
AtmoCONTROL
SOFTWARE MANUAL
100% ATMOSAFE. MADE IN GERMANY.
www.memmert.com | www.atmosafe.net

Willi-Memmert-Straße 90–96
We reserve the right to make changes

About this manual
This user manual describes the installation and use of the MEMMERT programming software
AtmoCONTROL. It is intended for use by trained personnel of the operator, who have the task
to the appliance. Incorrect use could result in damage to the appliance and/or to the chamber
with AtmoCONTROL and familiarise yourself with it.
This manual should always be kept in a place where those working with the software have
will work with the software are informed as to the whereabouts of this user manual. We rec-
The current version of AtmoCONTROL and this manual are available for download at
www.memmert.com/de/service/downloads/software/
www.memmert.com/de/service/downloads/software/

...........................................................................................................................
...........................................................................................................
....................................................................................................
......................................................................................................
...........................................................................................................
...........................................................................................................................
...............................................................................................................................
........................................................................................................................
........................................................................
.............................................................................................................................
.....................................................................................................
4.1.2 Creating a programme
4.1.4 Available parameters
4.2 Simulating the programme sequence (preview)
4.3 Saving, loading, transferring and running the programme
4.3.1 Saving the programme
4.3.2 Loading a saved programme
4.3.3 Transferring programme via Ethernet
4.3.4 Transferring a programme via USB storage medium
4.3.5 Selecting and starting a programme on the appliance
4.4.1 Programme example with clock timer
4.4.2 Programme example with door locking
4.4.3 Programme example sterilisation
4.4.4 Programme example loop

...........................................................................................................................
..............................................................................................................................
......................................................................................................................
...................................................................................................................................
..................................................................................................................

AtmoCONTROL is a software for programming and logging MEMMERT appliances of the
With AtmoCONTROL, you can
Appli-
ance
Tempe-
rature
Humi-
dity
Programmable main parameter
Pressure CO
2
O
2
Fan speed Air fl ap Light*

Category Minimum system requirements
VGA graphics and colour monitor
You must have administrator rights to be able to install AtmoCONTROL.
Working with AtmoCONTROL
AtmoCONTROL can be started in two ways:

1
Menu bar (see section 3.2.1)
Status bar (provides an overview of available appliances, see page 10)
Status bar (provides an overview of available appliances, see page 10)
Status bar (provides an overview of available appliances, see page 10)
Editor, simulation and protocol window (only for appliances listed on page 6,
otherwise only protocol window)
Editor, simulation and protocol window (only for appliances listed on page 6,
Editor, simulation and protocol window (only for appliances listed on page 6,
Programming mode switch (for editor/simulation/protocol, see pages 22 and 28)
otherwise only protocol window)
otherwise only protocol window)
2
3
6
5

Device Program Protocol Print Options Help
protocol data on USB
protocol data on USB
vice (see page 12)
vice (see page 12)
Save programme
Save programme
Export protocol data (see
Automatic sending of
Automatic sending of
emails (see page 30)
emails (see page 30)
Install device licence (see
Toolbar
The toolbar provides rapid access to the most important menu functions:
Load programme
Load programme
Connect online via Ethernet
1
Connect offline from USB device
2
Connect offline from database
3
Disconnect device
4
Disconnect all devices
5
New
6
Load
7
Save
8
Save As...
9
Upload to Device
10
Export to USB drive
11
12
13
Import...
Export...
15
Print
14
Language
USER-ID
16
Email options
17
Edit Backup Folder
18
19
20
21
1 2 3 4 5 6
About...
User Manual
Upload license file to device

The status bar gives an overview of the appliances logged on to AtmoCONTROL. Appliances
1 2
An appliance of type UF 260
(1), named "Laboratory" (2) by the user, is registered via Ether-
net (3) in AtmoCONTROL; current temperature 180.0 °C (4)
A description of how to set the IP address is
tered) in AtmoCONTROL as described in the following section.
Adding and disconnecting devices
Adding device connected via Ethernet
vices have at delivery (192.168.100.100). The IP address
UF 260plus Laboratory
180.0°C
3 4
i

A description of how to change the IP address of a device is provided in the user
Connecting device using USB storage
A description of how to read protocol data on a device is provided in the user
Connecting a device using
A window opens, in which you can open the
As soon as the appliance has
tion displayed at any time.
for your appliance later on,

file Log.txt is also transferred and saved in a subfolder of the directory
The log file is structured as shown in the
Alarm / event description
A detailed list of all event codes is given
Disconnecting devices
the status bar, select it and then click on
A B C D

temperature, pressure and humidity), which the appliance then implements from a definable
To be able to create a programme in AtmoCONTROL, the appliance which is to perform the
Two editor threads are always shown for appliances with humidity or pressure control,
want to see the parameter values for a specific point in time, you must change to the simula-
Elements to create a programme
Elements to create a programme
Elements to create a programme
Available parameters (functions)
Appliance selected
Appliance selected
Available parameters (functions)
Available parameters (functions)
Additional editor thread for appliances with humidity or pressure control
2
1
3
4

