mesabiotech SILARIS Dock Operator Manual

CONTENTS
1. COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2. WARNINGS AND PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
3. SYMBOLS AND ABBREVIATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
4. INTENDED USE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
5. DOCK DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
6. SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
6.1. DOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
6.2. AC ADAPTOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
7. DOCK MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
7.1 EXTERIOR CLEANING PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
7.2 DISPOSAL PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
7.3 SHIPPING RETURN PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
8. DOCK SET-UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
9. DOCK OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
9.1. INSERTING THE TEST CASSETTE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
9.2. RUNNING THE TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-10
10. INTERPRETATION OF RESULTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
11. RE-PROGRAMMING THE DOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
12. TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-12
13. OTHER WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
14. REORDER INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
1
1. COMPONENTS
The following items are supplied with the Silaris Dock.
1- Silaris Dock 1- AC Power Adaptor 1- AC Power Cord 1- Operator s Guide
2. WARNINGS AND PRECAUTIONS
• In Vitro Diagnostic Use Only
• Do not block the air vents located on the rear and sides of the Silaris Dock. A minimum of 2 inches (5 cm) of clearance must be maintained.
• Do not place the Silaris Dock in the vicinity of equipment that generates heat. Keep out of direct sunlight. The Dock must be maintained in an ambient temperature range of 15°C - 30°C (59°F - 86°F) to operate properly.
• To reduce the risk of damage to the unit, avoid use of sprays in the vicinity of the Silaris Dock.
• Use only the supplied power adapter with the Silaris Dock (MeanWell Part Number GSM25B12-P1J). This component has been tested and certified for use. If a replacement power supply is needed, please contact Sekisui Diagnostics Technical Support for assistance at 800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S).
• Dispose of the Dock as electronic waste in accordance to local regulations.
2
3. SYMBOLS AND ABBREVIATIONS
The following symbols and abbreviations are used throughout the Operators Guide:
AC – Alternating Electrical Current DC – Direct Electrical Current A – Amperes V – Electrical Voltage
SYMBOLS DEFINITION
~
Alternating Current
Direct Current
Manufacturer
MF
IVD
4. INTENDED USE
The Silaris Dock is a reusable platform, used in conjunction with a Silaris Test Cassette to run in-vitro diagnostic assays.
5. DOCK DESCRIPTION
The Silaris Dock is an electronic module which executes in-vitro diagnostic tests on compatible Silaris Test Cassettes. It consists of an electro-mechanical interface to a single Test Cassette. The Dock contains all electrical systems, controls and logic necessary to orchestrate in-vitro diagnostic tests within the inserted Test Cassette.
Upon insertion of a Test Cassette, the Dock will detect and identify the assay type. After the user transfers a clinical test sample into the Test Cassette and closes the Dock lid, embedded firmware in the Dock will control fluid flow of the sample into the various chambers of the Test Cassette, apply controlled voltage signals to the various Test Cassette heaters (monitored by sensors within the Dock), and provide visual status to the user with critical information such as estimated time to read, and various error states, should they be encountered.
Manufactured For
In Vitro Diagnostic
Electronic equipment marked with this symbol should be disposed in electronic recycling waste.
Biohazard – Potentially infectious materials. Precautions must be observed.
Catalog Number
Lot Number
Consult the Instructions For Use
Consult the instructions for use for important cautionary information such as warnings and precautions that cannot, for a variety of reasons, be presented on the medical device itself.
Caution: Federal Law restricts this device to sale by or on the order of a licensed practitioner.
Informational note
Nemko Electrical Safety
3
6. SPECIFICATIONS
6.1. DOCK
POWER INPUT 12VDC ± 1VDC, 1A maximum
Ambient Temperature: 15°C to 30°C (59°F to 86°F) OPERATING ENVIRONMENT Relative Humidity: 20%-80%, non-condensing Altitude: Sea level to 8,000 feet
DIMENSIONS 5.7 in (14.5 cm) H x 3.9 in (9.9 cm) W x 3.8 in (9.7 cm) D
TOTAL WEIGHT
(Dock, Power Supply, and Power Adaptor) < 1.4 lbs (619 g)
ACCEPTIBLE TILT The Dock must be installed on a level surface, with an incline not to exceed 10 degrees.
