
OPERATING MANUAL
ICH 256
Cooled incubator with compressor cooling and humidity control
Optional CO
supply or interior lighting
2

Willi-Memmert-Straße 90–96
When contacting customer service, always quote the product serial number on the nameplate
We reserve the right to make changes

This manual describes the assembly, function, transport and operation of ICH 256 cooled
work on the unit. Familiarise yourself with the safety regulations. Only perform work that is
tion is missing, ask your superior or contact the manufacturer. Do not do anything without
The ICH 256 cooled incubator is available with different configurations: If specific equipment
features or functions are available only with one of the configurations, this is indicated at the
to the enclosed, separate manual.
), please refer to the separate service manual.
This instruction manual belongs to the cooled incubator and should always be stored where
whereabouts of this instruction manual. We recommend that it is always stored in a protected
a different location, this manual must also go with it.

.........................................................................................................................
........................................................................................
................................................................................................................................
.............................................................................................................................
....................................................................................................................
.......................................................................................................................
....................................................................................................................
...........................................................................................................
.........................................................................................................................
..............................................................................................................
............................................................................................................................
..............................................................................................................................
..........................................................................................................................
4. Putting into Operation 23
4.1 Check the door and adjust if necessary
............................................................................
..........................................................................................................
..................................................................................................
......................................................................................................
............................................................................................................
...............................................................................................................
............................................................................................................
....................................................................................................
......................................................................................................................

........................................................................................
.............................................................................................................................
..............................................................................................
..............................................................................................................................

Terms and signs used
and regulations to avoid accidents and damage. These terms and signs are explained below.
Prohibition sign (forbidding an action)
Warning signs (warning of a danger)
Regulation signs (stipulate an action)
Other icons

and subject to many hours of testing in the factory. They contain the latest technology and
After removing covers, live parts may be exposed. You may receive
an electric shock if you touch these parts. Disconnect the mains plug
Any work inside the unit may only be performed by qualified electri-
When loading the chamber with an unsuitable load, poisonous or
explosive vapours or gases may be produced. This could cause the
chamber to explode, and people could be badly injured or poisoned.
The chamber may only be loaded with materials/test objects which
Additional safety regulations for cooled incubators with interior UV
when you open the door of the cooled incubator with interior UV

Additional safety regulations for cooled incubators with CO
can have a suffocating effect in high
concentrations. In normal operation, the cooled incubator gives
to its surroundings. You should therefore
ensure that the room in which it is installed is properly ventilated.
Always close the stop valve or pressure reducer on the gas bottle if
there is no gas bottle connected or if the bottle connected is empty.
can cause cold burns or frostbite.
gas to the eyes and skin.
water from penetrating as well as backflow into the gas bottles. It is
essential that you read the safety notes and regulations of the gas
is not a dangerous substance in terms of the German Hazardous Substances Ordinance
Warning signs for dangerous UV radiation in the working
area are attached to the doors or cooled incubators
with interior UV lighting (
These stickers must not be removed and must always
from Memmert customer service.
The cooled incubator may only be operated and maintained by persons who are of legal age,
and who have received relevant instructions. Personnel who are to be trained, instructed or
who are undergoing general training may only work with the appliance under the continuous
The cooled incubator may only be transported (with a forklift truck or manual pallet jack)

The owner of the cooled incubator
are qualified to do this, have received the respective instructions and are familiar with this
and repair work is properly carried out;
and its surroundings are kept clean and tidy;
safety goggles, safety shoes, protective gloves.
The manufacturer is not liable for any damage, danger or injuries that result from
The cooled incubator may only be used in a flawless condition. If you as the operator notice
and inform your superiors.
You can find information on eliminating malfunctions from page

In case of contact with CO
can cause accelerated breathing and headaches.
Anyone affected should breathe fresh air, using a breathing device independent of
Switching off the cooled incubator in an
Switch off the cooled incubator
by pressing the main switch

Design and Function
1
Design of ICH cooled incubators
Controller/control panel (see page
Push/turn control (see page
Controller/control panel (see page
Controller/control panel (see page
Push/turn control (see page
Push/turn control (see page
Standard feed-through
Standard feed-through
Cooling compressor (see page
Humidity sensor
Humidity sensor
Cooling compressor (see page
Cooling compressor (see page
Locking swivel castors
Locking swivel castors
Nameplate (see page
Nameplate (see page
Sliding grid
Sliding grid
for models with interior lighting, see page
illumination box (only
illumination box (only
Illumination box (only for models with
interior lighting, see page
Illumination box (only for models with
Illumination box (only for models with
interior lighting, see page
interior lighting, see page
Air is heated inside the cooled incubator by means of large-area all-round heating. Humidifica-
fan and mixed with the air current. Humidity is reduced by condensing on two Peltier cooling
automatically defrosted in cycles.
2
13
12
11
10
9
3
4
8
7
5
6

