Metek JOFRA STS-150 A User Manual

...because calibration is
a matter of confidence
User Manual
JOFRA STS-150 A
JOFRA STS-150 A
Copyright 2012 AMETEK Denmark A/S
List of contents
1.0 General information ...................... 4
2.0 Safety instructions ........................ 5
3.0 Introduction .................................... 7
4.0 Functionality .................................. 8
4.1 Functional description ................. 8
4.2 Connections ................................ 8
4.3 Serial number ............................ 10
5.0 Operation ..................................... 11
5.1 Operation area .......................... 11
6.0 Maintenance ................................. 12
7.0 Technical specifications ............ 13
128166 01 2012-11-12 3
1.0 General information
This manual is only effective for the following products:
JOFRA STS-150 A 912 (90°) JOFRA STS-150 A 915 (90°) JOFRA STS-150 A 935 (90°) JOFRA STS-150 A 966 (90°)
The products are manufactured by:
AMETEK Denmark A/S
Gydevang 32-34
3450 Allerød - Denmark
TEL: +45 48 16 80 00
FAX: +45 48 16 80 80
4 2012-11-12 128166 01
2.0 Safety instructions
Read this manual carefully before using the sensor!
In order to avoid any personal injuries and/or damage to the sensor all safety instructions and warnings must be observed.
Warning
Do not use in hazardous area. Handle carefully. Never exceed temperature range
Caution…
When measuring temperature in
The probe must always be protected
The probe must never be exposed to
128166 01 2012-11-12 5
fluids (e.g. at re-calibration) the enclosed protection tube must be used.
against any mechanical damage.
mechanical shock effects.
Avoid thermal shock  Any bending of the probe may cause
permanent damage
Never use power or tools to place
the probe.
6 2012-11-12 128166 01
3.0 Introduction
The JOFRA STS-150 A probes are specially designed for fast and traceable calibration and temperature measuring with your JOFRA equipment and are ready for use.
Please read this manual carefully before use, to obtain maximum value of your calibration system.
Warning
Do not use in hazardous area. Handle carefully. Never exceed temperature range
128166 01 2012-11-12 7
4.0 Functionality
4.1 Functional description
The sensors can be used for measuring temperatures in the range :
-90°C to 125°C / -130°F to 257°F (912)  -25°C to 155°C / -13°F to 311°F (915)  0°C to 350°C / 32°F to 662°F (935)  0°C to 660°C / 32°F to 1220°F (966)
The JOFRA STS-150 A probes may be supplied with certificates for a limited temperature range.
The resistance of the JOFRA STS-150 A probe is converted to temperature according to ITS-90 (calculated coefficients specif ic for the probe is stated on the certificate).
4.2 Connections
The pin-layout is as follows: Pin 1 : RTD I-
Pin 2 : RTD S­Pin 3 : RTD S+ Pin 4 : RTD I+ Pin 5 : Memory GND
8 2012-11-12 128166 01
Pin 6 : Memory I/O The figure below is shown from the connector side of the probe connector.
5
6
4
1
2
3
128166 01 2012-11-12 9
4.3 Serial number
The serial number is placed on the probe as shown on the figure below:
Serial number
10 2012-11-12 128166 01
5.0 Operation
5.1 Operation area
All the probes are intended for use in areas, which meet the following operating conditions:
Probe connector and cable : -20°C to 70°C (-4°F to 158°F) Storage temp. : -20°C to 70°C (-4°F to 158°F) Humidity : 0% to 90% RH Protection class : IP 50
Warning
Do not use in hazardous areas.
128166 01 2012-11-12 11
6.0 Maintenance
The probe does not require specific maintenan ce before or after use. The user may carry out the following procedure himself:
Cleaning the probe : Use alcohol or water and a soft cloth
Caution…
The probe must always be protected
against any mechanical damage.
The probe must never be exposed to
mechanical shock effects.
Avoid thermal shock  Any bending of the probe may cause
permanent damage
.
