Megger OTS80PB Quick Reference Manual

mM
OTS80PB/OTS60PB Oil Test Set
Testeur d’huile
Öltestgerät
Equipo de prueba de aceite
Quick reference guide
Schnellreferenz
Guía de referencia rápida
G SAFETY WARNINGS
Safety warnings and precautions must be read and under­stood before the OTS80PB/OTS60PB is used. They must be observed during use.
n The OTS80PB/OTS60PB must only be used in the way
intended and for the stated purposes described by Megger.
n The OTS80PB/OTS60PB must not be used if any part
of it is damaged.
n The OTS80PB/OTS60PB must be properly earthed. A
supplementary earth terminal is provided.
n Position the OTS80PB/OTS60PB so that power can be
easily removed.
n The OTS80PB/OTS60PB must not be used in wet
locations or with rain falling on the instrument.
n The OTS80PB/OTS60PB must only be used for testing
electrical grades of insulating liquids. No other objects or substances should be placed in the test chamber.
n The OTS80PB/OTS60PB must only be used with one
of Megger’s precision test vessels or VCM100D correctly fitted on the support horns in the test chamber.
n Replacement fuses must be of the correct type and
rating.
n If batteries are fitted, the 12 VDC charger must only
be powered from a negative earthed vehicle 12 VDC source using the Megger supplied lead.
n Calibration or repair must only be carried out by a
Megger qualified repair facility.
n The test chamber must be kept clean and free from lint
or other fibres.
n The OTS80PB/OTS60PB weighs between 16 kg and
23 kg. Care should be taken when lifting the instrument.
n It is prohibited to insert foreign objects into any gap on
the instrument before or during a test.
n Unit must be used in a well ventilated location. Small
quantities of Ozone may be created after prolonged testing.
Symbols used on this instrument
G Caution: refer to accompanying notes F Danger high voltage, risk of electrical shock
DC voltage
a AC voltage
c Equipment complies with current EU
Directives
u Earth terminal
f
Fuse
USB terminal
Equipment complies with ‘C tick’
requirements
Do not dispose of in the waste stream
Instrument setup
The OTS80PB/OTS60PB must only be used by suitably trained persons.
Out of the box
List of items
n OTS80PB/OTS60PB Oil test set instrument n 400 ml precision vessel assembly n Electrode set(s) as ordered & feeler gauge set n Impeller stirrer lid assembly if ordered n Magnetic bead stirrers (2) and retriever n User manual CD n Power lead as ordered n Carry case with shoulder strap if ordered
Accessories (supplied separately)
n Vessel assembly 150 ml n Transportation case n VCM100D voltage meter
Instrument preparation
n Place instrument on a solid surface n Connect ground terminal to a suitable earth connector n Fit selected electrodes to the threaded shafts in the
test vessel taking care not to scratch electrode surfaces
n Adjust the electrode gap with the feeler gauges
provided, then lock the adjustment wheels and recheck the gap
n Fit test vessel into instrument chamber n Connect line power n Turn on the instrument
Instrument controls
The OTS80PB/OTS60PB is controlled by four directional keys, ‘OK’ key and ‘TEST’ key.
Pressing the ‘TEST’ key will initiate a test. Directional and ‘OK’ keys are used to navigate and select options.
Display
The display is a 3.5” colour TFT display with backlight. Six tabbed windows are provided to facilitate instrument setup, test execution and other functions. Users navigate between top level tabbed windows using left and right arrows. Up and down arrows are used to highlight items and ‘OK’ toggles options positioned opposite each item.
Main tab icons
Home (Test) File Tools
Setup Information Language
Initial settings
It is advisable to set the following before initiating a test:
n Set date and time – time/date option n Adjust the display backlight for optimum legibility –
display setup option
n Select units – metric or imperial
Care of batteries
Customers should be mindful of the need to maintain battery capacity. As a minimum, lead acid batteries should be fully charged every 3 months. Customers with NiMH battery option are advised to operate from battery without line power periodically to exercise the battery. Batteries have a finite lifetime and are only replaceable at Megger authorised repair centres.
Care of instrument
n Always keep the instrument clean and free from dust
and fibrous material
n Cleanliness of the test chamber is very important n Cleaning the test vessel is mandatory and a thorough
clean must be done daily/weekly depending on frequency of testing
n Introducing fibres or other particles into the test
vessel will result in erroneous breakdown values. A small volume of a test sample can be used to rinse out remnants of old oil
n If there is any damage to the test vessel, chamber or
lid, the instrument should be repaired by a Megger authorised repair facility