In order to ensure optimal surgical safety, all operating table users should carefully read the operating table
1.WARNINGS, NOTES AND INSTRUCTIONS
!
!
!
!
!
!
!
!!!
!
!
and maintenance instructions before using the table. The entire surgical ward staff should be familiar with the
correct use of the PROMERIX operating table as well as all warnings and observations concerning it.
Warnings and observations found in this instruction manual are indicated as follows:
WARNING!Please observe to ensure patient safety.
CAUTION!Please observe in order to avoid causing damage to the equipment or its parts.
NOTE!Please observe in order to improve operating table properties.
WARNINGS!
PromeriX operating table should only be used in facilities made for medical purposes.
IP X4 classification is valid only when the mains cord is disconnected.
Connect the mains cord only to earthed power supply.
PromeriX shall be operated from it’s internal battery if the integrity of the protective earth conductor
arrangement is in doubt.
Place the patient in the longitudinal center line of the table top.
In NORMAL mode max. patient weight with sliding seat section is 275 kg ( 600 lbs ).
In NORMAL mode max. patient weight without sliding seat section is 325 kg ( 700 lbs ).
In REVERSE mode max. patient weight is 180 kg ( 400 lbs ).
Adjust the table top to the horizontal position ( 0-position ) with the hand control unit before transporting a
patient on the operating table.
Transporting of patients weighing up to 180 kg ( 400 lbs ) is only allowed when table top is in horizontal posi-
tion ( 0-position ) and the height of the table is max 900 mm ( 3 ft ).
5
Transporting of patients weighing 180-225 kg ( 400-500 lbs ) is only allowed when the table is in NORMAL
!
!!!!!!!
!!!
!
!!!
!
mode, in horizontal position ( 0-position ) and in the lowest position.
Transporting of patients weighing 225-325 kg ( 500-700 lbs ) is not allowed.
Use extreme caution when transporting the table with a patient on. Transporting of the table with a patient on
requires two persons. To maximize patient safety utilize proper restraint methods during transport.
To avoid endangering the patient's respiratory system, nerve pathways and circulatory system, the patient
should be positioned properly and kept under observation!
Check that table top sections catches lock correctly. Incorrect attachment of table sections is dangerous and
can cause personal injury or equipment damage!
Use only accessories recommended by the table manufacturer.
When adjusting the operating table, take care to avoid collision between accessories and the operatingtable.
Do not use worn or damaged accessories. When using the table, ensure that all accessories are properly
mounted and check the function of accessory locking and adjustments.
When adjusting the operating table, take care that the fingers, hands or other parts of the body of the patient
are not placed between edges of back, leg or seat section frames and pivoting points.
Use only permitted table top configurations illustrated in this manual.
Ensure before adjusting the operating table that table top will not hit external obstacles.
Ensure before adjusting the operating table that NORMAL/REVERSE mode has been chosen correctly and is
the same as patient's orientation on the table top. After use of main switch NORMAL mode is set as default.
Always follow manufacturer instructions when using diathermy or defibrillation equipment.
When using high frequency surgical equipment take care to prevent the patient coming into contact with
metal parts of the operating table or accessories. Do not place the patient on wet or damp surfaces or electri-
cally conductive pads. Hazard of burn injuries!
Combustible anaesthetic gases must not be used with the Promerix operating table.
6
Use potential equalization conductor with patient monitoring equipment. Connector placed at the table base.
WARNINGS, NOTES AND INSTRUCTIONS
!
!!!!
!!!
!
!
When recharging the battery first plug the mains cord to table appliance inlet, after that to wall socket.
Main switch can be used as an emergency stopping device ( demanded by the standards ). Normally the
switch lever is in up position. When the lever is turned down in OFF position all table functions are cancelled.
Table movement stops immediately.
When recharging is completed disconnect the mains cord first from the wall socket, after that from the table
appliance inlet.
Disconnect the mains cord from the table and turn the main switch to OFF position before cleaning of the
table.
Disconnect the mains cord from the table before any service procedures. 100-240 V ~ used in power unit
placed in table base. Hazard of electric shock!
Promerix operating table has been tested to EN 60601-1-2 to ensure proper electromagnetic compatibility.
Other products used in the vicinity of Promerix operating table should also comply with this standard. If they
do not comply, interference between products in unintended responses may occur. Please contact the
appropriate manufacturer if any problems arise.
