Mensor CPH7000 Specifications

Portable process calibrator Model CPH7000
Applications
Calibration service companies and service industry
Measurement and control laboratories
Quality assurance
On-site calibration
Special features
Manual pressure generation of -0.85 … +25 bar (12.3 ... +360 psi)
Accuracy: 0.025 % FS (incl. calibration certicate)
Generation/measurement of 0 ... 30 mA and voltage supply DC 24 V
Data logger with high measuring rate and large memory
A calibration assistant guides you through the calibration
Calibration technology
WIKA data sheet CT 15.51
Description
General information
The model CPH7000 process calibrator is a precise, portable calibrator for the calibration and checking of analogue pressure measuring instruments, pressure transmitters and process transmitters. Furthermore, pressure switches can be checked and the switching point determined. With the CPH7000, not only can transmitters be checked, but also simulated and tested.
Design
The CPH7000 optionally features an integrated reference pressure sensor and also a manual pressure generation, with which pressures of -0.85 ... +25 bar (12.3 ... +360 psi) can be generated. It is possible to simultaneously power an external transmitter via the electrical module and to measure (or also simulate) its output signal.
Portable process calibrator model CPH7000 with optional hand pump
Functions
The calibrator oers the possibility to set calibration routines quickly and easily, but also to run precongured calibration
routines and automatically save the measured values. Over the WIKA-Wireless interface, the completed calibration processes can be transmitted to a PC. This data can subsequently be evaluated and archived using WIKA-Cal software. Thus a completely paperless transmitter calibration is possible with just the CPH7000.
Accuracy
The CPH7000 is compensated over a wide temperature range of 10 ... 50 °C (50 ... 122 °F) and achieves an accuracy of 0.025 % of span. In order to avoid intricate calculations, the measured values can also be displayed directly in
customer-specic units.
03/2017
Data sheets showing similar products and accessories: Test pumps, hydraulic; model CPP series; see data sheet CT 91.05 Test pump, pneumatic; model CPP30; see data sheet CT 91.06 Calibration software; WIKA-Cal; see data sheet CT 95.10
Page 1 of 12WIKA data sheet CT 15.51
Features
For pressures greater than 25 bar (362.6 psi), there are the model CPT7000 external pressure sensors. Thus pressure measurement and calibration is possible in further pressure measuring ranges and accuracies. An optional atmospheric module records and documents the
Software
The calibration software for the CPH7000 is WIKA-Cal. WIKA-Cal, alongside PC-supported calibration and the
logger function, also oers the management of the calibration
and instrument data in an SQL database. The data transfer is
achieved completely wirelessly via WIKA-Wireless. environmental parameters important for a calibration, such as atmospheric pressure, air humidity and ambient temperature.
Certied accuracy
For the model CPH7000 process calibrator, the accuracy
Complete service case
The process calibrator, developed specically for
maintenance and service operations, is delivered in a
is certied in a factory calibration certicate accompanying
the instrument. On request, we can provide a DKD/DAkkS
calibration certicate.
portable case system and, depending on requirements, can be equipped with, for example, model CPT7000 external pressure sensors, a Pt100 temperature sensor or a portable system with storage bag.
Specications
Digital process calibrator model CPH7000
Display
Display Touchscreen colour display
Display resolution up to 5 digits; selectable
Pressure units mbar, bar, psi, Pa, kPa, hPa, MPa, mmHg, cmHg, inHg, mmH
Settings
Applications (Apps) Measure, calibrate, logger, switch test
Measuring rate 50 values/s
Menu languages English, German, Spanish, French, Italian, Russian, Arabic, Chinese (settable)
Connections
External pressure sensor
External ambient module
External temperature probe
Manual pressure generation
Voltage supply
Power supply internal Lithium-Ion rechargeable battery (typical charging time: < 5 h)
Battery life typical 8 hours
Permissible ambient conditions
Operating temperature -20 ... +50 °C (-4 ... +122 °F)
Storage temperature -20 ... +60 °C (-4 ... +140 °F)
Relative humidity 0 ... 85 % r. h. (non-condensing)
Shock and vibration 15 g per EN 60068-2-6
Reference conditions per IEC 61298-1
Atmospheric pressure 860 hPa < P < 1,060 hPa (12.5 psi < P < 15.4 psi)
Ambient temperature 18 °C < T < 28 °C, typ. 23 °C
Relative humidity 35 % r. h. < T < 95 % r. h. , typ. 55 % r. h.
