Memmert HCP150, HCP240, HCP50 Operating Manual

Page 1
HCP
OPERATING MANUAL
HUMIDITY CHAMBER HCP
Page 2
Manufacturer and customer service
MEMMERT GmbH + Co. KG Willi-Memmert Strasse 90-96 91186 Büchenbach Germany
Phone: +49 (0)9122 925-0 Fax: +49 (0)9122 14585 E-mail: sales@memmert.com Internet: www.memmert.com
Customer service: Service hotline: +49 (0)9171 9792 911 Service fax: +49 (0)9171 9792 979 E-mail: service@memmert.com
When calling for service enquiries, please state the appliance number given on the name­plate (see page 13).
Shipping address for repairs: Memmert GmbH + Co. KG Customer service Willi-Memmert Str. 90-96 91186 Büchenbach Germany
Please contact our customer service department before sending appliances for repair or before returning equipment, or the shipment may be refused.
© 2018 MEMMERT GmbH + Co. KG D39303 | Date 09/2018 Subject to change without notice
Page 3
D39303 | Date 09/2018 3
About this manual
About this manual
Purpose and target audience
This manual describes the design, function, transport, putting into operation, the actual op­eration and maintenance of HCP humidity chambers. It is intended for use by trained person­nel of the owner, who have the task of operating and/or maintaining the respective appliance.
If you are asked to work on the appliance, read this manual carefully before starting. Familiar­ise yourself with the safety regulations. Only perform work that is described in this manual. If there is something you do not understand, or certain information is missing, ask your man­ager or contact the manufacturer. Do not do anything without authorisation.
Versions
The appliances are available in different configurations and sizes. If specific equipment fea­tures or functions are available only for certain configurations, this is indicated at the relevant points in this manual.
The functions described in this manual refer to the latest firmware version. Due to individual configurations and sizes, illustrations in this manual may be slightly different
to the actual appearance. Function and operation are identical.
Other documents that have to be observed
When operating the appliance with the MEMMERT AtmoCONTROL PC software, observe the separate software manual. To open the AtmoCONTROL software manual, click on “Help” in the AtmoCONTROL menu bar.
Storage and resale
This operating manual belongs with the appliance and should always be stored where per­sons working on the appliance have access to it. It is the owner's responsibility to ensure that persons who are working on or are going to work on the appliance know where to find the operating manual. We recommend that it is always stored in a protected location close to the appliance. Make sure that the operating manual is not damaged by heat or humidity. If the appliance is resold or transported and then set up again at a different location, the operating manual must remain with it.
For the current version of this operating manual in PDF format, please go to http://www.memmert.com/en/service/downloads/user-manual/ .
Page 4
4 D39303 | Date 09/2018
Contents
1. For your safety 6
1.1 Terms and signs used........................................................................................................... 6
1.2 Product safety and dangers ................................................................................................ 7
1.3 Requirements of the operating personnel .......................................................................... 8
1.4 Responsibility of the owner ................................................................................................. 8
1.5 Intended use and improper use .......................................................................................... 8
1.6 Changes and alterations ...................................................................................................... 9
1.7 Behaviour in case of malfunctions and irregularities ..........................................................9
1.8 Switching off the appliance in an emergency .................................................................... 9
2. Construction and description 10
2.1 Design ................................................................................................................................ 10
2.2 Description and function .................................................................................................. 10
2.3 Working range ...................................................................................................................11
2.4 Material.............................................................................................................................. 12
2.5 Electrical equipment .......................................................................................................... 12
2.6 Connections and interfaces ............................................................................................... 12
2.7 Designation (nameplate) ................................................................................................... 13
2.8 Technical data .................................................................................................................... 14
2.9 Applied directives and standards ...................................................................................... 15
2.10 Declaration of conformity .................................................................................................15
2.11 Ambient conditions ...........................................................................................................16
2.12 Scope of delivery ...............................................................................................................16
2.13 Optional accessories .......................................................................................................... 16
3. Delivery, transport and setting up 17
3.1 For your safety ................................................................................................................... 17
3.2 Delivery .............................................................................................................................. 18
3.3 Transport ............................................................................................................................18
3.4 Unpacking ......................................................................................................................... 18
3.5 Storage after delivery ........................................................................................................18
3.6 Setting up .......................................................................................................................... 19
4. Putting into operation 21
4.1 Connect the appliance to the power supply ..................................................................... 21
4.2 Establish water supply .......................................................................................................21
4.3 Switching on ......................................................................................................................22
5. Operation and control 23
5.1 For your safety ................................................................................................................... 23
5.2 Operating personnel.......................................................................................................... 23
5.3 Opening the door .............................................................................................................. 23
5.4 Loading the appliance ....................................................................................................... 24
5.5 Operating the appliance .................................................................................................... 24
5.6 Monitoring function .......................................................................................................... 30
5.7 Graph ................................................................................................................................. 35
5.8 Ending operation ...............................................................................................................36
Contents
Page 5
D39303 | Date 09/2018 5
Contents
6. Malfunctions, warning and error messages 37
6.1 Warning messages of the monitoring function ................................................................ 37
6.2 Malfunctions, operating problems and appliance errors ................................................ 38
6.3 Power failure ...................................................................................................................... 40
7. Menu mode 41
7.1 Overview ............................................................................................................................ 41
7.2 Basic operation in menu mode using the example of language selection ....................... 42
7.3 Setup.................................................................................................................................. 43
7.4 Date and time .................................................................................................................... 45
7.5 Calibration ......................................................................................................................... 47
7.6 Program ............................................................................................................................. 51
7.7 Sounds ............................................................................................................................... 52
7.8 Protocol ............................................................................................................................. 53
7.9 USER ID .............................................................................................................................. 54
8. Maintenance and Servicing 55
8.1 Regular maintenance.........................................................................................................55
8.2 Cleaning ............................................................................................................................. 55
8.3 Repairs and Service ............................................................................................................ 55
9. Storage and disposal 56
9.1 Storage .............................................................................................................................. 56
9.2 Disposal ............................................................................................................................. 56
Index 57
Page 6
6 D39303 | Date 09/2018
For your safety
1. For your safety
1.1 Terms and signs used
In this manual and on the appliance itself, certain common terms and signs are used to warn you of possible dangers or to give you hints that are important in avoiding injury or damage. Observe and follow these notes and regulations to avoid accidents and damage. These terms and signs are explained below.
1.1.1 Terms used
WARNING
Warns of a dangerous situation that might lead to death or serious injuries
CAUTION
Warns of a dangerous situation that might lead to moderate or minor injuries
NOTICE
Warns of material damage
1.1.2 Signs used
Danger of electrocu-
tion
Fire hazard
Danger of burns
Danger from hot
steam
Do not tilt
Wear safety
boots
Disconnect the
mains plug
Observe information
in separate
manual
Page 7
D39303 | Date 09/2018 7
For your safety
1.2 Product safety and dangers
The appliances described in this manual are technically sophisticated, manufactured using high-quality materials and subject to many hours of testing in the factory. They reflect the state of the art and comply with recognised technical safety regulations. However, there are still risks involved, even when the appliances are used as intended. These are described below.
WARNING
After removing covers, live parts may be exposed. Touching these can lead to an electrical shock. Disconnect the mains plug before removing any covers. Work on the electrical system must only be performed by qualifi ed electricians.
WARNING
When loading the appliance with an unsuitable chamber load, poi­sonous or explosive vapours or gases may be produced. This could cause the appliance to explode, and people could be severely injured or poisoned. The appliance may only be loaded with materials/test objects which do not release any toxic or explosive vapours when heated (see also chapter Intended use and improper use on page
8).
WARNING
Hot steam can build up inside the appliance. You could be scalded on opening the door. Allow the appliance to cool before opening the door.
WARNING
Depending on operation, the surfaces in the interior of the appliance and the chamber load may still be very hot after the appliance is switched off. Touching these surfaces can cause burns. Wear heat­resistant protective gloves or wait until the appliance cools down before touching.
WARNING
Leaving the door open during operation can cause the appliance to overheat or pose a fi re hazard. Do not leave the door open during operation.
WARNING
Condensation in the electrical components may cause short circuits. After transporting or storing the device under humid conditions, re­move it from its packaging and let it ventilate for at least 24 hours in normal environmental conditions. Do not connect the device to the mains power during this time.
Page 8
8 D39303 | Date 09/2018
For your safety
1.3 Requirements of the operating personnel
The appliance may only be operated and maintained by persons who are of legal age and have been instructed accordingly. Personnel who are to be trained, instructed or who are un­dergoing general training may only work with the appliance under the continuous supervision of an experienced person.
Repairs may only be performed by qualified electricians. The regulations in the separate service manual must be observed.
1.4 Responsibility of the owner
The owner of the appliance
► is responsible for the flawless condition of the appliance and for it being operated in ac-
cordance with its intended use (see chapter 1.5);
► is responsible for ensuring that persons who are to operate or service the appliance are
qualified to do this, have been instructed accordingly and are familiar with the operating manual at hand;
► must know about the applicable guidelines, requirements and operational safety regula-
tions, and train staff accordingly;
► is responsible for ensuring that unauthorised persons have no access to the appliance; ► is responsible for ensuring that the maintenance plan is adhered to and that maintenance
work is carried out properly (see page 55);
► has to ensure that the appliance and its surroundings are kept clean and tidy, for example
through corresponding instructions and inspections;
► is responsible for ensuring that personal protective clothing is worn by operating person-
nel, e.g. work clothes and safety shoes.
1.5 Intended use and improper use
Humidity chambers may be used exclusively for temperature and climate testing of materials and substances in the context of the procedures and specifications described in this manual. Any other use is improper and may result in hazards and damage.
The appliance is not explosion-proof (does not comply with the German occupational health and safety regulation VBG 24). The appliance may only be loaded with materials and sub­stances which cannot form any toxic or explosive vapours at the set temperature and which cannot explode, burst or ignite.
