Meriam DAI2000, DGI1500, FDN0010, DGI1000, DGI0020 User manual

...
PRESSURE SENSOR MODULES
Pressure Measurement Modules for use with
MFT 4000, MFT 4010 and MFT 4020
Pressure Sensor Modules from Meriam Process Technology are available in a several types and ranges to provide the ultimate in flexibility and convenience for calibration standards. Meriam pressure modules are used with any MFT 4000 Series Multi-Function Tester by installing modules into any of three (3) MFT sensor bays. The modules are completely interchangeable and can also be hot-swapped, when needed, to facilitate most any pressure testing needs (restrictions on module installation and removal apply to Intrinsically Safe MFT units).
NIST traceable accuracy of ± 0.025% FS is available for most pressure ranges and types. This accuracy statement includes all effects of linearity, hysteresis, repeatability and temperature from 23° F to 122° F (no partial temperature compensation here!). A NIST certificate and calibration data are included with each module for quality assurance purposes. A reduced specification of ± 0.025% FS ± 0.004% FS per °F (Ref. Temp. = 70° F) is also available for most pressure ranges. This specification offers customers a cost savings with virtually the same measurement accuracy when used in conditioned or temperate environments.
Calibration coefficients for all modules are stored in non­volatile sensor module memory. This means any pressure module can be used in any Meriam MFT without manual programming of coefficients. When a module is installed or the MFT is powered up, the MFT senses the module bay connection, polls the module for the calibration coefficients and configures the display for pressure readout in the last engineering unit selected. Engineering units, including PSI, centimeters or inches of H 20°C, or 60°F), kg/cm inches of Hg (at 0°C ref. temp), are all selectable from the MFT keypad.
Meriam sensor modules are designed to be re-calibrated in the field with appropriate local calibration standards. The MFT’s Rcal feature allows customers to adjust zero (offset) and full scale (gain) plus up to seven (7) addition points on the calibration curve. Dead weight testers are generally used to apply a known pressure to a module. The Rcal feature allows the user to key in the applied pressure value, which lets the module know how to correct its calibration curve. Meriam also provides factory re-calibration services for pressure modules including complete documentation and certification.
2
, kPa, mBars, Bars, millimeters or
Certifications Available
CE Mark (standard)
NIST traceability certificate (standard)
Intrinsically Safe, MET Laboratories per
CSA C22.2 & UL Class I Division I (standard on all pressure modules)
O (at 4°C,
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Available Pressure Modules
Accuracy Model Pressure range Type ±.10% * FDN0010 0-10” H
FDN0020 0-20” H x x ±.025% * x x OR x x ±.025% x ±.004% x FS / °F x x x
DN0100 0-100” H2O Differential DN0200 0-200” H2O Differential DN0400 0-400” H2O Differential DN2000 0-2000” H2O Differential GI0020 0-20 PSIG Gauge GI0200 0-200 PSIG Gauge GI0500 0-500 PSIG Gauge GI1000 0-1000 PSIG Gauge GI1500 0-1500 PSIG Gauge GI2000 0-2000 PSIG Gauge AI2000 0-2000 mm Hg Absolute
O Differential
2
O Differential
2
DDN0400 Module
xAI0900 0-900 mm Hg Absolute xAI5200 0-100 PSIA Absolute ±.05% * GGI3000 0-3000 PSIG Gauge
* Includes all effects of linearity, hysteresis, repeatability and temperature from 23° F to 122° F
where x = D for ±0.025% FS with full temp comp