Position Hand-held lying face-up
1)
1)
1)
1)
inH2O (20 °C), inH2O (60 °F), inHg (0 °C), inHg (60 °F), kg/cm², kp/cm², lbf/ft², kN/m², atm,
Torr, micron, g/l (20 °C), kg/m³ (20 °C) as well as two user-dened units
max. 2, compatible with model CPT7000 reference pressure sensors
max. 1 ambient module
max. 1 temperature probe
-0.85 … +25 bar (-12.3 ... +360 psi)
3)
2)
2)
O, mH2O, inH2O (4 °C),
2
1) Available as an option
2) Temperature probe and ambient module use the same connection. Cannot be used at the same time.
3) Continuous operation (without backlighting, WIKA-Wireless deactivated and the electrical module gives no voltage/current).
Page 2 of 12 WIKA data sheet CT 15.51 ∙ 03/2017
Digital process calibrator model CPH7000
Communication
Interface WIKA-Wireless
4)
Case
Material PC + ABS blend
Ingress protection IP54 category 2
Dimensions see technical drawing
Weight approx. 1.9 kg (4.19 lbs.) without internal pump and reference sensor
approx. 2.5 kg (5.51 lbs.) incl. internal pump and reference sensor
4) Requires a PC with Bluetooth® 2.1 interface
Internal sensor technology
Pressure
Gauge pressure bar -1 ... +1 -1 ... 5 -1 ... 10 -1 ... 20 -1 ... 25
Absolute pressure bar abs. 0 ... 1 0 ... 1.6 0 ... 6 0 ... 10 0 ... 20 0 ... 25
Overpressure safety 3 times
Accuracy of the measuring chain
Resolution 5 digits
Pressure connection G 1/8 female thread (only with selected pneumatic unit)
Electrical safety:
Resistance to overvoltage Yes
Short-circuit resistant Yes
Reverse polarity protection Yes
Voltage-resistant Up to 60 V
Input impedance
Current measurement 20 Ω
Voltage measurement 1 MΩ
Current
Measuring range 0 … 30 mA
Supply 0 … 30 mA
Resolution 1 µA
Accuracy Measure: 0.01 % ±1 µA
Voltage
Measuring range DC 0 … 30 V
Supply DC 24 V
Resolution 1 mV
Accuracy 0.01 % or reading ±1 mV
Barometric reference
Measuring range 850 ... 1,100 mbar (12.3 ... 16 psi)
Accuracy ±1 mbar
1) Available as an option
5) The internal reference sensor is only available in combination with a pneumatic unit.
6) It is dened by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
7) Calibrated at 23 °C (74 °F) and in vertical mounting position.
8) In the event of interference caused by high-frequency electromagnetic elds in a frequency range of 200 ... 300 MHz, an increased deviation of up to 0.1 % is expected for the current
9) The barometric reference can be used to switch pressure types, absolute <=> gauge. With gauge
5)
psi -14.5 ... +14.5 -14.5 ... +70 -14.5 ... +145 -14.5 ... +300 -14.5 ... +360
psi abs. 0 ... 14.5 0 ... 23 0 ... 87 0 ... 145 0 ... 290 0 ... 360
6)
0.025 % of span
Supply: 0,01 % ±2 µA
1) 9)
measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range during a periodic
zero point adjustment.
output function.
pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry out a complete absolute pressure emulation.
