Mensor CPG2500 Operating Manual

Precision Pressure Indicator
CPG2500
Operating Instructions
Precision Pressure Indicator CPG2500 PN 0018908001B • 11/2014
Precision Pressure Indicator CPG2500
This Warning symbol indicates that danger of injury for persons and the environment and/or considerable damage
Warning
Caution
i
Notice
(mortal danger, danger of injury) will occur if the respective safety precautions are not taken.
This Caution symbol indicates danger for the system and material if the respective safety precautions are not taken.
This Notice symbol does not indicate safety notices but information for a better understanding of the facts.
2 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
Table of Contents
1. General Information 7
1.1 Warranty 7
1.2 Important Notice 7
1.3 FCC Radio Frequency Emission Notice 7
1.4 Trademarks and Copyrights 8
1.5 Software License Agreement 8
1.6 Mensor Service Plus 8
1.6.1 After the Warranty 8
1.6.2 Calibration Services 8
1.6.3 Certications and Accreditations 8
1.7 Packaging for Shipment 8
2. Safety Notices 9
2.1 User Responsibilities 9
2.2 General Safety Notices 9
2.3 Warnings and Caution Notices 10
3. General Description 11
3.1 Features 11
3.2 Turning On 12
3.3 Front Panel 12
3.4 Display 13
3.5 Chassis Assembly 13
4. Specications  15
4.1 Transducer Specication 15
4.2 Measurement & General Specications 16
5. Installation 17
5.1 Unpacking the Instrument 17
5.2 Dimensions (inches) 18
5.3 Mounting 19
5.4 Rear Panel 19
5.4.1 Pressure Connections 20
5.4.2 Pressure Port 20
5.4.3 Reference Port 20
5.5 Remote Communication Connections 20
5.6 Power Up 20
Operating Instructions - CPG2500 3
Precision Pressure Indicator CPG2500
6. Local Operation and Setup 21
6.1 General Operation 21
6.1.1 Setup Applications 21
6.1.2 Display Screen Features 21
6.2 Initial Setup 22
6.2.1 Contact and Version Information Application 22
6.2.2 Language Selection 22
6.3 Application Selection and parameter inputs 23
6.4 Applications 24
6.4.1 Measure Application 24
6.4.1.1 Pressure Mode / Emulation Mode 25
6.4.1.2 Auxiliary Displays 25
6.4.1.3 Zero Button 26
6.4.1.4 Tare Button 27
6.4.2 Settings Application 28
6.4.2.1 Languages 28
6.4.2.2 Brightness 29
6.4.2.3 Volume 29
6.4.2.4 User base units / Base units multiplier 30
6.4.2.5 Barometer Units 30
6.4.2.6 Conguration 31
6.4.3 Frames Application 32
6.4.3.1 Frame Format 32
6.4.3.2 Frames Channel 33
6.4.4 Transducer Application 34
6.4.4.1 Transducer Channel Selection 34
6.4.4.2 Transducer Delta Emulation 35
6.4.4.3 Transducer Filter 35
6.4.4.4 Transducer Resolution 36
6.4.4.5 Auxiliary Displays 37
6.4.4.6 Cal Function 38
6.4.5 Remote Application 39
6.4.5.1 Remote Command Set 39
6.4.5.2 Remote Communication Settings 40
6.4.6 Info Application 41
6.4.7 Leak Test Application 42
6.4.8 Troubleshooting Application 43
6.4.9 Service Application 44
6.4.10 Unlocked Service Application 45
4 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
7. Remote Operation 47
7.1 Command Set 47
7.2 IEEE-488 47
7.2.1 IEEE-488.2 Commands 47
7.3 Ethernet 48
7.4 Serial 49
7.4.1 Serial Cable Requirements 49
7.5 Mensor Command Set 50
7.6 Command and Query Format 50
7.7 Command Set Denitions 50
7.8 Output Formats 51
7.9 CPG2500 Commands and Queries 51
7.9.1 Units Command Syntax for Measurement Units 56
7.9.2 CPG2500 Error Codes 57
7.9.3 SCPI Commands and Queries 57
7.9.4 SCPI Commands Error Messages and Error Codes 59
7.9.5 GPIB Capability Codes 59
7.9.6 Interface Functions 59
8. Options 61
8.1 Barometer 61
8.2 Barometric Reference 61
8.2.1 Absolute Pressure Emulation 61
8.2.2 Emulation Mode Accuracy 62
8.2.3 Barometric Reference Calibration 62
8.2.4 Barometric reference Specications 62
8.3 Delta Channel 62
8.4 External Transducers (CPT6100 and CPT6180) 62
8.4.1 External Pressure Sensor Connection Cable 62
8.5 External Sensor Package (D-Sub) 63
8.6 Analog Output 63
8.6.1 Specications 63
8.7 Rack Mount Kit 63
8.8 Fittings 63
8.9 Remote Calibration 63
8.9.1 Remote Calibration Kit for Internal Transducers. 64
8.9.2 Barometric Reference Calibration Sled 64
8.9.3 External Calibration Procedures 65
8.9.3.1 Transducer Removal 65
8.9.3.2 Barometric Reference Removal 65
Operating Instructions - CPG2500 5
Precision Pressure Indicator CPG2500
9. Maintenance 67
9.1 Beyond the Warranty 67
