Metertech IDIP1 User Manual

Page 1
QUICK
START GUIDE
Page 2
2 CONTENTS
Welcom e iDip® Overview
Let’s
Select,
Data Management
Tips ...................................................
Troubleshooting ........................................
About
Reorder
This guide shows you the technical details on your iDip®, helps you and gets you started with tips for the tests you will use
You r eXa ct iDip® comes Cleaning Quick Start 6 (DPD- 3), pH, and Total
You will need Four (4) AAA batteries, a Phillips head screwdriver, compatible smart device
Compatible smart Apple iPad Apple iPhone Apple iPod touch
We are constantly adding new devices, check out our full list at sensafe.com/idip.
................................................ .2
........................................... .3
Get Started
Install Batteries.......................................... .4
Download Home Screen Test Result Screen
Select Customer........................................ .6
Tur n o n
Bluetooth ............................................. .7
Select Test Fill Cell
Zero Meter ............................................ .8
Tes t
Standard Strip Method................................... .9
App .......................................... .4
........................................... .5
Fill, Dip,
........................................ .5
Read
Meter .......................................... .6
............................................ .7
............................................... .8
Methods .......................................... .8
Standard Liquid Method ................................
R
esults/Notes ......................................... . 11
History ..............................................
Emailing
About eXact iDip® P
About Bluetooth® ......................................
About Built-in Sample Cell ...............................
Cleaning the Cell Compliance Testing Warrant y (2 years)
Technical Support .....................................
Results.......................................
hotometer ............................
......................................
....................................
.....................................
...............................................
WELCOME
WELCOME TO YOUR NEW
Brush
Guide
Strips each of: Free Chlorine (DPD-1), Combined/Total
devices:
®
(Mini, Mini Retina,
®
(4s,
5, 5c, 5s)
® (5th
EXACT IDIP®.
with:
Alkalinity
to
start testing your
gen)
rd
3
gen,
LET
US SHOW YOU AROUND.
daily.
Chlorine
water.
th
4
gen,
Air)
set it
and a
.10
.12 .12 .13 .13
.14 .14 .14 .14 .15 .15 .15 .16
up,
Page 3
YOUR NEW SPAS, PONDS & AQUARIUMS, FOOD & BEVERAGE PROCESS WATER, NVIRONMENTAL WATERS, AND MORE
EXACT IDIP® IS IDEAL FOR TESTING DRINKING WATER, POOLS &
IDIP® OVERVIEW 3
SAMPLE CELL
Built-in plastic 4ml cell collecting your water s ample
for
BLUETOOTH CONNECTED
LCD DISPLAY
BA
BASE
Install/replace batteries (IP67 rated
LOW
TTERY
RESULT
UNITS
waterproof)
here
CELL
COVER
C
overs the cell
mixing
and
light
situations
for
bright
SCREW
Unscrew remove base
to
LCD
DISPLAY
Displays results and test
name
READ
BUTTON
Starts test
timer
ZERO/ON
BUTTON
Turns the meter on and creates a baseline your water
testing
for
Page 4
4 LET’S GET STARTED
INSTALL “AAA” BATTERIES (NOT INCLUDED)
1. Use a Phillips head screwdriver to remove the screw from the base of your iDip®.
2. Remove the base.
3. Install 4 new AAA batteries as illustrated inside your iDip®. We recommend using high quality batteries.
4. Replace the base firmly with pressure while tightening the
screw. The meter will turn on automatically.
DOWNLOAD THE
Using your Smart Device, download the eXact iDip the eXact iDip to do is download the latest update to have the most current version with the latest tests and features available. We are constantly improving and welcome your suggestions to help make our product even better. Please be sure to visit www.sensafe.com/idip or send feedback to [email protected].
APP
®
app.
®
application is the brain for this system, all you have
Because
SCAN FOR APP
DOWNLOAD
Page 5
SETTINGS:
Opens menu easy access app features
to all
for
LET’S GET STARTED
HOME SCREEN
5
History:
Accesses
all previously saved results which can be sorted, edited, and
emailed
CUSTOMERS:
Attaches
results
to
people
and/or locations
from your smart device contacts
CALENDAR:
Displays schedule/ appointments
STORE:
Opens stor e unlock additional tests
TEST:
Initiates water testing
BLUETOOTH
TEST RESULT SCREEN
DEVICE
TESTING
PARAMETER
PARAMETER UNITS
ALL RESULTS ARE ATTACHED TO A CONTACT AND LOCATION. ENSURE YOUR CONTACT
INFORMATION (INCLUDING ADDRESS) IS ENTERED INTO YOUR DEVICE’S CONTACT LIST.
