Issue Number 1.
Date of issue; November 2020
Reprocessing Instructions
Surgical Handpieces & Micro Motors
The following instructions are applicable for Surgical Hand Pieces and Micro Motors supplied
by Mercian.
These instructions conform to the Medical Directive 93/42/EEC (Annex 1 section 13)
These instructions are intended for use only by persons with the required specialist
knowledge and training.
Precautions;
• The Handpiece must be allowed to cool to ambient conditions before being used
• Hand Pieces must be tested prior to every procedure to ensure they are rotating
freely to prevent them overheating during surgery.
• If any unusual noise or overheating is noticed withdraw from use immediately and
return for servicing.
• Do not cover up Micro Motor Air-vents with tape, this will result in overheating.
• Return for service to Mercian Annually
• As per HTM 0101 1.138 alkaline detergents in the pH range 8.0 to 10.0 are preferred
Limitations on Reprocessing:
• Repeated processing has minimal effect on these devices.
• End of life is normally determined by wear and damage in use.
• No specific limitations on the number of reprocessing cycles for the device.
Manual cleaning of Handpieces or Micro Motors:
• Clean & disinfect the exterior of the Handpiece/Micro Motor with a clean damp cloth
and suitable disinfection solution.
Note; Products containing acetone, chlorine, and bleach are not recommended as
disinfectants.
• Visually inspect the Handpiece & Micro Motor for any remaining debris or damage
N.B; Hand Pieces are not suitable for cleaning in an ultrasonic bath.