9.2 Spare Parts 67
10. Calibration 69
10.1 Environment 69
10.2 Pressure Standards 69
10.3 Media 69
10.4 Setup 70
10.5 Service Application 71
10.5.1 Service Application (unlocked) 72
10.6 Calibration Data 73
10.7 One Point Cal Application 73
10.8 Two Point Cal Application 74
10.9 Linearization 75
10.10 Head Pressure 77
11. Appendix 79
11.1 Conversion Factors, PSI 80
11.2 Conversion Factors, Millitorr 81
11.3 Conversion Factors, Pascal 82
6 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500

1 General Information

1.1 Warranty
All products manufactured by Mensor are warranted to be free of defects in workmanship and materials for
a period of one year from the date of shipment. No other express warranty is given, and no armation of
Seller, by words or actions, shall constitute a warranty. SELLER DISCLAIMS ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSES WHATSOEVER. If any defect in workmanship or material should develop under conditions of normal use and service within the warranty period, repairs will be made at no charge to the original purchaser, upon delivery of the product(s) to the factory, shipping charges prepaid. If inspection by Mensor or its authorized representative reveals that the product was damaged by accident, alteration, misuse, abuse, faulty installation or other causes beyond the control of Mensor, this warranty does not apply. The judgment of Mensor will be nal as to all matters con­cerning condition of the product, the cause and nature of a defect, and the necessity or manner of repair. Service, repairs or disassembly of the product in any manner, performed without specic factory permis­sion, voids this warranty.
MENSOR MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Mensor shall not be liable for errors contained herein or for incidental or con­sequential damages in connection with the furnishing, performance, or use of this material.
1.2 Important Notice
The product specications and other information contained in this manual are subject to change without
notice.
Mensor has made a concerted eort to provide complete and current information for the proper use of
the equipment. If there are questions regarding this manual or the proper use of the equipment, contact Mensor at:
Mensor WIKA Alexander Wiegand SE & Co. KG 201 Barnes Drive Alexander-Wiegand-Straße 30 San Marcos, Tx 78666 D-63911 Klingenberg / Germany Tel: 1.512.396.4200 Tel: (+49) 93 72/132-9986
1.800.984.4200 (USA only) web site: www.wika.de web site: www.mensor.com fax: (+49) 93 72/132-8767 fax: 512.396.1820 e-mail: testequip@wika.de e-mail: sales@mensor.com tech.support@mensor.com
1.3 FCC Radio Frequency Emission Notice
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the in­struction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to cor­rect the interference at his or her own expense.
USE SHIELDED CABLES TO CONNECT EXTERNAL DEVICES TO THIS INSTRUMENT TO MINIMIZE RF RADIATION.
Operating Instructions - CPG2500 7
Precision Pressure Indicator CPG2500
1.4 Trademarks and Copyrights
Mensor is a registered trademark of Mensor, LP. All other brand and product names are trademarks or registered trademarks of their respective companies. ©2007, Mensor, LP. All rights reserved.
1.5 Software License Agreement
This product contains intellectual property, i.e., software programs, that are licensed for use by the end user/customer (hereinafter “end user”).
This is not a sale of such intellectual property.
The end user shall not copy, disassemble or reverse compile the software program.
The software programs are provided to the end user “as is” without warranty
i
Notice
Mensor and its suppliers shall not be held to any liability for any damages suered or incurred by the end
user (including, but not limited to, general, special, consequential or incidental damages including dam-
ages for loss of business prots, business interruption, loss of business information and the like), arising
from or in connection with the delivery, use or performance of the software program.
of any kind, either express or implied, including, but not limited to, warranties
ofmerchantabilityandtnessforaparticularpurpose.Theentireriskofthe
quality and performance of the software program is with the end user.