MFT with T/C, DDN0400 &
VMA Modules installed
E for ±0.025% FS ±0.004% FS per °F
Specification
Temperature:
Storage: -40° F to 140° F (-40° C to 60° C) Operating: 23° F to 122° F (-5° C to 50° C)
Model codes and uses
x
DN: Differential Non-isolated – clean, dry, non­ corrosive gases only x
GI: Gauge Isolated – medias compatible with 316SS x
AI: Absolute Isolated – medias compatible with 316SS
Accessories
P/N A900529-00015 VMA Test Lead Kit - banana plugs on 9" breakouts both ends, assorted connectors
P/N B34686 961P pressure pump with integral variator, bleed valve, 1/8" FNPT connection, 0 - 145 PSIG P/N B34700 961V vacuum pump with integral variator, bleed valve, 1/8" FNPT connection, -650 mm Hg
P/N A34386 Low Pressure Connector Kit (not for use above 250 PSIG) P/N A34102 1/8" O.D. tubing to 1/8" MNPT connector (not for use above 250 PSIG) P/N A34112 1/8" O.D. tubing union tee (not for use above 250 PSIG) P/N A34099-1 1/4" MNPT X 1/8" FNPT reducer bushing, brass P/N A34103-2 10 ft of 1/8" O.D. nylon tubing (not for use above 250 PSIG) P/N A36856 Push to read equalizing manifold for DP modules, brass, 1/8"FNPT connections
P/N A900020-90502 High Pressure Connection Kit, 3’ hose (not for use above 9,000 PSIG) P/N A900020-90503 High Pressure Connection Kit, 1’ hose (not for use above 9,000 PSIG) P/N A900020-90501 5 pcs 1/4" MNPT x HP Quick Connect Fitting (not for use above 9,000 PSIG) P/N A900436-00001 1 pcs 1/8" MNPT HP Quick Connect fitting (not for use above 9,000 PSIG)
NIST Traceable Accuracy: See specification in Table above
Connections: 1/8” FNPT, 316ss
Pressure limits:
x
GI / xAI modules: 2X range / 316ss compatible
x
DN modules: 2X range when pressurized on high side only, 150 PSI when applied simultan­eously to high and low sides
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Pressure Module Units Resolution Table No. 1
10" WC 20" WC 200" WC 400" WC 2000" WC 20 PSIG 200 PSIG 500 PSIG 1000 PSIG Press. Unit
mm Hg
in. Hg
cm H2O
Bar
mBar
KPa
Kg/cm²
inH2O 4°C
inH2O 60°C
inH2O 20°C
PSI
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xxx.xx xxxx.x xxxx.x xxxx.x
x.xxxx x.xxxx xx.xxx xx.xxx xxx.xx xx.xxx xxx.xx xxx.xx xxx.xx
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xxx.xx xxxx.x xxxx.x xxxx.x
x.xxxx x.xxxx x.xxxx x.xxxx xx.xxx x.xxxx xx.xxx xx.xxx xx.xxx
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xxx.xx xxxx.x xxxx.x xxxx.x
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xx.xxx xxx.xx xxx.xx xxx.xx
x.xxxx x.xxxx x.xxxx x.xxxx xx.xxx x.xxxx xx.xxx xx.xxx xx.xxx
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xxx.xx xxxx.x xxxx.x xxxx.x
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xxx.xx xxxx.x xxxx.x xxxx.x
xx.xxx xx.xxx xxx.xx xxx.xx xxxx.x xxx.xx xxxx.x xxxx.x xxxx.x
x.xxxx x.xxxx xx.xxx xx.xxx xxx.xx xx.xxx xxx.xx xxx.xx xxx.xx
Pressure Module Units Resolution Table No. 2
1500 PSIG 2000 PSIG 3000 PSIG 900mm 2000mm 5200mm Press. Unit Hg Abs Hg Abs Hg Abs
mm Hg
in. Hg
cm H2O
Bar
mBar
KPa
Kg/cm²
inH2O 4°C
inH2O 60°C
inH2O 20°C
PSI
xxxx.x xxxxx xxxxx xxxx.x xxxx.x
xxx.xx xxxx.x xxxx.x xxx.xx xxx.xx xxx.xx
xxxx.x xxxxx xxxxx xxxx.x xxxx.x xxxx.x
xx.xxx xxx.xx xxx.xx xx.xxx xx.xxx xx.xxx
xxxx.x xxxxx xxxxx xxxx.x xxxx.x xxxx.x
xxx.xx xxxx.x xxxx.x xxx.xx xxx.xx xxx.xx
xx.xxx xxx.xx xxx.xx xx.xxx xx.xxx xx.xxx
xxxx.x xxxxx xxxxx xxxx.x xxxx.x xxxx.x
xxxx.x xxxxx xxxxx xxxx.x xxxx.x xxxx.x
xxxx.x xxxxx xxxxx xxxx.x xxxx.x xxxx.x
xxx.xx xxxx.x xxxx.x xxx.xx xxx.xx xxx.xx
xxxx.x
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...