MGF16BN Instructions
INSTALLATION AND PRECAUTIONS
Read these instructions before installing the gauge. The configuration options may be easier to set up before the gauge is installed.
NEVER insert objects into the gauge port or blow out with compressed air. Permanent damage will result to the sensor.
Use an appropriate screen or filter to keep debris out of gauge port.
Use fittings appropriate for the pressure range of the gauge.
Due to the hardness of 316L stainless steel, it is recommended that a thread sealant be used to ensure leak-free operation.
Do not apply vacuum to gauges not designed for vacuum operation. Permanent damage will result to the sensor.
POWER-UP
1. Press the front button.
2. The full-scale range is indicated and the display segments are tested.
3. The actual pressure and units are displayed.
POWER-UP WITH ZERO
1. Be sure the gauge port is exposed to normal atmospheric pressure and no pressure or vacuum is applied. The zeroing function is only activated at each power-up and the stored zero correction is erased when the gauge is shut off.
2. Press and hold the front button until This indicates that the gauge has been zeroed.
3. The full-scale range is indicated and the display segments are tested.
4. The actual pressure and units are displayed.
NORMAL OPERATION
Following the start-up initialization, the display indicates the pressure reading updated approximately 3 times per second. The auto shutoff timer starts when the gauge is pow­ered up or whenever the button is pushed, unless the gauge is set up with the auto shutoff timer turned off (0 AST).
If excessive vacuum is applied to a pressure-only gauge, the display will indicate – E r r until the vacuum is released. Applying vacuum to a gauge designed for pressure may damage the pressure sensor. If excessive pressure is applied (112.5% over range), an out-of-range indication of i – – – or i –.–.–.– will be displayed depending on model.
MINIMUM AND MAXIMUM READINGS
Minimum and maximum readings are continuously stored and updated whenever gauge is on. The stored readings can be manually cleared if desired. The high and low memory is also cleared whenever the gauge is off.
Press and hold the button for about 1 second until HI is displayed alternating with the units. The maximum stored value is displayed.
After HI is displayed, press and hold the button again for about 1 second until LO is displayed alternating with the units. The minimum stored value is displayed.
After LO is displayed, press and hold the button again for about 1 second until AP (Applied Pressure) is displayed. The high and low readings are kept in memory and the gauge returns to normal operation with the display indicating the current pressure.
Press and continue to hold the button until the display indicates Cir Hi/Lo (about 3 seconds total) and then release the button. Both HI and LO values are cleared and the gauge returns to the normal mode and displays the current pressure.
SHUT-DOWN
To shut off the gauge manually at any time, press and hold the button until the display indicates OFF (about 5 seconds) and then release.
When an auto shutoff time is set, the display indicates OFF five seconds prior to auto shutoff. The button can be pressed to keep the gauge on. The auto shutoff timer is reset whenever the button is pressed and released.
If the gauge is set up with the auto shutoff timer turned off (0 AST) it will stay on until manually shut off or until the batteries are depleted. Turn gauge off when not in use to conserve battery life.
DIMENSIONS
oooo
is displayed and then release the button.
OPERATION SUMMARY
Function Button Prompt—Release Button Result
Gauge On Press Gauge Range/Display Test Actual Pressure Gauge On & Zero Press/hold Hi Reading Press/hold HI Maximum Reading Lo Reading Press/hold Lo Minimum Reading Exit Hi/Lo Press/hold AP Actual Pressure Clear Hi/Lo Press/hold HI / LO / AP + Cir Actual Pressure Off, Clear Zero Press/hold HI / LO / AP + Ci
oooo
Zeroed Actual Pressure
r
+ OFF Gauge Off, Clear Zero
ENGINEERING UNIT SELECTION
Compound (inHg/PSIG) gauges can be changed to display single-unit vacuum/pressure readings (i.e. ±psig) when in the user configuration mode.
Standard PSIG units are mathematically converted to the newly selected engineering unit. When the gauge is powered up, the originally configured PSIG range is displayed and then the conversion with the selected engineering unit is displayed.
With the gauge powered up, press and hold the UP button. Release the button when the engineering units begin to flash.
Use the UP and DOWN buttons to scroll through the list of engineering units available for the pressure range of the sensor.