1.6 Mensor Service Plus
If you have problems and you don’t nd the answer in this manual, contact Mensor at 1.800.984.4200
(USA only) or 1.512.396.4200 for personal assistance, or at any of the contact addresses listed on the rear cover of this manual. We are ready to help.
1.6.1 After the Warranty
Mensor’s concern with the performance of this instrument is not limited to the warranty period. We pro-
vide complete repair, calibration and certication services after the warranty for a nominal fee.
1.6.2 Calibration Services
In addition to servicing our own products Mensor can perform a complete pressure calibration service, up to 20,000 psi, for all of your pressure instruments. This service includes an accredited calibration.
1.6.3 CerticationsandAccreditations
Mensor is registered to ISO 9001:2008. The calibration program at Mensor is accredited by A2LA, as complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1-1994 standards.
1.7 Packaging for Shipment
If the product must be shipped to a dierent location or returned to Mensor for any reason through a com­mon carrier it must be packaged properly to minimize the risk of damage.
The recommended method of packing is to place the instrument in a container, surrounded on all sides with at least four inches of shock attenuation material such as styrofoam peanuts.
8 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500

2 Safety Notices

2.1 User Responsibilities
To ensure safety, the user must make sure that:
• The system is used properly, no dangerous media are used and that all technical specications are
observed.
• The system is operated in perfect operating condition.
• This operation manual is legible and accessible to the user at the system’s location.
• The system is operated, serviced and repaired only by authorized and qualied personnel.
• The operator receives instruction on industrial safety and environmental protection, and is knowledge­able of the operating instructions and the safety notices contained therein.
2.2 General Safety Notices
The system should only be operated by trained personnel who are familiar with
i
Notice
Warning
this manual and the operation of the instrument.
WARNING: A condition for trouble-free and safe operation of this system is proper transport, proper storage, installation, assembly and proper use as well as careful operation and maintenance. Any operation not described in the following instructions should be prohibited. The system must be handled with care required for an electronic precision instru-
ment(protectfromhumidity,impacts,strongmagneticelds,staticelectricityand
extreme temperatures). Do not insert any objects into the instrument. The system is powered via the power cable with a voltage that can cause physi­cal injury. Even after disconnecting the system from the power supply, dangerous voltages can temporarily occur due to capacitance. Extreme care must be taken with pressure connections when using hazardous or toxic media. Repairs must only be performed by authorized service personnel.
i
Notice
Operating Instructions - CPG2500 9
Additional safety notices are found throughout this manual.
Precision Pressure Indicator CPG2500
2.3 Warnings and Caution Notices
WARNING: HIGH PRESSURE! High pressure gases are potentially hazardous. Energy stored in these gases and liquids can be released suddenly and with
Warning
Warning
Warning
Caution
extreme force. High pressure systems should be assembled and operated only by personnel who have been trained in proper safety practices.
WARNING: NOT EXPLOSION PROOF! Installation of this instrument in an area requiring devices rated as intrinsically safe is not recommended.
WARNING: POSSIBLE INJURY! The tubing, valves, and other apparatus attached to the gauge must be adequate for the maximum pressure which will be applied, otherwise physical injury to the operator or bystanders is possible.
CAUTION: USE THE PROPER PRESSURE MEDIUM! Use only clean, dry, non-corro-
sivegasesunlessotherwisespeciedbyMensor.Thisinstrumentisnotdesigned
for oxygen use.
CAUTION: As with most sensitive electronic equipment, switch the power switch
Caution
Additional Warning and Caution notices are found throughout this manual.
obeforeconnectingordisconnectingtoapowersourcetopreventdataloss.Do notpositiontheequipmentsothatitisdiculttodisconnecttheDCpowerjack.
CAUTION: ESD PROTECTION REQUIRED. The proper use of grounded work surfaces and personal wrist straps are required when coming into contact with exposed circuits (printed circuit boards) to prevent static discharge to sensitive electronic components.
10 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500

3 General Description

The CPG2500 Precision Pressure Indicator is a multi-channel pressure indicator designed to test and calibrate a variety of pressure devices in either absolute or gauge pressure modes. The CPG2500 can have two independent internal transducers, an optional barometric reference and an external remote pressure transducer.
