Mercoid SBLT2, SBLTX Data sheet

Series
6-45/64
(170.26)
Ø1
(Ø25.40)
1/2 NPT
SUBMERSIBLE CABLE
BREATHER
TUBE
BREATHER TUBE
SUBMERSIBLE CABLE
1/2 NPT
6-15/16
[176.21]
Ø1
[Ø25.40]
SBLT2
&
Submersible Level Transmitters
Perfect for Ground Water and Wells, Lightning Protected or Intrinsically Safe
SBLT2
SBLTX
SBLT2 and SBLTX Submersible Level Transmitters are manufactured for years of
trouble free service in the harshest applications. Both measure the height of liquid above the position in the tank referenced to atmospheric pressure. The transmitters consist of a piezoresistive sensing element, encased in a 316 SS housing. Bullet nose design protects the diaphragm from damage. The SBLT2 incorporates lightning and surge protection utilizing dual arrestor technology, grounded to case, eliminating both power supply surges and lightning ground strike transients (surge protection is not guaranteed and is not covered by warranty). The SBLTX is UL approved intrinsically safe for use in hazardous locations when used with proper barrier. Units come equipped with a 270-pound tensile strength shielded and vented cable. Ventilation tube in the cable automatically compensates for changes in atmospheric pressure above the tank. The vent is protected with a maintenance free filter eliminating particulate or water droplets from entering the transducers.
• Excellent chemical compatibility
• Lightning and surge protection on SBLT2 models
• Maintenance free vent filter
• UL approved intrinsically safe on SBLTX models
• Slim design for tight applications
APPLICATIONS
Well monitoring; Ground water monitoring; Environmental remediation; Surface water monitoring; Down hole; Water tanks.
Model
LEVEL
SBLT2-5-40-ETFE SBLT2-10-40-ETFE SBLT2-15-60-ETFE SBLT2-20-60-ETFE SBLT2-5-40 SBLT2-10-40 SBLT2-15-60 SBLT2-20-60 SBLT2-3.5M-5M SBLT2-5M-10M SBLT2-10M-18M
OPTIONS Intrinsically Safe Approval- Change model number from SBLT2 to SBLTX
Custom ranges or Cable Lengths- Contact the factory
Submersible
Level Transmitters,
Range psi (ft w.c.) [m w.c.]
5 (11.54) [3.52] 10 (23.09) [7.04] 15 (34.63) [10.56] 20 (46.18) [14.08] 5 (11.54) [3.52] 10 (23.09) [7.04] 15 (34.63) [10.56] 20 (46.18) [14.08]
4.97 (11.48) [3.5]
14.21 (32.81) [10]
25.58 (59.06) [18]
Cable Length ft (m)
40 (12.2) 40 (12.2) 60 (18.3) 60 (18.3) 40 (12.2) 40 (12.2) 60 (18.3) 60 (18.3)
16.40 (5)
32.81 (10)
59.06 (18)
Cable Type
ETFE ETFE ETFE ETFE Polyurethane Polyurethane Polyurethane Polyurethane Polyurethane Polyurethane Polyurethane
SPECIFICATIONS Service: Compatible liquids. Wetted Materials: 316 SS, 316L SS,
epoxy; Cable: Polyurethane or ETFE; Bullet nose: PVC.
Accuracy: ±0.25% of full-scale. Temperature Limit: SBLT2: 0 to
150°F (-18 to 66°C); SBLTX: 0 to 176°F (-18 to 80°C).
Compensated Temperature Range:
SBLT2: 0 to 140°F (-18 to 60°C); SBLTX: 0 to 176°F (-18 to 80°C).
Thermal Effect: ±0.02% full-scale/°F. Pressure Limit: 2X full-scale. Power Requirement: SBLT2: 13 to 30
VDC; SBLTX: 10 to 28 VDC.
Output Signal: 4 to 20 mA DC, 2-wire.
ACCESSORIES MTL5041, intrinsically safe galvanic isolator MTL7706, intrinsically safe zener barrier
A-297, Dessicant Filter for vent tube. Removes humidity for
protection of the sensor. Changes color to show saturation
A-625, 316 SS Cable Hanger use with NPT option for attaching chain for easy pulling out of application
Response Time: 50 ms. Max. Loop Resistance: 900 Ω at 30
VDC.
Electrical Connections: Wire pigtail. Mounting Orientation: Suspended in
tank below level being measured.
Weight: 2.2 lb (1.0 kg). Electrical Protection: SBLT2:
Lightning and surge protection; SBLTX: None. Agency Approvals: SBLT2: None; SBLTX: CE, cUL intrinsically safe for Class I, Div. 1, Groups A, B, C, D; Class II, Div. 1, Groups E, F, G; Class III Div. 1. (According to control drawing 01-700797-00).
344
DWYER INSTRUMENTS, INC. | www.dwyer-inst.com
Loading...