Merit Medical QuadraSphere Microspheres User Manual

QQuadraSphere® Microspheres
INSTRUCTIONS FOR USE
QuadraSphere
®
Microspheres are indicated for embolization of hypervascularized tumors and peripheral arteriovenous malformations.
DESCRIPTION:
QuadraSphere Microspheres are part of a family of embolic agents based on proprietary technologies. They are designed for controlled, targeted embolization. QuadraSphere Microspheres are biocompatible, hydrophilic, non-resorbable, expandable, and conformable microspheres. QuadraSphere Microspheres swell upon exposure to aqueous solutions. They are available in a range of sizes.
DEVICE PACKAGING:
QuadraSphere Microspheres are contained in a sterile, 10 ml Cyclic Olefin Copolymers (COC) vial, with a crimped cap, packaged in a sealed pouch. Contents: 25 mg (nominal) of dry QuadraSphere Microspheres per vial to be reconstituted in NaCl 0.9% aqueous solution before use (or aqueous solution of equivalent ionic concentration).
CONTRAINDICATIONS:
• Patients intolerant to vascular occlusion procedures
• Vascular anatomy or blood flow precluding correct catheter placement or embolic injection
CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician.
Dry (µm)
30-60
50-100 100-150 150-200
Z 1822r03/12 730082001
1
Z 1822 rev J 06/12 730082002/A
2
• Presence or suspicion of vasospasm
• Presence or likely onset of haemorrhage
• Presence of severe atheromatous disease
• Feeding arteries too small to accept the selected QuadraSphere Microspheres
• Presence of collateral vessel pathways potentially endangering normal territories during embolization
• High flow arteriovenous shunts or fistulae with luminal diameter greater than the selected size of QuadraSphere Microspheres
• Vascular resistance peripheral to the feeding arteries precluding passage of QuadraSphere Microspheres into the lesion
• Presence of arteries supplying the lesion not large enough to accept QuadraSphere Microspheres
• Do not use in pulmonary vasculature, coronary and central nervous system vasculature
• Known sensitivity to poly vinyl alcohol-co-sodium acrylate
WARNINGS:
• QuadraSphere Microspheres size must be chosen after consideration of the arteriovenous angiographic appearance. QuadraSphere Microspheres size should be selected to prevent passage from any artery to vein.
• Some of the QuadraSphere Microspheres may be slightly outside of the range, so the physician should be sure to carefully select the size of QuadraSphere Microspheres according to the size of the target vessels at the desired level of occlusion in the vasculature and after consideration of the arteriovenous angiographic appearance.
• Because of the significant complications of misembolization, extreme caution should be used for any procedures involving the extracranial circulation encompassing the head and neck, and the physician should carefully weigh the potential benefits of using embolization against the risks and potential complications of the procedure. These complications can include blindness, hearing loss, loss of smell, paralysis, and death.
• Serious radiation induced skin injury may occur to the patient due to long periods of fluoroscopic exposure, large patient, angled x-ray projections and multiple image recording runs or radiographs. Refer to your facility’s clinical protocol to ensure the proper radiation dose is applied for each specific type of procedure performed.
• Onset of radiation injury to the patient may be delayed. Patients should be counselled on potential radiation effects, what to look for and who to contact if symptoms occur.
• QuadraSphere Microspheres MUST NOT be reconstituted in sterile water for injection. Reconstitution in sterile water results in extensive swelling that renders the injection of QuadraSphere Microspheres very difficult or may prevent injection.
• Do not reconstitute QuadraSphere Microspheres with Lipiodol / Ethiodol.
• Pay careful attention for signs of mistargeted embolization. During injection carefully monitor patient vital signs to include SaO
(e.g. hypoxia, CNS changes). Consider
2
terminating the procedure, investigating for possible shunting, or increasing Microspheres size if any signs of mistargeting occur or patient symptoms develop.
• Consider upsizing the Microspheres if angiographic
3
evidence of embolization does not quickly appear evident during injection of the Microspheres.