through a sterile filter. The turbulence-free interior ventilation ensures a uniform gas distri-
content is reduced by introducing
at the top of the interior (see
) that contains fluorescent tubes (daylight
Intermediate test zone II 30
Long term test zone II 25
Long term test zone IVA 30
Intermediate test zone II 30
Long term test zone II 25
Long term test zone IVA 30
Long term test zone IVB 30
Alternate accelerated test 40
Alternate intermediate test 30
Intermediate test zone II 30
Long term test zone II 25
Long term test zone IVA 30
Long term test zone IVB 30
Alternate accelerated test 40
Alternate intermediate test 30
relative humidity [%]
relative humidity [%]
50
60
70
10
20
30
40
50
10
20
30
0
temperature[C°]
40
60
50
60
80
70
80
70
80
100
100
100
100
90
90
90
90
Working range relative humidity without condensation ICH256
Working range relative humidity without condensation ICH256
Working range relative humidity without condensation ICH256
Working range relative humidity without condensation ICH256
Testing conditions according to ICH Q1A
Testing conditions according to ICH Q1A
Testing conditions according to ICH Q1A
Testing conditions according to ICH Q1A
Long term test zone II 30°C/35%rh
Alternate long term test 25°C/40%rh
Alternate accelerated test 40°C/25%rh
Alternate long term test 25°C/40%rh
Alternate accelerated test 40°C/25%rh
Long term test zone II 30°C/35%rh
Long term test zone II 30°C/35%rh
Accelerated test 40°C/75%rh
Accelerated test 40°C/75%rh
Accelerated test 40°C/75%rh
valid at T
amb
~20°C

All-round large-area heating with additional thermal conduction layer
Alphanumeric text display
Two separate Pt100 temperature sensors DIN class A in a 4-wire circuit for control and
at the appliance, without the
The ICH 256 cooled incubator can be equipped with the following fittings:
The chamber load for the appliance must be carefully checked for chemical compatibility with
A material resistance table can be requested from the manufacturer.

Appliance fuse: Fusible link 250 V/15 A quick-blow
The temperature controller is protected with a miniature fuse 100 mA (200 mA at 115 V)
controller is equipped with a 6.3 A miniature fuse.
1
Connections on rear of appliance
Connections on rear of appliance
Connections on rear of appliance
connection (only for models with CO
connection (only for models with CO
connection (only for models with CO
ethernet connection, see page
Printer connection (see page
Fresh air connection (only for models with CO
with residual current circuit breaker).
This appliance is intended for operation on an electrical power system with a system imped-
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
3
4
CO
2
2
5
6

The appliance can be connected with the supplied
tor with gas bottle monitor (DIN 8546) to a CO
compressed gas bottle or directly to a central
The supplied tube can be used to connect the cooled incubator with the supplied water tank.
tests or similar. Any other use is improper, and may result in hazards and damage.
The incubators may not be used for drying, vaporising and branding paints or similar materi-
als the solvents of which could form an explosive mixture when combined with air. If there
gas bottles with a pressure regulator may be connected to the gas connection of the
appliance. The use of other gases or materials than CO

EC Declaration of Conformity
Manufacturer’s name and address:
This product complies with the provisions of the
Machinery Directive 2006/42/EC
Directive of the Council for harmonisation of
The tested product’s level of compliance with the essential protection requirements of the Directive is substantiated
by its compliance with the following standards:
EMC Directive 2004/108/EC
ouncil for harmonisation of the laws of
level of compliance with the essential protection requirements of the Directive is substantiated
by its compliance with the following standards:
Low Voltage Directive 2006/95/EC
ouncil for harmonisation of the laws of
equipment designed for use within certain voltage limits.
compliance with the essential protection requirements of the Directive is substantiated
by its compliance with the following standards:
(legally binding signature of the manufacturer)
documents have to be observed.
Äußere Rittersbacher Straße 38
the laws of Member States on the level of protection
EMC Directive 2004/108/EC
tates relating to electrical
irectives, however, does not provide any warranted properties. The
Product:
Type:
Size:
Rated voltage:
EN ICO 12100-1, -2: 2004
EN ISO 13850: 2007
Directive of the C
The tested product’s
DIN EN 61326-1: 2006-10
DIN EN 61000-3-11: 2001-04
Applied harmonised standards:
Directive of the C
The tested product’s level of
DIN EN 61 010DIN EN 61 010-2-010 (VDE 0411 part 2
Schwabach, Germany, 09/02/2011
This declaration states the
safety instructions in the supplied
91126 Schwabach
Cooled incubator
ICH
256
AC 230 V,
directives:
and revisions
for accident prevention.
EN ISO 13857: 2008
EN 60204-1: 2007
and revisions
compatibility.
Member S
EN 61000-3-
and revisions
-08 EN 61 010-1:2001
010):2004-06 EN 61 010-2-
with the above D
EN 61326-1: 2006
Member S
ely AC 115 V 50/60 Hz

) provides information about the appliance model, manufacturer and
technical data. It is attached to the front of the appliance, on the right beneath the door (see
1
Operating voltage
Operating voltage
Protection type
Protection type
Connection/performance values
Connection/performance values
Connection/performance values
Technical data
2
3
4
10
9
8
7
5
6
Technical data without interior
lighting
with interior
lighting