The appliance may not be used for the drying, evaporation and burning-in of paints or similar materials, the solvents of which could form an explosive mixture when combined with air. If there is any doubt as to the composition of materials, they must not be loaded into the appli­ance. Potentially explosive gas-air mixtures must not form, neither in the chamber nor in the direct vicinity of the appliance.
Page 9
D39303 | Date 09/2018 9
For your safety
1.6 Changes and alterations
No unauthorised changes or alterations may be made to the appliance. No parts may be added or inserted which have not been approved by the manufacturer.
Unauthorised changes or alterations result in the CE declaration of conformity losing its valid­ity, and the appliance may no longer be operated.
The manufacturer is not liable for any damage, danger or injuries that result from unauthor­ised changes or alterations, or from non-compliance with the provisions in this manual.
1.7 Behaviour in case of malfunctions and irregularities
The appliance may only be used in a flawless condition. If you as the operator notice irregu­larities, malfunctions or damage, immediately take the appliance out of service and inform your superior.
You can find information on troubleshooting from page 37.
1.8 Switching off the appliance in an emergency
Press the main switch at the ControlCOCKPIT (Fig. 1) and disconnect the power plug. This disconnects the appliance from the power supply at all poles.
ON
ONN
Fig. 1 Switch off the appliance by press­ing the main switch
Page 10
10 D39303 | Date 09/2018
Construction and description
2. Construction and description
2.1 Design
1 2
3 4
6
5
7
Fig. 2 Design of HCP humidity chambers
1 ControlCOCKPIT with capacitive function
keys and LCD displays (see page 25) 2 On/Off switch (see page 22) 3 Chamber fan
4 Stainless steel perforated sheet 5 Inner glass door 6 Nameplate (see page 13)
2.2 Description and function
Air is heated inside the appliance by means of large-area all-round heating. Humidification of the interior is achieved by evaporating water from a tank at a set rate by means of a hot-air generator on the rear side of the appliance. The sterile hot steam is introduced into the inte­rior underneath the fan and mixed with the air current. Dehumidification takes place through the metered supply of fresh air via a sterile filter.
Page 11
D39303 | Date 09/2018 11
Construction and description
2.3 Working range
The temperature-humidity diagram (Fig. 3) specifies at what range of temperature and humi­dity a permanent operation is possible.
NOTICE
If in operation for long periods at the upper level or outside the working range, puddles of water may form inside the chamber and water may force its way out of the door seal.
B
A
C
°C
% rh
Fig. 3 Temperature-humidity working range (possible at an ambient temperature of 22 °C ± 3 K, relative humidity < 50 %)
Range A:
In this range, temperature and humidity can be combined as you please. If the chamber is in operation at the upper limit or outside the working range for long periods, puddles of water may form inside the chamber and water may leak out through the door seal. In extreme ambi­ent conditions, the working range may be restricted.
Range B:
If the specified range is exceeded upwards, the hot steam fed in will immediately condense, due to the dew-point, at the coldest point in the appliance.
Range C:
At low temperatures and low relative air humidity, the effective range is heavily dependent on the degree of humidity of the chamber load.
Page 12
12 D39303 | Date 09/2018
Construction and description
2.4 Material
For the outer housing, MEMMERT uses stainless steel (Mat.No. 1.4016 – ASTM 430) and for the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through its high stability, optimal hygienic properties and corrosion resistance to many (but not all!) chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with the materials mentioned. A material resistance table can be requested from the manufacturer.
2.5 Electrical equipment
► Operating voltage and current consumption: See nameplate ► Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010 ► Protection type IP 20 acc. to EN 60 529 ► Interference suppression acc. to EN 55011 class B ► Appliance fuse: Safety fuse 250 V/15 A, quick-blow ► The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V)
2.6 Connections and interfaces
2.6.1 Electrical connection
This appliance is intended for operation on an electrical power system with a system imped­ance Z
max
of a maximum of 0.292 ohm at the point of transfer (service line). The operator must ensure that the appliance is operated only on an electrical power system that meets these requirements. If necessary, you can ask your local energy supply company what the system impedance is.
Observe the country-specific regulations when making connections (e.g. in Germany DIN VDE 0100 with earth leakage circuit breaker).
2.6.2 Communication interfaces
The communication interfaces are intended for appliances which meet the requirements of IEC 60950-1.
USB interface
The appliance is fitted by default with a USB interface in accordance with the USB specifica­tion. This way, you can
► transfer software stored on a USB storage
medium to the appliance (see page 51).
► export protocol logs from the appliance to
a USB storage medium (see page 53).
► transfer user ID data stored on a USB stor-
age medium to the appliance (see page
54).
The USB port is located on the right side of the ControlCOCKPIT (Fig. 4).
Fig. 4 USB interface
Page 13
D39303 | Date 09/2018 13
Construction and description
Ethernet interface
Via Ethernet interface, the appliance can be connected to a network, so that you can trans­fer programs created with the AtmoCONTROL software to the appliance and read out proto­cols. The Ethernet interface is located on the rear of the appliance (Fig. 5).
For identification purposes, each appliance con­nected must have its own unique IP address. Configuration of the IP address is described on page 43.
You will find a description of how to transfer programs via Ethernet in the enclosed AtmoCONTROL manual.
The appliance can be directly connected to a computer / laptop using an optional USB to Ethernet converter (see Optional accessories on page 16).
2.7 Designation ( nameplate)
The nameplate (Fig. 6) provides information about the appliance model, manufacturer and technical data. It is attached on the upper right behind the door (see page 10).
Typ: HCP240 F.-Nr.: H818.3008 230 V~ 50/60 Hz
8.0
1840 W
DIN12880-Kl.3.1 Nenntemp.: 18-90 °C
1 2
3 4
5
6
10
8
9
7
Fig. 6 Nameplate (example) 1 Type designation
2 Operating voltage 3 Applicable standard 4 Protection type 5 CE conformity
6 Address of manufacturer 7 Disposal note 8 Temperature range 9 Connection / power ratings 10 Appliance number
Fig. 5 Ethernet interface
Page 14
14 D39303 | Date 09/2018
Construction and description
2.8 Technical data
Appliance size 50 105 150 240
Appliance width D* [mm] 559 719 719 759 Appliance height E* [mm]
(varies due to adjustable feet)
795 851 1071 1181
Appliance depth F* (without door handle) [mm] 521 591 591 691 Depth of door handle [mm] 56 Chamber width A* [mm] 400 560 560 600 Chamber height B* [mm] 425 480 700 810 Interior depth C* [mm] (less 35 mm for fan) 330 400 400 500 Chamber volume [litres] 56 107 156 241 Net weight [kg] 55 75 90 110 Weight including packaging [kg] 74 100 116 145 Power [W] 1520 1720 1800 1840 Mains voltage [AC V] 230/115 Mains frequency [Hz] 50/60 max. number of sliding grids/shelves 5 6 10 12 max. load per sliding grid [kg] 15
max. load per appliance [kg] 75 90 120 140
Operating temperature range
7 °C above room temperature up to
90 °C
Setting temperature range [°C] +18 bis +90
Adjustment precision [°C] 0.1 Adjustment range of active humidity control [% rh]
(only for appliances with the corresponding con­figuration)
20 to 95
Setting accuracy humidity [%] 0.5
* See Fig. 7 on page 15.
Page 15
D39303 | Date 09/2018 15
Construction and description
40,8
8
E
B
243,6
(79,3)(79,3)(79,3)(79,3)
D
56
F
C
A
Fig. 7 Dimensions
2.9 Applied directives and standards
► Directive 2004/108/EC amended (Directive of the European Parliament and of the Council
on the approximation of the laws of the Member States relating to electromagnetic com­patibility). Standards complied with: DIN EN 61326:2004-05, EN 61326:1997, EN 61326/A1:1998, EN 61326/A2:2001 EN 61326/A2:2003
► Directive 2006/95/EC amended (Directive of the European Parliament and of the Council
on the harmonisation of the laws of Member States relating to electrical equipment de­signed for use within certain voltage limits). Standards complied with: DIN EN 61 010-1 (VDE 0411 part 1) DIN EN 61 010-2-010 (VDE 0411 part 2-010) EN 61 010-1:2001, EN 61 010-2-010
2.10 Declaration of conformity
You can download the EC declaration of conformity of the appliance online: English: http://www.memmert.com/en/service/downloads/ce-statement/ German: http://www.memmert.com/de/service/downloads/eg-konformitaetserklaerung/
Page 16
16 D39303 | Date 09/2018
Construction and description
2.11 Ambient conditions
► The appliance may only be used in enclosed areas and under the following ambient condi-
tions:
Ambient temperature
10 °C to 35 °C
Humidity rh max. 70 % non-condensing
Overvoltage category II
Pollution degree 2
Altitude of installation max. 2,000 m above sea level
► The appliance may not be used in areas where there is a risk of explosion. The ambient air
must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is not explosion-proof.
► Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sedimentation in the interior and, as a consequence, could result in short circuits or dam­age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust or aggressive vapours from developing should be taken.
2.12 Scope of delivery
► Power cable ► 1 or 2 stainless steel perforated sheets (load capacity: 15 kg each) ► 1 water tank ► 2 sponge plugs for silicone entry ports ► USB storage medium with software and AtmoCONTROL manual ► the operating instructions at hand ► Safety data sheet ► Calibration certificates ► Separately packaged fixing material for wall mounting (see page 20)
2.13 Optional accessories
► USB to Ethernet converter (Fig. 8). Makes it
possible to connect the Ethernet connection interface of the appliance (see page 13) to the USB port of a computer/laptop.
Fig. 8 USB to Ethernet converter
Page 17
D39303 | Date 09/2018 17
Delivery, transport and setting up
3. Delivery, transport and setting up
3.1 For your safety
WARNING
Because of the heavy weight of the appliance, you could injure yourself if you try to lift it. At least two people are needed to carry ICO 50 appliances, or four people for ICO 105 and ICO 150 appli­ances. Appliances larger than that may not be carried, but must be transported using a manual pallet jack or forklift truck.