When the desired units are displayed, press and release the front button to save the selec­tion and return to normal operation.
AUTO SHUTOFF TIME SELECTION
Auto shutoff time selection is done via internal buttons to help prevent accidental or unauthorized changes in the field. The selected shut off time is stored in non-volatile memory and will be retained even with the battery off or batteries removed.
Remove the rear cover to gain access to the two internal buttons located near the lower right and left corners of the circuit board.
With the gauge powered up, press and hold the DOWN button. Release the button when the auto shutoff time is displayed on the upper section.
The lower display segments will indicate AST M if the time displayed is in minutes, and AST H if it in hours.
An auto shutoff time of zero signifies that the auto shutoff feature is disabled.
Use the UP and DOWN buttons to select 0, or 1, 2, 5, 10, 15, 20 or 30 minutes, or 1, 2, 4, or 8 hours.
When the desired auto shutoff time is displayed, press and release the front button to save the selection and return to normal operation.
BATTERY REPLACEMENT
A low battery indication will be shown in the upper left-hand corner of the display when the battery voltage falls sufficiently. The battery should be replaced soon after the indica­tor comes on or unreliable readings may result or the gauge may fail to power up.
1. Remove the 6 Phillips-head screws on the back of the unit.
3. Always replace both batteries at the same time with high quality alkaline batteries. Install batteries with correct orientation. The negative (flat) end of each battery should be inserted first facing the battery holder spring.
4. Replace the back cover, including the rubber sealing gasket taking care not to pinch the power wires between the cover and the body of the case.
s
¼" NPT
XXXXX
i i i i i
8
8888
3.5"
3.0"
2
t
10920 Madison Ave
Cleveland, OH 44102
Phone: 216-281-1100
0.75"
2.0"
Fax: 216-281-0228
© 01-09Specifications are subject to change without notice.
Page 2
MGF16BN Configuration and Calibration Instructions
USER CONFIGURATION
Remove the rear cover to gain access to the buttons located near the lower right and left corners of the circuit board.
The display then indicates _ _ _ _ with the first underscore blinking, and with PASS on the character segments.
Note: The gauge will automatically revert to normal operation if no buttons are oper­ated for approximately 15 seconds. To cancel and return to normal operation, press and release the front button without entering any pass code characters.
Enter the user-modifiable pass code (3510 factory default)
1. Use the UP or DOWN buttons to set the left-most digit to 3.
2. Press and release the front button to index to the next position. The 3 will remain,
and the second position will be blinking.
3. Use the UP or DOWN buttons to select 5.
4. Press and release the front button to index to the next position.35 will remain, and
the third position will be blinking.
5. Use the UP or DOWN buttons to select i.
6. Press and release the front button to index to the next position. 35i will remain, and
the fourth position will be blinking.
7. Use the UP or DOWN buttons to select 0.
8. Press and release the front button to proceed with configuration procedures.
If an incorrect pass code is entered, the gauge will return to the start of the pass code entry sequence.
Factory/User configuration
The upper display section will be blank, and the lower section will display either USER_ or FCTRY.
If USER_ is selected, the existing user configuration will be retained and will be acces­sible for modification as described in the following steps. To select USER_, press and release the DOWN button. The lower section of the display will indicate USER_. Press and release the front button to continue with configuration.
If FCTRY is selected, the existing user configuration will be replaced by the configura­tion as it left the factory. To select FCTRY, press and release the UP button. The lower section of the display will indicate FCTRY. Press and release the front button to restore the factory configuration and restart the gauge.
High and low value capture configuration
The upper display section will be blank.
Use the UP and DOWN buttons to select from the following:
HI/LO Both highest and lowest values measured will be captured HI/-- Only highest value measured will be captured
--/LO Only lowest value measured will be captured
--/-- Capture feature is disabled
Press and release the front button to move on to the next parameter.
Gauge type configuration (vacuum/pressure models only)
This will only appear with 15, 100, or 200 psig ranges that were originally ordered as compound gauges.
The upper display section will be blank.
Use the UP and DOWN buttons to select from the following:
-/+EU Vacuum is indicated as negative pressure in the selected engineering units CMPND Vacuum is negative INHG, pressure is PSI
Press and release the front button to save the user configuration and restart the gauge.
