Metek RCT 10, RCT 11, RCT 12, RCT 13 Installation Manual

A Leader in
Level Measurement
For Assistance Call 1-800-527-6297
Outside North America + 215-674-1234
Installation and
Operating Instructions
RCT 10,11, 12 & 13
with HART® Protocol
using 409-T Electronics
U.S. and Canada: 1-800-553-9092 24-Hour Service: 1-800-527-6297 International: +1 215-674-1234 Fax: +1 215-674-2731 E-mail: drexelbrook.service@ametek.com Website: www.drexelbrook.com
AMETEK Drexelbrook makes no warranty of any kind with regard to the material contained in this
manual, including, but not limited to, implied warranties or fitness for a particular purpose. Drexelbrook shall not be liable for errors contained herein or for incidental or consequential damages in connection with the performance or use of material.
Copyright 2007 AMETEK Drexelbrook
EDO# 9-10-111 RCT 10, 11, 12 & 13-LM Issue #6
RCT Series Transmitter
with HART® Protocol
using 409-T Electronics
An ISO 9001 Certified Company
205 Keith Valley Road, Horsham, PA 19044 U.S. and Canada: 1-800-553-9092 24-Hour Service: 1-800-527-6297 International: +1 215-674-1234 Fax: +1 215-674-2731 E-mail: drexelbrook.info@ametek.com Website: www.drexelbrook.com
Table Of Contents
Table of Figures .............................................................................................................................. vii
Section 1: Introduction ............................................................................................................... 1
1.1 System Description ........................................................................................................... 1
1.2 Technology ........................................................................................................................ 1
1.3 Models Available ............................................................................................................... 2
1.4 Sensing Element Model Numbering ................................................................................. 4
1.5 Area Classifications .......................................................................................................... 4
Section 2: Installation ................................................................................................................. 5
2.1 Unpacking ......................................................................................................................... 5
2.2 Hazardous Location Installations ...................................................................................... 5
2.3 Mounting the Electronic Unit ............................................................................................. 5
2.4 Wiring the Electronic Unit ............................................................................................... 10
2.5 Wiring the Sensing Element ........................................................................................... 11
2.6 Spark (Static Electricity) Protection ................................................................................ 12
2.7 Surge Voltage (Lightning) Protection .............................................................................. 13
2.8 RFI (Radio Frequency Interference) Filters ..................................................................... 14
2.9 Electrostatic Filters ......................................................................................................... 16
2.10 Digital Integral Meter ....................................................................................................... 16
Section 3: Configuration and Calibration with Drexelbrook PC Software ........................... 19
3.1 General Description ........................................................................................................ 19
3.2 Model Number ................................................................................................................ 19
3.3 System Requirements .................................................................................................... 19
3.4 Installing The RS232 Modem .......................................................................................... 20
3.4.1 Installing The USB Modem ............................................................................................. 21
3.5 Install the Windows Version HARTWin 2.1 Software on Hard Drive ............................... 22
3.6 Description of Function Keys .......................................................................................... 22
3.7 Configuration................................................................................................................... 24
3.7.1 Level Configuration ......................................................................................................... 24
3.7.2 Vessel Configuration ....................................................................................................... 25
3.7.3 Lower and Upper Range Values (LRV and URV) ........................................................... 25
3.8 Calibration ....................................................................................................................... 26
3.8.1 Point Calibration .............................................................................................................. 27
3.8.2 Level Calibration ............................................................................................................. 28
3.8.3 Application Example ....................................................................................................... 28
3.9 PC Status Messages ...................................................................................................... 30
3.10 Set D/A Trim .................................................................................................................... 31
3.11 Strapping Table ............................................................................................................... 31
3.12 Digital Integral Meter Configuration ................................................................................ 33
3.13 Save/Print Entries ........................................................................................................... 33
3.14 Validation ........................................................................................................................ 34
3.14.1 Validation Design Concept .............................................................................................. 34
3.14.2 Validation Procedures ..................................................................................................... 34
3.14.3 Validation Results ........................................................................................................... 36
3.15 Calibration & Configuration via 401-44-3 Display/Keypad .............................................. 37
Section 4: Configuration and Calibration with Model 275 with
Drexelbrook Device Description ............................................................................ 39
4.1 Drexelbrook Device Description ...................................................................................... 39
4.2 Start-up ........................................................................................................................... 40
4.3 Configuration................................................................................................................... 41
4.4 Calibration ....................................................................................................................... 42
4.4.1 Point Calibration .............................................................................................................. 43
4.4.2 Capacitance Calibration .................................................................................................. 44
4.4.3 D/A Trim .......................................................................................................................... 44
4.4.4 Strapping Table ............................................................................................................... 46
Section 5: Configuration & Calibration with Model 275 Communicator without
a Device Description ............................................................................................... 47
5.1 Range/Span Control ....................................................................................................... 47
5.2 Basic Rules & Conventions of HART Configuration Software (Model 275) .................... 48
5.3 Tag ID.............................................................................................................................. 48
5.4 Set Up Procedures ......................................................................................................... 49
5.5 Reading Input and Output ............................................................................................... 52
5.6 Calibration Using Actual Tank Level ................................................................................ 53
5.7 Bench Calibration (if needed) ......................................................................................... 54
5.7.1 Bench Calibration Information Sheet .............................................................................. 55
5.8 Point Calibration .............................................................................................................. 56
5.8.1 Fine Tuning Calibration ................................................................................................... 56
5.8.2 Selecting Engineering Units ............................................................................................ 56
5.9 Handheld Calibrator Error Messages .............................................................................. 57
Section 6: Troubleshooting ...................................................................................................... 59
6.1 Identifying a Problem/Symptom ...................................................................................... 60
6.1.2 Error Codes .................................................................................................................... 60
6.2 Troubleshooting Loop Connection .................................................................................. 60
6.3 Rosemount Model 268 or 275 Calibrator Cannot Identify or Find Device ....................... 61
6.4 Rosemount Model 275 Calibrator with
Device Description Cannot Identify or Find Device ......................................................... 61
6.5 RCT Series Transmitter Does not Communicate with
Drexelbrook PC software ................................................................................................ 62
6.6 Troubleshooting Transmitter ............................................................................................ 63
6.6.1 Transmitter Drift Test ....................................................................................................... 64
6.7 Troubleshooting Sensing Element .................................................................................. 65
6.8 Troubleshooting Coaxial Cable ....................................................................................... 66
6.9 Static Electricity .............................................................................................................. 68
6.10 Radio Frequency Interference ......................................................................................... 68
6.11 Factory Assistance .......................................................................................................... 68
6.12 Field Service ................................................................................................................... 69
6.13 Customer Training ........................................................................................................... 69
6.14 Return Equipment ........................................................................................................... 69
6.15 RCT Series Troubleshooting Guide................................................................................. 71
Section 7: Specifications .......................................................................................................... 73
7.1 Transmitter Specifications ............................................................................................... 73
7.2 Coaxial Cable Specifications .......................................................................................... 74
7.3 Enclosure Specifications ................................................................................................. 74
7.4 Approvals Available ......................................................................................................... 75
7.4.1 ATEX Approvals - Install Per 420-0004-203-ID ............................................................... 75
7.4.2 CSA Approvals - Install Per 420-0004-225-CD ............................................................... 76
7.4.3 FM Approvals - Install Per 420-0004-224-CD ................................................................. 76
Section 8: Normal Maintenance ............................................................................................... 77
8.1 Viewport Cleaning ........................................................................................................... 77
Appendix A: Approval Drawings ................................................................................................A-1
A.1 ATEX APPROVAL DRAWINGS .....................................................................................A-1
A.2 CSA APPROVAL DRAWINGS ..................................................................................... A-12
A.3 FM APPROVAL DRAWINGS ....................................................................................... A-29
Table of Figures
Figure 1-2 RF Admittance Sensing Element with Cote-Shield ..................................................... 1
Figure 1-1 Capacitance Sensing Element .................................................................................... 1
Figure 2-1 Recommended Conduit Connection ........................................................................... 6
Figure 2-2 Mounting Dimensions ................................................................................................. 6
Figure 2-3 Installing Sensing Element .......................................................................................... 8
Figure 2-4 Common Installation Mistakes .................................................................................... 8
Figure 2-5 Installing Sensing Element in Agitated Vessel ............................................................ 9
Figure 2-6 Providing Ground Reference ....................................................................................... 9
Figure 2-7 RCT Series Wiring Connections Integral Mounting ................................................... 10
Figure 2-8 RCT Series Wiring Connections, Remote Mounting ................................................. 10
Figure 2-9 Spark Protection for Integral Sensing Elements ....................................................... 12
Figure 2-10 Spark Protection for Remote Sensing Elements ....................................................... 13
Figure 2-11 Surge Voltage Protection .......................................................................................... 14
Figure 2-12 Radio Frequency Interference (RFI) Filters ............................................................... 15
Figure 2-13 Electrostatic Filter ..................................................................................................... 16
Figure 2-14 Digital Meter in Housing with Viewport...................................................................... 17
Figure 3-1 RS232 Modem Assembly & Loop Connection .......................................................... 20
Figure 3-1a USB Modem Assembly & Loop Connection ............................................................. 21
Figure 3-2 Selecting COM ports during software installation ..................................................... 22
Figure 3-3 PC Software Menu Screen ....................................................................................... 23
Figure 3-4 Configure Transmitter from Menu screen .................................................................. 24
Figure 3-5 Level Configuration from Menu screen ..................................................................... 24
Figure 3-6 Vessel Configuration from Menu screen ................................................................... 25
Figure 3-7 LRV & URV Configuration from Menu screen ........................................................... 25
Figure 3-9 Range Span Jumpers ............................................................................................... 26
Figure 3-10 Point Calibration from Menu screen .......................................................................... 27
Figure 3-11 Level Calibration area of Menu screen ..................................................................... 28
Figure 3-12 Application Example Diagram ................................................................................... 29
Figure 3-13 PC Software Menu Screen View of Application Example ........................................ 29
Figure 3-13a PC Software Menu Screen View of Main Menu ....................................................... 30
Figure 3-14 Setting D/A Trim Menu Screen "Pop-Ups" ............................................................... 31
Figure 3-15 Menu Screen Transforms to Strapping Table ............................................................ 32
Figure 3-16 Configure Meter Pop-up from Menu Screen; values relate to LRV & URV ............... 33
Figure 3-17 Print Pop-up from Menu ............................................................................................ 33
Figure 3-18 Capacitance on Menu Screen ................................................................................... 34
Figure 3-19 Real Time View Pop-up from Menu Screen .............................................................. 35
Figure 3-20 Typical Printout of Transmitter Data .......................................................................... 36
Figure 4-1 Typical Transmitter Loop ........................................................................................... 40
Figure 6-1 Place capacitor on transmitter from probe to ground terminals ................................ 63
Figure 6-2 C-Box Adjustment ..................................................................................................... 64
Figure 6-3 Span Range Capacitance Values ............................................................................. 64
Figure 6-4 Transmitter Drift Test ................................................................................................. 65
Figure 6-5 Sensing Element Testing, Material Below the Sensing Element ............................... 66
Figure 6-6 Sensing Element Testing, Material Covering the Sensing Element .......................... 66
vii
SECTION 1
Introduction
SECTION 1: INTRODUCTION
1.1 System Description
The instructions in this manual are for the AMETEK Drexelbrook
RCT Series RCT Series for level measurement in liquids, slurries, interfaces, and granulars.
k
air
d
C
C =
A
k
k
k A
d
air
media
C
d
A
k
C
media
A
k
=
d
Figure 1-1 Capacitance
Sensing Element
1.2 Technology
In a simple capacitance sensing element, when the level rises and
R
Xc
Oscillator circuitry through phase shift
cancels small amounts of RF current
flow (both Resistive and Capacitive)
caused by coating
High Resistance
Xc
R
Coating
TFE Insulation Sensing Element
Figure 1-2
RF Admittance Sensing
Element with Cote-Shield
(More Coating is
Easier to Ignore)
Tank Wall
Little to No Resistance
Each AMETEK Drexelbrook RCT Series system consists of a RCT Series two-wire, 4-20 mA electronic unit and a 700 series sensing element. A 380 series connecting cable is also supplied for connection of the sensing element to the electronic unit.
The RCT Series system is an admittance-to-current transducer. A change in level produces a change in admittance which results in a change of current. It is termed a two-wire transmitter because the same two wires that are used to power the unit also indicate the change in level (4-20 mA).
material covers the sensing element, the capacitance within the circuit between the sensing element and the medium (conductive applications) or the sensing element and the vessel wall (insulating applications) increases. This is due to the dielectric constant (k) of the material which causes a bridge misbalance. The signal is
demodulated (rectied), amplied, and the output is increased. There
are drawbacks, however, especially when there is coating of the sensing element.
An RF Admittance level transmitter is the next generation. Although similar to the capacitance concept, RCT Series employs a radio frequency signal and adds the Cote-Shield™ circuitry within the Electronics Unit.
Built-in oscillator buffer and chopper drive circuits permit separate measurement of resistance and capacitance. Since the resistance and the capacitance of any coating are of equal magnitude (by physical laws), the error generated by a coating can be measured and subtracted from the total output.
This patented Cote-Shield™ circuitry is designed into the RCT Series and enables the instrument to ignore the effect of buildup or material coating on the sensing element. The sensing element is mounted in the vessel and provides a change in RF admittance indicating presence of material.
The Cote-Shield™ circuitry prevents the transmission of RF current through the coating on the sensing element. The only path to ground available for the RF current is through the material being measured.
The result is an accurate measurement regardless of the amount of coating on the sensing element. By far the most versatile technology available it works with all types of materials in a vast array of conditions, from cryogenics to high temperature, from vacuum to 10,000psi pressure.
