![](/html/48/4844/4844e8c9250a61840d44ef291a00ee12a8ad69deec82326545fcf6059b7c6afe/bg1.png)
CPG2400 Barometer
Applications
Barometric pressure monitoring
Weather stations
Replacement of Mercury manometers
Calibration of medical radiation devices
Features
Testing and
Calibration Technology
Data Sheet CPG2400 Barometer 10/2014
0.03% of reading accuracy
Range: 8 to 17 psia (~560 to 1170 mbar)
Alternate range: 17 to 34 inHg A @ 0C
Peak, null functions
Description
The CPG2400 Barometer is low cost single channel Barometer
with a range of 8 to 17 psia. With an accuracy of 0.03%of reading
it is considered an excellent replacement for Mercury manom-
eters. Alternative ranges can be specied in a variety of units
including inches of Mercury (inHg A), millibar (mbar) and kilo
pascals (kPa). Built in peak and null functions provide a convenient way to analyze barometric trends.
Applications
The CPG2400 Barometer is used as a desk top or rack mounted
barometric reference for weather stations, laboratories, clean
rooms, nuclear power plants, pharmaceutical processing, health
care equipment manufacturing and Calibration of medical radiation equipment. The CPG2400 Barometer is used wherever
there is a need to have a precise measurement of barometric
pressure in manufacturing, testing, calibration or monitoring of
atmospheric pressure.
CPG2400 Barometer
Functional Flexibility
The CPG2400 Barometer has a clear LCD display with an intuitive menu structure for easy navigation within the setup
screens. A membrane key pad provides the user the capability
to navigate within these menus to set display features such as
peak, null, pressure units. It also allows access to a password
protected calibration section for recalibration of the internal
sensor.
Communications
The CPG2400 Barometer can be operated from a remote
computer via a RS-232 or RS-485 serial port connection.
The serial port can be used to congure, read or calibrate the
sensor. This is useful when remote indication of the barometric
pressure is required. This is useful when remote indication of
the pressure is required. RS-232 is used for communication
over short distances and the RS-485 for longer distances and
multidrop applications.
Data Sheet CPG2400 Barometer • 10/2014 Page 1 of 2
Related Products:
CPG2500 Barometer
CPG2300 Barometer
![](/html/48/4844/4844e8c9250a61840d44ef291a00ee12a8ad69deec82326545fcf6059b7c6afe/bg2.png)
Specications CPG2400 Barometer
Uncertainty 0.03% R
Pressure range psia 8 to 17 psia (other unit ranges are available within the equivalent psi range)
Resolution Digits 5
Overpressure limits % FS 150
Compensated temperature degrees C 15 to 45
Storage temperature degrees C -20 to 70
Warm-up time Min. <1
Reading rate /sec ~4.6 sensor readings
Response time mS <252
Case size in. 2.6 high x 4.2 wide x 4.9 deep
Weight lbs. <2
Orientation Negligible, can be removed with re-zeroing
Communications RS-232 or RS-485 (100 ft cable length recommended)
Media compatibility Pressure port: Clean, dry, non-corrosive, non-combustible, non-oxidizing
gases for all rated ranges.
Ranges ≥ 7.5 psi: All other media compatible with aluminum, 316 stainless
steel, brass, Buna N, Viton, sealant, silicone grease and RTV. Not designed for
oxygen service. Cannot guarantee accuracy on media other than gases.
Reference port: Clean, dry, non-corrosive gases.
Power input 12 VDC, .125 A
Pressure interfaces 7/16 - 20 female SAE/MS. 1/8” FNPT adapters provided
Pressure units psi, inHg @0°C, inHg @60°F, inH2O @4°C, inH2O @20°C, bar, mbar, mmHg
@0°C, cmHg @0°C, Pa, hPa, kPa, MPa, kg/cm2, cmH2O @4°C, cmH2O
@20°C, MSW @0°C
Display Monochrome 128 x 64 LCD with white LED backlight
Total Uncertainty is the combined uncertainties of all compo-
nents of a measurement at the approximate 95% condence
level (K=2). Total uncertainty includes the uncertainties of the following: Calibration standard, repeatability, pressure hysteresis,
creep, linearity, and temperature eects over the compensated
temperature range.
Since product innovation is a continuous process at Mensor, we
reserve the right to change specications without notice.
Page 2 of 2 Data Sheet CPG2400 Barometer • 10/2014
Represented by:
The calibration program at Mensor is accredited by A2LA as
complying with both the ISO/IEC 17025:2005 and the ANSI/
NCSL Z540-1-1994 standards. All Mensor primary standards
are traceable to NIST. Mensor is registered to ISO9001:2008.
Mensor
201 Barnes Drive
San Marcos, Texas 78666
phone: 512.396.4200
1.800.984.4200
fax: 512.396.1829
e-mail: sales@mensor.com
http://www.mensor.com