The Silaris Dock is not to be used in direct sunlight.
OPERATIONAL LIFE 3,000 tests
6.2. AC ADAPTER
POWER INPUT 100-240VAC, 50/60Hz
POWER OUTPUT 12VDC, 2.08A, 25W maximum
NOTE: Use only the supplied power adapter with the Silaris Dock. This component has been tested and certified for use.
If a replacement power supply is needed, please contact Sekisui Diagnostics Technical Support for assistance at 800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S).
7. DOCK MAINTENANCE
To reduce the risk of damage to the Dock, avoid use of sprays in the vicinity of the Dock.
Prior to cleaning the Silaris Dock, disconnect the power supply. This will help to prevent damage to the unit.
Ensure that the Dock lid is closed prior to cleaning.
Do not open the Dock lid during the cleaning procedure.
Do not insert the cleaning wipe into the cassette slot. This may cause damage to the unit.
Do not use solvents other than those recommended in section 7.1 EXTERIOR CLEANING PROCEDURE.
Do not let liquid accumulate in or around Dock vents
Regular cleaning of the Silaris Dock is recommended. The Dock should be cleaned every day that it is used, and prior to disposal or shipment or servicing.
4
7.1 EXTERIOR CLEANING PROCEDURE
Cleaning of the exterior surfaces of the Silaris Dock and the immediate work area should be performed as necessary, no less than daily when in use.
1. Unplug the power cord from the wall outlet and Dock.
2. Close the Dock lid.
3. Using a cleaning solution (10% bleach or 70% ethanol), or a germicidal disposable wipe (such as CaviWipes™), gently wipe the outer surfaces of the Dock, removing any debris.
NOTE: Do not press the wipe against the open vents of the back or sides of the Dock.
4. Using a new cloth, wipe the front of the Dock twice top to bottom, then twice left to right. Follow this procedure for the back, top and bottom of the Dock.
5. Do not let liquid accumulate near any opening. Make sure no liquid enters the dock.
6. Leave the Dock “wet” for at least 3 minutes.
7. With a fresh dry cloth (such as a Kimwipe™) remove any residual moisture
8. Allow the unit to dry for at least 10 minutes prior to re-connecting the power cord for the AC Adapter.
7.2 DISPOSAL PROCEDURE
All electronic products must be disposed of separately from the municipal waste system.
A used Dock may be biologically contaminated. Prior to disposal, or shipping for services, the Dock must be
The Silaris Dock should be disposed in appropriate electronic recycling waste. For more information regarding
To decontaminate the Dock prior to disposal:
1. Unplug the DC power connector from the back of the Silaris Dock.
2. Open the lid of the Dock by depressing the lid latch button and pulling the lid to the open position.
3. If there is a Test Cassette inserted into the Dock, remove it by firmly grasping and pulling it upward.
4. Soak the entire Dock in a 10% bleach solution for at least 10 minutes. Ensure that the Dock is fully submerged in the solution, and that the entire interior of the Dock is flooded. It is recommended to maintain an upright orientation of the Dock in order to allow entrapped air to escape through the cassette slot.
5. Remove the Dock from the bleach solution. Turn the Dock upside down to drain the solution from the interior. Allow the Dock to dry.
6. Dispose of the Dock as electronic waste in accordance to local regulations.
NOTE: If for any reason, it becomes necessary to return a Dock to Sekisui Diagnostics, prior to packaging please carefully follow the exterior cleaning procedure. Do not follow this disposal Decontamination Procedure.
thoroughly cleaned.
disposal of the Dock, please contact Sekisui Diagnostics Technical Support at 800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S).