Warnings about use of small microspheres:
• Careful consideration should be given whenever use is contemplated of embolic agents that are smaller in diameter than the resolution capability of your imaging equipment. The presence of arteriovenous anastomoses, branch vessels leading away from the target area or emergent vessels not evident prior to embolization can lead to mistargeted embolization and severe complications.
Microspheres smaller than 100 microns will generally migrate distal to anastomotic feeders and therefore are more likely to terminate circulation to distal tissue. Greater potential of ischemic injury results from use of smaller sized microspheres and consideration must be given to the consequence of this injury prior to embolization. The potential consequences include swelling, necrosis, paralysis, abscess and/or stronger post-embolization syndrome.
• Post embolization swelling may result in ischemia to tissue adjacent to target area. Care must be given to avoid ischemia of intolerant, non targeted tissue such as nervous tissue.
PRECAUTIONS:
QuadraSphere Microspheres must only be used by physicians trained in vascular embolization procedures. The size and quantity of microspheres must be carefully selected according to the lesion to be treated and the potential presence of shunts. Only the physician can decide the most appropriate time to stop the injection of QuadraSphere Microspheres. Do not use if the vial, cap, or pouch appear damaged. For single patient use only - Contents supplied sterile ­Never reuse, reprocess, or resterilize the contents of a vial that has been opened. Reusing, reprocessing or resterilizing may compromise the structural integrity of the device and or lead to device failure, which in turn may result in patient injury, illness or death. Reusing, reprocessing or resterilizing may also create a risk of contamination of the device and or cause patient infection or cross infection including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness or death of the patient. All procedures must be performed according to accepted aseptic technique. QuadraSphere Microspheres MUST NOT be used in their original dry state. They must be reconstituted before use. QuadraSphere Microspheres swell in aqueous solution. The magnitude of swelling depends on the ionic concentration of the solution. The microspheres swell to approximately
four times their diameter in 0.9% NaCl aqueous solution and non-ionic contrast media, as compared to their initial dry diameter. QuadraSphere Microspheres are compressible and can be injected easily through microcatheters. However, injection of the QuadraSphere Microspheres before they are fully expanded could result in failure to reach the intended embolization target and possible embolization of a larger tissue area. Patients with known allergies to non-ionic contrast media may require corticosteroids prior to embolization. Additional evaluations or precautions may be necessary in
4
managing periprocedural care for patients with the following conditions:
• Bleeding diathesis or hypercoagulative state
• Immunocompromise
POTENTIAL COMPLICATIONS:
Vascular embolization is a high-risk procedure. Complications may occur at any time during or after the procedure, and may include, but are not limited to, the following:
• Paralysis resulting from untargeted embolization or ischemic injury from adjacent tissue oedema
• Undesirable reflux or passage of QuadraSphere Microspheres into normal arteries adjacent to the targeted lesion or through the lesion into other arteries or arterial beds, such as the internal carotid artery, pulmonary, or coronary circulation
• Pulmonary embolism due to arteriovenous shunting
• Ischemia at an undesired location, including ischemic stroke, ischemic infarction (including myocardial infarction), and tissue necrosis
• Capillary bed occlusion and tissue damage
• Vasospasm
• Recanalisation
• Blindness, hearing loss, and loss of smell
• Foreign body reactions necessitating medical intervention
• Infection necessitating medical intervention
• Complications related to catheterization (e.g. haematoma at the site of entry, clot formation at the tip of the catheter and subsequent dislodgement, and nerve and/or circulatory injuries which may result in leg injury)
• Allergic reaction to medications (e.g. analgesics)
• Allergic reaction to non-ionic contrast media or embolic material
• Vessel or lesion rupture and haemorrhage
• Death
• Additional information is found in the Warnings section
SWELLING BEHAVIOR:
QuadraSphere Microspheres swell during reconstitution with NaCl 0.9% aqueous solution and non-ionic contrast media. When hydrated in 100% NaCl 0.9% aqueous solution or non-ionic contrast medium, or 50% non-ionic contrast and 50% NaCl 0.9% aqueous solution, QuadraSphere Microspheres swell approximately 4 times their original dry diameter in approximately 10 minutes. For example, QuadraSphere Microspheres with a diameter of approximately 50-100 microns in their dry state will expand to approximately 200-400 microns during reconstitution as recommended below. Because of the inherent variability of the swelling process, some of the QuadraSphere Microspheres will be slightly outside of this range after reconstitution, so the physician should be sure to carefully select the size of QuadraSphere Microspheres according to the size of the target vessels at the desired level of occlusion in the vasculature and the nature of the aqueous solution.
