Megger OTS60SX Operating Manual

OTS60SX Semi Automatic Oil Test Set
User Guide Guide de l’utilisateur
M
Gebrauchsanleitung Guía del usuario
Safety Warnings 3 General Description 4 Preparing the Oil Test Set 5 Operation 5
Oil Sampling 5 Preparing the Oil Test Vessel 5 Control Panel 6 Breakdown Testing 6 Withstand Testing 6 Warning messages 7 Testing to National Standards 7
Sampling and Testing Oil for Dielectric strength 8 Oil Testing Specifications 10 Test Set Specification 10 Accessories 12 Repair and Warranty 13
2.
SAFETY WARNINGS
Safety warnings and precautions must be read and understood before the instrument is used. must be observed during use.
They
The oil test set must be properly earthed.
The test chamber must be kept scrupulously clean.
If the test chamber cover is cracked or damaged in any way the test set must not be used, but sent
for repair to an authorised agent.
The chamber door hinges are fitted with earth contact springs. These must not be damaged or
corroded.
The correct type of oil test vessel must be used, and correctly positioned on the support horns
before carrying out any testing.
Replacement fuses must be of the correct type and rating.
THE TEST SET MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.
NOTE
Symbols used on the Test Set
Caution: Refer to accompanying notes.
Risk of electric shock.
Equipment complies with relevant EU Directives
3.
General Description
The OTS60SX is a lightweight, semi-automatic, oil dielectric strength test set. The instrument is suitable for protected field use and can be powered from a range of mains supplies. The maximum 60 kV output allows tests to be performed on oil from a wide variety of electrical installations including transformers, circuit breakers and other equipment. The operation of the test set is extremely simple and the results are displayed on a bright LED display. A selection of vessels enables the instrument to be configured for a variety of tests.
The semi automatic operation allows all types of testing to be performed. An automatic one minute timer operates when the high voltage is paused for Withstand (proof) testing. Oil samples can also be tested according to breakdown specification by using a suitable sequence of tests.
The instrument is constructed in a strong, sheet steel case. Handles are provided for ease of transportation. A pouch is located on the side of the instrument to contain accessories such as the power cord, additional electrodes and the spacing gauge. An optional carrying case incorporates a shoulder strap.
A number of vessels are available suitable for testing to a wide range of national specifications. Three types of electrode are available; spherical (IEC type), mushroom (VDE/ASTM D1816) and cylindrical (ASTM D877). The oil sample can be stirred by a motor driven impeller available in selected vessels. The vessel is located in the top of the instrument and covered by a transparent polycarbonate door with a mesh screen so that the oil breakdown can be viewed. A safety interlock ensures that the high voltage is disconnected when the chamber door is opened.
Voltage Selector
Fuseholders
Mains power
input
Mains Input Panel
4.
Preparing the Oil Test Set
Operation
Ensure that the instrument is properly earthed. The test set must be connected to a socket (receptacle) with a protective earth (ground) conductor.
Before connecting the instrument to the mains power supply, ensure the voltage selector located on the mains input panel is set to the required voltage. The correct fuse value should be fitted for the supply voltage (See holders are located next to the supply inlet.
Power cord
If the power cord plug is not suitable for your type of socket outlets, do not use an adaptor. You should use a suitable alternative power cord, or if necessary change the plug by cutting the cord and fitting a suitable plug.
The colour code of the cord is:
Earth (Ground) Yellow / Green
Neutral Blue
Phase (Line) Brown
If using a fused plug, a 3 Amp fuse to BS 1362 should be fitted.
Note: A plug severed from the power cord should be destroyed, as a plug with bare conductors is hazardous in a live socket outlet.
Specification). The fuse
Oil Sampling
The relevant test specification will give advice on the best methods to use when sampling oil. Contamination of the oil may lead to misleading results. Always run off a small amount of oil before sampling and ensure that the flow of oil is steady.