50
105
150 240
CAUTION
You may get your hands or feet squashed when transporting and installing the appliance. Wear protective gloves and safe­ty boots. Only grasp the appliance at the sides of the base:
9
Page 18
18 D39303 | Date 09/2018
Delivery, transport and setting up
3.2 Delivery
The appliance is packed in cardboard and is delivered on a wooden palette.
3.3 Transport
The appliance can be transported in the following ways:
► With a forklift truck; move the forks of the truck entirely under the pallet. ► On a manual pallet jack
3.4 Unpacking
NOTICE
► To avoid damage, do not unpack the appliance until you reach the installation site.
Remove the cardboard packaging by pulling it upwards or carefully cutting along an edge.
3.4.1 Checking for completeness and transport damage
► Check the delivery note to ensure that the delivery is complete. ► Check the appliance for damage.
If you notice deviations from the delivery note, damage or irregularities, do not put the appli­ance into operation but inform the haulage company and the manufacturer.
3.4.2 Remove the transportation lock
Remove the transportation lock. This is located between the door hinge, door and frame and can be removed after opening the door.
3.4.3 Disposing of packaging material
Dispose of the packaging material (cardboard, wood, foil) in accordance with the applicable disposal regulations for the respective material in your country.
3.5 Storage after delivery
If the appliance is first to be stored after delivery: Read the storage conditions from page
56.
Page 19
D39303 | Date 09/2018 19
Delivery, transport and setting up
3.6 Setting up
3.6.1 Preconditions
The installation site must be flat and horizontal and must be able to reliably bear the weight of the appliance (see Technical data on page 14). Do not place the appliance on a flamma­ble surface.
Depending on the model (see nameplate), a 230 V or 115 V power connection must be avail­able at the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm. The clear­ance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm (Fig. 9). Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times.
FP FP
FP
FP
Fig. 9 Minimum clearance from walls and ceiling
Page 20
20 D39303 | Date 09/2018
Delivery, transport and setting up
3.6.2 Tilt protection
Due to its centre of gravity, the appliance can fall over to the front and injure you or other people. Always attach the appliance to a wall with the tilt protection included in the delivery. In case there is not enough space, do not put the appliance into operation and do not open the door. Contact the Memmert service (see page 2).
1. Screw the tilt protection
onto the back of the ap­pliance as illustrated.
2. Bend the tilt protec-
tion upwards by 90° in the desired distance to the wall (observe the minimum distance to the wall, see Fig. 9).
3. Drill a hole, insert a
dowel and screw the tilt protection to a suitable wall.
Page 21
D39303 | Date 09/2018 21
Putting into operation
4. Putting into operation
NOTICE
When putting the appliance into operation for the first time, do not leave it unattended until it has reached a steady state.
4.1 Connect the appliance to the power supply
WARNING
Condensation in the electrical components may cause short circuits. After transporting or storing the device under humid conditions, re­move it from its packaging and let it ventilate for at least 24 hours in normal environmental conditions. Do not connect the device to the mains power during this time.
Observe the country-specific regulations when making connections (e.g. in Germany DIN VDE 0100 with earth leakage circuit breaker). Observe the connection and power ratings (see nameplate and "Technical Data" on page 14). Make sure to establish a safe PE conductor connection.
Plug the provided power cable into the rear of the appli­ance and connect it to the power supply ( Fig. 10 ). Place the power cable so that
► it is easily accessible at all times and can be pulled off quickly, for example in case of inter-
ference or an emergency;
► it does not represent a trip hazard; ► it cannot come into contact with any hot parts.
4.2 Establish water supply
4.2.1 Water specifi cations
For steam creation, use only: demineralised/desalinated water (Aqua Dem) with a conduc­tivity 10 μS/cm. Otherwise, possible calcification in the steam generators, steam piping and steam generator pumps could affect the correct functioning of the appliance. Only use chlorine-free water with a pH value of > 5 and < 7.
Fig. 10 Connect the power cable to the rear of the appliance
Page 22
22 D39303 | Date 09/2018
Putting into operation
4.2.2 Filling and connecting the water tank
Fill the supplied water tank with water and use the enclosed tube to connect it to the “H
2
O” connection
on the rear of the chamber (Fig. 11).
4.3 Switching on
Switch on the appliance by pressing the main switch on the front of the appliance ( Fig. 12 ).
The start-up process is shown by three animated white dots
. If the dots are any other colour, an error
has occurred (see page 39).
The appliance displays are in English by default when the appliance is switched on for the first time. You can change the language as described from page 42 . However, to get a basic overview of operating the appliance, you should read the following chapter first.
ON
N
N
Fig. 12 Switching on the appliance
Fig. 11 Water connections on the rear of the appliance
Page 23
D39303 | Date 09/2018 23
Operation and control
5. Operation and control
5.1 For your safety
WARNING
Leaving the door open during operation can cause the appliance to overheat or pose a fi re hazard. Do not leave the door open during operation.
WARNING
Hot steam can build up inside the appliance. You could be scalded on opening the door. Allow the appliance to cool before opening the door.
WARNING
Depending on operation, the surfaces in the interior of the appliance and the chamber load may still be very hot after the appliance is switched off. Touching these surfaces can cause burns. Wear heat­resistant protective gloves or wait until the appliance cools down before touching.
5.2 Operating personnel
The appliance may only be operated by persons who are of legal age and have been instruct­ed accordingly. Personnel who are to be trained, instructed or who are undergoing general training may only work with the appliance under the continuous supervision of an experi­enced person.
5.3 Opening the door
► To open the door, pull the door handle to the side (to the left or to the right, depending
on the door variation, see Fig. Fig. 13 , A).
► To close the appliance, push the door closed and the door handle to the side (B).
A
B
Fig. 13 Opening and closing the door
Page 24
24 D39303 | Date 09/2018
Operation and control
5.4 Loading the appliance
WARNING
When loading the appliance with an unsuitable load, poisonous or explosive vapours or gases may be produced. This could cause the appliance to explode, and people could be severely injured or poi­soned. The appliance may only be loaded with materials which do not form any toxic or explosive vapours when heated up and can­not ignite (see also Intended use and improper use on page 8). If there is any doubt as to the composition of materials, they must not be loaded into the appliance.
NOTICE
► Check the chamber load for chemical compatibility with the materials of the appliance (see
page 12).
Insert the sliding steel grids or sliding shelves. The maximum number of grids/shelves and the load capacity are specified in the technical data overview from page 14.
The appliance must not be loaded too tightly, so that proper air circulation in the interior is guaranteed. Do not place any of the chamber load on the bottom, touching the side walls or right below the ceiling of the chamber (Fig. 14, see also the "correct loading" sticker on the appliance).
In case of improper loading (not enough space between the items), the set tempera­ture may be exceeded or it may take longer until it is reached.
5.5 Operating the appliance
5.5.1 ControlCOCKPIT
In manual mode, the desired parameters are entered in the ControlCOCKPIT on the front of the appliance (Fig. 15 ). You can also make basic settings here (menu mode). Additionally, warning messages are displayed, e.g. if the temperature is exceeded. In program mode, the parameters defined, the program description, the program segment currently active and pro­gram duration remaining are displayed (for a more detailed description, see page 28).
Fig. 14 Correct placement of the chamber load
Page 25
D39303 | Date 09/2018 25
Operation and control
ON
N
N
TIMER
End 14: 45
44h:44m
Holz trocknen aufheizen 09:12h
Fr 20.10.2010 20:31
min
000°C
ALARM
of °C
max
000°C
auto off
99K
-
+
012
%rh°C
GRAPH
off
ON
LIGHTONLIGHT
ON
N
N
HUMIDITY
30%rh
Set 30 %rh
22.4
37.0
TEMP
Set
°C
°C
°C22.4
TEMP
°CSet 37.0
ALARM
max
38.5°C
auto
0.0K
min
35.5°C
+
-
312
15 169 10 13 1445 6 7 8
TIMER
30m04h
Ende
13:30 23.11.
Manual mode
12.09.2012 13:44
11 12
Fig. 15 ControlCOCKPIT
1 Activation key for temperature setpoint
adjustment 2 Setpoint and actual temperature display 3 Switch to menu mode (see page 41)
4 Activation key digital backwards counter
with target time setting, adjustable from 1
minute to 99 days 5 Main switch 6 Display digital backwards counter with
target time setting, adjustable from 1 min-
ute to 99 days 7 Display humidity setpoint and actual
value
8 Humidity control activation key 9 Turn control for setpoint adjustment 10 Confirmation key (accepts setting made
with the turn control) 11 Activation key for the appliance state 12 Appliance state and program display 13 Activation key setting the temperature
and humidity monitoring 14 Display of the temperature and humidity
monitoring 15 Graphical representation of setpoint and
actual values 16 Activation key for graphical representation
5.5.2 Basic operation
In general, all settings are made according to the following pattern:
1. Activate the desired parameter (e.g. tem­perature). To do so, press the correspond­ing activation key on the left or right or the respective display. The activated dis­play is lined in colour, the other displays are dimmed. The set value is highlighted in colour.
.5°C100
TEMP
TIME
R
22.4°C
TEMP
37.0°CSet
2. By turning the turn control to the left or right, adjust the set value (e.g. to
37.0 ºC).
22.4°C
TEMP
37.0°CSet
Page 26
26 D39303 | Date 09/2018
Operation and control
3. Save the set value by pressing the confir­mation key. The display returns to normal and the appliance begins adjusting to the defined set value.
23.2°C
TEMP
Set 37
.0°C
Additional parameters can be set accordingly.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati­cally restores the former values.
If you want to cancel the setting procedure, press the activation key on the left or right of the display that you want to exit. The appliance restores the former values. Only the settings that you have confirmed by pressing the confirmation key before cancelling the setting procedure are accepted.
