxVol to position
Ultrason
the cursor under the first digit. Use È or Ç to change the digits or decimal point to enter the actual maximum volume you will be measuring. Press Æ to return to
MxVol and È .
You can enter values up to 6 digits (e.g. 150,000 liters). When the measured tank volume exceeds 4 digits, the Ultrason will automatically display a 'multiplier' on the lower alphanumeric display.
AMqua etrix
= 150,000 Liters
UNITS 1000L
FLOW mode will offer the additional selection of time units:
s - seconds
min - minutes
hr - hours
d - day
Note: The Ultrason will display value of 9,999,999. Use to
Store and Æ to Yes. Then press È or Ç to store all your UNITS/MODE selections.
US MGD, Imp MGD or m3/d so that Units will be 9,999,999 or less. Press È
Er:ILLEGAL UNITS if your choice of Units exceeds a maximum
Page 8
AquaMetrix
FLUME SELECTION
The
FLUME SELECTION menu will only appear in FLOW mode.
Use È and Æ to select the correct Flume or Weir.
Ultrason
UNITS/MODE
Use È and Æ again to select the Size then È to Press È or Ç to store flume selection and get to
FLOW MODE ONLY
FLUME SELECTION
V-Notch More> P. Bowlus More> L.Lagco More> Rect. Weir >
22.5 30 45 60 09
Parshall More>
2" 3" 6" 9" >
12" 18" 24" 36">
4" 6" 8" 10">
12" 15" 18">
21" 24" 42">
6" 8" 10" 12" >
18" 24" 30" 36">
Store? Yes
Size 1.000 ft
Store? and Æ to Yes.
CALIBRATION mode.
Rect.Weir w/EC >
Size 1.000 ft
Custom More>
K = 1.0000
N = 1.0000
CUSTOM FLUMES: Experimental or empirical data for any flume or weir can be reduced to an equation of the form
Q is flow volume K is as scaling factor which includes flume size and units of flow measurement n describes the flume non-linearity H is the level being measured
Use the AquaMetrix Utility program Select
n: enter an 'n' constant.
CUSTOM for flumes or weirs not listed in the Menu. At the K: prompt enter a 'K' constant. At the
Q=KHn where:
FIND K&n.exe to determine K and n constants from entered data.
CALIBRATION
Page 9
AquaMetrix
Ultrason
CALIBRATION
- for Level/Inventory Applications
1. Before starting the calibration determine:
a) MAX RANGE = ________________
(Maximum range = distance from Sensor to Zero level)
b) MAX LEVEL = ________________
(Maximum level of product being measured)
c) MIN RANGE = ________________
(Distance from sensor to Max Level) Minimum range = MAX RANGE - MAX LEVEL (must be at least 8" / 203 mm depending on sensor
model)
2. Check the maximum range with the sensor installed by:
a) Use Æ to get to
Range, then Store? Æ Yes
UNITS/MODE then È to get to
(NOTE: liquid must be at zero level)
or
b) Carefully measure distance from sensor to zero
level with tape measure.
TANK INVENTORY LEVEL MODE
0%
0%
MIN RANGE
MIN RANGE
100%
TANK INVENTORY RANGE MODE
SENSOR
MAX RANGE
SENSOR
MAX RANGE
MAX LEVEL
MAX VOLUME
ZERO LEVEL
MAX LEVEL
MINIMUM RANGE:
CALIBRATION
At
- Press È to get to MinRg
- Press Æ to move the cursor under the digits
-UseÈ or Ç to set the minimum range
MinRg must be ³12" (30.5 cm) for US2-1A1A
Note: sensors and ³ 16" (40.6 cm) for US2-2A1B sensors except
in
Hrt mode. For correct tank volume calculation in Hrt
mode MinRg must be the actual distance from the end of the sensor to the top of the tank (any value ³0). In
Level
or Range modes the Ultrason will only accept MinRg values ³ 8" (20.3 cm).