1
RCT 10, 11, 12, 13 Series
6 E 700-0005-285 6 F 700-0005-385 6 G 700-0205-002 6 H 700-0205-005 6 I 700-0205-015 6 J 700-0205-018 6 K 700-0205-020 6 L 700-0205-075 6 M 700-0205-079
7 0 700-0008-122 7 1 700-0008-124 Hast C 7 2 700-0008-126 CS 7 3 700-0008-126 316SS 7 4 700-0008-134 7 5 700-0008-136 7 6 700-0008-144 7 7 700-0008-144 91-77 7 8 700-0008-144 91-299 7 9 700-0008-174 CS 7 A 700-0008-174 316SS 7 B 700-0008-174 Hast C 7 C 700-0008-222 7 D 700-0008-222 91-269 7 E 700-0008-242 7 F 700-0018-122 7 G 700-0018-124 Hast C 7 H 700-0018-126 CS 7 I 700-0018-126 316SS 7 J 700-0018-134 7 K 700-0018-136 7 L 700-0008-144 7 M 700-0018-144 91-77 7 N 700-0018-144 91-299 7 O 700-0018-174 CS 7 P 700-0018-174 316SS 7 Q 700-0018-174 Hast C 7 R 700-0018-222 7 S 700-0018-222 91-269 7 T 700-0018-24
2
8 0 700-0001-051 8 1 700-0001-052 8 2 700-0001-053 8 3 700-0001-054 8 4 700-0001-061 8 5 700-0001-062 8 6 700-0001-063 8 7 700-0001-064 8 8 700-0002-041 8 9 700-0002-042 8 A 700-0002-043 8 B 700-0002-044 8 C 700-0002-051 8 D 700-0002-052 8 E 700-0002-053 8 F 700-0002-054 8 G 700-0002-061 8 H 700-0002-062 8 I 700-0002-063 8 J 700-0002-064
9 0 700-0001-034 9 1 700-0001-040
9 2 700-0001-044 9 3 700-0001-045 9 4 700-0001-324 9 5 700-0002-027 90-635 9 6 700-0002-029 9 7 700-0002-033 9 8 700-0002-040 9 9 700-0002-057 90-523 9 A 700-0002-057 91-006 9 B 700-0002-057 91-232 9 C 700-0002-059 9 D 700-0002-321 90-674 9 E 700-0003-009 9 F 700-0004-031 9 G 700-0004-050 9 H 700-0005-018 91-26 9 I 700-0005-018 91-195 9 J 700-0005-348 91-59 9 K 700-0009-002 9 L 700-0009-024 9 M 700-0009-057 9 N 700-0011-001 9 O 700-0011-003 9 P 700-0011-004 9 Q 700-0011-015 9 R 700-0021-001 9 S 700-0021-002 9 T 700-0021-003 9 U 700-0021-007 9 V 700-9000-494 N N System without Sensing
Element
1.3 Models Available
RCT RF Series
Technology
R RF
Measurement Type
C Continuous
Input
T Two Wire
Output Type
1 4-20 mA HART
R C T 01
Application Type
2 100 Khz 0 Degree
3 100 Khz 45 Degree
Digital Display
0 Without Meter
1 With Meter and Keypad
Reserved
Approvals
0 No Approval (Europe)
1 No Approval (NAFTA)
F FM Approval
C CSA Approval
A ATEX Approval
Surge/Noise Suppression
0 No additional filtering required
1 Loop RFI Filter
2 Probe RFI Filter
3 Loop and Probe RFI Filter
4 Loop Surge Filter
5 Probe Surge Filter
6 Loop and Probe Surge Filter
7 RFI and Surge Protection
8 Desalter Filter
Remote/Integral & Cable
0 Integral
1 Remote - no cable
2 Remote - 10 feet cable GP
3 Remote - 25 feet cable GP
4 Remote - 35 feet cable GP
5 Remote - 50 feet cable GP
6 Remote - 75 feet cable GP
7 Remote - 100 feet cable GP
8 Remote - 25 feet Tri-ax
9 Remote - 50 feet Tri-ax
A Remote - 75 feet Tri-Ax
B Remote - 100 feet Tri-Ax
C Remote - 10 feet Hi-temp GP composite
D Remote - 25 feet Hi-temp GP composite
E Remote - 35 feet Hi-temp GP composite
F Remote - 50 feet Hi-temp GP composite
GHRemote - 75 feet Hi-temp GP composite
Remote - 100 feet Hi-temp GP composite
Continued on Next Page
2
Sensing Element
0 0 700-1202-001 0 1 700-1202-012 0 2 700-1202-014 0 3 700-1202-018 0 4 700-1202-041 0 5 700-1202-041 (91-303) 0 6 700-1202-031 0 7 700-1202-010 0 8 700-1202-041 w/conc 0 9 700-1202-033 1 0 700-0001-018 1 1 700-0201-005 1 2 700-0201-005 Hast. C 1 3 700-0201-036 1 4 700-0202-002 1 5 700-0202-043 1 6 700-0002-360 1 7 700-0202-036 1 8 700-0001-022 1 9 700-0001-016 1 A 700-0001-024 Hast. C 1 B 700-0001-026 CS 1 C 700-0001-026 316SS 1 D 700-0001-074 CS 1 E 700-0001-074 316SS 1 F 700-0001-074 Hast. C 1 G 700-0001-344 1 H 700-0001-344 91-175
2 0 700-0209-002 2 1 700-209-002 91-112 2 2 700-0209-024 2 3 700-0002-023 2 4 700-0002-024 316SS 2 5 700-0002-024 Hast C 2 6 700-0002-027 2 7 700-0002-027 91-220 2 8 700-0002-028 2 9 700-0002-036 2 A 700-0002-037 2 B 700-0002-056 2 C 700-0002-057 2 D 700-0002-057 91-120 2 E 700-0002-224 316SS 2 F 700-0002-224 Hast C 2 G 700-0002-227 2 H 700-0002-321 2 I 700-0002-363
2 9 700-0002-036 2 A 700-0002-037 2 B 700-0002-056 2 C 700-0002-057 2 D 700-0002-057 91-120 2 E 700-0002-224 316SS 2 F 700-0002-224 Hast C 2 G 700-0002-227 2 H 700-0002-321 2 I 700-0002-363
3 0 3 1 700-0029-001 3 2 700-0029-002 3 3 700-0029-003 3 4 700-0029-004
3 5 700-0029-005 3 6 700-0029-021 3 7 700-0029-022 3 8 700-0029-023 3 9 700-0029-024 3 A 700-0029-025 3 B 700-0029-066 3 C 700-0029-102 3 D 700-0029-103 3 E 700-0029-104 3 F 700-0029-105 3 G 700-0029-106
4 0 700-0005-018 4 1 700-0005-018 91-13 4 2 700-0005-019 4 3 700-0005-028 4 4 700-0005-029 4 5 700-0005-035 4 6 700-0005-035 91-130 4 7 700-0005-048 4 8 700-0005-054 4 9 700-0005-085 4 A 700-0005-085 Hast C 4 B 700-0005-096 4 C 700-0005-348 4 D 700-0005-354 4 E 700-0205-078
5 0 700-0207-001 5 1 700-0207-002 5 2 700-0207-003 5 3 700-0207-004 5 5 700-0207-006 5 6 700-0201-025 5 7 700-0201-026 5 8 700-0201-027 5 9 700-0201-051 5 A 700-0201-052 5 B 700-0201-058
5 C 700-0201-059 5 D 700-0202-053 5 E 700-0202-054 5 F 700-0202-056
R C T 01
6 A 700-0005-009 6 B 700-0005-036 6 C 700-0005-045 6 D 700-0005-095 6 E 700-0005-285 6 F 700-0005-385 6 G 700-0205-002 6 H 700-0205-005 6 I 700-0205-015 6 J 700-0205-018 6 K 700-0205-020 6 L 700-0205-075 6 M 700-0205-079
7 0 700-0008-122 7 1 700-0008-124 Hast C 7 2 700-0008-126 CS 7 3 700-0008-126 316SS 7 4 700-0008-134 7 5 700-0008-136 7 6 700-0008-144 7 7 700-0008-144 91-77 7 8 700-0008-144 91-299 7 9 700-0008-174 CS 7 A 700-0008-174 316SS 7 B 700-0008-174 Hast C 7 C 700-0008-222 7 D 700-0008-222 91-269 7 E 700-0008-242 7 F 700-0018-122 7 G 700-0018-124 Hast C 7 H 700-0018-126 CS 7 I 700-0018-126 316SS 7 J 700-0018-134 7 K 700-0018-136 7 L 700-0008-144 7 M 700-0018-144 91-77 7 N 700-0018-144 91-299 7 O 700-0018-174 CS 7 P 700-0018-174 316SS 7 Q 700-0018-174 Hast C 7 R 700-0018-222 7 S 700-0018-222 91-269 7 T 700-0018-24
2
8 0 700-0001-051 8 1 700-0001-052 8 2 700-0001-053 8 3 700-0001-054 8 4 700-0001-061 8 5 700-0001-062 8 6 700-0001-063 8 7 700-0001-064 8 8 700-0002-041 8 9 700-0002-042 8 A 700-0002-043 8 B 700-0002-044 8 C 700-0002-051 8 D 700-0002-052 8 E 700-0002-053 8 F 700-0002-054 8 G 700-0002-061 8 H 700-0002-062 8 I 700-0002-063 8 J 700-0002-064
9 0 700-0001-034 9 1 700-0001-040
9 2 700-0001-044 9 3 700-0001-045 9 4 700-0001-324 9 5 700-0002-027 90-635 9 6 700-0002-029 9 7 700-0002-033 9 8 700-0002-040 9 9 700-0002-057 90-523 9 A 700-0002-057 91-006 9 B 700-0002-057 91-232 9 C 700-0002-059 9 D 700-0002-321 90-674 9 E 700-0003-009 9 F 700-0004-031 9 G 700-0004-050 9 H 700-0005-018 91-26 9 I 700-0005-018 91-195 9 J 700-0005-348 91-59 9 K 700-0009-002 9 L 700-0009-024 9 M 700-0009-057 9 N 700-0011-001 9 O 700-0011-003 9 P 700-0011-004 9 Q 700-0011-015 9 R 700-0021-001 9 S 700-0021-002 9 T 700-0021-003 9 U 700-0021-007 9 V 700-9000-494 N N System without Sensing
Element
1.3 Models Available (Continued)
Sensing Element
0 0 700-1202-001 0 1 700-1202-012 0 2 700-1202-014 0 3 700-1202-018 0 4 700-1202-041 0 5 700-1202-041 (91-303) 0 6 700-1202-031 0 7 700-1202-010 0 8 700-1202-041 w/conc 0 9 700-1202-033 1 0 700-0001-018 1 1 700-0201-005 1 2 700-0201-005 Hast. C 1 3 700-0201-036 1 4 700-0202-002 1 5 700-0202-043 1 6 700-0002-360 1 7 700-0202-036 1 8 700-0001-022 1 9 700-0001-016 1 A 700-0001-024 Hast. C 1 B 700-0001-026 CS 1 C 700-0001-026 316SS 1 D 700-0001-074 CS 1 E 700-0001-074 316SS 1 F 700-0001-074 Hast. C 1 G 700-0001-344 1 H 700-0001-344 91-175
2 0 700-0209-002 2 1 700-209-002 91-112 2 2 700-0209-024 2 3 700-0002-023 2 4 700-0002-024 316SS 2 5 700-0002-024 Hast C 2 6 700-0002-027 2 7 700-0002-027 91-220 2 8 700-0002-028 2 9 700-0002-036 2 A 700-0002-037 2 B 700-0002-056 2 C 700-0002-057 2 D 700-0002-057 91-120 2 E 700-0002-224 316SS 2 F 700-0002-224 Hast C 2 G 700-0002-227 2 H 700-0002-321 2 I 700-0002-363
Sensing Element
3 0 3 1 700-0029-001 3 2 700-0029-002 3 3 700-0029-003 3 4 700-0029-004
3 5 700-0029-005 3 6 700-0029-021 3 7 700-0029-022 3 8 700-0029-023 3 9 700-0029-024 3 A 700-0029-025 3 B 700-0029-066 3 C 700-0029-102 3 D 700-0029-103 3 E 700-0029-104 3 F 700-0029-105 3 G 700-0029-106
4 0 700-0005-018 4 1 700-0005-018 91-13 4 2 700-0005-019 4 3 700-0005-028 4 4 700-0005-029 4 5 700-0005-035 4 6 700-0005-035 91-130 4 7 700-0005-048 4 8 700-0005-054 4 9 700-0005-085 4 A 700-0005-085 Hast C 4 B 700-0005-096 4 C 700-0005-348 4 D 700-0005-354 4 E 700-0205-078
5 0 700-0207-001 5 1 700-0207-002 5 2 700-0207-003 5 3 700-0207-004 5 5 700-0207-006 5 6 700-0201-025 5 7 700-0201-026 5 8 700-0201-027 5 9 700-0201-051 5 A 700-0201-052 5 B 700-0201-058
5 C 700-0201-059 5 D 700-0202-053 5 E 700-0202-054 5 F 700-0202-056
Sensing Element
6 0 700-0204-038 6 1 700-0204-002 6 2 700-0204-048 6 3 700-0204-022 6 4 700-0204-024 6 5 700-0204-049 6 A 700-0005-009 6 B 700-0005-036 6 C 700-0005-045 6 D 700-0005-095 6 E 700-0005-285 6 F 700-0005-385 6 G 700-0205-002 6 H 700-0205-005 6 I 700-0205-015 6 J 700-0205-018 6 K 700-0205-020 6 L 700-0205-075 6 M 700-0205-079
7 0 700-0008-122 7 1 700-0008-124 Hast C 7 2 700-0008-126 CS 7 3 700-0008-126 316SS 7 4 700-0008-134 7 5 700-0008-136 7 6 700-0008-144 7 7 700-0008-144 91-77 7 8 700-0008-144 91-299 7 9 700-0008-174 CS 7 A 700-0008-174 316SS 7 B 700-0008-174 Hast C 7 C 700-0008-222 7 D 700-0008-222 91-269 7 E 700-0008-242 7 F 700-0018-122 7 G 700-0018-124 Hast C 7 H 700-0018-126 CS 7 I 700-0018-126 316SS 7 J 700-0018-134 7 K 700-0018-136 7 L 700-0008-144 7 M 700-0018-144 91-77 7 N 700-0018-144 91-299 7 O 700-0018-174 CS 7 P 700-0018-174 316SS 7 Q 700-0018-174 Hast C 7 R 700-0018-222 7 S 700-0018-222 91-269
Introduction
Sensing Element
3
RCT 10, 11, 12, 13 Series
1.4 Sensing Element Model Numbering
System Reference Number
18 Water-like
1C Concentric water-
24 Low viscosity
26 Interface of liquids
2A Thick conducting
2C Conducting liquids
40 Agitated
42 Heavy-duty for
48 Long lengths of
4E Insulating liquids
*“X” is a fluorocarbon-type insulation
Typical Application Type
conductive liquids
like insulating liquids
conducting liquids
containing ketones and esters
liquids
and interfaces
conducting liquids and granulars
abrasive granulars
conducting liquids
and granulars
Sensing Element Model
700-1-22 TFE-covered
700-1-26 TFE-covered
700-2-24 TFE-covered
700-2-27 FEP-covered
700-2-37 “X”*-covered
700-2-57 “X”*-covered
700-5-18 “X”*-covered
700-5-19 Urethane-
700-5-54 PFA-covered
700-205-78 “X”* covered
Standard Material of Construction
rod
rod with carbon steel concentric shield
rod
rod
rod
rod
cable
covered cable
cable
cable
OD and Standard Mounting
Rod 3/8" OD 3/4" NPT
Concentric Shield
1.66" OD 1 1/2" NPT
Rod 3/4" OD 3/4" NPT
Rod .56" OD 3/4" NPT
Rod .54" OD 3/4" NPT
Rod .84" OD 1" NPT
Cable 5/16" OD 3/4" NPT
Cable 3/4" OD 2" NPT
Cable .093" OD 3/4" NPT
Cable 5/16" OD 1" NPT
Temperature Pressure Limits
100oF @ 1000 PSI 300oF @ 500 PSI
100oF @ 1000 PSI 300oF @ 500 PSI
100oF @ 1000 PSI 450oF @ 500 PSI
100oF @ 1000 PSI 300oF @ 500 PSI
100oF @ 1000 PSI 250oF @ 500 PSI
100oF @ 1000 PSI 250oF @ 500 PSI
100oF @ 1000 PSI 250oF @ 500 PSI
150oF @ 5 PSI
100oF @ 1000 PSI 300oF @ 500 PSI
250oF @ 5 PSI
1.5 Area Classifications
The standard electronic unit mounted in the durable
See Section 1.4 for detailed specications of sensing
The electronic unit and sensing element are connected by
housing (RCTXXX1, 2) is dual-rated and meets the following conditions:
• Type NEMA 4X Waterproof / Corrosion Resistant
• IP 65
elements that are most often recommended with a RCT Series system. Contact the factory or your local representative if additional information is required.
a three-terminal coaxial cable. Drexelbrook cables are available in See Section 7.2 for Specications:
• General Purpose:
• Triax:
• Composite: (rst 10 feet high temperature)
4
Installation
SECTION 2: INSTALLATION
2.1 Unpacking
Carefully remove the contents of the carton and check
each item against the packing list before destroying any packing material. If there is any shortage or damage, report it immediately to the factory.
2.2 Hazardous Location Installations
Installation in hazardous areas must comply with the
control drawings See Appendix A. Always install to
the NEC and/ or local requirements/ codes/ directives
as mandated by the authority having jurisdiction. The
cable ttings supplied are weather resistant only. Before
using Intrinsic Safety Barriers, read manufacturer's instruction for barrier operation. The electronic unit is rated T4 and may not be used with materials with an auto
ignition temperature of less than 135Cº. Substitution
of components may impair intrinsic safety. To prevent
ignition of ammable or combustible atmospheres,
disconnect power before servicing.
When installing the transmitter in an enclosure not
supplied by AMETEK Drexelbrook, the enclosure
must be certied for the area in which it is to be
installed.
2.3 Mounting the Electronic Unit
The RCT (409-T) Series system was designed for eld
mounting, but it should be mounted in a location as free as possible from vibration, corrosive atmospheres, and any possibility of mechanical damage. For convenience at start-up, mount the instrument in a reasonably accessible location. Ambient temperatures should be between -40oF
and 167oF (-40°C and 75°C).
See section 7.4 for customer supplied enclosure requirements for 409 Series.
5
RCT 10, 11, 12, 13 Series
WRONG
CORRECT
CONDUIT
BREATHER
DRAIN
WRONG
CORRECT
* *
Allows Moisture Infiltration
Use only cable supplied by AMETEK Drexelbrook
All conduit connections are sealed. Gaskets are in place.
Fill Pipe Ends with silicone sealant.
*
2.3 Mounting the Electronic Unit (Continued)
Figure 2-1 shows the recommended conduit installation.
Figure 2-2 shows the dimensions when installing conduit
to the electronic unit.
Figure 2-1
Recommended Conduit
Connection
Figure 2-2
Mounting Dimensions
Inches (mm)
6
Installation
2.3 Mounting the Electronic Unit (Continued)
The mounting location for the sensing element is often
determined by whether there is a suitable location inside a vessel. An external side arm or stilling well can be considered.
The following sensing element mounting and installation instructions should be followed so that the equipment will operate properly and accurately:
A. In applications requiring an insulated sensing
element, use particular care during installation. There is always the danger of puncturing the insulation, especially with the thin-walled sensing elements.
B. Sensing elements should be mounted in such a
manner that they are not in the direct stream of
a lling nozzle or chute. If this is not possible, a deecting bafe should be installed between the sensing element and the ll.
C. Do not take the sensing element apart or loosen the
packing glands. Follow instructions in Figure 2-3.
D. Avoid installing the sensing element with any of
the common mistakes shown in Figure 2-4.
E. If a stilling well is used, ensure that "vent" holes
are large enough to allow free passage of both
air and process material. Holes should be 5/8" or
larger, 120
o
apart, every 2-3 feet along the length of
the stilling well.
F. Sensing elements that are mounted in agitated
vessels usually require brackets and supports to control the position of the sensing element during agitation. See Figure 2-5.
G. For non-metallic vessels without Drexelbrook self-
grounding sensing elements, choose one of the grounding recommendations shown in Figure 2-6.
7
RCT 10, 11, 12, 13 Series
2.3 Mounting the Electronic Unit (Continued)
Packing Gland Assembly Do Not Turn One Relative to The Other
Figure 2-3
Installing Sensing Element
LACK OF PROPER GROUND
(EARTH) CONNECTION
BETWEEN SENSOR MOUNTING
AND VESSEL WALL
Place Wrench Here When Tightening Condulet
Place Wrench Here to Install or Remove From Vessel
Figure 2-4
Common Installation Mistakes
STILLING
WELL
LACKS
VENT
HOLES
PROBE
CONTACTS
SIDE OF STILLING
WELL
PROBE
MAY
FLEX
INSIDE
PIPE
SLUDGE
DAMAGE
MAY
OCCUR
HERE
SLUDGE
MAY CLOG
PIPE
AVOID
FILL
STREAM
NOZZLE DIAMETER TOO SMALL OR LENGTH TOO LONG
DAMAGE
MAY
OCCUR
HERE
FILL LINE
WALL BUILD-UP TOUCHES SENSOR
8
2.3 Mounting the Electronic Unit (Continued)
Installation
RECOMMENDED RIGID
SENSOR INSTALLATION
Figure 2-5
Installing Sensing Element
in Agitated Vessel
ROD
STYLE
SENSOR
INSULATED SUPPORT BUSHINGS PART # 713-XX-X
CUSTOMER
SUPPLIED
SUPPORT BRACKET
If the vessel is non metallic see grounding sketch below. with factory supplied concentric shield or ground rod do not need additional grounding.
