Meilhaus Electronic Logian-16 User guide

Logian-16
User Manual
Meilhaus Electronic GmbH
Logian-16 - Logic Analyzer
User Guide
WWW.MEILHAUS.COM
Meilhaus Electronic GmbH Fischerstrasse 2 DE-82178 Puchheim Germany Tel.: +49 (0) 89 / 89 01 66 – 0 Fax.: +49 (0) 89 / 89 01 66 – 77
sales@meilhaus.com support@meilhaus.com
Logian-16 Users Manual Page 1
Logian-16
User Manual
Index
Preface ......................................................................................................... 3
1 Features of the Logic Analyzer .............................................................. 4
1.1 Package Contents........................................................................................4
1.2 Introduction..................................................................................................6
1.3 Hardware Specifications..............................................................................10
1.4 System Requirements................................................................................. 11
1.4.1 Operating System Requirements......................................................11
1.4.2 Hardware System Requirements....................................................... 11
1.5 Device Maintenance and Safety..................................................................12
2 Installation ............................................................................................14
2.1 Software Installation...................................................................................14
2.2 Hardware Installation..................................................................................16
2.3 Tips and Advice..........................................................................................18
2.4 Flow of software operation..........................................................................20
3 Introduction to Logic Analysis............................................................. 21
3.1 Logic Analysis ............................................................................................21
Task 1. Clock Source (Frequency) and RAM Size set up....................21
Task 2. Trigger Properties Setup.......................................................23
Task 3. Signal/Bus Trigger Edge Setup .............................................28
Task 4. Run to Acquire Data..............................................................28
3.2 Bus Logic Analysis......................................................................................30
Logian-16 Users Manual Page 2
Logian-16
User Manual
Preface
This Quick Start Guide is designed to help new and intermediate users navigate and perform common tasks with the Logic Analyzer. Despite its simple packaging and interface, the Logic Analyzer is a sophisticated measurement and analysis tool. It is also a highly sensitive electrical current sensing device. Users must carefully read instructions and procedures pertaining to installation and operation. Any instrument connected to the unit should be properly grounded. A pair of anti-static gloves is strongly recommended when performing a task with the device. To ensure accuracy and consistency of output data, use of the bundled components is strongly recommended.
User opinions are very important to us. Please contact our engineering team by telephone, fax or email with your questions or feedback. Thank you for choosing the Logic Analyzer.
Logian-16 Users Manual Page 3
Logian-16
PinTesting Cable
Pin Testing Cable
User Manual
1 Features of the Logic Analyzer
In this chapter, users will learn about the package contents, description, hardware specifications, system requirements, and safety issues of the Logic Analyzer. Though this chapter is purely informative, we highly recommend reading this carefully to ensure safety and accuracy when performing any operation with the Logic Analyzer.
1.1 Package Contents
Verify the package contents before discarding packing materials. The following components should be included with your product. For assistance, please contact our nearest distributor.
Table 1-1: Parts list for retail packages
Models ODM-16128U
Logic Analyzer 1
16-Pin Testing
Cable
8-Pin Testing Cable 2
Probe 20
USB Cable 1
Getting Started
Guide
Driver CD** 1
1-
(White)
2-
(Black)
* This Driver CD consists of a multilingual software interface program, as well as a
multilingual Users Manual.
0
1
1
1
Logian-16 Users Manual Page 4
Logian-16
User Manual
Fig. 1-1: Logic Analyzer
Fig. 1-3: Probes
(varied depending on models)
Fig. 1-5: Getting Started Guide
16-Pin x 1
Fig. 1-2: Testing Cables
8-Pin x 2
Fig. 1-4: USB Cable
Fig. 1-6: Driver CD
Fig. 1-7: 1-Pin External Clock Wire
(White)
Logian-16 Users Manual Page 5
Fig. 1-8: 2-Pin Ground Wire
(Black)
Logian-16
User Manual
1.2 Introduction
Logic Analyzer models ODM-16128U, all share the same external features as illustrated in the following figures.
Power
Trigger
Read
Run
Start
Signal Connector
Fig. 1-9: A view of the Logic Analyzer ODM-A Series. see Fig 1-12 for detailed information
on the Signal Connectors.
Fig. 1-10: Side view of the Logic Analyzer, which draws its power
from the USB connection.