Technical data without interior
Temperature recording is done by means of Pt100 in a 4-wire
Adjustment range normal operation: -10 °C to 60 °C, for mod-
supply or interior lighting 0 °C to 60 °C
Adjustment precision: 0.1 °C
Adjustment range: 10 to 80 % rh
Adjustment precision: 1 % rh
through an NDIR measuring pro-
Adjustment range: 0 to 20 %
Adjustment precision: 0.1 %
lighting
with interior
lighting

D
Dimensions of ICH cooled incubators
Ambient conditions
The cooled incubator may only be used in enclosed rooms and under the following ambient
Ambient temperature: 5 ºC to 28 ºC
Altitude of installation: max. 2000 m above sea level
The cooled incubator may not be used in areas where there is a risk of explosions. The
ambient air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The
age to electrical parts. For this reason, sufficient measures are to be taken to prevent large
Accessories
Water supply tank and connection tube
Additional for cooled incubators with CO
A
F
B
E
C
►

You may injure your hands or feet when transporting and
for this (e.g. forklift truck licence). The cooled incubator may not be
transported with a crane.
The cooled incubator could fall over and seriously injure you. Never
tilt the cooled incubator and transport it only in an upright position.
The cooled incubator may only be set up on the floor, and never on
tables or similar. A table could collapse or the appliance could fall
off and seriously injure or kill
Transport
The cooled incubator can be transported in three ways:
a forklift truck; move the forks of the truck entirely under the test chamber

The cooled incubator is delivered in cardboard packaging on a pallet.
from the delivery note, damage or irregularities, do not put the cooled incubator into
protective foil) in accordance with the legally
applicable disposal regulations for cardboard packaging in your country.

The cooled incubator may only be placed on the ground, not on a table. When doing this,
The installation site must be level
and must be able to reliably carry the weight of the cooled incubator (see page
The front swivel castors can be locked with a catch. To ensure stability, always turn the front
The distance between the wall and the rear of the cooled incubator must be at least 15 cm.
The clearance from the ceiling must not be less than 20 cm and the side clearance from the
wall must not be less than 8 cm (
). The ventilation slits of the
the appliance may not be adjusted.
Minimum clearance from walls and ceiling
≥ 20 cm
.
≥ 8 cm
≥ 8 cm
≥ 15 cm

The cooled incubator is intended for operation on an electrical power system with a
at the point of transfer (service line) of a maximum of 0.292 ohms. The op-
that meets these requirements.
) whose interfaces comply with the requirements for safety extra-low voltage (e.g.
accordance with VDE 0510/DIN EN
adhered to (production conductivity ≤
with VDE 0510 is available in larger
trade. The standard VDE 0510/DIN EN
H2O

can have a suffocating effect in high
concentrations. In normal operation, the cooled incubator gives
to its surroundings. You should therefore
ensure that the room in which it is installed is properly ventilated.
can cause cold burns or frostbite. Avoid
gas to the eyes and skin.
gas bottles away from open flames. Store gas bottles at lower than
as well as backflow into the gas bottles. It is essential that you read
the safety notes and regulations of the gas suppliers.
Attach the supplied pressure tube to the gas
CO
2

when opening the door of the cooled incubator with interior UV
The cooled incubator may only be operated and by persons who are of legal age and have re-
To open the doors, pull out the knobs (
To close, press in the door knobs.
Switching on appliance
The cooled incubator is switched on and off by pressing the
Switch on cooled incubator
Switch off cooled incubator
close
open

tubes prior to every test.
equipped with UV lighting: Wear
fluorescent tubes are working. If not: Replace the respective tube set (refer to
and reading off the measured illuminance. Replace the fluorescent tubes if the light
Loading the cooled incubator
When loading the appliance with an unsuitable load, poisonous or
explosive vapours or gases may be produced. This could cause the
cooled incubator to explode, and people could be seriously injured
or poisoned. The chamber may only be loaded with materials which
which cannot ignite. If there is any doubt as to the composition of
the chamber load for chemical compatibility with the materials of the cooled
), since considerable damage could otherwise occur to the
explosion-proof (it does not comply with the German workplace
when combined with air. Potentially explosive gas-air mixtures must not form, neither in the
ance could lead to sedimentation in the chamber interior and as a consequence, could result
taken to prevent large clouds of dust or aggressive vapours from developing.

