Generation/measurement of 0 ... 30 mA and voltage
supply DC 24 V
■
Data logger with high measuring rate and large memory
■
A calibration assistant guides you through the calibration
Calibration technology
WIKA data sheet CT 15.51
Description
General information
The model CPH7000 process calibrator is a precise, portable
calibrator for the calibration and checking of analogue
pressure measuring instruments, pressure transmitters and
process transmitters. Furthermore, pressure switches can
be checked and the switching point determined. With the
CPH7000, not only can transmitters be checked, but also
simulated and tested.
Design
The CPH7000 optionally features an integrated reference
pressure sensor and also a manual pressure generation, with
which pressures of -0.85 ... +25 bar (12.3 ... +360 psi) can be
generated. It is possible to simultaneously power an external
transmitter via the electrical module and to measure (or also
simulate) its output signal.
Portable process calibrator model CPH7000 with
optional hand pump
Functions
The calibrator oers the possibility to set calibration routines
quickly and easily, but also to run precongured calibration
routines and automatically save the measured values. Over
the WIKA-Wireless interface, the completed calibration
processes can be transmitted to a PC. This data can
subsequently be evaluated and archived using WIKA-Cal
software. Thus a completely paperless transmitter calibration
is possible with just the CPH7000.
Accuracy
The CPH7000 is compensated over a wide temperature
range of 10 ... 50 °C (50 ... 122 °F) and achieves an accuracy
of 0.025 % of span. In order to avoid intricate calculations,
the measured values can also be displayed directly in
customer-specic units.
∙ 03/2017
Data sheets showing similar products and accessories:
Test pumps, hydraulic; model CPP series; see data sheet CT 91.05
Test pump, pneumatic; model CPP30; see data sheet CT 91.06
Calibration software; WIKA-Cal; see data sheet CT 95.10
Page 1 of 12WIKA data sheet CT 15.51
Features
For pressures greater than 25 bar (362.6 psi), there are the
model CPT7000 external pressure sensors.
Thus pressure measurement and calibration is possible in
further pressure measuring ranges and accuracies.
An optional atmospheric module records and documents the
Software
The calibration software for the CPH7000 is WIKA-Cal.
WIKA-Cal, alongside PC-supported calibration and the
logger function, also oers the management of the calibration
and instrument data in an SQL database. The data transfer is
achieved completely wirelessly via WIKA-Wireless.
environmental parameters important for a calibration, such as
atmospheric pressure, air humidity and ambient temperature.
Certied accuracy
For the model CPH7000 process calibrator, the accuracy
Complete service case
The process calibrator, developed specically for
maintenance and service operations, is delivered in a
is certied in a factory calibration certicate accompanying
the instrument. On request, we can provide a DKD/DAkkS
calibration certicate.
portable case system and, depending on requirements, can
be equipped with, for example, model CPT7000 external
pressure sensors, a Pt100 temperature sensor or a portable
system with storage bag.
5) The internal reference sensor is only available in combination with a pneumatic unit.
6) It is dened by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
7) Calibrated at 23 °C (74 °F) and in vertical mounting position.
8) In the event of interference caused by high-frequency electromagnetic elds in a frequency range of 200 ... 300 MHz, an increased deviation of up to 0.1 % is expected for the current
9) The barometric reference can be used to switch pressure types, absolute <=> gauge. With gauge
measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range during a periodic
zero point adjustment.
output function.
pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry out a complete absolute pressure emulation.
