Mensor CPT6010 Specifications

CPT6010 Digital Pressure Transducer
Applications
High accuracy transfer standard Pressure standard in test & calibration stands Wind tunnels Hydrology, Oceanography Aerospace, Metrology
Features
Testing and Calibration Technology
Data Sheet CPT6010 • 04/2015
Absolute: 0-7.5 psia to 0-6000 psia
Gauge: 0-5 psig to 0-6000 psig
Pneumatic or Hydraulic RS-232 or RS-485 communication 303 stainless steel housing
Description
General
The CPT6010 Digital Pressure Transducer is a self-contained pressure sensing device that provides high accuracy pressure measurements in both the single and dual calibration models. This transducer incorporates a low hysteresis silicon sensor with electronically compensated pressure linearity over the specified temperature range.
The CPT6010 Transducer is characterized over the full pressure and temperature range to achieve 0.020% FS uncertainty. This uncertainty specification includes linearity, hysteresis, repeat­ability and temperature errors. Also featured is an output which is updated at a rate of 50 readings per second (20 ms).
Applications
The CPT6010 is used in OEM Applications where a high ac­curacy pressure sensing is required. It can also be used as a
Digital Pressure Transducer Model CPT6010
transfer standard or in pressure calibration and testing areas of production facilities.
Functional Flexibility
The CPT6010 Digital Pressure Transducer is a high accuracy pressure measurement component that uses either RS-232 or RS-485 to communicate with a host computer over long distances. Any MS-DOS compatible PC with an available serial output port can serve as the host controller.
System designers will appreciate the flexibility offered by having highly accurate pressure transducers that are not tied to a front panel and which may be located remotely. For remote operation the transducer equipped with RS-485 can be located up to 4,000 feet from the host. A simple cable can accommodate both the power and the two-way communications requirements.
Data Sheet CPT6010 • 04/2015 Page 1 of 2
Related Products: CPT6100 and CPT6180 Digital Pressure Transducer Pressure Systems
Specifications
Total Uncertainty
Calibration Stability after warmup
Calibration
Pressure Ranges
Special Pressure Ranges
Pressure Units
Resolution
Overpressure Limit
Compensated Temp. Range
Warm-up
Reading Rate
Communications
Max. Transmission
Multi-drop Capacity
0.020%FS. Total uncertainty (k=2) includes hysteresis, linearity, repeatability, reference standard, drift and temperature effects over the calibrated range for the calibration interval speci-
ed with periodic re-zeroing.
Less than 0.02% FS for six months.
Calibration Interval: 180 days Cal Uncertainties: 0.020% FS Calibration adjustment: Zero and Span. (Zero and span may be reset via the serial interface
without affecting the linearity.)
Abs: 0-7.5 psia up to 0-6000 psia. Gauge: 0-5 psig up to 0-6000 psig.
Vacuum or bidirectional ranges. Metric pressure units also available.
Selected from a list of 35: psi, inHg @0°C and 60°F, inH2O @4°C, 20°C and 60°F, ftH2O @ 4°C, 20°C and 60°F, mTorr, inSW @ 0C, ftSW @ 0C, ATM, bars, mbars, mmH2O @ 4°C, cm H2O @ 4°C, MH2O @ 4°C, mmHg @ 0°C, cmHg @ 0°C, Torr, hPa, mPa, kPa, Pa, D/cmsq, G/cmsq, Kg/ cmsq, mSW @ 0°C, OSI, PSF, TSF, TSI, µHg @
0°C, %fs. All seawater units are 3.5% salinity.
Up to 1 ppm, depending on measurement units and range.
150% FS or greater, depending on range
15 to 45 °C
10 minutes to rated accuracy
50 Hz
RS-232 or RS-485. LabVIEW®1 drivers are avail­able.
4000 feet (RS-485)
The maximum number of RS-485 Series 6000
transducers which can be connected to a single
host computer is 31.
Mechanical Shock
Case Size
Weight
Media Compatibility
Fittings
Power
Compliance
Options
1
LabVIEW® is a trademark of National Instruments Corporation.
5g max.
1.75” wide x 6.0” long (4.45 x 15.24 cm), not
including pneumatic and electrical.
Approximately 12 ounces (28.3 grams).
All media compatible with 316L stainless steel.
Pressure Port: 1/4 inch male NPT
Reference Port: 1/16 inch barb (gauge instru­ments only)
9-18 VDC, 45mA @ 12 VDC 6 pin Bendix connector #PT02E-10-6P
Conforms to CE standards EN 50081-1, EN 50082-1, EN 50081-2, and EN 50082-2.
Relief Valves —up to 1000 psig Custom ranges.
The calibration program at Mensor is accredited by A2LA as
complying with both the ISO/IEC 17025:2005 and the ANSI/ NCSL Z540-1-1994 standards. All Mensor primary standards are traceable to NIST. Mensor is registered to ISO9001:2008.
Since product innovation is a continuous process at Mensor,
we reserve the right to change specications without notice.
Represented by:
Mensor
201 Barnes Drive San Marcos, Texas 78666 Toll Free: 800-984-4200 Tel: 512-396-4200 Fax: 512-396-1820 Email: sales@mensor.com
Data Sheet CPT6010 • 04/2015Page 2 of 2