RESULTS:
Accesses
temporary results that have been saved
to
history
not
BOTTLE LABEL
ABBREVIATION
TEST RESULT
your
to
Page 6
6
1
SELECT,FILL,DIp,RE
a. Select ‘Customers’ from the ‘Home’ screen. b. Tap ‘Add Customers From c. Select a contact from your
Contacts.’
list.
AD
2 3
a
Power on the
Tap
‘ ’
and select ‘Bluetooth Test’ on the slide
screen.
b
eXact iDip® hand-held
c
photometer.
out
Page 7
Select your iDip from the bottom has connected and tap ‘OK’. *
*
Refer
to
also be the serial number for the bluetooth
SELECT,FILL,DIp,RE
the back
of
your iDip® to
determine the serial number for your device. This
connection.
Tap
‘
sELECT
and app will both display the test being
Test’ at the top (e.g. Alkalinity, Total). The iDip
of
the screen. Verify
run.
AD
7
it
will
4
5
Page 8
8
6
SELECT,FILL,DIp,RE
Rinse the cell 3 times with the water sample tested and
7
Press and the iDip meter is ready to test. Cell Cover on CELL
Standard Strip Method Tests
•
Alkalinity,
Total
•
Bromide (a s
• Bromin e
•
Calcium (as
•
Chloride (as
•
Chlorine
•
Chlorine,
•
Chlorine,
•
Chlorine, Total - High (120
•
Chromium (VI) (240
•
Copper (120
•
Hardness, Total
•
Hardness, Total
•
Hydrogen Peroxide - Low (120
Standard Liquid Method Tests
•
Cyanuric Acid (5 drops / 60
For non-standard www.sensafe.com/idip/.
NaBr)
CaCO3)
NaCl)
Dioxide
Free Total
sec)
sec)
- Low
- High
test methods and detailed
fiLL
to
the top
to
begin test.
®
display reads 0
For Standard Liq uid Method
before zeroing.
(count-up if required):
•
•
• Iodine
•
•
•
• Ozone
sec)
sec) •
•
• Permanganate
• pH
•
• Sulfate
• Sulfide
sec)
(# of drops / count-up if required):
Hydrogen Peroxide - Mid (100 Hydrogen Peroxide
Magnesium Nitrate (as NO3) (600 Nitrite (as NO2) (360
Peracetic
Phosphate (120
Metals (2 drops / 120
Hardness
Acid
test instructions visit
AD
PPM
- High
sec)
sec)
sec)
sec)
to be
indicating the
, place
sec)
Page 9
a
Remove one Alkalinity) and set in a dry, convenient place.
PRESS
simultaneously pads in the sample then use a gentle constant back and forth motion (2 strokes per second) until th timer displays Wait for count-up time if required.
READ
and
in tests, repeat step s 5-9. To save your results and make available emailing, continue on page
SELECT,FILL,DIp,RE
STANDARD STRIp METhOD
eXact Strip Micro (e.g. Total
to initiate a 20 second countdown and
DIP
the test strip by submerging all
“1”.
Remove and discard the s
result displayed on the iDip the app. To run add
itiona
11.
®
l
for
10
8a
e
tri
p.
AD
9
a
9
Page 10
10
8b
9b
10
SELECT,FILL,DIp,RE
STANDARD LIQuID METhOD
Shake the bottle of Acid) and add drops according to the list on page Keep reagent
Place th e Cell C o v e r o nto t h e CE L L . Press and a 20 second countdown begins. Turn the meter upside-down repetitively during the 20 seconds. When the timer displays
®
iDip
on a flat surface. Wait for count-up ti
b
eXact Reagent
bottle vertical while adding drops.
READ
result displayed on the iDip
and
in
the app. To run add tests, repeat step s 5-10. To save your results and make available emailing, continue on next page.
AD
(e.g. Cyanuric
“1”,
place the
me.
itiona
l
®
for
8.
Page 11
When all tests have been performed, tap ‘Results’ at the of
the screen. To notes tap the test
result.
bottom
add
desired
Type notes in the ‘Notes’ box, which are automatically saved
Go back to ‘Results’ screen and tap ‘Save’ to store into ‘History’
DATA MANAGEMENT
.