7)
8)
WIKA data sheet CT 15.51 ∙ 03/2017 Page 3 of 12
Power supply unit
Input voltage AC 100 ... 240 V, 50 ... 60 Hz
Output voltage DC 12 V
Nominal Output current 2,500 mA
Permissible ambient conditions
Operating temperature 0 ... 40 °C (32 ... +104 °F); up to 90 % r. h. (non-condensing)
Storage temperature -40 ... +70 °C (-40 ... +158 °F)
Relative humidity 20 ... 80 % r. h. (non-condensing)
Temperature probe Pt100
1)
Measuring range -50 ... +250 °C (-58 ... +482 °F)
Accuracy 1/10 DIN, classe B ±0.1 °C
Probe length 200 mm (7.87 in)
Probe diameter 3 mm (0.12 in)
Cable length 1 m (3.28 ft)
Connection to CPH7000 max. 1 temperature probe
2)
User-dened RTD probe Input of the coecients of R0, A, B and C
Ambient module
1)
Measuring range
Relative humidity 0 ... 100 % r. h.
Temperature -40 ... +125 °C (-40 ... +257 °F)
Accuracy
Relative humidity ±1.8 % r. h.
Temperature ±0.2 °C (0.36 °F)
Connection to CPH7000 max. 1 ambient module
WIKA-Wireless
4)
2)
Frequency range 2,400 ... 2,500 MHz
HF output power max. 2 dBm (+ 2 dBi)
Number of channels 79
Channel spacing 1/2 MHz
Bandwidth 80 MHz
Output power 4 dBm / 10 mW
1) Available as an option
2) Temperature probe and ambient module use the same connection. Cannot be used at the same time.
4) Requires a PC with Bluetooth
®
2.1 interface
Page 4 of 12 WIKA data sheet CT 15.51 ∙ 03/2017
Reference pressure sensor model CPT7000
Pressure range
Gauge pressure bar -0.25 ... +0.25 -0.4 ... +0.4 -0.6 ... +0.6 -1 ... 0 -1 ... +0.6
-1 ... +1 -1 ... +1.5 -1 ... +2.5 -1 ... +3 -1 ... +5
-1 ... +9 -1 ... +10 -1 ... +15 -1 ... +24 -1 ... +25
-1 ... +39 -1 ... +40
0 ... 0.25
10)
0 ... 2.5 0 ... 4 0 ... 6 0 ... 10 0 ... 16
0 ... 25 0 ... 40 0 ... 60 0 ... 100 0 ... 160
0 ... 250 0 ... 400 0 ... 600 0 ... 700 0 ... 1.000
Gauge pressure psi -4 ... +5 -6 ... +6 -8 ... +8 -14.5 ... 0 -14.5 ... +8
-14.5 ... +14.5 -14.5 ... +23 -14.5 ... +36 -14.5 ... +44 -14.5 ... +70
-14.5 ... +130 -14.5 ... +145 -14.5 ... +220 -14.5 ... +350 -14.5 ... +360
-14.5 ... +565 -14.5 ... +580
0 ... 4 0 ... 6 0 ... 8 0 ... 14.5 0 ... 23
0 ... 36 0 ... 60 0 ... 90 0 ... 145 0 ... 230
0 ... 360 0 ... 580 0 ... 870 0 ... 1,450 0 ... 2,320
0 ... 3,630 0 ... 5,800 0 ... 8,700 0 ... 10,150 0 ... 14,500
Absolute pressure bar abs. 0 ... 0.25
10)
0 ... 2.5 0 ... 4 0 ... 6 0 ... 10 0 ... 16
0 ... 25 0 ... 40
psi abs. 0 ... 4 0 ... 6 0 ... 8 0 ... 14.5 0 ... 23
0 ... 36 0 ... 60 0 ... 90 0 ... 145 0 ... 230
0 ... 360 0 ... 580
Overpressure safety 3 times; < 25 bar
2 times; > 25 bar ≤ 600 bar
1.5 times; > 600 bar
Process connection
Selectable versions
G ⅜ B
G ¼ B
G ¼ female
G ½ B
G ½ male to G ¼ female
G ½ B ush with O-ring from NBR
G ½ B ush with O-ring from EPDM
G 1 B ush with O-ring from NBR
G 1 B ush with O-ring from EPDM
¼ NPT
½ NPT
½ NPT male to ¼ NPT female
½ NPT female
M16 x 1.5 female with sealing cone
M18 x 1.5 male to G ¼ female
M20 x 1.5
M20 x 1.5 female with sealing cone
9/16-18 UNF female F 250-C
R ½ per ISO7 (DIN 2999)
Sensor data
Accuracy
6)
0.025 % of span
7)
Resolution 5 digits
Compensated range 10 ... 60 °C (50 ... 140 °F)
6) It is dened by the total measurement uncertainty, which is expressed with the coverage factor (k = 2) and includes the following factors: the intrinsic performance of the instrument, the
measurement uncertainty of the reference instrument, long-term stability, inuence of ambient conditions, drift and temperature eects over the compensated range during a periodic
zero point adjustment.