Figure 3.1 - Desk top version
3.1 Features
Here is a short list of signicant features designed into the CPG2500:
Up to two removable / interchangeable, highly stable, temperature compensated, internal pressure
transducers. (Option to connect two external transducers via a rear mounted RS-232 port in place of the two internal transducers)
An optional remote transducer in addition to the two internal transducers.
An optional removable / interchangeable internal high accuracy barometric reference transducer
providing gauge pressure emulation for absolute ranges and absolute pressure emulation for gauge ranges.
Optional Analog output for each internal sensor.
7” Color LCD display with touch screen
Delta function when two or more transducers are installed
Multiple languages; change the language for on-screen text and number/date formats by simply
touching one of the “national ag” icons available in the setup screen.
Desk top or rack mount
Local Operation, or command and read remotely.
Operating Instructions - CPG2500 11
Precision Pressure Indicator CPG2500
3.2 Turning On
You can conrm that your CPG2500 is operational right now. Apply power to the power connector on
the rear of the instrument with the included power adapter, remove any plastic plugs from the rear panel pressure ports, and press the power switch to ON. The system will go through an initialization process, which takes about 45 seconds, and then a display will appear similar to the screen shown below.
Power Switch
ON/OFF
To see information about the conguration of your new CPG2500, touch the Information Application
(App) icon on the menu and a window will appear listing the Mensor customer service contact, model number and the sensors that are installed. Press the Measure App [
] to return to the main screen.
Information Application
3.3 Front Panel
The CPG2500 front panel includes a 7” color LCD display with touch screen. Operator input is accom-
plished by pressing the words or App icons presented on the display. There is a single discrete on/o
button and a USB on the right hand side. The front panel also shows the model number designation and brand logos.
12 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
3.4 Display
The display is made up of two sections. In the main screen (“Measure Application”), the left two thirds
shows the transducer channels that are congured for display, along with the pressure reading, units,
mode (absolute or gauge), channel label, an auto zero or tare button (if enabled) and any auxiliary indica­tions that have been chosen. The right one third of the screen has Application Icons (“Apps”) for setting viewing options, sensor settings, communication settings plus a “Page Down” button [ pressed, shows a second and third page of icons for calibration, remote communication error checking and a leak test application.
] that, when
Channel label
Page Down
and range
Auto Zero
Reading
Units and Mode
Auxiliary Indications
Transducer Channels Settings Apps
Buttons, Labels and Windows: The CPG2500 touch screen has many buttons with relevant graphic icons or text which, when pressed, will open a related window where changes can be made or informa­tion viewed. Some of these buttons will toggle from one state to another, others present choices or dis­play a numerical data entry screen. Text or icons that are displayed, but do not respond to being touched, are called labels or windows. Operators will quickly become accustomed to the particular characteristics of the frequently used buttons.
Main Screen: The main screen or “Measure Application”, appears after power-up. This screen contains
the channel frames and settings button. It will remain as congured after a power cycle.
Transducer Channel Frame(s): The transducer channel frames (left 2/3 of the screen) contain informa-
tion specic to a channel. Up to three channels can be displayed at once, two are shown in the picture
above. The channel frames are color coded with channel A - green, channel B - blue and the remote channel - red. The optional barometer channel is sky-blue and the delta channel is yellow. If only one channel is installed, a full frame will be displayed in the color of the channel connected. The channel frame contains the pressure reading, units, and mode(absolute or gauge) plus any auxiliary displays that have been chosen.
3.5 Chassis Assembly
The chassis assembly is the housing for the system and the removable transducers. The system has no
user-serviceable parts, and therefore the chassis should not be opened except by qualied repair per­sonnel at Mensor or certied service locations for any reason except to change the removable transduc-
ers.
Operating Instructions - CPG2500 13
Precision Pressure Indicator CPG2500
NOTES
14 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
4 Specications
Accuracy specications presented herein are obtained by comparison with primary standards traceable to a national metrology institute or recognized international standard organization. These specications
are obtained in accordance with the ISO Guide to the Expression of Uncertainty in Measurement (GUM). The calibration program at Mensor is accredited by the American Association of Laboratory Accreditation (A2LA) as complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1-1994 standards. If there is an exception to the requirements and recommendations of Z540 during a calibration the excep-
tion is noted on the individual calibration certicate.
Mensor reserves the right to change specications without notice.
