Mensor CPT6020 Specifications

Precision Pressure Transducer Basic Version Model CPT6020
Applications
Pressure calibration
High accuracy pressure monitoring
Aerospace
Special Features
Accuracy: 0.020% FS
Range: 0-10 in. H2O (25 mbar) up to 15,015 psi (1001 bar)
Temperature compensation: -20 to 75 oC
RS-232 or RS-485 communication
Compact rugged design
Calibration technology
WIKA data sheet CT 25.13
Description
The CPT6020 Precision Pressure Transducer is a self-contained pressure sensing device that provides high accuracy pressure measurements. This transducer incorporates a low hysteresis silicon sensor with electronically compensated pressure linearity over the compensated temperature range. The CPT6020 transducer is characterized over the full pressure and temperature
range to achieve 0.020% FS accuracy. This specication includes
linearity, hysteresis, repeatability and temperature errors. Also featured is an output which is updated at a rate of 50 readings per second (20 ms).
Application
The CPT6020 is used in OEM applications where a high accura-
cy pressure sensing is required. Examples are: ow calibrators,
humidity calibrators, pressure controllers, aerospace wind tunnel calibration, automotive sensor testing, hydrology, oceanography, in the aviation and space industries in general, or wherever high accuracy pressure measurement and long-term calibration stability are valued. It can also be used as a transfer standard or in pressure calibration and testing areas of production facilities.
Precision Pressure Transducer, Model CPT6020
Functions
The model CPT6020 Precision Pressure Transducer has an
RS-232 or RS-485 interface. The RS-485 interface oers multi-drop
capability with cabling that includes power and communications.
Four dierent baud rates can be selected and the transducer can
be located up to 4,000 feet from the host. System designers appre­ciate these high accuracy remotely accessible transducers that
are not tied to a front panel. Each transducer can be congured for
gauge or absolute pressure types. With a recalibration time of 185
days and a high resolution of 8 signicant digits, the CPT6020 is exible enough to be used in a wide variety of applications.
Design
The 316L SS construction and IP67 rating are an asset when utiliz-
ing in corrosive or wet environments. Its compact design oers an
advantage in miniaturization of product design in many OEM appli­cations. The pressure connection and housing can be customized
to t your application. Standard ttings are easily changed using
the AN-4 or the Autoclave® F250C connection.
WIKA data sheet CT 25.13 ∙ PN 0019662001A ∙ 09/2018
Page 1 of 4
Specications Model CPT6020
Measurement Specification
Accuracy
Measuring ranges
Gauge pressure 0 ... 0.36 to 0 ... 1500 psig
Bi-directional pressure
Absolute pressure 0 ... 5 psia to 0 ... 15,015 psia
Calibration interval 185 days
Pressure units 39 and 1 user defined
CPT6020 as barometric reference
Measuring range 8 ... 17 psi abs. (552 ... 1,172 mbar abs.)
Accuracy
1) It is dened by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference
2) The negative portion of a bidirectional range has the same accuracy as the equivalent positive range.
General Specifications
Case
Orientation effects Negligible – completely removable with re-zeroing
Dimensions See technical drawings
Weight ~250 g (depending on range)
Display
Resolution
Warm-up time 15 min.
Connections
Pressure connections SAE J514/JIC or Autoclave® F250C (for pressure ranges > 6000 psi)
Overpressure limit 2X proof, 3X burst; static pressure < 50 psig
Materials, wetted parts Ranges ≤ 5 psi Silicon, 316 SS, glass filled resins, epoxy
Pressure media Ranges ≤ 5 psi – clean, dry, non-corrosive gases
Voltage supply
Power supply 9 to 18 VDC (12 VDC nominal)
Power consumption < 26 mA at 12 VDC +/-5%
Permissible ambient conditions
Compensated temperature range
Operating temperature range
Storage temperature range -40 to 85 oC
Humidity 0 ... 95% r.h. (non -condensing)
Operating altitude <3000 meters (10,000 feet)
Operating environment IP-67
Measure port internal volume <1 cc
Reference port internal volume ~ 40 cc
Communication
Interface RS-232 or RS-485
Baud rate
Measuring rate 50 values/second, default - (factory adjustable)
1)
0.020% FS
(0 ... 25 mbar to 0 ... 100 bar)
2)
± 0.18 to -15 ... 1500 psig
(± 12.5 mbar to -1 ... 100 bar)
(0 ... 350 mbar to 0 ... 1001 bar abs.)
1)
instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range with recommended zero point adjustment every 30 days.
0.020% of reading
100 ppb or better
Ranges > 5 to 1500 psi 316 SS
Ranges >1500 psi 316 SS, Fluorocarbon Rubber
Ranges > 5 psi – media compatible with the listed wetted parts.
-20 to 75 oC (-4 to 167 oF)
-40 to 85 oC
(-40 to 185 oF)
(-40 to 185 oF)
Default 57,600 baud - 9600, 19200 and 115200 user selectable
Page 2 of 4
WIKA data sheet CT 25.13 ∙ PN 0019662001A ∙ 09/2018
≤1500 psi
1500 to 6000 psi
6000 to 15000 psi
Certificates
Compliance CISPR 11 & A1, IEC 61000-4-2, 3, 4, 5, 6, 11, IEC 61000 A1, A2, IEC 61010-1
Calibration
3) Calibration in a vertical position.
3)
A2LA calibration certicate
Dimensions in [mm] in.
6 PIN M8 CONNECTOR
1/16" BARBED
REFERENCE PORT
31.8mm
1.25in
85.7mm
3.37in
6 PIN M8 CONNECTOR
117.6mm
4.63in
SEAL SCREW
31.8mm
1.25in
85.7mm
3.37in
122.9mm
4.84in
6 PIN M8 CONNECTOR
SEAL SCREW
31.8mm
1.25in
95.0mm
3.74in
18.9mm .74in
SAE J514 37° FLARE PORT
7/16-20 THREAD
18.8mm .74in
SAE J514 37° FLARE PORT
7/16-20 THREAD
Note: Reference port is for gauge type sensors only; for absolute type, and sealed gauge type, the port is plugged.)
25.3mm
1.00in
AUTOCLAVE F250C
FEMALE PORT
Page 3 of 4WIKA data sheet CT 25.13 ∙ PN 0019662001A ∙ 09/2018
Scope of Delivery
Accessories
Precision Pressure Transducer model CPT6020
Operating instructions
ISO 17025, A2LA accredited calibration certicate
Pressure adapter (as specied)
5 ft. connection cable with ying leads
Power supply & communication cable
Pressure adapters
Ordering Information
CPT6020 / Instrument version / Pressure unit / Type of pressure / Start of measuring range / End of measuring range / Accuracy / Type of certicate / Mounting position / Interface / Baud rate / Output mode / Pressure adapter / Additional order information
© 2011 WIKA Alexander Wiegand SE & Co. KG, all rights reser ved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
WIKA data sheet CT 25.13 ∙ PN 0019662001A ∙ 09/2018
09/2018 UM
Mensor
201 Barnes Drive San Marcos, Texas 78666 Tel. (512) 396-4200 Fax (512) 396-1820 sales@mensor.com www.mensor.com
Page 4 of 4
Loading...