Safety
The “MEphisto Scope 1” has a maximum input voltage of ±10 VDC.
Measuring higher voltages than this directly may damage the instrument and invalidate the warranty. The instrument is not galvanically
isolated - i.e. the ground of the internal circuitry connects directly to the
ground of the host pc via the USB supply. To measure high voltages such as mains
supplies use a high voltage differential probe for your personal safety and to prevent
damage to the instrument. Please follow the appropriate standards, installation
instructions and national safty standards. Meilhaus Electronic GmbH assumes no
responsibility for damage in case of faulty installation, operation or handling.
What‘s in the Box?
Please make sure, that your “MEphisto Scope 1” box includes the following items:
• 1 x USB scope “MEphisto Scope 1” (UM202) hardware.
• CD with driver and DAQ software “MEphistoLab”.
• USB cable (type A - type B)
• Getting Started Guide.
Accessories (optional)
“ME-Probe LF Pair”
2 passive modular probes (1:1/1:10
selectable) with accessories. Bandwidth:
15/150 MHz, 47/15.5 pF, 1/10 MΩ,
600 Vp CAT I.
“ME-UB” series
Connect up to 3 “ME-UB” boxes to the
26pin D-sub with a tripple break-out
cable “ME AK-DUB/UM202”.
2 Channel USB Scope “MEphisto Scope 1” (UM202)
Getting Started Guide
ME-UB15
15-pin terminal
ME-UBOI
8 opto-isolated
inputs (2.5…60V)
ME-UBOO
8 opto-isolated
outputs (up to 60V)
ME-UBRE
8 relays, changeovers (5 A/240 VAC)
Support
Should you experience problems with the device or software, please send an E-Mail
to support@meilhaus.com.
Newest infos and software updates on our web site: www.meilhaus.com.
Meilhaus Electronic GmbH
Fischerstrasse 2
82178 Puchheim/Germany
E-Mail sales@meilhaus.com
Web www.meilhaus.com
+49 - 89 - 89 01 66-0
FAX +49 - 89 - 89 01 66-77
Release: 06.05.2005
All trademarks acknowledged.
Information contained in this
document is subject to change
without notice.
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
DIO_17
DIO_18
DIO_19
DIO_20
DIO_21
DIO_22
DIO_23
DIO_24
(EXT_TRIG)
DIO_9
DIO_10
DIO_11
DIO_12
DIO_13
DIO_14
DIO_15
DIO_16
PC_GND
1
2
3
4
5
6
7
8
9
DIO_1
DIO_2
DIO_3
DIO_4
DIO_5
DIO_6
DIO_7
DIO_8
PC_GND
“MEphisto Scope 1” - Features
“MEphisto Scope 1” combines the functionality of an oscilloscope, a digital volt
meter (DVM), as well as 24 discrete digital I/Os in a single compact device. The
module is powered from the PC via the USB interface; no external power supply is
necessary.
Both analog inputs have BNC connectors for standard oscilloscope probes. Each
channel has it’s individual 16 Bit A/D converter which gives you simultaneous
sampling rates of 1 MS/s on both channels. The input range is ±10 V. Multiple
trigger options can be selected to start an analog acquisition. The 24 discrete
digital I/O channels (TTL) can be congured as input or output individually.
The “MEphistoLab” software which is included with the device, is a virtual
instrument panel, that gives you easy and quick access to all functions of the device
under Windows 98SE, Me, 2000 and XP. The software requires the installation of
the Agilent VEE runtime software version 7.0 (included on CD-ROM).
Hardware Specifications
Analog Inputs
Number, resolution 2 x 16 bit
Total sampling rate 2 MS/s (2 channels)
Simultaneous channels 2
Input voltage range ±100 mV … ±10 V
Analog bandwidth (-3 dB) 5 MHz (no anti aliasing lter)
Sample memory 256 kS
Trigger types Level, window, edge, slope (dU/dt), external, manual
Digital Inputs/Outputs
Number 24 I/Os
Conguration each bit programmable as input or output
Signal level +5 V (CMOS)
General Data
PC inferface USB 2.0 full-speed (USB 1.1 compatible)
Power supply +5 V/0.85 W via USB interface
Reference ground PC ground (PC_GND)
Connectors Analog channels: 2 x BNC connectors (front panel)
Digital channels: 26-pin D-sub female (rear panel),
USB connector: Type B (rear panel)
Size (W x H x D) 108 mm x 35 mm x 136 mm (incl. connectors)
Operating temperature 0…70 °C
Getting Started
1. Connect the “MEphisto Scope 1” to a USB port of your PC using the USB cable
included with the device. ⇒ The hardware wizard will automatically detect the
new USB device and will display the message “New hardware found”. The following description is to be orientated towards Windows XP but is also valid for
2. Insert the installation CD-ROM included with the package into the CD drive of
older Windows versions in an analog way. However the dialogs may differ slightly
depending on your Windows version.
your computer.
3. In the hardware wizard select the option “Search for a suitable driver for my
device (recommended)” and click “Next”. If Windows displays a message, telling
you the driver is not Microsoft certied, just skip this window.
⇒ The system driver will be installed and the entry “MEphisto Scope 1” will be
added in the device manager under the device class “Meilhaus DAQ Boards”.
4. End the installation of the system driver by „Finish“ and continue with the installation of the “MEphistoLab” software as follows:
5. With the CD still inserted in your drive, navigate to the directory [CD drive]:
\software\MEphistoLab\disk1 and start the installation of the “MEphistoLab”
software by double-clicking the le “setup.exe”.
! Please read the dialogs of the installation program carefully. If required, the
Agilent VEE runtime software will be installed automatically.
⇒ After the installation has nished you will nd the “MEphistoLab” software in
your Windows “Start” menu under “Programs - Meilhaus MEphistoLab”.
Digital I/O
• 24 digital I/Os
• Each bit can be congured
as input or output
• Signal level: +5 V (CMOS)
Pinout of the 26-pin D-sub
female connector:
Volt Meter
• 2 channel voltmeter with
pointer meter display
• Voltage range:
±100 mV, ±1 V, ±10 V
• Hold function
Oscilloscope
• 2 channel USB scope
• Time base: 1 µs…2 s
• Voltage: 200 mV…20 V
• Continuous or “single shot”
measurement
• X and Y cursor display
• Time/spectrum graph
• Trigger types: Level, edge,
window, slope (dU/dt),
external (pin 26), manual
• „Print Screen“ functiom
• Export to Microsoft Excel™
compatible CSV format
• Conguration can be stored