5.5.3 Operating modes
The appliance can be operated in different modes:
► Manual mode: The appliance runs in permanent operation at the values set on the Con-
trolCOCKPIT. Operation in this mode is described in chapter 5.5.4 .
► Operation with digital backwards counter with target time setting, adjustable from 1
minute to 99 days (timer): The appliance runs at the values set until the timer has elapsed. Operation in this mode is described in chapter 5.5.5 .
► Program mode: The appliance automatically runs program sequences which have been
defined using AtmoCONTROL software at a computer / laptop and then transferred to the appliance from a USB stick or via Ethernet. Operation in this mode is described in chapter
5.5.6 .
► via remote control (AtmoREMOTE)
The status display shows you which operating mode or operating state the appliance is currently in. The current operating state is highlighted in colour and indicated by the text display:
Appliance is in program mode
Program is stopped
Appliance is in manual mode
The example on the right shows the appliance in manual mode, identified by the coloured hand symbol.
► When the appliance is in timer mode, Timer active
is displayed:
► When the appliance is in remote control mode, the
following symbol
appears in the temperature
display:
T
T
Manual mode
12.09.2012 13:44
Timer active
12.09.2012 13:44
23.2°C
TEMP
Set 38
.0°C
Page 27
D39303 | Date 09/2018 27
Operation and control
5.5.4 Manual mode
In this operating mode, the appliance runs in permanent operation at the values set on the ControlCOCKPIT.
Adjustment options
As described in chapter 5.5.2 , you can set the following parameters after pressing the cor­responding activation key (in any sequence):
Temperature Adjustment range: +18 °C to +90 °C
Heating operation is indicated by the
symbol.
You can select °C or °F as the temperature units displayed (see page 44).
22.4
37.0
TEMP
Set
°C
°C
Humidity Adjustment range: 20 to 95 % rh
Humidification is indicated by the
symbol.
Dehumidification is indicated by the
symbol.
When the appliance heats up, the humidity is dynamically adjusted to approach the setpoint depending on the dew point of the interior temperature.
HUMIDITY
70
%rh
Set
34%
rh
5.5.5 Operation with digital backwards counter with target time setting,
adjustable from 1 minute to 99 days ( timer)
In timer operation, you can adjust the time the appliance runs at the set values. The appliance has to be in manual operating mode for this.
1. Press the activation key to the left of the timer display. The timer display is activated.
TIMER
-
Ende
9:00 23.11.
--h- m
2. Turn the turn control until the desired duration is displayed – in this example 4 hours 30 minutes. The approximate end time is shown beneath, in a smaller font.
TIMER
End
0
13:30 23.11.
04 mh 3
Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes) format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum duration adjustable is 99 days 00 hours.
3. Press the confirmation key to confirm.
Page 28
28 D39303 | Date 09/2018
Operation and control
The display now shows the remaining time in a large font and the approximate end time in a smaller font beneath. The status display shows
Timer active.
TIMER
30m04h
End
13:30 23.11.
Timer active
12.09.2012 13:44
4. Now, as described under 5.5.2, set the individual values which you want the appliance to operate at. The set values can be changed at any time while the timer elapses. The changes are effective immediately.
In
Setup, you can choose if the timer should be setpoint-dependent or not. This deter-
mines whether the timer should not start until a tolerance band around the set tempera­ture is reached or if it should start immediately after activation (see page 44). The symbol on the timer display indicates that the timer is set to setpoint-dependent.
Once the timer has finished, the display shows 00h:00m. All functions (heating etc.) are switched off. In addition, an acoustic alarm sounds, which can be turned off by pressing the confirmation key.
Note that condensation may occur in the interior after you switch off the heating.
To deactivate the timer, open the timer display by pressing the activation key again and then turning the turn control to reduce the timer setting until --:-- is displayed. Press the confirmation key to confirm.
5.5.6 Program mode
In this operating mode, programs saved in the appliance can be started with different combinations of individual parameters (temperature, humidity, etc.) at staggered intervals, which the appliance then automatically processes in sequence. These programs are not cre­ated directly at the appliance but externally at a computer / laptop and using AtmoCONTROL software. Transfer to the appliance is possible using the provided USB storage medium or via Ethernet.
A description of how to create and save programs can be found in the separate AtmoCONTROL software manual.
Starting a program
1. Press the activation key next to the status display. The current operating mode is highlighted automatically, in this example
Manual mode ( ).
Holz trocknen aufheizen 09:12h
Fr 20.10.2010 20:31
off
ON
LIGHT
ON
LIGHT
Manueller Modus
12.09.2012 13:44
Manual mode
12.09.2012 13:44
Activate
2. Turn the turn control until the  start symbol is highlighted. The current program is displayed, in this example
Test012.
13:44
Test 012 Ready
12.11.2012
TIMER
00m00h
End
13:30 23.11.
TIMER
--m--h
End
9:00 23.11.
Page 29
D39303 | Date 09/2018 29
Operation and control
Only the program currently selected in menu mode and shown in the display can be used. If you want to process another program, you need to activate it in menu mode first (see description starting on page 51).
3. To start the program, press the confirma­tion key. The program is executed. The display shows:
► the program name ► the program segment description, in this
example
Ramp 1
► the current run (in case of loops)
13:44
Test 012 Ramp 1
12.11.2012
You cannot change any parameters (e.g. the temperature) at the appliance while a program is running. However, the displays
ALARM and GRAPH can still be used.
Cancel program
You can cancel an active program at any time.
1. Press the activation key next to the status display. The status display is automatically highlighted.
LIGHT
100
%
manueller Betrieb
IGHT
%
Fr 20.10.2010 20:31
ALARM
max
400°C
80
GRAPH
8
0
13:44
Test 012 Ramp 1
12.11.2012
2. Turn the turn control until the
stop
symbol is highlighted.
13:44
Cancel program
Test 012
12.11.2012
3. Press the confirmation key to confirm. The program is cancelled.
13:44
End
Test 012
12.11.2012
A cancelled program cannot be resumed at the point it was cancelled. It must be restarted from the beginning.
End of program
End
is shown on the display to indicate that
the program has finished.
13:44
End
Test 012
12.11.2012
Page 30
30 D39303 | Date 09/2018
Operation and control
You can now
► restart the program as described ► select another program to run in menu mode (see page 51) and run it as described. ► return to manual mode. To do so, reactivate it by
pressing the activation key next to the status display, then turn the turn control until the hand symbol
is
highlighted in colour and press the confirmation key.
5.6 Monitoring function
5.6.1 Temperature monitoring
The appliance is equipped with a multiple overtemperature protection in accordance with DIN 12 880. This serves to avoid damage to the chamber load and/or appliance in case of a malfunction:
► electronic temperature monitoring (TWW) ► automatic temperature monitor ( ASF) ► mechanical temperature limiter (TB)
The monitoring temperature of the electronic temperature monitoring is measured via a separate Pt100 temperature sensor in the interior. Temperature monitoring settings are made via the
ALARM display. The settings made apply to all operating modes.
If temperature monitoring has been triggered, this is indicated on the tempera­ture display: the actual temperature is highlighted in red and a
warning symbol is shown ( Fig. 16 ). The type of temperature monitoring triggered (TWW in this example) is shown beneath the temperature.
If the acoustic alarm has been activated in menu mode (
Sound see page 52,
indicated by the speaker symbol
on the alarm display), the alarm is additionally signalled by an intermittent acoustic signal, which can be turned off by pressing the confirma­tion key. Information on what to do if this happens can be found in chapter Malfunctions, warning and error messages from page 37.
Before reading how to adjust temperature monitoring (from page 32), please read the description of the individual monitoring functions here.
Electronic temperature monitoring ( TWW)
The manually set monitoring temperature min and max of the overtemperature control is monitored by an adjustable over/undertemperature controller (TWW) of protection class 3.3 according to DIN 12 880. If the manually set monitoring temperature
max is exceeded, the
TWW takes overtemperature control and begins to regulate the monitoring temperature (Fig.
17).
Manual mode
12.09.2012 13:44
ALARM
max
38.5°C
1.0K
min
35.5°C
auto
+
-
End 14:45
44h:44m
mi
n
00
0
OLIO
LI
FEUCHTE
30.0%rh
Set 30.0%rh
°C22.4
°CSet 37.0
TIMER
30m04h
Ende
23.Nov 13:30
°C
TEMP
TWW Set 38.5 °C
38.9
22.4
37.0
TEMP
Set
°C
°C
°C22.4
TEMP
°CSet 37.0
°C
TEMP
TWW Set 38.5 °C
38.9
Fig. 16 Temperature monitoring triggered
Page 31
D39303 | Date 09/2018 31
Operation and control
Setting MAX
Set temperature
t
°C
Controller error
Emergency operation
Fig. 17 Schematic diagram of how TWW temperature monitoring works
Automatic temperature monitor ( ASF)
ASF is a monitoring device that automatically follows the set temperature setpoint within an adjustable tolerance band (Fig. 18).
The ASF – if switched on – is automatically activated as soon as the actual temperature value reaches 50 % of the set tolerance band of the setpoint (in the example: 50 °C ± 1 K) for the first time (section A).
When the temperature violates the set tolerance band around the setpoint (in the example in Fig. 18: 50 °C ± 2 K) – e.g. if the door is opened during operation (section B of illustration) – the alarm is set off. The ASF alarm is automatically terminated as soon as 50 % of the set tolerance band of the setpoint (in the example: 50 °C ± 1 K) are reached again (section C).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (in this example: The setpoint is changed from 50 °C to 25 °C, section D), until it reaches the tolerance range of the new temperature setpoint (section E).
52°C
48°C
ASF active
52°C
48°C
ASF alarm
ASF active ASF active
27°C
23°C
°C
50°C
25°C
t
ABCDE
AUTO AUTO AUTO
Fig. 18 Schematic diagram of how the ASF temperature monitoring works
Page 32
32 D39303 | Date 09/2018
Operation and control
Mechanical temperature monitoring: Temperature limiter (TB)
The appliance is equipped with a mechanical temperature limiter (TB) of protection class 1 in accordance with DIN 12 880. If the electronic monitoring unit fails during operation and the default maximum temperature is exceeded by at least 20 °C, the tempera­ture limiter, as the final protective measure, switches off the heating permanently.