4.1 TransducerSpecication
Standard Reference Transducers, Model CPR2550
Accuracy
Gauge Pressure 0 ... .36 to 0 ... <1 psi 0... 1 to 0 ... 10,000 psi 0 ...14.5 to 0 ... 6000 psi
Bi-Directional Pressure -0.18 to 0.18 to >-0.5 ... <0.5 -0.5 ... 0.5 to -15.5 ... 10,000 psi -14.5 ... 145 to -14.5 ... 6000 psi
Absolute Pressure 0 ... 7.5 to 0 ... 10,000 psia 0 ... 14.5 to 0 ...6015 psia
Calibration Interval 180 days 365 days
Media Compatibility
Metals in contact with media 6000/7000 series Aluminum, 316 SS, brass
Non-metals in contact with media PTFE (Teon®), Urethane, Silicone, RTV, Silicone grease, PVC, Epoxy, Buna-N, uoroelastomers
Sensor
Reading rate 33 readings/second
Calibration adjustments Internal zero adder and span multiplier, up to 11 point linearization for each sensor
(1)
0.03%Full Span 0.01%Full Span 0.01%IS-50
(3)
(Viton®)
365 days
(2)
Premium Reference Transducers, Model CPR2580
Accuracy
Gauge Pressure
Absolute Pressure
Calibration Interval
(1)
0.008% IS-33
0 ... 12 to 0 ...16.5 psig
0 ... 17.5 to 0 ... 33 psig 0 ... 80 to 0 ... 110 psig 0 ... 120 to 0 ... 220 psig
0 ... 12 to 0 ...16.5 psia
0 ... 18.4 to 0 ... 33 psia
0 ... 36 to 0 ... 50 psia
0 ... 80 to 0 ... 110 psia 0 ... 160 to 0 ... 220 psia 0 ... 240 to 0 ... 500 psia
365 days 365 days 365 days 365 days
Media Compatibility
Metals in contact with media 6000/7000 series aluminum, 316 SS, brass, inconel
Non-metals in contact with media PTFE (Teon®), Urethane, Silicone, RTV, Silicone grease, PVC, Epoxy, Buna-N, uoro-
Sensor
Reading rate 10 readings/second
Calibration adjustment Internal Zero adder and Span multiplier, up to 11 point linearization for each sensor
(1)
Accuracy is dened by total measurement uncertainty with the coverage factor (k=2) and includes the intrinsic performance of the instrument (linearity, hysteresis, repeatability), the measure-
ment uncertainty of the reference instrument, long term stability, inuences of ambient conditions, drift and temperature eects over the calibrated range, with periodic zero point adjustment.
(2)
The Intelliscale (IS) specication is a percent of reading specication combined with a percent of full scale specication. For example the designation “0.01% IS-50”, simply means that the
uncertainty of the sensor is 0.01% of reading above 50% of Full Scale, and 0.01% of 50% of Full Scale below 50% of the Full Scale. The 0.008% IS-33 specication is the same except the
uncertainty is 0.008% and the break over from percent of reading to percent of Full Scale is at 33% of Full Scale. Same rules apply for bidirectional ranges.
(3)
Gauge and Bidirectional ranges with full scale or full span range less than 14.5 psi have a calibration interval of 180 days, otherwise the calibration interval is 365 days.
(2)
0.008% IS-50
0 ... 700 to 0 ... 1100 psia 0 ... 1400 to 0 ... 3300 psia 0 ... 4200 to 0 ... 6015 psia
(2)
0.01% FS 0.014% FS
0 ... 8000 to 0 ... 11,000 psia 0 ... 12,000 to 0 ... 22,000 psia
0 ... 24,000 to 0 ... 31,500 psia 0 ... 32,000 to 0 ... 42,000 psia
elastomers (Viton®)
Operating Instructions - CPG2500 15
Precision Pressure Indicator CPG2500
4.2 Measurement&GeneralSpecications
Basic Instrument
Instrument
Instrument version Standard: Table top with tilt feet
Option:
-19” rack-mounting for single instrument mount.
-19” rack-mounting for dual instrument mount.
