Applications
Portable Field Calibration Calibration Labs Semiconductor Manufacturers Utilities Pharmaceutical / Medical / Health Care
Features
0.015% of full scale (ranges ≥ 1 psig)
Dual sensor Optional loop powered milliamp measurement Dynamic Temperature Compensation from 0-50oC RS-232 Communications 17 selectable pressure units and 1 user-defined Calibration password protected Display of Max. and Min. readings Null capability Unique and simple user interface Absolute, Gauge, Bidirectional or Vacuum CE compliant
CPG2300 Portable Digital Pressure Gauge
Testing and Calibration Technology
Data Sheet CPG2300 • 10/2014
Description
Applications
The CPG2300 Portable Digital Pressure Gauge is used in diverse applications from rugged field calibration of pressure transmitters used in gas transmission, to testing and calibration of transducers in clean rooms and nuclear power plants. Wherever there is a need for a high level of accuracy in a handheld pressure calibration device.
Functional Flexibility
The CPG2300 Portable Digital Pressure Gauge has one of the best uncertainty specifications of any portable pressure calibration instrument. With an uncertainty of 0.015% FS, technicians are able to achieve the test uncertainty ratios required for modern transmitters. Temperature compensation from 0 to 50oC, permits use in most environments, without degradation of accuracy. Seventeen selectable pressure units and one user defined unit allows flexibility when calibrating.
Data Sheet CPG2300 • 10/2014 Page 1 of 2
Related Products: CPG2500 Digital Pressure Gauge CPG2400 Digital Pressure Gauge
The CPG2300 Portable Digital Pressure Gauge can have one or two independent precision pressure sensors which can be displayed simultaneously. Channel A can be ordered with full scale ranges up to 6000 psi. Channel B can be ordered with ranges from 3 to 1000 psig or 7.5 to 1000 psia.
The user interface is through a monochrome 128 x 64 LCD, with white LED back light. A tactile membrane keypad makes navigation within the intuitive menu easy. Options to display range A and B, or range A or B and the analog measurement channel (mA), plus the ability to power a control loop make the DPG 2300 ideal for field calibration of pressure and differential pressure transmitters. Built in functionality to display min/max peak pressure and rate per second per hour and per three hour plus a null function give the DPG 2300 additional functionality for trouble shooting process pressure transmitters or monitoring barometric pressure.
Specifications
Total Uncertainty
Calibration Stabil­ity (after warm up)
Calibration interval
Calibration adjustment
Channel A pressure ranges:
Channel B pressure ranges:
Pressure range: Bi-directional, Vacuum
Pressure units
Resolution
Overpressure limit
Compensated temperature
Operating temp.
Storage temp.
Warm-up
Battery life
Recharge time
Reading rate
Response time
Communications
Case size
Ranges 1 psig 0.015% FS
FS
Ranges < 1 psig 0.030% FS
Better than 0.015% FS (0.03% for FS ranges < 5 psig) for 180 days with periodic re-zeroing.
180 days
Zero and Span may be reset without affecting linearity.
0 ... 0.36 up to 0 ... 6000 psig 0 ... 7.5 up to 0 ... 6000 psia
0...3 up to 0...1000 psig
0...7.5 up to 0...1000 psia
psig: -0.18 to +0.18 min, -atm to 6,000 max
psi, in.Hg @ 0°C and 60°F, in.H2O @ 4°C and 20°C, ft.H2O @ 4°C, 20°C and 60°F, mTorr, inSW @ 0°C, ftSW @ 0°C, atm, bars, mbars, mmH2O @ 4°C, cmH2O @ 4°C, MH2O @ 4°C, mmHg @ 0°C, cmHg @ 0°C, Torr, hPa, mPa, kPa, Pa, D/cmsq, g/cmsq, kg/cmsq, mSW @ 0°C, PSI, PSF, TSF, TSI, mHg @ 0°C, %FS. All seawater units are 3.5% salinity.
5 digits
150% FS or greater, depending on range
0 ... 50
°C
°C 0...50
-20 ... 70
°C
min < 1 to rated accuracy
hr. 20
hr. < 8
1/
~4.6
sec
ms <250 for FS pressure step
RS-232, 9600 baud, N, 8, 1
in. T-Shaped, 8.6 H x 4.3 W x 1.6 D
Fittings
Power
Options
Display
CE
Total Uncertainty includes the uncertainties of the following: Calibration
standard, repeatability, pressure hysteresis, creep, linearity, and tempera­ture effects over the compensated temperature range.
1/4 inch FNPT pressure ports, nom. 1/16 inch hose barb for ref ports.
+9 to +14.5 VDC, 830 mA max for battery recharge.
Relief valves (mounted externally). 4-20 mA measurement with 24 VDC loop power source; current measurement ac­curacy 0.015%R.
Monochrome 128 x64 LCD with white LCD backlight.
Compliant to EN50081 and EN50082.
Order Number
Choose the number associated with the desired option.
Model Number: 23 _ _ , Range A:____, Range B:____
1 - Single Range 2 - Dual Range
0 - Without 4-20 mA Meas. 1 - With 4 - 20 mA Meas.
Accessories
Hand pump
Canvas case with belt clip
Since product innovation is a continuous process at Mensor, we reserve
the right to change specications without notice.
The calibration program at Mensor is accredited by A2LA as comply­ing with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1­1994 standards. All Mensor primary standards are traceable to NIST. Mensor is registered to ISO9001:2008.
Analog cables
Weight
Media compatibility
Page 2 of 2
lb. Approximately 1.5
Pressure port: Clean, dry, non-corrosive, non-combustible, non-oxidizing gases for all rated ranges. Ranges 5 psi: All other media compat­ible with aluminum, 316 stainless steel, brass, Buna N, Viton, sealant, silicone grease and RTV. Not designed for oxygen service. Cannot guarantee accuracy on media other than gases. Reference port: Clean, dry, non-corrosive gases.