Portable and mobile RF-communications equipment can affect Promerix operating table.
7
CAUTIONS!
WARNINGS, NOTES AND INSTRUCTIONS
!
!
!
!
!
!
!
!
!
!
!
!
!
!
If the operating table has been in the cold, allow it to warm up at room temperature for at least 6 hours before
recharging the battery or switching on, to allow any condensation formed to evaporate.
Recharge the battery prior to use. See the starting instructions attached to table base panel.
OVERRIDE panel is only for returning the table top to the horizontal position in the case of electronic mal-
function. Override makes it possible to adjust table without the controlling processor and any of the pro-
grammed restrictions are not functioning.
Downwards adjusted leg or back section may hit the table base or column casings depending on the used
height or tilt angle adjustments. Can cause damage to the operating table and creating pinching hazard!
Do not place any objects on the base of the operating table! Danger of damage during adjustments!
The best place to store the hand control unit especially during transport is the lever under the head section.
Head and devided leg section 60475 gas springs can be disposed of as metal waste after nitrogen gas and
oil has been removed. Instructions for gas releasing available from your sales representative.
Operating table has been classified as splash proof equipment. Use of pressurized water is not allowed.
Cleaning and disinfecting should be made according to this manual.
Do not pull or push the shields of back and leg section joints. Check during adjustment that there is no fo-
reign obstacles between shields.
It is possible to adjust the SLIDE adjustment only when back or leg section are in horizontal or above hori-
zontal position. Slide adjustment will stop automatically before collision if back or leg section in under the
horizontal position.
SLIDE adjustment is only possible when table top is in horizontal position. Use of SLIDE adjustment is
restricted with tilted table top.
Mattresses should be cleaned only with neutral detergent ( ph 7-8 ).
The antistatic properties of the operating table require the use of original brand mattresses and antistatic
flooring.
8
NOTES!
WARNINGS, NOTES AND INSTRUCTIONS
!
It’s recommended to recharge the battery overnight after day’s use. Table will be always ready for use and
the battery will have longer service life.
If the main switch is turned to OFF position ( downwards ), all table functions are blocked including recharg-
ing of the battery. Check that the green led above the charging inlet lights up when you connect the mains
cord to the table.
When mains cord is connected to operating table the cooling fan of the power unit is activated causing a
humming sound. After disconnecting the mains cord the cooling fan still runs for few seconds.
Turn the main switch to OFF position ( downwards ) when the table is not in use or stored for longer period of
time. This eliminates the possibility of unintended use and is also saving the battery capacity.
Activate the floor lock before adjusting the table. Only trendelenburg and back rest adjustments are function-
ing if floor lock is not activated.
Cardboard packing material should be recycled. Wood and plastics are energy waste.
Batteries contain lead and if replaced, need to be recycled in accordance with local environmental regula-
tions.
Velcro tapes on table top plates can be cleaned with pressurized air. Velcro tape can also be replaced; spare
parts and instructions available from Merivaara service dpt.
Floor lock can be deactivated with a release knob placed at the table base if normal deactivating with hand
control unit or override panel can’t be done. Turn the knob 90° clockwise and the table will descend on to its
castors. Turn the knob back to its original position after releasing.
9
Merivaara Corporation conforms to all its operations ISO 9001 Quality Management and ISO 14001 Environ-
2.GENERAL
Expertise is essential
The patient is the most important part of treatment. This is precisely why
the equipment used in treatment must be absolutely
safe and convenient to use. As a health care
professional, you deserve the very best tools,
allowing you to concentrate on your own field of
expertise. Merivaara is an expert in providing
hospital equipment.
Merivaara products have been designed to function
efficiently and flexibly during the various stages of
treatment. They assist you in the performance of
your work, without distracting you from the task at
hand. Our modular equipment system includes
state-of-the-art equipment for hospital procedures
and hospital room environments as well as for
nursing homes and home care applications.
For more information on Merivaara products,
contact our Sales Office. For matters related to
equipment servicing, contact the Merivaara Service Department.
mental Management System Standards. In addition the Promerix operating table meets the following stan-
dards: EN 60601-2-46, EN 60601-1-2 ( EMC ) and EN 60601-1. The table complies with directive 93/42/EEC (
MDD ) product class I, and bears a CE marking based on this classification.
