![](/html/40/4045/4045ac16a805d269290db314a04a241eee0bebe75fdf8d1764c8e5330e69b959/bg1.png)
Precision Pressure Transducer
Basic Version
Model CPT6020
Applications
■
Pressure calibration
■
High accuracy pressure monitoring
■
Pressure sensing in critical applications
■
Aerospace
Special Features
■
Accuracy: 0.020% FS
■
Range: 0-10 in. H2O (25 mbar) up to 15,015 psi (1001 bar)
■
Temperature compensation: -20 to 75 oC
■
RS-232 or RS-485 communication
■
Compact rugged design
Calibration technology
WIKA data sheet CT 25.13
Description
The CPT6020 Precision Pressure Transducer is a self-contained
pressure sensing device that provides high accuracy pressure
measurements. This transducer incorporates a low hysteresis
silicon sensor with electronically compensated pressure linearity
over the compensated temperature range. The CPT6020
transducer is characterized over the full pressure and temperature
range to achieve 0.020% FS accuracy. This specication includes
linearity, hysteresis, repeatability and temperature errors. Also
featured is an output which is updated at a rate of 50 readings per
second (20 ms).
Application
The CPT6020 is used in OEM applications where a high accura-
cy pressure sensing is required. Examples are: ow calibrators,
humidity calibrators, pressure controllers, aerospace wind tunnel
calibration, automotive sensor testing, hydrology, oceanography,
in the aviation and space industries in general, or wherever high
accuracy pressure measurement and long-term calibration stability
are valued. It can also be used as a transfer standard or in pressure
calibration and testing areas of production facilities.
Precision Pressure Transducer, Model CPT6020
Functions
The model CPT6020 Precision Pressure Transducer has an
RS-232 or RS-485 interface. The RS-485 interface oers multi-drop
capability with cabling that includes power and communications.
Four dierent baud rates can be selected and the transducer can
be located up to 4,000 feet from the host. System designers appreciate these high accuracy remotely accessible transducers that
are not tied to a front panel. Each transducer can be congured for
gauge or absolute pressure types. With a recalibration time of 185
days and a high resolution of 8 signicant digits, the CPT6020 is
exible enough to be used in a wide variety of applications.
Design
The 316L SS construction and IP67 rating are an asset when utiliz-
ing in corrosive or wet environments. Its compact design oers an
advantage in miniaturization of product design in many OEM applications. The pressure connection and housing can be customized
to t your application. Standard ttings are easily changed using
the AN-4 or the Autoclave® F250C connection.
WIKA data sheet CT 25.13 ∙ PN 0019662001A ∙ 09/2018
Page 1 of 4
![](/html/40/4045/4045ac16a805d269290db314a04a241eee0bebe75fdf8d1764c8e5330e69b959/bg2.png)
Specications
Model CPT6020
Measurement Specification
Accuracy
Measuring ranges
Gauge pressure 0 ... 0.36 to 0 ... 1500 psig
Bi-directional pressure
Absolute pressure 0 ... 5 psia to 0 ... 15,015 psia
Calibration interval 185 days
Pressure units 39 and 1 user defined
CPT6020 as barometric reference
Measuring range 8 ... 17 psi abs. (552 ... 1,172 mbar abs.)
Accuracy
1) It is dened by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference
2) The negative portion of a bidirectional range has the same accuracy as the equivalent positive range.
General Specifications
Case
Orientation effects Negligible – completely removable with re-zeroing
Dimensions See technical drawings
Weight ~250 g (depending on range)
Display
Resolution
Warm-up time 15 min.
Connections
Pressure connections SAE J514/JIC or Autoclave® F250C (for pressure ranges > 6000 psi)
Overpressure limit 2X proof, 3X burst; static pressure < 50 psig
Materials, wetted parts Ranges ≤ 5 psi Silicon, 316 SS, glass filled resins, epoxy
Pressure media Ranges ≤ 5 psi – clean, dry, non-corrosive gases
Voltage supply
Power supply 9 to 18 VDC (12 VDC nominal)
Power consumption < 26 mA at 12 VDC +/-5%
Permissible ambient conditions
Compensated temperature range
Operating temperature range
Storage temperature range -40 to 85 oC
Humidity 0 ... 95% r.h. (non -condensing)
Operating altitude <3000 meters (10,000 feet)
Operating environment IP-67
Measure port internal volume <1 cc
Reference port internal volume ~ 40 cc
Communication
Interface RS-232 or RS-485
Baud rate
Measuring rate 50 values/second, default - (factory adjustable)
1)
0.020% FS
(0 ... 25 mbar to 0 ... 100 bar)
2)
± 0.18 to -15 ... 1500 psig
(± 12.5 mbar to -1 ... 100 bar)
(0 ... 350 mbar to 0 ... 1001 bar abs.)
1)
instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range with recommended zero point adjustment every 30 days.
0.020% of reading
100 ppb or better
Ranges > 5 to 1500 psi 316 SS
Ranges >1500 psi 316 SS, Fluorocarbon Rubber
Ranges > 5 psi – media compatible with the listed wetted parts.
-20 to 75 oC (-4 to 167 oF)
-40 to 85 oC
(-40 to 185 oF)
(-40 to 185 oF)
Default 57,600 baud - 9600, 19200 and 115200 user selectable
Page 2 of 4
WIKA data sheet CT 25.13 ∙ PN 0019662001A ∙ 09/2018