Memmert VO29 Data Sheet

Page 1/4 | 12.07.2018
Wisconsin Oven Distributors, LL C | Tel. 262-594-3941| E-Mail info@memmertusa.com

Product specification

Vacuum Ovens
VO29
Digital pressure control ensures rapid and gentle vacuum drying, its speed controlled vacu um pu mp (accessory) sav es around 70% energy.
The direct contact between the load and the heatable and removable thermoshelves in the chamber of the Memmert vacuum oven VO ensures rapid and uniform temperature control of food, cosmetics, watches, books, PCBs or injection molds, without the loss of heat. On this page, you can find all the essential technical data on our vacuum drying oven. Our customer relations team will be pleased to help if you want further information.
Page 2/4 | 12.07.2018
Wisconsin Oven Distributors, LL C | Tel. 262-594-3941| E-Mail info@memmertusa.com
Memmert Vacuum Ovens VO29
Temperature
Temperature temperature measured through 4-wire Pt100 sensor individually for each thermoshelf Working temperature range min. 5°C above ambient up to +200°C
Setting accuracy
0.1 °C
temperature
Pressure (Vacuum)
Vacuum range 5 to 1100 mbar Pressure control Digital electronic pressure control for a speed-controlled vacuum pump. Tubing for vacuum and air
are made of material 1.4571 (ASTM 316 Ti). Programmable, digitally controlled inlet for air.
Permitted final vacuum 0.01 mbar Maximum leakage rate 0.01 bar/h Pump control speed control, optimized rinsing procedures for the pump membranes as well as signal output for
pump ON/OFF
Connection Vacuum connection with small flange DN16, and gas inlet with small flange DN16

Control technology

adjustable parameters temperature (Celsius or Fahrenheit), pressure (vacuum), timer Language setting German, English, Spanish, French, Polish, Czech, Hungarian, Italian ControlCOCKPIT TwinDISPLAY. Adaptive multifunctional digital PID-microproc essor controller with 2 high-definition
TFT-color displays.
Timer Digital backwards counter with target time setting, adjustable from 1 minute to 99 days Function SetpointWAIT the process time does not start until the set temperature is reached

Communication

Interface Ethernet LAN, USB Documentation program stored in case of power failure
Programming AtmoCONTROL software on a USB stick for programming, managing and transferring programs
via Ethernet interface or USB port

Safety

Temperature control mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating
approx. 10°C above nominal temperature
AutoSAFETY additionally integrated over- and undertemperature protection "ASF", automatic ally following the
setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature, compressor in case of undertemperature
Autodiagnostic system integral fault diagnostics for temperature and humidity control Alarm visual and acoustic