Measure ground continuity from housing ground screw to metal wall of vessel. A good ground will measure less than 5 ohms.
RECOMMENDED FLEXIBLE
SENSOR INSTALLATION
FLEXIBLE
CABLE TYPE
SENSOR
AVOID WEIGHTED
FLEXIBLE SENSORS IN
HEAVILY AGITATED
VESSELS
FLEXIBLE
BOTTOM ANCHOR
PART# 727-XX-X
CUSTOMER
SUPPLIED SUPPORT
sensors
Figure 2-6
Providing Ground
Reference*
*This is a sensing
element ground
reference and possibly
different from an
electrical power ground.
GROUND
CLAMP
METAL
SENSOR
D
RUN 14 AWG. WIRE FROM FABRICATED GROUND TO GREEN SENSOR GROUND SCREW
CLAMP
PLASTIC
A B
TANK
WEIGHT
F
PUMP
E
A
GROUND ROD A ground rod can be fabricated out of any metal compatible with the process. Use material that is at least 3/8inch dia. such as pipe, All-Thread, or tubing. The ground rod must be parallel and relatively close to the sensor.
B
GROUND WIRE 1/4 inch or larger dia. stainless steel rope that is anchored or weighted can be used.
C
C
MOUNT SENSOR IN METAL PIPE Mounting the level sensor inside a metal pipe provides an excellent ground reference.
D
METAL PIPING Metal piping that connects to the tank bottom can be used as a ground reference. Use only if the process material
is greter than 1000 uMHOS/cm such as acids and caustic.
E
SUBMERGED METAL STRUCTURE Use any constantly submerged metal object such as: pumps, agitators, or thermowells.
greter than 1000 uMHOS/cm such as acids and caustic.
F
METAL FLANGE A submerged metal flange or orifice plate can be used.
Use only if the process material is greater thatn 1000 uMHOS/Cm. such as acids and caustic.
Use only if the process material is
9
RCT 10, 11, 12, 13 Series
2.4 Wiring the Electronic Unit
Center Wire (Blue)
GROUND PROBEPAD SHD
Shield Wire (Red)
Two Terminal
Sensing Element
RCT Series Wiring Connections Integral Mounting
Three Terminal Sensing Element
Three Terminal Sensing Element
Figure 2-7
GROUND PROBEPAD SHD
GROUND PROBEPAD SHD
YEAR OF MANUFACTURE
GND
SERIAL NO. MODEL NO.
YEAR OF MANUFACTURE
GND
SERIAL NO. MODEL NO.
V -
+ To Signal
- To Signal
V +M -
V -
V +M -
Center Wire
(Blue)
Shield Wire (Red)
Two Terminal Sensing Element
Center Wire
(Blue)
Shield Wire (Red)
to be cut by field installer
RCT Series Wiring Connections, Remote Mounting
+ To Signal
- To Signal
1. All devices must be wired in series. Voltage driven devices require a series voltage dropping resistor.
2. Signal Terminals + & - can operate with minimum
of 12.0 VDC with 20mA owing.
3. If the eld wiring is to be in hazardous areas, then
suitable safety barriers are required between the power
supply and the eld to provide for intrinsically safe
wiring.
4. Clipped Shield Wire must NOT touch housing!
Figure 2-8
10
2.4 Wiring the Electronic Unit (Continued)
Integral units are pre-wired at the factory. Figure 2-7 shows the wiring of the integral unit.
For remote units, the signal connections are made to the four-terminal block on the front of the chassis. Due to the low power consumption of the instrument, the wiring
need only be light gauge (e.g. 20 AWG). Twisted shielded
pair cables are recommended for lengths over 200 feet.
The cable from the sensing element is connected to the
ve-terminal strip on the back side of the instrument
chassis. The cable connections are probe (prb) or center wire (cw), ground (gnd), and shield (shd). See Figure 2-8 for wiring connections of the remote unit.
Only coaxial cables supplied by Drexelbrook should be used to connect the transmitter to the sensing element. Use of other cables can result in unstable performance.
2.5 Wiring the Sensing Element
CAUTION !
Before using Intrinsic Safety Barriers, read manufacturer's instruction for barrier operation.
The 409-T has a built-in current limiter which holds the signal current to a maximum of 28 mA.
The cable connections to the sensing element are shown in Figure 2-8
Installation
• Do not connect the cable to the sensing element
until after the sensing element has been installed in the vessel and the condulet housing has been secured.
• If the sensing element does not have a shield
connection, (the most common condition for a 2­terminal sensing element) be sure to clip and /or
tape the shield wire at the sensing element end of the cable only. See Figure 2-8.
Only coaxial cables supplied by Drexelbrook should be
used to connect the transmitter to the sensing element. Use of other cables can result in unstable performance.
11
RCT 10, 11, 12, 13 Series
2.6 Spark (Static Electricity) Protection
Spark Protection for Integral Sensing Elements
If spark protection is supplied for an integral sensing
element, use the following instructions for installing the spark protection.
A. Attach the mounting link on the spark protector to
the sensing element center connection screw.
B. Connect the green wire from the spark protector to
the housing/chassis mounting screw.
C. Connect the center wire connector (Blue) of the
spark protector to the "probe" connection on the transmitter.
D. Connect the shield connector (Red) of the spark
protector to the shield (sh) connector on the transmitter.
Cote-Shield Wire (Red)
Center Wire (Blue)
Figure 2-9
Spark Protection for Integral
Sensing Elements
Ground (Green)
Spark Protector
377-1-24
Fastened Under Screw to
Housing
Spark Protection for Remote Sensing Elements
If spark protection is supplied for a remote sensing
element, use the following instructions for installing the spark protection.
A. Attach the mounting link on the spark protector to
the sensing element center connection screw.
B. Connect the green wire from the spark protector to
the ground screw.
C. Feed the coax cable into the condulet.
12
Installation
2.6 Spark (Static Electricity) Protection (Continued)
D. Connect the coax cable center wire (cw) to the spark
protector and the ground wire (gnd) to the ground screw as shown in Figure 2-10.
E. Connect the shield wire to the Cote-Shield terminal
(sh).*
* For sensing elements that do not have shield
connections, clip the shield wire as shown in
Figure 2-8.
Make sure the transmitter has a ground attached
either on sensing element side or loop side of the unit.
Unless the assembly is attached to a metallic vessel,
chassis of transmitter is not grounded.
Figure 2-10
Spark Protection for Remote
Sensing Elements
2.7 Surge Voltage (Lightning) Protection
Optional surge protection is sometimes supplied with
transmitters that are expected to be exposed to surge voltages or surges due to lightning near the two-wire loop.
A Drexelbrook Model 377-4-12 Surge Voltage Protection
affords a great deal of protection to the transmitter but is not absolute in its protection against a very close lightning strike. Refer to Figure 2-11 to properly
connect the Surge Voltage Protection. Be sure that the
transmitter housing is well connected to a good ground.
13
RCT 10, 11, 12, 13 Series
2.7 Surge Voltage (Lightning) Protection (Continued)
GND
SERIAL NO. MODEL NO.
YEAR OF MANUFACTURE
V +M -
V -
Transient Surge Protector DE Part No. 377-4-18
Figure 2-11
Surge Voltage Protection
+ To Signal (12 to 30 VDC)
- To Loop
2.8 RFI (Radio Frequency Interference) Filters
When installing the RCT Series transmitter, follow these
recommendations to avoid problems with Radio
Frequency Interference (RFI).
Choose a location to mount the electronic unit at
least 6 feet (2M) from a walkway where personnel
using walkie talkies may pass.
If the vessel is non-metallic, select, if possible, a shielded (concentric) sensor. If unsure about suitability, contact the Drexelbrook Applications department for a recommendation.
For remotely-mounted electronic units connect the sensor to the electronic unit by placing the coaxial cable in grounded metal conduit. Integrally mounted electronic unit sensor connections are already shielded.
Use Twisted Shielded Pair wiring for all loop wiring connections. Loop connection wiring should also be in grounded metallic conduit.
Where possible, use of cast aluminum housings without windowed openings for the electronic unit is recommended. If local close-coupled indicators
are used, install a loop lter between the indicator
and the electronic unit.
Ground the electronic unit and housing with a minimum
of 14 gauge wire to a good earth ground. Make sure that conduits entering and leaving the housing have a good electrical ground connection to the housing
14
Installation
2.7 RFI Filters (Continued)
If the recommendations listed are followed, it is usually
not necessary to add RFI ltering to protect against signal strengths of 10 Volts/ Meter or less. This degree of protection is usually sufcient to protect against walkie
talkies that are used 3 feet (1M) or more from a typical
electronic unit. If greater protection is required, or lters have already been provided, install RFI lters as shown
in Figure 2-12.
CE Mark Certication:
3-Terminal Coaxial Cable - Systems with remote mounted
electronics require the use of a Probe RFI lter (only)
if the sensing element is connected with 3-terminal
coaxial cable, installed in accordance with gure 2-14, to maintain CE Mark certication.
Triaxial Cable - Systems with remote mounted electronics
that connect to the sensing element via a triaxial cable
do not need a Probe RFI lter to maintain CE Mark certication.
GROUND PROBEPAD SHD
YEAR OF MANUFACTURE
SHLD
GND
CW
Figure 2-12
Radio Frequency Interference (RFI) Filters
GND
SERIAL NO. MODEL NO.
V +M -
V -
+ To Signal
- To Signal
+
POS
GND
-
NEG
15
RCT 10, 11, 12, 13 Series
2.9 Electrostatic Filters
In applications such as desalters or treaters and other
coalescers with electrostatic grids, it is customary for
Drexelbrook to supply a special lter on the sensing element. The purpose of the lter is to remove voltage
that may be imposed in the sensor from the high voltage
grids. Some earlier applications have the lter located at
the transmitter instead of the sensing element; either is acceptable.
Connect the electrostatic lter Drexelbrook Part Number 385-0028-004 as shown in Figure 2-13.
Typical Mounting For Typical Sensing Element
Figure 2-13
Electrostatic Filter
(385-0028-004)
2.10 Digital Integral Meter
An optional digital integral meter (DIM) (401-44-3) can be
used with the RCT Series electronic unit for local digital
loop indication. When purchased with the RCT Series
instrument, a housing with viewport is supplied. The meter display is visible through the viewport. If the meter
is added as a retrot to an existing installation, a new housing dome with viewport (260-2-222) is required and supplied as part of the retrot package.
To install the meter:
Remove the "perforated knockouts" from the top label of the transmitter to expose two threaded holes and ribbon cable socket
• plug the mini ribbon cable into the socket
• secure meter to top of electronic unit with screws
16
Installation
2.10 Digital Integral Meter (Continued)
Integral meter can be rotated 90° to allow for proper
viewing orientation:
• remove hold-down screws
• remove black cover screws
• move cover screws to original hold-down screw
location
remount meter in new orientation
The meter is not inserted into the 4-20 mA loop. It
receives power and data directly from the Drexelbrook smart transmitter via attached mini ribbon cable. See
Figure 2-14.
If the Ribbon Cable is attached while the transmitter is
Powered Up, the transmitter may have to be restarted
before the DIM will display correctly.
The meter can be congured using the Drexelbrook PC
software, the Model 275 Calibrator, or from the DIM
itself.
Figure 2-14
Digital Meter in Housing with
Viewport
17
SECTION 3
Conguration & Calibration
SECTION 3: CONFIGURATION AND CALIBRATION WITH
DREXELBROOK PC SOFTWARE
This section instructs the user how to use the Drexelbrook
401-700-20/40 Series PC calibrator software to congure and calibrate the RCT Series(RF Admittance)
Transmitter.
3.1 General Description
The 401-700-20/40 software package allows the use of
any Windows® 9X/NT/2000/XP-based personal, laptop,
or notebook computer to calibrate the HART Protocol transmitter.
The PC software can be used in place of the Rosemount
268 or 275 handheld calibrators used for multi-Process Variable (PV) transmitters.
3.2 Model Number
4 0 1 - 0 7 0 0 - 0 2 X / 4 X 2X= 1 PC Software Package includes:
RS232 Modem Assembly 401-0700-004 (Figure 3.1).
2X=2 PC Software Package includes:
Contents in 401-0700-021, HART® 6.0 (DOS version) on a 3½" Floppy Disk, and HARTWin™ version 2.1 or greater
on a CD-ROM.
4X=1 PC Software Package includes:
USB Modem Assembly 401-0700-007(Figure 3.1a).
4X=2 PC Software Package includes:
Contents in 401-0700-41, Utilities and Drivers on a CD-
ROM, and HARTWin™ version 2.3 or greater on a CD-
ROM.
4 0 1 - 0 7 0 0 - 0 0 6
HART® 6.0 (DOS version) on a 3½" Floppy Disk
4 0 1 - 0 7 0 0 - 0 3 1
HARTWin™ version 2.X on a CD-ROM.
3.3 System Requirements
PC Requirements
Windows® 95, 98, ME, 2000, XP.
The USB modem is not compatible with Windows® 95, 98
First Edition, or NT. It is recommended that the software
be installed on a hard drive with 20 megabytes or more of space available.
Input to Modem
RS232 or USB Port, from one of the COM serial ports (COM1, COM2, etc.). The PC provides operating power
Output (to Transmitter being Calibrated)
for the modem but not for the transmitter.
4-20 mA in HART® Protocol.
19
RCT 10, 11, 12, 13 Series
3.4 Installing The RS232 Modem
Refer to Figure 3-1 for a connection diagram and use
the following procedure to install the hardware that is
necessary to run the PC software.
A. Connect the RS232 Drexelbrook Modem
401-700-004 to one of the COM serial ports (COM1, COM2, etc.) of the computer.
B. Connect the Modem's 4-20 loop connectors to the
transmitter loop.
C. Turn on the computer.
HART COMPATIBLE
401-0700-004
Figure 3-1
RS232 Modem Assembly & Loop Connection
20
Conguration & Calibration
3.4.1 Installing The USB Modem
Refer to Figure 3-1a for a connection diagram and use
the following procedure to install the hardware that is
necessary to run the PC software.
A. Turn on the computer
B. Install Modem Software:
It is highly recommended the USB drivers be installed
BEFORE you plug in the modem.
Install the USB Drivers by inserting the Modem
Installation Disk into CD Drive of the computer.
If program does not "Auto-Run", select "D:\setup"
(where D is the letter representing the CD Drive)
Be Sure to Select the USB interface in the setup
prompt.
Follow any "On-Screen" Instructions.
C. Connect the Drexelbrook Modem 401-700-007 to a
USB port on the computer. With the USB drivers already installed, the computer will detect the modem
and assign a COM PORT number.
D. Connect the Modem's 4-20 loop connectors to the
transmitter loop.
HART COMPATIBLE
Notes:
1. Modem will operate from 32º F to 122º F (0º C to 50º C). It can be stored from -40º F to +185º F (-40º C to +85º C). 0% to 95% relative humidity - non condensing.
2. Servic Department 1-800-527-6297 or 1-215-674-1234
401-0700-007
US B
Figure 3-1a
USB Modem Assembly & Loop Connection
21
RCT 10, 11, 12, 13 Series
3.5 Install the Windows Version HARTWin 2.1 Software on Hard Drive
Installation is quite simple.
A. Place the 401-700-031 CD into the CD drive
(usually drive D:).
B. If program does not "Auto-Run", select "D:\setup"
(where D is the letter representing the CD Drive).
C. Follow "On-Screen" instructions in Setup to create
program le.
D. Once loaded, double click "HartWin" icon and the
program should run under its own window.
Figure 3-2
Selecting COM ports during
software installation
E. Select communication port [Com 1, Com 2, etc.]
and then click “OK.” See Figure 3-2
F. If you are not sure which communication port you
are using (such as when rst using a USB modem), select “Search Ports,” then OK. The software
automatically will seek out the correct one. In either case the software begins to communicate with the HART protocol transmitter and returns with a view (below) containing “name plate data,”
Tag ID and all default or existing conguration
information. This is the same as if you clicked on the Read Transmitter function button.
G. The next view, shown in Figure 3-3, appears
automatically, displaying current transmitter database for calibration set-up for your selected
Tag ID. The Scratch Pad will automatically show
the last message (last user, last calibration, etc.) up to 32 characters. If this is a new transmitter, the
Tag ID is user-dened. Serial number, transmitter
software version, range, etc. is automatically entered from the “name plate data” embedded in the transmitter:
3.6 Description of Function Keys
Figure 3-3 shows a PC calibration software menu screen.
Read Transmitter [F3 on keyboard]
The following paragraphs describe the function buttons.
The data elds are described in Section 3.7-Conguration.
Reads all pertinent data from the transmitter and displays it on the screen. The Read function also updates the real time window. Keep in mind that it takes several seconds to load the information from the transmitter.
When the load is complete, the screen shows the database parameters, except any user-dened strapping table
information. This command is also used when connecting to another transmitter.
22
Conguration & Calibration
3.6 Description of Function Keys (Continued)
Write to Transmitter [F5 on keyboard]
Sends new or edited conguration data to the transmitter. Data elds that have been edited but not sent to the
transmitter are displayed in red.
Real Time View [F4 on keyboard]
Displays the real time values of level, capacity, distance, temperature, loop current, percentage, and status.