The chamber must not be loaded too tightly, so that proper
air circulation in the working
walls or right below the ceiling (heating ribs) of the working chamber. To guarantee optimum
air circulation, push in the sliding grids so that the gaps between the door, sliding grid and
Correct and incorrect chamber loading
Connecting gas supply

The desired parameters are entered on the operating panel of the controller on the front of
). Basic settings, as well as those for time and pressure, can also be made
1
Operating mode display (see page
Display appliance is heating up
Operating mode display (see page
Operating mode display (see page
Display: appliance is cooling down
Alarm display (see chapter
Display: appliance is cooling down
Display: appliance is cooling down
Humidity display
Humidity display
setpoint (only for models
Setting the CO
Setting the CO
setpoint (only for models
setpoint (only for models
Display: appliance is humidifying
Display: water tank empty
Display: appliance is humidifying
Display: appliance is humidifying
Display: water tank empty
Display: water tank empty
Display: fan speed
Display: fan speed
rior lighting) Display: interior lighting
All operating functions are selected by turning the push-turn control to the
Tu
Mo
on
off
Sa Su
Fr
Th
We
h
2 3 4 5 6 7 8 9
t3
t2
t1
set
t4
loop
SETUP
PRINT
on
off
STERI DEFRO
4
3
2
1
push
°C
MIN
AUTO
card
IN 1
OUT
IN 2
IN 1
OUT
°C
MAX
IN 2
%
mb
IN 1
OUT
IN 2
OUT
2
2
rh
CO
mb
10111213141516
set
set

temperature), then all other parameters go dark and the selected one flashes.
With the SET key held down, set the desired value (e. g.
value, flashing. Then, the current temperature is displayed and the cooled
The control returns automatically to the main menu if neither the push-turn control
Quick adjustment of temperature:
Quick adjustment of temperature:
The appliance flashes briefly, displaying the temperature setpoint. Then, the current tempera-
ture appears on the display and the controller begins to adjust the temperature to the set
Week time switch: The cooled incubator runs at the set values only at certain times. Opera-
tion in this mode is described from page
values are programmed (so-called ramps), which the cooled incubator automatically works
through one after another. Operation in this mode is described from page
) and printouts (PRINT, see page
Operating modes
set
set
set
PRINT
SETUP

Setting the operating mode
values for operating the appliance can be selected. The settings have an immediate effect on
the functions of the appliance.
that it comes out of the appliance (see
with the push-turn control as described
As described above, set the individual parameters with the push-turn control and the SET
supply and without interior
Adjustment range: 10 % to 100 % in steps
set
SETUP
PRINT
°C
°C
MIN
MAX
AUTO
2

The appliance should heat up to 37 °C with a humidity of 80 % rh and a fan speed of 40 %.
The monitoring function should trigger at 38.5 °C and at 36.0 °C: For a model with CO
content should be 5.0 %. For a model with interior lighting, daylight should be
50 °C
37 °C
80 %
50 %
40 %
80 %
rh
50 %
rh
15 %
10 %
5 %
DL
t

°C with the push-turn control.
temperature appears on the display and the control-
Turn the push-turn control to the right, up to the moni-
toring temperature and until the MIN or MAX symbol
flashes. Hold down the SET key and, with the push-turn
the AUTO symbol flash. Hold down the SET key and, with
the push-turn control, select
The tolerance band is set in the SETUP menu (see
Turn the push-turn control to the right until the humid-
The humidification process is indicated by the
PRINT
SETUP
°C
°C
MIN
MAX
AUTO
%rh

Turn the push-turn control to the right until the CO2 dis-
the CO2 setpoint. The current CO2 actual value appears
Turn the push-turn control to the right until
with the push-turn control. Release
the SET key. The daylight lighting is activated.
The cooled incubator is now running in permanent operation with the set values.
and cooling functions, along with the CO
and humidity supply are switched off here and the
The sequence of the week time switch repeats itself each week.
Adjustment range: Monday to Sunday
Mo
Adjustment range: Working days Mo-Fr
Appliance not switched on on this day
Adjustment range: one minute after the switch-
Tu
We
Th
2
SETUP
PRINT
Sa
Su
Fr
Mo
Mo
on
off
on
off
on
off
Tu
Tu
We
We
Th
Th
Sa
Sa
Su
Su
Fr
Fr

the desired time blocks and days.
ture setpoint can be directly accessed by briefly pressing the SET key. By turning the control
to the right, you get to the temperature monitoring, humidity and CO
the left, you can return to the settings for the individual time blocks.
The appliance should switch on from Mo-Fr (workdays group) at 9.30 (am) and switch off at
Fr
Operation with week time switch (example)
week time switch operating mode
turn control, while the SET key is held down.
time switch operating mode.
Sa
PRINT
Tu
We
Mo
Th
Su
SETUP

Turning the push-turn control to the left, select "
time with the push-turn switch to
Mo
time with the push-turn switch to
With the push-turn control, select "
time with the push-turn switch to
With the push-turn control, select "
time with the push-turn switch to
automatically one after another.
Setting the programme operating mode
Setting the programme operating mode
function with the push-turn.
function with the push-turn.
on
off
Tu
We
Th
Fr
Sa Su
Mo
Mo
Mo
on
off
Tu
Tu
on
off
Tu
on
off
We
We
We
Th
Th
Th
Fr
Fr
Fr
PRINT
Sa Su
Sa
h
Sa
h
Su
Su
SETUP
PRINT
SETUP