Logian-16 Users Manual Page 6
Logian-16
User Manual
Fig. 1-11: Rear view of the Logic Analyzer ODM-A Series.
Logian-16 Users Manual Page 7
User Manual
Table 1-2: List of functional pins in each model.
Models ODM-16128U
Port A ( A0~A7) X
Port B ( B0~B7) X Port C ( C0~B7) Port D ( D0~B7)
R_O X T_O X S_O X
CLK X GND X VDD X
IOA X IOB X IOC X
GND X
Table 1-3: Definitions and Functions of pins for all models.
CLK Clock Connects a given external module to be analyzed.
GND Ground
Table 1-4: Definitions and Functions of pins for advanced models (1).
R_O Read (Out)
T_O Trigger (Out)
S_O Start (Out)
Two pins used for grounding the Logic Analyzer with a given external module to be analyzed.
When the Logic Analyzer is about to upload data from memory to the PC, the R_O will send a Rising Edge signal of DC3.3V. When the upload is finished, a Falling Edge signal is sent. When a trigger condition is established, the T_O will send a Rising Edge signal of DC3.3V. When memory is full, a Falling Edge signal is sent. When a user initiates a sampling task by clicking the RUN icon in the window or clicking the START button on the device, the R_O will send a Rising Edge signal of DC3.3V. When the Logic Analyzer finishes uploading, a falling edge signal is sent.
Logian-16
Logian-16 Users Manual Page 8
Logian-16
User Manual
Logian-16 Users Manual Page 9
Table 1-5: Definitions and Functions of pins for advance models (2).
VDD
IOA Ext. I/O Module A IOB Ext. I/O Module B Same as IOA.
IOC Ext. I/O Module C Same as IOA.
GND Ground Grounds external devices, in sequence
Voltage Drain
(Semiconductor)
Provides +3.3 V for external modules by draining voltage from the Logic Analyzer. Transmits signals from an external model or device not being tested.
1.3 Hardware Specifications
Table 1-6: Hardware specifications of ODM-A Series.
Items\Type ODM-16128U Interface USB 2.0 (1.1) Operating System Win 95/NT/98SE/ME/2000/XP/2003 Power Supply USB 1.1 (USB 2.0 Recommended) Channels 16 Bandwidth 75MHz Memory 4M Bits
Memory Depth (Per
Channel)
Internal Clock Rate
(asynchronous)
Max External Clock
(synchronous)
Trigger Channels 16 Channels Trigger Condition Edge/Pattern
Pre-Trigger/ Post-
Trigger
Trigger Level 1 Level Trigger Count 1-65535 Max Trigger Page Max 8192 Enable Channel 16
Buses Data Decoded
Enable Delay
Compression
Logian-16 Users Manual Page 10
128kBit
100 Hz ~ 200 MHz
Max 100MHz
Yes
Yes
Start: Edge and Pattern
End: 1-65535
16 Channel Compression
Logian-16
User Manual
Logian-16
User Manual
1.4 System Requirements
This section discusses basic operating system and hardware requirements for the Logic Analyzer. Software and hardware capabilities may vary depending on PC configuration,. This manual assumes proper installation of a supported operating system as listed below.
1.4.1 Operating System Requirements
In this sub-section, we share our experiences testing the Logic Analyzer on the following Microsoft Windows operating systems. Since the Logic Analyzer requires operating system support of the USB protocol, Windows 95r2 and earlier OS versions are incompatible.
1) Windows NT 4.0 Workstation & Server, Service Pack 6 – not recommended
2) Windows 98, 98 Second Edition - supported
3) Windows ME – not recommended
4) Windows 2000 Professional, Server Family - supported
5) Windows XP Home, Professional Editions (32-Bit versions) - supported
6) Windows Server 2003 Standard Edition, Enterprise Edition, Small Business Edition. (32-Bit Versions) – not supported
1.4.2 Hardware System Requirements
CPU
Windows NT, 98, 98 SE
166 MHz, or above.
Windows 2000, XP
300 MHz, or above (strongly suggest 900 MHz, or above).
We have tested various 32-Bit and 64-Bit CPUs. Overall, we find that all 32-Bit CPUs work very well with Logic Analyzer software. Moreover, we find that AMD’s 64-Bit CPUs, except Opteron, with a 64-Bit Windows operating system, work just fine with Logic Analyzer; no significant problems occur.