- Press Æ to return to
MinRg
MAXIMUM RANGE:
Press È to get to
MaxRg
Repeat as for MinRg
100%
TANK INVENTORY HORIZONTAL ROUND TANK (HRT) MODE
MIN RANGE
MAX
VOLUME
MAX RANGE
ZERO LEVEL
Page 10
AquaMetrix
CALIBRATION
- for Open Channel Flow
Ultrason
FLUME SELECTION
CALIBRATION
MinRg 1.333 ft
MaxRg 32.00 ft
4mA @ 0.00 %
20mA @ 100.00 %
Damping 20 %
RejTime 08 s
Store? Yes
1. Before starting the calibration determine:
a) MAX RANGE = ____________________
(Maximum range = distance from the Sensor to Zero level point)
b) MAX LEVEL = ____________________
(Maximum level of flow through flume or weir)
c) MIN RANGE = __________________
(Distance from sensor to Max Level) Minimum range = MAX RANGE - MAX LEVEL (must be at least 8" / 20.3 cm depending on sensor
model).
RELAY PARAMETERS
PALMER BOWLUS
SENSOR
MAX LEVEL
MIN RANGE
MAX RANGE
ZERO LEVEL
2. Check
a) Use Æ to get to
the maximum range with the sensor installed by:
UNITS/MODE then È to get to
Range, then Store? Yes
(NOTE: flow must be at zero flow)
or
b) Carefully measure distance from sensor to zero level
with tape measure.
Page 11
V-NOTCH
SENSOR
MIN RANGE
MAX LEVEL
MAX RANGE
ZERO LEVEL
AquaMetrix
4-20mA CURRENT LOOP OFFSET
Some applications may require the 4-20mA output to be offset so that 4mA or 20mA corresponds to a Level, Range or Flow other than Zero and Full Scale.
4mA at - use Ç and È to set % output for 4mA. It is adjustable from -5% (3.8mA) up to 15% lower
than the 20mA setting. Adjustment resolution is 0.05% (0.01mA).
20mA at - use Ç and È to set % output for 20mA (down to 15% greater than the 4mA setting and up
to 300%). Adjustment resolution is 0.05% (0.01mA).
DAMPING
Minimum damping allows fast response to level changes. Increasing damping slows the Ultrason's response to level changes and is ideal to smooth the display and outputs in turbulent conditions. Damping value is shown in percent (0-99%). Some experimentation may be required to select the optimum damping value. A value of 20% is recommended for most applications. For fast level changes (up to ½ inch/sec - 13 mm/sec), a
Damping value of 1% is recommended. Maximum is 99%.
Ultrason
If the Ultrason is unable to obtain repeated echoes the Damping setting will be automatically set to 1% by the instrument.
REJECTION TIME (RejTime)
Rejection Time adjustment is a powerful Ultrason feature to suppress false echoes and the effects of transient targets such as agitators, splashing, or turbulence. Rejection Time is shown in seconds. The time value represents the number of seconds the Ultrason must receive repeated echoes from a new target before indicating its level on the display and outputs. The factory default is 8 seconds and should be ideal for most applications. Decreasing this value allows the Ultrason to respond more quickly to fast level changes, and increasing the time value allows the Ultrason to reject false echoes from slow sweeping agitators, tank filling streams, splashing, severe turbulence and steam. Minimum setting is 1 second (for fast level changes up to 1/2 inch/sec - 13 mm/sec). Maximum setting is 85 seconds. If the Ultrason is unable to obtain repeated echoes the by the instrument.
RejTime setting will be automatically set to 1 second
Page 12
AquaMetrix
RELAY PARAMETERS
Ultrason
Each relay can be individually calibrated. Press È to ÆTemp Æ Level Æ Flow.
Off
Pulse
CALIBRATION
mode means the relay will not be energized
Appears in Flow mode only. Press È and set digits to the flow volume increment required
RELAY PARAMETERS
R1on 0.00 ft3
R2 Function Off
Pulse
Pulse
TempR1 Function Off
R1on 0.00 C°
R1 0.00
OFF
Temp
between relay pulses. Use this feature for remote samplers, chlorinators or totalizers. Minimum time between pulses is 1 second and pulse duration is 350 milliseconds.
R Function and Æ to select: Off Æ Pulse
SPECIAL FUNCTION
Level Flow
R1 Mode Pump LoAlm
° C
R1 on 0.00 ft R1 on 0.00ft3/s
R1 OFF 0.00 ft
R1 LOE Off
Level Flow
On
HiAlm
Hold
Temp References temperature reading from the Ultrason's ultrasonic sensor (air temperature at
the sensor location). Press È and Æ and set the relay On and Off set points.