7) Calibrated at 23 °C (74 °F) and in vertical mounting position
10) Expanded accuracy of 0.1 % FS
11) Expanded accuracy of 0.075 % FS
0 ... 0.4
0 ... 0.4
11)
11)
0 ... 0.6 0 ... 1 0 ... 1.6
0 ... 0.6 0 ... 1 0 ... 1.6
WIKA data sheet CT 15.51 ∙ 03/2017 Page 5 of 12
Reference pressure sensor model CPT7000
Material
®
Wetted parts Stainless steel (with measuring ranges ≤ 25 bar (≤ 360 psi) Elgiloy
Internal transmission uid Synthetic oil (only for measuring ranges up to 25 bar (360 psi))
Permissible ambient conditions
Medium temperature -20 ... +60 °C (-4 ... +140 °F)
Operating temperature -20 ... +60 °C (-4 ... +140 °F)
Storage temperature -20 ... +80 °C (-4 ... +176 °F)
Relative humidity 0 ... 95 % r. h. (non-condensing)
Temperature compensation 10 ... 60 °C (50 ... 140 °F)
Temperature coecient Zero point = 0.1 % / 10 K
Span = 0.1 % / 10 K
Case
Material Stainless steel
Connection to the CPH7000 Option: External operation via 1 m or 3 m (3.28 ft or 9.84 ft) connection cable (plug-and-play)
Ingress protection IP65 / IP67 when connected
Dimensions see technical drawing
Weight approx. 230 g (0.5 lbs.)
in addition)
Approvals
Logo Description Country
EU declaration of conformity for CPH7000
EMC directive EN 61326-1:2013 emission (group 1, class B) and interference immunity (portable measuring equipment)
R&TTE directive EN 300 328 harmonised frequency range 2,400 … 2,500 MHz is used; Bluetooth® Classic, max. transmission power 10 mW. The instrument may be used without limitations in the EU and also CH, NO and LI.
Protection of health and safety EN 62479
EU declaration of conformity for CPT7000
EMC directive EN 61326 emission (group 1, class B) and interference immunity (industrial application)
Pressure equipment directive
PS > 200 bar, module A, pressure accessory
European Union
European Union
Certicates
Certicate
Calibration Standard: 3.1 calibration certicate per EN 10204
Option: DKD/DAkkS calibration certicate
Recommended recalibration interval 1 year (dependent on conditions of use)
Approvals and certicates, see website
Page 6 of 12 WIKA data sheet CT 15.51 ∙ 03/2017
Dimensions in mm (in)
Digital process calibrator model CPH7000
Reference pressure sensor model CPT7000
WIKA data sheet CT 15.51 ∙ 03/2017 Page 7 of 12
Application icons (app)
The home screen is very clearly subdivided into application­oriented apps:
Measure:
Display of 3 dierent measurements
Logger:
Simultaneous recording of up to 3 signals
Info:
All instrument information available at a glance
Remote:
WIKA-Wireless radio transmission settings
Calibrate:
Setting of calibrations using calibration assistant
Switch test:
Testing of pressure switches (NC or NO)
Settings:
General instrument settings
Service:
All service data at a glance
Special operating modes
Operating mode: Measure
Features
3 dierent measurements in one view
30 pressure units + 2 programmable units
Resolution: Up to 4 decimal places
Graphical display via bargraph
Optionally settable functions: Min/Max/Tare/Filter/Alarm min/Alarm max/Mean value/Rate/Sensor temperature
Applications
Measurement of operating/process pressures
Comparative measurements with test items (power supply and display for the test item through the CPH7000)
Maximum and Minimum memory (e.g. for leak testing)
Alarm function for safety testing
For further information see the operating instruction.