Point Calibration [F6 on keyboard]
Calibrates the HART® protocol transmitter using Point
calibration. See Section 3.8 Calibration. Enter the low point and high point of level for an accurate calibration.
D/A Trim
Allows a eld reference meter to be connected to the
transmitter for adjusting transmitter output current. See Section 3.10.
Strapping Table
Displays the values of the input to level and output to volume in percent in a 21-point table. Allows points to be changed to accommodate irregularly shaped vessels. See Section 3.11.
Congure Meter
Congures the optional Digital Integral Meter (440-44-3)
used for local indication. See Section 3.12.
True Level Calibration (grayed out)
This button is inactive for the RCT Series Transmitter. It is for the extra features that come with the True Level model transmitter.
Figure 3-3
PC Software Menu Screen
automatically communicates
all "name plate data" from
transmitter
23
RCT 10, 11, 12, 13 Series
3.7 Configuration
Refer to Figure 3-3 PC Software Menu Screen.
Conguration involves downloading information to
the HART protocol transmitter that is specic to the
Figure 3-4
Congure Transmitter from
Calibration requires that application information and
Menu screen
application and vessel that is being measured.
two points of level and/or capacitance be supplied to the
transmitter from the calibration software.
A. Begin conguration by using Tag ID (8 characters)
to identify the unit or vessel. Use the Scratchpad (32 characters) to record the date of calibration or
other similar notes. Press Tab or Enter on your
keyboard.
3.7.1 Level Configuration
B. Select Level or Vessel in the Analog Loop
Assign selection box. Press Tab or Enter on your
keyboard.
Level conguration sets the output to follow the level of the material being measured.
Vessel conguration sets the output to follow the strapped volume or weight in the vessel. For example, gallons in a horizontal vessel.
C. Edit Damping Time from 0-90 seconds, if desired.
D. Click on Write to Transmitter.
E. Move to Level Conguration section of menu.
A. Select Level Units. The default is feet. Choose the
units that correspond to the level measurement.
Figure 3-5
Level Conguration from
Menu screen
B. Edit the Maximum Level to agree with the actual
tank height (not the length of the sensing element).
C. Click on Write to Transmitter and move to the Vessel Conguration section of the menu.
24
3.7.2 Vessel Configuration
Figure 3-6
Vessel Conguration from
Menu screen
Conguration & Calibration
A. Select Vessel Units. The default is gallons. Press
Enter and choose the units that correspond to the vessel measurement. The units include both
weight and volume outputs. Press Tab or Enter on
your key board to continue.
B. Edit the Maximum Capacity of the vessel. Enter
the corresponding value of weight or volume equal to the Maximum Level. Enter 100 for percent if the weight or volume units are not known or
needed. Press Tab or Enter on your keyboard to
continue.
C. Select Vessel Type. Available options include:
• Vertical Tank (Vertical)
• Horizontal cylinder with at ends (Hrzcy/Fl)
• Horizontal cylinder with dished ends (HrzCy/Ds)
• Horizontal cylinder with hemispherical ends
(Hrzcy/Hm)
• Spherical (Sphere)
• The default is Vertical. Press Enter and choose
the type of vessel.
D. Click on Write to Transmitter.
E. Move on to Range Values (URV & LRV) section of
menu.
3.7.3 Lower and Upper Range Values (LRV and URV)
Enter the LRV and URV to set the current (mA) window of
the vessel.
Figure 3-7
LRV & URV Conguration
from Menu screen
A. Edit LRV (Lower Range Value) to display the
output you want to see when the transmitter generates 4 mA current. The default LRV is 0 feet.
B. Edit URV (Upper Range Value) to display the
output you want to see when the transmitter generates 20 mA current. The default URV is 100 feet for the RCT Series.
C. Click on Write to Transmitter. Conguration is
now complete.
25
RCT 10, 11, 12, 13 Series
3.8 Calibration
There are two methods for calibrating the transmitter
using the PC software:
Point Calibration (menu button selection):
Uses the two known level points in the vessel for
calibration. The further apart the two points are for the calibration the better the accuracy of the overall measurement. Always initiate the point calibration process by selecting the Point Calibration button
on the PC menu screen and following the prompts
in the pop-up window.
Level Calibration (lower right of window):
See Figure 3-3
Uses capacitance values obtained from the
AMETEK Drexelbrook Service department (or a previous calibration or identical application) for the zero and span calibration data. Call 1-800-
527-6297. Please provide your DE purchase order
number, transmitter serial number, vessel and application data to the Service Engineer. Level calibration is done using the Level Calibration
data elds on the PC menu screen.
It is permissible or sometimes even recommended that
May use both Point & Level Calibrations
both methods be used in order to establish a calibration
standard. For example, if the vessel was already lled before the calibration was attempted and it is difcult or
impossible to lower the level to establish the second point, it would be best to use a calculated zero capacitance for
the low point and actual level for the high point. While
this wouldn’t be as accurate as two known level points, it will be reasonably accurate until an actual low point calibration can be established. The Service department will help in calculating high or low capacitance values.
Because calibration involves determining two known
points of capacitance, a span (or range) jumper provides an adjustment for the change in capacitance required to produce full scale current.
• The Range Span Jumper is located on the back of the transmitter chassis. See Fig. 3-9.
• Each Range Span position on the RCT Series advances the range in inches or feet to
approximately ve times the previous setting.
See Table 3-1
Figure 3-9
Range Span Jumpers
26
Conguration & Calibration
RCT Series Span Range Setting Chart
Sensing Element Length Vs. Span Position Number / Maximum pf
Jumper Position = 1 2 3 4 5 6 Maximum pF = 20 100 450 2000 10000 40000
System # Maximum Sensing Element Length In Feet
Conducting Liquids:
RCTX3XXXXX18 N/A N/A 5.4 20 N/A N/A RCTX3XXXXX74 N/A N/A 9.2 40 200 920 RCTX3XXXXX45 N/A N/A 1.2 5.3 20 N/A RCTX3XXXXX70 N/A N/A 1.4 6.25 31 140 RCTX3XXXXX40 N/A N/A 7.8 20 N/A N/A RCTX3XXXXX47 N/A N/A N/A N/A 4.1 14 RCTX3XXXXX22 N/A N/A 5.4 20 N/A N/A RCTX3XXXXX65 N/A N/A 5.4 20 N/A N/A RCTX3XXXXX72 N/A N/A N/A 4.2 21 95
Interface Application:
RCTX2XXXXX45 N/A N/A 1.2 5.3 20 N/A RCTX2XXXXX43 N/A N/A 1.0 4.8 20 N/A RCTX2XXXXX47 N/A N/A N/A N/A 4.0 14
Insulating K = 1.5-5:
RCTX2XXXXX22 N/A 2.7 12.5 20 N/A N/A RCTX2XXXXX18 N/A 8.3 20 N/A N/A N/A RCTX2XXXXX74 N/A 8.0 37.5 166 N/A N/A RCTX2XXXXX45 N/A 3.6 16.2 20 N/A N/A RCTX2XXXXX70 N/A 3.8 17.1 76 N/A N/A RCTX2XXXXX40 N/A 6.0 20 N/A N/A N/A RCTX2XXXXX43 N/A 5.0 20 N/A N/A N/A RCTX2XXXXX47 N/A 5.0 14 N/A N/A N/A
Table 3-1
Span Range Position for a number of common Sensing Elements
3.8.1 Point Calibration
The Point to Point method of calibration is the most
accurate way to calibrate the transmitter with two level points. The high or low level must be known and should be held steady for accurate calibration. They may be any two points at more than 10% apart, and need not be the
Figure 3-10
Point Calibration from Menu
The Point Calibration pop-up window is accessed by
screen
4mA or 20 mA points.
clicking on the menu "button" Point Calibration. Either a high point or a low point can be entered rst.
A. Type in current value as the high point of the two
point calibration.
B. Click on Hi Point or press Enter (or Tab) on the
27
RCT 10, 11, 12, 13 Series
3.8.1 Point Calibration (Continued)
keyboard. High point calibration is now complete.
C. Move level in vessel a minimum of 10%. D. Type in that value for Low Point of the two point
calibration.
E. Click on Low Point or press Enter (or Tab) on
keyboard. Low point calibration is now complete.
3.8.2 Level Calibration
Level calibration uses zero and span capacitance values
as the calibration data . These values can be obtained from the AMETEK Drexelbrook Service department (or from a previous calibration or identical application).
Figure 3-11
Level Calibration area of
Menu screen
Please be prepared when you call (1-800-527-6297) with
the purchase order number and the serial number of the transmitter.
A. Go to Level Calibration area of the menu. B. Enter Lower Level value. Press Tab or Enter. C. Enter Lower Capacitance value. Press Tab or
Enter.
D. Enter Upper Level value. Press Tab or Enter. E. Enter Upper Capacitance value. Press Tab or
Enter.
F. Click on Write to Transmitter.
3.8.3 Application Example
Example of an application using the PC software.
(Application Data)
• Vertical Tank
• No Damping
• Caustic or Acid Material in Tank Sensing Element: 700-5-54, Model Code: 74 (See Section 1.3)
• Span Range Switch factory set to 4
• Maximum Capacity of Vessel = 1200 gallons
• Maximum Size of Vessel = 20 feet
• 4 mA (LRV) = 0 gallons
• 20 mA (URV) = 1185 gallons [19.5 feet]
• Point Cal was done using two known level points:
• Lo Cal = 3 feet [selected level]
• Hi Cal = 16 feet [current level]
28
3.8.3 Application Example (Continued)
Conguration & Calibration
Figure 3-12
Application Example Diagram
Maximum Level 20 feet
URV = 1185 gallons
20 mA
*LRV may either reference the bottom of the vessel, bottom of the sensor, or an elevated point on the sensor.
[19.5 feet]
High Point 16 feet Hi Cal
Low Point 3 feet Lo Cal
LRV Alternate*
[0 feet]
LRV 4 mA Point
Maximum Capacity 1200 gallons
Figure 3-13
PC Software Menu Screen
View of Application Example
29
RCT 10, 11, 12, 13 Series
3.9 PC Status Messages
Status Message: SPAN TOO SMALL
Difference between URV and LRV is less than 10% of range.
Figure 3-13a
PC Software Menu Screen
View of Main Menu
Example: For 0 to 10 foot calibration points: LRV=3.0
feet and URV=3.8 feet. When calibration points are too
close together, overall accuracy of calibration is adversely affected. Action: The calibration points should be farther apart.
Calibration Status Message: RAISE SPAN JUMPER
Based on LRV, URV, and capacitance calibration data, the estimated 100% capacitance exceeds selected range by greater than 10%. Example: A unit in Range 4 (2000 pF) projects
maximum capacitance equal to 2500 pF. Error message is
displayed. Action: Raise Range jumper (Section 3.8, Table 3-1) to
position 5 for this example.
Calibration Status Message: LOWER SPAN JUMPER
Based on LRV, URV, and capacitance calibration data, the estimated 100% capacitance value is less than 10% of the maximum pF for the range below the selected range. Example: A unit in Range Span 4 (2000 pF) projects maximum capacitance equal to 400 pF. Action: Lower Range jumper (Section 3.8, Table 3-1) to
position 3 (Max. pF of 450 pF) for this example.
Real-time Status Message: UNDERRANGE
Present capacitance/milliampere value is less than -5% of
range. Examples: Center wire connection is broken. Sensing element is not operating. An elevated Zero is used and actual level is below 4mA point. Vessel has lost its RF ground reference. Action: Check connections and ground. Recalibrate if this level is in operational range of process.
Real-time Status Message: OVERRANGE
Present capacitance/ milliamp value over 105% of system
span. Examples: Actual level exceeds span point on sensing element. Cut in sensing element insulation or shorted coax. Action: Check sensing element and coax. Re-calibrate if this level is in operational range of process.
30
Conguration & Calibration
3.10 Set D/A Trim
D/A Trim is NOT a calibration! This is a pre
calibrated alignment to precision factory settings and is rarely in need of change. The procedure is intended only as a slight "meter" adjustment to a known external reference.
The Digital to Analog (D/A) Trim adjusts the transmitter
mA (current) output. Since the smart transmitter performs a digital to analog conversion, there may be a discrepancy in the 4-20 mA output loop as measured with a reliable external milliampere meter.
For example: perhaps after calibration you observe that
the tank is empty and a hand-held mA meter reads only
3.94 mA, while the Real Time View in the PC Menu shows
4.00 mA. By adjusting the D/A trim, you may digitally
manipulate the output current to equal 4.00. You may
Figure 3-14
Setting D/A Trim Menu Screen
To make these adjustments, click on D/A Trim on the PC
"Pop-Ups"
also wish to adjust the high end to 20.00 mA.
software Menu Screen and follow the pop-up window instructions:
3.11 Strapping Table
The strapping table is a 2-point to 21-point table
used by the RCT Series to cause the output current to
follow a specied relationship to the level. There are
certain strapping tables that are already built in to the transmitter software. These are: Linear (vertical
tank); Horizontal Tank with at ends; Horizontal Tank
with dished ends; Horizontal Tank with hemispherical
ends; and Spherical Tank. These predened tables are
automatically created by selections made with Vessel
Conguration assignments during Conguration
procedure in Section 3.7.2, and viewed by clicking the Strapping Table "button" on the Main menu.
If the output-to-level relationship is not dened by one
of these tables, you may create a table in the Strapping Table program. To create a non-linear relationship, you will need at least 3 points and may use as many as 21
points. A 21-point table will dene the relationship with
more accuracy. A common example for a simple table would be a Cone Bottom Vertical tank which would require 3 points—the bottom, straight-side break point,
and the top. On the other hand, an open channel ow application could benet from using all 21 available
points.
31
RCT 10, 11, 12, 13 Series
3.11 Strapping Table (Continued)
A. Plan your table by lling out table 3-2. You may
use the rst column which lists every 5% between 0 and 100%, or you may ll in your own values in
column 2.
B. Fill out column 3 with output values corresponding
to those listed in column 1 or 2.
C. "Click" on Strapping Table button to access table:
D. Enter the values you calculated into the screen
view presented.
E. "Click" on Write Strapping Table.
F. "Click" on Exit when completed.
Point Number Level Standard Level Optional Output Value
Preset Values Values In Selected % Level % Level Units
1. 0
2. 5
3. 10
4. 15
5. 20
6. 25
Table 3-2
RCT Series
Strapping Table
7. 30
8. 35
9. 40
10. 45
11. 50
12. 55
13. 60
14. 65
15. 70
16. 75
17. 80
18. 85
19. 90
20. 95
21. 100
Figure 3-15
Menu Screen Transforms to
Strapping Table
By clicking on Last Read Values, this view may also be
used to review existing strapping tables previously entered.
32
Conguration & Calibration
3.12 Digital Integral Meter Configuration
The optional Digital Integral Meter (DIM)(401-44-3) is
used for local display. It can be viewed through a glass viewport in the transmitter housing cover. See Figure
2-16
The meter can be congured to read any engineering
units, e.g. 4-20 mA, gallons, inches, feet, etc. Status messages are also displayed on the meter. Refer to
Figure 3-16
Congure Meter Pop-up from
To congure the meter, "Click" on Congure Meter in
Menu Screen; values relate to
LRV & URV
The meter is congured by:
Factory default settings are: Minimum Value = 0.00 Maximum Value = 100.00
Section 2.9 for meter installation.
menu screen for the pop-up;
• setting the minimum value equal to the value to be
displayed at the LRV and,
• setting the maximum value equal to the value to be
displayed at the URV.
To set the meter display range equal to calibration
range:
Minimum Value = LRV Maximum Value = URV
To set the meter display range equal to percent of
level:
Minimum Value = 0 Maximum Value = 100
When a smart transmitter is powered down or the ribbon
cable is disconnected, there is a 1 minute delay before the DIM begins to display upon return of power.
If the display becomes distorted:
• Remove power from the smart transmitter,
• Wait one minute,
• Reapply power to restart the meter.
3.13 Save/Print Entries
In addition to your own convenience, many regulatory
agencies are requiring a record of the values being used during certain processes. All of the values developed
in this conguration and calibration procedure may
be saved to be reloaded into another (or replacement) transmitter. All of the values may likewise be printed out
as hard copy, including the Serial Number, transmitter software version, Tag ID, Scratch Pad, Level and Vessel
Figure 3-17
Print Pop-up from Menu
Congurations, Level Calibration, all of the Real Time
View numbers, and all of the Strapping Table entries.
33
RCT 10, 11, 12, 13 Series
3.13 Save/Print Entries (Continued)
Pop-up screens come from selections in the FILE pull
down at the top left of the PC menu Screen.
Copies are saved in both .slt le and .txt les.
The .slt le will download into a transmitter through the
OPEN command. The text le may be printed out, or
reformatted.
PRINT command provides a pre-formatted hard copy.
3.14 Validation
More and more industries are requiring formal validation
of their processes for their customers as well as for various government regulatory agencies. The RCT Series Transmitter has this capability built in.
3.14.1 Validation Design Concept
Smart RF Continuous Level systems derive their input
information from a sensing element that provides a capacitance value to the RF Transmitter. The RF Transmitter output signal is derived from this capacitance value, based on the capacitance span of the transmitter during initial calibration.
If the RF Transmitter's minimum and maximum
capacitance values are known, and remain unchanged,
the effects of a specic capacitance value within this
range can be accurately predicted. If a known capacitance
(which can be NIST-traceable) within this range produces
repeatable results and the minimum and maximum values remain unchanged the RF Level system can be assumed to be operating correctly.
With a known capacitance input, the output signal
would not be repeatable if the calibration information is altered, or if the RF transmitter was not operating within
specications. Repeatable calibration information can be maintained through the use of the Save/Print capability
built into the RCT Series Transmitter.