Adjusting temperature monitoring
1. Press the activation key to the left of the
ALARM display. The temperature moni-
toring setting is automatically activated (
).
min
000°C
ALARM
max
000°C
auto off
99K
-
+
ALARM
max
40.0°C
auto
0.0K
min
15.0°C
+
-
2. Save the selection by pressing the con-
firmation key. The
min setting (under-
temperature protection) is automatically activated.
ALARM
0.0K
min
°C
auto
max
°C
34 0.
40 0.
+
-
1. By turning the turn control, adjust the
desired lower alarm limit value, in the example on the right 35.5 °C.
The lower alarm limit value cannot be set higher than the top one. If no undertem­perature protection limit is required, set the lowest temperature.
ALARM
0.0K
min
°C
auto
max
°C
35 5.
40 0.
+
-
2. Press the confirmation key to confirm.
The
max display (overtemperature pro-
tection) is activated.
ALARM
0.0
K
min
°C
auto
max
°C
35 5.
40 0.
+
-
3. By turning the turn control, adjust the
desired upper alarm limit value, in the example on the right 38.5 °C.
The monitoring temperature must be set sufficiently high above the maximum set temperature. We recommend 0.5 to 1 K.
ALARM
0.0
K
min
°C
auto
max
°C35.5 38 5.
+
-
4. Accept the upper alarm limit value by
pressing the confirmation key. The setting of the automatic temperature monitor (ASF) is automatically activated (
auto).
ALARM
max
38.5 °C
0.0K
min
35.5°C
auto
+
-
TEMP
TB Set 80.0 °C
128
.4°C
Page 33
D39303 | Date 09/2018 33
Operation and control
5. With the turn control, select ON () or
OFF ().
ALARM
max
38.5 °C
0.0K
min
35.5°C
auto
+
-
6. Press the confirmation key to confirm.
The ASF tolerance band setting is acti­vated.
ALARM
K
min
°C
auto
max
38.5°C
35 5.
0.0
+
-
7. With the turn control, adjust the desired
tolerance band. We recommend 1 to 3 K.
ALARM
K
min
°C
auto
max
38.5°C
35 5.
1.0
+
-
8. Press the confirmation key to confirm.
Temperature monitoring is now active.
ALARM
max
38.5°C
1.0K
min
35.5°C
auto
+
-
5.6.2 Humidity monitoring
If humidity monitoring was triggered, this is indicated by the humidity display: the actual humidity is highlighted in red and a warning symbol
is shown ( Fig. 19 ). If the acoustic alarm has been activated in menu mode (
Sound, see page 52, as indicated
by the speaker symbol
), the alarm is additionally signalled by an intermittent acoustic signal. Information on what to do if this happens can be found in chapter Malfunctions, warning and error messages from page 41.
Adjusting humidity monitoring
1. Press the activation key to the left of the
ALARM display. The temperature moni-
toring setting is automatically activated.
min
000°C
ALARM
max
000°C
auto off
99K
-
+
ALARM
max
40.0°C
auto
0.0K
min
15.0°C
+
-
m
min
000
°
O
LIG
IG
O
LIG
FEUCHTE
30.0%rh
Set 30.0%rh
0C°C
°C22.4
MP
°CSet 37.0
a
m
m
:30
HUMIDITY
Set 70
.0%rh
%rh
75.4
FEUCHTE
30.0%rh
Set 30.0%rh
°CSet 37.0
HUMIDITY
Set 70
.0%rh
%rh
75.4
Fig. 19 Humidity monitoring triggered
Page 34
34 D39303 | Date 09/2018
Operation and control
2. Turn the turn control until the humidity
monitoring entry
is highlighted.
ALARM
min
40.0 %rh
max
60.0 %rh
3. Accept the selection by pressing the con-
firmation key. The lower humidity alarm limit is automatically highlighted.
ALARM
min
40.0%rh
max
60.0 %rh
4. By turning the turn control, adjust the
desired lower alarm limit, in the example on the right 50 % rh.
ALARM
min
50.0%rh
max
60.0 %rh
5. Accept the selection by pressing the con-
firmation key. The upper humidity alarm limit is automatically highlighted.
ALARM
min
50.0%rh
max
60.0 %rh
6. By turning the turn control, adjust the
desired upper alarm limit, in the example on the right 70 % rh.
ALARM
min
50.0%rh
max
70.0 %rh
7. Accept the selection by pressing the
confirmation key and leave the
Alarm
display by pressing the activation key on the side. Humidity monitoring is now active.
ALARM
min
50.0 %rh
max
70.0 %rh
Page 35
D39303 | Date 09/2018 35
Operation and control
5.7 Graph
The GRAPH display provides an overview of the chronological sequence of the setpoint values and actual values for temperature and humidity content as a curve.
Press the activation key to the right of the
GRAPH display. The
display is enlarged and the tem­perature profile shown.
0 4 8 12162024
°C
40 20
60
80
100
Fr 20.10.2010 20:34
12.09.2012
14.00 16.00 18.00
38
39
40
To display the setpoint and actual values for the humidity: Press the activation key next to the param­eter selection.
0 4 8 12162024
°C
40 20
60
80
100
Fr 20.10.2010 2
0
12.09.2012
14.00 16.00 18.00
38
39
40
Select the humidity symbol with the turn control. Accept the selec­tion by pressing the confirmation key.
12.09.2012
38
39
40
► To change the time frame to
be displayed: Press the activa­tion key next to the  ar­row symbols. The time frame to be displayed can now be changed by turning the turn control.
0 4 8 12162024
°C
40 20
60
80
100
Fr 20.10.2010 20
:
12.09.2012
14.00 16.00 18.00
38
39
40
► To zoom the graph in or out:
Press the activation key next to the magnifying glass sym­bol. Select whether you want to zoom in or out (+/–) with the turn control and confirm your selection by pressing the confirmation key.
.2010 20:34
2
To close the graphical representation, press the activation key you used to activate it again.
Page 36
36 D39303 | Date 09/2018
Operation and control
5.8 Ending operation
WARNING
Hot steam can build up inside the appliance. You could be scalded on opening the door. Allow the appliance to cool before opening the door.
WARNING
Depending on operation, the surfaces in the interior of the appliance and the chamber load may still be very hot after the appliance is switched off. Touching these surfaces can cause burns. Wear heat­resistant protective gloves or wait until the appliance cools down before touching.
1. Switch off active appliance functions
(turn back the set values).
2. Remove the chamber load.
3. Check the water tank and fill up if neces-
sary (see page 22).
4. Switch off the appliance with the main
switch (Fig. 20).
ON
N
N
Fig. 20 Switching off the appliance
Page 37
D39303 | Date 09/2018 37
Malfunctions, warning and error messages
6. Malfunctions, warning and error messages
WARNING
After removing covers, live parts may be exposed. Touching these can lead to an electrical shock. Do not try to rectify appliance er­rors yourself by opening the appliance, but contact the MEMMERT customer service department (see page 2) or an authorised service point.
6.1 Warning messages of the monitoring function
If the acoustic alarm has been activated in menu mode (Sound, see page 52, as indicated by the speaker symbol
), the alarm is additionally signalled by an intermittent acoustic signal. If the confirmation key is pressed, the acoustic alarm can be temporarily switched off until the next alarm event occurs.
6.1.1 Temperature monitoring
Description Cause Action See
Temperature alarm and "ASF" are displayed
TEMP
ASF Set 38.5 °C
40.4°C
Automatic tem­perature monitor (ASF) was trig­gered.
Check if the door is closed. Close the door.
Extend the ASF tolerance band If the alarm continues: Contact
customer service
Page 32 Page 60
Temperature alarm and "TWW" are displayed
TEMP
TWW Set 38.5 °C
42
.4°C
The adjustable temperature controller (TWW) has assumed heating control.
Increase the difference between the monitoring and setpoint temperature – by either increas­ing the max value of the tem­perature monitoring or decreas­ing the setpoint temperature.
If the alarm continues: Contact customer service
Page 32
Page 60
Temperature alarm and "TB" are displayed
TEMP
TB Set 80.0 °C
128
.4°C
The mechanical temperature lim­iter (TB) perma­nently switched off heating.
Switch off the appliance and leave to cool down.
If the error occurs again, contact the customer service. Page 2
Page 38
38 D39303 | Date 09/2018
Malfunctions, warning and error messages
6.1.2 Humidity monitoring (only for appliances in the corresponding confi guration)
Description Cause Action See
Error display symbol
HUMIDITY
Set 55
.0%rh
%rh
55
.4
Water tank empty
Fill the water tank with deminer­alised/distilled water and press the confirmation key
Page 22
Alarm display (
MaxAl)
HUMIDITY
MaxAl Set 70
.0%rh
%rh
75
.4
Upper humidity limit exceeded
Open the door for 30 sec. and wait to see if the appliance reli­ably adjusts to the setpoint.
If the error occurs again, contact customer service. Page 2
Alarm display (
MinAl)
HUMIDITY
MinAl Set 60
.0%rh
%rh
55
.4
Humidity below lower limit
Check if the door is closed. Check the water supply and the
filling level of the water tank. If required, refill the water tank/ water trays.
If the error occurs again, contact customer service.
Page 22
Page 2
6.2 Malfunctions, operating problems and appliance errors
Description Cause Action See
Displays are dark External power supply
was interrupted
Check the power supply
Page 21
Miniature fuse, appliance fuse or power module faulty
Contact customer service
Page 60
Displays cannot be activated
Appliance locked by USER IDUnlock with USER ID Page 54
The appliance is in pro­gram, timer or remote control mode (mode "Write" or "Write + Alarm")
Wait until the end of the program or timer mode or switch off the remote control
Displays suddenly look different
Appliance is in "wrong" mode
Change to operating or menu mode by pressing the
MENU
key
Page 39
D39303 | Date 09/2018 39
Malfunctions, warning and error messages
Description Cause Action See
Error message T:E-3 in the temperature display
TEMP
T:E-3 Set 37.0 °C
37.4°C
Temperature operating sen­sor is defective. The moni­toring sensor takes over the measurement function.