Memory
Windows NT, 98, 98 SE
128 MB or above (64 MB minimum). Windows 2000, XP
256 MB or above (128 MB minimum).
Hard Drive
At least 100 Mb available space.
USB
USB 1.1 compatible (recommend USB 2.0).
Display Devices (recommended)
1. 17 monitor with 1024x786 resolution or higher.
Logian-16 Users Manual Page 11
User Manual
Do not use harsh chemicals or cleaners containing substances
2. 8MB SDRAM on Video Card.
1.5 Device Maintenance and Safety
Follow these instructions for proper operation and storage of the Logic Analyzer.
Table 1-7: General Advice
Cautions Do not place heavy objects on the Logic Analyzer.
Avoid hard impacts and rough handling.  Protect the Logic Analyzer from static discharge.
Cleaning Use a soft, damp cloth with a mild detergent to clean.
Working Voltage DC 4.5 V DC 5.0 V DC 5.5 V Current at Rest 200 mA Current at Work 400 mA Power at Rest 1 W Power at Work 2W Error in Phase Off* ± 1.5 nS
V
of Testing Channels ± DC 30 V
input
V
Reference
Input Resistance 500K/10pF Working Temperature 5°C 70°C Storage Temperature -40°C 80°C
Table 1-9* refer to the User Manual for error analysis calculation.
Do not disassemble the Logic Analyzer; this will void the
warranty and could affect its operation.
Do not immerse or spray any liquid on the Logic Analyzer.. 
such as benzene, toluene, xylene or acetone.
Table 1-8: Electrical Specifications
Items Minimum Typical Maximum
DC -6 DC +6 V
Logian-16
Logian-16 Users Manual Page 12
This is a Class A product which may cause radio interference in a
occurs. Occasionally, however, temporary conductivity caused by
WARNING
Logian-16
User Manual
: Operating Environment
Avoid direct sunlight  Use in a dust free, non-conductive environment (see Note )  Relative Humidity: < 80%  Altitude: < 2000m  Temperature: 0 ~ 40 degrees C
domestic environment. Note: EN 61010-1:2001 specify degrees of pollution and their
requirements. Logic Analyzer falls under Level 2. Pollution refers to addition of foreign matter, solid, liquid or
gaseous (ionized gases), that may produce a reduction of dielectric strength or surface resistively’.
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs. This pollution has no effect.
Pollution Degree 2: Normally only non-conductive pollution condensation must be expected. Pollution Degree 3: Conductive pollution occurs or dry,
non-conductive pollution occurs which becomes conductive due to condensation. In such conditions, equipment is normally protected against exposure to direct sunlight, precipitation and wind, but neither temperature nor humidity is controlled.
Storage
Environment
Relative Humidity: < 80% Temperature: 0 ~ 50 Degrees C
Conclusion
After reading this section, users should have a basic grasp of the Logic Analyzer. A
complete understanding of the Safety and Care Recommendations section is a critical prerequisite of any further operation as presented in the User Manual.
Logian-16 Users Manual Page 13
Logian-16
User Manual
2 Installation
This chapter describes installation of the Logic Analyzer hardware and software.
Software installation steps must be followed precisely to ensure successful installation.
2.1 Software Installation
In this section, users will learn how to install the software interface and drivers. As with proper installation of many USB devices, the Logic Analyzer application and driver software must be installed prior to connection of the hardware. The following steps illustrate an installation of the Logic Analyzer. The other two models mentioned in Chapter 1 would follow identical procedures.
Step 1. Insert the driver CD-ROM in the PC CD drive. Step 2. Execute the installation program. Go to the START menu, click START,
click Run, click Browse, select Setup.exe file in the appropriate model folder and then click OK. It is recommended that all other programs are closed while installation proceeds.
Step 3. Choose the desired language. Step 4. Click Next to proceed with the Install wizard
Step 5. Select I accept the term in this license agreement, ” and click Next. Step 6. Enter User and Organization name.
Step 7. Choose the setup type. We recommend Complete for most users. Step 8. Click Install to confirm settings and begin actual installation.
Step 9. Click Yes to acknowledge the Microsoft Digital Signature message and
continue the installation.
Step 10. Click Finish to complete the installation. Step 11. Click Yes to restart the PC.
Logian-16 Users Manual Page 14
Logian-16
User Manual
Step1.
Step 2.
Step 3.
Step 5.
Step 6.
Step 7.
Step 4.
Logian-16 Users Manual Page 15
Step 8.
Logian-16
User Manual
Hardware Installation
Hardware installation simply involves connecting the Logic Analyzer to your
computer with the included USB Cable as shown in Figures 2-4 and 2-5.
1. Plug the fixed end of the cables into the LA (Fig.2-1).
Fig. 2-1
Fig. 2-2
2. Plug the loose ends into the connectors on the circuit board to be analyzed (Fig.2-2).
Note: The following sequence must be observed when connecting the connectors into the circuit board: A0 = Brown, A2 = Red, A3=Orange and so on.
3. The circuit board must be grounded to the Logic Analyzer with the connecting cables (Fig.2-3).
Fig. 2-3
Step 1 - Plug the thin male end of the USB cable into the laptop or PC. Step 2 - Plug the square female end into the logic analyzer.
Logian-16 Users Manual Page 16
Logian-16
User Manual
Fig. 2-4
4. Plug the square end of the USB cable into the Logic Analyzer (Fig.2-4).
5. Plug the thin end into the computer (Fig.2-5).
Fig. 2-5
At this point, the computer should be able to detect the Logic Analyzer and finalize the installation for hardware connection. For further information, refer to the Troubleshooting and Frequently Asked Questions (FAQ) chapters in the User Manual.
Fig. 2-6: An assembly of Laptop, Logic Analyzer, and a testing board.
Logian-16 Users Manual Page 17
Logian-16
Take the loose end of the cable and
User Manual
2.2 Tips and Advice
1. When testing a circuit board, make sure that the internal sampling frequency (within
the Logic Analyzer) is at least four times the external board frequency.
2. If the signal connector does not work well with the pins on the test board, try using the
supplied probes.
Fig. 2-7: Probes supplied with us .
3. 3-1.
insert it into the clip.
Fig. 2-8
3-2. Compress the probe as shown to
reveal 2 metal prongs (Fig.2-8).
3-3. Place the metal prongs on a metal
connector on the motherboard and release the fingers so that the prongs grip the metal connector (Fig.2-9).
Fig. 2-9
Logian-16 Users Manual Page 18
Logian-16
User Manual
4. The Logic Analyzer will connect to Web server for software updates if an internet
connection is available.
5. Unwanted signals can be filtered out using the Enable or Enable Delay functions.
6. When measuring for a long period, Compression makes memory more efficient.
7. Trigger condition depends on the test board. If triggering does not work well, try
narrowing the trigger conditions and optimize them repeatedly.
8. If a test board has a lower frequency than Logic Analyzer, sample signals according
to the external clock.
9. When clocking by an external clock, filter extra signals with the Enable function.
10. Unused channels may be removed from the Bus/Signal display using Bus/Signal
(Menu) Channel Setup.
Logian-16 Users Manual Page 19
2.3 Flow of software operation
Logian-16
User Manual
Logian-16 Users Manual Page 20
Logian-16
User Manual
Introduction to Logic Analysis
Chapter 3 gives detailed instructions in performing two basic analysis operations and five advance analysis applications with the Logic Analyzer. The basic analytical operations are the Logic Analysis and the Bus Analysis, which are fundamental to all further applications.
2.4 Logic Analysis
Logic Analysis is meant for a single signal analysis. Section 3.1 gives detailed instructions on the softwares basic setup.
Basic Software setup of the Logic Analysis
Task 1. Clock Source (Frequency) and RAM Size set up
Step 1. Click icon or Click Sampling Setup from Bus/Signal on the menu bar,
the dialogue as shown in Fig 2-1 will appear.
Fig 3-1 – Clock Source
Logian-16 Users Manual Page 21
Step 2. Internal Clock (Asynchronous Clock)
Click on Internal Clock, and then select the Frequency from the pull down menu to set up the frequency of the device under test (DUT). The frequency of the Internal Clock must be at least four times higher then the frequency of the Oscillator on the DUT. Or, select the frequency from the pull down menu on Tool Bar as Fig 3-2 shows.
Tip: Connect the signal output pin of the tested board to the Signal connector of
Logic Analyzer to measure it using the internal clock of Logic Analyzer.
Logian-16
User Manual
Fig 2-2 – Clock source drop down menu
External Clock (Synchronous Clock) Click on External Clock, and then select Rising Edge” or Falling Edge” as the
trigger condition of the DUT. In the Frequency column, type the frequency of the oscillator on the DUT.
Tip: The External Clock is applied when the frequency of the oscillator on the
tested board is less then 100MHz. Connect the output pin of oscillator on the tested board to the CLK pin of Logic Analyzer as shown in Fig 3-3.
Logian-16 Users Manual Page 22
Step 3. RAM size
Click on the RAM size from the pull down menu on the Sampling Setup dialogue as shown in Fig 3-3.
Logian-16
User Manual
Fig 3-3 – RAM Size
Tip: The relationship between RAM size, Enable, Compression and channels
as shown in Table 3-1 and Fig 3-3.
Table 3-1 RAM size vs Enable, and RAM size vs Compression and channels
Model No.
ODM-16128U
RAM
sizes/
channel
2K ~
128K
Channels
available
16 channels Available 256K 16 channels Disable
Compression
Mode &
Enable Mode
RAM
size/
channel
Channels
available
Task 2. Trigger Properties Setup
Step 1. Click icon or Click Properties from the Trigger on the Menu Bar. The
dialogue will appear as shown in Fig 2-4.
Fig 2-4 – Trigger Properties
Compression
Mode &
Enable Mode
Step 2. Trigger Level Setup
Click the drop down menu of Trigger Level on Ports A , B , C and D to select
Logian-16 Users Manual Page 23
the Trigger Level as the voltage level that a trigger source signal must reach before the trigger circuit initiates a sweep.
Tip: There are four commonly used preset voltages for Trigger Level, TTL,
CMOS (5V), CMOS (3.3V), and ECL. Users also may define their own voltage from -6V to +6V to fit with their DUT. Port A represents the pins from A0 ~ A7 on the signal connector of the Logic Analyzer, and so do Ports B, C and D. The voltage of each port may be configured independently.
Fig 2-5 – Trigger Properties Error
Step 3. Trigger Count.
Type the numbers or select the number from the pull down menu of the Count
on the Tool Bar or Click the pull down menu of the Trigger
Count on the Trigger Properties dialogue as shown in Fig 2-5 The system will be triggered where the Trigger Count is set as shown in
Figs 3-6,3-7 and Fig 3-8.
Logian-16
User Manual
Fig 3-6 – Trigger Count Drop down Menu
Logian-16 Users Manual Page 24
Fig 3-7 – Trigger Count Screen shot 1
1
2
3 5 4
Fig 3-8 – Trigger Count Screen shot 2
Step 4. Trigger Page/ Delay Time and Clock
The Trigger Page and The Delay Time and Clock cant be applied at the same time.
1. Trigger Page:
Click Trigger Page, then Type the numbers or select the numbers from the pull down menu of the Page on the Tool Bar or Click the pull down menu of the Trigger Page on the Trigger Delay” page of the Trigger Properties dialogue as shown in Figs 3-9,3-10 and 3-11. The page numbers selected will be displayed on the screen.
Logian-16
User Manual
Tip: The Trigger bar (T bar) will not be displayed when the set up of the Trigger page is more than 1
Logian-16 Users Manual Page 25
Logian-16
User Manual
Fig 3-9 – Trigger Page
Fig 3-10 – Trigger Position
Fig 3-11 – Trigger Position and screen
2. Delay Time and Clock
Click the Delay Time and Clock, then type the numbers into the column of the Trigger Delay Time or type numbers into the Trigger Delay Clock at the Trigger Delay” page of the Trigger Properties dialogue as 0. Or type the numbers into the column of Trigger Delay on the Tool Bar. The system will display the wave start.
Tip: The formula of Delay Time and Clock is Trigger Delay Time = Trigger
Delay Clock * (1/ Frequency).
Logian-16 Users Manual Page 26
Logian-16
User Manual
Step 5. Type the percentages or select the percentages from the pull down menu of the
on the Tool Bar or Click the pull down menu of the Trigger Position on the Trigger Delay page of the Trigger Properties dialogue as shown in Figs 3-12, 3-13, 3-14, and 3-15, The Trigger Position percentages selected will be displayed where counted from the right side of the screen of the system.
Fig 3-12 – Trigger Position Drop down menu
Fig 3-13 – Trigger Position 0%
Fig 3-14 – Trigger Position 10%
Fig 3-15 – Trigger Position 70%
Logian-16 Users Manual Page 27
User Manual
Step 6. Click OK to confirm the setup and exit the Trigger Properties dialogue.
Task 3. Signal/Bus Trigger Edge Setup
Highlight a designated signal, and then set its required trigger edge.
1. Left click to set the signal trigger edge as shown in Fig 3-16.
2. Right click to set the signal trigger edge as shown in Fig 3-17.
3. Click Trigger on the menu bar and choose a trigger from the list of triggers as shown in Fig 3-18.
Fig 3-16 – Trigger Left click Fig 3-17 – Trigger Right Click
Logian-16
Fig 3-18 – Trigger menu
Task 4. Run to Acquire Data
1. Single Run
Click the Single Run icon from the tool bar or press START button on top of Logic Analyzer (or press F5), then activate the signal from the DUT to the Logic Analyzer to acquire the data shown in the wave display area.
2. Repetitive Run
Click the Repetitive Run icon from the tool bar, then activate continuous signal to the Logic Analyzer to acquire the repetitive Data, and then click the stop icon to end the repetitive run.
Tip: Click icon to view all the data, and then select the wave analysis tools to
analyze the waves.
Logian-16 Users Manual Page 28
Logian-16
User Manual
Fig 3-19 –Click icon to view all the data
3. Stop to End Run
Click the Stop icon to End the Run.
Tip: If the status stays displays “Waiting…” with no signal output as shown in Fig 3-20,
click the stop icon to end the run, check the setup again, and try the run process again.
Fig 3-20 – Waiting Status
Logian-16 Users Manual Page 29
Logian-16
User Manual
2.5 Bus Logic Analysis
Section 4.2 presents detailed instructions about logic analysis with a set of grouped
signals, which is known as Bus Logic Analysis.
Basic Software setup of the Bus Logic Analysis Step 1. Set up the RAM size, Frequency, Trigger level and Trigger position as
described in Section 3.1.
Step 2. Group Signals into a Bus
Click Channels setup on Bus/Signal of menu bar, or click icon. The dialogue window shown in Fig 3-21 will appear.
Fig 3-21 – Channel setup
Rename the bus and set up the signals of the bus as shown in Fig 3-22.
Fig 3-22 – Renaming Bus
1. Click the column with blue, then type the given name of the bus, and then press enter to confirm it.
2. Go to the relative channels show as shown in the example and, go to numbers 1, 2,3,4,5 which are located on column A and row Bus1. Click them, to become purple, to set these segments of signals.
Logian-16 Users Manual Page 30
3. Click OK to get the result as shown in Fig 3-23.
Tip: Channels Setup
In Dialogue of Channels Setup, there isnt only Add Bus/Signal, but also Delete Bus/Signal, Delete All, Restore Defaults functions provided.
1. Delete Bus/Signal, first highlight the bus or channels on area 6 of Fig 3-23, then click Delete Bus/Signal to delete it.
2. Delete All, click Delete All to delete all bus/signals on area 6 of Fig 3-23.
3. Restore Defaults, click Restore Defaults to restore the dialogue of Channels Setup as shown in Fig 3-21.
Step 3. Set Trigger condition
1. Highlight the bus which will be triggered then click icon or select Bus from the Trigger of the Menu bar, the dialogue window as shown in Fig. 3-24 will appear.
Logian-16
User Manual
Fig 2-23 – Channel setup Window
Fig 3-24 – Bus Trigger Setup
Tip: or double click on trigger column of the bus as shown in Fig 3-25.
Logian-16 Users Manual Page 31
Logian-16
User Manual
Fig 3-25 – Trigger Column
2. Set Binary, Hexadecimal or Decimal as the signal of the bus to represent the value(see Fig 3-24).
3. Set = = and type the value of bus into value column to set the trigger condition of the bus.
4. Click OK,
Step 4. Click run and activate the signal from the tested board to the system to get
the result as shown in Fig 3-26.
Tip: Click icon to view all data, and then select the wave analysis tools to
analyze the waves.
Set Value is 5E as Hexadecimal, and set Operator equal to = =, then click OK, Click run and activate the signal from the tested board to the system to get the result as the trigger happens on 0X5E.
Fig 3-26 – Bus Trigger Setup
Logian-16 Users Manual Page 32
Loading...