3.14.2 Validation Procedures
Figure 3-18
Capacitance on Menu Screen
A. Drexelbrook Laptop software must be used.
Connect the laptop to the smart level transmitter signal loop to be validated and start the software according to the instructions provided at the beginning of this Section 3.
B. At the Main conguration screen observe the
Level Calibration, Lower Capacitance, and Upper Capacitance values and the Lower
34
3.14.2 Validation Procedures (Continued)
Level and Upper Level values. Select an NPO Capacitor (which can be NIST traceable, if desired
or required) that falls somewhere mid-range. Example: See Fig. 3-18. If Lower Capacitance
is 50pF and Upper Capacitance is 2000pF, that
corresponds to a Lower Level and Upper Level of 0-
10 feet. Select an NPO Capacitor of approximately 1000pF. [Drexelbrook 401-6-8 Capacitor Substitute Box may also be used; it is traceable to NIST].
C. Connect the NPO capacitor selected from the last
step to the Probe and Ground connections at the transmitter (with coaxial cable from sensing element disconnected).
D. Select Real Time View on the PC software Menu
Screen (F4 on your keyboard) See Fig. 3-19. The display should show the Capacitance as the
value of the NPO Capacitor (within the capacitors
tolerance), and the LEVEL should display close to the mid-range of the Lower and Upper Level from the Level Calibration eld. The Loop Current
Figure 3-19
Real Time View Pop-up from
Menu Screen
* Every capacitor manufactured will generate a slightly different capacitance value
within its specied tolerance. By
marking the capacitor and using only this capacitor for testing and validating the AMETEK Drexelbrook RCT Series Transmitter, the system should produce repeatable results
within transmitter specications.
and the Percentage will also reect the values that are generated by the NPO Capacitor. Add to
the scratch pad of the Menu Screen the value of
the NPO capacitor that you used. If desired, this information can be printed out for le or record
purposes. Mark or Tag this capacitor* to correspond
to this specic transmitter. Put the capacitor in
a safe location for use in subsequent testing and validation.
E. By placing the same exact capacitor* on the RF
transmitter's Probe and Ground terminals and observing the signal output generated by this
capacitor, it can be veried that the transmitter
is operating properly and that the calibration information is the same as during the initial set up.
Conguration & Calibration
35
RCT 10, 11, 12, 13 Series
3.14.3 Validation Results
See Fig. 3-20. If the information that is shown (or printed) matches the
initial readings within system specications, then it can be veried that the calibration and conguration is as originally set. It can also be veried that the transmitter’s
response falls within acceptable tolerances. The system
has passed validation tests. Using the Save/Print feature
built into the transmitter allows the ability to comply with the record-keeping needed for many processes by regulatory agencies.
AMETEK Drexelbrook
205 Keith Valley Road
Horsham, PA 19044
Telephone: 215-674-1234
FAX: 215-674-2731
Service: 800-527-6297
Tag-ID: LT 101 Serial Number: 1172 Scratch PAD: 1000 pf NPO validation capacitor Software Version: 3.1 Analog Loop Assign: LEVEL Range Position: 4 Damping Time: 0 sec. Type: 30
Level Configuration Vessel Configuration Level Calibration
Level Units: feet Vessel Units: gallons Lower Level: 0 ft Maximum Level: 10.00 ft Maximum Capacity: 1200.00 gal Upper Level: 10 ft Level Type: Standard Vessel Type: Vertical Lower Capacitance: 50 pf
LRV (4mA): 0.00 ft URV (20 mA): 10.00ft
Upper Capacitance: 2000 pf
Real Time View
Level: 10 ft Vessel: 600 gal Capacitance: 1000 pF Loop Current: 12.00 mA Percentage: 50 % Status: OK
Strapping Table
Vessel Type: Vertical Number of Points: 21
Level Volume
In Percent Out Percent
0.00 0.00 o.oo 0.00
1.00 5.00 60.00 5.00
2.00 10.00 120.00 10.00
3.00 15.00 180.00 15.00
4.00 20.00 240.00 20.00
5.00 25.00 300.00 25.00
6.00 30.00 360.00 30.00
7.00 35.00 420.00 35.00
8.00 40.00 480.00 40.00
9.00 45.00 540.00 45.00
10.00 50.00 600.00 50.00
11.00 55.00 660.00 55.00
12.00 60.00 720.00 60.00
13.00 65.00 780.00 65.00
14.00 70.00 840.00 70.00
15.00 75.00 900.00 75.00
16.00 80.00 960.00 80.00
17.00 85.00 1020.00 85.00
18.00 90.00 1080.00 90.00
19.00 95.00 1140.00 95.00
20.00 100.00 1200.00 100.00
Typical Printout of Transmitter Data
Figure 3-20
36
Conguration & Calibration
3.15 Calibration & Configuration via 401-44-3 Display/Keypad
EnterUp/Down Enter
Enter
View Lower Cap
View Lower Level
Enter
View Upper Level
Enter
Enter
View Upper Cap
Enter
Enter
Enter
View Min Value
Enter
Min Value
0 - 99900
Edit
Enter
Enter
Up/Down
Enter
View Max Value
Enter
Up/Down
Max Value
Up/Down
Enter Enter
Edit 0 - 99900
Enter
Enter
Up/Down
Lower Level
Enter
Edit 0 - 99900
Enter
Enter
Up/Down
Lower Level
Enter
Enter
Up/Down
View %Out Scale
Enter
Up/Down
Up/Down
Enter
Sc.PO (Scale Percent Out) See?
Enter
Enter
Edit 1.0 or 100.0
Enter
Edit
Up/Down
Enter
Edit
Up/Down
Enter
Up/Down
Up/Down
C.CAL (Cap Cal) See?
Up/Down
Enter
Enter
View Damping
Enter
Enter
dELy (Damping Time) See?
Up/Down
Up/Down
Enter
Enter
Edit 0 - 90
Enter
Edit
Up/Down
Enter
Enter
S.S.HH
Enter
Enter
Sr.Hr
Up/Down
CAP (Capacitance) View Capacitance
(Software Rev/Hardware Rev)
Enter
Enter
Yes
4.0?
20.0?
Value =
Value =
Enter
Enter
Enter
I.AdJ (DAC Trims)
Yes
No
Edit 19.0-21.0
Up/Down
Enter
Enter
No
Edit 3.5 - 4.5
Return
Up/Down
Lower Cap
Edit 0 - 99900
Enter
Lower Cap
Enter
View Type
Enter
Enter
tyPE (Type) See?
Up/Down Enter
Upper Level
Up/Down
Edit 0 - 99900
Enter
Up/Down Enter
Upper Level
Up/Down
Enter
Edit 00 or 30
Enter
Edit
Up/Down
Up/Down
Up/Down
Up/Down
Upper Cap
Enter Enter
Edit 0 - 99900
Enter
Up/Down
Upper Cap
Enter
View Vessel Type
Enter
Enter
StrP (Vessel Type) See?
Return
Enter
Return
Edit
Up/Down
Up/Down
Enter
Sphere
Standard
HCyl - Flat
HCyl - Dished
HCyl - Hemisphere
Enter
Edit
Up/Down
Escape Menu
Up/Down
Enter
Device
Restarts
Enter
Enter
Up/Down
L.FAC (Load Defaults) Sure?
rEt (Return to Main Menu)
Min Value
Enter
Up/Down
Max Value
Up/Down
Return
Enter
Up/Down
Edit
Up/Down Up/Down
Enter
Up/Down
CFg.d (Configure Display) See?
View LRV
Enter
Enter
Lrng (Lower Range Value) See?
Edit 0 - 99900
Enter
Edit
Up/Down
Up/Down
View URV
Edit 0 - 99900
Enter
Enter
Edit
Up/Down
Up/Down
Enter
Up/Down
Urng (Upper Range Value) See?
Up/Down
View Low Point
Edit 0 - 99900
Enter
Enter
Cal?
Up/Down
Up/Down
Enter
Up/Down
Up/Down
LOPt (Low Point Cal) See?
View High Point
Enter
Enter
HIPt (High Point Cal) See?
Edit 0 - 99900
Enter
Cal?
Up/Down
Up/Down
Enter
donE (Leave Menu)
Enter
Exit Main Menu
Up/Down
Up/Down
diAg (Diagnostic Menu)
37
SECTION 4
Conguration & Calibration
SECTION 4: CONFIGURATION AND CALIBRATION WITH
MODEL 275 WITH DREXELBROOK DEVICE DESCRIPTION
4.1 Drexelbrook Device Description
The Drexelbrook Device Description for the Rosemount
275 handheld calibrator makes it easy to calibrate a
RCT Series. The device description software is stored in the memory module (located in the back portion) of the calibrator.
To determine if your Model 275 Handheld calibrator
has the Drexelbrook Device Description loaded, do the following steps:
• Turn on the calibrator.
• From the top screen, push 1. Ofine.
• Push 1. New Conguration
• A Table of Contents is shown that lists all the
Manufacturers in alphabetical order.
• Drexelbrook is third on the list in the current
release.
• Select the Manufacturer (Drexelbrook) and a list of supported devices is displayed (RCT Series).
• Return to top screen, by backing up from arrow
keys.
The Memory Module with the device description can
be purchased as a direct replacement either from
Drexelbrook (401-700-25) or from the local Fisher-
Rosemount Service Center.
Appendix A describes conguration of the RCT Series
transmitter with a Rosemount 275 handheld calibrator
without a device description.
39
RCT 10, 11, 12, 13 Series
Typical Transmitter Loop
Nominal
24Vdc Supply
17-30 Vdc
ABC7 DEF8
GHI9
ABC7
DEF8
GHI9
ABC7
DEF8 GHI9
ABC7
DEF8 GHI9
ABC7
DEF8
GHI9
ABC7
DEF8
GHI9
ABC7
DEF8
GHI9
+
Resistance added if necessary to make total loop resistance at least 250 ohms during the communication process.
2-wire twisted
shielded pair
(recommended)
4.23
Optional loop
powered indicator
e.g. DLM4000 series
Calibrator or PC Modem may be connected anywhere on the transmitter side of the 250 ohm resistance. Voltage at the transmitter terminals must be at least 12 volts with 20 mA of loop current.
Other possible loop devices such as a setpoint controller.
Optional Safety Barrier HART® Compatible
4.2 Start-up
After the RCT Series transmitter is installed and loop
power is applied, per Section 2, do the following:
1. Connect the Rosemount Model 275 as shown in
Figure 4-1.
2. Turn on the Calibrator and look for the ONLINE screen to appear. ONLINE means that the 275
Calibrator has recognized the RCT Series and is
ready for Conguration and Calibration.
3. You must start the process by doing the
Conguration rst-- followed by Calibration. There are also instructions for conguring the Strapping Tables and for doing a D/A Trim to make the loop
output agree with a calibration standard for loop current.
Figure 4-1
Typical Transmitter Loop
40
4.3 Configuration
UNIV. III: <TAG>
LRV
NNNNN
UNIV. III: <TAG>
MAX LEVEL
NNNNN
UNIV. III: <TAG>
LEVEL UNITS
AAAAA
UNIV. III: <TAG>
LEVEL TYPE
AAAAA
<select from list>
XXXXX
XXXXX
Conguration & Calibration
UNIV. III: <TAG>
URV
NNNNN
XXXXX
UNIV. III: <TAG>
DAMP TIME
NNNNNs
XXXXX
UNIV. III: <TAG>
CURRENT ANALOG
AAAAA
UNIV. III: <TAG>
LOOP ASSIGN IS
SELECT ANALOG
LOOP ASSIGN
1. LEVEL
2. VESSEL
3. EXIT
UNIV. III: <TAG>
VESSEL UNITS
AAAAA
<select from list>
UNIV. III: <TAG>
VESSEL TYPE
AAAAA
<select from list>
UNIV. III: <TAG>
MAX CAPACITY
NNNNNu
XXXXX
Required only if weight
or volume readings are
desired.
UNIV. III: <TAG>
VESSEL CONFIG
1. VESSEL UNITS AAA
2. VESSEL TYPE AAA
3. MAX CAPACITY NNNu
4. DAMP TIME NNNs
5. LRV NNNu
6. URV NNNu
7. CHG ANLG LOOP ASGN
UNIV. III: <TAG>
ENTER MINIMUM METER VALUE
NNNNNNNNNN
XXX XXXX XXX
ENTER METER SPAN VALUE
NNNNNNNNNN
XXX XXXX XXX
UNIV. III: <TAG>
UNIV. III: <TAG>
LEVEL CONFIG
1. LEVEL TYPE AAA
2. LEVEL UNITS AAA
3. MAX LEVEL NNNu
4. LRV NNNu
5. URV NNNu
6. DAMP TIME NNNs
7. RANGE POSITION N
8. CHG ANLG LOOP ASGN
Required only if weight
or volume readings are
desired.
UNIV. III: <TAG>
ONLINE
1. LEVEL NNNu
2. VESSEL NNNu
3. CAP NNNpF
4. REF. CAP. NNNpF
5. AO NNNmA
6. % OF RNG NNN%
UNIV. III: <TAG>
CONFIGURATION
1. LEVEL CONFIG
2. VESSEL CONFIG
3. METER CONFIG
4. STRAPPING TABLE
5. CALIBRATION
UNIV. III: <TAG>
BASIC SETUP
1. TAG
2. LEVEL UNITS
3. VESSEL UNITS
4. DEVICE INFORMATION
5. VESSEL TYPE
6. DAMP TIME
UNIV. III: <TAG>
DETAILED SETUP
1. HART OUTPUT
2. SIGNAL CONDITION
3. CALIBRATION
DEVICE INFORMATION
UNIV. III: <TAG>
REVIEW
1. DISTRIBUTOR
2. MODEL
MANUFACTURER
TE
6. DESCRIPTOR
3. SERIAL #
4. TAG
5. DA
LEVEL UNITS
VESSEL UNITS
VESSEL TYPE
LRU
URV
7. SCRATCHPAD
8. DEV ID
9. REVISION #'S
DAMP TIME
RANGE POSITION
ANLG LOOP
POLL ADDR
UNIV. III: <TAG>
HART OUTPUT
TYPE
DISTRIBUTOR
MODEL
1. POLL ADDR
2. NUM REQ PREAMS
UNIV. III: <TAG>
SERIAL #
TAG
DATE
DESCRIPTOR
SCRATCH PAD
DEV ID
SIGNAL CONDITION
1. CHG ANLG LOOP ASGN
2. LRV NNNu
3. URV NNNu
4. DAMP TIME NNNs
5. % OF RNG NNN%
UNIVERSAL REV
FLD DEV REV
SOFTWARE REV
HARDWARE REV
Configuration
Start
UNIV. III: <TAG>
1. PROCESS VARIABLES
UNIV. III: <TAG>
ONLINE
1. DEVICE SETUP
2. LEVEL NNNu
2. CONFIGURATION MENU
3. BASIC SETUP
4. DETAILED SETUP
5. REVIEW
3. VESSEL NNNu
4. CAP NNNpF
5. REF. CAP. NNNpF
6. AO NNNmA
7. % OF RNG NNN%
8. LRV NNNu
9. URV NNNu
41
UNIV. III: <TAG>
ANLG LOOP AAA
STATUS
STATUS
1. UNIV III STATUS
RANGE POSITION N
UNIV. III: <TAG>
UNIV. III STATUS
SPAN TOO SMALL AAAAA
DECREMENT SPAN AAAAA
INCREMENT SPAN AAAAA
OVERRANGE AAAAA
UNDERRANGE AAAAA
LEGEND
AAA - ALPHABETICAL DISPLAY
NNN - NUMERIC DISPLAY
XXX - FIELD MAY BE EDITTED
u - UNITS
pF - PICOFARADS
MUST PERFORM THIS FUNCTION
mA - MILLIAMPS
s - SECONDS
% - PERCENT
OPTIONAL FUNCTION
OTHER FUNCTION
RCT 10, 11, 12, 13 Series
4.3 Configuration (Continued)
Following is the keystroke sequence for Conguration
using the Model 275 Calibrator.
Select Device Setup. Select Conguration Menu. Select Level Cong.
Select Level Type—edit Level Type—return to Level
Cong screen.
Select Level Units—edit Level Units—return to Level
Cong screen.
Select Max Level—edit Max Level—return to Level
Cong screen.
Select LRV—edit LRV—return to Level Cong screen.
Select URV—edit URV—return to Level Cong screen.
Select Damp Time—edit Damp Time—return to Level
Cong screen.
Select Chg Anlg Loop Assign—edit Current Loop
Assign. If current loop assign is Level and is correct, go to
next screen and select Exit. Proceed to 4.3 Calibration.
If Vessel conguration is to be selected, choose Vessel, go
back to Cong screen and select Vessel Cong. Edit all values as done for Level Cong. Select Exit and proceed
to 4.4 Calibration.
4.4 Calibration
There are two methods of calibrating the RCT Series
transmitter: Point Calibration or Capacitance Calibration.
Point calibration uses the actual level in your vessel for
calibration. The further apart the two points are for the calibration, then the better the accuracy of the overall measurement.
Capacitance calibration uses values obtained from the
Drexelbrook Service department (or a previous calibration or identical application) for the zero and span calibration
data. Call 1-800-527-6297 for assistance. Please provide
the purchase order number, transmitter serial number and application information to the Service Engineer.
It is permissible or sometimes even recommended that
both methods be used in order to establish a calibration
standard. For example, if the vessel was already lled before the calibration was attempted and it is difcult or
impossible to lower the level to establish the second point,
42
Conguration & Calibration
4.4 Calibration (Continued)
it would be best to use a calculated zero capacitance for
the low point and actual level for the high point. While
this wouldn’t be as accurate as two known level points, it will be reasonably accurate until an actual low point can be established. The Service department will help in calculating high or low capacitance values.