Level
Range Volume
Appears in Level, Range, Volume or HRT modes only. Press È to R Function and Æ to select
Pump mode - press È and Æ and set relay On and Off set points.
LoAlm mode - press È and set relay On point. Relay will be On with levels below
Pump, LoAlm (low alarm) or HiAlm (high alarm).
this point.
HiAlm mode - press È and set relay On point. Relay will be On with levels above
this point.
LOE - press Æ to select Off ÆOn Æ Hold. In a loss of echo condition, the
(Loss Of Echo) relay will turn Off (de-energize), On (energize) or Hold its current state.
Each relay can be individually configured. Press È.
To calibrate a relay for echo loss alarm only, set the relay On and Off set points to exactly the same value, and then set
LOE to On mode. The relay will now energize only when an
echo loss condition occurs.
Page 13
AquaMetrix
Ultrason
Flow
Appears in Flow mode only. Press È to R Function and Æ to select Pump, LoAlm (low alarm) or HiAlm (high alarm).
Pump mode - press È and Æ and set relay On and Off set point.
LoAlm mode - press È and set relay On point. Relay will be energized with levels
below the On point and de-energized with levels above the On point.
HiAlm mode - press È and set relay On point. Relay will be energized with levels
above the On point and de-energized with levels below the On point.
LOE - press Æ to select Off ÆOn Æ Hold. In a loss of echo condition, the
(Loss Of Echo) relay will turn Off (de-energize), On (energize) or Hold its current state.
Each relay can be individually configured. Press È.
ALT permits relays calibrated in Pump mode to be
alternated automatically. Pump alternation allows even pump wear in pumping station applications.
ALT (-None-) indicates that no relays have
been programmed in Pump mode.
ALT 123 - Press Æ to place cursor (underline)
under the relay number to be programmed for alternation and press Ç or È to highlight. Repeat until all relays to be alternated are highlighted.
PUMP ALTERNATION (R1, R2, R3)
LAG
PUMP 1
2
3
1
1st CYCLE
2nd CYCLE
3rd CYCLE
...ETC
LEAD PUMP
1
2
3
LAG
PUMP 2
3
1
2
Press È to
Store? Yes and Æ to Yes. Press Ç or È to save your Relay settings.
Page 14
AquaMetrix
SPECIAL FUNCTIONS
SLT32 V2.26 shows software version installed
Ultrason
View Codes? Yes
Select Yes and then È to display an instrument performance code. Faults detected by the instrument are stored in the code and will be requested by AquaMetrix if you call for service or assistance.
Tag Enter Instrument Tag number (0-9999)
Date MM DD/YYYY
(with optional Data Logger) Press Æand Ç or È to change Month, Day Year.
Time HH:MM:SS
(with optional Data Logger) Press Æ and Ç or È to change Hours, Minutes, Seconds
LOE TIME
Press Æ and Ç or È to change the number of seconds without receiving an echo before the Ultrason displays
ECHO LOSS, and Control
relays change state as calibrated under Relay Parameters.
Factory default is 30 seconds and is recommended for most applications, Minimum is 10 seconds and maximum is 60 seconds.
SPECIAL FUNCTION
SLT32 V 2.26
VYiew Codes? es
1234567890123456
Tag 01
Date May 31/1999
Time 23:39:51
LOE Time 30 s
Reset Tot? Yes
Temp 20.3 Cº
Min Temp 20.0 Cº
Max Temp 23.0 Cº
DISPLAY FCºº
ISB?: N Yeso
Simul 0.00%
Reset Tot? Yes (FLOW mode only) Position the cursor under
Yes to reset the totalizer. Store this selection
at the
Store? Yes prompt.
Temp 25.0°C Indicates current temperature at the sensor
head. Press Æ to
Temp
to display the minimum and maximum
Min Temp and Æ to Max
temperatures the Ultrason has sensed. Press Æ to Display °
C °F. Position the cursor under °C
4mA ADJ 4.000
20mA ADJ 20.000
New Password: 00
Com 24 48 96 192
Store? Yes
for celcuis display or °F for Farenheight display. Press È to return to
ISB?: No Yes Select Yes only when the instrument is equipped with an optional Intrinsic
Temp.
Safety Barrier for sensor mounting in hazardous locations. This setting compensates for the electrical characteristics of an Intrinsic Safety Barrier.
Page 15
Loading...
+ 33 hidden pages