5
7.3 SHIPPING RETURN PROCEDURE
Cleaning of the Silaris Dock must be completed prior to the return to Sekisui Diagnostics. Cleaning must be completed according the instructions provided. Please complete the information below and attach the completed record to the dock prior to return.
Please complete the cleaning procedure following the instructions below:
1. Unplug the power cord from the wall outlet and Dock.
2. Close the Dock lid.
3. Using a cleaning solution (10% bleach or 70% ethanol), or a germicidal disposable wipe (such as CaviWipes™), gently wipe the outer surfaces of the Dock, removing any debris. Do not press the wipe against the open vents of the back or sides of the Dock.
4. Using a new cloth, wipe the front of the Dock twice top to bottom, then twice left to right. Follow this procedure for the back, top and bottom of the Dock.
5. Do not let liquid accumulate near any opening. Make sure no liquid enters the dock.
6. Leave the Dock “wet” for at least 3 minutes.
7. With a fresh dry cloth (such as a Kimwipe™) remove any residual moisture.
The information below will need to be completed and returned along with the Dock, AC Power Cord, AC Adaptor using the shipping materials provided.
DOCUMENTATION FOR COMPLETION OF CLEANING PROCEDURE
SILARIS DOCK INFORMATION
Dock Lot Number:
CONTACT & FACILILTY INFORMATION
Contact Name: Title:
Contact Phone: Email:
Facility Name:
Street Address:
City, State, Zip Code:
DECONTAMINATION INFORMATION
The Silaris Dock has been cleaned as described above. The dock surfaces have been wiped with an appropriate disinfecting solution and left for at least 3 minutes.
Performed By:
Signature Date
Please contact Sekisui Diagnostics Technical Support at 800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S) for further assistance.
6
8. DOCK SET-UP
Do not block the air vents located on the rear and sides of the Silaris Dock. A minimum of 2 inches (5 cm) of clearance must be maintained.
Do not place the Silaris Dock in the vicinity of equipment that generates heat. Keep out of direct sunlight. The Dock must be maintained in an ambient temperature range of 15°C to 30°C (59°F to 86°F) to operate properly.
Place the Silaris Dock on a level, stable table or laboratory bench.
NOTE: To ensure proper function, allow at least 2 inches (5 cm) of clearance around the Dock. Do not place it near equipment that will give off heat.
Insert the round DC connector of the adapter into the Dock as shown.
Connect the power cord to the AC adapter, and then insert the
power cord plug into the electrical outlet.
This system does not use a ground connection.
Depending on the type of AC outlet, there may be an unused ground pin showing. This is normal and expected.
Once the Silaris Dock is connected to power, it will initialize automatically.
The unit will beep, followed by a series of brief messages on the display.
Immediately after displaying the final message, the unit will beep twice, which indicates that the Dock is ready to use.
NOTE: The images shown are for example purposes only. The firmware P/N, version, and ambient temperature may differ.
MESA BIOTECH, INC,
MOLECULAR POC
FIRMWARE
P/N 210405 v. B
AMBIENT
TEMPERATURE: 25C
DOCK READY
INSERT CASSETTE
7
9. DOCK OPERATION
This section provides detailed instruction pertaining to the operation of the Silaris Dock only. It is not intended to be a comprehensive instruction for performing diagnostic tests. Such instructions are found packaged with the specific Silaris Test Kits.
The Silaris Dock is capable of running a number of Silaris Tests. For the purpose of illustration, this section includes Dock messages specific to the Silaris Influenza A&B Test.
Refer to the test specific Instructions For Use for sample collection and preparation.
The Silaris Test Cassettes are for use with the Silaris Dock only.
9.1 INSERTING THE TEST CASSETTE
Each Test Cassette may only be used once. Sensors in the Dock will not allow the assay to run if the Test Cassette has already been inserted.