4.4.1 Point Calibration
Following is the keystroke sequence for Point Calibration
using the Model 275 Calibrator.
Select Device Setup. Select Conguration Menu. Select Calibration.
Select Point Cal—select either Low Point Cal or High
Point Cal depending on whether you plan to raise or lower the level for your second point—edit value to agree
with the present actual level and return to Point Cal
screen.
Exit—Calibration is complete.
To D/A Trim
UNIV.III: <Tag> POINT CAL
1. LOW POINT CAL
2. HIGH POINT CAL
UNIV.III: <Tag>
From configuration
screen
CALIBRATION
1. STRAPPING TABLE
2. D/A TRIM
3. POINT CAL
4. CAPACITANCE CAL
5. TRUE LEVEL
6. CHNG ANLG LOOP ASGN
UNIV.III: <Tag> CAPACITANCE CAL
1. LOWER LEV NNNu
2. LOWER CAP NNNpF
3. UPPER LEV NNNu
4. UPPER CAP NNNpF
UNIV.III: <Tag> ENTER LOW POINT CAL NNNu XXXu
UNIV.III: <Tag> ENTER HI POINT CAL NNNu XXXu
UNIV.III: <Tag> LOWER LEV NNNu XXXu
UNIV.III: <Tag> LOWER CAP NNNpF XXXpF
UNIV.III: <Tag> UPPER LEV NNNu XXXu
43
UNIV.III: <Tag> UPPER CAP NNNpF XXXpF
RCT 10, 11, 12, 13 Series
4.4.2 Capacitance Calibration
Capacitance calibration uses zero and span capacitance
values as the calibration data. These values can be obtained from the Drexelbrook Service department (or a previous calibration or identical application). Call 1-800-
527-6297 for assistance. Please provide your DE purchase
order number, transmitter serial number, and application information to the Service Engineer.
Following is the keystroke sequence for Capacitance
Calibration using the Model 275 Calibrator.
Select Device Setup. Select Conguration Menu. Select Calibration.
Select Capacitance Cal—select either Lower Level or
Upper Level depending on whether the next value will be higher or lower for the second point—edit capacitance value and level as a pair—return to Capacitance Cal screen.
Exit—Calibration is complete.
4.4.3 D/A Trim
Refer to the D/A Trim diagram for the D/A Trim sequence
and Strapping Table conguration.
44
D/A Trim Diagram
UNIV.III: <Tag>
POINT NUMBER: X
UNIV.III: <Tag>
POINT NUMBER: X
UNIV.III: <Tag>
ENTER POINT
ENTER OUTPUT
CAPACITY VALUE:
XXX
XXX
ENTER INPUT
LEVEL VALUE:
XXX
XXX
UNIV.III: <Tag>
NUMBERXX
UNIV.III: <Tag>
POINT NUMBER: X
ENTER OUTPUT
CAPACITY VALUE:
XXX
XXX
POINT NUMBER X
ENTER INPUT
LEVEL VALUE:
XXX
XXX
UNIV.III: <Tag>
POINT NUMBER: X
ENTER OUTPUT AS A
PERCENT OF MAXIMUM
CAPACITY:
XXX
XXX
UNIV.III: <Tag>
POINT NUMBER: X
ENTER INPUT AS A
UNIV.III: <Tag>
% OF MAXIMUM LEVEL:
XXX
XXX
Conguration & Calibration
UNIV.III: <Tag>
RETURNING FLD
DEV TO ORIGINAL
OUTPUT
UNIV.III: <Tag>
FLD DEV OUTPUT
20mA EQUAL TO
REFERENCE
METER?
1. YES 2. NO
POINT NUMBER: X
ENTER OUTPUT AS A
% OF MAXIMUM
CAPACITY:
XXX
XXX
UNIV.III: <Tag>
ENTER METER
VALUE
XXXX
ENTER POINT
NUMBERXX
UNIV.III: <Tag>
UNIV.III: <Tag>
SELECT EDIT
METHOD
1. EDIT 1 POINT
2. EDIT ALL POINTS
3. EXIT
UNIV.III: <Tag>
UNIV.III: <Tag>
ENTER NUMBER OF
POINTS FOR TABLEXX
UNIV.III: <Tag
D/A Trim is NOT a calibration! This is a pre-calibrated alignment to precision factory
UNIV.III: <Tag>
POINT NUMBER: X
ENTER INTPUT AS A
% OF MAXIMUM
LEVEL:
XXX
XXX
SELECT EDIT
METHOD
1. EDIT 1 POINT
2. EDIT ALL
POINTS
SELECT EDIT METHOD
!.EDIT ACTUAL POINTS
2.EDIT POINTS IN %
3. EXIT
UNIV.III: <Tag>
TABLE EDIT/REVIEW:
1.CHANGE # POINTS
2. EDIT POINTS
3. REVIEW POINTS
4. SEND POINTS
UNIV.III: <Tag>
POINT #X:
5. EXIT
INPUT: XXX (UNIT)
% INPUT: XXX%
OUTPUT: XXX
(UNITS)
% OUTPUT: XXX%
UNIV.III: <Tag>
POINT NUMBER X HAS
BEEN MODIFIED. WAIT
AS POINT IS WRITTEN
TO TRANSMITTER.
UNIV.III: <Tag>
SETTING VESSEL TYPE
TO USER-DEFINED
D/A Trim
UNIV.III: <Tag>
SETTING FLD DEV
OUTPUT TO 20mA
UNIV.III: <Tag>
FLD DEV OUTPUT 4mA
EQUAL TO REFERENCE
METER?
1. YES 2. NO
UNIV.III: <Tag>
ENTER METER
VALUE
XXXX
UNIV.III: <Tag>
SETTING FLD DEV
OUTOUT TO 4mA
settings and is rarely in need of change. The procedure is intended only as a slight "meter"
adjustment to a known external reference.
Strapping Table
From configuration
screen
UNIV.III: <Tag>
45
CALIBRATION
1. STRAPPING TABLE
2. D/A TRIM
3. POINT CAL
4. CAPACITANCE CAL
5. TRUE LEVEL
6. CHNG ANLG LOOP
ASGN
UNIV.III: <Tag>
CONNECT
REFERENCE
METER
RCT 10, 11, 12, 13 Series
4.4.4 Strapping Table
The strapping table is a 2-point to 21-point table used by
the RCT Series to cause output current to follow specied
relationship to level. There are certain strapping tables that are already built in to the transmitter software. These are: Linear (vertical tank); Horizontal Tank with
at ends; Horizontal Tank with dished ends; Horizontal
Tank with hemispherical ends; and Spherical Tank.
These predened tables are automatically created by selections made with Vessel Cong assignment during Conguration procedure in Section 4.3.
If output-to-level relationship is not dened by one of
these tables, you may create a table in Strapping Table program. To create a non-linear relationship, you will need at least 3 points and may use as many as 21 points.
A 21-point table will dene relationship to about a 0.1%
accuracy. Common example for a simple table would be Cone Bottom Vertical tank which would require 3 points—bottom, straight-side break point, and top. Open
channel ow application, however, could benet from
using all 21 available points.
• Plan your table by lling out form below. You may
use rst column which lists every 5% between 0 and 100%, or you may ll in your own values in
column 2.
• Fill out column 3 with output values corresponding
to those listed in column 1 or 2.
Point Number Level Standard Level Optional Output Value
Preset Values Values In Selected % Level % Level Units
1. 0
2. 5
3. 10
4. 15
5. 20
6. 25
7. 30
8. 35
9. 40
10. 45
11. 50
12. 55
13. 60
14. 65
15. 70
16. 75
17. 80
18. 85
19. 90
20. 95
21. 100
46
Conguration & Calibration
SECTION 5: Configuration & Calibration of RCT Series with
Model 275 Communicator without a Device Description
The Model 275 Communicator can be used to enter and
download conguration/ calibration values to the RCT
Series transmitter.
5.1 Range/Span Control
Denitions:
• LRV - Lower Range Value of control loop.
• LSL - Lower Sensor Limit corresponding to
minimum level point.
• PV - Process Variable; level value from transmitter.
• SV - Secondary Variable; could be volume or
weight-based on PV.
• URV - Upper Range Value of control loop.
• USL - Upper Sensor Limit corresponding to
maximum level point.
The rst step in conguration is to identify the span
range position of the instrument. There is a jumper on the back of the electronics chassis: Range Span See Fig. 3-9. The Range Span provides continuous adjustment of the change in capacitance required to produce full scale current. Each Range Span position advances the
range in inches or feet to approximately ve times the
previous setting. Table A-1 gives the range span position for a number of common sensing elements and range of measurement.
Probe Length Vs. Span Position Number / Maximum Pf
Table 5-1
Range/Span
Setting Chart
Jumper Position: 1 2 3 4 5 6
Max Pf 20 100 450 2000 10000 40000
System Maximum Probe Length In Feet
Conducting Liquids
RCTX3XXXXX18 N/A N/A 5.4 20 N/A N/A RCTX3XXXXX74 N/A N/A 9.2 40 200 920 RCTX3XXXXX45 N/A N/A 1.2 5.3 20 N/A RCTX3XXXXX70 N/A N/A 1.4 6.25 31 140 RCTX3XXXXX40 N/A N/A 7.8 20 N/A N/A RCTX3XXXXX47 N/A N/A N/A N/A 4.1 14 RCTX3XXXXX22 N/A N/A 5.4 20 N/A N/A RCTX3XXXXX65 N/A N/A 5.4 20 N/A N/A RCTX3XXXXX72 N/A N/A N/A 4.2 21 95
Interface Application
RCTX2XXXXX45 N/A N/A 1.2 5.3 20 N/A RCTX2XXXXX43 N/A N/A 1.0 4.8 20 N/A RCTX2XXXXX47 N/A N/A N/A N/A 4.0 14
Insulating K=1.5-5
RCTX2XXXXX22 N/A 2.7 12.5 20 N/A N/A RCTX2XXXXX18 N/A 8.3 20 N/A N/A N/A RCTX2XXXXX74 N/A 8.0 37.5 16 N/A N/A RCTX2XXXXX45 N/A 3.6 16.2 20 N/A N/A RCTX2XXXXX70 N/A 3.8 17.1 76 N/A N/A RCTX2XXXXX40 N/A 6.0 20 N/A N/A N/A RCTX2XXXXX43 N/A 5.0 20 N/A N/A N/A RCTX2XXXXX47 N/A 5.0 14 N/A N/A N/A
47
RCT 10, 11, 12, 13 Series
5.2 Basic Rules & Conventions of HART Configuration Software (Model 275)
To become familiar with the operation of the Model 275
Communicator, it is best to review the Operating Manual that comes with the unit. This paragraph reviews some
of the basic characteristics of the Model 275 conguration/
calibration software. The following pages show the decision tree menus.
Arrow Keys: Use the arrow keys to move through the
software menus. An arrow at the beginning of the menu item indicates the next progressive step.
Alphanumeric Keys: Use the alphanumeric keys to
enter data. Data elds are characterized by a blinking
cursor.
Function Keys: The function keys, F1 through F4,
indicate the specic actions that are available to complete
the software function. The action words that appear above each key change as you move through the menus and select the software task.
Heart Symbol: When a appears in the upper right
corner of the screen, communication is occurring between the communicator and the transmitter.
5.3 Tag ID
To name or identify the transmitter, use the 8-character
Tag ID. The Tag ID is entered or changed in the following menus.
a) select the ONLINE MAIN menu
b) 1 DEVICE SETUP ⇒ ⇓ ⇓ c) 3 BASIC SETUP d) 1 TAG e) change (use alphanumeric keys) Tag ID
f) F4 ENTER g) F2 SEND
h) F4 OK i) F4 OK or:
a) select the ONLINE MAIN menu
b) 1 DEVICE SETUP⇒ ⇓ ⇓ c) 3 BASIC SETUP⇒ ⇓ ⇓ ⇓ d) 4 DEVICE INFORMATION⇒ ⇓ ⇓ ⇓ e) 4 TAG f) change (use alphanumeric keys) Tag ID
g) F4 ENTER h) F2 SEND
i) F4 OK j) F4 OK k) F3 HOME
48
5.4 Set Up Procedures
Conguration & Calibration
49
RCT 10, 11, 12, 13 Series
5.4 Set Up Procedures (Continued)
50
5.4 Set Up Procedures (Continued)
Conguration & Calibration
51
RCT 10, 11, 12, 13 Series
5.5 Reading Input and Output
Within the Model 275 conguration, there are several
different menus that allow you to view the Process Variable (PV) and the Analog Output (AO1).
View PV Input:
a) select the ONLINE MAIN menu
b) 1 DEVICE SETUP ⇒ ⇓ ⇓ ⇓ c) 4 DETAILED SETUP d) 1 SENSORS e) 1 PV
f) view Process Variable
g) EXIT F4 h) HOME F3
or:
a) select the ONLINE MAIN menu b) ⇒2 PV c) view Process Variable
d) F4 EXIT e) 2 PV f) 3 PV AO g) view Analog Output h) F4 EXIT
or:
a) select the ONLINE MAIN menu
b) 1 DEVICE SETUP c) 1 PROCESS VARIABLES d) 1 SNSR
e) view Process Varible
f) F4 EXIT g) 1 SNSR h) 2 AI % RANGE
i) view PV%
j) F4 EXIT k) 2 AI % RANGE l) 3 AO1 m) view Analog Output n) EXIT F4 o) HOME F3
52
Conguration & Calibration
5.6 Calibration Using Actual Tank Level
When calibrating using a handheld without a device
description, tank level must be moved. Calibration is a two-step process. A low point calibration and high point calibration are required. Calibration points do not have to be an empty tank or a full tank. (example: low point performed at 20% and high point performed at 80 % will work). You also may perform high point cal before low point cal.
From main screen go to:
1 Device setup 2 Diag/Service 3 Calibration 4 Enter values
You now have four choices:
1 PV LRV 2 PV URV 3 PV USL 4 PV LSL
Choose PV LRV if your vessel currently has a low level. Choose PV URV if your vessel currently has a high level.
Edit the displayed value to equal the actual tank level,
then:
ENTER SEND
OK OK
1 Apply values OK
You now have two choices:
1 4mA 2 20 mA
Choose 20mA if performing a high point calibration. Choose 4mA if performing a low point calibration.
Continue: OK
ENTER
3 EXIT OK HOME
Observe on the home screen that your PV URV or PV
This completes the rst calibration point of the two-step
LRV value has been changed by the calibration procedure. You must manually change these values back. Re-enter
your desired PV URV and PV LRV at this time.
process. For the second point, change the level in your tank and repeat the entire above procedure.
53
RCT 10, 11, 12, 13 Series
5.7 Bench Calibration (if needed)
When performing a bench calibration tank level will be
simulated using a Drexelbrook 401-6-8 C-box. Calibration
is a two-step process. A low point calibration and high point calibration are required
Hook up a Drexelbrook c-Box in place of the sensing
element. Dial up the capacitance value that corresponds
to an empty tank. For specic information on how to wire
and use the C-box calibrator see the instructions that came with the calibrator.
From main screen go to:
1 Device setup
2 Diag/Service 3 Calibration 4 Enter values
You now have four choices
1 PV LRV 2 PV URV 3 PV USL 4 PV LSL
Choose PV LRV if performing a low point calibration. Choose PV URV if performing a high point calibration.
Edit the displayed value to equal the actual tank
level,then:
ENTER SEND
OK OK
1 Apply values OK
You now have two choices:
1 4mA 2 20mA
Choose 20mA if performing a high point calibration. Choose 4 mA if performing a low point calibration.
Continue: OK
ENTER
3 EXIT OK HOME
Observe on the home screen that your PV URV or PV
This completes the rst calibration point of the two-step
LRV value has been changed by the calibration procedure. You must manually change these values back. Re-enter
your desired PV URV and PV LRV at this time.
process. Change the value on the C-box to correspond to
the Picofarads of a full tank and repeat the entire above
procedure for a high point calibration.
54
Conguration & Calibration
5.7.1 Bench Calibration Information Sheet
Company
City State Customer P.O. Number Item DE# Tag No.
Filled out by: Date Phone Fax
Material being Measured - Fill out any known information
Name of Material: Level Measurement Interface Measurement Dielectric Constant: (K) Conductivity: (g) Other:
Installation Details
NPT Thread Mount Flange Mount- if flange mount
B = inches
E = A = inches H = inches D = inches H = inches IL = inches
Vessel Shape Vertical Cylinder
Horizontal Cylinder Other
B
E
Unless specified otherwise, calibration values of zero and span capacitances will be based on 4-20 mA being over the entire range of 'H'. If other
A
H
URV
IL
values are desired specify LRV and URV in inches.
LRV = inches
LRV
URV = inches
D
Calculations by Drexelbrook
LRV ________ Calculated Zero Capacitance ________ pF URV ________ Calculated Span Capacitance ________ pF Calculated by: ____________________ Date _________ Phone 800-527-6297
Fax 215-443-5117
Appx-A-smh.pmd Page 0 of 1 Created 07/31/1997 by ELS Revised 08/07/2002 12:21 PM
55
RCT 10, 11, 12, 13 Series
5.8 Point Calibration
Equipment Required:
• RCT Series HART® Smart Transmitter
• Model 275 HART® Calibrator
• 24V Power source
• Approximately 250 ohm total loop resistance
• Two known process levels applied to the probe
This procedure uses an example of a point calibration for
full-scale (20mA) = 35 ft. and zero (4mA) = 1.5 ft.