To create a programme, drag the individual parameter icons onto the editor thread one after
the zoom icons in the toolbar list (see section 3.2.2 on page 9) or with the mouse wheel,
you can zoom in or out of the display or have the entire programme displayed.
Ramp 05
Drag the parameter icon (in this case a tem-
perature change) onto the editor thread while
holding the mouse button down
Drag other parameters – light and a humidity
change in this case – onto the editor thread.
A red insertion mark helps you to find the
The temperature icons (change/hold temperature) may only be placed on the upper editor
thread, humidity and pressure icons only on the lower one.
The meaning of the individual icons and the adjustment options are described from page
Removing a parameter icon from the editor thread
Removing a parameter icon from the editor thread
with the mouse button depressed, to the recycle bin symbol.
10h:00m
11h:30m
Ramp 05
10h:00 m
11h:30 m
LIGHT
21.0 °C 180.0°C
180.0°C
21.0 °C 180.0°C
Increase
40.7%rh 70.6%rh
180.0°C
2h:30 m
0%

To adjust values, click successively on the corresponding fields – in
the example on the right, the setpoint temperature. The value is
The adjustment range depends on the appliance for which the
The main parameters have additional adjustment options, which
fold down icon (Fig. 6, No. 1).
Further adjustment options fold down after clicking on the arrow icon on the right
Further adjustment options fold down after clicking on the arrow icon on the right
Further adjustment options fold down after clicking on the arrow icon on the right
- - h:- - m
- - - .- °C21.0 °C
Ramp 01
21.0 °C
2
Tolerance band
1
0,1 K/min
11h:30m
37.0°C
Tolerance band
0,1 K/min
3

Which parameters are available to adjust the programme depends on the appliance for
which a programme is to be created. Only those parameters are available that the appli-
Depiction
Temperature to be maintained
Tolerance value above/below
Alarm if limits are exceeded
Target (setpoint) temperature
Tolerance value above/below
When run, this is displayed in the
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
Tolerance band
Tolerance band
0,1 K/min

Depiction
Tolerance value above/below
Alarm if limits are exceeded
Target (setpoint) humidity
Tolerance value above/below
When run, this is displayed in the
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
Hold
12h:00 m
70.0%rh
2h:30 m
40.7%rh 70.6 %rh
Tolerance band
+
0.0%rh
-
5.0%rh
on off
Alarm
Safe
Tolerance bandIncrease
+
5.0%rh
-
5.0%rh
on off
SPWT

Depiction
Tolerance value above/below
Alarm if limits are exceeded
Target (setpoint) pressure
Tolerance value above/below
When run, this is displayed in the
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
in icon bar
Meaning Depiction on
editor thread
Function and
adjustment options
hold p
decrease
1100mb
2h:30 m
500mb
2h:30 m
500mb
Tolerance band
+
20mb
-
10mb
on off
Alarm
Safe
Tolerance band
+
20mb
-
10mb
on off
SPWT

tive position – and remains active until it is changed by the insertion of a new parameter icon
Depiction
Appliance emits an acoustic signal at the
was inserted, for example if a specific setpoint
value is reached or the programme is finished.
in icon bar
Meaning Depiction on
editor thread
CO
2
50%
15
0
O2
CO2
12.0%
FAN
100%
FLAP
100%
LIGHT
Adjustment options/
comments
100%
UV LIGHT
on off
HORN

Depiction
Adjustment options: open/close
Activates the defrosting function of the appli-
The programme is repeated each week at the
finished on both editor threads.
in icon bar
15
Meaning Depiction on
editor thread
DOOR
openclose
SWITCH
A
openclose
DEFROST
CLOCK TIMER
Mo
Tu
We
Th
Fr
Sa
Su
11h:30 m
CALENDAR
Day
Month
Year
2012
- - - -h: m
22h:30 m
Adjustment options/
comments
18
05
SYNC
and or

Depiction
The programme jumps back from the insertion
the range that is to be repeated.
in icon bar
Meaning Depiction on
editor thread
Adjustment options/
comments
STAND BY
JUMP
TARGET
STANDBY
LOOP
JUMP
4 X
TARGET
JUMP
TARGET
LOOP
LOOP
2 X
4 X

While creating the programme, you can display the prospective progression of all parameters
Programme preview diagram (simulation)
you want to alter the programme.
Saving, loading, transferring and running the programme
The name with which you save the programme is later

Transferring programme via Ethernet
To be able to transfer a programme via Ethernet, the appliance and computer must be
Transferring a programme via USB storage medium
With appliances that have humidity control, make sure that the
vals, especially for programmes that run for long periods. The same applies for appliances