You can now select and modify the following parameters in turn (see also the adjustment
STERI
Adjustment range: Monday to Sunday, workdays Mo-Fr, weekends Sa-Sun, every day Mon-Sun
hours. In the example shown: Duration of first ramp seg-
Tu
Mo
on
off
Sa Su
Fr
Th
We
t1
h
t3
t2
PRINT
4
t4
3
loop
2
1
SETUP
DEFRO
°C
MAX
MIN
AUTO
IN 1
OUT
IN 2
°C
%rh
mb
STERI
DEFRO
STERI
DEFRO
°C
MAX
MIN
AUTO
DEFRO
MAX
MIN
AUTO
°C
MAX
MIN
AUTO
Tu
Mo
on
off
Tu
Mo
on
off
Tu
Mo
on
off
Sa Su
Fr
Th
We
t1
h
Sa Su
Fr
Th
We
t1
h
Sa Su
Fr
Th
We
t1
h
t3
t2
PRINT
t3
t2
PRINT
t3
t2
PRINT
4
t4
3
loop
2
1
SETUP
4
t4
3
loop
2
1
SETUP
STERI
4
t4
3
loop
2
1
SETUP
IN 1
OUT
IN 2
°C
IN 1
°C
%rh
mb
OUT
IN 2
mb
°C

STERI
Adjustment range: 10 % to 100 % in 10-% steps In the example shown: fan speed 60 % (six
content of first ramp segment
content of first ramp segment
Tu
Mo
on
off
Tu
Mo
on
off
Tu
Mo
on
off
Sa Su
Fr
Th
We
t1
h
Sa Su
Fr
Th
We
t1
h
Sa Su
Fr
Th
We
t1
h
t3
t2
t2
t2
t4
loop
SETUP
PRINT
t3
t4
loop
SETUP
PRINT
t3
t4
loop
SETUP
PRINT
4
3
2
1
STERI
4
3
2
1
STERI
4
3
2
1
DEFRO
DEFRO
DEFRO
°C
MIN
°C
MIN
°C
MIN
AUTO
AUTO
AUTO
°C
MAX
%rh
°C
mb
MAX
%rh
°C
mb
MAX
CO
2
STERI
Tu
Mo
on
off
Sa Su
Fr
Th
We
t1
h
t3
PRINT
t4
loop
SETUP
t2
DEFRO
4
3
2
1
°C
MIN
AUTO
%rh
°C
mb
MAX
CO
2

STERI
These commands thus control the programme sequence:
Turn the push-turn control to the right until
appears the display and then press the SET
After releasing the SET key ...
as described above, or an existing
Tu
Mo
on
off
Tu
Mo
on
off
Tu
Mo
on
off
Sa Su
Fr
Th
We
t1
h
Sa Su
Fr
Th
We
t1
h
Sa Su
Fr
Th
We
t1
h
t3
t2
t2
t2
t4
loop
SETUP
PRINT
t3
t4
loop
SETUP
PRINT
t3
t4
loop
SETUP
PRINT
4
3
2
1
STERI
4
3
2
1
STERI
4
3
2
1
DEFRO
DEFRO
DEFRO
AUTO
AUTO
AUTO
%rh
°C
mb
MAX
%rh
°C
mb
MAX
%rh
°C
mb
MAX
°C
MIN
°C
MIN
°C
MIN
CO
CO
CO
2
2
2

These commands thus control the programme sequence:
Wait until the setpoint temperature has been reached.
The appliance starts the next programme segment only when the
Wait until setpoint humidity has been reached.
Appliance starts the next programme segment only when the
time has already elapsed.
Wait until setpoint temperature and setpoint humidity have been
when the programmed setpoint temperature and programmed
The programme entered is repeated after it has run through all
°C
Schematic example of the use of ramp segment close statements
Delayed
programme start
Close
command
ramp
segment
No. 1
spwt (t)
Segment1
command
segment
Segment2
Close
ramp
No. 2
next
Close
command
ramp
segment
No. 3
spwt (tH)
Segment3
Segment4
command
ramp
segment
No. 4
next
command
Segment5
Close
ramp
segment
No. 5
end

a fan speed of 30 %, and reach a relative humidity of 70 % rh. For models with CO
content should be set to 5 % and for models with interior lighting, daylight should be
the setpoint values at a fan speed of 50 % for 45 minutes. For models with lighting, the test
Afterwards, the appliance should cool down to 20 °C and reach 50 % rh humidity within 1
content should be adapted to
the natural content of the ambient air and for models with interior lighting, daylight should
50 °C
Settings example programme mode
Ramp 1 Ramp 2 Ramp 3
no
combination
possible
DL
UV
37 °C
20 °C
80 %
50 %
30 %
70 % rh
50 % rh
15 %
10 %
5 %
overtemperature limit
0.01 h
0.45 h
1.00 h
t
Mo 8.00 h