5.8.1 Fine Tuning Calibration
When a known level is available that is closer to an
The procedure SET UP ZERO AND FULL SCALE LIMITS must be performed to set up the
vessel zero and full-scale limits
ENTER THE UPPER CALIBRATION POINT and/ or
sometime prior to performing
ENTER THE LOWER CALIBRATION POINT with the
a point calibration. This informs the RCT Series of the tank level limits and only
The order of execution between the upper and lower
needs to be performed once for a given installation.
endpoint than a previous calibration point, it may be used as a new calibration point to increase accuracy. This is done in the following steps:
new known level applied.
calibration procedures does not matter. They can be done at different times.
5.8.2 Selecting Engineering Units
The examples in this procedure use the default units of
feet. For other units, use the following procedure:
A. Connect the handheld communicator to the 4-20
mA loop per Figure 4-1.
B. Power the loop and proceed with following steps.
Do See Function Comments
1/0 Online (Generic) Power On 1 Device Setup Select Device Setup 3 Basic Setup Select Basic Setup 4 PV Snsr unit Select PV Units
and Available Units
Use the arrow keys to scroll through the available units and press [ENTER] when the desired unit is displayed.
[ENTER] Basic Setup Select highlighted unit
Home Online (Generic) Desired units are now selected.
56
Conguration & Calibration
5.9 Handheld Calibrator Error Messages
Error Message: Applied Process Too High
The applied process at 100% exceeds the capacitance
range setting by at least 5% of span.
Example: URV = 10 ft., Applied Process = 12 ft. will
generate this error.
Action: The Span Select Jumper Must Be Raised. This allows the process output at 100% to be within the
range of the span range capacitance.
Error Message: Span Too Small
The difference between the URV and LRV is less than
10% of the range.
Example: For 0 to 10 ft. calibration points: LRV = 3.0
ft., URV = 3.8 ft. will generate this error.
Action: The Calibration Points Should Be Farther Apart. When the calibration points are too close together, the
overall accuracy of the calibration is adversely affected.
Error Message: Upper Range Value Too High
The process capacitance at 100% exceeds the present
capacitance range jumper setting.
Example: For 0 to 10 ft. calibration: span select jumper
= 100 pF, point cal. LRV = 0 ft. @ 10 pF, URV = 7 ft. @ 90 pF generates this error because the process would be 124 pF @ 10 ft.
Action: The Span Select Jumper Must Be Raised. This allows the process output at 100% to be within the
range of the span range capacitance.
NOTE: Respond to the prompt which is displayed immediately
after the error message that reads “Restore Device Value?” with <N>.
Error Message: Upper Range Value Too Low
The Full scale projected from the entered URV and LRV
is less than 5% of the present capacitance range jumper
setting.
Example: For 0 to 50 ft. calibration: span select jumper
= 1000 pF, LRV = 0 ft. @ 10 pF, URV = 40 ft. @ 80 pF generates this error.
Action: The Span Select Jumper Must Be Lowered.
This allows the URV-LRV span to be greater than 5% of
NOTE: Respond to the prompt which is displayed immediately
the capacitance range jumper setting.
after the error message that reads “Restore Device Value?” with <N>.
57
SECTION 6
Troubleshooting
SECTION 6: TROUBLESHOOTING
Problem/Symptom Tests in order of probability Reference Section(s) Comments
Rosemount 268 or 275 Calibrator gives error message that no device was found
Rosemount 275 Calibrator gives error message that device could not be identified
Can’t communicate with transmitter using Drexelbrook PC Software
0 mA output all the time (no measurable output current at any time)
More than 20 mA output all the time (output current always exceeds 20 mA)
Output drifts (output accuracy varies slowly over time…e.g. hours or days)
Output erratic - (output jumps around noticeably in terms of seconds or minutes)
Output intermittent (output jumps quickly usually between >0mA and some "on scale" value
Inaccurate readings (Level readings are incorrect compared to actual known level)
Reading does not change with level
Output goes in opposite direction from level change
Application-related Problems
Product Bridging When process material fills what was originally airspace between the sensor and a nozzle
Corrosion of metal parts TFE and stainless steel in the sensor's pressure seal have widely different coefficients
Check calibrator connections Check for 250 resistance (min.) Check voltage at transmitter Check transmitter
Check calibrator connections Check for 250resistance (min.) Check voltage at transmitter Check transmitter
Check calibrator connections Check for 250resistance (min.) Check voltage at transmitter Check transmitter Try another modem
Check voltage at transmitter Check polarity of loop Test Transmitter
Check for moisture in head of sensor Test Sensing Element Test Transmitter Check Calibration
Test transmitter without sensing element (drift test) Verify proper sensing element ground reference
Test Transmitter Check process level Check for Static Discharge Check for radio interference
Check Signal Loop Connections 6.7
Check calibration Test transmitter Check method of comparison
Check cables Check sensing element Test Transmitter
Check calibration Section(s) 3, 4, and 5 Probable high point
Comments
or the vessel, it no longer behaves like a coating. It measures like actual level. Contact Drexelbrook.
of expansion that sometimes permit pressure leaks to occur. Re-torqueing the packing assembly can usually fix the problem. Contact the Service department for the proper torque values and procedure.
6.2 and 6.3
6.6
6.2 and 6.3
6.6
6.2 and 6.5
6.6
6.2 (6.3, 6.4, or 6.5)
6.6
6.7
6.6 Section(s) 3, 4, and 5
6.6
Fig. 2-6
6.6
6.9
6.10
6.8
Section(s) 3, 4, and 5
6.6
6.8
6.7
6.6
Often a result of loop connection problems
Often a result of loop connection problems
Often a result of loop connection problems
Probable loop problem Faulty connection in loop
Erratic readings often show actual process conditions. Look for bubbles or stratification, etc.
Intermittent Loop Connection
Have you verified actual level? (At times even sight gauges can be misleading.)
Be sure that level is really changing. Possible plugged or unvented stilling well.
cal/low point cal reversal or inverted interface application.
Table 6-1 Problem / Symptom Chart
59
RCT 10, 11, 12, 13 Series
6.1 Identifying a Problem/Symptom
Use Table 6-1 as a guide to nd and correct a problem
when it occurs. Most problems are not related to transmitter failure. It is important to be methodical when tracking down a problem. If you experience a problem that you cannot solve using this guide, call the
Drexelbrook 24-hour Service Hot line at 1-800-527-6297 or 215-674-1234. You may also E-mail us at the Internet
address: drexelbrook.service@ametek.com. Further
service information may be found at our World Wide Web
address www.drexelbrook.com.
When you contact us, be prepared to give the service
person as much information as you can about the model numbers, application requirements, and the materials being measured. At the end of this section, a form is available to organize the information that will help us
resolve the problem. Prior to your call, a copy of the
completed form can be faxed directly to the Service
department at (215) 443-5117.
6.1.2 RCT Integral Digital Meter Display (401-0044-003) Error Codes
Error Message Description Probable Cause Corrective Action Err 1
Err 2 Err 3
Err4
EEPROM Failure / Checksum Error
Not Used
Analog to Digital Converter Response Failure
Probe Value too High Probe Resistance to
Transmitter Failure Restore Factory Defaults
See Section 3.15
Transmitter Error Contact Factory Service
1-800-527-6297
Check for Electrical Short
Ground too Low
See Section 6.7 & 6.8
6.2 Troubleshooting Loop Connection
Specic transmitter loop connections will vary from
installation to installation but in general will be connected in a similar manner to typical transmitter loop in Figure 4-1. When troubleshooting the loop connection, verify the following items.
• Loop devices are wired in series.
• There is at least 250 ohms total loop resistance.
• There is at least 12 Vdc available for the transmitter when a loop current of 20 mA is owing.
60
Troubleshooting
Is loop
current between
3.7 and 22 mA?
Is there a
minmum of 250
ohms loop
resistance?
1. Check calibrator connections to loop per loop drawing
Figure 4-1.
2. Check for "noise" and ripple on loop with oscilloscope. Maximum noise level per HART Foundation is 1.2 mV rms (500 - 10 kHz). Maximum ripple (47 - 125 hZ) specification is .2V p-p. Line noise can sometimes be traced to things like motor speed controller wiring in close proximity with transmitter. Noise can sometimes be overcome by increasing loop resistance thereby increasing calibrator signal. Noise effects can also be reduced by connecting calibrator directly at transmitter. Generally noise is only a problem when calibrator is communicating with transmitter and does not affect normal operation of transmitter.
3. Check voltage at transmitter it should be at least 12 V when 20 mA is flowing in loop.
4. If wiring is correct, go to
Section 6.6
to test transmitter.
Calibrator does
not
communicate
with
transmitter
Yes
Yes
Check:
1. If current greater than 22 mA, disconnect at sensing element and re-check. If current returns to normal, check sensing element using checkout procedure in
Section 5.7
. If current does not return to normal,
test transmitter with procedure in
Section 5.6
.
2. If loop current is 0 mA, check polarity of wiring at transmitter. If OK check for open loop.
3. If loop current is between 0 and 3 mA transmitter is likely bad. Test with procedure in
Section 5.6
.
4. Check voltage at transmitter, it should be at least 12 volts when 20 mA is flowing in loop.
5. Disconnect the three probe cable connections at transmitter and retry. If it now communicates, check cable and probe as described in
Sections 5.7 & 5.8
.
1. Check Power Supply source resistance. The 250 ohms may be built in as with the Drexelbrook 401­ 500 series or most DCS inputs.
2. If not, add enough resistance for loop to be at least 250 ohms.
Start
No
No
6.3 Rosemount Model 268 or 275 Calibrator cannot identify or find device
If the Rosemount 268 or 275 calibrator (generic) gives
error message that no device description was found, use
the following owchart to troubleshoot.
6.4 Rosemount Model 275 Calibrator with device description cannot identify or find device
This condition may be the result of trying to calibrate
a transmitter with software less than version 3.0 (transmitters shipped prior to approximately January
1997). There are two options you can use to identify the
device and delete the error message.
Option 1 - Congure the transmitter as a "generic"
device per the instructions in this manual.
Option 2 - Contact the Drexelbrook Service
department (1-800-527-6297) about upgrading your
transmitter to the latest software.
61
RCT 10, 11, 12, 13 Series
Is loop current
between 3.7 and
22 mA?
Is there a
minmum of 250
ohms loop
resistance?
Are you connected
to a DCS?
Is It Polling?
Make it stop polling or power transmitter from a separate source - then re-check operation.
Yes
Yes
Yes
Check:
1. Is correct COM Port selected at startup?
2. Is there any software running that would re-direct COM Port such as Windows, mouse drivers, terminal emulation software, or TSR's. (This can be tested by booting from Drexelbrook Calibration software in the A: drive)
3. Possible bad RS-232 cable or defective modem.
4. Check modem connections to loop per loop drawing on
Figure 3-1
.
5. Go to
Section 5.6
to test transmitter.
Modem does
not
communicate
with
transmitter.
No
No
Start
No
Are you using
the Drexelbrook
supplied modem?
Check:
1. Is modem non-isolated from ground and/or is your laptop plugged into AC power.? If so you may have ground problem. Contact Service department.
2. Some laptops don't provide enough voltage to drive modem correctly from
COM Port. Check with modem supplier or try a
different laptop.
3. Checkout "Yes" response tests listed below.
No
Yes
Check:
1. If current is greater than 22 mA disconnect at sensing element and re-check. If current returns to normal, check sensing element using checkout procedure in
Section 5.7
. If current does not return to normal,
test transmitter with procedure in
Section 5.6
.
2. If loop current is 0 mA, check polarity of loop at transmitter. If it is OK, check for open loop.
3. If loop current is between 0 and 3 mA transmitter is likely bad. Test with procedure in
Section 5.6
.
4. Check voltage at transmitter it should be at least 12 volts when 20 mA is flowing in the loop
5. Disconnect the three probe cable connections at transmitter and retry. If it now communicates, check cable and probe as described in
Sections 5.7 & 5.8
.
No
Yes
1. Check Power Supply and loop source resistance (the 250 ohms may be built in as with the Drexelbrook 401-500 series or most DCS inputs).
2. If not there add enough resistance for loop to be at least 250 ohms.
1. Check calibrator connections to loop per loop drawing
Figure 4-1
.
2. Check for "noise" and ripple on loop with oscilloscope. Maximum noise level per HART Foundation is 1.2 mV rms (500 - 10 kHz). Maximum ripple (47 - 125 Hz) specification is .2V p-p. Line noise can sometimes be traced to things like motor speed controller wiring in close proximity with transmitter. Noise can sometimes be overcome by increasing loop resistance thereby increasing calibrator signal. Noise effects can also be reduced by connecting calibrator directly at transmitter. Generally noise is only a problem when calibrator is communicating with transmitter and does not affect normal operation of transmitter.
3. Check voltage at transmitter. It should be at least 12 volts when 20 mA is flowing in the loop.
4. If wiring is OK, go to
Section 6.6
to test transmitter.
6.5 RCT Series transmitter does not communicate with Drexelbrook PC software
62
Troubleshooting
6.6 Troubleshooting Transmitter
To troubleshoot the transmitter, use one of the following
Always Install to Local
Codes / Requirements /
Directives as Mandated
by the Authority Having
Jurisdiction
Install Per
420-0004-203-ID
TRANSMITTER TEST - Using a PC or 275 Handheld
WARNING: Substitution of Components May Impair Intrinsic Safety.
WARNING: To Prevent Ignition of Flammable or Combustible Atmospheres, disconnect Power Before Servicing.
tests, depending on the device used with your calibration.
Some of the following tests require the use of high quality
xed capacitors in the picofarad ranges or a Drexelbrook calibrator box (C-box 401-6-81). Contact the Drexelbrook
Service department if you need a C-box. Fixed
temperature stable capacitors (NPO types) can often be
found at many electronic supply houses.
with device Description
1. Determine the span range currently selected on the electronic unit. See Figure 3-9
2. Using the MAX pF values listed in Figure 6-3, select a capacitance value near the midpoint of the MAX pF range. For example, position #4 has a MAX pF range of 2000 pF. Select a 1000 pF capacitance for this test.
3. Remove all three connections of the coaxial cable at the transmitter's probe terminals.
CAP
1 2 3 4 5 6
RANGE SPAN
GROUND PROBEPAD SHD
Figure 6-1
SPARE
4. Place capacitor on transmitter from PROBE to GND (ground) terminals See Figure 6-1.
5a. Using a PC - Using the real-time view (F4), verify
that the displayed capacitance value is within the value and tolerance printed on the test capacitor. If using a Drexelbrook C-box, be sure to add the standing capacitance of the box (10pF low range,
20 pF normal range).
5b. Using a Rosemount 275 handheld with device
description - Viewing the HOME screen, verify that the displayed capacitance value is within the value and tolerance printed on the test capacitor. If using a Drexelbrook C-box, be sure to add the standing capacitance of the box (10pF low range, 20pF normal range).
6. If the displayed capacitance value is within the
stated tolerance, the unit is working. If the displayed capacitance value is not accurate, call
1-800-527-6297.
63
RCT 10, 11, 12, 13 Series
6.6 Troubleshooting Transmitter (Continued)
TRANSMITTER TEST - Using a 268 or 275 Handheld
WITHOUT a Device Description (Generic Mode)
Transmitter
1. Visually verify the span range of the electronic unit.
1 2 3 4 5 6
RANGE SPAN
SPARE
2. Using the C-box, adjust the capacitance until 4 mA is achieved. Record value. See Figure 6-2
GROUND PROBEPAD SHD
C-Box
Figure 6-2
3. Adjust C-box capacitance until 20 mA is achieved. Record value.
4. Using Figure 6-3, verify that the capacitance value recorded at 20 mA is less than the MAX pF value
for the Span Range Position of the electronic unit.
5. Verify that the loop conguration is Level Cong (signal output linear to level vs. Vessel Cong).
6. Subtract pF value recorded for 4 mA from the pF
value recorded for 20 mA. Divide this number in half.
Example: 4 mA = 120 pF 20 mA = 800 pF
800 pF – 120 pF = 680 pF 680 pF ÷ 2 = 340 pF
7. Adjust the capacitance box (C-box) to the
number gured in step 6. The signal should read approximately 50%.
8. If the display reads 50%, the unit is working
properly. If the display is not accurate, contact the
Service department at 1-800-527-6297 for further
assistance.
6.6.1 Transmitter Drift Test
If symptoms point toward calibration drift, it is important
SPAN JUMPER POSITION RANGE 1 2 3 4 5 6
MAX pF 20 100 450 2000 10000 40000
Figure 6-3
Span Range Capacitance Values
to determine if the apparent drift is coming from the transmitter, the sensing element, or the application of the equipment. The following test determines if the transmitter is stable. In most cases, no drift will be found in the transmitter.
64
6.6.1 Transmitter Drift Test (Continued)
1. Remove coaxial cable from the transmitter
terminals.
Transmitter
2. Without changing any data stored in the
GND
SERIAL NO. MODEL NO.
YEAR OF MANUFACTURE
V +M -
V -
mA Meter
on Record
transmitter, connect a Drexelbrook capacitance
substitution box (401-6-8) or an NPO test capacitor from the probe terminal to the GND terminal on
the transmitter. (Select a capacitance value that produces between 4 and 20 mA of loop current.)
3. Observe the loop current over a 12-hour period to
conrm the stability of the unit. If the readings
remain stable for this period, then the problem
24 VDC
is not in the transmitter. If the loop current has changed more than 1% during the test period, then
the unit is defective. Please contact the Service
Figure 6-4
department for further instructions regarding repair or replacement.