7)
8)
WIKA data sheet CT 15.51 ∙ 03/2017 Page 3 of 12
Power supply unit
Input voltageAC 100 ... 240 V, 50 ... 60 Hz
Output voltageDC 12 V
Nominal Output current2,500 mA
Permissible ambient conditions
Operating temperature0 ... 40 °C (32 ... +104 °F); up to 90 % r. h. (non-condensing)
Storage temperature-40 ... +70 °C (-40 ... +158 °F)
Relative humidity20 ... 80 % r. h. (non-condensing)
Temperature probe Pt100
1)
Measuring range-50 ... +250 °C (-58 ... +482 °F)
Accuracy1/10 DIN, classe B ±0.1 °C
Probe length200 mm (7.87 in)
Probe diameter3 mm (0.12 in)
Cable length1 m (3.28 ft)
Connection to CPH7000max. 1 temperature probe
2)
User-dened RTD probeInput of the coecients of R0, A, B and C
Ambient module
1)
Measuring range
Relative humidity0 ... 100 % r. h.
Temperature-40 ... +125 °C (-40 ... +257 °F)
Accuracy
Relative humidity±1.8 % r. h.
Temperature±0.2 °C (0.36 °F)
Connection to CPH7000max. 1 ambient module
WIKA-Wireless
4)
2)
Frequency range2,400 ... 2,500 MHz
HF output powermax. 2 dBm (+ 2 dBi)
Number of channels79
Channel spacing1/2 MHz
Bandwidth80 MHz
Output power4 dBm / 10 mW
1) Available as an option
2) Temperature probe and ambient module use the same connection.
Cannot be used at the same time.
6) It is dened by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range during a periodic
zero point adjustment.
7) Calibrated at 23 °C (74 °F) and in vertical mounting position
Internal transmission uidSynthetic oil (only for measuring ranges up to 25 bar (360 psi))
Permissible ambient conditions
Medium temperature-20 ... +60 °C (-4 ... +140 °F)
Operating temperature-20 ... +60 °C (-4 ... +140 °F)
Storage temperature-20 ... +80 °C (-4 ... +176 °F)
Relative humidity0 ... 95 % r. h. (non-condensing)
Temperature compensation10 ... 60 °C (50 ... 140 °F)
Temperature coecientZero point = 0.1 % / 10 K
Span = 0.1 % / 10 K
Case
MaterialStainless steel
Connection to the CPH7000Option: External operation via 1 m or 3 m (3.28 ft or 9.84 ft) connection cable (plug-and-play)
Ingress protectionIP65 / IP67 when connected
Dimensionssee technical drawing
Weightapprox. 230 g (0.5 lbs.)
in addition)
Approvals
LogoDescriptionCountry
EU declaration of conformity for CPH7000
■
EMC directive
EN 61326-1:2013 emission (group 1, class B) and interference immunity (portable
measuring equipment)
■
R&TTE directive
EN 300 328 harmonised frequency range 2,400 … 2,500 MHz is used; Bluetooth®
Classic, max. transmission power 10 mW. The instrument may be used without
limitations in the EU and also CH, NO and LI.
■
Protection of health and safety
EN 62479
EU declaration of conformity for CPT7000
■
EMC directive
EN 61326 emission (group 1, class B) and interference immunity (industrial application)
■
Pressure equipment directive
PS > 200 bar, module A, pressure accessory
European Union
European Union
Certicates
Certicate
CalibrationStandard: 3.1 calibration certicate per EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval1 year (dependent on conditions of use)
Approvals and certicates, see website
Page 6 of 12WIKA data sheet CT 15.51 ∙ 03/2017
Dimensions in mm (in)
Digital process calibrator model CPH7000
Reference pressure sensor
model CPT7000
WIKA data sheet CT 15.51 ∙ 03/2017 Page 7 of 12
Application icons (app)
The home screen is very clearly subdivided into applicationoriented apps:
Measure:
Display of 3 dierent measurements
Logger:
Simultaneous recording of up to 3 signals
Info:
All instrument information available at a glance
Remote:
WIKA-Wireless radio transmission settings
Calibrate:
Setting of calibrations using calibration assistant
Switch test:
Testing of pressure switches (NC or NO)
Settings:
General instrument settings
Service:
All service data at a glance
Special operating modes
Operating mode: Measure
Features
■
3 dierent measurements in one view
■
30 pressure units + 2 programmable units
■
Resolution: Up to 4 decimal places
■
Graphical display via bargraph
■
Optionally settable functions: Min/Max/Tare/Filter/Alarm
min/Alarm max/Mean value/Rate/Sensor temperature
Applications
■
Measurement of operating/process pressures
■
Comparative measurements with test items (power supply
and display for the test item through the CPH7000)
■
Maximum and Minimum memory (e.g. for leak testing)
■
Alarm function for safety testing
For further information see the operating instruction.