Dimensions See technical drawing
Weight 12.5 lbs./ 5.7 kg (with all internal options)
Warm-up time Approximately 15 minutes
Display
Screen 7” color LCD display
Resolution Selectable from 4 to 6 digits, depending on range and units
Data entry Touch screen keypad
Measurement Units psi, psf, osi, atm, inH20@4C, inH20@20C, inH20@60F, mbar, bar, Dy/cm2, pascal,
hPa, kPa, MPa, inHg@0C, inHg@60F, mTorr, Torr, mmHg@0C, cmHg@0C, mHg@0C, mmH20@4C, cmH20@4C, mH20@4C, mmH20@20C, cmH20@20C, mH20@20C, mSW, ftH20@4C, ftH20@20C, ftH20@60F, inSW, ftSW, tsi, tsf, g/cm
Range, + plus 2 user dened units (multiplier from psi, bar or pascal)
Rate Units /sec., /min., /hr., /3-hr
Languages English, German, Spanish, French, Italian, Portuguese, Polish, Russian, Chinese,
Japanese, Korean
Measurement lters O, Low, Normal (default), High
Connections
Number of integrated transducer (selectable)
Pressure connections Ranges ≤ 6015 psi: 7/16 - 20 female SAE/MS (adapters provided)
Pressure adaptors Standard: 1/8 in. FNPT
Overpressure limits 110 % FS typical, optional external relief valves are available
Voltage supply
Power input requirements 100-120 or 200-240 VAC, 50-60Hz, 24VA max
Switching power supply Output: 12 VDC, 1.7 A (includes 4 region specic plugs adapters)
Permissible ambient conditions
Storage temperature range 0 to 70 deg C
Operating environment 0 ... 95 % RH (relative humidity, non-condensing)
Operating temperature range 15 ... 45 deg C
Operating Altitude <3048 meters (10,000 ft)
Communications
Remote interface IEEE 488, RS-232, USB and Ethernet
Command sets Mensor, WIKA SCPI
Standard: 1 reference transducer Optional: 2nd reference transducers, external transducer, internal barometric reference
Ranges > 6015 psi: Autoclave F250C/HIP HF4
Optional: 1/4 inch tube, 6 mm tube, 1/4 in FNPT, 1/8 in. FBSP.
2
, kg/cm2, kg/m2, % of
CE conformity and certicates
CE compliance EN61326-1:2013 electromagnetic compliance
EN61010-1:2010 safety/CB scheme
Calibration Calibration certicate per ISO/IEC 17025:2005. Accreditation is by the American
Association for Laboratory Accreditation (A2LA).
16 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500

5 Installation

5.1 Unpacking the Instrument
In addition to functional testing, each unit is inspected for appearance prior to leaving the factory. Upon receipt, please examine the instrument for shipping damage. Report any apparent damage to the carrier immediately.
In addition to this manual you should have:
• CPG2500 Precision Pressure Indicator;
• Power Supply;
• Fitting adapters ordered;
• Any accessories ordered;
An envelope containing the calibration certicate.
Operating Instructions - CPG2500 17
Precision Pressure Indicator CPG2500
5.2 Dimensions (inches)
Desk top case
Front view
Single mount - Front view
Side view
19” rack mount
Dual mount - Front view
Rear Panel
Remote Transducer
connection
RS-232
USB (device)
USB (host)
Ethernet port
IEEE-488
Power
Measure port Channel B (7/16-20 UNF)
Baro ref port (baro ref option)
Measure port Channel A (7/16-20 UNF)
18 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
5.3 Mounting
The instrument can be set up on a desk top or it can be rack-mounted. Rack mount hardware is optional on the CPG2500 (see Section 5.2 Dimensions and Section 8, Options).
The special sensors used in the CPG2500 are relatively insensitive to tilt and vibration. However to fur­ther assure stability and accuracy, avoid mounting the instrument on surfaces subject to excessive motor or machinery vibration.
5.4 Rear Panel
Two slots are available on the rear of the CPG2500. These slots can be lled with a removable transducer
or a remote transducer interface connection. Gauge Transducers will have a reference and a pressure port. Absolute transducers will have a plug in the reference port. If a barometer is installed a hose barb tting will be installed in the upper right. Transducers with analog output will have an electrical connec­tion. Positioned on the left side is a remote transducer connection, the RS-232, Ethernet, IEEE-488, USB device connections for communication, the USB host connection and the 12 VDC power input.