.
1
Page 12
12 DATA MANAGEMENT
In
History, you select, and email results.
To email, you either tap an result, or use ‘Select’ button access multiple data
points.
can
edit,
your
can
individual
the
to
Press the blue envelope icon the top
if
multiple tests are selected. Additional notes can be added
the email body. Add recipients to your e-mail and tap send to complete.
if
a single result is selected. Tap ‘Email’ at
to
Page 13
y
)
•
The meter has a 3 minute automatic shutdown timer by default. To change this, go
Settings in the slide out menu.
•
Before testing, rinse the meter sample cell with the sample water
•
Always fill the cell
•
Test immediately after filling the cell with the water
•
To obtain optimal accuracy when testing outdoors (sunlight), use the Cell Cover when
zeroing and reading the sample.
•
Due
to noticeably discarded. Using these strips may give unreliable
•
Meter is not compatible for use with powder pillows, tablets, and
manufacturers.
•
Dip strip for entire countdown.
•
Each
eXact Strip Micro is valid for ONLY one test. Discard
•
Dry the outside of the meter before storage.
•
Remove batteries before storing for
•
Store the meter and test materials out of direct sunlight and away from chemical
storage areas.
•
Minimize exposure of meter and test reagents to heat above 90°F
•
When installing batteries, make sure that the O-ring is still attached
before tightening.
•
Cleaning the cell with water and brush after each test is recommended for best accuracy and prior to storage of unit. Use Distilled White Vinegar to clean Phosphate or Iron to remove reagent deposits from
Issue
No
response
from
iDip
Dim screen on “Er1” on iDip LCD
“Er2” on iDip LCD No connection
“LO” on iDip LCD while
zeroing
“HI”
on iDip LCD
while
reading
“LO” on iDip LCD while
reading
“AbS” on iDip LCD Start-up
to
capacity
the strip slitting process, you may find one or two strips that are
smaller or larger in width than the normal strips in the bottle. These should be
TROubLEShOOTING
Cause
Low Chip
iDip
Low No result sent
iDip from
Lost connection Low
Dirty Cloudysample Bad LED
Result above detection level
Result detection level
(4ml).
batter failure
battery
battery
cell
below
TIpS
results.
prolonged periods.
cell wall.
to
app
to app
toapp
screen
3 times.
sample.
liquids from other
strip after use.
the cell after
(32°C).
to the screw
testing for
Solution
Replace
batteries
Contact
ITS.
Replace
batteries
Disconnect/reconnect iDip (see pg.
Connect
to
Reconnect Replace
batteries Clean cell(seepg.15 Dilute sample Contact
ITS. Rerun test
Rerun test
Continue with
7)
iDip (see pg. toiDip
oruse sieve
to
verify
to
verify
testing
13
to
to
7)
result
result
Page 14
14 AbOuT
ACCURACY OF THE EXACT IDIP
Combined with your smart device, the eXact iDip® is designed to test your water for multiple water quality paramete rs. Download the free eXact iDip® app and sync to your smart device running Bluetooth® Smart Technology.
All tests have been calibrated using ce rtified reference standards and analytical spectrophotometric methods. The algorithms in the app reflect the best correlation of the eXact iDip® against the AWWA, USEPA, DIN and ISO reference test methods for chlorine. Independent studies verify the eXact iDip® repeatedly agrees with a USEPA compliant reference method greater than 99% (R2= 0.99948, 0-5.00 ppm). The eXact iDip® has been factory calibrated and will stay valid because of the exceptional quality. We are so confident in the eXact iDip®, we offer an industry leading 2-year warranty.
We built the eXact iDip® to be easy, accurate and enviro nmentally friendly. We have achieved this by utilizing our patented eXact Strip Micro Technology, which uses at lea st 60% less water and chemistry than alternative methods. Instead of using a 10mL water sample, eXact Strip Micro uses a 4mL water sample. The accuracy of the meter is maintained by designing the sample cell with a 11mm path length.
®
BLUETOOTH TECHNOLOGY
Bluetooth® is a low-power wireless networking standard which uses short radio waves to allow electronic devices to communicate with each other wirelessly. The eXact iDip® comes standard with the latest Bluetooth® 4.0 technology (www.bluetooth.com/Pages/Bluetooth­Smart. aspx), a class 2 device with a wireless working distance of up to 30 feet (10 meters) and a 2.1 Mbps data transfer rate. This allows a seamless transfer of data between a smart device and the eXact iDip®.