5 841
maintain this temperature (infinitely
) (4) until it is changed: also Monday to Friday (5) at 6
pm (6) to 50 °C (7) – again continued (infinitely
maintain this temperature (infinitely
maintain this temperature (infinitely
) (4) until it is changed: also Monday to Friday (5) at 6
) (4) until it is changed: also Monday to Friday (5) at 6
) (8) until it is changed again in the morning
120 °C
50 °C
20 °C
2 3
6 7
347
8
Mo. Tu. We. Th. Fr. Sa. Su.
1

3 4
lowered (5) for 30 minutes (4) to 45.0
on" (8) ensures that the door is opened only when the temperature really has dropped to
1 2 5
Tolerance band
1,7 K/min
8 6
7
95 °C
45 °C
20 °C
1
6
3
2
5
12:00 h ≥ 0:30 h

4
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
the appliance heats up to 180.0
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
At the beginning, the fan is switched on to 100 % (1) and the air flap is closed (0 %) (2). Then,
ting „Safe“ (5) ensures that the sterilisation time does not start (6) before the set tolerance band
the appliance heats up to 180.0
the appliance heats up to 180.0
6
Tolerance band
1 2 3
5
1
0 % 100 % 0 %
2
100 % 0 %
6
180 °C
4
3
20 °C
3:00 h
7

41
First, the appliance heats up to 250.0 °C (2) for one hour (1). Then, the fan begins to run at 80
fan is switched off (6). This sequence is repeated from the jump target (7) ten times (8).
250 °C
20 °C
27 3 5
6
0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0% 80% 0%
235
1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h 1 h
1 4
8
6 8

To be able to import a protocol via network, the appliance and computer must be con-
Protocol display (example)
Importing protocol from USB data medium
At the appliance, protocols can be exported to an USB storage medium and imported in
the user manual for the appliance.
files on an USB storage medium (see page 11).

xport”, you can export a freely
types *.csv or *.xlsx (Excel), which can be processed in
tions and protocols – depending on what is currently
You can set the language of the user
With the appliances listed in the table on page 6, it
AtmoCONTROL cannot generate a USER-ID file,
figuration in AtmoCONTROL is also not possible.
There can be only one USER-ID on a USER-ID USB stick. The settings in this file then apply for
A USER-ID identifier on a USER-ID USB stick for one (or several) serial numbers can be pur-
User-ID

with the USER-ID file into the
A window appears with the functions of the logged on appliance
that can be blocked (depending on the appliance type).
AtmoCONTROL can automatically send an email to a freely definable recipient if an alarm
was triggered, e.g. if the temperature is exceeded. To make the corresponding settings, select
Settings for the automatic sending of emails in case of alarm

Backup folder
You can set up a backup folder in which
AtmoCONTROL saves backup copies of
You can either use the default folder or

Event codes of the log fi le Log.txt
Error codes for Generaon 2012 appliances Status: October 09, 2012
Graphical user interface error
Error in calibraon file (user)
Error in calibraon file (calibraon system)
Warning at 95% disk space usage
210: Failed To Copy Protocol
Error copying the protocol
Error restoring LastState
212: Max Count Of Profiles
Maximum number of profiles on SD card reached
Temperature limiter has triggered
Triac optocoupler heang 1 power module 1
defecve
Triac optocoupler heang 2 power module 1
defecve
Triac optocoupler heang 1 power module 2
defecve
Triac optocoupler heang 2 power module 2
defecve
Triac optocoupler heang 1 power module 3
defecve
Triac optocoupler heang 2 power module 3
defecve
Triac heating 1 power module 1 defecve
Triac heang 2 power module 1 defecve
Triac heang 1 power module 2 defecve
Triac heang 2 power module 2 defecve
In case of a hardware error of the oven, the controller displays the following error codes in a status /
error window, as well as logs them in the “Log.txt” file in the “Config” directory on the SD card.
The posion of the error number further specifies the posion of the error.
The posion of the red digit
indicates the defecve stage
of the appliance.

Triac heang 1 power module 3 defecve
Triac heang 2 power module 3 defecve
Power module 1 is not responding
Power module 2 is not responding
Power module 3 is not responding
Humidity power module is not responding
Humidity below minimum value
Humidity maximum value exceeded
Temperature sensor defecve
Monitoring sensor defecve
Temperature below minimum value
Temperature maximum value exceeded
Temperature tolerance band violated
Automac lights switch-off
CO2 supply interrupted / gas cylinder empty
Noficaon of gas cylinder change
N2 supply interrupted / gas cylinder empty
801: Program Start
Pressure below alarm limit
Pressure alarm limit exceeded
Voltage below minimum limit
Programme cannot be loaded

AtmoCONTROL
D24042 | Date 09/2014
englisch
Memmert GmbH + Co. KG
Willi-Memmert-Straße 90-96 | D-91186 Büchenbach
Tel. +49 9122 925-0 | Fax +49 9122 14585
E-Mail: sales@memmert.com
facebook.com/memmert.family
Die Experten-Plattform: www.atmosafe.net