► The appliance can
temporarily be kept in service
► Contact customer
service as soon as possible
Page 2
Error message
AI E-3 in
the temperature display
TEMP
AI E-3 Set 37.0 °C
37.4°C
Temperature monitoring sensor is defective. The operating sensor takes over the measurement function.
► The appliance can
temporarily be kept in service
► Contact customer
service as soon as possible Page 2
Error message
E-3 in
the temperature display
TEMP
Set 45.0 °C
E-3 °C
Operating and monitoring sensor defective
► Switch off appli-
ance
► Remove the cham-
ber load
► Contact customer
service
Page 2
Error message
E-6 in
the humidity display
HUMIDITY
Set 50
.0%rh
%rh
E-6
Humidity sensor defective ► No humidity con-
trol possible
► Contact customer
service
Page 2
When switching on the appliance, the start animation is displayed in another colour than white
► Cyan :
Not enough storage space on the SD card
► Red :
The system files could not be loaded
► Orange : The
fonts and images could not be loaded
Contact customer service
Contact customer service
Download the firm­ware update from memmert.com and install it
Page 2
Page 2
Page 40
40 D39303 | Date 09/2018
Malfunctions, warning and error messages
6.3 Power failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set. The time and duration of the power failure are documented in the log memory.
In timer or program mode
In case of an interruption of the power supply of less than 60 minutes, the current program is continued from the point at which it was interrupted. For longer interruptions of the power supply, all appliance functions (heating, fan etc.) are switched off.
In remote control mode
The previous values are restored. If a program has been initiated via remote control, it is con­tinued.
Page 41
D39303 | Date 09/2018 41
Menu mode
7. Menu mode
In menu mode, you can make basic settings, load programs and export protocols, as well as calibrate the appliance.
Caution:
Before changing menu settings, read the description of the respective functions on the fol­lowing pages to avoid possible damage to the appliance and/or chamber load.
To enter menu mode, press the MENU key.
To exit the menu mode at any time, press the MENU key again. The appliance then returns to operating mode. Only changes accepted by pressing the confirmation key are saved.
7.1 Overview
Press the MENU key to change between the displays in menu mode:
SETUP
SIGNALTÖNE
PROG
ZEIT UND DATUM
CALIB
USER ID
PROTOCOL
LANGUAGE
TIME
SOUND
ON
N
N
12 34 512141513
67 18 1989 1011 16 17
Fig. 21 ControlCOCKPIT in menu mode
1 Language selection activation key 2 Language selection display 3 Date and time display 4 Date and time setting activation key 5 Exit menu mode and return to operating
mode
6 Setup activation key (basic appliance set-
tings) 7 Setup display (basic appliance settings) 8 Adjustment display 9 Adjustment activation key
10 Turn control for adjustment 11 Confirmation key (accepts setting made
with the turn control) 12 Program selection activation key 13 Program selection display 14 Protocol display 15 Protocol activation key 16 Acoustic signal adjustment activation key 17 Acoustic signal adjustment display 18 USER ID display 19 USER ID display activation key
Page 42
42 D39303 | Date 09/2018
Menu mode
7.2 Basic operation in menu mode using the example of language selection
In general, all settings in menu mode are done just like in operating mode: Activate the respective display, use the turn control for setting and press the confirmation key to accept the change. A more detailed description is provided in the following, using the example of language selection.
1. Activate the desired parameter (in this
example the language). To do so, press the corresponding activation key on the left or right or the respective display. The activated display is enlarged.
ENGLISH
DEUTSCH
ESPANOL POLSKI CESTINA
FRANCAIS
If you want to exit or cancel the settings, again press the activation key which you have used to activate the display. The appliance returns to the menu overview. Only the settings that you have con­firmed by pressing the confirmation key before cancelling the setting procedure are accepted.
SETUP
SIGNALTÖNE
CALIB
LANGUAGE
TIME
ON
N
N
2. With the turn control, select the desired
new setting, e.g. Español (Spanish).
ENGLISH DEUTSCH
MAGYAR ITALIANO
ESPANOL
FRANCAIS
POLSKI CESTINA
3. Save the setting by pressing the confir-
mation key.
ENGLISH DEUTSCH
MAGYAR ITALIANO
ESPANOL
FRANCAIS
POLSKI CESTINA
4. To return to the menu overview, press the
activation key again.
SETUP
SIGNALTÖ
N
CALI
B
LANGUAGE
TIM
You can now
► activate another menu function by press-
ing the corresponding activation key or
► return to operating mode by pressing the
MENU key.
Balance
Einheit
IP Adresse
Alarm Temp Timer Mode
Unit
IP address
Subnet mask
Alarm Temp Timer Mode
Page 43
D39303 | Date 09/2018 43
Menu mode
All other settings can be made accordingly. The settings possible are described in the follow­ing sections.
If no new values are entered or confirmed for approx. 30 seconds, the appliance automati­cally restores the former values.
7.3 Setup
7.3.1 Overview
In the SETUP display, you can set the following parameters:
► the IP address and subnet mask of the appliance's Ethernet interface (for connection to a
network)
► The units of the temperature display (°C or °F, see page 44) ► How the digital backwards counter with target time setting (Timer Mode, see page 44) ► Remote control (see page 45) ► Gateway (see page 45)
If the Setup menu contains more entries than can be displayed, this is indicated by the display "1/2". This means that there is a second "page" of entries.
To display the hidden entries, use the turn control to scroll beyond the lowest entry. The page display changes to "2/2".
7.3.2 IP address and subnet mask
If you want to operate one ore more appliances in a net­work, each appliance must have its own unique IP address for identification. By default, each appliance is delivered with the IP address 192.168.100.100.
LAN 1: 192.168.1.233
192.168.1.216
LAN 2: 192.168.1.215
LAN 3: 192.168.1.241
2.
8.
1
Editor
Programmname
Simulation Protokoll
-
+-+
INP 250 Test 01
180
.0°C
i
HPP 250 Labor
i
37
.0°C
44
.4%rh
15
STAND BY
Programm negnulletsniEllokotorPtäreG efliHnekcurD
AtmoCONTROL
Fig. 22 Operation of several appliances in a network (schematic example)
Setup
Unit
IP adress 255.145.136.22
5
Subnet mask 255.255.0 .0
Alarm Temp
°C F
Timer mode
TWW TWB
1/2
Page 44
44 D39303 | Date 09/2018
Menu mode
1. Activate the
SETUP display. The entry IP
address is automatically highlighted.
SETUP
Balance
Einheit
IP Adresse
+30%
192.168.100.100
Alarm Temp
°C F TWW TWB
Timer Mode
SETUP
Unit
IP address 192.168.100.100
Subnet mask 255.255.0. 0
Alarm Temp
°C F TWW TWB
Timer Mode
2. Accept the selection by pressing the con-
firmation key. The first three digits of the IP address are automatically selected.
SETUP
Einheit
IP-Adresse 192.168.1 0 0 . 100
Subnetmask 255. 255.0 .0
Alarm Temp
°C F TWW TWB
Timer Mode
3. With the turn control, set the new num-
ber, e.g. 255.
SETUP
Unit
IP address 255.168.100. 100
Subnet mask 255.255.0.0
Alarm temp
°C F TWW TWB
Timer mode
4. Accept the selection by pressing the
confirmation key. The next three digits of the IP address are automatically selected. Setting these is done according to the description above.
SETUP
Unit
IP address 255.168.100.100
Subnet mask 255.255.0.0
Alarm temp
°C F TWW TWB
Timer mode
5. After setting the last three digits, accept
the new IP address by pressing the con­firmation key. The selection returns to the overview.
The subnet mask is set accordingly.
SETUP
Unit
IP address 255.145.136.225
Subnet mask 255.255.0.0
Alarm temp
°C F TWW TWB
Timer mode
7.3.3 Unit
Here you can choose whether the temperature is displayed in °C or °F.
7.3.4 Timer Mode
Here, you can choose whether the digital backwards counter with target time setting (timer, see page 27) should run setpoint­dependent or not – this determines whether the timer should not start until a tolerance band of ±3 K around the setpoint tempera­ture is reached (Fig. 23, B) or whether it should start right after activation (A).
SETUP
Unit
IP address 255.145.136.225 Subnet mask 255.255.0.0
Alarm temp
°C °F
TWW TWB
Timer mode
Unit
IP address 255.145.136.225 Subnet mask 255.255.0.0
Alarm temp
°C F TWW TWB
Timer mode
Page 45
D39303 | Date 09/2018 45
Menu mode
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 23 Timer mode A Timer independent of setpoint: Timer starts immediately after activation B Timer setpoint-dependent: Timer does not start until tolerance band is reached
7.3.5 Remote control (AtmoREMOTE)
In the setup entry remote control, you can set whether the appliance should be controlled via remote control and if so, in which mode. These settings are available:
Off Read only Write Write + Alarm
If the appliance is in remote control mode, the symbol appears in the temperature display. In the settings
Write
and Write + Alarm, the appliance cannot be controlled at the ControlCOCKPIT until the remote control has been switched off (setting
Off) or set to Read only.
In order to use the remote control function, program­ming skills and special libraries are required.
7.3.6 Gateway
The setup entry gateway is used to connect two networks with different protocols.
The gateway is set the same way as the IP address (see page
43).
7.4 Date and time
In the TIME display, you can set the date and time, time zone and daylight saving time.
NOTICE
Always set the time zone (and daylight saving time yes/no) before you set the date and time. Avoid changing the set time after that since this can lead to gaps or overlapping when recording measured values. If you still need to change the time, you should not run a program immediately before or after doing so.