Remove the Test Cassette from the foil pouch by tearing off the top of the pouch starting at the notches in the pouch.
Open the lid of the Dock by depressing the black lid latch button and pulling the lid to the open position.
O
C K
R
E A
D
YD N S
E R
T
C A
S S
E
T
T EI
Insert the Test Cassette into the slot with the Test Cassette label facing toward the front of the Dock.
NOTE: The sample port (covered with a foil tab) on the top of the Test Cassette should be accessible to the user.
Fully depress the Test Cassette into the Dock until it ‘clicks’ into position.
The Silaris Dock will measure the temperature of the Test Cassette and display it.
YDOCK READ
EINSERT CASSETT
CASSETTE
TEMPERATURE: 24C
NOTE: Once Test Cassette is seated into the Dock, the user will have only 5 minutes to start the test before the Dock times out.
If the Test Cassette has been properly inserted, the display will identify the type of assay that is supported by the Dock.
If the display shows an error code, refer to
FLU A/FLU B
CASS. INSERTED
section 12. TROUBLESHOOTING for additional information.
After a few seconds the display will prompt the user to add the sample to the Test Cassette sample port.
The Test Cassette is now ready for the addition of the sample.
To avoid contamination, do not remove the foil tab covering the Test Cassette
ADD SAMPLE
THEN CLOSE LID
sample port until you are ready to add the sample.
8
9.2. RUNNING THE TEST
Per the Test Kit instructions, remove the foil tab from the Test Cassette.
Immediately add the sample to the Test C assette sample port.
The Silaris Dock will beep upon sensing the sample, and display
“SAMPLE LOADED CLOSE LID.”
NOTE: The Dock will beep once every second until the lid is closed.
Close the lid, pressing it into place over the top of the Test Cassette until it clicks into its
locked position.
The Dock will then display “SAMPLE LOADED LID CLOSED.”
The Dock will verify that the Test Cassette has been sealed. Once verified, it will initiate the test and display “CASSETTE SEALED TEST STARTED.”
During the test, the Dock will display an estimated time to completion.
NOTE: The remaining time is an estimate of the time required to run the assay. The actual time remaining is determined by a number of factors, including ambient temperature, altitude, and the type of assay being run.
During a test run, you will periodically hear the sound of a fan and flow of air
through the Dock. This is perfectly normal behavior, and is not indicative of a Dock malfunction.
After starting the test, avoid bumping, tilting, or otherwise
disturbing the Silaris Dock, as this may cause the test to abort.
Allow the assay to run to completion before moving or disturbing the Silaris Dock.
When the assay is finished, the remaining time will be 00:00, and the Silaris Dock will display
“TEST COMPLETE READ RESULTS”.
NOTE: The Dock will beep every two seconds until the Test Cassette is removed.
SAMPLE LOADED
CLOSE LID
SAMPLE LOADED
LID CLOSED
CASSETTE SEALED
TEST STARTED
TEST RUNNING
REMAINING: 25:12
TEST COMPLETE
READ RESULTS
Remove the Test Cassette by pressing the latch button and pulling the lid open.
Firmly grasp the top of the Test Cassette and pull straight up until the Test Cassette is
removed from the Dock.
The results of the assay may be immediately read.
Refer to test specific Instructions For Use for interpretation of results for a specific test.
TEST COMPLETE READ RESULTS
Continued on page 10
9
9.2. RUNNING THE TEST
Once the Test Cassette has been removed, the Dock will display the number of tests that have been run and the number remaining.
NOTE: If fewer than 100 tests remain, the Silaris Dock will beep three times, and the display will remind the user to replace the unit.
RUN COUNT 113
REMAINING: 2887
REPLACE DOCK
IN 56 TESTS
After a few seconds, the Dock will display “DOCK READY INSERT CASSETTE”. This indicates that the Silaris Dock is ready to run an addi- tional Test Cassette if needed.