CALIBRATION
The gauge is calibrated at the factory using equipment traceable to NIST. There is no need to calibrate the gauge before putting it in service.
Calibration should only be performed by qualified individuals using appropriate calibra­tion standards and procedures. The calibration equipment should be at least four times more accurate than the gauge being calibrated. The calibration system must be able to generate and measure pressure/vacuum over the full range of the gauge. A vacuum pump able to produce a vacuum of 10 microns (0.01 torr or 10 millitorr) or lower is required for vacuum gauges. Remove the rear cover to gain access to the UP and DOWN buttons located near the lower right and left corners of the circuit board.
With the gauge off, press and hold the DOWN button. Then press the front button. Release all buttons when the display indicates CAL.
The display begins by indicating the full-scale positive pressure rating of the gauge in the engineering units as configured by the factory, and then proceeds to show all display segments.
Before the gauge enters the Calibration Mode, the display initially indicates _ _ _ _ with the first underscore blinking, and with PASS on the character segments.
Note: The gauge will automatically revert to normal operation if no buttons are oper-
ated for approximately 15 seconds. To cancel and return to normal operation, press and release the front button without entering any pass code characters.
Enter the user-modifiable pass code (3510 factory default)
1. Use the UP or DOWN buttons to set the left-most digit to 3.
2. Press and release the front button to index to the next position. The 3 will remain,
and the second position will be blinking.
3. Use the UP or DOWN buttons to select 5.
4. Press and release the front button to index to the next position.35 will remain, and
the third position will be blinking.
5. Use the UP or DOWN buttons to select i.
6. Press and release the front button to index to the next position. 35i will remain, and
the fourth position will be blinking.
7. Use the UP or DOWN buttons to select 0.
8. Press and release the front button to proceed with configuration procedures.
If an incorrect pass code is entered, the gauge will return to the start of the pass code entry sequence.
The calibration may be performed in any of the available engineering units as well as percent (PCT). For greatest accuracy, use the UP and DOWN buttons to select engineer­ing units for calibration with highest resolution. Press and release the front button when the appropriate engineering units are displayed. Suggested units are listed below.
Sensor Suggested range and units for calibration
5 PSI 5.000 PSI 15 PSI 775.7 MMHG (TORR) 30 PSI 61.08 INHG 50 PSI 50.00 PSI 60 PSI 60.00 PSI 100 PSI 7.031 KG/CM2 200 PSI 407.2 INHG 300 PSI 610.8 INHG 500 PSI 500.0 PSI 1000 PSI 70.31 KG/CM2 2000 PSI 4072 INHG 3000 PSI 6108 INHG 5000 PSI 5000 PSI Any 10000 PCT (percent)
The display will then indicate the currently applied pressure in the engineering units selected for calibration.
Gauge reference pressure gauges
Apply zero pressure. This may be done by venting to atmosphere. The character seg­ments of the display will alternate between ZERO and CAL. Adjust for a display indication of zero using the UP and the DOWN buttons. *See note below.
Apply full-scale pressure. The character segments of the display will alternate between +SPAN and CAL. Adjust for a display indication of full-scale pressure using the UP and the DOWN buttons. *See note below.
Apply 50% full-scale pressure. The character segments of the display will alternate between +MID and CAL. Adjust for a display indication equal to 50% of full-scale pressure using the UP and the DOWN buttons. *See note below.
Compound gauges (vacuum/pressure ranges)
Warning: application of vacuum to non-vacuum models may result in irreparable dam­age to the sensor.
For bipolar and –30.00inHg/+15.00psig compound range models only, apply 50% full­scale vacuum. The character segments of the display will alternate between -MID and CAL. Adjust for a display indication equal to 50% of full-scale vacuum using the UP and the DOWN buttons. *See note below.
* Note: Each time one of the calibration buttons is pressed and released quickly, a small change is made to the digitized pressure signal. It may take more than one of these small changes to result in a single digit change on the display. To make larger changes, press and hold the appropriate calibration button. After about one second, the display will begin to change continuously. Release the button to halt the continuous change. Then make fine adjustments by pressing and quickly releasing the calibration buttons as previously described.