Remote Control Off
Gateway 192.168.5 .1
23.2°C
TEMP
Set 38
.0°C
Remote Control Off
Gateway 192.168.5 .1
Page 46
46 D39303 | Date 09/2018
Menu mode
1. Activate the time setting. To do so, press
the activation key on the right side of the
TIME display. The display is enlarged
and the first adjustment option (
Date)
automatically highlighted.
Date and time
Time Time zone Daylight savings
12.05.2012
12:00 GMT 01:00
Date
2. Turn the turn control until Time zone is
highlighted.
Time
Daylight savings
12.05 . 2012 12:00 GMT 01:00
Date
Time zone
3. Accept the selection by pressing the
confirmation key.
Time
Daylight savings
12.05 . 2012 12:00 GMT
Date
Time zone
01:00
4. Set the time zone of the installation site
with the turn control, e.g. 00:00 for Great Britain and 01:00 for Germany, France or Spain. Accept the selection by pressing the confirmation key.
Time
Daylight savings
12.05 . 2012 12:00 GMT
Date
Time zone
00:00
5. With the turn control, select the Daylight
savings
entry.
Time
12.05 . 2012 12:00 GMT 00:00
Date
Time zone
Daylight savings
6. Accept the selection by pressing the
confirmation key. The adjustment options are highlighted.
Time
12.05 . 2012 12:00 GMT 00:00
Date
Time zone
Daylight savings
7. Set daylight savings to off () or on
(
) with the turn control – in this case
on (
). Save the setting by pressing the
confirmation key.
Time
12.05 . 2012 12:00 GMT 00:00
Date
Time zone
Daylight savings
Daylight saving time and standard time are not changed automatically. For this reason, please keep in mind to adjust them at the beginning of each period.
Page 47
D39303 | Date 09/2018 47
Menu mode
8. Now, set date (day, month year) and time
(hours, minutes) in the same way. Accept the selection by pressing the confirma­tion key.
Time Time zone Daylight savings
..2012 12:00 GMT 00:00
Date 27 05
7.5 Calibration
NOTICE
To guarantee perfect control, we recommend to calibrate the appliance once a year.
7.5.1 Temperature calibration
The appliances are temperature calibrated and adjusted at the factory. In case readjustment should be necessary later on – for example due to influence of the chamber load – the ap­pliance can be calibrated customer-specifically using three calibration temperatures of your choice:
► Cal1 Temperature calibration at low temperature ► Cal2 Temperature calibration at medium temperature ► Cal3 Temperature calibration at high temperature
For temperature calibration, you will need a calibrated reference measuring device.
CAL 1
+0,5°C
CAL 3
+0,4°C
CAL 2
-0,4°C
factory calibration
20°C 30°C 40°C 50°C
0°C
10°C
Fig. 24 Schematic example of temperature calibration
Example: Temperature deviation at 42 °C should be corrected.
Page 48
48 D39303 | Date 09/2018
Menu mode
1. Press the activation key to the right of the
CALIB display. The display is enlarged
and the temperature adjustment option is automatically selected.
Cal1
°C -0,2
K
Cal2 100.0 °C +0,1 K Cal3 180.0 °C -0,2 K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
CK0,2- CK0,1+ CK0,2-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
40.0
60.0
30.0
Temperature
Humidity
2. Press the confirmation key repeatedly, until the compensation temperature Cal2 is selected.
CK0,2- CK0,1+ CK0,2-
e
Cal1
Cal2
Cal3
60.0
30.0
40.0
3. With the turn control, set the compensa­tion temperature Cal2 to 42 °C.
CK0,2- CK0,1+ CK0,2-
Cal1
Cal2
Cal3
60.0
30.0
42.0
4. Save the setting by pressing the confir­mation key. The corresponding calibra­tion value is automatically highlighted.
CK0,2-
CK
CK
0,2-
e
Cal1
Cal2
Cal3
60.0
30.0
42.0 0,1+
5. Set the calibration value to 0.0 K and ac­cept the setting by pressing the confirma­tion key.
CK0,2-
CK
CK
0,2-
e
Cal1
Cal2
Cal3
60.0
30.0
42.0 0,0
6. Position the sensor of a calibrated refer­ence instrument centrally in the appli­ance's working chamber.
7. Close the door and, in manual mode, adjust the set temperature to 42 °C.
21.4°C
TEMP
42.0 °CSet
8. Wait until the appliance reaches the set temperature and displays 42 °C. The ref­erence instrument should display 43.6 °C.
42.0°C
TEMP
Set 42
.0°C
43.6 °C
Page 49
D39303 | Date 09/2018 49
Menu mode
9. In the SETUP, adjust the calibration value Cal2 to +1.6 K (reference value measured minus value displayed) and save the set­ting by pressing the confirmation key.
CK0,2-
CK
CK
0,2-
e
Cal1
Cal2
Cal3
60.0
30.0
42.0 1,6+
10. After the calibration procedure, the temperature measured by the reference instrument should now also be 42 °C.
42.0°C
TEMP
Set 42
.0°C
42,0 °C
With Cal1, a calibration temperature below Cal2 can be programd accordingly, and with Cal3, a temperature above. The minimum interval between the Cal values is 10 K.
If all compensation correction values are set to 0.0 K, the factory calibration settings are restored.
7.5.2 Humidity calibration
Humidity control can be adjusted according to customer requirements by means of three freely selectable balance points. For each selected calibration point, a positive or negative compensation correction value between –10% and +10% can be set (Fig. 25).
For humidity adjustment, you will need a calibrated reference measuring device.
rh 20
+3,0 %
rh 90
-1,5 %
60%
40%
0%
20%
80% 100%
rh 50
+1,5 %
Factory calibration
Fig. 25 Humidity adjustment (example)
Example: Humidity deviation at 60 % should be corrected:
1. Press the activation key to the right of the
CALIB display. The display is enlarged
and the temperature adjustment option is automatically selected.
Cal1
°C -0,2
K
Cal2 100.0 °C +0,1 K Cal3 180.0 °C -0,2 K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
CK0,2- CK0,1+ CK0,2-
Calibration
Last updated: 12.10.2012 12:00:00
Cal1
Cal2
Cal3
40.0
60.0
30.0 Temperature
Page 50
50 D39303 | Date 09/2018
Menu mode
2. Turn the turn control until
Humidity is
highlighted.
%rh
%rh
%0,5- %1,0+
Temperature
Cal1
Cal2
40.0
50.0
Humidity
%rh
%
1,0+Cal3
80.0
3. Press the confirmation key repeatedly, until the calibration point Cal2 is selected.
%rh
%rh
%0,5- %1,0+
Temperature
Cal1
Cal2
40.0
Humidity
50.0
%rh
%
1,0+
Cal3 80.0
4. With the turn control, set the calibration point Cal2 to 60% rh.
%rh
%rh
%0,5- %1,0+
Temperature
Cal1
Cal2
40.0
Humidity
60.0
%rh
%
1,0+Cal3
80.0
5. Save the setting by pressing the confir­mation key. The corresponding calibra­tion value is automatically highlighted.
%rh
%rh
%0,5-
%
JUSTIEREN
Temperature
Cal1
Cal2
40.0
Humidity
60.0 1,0+
%rh
%
1,0+Cal3
80.0
6. Set the calibration value to 0.0% and ac­cept the setting by pressing the confirma­tion key.
%rh
%rh
%0,5-
%
Temperature
Cal1
Cal2
40.0
Humidity
60.0 0,0
%rh
%
1,0+Cal3 80.0
7. Position the sensor of the calibrated refer­ence instrument centrally in the working chamber of the appliance.
8. Close the door and, in manual mode, adjust the set humidity to 60% rh.
HUMIDITY
29.5%rh
Set 60.0%rh
9. Wait until the appliance reaches the set humidity and displays 60% rh. The refer­ence instrument should display 58.5% rh.
HUMIDITY
60.0%rh
Set 60.0%rh
58.5 %rh
Page 51
D39303 | Date 09/2018 51
Menu mode
10. In the SETUP, adjust the compensation correction value Cal2 to -1.5% (actual value measured minus setpoint humid­ity) and save the setting by pressing the confirmation key.
%rh
%rh
%0,5-
%
Temperature
Cal1
Cal2
40.0
Humidity
60.0 -1.5
%rh
%
1,0+Cal3
80.0
11. After the calibration procedure, the hu­midity measured by the reference instru­ment should now also be 60% rh.
HUMIDITY
60.0%rh
Set 60.0%rh
60.0 %rh
7.6 Program
In the Program display, programs created using the AtmoCONTROL software can be trans­ferred to the appliance and saved on a USB storage medium. Here, you can also select the program provided for use (see page 28 ) and delete programs.
To load a program from a USB stor­age medium: Connect the USB storage medium with the saved program(s) to the interface on the right side of the Control­COCKPIT.
1. Activate the program display. To do so, press the activation key on the left side of the
Prog display. The display is enlarged
and the entry
Select automatically
highlighted. The programs available for activation are shown on the right. The program currently available for use – in this example
Test 012 – is highlighted in
orange.
Program
Delete
Test 012
Test 013 Test 014 Test 015 Test 016
Select
2. Access the Select function by pressing the confirmation key. All programs avail­able are displayed, including the ones saved on the USB data storage medium (identified by the
USB symbol). The program currently available for use is highlighted in orange.
Select
Delete
Test 012
Test 022 Test 013 Test 014 Test 022
3. With the turn control, select the program you want to make available for use.
Select
Delete
Test 012
Test 022 Test 013 Test 014
Test 023
Page 52
52 D39303 | Date 09/2018
Menu mode
4. Accept the selection by pressing the confirmation key. The program is now loaded, which is indicated by the transfer symbol.
Select
Delete
Test 012
Test 022 Test 013 Test 014
Test 023
5. As soon as the program is ready, the selection returns to
Select. To start the
program: Return to operating mode by pressing the MENU key, as described on page 28.