DOCK READY
INSERT CASSETTE
10. INTERPRETATION OF RESULTS
Refer to test specific Instructions For Use for interpretation of results for a specific test.
11. RE-PROGRAMMING THE DOCK
Periodically new tests are developed which require new programming for the Silaris Dock. In order to support these new tests, the Dock firmware must be updated. This is a very simple process, by following these easy steps:
Apply power to the Silaris Dock as indicated in section 8. DOCK SET-UP. The Dock will automatically initialize and display “DOCK READY INSERT CASSETTE.”
Insert the latest Programming Cassette (denoted by red top) into the Dock as instructed in section 9.1 INSERTING THE TEST CASSETTE. The Dock
DOCK READY
INSERT CASSETTE
PROGRAMMING
CASSETTE READY
will sense the Programming Cassette and display: “PROGRAMMING CASSETTE READY,” followed by “PROGRAMMING WAIT FOR REBOOT.”
PROGRAMMING
WAIT FOR REBOOT
10
The Dock’s display will turn off for a few seconds while the unit reboots and reprograms its embedded firmware.
Once the display turns on, verify that the new firmware revision is correctly reported.
NOTE: Image shown is for illustration purposes only. Firmware P/N and version may differ.
The display will prompt you to remove the Programming Cassette by displaying, “REMOVE PROGRAMMING CASSETTE.”
Once the Programming Cassette is removed, the Silaris Dock will restart with the new firmware installed.
FIRMWARE P/N
210405 v. B
REMOVE PROGRAMMING
CASSETTE
12. TROUBLESHOOTING
The Silaris Dock performs a number of different environmental tests before and during an assay run to ensure that the conditions will allow a successful assay. If any of these tests fail, an error code is displayed along with a brief notification of the type of failure.
For example, if the Dock detects that a used Test Cassette has been inserted, an error code will be displayed for several seconds.
ERROR CODE 05
This screen will be followed by a brief message describing the error.
CASSETTE ALREADY
USED! REPLACE!
Listed below are the different error codes with an explanation of the causes and requested actions for each.
Error Code
Displayed Error
Description
Cause/Requested Actions
The maximum operational altitude is 8,000 feet (2,438 meters) above sea
01
ALTITUDE OUT OF
VALID RANGE
level. The Dock will check the altitude on its initial power up and not continue if the altitude range has been exceeded. To continue, the Dock must be re-located to an altitude below 8,000 feet.
DOCK: The valid ambient operating temperature range of the Dock is 15°C
02
TEMPERATURE OUT OF
VALID RANGE
to 30°C (59°F to 86°F). This is checked during power up and whenever the Dock is reset. Move the Dock to an acceptable ambient temperature environment and let it sit for at least 5 minutes to acclimate.
TEST CASSETTE: The valid operating temperature range of the Test Cassette is 15°C to 30°C (59°F to 86°F). This is checked as soon as a Test Cassette is inserted into the Dock. Insure that the Test Cassette temperature
03
TEMPERATURE OUT OF
VALID RANGE
is valid, and re-insert it into the Dock.
05
04
UNKNOWN TYPE OF
TEST CASSETTE
CASSETTE ALREADY USED!
REPLACE!
This error is caused when the Test Cassette is not recognized by the Dock. The most likely reason for this would be that the latest program revision has not been installed into the Dock. See Section 11. RE-PROGRAMMING THE DOCK.
It is also possible that the Test Cassette is not firmly seated into the Dock. In this case, remove the Test Cassette, wait a couple of seconds, and re-insert it, pressing it firmly into position.
Each Test Cassette may only be used one time. Even if a test does not complete, that Test Cassette may not be used again if the previous test was started or if the lid was closed.
Continued on page 12
11
12. TROUBLESHOOTING
Error Code
06
07
08
09
10
11
12
13
14
15
Displayed Error
Description
CASS. NOT SEATED PROPERLY. RESEAT
TEST ABORTED!