12 2012-11-12 128166 01
7.0 Technical specifications
Sensor type : Platinum sensor Pt100. STS-150 A = 0.00385
Probe length : (912) : 210/193 mm (90° angle)
(8.27/7.60 in) (915) : 180/163 mm (90° angle) (7.09/6.42 in) (935) : 165/148 mm (90° angle) (6.50/5.83 in) (966) : 201/183 mm (90° angle) (7.91/7.20 in)
Temperature range : (912) : -90°C to 125°C / -130°F to
257°F (915) : -25°C to 155°C / -13°F to 311°F (935) : 0°C to 350°C / 32°F to 662°F (966) : 0°C to 660° / 32°F to 1220°F
Accuracy : Repeatability : 0.004°C (0.007°F) Hysteresis
1)
: 0.01°C @ 0°C
(0.02°F@32°F)
128166 01 2012-11-12 13
Stability2) : typ. 0.016°C @ 0°C (0.029°F@32°F)
Self heating effect : 0.06°C/mW / 0.11°F/mW Diameter : OD4 mm Immersion depth :
(912/915/935)
: 60 mm (2.36 in)
(966) : 100 mm (3.94 in)
Media compatibility : INCONEL 600
1)When used in the range -90°C to 125°C / -130°F to 257°F (STS­150 A 912), –25°C to 155°C / -13°F to 311°F (STS-150 A 915
to 350°C / 32°F to 662°F (STS-150 A 935) or 0°C to 660°C / 32°F to 1220°F (STS-150 A/B 966).
2)Stability when exposed to maximum temperature for 100 hours. Stability will depend on actual use of the probe.
), 0°C
Response time : (912/915/935)
(966) :
Recommended meas. current : 1 mA
14 2012-11-12 128166 01
: (50%)
(90%)(50%)
(90%)
= 7 sec. =18 sec.
= 8 sec. =26 sec.
Connection : (912/915/935)
: Cable with REDEL connector
(966) : LEMO plug with build in memory is standard
Certificate: If the STS-150 A reference probe is supplied with a
certificate, the calibration is carried out as recommended below according to the ITS 90 temperature scale.
The STS-150 A type 912 probe is as standard calibrated in the range : –90°C to 125°C (-130°F to 257°F).
The STS-150 A type 915 probe is as standard calibrated in the range : –25°C to 155°C (-13°F to 311°F).
The STS-150 A type 935 probe is as standard calibrated in the range : 0°C to 350°C (32°F to 662°F)
128166 01 2012-11-12 15
The STS-150 A type 966 probe is as standard calibrated in the range : 0°C to 660°C (32°F to 1220°F)
It is recommended to calibrate in minimum 3 – 6 calibration points (depending on the temperature range) above 0°C and in minimum 2 calibration points beneath 0°C.
16 2012-11-12 128166 01
UK
AMETEK Calibration Instruments
Tel +44 (0)1243 833 302 jofra@ametek.co.uk
France
AMETEK S.A.S.
Tel +33 (0)1 30 68 89 40 general.lloyd-instruments@ametek.fr
Germany
AMETEK GmbH
Tel +49 (0)2159 9136 510 info.mct-de@ametek.de
Denmark
AMETEK Denmark
Tel +45 4816 8000 jofra@ametek.com
USA
AMETEK Mansfield & Green
Tel +1 (800) 527 9999 cal.info@ametek.com
India
AMETEK Instruments India Pvt Ltd.
Tel +91 22 2836 4750 jofra@ametek.com
Singapore
AMETEK Singapore Pte Ltd
Tel +65 6484 2388 jofra@ametek.com
China
AMETEK Inc. - Shanghai
Tel +86 21 5868 5111
AMETEK Inc. - Beijing
Tel +86 10 8526 2111
AMETEK Inc. - Guangzhou
Tel +86 20 8363 4768 jofra.sales@ametek.com.cn
www.jofra.com
Information in this document is subject to change without notice. ©2012 by AMETEK, Inc., www.ametek.com. All rights reserved.
AMETEK Calibration Instruments
is one of the world’s leading manufacturers and developers
of calibration instruments for temperature, pressure and
process signals as well as for temperature sensors both
from a commercial and a technological point of view.
JOFRA Temperature Instruments
Portable precision thermometers. Dry-block and liquid
bath calibrators: 5 series, with more than 25 models and
temperature ranges from -90° to 1205°C / -130° to 2200°F.
All featuring speed, portability, accuracy and advanced
documenting functions with JOFRACAL calibration software.
JOFRA Pressure Instruments
Convenient electronic systems ranging from -25 mbar to
1000 bar (0.4 to 15,000 psi) - multiple choices of pressure
ranges, pumps and accuracies, fully temperature-
compensated for problem-free and accurate field use.
JOFRA Signal Instruments
Process signal measurement and simulation for easy control
loop calibration and measurement tasks - from handheld
field instruments to laboratory reference level bench top
instruments.
JOFRA / JF Marine Instruments
A complete range of calibration equipment for temperature,
pressure and signal, approved for marine use.
FP Temperature Sensors
A complete range of temperature sensors for industrial and
marine use.
M&G Pressure Testers
Pneumatic floating-ball or hydraulic piston dead weight
testers with accuracies to 0.015% of reading.
M&G Pumps
Pressure generators from small pneumatic “bicycle” style
pumps to hydraulic pumps generating up to 1,000 bar
(15,000 psi).