Delete
Test 012 Test 022 Test 013 Test 014
Test 015
Test 023
Select
If you connected a USB storage medium, you can now remove it. To delete a program, select
Delete with the turn control and select the program to be de-
leted the same way you can select a program for activation.
7.7 Sounds
In the SOUND display, you can define whether or not the appliance should emit acoustic signals and, if yes, define on which events it should do so:
► on the press of a key ► at the end of a program ► On alarm ► if the door is open
1. Activate the acoustic signal adjustment. To do so, press the activation key on the left side of the
SOUND display. The
display is enlarged. The first category (in this case
Keysound) is automatically
highlighted. On the right, the current set­tings are shown on.
Sound
Keysound
At the end
If door open
On alarm
If you want to edit another list entry: Turn the turn control until the respective entry – e.g.
if door open – is highlighted
in colour.
Keysound At the end
If door open
On alarm
Sound
2. Save the selection by pressing the confir­mation key. The adjustment options are automatically highlighted.
Keysound At the end
If door open
On alarm
Sound
Page 53
D39303 | Date 09/2018 53
Menu mode
3. With the turn control, select the desired setting – in this example OFF ().
Keysound At the end
If door open
On alarm
Sound
4. Save the setting by pressing the confir­mation key.
If an acoustic alarm sounds, it can be turned off by pressing the confirmation key.
Keysound At the end
If door open
On alarm
Sound
7.8 Protocol
The appliance continually logs all relevant measured values, settings and error messages at 1-minute intervals. The internal log memory is of the continuous memory type. The logging function cannot be switched off and is always active. The measured data are stored in the appliance, safe from manipulation. If the power supply is interrupted, the time of the power failure and voltage recovery are stored in the appliance.
You can export the protocol data for different periods to a USB storage medium via the USB port or, via Ethernet, import them to the AtmoCONTROL software for graphical representa­tion, print-out or storage.
The log memory of the appliance is not modified or deleted by reading it out.
1. Connect the USB storage medium to the USB port on the right of the Control­COCKPIT.
2. Activate the protocol. To do so, press the activation key on the right side of the
PROTOCOL display. The display is
enlarged and the period
This month au-
tomatically highlighted. To select another logging period, use the turn control.
Protocol
12.10.2012 12:00
ontrol range
3. Save your selection by pressing the confir­mation key. The transfer starts and a status symbol indicates the progress.
Protocol
This Year Complete control range
This Month
4. As soon as the transfer is complete, a check mark appears in front of the period selected. You can now remove the USB storage medium.
Protocol
This Year Complete control range
This Month
Page 54
54 D39303 | Date 09/2018
Menu mode
For a description of how to import and process protocol data in AtmoCONTROL or read it out via Ethernet, please refer to the separate AtmoCONTROL manual.
7.9 USER ID
7.9.1 Description
With the USER ID function, you can lock the settings of individual (e.g. temperature) or all parameters, so that they cannot be changed at the appliance by accident or unauthorised persons. You can also lock setting options in menu mode (e.g. adjustment or date and time settings) this way.
If adjustment options are locked, this is indicated by the lock symbol in the respective display (Fig.
26).
USER ID data is entered in the AtmoCONTROL software and saved on the USB storage medium. The USB stor­age medium is thus acting as a key: Parameters can only be locked or unlocked if it is connected.
A description of how to create a USER ID in AtmoCONTROL is provided in the separate AtmoCONTROL manual.
7.9.2 USER ID activation and deactivation
1. Connect the USB storage medium with the USER ID data to the USB port on the right of the ControlCOCKPIT.
2. Activate the USER ID. To do so, press the activation key on the right side of the
USER ID display. The display is enlarged
and the entry
Activate automatically
highlighted.
USER-ID
Deactivate
Activate
3. Confirm the activation by pressing the confirmation key. The new USER ID data are transferred from the USB storage medium and activated. As soon as activa­tion is complete, a check mark appears in front of the corresponding entry.
USER-ID
Deactivate
Activate
4. Remove the USB storage medium. Locked parameters are indicated by the lock symbol on the respective display (Fig. 26 ).
T
o unlock the appliance, connect the USB storage medium, activate the
USER ID display and
select the entry
Deactivate.
22.4
TEMP
°C
°
Fig. 26 Temperature adjustment at appliance locked (example)
Page 55
D39303 | Date 09/2018 55
Maintenance and Servicing
8. Maintenance and Servicing
WARNING
Danger due to electric shock. Disconnect the mains plug before any cleaning or maintenance work.
8.1 Regular maintenance
Annually:
► Check the sterile filter in the control unit and the water pump filter in the rear panel and
replace them if they are dirty.
► To guarantee perfect control, we recommend calibrating the appliance once a year (see
page 47).
Every two years:
► Replace the sterile filter in the control unit and the water pump filter.
8.2 Cleaning
8.2.1 Interior and metal surfaces
Regular cleaning of the easy-to-clean interior prevents build up of material remains that could impair the appearance and functionality of the stainless steel chamber over time.
The metal surfaces of the appliance can be cleaned with normal stainless steel cleaning agents. Make sure that no rusty objects come into contact with the interior or with the stain­less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots should appear on the surface of the interior due to impurities, the affected area must be im­mediately cleaned and polished.
8.2.2 Plastic parts
Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic or solvent-based cleaning agents.
8.2.3 Glass surfaces
Glass surfaces can be cleaned with a commercially available glass cleaner.
8.3 Repairs and Service
Only authorised customer service points should carry out maintenance work.
Page 56
56 D39303 | Date 09/2018
Storage and disposal
9. Storage and disposal
9.1 Storage
The appliance may only be stored under the following conditions:
► in a dry and enclosed, dust-free room ► frost-free ► disconnected from the power supply and water supply
Disconnect the tube of the water supply tank and empty it.
9.2 Disposal
This product is subject to the Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) of the European Parliament and of the Council. This appliance was placed on the market after August 13
th
, 2005 in countries which have already integrated this Directive into their national laws. It may not be disposed of in normal household waste. For disposal, please contact your dealer or the manufacturer. Any ap­pliances that are infected, infectious or contaminated with materials hazardous to health are excluded from return. Please also observe all other regulations applicable in this context.
Before disposing of the appliance, please render the door locking mechanism unusable, for example to prevent playing children from be­ing locked inside the appliance.
There is a lithium battery in the ControlCOCKPIT of the appliance. Remove it and dispose of it in accordance with the regulations in your country (Fig. 27).
Fig. 27 Removing the lithium battery
Page 57
D39303 | Date 09/2018 57
Index
Index
A
Accessories 16 Activation button 25 Alarm 30, 31, 33, 37 Ambient conditions 16 Ambient temperature 16 Appliance error 38 ASF 30, 31 AtmoCONTROL 3, 13, 16, 26,
28, 51, 53, 54
B
Basic appliance settings 41 Basic settings 41
C
Calibration 47 Cancel programme 29 Carrying 17 Chamber load 24 Changes 9 Cleaning 55 Communication interfaces
12
Compensation correction
values 49 Connections 12 ControlCOCKPIT 24, 25 Customer service 2
D
Date 45 Declaration of conformity 15 Delete programme 52 Delivery 17, 18, 21 Digital backwards counter
with target time setting
27 Dimensions 15 Directives 15 Disposal 56 Door 23
E
Electrical connection 12 Electrical equipment 12 Electronic temperature moni-
toring 30 Emergency 9 Ending operation 36
End of programme 29 Error message 39 Error messages 37 Ethernet 13
F
Forklift truck 18
G
Gateway 45 Graph 35
H
Hazards 7 Hot-air generator 10 Humidity 27 Humidity adjustment 49 Humidity monitoring 33,
34, 38
Humidity profile 35
I
Installation site 19 Intended use 8 Interfaces 12 IP address 43
K
Keysound 52
L
Language selection 42 Loading the appliance 24 Log memory 40, 53
M
Maintenance 55 Malfunctions 9, 37, 38 Manufacturer 2 Material 12 Mechanical temperature
monitoring 32 Menu 41 Menu mode 41 Minimum clearances 19 Monitoring function 30 Monitoring temperature 30
N
Network 13, 43
O
Operating modes 26 Operating personnel 8 Operating problems 38 Operation 23
P
Packaging material 18 Parameter adjustment 25,
42 Power failure 40 Power supply 21, 53 Product safety 7 Programme 51 Programme mode 26, 28 Protocol 53 PT100 temperature sensor
30 Putting into operation 21
R
Regular maintenance 55 Remote control 45
S
Safety regulations 6, 10 Service 55 Servicing 55 Setting up 17, 19 Setup 43 Sound 47, 52 SOUND 52 Speaker symbol 30, 33, 37 Standards 15 Starting a programme 28 Sterilisation programmes 52 Storage after delivery 18 Switching off 36 Switching on 22
T
Taking out of service 56 TB 32 Technical data 14 Temperature 27 Temperature calibration 47 Temperature comparison 47 Temperature deviation 47 Temperature limiter 32 Temperature monitor 30, 31
Page 58
58 D39303 | Date 09/2018
Index
Temperature monitoring 30 Temperature sensor 30 Tilt protection 20 Time 45 Timer 27 Timer mode 44 Transport 17, 18 Transport damage 18 Turn control 25 TWW 30 TWW temperature monitor-
ing 31
Type plate 13
U
Unit 44 Unpacking 18 USB port 12, 53 USER ID 54
W
Warning messages 12, 37 Water specifications 21 Water supply 21 Weight 14 What to do in case of ac-
cidents 9
Page 59
Page 60
Memmert GmbH + Co. KG Willi-Memmert-Straße 90-96 | D-91186 Büchenbach Tel. +49 9122 925-0 | Fax +49 9122 14585 E-Mail: sales@memmert.com facebook.com/memmert.family Die Experten-Plattform: www.atmosafe.net
Humidity chamber HCP
D39303| Date 09/2018
English