Representation of possible
measuring channels
Selection of the type of
measurement or calibration
Page 8 of 12WIKA data sheet CT 15.51 ∙ 03/2017
Operating mode: Calibrate
Menu screen of the
application “Calibration”
Representation of calibration
results as table
Features
■
Calibration assistant
■
Supply with pressure, current and voltage
■
Calibration protocol is automatically saved
Applications
■
On-site calibration of pressure sensor and pressure
measuring instruments (without PC)
■
A calibration assistant guides you easily through the
calibration (following DKD/DAkkS). With this, the data
sets, including date and time, are recorded within the
CPH7000.
■
Prior to calibration, the calibration routines can be set
directly on the instrument or uploaded via WIKA-Cal
software.
■
Up to 100 calibrations can be stored
■
Re-calibrations possible
PC software available
Communication with WIKA-Cal calibration software via WIKAWireless
Representation of calibration
results as graph
Operating mode: Switch test
Switch test with external voltage
supply
Menu screen of the
application “Switch test”
Features
■
Pressure display on the closing and opening of the switch
■
Automatic calculation of the hysteresis
V
outVin
Switch test with voltage supply
DC 24 V of the CPH7000
+
+
s
-
-
mA GND
V
outVin
mA GND
Applications
■
On-site functional check of pressure switches (without PC)
■
Determination of the switch point accuracy and
repeatability
■
Determination of the switch point hysteresis
WIKA data sheet CT 15.51 ∙ 03/2017 Page 9 of 12
Operating mode: Logger
Menu screen of the
application „Logger“
Representation of logger
results as table
Features
■
Logging of max. 3 signals/measured value at the same
time
■
Automatic or manual data acquisition
■
Direct display as graph or table
■
Logger protocols are automatically saved
Applications
■
Logging of current, voltage, pressure and temperature
■
The logger menu guides one, step by step, through the
logger process. The data sets, incl. date and time, are
automatically saved in the CPH7000.
■
Before logging the data, routines can be set directly on the
instrument or uploaded via WIKA-Cal software.
■
Re-logging possible
PC software available
Communication with WIKA-Cal calibration software via
WIKA-Wireless
Representation of logger
results as graph
Operating mode: Simulation of transmitter signals
-+
Test item
4 ... 20 mA
V
outVin
Test
item
mA GND
+-
Features
Manual or automatic current source function
+
-
Applications
The CPH7000 can be connected in place of a transmitter
within a current loop and used as a current source. The
transmitter output signals from 0 ... 24 mA can be simulated
through manual input or automatically using the ramp and
step functions.
V
outVin
Page 10 of 12WIKA data sheet CT 15.51 ∙ 03/2017
mA GND
WIKA-Cal calibration software
Easy and fast creation of a high-quality calibration certicate
The WIKA-Cal calibration software is used for generating
calibration certicates or logger protocols for pressure
measuring instruments and is available as a demo version for
a cost-free download.
A template helps the user and guides him through the
creation process of a document.
In order to switch from the demo version to a full version of
the respective template, a USB key with the template has to
be purchased.
The pre-installed demo version automatically changes to
the selected full version when the USB key is inserted and
is available as long as the USB key is connected to the
computer.
■
Creation of calibration certicates for mechanical and
electronic pressure measuring instruments
■
A calibration assistant guides you through the calibration
■
Automatic generation of the calibration steps
■
Generation of 3.1 certicates per DIN EN 10204
■
Creation of logger protocols
■
User-friendly interface
■
Languages: German, English, Italian and more due with
software updates
For further information see data sheet CT 95.10
Calibration certicates can be created with the Cal-Template and logger protocols can be created with the Log-Template.