6.7 Troubleshooting Sensing Element
Troubleshooting
Troubleshooting sensing element requires use of
an analog ohmmeter. Digital meters do not properly
measure resistance for the purpose of this test. An analog ohmmeter typically provides more current when measuring resistance, which is required to detect a pinhole or crack in the sensing element insulation. In addition, digital meters frequently give erroneous results due to a battery-like effect when dissimilar metals contact conductive liquids.
CAUTION: Sensing element is intrinsically
safe. Therefore, when using this product, it is recommended that all service activity comply with appropriate guidelines.
Remove sensing element from vessel to a safe area.
Test outlined in steps 1 and 2 can be performed
in a metal test vessel lled with high conductivity
water. Depending on locality, tap water may not be suitable. If not, a spoonful of table salt may work.
In the following tests, if it is not possible to raise or lower
level in vessel, sensing element may be suspended in a
metal pipe or other container that is lled with conductive
water (see above note) and connected to grounded sensing element condulet. If container is not metallic, then a ground wire or rod is needed to be placed into the water and referenced to sensing element condulet or mounting devices.
65
RCT 10, 11, 12, 13 Series
6.7 Troubleshooting Sensing Element (Continued)
Testing the Sensing Element - Step 1 (Figure 6-5)
With the material below the sensor, and the coaxial
cable disconnected at the sensing element, measure the resistance from the sensing element center connector to ground connector (or condulet). The ohmmeter should
be set to R x 10000 scale. The reading should be innite
(open circuit). Readings of less than one meg-ohm indicate excessive electrical leakage, probably due to product leakage or condensation in the packing seal or condulet. Contact the Service department for recommended repairs.
Figure 6-5
Sensing Element Testing,
Material Below
the Sensing Element
Testing the Sensing Element - Step 2 (Figure 6-6)
Raise the level in the vessel to cover as much of the sensor as possible. Repeat the measurement made in step 1. Readings of 1 meg-ohm or less indicate a pinhole or crack in the sensing element insulation. Failed insulation is
not eld repairable. Consult the Service department for
further assistance.
Figure 6-6
Sensing Element Testing,
Material Covering
the Sensing Element
6.8 Troubleshooting Coaxial Cable
If there is water or other conductive material in the
conduit it can change the electrical properties of the coax cable and cause the system to perform poorly. Moisture in the conduit may not be detected by the following test.
66
6.8 Troubleshooting Coaxial Cable (Continued)
The only sure way is to inspect the coax and associated conduit for trapped water.
1. Disconnect all three wires of the coaxial cable at
the electronic unit.
2. Disconnect all wires at the sensing element end of
the coax.
3. Using an ohmmeter, measure between two of the
coaxial cable conductors. Note any reading. Repeat
for all three conductors. All readings should show
an open circuit, (innite resistance).
Troubleshooting
OHMMETER
8
0
4. Check for continuity of each conductor. Short out
two of the coaxial cable conductors and measure these two conductors at the other end. A reading close to 0 ohms should be shown.
CENTER - GROUND
CENTER - SHIELD SHIELD - GROUND
SHORTED WIRES SHOULD READ 0 OHMS
CHECK FOR CONTINUITY
"0" "0" "0"
OHMS OHMS OHMS
SHORT OUT
TWO CONDUCTORS
67
RCT 10, 11, 12, 13 Series
6.9 Static Electricity
Static electricity can cause the 4-20 mA output to appear
to jump around in an erratic fashion with a time period of few seconds. Applications that are prone to static electricity include insulating liquids that may be agitated or pumped and granulars that may be air-conveyed at high rates of speed. Conductive liquids and conductive granulars tend not to generate static electricity. In addition to causing erratic readings, static electricity can cause instrument failure. If you ever get a static shock from a sensing element, you need spark protection. (See section 2.5)
Drexelbrook normally supplies static electricity discharge
devices (spark protectors) with its sensing elements that are going to be used in these types of applications. If you need a spark protector, contact the Drexelbrook Service department.
6.10 Radio Frequency Interference
All Drexelbrook transmitters have a signicant amount
of RFI protection built in. There are situations, however,
where the standard protection is inadequate. RFI lters
are available to provide additional protection for both
the sensor and the 4-20 mA loop from unusually difcult sources of interference. Proper grounding and careful
attention to installation practices can usually make them unnecessary. Some recommended installation practices include:
1. Use twisted shielded cable for the 4-20 mA loop.
2. Use grounded metal conduit for all entrances to the transmitter housing.
3. Ground the transmitter housing to a good earth
ground.
4. Use concentric shield sensors in non-metallic vessels.
If RFI continues to be a problem, contact the Drexelbrook
service department for the proper lters and assistance.
6.11 Factory Assistance
AMETEK Drexelbrook can answer any questions about
your level measurement system.
For Technical Support: 1-800-527-6297
If you require assistance and attempts to locate the
All other inquiries: Call Customer Service at
1-800-553-9092 (US and Canada) , or + 215-674-1234
(International).
problem have failed:
68
Troubleshooting
6.11 Factory Assistance (Continued)
• Contact your local Drexelbrook representative,
• For Technical Assistance call toll-free:
1-800-527-6297 (US and Canada) or + 215-674-1234 (International),
• FAX: + 215-443-5117,
• E-mail: drexelbrook.service@ametek.com
Please provide the following information:
• Instrument Model Number
• Sensing Element Model Number and Length
• Original Purchase Order
• Material being measured
• Temperature
• Pressure
• Agitation
• Brief description of the problem
• Checkout procedures that have failed
6.12 Field Service
Trained eld servicemen are available on a time-plus-
expense basis to assist in start-ups, diagnosing difcult
application problems, or in-plant training of personnel. Contact the service department for further details.
6.13 Customer Training
Periodically, AMETEK Drexelbrook instrument training
seminars for customers are held at the factory. These sessions are guided by Drexelbrook engineers and specialists, and provide detailed information on all aspects of level measurement, including theory and practice of instrument operation. For more information about these valuable workshops, write to AMETEK
Drexelbrook, attention: Communications/ Training Group, or call direct + 215-674-1234.
6.14 Return Equipment
Any equipment being returned for repair or credit
must be pre-approved by the factory.
In many applications, sensing elements are exposed to
hazardous materials.
69
RCT 10, 11, 12, 13 Series
6.14 Return Equipment (Continued)
• OSHA mandates that our employees be informed
and protected from hazardous chemicals.
• Material Safety Data Sheets (MSDS) listing the
hazardous materials to which the sensing element has been exposed MUST accompany any repair.
• It is your responsibility to fully disclose all
chemicals and decontaminate the sensing element.
To obtain a return authorization (RA#), contact the
Service department at 1-800-527-6297 (US and Canada) or + 215-674-1234 (International).
Please provide the following information:
• Model Number of Return Equipment
• Serial Number
• Original Purchase Order Number
• Process Materials to which equipment has been
exposed
• MSDS sheets for any hazardous materials
• Billing Address
• Shipping Address
• Purchase Order No. for Repairs
Please include a purchase order even if the repair is under
warranty. If repair is covered under warranty, you will not be
charged.
Ship equipment freight prepaid to: AMETEK Drexelbrook
205 Keith Valley Road Horsham, PA 19044-1499
COD shipments will not be accepted.
70
AMETEK Drexelbrook™
Universal III Mark II Troubleshooting Guide
Service Department (800) 527-6297 FAX(215) 443-5117
Phone #
6.15 RCT Series Troubleshooting Guide
AMETEK Drexelbrook
RCT Series Troubleshooting Guide
Service Department (800) 527-6297 FAX (215) 443-5117
Service Dept. Contact
Troubleshooting
Customer Name Company
Fax #
Purchase Order # DE #
Electronic Unit Model # Serial # Range Jumper Position
Sensing Element Model # Serial # Insertion Length Mounting
Process Material Temp. Press. Other
Provide as much of the following information as possible. All of the information is available from the Drexelbrook Calibration Software, or from a Rosemount Model 275 with Drexelbrook Device Description (DD) installed. Information with an asterisk is available from a Rosemount 268 or 275 in the Generic mode.
AMETEK Drexelbrook
*Tag ID
*Scratch Pad
Analog Loop Assign
*Damping Time
Level Configuration Vessel Configuration
Level Units
Maximum Level
Level Type
*LRV (4mA)
*URV (20mA)
HART® Protocol Software Version
*Serial Number
Software Version
Range Position
Type (00/30)
Vessel Units
Maximum Capacity
Vessel Type
City/State
Vessel Sketch
True Level Reference
Lower Reference Cap.
Upper Reference Cap.
Press F4 For Real-Time View
Level
Vessel
Capacitance
Reference
Loop Current
Percentage
Status
Capacitance Calibration
Lower Level
Lower Capacitance
Upper Level
Upper Capacitance Status:
Detailed description of problem:
Show principal tank dimensions, including vessel construction, mounting
location, nozzle, LRV, URV, present level, etc.
71
SECTION 7
SECTION 7: SPECIFICATIONS
7.1 Transmitter Specifications
Power Requirement
• 12 to 30 VDC
• Minimum of 12 VDC at 20 mA
Input Range
• 409-T: 1.0 to 40,000 pF
Output Range
• 4-20 mA
Accuracy
• ± 0.25% of range. Accuracy includes the combined
effects of linearity, hysteresis, and repeatability. It refers to the transmitter only and is measured
at reference conditions of 25 degrees C ±1°, 10­55% R.H. and 24 ±1.2 Vdc, using an admittance
standard (applied to the transmitter sensor terminals) in place of the sensor.
Specications
Load Resistance
• Maximum Load Resistance = 750 ohms
• Minimum Load Resistance
With HART® Communications = 250 ohms Without HART® Communications = 0 ohms
Temperature Effect
• ±1% of range per 50°F (30°C).
Supply Voltage Effect
• < 0.1% from 12 to 30 VDC.
Effect of Load Resistance
• < 0.1% for full resistance range at 24 VDC supply.
Response to Step Change
• < 1 second standard (to 90% of nal value); 0-90
seconds available with delay.
Fail-Safe
• Low-Level Fail-Safe (LLFS) standard.
Also called direct-acting because current increases as the level increases.
• High-level Fail-Safe (HLFS).
Also called reverse-acting because current decreases as level increases.
There Are No Devices That Are Absolutely “Fail-Safe".
Fail-safe means that in the event of the most probable failures, the instruments will fail safely. Probable failures include things such as loss of power and transistor and component failures. If your application needs absolute fail-safe, a back­up instrument should be installed.
Ambient Temperature
• -40°F to +167°F (-40°C to 75°C)
73
RCT 10, 11, 12, 13 Series
7.1 Transmitter Specifications (Continued)
Calibration Adjustments
• Range Span, 6 positions (back panel)
Lowest Permitted Resistance (bare sensing element to ground) causing 5% error in each
model:
• 600 ohms - 409-TX0, 2
• 100K ohms - 409-TX1, 3
Intrinsic Safety
• Sensing element and cable: Designed to be intrinsically safe for Class I Groups A, B, C and D; *Class II Groups E, F, and G, Class III, Div. 1.
Electronics and signal wires: Intrinsically safe for
Class I Groups A, B, C, and D, Class II Groups E, F and G Div. 1 when powered by an intrinsically safe
power supply. EEx ia Zone 0.
* When provided in the aluminum enclosure
(RCT Series)
Nonincendive Equipment
• Class I Div. 2, Groups A, B, C, and D; Class II, Div 2, Groups E, F, and G, Class III Div 2
Sensing Element Cable Length
• 100 feet maximum.
7.2 Coaxial Cable Specifications
General Purpose 380-XXX-12
• .51" (13mm) OD at largest point, 160°F (70°C) temperature limit.
Composite Cable 380-XXX-18
(rst 10 feet high temperature)
• .62" (16mm) OD at largest point, 450°F (230°C) temperature limit for rst 10 feet.
• 160°F (70°C) temperature limit for remainder.
CE Mark Triaxial Cable 380-XXX-520
• .62" (16mm) OD at largest point,
-55°F to 176°F (-55°C to 80°C) temperature limit.
7.3 Enclosure Specifications
• The RCT transmitter and remote sensor enclosures
are constructed of aluminum and are rated both
NEMA 4X and IP65. Conduit entries are 3/4 NPT
with M20 adapters available The enclosures are powder coated to provide additional corrosion resistance.
74
7.4 Approvals Available
7.4.1 ATEX Approvals - Install Per 420-0004-203-ID
Integral Systems
Specications
205 KEITH VALLEY ROAD HORSHAM, PA 19044
Amb Temp -40˚C...+75˚C Ui = 30V Ii = 140 mA Pi = 1W Ci = 0.001 uF Li = 0mH
12-30 VDC 4-20mA T135˚C
INSTALL PER 420-004-203-ID
205 KEITH VALLEY ROAD HORSHAM, PA 19044
MODEL NO. 409-TXXX
SERIES LEVEL CONTROL
II 1 G EEx ia IIC T4
0344
KEMA NO. 03 ATEX1249 X
Ui : 30V Ii : 140 mA Pi : 1W Ci : 0.001uF Li : 0mH
INSTALL PER 420-0004-203-ID
270-0101-812
RCT Series
Level Control
0344
II 1 GD EEx ia IIC T4 T135˚C
KEMA NO. 03 ATEX1249 X
SERIAL NO. ACA-XXXXXX
ENCLOSURE TYPE IP65
270-0101-804
Amb Temp:
-40˚C...+75˚C
12-30VDC 4-20mA
IP650344
II 1 GD EEx ia IIC T4
Remote Systems
IP650344
II 1 GD EEx ia IIC T4
Temperature Class Process Temperature T6 85° C T5 100° C T4 135° C
T3 200° C
Chassis Only
II 1 G EEx ia IIC T4
For indoor locations enclosure must be minimum IP20 rated.
For outdoor locations enclosure must be minimum IP44 rated.
A higher degree may be required for harsh environments. See Note 1
RCT SERIES REMOTE SENSING ELEMENT ENCLOSURE
YEAR OF MANUFACTURE - SEE INSIDE TRANSMITTER ENCLOSURE
II 1 G EEx ia IIC T3...T6, T135˚C
BELONGS TO KEMA 03 ATEX1249X
0344
Amb Temp -40˚C...+75˚C Ui = 30V Ii = 140 mA Pi = 1W Ci = 0.001 uF Li = 0mH
270-0101-847
Remote Sensor
Note 1: Always Install to Local Codes / Requirements / Directives as Mandated by the Authority
Having Jurisdiction
Note 2: For applications in explosive atmospheres caused by air / dust mixtures, a degree of
ingress protection of ot least IP 60 according to EN 60529 will only be achieved when certified cable and conduit entries are used providing a degree of ingress protection of at least IP65 according to EN60529
Note 3: The aluminum enclosure must be protected from mechanical friction and impact that
could cause ignition capable sparks.
WARNING: Substitution of Components May Impair Intrinsic Safety.
WARNING: To Prevent Ignition of Flammable or Combustible Atmospheres,
disconnect Power Before Servicing
75
RCT 10, 11, 12, 13 Series
7.4.2 CSA Approvals - Install Per 420-0004-225-CD
Intrinsically Safe
Class I, Div. 1, Groups A, B, C and D; Class II, Div. 1, Group E, F and G; Class III
Class I, Zone 0: Ex ia IIC
Class I, Div. 2, Groups A, B, C and D; Class II, Div. 2, Group E, F and G; Class III
Class I, Zone 2: Ex na IIC
For Ambient temperatures above 70º C, use wiring rated 75º C or higher.
7.4.3 FM Approvals - Install Per 420-0004-224-CD
Intrinsically Safe:Class I, II, III, Div. 1, Groups A, B, C,
D, E, F, G Nonincendive: Class I, Div. 2, Groups A, B, C, D Special Protection: Class II, Div. 2 , Groups F, G Dust-Ingnitionproof: Class I, II, Div. 1, Groups E, F, G
76
Normal Maintenance
SECTION 8: NORMAL MAINTENANCE
8.1 Viewport Cleaning
The viewport (if supplied) is made of Borosilicate glass
and can be cleaned with any common glass cleaning
product (e.g.: Windex™, Isopropyl alcohol, etc.) that is suitable for the Class and Division rating of the specic
system installation.
77
APPENDIX A
APPENDIX A: APPROVAL DRAWINGS
A.1 ATEX APPROVAL DRAWINGS
Appendix A
A-1
RCT 10, 11, 12, 13 Series
A.1 ATEX APPROVAL DRAWINGS (Continued)
A-2
A.1 ATEX APPROVAL DRAWINGS (Continued)
Appendix A
A-3
RCT 10, 11, 12, 13 Series
A.1 ATEX APPROVAL DRAWINGS (Continued)
A-4
A.1 ATEX APPROVAL DRAWINGS (Continued)
Appendix A
A-5
RCT 10, 11, 12, 13 Series
A.1 ATEX APPROVAL DRAWINGS (Continued)
A-6
A.1 ATEX APPROVAL DRAWINGS (Continued)
Appendix A
A-7
RCT 10, 11, 12, 13 Series
A.1 ATEX APPROVAL DRAWINGS (Continued)
A-8
A.1 ATEX APPROVAL DRAWINGS (Continued)
Appendix A
A-9
RCT 10, 11, 12, 13 Series
A.1 ATEX APPROVAL DRAWINGS (Continued)
A-10
A.1 ATEX APPROVAL DRAWINGS (Continued)
Appendix A
A-11
RCT 10, 11, 12, 13 Series
A.2 CSA APPROVAL DRAWINGS
A-12
A.2 CSA APPROVAL DRAWINGS (Continued)
Appendix A
A-13
RCT 10, 11, 12, 13 Series
A.2 CSA APPROVAL DRAWINGS (Continued)
A-14
Loading...