Figure 5.4 A - Two removable gauge transducers installed
Figure 5.4 B - Two remote transducer interface connections installed
Operating Instructions - CPG2500 19
Precision Pressure Indicator CPG2500
5.4.1 Pressure Connections
Whenmakingupaconnectiontoano-ringadapterttingportuseaback-up
wrench to prevent over-stressing the threads in the manifold block.
i
Notice
For pressure range less than 6015 psi, the pressure ports on the rear are female 7/16 - 20 SAE/MS
straight threads per MS16142 and SAE J514 table 14. Connected adaptors require a tube tting boss seal with an o-ring per MS33656. Mensor can provide a variety of adapter ttings (see Section 8 Op­tions) with the instrument. Do not use sealant on ttings sealed with an o-ring. The integrity of each seal
is particularly important since even microscopic leaks can cause errors in pressure measurements. For a
pressure range greater than 6015 psi, a xed Autoclave F250C/HIP HF4 tting is attached
5.4.2 Pressure Port
The CPG2500 will precisely measure the pressure connected to the pressure port up to the full scale range of the transducer installed.
5.4.3 Reference Port
The reference port is available on gauge units that have sensors that are not sealed gauge units. For these units this port is available to connect to the reference side of the sensor. This port is normally left open to atmosphere but may be connected to a stable reference pressure. In an absolute pressure trans­ducer this port is plugged.
5.5 Remote Communication Connections
See Section 7, Remote Operation, for connections and commands for operation over IEEE-488, Ether­net, USB or RS-232 ports.
5.6 Power Up
Apply power to the power connector on the rear of the instrument using the power adaptor included, and switch the power switch on the front of the unit ON. The instrument will go through an initialization pro­cess and system check. As soon as the system check is completed the system will default to a screen similar to the one shown in Section 6.1.2 - Display Screen Features. The main measurement screen may
be congured in many dierent ways but initially it will be in a default conguration. Subsequently, the unit will power up in the conguration that it was in when last powered o. Allow at least 15 minutes of warm
up before performing critical pressure measurements.
20 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500

6 Local Operation and Setup

6.1 General Operation
This section describes the procedures for operating the CPG2500 from the front panel. Instructions for operating the device remotely from an external computer are covered in Section 7, Remote Operation. By following the procedures provided in these two sections and Section 10, Calibration, you can expect your CPG2500 to deliver maximum accuracy and dependability for many years of useful service.
6.1.1 Setup Applications
Conguration of the CPG2500 is achieved by changing settings accessed through the Application (“App”)
buttons. Local operation is accomplished by observing the data presented in the display. The appearance and functionality of the display can be changed by pressing the App button for the related function. After an app has been chosen, a set of related parameters will appear on the left. After choosing one of these parameters, a set of selections related to that parameter will appear on the right or a data entry keypad. The desired selection or data can be entered here.
6.1.2 Display Screen Features
The screen shown below provides a brief description of the features shown on a 2 channel display after initialization. The left two thirds of the display contains the area where information is displayed (in this case the Measure Application) and the right one third contains the selection icons for each application. Channels are color coded: Channel A is green, B is blue, C (remote) is red and the barometer is sky-blue. Color coding persists throughout all channel-specic screens. A zero or tare button and auxiliary displays (auxiliary units, rate and peak) will appear in the Measure App if activated. All of the CPG2500 screen features are described in more detail throughout this manual.
Range
Reading
Pressure Units
and Mode
App Display / Parameter Selection
Figure 6.1.2 - Display Screen Features
App Selection /
Data Entry
Operating Instructions - CPG2500 21
Precision Pressure Indicator CPG2500
6.2 Initial Setup
Section 6.3.1 and 6.3.2 are provided rst so that the operator can initially check the information screen to
verify the installed components and to change the language if needed.
6.2.1 Contact and Version Information Application
Press this application button to display Mensor contact, installed sensor, instrument and software version information.
6.2.2 Language Selection
Pressing the settings application button will open a screen where the language, display
brightness, volume, user base units/multiplier and conguration loading/saving, can be
changed. The current language selections available are shown in the table below. Additional language choices will appear on the screen after pushing the Page Down button [
Language Country
English USA
German Germany
French France
Spanish Spain
Italian Italy
Chinese China
English Great Britain
German Switzerland
French Switzerland
Spanish Mexico
Russian Russia
Korean Korea
English Canada
French Canada
Polish Poland
Japanese Japan
Figure 6.2.1 - Information
]:
22 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
6.3 Application Selection and parameter inputs
The application selection area on the right one third of the screen (see Figure 6.1.2 - Display Screen Fea­tures) is the area where setup, information and calibration Apps can be chosen. A second and third page of application selections can be accessed by pressing the page down button [ ]. A series of vertically placed circles on the center right indicates the active page by a larger circle. As each App is chosen, related application parameters will appear on the left two thirds of the screen along with the name of the application, and a reduced size icon in the top title section . When a parameter is chosen, related selec­tions, sliding scales or a data entry key pad will appear in the input area on the right where the application selection buttons were previously displayed. An example of each type of input is shown below. To return to the App selection menu, simply press the menu button [ use of each selection and menu is intuitively apparent and will become second nature with minimal expo­sure to the menu structure.