Preparing the Oil Test Vessel
Optional vessels are available for testing to national specifications (See table and Accessories). Some specifications require stirring of the oil sample as well as different test electrode shapes. Once the correct vessel has been chosen for the required test the electrode gap should be set. This is achieved by adjusting the threaded nut on the side of the vessel. Small adjustment of the gap can be made by hand tightening the nuts and rotating the spindle. A slot is provided on the end of the spindle for this purpose.
A set of individual 0,5mm feeler gauges is provided for accurate measurement of the gap. Gauges can be combined to measure from 1mm to 4mm. (see page 12)
Clean the vessel in accordance with the instructions given in the relevant test specification. Allowing space for the lid to be fitted, fill the vessel with the oil sample. The amount of oil required will depend upon the type of lid used. Load the vessel in the test chamber, connecting the motor power cable, if fitted, and carefully shut the door so that the safety interlock switch operates.
5
Main
controls
HV on
indicator
Control Panel
LED
Display
Ramp rate
selection
Mains on/off
switch
Control Panel
The control panel contains the supply on/off switch, the high voltage display, voltage ramp rate selection and the three control keys. The operation of the instrument is very simple. When the instrument is turned on, the on/off switch is illuminated, and the software edition code will be flashed up on the display.
Breakdown Testing
The suitable ramp rate for the test to be performed should be selected using the ‘s' key. The START key ‘ ‘ will then turn on the high voltage. The red H.V. indicator will light to show the output voltage is on. The voltage will then increase at the selected ramp rate and the corresponding value will be shown on the display. If oil breakdown occurs the instrument will detect this and immediately shut off the high voltage. The breakdown voltage value will be left on the display until the next test is started. If no breakdown occurs, the high voltage will rise to 60 kV. At this point the test voltage is cut off and this maximum value left on the display.
A test can be stopped at any time by pressing the button. This will remove the high voltage from the oil sample and leave the display at
Withstand Testing A Withstand (proof) test can be carried out by pressing the PAUSE ‘ ‘ key. This will maintain the high voltage at the current
value for one minute. After this time the high voltage will continue to rise until one of the following:-
1) Oil breakdown occurs
2) The STOP
0,0 kV.
key is pressed
n'
s
STOP
n
6.
3) The output voltage reaches the maximum value of the test set.
During withstand testing the ramp rate can be changed while the pause is in progress. This can allow the withstand test value to be accurately set by adjusting the ramp rate to the lowest value.
Warning Messages
Door Open
If the instrument overheats, the words ‘ ‘ and ‘ ‘ will be alternately displayed and the operation disabled. This will reset automatically, when cooled, after a few minutes.
Dismantling the vessel
1. Loosen both white locking screws.
2. Hold each electrode in turn, and unscrew, and remove the sliding arm from each electrode.
3. Remove both electrodes from the vessel.
4. Wash hands, and clean and dry the vessel as specified in ASTM D.877.
Fill level
Electrode
Locking
screw
Viton
(Oil resistant)
‘O’ Ring
Testing to National Standards
The
OTS60SX
can be used to test to any international standard. A table of the standards is given on page 10. To perform these tests it is necessary to select the correct vessel so that the specified electrodes and suitable stirring are used. Most standards require repetitive breakdown tests to be carried out and the average taken. Note that sometimes the first breakdown result is omitted from the average calculation. If continuous stirring is not required the stirrer motor supply can be disconnected when appropriate by opening the chamber door.
100 ml Oil Test Vessel
The D877 Low Volume Test Vessel is an optional accessory for use with the Megger®
OTS60SX
Oil Test Set.
The test vessel must be kept scrupulously clean. Before each use, dismantle the test vessel, clean and dry, re-assemble, and set the electrode gap to 2,54 mm (0.100 inch).
Sliding arm
Re-assembling the vessel
1. With clean, uncontaminated hands, place the cleaned electrodes in the vessel.
2. Position each electrode in turn with the fingers, and screw in a sliding arm through the vessel into the each electrode in turn.
3. Centre the electrodes, and set the gap, using the appropriate gauge.
4. Ensuring that the Viton ‘O’ rings are not fouled, hand tighten both locking screws.
Megger Limited
also supply a range of automatic oil test sets. These have the parameters of a wide range of specifications programmed into the instrument for unattended operation.