Representation of possible measuring channels
Selection of the type of measurement or calibration
Page 8 of 12 WIKA data sheet CT 15.51 ∙ 03/2017
Operating mode: Calibrate
Menu screen of the application “Calibration”
Representation of calibration results as table
Features
Calibration assistant
Supply with pressure, current and voltage
Calibration protocol is automatically saved
Applications
On-site calibration of pressure sensor and pressure measuring instruments (without PC)
A calibration assistant guides you easily through the calibration (following DKD/DAkkS). With this, the data sets, including date and time, are recorded within the CPH7000.
Prior to calibration, the calibration routines can be set directly on the instrument or uploaded via WIKA-Cal software.
Up to 100 calibrations can be stored
Re-calibrations possible
PC software available
Communication with WIKA-Cal calibration software via WIKA­Wireless
Representation of calibration results as graph
Operating mode: Switch test
Switch test with external voltage supply
Menu screen of the application “Switch test”
Features
Pressure display on the closing and opening of the switch
Automatic calculation of the hysteresis
V
outVin
Switch test with voltage supply DC 24 V of the CPH7000
+
+
s
-
-
mA GND
V
outVin
mA GND
Applications
On-site functional check of pressure switches (without PC)
Determination of the switch point accuracy and repeatability
Determination of the switch point hysteresis
WIKA data sheet CT 15.51 ∙ 03/2017 Page 9 of 12
Operating mode: Logger
Menu screen of the application „Logger“
Representation of logger results as table
Features
Logging of max. 3 signals/measured value at the same time
Automatic or manual data acquisition
Direct display as graph or table
Logger protocols are automatically saved
Applications
Logging of current, voltage, pressure and temperature
The logger menu guides one, step by step, through the logger process. The data sets, incl. date and time, are automatically saved in the CPH7000.
Before logging the data, routines can be set directly on the instrument or uploaded via WIKA-Cal software.
Re-logging possible
PC software available
Communication with WIKA-Cal calibration software via WIKA-Wireless
Representation of logger results as graph
Operating mode: Simulation of transmitter signals
- +
Test item
4 ... 20 mA
V
outVin
Test
item
mA GND
+ -
Features
Manual or automatic current source function
+
-
Applications
The CPH7000 can be connected in place of a transmitter within a current loop and used as a current source. The transmitter output signals from 0 ... 24 mA can be simulated through manual input or automatically using the ramp and step functions.
V
outVin
Page 10 of 12 WIKA data sheet CT 15.51 ∙ 03/2017
mA GND
WIKA-Cal calibration software
Easy and fast creation of a high-quality calibration certicate
The WIKA-Cal calibration software is used for generating
calibration certicates or logger protocols for pressure
measuring instruments and is available as a demo version for a cost-free download. A template helps the user and guides him through the creation process of a document. In order to switch from the demo version to a full version of the respective template, a USB key with the template has to be purchased. The pre-installed demo version automatically changes to the selected full version when the USB key is inserted and is available as long as the USB key is connected to the computer.
Creation of calibration certicates for mechanical and
electronic pressure measuring instruments
A calibration assistant guides you through the calibration
Automatic generation of the calibration steps
Generation of 3.1 certicates per DIN EN 10204
Creation of logger protocols
User-friendly interface
Languages: German, English, Italian and more due with software updates
For further information see data sheet CT 95.10
Calibration certicates can be created with the Cal-Template and logger protocols can be created with the Log-Template.
Cal Demo
Generation of calibration certicates limited to 2 measuring points, with
automatic initiation of pressures via a pressure controller.
Cal Light
Generation of calibration certicates with no limitations on measuring
points, without automatic initiation of pressures via a pressure controller.
Calibration certificate
Kalibrierzeugnis
General
Allgemein
Calibration certificate
Kalibrierschein
Customer Demo customer
Kunde
Demo street DE-12345 Demo city
Customer order No.