ABOUT THE BUILT IN SAMPLE CELL
The built-in sample cell is made of transparent plastic; the sturdy cell design will last for over 20,000 readings. Scratches on the cell will not compromise the accuracy of your results because of the cell’s fixed position.
CLEANING THE CELL
Fill the cell with water and use the enclosed cleaning bru sh to scrub the CELL thoroughly. Rinse the CEL L. You can use dish soap or distilled white vinegar for cleaning the CELL if needed. NEVER use solvents such as acetone.
SCAN FOR FLYER
SCAN FOR TEST DEMO VIDEO
Page 15
AbOuT
15
COMPLIANCE TESTING FOR FREE AND TOTAL CHLORINE
This DPD test system because this test G4, ISO 7393/2) accepted for testing requirements for Free and Total Chlorine. The eXact iDip® meter uses a wavelength requirement between 490 and 530nm. The eXact Strip CL (DPD-1) uses the same reagents AWWA (American Water Works Association) method 4500-Cl G. USEPA does not “approve” commercial DPD delivery systems as reagent powder pillows, tablets,dispensers, or eXact Strip DPD delivery devices. The eXact Strip Micro CL (D P D -1) for Free and the eXact Strip Micro CL (D PD-3) or the eXact Strip Micro (DPD - 4) for Total Chlorine m eet yo ur re por t abl e testing because the eXact Strip Micro CL delivers the same in
identical proportions (see table below); therefore, the system compliant. Likewise, AWWA proportions are followed as required Total Chlorine measurements using Potassium
COMp
(
fREE Ch
Anhydrous DPD sulfate Anhydrous Na HPO Anhydrous KH PO Na
EDTA 1.1% 1.1%
2
is
accepted
is
USEPA (D IN Standard 38
of
525nm; and the
is
that the colorimeter wavelength
ONENT
LORiNE
)
4
2 4 2
by
most health departments
compliance
It
should
aWWa
4500-CL
1.5% 1.5%
33.4% 33.4%
64.0% 64.0%
g
408
is
Micro be
Standard
understood that
chemicals
Iodide.
Methods
4500 - Cl
the
such
Chlorine,
CL
requirements
is
for
EXact
DpD-1
WARRANTY (2 YEARS)
Registration of your eXact photometer must be received within 30 days from of purchase to activate the warranty. The eXact photometer is warranted to free from defects in materials and workmanship for a period of two (2) years the date of purchase by the customer. ITS will repair or replace, at its product which is deemed to be faulty due to manufacturing defect. Wa rranty does not cover product damage caused by abuse (such as crushing a tablet in cell), battery corrosion damage, or improper use. If the meter is faulty or defective cont act ITS by phone (+1-803- 32 9 - 9712 Ext. 0) or email to describe the problem and obtain a return authorization form before returning photometer to IT S. Damage caused by improper packing of the photometer for shipment to ITS will not be covered by the warranty. Customer is responsible shipping charges to ITS . ITS pays postage whe n photom eter is returned to customer. A maximum processing fee of $75 will be charged for repair or replacement of registered photometers and damages not covered by this warrant y. Registration available over the phone (+1-803-329-9712 Ext. 0) or online at
com/micro/warranty/ (Personal data is
of the photometer will not extend or renew the period of guarantee. This does not affect your statutory rights. The warranty is not
kept confidential). The repair
http://www.sensafe.
transferable.
date
be
from
discretion,
the
otherwise
for
or repl
acement
warranty
TECHNICAL SUPPORT
Please visit www.sensafe.com/idip for the latest technical information and how-to videos.