7
Sampling and Testing Oil for Dielectric Strength
The following notes should be read in conjunction with the relevant oil testing specifications.
Causes of Bad Oil
Transformers and switchgear oil may be rendered unsuitable for further use due to four main reasons:-
1. Low dielectric strength.
2. High acidity.
3. High sludge content.
4. Excessive free water content. Low dielectric strength may result from many causes, the most
common of which is foreign particles or fibres and water in combination. Individually their effect may be relatively small, but together a contamination of only a few parts in a million can cause considerable lowering of the breakdown voltage of the oil. High acidity, sludge and free water should not be tolerated, but will not necessarily reduce the dielectric strength below acceptable or specified levels.
Appearance of the Sample
Only an experienced person can judge the condition of an insulating oil from its appearance, but a general guide may be obtained from the following observations:-
1. A cloudy appearance may indicate that sludge has been formed.
2. A dark yellow colour could be a sign of overheating.
3. A blackish colour often results from an arc having taken place with either carbonisation of the oil or of the insulation within the equipment.
4. A green colour may be due to copper salts dissolving in the oil.
Cleanliness of the Apparatus
The necessity for scrupulous cleanliness in the apparatus and during the process of sampling, cannot be over emphasised. The measurement of the dielectric strength of the sample is as dependent on the cleanliness of the test cell and the sampling apparatus as the condition of the oil itself.
After cleaning the apparatus, it should under no circumstances be wiped, dried or even handled with a dusty or fibrous cloth as loose dust or fibres are liable to cause contamination.
Precautions Necessary During Sampling
Since the oil taken from a drain cock will inevitably contain an excess of settled out solid impurities, and will not necessarily be representative of the bulk of the oil, it should be run to waste until clear.
Always run off a quantity of oil into a glass bottle or a test tube for an initial check on the appearance of the oil. Attention to this point may prevent contamination of a clean test vessel.
If necessary, thoroughly clean and dry the sample point using a suitable solvent. Any cloth used should be lint free. Open the sample cock and drain to waste enough oil to ensure that the sample cock is fully flushed and the sample is representative of the bulk of oil to be tested. This quantity will depend upon the size of the transformer or main container. About 2 litres is usually suitable.
When sampling from a drum or supply container, the oil should flow at a steady rate into the test vessel and after being swilled around the sides it should be discarded. Without altering the rate of flow of the oil the quantity required for the test should then be
8
run off while taking extreme care to prevent the ingress of atmospheric dust, cloth fibres or moisture.
Do not use a syphon. In the absence of a useable sample point, use a ‘thief’.
Shield the sample from direct light until ready to be tested. Turbulence and air bubbles should be avoided when pouring the
oil. Relevant national specifications should be observed. Particular care to prevent contamination is necessary when a
test is to be made on a bulk delivery tanker, and on all occasions when a high value of dielectric strength is anticipated. The procedure suggested above should be rigorously applied.