This ramp programme can only be set for cooled incubators equipped with CO
The respective descriptions in this example are therefore not relevant for these appli-
ter, using the "Celsius" software.
while the SET key is held down.
After releasing the SET key, the control is in the pro-
After releasing the SET key, the controller is in the pro-
Weekday for delayed programme start:
by turning the push-turn control
while the SET key is held down.
Mo
Turn the push-turn control further to the right until the
Turn the push-turn control to the right until the tem-
with the push-turn control.
PRINT
h
h
SETUP

Turn the push-turn control to the right until the fan
mb
IN 1
IN 2
OUT
IN 1
IN 2
OUT
°C
°C
MIN
AUTO
MAX
rh
DEFROSTERI
%
Turn the push-turn control to the right until the humidity
with the push-turn control.
%
content of the first ramp segment (only
supply, otherwise please con-
Turn the push-turn control to the right until the CO
with the push-turn control.
with lighting, otherwise please continue with item11):
Turn the push-turn control to the left until the
Turn the push/turn control to the right until a segment
with the push/turn control.
Turn the push-turn control further to the right until the
Turn the push-turn control to the right until the tem-
with the push-turn control.
CO
2
h
°C

Turn the push-turn control to the right until the fan
Turn the push-turn control to the right until the humid-
with the push-turn control.
%
content of the second ramp segment
Turn the push-turn control to the right until the CO
with the push-turn control.
Turn the push-turn control to the left until the
Turn the push-turn control to the left until the
Turn the push-turn control to the right until a segment
with the push/turn control.
Turn the push-turn control to the right until the tem-
rh
CO
2
h
°C

Turn the push-turn control to the right until the humidity
with the push-turn control.
%
content of the second ramp segment
Turn the push-turn control to the right until the CO
with the push-turn control.
for models with lighting, otherwise please continue with
Turn the push-turn control to the left until the
Turn push-turn control to the right until a close state-
Turn push-turn control to the right until
the display, and press the SET key briefly to confirm.
Turn the push-turn control to the right and adjust the
temperature monitoring (for more detailed information
Activating the programme:
Turn the push-turn control to the right until the stop
the push-turn control. Release the SET key, and the pro-
The cooled incubator can optionally be used, controlled and programmed with a PC/laptop.
The control of the appliance with the Memmert computer software “Celsius“ is
CO
2
°C
MIN
MAX

Warning messages during operation: See page
controller is initially deactivated. The CO
gas is introduced into the chamber via a sterile filter. The
CO
After the setpoint temperature has been reached,
concentration is displayed in % depending
concentration exceeds the
and wait to see if the controller steadily adjusts to
the setpoint. If the error occurs again, contact the
This monitoring function only starts
The pressure in the gas bottles is a constant
approx. 57 bar at 20 °C ambient tempera-
ture. The pressure in the bottle cannot be
2

The active humidity control guarantees that setpoint humidity is quickly reached, without the
after the setpoint temperature is reached, the humidification and dehumidification control
that it clicks into the appliance (
Switch off cooled incubator

Warning Messages and Malfunctions
Warning messages
The warning messages also set off an intermittent acoustic signal:
This can be temporarily switched off by pressing the SET key.
Error in the temperature control system (see also chapter "
Error in the temperature control system (see also chapter "
Error in the dehumidification system:
Error in the dehumidification system:
water supply tank with distilled
the setpoint. If the error occurs
again, contact the customer
system (only for models with CO
system (only for models with CO

Warning Messages and Malfunctions
the gas bottle is empty, replace
After removing covers, live parts may be exposed. You may
Error Possible cause Remedy

Warning Messages and Malfunctions
Error Possible cause Remedy
water, then reset humidity setpoint back to
the controller steadily adjusts to the setpoint.
The error can be rectified by the controller
ambient temperature + 8 °C
Wait until the appliance has cooled down.

Warning Messages and Malfunctions
and week time switch operating modes
and week time switch operating modes
The time and duration of the power failure are documented in the
After a power failure of less than 60 minutes, the current programme is continued from
the point at which it was interrupted. The time and duration of the power failure are
values (see table below).
The programme can only be continued from the computer. The time and duration of the
Parameters Default value

Advanced Functions
The cooled incubator is equipped with a parallel printer port as used in computers. Standard
The controller has an internal
). The log data can be printed out in
this mode via the connected printer.
turn control, as described on page 27. By turning
the push-turn control and holding down the SET
PRINT
SETUP

and changed, as described in chapter Basic operation on page 27:
The controller features a calendar that automatically
Tu
Acoustic signal at programme end
Acoustic signal at programme end
alarm, e.g. over/undertemperature
alarm, e.g. over/undertemperature
Automatic defrosting system
Automatic defrosting system
PRINT SETUP

for logging purposes in accordance with GLP. Date and clock time are specified on the log
Temperature monitoring
temperature sensor in the
for the appliance and surroundings.
The appliance is equipped with a double overtemperature protection (mechanical/electronic)
1
alarm icon lit up: TB
alarm icon lit up: TB
alarm flashing: TWW
alarm flashing: TWW
Automatic temperature monitor (ASF, see page
Overtemperature protection (TWW, TWB, see page
1
°C
6
5
2
3
MIN
AUTO
MAX
4