2.1Intended use
Promerix Operating Table is intended for general surgical use. It is also well suited for:
• Day surgery
• ENT surgery
• Plastic surgery
• Orthopedics
• Arthroscopy
• Gynecology
• Urology
• Pediatrics
• Neurosurgery
10
2.2Operating characteristics
Promerix is transportable electrohydraulic battery and mains operated operating table. It is microprocessor
controlled and it‘s function parameters are reprogrammable with special PC-operated service program. Table
can be connected to 100….240 VAC mains power and its secondary voltage is 24 VDC. The table electronics
are safely earthed and it has equipotential connection.
Table top consists of fixed or sliding seat section and detachable 25 and 40-sections. You can rearrange the
sections as you wish depending on the surgical concept and operation. Look page 14. The width of the table
top is 540 mm, ( 21 1/4 inch ) without rails and length 2100 mm ( 82 inch ). Table top plates are x-ray translu-
cent material.
Promerix can be adjusted with cable connected hand control unit, IR-hand control unit or foot switch. Operat-
ing table can also be adjusted with override panel in case of electronical malfunction ( hand control unit or
table microprocessor defective ).
Safe and fault-free use and maintenance of the equipment requires careful acquainting to these instructions.
When mounting accessories to the equipment, the instructions provided with them must be followed closely.
Always keep the instructions for accessories together with this manual.
11
3.1Cleaning instructions
3.CLEANING
!
Disconnect the mains cord from the table and turn the main switch to OFF position before cleaning of the
table.
Operating table has been classified as a splash proof equipment. Use of pressurized water is not allowed.
Cleaning and disinfecting should be made according to this manual.
Do not pull or push the shields of back- and leg section joints. Check during adjustment that there is no for-
eign obstacles between the shields.
Surface of the Promerix mattresses is electrically conductive. Cleaning of the mattresses should be done with
neutral detergent( pH 7-8 ) and warm water applied with a single use wipe.
3.2Cleaning
•Remove accessories and mattresses.
•Clean by wiping down with a mild alkaline detergent ( pH 7-8 ).
3.3Disinfection
•Remove accessories and mattresses.
•Disinfect only when necessary.
•Wipe down the equipment with the surface disinfectant used at the facility in accordance with manu-
facturer instructions, unless the surface disinfectant contains phenols and alcohol, which can corrode
plastic parts and mattresses.
NOTE!Dry the operating table carefully
immediately after cleaning or disinfecting.
3.4Mattresses and pads
Read the care instructions for mattresses and pads.
12
4.1Safety during maintenance procedures
4.MAINTENANCE INSTRUCTIONS
!!!!!
Maintenance allowed only to persons specialized to Promerix service work.
Disconnect the mains cord from the table before any service procedures. 100-240 V ~ is used power unit
placed in table base. Hazard of electric shock!
Turn the main switch to OFF position before connecting or disconnecting any of electronic connections or
components in order to protect them against voltage surges.
Static charges can cause sparks harming sensitive electronic components. Ground yourself to metallic parts
of table before touching electronic components.
Disconnecting any of the hydraulic hoses can cause uncontrolled and dangerous movement of the table.
Support the table parts before disconnecting any hydraulic hoses!
When mains cord is connected to operating table the cooling fan of the power unit is activated causing a
humming sound. After disconnecting the mains cord the cooling fan still runs for few seconds.
4.1.1 Daily service
•During normal cleaning, check-up general condition of the operating table. Contact service personnel
if needed.
•We recommend that you charge the table battery every day. This is to ensure sufficient battery charge
level and long service life.
•Check the mains cord condition.
•Promerix operating table must be cleaned carefully after each surgical procedure!
4.1.2 Monthly service
•Clean and lubricate guide pins, latches, spring-loaded pins and pivoting points with light machine oil
or spray vaceline.
•Check that castors roll smoothly and floor lock is functioning correctly.
•Test table functions by adjusting them to their extreme positions.
13
•Check hand control unit and foot switch cables, connector and plug.
!
4.1.3 Annual service
•Perform all daily and monthly service actions.
•Check the functioning of table section's latches.
•Check all cables and connectors, regarding injuries (kinks, cracks, wear)
•Visual check at the hydraulic hose connections and piston rods, regarding external leakage.