For additional technical support, call (803) 329-9712 or send us an email at
the
return
non-
is
G
Page 16
®
T
16 REORDER
US Designed and Patent-Protected by Industrial Test Systems, I nc. 1875 Langston Street, Rock Hill, SC USA Manufactured by Metertech Inc. 63-2, Cheng Gong Road, Sec. 1, NanGang, Taipei, Taiwan, R.O.C. 2783-2854 Fax:+886-2-2783-1764
Apple, the Apple logo, iPad, iPhone, and iPod touch a re trademarks EXACT and IDI The Bluetooth
AVAILABLE TESTS & REAGENT S
Test
Parameter
THH
™
Tot al
(DPD-1)
(VI)
(Cu+2)
Acid
(TPTZ)**
(DPD-4)
Carrying
NaBr)
Hardness
3
)
2
S)
2
Part #
486101
Kit
CY
Kit
Kit
(DPD-1)**
(DPD-1)
Range
(DPD-4)
)
(DPD-4)
(DPD-1)
)
3
Salt
(DPD-3)
LR MR HR
486211 486212 486213
Metals
Kit
486214
Glycine
Case 481661-ID N/A N/A
486641 486659 486636 486629 486757 486757 486633 486637 486672 486638
*
**
486670 486614 486632
481652-II
CaCO3)
CaCO3)
E
486656 486630 486616 486648 486676 486627 486650 486610 486606 486604 486653 486655 486623 486634 486674 486626
486639-II
486814 486608 486646
N/A
in
case quantities (12 units per
-mail:info@metertech.com.tw
of
Industrial Test Systems, Inc. Rock Hill, SC USA.
by
Bluetooth SIG and any use
eXact iDip
(1) Bulk meter, 6 foils of each: FC, CC, AL, pH, (1) cleaning Brush, (1) instruction booklet
eXact Pool Refill
AL, pH, DPD-1, DPD-3, CA,
eXact Well Refill
pH, Fe, NO , AL,
3
eXact Tap Refill
pH, AL, THH, DPD-1, DPD-4, HR Cl,
eXact Process Refill
pH, DPD-1, DPD-4, HR Cl, MR H O ,
2
2
eXact iDip Alkalinity, Bromide (as Bromine
Calcium (as CaCO Chloride (as NaCl) Chloride, High Range (as NaCl)** Chlorine Dioxide Chlorine, Free Chlorine, High Chlorine, Combined Chlorine, Total Chromium Copper Cyanuric Hardness, Total HR (as Hardness, Total LR (as Hydrogen Peroxide Hydrogen Peroxide Hydrogen Peroxide Iodine
(DPD-1)
Iron, Total Magnesium Manganese** Metals Molybdate** Nitrate (as NO Nitrite (as NO Ozone Peracetic Acid Permanganate pH Phosphate Sulfate Sulfide (as H
urbidity**
*Combined Chlorine DPD- 3 Test requires Free Chlorine DPD- 1 ( 48663 7) For resellers and distributors - products sold **Test uses a non-standard test method. Visit sensafe.com/idip for
®
P
are registered trad em arks
®
word mark and logos are owned
As As As As As
11 - 200 19 - 400
0.07 - 17 19 - 550
3 - 600
347 - 9975
0.04 - 15
0.05 - 12
1 - 280
0.05 - 12
0.05 - 12
0.01 - 2
0.6 - 11 3 - 110
60 - 900
1 - 125
0.02 - 3.5 1 - 130
16 - 4200
Coming
0.03 - 8 2 - 400
0.03 - 2.6 1 - 1.75
0.02 - 5
0.25 - 32
0.02 - 4
0.01 - 2
0.05 - 11
0.02 - 6
6.0 - 8.5
0.02 - 5
1 - 270
Coming 24 - 780
details.
of
Apple Inc., registered
of
such marks
Range
Below 6 Below Below Below Below
ppm 100 ppm 25 ppm ppm
ppm
ppm ppm ppm
ppm 50
ppm ppm
ppm 50 ppm
ppm
ppm
ppm 100
ppm
ppm
ppm 100
Soon
ppm
ppm
ppm ppm ppm 50
ppm 50 ppm 50 ppm 100
ppm ppm
pH 100
ppm
ppm Soon
NTU N/A
to
be run
first.
case).
in
the U.S. and other c
is
under
license.
As As As As
# of
Tests
Below Below Below Below
100
50 25
25 100 100
100 100
50
60
50
50
50
100
50
50
24
24
100 100
50
50
50
T
el.:+886-2-
ountries.
R050914
Page 17
FEDERAL COMMUNICATIONS COMMISSION INTERFERENCE STATEMENT
This equipment has been tested and found to comply with the limit s for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency e nergy and, if not installed and used in accordance with the instructions, may caus e harmful interference to radio communications. However, there is no gua rantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment of f and on, the user is encouraged to try to correct the interference by one or more of the following measures:
-- Reorient or relocate the receiving antenna.
-- Increase the separation between the equipment and receiver.
-- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-- Consult the dealer or an experienced radio/TV technician for help.
CAUTION: Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Loading...