Mechanical temperature monitoring:
The cooled incubator is equipped with a mechanical temperature limiter (TB) of protection
temperature is exceeded by approx. 20 °C, the temperature limiter, as the final protective
Error rectification after the TB has been triggered:
Error rectification after the TB has been triggered:
). Press until you hear a „click“
Temperature monitoring can be adjusted indepen-
The manually set monitoring temperature
the overtemperature control is monitored by an
adjustable over/undertemperature controller (TWW) protection class 3.3 acc. to DIN 12880.
is exceeded, the TWW takes over
temperature control and begins to regulate the monitoring temperature (
The alarm icon flashes as a warning
°C
Schematic diagram of how the TWW
temperature monitoring works
The monitoring temperature must always be set sufficiently high above the
alarm is switched on in the SETUP, the TWW
alarm is additionally signalled by
an intermittent tone. If the SET key is pressed, the acoustic alarm can be temporarily switched
Emergency operation
Setting MAX
Set
temperature
Controller error
t

Adjustment range: up to max. 10 °C above nominal temperature
with the push-turn control (e.g. to
MIN
Adjustment range: 10 °C below the minimum temperature of
the appliance up to 10 °C above nominal temperature of the
appliance (for details of the nominal temperature, see nameplate).
The lower alarm limit value cannot be set higher than the top
MIN
°C
MAX
AUTO
°C
MAX
AUTO

ASF is a monitoring device that automatically follows the set temperature setpoint within an
adjustable tolerance band (
The ASF is activated – if switched on – automatically if the actual temperature value reaches
first time (section A). The activation of the ASF is shown by the brightly lit
When the temperature violates the set tolerance band around the setpoint (in the example in
: 37 °C ± 3 °C) – e. g. if the door is opened during operation (section B of illustration) –
alarm is set off. This is shown by the
alarm is switched on in the SETUP, the ASF
alarm is additionally signalled by an
alarm is automatically triggered as soon as 50 % of the set tolerance band of the
tolerance range of the new temperature setpoint (section E).
°C
Schematic diagram of how the ASF temperature monitoring works
Switching on the automatic temperature monitor:
Switching on the automatic temperature monitor:
with the push-turn control.
Switching off the automatic temperature monitor:
Switching off the automatic temperature monitor:
with the push-turn control.
The tolerance band for the ASF can be set in the SETUP
37 °C
40 °C
34 °C
A B C D E
40 °C
33 °C
27 °C
ASF active
34 °C
AUTO AUTO AUTO
ASF alarm
ASF active ASF active
MIN
MIN
t
°C
MAX
AUTO
°C
MAX
AUTO

The cooled incubator is equipped by default with a
). With this interface, it is possible to control and log the
appliance remotely from the computer. This is done with the help of the
To do this, the cooled incubator must be given a unique
). Using this, the respective cooled incubator can be selected and programmed from
the computer. The default setting is
The cooled incubator can optionally be equipped with an
ethernet interface to connect it to
a network. For identification purposes, each cooled incubator connected must have its own
XTADMIN", which can be found on the “Celsius“ CD-ROM, can be used to
192.168.1.216
One or more cooled incubators are connected to a network via
LAN 1: 192.168.1.233
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241

Log memory
The controller continually logs all relevant measured values, settings and error messages at
automatically with new data.
The logging function cannot be switched off and is always active. The measured data are
about three months in permanent operation.
Reading in the log memory to the computer via
Reading in the log memory to the computer via
The log data can be read out either via the
and from there be displayed graphically, printed out and stored.
Automatic defrosting system
The integrated automatic defrosting system for the
the cooled incubator at low temperatures and in permanent operation. The time values for
the automatic defrosting system can be set in the submenu
SETUP
working chamber. Afterwards, the smooth surface of the interior can easily be cleaned.

at regular intervals. If you would like to further reduce this detraction in performance, you can
for defrosting, 12 hours, is not sufficient. If this is the case, you should set a more frequent
Automatic defrosting is disabled with the parameter
tures, this causes the cooling unit to ice over over time. Regular defrosting needs to be carried
The device number of the appliance and a
To use it, insert the card into the chip
against all adjustments once the chip
Calibration
The cooled incubator can be calibrated customer-specifically using three calibration
temperatures of your choice:
Temperature calibration at low temperature (e.g. 5 °C)
Temperature calibration at medium temperature (e.g. 25 °C)
Temperature calibration at high temperature (e.g. 50 °C)
access
authority
card
Name:
ID:
_____________________
_____________________