] above the input area. The purpose and
App Title
Page Down
Menu Button
Input Title
Operating Instructions - CPG2500 23
Precision Pressure Indicator CPG2500
6.4 Applications:
6.4.1 Measure Application
The Measure App is the normal operation screen referred to in previous instruments as the
‘main screen’. This application is dierent from the others in that it is not use to setup the con­guration but is used to monitor the pressure applied to the installed transducers.
The screen in gure 6.4.1-A shows the basic Measure App in a three channel instrument. The units but-
ton is always present in each channel. When the units button is pressed a selection of imperial and metric
units will be displayed on the right (gure 6.4.1-B); notice that the Units button has a lighter background
when the selection menu is active. Any unit of measure can be selected for any channel. If a barometric reference is installed, the Mode button, described below, will toggle from Gauge to Absolute mode when pressed.
Units Button
Mode Button
Figure 6.4.1-A Basic Measure App
Figure 6.4.1-B - Units Change
24 Operating Instructions - CPG2500
Precision Pressure Indicator CPG2500
6.4.1.1 Pressure Mode / Emulation Mode
The Mode button is only active if there is an optional Barometric reference installed. Otherwise, the mode
button becomes a label (gure 6.4.1.1) indicating the native mode of the transducer (absolute or gauge).
When an optional barometric reference is installed, a native gauge transducer can emulate absolute pressure using the barometric reference. Alternatively, a native absolute transducer can emulate gauge pressure. Emulation can be activated simply by pressing the Mode button.
Without a Barometer the Mode button becomes a label
Figure 6.4.1.1 - Mode Label
6.4.1.2 Auxiliary Displays
The screen in gure 6.4.1.2 shows all of the possible auxiliary display items that can be included in the
Measure App, in addition, each channel can contain one of two possible calibration functions. The aux­iliary display item includes “Alternate Units”, “Peak” and “Rate”. Calibration functions are either a “Zero” button or a “Tare” button. These auxiliary features will appear in the Measure App when selected from the Transducer App (section 6.4.4.4).
Figure 6.4.1.2 - Three channel Measure App with auxiliary displays
Operating Instructions - CPG2500 25
Precision Pressure Indicator CPG2500
Auxiliary buttons can be placed in three dierent parts of the screen depending on how they are set in the Transducer Application. Each auxiliary display can be modied by pressing the displayed button.
Peak: Pressing the Peak button will reset the upper and lower peak value to the current reading, subse­quent negative or positive divergence from that reading will be recorded in the button.
Rate: Pressing the Rate button will display a choice of time rate units for the rate denominator.
Units: Pressing the Auxiliary Units button will display the same set of units available for the primary units.
Pressing any of these units will change the auxiliary units to that chosen unit.
6.4.1.3 Zero Button
If the Zero Calibration function has been chosen in the Transducer App (section 6.4.4), then the Zero Cal Button [ Cal Button is pressed, a keyboard will appear to allow a single point calibration. If the channel is measur­ing gauge pressure, pressing the button will set the current reading to zero. If the channel is in emulation mode (absolute or gauge) then the value will not be saved to the sensor but only as a temporary adjust­ment while in emulation mode. After exiting the emulation mode or after a power cycle, the temporary adjustment will be cleared. The zero adjustment for a channel not in emulation mode will be saved to the sensor as if single point calibration had been performed.
] will appear in the Measure App. If the channel is measuring absolute pressure, and the Zero
Figure 6.4.1.4 shows two channels displayed, the zero cal function has been enabled for both channels. The screen on the left shows both channels with zero buttons. The screen on the right shows the same two channels, but the zero button on the absolute channel has been pressed, showing the keypad en­abled to accept a new single point calibration value.
Figure 6.4.1.3 - Zero Button, Gauge - Absolute
The background color of the zero button will momentarily change to a lighter color as the zero calibration is performed then will revert back to a darker color when complete.
26 Operating Instructions - CPG2500
Loading...
+ 58 hidden pages