CASSETTE REMOVED
INADEQUATE SAMPLE
REMOVE UNATTENDED CASS.
CASSETTE SEAL FAILED. REPLACE
DOCK EXPIRED
REPLACE DOCK
TEST ABORTED!
TILT ERROR!!
TEST ABORTED!
POWER IS LOW
TEST ABORTED!
FAN/COOLING FAIL
TEST ABORTED!
FLOW ERROR
Cause/Requested Actions
When inserting the Test Cassette into the Dock, it must be pressed firmly into place. If this error occurs, remove the Test Cassette, wait a couple of seconds, and re-insert it, ensuring that the Test Cassette has come to a firm stop. If this error continues after trying multiple Test Cassettes, contact Sekisui Diagnostics Technical Support at 800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S) for further assistance.
This error occurs when a Test Cassette has been removed before the assay test has been completed. To avoid this error and guarantee accurate testing, the Test Cassette must not be removed until the test is complete. See Section
9.2 RUNNING THE TEST.
This error will occur if the lid is closed before the sample is loaded into the Test Cassette. To avoid this error, never close the lid before sample addition. If a sample was added but not detected, discard that Test Cassette and insert a new Test Cassette. Ensure the sample is at room temperature. Ensure sufficient sample volume is added to the Test Cassette.
Each Test Cassette must be used in a timely manner when removed from its protective pouch. If the Test Cassette has been sitting in the Dock for more than 5 minutes, the test will abort and the Test Cassette must be discarded.
If this error occurs, replace the Test Cassette and run the test again. If this failure repeats, contact Sekisui Diagnostics Technical Support at 800­332-1042 (U.S Only) or 781-652-7800 (Outside U.S) for further assistance.
When the Dock has completed 3,000 runs, this error will appear. Contact your Sekisui Diagnostics or Distribution Representative to purchase a replacement Dock.
The Dock must be kept level at all times when operating. Do not bump, lift, or tilt the Dock during operation. If this error occurs the test is aborted and must be re-run with a new Test Cassette.
The Dock’s power supply voltage has fallen below acceptable limits. The run is aborted and another run is not allowed to start until the power supply voltage is within acceptable limits.
The Dock is not temperature cycling as expected. This may be caused by a fan malfunction, or a problem with the liquid flow inside the Test Cassette. The run is aborted and must be re-run with a new Test Cassette. If this error happens repeatedly, the Dock must be replaced.
The Dock is not heating properly during temperature cycling. This is indicative of a problem inside the Test Cassette with fluid flow. The test is aborted and must be re-run with a new Test Cassette.
12
13. OTHER WARNINGS
The following messages are informational, and do not indicate an error in the Dock or test.
While opening the lid during a test will not pause or abort the test, it is not recommended. If this warning occurs, close the lid and let it remain closed until the completion of the test.
When 100 or fewer tests of the recommended life time usage of 3,000 tests remain, this warning will be displayed after every completed test. Contact your Sekisui Diagnostics or Distribution Representative to purchase a replacement Dock before this number reaches zero.
CLOSE LID
REMAINING: 21:15
REPLACE DOCK
IN 99 TEST
14. REORDER INFORMATION
Silaris Dock Catalog # 1026
Please contact Sekisui Diagnostics Technical Support at 800-332-1042 (U.S Only) or 781-652-7800 (Outside U.S) for further assistance.
13
Sekisui Diagnostics, LLC
MF
6659 Top Gun Street San Diego, CA 92121, USA Tel: (781) 652-7800 www.sekisuidiagnostics.com
Mesa Biotech, INC. 6190 Cornerstone Court East Suite 200 San Diego, CA 92121, USA www.mesabiotech.com
© 2018 Sekisui Diagnostics, LLC. All rights reserved. Silaris® is a registered U.S. trademark of Sekisui Diagnostics, LLC. Because every result matters is a trademark of Sekisui Diagnostics, LLC.