), a positive or negative compensation correction value
Default
Temperature calibration (example)
accompanying compensation correction value to
temperature is too low, the compensation correction value must be set with a negative
Temperature deviation at 35 °C is to be corrected.
and set the corresponding
With a calibrated reference instrument and at a defined setpoint temperature of
calibration
+0,3 °C
CAL. 1
+0,2 °C
CAL. 2
-0,4 °C
h
°C
°C

h
After the calibration procedure, the reference instrument should display
°C, the factory calibration settings are
The cooled incubator can be calibrated for the individual customer by means of two balance
tive compensation correction value can be set between –5 % and +5 %.
Humidity calibration (example)
With a reference instrument, measure the deviation in the stationary state in the selected
The procedure can be performed with humidity balance points of 20 % rh and 80 % rh.
rh 20
+3,0 %
Default calibration
°C
°C
rh 80
-2,0 %
0%
20%
40%
60%
80% 100%

and set the accompanying compensation
IN 1
With a calibrated reference instrument, an actual humidity of 78 % rh is measured at
After the calibration procedure, the reference instrument should display 80.0 %.
a further comparison can be programmed at 20 % relative humidity.
% rh, the factory calibration settings
OUT
Tu
Mo
on
off
Tu
Mo
on
off
Sa Su
Fr
Th
We
Sa Su
Fr
Th
We
t3
t2
t1
t2
t1
t4
loop
SETUP
PRINT
t3
t4
loop
SETUP
PRINT
STERI DEFRO
4
3
2
1
4
3
2
1
STERI DEFRO
°C
°C
°C
MAX
MIN
AUTO
°C
MAX
MIN
AUTO
rh
%
mb
OUT
IN 2
rh
%
mb
IN 2
2
CO
mb
IN 1
OUT
IN 2
2
CO
mb
co2 10
Default calibration
+1,5 %
co2 15
-0,7 %
co2 5
-0,4 %
0%
10%
20%

balance point in the SETUP (
With a reference instrument, measure the deviation in the stationary state in the selected
The procedure can be performed for the CO
balance points 5 %, 10 % and 15 %.
deviation at 10 % should be corrected:
balance point in the SETUP to
and set the accompanying compensation
IN 1
With a calibrated reference instrument and at a defined setpoint CO
content of 11.5 % is measured.
The reference instrument should display 10 % after the calibration procedure.
IN 1
OUT
Tu
Mo
on
off
Tu
Mo
on
off
Sa Su
Fr
Th
We
Sa Su
Fr
Th
We
t3
t2
t1
t2
t1
t4
loop
SETUP
PRINT
t3
t4
loop
SETUP
PRINT
STERI DEFRO
4
3
2
1
4
3
2
1
STERI DEFRO
°C
°C
°C
MAX
MIN
AUTO
°C
MAX
MIN
AUTO
IN 2
rh
%
mb
IN 1
OUT
IN 2
rh
%
mb
OUT
IN 2
2
CO
mb
IN 1
OUT
IN 2
2
CO
mb

appearance and functionality of the stainless steel
To clean the interior, the fan cover can be removed by
the chamber can be cleaned with normal stainless steel
the working chamber or with the stainless steel housing.
working chamber due to impurities, immediately clean
and polish the affected area.
To make cleaning easier, the protective cover can be

function and long lifetime of the
After the screws (2) have been
front (3) can be removed and the
After the screws have been loos-
to clean the refrigeration module
from both sides if necessary.
that the hinge screws are not loose.
). Replace them if they are
Additional maintenance every two years, for models with CO
Additional maintenance every two years, for models with CO
4
1
3
2

fluorescent tubes only with
fluorescent tubes of the same type; e.g. UV
fications, please refer to chapter "
catch and pull off the plug
Open the bayonet
Open the bayonet
Pull out the illumination box
Turn the illumination box around and place it carefully on a smooth surface so that the
fluorescent tubes are on top (
Turn around the illumination
Remove Allen screws and retaining plate
and take out the glass cover.
Remove Allen screws and retaining plate
Remove Allen screws and retaining plate

2
Remove the plastic fixing and slide to the middle.
Pull the tubes carefully out of the holder.
Remove the plastic fixing and slide to the middle.
Remove the plastic fixing and slide to the middle.
Attach the glass cover and fix it with the restraining plate. To do so, screw in the two Allen
1
1
2

Adjusting door
A well-closing door is indispensable for cooled
appliances optimally guarantee the tight closing
the door closes exactly despite this, an adjust-
The top section (1) of the door hinge can be
the two screws (2) at the top and bottom of the
Adjust the door by turning the eccentric (3)
Apply the locking paint to the headless
Tighten the screws again.
voltage-carrying parts may
an electric shock if you touch
these parts. Disconnect the
2
Adjusting door
Adjusting door
Eccentric tappet
Eccentric tappet
1
3
4
5
6
7

Storage
The cooled incubator may only be stored under the following conditions:
This product is subject to the Directive 2002/96/EC on Waste Electrical Electronic
Any appliances that are infected, infectious or contaminated with
The appliance may not be left at public or communal recycling or col-

Temperature calibration 59
Temperature comparison 59
Temperature monitoring 53
Temperature protection 49
What to do in case of ac-

Memmert GmbH + Co KG | PO Box 1720 | D-91107 Schwabach, Germany | Phone +49 (0) 9122-925-0 | Fax +49 (0) 9122-145-85 | E-Mail: service@memmert.com | www.memmert.com
28.06.2013
ICH englisch
D23989