CATHETER COMPATIBILITY:
QuadraSphere Microspheres can be injected with
5
microcatheters with the following specifications:
Approximate
Dry (µm)
Reconstituted Size
range (µm)
30-60 120-240
50-100 200-400 100-150 400-600 150-200 600-800
INSTRUCTIONS:
QuadraSphere Microspheres must be reconstituted with 100% NaCl 0.9% aqueous solution or non-ionic contrast medium, or 50% non-ionic contrast medium and 50% NaCl
0.9% aqueous solution.
• Carefully select the size of QuadraSphere Microspheres according to the size of the target vessels at the desired level of occlusion in the vasculature and the nature of the aqueous solution. See the description of “Swelling Behavior”.
• QuadraSphere Microspheres may be present outside the vial. Therefore, the vial must be aseptically handled away from the main sterile field.
• Ensure the compatibility of the QuadraSphere Microspheres with the intended size of catheter to be used. See the table above.
• Inspect the packaging to confirm that it is intact. Remove the vial from the pouch. The external surface of the vial is sterile.
Catheter Size ID
(in.)
0.0210.0210.0240.027
PREPARATION FOR EMBOLIZATION:
The approximate reconstitution time is 10 min.
• Fill a 10ml syringe with 100% NaCl 0.9% aqueous solution or non-ionic contrast medium (or 50% NaCl 0.9% aqueous solution and 50% contrast). Connect the syringe to a needle of 20 gauge diameter or larger.
• To ensure proper reconstitution of the QuadraSphere Microspheres, grasp the vial horizontally in your fingertips and roll the vial several times. This will transfer the dry contents of the vial to the sidewall.
Note:
Pull back only the flip-top cap; do not remove the
crimp ring or the stopper from the vial.
• Carefully insert the needle from the syringe through the stopper of the vial. Continue rolling the vial in your fingertips and inject the full amount (10ml) of reconstitution medium into the vial, then place the vial vertically and carefully remove the syringe with the needle attached.
Note:
The vial is hermetically closed. Proper aspiration
and/or venting techniques, as approved by the healthcare facility, may be used for easier injection of reconstitution medium into vial. If aspiration of air from the vial is
performed prior to reconstitution, exercise caution not to remove the spheres from the vial.
• To ensure a homogeneous reconstitution of the QuadraSphere Microspheres, gently invert the vial back and forth so that the liquid contacts the stopper 5-10 times.
6
Note:
STERILIZE
2
2
Vigorous shaking may introduce micro bubbles,
which can cause the microspheres to aggregate.
• Wait a minimum of 10 minutes
to allow the QuadraSphere
Microspheres to reconstitute and expand fully.
• Use a 30ml syringe and 20 gauge or larger needle to aspirate the contents of the vial. Rotate the vial to a vertical position with the bottom of the vial facing upward. Pull the needle back so that it is submerged in the liquid but not occluded by the stopper. Gently aspirate the entire contents of the vial into the syringe.
Note:
If the air was previously aspirated from the vial, gentle injection of air using the syringe prior to aspirating the contents of the vial will ensure an easier aspiration of vial contents into the syringe. If all contents are not withdrawn, introduce an additional volume of air and repeat the aspiration process. It is possible to add an additional amount of non-ionic contrast or NaCl 0.9% aqueous solution into the syringe in order to get a higher dispersion of microspheres.
• If microspheres were reconstituted using 100% NaCl
0.9%, non-ionic contrast medium must be added to the syringe containing the QuadraSphere Microspheres for visualization under fluoroscopy. If non-ionic contrast medium was used to reconstitute the microspheres, additional non-ionic contrast medium may be added.