Kundenbestellnummer
Order date
Bestelldatum
Date of calibration 01/02/2017
Datum der Kalibrierung
Operator Demo User
Bearbeiter
Environmental conditions
Umgebungsbedingungen
Temperature (20,0 ± 0,1) °C
Temperatur
Atmospheric pressure (1013,0 ± 0,1) hPa
Atmosphärischer Luftdruck
Rel. humidity (40 ± 1) %
Rel. Luftfeuchte
Measurement conditions
Messbedingungen
Working standard (WS) Pressure controller
Gebrauchsnormal (GN) Druckcontroller
Model Demo CPC
Typ
Calibration mark Demo 12345
Kalibrierzeichen
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy 0.02 % FS
Genauigkeit
Demo laboratory
Tel.: 01234/56789
Demo street 123
Fax: 01234/56789
12345 Demo city
E-Mail: demo@demo.com
Demo country
www.demo.com
Certificate No. 00000122
Zeugnis-Nr.
Device under test (DUT)
Kalibriergegenstand (KG)
Electrical gauge
Elektrisches Druckmessgerät
Model D-10
Typ
Serial number demo1234
Seriennummer
Scale range (0 ... 10) bar g
Anzeigebereich
Accuracy 0.05 % FS
Genauigkeit
Resolution (scale division) 0.0001 bar
Auflösung (Skalenwert)
Measuring point No. demo tag
Messstellen-Nr
Place of calibration
Ort der Kalibrierung
g = (9.80946 ± 0.01) m/s²
Demo place
Musterort
Calibration certificate
Kalibrierzeugnis
Measurement results
Messergebnisse
Reference value WS Reading DUT
Statement of compliance
Konformitätsaussage
Object keeps the specifications!
Der Kalibriergegenstand hält die Herstellerspezifikationen ein!
Demo laboratory
Page 1 / 2
Demo street 123
Seite
12345 Demo city Demo country
Referenzwert GN Ablesung KG
bar bar
0.00000 0.0000 0.0010
10.00000 10.0010 10.0020
Tel.: 01234/56789 Fax: 01234/56789 E-Mail: demo@demo.com www.demo.com
Certificate No. 00000122
Zeugnis-Nr.
Log Demo
Creation of data logger test reports, limited to 5 measured values.
Page 2 / 2
Seite
Log
Creation of data logger test reports without limiting the measured values.
WIKA data sheet CT 15.51 ∙ 03/2017 Page 11 of 12
Scope of delivery
Process calibrator model CPH7000
Power supply unit
Operating instructions
Service case with 2 connection cables (4 mm plugs)
3.1 calibration certicate per DIN EN 10204
Options
DKD/DAkkS calibration certicate
Accessories
Connection adapters
Various pressure adapters
Measuring equipment
External pressure sensors model CPT7000
Ambient module
Barometer
Pt100 measuring probe
Carrying system
Belt pocket for accessories
Process calibrator model CPH7000
Software
WIKA-Cal calibration software
Service case with process calibrator and accessories
Ordering information
CPH7000 / Version / Pressure generation / Unit / Pressure type / Measuring range / Accuracy / Type of certicate / Barometer /
Barometer calibration / Atmospheric module / Atmospheric module calibration / Temperature probe / Temperature probe calibration / Electrical module calibration / Communication / Software / Pressure connection set / Carrying system / Transport case / Further approvals / Additional ordering information
CPT7000 / Version / Unit / Pressure type / Measuring range / Process connection / Medium temperature / Wetted parts /
Special design for media / Accuracy / Type of certicate / Sensor cable / Further approvals / Additional ordering information
© 12/2016 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specications given in this document represent the state of engineering at the time of publishing. We reserve the right to make modications to the specications and materials.
Page 12 of 12 WIKA data sheet CT 15.51 ∙ 03/2017
WIKA Alexander Wiegand SE & Co. KG
Alexander-Wiegand-Straße 30 63911 Klingenberg/Germany Tel. +49 9372 132-0 Fax +49 9372 132-406 info@wika.de www.wika.de
03/2017 EN
Loading...