Save Calibration
Once the adjustments are complete, press and hold the front button until the display indicates ---- then release the button to store the calibration parameters in non-volatile memory and restart the gauge.
Verify the pressure indications at 0%, 25%, 50%, 75% and 100% of full scale.
© 01-09
Page 3
MGF16BN User Defined Pass Code Configuration
USER DEFINED PASS CODE CONFIGURATION
Remove the rear cover to gain access to the buttons located near the lower right and left corners of the circuit board.
Battery powered models: With the gauge off, press and hold the UP button. Then press the front button. Release all buttons when the display indicates CFG.
The display initially indicates _ _ _ _ with the first underscore blinking, and with PASS on the character segments.
To cancel and return to normal operation, press and release the POWER button without entering any pass code characters.
Enter access code 1220
1. Use the UP and DOWN buttons to set the left-most digit to i.
2. Press and release the front button to index to the next position. The i will remain, and
the second position will be blinking.
3. Use the UP and DOWN buttons to select 2.
4. Press and release the front button to index to the next position. i2 will remain, and
the third position will be blinking.
5. Use the UP and DOWN buttons to select 2.
6. Press and release the front button to index to the next position. i 2 2 will remain, and
the fourth position will be blinking.
7. Use the UP and DOWN buttons to select 0.
8. Press and release the front button to proceed.
Change access code
Once the correct access code has been entered, the display will indicate the existing user­defined pass code with UDPCD on the character segments.
1. Operate the UP or DOWN button to select the first character of the new pass code.
2. When the correct first character is being displayed, press and release the front button to proceed to the next pass code character.
3. Repeat above until the entire pass code is complete.
4. To exit the User-Defined Pass Code change mode, press and hold the front button. Release the button when the display indicates ---- to restart the gauge.
5. Replace the rear cover taking care not to pinch the power wires between the cover and the case.
10920 Madison Ave
Cleveland, OH 44102
Phone: 216-281-1100
Fax: 216-281-0228
© 01-09
Page 4
MGF16BN Battery-Powered Digital Pressure Gauges
l ±0.25% Test Gauge Accuracy
l 316L Stainless Steel Wetted Parts
l NEMA 4X
l Capture Minimum and Maximum Readings
l Push Button Zero
l Selectable Engineering Units
l Selectable Auto Shutoff Times
The Meriam Instrument MGF16BN digital pressure gauge successfully addresses the modern demands of process pressure measurement. Pressure ranges are available to 5,000 PSIG at standard accuracy of ±0.25% of full scale and optional accuracy of ±0.1% of full scale. Compound gauges that can be used for positive pressure measurements and vacuum measurements are also available.
The display is a 4½ digit LCD with 0.5" high numerals and a lower alphanumeric display for engineering units and to aid in set up. Each gauge includes up to 15 field selectable engineering units depending on the sensor range. The selectable units include PSIG, kiloPascals, megaPascals, millibar, bar, atmospheres, kilograms per square centimeter, grams per square centimeter, centimeters of water, ounces per square inch, feet of water, inches of water, millimeters of mercury, Torr, and inches of mercury.
Other standard features include min/max capture, zeroing, pass code program lock-out, adjustable display shutoff time and field recalibration. The housing is ABS/polycarbon­ate and is NEMA 4X rated. Wetted components are 316L stainless steel for compatibility
with many process fluids. Process connection is via a 1⁄4" MNPT 316L stainless steel
fitting. The gauge is powered by two AA alkaline batteries.
Typical applications are instrumentation and control in the process industries, hydraulics, pneumatics, industrial gases and OEM’s.