Cal Demo
Generation of calibration certicates limited to 2 measuring points, with
automatic initiation of pressures via a pressure controller.
Cal Light
Generation of calibration certicates with no limitations on measuring
points, without automatic initiation of pressures via a pressure
controller.
Calibration certificate
Kalibrierzeugnis
General
Allgemein
Calibration certificate
Kalibrierschein
Customer Demo customer
Kunde
Demo street
DE-12345 Demo city
Customer order No.
Kundenbestellnummer
Order date
Bestelldatum
Date of calibration 01/02/2017
Datum der Kalibrierung
OperatorDemo User
Bearbeiter
Environmental conditions
Umgebungsbedingungen
Temperature (20,0 ± 0,1) °C
Temperatur
Atmospheric pressure (1013,0 ± 0,1) hPa
Atmosphärischer Luftdruck
Rel. humidity (40 ± 1) %
Rel. Luftfeuchte
Measurement conditions
Messbedingungen
Working standard (WS) Pressure controller
Gebrauchsnormal (GN) Druckcontroller
ModelDemo CPC
Typ
Calibration markDemo 12345
Kalibrierzeichen
Scale range(0 ... 10) bar g
Anzeigebereich
Accuracy0.02 % FS
Genauigkeit
Demo laboratory
Tel.: 01234/56789
Demo street 123
Fax: 01234/56789
12345 Demo city
E-Mail: demo@demo.com
Demo country
www.demo.com
Certificate No.00000122
Zeugnis-Nr.
Device under test (DUT)
Kalibriergegenstand (KG)
Electrical gauge
Elektrisches Druckmessgerät
ModelD-10
Typ
Serial number demo1234
Seriennummer
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy0.05 % FS
Genauigkeit
Resolution (scale division) 0.0001 bar
Auflösung (Skalenwert)
Measuring point No. demo tag
Messstellen-Nr
Place of calibration
Ort der Kalibrierung
g = (9.80946 ± 0.01) m/s²
Demo place
Musterort
Calibration certificate
Kalibrierzeugnis
Measurement results
Messergebnisse
Reference value WS Reading DUT
Statement of compliance
Konformitätsaussage
Object keeps the specifications!
Der Kalibriergegenstand hält die Herstellerspezifikationen ein!
Creation of data logger test reports, limited to 5 measured values.
Page 2 / 2
Seite
Log
Creation of data logger test reports without limiting the measured
values.
WIKA data sheet CT 15.51 ∙ 03/2017 Page 11 of 12
Scope of delivery
■
Process calibrator model CPH7000
■
Power supply unit
■
Operating instructions
■
Service case with 2 connection cables (4 mm plugs)
■
3.1 calibration certicate per DIN EN 10204
Options
■
DKD/DAkkS calibration certicate
Accessories
Connection adapters
■
Various pressure adapters
Measuring equipment
■
External pressure sensors model CPT7000
■
Ambient module
■
Barometer
■
Pt100 measuring probe
■
Carrying system
■
Belt pocket for accessories
Process calibrator model CPH7000
Software
■
WIKA-Cal calibration software
Service case with process calibrator and accessories
Ordering information
CPH7000 / Version / Pressure generation / Unit / Pressure type / Measuring range / Accuracy / Type of certicate / Barometer /
Barometer calibration / Atmospheric module / Atmospheric module calibration / Temperature probe / Temperature probe
calibration / Electrical module calibration / Communication / Software / Pressure connection set / Carrying system / Transport
case / Further approvals / Additional ordering information
CPT7000 / Version / Unit / Pressure type / Measuring range / Process connection / Medium temperature / Wetted parts /
Special design for media / Accuracy / Type of certicate / Sensor cable / Further approvals / Additional ordering information
The specications given in this document represent the state of engineering at the time of publishing.
We reserve the right to make modications to the specications and materials.