9
Oil Testing Specifications
2.54mm
Continuous/None
Continuous/None
10
Specification
Output voltage 0 - 60 kV rms Test Frequency 61,8 Hz Transformer rating 500 VA Output disconnection Within 1 ms of detection of breakdown
Power supply 115 V (99 - 132 V), 230 V (198 - 264 V) (switch selectable) 50/60 Hz, 80 VA Fuses 500 mA (T) HBC to IEC127/5 for 220/240 V supply
1 A (T) HRC to IEC127/5 for 110/120 V supply Mains power cord fused plug (when applicable): 3 Amp to BS1362
Operation Semi automatic with 0,5, 2 or 3 kV/s selectable rate of voltage rise
Pause function with automatic 1 minute timer for Withstand testing
Display 0,0 - 60,0 kV LED digital display
HV on LED Ramp rate LED
Measurement Accuracy 2% ± 3 digits Controls Start, Pause, Stop, Ramp Rate select and Power On/Off switch Safety The instrument meets the requirements of IEC 61010-1 EMC In accordance with IEC 61326-1 Operational uncertainties Refer to www.megger.com Temperature Range Operating: 0 to 40 ºC
Humidity Range Operating: 80% RH at 40 ºC
Dimensions 336 mm (13,2 in) (H) x 400 mm (5,7 in) (L) x 235 mm (9,3 in) (D) Weight 17,5 kg (38,5 lb) Cleaning Wipe disconnected test set with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA)
Storage: -40 to +70 ºC
Storage: 93% RH at 40 ºC, 95% RH at 25 ºC
11
Accessories
Supplied with Instrument Part Number
User Guide 6172-120 Spacing Gauge (0,5, 1,0, 1,5, 2,0, 2,5, 2,54 & 4mm) 6132-009 Accessory Pouch 6320-232 Fuses:
2 x 500 mA (T) for 230 V supply (fitted to the instrument) 25950-014
2 x 1 A (T) for 115 V supply 25950-004 Mains lead (fit your own plug) 25424-860 Mains lead with US plug 25970-002
Optional Accessories
Carrying Case 6420-106
Test Vessels
0,5 l for IEC156 (spherical electrodes) 6320-233 0,5 l for IEC156 with stirrer (spherical electrodes) 6320-236 0,5 l for VDE0370 (mushroom electrodes) 6320-234 0,5 l for ASTM D1816 with stirrer
(mushroom electrodes) 6320-237
0.15 litre vessel with cylindrical electrodes
for ASTMD 877 6111-356
Spare electrodes for 0.5 L vessels Part Number
Spherical (12,7 mm, 0,5 in. dia.) 6220-484 Mushroom (36 mm dia.) 6220-580 Cylindrical (25,4 mm, 1 in. dia.) 6220-483 Cylindrical with 0,5 mm edge radius 6220-538
Guage Indentification
1mm 2mm 2.5mm 2.54mm 4mm
Rings 1 2 3 4
5152-293 5152-294 5152-296 5152-319 5152-296
12
Repair and Warranty
The instrument circuit contains static sensitive devices, and care must be taken in handling the printed circuit board. If the protection of an instrument has been impaired it should not be used, and be sent for repair by suitably trained and qualified personnel. The protection is likely to be impaired if, for example, the instrument shows visible damage, fails to perform the intended measurements, has been subjected to prolonged storage under unfavourable conditions, or has been exposed to severe transport stresses.
New Instruments are Guaranteed for 1 Year from the Date of
Purchase by the User.
Any unauthorized prior repair or adjustment will
Note:
automatically invalidate the Warranty.
Instrument Repair and Spare Parts
For service requirements for Megger Instruments contact:
Megger LImited or Megger
Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent, CT17 9EN. Norristown, PA 19403 England. U.S.A.
Tel: +44 (0) 1304 502243 Tel: +1 (610) 676-8579 Fax: +44 (0) 1304 207342 Fax: +1 (610) 676-8625
Approved Repair Companies
A number of independent instrument repair companies have been approved for repair work on most Megger instruments, using genuine Megger spare parts. Consult the Appointed Distributor / Agent regarding spare parts, repair facilities and advice on the best course of action to take.
Returning an Instrument for Repair
If returning an instrument to the manufacturer for repair, it should be sent freight pre -paid to the appropriate address. Acopy of the Invoice and of the packing note should be sent simultaneously by airmail to expedite clearance through Customs. A repair estimate showing freight return and other charges will be submitted to the sender, if required, before work on the instrument commences.
or an approved repair company.
13
Contenu User Guide - p2 Gebrauchsanleitung - s24 Guía del usuario - p34
Avertissments relatifs a la securité 15
Description générale Préparation du boîtier d'essai d'huile 17
Fonctionnement 17
Echantillonnage d'huile 17
Préparation du récipient d'huile 17
Panneau de commande 18
Essai de claquage 18
Essai de résistance (Epreuve) 18
Témoins d’affichage 19
Essais selon les normes nationales 19
Echantillonnage et essai de résistance diélectrique des huiles 20
Spécifications essai des huiles 22
Réparation et Garantie 23
16
14
Loading...
+ 30 hidden pages