•Check of the oil level in the reservoir of the oil pump, if the level is too low fill up with filtrated
mineral oil (ARAL VITAM DE32).
•Oil change: every 5 years
All electrical repairs must be performed by a licensed electrician.
4.2 Troubleshooting and checklist
POSSIBLE SITUATIONS:
PROBLEM: Table will not activate when pressing ON/OFF-buttonPROCEDURE:
CAUSE:-Main Switch in OFF positionTurn to ON position
-Hand Control Unit not connected properlyCheck the connection
-Faulty Hand Control UnitReplace the hand control unit
-Broken Hand Control Unit cableReplace the cable
-Empty batteriesConnect mains cable
and recharge batteries
-Battery fuse blownCheck and replace
-Faulty MPCReplace MPC
14
PROBLEM: Table adjustment not functioning when pressing function
!
!
button, table activated
CAUSE:-Floor lock not activatedPress activation button
-Making adjustment table program restrictsCheck the table posi-
-Adjustment already in its end positionCheck the table posi-
PROCEDURE:
tioning
tioning
-Faulty hand control unit or IR-control unit not programmed cor-
rectly.
PROBLEM:Table movement won’t stop although function button is
released
CAUSE:-Severe functioning failure of electronics. Malfunctions are
always recorded in the Error Log.
-Faulty hand control unitMAIN SWITCH to OFF
-Valve is mechanically stuck or stays open constantlyContact service, new
-Short circuit in the transistor controlling the solenoidContact service
-Damaged solenoid or sensor wiresContact service
Check the control unit
PROCEDURE:
MAIN SWITCH to OFF
position!
Use service program
manual to resolve error
code.
position!
Contact service
valveblock or
solenoid is needed
PROBLEM:Back section moves when leg section button is pressed or
vice versa
CAUSE:-Normal/Reverse – mode wrongly selectedCheck the selected
PROBLEM:Audial signal when using IR-hand control unitPROCEDURE:
CAUSE:- Indicates that the operating table floor lock is not activated!Press the floor lock but-
- Cause no marking into the Error LogReset by pressing the
PROCEDURE:
mode
ton steadily
ON/OFF button.
!
15
PROBLEM:Table not responding to IR-hand control unit commandsPROCEDURE:
!
CAUSE:-Control unit wrongly aimedAim towards the table
-Distance between table and controlCheck the distance
-IR-receiver blockedCheck
-IR-control unit battery depletedReplace batteries
-IR-control unit wrongly codedSet codes again
-Emergency situations use override panel buttons!
PROBLEM:Table motion stops during adjustmentPROCEDURE:
CAUSE:- A simultaneous commands will stop the table movementRelease another of the
pressed buttons
- Can cause if foot control and hand control units are pressed
at the same time.
PROBLEM:One piece leg section can not be adjustedPROCEDURE:
CAUSE:-One piece leg sections not attached to seat section correctly Check the latches
-Faulty inductive sensorContact service
PROBLEM:Hydraulic unit starts but no movement occursPROCEDURE:
CAUSE:-Valve block solenoid defective or looseContact service
-Table adjustment in its end positionCheck the table posi-
PROBLEM:The table changes position by itselfPROCEDURE:
Use one control unit at
the time
tioning
- Control unit buttons are accidentally pressed or pinchedCheck the control unit
16
CONTACT:
Further information of PromeriX service, product support and spare parts are available from your local dealer
or from Merivaara Corp.
Merivaara service+358 3 3394 6152
Merivaara service fax+358 3 3394 6249
E-mailmerivaara@merivaara.fi
Internetwww.merivaara.com
17
5.1Identification plate
5.TECHNICAL DATA
Primary voltage and frequency
Safe max. working load
B-type applied part
CE approved device
Splash proof
Consult accompanying documents
Ingress Protection
=180 kg
275
=180 kg
325
Maximum load without slide ( including patient, mattresses and accessories )
ID-plate
Identification plate is located in the joint, under of the seat section.
5.1.1Illustrated designations and symbols
~ Alternating current
!
Equipotentiality
Protective grounding
B-type applied part
Attention, consult accompanying documents
Maximum load with slide ( including patient, mattresses and accessories )
18
5.2Properties and materials
5.2.1Surface materials
Surface materials used in Promerix table with possibility to skin contact: