Warning
Safety sheet
. . . . . . . . . .
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . .
. . . . . . . . .
Read First
Safety Information
To ensure safe operation and service of the Meter, follow these instructi ons. Failure to observe warnings can result in severe injury or death.
- Avoid working alone so assistance can be rendered.
- To enhance safety, test leads should be disconnected from instrument when not in use.
- Do not use test leads or the Clamp Meter if they look damaged.
- Do not use the Meter if the Clamp Meter is not operating properly or if it is wet.
- Use the Clamp Meter only as speci fied in the Instruction card or the protection provided by the Clamp Meter might be impaired.
- Special precautions are necessary when operating in situations where exposed live parts at dangerous voltages may be encountered. Personal protective equipment (not supplied with this instrument) should be used.
- The test leads sould be disconnected from the instrument when making a current measurement.
- Use caution with voltages above 30 V ac rms, or 60 V dc. These voltages pose a shock hazard.
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or
reinforced insulation Battery
Earth
AC measurement
Conforms to EU directives
If the meter is used in the vicinity of equipment which generates electromag­netic interference, the display may become unstable or the measurements show may be subject to large errors.
Caution
Off / On
Auto Power Off
30min
Auto Power Off disable :
Press buttons (except Hold button) than switch the rotary knob to power on the Meter.
AC V / DC V
Resistance
Continuity
Page 2
ACA / DCA
CAT.Ⅲ.600V with respect to
A
0
9
5
earth for the jaw. Tactile barrier for
Hand Guard. Do not hold the
meter in front of the Tactile Barrier.
I
I
I
I
X
Hz
X
I + (-I) = 0
Zero
o REL: Meter save the displayed offset value after one press of the ZERO button . T h e LCD is display­ ing the relative value. p REL (flashing): A se cond press of the ZERO button saves the offset value, with the present va lue then displayed. n Normal : Press and hold ZERO for
2 sec to return to normal opera­ tion and cancel the offset value.
Data Hold
Min / Max Hold
o MAX: Meter is saving the maxi­ mum and minimum value. Maxi­ mum value is displayed. p MIN: Meter is saving the maxi mum and minimum value. Mini­ mum value is displayed. q MAX MIN (flashing): Meter is saving the maximum and mini­ mum values. Present value is displayed. n Normal : Press and ho ld MIN MAX to return to normal operation.
Battery Replacement
Peak Hold
o P
: Meter is saving the peak
MAX
maximum and minimum value. Peak maximu m value is displayed. p P
: Meter is saving the peak
MIN
maximum and minimum value. Peak minimum value is displayed. q CA L : Press and hold PEAK but-
ton3 sec to calibrate the Meter itself for accurate meas-
urement. n Normal : Press and hold PEAK button to return to norm al opera­ tion.
z
H
0
5
A
C
A
A
0
Low battery :
” is displayed or displays “” a flash when press
button.
Back Light
Back light Automatic off after 60 seconds.
Page 3
Specifications
1-1 General Specifications
LCD display digits :
3 3/4 digit large scale LCD readout.
Display count : 4000 counts. Measuring rate : 1.5 times / sec. Overrange display :
“OL” is displayed for “Ω” functions, shows the real value for “A” and “V”
function.
Automatic power off time :
Approximately 30 minutes after power on.
Low battery indicator :
is displayed. Replace the battery when the indicator appears in the
display.
Power requirement :
9V PP3 / 6LR61 battery. Rechargeable batteries are not suitable for use with this instrument.
1-2 Environmental Conditions
Indoor Use. Calibration :
One year calibration cycle.
Operating temperature :
0°C ~ 30°C (80% RH) 30°C ~ 40°C (75% RH) 40°C ~ 50°C (45% RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (b atteri es not fitted).
Overvoltage category :
IEC 61010-1 600V CAT.Ⅲ. CAT. equipment is designed to
protect against the transients in the equipment in fixed installations, such as distribution panels, feeders and short branch circuits and light­ing systems in large buildings.
Operating altitude : 2000m (6562 ft) Conductor Size : 35mm diameter. Pollution degree : 2 EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T­28800E (5 ~ 55 Hz, 3g maximum). Drop Protection : 1.2m drop to hardwood on concrete floor.
1-3 Electrical Specifications
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C < 80%RH.
Temperature coefficient :
0.2 x (Specified accuracy) / °C, < 18°C, > 28°C .
Voltage
Function Range Accuracy
Overload protection : 600V rms Input impedance : 1MΩ // less
than 100pF.
AC Conversion Type :
AC Conversion are average sensing rms indication calibrated to the rms value of a sine wave input.
Resistance & Continuity
Overload protection : 600V rms Max. open circuit voltage : 3V Continuity check : Internal sounds
activates if the resistance of the circuit under test is less than 30Ω approximately.
0~400.0Vrms
V
400~600Vrms
V
Function Range Accuracy
Ω 400.0Ω ±(1% + 3 dgt)
0~400.0 V
400~600 V
±(1.0%+ 5 dgt) 50Hz ~ 500Hz
±(0.7% + 2 dgt)
AC Cur rent
Function Range Accuracy
A
(50~60Hz)
A
(61~400Hz)
0~400.0A
400~600A
0~400.0A
400~600A
±(1.5%+ 5 dgt) *1 ±(2.0%+ 7 dgt) ±(2.0%+ 5 dgt) *1
±(2.5%+ 7 dgt)
*1 : Over 80% of full scale add ±1.6 % Overload protection : 600A rms
AC Conversion Type and additional accuracy is same as AC Voltage.
Position Error : ±1% of reading.
DC Current
Function Range Accuracy
A
0~400.0A
400~600A
±(1.2%+ 5 dgt) ±(1.9%+ 7 dgt)
Overload protection : 600A rms Position Error : ±1% of reading. Addition erro r a cco rd ing to rema -
nence : 1% max. of current crest. Auto Power Off (APO)
The meter will automatically shut itself off after approximately 30 minutes after power on.
Frequency : Hz
Range Resolution Accuracy
20~400Hz 1Hz
±(0.1% + 2 dgt)
Overload protection : 600A rms. Sensitivity : 3A rms for ACA (A)
(>400Hz Unspecified) Peak Hold : ± (3% + 15dgt) *>600Vpeak Unspecified.
*>600Apeak Unspecified.
Min/Max Hold : add ± 15 dgt to accuracy for ACA / DCA.
* Automatically switch to the low resolution range at Peak Hold and MIN MAX Hold.
Maintenance
Do not attempt to repair this Meter. It contains no user-se r viceable parts. Repair or serving should only be performed by qualified personal.
Cleaning
Periodically wipe the case with a dry cloth and detergent do not use abrasives or solvents.
Page 4
Limited Warranty
This Meter is warranted to the orig inal purchaser against defects in ma terial and workmanship for 1 year from the date of purchase. During this warran ty period, manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthoriz­ed repair, alteration, contamination, or abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.