ACCURACY
Standard: ±0.25% of full scale ±1 least significant digit includes linearity, hysteresis, repeatability Optional: -HA ±0.1% FS ±1 LSD
-NC NIST traceable test report and calibration data
DISPLAY
3 readings per second nominal display update rate 4½ digit LCD, 0.5" H, 5 character 0.25" H alphanumeric lower display
CONTROLS & FUNCTIONS
Front button turns gauge on or off, zeros gauge, and cycles through min/max functions Internal push buttons for selection of engineering units, selection of auto shutoff times, and calibration
CALIBRATION
Pass code protected calibration Non-interactive zero, span, and linearity, ±10% of range
AUTOMATIC SHUTOFF
5 minute default User selectable times ranging from 1 minute to 8 hours or front button on/off
POWER
2 AA alkaline batteries Approximately 2000 hours battery life
WEIGHT
Gauge: 9 ounces (approximately) Shipping weight: 1 pound (approximately)
MATERIAL
ABS/polycarbonate NEMA 4X case, rear gasket, polycarbonate label
CONNECTION SIZE
1/4" NPT male, 316L SS
MATERIAL, MEDIA COMPATIBILITY
All wetted parts are 316L SS, compatible with most liquids and gases
OVERPRESSURE
3000 psig range: 5000 psig 5000 psig range: 7500 psig All others: 2 x pressure range
112.5% FS out-of-range display: i – – – or i –.–.–.– depending on model
BURST PRESSURE
4 times sensor pressure rating, or 10,000 psi, whichever is less
ENVIRONMENTAL
Storage temperature: –40 to 203°F (–40 to 95°C) Operating temperature: –4 to 185°F (–20 to 85°C) Compensated temperature: 32 to 158°F (0 to 70°C)
Low battery symbol on display when batteries must be replaced
RANGES ±0.25% FS MODELS ±0.1% FS MODELS SELECTABLE ENGINEERING UNITS
–30.0 inHg to 15.0 PSIG
–30.0 inHg to 100.0 PSIG
–30.0 inHg to 200.0 PSIG
0 - 5.000 PSIG
0 - 15.00 PSIG
0 - 60.00 PSIG
0 - 100.0 PSIG
0 - 200.0 PSIG
0 - 300.0 PSIG
0 - 500.0 PSIG
0 - 1000 PSIG
0 - 2000 PSIG
0 - 3000 PSIG
0 - 5000 PSIG
Optional for all Models
MGF16BN–30V15PSIG
MGF16BN–30V100PSIG
MGF16BN–30V200PSIG
MGF16BN5PSIG MGF16BN5PSIG-HA
MGF16BN15PSIG MGF16BN15PSIG-HA
MGF16BN60PSIG MGF16BN60PSIG-HA
MGF16BN100PSIG MGF16BN100PSIG-HA
MGF16BN200PSIG MGF16BN200PSIG-HA
MGF16BN300PSIG MGF16BN300PSIG-HA
MGF16BN500PSIG MGF16BN500PSIG-HA
MGF16BN1000PSIG MGF16BN1000PSIG-HA
MGF16BN2000PSIG MGF16BN2000PSIG-HA
MGF16BN3000PSIG MGF16BN3000PSIG-HA
MGF16BN5000PSIG MGF16BN5000PSIG-HA
NC NC
InHg/PSI, can be user reconfigured as a bipolar gauge with selectable units
InHg/PSI, can be user reconfigured as a bipolar gauge with selectable units
InHg/PSI, can be user reconfigured as a bipolar gauge with selectable units
PSIG, kPa, mbar, bar, atm, kg/cm
PSIG, kPa, MPa, mbar, bar, atm, kg/cm2, g/cm2, cmH2O, oz/in2, ftH2O, inH2O, mmHg, torr, inHg
PSIG, kPa, MPa, mbar, bar, atm, kg/cm2, g/cm2, cmH2O, oz/in2, ftH2O, inH2O, mmHg, torr, inHg
PSIG, kPa, MPa, mbar, bar, atm, kg/cm
PSIG, kPa, MPa, bar, atm, kg/cm2, oz/in2, ftH2O, inH2O, inHg
PSIG, kPa, MPa, bar, atm, kg/cm
PSIG, kPa, MPa, bar, atm, kg/cm2, ftH2O, inHg
PSIG, kPa, MPa, bar, atm, kg/cm
PSIG, MPa, bar, atm, kg/cm2, ftH2O, inHg
PSIG, MPa, bar, atm, kg/cm2, ftH2O, inHg
PSIG, MPa, bar, atm, kg/cm
2,
g/cm2, cmH2O, oz/in2, ftH2O, inH2O, mmHg, torr, mmH2O, inHg
2,
g/cm2, cmH2O, oz/in2, ftH2O, inH2O, mmHg, torr, inHg
2,
oz/in2, ftH2O, inHg
2,
ftH2O, inHg
2
NIST traceability certificate with 5 test points and data