DELIVERY INSTRUCTIONS:
• Carefully evaluate the vascular network associated with the target lesion utilizing high resolution imaging.
Note:
It is important to determine if any arteriovenous
shunts are present before beginning embolization.
• Using standard techniques, position the delivery catheter within the target vessel and the catheter tip as close as possible to the embolization target.
• Use an injection syringe no larger than 3ml for the delivery of QuadraSphere Microspheres. Use of a 1ml injection syringe is recommended.
• Aspirate 1ml of the QuadraSphere Microspheres mixture into the injection syringe.
• Two methods for embolic aliquot sequestering for injection may be used:
- Option 1:
Connect a 3 way-stopcock to the 30ml syringe containing the QuadraSphere Microspheres to the infusion micro catheter and use a 1ml syringe for injection through the open port of the 3 way-stopcock.
- Option 2:
Serial aliquots of QuadraSphere Microspheres can be drawn from the 30ml syringe into a 1ml injection syringe through a 3 way-stop cock that is not attached to the infusion catheter. The 1ml syringe containing each aliquot can be attached independently to the infusion microcatheter and injected.
• Invert the 30ml syringe back and forth to maintain the homogenous suspension of the QuadraSphere Microspheres mixture.
• Under continuous fluoroscopic guidance, inject the aliquot of QuadraSphere Microspheres in a slow, non forceful, pulsatile manner over a time period of approximately 1 minute per ml of microspheres solution. Always inject under free-flow conditions and monitor for reflux.
Reflux of embolic spheres can induce immediate
Note:
ischemia of untargeted tissues and vessels.
• When stasis in the feeding pedicle occurs while delivering
7
the QuadraSphere Microspheres, wait a minimum of 5
STERILIZE
2
2
minutes then perform a selective angiogram after the full 5 minutes wait to verify the cessation of antegrade flow.
• If cessation of antegrade flow has not occurred, continue infusion under fluoroscopic guidance until the desired devascularization is obtained.
• After the QuadraSphere Microspheres infusion is completed, remove the catheter while maintaining gentle aspiration to avoid dislodging any residual QuadraSphere Microspheres that may still be in the catheter lumen. Discard the catheter after removal and do not reuse.
• Discard any open vial or unused QuadraSphere Microspheres.
CAUTION:
In the event that the catheter becomes obstructed or significant infusion resistance is encountered during injection, do not attempt to flush the catheter with excessive pressure because reflux of embolic material may occur resulting in untargeted embolization. Remove the catheter while applying gentle aspiration and discard.
CONSERVATION AND STORAGE:
QuadraSphere Microspheres must be stored in a dry, dark place in their original vials and packaging. Use by the date indicated on the labels of the outer box.
When the procedure of reconstitution is completed, store the solution of QuadraSphere Microspheres in 2 to 8°C
conditions and use within 24 hours, IF not used
immediately. Do not store QuadraSphere Microspheres after
contrast medium has been added.
Information on packaging:
Symbol
LOT
REF
Manufacturer: Name & Address
Use by date: year-month
Batch code
Catalogue number
Do not resterilize
Do not use if package is damaged
Keep away from sunlight
Keep dry
Designation
STERILE
Do not re-use
Caution - Refer to Instructions For Use
Non-pyrogenic
Sterilized using irradiation
R
Size of dry microspheres / Size of hydrated microspheres
8
Biosphere Medical, S.A.
Parc des Nations - Paris Nord 2
383 rue de la Belle Etoile
95700 Roissy en France
France
Distributed in USA by:
Merit Medical Systems, Inc.
Customer Service
1-800-356-3748
781-681-7900
QuadraSphere Microspheres is a registered trademark of BioSphere Medical.
All serious or life-threatening adverse events or deaths associated with use of QuadraSphere Microspheres should be reported to the U.S. Food and Drug Administration under the MedWatch program and to the device manufacturer. Information about the MedWatch program and forms for reporting adverse events can be obtained at www.fda.gov/safety/medwatch/howtoreport/ucm053074.htm or by calling toll free 888-463-6332. Reports to BioSphere Medical, Inc. can be made by calling toll free 800-394-
0295.
Loading...